THE LOCATOR
OVERDENTURE
IMPLANT SYSTEM.
®

FOUR DECADES OF ATTACHMENT
KNOWLEDGE INCORPORATED
INTO NARROW DIAMETER
OVERDENTURE IMPLANTS.
The LOCATOR Overdenture Implant System (LODI) is comprised
of 2.4mm and 2.9mm narrow diameter dental implants (available in
10, 12 and 14mm lengths) with a detachable LOCATOR Attachment
that is available in a 2.5mm and 4mm cuff height. The LODI is used
to restore masticatory function for the patient and may be suitable
for immediate function if sufficient primary stability of the implant
is achieved at the time of placement.

IMPORTANT: This document contains the most current instructions for use.
Please read and retain.
TABLE OF CONTENTS
LOCATOR® overdenture implant system
2	Indications
2	Contraindications
2	Caution
2	 Storage and Handling
2	 Single-Use Devices
2	Sterilization
3	 Cleaning Instructions for Instruments (and Individually Packaged Replacement Attachments)
3	 Surgical Tray Cleaning Instructions
3	 Inspection and Maintenance of Cleaned Instruments
3	 Steam Sterilization Instructions
4	 Torque Indicating Ratchet Wrench Cleaning
5	 Warnings and Precautions
DRILLING DEPTH CONTROL
6	 Drill Laser Depth Markings
6	 Drill Stops
6	 Final Drill Diameter and Depth for Various Bone Types
7	 Drilling Sequence Examples
7	 Placement of a 2.4mm x 12mm Implant Flapless Surgical Procedure
8	 Placement of a 2.9mm x 12mm Implant Flapless Surgical Procedure
LOCATOR MALES & EXTENDED RANGE MALES
LOCATOR CORE TOOL
PRE-SURGICAL TREATMENT PLANNING
IMPLANT PLACEMENT
LOCATOR ATTACHMENT PLACEMENT
PROCESSING LOCATOR DENTURE CAPS & MALES INTO THE DENTURE
18	
20	
21	
21	
22	

Direct Technique: Chairside Processing
Indirect Technique: Laboratory Processing
Lab Steps
Bite Records
Delivery

EXPLANATION OF SYMBOLS ON OUTER PACKAGING LABELS

1
LOCATOR OVERDENTURE IMPLANT SYSTEM
®

INDICATIONS

STERILIZATION

The LOCATOR Overdenture Implant System is
designed to retain overdentures or partial dentures in
the mandible or maxilla.

The LOCATOR Overdenture Implant is packaged
with the LOCATOR Attachment and together are
supplied STERILE (subjected to radiation (gamma)
as a means of sterilization).

CONTRAINDICATIONS
Not appropriate where a totally rigid connection is
required. Use of a single implant with divergence of
greater than 20 degrees is not recommended. Dental
implants should not be used in patients with serious
medical problems or in a poor general state of health.
Patients with medical problems such as; uncontrolled
bleeding disorders, drug or alcohol abuse, weakened
immune system, titanium allergy or uncontrollable
endocrine disorders should be carefully evaluated
prior to treatment.

CAUTION

All other restorative components, instruments,
and replacement LOCATOR Attachments (sold
separately) are supplied NON-STERILE.
The nylon males may be sterilized using a liquid
chemical sterilant following the manufacturer’s
recommendations to achieve sterilization.
Note: The Liquid Chemical Sterilant must be
approved for Sterilization not just high-level
disinfection and must be compatible with nylon
material.

Federal (U.S.A.) law restricts this device to sale by or
on the order of a licensed dentist.

Storage and Handling
The LOCATOR Overdenture Implant System in its
undamaged, original packaging is not subject to
any special considerations for storage or handling
(during transport and storage). Only sterile titanium
or stainless instruments/tools should be used to
handle and deliver the implant to the surgical site.

SINGLE-USE DEVICES
The LOCATOR Overdenture Implant System is a
single-use device.
LOCATOR Overdenture Implant: A previously used
LOCATOR Overdenture Implant could contain patient
contamination build-up. Therefore, the inadvertent
re-use of this device could result in infection leading to
lack of integration (of the implant to the bone).
LOCATOR Males: The inadvertent re-use of LOCATOR
nylon males could cause loss of retention of the
overdenture due to wear from previous use or damage
during removal with the LOCATOR Core Tool.
LOCATOR Attachments: The inadvertent re-use
of LOCATOR Attachments could contain patient
contamination build-up and subsequent wear of the
retention bands. This would result in the device to
perform with improper fit and function which would
result in loss of retention of the prosthesis.

