SlideShare una empresa de Scribd logo
1 de 56
Effects of Radiation
and Accelerated Aging
Part 6: Radiation Sterilization Issues
ATR Spectrum
Reflection from smear
on aluminum sheet
The bloom visible on this catheter
was analyzed both by contacting
its surface with the ATR Objective
and by smearing the bloom onto
aluminum sheet and analyzing
the smear by Reflectance.
The smear shows only the DEHP
while the ATR sees both the
DEHP and the PVC catheter.
Radiation Increases PVC Blooming
Of Cytotoxic DEHP Plasticizer
Radiation Generates Free Radicals in Polypropylene
Free Radicals Generate Carbonyl Groups in Polymers
Test Method Development
Decreased Ductility Can Cause Devices To Break During Use.
Radiation may increase the force to strain the device, but the strain
at break decreases. Devices seldom fail because they’re too stiff.
They fail because they break.
A Syringe barrel wall may be unbreakable
during use despite receiving the highest
radiation dose allowed. Syringe tips may
be much more fragile. Customer complaint
history and product history may indicate
what areas of a product are areas of concern;
what are the products’ weakest links. It is
these modes of failure that must be tested.
Testing must mimic the mode of failure
expected during customer use. Breakage
tests must be performed at a speed of
deformation similar to that experienced
by the product during customer use.
Testing samples subjected to Accelerated
Aging will provide data quickly that predicts
the future behavior of products. The Test
Protocols will define the Accelerated Aging
to be used.
Measuring Radiation Damage using Micro-FTIR
Mounting Aluminum Sheet On 3X2 Glass Slide
Shavings off Polypropylene Barrel’s Surface Analyzed Herein:
Shavings Obtained with Razor Blade
Close-ups of Polypropylene Shavings Analyzed Herein
0.2
0.4
0.6
0.8
1.0
1.2
1.4
1.6
1.8
2.0
Absorbance
1000150020002500300035004000
Wavenumbers (cm-1)
Same Polypropylene Film Before and After 40 kGYs of Cobalt Radiation
Blue is after
0.10
0.15
0.20
0.25
0.30
0.35
0.40
0.45
0.50
0.55
0.60
0.65
0.70
Absorbance
20002500300035004000
Wavenumbers (cm-1)
Same Polypropylene Film Before and After 40 kGYs of Cobalt Radiation
Blue is after
Hydroperoxides Carbonyls
0.08
0.10
0.12
0.14
0.16
0.18
0.20
0.22
0.24
0.26
Absorbance
16001700180019002000
Wavenumbers (cm-1)
Same Polypropylene Film Before and After 40 kGYs of Cobalt Radiation
Blue is after
Scissioned Polymer Chains Oxidize
IR Spectrum Shows Radiation Induced Oxidation of Polypropylene
as per J. Donohue MDDI
Polypropylene Oxidation from 20 and 40 kGYs of Cobalt Radiation
The “Dark Reaction” of Irradiated Polypropylene
Oxidative Degradation Continues Long After Irradiation Has Ceased
This is the Reaction that is Accelerated by Accelerated Aging
Dr. Apostolou and I “wrote the book” on Accelerated Aging Methods that work
Oxygen Can More Easily Penetrate and React
with the Polymer in a Thin Film
This Post Rad Oxidation is Not Just Peroxide Scissions.
Ambient Oxygen Continues to React with the Polymer.
This is Proven by this Vacuum Oven Aging.
Polyethylene Undergoes Similar Oxidation
when Irradiated with 20 and 40 kGYs
A Carbonyl Index Can be Defined to Measure this Oxidation
Statistical Results for Micro-FTIR Dosimetry of Gamma vs Control
The Micro-FTIR Method is Accurate, Precise, and Robust
0.10
0.20
0.30
0.10 0.20 0.30
0.00
0.00 0.40
X = Thickness of slice (area of 1304 cm-1 Absorbance)
Area of C=O
Absorbance
divided by X
(Carbonyl Index)
Radiation Damage (= Dose) Measured for Thin Surface
Shavings of Sterilized Polypropylene Medical Devices
Shavings of samples with 0 Mrads
Shavings of samples with 3.5 Mrads
Shavings of sterile product
Shavings of underdosed product
FTIR Microscopy can determine a competitor’s dose or detect underdosed non-sterile devices.
Stability of Fina Syndiotactic and Isotactic
Polypropylenes to Cobalt Radiation
and Accelerated Aging
Ziegler-Natta and Metallocene Catalysts
Controlled Orientation of Monomer Approach To Active Site
Yields Controlled Stereoregularity of Polymer Chain Formed.
For Z-Ns, Solid Catalysts Control Approach to Active Site.
For Metallocenes, Molecular Structure Controls Approach.
Stereostructure of
Isotactic Polypropylene
Hydroperoxide Formation
by “Backbiting” Oxidation
Strings of Close, Unstable,
Pendant Hydroperoxides
