PULSEOXIMETRY
DR.RIYAS A
Dr SMCSI MEDICAL COLLEGE
KARAKONAM
Pulse oximetery
Introduction
• The maintenance of optimal O2 delivery is the
core concern during anaesthesia
• “Oxygen lack not only stops the machine but
wrecks the machinery ”– J.S. Haldane.
• Monitoring of oxygenation using
pulseoximeter avoids many catastrophies
DEFINITION
• Pulse oximetry, sometimes called the fifth vital
sign, is a noninvasive method of measuring
hemoglobin saturation (SpO2) by using a light
signal transmitted through tissue
• A low SpO2 can provide warning of hypoxemia
before other signs such as cyanosis or a change in
heart rate are observed
HISTORY
• MATHEES- father of oximetry
• 20 papers in1934 –1944
• HERTZMAN 1937 –use of photoelectric finger
plethsmography
• 1975 –concept of pulse oximetry –Japan
• YELDERMAN &NEW -1983 –Nellcor pulse
oximeter
Continuation…
• The American Society of Anesthesiologists
(ASA) and American Association of Nurse
Anesthetists have made assessment of
oxygenation a standard for intraoperative
and postoperative monitoring. In 2005, an
audible alarm for the pulse oximeter was
added to the ASA monitoring standard
PRINCIPLES
• ABSORBTION SPECTRO PHOTOMETRY
• BEER LAMBERT LAW
• LAMBERT’S LAW states that when a light falls on a
homogenous substance,intensity of transmitted light
decreases as the distance through the substance increase
• BEER’S LAW states that when a light is transmitted
through a clear substance with a dissolved solute ,the
intensity of transmitted light decreases as the concentration
of the solute increases
CONTD..
• Substances have a specific pattern of absorbing
specific wavelength –Extinction coefficient
Uses two lights of wavelengths
• 660nm –deoxy Hb absorbs ten times as oxy hb
• 940 nm – absorption of oxyHb is greater
• Lab oximeters use 4 wavelengths to measure 4
species of haemoglobin
• It=I oe–Ecd
[Ecd –absorbance]
wavelength
Extinction
coefficient
660nm 940nm
MetHb
Oxy Hb
Deoxy
HbCOHB
DESIGN OF PULSEOXIMETER
• 2 WAVELENGTHS-
660nm [red] & 940nm[infra red]
The ratio of absorbencies at these two wavelengths is
calibrated empirically against direct measurements of
arterial blood oxygen saturation (Sa
O2
) in volunteers, and the
resulting calibration algorithm is stored in a digital
microprocessor within the pulse oximeter.
• LED & PHOTODETECTOR
• Newer types of LED is based on aluminium gallium
arsenide system
• Signal processed in the micro processor
• Senses only the pulsatile flow
Operating principles
• The pulse oximeter estimates SpO2 from the
differential absorption of red and infrared
light in tissue . Reduced hemoglobin
absorbs more light in the red band than
oxyhemoglobin .Oxyhemoglobin absorbs
more light in the infrared band.
reflectance pulseoxymetry
Continuation….
• Reflectance oximetry relies on light that is
reflected (backscattered) to determine
oxygen saturation
• Reflection originates from nonhomogeneity
in the optical path, that is, at the interfaces
between materials with different reflective
indices
Oxygen desaturation
• Saturation is defined as ratio of O2 content
to oxygen capacity of Hb expressed as a
percentage.
• Desaturation leads to Hypoxemia – a
relative deficiency of O2 in arterial blood.
PaO2 < 80mmHg – hypoxemia
CONTD…
• Oxygen saturation will not decrease until
PaO2 is below 85mmHg.
• At SaO2 of 90% PaO2 is already 60mmHg.
• Rough guide for PaO2 between saturation
of 90%-75% is PaO2 = SaO2 - 30.
• SaO2< than 76% is life threatening.
Fractional Saturation
• This is Ratio of oxygenated Haemoglobin to
sum of all haemoglobin in blood.
Fractional saturation = HbO2
-
--------------------------
HbO2+ Hb+ Met Hb +CO Hb
Functional Saturation
• This is a measure of ratio between HbO2 and
sum of oxygenated and reduced Hb.
• Functional Saturation= HbO2
------------
HbO2 + Hb
Types of Hypoxemia
1.Hypoxic hypoxemia
↓ PaO2 ↓ SaO2 – Normal Hb
2.Anaemic hypoxemia
↓Hb , Normal PaO2 & SaO2
3.Toxic hypoxemia
↓SaO2,Normal PaO2
• PaO2 [mmHg] SaO2 [%]
• Normal 97 to ≥80 97 to ≥95
• Hypoxia < 80 < 95
• Mild 60-79 90-94
• Moderate 40 – 59 75 – 89
• Severe <40 < 75
physiology
• Efficient oxygen transport relies on the
ability of hemoglobin to reversibly load and
unload oxygen. The relationship between
oxygen tension and oxygen binding is seen
in the oxyhemoglobin dissociation curve
which plots the hemoglobin oxygen
saturation against the oxygen tension.
Relationship between PaO2 and
SaO2
• Oxygen – haemoglobin dissociation curve
The sigmoid shape of the curve is essential for
physiologic transport
PO2
100
 20
80mmHg
Hb% saturation
97.5
 1.5
96%
60
 20
40mmHg
91
 17
74%
Cardiovascular response to
hypoxemia
SPO2% HR BP Stroke
volume
SVR Response
>80    No
change
Reflex
60-80   No
change
 Direct
<60     Direct
Uses of pulseoximetry
• Monitoring oxygenation
During anaesthesia
in ICU, PACU
during transport
• Monitoring oxygen therapy
• Assesment of perfusion
• Monitoring vascular volume
• Sleep studies -24-h ambulatory recordings of SpO2 is
useful for screening for daytime sleep sequelae associated
with the potential risk of this pathology in OSAS during
social activities.
Continu…
• Emergency department
• Reflectance pulse oximeters can be useful
for assessing fetal status during labor and
delivery by applying a forehead probe
Continuatio…
• Pulse oximetry has been used during
cardiopulmonary resuscitation .Because of
artifacts and lag times, it is more useful in
primary respiratory arrest than in cardiac
arrest. It is useful in assessing oxygenation
during newborn resuscitation
• Out-of-hospital Use
Pulse oximetry is useful in the prehospital
setting, including when transporting
patients by helicopter or ambulance
Controlling Oxygen Administration
Monitoring Peripheral
Circulation
• Pulse oximetry can detect arm positions that
compromise circulation .The pulse oximeter
