TONOMETRY
H.Kangari O.D.
Different technique in measurement of the
Intra Ocular Pressure(IOP):
 Goldmann Applination Tonometry
 Perkins Tonometer
 Non-Contact Tonometry
 Digital Evaluation


TONOMETRY/H.KangariO.D.

2
Goldmann Applination Tonometry is the
most accurate method of measuring
Intraocular Pressure.
 Measuring the intraocular pressure is
important in diagnosis and management
of Glaucoma and Ocular Hypertension.


TONOMETRY/H.KangariO.D.

3
CONTRAINDICATIONS TO IOP
MEASUREMENT WITH GOLDMAN
TECHNIQUE
1)Active Infection
 2)Patients with recurrent corneal erosions
 3) Corneal abrasions that are not totally
healed


TONOMETRY/H.KangariO.D.

4


The Intraocular
Pressure is
determined by the
amount of force
required to flatten a
constant corneal
surface area

TONOMETRY/H.KangariO.D.

5
TONOMETRY/H.KangariO.D.

6
Major components of the tonometer







1)Contact Prism
2)Contact Probe
3)Housing
4)Calibrated
measuring drum
5)Tonometer Arm

TONOMETRY/H.KangariO.D.

7
TONOMETRY/H.KangariO.D.

8
TONOMETRY/H.KangariO.D.

9
CHECKING THE
INSTRUMENT







To calibrate the instrument a control weight bar
is used. Only a trained technician can calibrate
the tonometer.
To grossly check the calibration:
Turn the measuring drum below zero (the arm
should move backward)
Then turn the measuring drum past the zero
mark and the arm should move forward.
The arm should move in the same degree in
both direction, above and below the zero mark.
TONOMETRY/H.KangariO.D.

10
PROBE STERILAZATION








Always clean the tonometer probe before and
after each use.
Remove the probe from contact holder (hold the
arm before you remove the probe)
Wet a tissue with alcohol and apply to probe in a
circular fashion for 5 seconds.
Rinse the probe with Saline Solution and blot dry
with a tissue.
TONOMETRY/H.KangariO.D.

11
PROBE STERILAZATION
OR insert the probe into Hydrogen
Peroxide 3% for 5 to 10 minutes.
 Rinse the probe with Saline Solution and
blot dry with a tissue.


TONOMETRY/H.KangariO.D.

12
CLEANING & STERILIZING THE
PROBE


some factors to be considered such as the following:



(1) 3% H2O2 will ruin the probes in a short period of time and if not properly
dried can cause marked pain and a secondary anterior uveitis that is very
difficult to resolve. Minimum of 5 minutes maximum of 10 minutes.



(2) 70% Isopropyl alcohol will also destroy the probes and if not properly
dried can also cause marked pain and a secondary anterior uveitis that is
very difficult to resolve.
When Using Either Of The Above Recommended Procedures Make Sure
You Always Rinse The Probe Thoroughly With Saline Solution Then Dry It
Completely Before Using It On The Patient's Eye. Never Leave The Probe
In The Hydrogen Peroxide For Longer Than 10 Minutes.



TONOMETRY/H.KangariO.D.

13




Hold the arm, while
inserting the probe.
ALIGN the probe- if the
patient has less than 3.00
D of astigmatism, position
the probe so that the
patient’s minus cylinder
axis is aligned with the
WHITE LINE on the prism
holder.
TONOMETRY/H.KangariO.D.

14


ALIGN the probe- if the
patient has greater than
3.00 D of astigmatism,
position the probe so that
the patient’s minus
cylinder axis is aligned
with the RED LINE on the
prism holder.



KEEP in mind, as you
rotate the probe
according to cylinder axis
the mires will tilt with the
direction of axis.
TONOMETRY/H.KangariO.D.

15
PATIENT PREPARATION







After you have completed a full refractive,
binocular and slit lamp evaluation proceed to
tonometry.
Anesthetize the cornea. Inform the patient that
you are going to put a drop in their eye which will
burn and their lids will feel heavy afterwards.
Place a drop of Tetracaine 0.5% in each eye
Wet a strip of fluorescein with sterile saline and
place it is superior or inferior bulbar conjunctiva.
TONOMETRY/H.KangariO.D.

16
Position the patient comfortably at slit
lamp.
 With the COBALT BLUE filter scan the
cornea
 Place the measuring drum at 10 mmHg
 Use Low Magnification


TONOMETRY/H.KangariO.D.

17



Give proper instructions to the patient:
“ I want you to keep your forehead and Chin in
the rest all the time. I want you to look at this
point ( based on the eye being tested give an
appropriate place for the patient to look at ).
Keep your eyes open as wide as you can and try
not to blink. I am going get close to your eye with
this blue light.

TONOMETRY/H.KangariO.D.

18
APPLINATION
From the outside instrument get as close
as you can to the cornea and center the
probe on the cornea.
 If you now look in through the microscope,
you will see two faint blue mires (semi
circles)
 If you don’t see this, AGAIN center the
probe from the outside of the instrument.


TONOMETRY/H.KangariO.D.

