JOURNAL CLUB

  Vikraman.S
Quality Assurance for Cyberknife
Questions ?
Questions ?
•   Mould ?
•   Reference CT Marker ?
•   Patient Setup ?
•   Patient Collision system ?
•   Isocentre ?
•   SSD ?
•   Field Size ?
•   Reference field size ?
•   Dosimetry ?
•   Beam Data ?
•   QA ?
•   Setup verification ?
•   3D Cone beam CT verification ?
•   DRR Verification ?
•   Delivery QA ?
•   PDD,TPR,Output Factors,Kq ?
Introduction
• Fundamental of SRS is to accurate placement of intended radiation dose .
• Small errors can lead to inaccurate estimate of accumulate dose
• Consist of Linear accelerator, Manipulator arms and couch/Robotic couch
• Manipulator arms calibrated to direct the beams at the intersection of two
  orthogonal beams.
• Movements of the robotic manipulator controlled by computer in turn
  technologist
• TPS – Montecarlo /Ray searching ,Inverse optimization
• Isocentre /Non Isocentre
Targeting System
                                            Linear
                                            accelerator



         X-ray sources

Synchrony™
   camera

                     Manipulator               Robotic Delivery System




                                                          Image
                                                          detectors




             Treatment couch
Physicist
Record Keeping
• Quality Assurance and Quality control
• Good record keeping can increase work efficiency and reduce the risk of
  making errors
• Recommended – Every QA test there should be written guideline which
  clearly defines the objective ,lists the action levels for the test and
  corrective actions to be taken when these levels are exceeded.
• Essential to keep hand written /Electronic record in a well organized file
• Clear definition of steps for action levels
Terminology
• AQA – Auto QA
• DQA- Delivery QA
• E2E – End to End test
• EMO – Emergency motion off
• EPO – Emergency power off
• Geometric Isocentre – A point in space defined by the
  position of isocrystal
• Treatment Isocentre – common crossing point of cyberknife
  beams and need not to be geometric isocentre
• Isocrystal - A light sensitive detector of about 1.5 mm
  diameter mounted at the tip of rigid post.
Collision safeguards
• Robot and Room Safety – Design ,Installation and quality assurance
• Electrical Safety / Patient and Robot movement
• Modified industrial robot to support and position a linear accelerator
  weighing 160 Kg.
• There are no inherent mechanical restrictions on robot movement
Zones of motion restrictions




• Mock Treatment using - Simulation/Demonstration mode
EPO & EMO
Daily QA

•   Warm up 2000 MU for closed chambers
•   Warm up 6000 MU for open chambers
•   4 Hr downtime needs warm up
•   Daily output measurements by farmer chamber with buildup
•   Field size & MU distance ? 80 cm
•   Recommended to keep incorrect collimator to check the
    interlock
Monthly QA

•   Output
•   Energy Constancy
•   No Flatness
•   Symmetry - largest collimator to be used
•   Symmetry should be measured at a depth of 5 cm in INPLANE & CROSSPLANE
•   Radiochromic film – 10,15,25 mm radii
•   1 % tolerance with TPS commissioning
Annual QA
•   Detector ?
•   Protocol ?
•   Reference field ?
•   KQ
•   Equivalent field size
•   TPR 20/10
•   BJR
•   0.3 % error by using 0.68 Beam Quality
•   SSD 100 cm
•   6 cm equivalent square field size 6.75 X 6.75 mm
•   Detector for absolute calibration is > 10 -25mm active length
•   TLD program is recommended
•   1mm accuracy with laser central laser
•   Task group recommends laser less than 1mm accuracy
•   Gimbal Mounted laser system helps to adjust laser with less than 0.5 mm accuracy
•   Techniques to adjust to less 0.5 mm accuracy =
      Adjust the laser at 160 cm SAD to 1mm = 0.5 mm accuracy at 80 cm SAD
•   TPR
•   PDD
•   OCR
•   Output for different collimators
•   Diode detectors
•   For < 2cm chambers, micro chambers, not recommended for output factors
•   MU linearity to be checked < 1%
IMAGING ?
• No recommendations for IMAGING QA from Accuray,
• No Baseline for IMAGING QA apart from ATP
• Task group recommends
  Imaging geometry
  X ray generators and sources
  Amorphous Silicon detector
  Patient dose due to image guidance


