SlideShare una empresa de Scribd logo
1 de 6
Descargar para leer sin conexión
Page | 1
BioLaz®
Real-Time Microbial Monitor Response to Cleanroom
Cleaning Agents Provided by Shield Medicare
There are many cleaning agents that can be used
within a pharmaceutical cleanroom. Here we report
whether Klercide cleaning agents used in a simulated
Class A cleanroom environment could potentially
cause false positive Bio-Counts in the BioLaz Real-
Time Microbial Monitor, a sensor that detects
presence of microbiological organisms through the
use of fluorescence.
Figure 1. BioLaz Real-Time Microbial Monitor
Two different testing methods were conducted on
the chemicals.
 Those supplied in original packaging of a spray
bottle were tested as a spray bottle as well as
having the chemical nebulized.
 Those only in a liquid container were nebulized to
create a mist that was measured. Liquids were not
directly injected into the BioLaz Real-Time
Microbial Monitor.
The following cleaning agents were tested:
 Klercide-CR®
Sterile Biocide A
 Klercide-CR Sterile Filtered Biocide B
 Klercide-CR Sterile Filtered Biocide C
 Klercide-CR Sterile Biocide D
 Klercide-CR Sterile Biocide E
 Klercide-CR Sterile Filtered Biocide F
 Klercide 70/30®
Sterile IPA
 Klercide 70/30 Sterile Denatured Ethanol
Methods
Environmental Testing Chamber
To ensure a consistent environment, all testing was
performed inside an environmental testing chamber
that utilized a HEPA filter to recreate a Grade A
environment. During periods of manually
administering the spray the chamber door was
opened to allow delivery of the mist. This human
interaction was done in a manner to minimize the
amount of ambient air from the test laboratory
introduced into the test chamber.
Chemicals Supplied in a Trigger Spray Bottle
Prior to testing and prior to exposure to each cleaning
agent, the BioLaz Real-Time Microbial Monitor was
allowed to establish a zero count baseline value
inside the testing chamber. A mist generated from
one and two sprays was manually introduced into the
testing chamber. This mist was allowed to be drawn
into the BioLaz Real-Time Microbial Monitor by
normal sampling methods, i.e., the sample was drawn
in from the surrounding air that contained the cleaning
agent mist and the filtered chamber air. It was
important not to spray directly into the inlet as this is
an aerosol device and not designed to be exposed to
liquid ingress.
A total of two tests were run for each chemical
sprayed.
The same battery of tests was performed
simultaneously with a Lasair®
III Particle Counter in
addition to the BioLaz Real-Time Microbial Monitor.
The testing method was the same as described
above. This testing was done to assure droplets
generated by the spray were of a proper size and
quantity to be measured by the BioLaz Real-Time
Microbial Monitor.
Figure 2. Lasair III Aerosol Particle Counter
Page | 2
Figure 3. HEPA filtered test chamber for manual
administration of mist
All Chemicals
Prior to testing and prior to exposure to each
cleaning agent, the BioLaz Real-Time Microbial
Monitor was allowed to establish a zero count
baseline value inside the test chamber by running a
sample of nebulized deionized (DI) water.
An aseptic nebulizer was filled with either the
cleaning agent or DI water and a fine mist was
sprayed into the filter chamber. This mist was
allowed to be drawn into the BioLaz Real-Time
Microbial Monitor by normal sampling methods, i.e.,
the sample was drawn in from the surrounding air
containing the cleaning agent mist and the filtered
chamber air. It was important not to spray directly
into the inlet, as this is an aerosol device and not
designed to be exposed to liquid ingress.
Each test consisted of a three-minute sample of DI
water followed by a ten-minute sample of the
chemical.
Chemicals in spray bottles were tested with the
chemical nebulized in addition to testing with the
spray container.
The chamber door remained closed at all times
when the nebulizing process occurred.
Figure 4. HEPA filtered test chamber for nebulized
administration of mist
Results
The DI water showed no Bio-Counts for any tests.
During periods when the cleaning agent mist was
generated, the non-biological channel and the
biological channel were monitored with the BioLaz
Real-Time Microbial Monitor. This compared the
liquid droplets being generated and sampled to
those that were recognized due to their fluorescence
as biological particles.
Table 1. Nebulized Samples: Summary of Counts*
 