that is attached to a toe can help to warn of
decreased perfusion at the foot in patients in
the lithotomy position
Contin…
• Locating Arteries:
• Determining Systolic Blood Pressure
• Avoiding Hyperoxemia
• Monitoring Vascular Volume and
Sympathetic Tone
Advantages
• Simple to use
• Non-invasive
• Require no warm up time
• Especially in African &Asian patients
• Cost-effectiveness over ABG
• Accuracy
• Independence from Gases and Vapors
• Fast Response Time
• Continuous Measurements
• Separate Respiratory and Circulatory
Variables
• Convenience
• Fast Start Time
• Tone Modulation
• User-friendliness
• Light Weight and Compactness
• Probe Variety
• Battery Operated
Disadvantages
Decrease in PAO2 before fall in SPO2
Due to the shape of ODC
SPO2 94% - PAO2 75%
Continuat…
• Difficulty in Detecting High Oxygen Partial
Pressures
• Erratic Performance with Dysrhythmias
Limitations
 Shivering patient -motion artefacts
 High intensity ambient light
 Perfusion of the patient
ear probe may be more reliable
 Abnormal pulses –erratic perfomance
 Penumbra effect
- light shunted directly to photo dectector
Contd…
 Carboxy Hb – Produces SpO2 > than true O2
saturation
10-20% in heavy smokers
At 660nm its absorbtion similar to
oxyhaemoglobin ,so over reads
For every 1%COHB 1% increase in
pulseoximeter reading
 Methaemoglobinemia – absorbs equal
amount of red &infra red light - Cause SpO2
to move towards 85%
Endo / exogenous dyes interfere
-Indocyanine green, indigocarmine
,methylene blue
 Blue ,Black ,Green nail polishes
 Diathermy leads to disturbance in monitor
- can be corrected by placing
grounding plate near surgical field
probe as far away from surgical field
& electro cautery
CONTD….
Problems
• False positives and negatives
• Burn injury
• Pressure injuries
Other types of oximetry
• Reflectance oximetry
sense back scattered light
LED placed to the side of photo diode
Dis adv - noise interference & more costly
Adv - signal in low perfusion is better
Contd…
Masimo signal extraction technology
Mixed venous blood o2 saturation
- by special PAC
 Noninvasive brain oximetry
- senses regional o2 saturation in brain[rSO2]
- sensor in forehead emits light of specific
wavelength & measure the reflected light
Transcutaneous oximetry
Sites
• Finger
• Toe
• Ear lobe
• Nose
• Tongue
• Cheek
• esophagus
Other sites
• Pharyngeal pulse oximetry by using a pulse
oximeter attached to a laryngeal mask may
be useful in patients with poor peripheral
perfusion
continua
• Flexible probes may work through the
palm, foot, penis, ankle, lower calf, or even
the arm in infants
Contin…
• Pulse oximetry may be used to monitor fetal
oxygenation during labor by attaching a
reflectance pulse oximetry probe to the
presenting part A disadvantage is that the
probe has to be placed blindly and may be
positioned over a subcutaneous vein or
artery, which will affect the reliability of the
readings .
fixation
• Proper probe placement is crucial for good
performance. A malpositioned probe can
result in false-positive and false-negative
alarms
• The probe should be protected from bright
light
Patient Complications
• Corneal Abrasions
• Pressure and Ischemic Injuries
• Burns
• Electric Shock
• Carbon Monoxide Monitoring
Oxymetry standards
• There must be a means to limit the duration
of continuous operation at temperatures
above 41°C.
• The accuracy must be stated over the range
of 70% to 100% SpO2. If the manufacturer
claims accuracy below 65%, the accuracy
must be stated over the additional range.
Continu…
• If the manufacturer claims accuracy during
motion, this and the test methods used to
establish it must be disclosed in the
instructions for use.
• If the manufacturer claims accuracy during
conditions of low perfusion, this and the
test methods used to establish it must be
disclosed in the instructions for use.
.
Continuatio…
• There must be an indication when the SpO2
or pulse rate data is not current.
• An indication of signal inadequacy must be
provided if the SpO2 or pulse rate value
displayed is potentially incorrect
Continuation…
• If the pulse oximeter is provided with any
physiologic alarm, it must be provided with
an alarm system that monitors for
equipment faults, and there must be an
alarm for low SpO2 that is not less than 85%
SpO2 in the manufacturer-configured alarm
preset. An alarm for high SpO2 is optional.
Take home message
• Remember at SaO2 of 90%, PaO2 is
already 60mmHg
• SaO2 < than 76% is life threatening
•Thank you
Case Study
• Consider Mrs. W., a patient scheduled to undergo an
endoscopy. A complete blood count (CBC) shows her
total hemoglobin level is normal – 15 mg/dl. A pulse
oximeter shows that her oxygen saturation level is
97%.
• To determine her overall oxygen-carrying capacity,
multiply 1.34 ml (the amount of oxygen each gram of
hemoglobin carries) by the hemoglobin level and then
by the SpO2
. 1.34 x 15 x 0.97.
• The total amount of oxygen carried in Mrs. W.’s
hemoglobin is 19.50 ml/dl (hemoglobin oxygen content),
which falls within the normal range of 19-20 ml/dl.
Contd…
• Mr. B., who’s had shortness of breath for two days.
His lungs are clear, and although he has not specific
pulmonary symptoms, he looks malnourished and is
anaemic.
• A spot check with a pulse oximeter shows a SpO2
of
97%, the same as Mrs. W., you discover that Mr. B.’s
hemoglobin oxygen content is 14.30 ml/dl, far below
normal values.
• Because Mr. B. has fewer hemoglobin molecules, the
total amount of oxygen available to his tissues is low,
resulting in shortness of breath. What hemoglobin he
does have is full of oxygen, producing normal and
misleading SpO2
.
• These two cases illustrate a major
cautionary note: Never let normal
SpO2
readings lull you into a false
sense of security! Always interpret
SpO2
values in the context of the
patient’s total hemoglobin level.
Pulseoximetry