   Dose delivery accuracy depends on
                 IMAGING
Imaging core technology for frameless SRS
Fixed Imaging
• Fixed sources
• Fixed detectors
Imaging geometry
X ray generators and sources
•   2.5 mm Al filter
•   40 -125 KV
•   25-300 mA
•   1-500 ms




• AAPM Report no 14 part 3 and No 74 of Task group 12
Amorphous Silicon detectors
• AAPM Report No 75 for Silicon detectors
• 1024 x 1024 resolution
• 41 x 41 cm2
Patient dose due to image guidance
•   G4 Source to Isocentre separation = 225 cm
•   Isocentre to detector distance = 141 cm
•   Active area = 41 x 41 cm2
•   0.1 – 0.70 mGy per image .
    Exit dose & Entrance dose ?

AAPM Report No 75 is recommended to estimate the Entrance dose
TPS
•   Montecarlo
•   Ray tracing
•   Software Testing based on AAPM report 53
•   Secondary MU is part of QA as per AAPM report 114
•   2 % tolerance with cumulative of all beams at reference point
•   20 % lower dose in MC vs. Ray tracing ( Depends on the tumor size and
    location)
•   Custom CT model
•   Inhomgeneity corrections
•   Modelled output factors tolerance is 0.5 % ( SRS)
•   Recommended to use Anthromorphic phantoms with films for DQA
•   Delivery QA is recommended for every patient for 90 % pass rate with 2%
    & 2mm gamma criteria.
•   Aim for less peripheral dose while planning
Tracking System

• Bony Structure tracking
• Fiducial Tracking
• Soft tissue Tracking
  Bony - 6D Skull tracking & Xsight Spine
  Soft - X Sight Lung without need of Fiducial
  Fiducial – With marker
  Synchrony
  In Tempo Adaptive imaging
Auto QA
•  Isocentre targeting Accuracy
•  AQA Phantom
•  Daily QA
•  CT Scan 2cm
 Acrylic phantom
 replaced with
 Same size of Metal Phantom
• Import into TPS
• Expose to AP & Lat
• 1mm tolerance from baseline
E2E
DQA
Nodes & Paths
•Questions ?

More Related Content

What's hot

telecobalt machine.ppt
telecobalt machine.ppttelecobalt machine.ppt
telecobalt machine.pptAtul Verma
 
challenges of small field dosimetry
challenges of small field dosimetrychallenges of small field dosimetry
challenges of small field dosimetryLayal Jambi
 
Importance of Planning CT Simulation(3D) in Radiothrapy/Radiation oncology.
Importance of Planning CT Simulation(3D) in Radiothrapy/Radiation oncology.Importance of Planning CT Simulation(3D) in Radiothrapy/Radiation oncology.
Importance of Planning CT Simulation(3D) in Radiothrapy/Radiation oncology.Saikat Roy
 
Particle beam – proton,neutron & heavy ion therapy
Particle beam – proton,neutron & heavy ion therapyParticle beam – proton,neutron & heavy ion therapy
Particle beam – proton,neutron & heavy ion therapyAswathi c p
 
Cyberknife At Saint Raphael’s Campus_revB
Cyberknife At Saint Raphael’s Campus_revBCyberknife At Saint Raphael’s Campus_revB
Cyberknife At Saint Raphael’s Campus_revBJustin Vinci
 
Brachytherapy dosimetry
Brachytherapy dosimetryBrachytherapy dosimetry
Brachytherapy dosimetrySabari Kumar
 
Imrt Treatment Planning And Dosimetry
Imrt Treatment Planning And DosimetryImrt Treatment Planning And Dosimetry
Imrt Treatment Planning And Dosimetryfondas vakalis
 
Beam Directed Radiotherapy - methods and principles
Beam Directed Radiotherapy - methods and principlesBeam Directed Radiotherapy - methods and principles
Beam Directed Radiotherapy - methods and principlesSantam Chakraborty
 
Electron arc therapy
Electron arc therapyElectron arc therapy
Electron arc therapyVinay Desai
 

What's hot (20)

telecobalt machine.ppt
telecobalt machine.ppttelecobalt machine.ppt
telecobalt machine.ppt
 