BioLaz 1 ft
3
/min
Normalized
 
Total
Bio-
Counts
Total
Particles
> 0.5 µm
Klercide-CR
Sterile Biocide A (Lot A Bottle #1)
nebulized
0.1 387.0
Sterile Filtered Biocide B
nebulized
0.1 1,565.5
Sterile Filtered Biocide C
nebulized
0 1.3
Sterile Biocide D
nebulized
0.1 40.6
Sterile Biocide E
nebulized
0 645.4
Sterile Filtered Biocide F
nebulized
0 0.5
Klercide 70/30
Sterile IPA
nebulized
0 0.2
Sterile Denatured Ethanol
nebulized
0 0
DI water
nebulized
0 10.1
* Normalized counts may result in partial particle count calculation
Klercide-CR Sterile Biocide A exhibited high biological
counts when using the trigger spray, however these
counts were not seen when the nebulizer was used.
Additional testing was performed using a second
bottle of the same lot (Lot A) as well as a bottle from a
second lot (Lot B). Again, high biological counts were
seen when using the trigger spray but not when the
nebulizer was used.
Additional testing was performed to determine if the
trigger spray bottle containing the Klercide-CR Sterile
Biocide A could be contributing to the fluorescence
seen in the previous testing. Three spray bottles were
purchased and filled with DI water. Testing was
performed as before utilizing one and two sprays of
the DI water from each spray bottle. The DI water was
then discarded and Klercide-CR Sterile Biocide A was
poured into the appropriate spray bottle. Testing was
performed as before utilizing one and two sprays of
the Klercide-CR Sterile Biocide A. Biological counts
were not seen when using the purchased spray
bottles containing Klercide-CR Sterile Biocide A.
Page | 3
Table 2. Summary of Sprayed Samples*
BioLaz 1 ft
3
/min
Normalized
Total
Bio-
Counts
Total
Particles
> 0.5 µm
Klercide-CR Sterile Biocide A (Lot A Bottle #1)
1 spray 0 122.0
1 spray repeated 18.0 389.0
2 sprays 75.0 908.0
2 sprays repeated 32.5 614.5
Klercide-CR Sterile Filtered Biocide B
1 spray 0 44
1 spray repeated 0 34
2 sprays 1 58
2 sprays repeated 0 33
Klercide-CR Sterile Filtered Biocide C
1 spray 0 23.0
1 spray repeated 0 34.0
2 sprays 0 92.0
2 sprays repeated 0 24.5
Klercide-CR Sterile Biocide E
1 spray 0 40
1 spray repeated 0 46
2 sprays 1 145
2 sprays repeated 0 146
Klercide-CR Sterile Filtered Biocide F
1 spray 0 11
1 spray repeated 0 36
2 sprays 0 66
2 sprays repeated 0 38
Klercide 70/30 Sterile IPA
1 spray 0 29
1 spray repeated 0 35
2 sprays 0 29
2 sprays repeated 0 20
Klercide 70/30 Sterile Denatured Ethanol
1 spray 2 42
1 spray repeated 0 24
2 sprays 3 197
2 sprays repeated 0 31
* Normalized counts may result in partial particle count calculation
Page | 4
Table 3. Details of Sprayed Samples*
BioLaz 1 ft3
/min
Normalized
Lasair III 1 ft
3
/min
Normalized
Total
Bio-
Counts
Total
Particles
> 0.5 µm
Total
Particles
0.3 µm
Total
Particles
0.5 µm
Total
Particles
1 µm
Total
Particles
5 µm
Total
Particles
10 µm
Total
Particles
25 µm
Klercide-CR Sterile Biocide A (Lot A Bottle #1)
1 spray 7.8 466.2 49,896 52,722 128,610 8,292 8,046 2,922
1 spray repeated 11.7 252.5 38,979 35,271 110,178 7,857 7,155 2,316
2 sprays 132.1 3,760.7 64,350 64,074 64,074 15,666 17,922 7,788
2 sprays repeated 132.1 1,639.5 86,406 92,388 283,020 23,022 36 6,198
Klercide-CR Sterile Biocide A (Lot A Bottle #2)
1 spray 42.7 505.0 25,110.0 23,532.0 42,816.0 5,406.0 948.0 153.0
1 spray repeated 93.2 1,274.3 41,364.0 41,346.0 74,727.0 5,706.0 2,154.0 1,092.0
2 sprays 334.1 3,605.3 27,492.0 27,489.0 49,659.0 49,659.0 1,935.0 1,257.0
2 sprays repeated 336.4 2,796.4 68,227.8 69,567.6 130,719.6 13,375.8 3,223.8 2,415.6
Klercide-CR Sterile Biocide A (Lot B)
1 spray 233.1 2,898.2 37,380.0 36,654.0 53,064 3,726.0 432 168
1 spray repeated 101.0 1,274.3 35,184.0 33,468.0 47,766 3,768.0 738 210
2 sprays 784.8 12,804.9 120,240.0 138,804.0 224,532 12,222.0 1,092 558
2 sprays repeated 459.9 6,210.6 58,215.6 64,375.8 97,392 5,503.8 1,200 864
Klercide-CR Sterile Biocide B
1 spray 0 357.4 36,462 24,912 173,640 28,020 11,760 3,288
1 spray repeated 0 528.4 46,590 41,136 417,414 36,882 10,710 2,682
2 sprays 0 2,292.2 66,942 54,552 296,898 53,634 20,238 6,942
2 sprays repeated 15.5 955.7 53,976 37,128 308,124 62,622 19,068 6,168
Klercide-CR Sterile Biocide C
1 spray 0 139.9 20,328 11,976 3,150 2,448 2,310 960
1 spray repeated 0 38.9 19,770 10,260 5,346 2,958 2,352 1,056
2 sprays 7.8 46.6 47,250 26,562 8,994 5,226 4,488 1,374
2 sprays repeated 0 233.1 48,126 36,270 28,692 6,612 5,550 2,172
Klercide-CR Sterile Biocide E
1 spray 0 62.2 13,782 12,342 25,902 4,914 2,142 612
1 spray repeated 0 116.6 18,510 20,358 36,936 7,272 2,502 978
2 sprays 7.8 1,010.1 34,488 33,078 63,906 12,030 4,170 1,848
2 sprays repeated 7.8 520.6 36,000 36,666 69,810 12,306 5,502 2,904
Klercide-CR Sterile Filtered Biocide F
1 spray 0 85.5 57,126 71,148 89,022 4,080 7,506 1,650
1 spray repeated 0 69.9 71,748 84,240 98,706 4,230 7,902 3,570
2 sprays 0 567.2 90,432 97,398 161,412 13,350 22,392 5,808
2 sprays repeated 23.3 1,429.7 95,658 105,930 137,148 6,000 11,994 5,850
Klercide 70/30 Sterile IPA
1 spray 0 69.9 83,886 58,734 25,878 1,974 3,198 966
1 spray repeated 0 38.9 69,036 44,544 16,782 1,236 1,806 594
2 sprays 0 116.6 105,666 60,684 22,152 2,328 4,902 2,832
2 sprays repeated 0 124.3 89,478 52,590 19,308 2,850 6,204 4,416
Klercide 70/30 Sterile Denatured Ethanol
1 spray 0 132.1 71,094.0 49,812.0 31,338.00 1,596.0 2,352 1,884
1 spray repeated 0 7.8 15,502.5 10,414.5 5,737.56 250.5 474 327
2 sprays 0 38.9 65,544.0 44,640.0 25,338.00 3,414.0 6,528 3,354
2 sprays repeated 0 85.5 58,140.0 37,410.0 20,226.00 2,628.0 4,986 2,838
* Normalized counts may result in partial particle count calculation
Page | 5
Table 4. Details of Nebulized Samples*
BioLaz 1 ft
3
/min
Normalized
Lasair III 1 ft
3
/min
Normalized
Total
Bio-
Counts
Total
Particles
> 0.5 µm
Total
Particles
0.3 µm
Total
Particles
0.5 µm
Total
Particles
1 µm
Total
Particles
5 µm
Total
Particles
10 µm
Total
Particles
25 µm
Klercide-CR Sterile Biocide A (Lot A Bottle #1)
Nebulized 0 0 2,212,418 491,731 3,784.2 0.6 0 0
Klercide-CR Sterile Biocide A (Lot B)
Nebulized 0 0 2,522,100 781,561 20,766.0 0 0 0
* Normalized counts may result in partial particle count calculation
Table 5. Details of Additional Testing of Biocide A in New Bottles*
BioLaz 1 ft
3
/min
Normalized
Lasair III 1 ft
3
/min
Normalized
Total Bio-
Counts
Total
Particles
> 0.5 µm
Total
Particles
0.3 µm
Total
Particles
0.5 µm
Total
Particles
1 µm
Total
Particles
5 µm
Total
Particles
10 µm
Total
Particles
25 µm
Bottle A1 DI Water
1 spray 0  15.5 11,736 2,688 3,480 2,796  7,452 3,168
1 spray repeated 15.5  186.5 8,412 2,268 3,144 2,424  6,072 2,472
2 sprays 25.9  56.9 4,584 848 4,536 3,044  63,800 2,264
2 sprays repeated 0  0 8,088 1,764 7,668 6,816  14,352 5,016
Biocide A Lot A Bottle #1 
1 spray 0  0 63,564 62,256 108,192 6,864  2,592 1,548
1 spray repeated 0  0 66,300 59,640 138,168 10,152  7,488 3,264
2 sprays 0  0 93,048 82,932 259,800 19,884  15,864 6,792
2 sprays repeated 0  0 103,848 98,748 294,420 28,980  33,660 9,888
Bottle A2 DI Water
1 spray 0  15.5 5,940 1,656 2,484 1,476  4,524 2,244
1 spray repeated 0  46.6 3,564 852 648 456  1,116 804
2 sprays 15.5  31.1 21,756 4,332 18,732 11,424  26,628 9,972
2 sprays repeated 15.5  77.7 1,416 300 84 0  0 0
Biocide A Lot A Bottle #2
1 spray 0  0 43,092 35,844 143,484 9,300  9,252 4,164
1 spray repeated 0  0 28,212 23,736 32,304 4,296  4,824 2,520
2 sprays 0  0 59,088 50,388 139,056 13,980  14,280 5,784
2 sprays repeated 0  0 80,472 73,632 167,676 11,976  7,296 3,768
Bottle B DI Water
1 spray 15.5  15.5 7,188 2,124 2,784 2,688  5,340 1,368
1 spray repeated 0  15.5 1,212 468 24 36  12 0
2 sprays 0  0 9,120 2,268 3,684 2,388  5,004 2,940
2 sprays repeated 0  0 32,820 9,780 13,020 9,600  22,080 11,640
Biocide A Lot B1
1 spray 0  0 12,864 9,312 20,208 2,472  456 540
1 spray repeated 0  0 23,748 18,732 54,276 4,572  4,488 2,052
2 sprays 0  0 53,088 49,392 129,756 11,196  8,148 2,784
2 sprays repeated 0  0 56,592 43,764 102,756 9,324  4,320 1,500
* Normalized counts may result in partial particle count calculation
Conclusion
For chemicals the BioLaz Real-Time Microbial Monitor
did not indicate Bio-Counts when subjected to the
mists of the cleaning agents listed in the test. However
if external contamination such as certain brighteners
are used, there may be fluorescence characteristics.
This is why it is important to test under production
conditions to determine what baseline conditions are.
Additionally, the system was able to establish a zero
count baseline level at the end of each exposure
period, indicating there was no damage or residual
effect to the instrument. Care should be taken when
using Klercide-CR® Sterile Biocide A in the provided
trigger spray bottle as it appears to exhibit
fluorescence as measured by the BioLaz Real-Time
Microbial Monitor, possibly due to materials in the
container, not the chemical.
Page | 6
_________________________________________________________________________________________________________________
Author: Elizabeth Bennett
ebennett@pmeasuring.com
303.443.7100 x221
Elizabeth joined Particle Measuring Systems in 2012 as a Microbiology Application Scientist. She has a B.S.
in Microbiology from Colorado State University.
BioLaz®
and Lasair®
are registered trademarks of Particle Measuring Systems, Inc.
All other trademarks are the property of their respective owners.
© 2013 Particle Measuring Systems. All rights reserved.
Reproduction or translation of any part of this work without the permission of the copyright owner is unlawful. Requests for permission or
further information should be addressed to Particle Measuring Systems, Inc. at 1-800-238-1801
Technical Paper 52
5/16/2013