19
When you see two faint mires, move in the
instrument with the joy stick to touch the
cornea.
 When you touch the cornea, the mires will
turn fluorescent green.


TONOMETRY/H.KangariO.D.

20


If semicircles move freely with the rotation
of the measuring drum, the initial contact
is correct.

TONOMETRY/H.KangariO.D.

21




The width and height
of two rings must be
equal and centered in
the field of view.
Rotate the measuring
drum, until the inner
borders of two
semicircles just touch.

TONOMETRY/H.KangariO.D.

22
Back off the tonometer and read the
pressure from the measuring drum.
 The scales of measuring drum is marked
from 0-8 grams of force. To convert to
mmHg, multiply the result by 10.
 After completion of the procedure,
reevaluate the corneal integrity with cobalt
blue light.


TONOMETRY/H.KangariO.D.

23


Correct position of
mires when
measuring the IOP

TONOMETRY/H.KangariO.D.

24




If this is the left eye,
the contact of the
probe is too much to
temporal side.
Move the probe to
nasal side

TONOMETRY/H.KangariO.D.

25
Top mire is superiorly
positioned.
The probe contact is too
inferior.
Move the probe
superiorly.

TONOMETRY/H.KangariO.D.

26




The contact is too
low.
Move the instrument
upward.

TONOMETRY/H.KangariO.D.

27


?

TONOMETRY/H.KangariO.D.

28


?

TONOMETRY/H.KangariO.D.

29







Inappropriate contact
The instrument has been
moved too much toward
the patient.
The mires will not move
by changing the drum.
Move backward and
reapplinate.

TONOMETRY/H.KangariO.D.

30



Mires are too thin.
Add more flourescein.

TONOMETRY/H.KangariO.D.

31


After applination, this
type of staining is
indicate of applying
too much force.

TONOMETRY/H.KangariO.D.

32
DIURNAL VARIATION






Generally, the variation in intraocular pressure over a 24
hour period is considered to be 3 to 5 mmHg with the
highest readings being about (6:00 a.m.).
However, there is recent evidence that known glaucoma
patient's highest pressure findings are in the afternoon.
Therefore, it would be best to monitor any questionable
patients. Take diurnal pressure measurements during
the day looking for any pressure spikes with variations
greater than 5 mmHg. Example: 14 mmHg O.U. @ 8:30
a.m. and 21 mmHg O.U. @ 3:30 p.m. is diagnostic.
Differences in pressure readings between the two eyes
of 3 mmHg or more must be questioned, this is not
normal.
TONOMETRY/H.KangariO.D.

33
Latest information:




Intraocular pressures (IOP) following Laser In Situ Keratomileusis
(LASIK) and Photorefractive Keratectomy (PRK) are
underestimated.
It depends on the patient's Pre-Op refractive prescription and how
much tissue needs to be removed. There is about 10 microns of
tissue removed per diopter of refractive error. A patient with a
refractive error of 3 diopters of myopia will have approximately 30
microns of tissue removed. For this patient the IOP findings would
be underestimated by about 2mm Hg and more for higher refractive
errors. This underestimation has been reported for Goldmann
applanation tonometry. It is an important new finding and appears to
be related to changes in corneal thickness. This might help explain
normal tension glaucoma where these patients may have thinner
corneas.

TONOMETRY/H.KangariO.D.

34
PERKINS TONOMETER


For patients with physical constraints who
cannot be positioned in the slit lamp or
bedriden patients Perkins tonometer (a
hand held tonometer can be used)

TONOMETRY/H.KangariO.D.

35
TONOMETRY/H.KangariO.D.

36
DIGITAL INTRAOCULAR
PRESSURE ASSESSMENT
This technique is used in those individual,
where the slit-mounted instrument or the
hand-held tonometer cannot be used.
 In very young children, extremely anxious
patient or developmentally delayed
individuals.
 The firmness of the globe is subjectively
evaluated as soft, meduim, or hard. The
harder the globe the higher the IOP.


TONOMETRY/H.KangariO.D.

37
Digital IOP assessment is contraindicated
for eyes with a recent history of blunt
trauma, penetrating ocular injury, or
intraocular surgery.
 Sedation might be recommended for more
accurate measurement of the pressure.


TONOMETRY/H.KangariO.D.

38


Ask the patient to look
down and gently rest
the tip of the fingers
on the center of the
upperlid

TONOMETRY/H.KangariO.D.

39




Indent the globe
slightly with one finger
tip.
The opposite fingertip
will rebound slightly

TONOMETRY/H.KangariO.D.

40
NON-CONTACT TONOMETER

TONOMETRY/H.KangariO.D.