Total body irradiation
Total body irradiationTotal body irradiation
Total body irradiation
 
challenges of small field dosimetry
challenges of small field dosimetrychallenges of small field dosimetry
challenges of small field dosimetry
 
Tomotherapy
TomotherapyTomotherapy
Tomotherapy
 
Qc
QcQc
Qc
 
Importance of Planning CT Simulation(3D) in Radiothrapy/Radiation oncology.
Importance of Planning CT Simulation(3D) in Radiothrapy/Radiation oncology.Importance of Planning CT Simulation(3D) in Radiothrapy/Radiation oncology.
Importance of Planning CT Simulation(3D) in Radiothrapy/Radiation oncology.
 
linac QA.pptx
linac QA.pptxlinac QA.pptx
linac QA.pptx
 
Particle beam – proton,neutron & heavy ion therapy
Particle beam – proton,neutron & heavy ion therapyParticle beam – proton,neutron & heavy ion therapy
Particle beam – proton,neutron & heavy ion therapy
 
Cyberknife At Saint Raphael’s Campus_revB
Cyberknife At Saint Raphael’s Campus_revBCyberknife At Saint Raphael’s Campus_revB
Cyberknife At Saint Raphael’s Campus_revB
 
Brachytherapy dosimetry
Brachytherapy dosimetryBrachytherapy dosimetry
Brachytherapy dosimetry
 
Treatment plannings i kiran
Treatment plannings i   kiranTreatment plannings i   kiran
Treatment plannings i kiran
 
Imrt Treatment Planning And Dosimetry
Imrt Treatment Planning And DosimetryImrt Treatment Planning And Dosimetry
Imrt Treatment Planning And Dosimetry
 
Cobalt & linac
Cobalt & linacCobalt & linac
Cobalt & linac
 
Beam Directed Radiotherapy - methods and principles
Beam Directed Radiotherapy - methods and principlesBeam Directed Radiotherapy - methods and principles
Beam Directed Radiotherapy - methods and principles
 
Linear Accelerators
Linear AcceleratorsLinear Accelerators
Linear Accelerators
 
Electron arc therapy
Electron arc therapyElectron arc therapy
Electron arc therapy
 
Beam Direction
Beam DirectionBeam Direction
Beam Direction
 
Icru 38
Icru   38Icru   38
Icru 38
 
Isodose lines
Isodose linesIsodose lines
Isodose lines
 
Mlc
MlcMlc
Mlc
 

Similar to Task group report 135 cyberknie

QUALITY-ASUURANCE-OF-DIAGNOSTIC-X-RAY-EQUIPMENT.compressed (1).pptx
QUALITY-ASUURANCE-OF-DIAGNOSTIC-X-RAY-EQUIPMENT.compressed (1).pptxQUALITY-ASUURANCE-OF-DIAGNOSTIC-X-RAY-EQUIPMENT.compressed (1).pptx
QUALITY-ASUURANCE-OF-DIAGNOSTIC-X-RAY-EQUIPMENT.compressed (1).pptxAkashSharma640477
 
LINAC COMMSSN.ppt
LINAC COMMSSN.pptLINAC COMMSSN.ppt
LINAC COMMSSN.pptBBAdhikari
 
Nuclear Medicine Instrumentation and quality control presentation
 Nuclear Medicine Instrumentation and quality control presentation Nuclear Medicine Instrumentation and quality control presentation
Nuclear Medicine Instrumentation and quality control presentationMumba Chilimboyi
 
Anjali (qa qc on cr dr)
Anjali (qa qc on cr dr)Anjali (qa qc on cr dr)
Anjali (qa qc on cr dr)globalsoin
 
IAA-LA2-10-01 Spectral and Radiometric Calibration Procedure for a SWIR Hyper...
IAA-LA2-10-01 Spectral and Radiometric Calibration Procedure for a SWIR Hyper...IAA-LA2-10-01 Spectral and Radiometric Calibration Procedure for a SWIR Hyper...
IAA-LA2-10-01 Spectral and Radiometric Calibration Procedure for a SWIR Hyper...Christian Gabriel Gomez
 
Components of CT Scan Machine
Components of CT Scan MachineComponents of CT Scan Machine
Components of CT Scan MachineSyed Jamal
 