Más contenido relacionado

La actualidad más candente

laryngoscope class by Dr Sandeep Singh Jadon ppt.pptx
laryngoscope class by Dr Sandeep Singh Jadon ppt.pptxlaryngoscope class by Dr Sandeep Singh Jadon ppt.pptx
laryngoscope class by Dr Sandeep Singh Jadon ppt.pptxSandeep Singh Jadon
 
ambu bag.pptx
ambu bag.pptxambu bag.pptx
ambu bag.pptxLway1
 
Anaesthesia machine
Anaesthesia machineAnaesthesia machine
Anaesthesia machineimran80
 
CENTRAL VENOUS PRESSURE MONITORING- IACTACON 2018
CENTRAL VENOUS PRESSURE MONITORING- IACTACON 2018CENTRAL VENOUS PRESSURE MONITORING- IACTACON 2018
CENTRAL VENOUS PRESSURE MONITORING- IACTACON 2018Dr. Harshil Joshi
 
Non Invasive and Invasive Blood pressure monitoring RRT
Non Invasive and Invasive Blood pressure monitoring RRTNon Invasive and Invasive Blood pressure monitoring RRT
Non Invasive and Invasive Blood pressure monitoring RRTRanjith Thampi
 
Oral Airway Presentation
Oral Airway PresentationOral Airway Presentation
Oral Airway PresentationAdam Divine
 
#Part 1 venturi mask
#Part 1 venturi mask#Part 1 venturi mask
#Part 1 venturi maskNisar Arain
 
Venturi Mask for Oxygen Delivery Administration
Venturi Mask for Oxygen Delivery AdministrationVenturi Mask for Oxygen Delivery Administration
Venturi Mask for Oxygen Delivery AdministrationSmart Medical Buyer
 
Endotracheal tubes
Endotracheal tubesEndotracheal tubes
Endotracheal tubesPratik Kumar
 
Facemask , oral and nasal airways
Facemask , oral and nasal airwaysFacemask , oral and nasal airways
Facemask , oral and nasal airwaysDR SHADAB KAMAL
 
Central venous pressure monitoring
Central venous pressure monitoring Central venous pressure monitoring
Central venous pressure monitoring DR .PALLAVI PATHANIA
 
Bag and Mask Ventilation By Sakun Rasaily @Ram K Dhamala
Bag and Mask Ventilation By Sakun Rasaily @Ram K DhamalaBag and Mask Ventilation By Sakun Rasaily @Ram K Dhamala
Bag and Mask Ventilation By Sakun Rasaily @Ram K Dhamalaramdhamala11
 
Central venous pressure
Central venous pressureCentral venous pressure
Central venous pressureDavis Kurian
 
Supraglottic airway device
Supraglottic airway deviceSupraglottic airway device
Supraglottic airway deviceDebojyoti Dutta
 

La actualidad más candente (20)

laryngoscope class by Dr Sandeep Singh Jadon ppt.pptx
laryngoscope class by Dr Sandeep Singh Jadon ppt.pptxlaryngoscope class by Dr Sandeep Singh Jadon ppt.pptx
laryngoscope class by Dr Sandeep Singh Jadon ppt.pptx
 
Laryngeal mask-airway
Laryngeal mask-airwayLaryngeal mask-airway
Laryngeal mask-airway
 
ambu bag.pptx
ambu bag.pptxambu bag.pptx
ambu bag.pptx
 
Anaesthesia machine
Anaesthesia machineAnaesthesia machine
Anaesthesia machine
 
Oropharyngeal Airway.pptx
Oropharyngeal Airway.pptxOropharyngeal Airway.pptx
Oropharyngeal Airway.pptx
 
CENTRAL VENOUS PRESSURE MONITORING- IACTACON 2018
CENTRAL VENOUS PRESSURE MONITORING- IACTACON 2018CENTRAL VENOUS PRESSURE MONITORING- IACTACON 2018
CENTRAL VENOUS PRESSURE MONITORING- IACTACON 2018
 
Non Invasive and Invasive Blood pressure monitoring RRT
Non Invasive and Invasive Blood pressure monitoring RRTNon Invasive and Invasive Blood pressure monitoring RRT
Non Invasive and Invasive Blood pressure monitoring RRT
 
Oral Airway Presentation
Oral Airway PresentationOral Airway Presentation
Oral Airway Presentation
 
#Part 1 venturi mask
#Part 1 venturi mask#Part 1 venturi mask
#Part 1 venturi mask
 
Venturi Mask for Oxygen Delivery Administration
Venturi Mask for Oxygen Delivery AdministrationVenturi Mask for Oxygen Delivery Administration
Venturi Mask for Oxygen Delivery Administration
 
CAPNOGRAPHY
CAPNOGRAPHYCAPNOGRAPHY
CAPNOGRAPHY
 
Machine boyles
Machine boylesMachine boyles
Machine boyles
 
Endotracheal tubes
Endotracheal tubesEndotracheal tubes
Endotracheal tubes
 
supraglottic airway devices
supraglottic airway devicessupraglottic airway devices
supraglottic airway devices
 
Facemask , oral and nasal airways
Facemask , oral and nasal airwaysFacemask , oral and nasal airways
Facemask , oral and nasal airways
 
Central venous pressure monitoring
Central venous pressure monitoring Central venous pressure monitoring
Central venous pressure monitoring
 
HME
HME HME
HME
 
Bag and Mask Ventilation By Sakun Rasaily @Ram K Dhamala
Bag and Mask Ventilation By Sakun Rasaily @Ram K DhamalaBag and Mask Ventilation By Sakun Rasaily @Ram K Dhamala
Bag and Mask Ventilation By Sakun Rasaily @Ram K Dhamala
 
Central venous pressure
Central venous pressureCentral venous pressure
Central venous pressure
 
Supraglottic airway device
Supraglottic airway deviceSupraglottic airway device
Supraglottic airway device
 

Destacado

Pulse oximeter - SpO2
Pulse oximeter - SpO2Pulse oximeter - SpO2
Pulse oximeter - SpO2Nguyễn Linh
 
Pulse oximetry slidefinal abc
Pulse oximetry slidefinal abcPulse oximetry slidefinal abc
Pulse oximetry slidefinal abcJibi Life
 
Capnography in ems.ppt
Capnography in ems.pptCapnography in ems.ppt
Capnography in ems.pptDo Harm
 
08 capnometry and pulse oximetry
08 capnometry and pulse oximetry08 capnometry and pulse oximetry
08 capnometry and pulse oximetryDang Thanh Tuan
 