Más contenido relacionado

La actualidad más candente

Leak tests in parenteral preparations s majzoob-20-july2015
Leak tests in parenteral preparations s majzoob-20-july2015Leak tests in parenteral preparations s majzoob-20-july2015
Leak tests in parenteral preparations s majzoob-20-july2015Sayeh Majzoob
 
Evaluation of parenteral products
Evaluation of parenteral productsEvaluation of parenteral products
Evaluation of parenteral productsRahulDev258
 
Bod analysis schematic diagram
Bod analysis schematic diagramBod analysis schematic diagram
Bod analysis schematic diagramJanine Samelo
 
Water analysis and testing
Water analysis and testingWater analysis and testing
Water analysis and testingNEHA MISHRA
 
Sterility testing 112070804014
Sterility testing 112070804014Sterility testing 112070804014
Sterility testing 112070804014Patel Parth
 
Testing of disinfectants
Testing of disinfectantsTesting of disinfectants
Testing of disinfectantsK Manivannan
 
quality control tests for parenterals
quality control tests for parenteralsquality control tests for parenterals
quality control tests for parenteralsAhmad Ali
 
Usp 40 -71 sterility tests
Usp 40 -71 sterility testsUsp 40 -71 sterility tests
Usp 40 -71 sterility testsBhargava Gottam
 
Filter validation
Filter validationFilter validation
Filter validationVINOTH R
 
Sterile dosage form
Sterile dosage formSterile dosage form
Sterile dosage formalaa deabes
 
Microbial analysis of water system and endotoxin estimation
Microbial analysis of water system and endotoxin estimationMicrobial analysis of water system and endotoxin estimation
Microbial analysis of water system and endotoxin estimationashapatel676
 
aseptic process technology
 aseptic process technology aseptic process technology
aseptic process technologyPRANJAY PATIL
 
in process quality control test for ophthalmic and parenterals
in process quality control test for ophthalmic and parenterals in process quality control test for ophthalmic and parenterals
in process quality control test for ophthalmic and parenterals Henisha Patel
 