Commissioning of a Surface Guided Radiotherapy System
Commissioning of a Surface Guided Radiotherapy System Commissioning of a Surface Guided Radiotherapy System
Commissioning of a Surface Guided Radiotherapy System SGRT Community
 
Qualification of analytical instruments
Qualification of analytical instruments Qualification of analytical instruments
Qualification of analytical instruments PUNEET NIRMAL
 
Computed tomograpy
Computed tomograpyComputed tomograpy
Computed tomograpyAthuliya K
 
QC of gamma camera using different phantoms and their specification
QC of gamma camera using different phantoms and their specificationQC of gamma camera using different phantoms and their specification
QC of gamma camera using different phantoms and their specificationPreetrana12
 
Track 4 session 3 - st dev con 2016 - pedestrian dead reckoning
Track 4   session 3 - st dev con 2016 - pedestrian dead reckoningTrack 4   session 3 - st dev con 2016 - pedestrian dead reckoning
Track 4 session 3 - st dev con 2016 - pedestrian dead reckoningST_World
 
Inline freeness transmitter
Inline freeness transmitterInline freeness transmitter
Inline freeness transmitterMiia Moffitt
 
Cyberknife Treatment Delivery Radiation Therapist Perspective
Cyberknife Treatment Delivery  Radiation Therapist PerspectiveCyberknife Treatment Delivery  Radiation Therapist Perspective
Cyberknife Treatment Delivery Radiation Therapist PerspectiveSubrata Roy
 
QUALITY-ASSURANCE---MAMMOGRAPHY.pdf
QUALITY-ASSURANCE---MAMMOGRAPHY.pdfQUALITY-ASSURANCE---MAMMOGRAPHY.pdf
QUALITY-ASSURANCE---MAMMOGRAPHY.pdfAngelina Protik
 
Qualification of HPLC & LCMS.pptxfjddjdjdhdjdjj
Qualification of HPLC & LCMS.pptxfjddjdjdhdjdjjQualification of HPLC & LCMS.pptxfjddjdjdhdjdjj
Qualification of HPLC & LCMS.pptxfjddjdjdhdjdjjPratik434909
 
Qualification of HPLC & LCMS.pptdjdjdjdjfjkfx
Qualification of HPLC & LCMS.pptdjdjdjdjfjkfxQualification of HPLC & LCMS.pptdjdjdjdjfjkfx
Qualification of HPLC & LCMS.pptdjdjdjdjfjkfxPratik434909
 

Similar to Task group report 135 cyberknie (20)

QUALITY-ASUURANCE-OF-DIAGNOSTIC-X-RAY-EQUIPMENT.compressed (1).pptx
QUALITY-ASUURANCE-OF-DIAGNOSTIC-X-RAY-EQUIPMENT.compressed (1).pptxQUALITY-ASUURANCE-OF-DIAGNOSTIC-X-RAY-EQUIPMENT.compressed (1).pptx
QUALITY-ASUURANCE-OF-DIAGNOSTIC-X-RAY-EQUIPMENT.compressed (1).pptx
 
LINAC COMMSSN.ppt
LINAC COMMSSN.pptLINAC COMMSSN.ppt
LINAC COMMSSN.ppt
 
Nuclear Medicine Instrumentation and quality control presentation
 Nuclear Medicine Instrumentation and quality control presentation Nuclear Medicine Instrumentation and quality control presentation
Nuclear Medicine Instrumentation and quality control presentation
 
LTC Laser Profilometry
LTC Laser ProfilometryLTC Laser Profilometry
LTC Laser Profilometry
 
Anjali (qa qc on cr dr)
Anjali (qa qc on cr dr)Anjali (qa qc on cr dr)
Anjali (qa qc on cr dr)
 
14184864.ppt
14184864.ppt14184864.ppt
14184864.ppt
 
IAA-LA2-10-01 Spectral and Radiometric Calibration Procedure for a SWIR Hyper...
IAA-LA2-10-01 Spectral and Radiometric Calibration Procedure for a SWIR Hyper...IAA-LA2-10-01 Spectral and Radiometric Calibration Procedure for a SWIR Hyper...
IAA-LA2-10-01 Spectral and Radiometric Calibration Procedure for a SWIR Hyper...
 