Cerebral oximetry as a real time monitoring tool to assess quality
Cerebral oximetry as a real time monitoring tool to assess qualityCerebral oximetry as a real time monitoring tool to assess quality
Cerebral oximetry as a real time monitoring tool to assess qualityEmeka Onuorah
 
Insight into Monitoring of the surgical patient
Insight into Monitoring of the surgical patientInsight into Monitoring of the surgical patient
Insight into Monitoring of the surgical patientlmdm11
 
Capnography vs plethysmography
Capnography vs plethysmographyCapnography vs plethysmography
Capnography vs plethysmographySourav Mondal
 

Destacado (17)

Pulse Oximetry
Pulse OximetryPulse Oximetry
Pulse Oximetry
 
Pulseoximetry
PulseoximetryPulseoximetry
Pulseoximetry
 
use of pulse oximetry
use of pulse oximetryuse of pulse oximetry
use of pulse oximetry
 
Pulse oximeter - SpO2
Pulse oximeter - SpO2Pulse oximeter - SpO2
Pulse oximeter - SpO2
 
Pulse oximetry slidefinal abc
Pulse oximetry slidefinal abcPulse oximetry slidefinal abc
Pulse oximetry slidefinal abc
 
pulse oximetry final
pulse oximetry finalpulse oximetry final
pulse oximetry final
 
Capnography in ems.ppt
Capnography in ems.pptCapnography in ems.ppt
Capnography in ems.ppt
 
08 capnometry and pulse oximetry
08 capnometry and pulse oximetry08 capnometry and pulse oximetry
08 capnometry and pulse oximetry
 
Cerebral oximetry as a real time monitoring tool to assess quality
Cerebral oximetry as a real time monitoring tool to assess qualityCerebral oximetry as a real time monitoring tool to assess quality
Cerebral oximetry as a real time monitoring tool to assess quality
 
Capnography
CapnographyCapnography
Capnography
 
Insight into Monitoring of the surgical patient
Insight into Monitoring of the surgical patientInsight into Monitoring of the surgical patient
Insight into Monitoring of the surgical patient
 
Spectrum analyzer
Spectrum analyzerSpectrum analyzer
Spectrum analyzer
 
Capnography vs plethysmography
Capnography vs plethysmographyCapnography vs plethysmography
Capnography vs plethysmography
 
Pulse oximeter
Pulse oximeterPulse oximeter
Pulse oximeter
 
Fetal circulation
Fetal circulationFetal circulation
Fetal circulation
 
VAPORIZERS!
VAPORIZERS!VAPORIZERS!
VAPORIZERS!
 
Pulseoximetry fianl
Pulseoximetry fianlPulseoximetry fianl
Pulseoximetry fianl
 

Similar a Pulseoximetry

Similar a Pulseoximetry (20)

Ppt0000002
Ppt0000002Ppt0000002
Ppt0000002
 
Pulse oximetry capnography
Pulse oximetry capnography Pulse oximetry capnography
Pulse oximetry capnography
 
Pulse oximetry
Pulse oximetryPulse oximetry
Pulse oximetry
 
Pulse Oxymetry , Inspired & Expired Gas Monitoring
Pulse Oxymetry ,  Inspired & Expired Gas MonitoringPulse Oxymetry ,  Inspired & Expired Gas Monitoring
Pulse Oxymetry , Inspired & Expired Gas Monitoring
 
Capnography
Capnography Capnography
Capnography
 
Monitoring Modality in anesthesia
Monitoring Modality  in anesthesiaMonitoring Modality  in anesthesia
Monitoring Modality in anesthesia
 
BASIC ANAESTHETIC MONITORING
BASIC ANAESTHETIC MONITORING BASIC ANAESTHETIC MONITORING
BASIC ANAESTHETIC MONITORING
 
Pulse oximetry
Pulse oximetryPulse oximetry
Pulse oximetry
 
Capnography
Capnography Capnography
Capnography
 
Oxygen reserve index
Oxygen reserve indexOxygen reserve index
Oxygen reserve index
 
Nibp and oxygen analyser
Nibp and oxygen analyserNibp and oxygen analyser
Nibp and oxygen analyser
 
Monitoring intraoperatif dan terapi cairan.pptx
Monitoring intraoperatif dan terapi cairan.pptxMonitoring intraoperatif dan terapi cairan.pptx
Monitoring intraoperatif dan terapi cairan.pptx
 
Oximeter The Life Saving Instrument
Oximeter The Life Saving InstrumentOximeter The Life Saving Instrument
Oximeter The Life Saving Instrument
 
4.Oxygen therapy.pptx
4.Oxygen therapy.pptx4.Oxygen therapy.pptx
4.Oxygen therapy.pptx
 
Op oxygen sensors
Op oxygen sensorsOp oxygen sensors
Op oxygen sensors
 
Oxygen therapy
Oxygen therapyOxygen therapy
Oxygen therapy
 
photoplethysmogram
photoplethysmogram photoplethysmogram
photoplethysmogram
 
Respiratory monitoring
Respiratory monitoringRespiratory monitoring
Respiratory monitoring
 
respiratory monitoring backup.ppt
respiratory monitoring backup.pptrespiratory monitoring backup.ppt
respiratory monitoring backup.ppt
 
Moderate sedation monitoring
Moderate sedation monitoring Moderate sedation monitoring
Moderate sedation monitoring
 

Más de drriyas03

Acute renal failure patho physiology & anaesthetic management
Acute renal failure patho physiology & anaesthetic managementAcute renal failure patho physiology & anaesthetic management
Acute renal failure patho physiology & anaesthetic managementdrriyas03
 
Hypothermia and anaesthesia implication
Hypothermia and anaesthesia implicationHypothermia and anaesthesia implication
Hypothermia and anaesthesia implicationdrriyas03
 
Ionotropes and vasopressors
Ionotropes and vasopressorsIonotropes and vasopressors
Ionotropes and vasopressorsdrriyas03
 
Sepsis 2012 andbeyond
Sepsis 2012 andbeyondSepsis 2012 andbeyond
Sepsis 2012 andbeyonddrriyas03
 
Fluid management in the paediatric patient anaesthetist consideration...
Fluid management in the paediatric patient anaesthetist consideration...Fluid management in the paediatric patient anaesthetist consideration...
Fluid management in the paediatric patient anaesthetist consideration...drriyas03
 
Airway assessment and pedictors of difficult airway....must know for anaesthe...
Airway assessment and pedictors of difficult airway....must know for anaesthe...Airway assessment and pedictors of difficult airway....must know for anaesthe...
Airway assessment and pedictors of difficult airway....must know for anaesthe...drriyas03
 