Free Radical Degradation of Isotactic Polypropylene
0.2
0.4
0.6
0.8
1.0
1.2
1.4
1.6
1.8
Absorbance
1000150020002500300035004000
Wavenumbers (cm-1)
Isotactic Polypropylene Before and After 38 kGy & 17 Days @ 70 C
Hydroperoxides Carbonyls
The Isotactic Polymer is extensively Oxidized by Irradiation.
It sizzles like bacon when it is melted.
0.2
0.4
0.6
0.8
1.0
1.2
1.4
1.6
1.8
2.0
Absorbance
1000150020002500300035004000
Wavenumbers (cm-1)
Syndiotactic Polypropylene Before and After 38 kGy & 17 Days @ 70 C
The Syndiotactic Polymer exhibits very little Rad-induced Oxidation
0.08
0.10
0.12
0.14
0.16
0.18
0.20
0.22
0.24
Absorbance
16001700180019002000
Wavenumbers (cm-1)
IPP & SPP: 38 kGy & 70 C Aging Study
Accelerated Aging Increases IPP Oxidation
but has Very Little Effect on SPP
IPP @ 70 C: 17 Days
88 Hrs
16 Hrs
0 Hrs
IPP 0 Dose
SPP 0 Dose
SPP @ 70 C:
0 to 17 Days
0.08
0.10
0.12
0.14
0.16
0.18
0.20
0.22
0.24
Absorbance
16001700180019002000
Wavenumbers (cm-1)
IPP & SPP: 38 kGy & 40 C Aging Study
IPP @ 40 C: 17 Days
88 Hrs
16 Hrs
0 Hrs
IPP 0 Dose
SPP 0 Dose
SPP @ 40 C:
0 to 17 Days
Accelerated Aging Increases IPP Oxidation
but has Very Little Effect on SPP
Carrier Gas Flowpath to Mass Spec Detector
Pure He flowing through
glass lined steel tube
that contains sample.
Tube is injected into
GC Inlet and heated.
Volatiles separate on
column and are analyzed
by the MS.
Radiation Sterilized
Syndiotactic Polypropylene
Generates an Order of
Magnitude Less Volatiles
than an Equal Mass of
Irradiated Isotactic Polypropylene
THDSB/GC/MS Analyses
of Post-Rad Volatiles
The Close, Unstable, Pendant Hydroperoxides Explode Like a String of Firecrackers
CO2
acetaldehyde
acetone
acetic acid
4-hydroxy4-methylpentanone
2,4-dimethylfuran
cyclopropylacetone
allylacetone
acetic anhydride
3,5,5-trimethylfuranone
acetoacetone
THDSB/GC/MS IDENTIFICATION
OF IPP POST-RAD VOLATILES
Heated Irradiated Isotactic Polypropylene Degrades
into Volatiles Based on C-C-O Units that Reveal the
Chemical Mechanisms of its Oxidation
Thermally Degraded Post Rad Isotactic Polypropylene
Emits Low Mass Scission Products Based on C-C-O Units …
Air & Water
Desorbed Out
Of Tube
Acetone
Acetic Acid
… Because Similar Oxidized Structures Degrade into Similar Volatiles
THDSB/GC/IR Analyses of Post-Rad Volatiles
Thermally Degraded Post Rad Syndiotactic Polypropylene
Emits a Far More Random Mix of Scission Products …
Air & Water
Desorbed Out
Of Tube
2-Hydroxy-Propionic Acid
3-Methyl-2,4-Pentanediol
4-Butyl-Gamma-Octanolactone
… Because a More Random Dispersion of Oxidized Structures Yields
a More Random Mix of Volatile Degradation Products
THDSB/GC/FTIR Analyses of Post-Rad Volatiles
Phenolic Antioxidants Protect Against Radiation Damage by
Scavenging the Free Radicals Formed in the Polymer by Radiation
Part 7: Additives; their Analysis and Issues
But Phenolic Antioxidants Turn Plastic Yellow When Irradiated
Hindered Amine Light Stabilizers Form Cytotoxic Hydroxylamines
as they Protect the Polymer from Radiation Damage
Without Discoloration
Millad 3988 Clarifies Polypropylene
Molding Heat Causes Hydrolysis, Releasing Benzaldehyde Derivatives
This Causes the Polymer to Emit a “Cherry Candy” Smell
Millad makes Polypropylene more brittle
Clarified Polypropylene Crystallized at 130 C from the Melt
Nucleation Determines Morphology
ClarifiedNot Clarified
Sublimation depletes Boundary of MilladPolypropylene Spherulites grow
Millad prevents
Spherulite growth
Millad Forms Thermally Unstable Precipitate if it is Overheated During
Injection Molding. Precipitate’s Sizzling Decomposition into Gaseous,
Superheated Aldehyde Strips Char out of Molding Press and Into Molded Parts.
Heated excised precipitate chunks undergoing thermal decomposition on Hotplate
Sizzling Decomposition into Aldehyde and Arylate
>Tiny orange spots were
scattered across the
Polypropylene matrix
>Any attempt to get the spectrum
of more than one orange spot at
a time yielded only a spectrum
of the Polypropylene matrix
>But the IR-Plan can can zoom in
on a single tiny orange spot to
yield the Arylate spectrum
shown …
… and Thermal Desorption of the
degrading material into the
GC/MS can show the formation
of the Aldehydes and Alcohol
Intermediates.
Thermal Decomposition of Precipitate Forms Aldehyde.
Cannizzaro Reaction of Aldehydes Forms Acid and Alcohol.