41
TONOMETRY/H.KangariO.D.

42
TONOMETRY/H.KangariO.D.

43
TONOMETRY/H.KangariO.D.

44

Más contenido relacionado

La actualidad más candente

La actualidad más candente (20)

Keratometry
KeratometryKeratometry
Keratometry
 
Objective refraction
Objective refractionObjective refraction
Objective refraction
 
Trial set
Trial setTrial set
Trial set
 
Maddox rod
Maddox rodMaddox rod
Maddox rod
 
Jackson cross cylinder
Jackson cross cylinderJackson cross cylinder
Jackson cross cylinder
 
Synaptophore
SynaptophoreSynaptophore
Synaptophore
 
A scan biometry
A scan biometryA scan biometry
A scan biometry
 
JCC -Jackson Cross Cylinder
JCC -Jackson Cross CylinderJCC -Jackson Cross Cylinder
JCC -Jackson Cross Cylinder
 
Soft toric Contact Lens
Soft toric Contact LensSoft toric Contact Lens
Soft toric Contact Lens
 
Fundus Fluoroscein Angiography
Fundus Fluoroscein AngiographyFundus Fluoroscein Angiography
Fundus Fluoroscein Angiography
 
Binocular balancing
Binocular balancing Binocular balancing
Binocular balancing
 
Slit lamp biomicroscopy
Slit lamp biomicroscopySlit lamp biomicroscopy
Slit lamp biomicroscopy
 
Corneal topography
Corneal topographyCorneal topography
Corneal topography
 
Corneal Topography
Corneal TopographyCorneal Topography
Corneal Topography
 
Near point of convergence
Near point of convergenceNear point of convergence
Near point of convergence
 
Hess chart, diplopia chart, cover tests
Hess chart, diplopia chart, cover testsHess chart, diplopia chart, cover tests
Hess chart, diplopia chart, cover tests
 
Accommodation/ Accommodation of Eye, Measurement of Accommodation of Eye (hea...
Accommodation/ Accommodation of Eye, Measurement of Accommodation of Eye (hea...Accommodation/ Accommodation of Eye, Measurement of Accommodation of Eye (hea...
Accommodation/ Accommodation of Eye, Measurement of Accommodation of Eye (hea...
 
Retinoscopy
RetinoscopyRetinoscopy
Retinoscopy
 
Aniseikonia
AniseikoniaAniseikonia
Aniseikonia
 
corneal Pachymetry
 corneal Pachymetry corneal Pachymetry
corneal Pachymetry
 

Destacado (18)

Toric rgp fit
Toric rgp fitToric rgp fit
Toric rgp fit
 
optometric prescription
optometric prescriptionoptometric prescription
optometric prescription
 
Intermittent exotropia
Intermittent exotropiaIntermittent exotropia
Intermittent exotropia
 
Astigmatism 2
Astigmatism 2Astigmatism 2
Astigmatism 2
 
Auto perimetry
Auto perimetryAuto perimetry
Auto perimetry
 
Spectacle prescription fulfillment in medical optometry cope approved
Spectacle prescription fulfillment in medical optometry cope approvedSpectacle prescription fulfillment in medical optometry cope approved
Spectacle prescription fulfillment in medical optometry cope approved
 
Contact Lens-Induced Acute Red Eye(CLARE)
Contact Lens-Induced Acute Red Eye(CLARE)Contact Lens-Induced Acute Red Eye(CLARE)
Contact Lens-Induced Acute Red Eye(CLARE)
 
Presbyopia
PresbyopiaPresbyopia
Presbyopia
 
Blepharitis
BlepharitisBlepharitis
Blepharitis
 
Marketing an optometric practice
Marketing an optometric practiceMarketing an optometric practice
Marketing an optometric practice
 
Primetry
PrimetryPrimetry
Primetry
 
Amblyopia treatment dr mirzajani
Amblyopia treatment dr mirzajaniAmblyopia treatment dr mirzajani
Amblyopia treatment dr mirzajani
 
Optometry, a focus on vision
Optometry, a focus on visionOptometry, a focus on vision
Optometry, a focus on vision
 
Myopia
MyopiaMyopia
Myopia
 
Non surgical management of strabismus .ppt
Non surgical management of strabismus .pptNon surgical management of strabismus .ppt
Non surgical management of strabismus .ppt
 
keratometry
 keratometry keratometry
keratometry
 
Funduscopy
Funduscopy Funduscopy
Funduscopy
 
Biometry
BiometryBiometry
Biometry
 

Similar a Goldman applanation tonometry

TONOMETRY IN OPTOMETRY & OPHTHALMOLOGY .PPTX
TONOMETRY IN OPTOMETRY & OPHTHALMOLOGY .PPTXTONOMETRY IN OPTOMETRY & OPHTHALMOLOGY .PPTX
TONOMETRY IN OPTOMETRY & OPHTHALMOLOGY .PPTXANUJA DHAKAL
 
Auto-refractometer, Tonometer and DO.pptx
Auto-refractometer, Tonometer and DO.pptxAuto-refractometer, Tonometer and DO.pptx
Auto-refractometer, Tonometer and DO.pptxKhushikansal1
 
tonometer.pptx
tonometer.pptxtonometer.pptx
tonometer.pptxVizo7
 
Tonometry by dr. mamta meena
Tonometry by dr. mamta meenaTonometry by dr. mamta meena
Tonometry by dr. mamta meenaAIMS Education
 
Tonometry Procedure
Tonometry ProcedureTonometry Procedure
Tonometry ProcedureSYR1698
 