18 main arteries & veins of neck for anaesthesia
18 main arteries & veins of neck for anaesthesia18 main arteries & veins of neck for anaesthesia
18 main arteries & veins of neck for anaesthesiadrriyas03
 
Cardiac cycle
Cardiac cycleCardiac cycle
Cardiac cycledrriyas03
 
Ppt cvs phsiology a small review for anaesthetist
Ppt cvs phsiology a small review for anaesthetistPpt cvs phsiology a small review for anaesthetist
Ppt cvs phsiology a small review for anaesthetistdrriyas03
 

Más de drriyas03 (10)

Antifungals
AntifungalsAntifungals
Antifungals
 
Acute renal failure patho physiology & anaesthetic management
Acute renal failure patho physiology & anaesthetic managementAcute renal failure patho physiology & anaesthetic management
Acute renal failure patho physiology & anaesthetic management
 
Hypothermia and anaesthesia implication
Hypothermia and anaesthesia implicationHypothermia and anaesthesia implication
Hypothermia and anaesthesia implication
 
Ionotropes and vasopressors
Ionotropes and vasopressorsIonotropes and vasopressors
Ionotropes and vasopressors
 
Sepsis 2012 andbeyond
Sepsis 2012 andbeyondSepsis 2012 andbeyond
Sepsis 2012 andbeyond
 
Fluid management in the paediatric patient anaesthetist consideration...
Fluid management in the paediatric patient anaesthetist consideration...Fluid management in the paediatric patient anaesthetist consideration...
Fluid management in the paediatric patient anaesthetist consideration...
 
Airway assessment and pedictors of difficult airway....must know for anaesthe...
Airway assessment and pedictors of difficult airway....must know for anaesthe...Airway assessment and pedictors of difficult airway....must know for anaesthe...
Airway assessment and pedictors of difficult airway....must know for anaesthe...
 
18 main arteries & veins of neck for anaesthesia
18 main arteries & veins of neck for anaesthesia18 main arteries & veins of neck for anaesthesia
18 main arteries & veins of neck for anaesthesia
 
Cardiac cycle
Cardiac cycleCardiac cycle
Cardiac cycle
 
Ppt cvs phsiology a small review for anaesthetist
Ppt cvs phsiology a small review for anaesthetistPpt cvs phsiology a small review for anaesthetist
Ppt cvs phsiology a small review for anaesthetist
 

Último

ISYU TUNGKOL SA SEKSWLADIDA (ISSUE ABOUT SEXUALITY
ISYU TUNGKOL SA SEKSWLADIDA (ISSUE ABOUT SEXUALITYISYU TUNGKOL SA SEKSWLADIDA (ISSUE ABOUT SEXUALITY
ISYU TUNGKOL SA SEKSWLADIDA (ISSUE ABOUT SEXUALITYKayeClaireEstoconing
 
ACC 2024 Chronicles. Cardiology. Exam.pdf
ACC 2024 Chronicles. Cardiology. Exam.pdfACC 2024 Chronicles. Cardiology. Exam.pdf
ACC 2024 Chronicles. Cardiology. Exam.pdfSpandanaRallapalli
 
Visit to a blind student's school🧑‍🦯🧑‍🦯(community medicine)
Visit to a blind student's school🧑‍🦯🧑‍🦯(community medicine)Visit to a blind student's school🧑‍🦯🧑‍🦯(community medicine)
Visit to a blind student's school🧑‍🦯🧑‍🦯(community medicine)lakshayb543
 
Difference Between Search & Browse Methods in Odoo 17
Difference Between Search & Browse Methods in Odoo 17Difference Between Search & Browse Methods in Odoo 17
Difference Between Search & Browse Methods in Odoo 17Celine George
 
Procuring digital preservation CAN be quick and painless with our new dynamic...
Procuring digital preservation CAN be quick and painless with our new dynamic...Procuring digital preservation CAN be quick and painless with our new dynamic...
Procuring digital preservation CAN be quick and painless with our new dynamic...Jisc
 
THEORIES OF ORGANIZATION-PUBLIC ADMINISTRATION
THEORIES OF ORGANIZATION-PUBLIC ADMINISTRATIONTHEORIES OF ORGANIZATION-PUBLIC ADMINISTRATION
THEORIES OF ORGANIZATION-PUBLIC ADMINISTRATIONHumphrey A Beña
 
Earth Day Presentation wow hello nice great
Earth Day Presentation wow hello nice greatEarth Day Presentation wow hello nice great
Earth Day Presentation wow hello nice greatYousafMalik24
 
Computed Fields and api Depends in the Odoo 17
Computed Fields and api Depends in the Odoo 17Computed Fields and api Depends in the Odoo 17
Computed Fields and api Depends in the Odoo 17Celine George
 
Full Stack Web Development Course for Beginners
Full Stack Web Development Course  for BeginnersFull Stack Web Development Course  for Beginners
Full Stack Web Development Course for BeginnersSabitha Banu
 
Grade 9 Q4-MELC1-Active and Passive Voice.pptx
Grade 9 Q4-MELC1-Active and Passive Voice.pptxGrade 9 Q4-MELC1-Active and Passive Voice.pptx
Grade 9 Q4-MELC1-Active and Passive Voice.pptxChelloAnnAsuncion2
 
Like-prefer-love -hate+verb+ing & silent letters & citizenship text.pdf
Like-prefer-love -hate+verb+ing & silent letters & citizenship text.pdfLike-prefer-love -hate+verb+ing & silent letters & citizenship text.pdf
Like-prefer-love -hate+verb+ing & silent letters & citizenship text.pdfMr Bounab Samir
 
ANG SEKTOR NG agrikultura.pptx QUARTER 4
ANG SEKTOR NG agrikultura.pptx QUARTER 4ANG SEKTOR NG agrikultura.pptx QUARTER 4
ANG SEKTOR NG agrikultura.pptx QUARTER 4MiaBumagat1
 
How to do quick user assign in kanban in Odoo 17 ERP
How to do quick user assign in kanban in Odoo 17 ERPHow to do quick user assign in kanban in Odoo 17 ERP
How to do quick user assign in kanban in Odoo 17 ERPCeline George
 
USPS® Forced Meter Migration - How to Know if Your Postage Meter Will Soon be...
USPS® Forced Meter Migration - How to Know if Your Postage Meter Will Soon be...USPS® Forced Meter Migration - How to Know if Your Postage Meter Will Soon be...
USPS® Forced Meter Migration - How to Know if Your Postage Meter Will Soon be...Postal Advocate Inc.
 