La actualidad más candente (18)

Leak tests in parenteral preparations s majzoob-20-july2015
Leak tests in parenteral preparations s majzoob-20-july2015Leak tests in parenteral preparations s majzoob-20-july2015
Leak tests in parenteral preparations s majzoob-20-july2015
 
Evaluation of parenteral products
Evaluation of parenteral productsEvaluation of parenteral products
Evaluation of parenteral products
 
101396 , Iftakhar
101396 , Iftakhar101396 , Iftakhar
101396 , Iftakhar
 
Bod analysis schematic diagram
Bod analysis schematic diagramBod analysis schematic diagram
Bod analysis schematic diagram
 
Water analysis and testing
Water analysis and testingWater analysis and testing
Water analysis and testing
 
Sterility testing 112070804014
Sterility testing 112070804014Sterility testing 112070804014
Sterility testing 112070804014
 
Microdilution assay
Microdilution assayMicrodilution assay
Microdilution assay
 
Quality control of parentrals
Quality control of parentralsQuality control of parentrals
Quality control of parentrals
 
Testing of disinfectants
Testing of disinfectantsTesting of disinfectants
Testing of disinfectants
 
quality control tests for parenterals
quality control tests for parenteralsquality control tests for parenterals
quality control tests for parenterals
 
Usp 40 -71 sterility tests
Usp 40 -71 sterility testsUsp 40 -71 sterility tests
Usp 40 -71 sterility tests
 
Filter validation
Filter validationFilter validation
Filter validation
 
Sterile dosage form
Sterile dosage formSterile dosage form
Sterile dosage form
 
Microbial analysis of water system and endotoxin estimation
Microbial analysis of water system and endotoxin estimationMicrobial analysis of water system and endotoxin estimation
Microbial analysis of water system and endotoxin estimation
 
aseptic process technology
 aseptic process technology aseptic process technology
aseptic process technology
 
in process quality control test for ophthalmic and parenterals
in process quality control test for ophthalmic and parenterals in process quality control test for ophthalmic and parenterals
in process quality control test for ophthalmic and parenterals
 
Ipqc test for injectables
Ipqc test for injectablesIpqc test for injectables
Ipqc test for injectables
 
Rapid sterility testing
Rapid sterility testingRapid sterility testing
Rapid sterility testing
 

Destacado

Destacado (8)

Ispe Article
Ispe ArticleIspe Article
Ispe Article
 
Knapp
KnappKnapp
Knapp
 
Pda vi berlin 2014 ds aldrich
Pda vi berlin 2014 ds aldrichPda vi berlin 2014 ds aldrich
Pda vi berlin 2014 ds aldrich
 
PDA Presentation
PDA PresentationPDA Presentation
PDA Presentation
 
Ipqc for parenterals
Ipqc for parenteralsIpqc for parenterals
Ipqc for parenterals
 
Compendial requirements for particle testing 2014
Compendial requirements for particle testing 2014Compendial requirements for particle testing 2014
Compendial requirements for particle testing 2014
 
Glass manufacture
Glass manufactureGlass manufacture
Glass manufacture
 
Pda Visual Inspection 2009 Aldrich
Pda Visual Inspection 2009 AldrichPda Visual Inspection 2009 Aldrich
Pda Visual Inspection 2009 Aldrich
 

Similar a TP52_BioLaz_ChemTesting_SheildMedicare_051613

Anti Microbiological Assay Test or Antibiotic Assay Test of Pharmaceutical Pr...
Anti Microbiological Assay Test or Antibiotic Assay Test of Pharmaceutical Pr...Anti Microbiological Assay Test or Antibiotic Assay Test of Pharmaceutical Pr...
Anti Microbiological Assay Test or Antibiotic Assay Test of Pharmaceutical Pr...ijtsrd
 
D7 Laundry Sanitizer and Disinfectant Test
D7 Laundry Sanitizer and Disinfectant Test D7 Laundry Sanitizer and Disinfectant Test
D7 Laundry Sanitizer and Disinfectant Test Casey Latto
 
Use of bio coagulant in wastewater treatment_Kanoj Neeraj_2013
Use of bio coagulant in wastewater treatment_Kanoj Neeraj_2013Use of bio coagulant in wastewater treatment_Kanoj Neeraj_2013
Use of bio coagulant in wastewater treatment_Kanoj Neeraj_2013India Water Portal
 
Evaluation of the efficiency of sterilization methods. Sterility indicators
Evaluation of the efficiency of sterilization methods.Sterility indicatorsEvaluation of the efficiency of sterilization methods.Sterility indicators
Evaluation of the efficiency of sterilization methods. Sterility indicatorsMs. Pooja Bhandare
 
Evaluate the Disinfectant Activity of Some Commercial Preparations by Rideal ...
Evaluate the Disinfectant Activity of Some Commercial Preparations by Rideal ...Evaluate the Disinfectant Activity of Some Commercial Preparations by Rideal ...
Evaluate the Disinfectant Activity of Some Commercial Preparations by Rideal ...SriramNagarajan15
 
Evaluate the Disinfectant Activity of Some Commercial Preparations by Rideal ...
Evaluate the Disinfectant Activity of Some Commercial Preparations by Rideal ...Evaluate the Disinfectant Activity of Some Commercial Preparations by Rideal ...
Evaluate the Disinfectant Activity of Some Commercial Preparations by Rideal ...pharmaindexing
 
Pharmaceutical Plant Environmental Monitoring
Pharmaceutical Plant Environmental MonitoringPharmaceutical Plant Environmental Monitoring
Pharmaceutical Plant Environmental MonitoringFarrukh Mehmood
 
Study of-related-factors-about-positive-biological-monitoring-of-steam-steril...
Study of-related-factors-about-positive-biological-monitoring-of-steam-steril...Study of-related-factors-about-positive-biological-monitoring-of-steam-steril...
Study of-related-factors-about-positive-biological-monitoring-of-steam-steril...eko_apt
 
MicroPRO, A Rapid Microbiology Method Based on Flow Cytometry
MicroPRO, A Rapid Microbiology Method Based on Flow CytometryMicroPRO, A Rapid Microbiology Method Based on Flow Cytometry
MicroPRO, A Rapid Microbiology Method Based on Flow Cytometryguest32bcc5
 