Components of CT Scan Machine
Components of CT Scan MachineComponents of CT Scan Machine
Components of CT Scan Machine
 
Commissioning of a Surface Guided Radiotherapy System
Commissioning of a Surface Guided Radiotherapy System Commissioning of a Surface Guided Radiotherapy System
Commissioning of a Surface Guided Radiotherapy System
 
Qualification of analytical instruments
Qualification of analytical instruments Qualification of analytical instruments
Qualification of analytical instruments
 
Computed tomograpy
Computed tomograpyComputed tomograpy
Computed tomograpy
 
QC of gamma camera using different phantoms and their specification
QC of gamma camera using different phantoms and their specificationQC of gamma camera using different phantoms and their specification
QC of gamma camera using different phantoms and their specification
 
Track 4 session 3 - st dev con 2016 - pedestrian dead reckoning
Track 4   session 3 - st dev con 2016 - pedestrian dead reckoningTrack 4   session 3 - st dev con 2016 - pedestrian dead reckoning
Track 4 session 3 - st dev con 2016 - pedestrian dead reckoning
 
Inline freeness transmitter
Inline freeness transmitterInline freeness transmitter
Inline freeness transmitter
 
Cbct
CbctCbct
Cbct
 
unit 1.1.pptx
unit 1.1.pptxunit 1.1.pptx
unit 1.1.pptx
 
Cyberknife Treatment Delivery Radiation Therapist Perspective
Cyberknife Treatment Delivery  Radiation Therapist PerspectiveCyberknife Treatment Delivery  Radiation Therapist Perspective
Cyberknife Treatment Delivery Radiation Therapist Perspective
 
QUALITY-ASSURANCE---MAMMOGRAPHY.pdf
QUALITY-ASSURANCE---MAMMOGRAPHY.pdfQUALITY-ASSURANCE---MAMMOGRAPHY.pdf
QUALITY-ASSURANCE---MAMMOGRAPHY.pdf
 
Qualification of HPLC & LCMS.pptxfjddjdjdhdjdjj
Qualification of HPLC & LCMS.pptxfjddjdjdhdjdjjQualification of HPLC & LCMS.pptxfjddjdjdhdjdjj
Qualification of HPLC & LCMS.pptxfjddjdjdhdjdjj
 
Qualification of HPLC & LCMS.pptdjdjdjdjfjkfx
Qualification of HPLC & LCMS.pptdjdjdjdjfjkfxQualification of HPLC & LCMS.pptdjdjdjdjfjkfx
Qualification of HPLC & LCMS.pptdjdjdjdjfjkfx
 

Recently uploaded

See the 2,456 pharmacies on the National E-Pharmacy Platform
See the 2,456 pharmacies on the National E-Pharmacy PlatformSee the 2,456 pharmacies on the National E-Pharmacy Platform
See the 2,456 pharmacies on the National E-Pharmacy PlatformKweku Zurek
 
Call Girls Service Chennai Jiya 7001305949 Independent Escort Service Chennai
Call Girls Service Chennai Jiya 7001305949 Independent Escort Service ChennaiCall Girls Service Chennai Jiya 7001305949 Independent Escort Service Chennai
Call Girls Service Chennai Jiya 7001305949 Independent Escort Service ChennaiNehru place Escorts
 
call girls in green park DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️
call girls in green park  DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️call girls in green park  DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️
call girls in green park DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️saminamagar
 
Call Girls Whitefield Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Whitefield Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Whitefield Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Whitefield Just Call 7001305949 Top Class Call Girl Service Availablenarwatsonia7
 
Bangalore Call Girls Marathahalli 📞 9907093804 High Profile Service 100% Safe
Bangalore Call Girls Marathahalli 📞 9907093804 High Profile Service 100% SafeBangalore Call Girls Marathahalli 📞 9907093804 High Profile Service 100% Safe
Bangalore Call Girls Marathahalli 📞 9907093804 High Profile Service 100% Safenarwatsonia7
 
Ahmedabad Call Girls CG Road 🔝9907093804 Short 1500 💋 Night 6000
Ahmedabad Call Girls CG Road 🔝9907093804  Short 1500  💋 Night 6000Ahmedabad Call Girls CG Road 🔝9907093804  Short 1500  💋 Night 6000
Ahmedabad Call Girls CG Road 🔝9907093804 Short 1500 💋 Night 6000aliya bhat
 