Incoming and Outgoing Shipments in 3 STEPS Using Odoo 17
Incoming and Outgoing Shipments in 3 STEPS Using Odoo 17Incoming and Outgoing Shipments in 3 STEPS Using Odoo 17
Incoming and Outgoing Shipments in 3 STEPS Using Odoo 17Celine George
 
Judging the Relevance and worth of ideas part 2.pptx
Judging the Relevance  and worth of ideas part 2.pptxJudging the Relevance  and worth of ideas part 2.pptx
Judging the Relevance and worth of ideas part 2.pptxSherlyMaeNeri
 
DATA STRUCTURE AND ALGORITHM for beginners
DATA STRUCTURE AND ALGORITHM for beginnersDATA STRUCTURE AND ALGORITHM for beginners
DATA STRUCTURE AND ALGORITHM for beginnersSabitha Banu
 
4.18.24 Movement Legacies, Reflection, and Review.pptx
4.18.24 Movement Legacies, Reflection, and Review.pptx4.18.24 Movement Legacies, Reflection, and Review.pptx
4.18.24 Movement Legacies, Reflection, and Review.pptxmary850239
 

Último (20)

ISYU TUNGKOL SA SEKSWLADIDA (ISSUE ABOUT SEXUALITY
ISYU TUNGKOL SA SEKSWLADIDA (ISSUE ABOUT SEXUALITYISYU TUNGKOL SA SEKSWLADIDA (ISSUE ABOUT SEXUALITY
ISYU TUNGKOL SA SEKSWLADIDA (ISSUE ABOUT SEXUALITY
 
ACC 2024 Chronicles. Cardiology. Exam.pdf
ACC 2024 Chronicles. Cardiology. Exam.pdfACC 2024 Chronicles. Cardiology. Exam.pdf
ACC 2024 Chronicles. Cardiology. Exam.pdf
 
Raw materials used in Herbal Cosmetics.pptx
Raw materials used in Herbal Cosmetics.pptxRaw materials used in Herbal Cosmetics.pptx
Raw materials used in Herbal Cosmetics.pptx
 
Visit to a blind student's school🧑‍🦯🧑‍🦯(community medicine)
Visit to a blind student's school🧑‍🦯🧑‍🦯(community medicine)Visit to a blind student's school🧑‍🦯🧑‍🦯(community medicine)
Visit to a blind student's school🧑‍🦯🧑‍🦯(community medicine)
 
Difference Between Search & Browse Methods in Odoo 17
Difference Between Search & Browse Methods in Odoo 17Difference Between Search & Browse Methods in Odoo 17
Difference Between Search & Browse Methods in Odoo 17
 
Procuring digital preservation CAN be quick and painless with our new dynamic...
Procuring digital preservation CAN be quick and painless with our new dynamic...Procuring digital preservation CAN be quick and painless with our new dynamic...
Procuring digital preservation CAN be quick and painless with our new dynamic...
 
THEORIES OF ORGANIZATION-PUBLIC ADMINISTRATION
THEORIES OF ORGANIZATION-PUBLIC ADMINISTRATIONTHEORIES OF ORGANIZATION-PUBLIC ADMINISTRATION
THEORIES OF ORGANIZATION-PUBLIC ADMINISTRATION
 
Earth Day Presentation wow hello nice great
Earth Day Presentation wow hello nice greatEarth Day Presentation wow hello nice great
Earth Day Presentation wow hello nice great
 
Computed Fields and api Depends in the Odoo 17
Computed Fields and api Depends in the Odoo 17Computed Fields and api Depends in the Odoo 17
Computed Fields and api Depends in the Odoo 17
 
Full Stack Web Development Course for Beginners
Full Stack Web Development Course  for BeginnersFull Stack Web Development Course  for Beginners
Full Stack Web Development Course for Beginners
 
Grade 9 Q4-MELC1-Active and Passive Voice.pptx
Grade 9 Q4-MELC1-Active and Passive Voice.pptxGrade 9 Q4-MELC1-Active and Passive Voice.pptx
Grade 9 Q4-MELC1-Active and Passive Voice.pptx
 
Like-prefer-love -hate+verb+ing & silent letters & citizenship text.pdf
Like-prefer-love -hate+verb+ing & silent letters & citizenship text.pdfLike-prefer-love -hate+verb+ing & silent letters & citizenship text.pdf
Like-prefer-love -hate+verb+ing & silent letters & citizenship text.pdf
 
ANG SEKTOR NG agrikultura.pptx QUARTER 4
ANG SEKTOR NG agrikultura.pptx QUARTER 4ANG SEKTOR NG agrikultura.pptx QUARTER 4
ANG SEKTOR NG agrikultura.pptx QUARTER 4
 
How to do quick user assign in kanban in Odoo 17 ERP
How to do quick user assign in kanban in Odoo 17 ERPHow to do quick user assign in kanban in Odoo 17 ERP
How to do quick user assign in kanban in Odoo 17 ERP
 
USPS® Forced Meter Migration - How to Know if Your Postage Meter Will Soon be...
USPS® Forced Meter Migration - How to Know if Your Postage Meter Will Soon be...USPS® Forced Meter Migration - How to Know if Your Postage Meter Will Soon be...
USPS® Forced Meter Migration - How to Know if Your Postage Meter Will Soon be...
 