Tonometry & its recent advances
Tonometry & its recent advancesTonometry & its recent advances
Tonometry & its recent advanceslijulk
 
Gonioscopy and methods to assess anterior segments
Gonioscopy and methods to assess anterior segmentsGonioscopy and methods to assess anterior segments
Gonioscopy and methods to assess anterior segmentsBipin Bista
 
tonometry-140624223245-phpapp02.pptx
tonometry-140624223245-phpapp02.pptxtonometry-140624223245-phpapp02.pptx
tonometry-140624223245-phpapp02.pptxaiswaryavenugopalann
 
CLMMRH Ophthalmology Case Pres Reporting
CLMMRH Ophthalmology Case Pres ReportingCLMMRH Ophthalmology Case Pres Reporting
CLMMRH Ophthalmology Case Pres ReportingPaulineMAntolo
 

Similar a Goldman applanation tonometry (20)

Iop
IopIop
Iop
 
Tonometry
TonometryTonometry
Tonometry
 
Tonometry
TonometryTonometry
Tonometry
 
Tonometry
TonometryTonometry
Tonometry
 
Tonometer
TonometerTonometer
Tonometer
 
Different types of Tonometry
Different types of TonometryDifferent types of Tonometry
Different types of Tonometry
 
TONOMETRY IN OPTOMETRY & OPHTHALMOLOGY .PPTX
TONOMETRY IN OPTOMETRY & OPHTHALMOLOGY .PPTXTONOMETRY IN OPTOMETRY & OPHTHALMOLOGY .PPTX
TONOMETRY IN OPTOMETRY & OPHTHALMOLOGY .PPTX
 
Tonometry and tonography
Tonometry and tonographyTonometry and tonography
Tonometry and tonography
 
Auto-refractometer, Tonometer and DO.pptx
Auto-refractometer, Tonometer and DO.pptxAuto-refractometer, Tonometer and DO.pptx
Auto-refractometer, Tonometer and DO.pptx
 
TONOMETRY
TONOMETRYTONOMETRY
TONOMETRY
 
tonometer.pptx
tonometer.pptxtonometer.pptx
tonometer.pptx
 
Tonometry
 Tonometry Tonometry
Tonometry
 
Tonometry by dr. mamta meena
Tonometry by dr. mamta meenaTonometry by dr. mamta meena
Tonometry by dr. mamta meena
 
Tonometry Procedure
Tonometry ProcedureTonometry Procedure
Tonometry Procedure
 
Tonometry & its recent advances
Tonometry & its recent advancesTonometry & its recent advances
Tonometry & its recent advances
 
Tonometry.pptx
Tonometry.pptxTonometry.pptx
Tonometry.pptx
 
Gonioscopy and methods to assess anterior segments
Gonioscopy and methods to assess anterior segmentsGonioscopy and methods to assess anterior segments
Gonioscopy and methods to assess anterior segments
 
Part_1a.ppt
Part_1a.pptPart_1a.ppt
Part_1a.ppt
 
tonometry-140624223245-phpapp02.pptx
tonometry-140624223245-phpapp02.pptxtonometry-140624223245-phpapp02.pptx
tonometry-140624223245-phpapp02.pptx
 
CLMMRH Ophthalmology Case Pres Reporting
CLMMRH Ophthalmology Case Pres ReportingCLMMRH Ophthalmology Case Pres Reporting
CLMMRH Ophthalmology Case Pres Reporting
 

Más de Hossein Mirzaie

Mba monograph13 0412_marketinganoptometricpractice
Mba monograph13 0412_marketinganoptometricpracticeMba monograph13 0412_marketinganoptometricpractice
Mba monograph13 0412_marketinganoptometricpracticeHossein Mirzaie
 
Peripheral refraction in myopia
Peripheral refraction in myopiaPeripheral refraction in myopia
Peripheral refraction in myopiaHossein Mirzaie
 
Soft contact lens changes through history (in persian)
Soft contact lens changes through  history  (in persian)Soft contact lens changes through  history  (in persian)
Soft contact lens changes through history (in persian)Hossein Mirzaie
 
Chapter 27-ophthalmic-care
Chapter 27-ophthalmic-careChapter 27-ophthalmic-care
Chapter 27-ophthalmic-careHossein Mirzaie
 
Primary level eyecare services
Primary level eyecare servicesPrimary level eyecare services
Primary level eyecare servicesHossein Mirzaie
 
7 criticalmistakesthatoptometristsmakeintheirpractices
7 criticalmistakesthatoptometristsmakeintheirpractices7 criticalmistakesthatoptometristsmakeintheirpractices
7 criticalmistakesthatoptometristsmakeintheirpracticesHossein Mirzaie
 

Más de Hossein Mirzaie (12)

Mba monograph13 0412_marketinganoptometricpractice
Mba monograph13 0412_marketinganoptometricpracticeMba monograph13 0412_marketinganoptometricpractice
Mba monograph13 0412_marketinganoptometricpractice
 
Peripheral refraction in myopia
Peripheral refraction in myopiaPeripheral refraction in myopia
Peripheral refraction in myopia
 