Aspirin presentation slides by Dr. Rewas Ali
Aspirin presentation slides by Dr. Rewas AliAspirin presentation slides by Dr. Rewas Ali
Aspirin presentation slides by Dr. Rewas AliRewAs ALI
 
Call Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service Availablenarwatsonia7
 
VIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service Mumbai
VIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service MumbaiVIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service Mumbai
VIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service Mumbaisonalikaur4
 
call girls in Connaught Place DELHI 🔝 >༒9540349809 🔝 genuine Escort Service ...
call girls in Connaught Place  DELHI 🔝 >༒9540349809 🔝 genuine Escort Service ...call girls in Connaught Place  DELHI 🔝 >༒9540349809 🔝 genuine Escort Service ...
call girls in Connaught Place DELHI 🔝 >༒9540349809 🔝 genuine Escort Service ...saminamagar
 
Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...
Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...
Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...narwatsonia7
 
Asthma Review - GINA guidelines summary 2024
Asthma Review - GINA guidelines summary 2024Asthma Review - GINA guidelines summary 2024
Asthma Review - GINA guidelines summary 2024Gabriel Guevara MD
 
Call Girls Hebbal Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hebbal Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Hebbal Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hebbal Just Call 7001305949 Top Class Call Girl Service Availablenarwatsonia7
 
Call Girl Koramangala | 7001305949 At Low Cost Cash Payment Booking
Call Girl Koramangala | 7001305949 At Low Cost Cash Payment BookingCall Girl Koramangala | 7001305949 At Low Cost Cash Payment Booking
Call Girl Koramangala | 7001305949 At Low Cost Cash Payment Bookingnarwatsonia7
 
Call Girl Lucknow Mallika 7001305949 Independent Escort Service Lucknow
Call Girl Lucknow Mallika 7001305949 Independent Escort Service LucknowCall Girl Lucknow Mallika 7001305949 Independent Escort Service Lucknow
Call Girl Lucknow Mallika 7001305949 Independent Escort Service Lucknownarwatsonia7
 
Hemostasis Physiology and Clinical correlations by Dr Faiza.pdf
Hemostasis Physiology and Clinical correlations by Dr Faiza.pdfHemostasis Physiology and Clinical correlations by Dr Faiza.pdf
Hemostasis Physiology and Clinical correlations by Dr Faiza.pdfMedicoseAcademics
 
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.Artifacts in Nuclear Medicine with Identifying and resolving artifacts.
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.MiadAlsulami
 
Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...
Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...
Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...Miss joya
 
Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...
Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...
Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...narwatsonia7
 
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy Girls
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy GirlsCall Girls In Andheri East Call 9920874524 Book Hot And Sexy Girls
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy Girlsnehamumbai
 

Recently uploaded (20)

See the 2,456 pharmacies on the National E-Pharmacy Platform
See the 2,456 pharmacies on the National E-Pharmacy PlatformSee the 2,456 pharmacies on the National E-Pharmacy Platform
See the 2,456 pharmacies on the National E-Pharmacy Platform
 
Call Girls Service Chennai Jiya 7001305949 Independent Escort Service Chennai
Call Girls Service Chennai Jiya 7001305949 Independent Escort Service ChennaiCall Girls Service Chennai Jiya 7001305949 Independent Escort Service Chennai
Call Girls Service Chennai Jiya 7001305949 Independent Escort Service Chennai
 
call girls in green park DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️
call girls in green park  DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️call girls in green park  DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️
call girls in green park DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️
 
Call Girls Whitefield Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Whitefield Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Whitefield Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Whitefield Just Call 7001305949 Top Class Call Girl Service Available
 
Bangalore Call Girls Marathahalli 📞 9907093804 High Profile Service 100% Safe
Bangalore Call Girls Marathahalli 📞 9907093804 High Profile Service 100% SafeBangalore Call Girls Marathahalli 📞 9907093804 High Profile Service 100% Safe
Bangalore Call Girls Marathahalli 📞 9907093804 High Profile Service 100% Safe
 
Ahmedabad Call Girls CG Road 🔝9907093804 Short 1500 💋 Night 6000
Ahmedabad Call Girls CG Road 🔝9907093804  Short 1500  💋 Night 6000Ahmedabad Call Girls CG Road 🔝9907093804  Short 1500  💋 Night 6000
Ahmedabad Call Girls CG Road 🔝9907093804 Short 1500 💋 Night 6000
 