Incoming and Outgoing Shipments in 3 STEPS Using Odoo 17
Incoming and Outgoing Shipments in 3 STEPS Using Odoo 17Incoming and Outgoing Shipments in 3 STEPS Using Odoo 17
Incoming and Outgoing Shipments in 3 STEPS Using Odoo 17
 
Judging the Relevance and worth of ideas part 2.pptx
Judging the Relevance  and worth of ideas part 2.pptxJudging the Relevance  and worth of ideas part 2.pptx
Judging the Relevance and worth of ideas part 2.pptx
 
DATA STRUCTURE AND ALGORITHM for beginners
DATA STRUCTURE AND ALGORITHM for beginnersDATA STRUCTURE AND ALGORITHM for beginners
DATA STRUCTURE AND ALGORITHM for beginners
 
4.18.24 Movement Legacies, Reflection, and Review.pptx
4.18.24 Movement Legacies, Reflection, and Review.pptx4.18.24 Movement Legacies, Reflection, and Review.pptx
4.18.24 Movement Legacies, Reflection, and Review.pptx
 
TataKelola dan KamSiber Kecerdasan Buatan v022.pdf
TataKelola dan KamSiber Kecerdasan Buatan v022.pdfTataKelola dan KamSiber Kecerdasan Buatan v022.pdf
TataKelola dan KamSiber Kecerdasan Buatan v022.pdf
 