Soft contact lens changes through history (in persian)
Soft contact lens changes through  history  (in persian)Soft contact lens changes through  history  (in persian)
Soft contact lens changes through history (in persian)
 
charles bonnet syndrome
charles bonnet syndromecharles bonnet syndrome
charles bonnet syndrome
 
3D vision syndrome
3D vision syndrome3D vision syndrome
3D vision syndrome
 
Staining grid lecture
Staining grid lectureStaining grid lecture
Staining grid lecture
 
Prism.ppt
Prism.pptPrism.ppt
Prism.ppt
 
Eyeglass frames
Eyeglass framesEyeglass frames
Eyeglass frames
 
Chapter 27-ophthalmic-care
Chapter 27-ophthalmic-careChapter 27-ophthalmic-care
Chapter 27-ophthalmic-care
 
Lens power measurement
Lens power measurementLens power measurement
Lens power measurement
 
Primary level eyecare services
Primary level eyecare servicesPrimary level eyecare services
Primary level eyecare services
 
7 criticalmistakesthatoptometristsmakeintheirpractices
7 criticalmistakesthatoptometristsmakeintheirpractices7 criticalmistakesthatoptometristsmakeintheirpractices
7 criticalmistakesthatoptometristsmakeintheirpractices
 

Último

Sonagachi Call Girls Services 9907093804 @24x7 High Class Babes Here Call Now
Sonagachi Call Girls Services 9907093804 @24x7 High Class Babes Here Call NowSonagachi Call Girls Services 9907093804 @24x7 High Class Babes Here Call Now
Sonagachi Call Girls Services 9907093804 @24x7 High Class Babes Here Call NowRiya Pathan
 
Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...
Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...
Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...narwatsonia7
 
Call Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service Availablenarwatsonia7
 
College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...
College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...
College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...Miss joya
 
Book Call Girls in Yelahanka - For 7001305949 Cheap & Best with original Photos
Book Call Girls in Yelahanka - For 7001305949 Cheap & Best with original PhotosBook Call Girls in Yelahanka - For 7001305949 Cheap & Best with original Photos
Book Call Girls in Yelahanka - For 7001305949 Cheap & Best with original Photosnarwatsonia7
 
Call Girls Thane Just Call 9910780858 Get High Class Call Girls Service
Call Girls Thane Just Call 9910780858 Get High Class Call Girls ServiceCall Girls Thane Just Call 9910780858 Get High Class Call Girls Service
Call Girls Thane Just Call 9910780858 Get High Class Call Girls Servicesonalikaur4
 
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy Girls
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy GirlsCall Girls In Andheri East Call 9920874524 Book Hot And Sexy Girls
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy Girlsnehamumbai
 
Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...
Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...
Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...Miss joya
 
Bangalore Call Girls Marathahalli 📞 9907093804 High Profile Service 100% Safe
Bangalore Call Girls Marathahalli 📞 9907093804 High Profile Service 100% SafeBangalore Call Girls Marathahalli 📞 9907093804 High Profile Service 100% Safe
Bangalore Call Girls Marathahalli 📞 9907093804 High Profile Service 100% Safenarwatsonia7
 
call girls in green park DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️
call girls in green park  DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️call girls in green park  DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️
call girls in green park DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️saminamagar
 
Call Girls Service In Shyam Nagar Whatsapp 8445551418 Independent Escort Service
Call Girls Service In Shyam Nagar Whatsapp 8445551418 Independent Escort ServiceCall Girls Service In Shyam Nagar Whatsapp 8445551418 Independent Escort Service
Call Girls Service In Shyam Nagar Whatsapp 8445551418 Independent Escort Serviceparulsinha
 
call girls in Connaught Place DELHI 🔝 >༒9540349809 🔝 genuine Escort Service ...
call girls in Connaught Place  DELHI 🔝 >༒9540349809 🔝 genuine Escort Service ...call girls in Connaught Place  DELHI 🔝 >༒9540349809 🔝 genuine Escort Service ...
call girls in Connaught Place DELHI 🔝 >༒9540349809 🔝 genuine Escort Service ...saminamagar
 
Russian Call Girls Chickpet - 7001305949 Booking and charges genuine rate for...
Russian Call Girls Chickpet - 7001305949 Booking and charges genuine rate for...Russian Call Girls Chickpet - 7001305949 Booking and charges genuine rate for...
Russian Call Girls Chickpet - 7001305949 Booking and charges genuine rate for...narwatsonia7
 
Call Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
Call Girl Bangalore Nandini 7001305949 Independent Escort Service BangaloreCall Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
Call Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalorenarwatsonia7
 
Low Rate Call Girls Mumbai Suman 9910780858 Independent Escort Service Mumbai
Low Rate Call Girls Mumbai Suman 9910780858 Independent Escort Service MumbaiLow Rate Call Girls Mumbai Suman 9910780858 Independent Escort Service Mumbai
Low Rate Call Girls Mumbai Suman 9910780858 Independent Escort Service Mumbaisonalikaur4
 