TESTING OF DISINFECTANT CLASSES OF DISINFECTANTS METHOD FOR TESTING DISINFEC...
TESTING OF DISINFECTANT CLASSES OF DISINFECTANTS  METHOD FOR TESTING DISINFEC...TESTING OF DISINFECTANT CLASSES OF DISINFECTANTS  METHOD FOR TESTING DISINFEC...
TESTING OF DISINFECTANT CLASSES OF DISINFECTANTS METHOD FOR TESTING DISINFEC...VeerendraMaravi
 
Anti-Adhesion and Anti-Biofilm Effectiveness of Disinfectants Used In Hemodia...
Anti-Adhesion and Anti-Biofilm Effectiveness of Disinfectants Used In Hemodia...Anti-Adhesion and Anti-Biofilm Effectiveness of Disinfectants Used In Hemodia...
Anti-Adhesion and Anti-Biofilm Effectiveness of Disinfectants Used In Hemodia...IJERA Editor
 
Decon7 Crushes Glutex in Study (Salmonella on Wood and Concrete Carriers)
Decon7 Crushes Glutex in Study (Salmonella on Wood and Concrete Carriers)Decon7 Crushes Glutex in Study (Salmonella on Wood and Concrete Carriers)
Decon7 Crushes Glutex in Study (Salmonella on Wood and Concrete Carriers)Casey Latto
 
Citrex 0610 I En Revised Spa 1 Citrex Mic Mbc
Citrex 0610 I En Revised Spa 1 Citrex Mic MbcCitrex 0610 I En Revised Spa 1 Citrex Mic Mbc
Citrex 0610 I En Revised Spa 1 Citrex Mic MbcCITREX
 
A comparison between organic and conventionally grown carrots prepared on cut...
A comparison between organic and conventionally grown carrots prepared on cut...A comparison between organic and conventionally grown carrots prepared on cut...
A comparison between organic and conventionally grown carrots prepared on cut...Crystal Rose
 
Evaluation of disinfectant
Evaluation of disinfectantEvaluation of disinfectant
Evaluation of disinfectantIkenna Godwin
 
IPQC & FPQC Parenteral Formulation
IPQC & FPQC Parenteral Formulation IPQC & FPQC Parenteral Formulation
IPQC & FPQC Parenteral Formulation roshankhetade2
 
Sterility testing products (solids, liquids, ophthalmic and other sterile pro...
Sterility testing products (solids, liquids, ophthalmic and other sterile pro...Sterility testing products (solids, liquids, ophthalmic and other sterile pro...
Sterility testing products (solids, liquids, ophthalmic and other sterile pro...Ms. Pooja Bhandare
 
part6-230110081003-34eff4a5 (1).pdf
part6-230110081003-34eff4a5 (1).pdfpart6-230110081003-34eff4a5 (1).pdf
part6-230110081003-34eff4a5 (1).pdfimshrikant
 

Similar a TP52_BioLaz_ChemTesting_SheildMedicare_051613 (20)

Unit 2 ppt new.pptx
Unit 2 ppt new.pptxUnit 2 ppt new.pptx
Unit 2 ppt new.pptx
 
Anti Microbiological Assay Test or Antibiotic Assay Test of Pharmaceutical Pr...
Anti Microbiological Assay Test or Antibiotic Assay Test of Pharmaceutical Pr...Anti Microbiological Assay Test or Antibiotic Assay Test of Pharmaceutical Pr...
Anti Microbiological Assay Test or Antibiotic Assay Test of Pharmaceutical Pr...
 
D7 Laundry Sanitizer and Disinfectant Test
D7 Laundry Sanitizer and Disinfectant Test D7 Laundry Sanitizer and Disinfectant Test
D7 Laundry Sanitizer and Disinfectant Test
 
Sterility Testing of Products
Sterility Testing of ProductsSterility Testing of Products
Sterility Testing of Products
 
Use of bio coagulant in wastewater treatment_Kanoj Neeraj_2013
Use of bio coagulant in wastewater treatment_Kanoj Neeraj_2013Use of bio coagulant in wastewater treatment_Kanoj Neeraj_2013
Use of bio coagulant in wastewater treatment_Kanoj Neeraj_2013
 
Evaluation of the efficiency of sterilization methods. Sterility indicators
Evaluation of the efficiency of sterilization methods.Sterility indicatorsEvaluation of the efficiency of sterilization methods.Sterility indicators
Evaluation of the efficiency of sterilization methods. Sterility indicators
 
Evaluate the Disinfectant Activity of Some Commercial Preparations by Rideal ...
Evaluate the Disinfectant Activity of Some Commercial Preparations by Rideal ...Evaluate the Disinfectant Activity of Some Commercial Preparations by Rideal ...
Evaluate the Disinfectant Activity of Some Commercial Preparations by Rideal ...
 
Evaluate the Disinfectant Activity of Some Commercial Preparations by Rideal ...
Evaluate the Disinfectant Activity of Some Commercial Preparations by Rideal ...Evaluate the Disinfectant Activity of Some Commercial Preparations by Rideal ...
Evaluate the Disinfectant Activity of Some Commercial Preparations by Rideal ...
 
Pharmaceutical Plant Environmental Monitoring
Pharmaceutical Plant Environmental MonitoringPharmaceutical Plant Environmental Monitoring
Pharmaceutical Plant Environmental Monitoring
 
Study of-related-factors-about-positive-biological-monitoring-of-steam-steril...
Study of-related-factors-about-positive-biological-monitoring-of-steam-steril...Study of-related-factors-about-positive-biological-monitoring-of-steam-steril...
Study of-related-factors-about-positive-biological-monitoring-of-steam-steril...
 
MicroPRO, A Rapid Microbiology Method Based on Flow Cytometry
MicroPRO, A Rapid Microbiology Method Based on Flow CytometryMicroPRO, A Rapid Microbiology Method Based on Flow Cytometry
MicroPRO, A Rapid Microbiology Method Based on Flow Cytometry
 
TESTING OF DISINFECTANT CLASSES OF DISINFECTANTS METHOD FOR TESTING DISINFEC...
TESTING OF DISINFECTANT CLASSES OF DISINFECTANTS  METHOD FOR TESTING DISINFEC...TESTING OF DISINFECTANT CLASSES OF DISINFECTANTS  METHOD FOR TESTING DISINFEC...
TESTING OF DISINFECTANT CLASSES OF DISINFECTANTS METHOD FOR TESTING DISINFEC...
 