Condensation of Acid and Alcohol Forms Arylate.
Arylate Forms The Orange Spots
Thin Layer Chromatography (TLC) can Separate Chemical Mixtures that the
GC/MS can’t: Chemicals that are non-Volatile or Thermally Unstable
Liquid carries chemicals in spot of extract up the plate, separating them
>TLC is done on a Plate Covered with
Fluorescent Silica
>UV Light makes the Separated
Chemicals from the Mixture
Visible
>The Regions of Silica Containing
these Chemicals are scraped off
the plate, separated from the Silica,
and Identified using the Analytical
Instrumentation
Under Visible Light
Under UV Light
After Scraping
0.0
0.5
1.0
1.5
2.0
2.5
3.0
3.5
4.0
4.5
Absorbance
1000150020002500300035004000
Wavenumbers (cm-1)
Measuring Durometer of FINISHED (competitive) Devices:
FTIR of DEHP Plasticized PVC Device vs Pure PVC
Pure PVC
Part 8: Tricks of the Trade
-0.4
-0.2
0.0
0.2
0.4
0.6
0.8
1.0
1.2
1.4
1.6
1.8
2.0
2.2
Absorbance
1450150015501600
Wavenumbers (cm-1)
FINISHED Device: Measuring Plasticized PVC Durometer
Hard Endotracheal Tube
Soft Nasogastric Tube
DEHP
PVC
Spectra “normalized” for equal plasticizer (DEHP) content show
that the harder PVC has a higher PVC to Plasticizer Ratio
Such a test can tell the Durometer used by a Competitor
Higher
PVC
conc.
0.0
0.1
0.2
0.3
0.4
0.5
0.6
0.7
0.8
0.9
1.0
1.1
Absorbance
1000150020002500300035004000
Wavenumbers (cm-1)
Reverse engineering competitive catheters
Micro-FTIR Spectrum of Pellethane
zoom in here next slide
0.15
0.20
0.25
0.30
0.35
0.40
0.45
0.50
0.55
0.60
Absorbance
1360138014001420
Wavenumbers (cm-1)
Micro-FTIR ID: FINISHED Device Pellethane Durometer
80A
90A
75D
>Many catheters have tips made from a softer grade
material than the shaft.
>FTIR can measure the Durometer of finished
Polyurethane devices quickly and easily
>Multiple runs below show that the method is robust
0.2
0.4
0.6
0.8
1.0
1.2
1.4
1.6
Absorbance
1000150020002500300035004000
Wavenumbers (cm-1)
Micro-FTIR Spectrum of Tecothane
zoom in here next slide
0.02
0.04
0.06
0.08
0.10
0.12
0.14
0.16
0.18
Absorbance
1360138014001420
Wavenumbers (cm-1)
Micro-FTIR ID: FINISHED Device Tecothane Durometer
74A
85A
55D
75D
0.05
0.10
0.15
0.20
0.25
0.30
0.35
0.40
0.45
0.50
0.55
0.60
0.65
0.70
0.75
0.80
Absorbance
1000150020002500300035004000
Wavenumbers (cm-1)
Micro-FTIR Spectrum of Tecoflex with 20% BaSO4
zoom in here next slide
0.1
0.2
0.3
0.4
0.5
0.6
0.7
0.8
0.9
1.0
1.1
1.2
Absorbance
130013201340136013801400
Wavenumbers (cm-1)
Micro-FTIR ID: FINISHED Device Tecoflex Durometer
80 A
85 A
100 A
65 D
60 D
NOTE:
This method is more accurate than the
manufacturer’s ability to control or measure
their Durometer. The Hardness Pucks that
were supplied by the manufacturer (after
measuring Durometer with a Hardness Tester)
and were supposed to be 60 D were instead
HARDER than the supposed 65 D pucks.
HIGH-GLOSS
ABS CRACKS
LOW-GLOSS ABS
RESISTS CRACKING
FTIR Shows Why High-Gloss ABS Has Less ESCR Than Low-Gloss ABS
More Styrene/Butadiene Rubber dispersed in the Acrylonitrile matrix
yields High Gloss ABS with less resistance to Environmental Stress Cracking
Packaging Materials
>There are a lot of “Tricks of the Trade” in Materials Analysis.
>For example, Most packaging films are laminates with outer
heat-sealable plies and an inner strength ply.
>The FTIR Microscope requires Liquid Nitrogen (LN2) to operate
and this LN2 can be used to cryo-fracture materials.
>Cryo-fractured laminated films can be separated easily into their
individual plies for material identification and also for accurate
thickness measurements free from thickness artifacts caused by
cutting techniques that can decrease the measured thicknesses.
The film contains K-Resin…
…and the stretched Heat Seal Ply is EVA
Device package bottom web was
cryo-fractured with liquid N2 and
the heat seal ply was stretched over
aluminum sheet.
Device package bottom web was
cryo-fractured with liquid N2 and
the protruding and overhanging
plys show a Surlyn center ply
sandwiched in EVA heat seal plys.
Surlyn center ply
protruding from
fractured laminated
film
Surlyn ply and EVA heat
seal ply together
Device package bottom web was
cryofractured with liquid N2 and
this two ply film separated into
a K Resin ply and a more flexible
EVA heat seal ply.
K Resin ply is stiffer
EVA heat seal ply