Aspirin presentation slides by Dr. Rewas Ali
Aspirin presentation slides by Dr. Rewas AliAspirin presentation slides by Dr. Rewas Ali
Aspirin presentation slides by Dr. Rewas Ali
 
Call Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service Available
 
VIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service Mumbai
VIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service MumbaiVIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service Mumbai
VIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service Mumbai
 
call girls in Connaught Place DELHI 🔝 >༒9540349809 🔝 genuine Escort Service ...
call girls in Connaught Place  DELHI 🔝 >༒9540349809 🔝 genuine Escort Service ...call girls in Connaught Place  DELHI 🔝 >༒9540349809 🔝 genuine Escort Service ...
call girls in Connaught Place DELHI 🔝 >༒9540349809 🔝 genuine Escort Service ...
 
Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...
Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...
Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...
 
Asthma Review - GINA guidelines summary 2024
Asthma Review - GINA guidelines summary 2024Asthma Review - GINA guidelines summary 2024
Asthma Review - GINA guidelines summary 2024
 
Call Girls Hebbal Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hebbal Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Hebbal Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hebbal Just Call 7001305949 Top Class Call Girl Service Available
 
Call Girl Koramangala | 7001305949 At Low Cost Cash Payment Booking
Call Girl Koramangala | 7001305949 At Low Cost Cash Payment BookingCall Girl Koramangala | 7001305949 At Low Cost Cash Payment Booking
Call Girl Koramangala | 7001305949 At Low Cost Cash Payment Booking
 
Call Girl Lucknow Mallika 7001305949 Independent Escort Service Lucknow
Call Girl Lucknow Mallika 7001305949 Independent Escort Service LucknowCall Girl Lucknow Mallika 7001305949 Independent Escort Service Lucknow
Call Girl Lucknow Mallika 7001305949 Independent Escort Service Lucknow
 
Hemostasis Physiology and Clinical correlations by Dr Faiza.pdf
Hemostasis Physiology and Clinical correlations by Dr Faiza.pdfHemostasis Physiology and Clinical correlations by Dr Faiza.pdf
Hemostasis Physiology and Clinical correlations by Dr Faiza.pdf
 
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.Artifacts in Nuclear Medicine with Identifying and resolving artifacts.
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.
 
Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...
Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...
Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...
 
Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...
Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...
Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...
 
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy Girls
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy GirlsCall Girls In Andheri East Call 9920874524 Book Hot And Sexy Girls
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy Girls
 