Anti-Adhesion and Anti-Biofilm Effectiveness of Disinfectants Used In Hemodia...
Anti-Adhesion and Anti-Biofilm Effectiveness of Disinfectants Used In Hemodia...Anti-Adhesion and Anti-Biofilm Effectiveness of Disinfectants Used In Hemodia...
Anti-Adhesion and Anti-Biofilm Effectiveness of Disinfectants Used In Hemodia...
 
Decon7 Crushes Glutex in Study (Salmonella on Wood and Concrete Carriers)
Decon7 Crushes Glutex in Study (Salmonella on Wood and Concrete Carriers)Decon7 Crushes Glutex in Study (Salmonella on Wood and Concrete Carriers)
Decon7 Crushes Glutex in Study (Salmonella on Wood and Concrete Carriers)
 
Citrex 0610 I En Revised Spa 1 Citrex Mic Mbc
Citrex 0610 I En Revised Spa 1 Citrex Mic MbcCitrex 0610 I En Revised Spa 1 Citrex Mic Mbc
Citrex 0610 I En Revised Spa 1 Citrex Mic Mbc
 
A comparison between organic and conventionally grown carrots prepared on cut...
A comparison between organic and conventionally grown carrots prepared on cut...A comparison between organic and conventionally grown carrots prepared on cut...
A comparison between organic and conventionally grown carrots prepared on cut...
 
Evaluation of disinfectant
Evaluation of disinfectantEvaluation of disinfectant
Evaluation of disinfectant
 
IPQC & FPQC Parenteral Formulation
IPQC & FPQC Parenteral Formulation IPQC & FPQC Parenteral Formulation
IPQC & FPQC Parenteral Formulation
 
Sterility testing products (solids, liquids, ophthalmic and other sterile pro...
Sterility testing products (solids, liquids, ophthalmic and other sterile pro...Sterility testing products (solids, liquids, ophthalmic and other sterile pro...
Sterility testing products (solids, liquids, ophthalmic and other sterile pro...
 
part6-230110081003-34eff4a5 (1).pdf
part6-230110081003-34eff4a5 (1).pdfpart6-230110081003-34eff4a5 (1).pdf
part6-230110081003-34eff4a5 (1).pdf
 