Más contenido relacionado

La actualidad más candente

Physical & health hazards
Physical & health hazardsPhysical & health hazards
Physical & health hazards
Amit Shah
 
Separation, analysis etc good
Separation, analysis etc goodSeparation, analysis etc good
Separation, analysis etc good
Dr Peter Payne
 

La actualidad más candente (20)

Wet and Dry Etching
Wet and Dry EtchingWet and Dry Etching
Wet and Dry Etching
 
SIGMA TEST & RESEARCH CENTRE
SIGMA TEST & RESEARCH CENTRESIGMA TEST & RESEARCH CENTRE
SIGMA TEST & RESEARCH CENTRE
 
Wafer preparation
Wafer preparationWafer preparation
Wafer preparation
 
Plasma Etching
Plasma Etching Plasma Etching
Plasma Etching
 
THE HALOGENS
THE HALOGENSTHE HALOGENS
THE HALOGENS
 
Etching
Etching Etching
Etching
 
Etching
EtchingEtching
Etching
 
Wafer cleaning
Wafer cleaningWafer cleaning
Wafer cleaning
 
Photolithography
PhotolithographyPhotolithography
Photolithography
 
AFMC Chemistry 1995
AFMC Chemistry  1995AFMC Chemistry  1995
AFMC Chemistry 1995
 
Qualitative analysis of group 4 cations
Qualitative analysis of group 4 cationsQualitative analysis of group 4 cations
Qualitative analysis of group 4 cations
 