Call Girl Koramangala | 7001305949 At Low Cost Cash Payment Booking
Call Girl Koramangala | 7001305949 At Low Cost Cash Payment BookingCall Girl Koramangala | 7001305949 At Low Cost Cash Payment Booking
Call Girl Koramangala | 7001305949 At Low Cost Cash Payment Bookingnarwatsonia7
 
Call Girl Lucknow Mallika 7001305949 Independent Escort Service Lucknow
Call Girl Lucknow Mallika 7001305949 Independent Escort Service LucknowCall Girl Lucknow Mallika 7001305949 Independent Escort Service Lucknow
Call Girl Lucknow Mallika 7001305949 Independent Escort Service Lucknownarwatsonia7
 
Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...
Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...
Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...Miss joya
 
Call Girls Electronic City Just Call 7001305949 Top Class Call Girl Service A...
Call Girls Electronic City Just Call 7001305949 Top Class Call Girl Service A...Call Girls Electronic City Just Call 7001305949 Top Class Call Girl Service A...
Call Girls Electronic City Just Call 7001305949 Top Class Call Girl Service A...narwatsonia7
 

Último (20)

Sonagachi Call Girls Services 9907093804 @24x7 High Class Babes Here Call Now
Sonagachi Call Girls Services 9907093804 @24x7 High Class Babes Here Call NowSonagachi Call Girls Services 9907093804 @24x7 High Class Babes Here Call Now
Sonagachi Call Girls Services 9907093804 @24x7 High Class Babes Here Call Now
 
Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...
Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...
Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...
 
Call Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service Available
 
College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...
College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...
College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...
 
Book Call Girls in Yelahanka - For 7001305949 Cheap & Best with original Photos
Book Call Girls in Yelahanka - For 7001305949 Cheap & Best with original PhotosBook Call Girls in Yelahanka - For 7001305949 Cheap & Best with original Photos
Book Call Girls in Yelahanka - For 7001305949 Cheap & Best with original Photos
 
Call Girls Thane Just Call 9910780858 Get High Class Call Girls Service
Call Girls Thane Just Call 9910780858 Get High Class Call Girls ServiceCall Girls Thane Just Call 9910780858 Get High Class Call Girls Service
Call Girls Thane Just Call 9910780858 Get High Class Call Girls Service
 
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy Girls
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy GirlsCall Girls In Andheri East Call 9920874524 Book Hot And Sexy Girls
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy Girls
 
Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...
Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...
Russian Call Girls in Pune Riya 9907093804 Short 1500 Night 6000 Best call gi...
 
Bangalore Call Girls Marathahalli 📞 9907093804 High Profile Service 100% Safe
Bangalore Call Girls Marathahalli 📞 9907093804 High Profile Service 100% SafeBangalore Call Girls Marathahalli 📞 9907093804 High Profile Service 100% Safe
Bangalore Call Girls Marathahalli 📞 9907093804 High Profile Service 100% Safe
 
call girls in green park DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️
call girls in green park  DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️call girls in green park  DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️
call girls in green park DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️
 
Call Girls Service In Shyam Nagar Whatsapp 8445551418 Independent Escort Service
Call Girls Service In Shyam Nagar Whatsapp 8445551418 Independent Escort ServiceCall Girls Service In Shyam Nagar Whatsapp 8445551418 Independent Escort Service
Call Girls Service In Shyam Nagar Whatsapp 8445551418 Independent Escort Service
 
call girls in Connaught Place DELHI 🔝 >༒9540349809 🔝 genuine Escort Service ...
call girls in Connaught Place  DELHI 🔝 >༒9540349809 🔝 genuine Escort Service ...call girls in Connaught Place  DELHI 🔝 >༒9540349809 🔝 genuine Escort Service ...
call girls in Connaught Place DELHI 🔝 >༒9540349809 🔝 genuine Escort Service ...
 
Escort Service Call Girls In Sarita Vihar,, 99530°56974 Delhi NCR
Escort Service Call Girls In Sarita Vihar,, 99530°56974 Delhi NCREscort Service Call Girls In Sarita Vihar,, 99530°56974 Delhi NCR
Escort Service Call Girls In Sarita Vihar,, 99530°56974 Delhi NCR
 
Russian Call Girls Chickpet - 7001305949 Booking and charges genuine rate for...
Russian Call Girls Chickpet - 7001305949 Booking and charges genuine rate for...Russian Call Girls Chickpet - 7001305949 Booking and charges genuine rate for...
Russian Call Girls Chickpet - 7001305949 Booking and charges genuine rate for...
 