TP52_BioLaz_ChemTesting_SheildMedicare_051613

  • 1. Page | 1 BioLaz® Real-Time Microbial Monitor Response to Cleanroom Cleaning Agents Provided by Shield Medicare There are many cleaning agents that can be used within a pharmaceutical cleanroom. Here we report whether Klercide cleaning agents used in a simulated Class A cleanroom environment could potentially cause false positive Bio-Counts in the BioLaz Real- Time Microbial Monitor, a sensor that detects presence of microbiological organisms through the use of fluorescence. Figure 1. BioLaz Real-Time Microbial Monitor Two different testing methods were conducted on the chemicals.  Those supplied in original packaging of a spray bottle were tested as a spray bottle as well as having the chemical nebulized.  Those only in a liquid container were nebulized to create a mist that was measured. Liquids were not directly injected into the BioLaz Real-Time Microbial Monitor. The following cleaning agents were tested:  Klercide-CR® Sterile Biocide A  Klercide-CR Sterile Filtered Biocide B  Klercide-CR Sterile Filtered Biocide C  Klercide-CR Sterile Biocide D  Klercide-CR Sterile Biocide E  Klercide-CR Sterile Filtered Biocide F  Klercide 70/30® Sterile IPA  Klercide 70/30 Sterile Denatured Ethanol Methods Environmental Testing Chamber To ensure a consistent environment, all testing was performed inside an environmental testing chamber that utilized a HEPA filter to recreate a Grade A environment. During periods of manually administering the spray the chamber door was opened to allow delivery of the mist. This human interaction was done in a manner to minimize the amount of ambient air from the test laboratory introduced into the test chamber. Chemicals Supplied in a Trigger Spray Bottle Prior to testing and prior to exposure to each cleaning agent, the BioLaz Real-Time Microbial Monitor was allowed to establish a zero count baseline value inside the testing chamber. A mist generated from one and two sprays was manually introduced into the testing chamber. This mist was allowed to be drawn into the BioLaz Real-Time Microbial Monitor by normal sampling methods, i.e., the sample was drawn in from the surrounding air that contained the cleaning agent mist and the filtered chamber air. It was important not to spray directly into the inlet as this is an aerosol device and not designed to be exposed to liquid ingress. A total of two tests were run for each chemical sprayed. The same battery of tests was performed simultaneously with a Lasair® III Particle Counter in addition to the BioLaz Real-Time Microbial Monitor. The testing method was the same as described above. This testing was done to assure droplets generated by the spray were of a proper size and quantity to be measured by the BioLaz Real-Time Microbial Monitor. Figure 2. Lasair III Aerosol Particle Counter
  • 2. Page | 2 Figure 3. HEPA filtered test chamber for manual administration of mist All Chemicals Prior to testing and prior to exposure to each cleaning agent, the BioLaz Real-Time Microbial Monitor was allowed to establish a zero count baseline value inside the test chamber by running a sample of nebulized deionized (DI) water. An aseptic nebulizer was filled with either the cleaning agent or DI water and a fine mist was sprayed into the filter chamber. This mist was allowed to be drawn into the BioLaz Real-Time Microbial Monitor by normal sampling methods, i.e., the sample was drawn in from the surrounding air containing the cleaning agent mist and the filtered chamber air. It was important not to spray directly into the inlet, as this is an aerosol device and not designed to be exposed to liquid ingress. Each test consisted of a three-minute sample of DI water followed by a ten-minute sample of the chemical. Chemicals in spray bottles were tested with the chemical nebulized in addition to testing with the spray container. The chamber door remained closed at all times when the nebulizing process occurred. Figure 4. HEPA filtered test chamber for nebulized administration of mist Results The DI water showed no Bio-Counts for any tests. During periods when the cleaning agent mist was generated, the non-biological channel and the biological channel were monitored with the BioLaz Real-Time Microbial Monitor. This compared the liquid droplets being generated and sampled to those that were recognized due to their fluorescence as biological particles. Table 1. Nebulized Samples: Summary of Counts*   BioLaz 1 ft 3 /min Normalized   Total Bio- Counts Total Particles > 0.5 µm Klercide-CR Sterile Biocide A (Lot A Bottle #1) nebulized 0.1 387.0 Sterile Filtered Biocide B nebulized 0.1 1,565.5 Sterile Filtered Biocide C nebulized 0 1.3 Sterile Biocide D nebulized 0.1 40.6 Sterile Biocide E nebulized 0 645.4 Sterile Filtered Biocide F nebulized 0 0.5 Klercide 70/30 Sterile IPA nebulized 0 0.2 Sterile Denatured Ethanol nebulized 0 0 DI water nebulized 0 10.1 * Normalized counts may result in partial particle count calculation Klercide-CR Sterile Biocide A exhibited high biological counts when using the trigger spray, however these counts were not seen when the nebulizer was used. Additional testing was performed using a second bottle of the same lot (Lot A) as well as a bottle from a second lot (Lot B). Again, high biological counts were seen when using the trigger spray but not when the nebulizer was used. Additional testing was performed to determine if the trigger spray bottle containing the Klercide-CR Sterile Biocide A could be contributing to the fluorescence seen in the previous testing. Three spray bottles were purchased and filled with DI water. Testing was performed as before utilizing one and two sprays of the DI water from each spray bottle. The DI water was then discarded and Klercide-CR Sterile Biocide A was poured into the appropriate spray bottle. Testing was performed as before utilizing one and two sprays of the Klercide-CR Sterile Biocide A. Biological counts were not seen when using the purchased spray bottles containing Klercide-CR Sterile Biocide A.
  • 3. Page | 3 Table 2. Summary of Sprayed Samples* BioLaz 1 ft 3 /min Normalized Total Bio- Counts Total Particles > 0.5 µm Klercide-CR Sterile Biocide A (Lot A Bottle #1) 1 spray 0 122.0 1 spray repeated 18.0 389.0 2 sprays 75.0 908.0 2 sprays repeated 32.5 614.5 Klercide-CR Sterile Filtered Biocide B 1 spray 0 44 1 spray repeated 0 34 2 sprays 1 58 2 sprays repeated 0 33 Klercide-CR Sterile Filtered Biocide C 1 spray 0 23.0 1 spray repeated 0 34.0 2 sprays 0 92.0 2 sprays repeated 0 24.5 Klercide-CR Sterile Biocide E 1 spray 0 40 1 spray repeated 0 46 2 sprays 1 145 2 sprays repeated 0 146 Klercide-CR Sterile Filtered Biocide F 1 spray 0 11 1 spray repeated 0 36 2 sprays 0 66 2 sprays repeated 0 38 Klercide 70/30 Sterile IPA 1 spray 0 29 1 spray repeated 0 35 2 sprays 0 29 2 sprays repeated 0 20 Klercide 70/30 Sterile Denatured Ethanol 1 spray 2 42 1 spray repeated 0 24 2 sprays 3 197 2 sprays repeated 0 31 * Normalized counts may result in partial particle count calculation