M.sc. laboratory manual organic chemistry binary mixture seprations
M.sc. laboratory manual organic chemistry binary mixture seprationsM.sc. laboratory manual organic chemistry binary mixture seprations
M.sc. laboratory manual organic chemistry binary mixture seprations
 
Unit 4 (Redox Titration)ppt analysis 1st sem
Unit 4 (Redox Titration)ppt analysis 1st semUnit 4 (Redox Titration)ppt analysis 1st sem
Unit 4 (Redox Titration)ppt analysis 1st sem
 
El Futuro del agua está claro.
El Futuro del agua está claro.El Futuro del agua está claro.
El Futuro del agua está claro.
 
IC Fabrication Process
IC Fabrication ProcessIC Fabrication Process
IC Fabrication Process
 
Physical & health hazards
Physical & health hazardsPhysical & health hazards
Physical & health hazards
 
Venkatesh
VenkateshVenkatesh
Venkatesh
 
Passivation
PassivationPassivation
Passivation
 
Separation, analysis etc good
Separation, analysis etc goodSeparation, analysis etc good
Separation, analysis etc good
 
AFMC Chemistry 2005
AFMC Chemistry  2005AFMC Chemistry  2005
AFMC Chemistry 2005
 

Similar a Micro ftir5-measuring radiation effects,durometer

Testing the degradation of ascorbic acid by tlc
Testing the degradation of ascorbic acid by tlcTesting the degradation of ascorbic acid by tlc
Testing the degradation of ascorbic acid by tlc
ceutics1315
 
Peroxides & Photovoltaic
Peroxides & PhotovoltaicPeroxides & Photovoltaic
Peroxides & Photovoltaic
Mohamed Adam K
 
POSTER GRANADA 2 MARCILLA ZOLLER BELTRAN v final2
POSTER GRANADA 2 MARCILLA ZOLLER BELTRAN v final2POSTER GRANADA 2 MARCILLA ZOLLER BELTRAN v final2
POSTER GRANADA 2 MARCILLA ZOLLER BELTRAN v final2
Agnes Zoller
 
Abrafati 2013 led - BASF
Abrafati 2013 led - BASFAbrafati 2013 led - BASF
Abrafati 2013 led - BASF
Roberto Caforio
 
Chemistry of man made fiber production
Chemistry of man made fiber productionChemistry of man made fiber production
Chemistry of man made fiber production
taameb
 
Sunlight induced removal of Rhodamine B from water through Semiconductor Pho...
Sunlight induced removal of Rhodamine B from water through  Semiconductor Pho...Sunlight induced removal of Rhodamine B from water through  Semiconductor Pho...
Sunlight induced removal of Rhodamine B from water through Semiconductor Pho...
Hariprasad Narayanan
 
SLIDE VIVA APRIL 2016
SLIDE VIVA APRIL 2016SLIDE VIVA APRIL 2016
SLIDE VIVA APRIL 2016
myrahalimi
 

Similar a Micro ftir5-measuring radiation effects,durometer (20)

Experiment 3: Thin-layer Chromatography and Column Chromatography
Experiment 3: Thin-layer Chromatography and Column ChromatographyExperiment 3: Thin-layer Chromatography and Column Chromatography
Experiment 3: Thin-layer Chromatography and Column Chromatography
 
Testing the degradation of ascorbic acid by tlc
Testing the degradation of ascorbic acid by tlcTesting the degradation of ascorbic acid by tlc
Testing the degradation of ascorbic acid by tlc
 
Dissolved Gas Analysis in Transformer Oil
Dissolved Gas Analysis in Transformer OilDissolved Gas Analysis in Transformer Oil
Dissolved Gas Analysis in Transformer Oil
 
Peroxides & Photovoltaic
Peroxides & PhotovoltaicPeroxides & Photovoltaic
Peroxides & Photovoltaic
 
POSTER GRANADA 2 MARCILLA ZOLLER BELTRAN v final2
POSTER GRANADA 2 MARCILLA ZOLLER BELTRAN v final2POSTER GRANADA 2 MARCILLA ZOLLER BELTRAN v final2
POSTER GRANADA 2 MARCILLA ZOLLER BELTRAN v final2
 