Pulseoximetry

  • 1. PULSEOXIMETRY DR.RIYAS A Dr SMCSI MEDICAL COLLEGE KARAKONAM
  • 3. Introduction • The maintenance of optimal O2 delivery is the core concern during anaesthesia • “Oxygen lack not only stops the machine but wrecks the machinery ”– J.S. Haldane. • Monitoring of oxygenation using pulseoximeter avoids many catastrophies
  • 4. DEFINITION • Pulse oximetry, sometimes called the fifth vital sign, is a noninvasive method of measuring hemoglobin saturation (SpO2) by using a light signal transmitted through tissue • A low SpO2 can provide warning of hypoxemia before other signs such as cyanosis or a change in heart rate are observed
  • 5. HISTORY • MATHEES- father of oximetry • 20 papers in1934 –1944 • HERTZMAN 1937 –use of photoelectric finger plethsmography • 1975 –concept of pulse oximetry –Japan • YELDERMAN &NEW -1983 –Nellcor pulse oximeter
  • 6. Continuation… • The American Society of Anesthesiologists (ASA) and American Association of Nurse Anesthetists have made assessment of oxygenation a standard for intraoperative and postoperative monitoring. In 2005, an audible alarm for the pulse oximeter was added to the ASA monitoring standard
  • 7. PRINCIPLES • ABSORBTION SPECTRO PHOTOMETRY • BEER LAMBERT LAW • LAMBERT’S LAW states that when a light falls on a homogenous substance,intensity of transmitted light decreases as the distance through the substance increase • BEER’S LAW states that when a light is transmitted through a clear substance with a dissolved solute ,the intensity of transmitted light decreases as the concentration of the solute increases
  • 8. CONTD.. • Substances have a specific pattern of absorbing specific wavelength –Extinction coefficient Uses two lights of wavelengths • 660nm –deoxy Hb absorbs ten times as oxy hb • 940 nm – absorption of oxyHb is greater • Lab oximeters use 4 wavelengths to measure 4 species of haemoglobin • It=I oe–Ecd [Ecd –absorbance]
  • 10. DESIGN OF PULSEOXIMETER • 2 WAVELENGTHS- 660nm [red] & 940nm[infra red] The ratio of absorbencies at these two wavelengths is calibrated empirically against direct measurements of arterial blood oxygen saturation (Sa O2 ) in volunteers, and the resulting calibration algorithm is stored in a digital microprocessor within the pulse oximeter. • LED & PHOTODETECTOR • Newer types of LED is based on aluminium gallium arsenide system • Signal processed in the micro processor • Senses only the pulsatile flow
  • 11. Operating principles • The pulse oximeter estimates SpO2 from the differential absorption of red and infrared light in tissue . Reduced hemoglobin absorbs more light in the red band than oxyhemoglobin .Oxyhemoglobin absorbs more light in the infrared band.
  • 13. Continuation…. • Reflectance oximetry relies on light that is reflected (backscattered) to determine oxygen saturation • Reflection originates from nonhomogeneity in the optical path, that is, at the interfaces between materials with different reflective indices
  • 14.
  • 15.
  • 16. Oxygen desaturation • Saturation is defined as ratio of O2 content to oxygen capacity of Hb expressed as a percentage. • Desaturation leads to Hypoxemia – a relative deficiency of O2 in arterial blood. PaO2 < 80mmHg – hypoxemia
  • 17. CONTD… • Oxygen saturation will not decrease until PaO2 is below 85mmHg. • At SaO2 of 90% PaO2 is already 60mmHg. • Rough guide for PaO2 between saturation of 90%-75% is PaO2 = SaO2 - 30. • SaO2< than 76% is life threatening.
  • 18. Fractional Saturation • This is Ratio of oxygenated Haemoglobin to sum of all haemoglobin in blood. Fractional saturation = HbO2 - -------------------------- HbO2+ Hb+ Met Hb +CO Hb
  • 19. Functional Saturation • This is a measure of ratio between HbO2 and sum of oxygenated and reduced Hb. • Functional Saturation= HbO2 ------------ HbO2 + Hb
  • 20. Types of Hypoxemia 1.Hypoxic hypoxemia ↓ PaO2 ↓ SaO2 – Normal Hb 2.Anaemic hypoxemia ↓Hb , Normal PaO2 & SaO2 3.Toxic hypoxemia ↓SaO2,Normal PaO2
  • 21. • PaO2 [mmHg] SaO2 [%] • Normal 97 to ≥80 97 to ≥95 • Hypoxia < 80 < 95 • Mild 60-79 90-94 • Moderate 40 – 59 75 – 89 • Severe <40 < 75
  • 22. physiology • Efficient oxygen transport relies on the ability of hemoglobin to reversibly load and unload oxygen. The relationship between oxygen tension and oxygen binding is seen in the oxyhemoglobin dissociation curve which plots the hemoglobin oxygen saturation against the oxygen tension.
  • 23. Relationship between PaO2 and SaO2 • Oxygen – haemoglobin dissociation curve The sigmoid shape of the curve is essential for physiologic transport
  • 24. PO2 100  20 80mmHg Hb% saturation 97.5  1.5 96% 60  20 40mmHg 91  17 74%
  • 25. Cardiovascular response to hypoxemia SPO2% HR BP Stroke volume SVR Response >80    No change Reflex 60-80   No change  Direct <60     Direct
  • 26. Uses of pulseoximetry • Monitoring oxygenation During anaesthesia in ICU, PACU during transport • Monitoring oxygen therapy • Assesment of perfusion • Monitoring vascular volume • Sleep studies -24-h ambulatory recordings of SpO2 is useful for screening for daytime sleep sequelae associated with the potential risk of this pathology in OSAS during social activities.
  • 27. Continu… • Emergency department • Reflectance pulse oximeters can be useful for assessing fetal status during labor and delivery by applying a forehead probe
  • 28. Continuatio… • Pulse oximetry has been used during cardiopulmonary resuscitation .Because of artifacts and lag times, it is more useful in primary respiratory arrest than in cardiac arrest. It is useful in assessing oxygenation during newborn resuscitation
  • 29. • Out-of-hospital Use Pulse oximetry is useful in the prehospital setting, including when transporting patients by helicopter or ambulance Controlling Oxygen Administration
  • 30. Monitoring Peripheral Circulation • Pulse oximetry can detect arm positions that compromise circulation .The pulse oximeter that is attached to a toe can help to warn of decreased perfusion at the foot in patients in the lithotomy position
  • 31. Contin… • Locating Arteries: • Determining Systolic Blood Pressure • Avoiding Hyperoxemia • Monitoring Vascular Volume and Sympathetic Tone
  • 32. Advantages • Simple to use • Non-invasive • Require no warm up time • Especially in African &Asian patients • Cost-effectiveness over ABG
  • 33. • Accuracy • Independence from Gases and Vapors • Fast Response Time • Continuous Measurements • Separate Respiratory and Circulatory Variables • Convenience
  • 34. • Fast Start Time • Tone Modulation • User-friendliness • Light Weight and Compactness • Probe Variety • Battery Operated
  • 35. Disadvantages Decrease in PAO2 before fall in SPO2 Due to the shape of ODC SPO2 94% - PAO2 75%
  • 36. Continuat… • Difficulty in Detecting High Oxygen Partial Pressures • Erratic Performance with Dysrhythmias
  • 37. Limitations  Shivering patient -motion artefacts  High intensity ambient light  Perfusion of the patient ear probe may be more reliable  Abnormal pulses –erratic perfomance  Penumbra effect - light shunted directly to photo dectector
  • 38. Contd…  Carboxy Hb – Produces SpO2 > than true O2 saturation 10-20% in heavy smokers At 660nm its absorbtion similar to oxyhaemoglobin ,so over reads For every 1%COHB 1% increase in pulseoximeter reading  Methaemoglobinemia – absorbs equal amount of red &infra red light - Cause SpO2 to move towards 85%
  • 39. Endo / exogenous dyes interfere -Indocyanine green, indigocarmine ,methylene blue  Blue ,Black ,Green nail polishes  Diathermy leads to disturbance in monitor - can be corrected by placing grounding plate near surgical field probe as far away from surgical field & electro cautery CONTD….
  • 40. Problems • False positives and negatives • Burn injury • Pressure injuries
  • 41. Other types of oximetry • Reflectance oximetry sense back scattered light LED placed to the side of photo diode Dis adv - noise interference & more costly Adv - signal in low perfusion is better
  • 42. Contd… Masimo signal extraction technology Mixed venous blood o2 saturation - by special PAC  Noninvasive brain oximetry - senses regional o2 saturation in brain[rSO2] - sensor in forehead emits light of specific wavelength & measure the reflected light Transcutaneous oximetry
  • 43. Sites • Finger • Toe • Ear lobe • Nose • Tongue • Cheek • esophagus
  • 44. Other sites • Pharyngeal pulse oximetry by using a pulse oximeter attached to a laryngeal mask may be useful in patients with poor peripheral perfusion
  • 45. continua • Flexible probes may work through the palm, foot, penis, ankle, lower calf, or even the arm in infants
  • 46. Contin… • Pulse oximetry may be used to monitor fetal oxygenation during labor by attaching a reflectance pulse oximetry probe to the presenting part A disadvantage is that the probe has to be placed blindly and may be positioned over a subcutaneous vein or artery, which will affect the reliability of the readings .
  • 47. fixation • Proper probe placement is crucial for good performance. A malpositioned probe can result in false-positive and false-negative alarms • The probe should be protected from bright light
  • 48. Patient Complications • Corneal Abrasions • Pressure and Ischemic Injuries • Burns • Electric Shock • Carbon Monoxide Monitoring
  • 49. Oxymetry standards • There must be a means to limit the duration of continuous operation at temperatures above 41°C. • The accuracy must be stated over the range of 70% to 100% SpO2. If the manufacturer claims accuracy below 65%, the accuracy must be stated over the additional range.
  • 50. Continu… • If the manufacturer claims accuracy during motion, this and the test methods used to establish it must be disclosed in the instructions for use. • If the manufacturer claims accuracy during conditions of low perfusion, this and the test methods used to establish it must be disclosed in the instructions for use. .
  • 51. Continuatio… • There must be an indication when the SpO2 or pulse rate data is not current. • An indication of signal inadequacy must be provided if the SpO2 or pulse rate value displayed is potentially incorrect
  • 52. Continuation… • If the pulse oximeter is provided with any physiologic alarm, it must be provided with an alarm system that monitors for equipment faults, and there must be an alarm for low SpO2 that is not less than 85% SpO2 in the manufacturer-configured alarm preset. An alarm for high SpO2 is optional.
  • 53. Take home message • Remember at SaO2 of 90%, PaO2 is already 60mmHg • SaO2 < than 76% is life threatening
  • 55. Case Study • Consider Mrs. W., a patient scheduled to undergo an endoscopy. A complete blood count (CBC) shows her total hemoglobin level is normal – 15 mg/dl. A pulse oximeter shows that her oxygen saturation level is 97%. • To determine her overall oxygen-carrying capacity, multiply 1.34 ml (the amount of oxygen each gram of hemoglobin carries) by the hemoglobin level and then by the SpO2 . 1.34 x 15 x 0.97. • The total amount of oxygen carried in Mrs. W.’s hemoglobin is 19.50 ml/dl (hemoglobin oxygen content), which falls within the normal range of 19-20 ml/dl.
  • 56. Contd… • Mr. B., who’s had shortness of breath for two days. His lungs are clear, and although he has not specific pulmonary symptoms, he looks malnourished and is anaemic. • A spot check with a pulse oximeter shows a SpO2 of 97%, the same as Mrs. W., you discover that Mr. B.’s hemoglobin oxygen content is 14.30 ml/dl, far below normal values. • Because Mr. B. has fewer hemoglobin molecules, the total amount of oxygen available to his tissues is low, resulting in shortness of breath. What hemoglobin he does have is full of oxygen, producing normal and misleading SpO2 .
  • 57. • These two cases illustrate a major cautionary note: Never let normal SpO2 readings lull you into a false sense of security! Always interpret SpO2 values in the context of the patient’s total hemoglobin level.