  • 4. Page | 4 Table 3. Details of Sprayed Samples* BioLaz 1 ft3 /min Normalized Lasair III 1 ft 3 /min Normalized Total Bio- Counts Total Particles > 0.5 µm Total Particles 0.3 µm Total Particles 0.5 µm Total Particles 1 µm Total Particles 5 µm Total Particles 10 µm Total Particles 25 µm Klercide-CR Sterile Biocide A (Lot A Bottle #1) 1 spray 7.8 466.2 49,896 52,722 128,610 8,292 8,046 2,922 1 spray repeated 11.7 252.5 38,979 35,271 110,178 7,857 7,155 2,316 2 sprays 132.1 3,760.7 64,350 64,074 64,074 15,666 17,922 7,788 2 sprays repeated 132.1 1,639.5 86,406 92,388 283,020 23,022 36 6,198 Klercide-CR Sterile Biocide A (Lot A Bottle #2) 1 spray 42.7 505.0 25,110.0 23,532.0 42,816.0 5,406.0 948.0 153.0 1 spray repeated 93.2 1,274.3 41,364.0 41,346.0 74,727.0 5,706.0 2,154.0 1,092.0 2 sprays 334.1 3,605.3 27,492.0 27,489.0 49,659.0 49,659.0 1,935.0 1,257.0 2 sprays repeated 336.4 2,796.4 68,227.8 69,567.6 130,719.6 13,375.8 3,223.8 2,415.6 Klercide-CR Sterile Biocide A (Lot B) 1 spray 233.1 2,898.2 37,380.0 36,654.0 53,064 3,726.0 432 168 1 spray repeated 101.0 1,274.3 35,184.0 33,468.0 47,766 3,768.0 738 210 2 sprays 784.8 12,804.9 120,240.0 138,804.0 224,532 12,222.0 1,092 558 2 sprays repeated 459.9 6,210.6 58,215.6 64,375.8 97,392 5,503.8 1,200 864 Klercide-CR Sterile Biocide B 1 spray 0 357.4 36,462 24,912 173,640 28,020 11,760 3,288 1 spray repeated 0 528.4 46,590 41,136 417,414 36,882 10,710 2,682 2 sprays 0 2,292.2 66,942 54,552 296,898 53,634 20,238 6,942 2 sprays repeated 15.5 955.7 53,976 37,128 308,124 62,622 19,068 6,168 Klercide-CR Sterile Biocide C 1 spray 0 139.9 20,328 11,976 3,150 2,448 2,310 960 1 spray repeated 0 38.9 19,770 10,260 5,346 2,958 2,352 1,056 2 sprays 7.8 46.6 47,250 26,562 8,994 5,226 4,488 1,374 2 sprays repeated 0 233.1 48,126 36,270 28,692 6,612 5,550 2,172 Klercide-CR Sterile Biocide E 1 spray 0 62.2 13,782 12,342 25,902 4,914 2,142 612 1 spray repeated 0 116.6 18,510 20,358 36,936 7,272 2,502 978 2 sprays 7.8 1,010.1 34,488 33,078 63,906 12,030 4,170 1,848 2 sprays repeated 7.8 520.6 36,000 36,666 69,810 12,306 5,502 2,904 Klercide-CR Sterile Filtered Biocide F 1 spray 0 85.5 57,126 71,148 89,022 4,080 7,506 1,650 1 spray repeated 0 69.9 71,748 84,240 98,706 4,230 7,902 3,570 2 sprays 0 567.2 90,432 97,398 161,412 13,350 22,392 5,808 2 sprays repeated 23.3 1,429.7 95,658 105,930 137,148 6,000 11,994 5,850 Klercide 70/30 Sterile IPA 1 spray 0 69.9 83,886 58,734 25,878 1,974 3,198 966 1 spray repeated 0 38.9 69,036 44,544 16,782 1,236 1,806 594 2 sprays 0 116.6 105,666 60,684 22,152 2,328 4,902 2,832 2 sprays repeated 0 124.3 89,478 52,590 19,308 2,850 6,204 4,416 Klercide 70/30 Sterile Denatured Ethanol 1 spray 0 132.1 71,094.0 49,812.0 31,338.00 1,596.0 2,352 1,884 1 spray repeated 0 7.8 15,502.5 10,414.5 5,737.56 250.5 474 327 2 sprays 0 38.9 65,544.0 44,640.0 25,338.00 3,414.0 6,528 3,354 2 sprays repeated 0 85.5 58,140.0 37,410.0 20,226.00 2,628.0 4,986 2,838 * Normalized counts may result in partial particle count calculation
  • 5. Page | 5 Table 4. Details of Nebulized Samples* BioLaz 1 ft 3 /min Normalized Lasair III 1 ft 3 /min Normalized Total Bio- Counts Total Particles > 0.5 µm Total Particles 0.3 µm Total Particles 0.5 µm Total Particles 1 µm Total Particles 5 µm Total Particles 10 µm Total Particles 25 µm Klercide-CR Sterile Biocide A (Lot A Bottle #1) Nebulized 0 0 2,212,418 491,731 3,784.2 0.6 0 0 Klercide-CR Sterile Biocide A (Lot B) Nebulized 0 0 2,522,100 781,561 20,766.0 0 0 0 * Normalized counts may result in partial particle count calculation Table 5. Details of Additional Testing of Biocide A in New Bottles* BioLaz 1 ft 3 /min Normalized Lasair III 1 ft 3 /min Normalized Total Bio- Counts Total Particles > 0.5 µm Total Particles 0.3 µm Total Particles 0.5 µm Total Particles 1 µm Total Particles 5 µm Total Particles 10 µm Total Particles 25 µm Bottle A1 DI Water 1 spray 0  15.5 11,736 2,688 3,480 2,796  7,452 3,168 1 spray repeated 15.5  186.5 8,412 2,268 3,144 2,424  6,072 2,472 2 sprays 25.9  56.9 4,584 848 4,536 3,044  63,800 2,264 2 sprays repeated 0  0 8,088 1,764 7,668 6,816  14,352 5,016 Biocide A Lot A Bottle #1  1 spray 0  0 63,564 62,256 108,192 6,864  2,592 1,548 1 spray repeated 0  0 66,300 59,640 138,168 10,152  7,488 3,264 2 sprays 0  0 93,048 82,932 259,800 19,884  15,864 6,792 2 sprays repeated 0  0 103,848 98,748 294,420 28,980  33,660 9,888 Bottle A2 DI Water 1 spray 0  15.5 5,940 1,656 2,484 1,476  4,524 2,244 1 spray repeated 0  46.6 3,564 852 648 456  1,116 804 2 sprays 15.5  31.1 21,756 4,332 18,732 11,424  26,628 9,972 2 sprays repeated 15.5  77.7 1,416 300 84 0  0 0 Biocide A Lot A Bottle #2 1 spray 0  0 43,092 35,844 143,484 9,300  9,252 4,164 1 spray repeated 0  0 28,212 23,736 32,304 4,296  4,824 2,520 2 sprays 0  0 59,088 50,388 139,056 13,980  14,280 5,784 2 sprays repeated 0  0 80,472 73,632 167,676 11,976  7,296 3,768 Bottle B DI Water 1 spray 15.5  15.5 7,188 2,124 2,784 2,688  5,340 1,368 1 spray repeated 0  15.5 1,212 468 24 36  12 0 2 sprays 0  0 9,120 2,268 3,684 2,388  5,004 2,940 2 sprays repeated 0  0 32,820 9,780 13,020 9,600  22,080 11,640 Biocide A Lot B1 1 spray 0  0 12,864 9,312 20,208 2,472  456 540 1 spray repeated 0  0 23,748 18,732 54,276 4,572  4,488 2,052 2 sprays 0  0 53,088 49,392 129,756 11,196  8,148 2,784 2 sprays repeated 0  0 56,592 43,764 102,756 9,324  4,320 1,500 * Normalized counts may result in partial particle count calculation Conclusion For chemicals the BioLaz Real-Time Microbial Monitor did not indicate Bio-Counts when subjected to the mists of the cleaning agents listed in the test. However if external contamination such as certain brighteners are used, there may be fluorescence characteristics. This is why it is important to test under production conditions to determine what baseline conditions are. Additionally, the system was able to establish a zero count baseline level at the end of each exposure period, indicating there was no damage or residual effect to the instrument. Care should be taken when using Klercide-CR® Sterile Biocide A in the provided trigger spray bottle as it appears to exhibit fluorescence as measured by the BioLaz Real-Time Microbial Monitor, possibly due to materials in the container, not the chemical.
  • 6. Page | 6 _________________________________________________________________________________________________________________ Author: Elizabeth Bennett ebennett@pmeasuring.com 303.443.7100 x221 Elizabeth joined Particle Measuring Systems in 2012 as a Microbiology Application Scientist. She has a B.S. in Microbiology from Colorado State University. BioLaz® and Lasair® are registered trademarks of Particle Measuring Systems, Inc. All other trademarks are the property of their respective owners. © 2013 Particle Measuring Systems. All rights reserved. Reproduction or translation of any part of this work without the permission of the copyright owner is unlawful. Requests for permission or further information should be addressed to Particle Measuring Systems, Inc. at 1-800-238-1801 Technical Paper 52 5/16/2013