Call Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
Call Girl Bangalore Nandini 7001305949 Independent Escort Service BangaloreCall Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
Call Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
 
Low Rate Call Girls Mumbai Suman 9910780858 Independent Escort Service Mumbai
Low Rate Call Girls Mumbai Suman 9910780858 Independent Escort Service MumbaiLow Rate Call Girls Mumbai Suman 9910780858 Independent Escort Service Mumbai
Low Rate Call Girls Mumbai Suman 9910780858 Independent Escort Service Mumbai
 
Call Girl Koramangala | 7001305949 At Low Cost Cash Payment Booking
Call Girl Koramangala | 7001305949 At Low Cost Cash Payment BookingCall Girl Koramangala | 7001305949 At Low Cost Cash Payment Booking
Call Girl Koramangala | 7001305949 At Low Cost Cash Payment Booking
 
Call Girl Lucknow Mallika 7001305949 Independent Escort Service Lucknow
Call Girl Lucknow Mallika 7001305949 Independent Escort Service LucknowCall Girl Lucknow Mallika 7001305949 Independent Escort Service Lucknow
Call Girl Lucknow Mallika 7001305949 Independent Escort Service Lucknow
 
Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...
Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...
Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...
 
Call Girls Electronic City Just Call 7001305949 Top Class Call Girl Service A...
Call Girls Electronic City Just Call 7001305949 Top Class Call Girl Service A...Call Girls Electronic City Just Call 7001305949 Top Class Call Girl Service A...
Call Girls Electronic City Just Call 7001305949 Top Class Call Girl Service A...
 