Abrafati 2013 led - BASF
Abrafati 2013 led - BASFAbrafati 2013 led - BASF
Abrafati 2013 led - BASF
 
Hydroxylamine Cleaning Chemistries
Hydroxylamine Cleaning ChemistriesHydroxylamine Cleaning Chemistries
Hydroxylamine Cleaning Chemistries
 
Chemistry of man made fiber production
Chemistry of man made fiber productionChemistry of man made fiber production
Chemistry of man made fiber production
 
Catalysis
CatalysisCatalysis
Catalysis
 
Stability Of Carbon 14 Labelled Compounds
Stability Of Carbon 14 Labelled CompoundsStability Of Carbon 14 Labelled Compounds
Stability Of Carbon 14 Labelled Compounds
 
Instrumental analysis
Instrumental analysisInstrumental analysis
Instrumental analysis
 
Catalysis
CatalysisCatalysis
Catalysis
 
EVALUATION SEMINAR ON IR INSTRUMENTATION
EVALUATION SEMINAR ON IR INSTRUMENTATIONEVALUATION SEMINAR ON IR INSTRUMENTATION
EVALUATION SEMINAR ON IR INSTRUMENTATION
 
ANDC TA PPT day1 07.01.2023.pptx
ANDC TA PPT day1 07.01.2023.pptxANDC TA PPT day1 07.01.2023.pptx
ANDC TA PPT day1 07.01.2023.pptx
 
30-21.pdf
30-21.pdf30-21.pdf
30-21.pdf
 
Analysis of Disinfection Byproducts by Ion Chromatography
Analysis of Disinfection Byproducts by Ion ChromatographyAnalysis of Disinfection Byproducts by Ion Chromatography
Analysis of Disinfection Byproducts by Ion Chromatography
 
Sunlight induced removal of Rhodamine B from water through Semiconductor Pho...
Sunlight induced removal of Rhodamine B from water through  Semiconductor Pho...Sunlight induced removal of Rhodamine B from water through  Semiconductor Pho...
Sunlight induced removal of Rhodamine B from water through Semiconductor Pho...
 
Lee - Organic Materials Chemistry - Spring Review 2013
Lee - Organic Materials Chemistry - Spring Review 2013Lee - Organic Materials Chemistry - Spring Review 2013
Lee - Organic Materials Chemistry - Spring Review 2013
 
SLIDE VIVA APRIL 2016
SLIDE VIVA APRIL 2016SLIDE VIVA APRIL 2016
SLIDE VIVA APRIL 2016
 
IR and FT-IR Spectroscopy
IR and FT-IR SpectroscopyIR and FT-IR Spectroscopy
IR and FT-IR Spectroscopy
 

Más de John Donohue

Más de John Donohue (12)

Micro ftir my-labforhire1
Micro ftir my-labforhire1Micro ftir my-labforhire1
Micro ftir my-labforhire1
 
Business card
Business cardBusiness card
Business card
 
Micro ftir8-particulates--electronics--fishhooks and bait
Micro ftir8-particulates--electronics--fishhooks and baitMicro ftir8-particulates--electronics--fishhooks and bait
Micro ftir8-particulates--electronics--fishhooks and bait
 
Micro ftir7- drug eluting coronary stents
Micro ftir7- drug eluting coronary stentsMicro ftir7- drug eluting coronary stents
Micro ftir7- drug eluting coronary stents
 
Micro ftir6-adhesive attacks polycarbonate
Micro ftir6-adhesive attacks polycarbonateMicro ftir6-adhesive attacks polycarbonate
Micro ftir6-adhesive attacks polycarbonate
 
Micro ftir4-measuring crystallinity
Micro ftir4-measuring crystallinityMicro ftir4-measuring crystallinity
Micro ftir4-measuring crystallinity
 
Micro ftir3 end-recalls and shutdowns
Micro ftir3 end-recalls and shutdownsMicro ftir3 end-recalls and shutdowns
Micro ftir3 end-recalls and shutdowns
 
Micro ftir2 medical-devices
Micro ftir2 medical-devicesMicro ftir2 medical-devices
Micro ftir2 medical-devices
 