2
LOCATOR OVERDENTURE IMPLANT SYSTEM
®

(Continued)
CLEANING INSTRUCTIONS FOR
INSTRUMENTS (AND INDIVIDUALLY
PACKAGED REPLACEMENT
Attachments)
Disassemble any instruments that can be
disassembled according to manufacturers’
instructions.
Soak instruments in enzymatic cleaning solution
(mixed according to manufacturers’ instructions) by
completely submerging them for 20 minutes. Scrub
instruments using a soft-bristled, nylon brush until all
soil has been removed.
Remove the instruments from the enzymatic
cleaning solution and rinse in tap water for a
minimum of 3 minutes. Make sure to thoroughly flush
internal holes/crevices of instruments (such as the
tissue punch, drill extender, implant drivers, and
disassembled core tool and ratchet torque wrench)
that have difficult to reach areas.
Note: Use of a syringe or water jet will improve
flushing of difficult to reach areas.
Place instruments in sonication bath (with
enzymatic cleaning solution prepared according to
manufacturers’ instructions) making sure that they are
completely submerged, and sonicate for 10 minutes.
Remove the instruments from the sonication bath,
and rinse for 3 minutes making sure to thoroughly
flush cleaning solution out of the holes/crevices and/
or difficult to reach areas.
Remove excess moisture from the instruments
with a clean, absorbent, and non-shedding wipe.

Surgical Tray cleaning instructions
Rinse the tray and tray insert with tap water.
Place the Surgical Tray and Insert in enzymatic
cleaning solution (mixed according to manufacturers’
instructions) and wipe off soil with a clean, absorbent,
non-shedding wipe. Allow the Surgical Tray and Insert
to soak in the cleaning solution for 20 minutes making
sure that they are completely submerged.

3

Surgical Tray cleaning instructions
(CONTINUED)
Remove the Surgical Tray and Insert from the
enzymatic cleaning solution and rinse in tap water
for a minimum of 3 minutes. Make sure to thoroughly
flush each piece to completely remove cleaning
solution residue.
Remove excess moisture from the Surgical Tray
and Insert with a clean, absorbent, and non-shedding
wipe.

INSPECTION AND MAINTENANCE OF
CLEANED INSTRUMENTS
Carefully inspect each instrument to ensure
that all visible contamination has been removed. If
contamination is noted, repeat the cleaning process.
Please note that if during inspection of instruments,
you see signs of wear, damage, or unrecognizable
color change, replace the instrument.
Re-assemble multi-part instruments and check
them for proper function (LOCATOR Core Tool and
the Torque Indicating Ratchet Wrench) Reference
the IFU that comes with each of these parts and
subsequent sections of this document for the proper
assembly process.

Steam Sterilization Instructions
The validation procedures require the use of FDAcleared sterilizers, sterilization trays, sterilization
wraps, biological indicators, chemical indicators
and other sterilization accessories labeled for the
sterilization cycle recommended. The health care
facility should monitor the sterilizer for the facility
according to an FDA recognized sterility assurance
standard such as ANSI/AAMI ST79:2006.
Place all instruments into the surgical tray.
For gravity cycle, place Surgical Kit in a 10” x 15”
Autoclave Bag, and for Pre-Vacuum Cycle double
wrap the kit with autoclave wrap material and secure
wrap with autoclave tape.
Autoclave Sterilization Parameters are listed below:
CYCLE TYPE
GRAVITY
PRE-VACUUM

TEMPERATURE
121ºC / 250ºF
132ºC / 270ºF

EXPOSURE TIME
50 MINUTES
4 MINUTES

DRYING TIME
15 MINUTES
20 MINUTES
LOCATOR OVERDENTURE IMPLANT SYSTEM
®

(Continued)
TORQUE INDICATING RATCHET WRENCH CLEANING
Intended Use: A dental torque wrench for placement
and adjustment of dental implants, attachments,
attachment screws and prosthetic screws during oral
surgery and prosthetic procedures. Scale Unit: Ncm.

Sterilzation: Autoclave/steam gravity sterilize for 50
minutes at 121ºc, dry for 15 minutes. For pre-vacuum
cycle, autoclave/steam sterilize for 4 minutes at 132ºC
with drying time of 20 minutes.

WARNING: Device must be autoclaved prior to use.
This device must not be cleaned using hydrogen
peroxide.

Note: Drying times may vary according to load.