Task group report 135 cyberknie

  • 1. JOURNAL CLUB Vikraman.S
  • 4. Questions ? • Mould ? • Reference CT Marker ? • Patient Setup ? • Patient Collision system ? • Isocentre ? • SSD ? • Field Size ? • Reference field size ? • Dosimetry ? • Beam Data ? • QA ? • Setup verification ? • 3D Cone beam CT verification ? • DRR Verification ? • Delivery QA ? • PDD,TPR,Output Factors,Kq ?
  • 5. Introduction • Fundamental of SRS is to accurate placement of intended radiation dose . • Small errors can lead to inaccurate estimate of accumulate dose • Consist of Linear accelerator, Manipulator arms and couch/Robotic couch • Manipulator arms calibrated to direct the beams at the intersection of two orthogonal beams. • Movements of the robotic manipulator controlled by computer in turn technologist • TPS – Montecarlo /Ray searching ,Inverse optimization • Isocentre /Non Isocentre
  • 6. Targeting System Linear accelerator X-ray sources Synchrony™ camera Manipulator Robotic Delivery System Image detectors Treatment couch
  • 8. Record Keeping • Quality Assurance and Quality control • Good record keeping can increase work efficiency and reduce the risk of making errors • Recommended – Every QA test there should be written guideline which clearly defines the objective ,lists the action levels for the test and corrective actions to be taken when these levels are exceeded. • Essential to keep hand written /Electronic record in a well organized file • Clear definition of steps for action levels
  • 9. Terminology • AQA – Auto QA • DQA- Delivery QA • E2E – End to End test • EMO – Emergency motion off • EPO – Emergency power off • Geometric Isocentre – A point in space defined by the position of isocrystal • Treatment Isocentre – common crossing point of cyberknife beams and need not to be geometric isocentre • Isocrystal - A light sensitive detector of about 1.5 mm diameter mounted at the tip of rigid post.
  • 10.
  • 11.
  • 12. Collision safeguards • Robot and Room Safety – Design ,Installation and quality assurance • Electrical Safety / Patient and Robot movement • Modified industrial robot to support and position a linear accelerator weighing 160 Kg. • There are no inherent mechanical restrictions on robot movement
  • 13. Zones of motion restrictions • Mock Treatment using - Simulation/Demonstration mode
  • 15. Daily QA • Warm up 2000 MU for closed chambers • Warm up 6000 MU for open chambers • 4 Hr downtime needs warm up • Daily output measurements by farmer chamber with buildup • Field size & MU distance ? 80 cm • Recommended to keep incorrect collimator to check the interlock
  • 16. Monthly QA • Output • Energy Constancy • No Flatness • Symmetry - largest collimator to be used • Symmetry should be measured at a depth of 5 cm in INPLANE & CROSSPLANE • Radiochromic film – 10,15,25 mm radii • 1 % tolerance with TPS commissioning
  • 17. Annual QA • Detector ? • Protocol ? • Reference field ? • KQ • Equivalent field size • TPR 20/10 • BJR • 0.3 % error by using 0.68 Beam Quality • SSD 100 cm • 6 cm equivalent square field size 6.75 X 6.75 mm
  • 18. Detector for absolute calibration is > 10 -25mm active length • TLD program is recommended • 1mm accuracy with laser central laser • Task group recommends laser less than 1mm accuracy • Gimbal Mounted laser system helps to adjust laser with less than 0.5 mm accuracy • Techniques to adjust to less 0.5 mm accuracy = Adjust the laser at 160 cm SAD to 1mm = 0.5 mm accuracy at 80 cm SAD • TPR • PDD • OCR • Output for different collimators • Diode detectors • For < 2cm chambers, micro chambers, not recommended for output factors • MU linearity to be checked < 1%
  • 19. IMAGING ? • No recommendations for IMAGING QA from Accuray, • No Baseline for IMAGING QA apart from ATP • Task group recommends Imaging geometry X ray generators and sources Amorphous Silicon detector Patient dose due to image guidance Dose delivery accuracy depends on IMAGING Imaging core technology for frameless SRS
  • 20. Fixed Imaging • Fixed sources • Fixed detectors
  • 22. X ray generators and sources • 2.5 mm Al filter • 40 -125 KV • 25-300 mA • 1-500 ms • AAPM Report no 14 part 3 and No 74 of Task group 12
  • 23. Amorphous Silicon detectors • AAPM Report No 75 for Silicon detectors • 1024 x 1024 resolution • 41 x 41 cm2
  • 24. Patient dose due to image guidance • G4 Source to Isocentre separation = 225 cm • Isocentre to detector distance = 141 cm • Active area = 41 x 41 cm2 • 0.1 – 0.70 mGy per image . Exit dose & Entrance dose ? AAPM Report No 75 is recommended to estimate the Entrance dose
  • 25. TPS • Montecarlo • Ray tracing • Software Testing based on AAPM report 53 • Secondary MU is part of QA as per AAPM report 114 • 2 % tolerance with cumulative of all beams at reference point • 20 % lower dose in MC vs. Ray tracing ( Depends on the tumor size and location) • Custom CT model • Inhomgeneity corrections • Modelled output factors tolerance is 0.5 % ( SRS) • Recommended to use Anthromorphic phantoms with films for DQA • Delivery QA is recommended for every patient for 90 % pass rate with 2% & 2mm gamma criteria. • Aim for less peripheral dose while planning
  • 26.
  • 27. Tracking System • Bony Structure tracking • Fiducial Tracking • Soft tissue Tracking Bony - 6D Skull tracking & Xsight Spine Soft - X Sight Lung without need of Fiducial Fiducial – With marker Synchrony In Tempo Adaptive imaging
  • 28. Auto QA • Isocentre targeting Accuracy • AQA Phantom • Daily QA • CT Scan 2cm Acrylic phantom replaced with Same size of Metal Phantom • Import into TPS • Expose to AP & Lat • 1mm tolerance from baseline
  • 29. E2E
  • 30. DQA