Micro ftir1 asbestos-test
Micro ftir1 asbestos-testMicro ftir1 asbestos-test
Micro ftir1 asbestos-test
 
Micro ftir0 lab-intro
Micro ftir0 lab-introMicro ftir0 lab-intro
Micro ftir0 lab-intro
 
D and A method slideshare
D and A method slideshareD and A method slideshare
D and A method slideshare
 
Infrared Microscopy and Microanalyses for hire: Analytical Lab for small comp...
Infrared Microscopy and Microanalyses for hire: Analytical Lab for small comp...Infrared Microscopy and Microanalyses for hire: Analytical Lab for small comp...
Infrared Microscopy and Microanalyses for hire: Analytical Lab for small comp...
 

Último

Top Rated Call Girls In chittoor 📱 {7001035870} VIP Escorts chittoor
Top Rated Call Girls In chittoor 📱 {7001035870} VIP Escorts chittoorTop Rated Call Girls In chittoor 📱 {7001035870} VIP Escorts chittoor
Top Rated Call Girls In chittoor 📱 {7001035870} VIP Escorts chittoor
dharasingh5698
 
Call Girls in Ramesh Nagar Delhi 💯 Call Us 🔝9953056974 🔝 Escort Service
Call Girls in Ramesh Nagar Delhi 💯 Call Us 🔝9953056974 🔝 Escort ServiceCall Girls in Ramesh Nagar Delhi 💯 Call Us 🔝9953056974 🔝 Escort Service
Call Girls in Ramesh Nagar Delhi 💯 Call Us 🔝9953056974 🔝 Escort Service
9953056974 Low Rate Call Girls In Saket, Delhi NCR
 
notes on Evolution Of Analytic Scalability.ppt
notes on Evolution Of Analytic Scalability.pptnotes on Evolution Of Analytic Scalability.ppt
notes on Evolution Of Analytic Scalability.ppt
MsecMca
 
Call Girls In Bangalore ☎ 7737669865 🥵 Book Your One night Stand
Call Girls In Bangalore ☎ 7737669865 🥵 Book Your One night StandCall Girls In Bangalore ☎ 7737669865 🥵 Book Your One night Stand
Call Girls In Bangalore ☎ 7737669865 🥵 Book Your One night Stand
amitlee9823
 
XXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXX
XXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXX
XXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXX
ssuser89054b
 
Call Now ≽ 9953056974 ≼🔝 Call Girls In New Ashok Nagar ≼🔝 Delhi door step de...
Call Now ≽ 9953056974 ≼🔝 Call Girls In New Ashok Nagar  ≼🔝 Delhi door step de...Call Now ≽ 9953056974 ≼🔝 Call Girls In New Ashok Nagar  ≼🔝 Delhi door step de...
Call Now ≽ 9953056974 ≼🔝 Call Girls In New Ashok Nagar ≼🔝 Delhi door step de...
9953056974 Low Rate Call Girls In Saket, Delhi NCR
 

Último (20)

Top Rated Call Girls In chittoor 📱 {7001035870} VIP Escorts chittoor
Top Rated Call Girls In chittoor 📱 {7001035870} VIP Escorts chittoorTop Rated Call Girls In chittoor 📱 {7001035870} VIP Escorts chittoor
Top Rated Call Girls In chittoor 📱 {7001035870} VIP Escorts chittoor
 
Thermal Engineering-R & A / C - unit - V
Thermal Engineering-R & A / C - unit - VThermal Engineering-R & A / C - unit - V
Thermal Engineering-R & A / C - unit - V
 
Unit 2- Effective stress & Permeability.pdf
Unit 2- Effective stress & Permeability.pdfUnit 2- Effective stress & Permeability.pdf
Unit 2- Effective stress & Permeability.pdf
 
Call Girls in Ramesh Nagar Delhi 💯 Call Us 🔝9953056974 🔝 Escort Service
Call Girls in Ramesh Nagar Delhi 💯 Call Us 🔝9953056974 🔝 Escort ServiceCall Girls in Ramesh Nagar Delhi 💯 Call Us 🔝9953056974 🔝 Escort Service
Call Girls in Ramesh Nagar Delhi 💯 Call Us 🔝9953056974 🔝 Escort Service
 