Cleaning: Press the driver to remove it from the
head of the wrench, and remove the head by pressing
a finger into the recess and gently pulling the head.
The three separated parts are now ready for cleaning
using the following procedure:
Soak torque wrench parts in enzymatic cleaning
solution (mixed according to manufacturer’s
instructions) by completely submerging it for 20
minutes. Scrub torque wrench parts using a soft
bristled, nylon brush until all soil has been removed.
Remove the torque wrench parts from the enzymatic
cleaning solution and rinse in tap water for a minimum
of 3 minutes.
Place torque wrench parts in sonication bath (with
the enzymatic cleaning solution prepared according
to manufacturer’s instructions) making sure they are
completely submered, and sonicate for 10 minutes.
Remove the torque wrench parts from the sonication
bath, and rinse in water for 3 minutes making sure to
thoroughly flush cleaning solution out of the holes/
crevices and/or difficult to reach areas.
Remove excess moisture from the torque wrench
parts with a clean, absorbent, and non-shedding wipe.

After sterilization, attach the head of the wrench
to the body by pushing the components together and
turning them in opposite directions until there is an
audible click.
Push the driver into the wrench until there is an
audible click. The arrow on the head of the wrench
shows the direction in which the wrench is functioning.
Turn the wrench in the direction of the arrow until
the desired torque is achieved.
WARNING: Before each use, make sure that the
functionalities are intact and that the first line on the
scale aligns with the arrow. The arm of the torque
wrench must not go beyond the end of the scale, as
this could result in inaccurate readings. If the torque
wrench is used as an ordinary wrench, without using
the torque scale, then it may not be sujected to a load
of more than 80Ncm.
WARNING: After overloading or if dropped or in other
ways mishandled, the wrench must no longer be used
since correct function can no longer be guaranteed.
Please refer to the LOCATOR Overdenture Implant
System Technique Manual available from the
manufacturer or your distributor for detailed surgical
procedure instructions. It is also available online at
www.zestanchors.com

4
LOCATOR OVERDENTURE IMPLANT SYSTEM
®

(Continued)
WARNINGS AND PRECAUTIONS
The LOCATOR Overdenture Implant System has
not been evaluated for safety and compatibility in
the MR environment. The LOCATOR Overdenture
Implant System has not been tested for heating or
migration in the MR environment.
Product (implant/attachment) from damaged
sterilized packaging must not be used on patients.
In the event that the sterilized packaging for the
LOCATOR Overdenture Implant System is damaged,
the damaged packaging (with the product) must be
returned to the manufacturer and a replacement will
be provided (if damage to sterilized packaging is
caused by product shipment).
The drill extender is to be used with surgical
drills only and should not be used in high torque
applications.
Avoid application of excessive bending load on
smaller diameter drills during drilling. Drills will dull
based on many factors including bone density,
handling, autoclave exposure, etc. Replace drills
when wear is noticeable to avoid excessive heat
being transferred to surrounding bone during
osteotomy preparation.
If the LOCATOR Overdenture Implant System is
subjected to inappropriate loading conditions, there
may be a potential risk of metal fatigue or localized
bone failure. The use of other tissue grafting
components or parts that are made from dissimilar
metals should not be used in or near the implant.
Patient evaluation including the determination of
the general health, oral hygiene habits and status,
motivation toward good dental care, and anatomic
acceptability prior to implant surgery is critical.
Thorough evaluation of the patient’s medical status
and health history is mandatory. Panaromic and
periapical radiographs as well as thorough oral
inspection and palpation are recommended to
determine anatomic landmarks, dental pathology,
and adequacy of bone. A cephalogram is suggested
for totally edentulous patients. Any oral condition
that adversely affects natural teeth, if uncorrected,
will have an adverse effect on the implants.

5

Periodontal disease, abnormal bone conditions,
severe bruxism, cross-bite situations, and extenuating
circumstances (e.g. excessive smoking, medical issues,
etc) that may adversely affect the procedure must
be evaluated and corrected if necessary, or use of the
implant may be contraindicated.
Based on the results of the patient’s pre-surgical
assessment, the clinician should select and order
the appropriate implant (determine correct
implant diameter and length based on bone
type), restorative parts, and tools. Refer to Drilling
Sequence section for further details. The clinician
should also determine if the patient is allergic to any
of the materials that will be used in the procedure as
part of the pre-surgical treatment planning. If during
patient evaluation, insufficient bone width, abnormal
bone defects or contours are detected, then the
placement of the implant may be contraindicated.
Patient motivation is a key factor in achieving
success with any implant. The patient must be
willing to practice the oral hygiene necessary for
implant maintenance. The clinician must provide
the patient with information regarding proper care
and maintenance of the implants. Also, they must
inform the patient that conditions such as excessive
smoking, improper/lack of maintenance may have
adverse effects.
The use of this or any surgical implant product
requires that the clinician be thoroughly familiar
with the product and the method for its use and
application. They must also be familiar with all the
instruments, and surgical procedures required
(as described in this document). The clinician must
also use reasonable judgment in deciding when
and where to use the product.
DRILLING DEPTH CONTROL
DRILL LASER DEPTH MARKINGS