Goldman applanation tonometry

  • 2. Different technique in measurement of the Intra Ocular Pressure(IOP):  Goldmann Applination Tonometry  Perkins Tonometer  Non-Contact Tonometry  Digital Evaluation  TONOMETRY/H.KangariO.D. 2
  • 3. Goldmann Applination Tonometry is the most accurate method of measuring Intraocular Pressure.  Measuring the intraocular pressure is important in diagnosis and management of Glaucoma and Ocular Hypertension.  TONOMETRY/H.KangariO.D. 3
  • 4. CONTRAINDICATIONS TO IOP MEASUREMENT WITH GOLDMAN TECHNIQUE 1)Active Infection  2)Patients with recurrent corneal erosions  3) Corneal abrasions that are not totally healed  TONOMETRY/H.KangariO.D. 4
  • 5.  The Intraocular Pressure is determined by the amount of force required to flatten a constant corneal surface area TONOMETRY/H.KangariO.D. 5
  • 7. Major components of the tonometer      1)Contact Prism 2)Contact Probe 3)Housing 4)Calibrated measuring drum 5)Tonometer Arm TONOMETRY/H.KangariO.D. 7
  • 10. CHECKING THE INSTRUMENT      To calibrate the instrument a control weight bar is used. Only a trained technician can calibrate the tonometer. To grossly check the calibration: Turn the measuring drum below zero (the arm should move backward) Then turn the measuring drum past the zero mark and the arm should move forward. The arm should move in the same degree in both direction, above and below the zero mark. TONOMETRY/H.KangariO.D. 10
  • 11. PROBE STERILAZATION     Always clean the tonometer probe before and after each use. Remove the probe from contact holder (hold the arm before you remove the probe) Wet a tissue with alcohol and apply to probe in a circular fashion for 5 seconds. Rinse the probe with Saline Solution and blot dry with a tissue. TONOMETRY/H.KangariO.D. 11
  • 12. PROBE STERILAZATION OR insert the probe into Hydrogen Peroxide 3% for 5 to 10 minutes.  Rinse the probe with Saline Solution and blot dry with a tissue.  TONOMETRY/H.KangariO.D. 12
  • 13. CLEANING & STERILIZING THE PROBE  some factors to be considered such as the following:  (1) 3% H2O2 will ruin the probes in a short period of time and if not properly dried can cause marked pain and a secondary anterior uveitis that is very difficult to resolve. Minimum of 5 minutes maximum of 10 minutes.  (2) 70% Isopropyl alcohol will also destroy the probes and if not properly dried can also cause marked pain and a secondary anterior uveitis that is very difficult to resolve. When Using Either Of The Above Recommended Procedures Make Sure You Always Rinse The Probe Thoroughly With Saline Solution Then Dry It Completely Before Using It On The Patient's Eye. Never Leave The Probe In The Hydrogen Peroxide For Longer Than 10 Minutes.  TONOMETRY/H.KangariO.D. 13
  • 14.   Hold the arm, while inserting the probe. ALIGN the probe- if the patient has less than 3.00 D of astigmatism, position the probe so that the patient’s minus cylinder axis is aligned with the WHITE LINE on the prism holder. TONOMETRY/H.KangariO.D. 14
  • 15.  ALIGN the probe- if the patient has greater than 3.00 D of astigmatism, position the probe so that the patient’s minus cylinder axis is aligned with the RED LINE on the prism holder.  KEEP in mind, as you rotate the probe according to cylinder axis the mires will tilt with the direction of axis. TONOMETRY/H.KangariO.D. 15
  • 16. PATIENT PREPARATION     After you have completed a full refractive, binocular and slit lamp evaluation proceed to tonometry. Anesthetize the cornea. Inform the patient that you are going to put a drop in their eye which will burn and their lids will feel heavy afterwards. Place a drop of Tetracaine 0.5% in each eye Wet a strip of fluorescein with sterile saline and place it is superior or inferior bulbar conjunctiva. TONOMETRY/H.KangariO.D. 16
  • 17. Position the patient comfortably at slit lamp.  With the COBALT BLUE filter scan the cornea  Place the measuring drum at 10 mmHg  Use Low Magnification  TONOMETRY/H.KangariO.D. 17
  • 18.   Give proper instructions to the patient: “ I want you to keep your forehead and Chin in the rest all the time. I want you to look at this point ( based on the eye being tested give an appropriate place for the patient to look at ). Keep your eyes open as wide as you can and try not to blink. I am going get close to your eye with this blue light. TONOMETRY/H.KangariO.D. 18
  • 19. APPLINATION From the outside instrument get as close as you can to the cornea and center the probe on the cornea.  If you now look in through the microscope, you will see two faint blue mires (semi circles)  If you don’t see this, AGAIN center the probe from the outside of the instrument.  TONOMETRY/H.KangariO.D. 19
  • 20. When you see two faint mires, move in the instrument with the joy stick to touch the cornea.  When you touch the cornea, the mires will turn fluorescent green.  TONOMETRY/H.KangariO.D. 20
  • 21.  If semicircles move freely with the rotation of the measuring drum, the initial contact is correct. TONOMETRY/H.KangariO.D. 21
  • 22.   The width and height of two rings must be equal and centered in the field of view. Rotate the measuring drum, until the inner borders of two semicircles just touch. TONOMETRY/H.KangariO.D. 22
  • 23. Back off the tonometer and read the pressure from the measuring drum.  The scales of measuring drum is marked from 0-8 grams of force. To convert to mmHg, multiply the result by 10.  After completion of the procedure, reevaluate the corneal integrity with cobalt blue light.  TONOMETRY/H.KangariO.D. 23
  • 24.  Correct position of mires when measuring the IOP TONOMETRY/H.KangariO.D. 24
  • 25.   If this is the left eye, the contact of the probe is too much to temporal side. Move the probe to nasal side TONOMETRY/H.KangariO.D. 25
  • 26. Top mire is superiorly positioned. The probe contact is too inferior. Move the probe superiorly. TONOMETRY/H.KangariO.D. 26
  • 27.   The contact is too low. Move the instrument upward. TONOMETRY/H.KangariO.D. 27
  • 30.     Inappropriate contact The instrument has been moved too much toward the patient. The mires will not move by changing the drum. Move backward and reapplinate. TONOMETRY/H.KangariO.D. 30
  • 31.   Mires are too thin. Add more flourescein. TONOMETRY/H.KangariO.D. 31
  • 32.  After applination, this type of staining is indicate of applying too much force. TONOMETRY/H.KangariO.D. 32
  • 33. DIURNAL VARIATION    Generally, the variation in intraocular pressure over a 24 hour period is considered to be 3 to 5 mmHg with the highest readings being about (6:00 a.m.). However, there is recent evidence that known glaucoma patient's highest pressure findings are in the afternoon. Therefore, it would be best to monitor any questionable patients. Take diurnal pressure measurements during the day looking for any pressure spikes with variations greater than 5 mmHg. Example: 14 mmHg O.U. @ 8:30 a.m. and 21 mmHg O.U. @ 3:30 p.m. is diagnostic. Differences in pressure readings between the two eyes of 3 mmHg or more must be questioned, this is not normal. TONOMETRY/H.KangariO.D. 33
  • 34. Latest information:   Intraocular pressures (IOP) following Laser In Situ Keratomileusis (LASIK) and Photorefractive Keratectomy (PRK) are underestimated. It depends on the patient's Pre-Op refractive prescription and how much tissue needs to be removed. There is about 10 microns of tissue removed per diopter of refractive error. A patient with a refractive error of 3 diopters of myopia will have approximately 30 microns of tissue removed. For this patient the IOP findings would be underestimated by about 2mm Hg and more for higher refractive errors. This underestimation has been reported for Goldmann applanation tonometry. It is an important new finding and appears to be related to changes in corneal thickness. This might help explain normal tension glaucoma where these patients may have thinner corneas. TONOMETRY/H.KangariO.D. 34
  • 35. PERKINS TONOMETER  For patients with physical constraints who cannot be positioned in the slit lamp or bedriden patients Perkins tonometer (a hand held tonometer can be used) TONOMETRY/H.KangariO.D. 35
  • 37. DIGITAL INTRAOCULAR PRESSURE ASSESSMENT This technique is used in those individual, where the slit-mounted instrument or the hand-held tonometer cannot be used.  In very young children, extremely anxious patient or developmentally delayed individuals.  The firmness of the globe is subjectively evaluated as soft, meduim, or hard. The harder the globe the higher the IOP.  TONOMETRY/H.KangariO.D. 37
  • 38. Digital IOP assessment is contraindicated for eyes with a recent history of blunt trauma, penetrating ocular injury, or intraocular surgery.  Sedation might be recommended for more accurate measurement of the pressure.  TONOMETRY/H.KangariO.D. 38
  • 39.  Ask the patient to look down and gently rest the tip of the fingers on the center of the upperlid TONOMETRY/H.KangariO.D. 39
  • 40.   Indent the globe slightly with one finger tip. The opposite fingertip will rebound slightly TONOMETRY/H.KangariO.D. 40