Generative AI or GenAI technology based PPT
Generative AI or GenAI technology based PPTGenerative AI or GenAI technology based PPT
Generative AI or GenAI technology based PPT
 
Intro To Electric Vehicles PDF Notes.pdf
Intro To Electric Vehicles PDF Notes.pdfIntro To Electric Vehicles PDF Notes.pdf
Intro To Electric Vehicles PDF Notes.pdf
 
Water Industry Process Automation & Control Monthly - April 2024
Water Industry Process Automation & Control Monthly - April 2024Water Industry Process Automation & Control Monthly - April 2024
Water Industry Process Automation & Control Monthly - April 2024
 
KubeKraft presentation @CloudNativeHooghly
KubeKraft presentation @CloudNativeHooghlyKubeKraft presentation @CloudNativeHooghly
KubeKraft presentation @CloudNativeHooghly
 
notes on Evolution Of Analytic Scalability.ppt
notes on Evolution Of Analytic Scalability.pptnotes on Evolution Of Analytic Scalability.ppt
notes on Evolution Of Analytic Scalability.ppt
 
Call Girls In Bangalore ☎ 7737669865 🥵 Book Your One night Stand
Call Girls In Bangalore ☎ 7737669865 🥵 Book Your One night StandCall Girls In Bangalore ☎ 7737669865 🥵 Book Your One night Stand
Call Girls In Bangalore ☎ 7737669865 🥵 Book Your One night Stand
 
Thermal Engineering -unit - III & IV.ppt
Thermal Engineering -unit - III & IV.pptThermal Engineering -unit - III & IV.ppt
Thermal Engineering -unit - III & IV.ppt
 
UNIT - IV - Air Compressors and its Performance
UNIT - IV - Air Compressors and its PerformanceUNIT - IV - Air Compressors and its Performance
UNIT - IV - Air Compressors and its Performance
 
Block diagram reduction techniques in control systems.ppt
Block diagram reduction techniques in control systems.pptBlock diagram reduction techniques in control systems.ppt
Block diagram reduction techniques in control systems.ppt
 
data_management_and _data_science_cheat_sheet.pdf
data_management_and _data_science_cheat_sheet.pdfdata_management_and _data_science_cheat_sheet.pdf
data_management_and _data_science_cheat_sheet.pdf
 
XXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXX
XXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXX
XXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXX
 
Call Now ≽ 9953056974 ≼🔝 Call Girls In New Ashok Nagar ≼🔝 Delhi door step de...
Call Now ≽ 9953056974 ≼🔝 Call Girls In New Ashok Nagar  ≼🔝 Delhi door step de...Call Now ≽ 9953056974 ≼🔝 Call Girls In New Ashok Nagar  ≼🔝 Delhi door step de...
Call Now ≽ 9953056974 ≼🔝 Call Girls In New Ashok Nagar ≼🔝 Delhi door step de...
 
(INDIRA) Call Girl Meerut Call Now 8617697112 Meerut Escorts 24x7
(INDIRA) Call Girl Meerut Call Now 8617697112 Meerut Escorts 24x7(INDIRA) Call Girl Meerut Call Now 8617697112 Meerut Escorts 24x7
(INDIRA) Call Girl Meerut Call Now 8617697112 Meerut Escorts 24x7
 
chapter 5.pptx: drainage and irrigation engineering
chapter 5.pptx: drainage and irrigation engineeringchapter 5.pptx: drainage and irrigation engineering
chapter 5.pptx: drainage and irrigation engineering
 
Call Girls Pimpri Chinchwad Call Me 7737669865 Budget Friendly No Advance Boo...
Call Girls Pimpri Chinchwad Call Me 7737669865 Budget Friendly No Advance Boo...Call Girls Pimpri Chinchwad Call Me 7737669865 Budget Friendly No Advance Boo...
Call Girls Pimpri Chinchwad Call Me 7737669865 Budget Friendly No Advance Boo...
 
Work-Permit-Receiver-in-Saudi-Aramco.pptx
Work-Permit-Receiver-in-Saudi-Aramco.pptxWork-Permit-Receiver-in-Saudi-Aramco.pptx
Work-Permit-Receiver-in-Saudi-Aramco.pptx
 

Micro ftir5-measuring radiation effects,durometer