DRILL STOPS

6mm
Depth

8mm
Depth

10mm
Depth

12mm
Depth

14mm
Depth

FINAL DRILL DIAMETER AND DEPTH FOR VARIOUS BONE TYPES
BONE TYPE

D1

2.4MM
implant diameter
Final Drill
DRILL DEPTH
Diameter

2.1mm

Full

D2 / D3

D4

2.9MM
implant diameter
Final Drill
DRILL DEPTH
Diameter

2.4mm

Full

1.6mm

Depth 4mm
less than
implant
length

2.1mm

Depth 4mm
less than
implant
length

1.6mm

Depth 4mm
less than
implant
length

2.1mm

Depth 4mm
less than
implant
length

Bone type is a general classification. The overall bone quality must be assessed by the clinician through
treatment planning and at the time of surgery in order to create the appropriate osteotomy size to achieve
the desired insertion torque.

6
DRILLING SEQUENCE EXAMPLES
PLACEMENT OF A 2.4MM x 12MM IMPLANT FLAPLESS SURGICAL PROCEDURE
2.4MM LASER DEPTH MARKINGS

2.4MM DRILL STOPS

2mm Soft Tissue

1.2mm Pilot
Full Depth
14mm Line

1.6mm
Full Depth
14mm Line

2.1mm
Full Depth
10mm Line

Tap Speed
Until Stop

Ratchet to
Full Depth

1.2mm Pilot
Full Depth
12mm
Drill Stop

1.6mm
Full Depth
12mm
Drill Stop

2.1mm
Full Depth
8mm
Drill Stop

Tap Speed
Until Stop

Ratchet to
Full Depth

2mm Soft Tissue

1.2mm Pilot
1.6mm
Full Depth 4mm Short
14mm Line 10mm Line

Tap Speed
Until Stop

Ratchet to
Full Depth

1.2mm Pilot
1.6mm
Full Depth 4mm Short
12mm
8mm
Drill Stop
Drill Stop

Tap Speed
Until Stop

Ratchet to
Full Depth

1.2mm Pilot
1.6mm
Full Depth 4mm Short
12mm
8mm
Drill Stop
Drill Stop

Tap Speed
Until Stop

Ratchet to
Full Depth

2mm Soft Tissue

7

1.2mm Pilot
1.6mm
Full Depth 4mm Short
14mm Line 10mm Line

Tap Speed
Until Stop

Ratchet to
Full Depth
DRILLING SEQUENCE EXAMPLES
(Continued)
PLACEMENT OF A 2.9MM x 12MM IMPLANT FLAPLESS SURGICAL PROCEDURE
2.9MM LASER DEPTH MARKINGS

2.9MM DRILL STOPS

2mm Soft Tissue

1.2mm Pilot
Full Depth
14mm Line

1.6mm
Full Depth
14mm Line

2.4mm
Full Depth
14mm Line

Tap Speed
Until Stop

Ratchet to
Full Depth

1.2mm Pilot
Full Depth
12mm
Drill Stop

1.6mm
Full Depth
12mm
Drill Stop

2.4mm
Full Depth
12mm
Drill Stop

Tap Speed
Until Stop

Ratchet to
Full Depth

1.2mm Pilot
Full Depth
12mm
Drill Stop

1.6mm
Full Depth
12mm
Drill Stop

2.1mm
4mm Short
8mm
Drill Stop

Tap Speed
Until Stop

Ratchet to
Full Depth

1.2mm Pilot
Full Depth
12mm
Drill Stop

1.6mm
Full Depth
12mm
Drill Stop

2.1mm
4mm Short
8mm
Drill Stop

Tap Speed
Until Stop

Ratchet to
Full Depth

2mm Soft Tissue

1.2mm Pilot
Full Depth
14mm Line

1.6mm
Full Depth
14mm Line

2.1mm
4mm Short
10mm Line

Tap Speed
Until Stop

Ratchet to
Full Depth

2mm Soft Tissue

1.2mm Pilot
1.6mm
Full Depth Full Depth
14mm Line 14mm Line

2.1mm
4mm Short
10mm Line

Tap Speed
Until Stop

Ratchet to
Full Depth

8
LOCATOR MALES & EXTENDED RANGE MALES
®

LOCATOR MALES

EXTENDED RANGE MALES

The unique Dual Retention innovation provides
the LOCATOR Attachment with a greater retention
surface area than ever before available with
other attachments.

Allows you to restore a non-parallel implant with
up to 20 degrees of angulation. This calculates to
an extensive 40 degrees of divergence between
two implants.

1.5 lbs
3 lbs
5 lbs
Dual retention, pivoting action provides resiliency
to maximize stability and longevity.

0 lbs

1 lbs
2 lbs
4 lbs
Extended Range Male’s pivoting action allows
for insertion with up to 40° total divergence.

a lodi male processing pack
is INCLUDED WITH EACH IMPLANT
Each processing pack has what you need to select
retention levels and address draw correction;
improving ease of denture placement and removal.

Denture
Cap
LODI Male
Processing Pack

9

Blue
1.5 lbs

Pink
3 lbs

Spacer Ring

Red
1 lbs
LOCATOR CORE TOOL
®

LOCATOR CORE TOOL
The LOCATOR Implant Attachment System features a Core Tool that contains 3-tools-in-1. This convenient
tool is used to carry and place the LOCATOR Attachment onto the implant, remove the LOCATOR Male,
and insert the male into the LOCATOR Denture Cap. Insert drivers for various types of torque wrenches are
available to achieve 30Ncm of torque.

REMOVAL
The Male Removal Tool has a sharp edge
on the end to catch and remove the
male from the Denture Cap.

INSERTION
The Male Seating Tool is used
to seat the LOCATOR Male.

PLACEMENT
The LOCATOR Abutment Driver with the
Abutment Holder Sleeve carries the attachment
securely and places it onto the implant.

Loosen the Male Removal Tool a full 3 turns counter
clockwise (you will see a visible gap).
To remove a LOCATOR nylon male from the Denture
Cap; simply insert the tip into the cap/male assembly
and push straight in to the bottom of the nylon male.
Then tilt the tool so that the sharp edge of the tip
will grab hold of the male and pull it out of the
Denture Cap.
To discard the nylon male from the tip on the Core
Tool; point the tool down and away from you and
tighten the Male Removal Tool clockwise back onto
the Core Tool. This will activate the removal pin and
dislodge the nylon male from the tip end of the Male
Removal Tool.
Separate the Male Removal Tool section from the
LOCATOR Core Tool and use the Male Seating Tool
end of the remaining two sections to place a new
nylon male into the empty Denture Cap.

10
PRE-SURGICAL TREATMENT PLANNING
Evaluate available bone width for implant positions
by using the index finger/thumb technique or a ridge
mapping instrument.

Measure the gingiva depth at each implant location
using a perio probe to determine the proper depth
measurement line on the osteotomy preparation drills
and to select the LOCATOR® Attachment cuff height of
either 2.5mm or 4mm.

3A-3B

Radiograph the arch using a panoramic
radiograph or CBCT to determine available bone
height for implant positions. A radiographic template
with measurement balls may be fabricated to assist
with determining dimensions.

A

B

11
PRE-SURGICAL TREATMENT PLANNING
(Continued)
Determine if the patient’s existing denture(s) will be
used or if new ones will be fabricated. If a new denture
is made, follow the conventional process. Have the
patient wear the new denture for 2 weeks prior to
implant placement.

A surgical guide for implant placement may be
fabricated from the patient’s existing or new denture
prior to surgery.

12
IMPLANT PLACEMENT
After patient selection and evaluation protocols have been
completed, the number of implants required is determined
and discussed with the patient. The patient’s denture is
then fabricated or modified, followed by identification of
appropriate implant sites. For mandibular placement, the
position of the nerve and bone quality must be taken into
consideration. For maxillary placement, bone volume, bone
quality and sinus location must be taken into consideration.
EXAMPLE IS PLACEMENT OF FOUR 2.9MM X 10MM IMPLANTS IN TYPE D1 Bone

Using the surgical guide or by free hand, mark the
implant osteotomy locations by drilling through the
gingiva and into the bone crest approximately 6mm
using the 1.2mm Pilot Drill. Note the gingival depth. The
recommended drill speed is 1200-1500 rpm.

Remove the gingiva cores at each site using the
Rotary Tissue Punch. Place the guide pin portion into the
pilot holes and rotate to cut away the gingiva. Rotate the
Tissue Punch to the laser depth mark indicated from the
gingiva depth measurement. The recommended speed is
up to a maximum of 800 rpm.

13
IMPLANT PLACEMENT
(Continued)
EXAMPLE IS PLACEMENT OF
FOUR 2.9MM X 10MM IMPLANTS
IN TYPE D1 Bone • CONTINUED

3A-3B

A

Place the 1.2mm diameter (small) end of the
Direction Indicator into the pilot drill osteotomies to
verify the proper angulation. Place the proper length drill
stop onto the 1.2mm Pilot Drill according to the desired
drill depth. Or, drill to the proper laser depth marking on
the drill calculated by adding the desired drill depth and
tissue depth. The recommended drill speed is 1200-1500
rpm. Continue osteotomy preparation to the desired
depth at each implant site.

B

4A-4B

Place the proper length drill stop onto the
1.6mm drill according to the desired drill depth. Or,
drill to the proper laser depth marking on the drill
calculated by adding the desired drill depth and tissue
depth. The recommended drill speed is 1200-1500 rpm.
Continue osteotomy preparation to the desired depth
at each implant site.

A

B

14
IMPLANT PLACEMENT
(Continued)
EXAMPLE IS PLACEMENT OF
FOUR 2.9MM X 10MM IMPLANTS
IN TYPE D1 Bone • CONTINUED

5A-5B

Place the 1.6mm diameter end of the
Directional Indicator into the osteotomies to verify the
proper angulation. Place the proper length drill stop
onto the 2.4mm drill according to the desired drill
depth. Or, drill to the proper laser depth marking on the
drill calculated by adding desired drill depth and tissue
depth. The recommended drill speed is 1200-1500 rpm.
Continue osteotomy preparation to the desired depth
at each implant site.

A

B

6A-6B

Remove the
implant package from the
box and peel back the
tyvek seal from the plastic
tray. Drop the implant vial
on the sterile tray. The
contents of the plastic
tray are sterile and should
only contact components
within the sterile field.

A

B

7A-7B

Remove the cap from the implant vial and
do not discard. The LOCATOR® Attachment is in the cap.
Set the drilling unit speed at 50rpm and the placement
torque at 35Ncm. Place the Implant Driver in the
handpiece. Insert the Implant Driver onto the hex on the
top of the implant and press down to frictionally engage.
The bottom of the driver should contact the attachment
seating surface and fully engage the entire length of the
implant hex.

A

15

B
IMPLANT PLACEMENT
(Continued)
EXAMPLE IS PLACEMENT OF
FOUR 2.9MM X 10MM IMPLANTS
IN TYPE D1 Bone • CONTINUED
Remove the implant from the vial in a straight
out motion.

Carry the implant to the mouth, place it into the
osteotomy and tap into place at 50 rpm. Avoid sudden
movement and/or touching an external object which
may dislodge the implant from the driver.
Warning: Discard and do not use an implant that has
been dropped in a non-sterile area and replace with a
new sterile implant.

10A-10B

When the
drilling unit stops tapping
the implant, place
the Torque Indicating
Ratchet Wrench Insert
into the Torque Ratchet
Wrench. Place the
Implant Driver (short or
long) into the insert.

11

Position the Implant Driver onto the hex on the
top of the implant and verify that it is fully engaged.
Slowly ratchet the implant to full depth. If final seating
torque measures 30Ncm or above, the implant may be
immediately loaded at the discretion of the clinician.
If the final seating torque measures below 30Ncm,
the denture acrylic should be relieved and a soft liner
placed around the LOCATOR® Attachments during the
integration period. Implant insertion torque should not
exceed 70Ncm. If 70Ncm of torque is reached, prior
to full seating, the implant should be removed and
the osteotomy should be enlarged.

16
LOCATOR AttachmenT PLACEMENT
®

1A-1B Open the flip cap on the top of the vial
cap and remove the LOCATOR Attachment. Place
the Abutment Holder Sleeve onto the LOCATOR
Abutment Driver. Place the attachment into the
Abutment Holder Sleeve to securely carry it to
the mouth.

A

B

2A-2B Thread the LOCATOR Attachment onto
the implant until finger tight. If the implant placement
torque was 30Ncm or greater, the attachments may be
tightened to the recommended torque level of 30Ncm.
If the implant placement torque did not reach 30Ncm,
the attachments should only be hand tightened. Place
the LOCATOR Attachment Torque Driver Insert into
the Torque Ratchet Wrench. Insert the driver into the
attachment and verify it is fully engaged. Torque the
attachments to 30Ncm.
A

B
If the implant placement torque was 30Ncm or
greater, the implants may at the discretion of the clinician
be immediately loaded. Continue on with steps for
processing the LOCATOR Denture Caps and Males into
the denture. If the implant placement torque was less
than 30Ncm, relieve the denture acrylic and place a soft
liner in the denture around the LOCATOR Attachments
during the integration period.

17
PROCESSING LOCATOR DENTURE CAPS AND MALES
INTO THE DENTURE
®

DIRECT TECHNIQUE: CHAIRSIDE PROCESSING
(NEW OR EXISTING DENTURE)
If the implants were not immediately loaded, the
LOCATOR Attachments must be torque tightened.
Place the LOCATOR Abutment Torque Driver Insert
into the Torque Ratchet Wrench. Insert the driver into
the attachment and verify that it is fully engaged.
Torque the attachments to 30Ncm.

Place a White Block Out Spacer Ring around each
attachment and press it down to the tissue. Place a
Denture Cap with a Black Processing Male inside of it
onto each attachment and press down firmly.

Apply fit check marking paste inside of the denture.
Insert it into the mouth in position over the Denture Caps
to mark the areas where the denture will need to be
relieved to allow space for the caps to be picked up.

Relieve the areas marked with an acrylic bur. Try in
the denture to verify that the Denture Caps are not in
contact with the acrylic in any area. Cut lingual/palatal
vent windows in the denture to visualize full seating and
for an excess acrylic vent.

18
PROCESSING LOCATOR DENTURE CAPS AND MALES
INTO THE DENTURE (CONTINUED)
®

DIRECT TECHNIQUE: CHAIRSIDE PROCESSING
(CONTINUED)
An autopolymerizing or light cure acrylic resin
may be used to pick up the caps. Dry the Denture
Caps. Apply a small amount of acrylic around the
circumference of each cap. Place acrylic into the relief
areas of the denture and seat it over the caps and onto
the tissue. Have the patient close into occlusion and hold
while the acrylic sets.

Disengage the denture from the LOCATOR
Attachments and remove from the mouth. Verify that
the Denture Caps have been securely picked up in the
denture. Fill any voids and polish the denture.

7A-7B

See the LOCATOR Core Tool instructions on
page 10. Remove the Black Processing Male using the
Male Removal Tool. Place the selected final male into
each Denture Cap using the Male Insertion Tool. It is
recommended to place the least retentive male to begin
with. See the male retention chart on page 9.

A

B
Place the denture in the mouth and press down
to engage the males on the LOCATOR Attachments.
Verify the occlusion. Instruct the patient on how to
remove and insert the denture. If the retention is not
satisfactory, remove the males and replace with the
next level of retention. See the male retention chart
on page 9. Instruct the patient on proper home care
maintenance and required recall visits.

19
PROCESSING LOCATOR DENTURE CAPS AND MALES
INTO THE DENTURE (CONTINUED)
®

INDIRECT TECHNIQUE:
LABORATORY PROCESSING (Appt. 1)
If the implants were not immediately loaded, the
LOCATOR® Attachments must be torque tightened.
Place the LOCATOR Abutment Torque Driver Insert into
the Torque Ratchet Wrench. Insert the driver into the
attachment and verify that it is fully engaged. Torque the
attachments to 30Ncm.

A stock or custom impression tray may be used.
Ensure on each that there is enough space for the 4mm
height of the LOCATOR Impression Copings.

4mm

Place a LOCATOR Impression Coping on each
attachment and press down firmly. Syringe a medium
body impression material around the circumference of
each coping. Fill the impression tray and insert it over the
copings and onto the tissue. Allow the material to set.
Remove the impression and verify that the copings have
been securely picked up inside of it.

Press the LOCATOR Analogs into each impression
coping. Send the impression to the laboratory.

20
PROCESSING LOCATOR DENTURE CAPS AND MALES
INTO THE DENTURE (CONTINUED)
®

INDIRECT TECHNIQUE:
LABORATORY PROCESSING (CONTINUED)
LAB STEP 1
Verify that the analogs are secure in the impression
copings. Pour die stone into the impression to fabricate
the master cast.

Fabricate the baseplate and wax rim on the cast
for the bite registration. The Denture Caps with Black
Processing Males may be processed into the baseplate to
provide stabilization during record making.

Bite Records (Appt. 2)
Place the wax rim into the mouth and make the bite
records. Take an impression of the opposing arch and
pour the cast. Select a shade for the denture teeth.

Lab Step 2
Articulate the master cast with the opposing using
the bite records. Set the denture teeth on the baseplate
and fabricate a wax denture for try in.

21
PROCESSING LOCATOR DENTURE CAPS AND MALES
INTO THE DENTURE (CONTINUED)
®

Denture Try In (Appt. 3)
Place the wax denture into the mouth and verify the
esthetics, phonetics and occlusion.

Lab Step 3
Wax and flask the denture for processing. Separate
the flask and boil away all wax. Place the Denture Caps
with Black Processing Males on the analogs and press
down firmly. Place the cast back into the flask and verify
that there is no contact with the teeth. Close the flask
and process the denture. Remove the denture from the
flask, finish and polish.

2A-2B

See the LOCATOR Core Tool instructions on
page 10. Remove the Black Processing Male using the
Male Removal Tool.Place the selected final male into
each Denture Cap using the Male Insertion Tool. It is
recommended to place the least retentive male to begin
with. See the male retention chart on page 9.

A

B

DELIVERY (Appt. 4)
Place the denture in the mouth and press down
to engage the males on the LOCATOR Attachments.
Verify the occlusion. Instruct the patient on how to
remove and insert the denture. If the retention is not
satisfactory, remove the males and replace with the
next level of retention. See the male retention chart
on page 9. Instruct the patient on proper home care
maintenance and required recall visits.

22
EXPLANATION OF SYMBOLS ON
OUTER PACKAGING LABELS
SYMBOL

STANDARD

ISO 980

EXPLANATION
OF SYMBOL

Symbol for
“DO NOT REUSE”

LOCATION OF
SYMBOL

ON THE
FOLLOWING LABELS:

For single-use. Use only once.

Symbol for
“BATCH CODE”
ISO 980

1) L9127-XXXX 2) L7409-XXXX
3) L9108-XXXX

ON THE
FOLLOWING LABELS:

This symbol shall be accompanied
by the manufacturer’s batch code
or lot code. The batch/lot code
shall be adjacent to the symbol.

4) L9127-XXXX 5) L7409-XXXX
6) L9108-XXXX 7) L9115-07420
8) L9912-07362 9) L9109-XXXX

Symbol for
“MANUFACTURER”
ISO 980

ISO 980

ON THE
FOLLOWING LABELS:

This symbol shall be accompanied
by the manufacturer name
(Zest Anchors) and address
(2061 Wineridge Place, Escondido,
CA 92029); adjacent to the symbol.

10) L9127-XXXX 11) L7409-XXXX
12) L9108-XXXX 13) L9115-07420
14) L9912-07362 15) L9109-XXXX

Symbol for
“STERILIZED USING
IRRADIATION”

ON THE
FOLLOWING LABELS:

NOTE: Refers to Implant/Attachment
sterilized packaging only.

Symbol for
“CATALOGUE NUMBER”
ISO 980

23

16) L9127-XXXX 17) L7409-XXXX

ON THE
FOLLOWING LABELS:

The product catalogue number
shall be after or below the symbol
adjacent to it.

18) L9127-XXXX 19) L7409-XXXX
20) L9108-XXXX
NOTES
________________________________________________________________________________________________________________________________________________________________________________________
________________________________________________________________________________________________________________________________________________________________________________________
________________________________________________________________________________________________________________________________________________________________________________________
________________________________________________________________________________________________________________________________________________________________________________________
________________________________________________________________________________________________________________________________________________________________________________________
________________________________________________________________________________________________________________________________________________________________________________________
________________________________________________________________________________________________________________________________________________________________________________________
________________________________________________________________________________________________________________________________________________________________________________________
________________________________________________________________________________________________________________________________________________________________________________________
________________________________________________________________________________________________________________________________________________________________________________________
________________________________________________________________________________________________________________________________________________________________________________________
________________________________________________________________________________________________________________________________________________________________________________________
________________________________________________________________________________________________________________________________________________________________________________________
________________________________________________________________________________________________________________________________________________________________________________________
________________________________________________________________________________________________________________________________________________________________________________________
________________________________________________________________________________________________________________________________________________________________________________________
________________________________________________________________________________________________________________________________________________________________________________________
________________________________________________________________________________________________________________________________________________________________________________________
________________________________________________________________________________________________________________________________________________________________________________________
________________________________________________________________________________________________________________________________________________________________________________________
________________________________________________________________________________________________________________________________________________________________________________________
________________________________________________________________________________________________________________________________________________________________________________________
________________________________________________________________________________________________________________________________________________________________________________________
________________________________________________________________________________________________________________________________________________________________________________________
________________________________________________________________________________________________________________________________________________________________________________________
________________________________________________________________________________________________________________________________________________________________________________________
________________________________________________________________________________________________________________________________________________________________________________________
________________________________________________________________________________________________________________________________________________________________________________________
________________________________________________________________________________________________________________________________________________________________________________________
________________________________________________________________________________________________________________________________________________________________________________________
________________________________________________________________________________________________________________________________________________________________________________________

24
ZEST ANCHORS
2061 Wineridge Place
Escondido, CA 92029
USA

Tel: 1.855.868.LODI (5634)
Fax: 760.743.7975
Email: sales@zestanchors.com
www.zestanchors.com

Distributed by:

©2012 ZEST Anchors LLC. All rights reserved. ZEST and LOCATOR are registered trademarks of ZEST IP Holdings, LLC.

L8019-TM

REV C 2/2013

Más contenido relacionado

La actualidad más candente

soloLAB operation and build maintenance manual
soloLAB operation and build maintenance manualsoloLAB operation and build maintenance manual
soloLAB operation and build maintenance manualSolo Containment
 
Pinnacle EDT Training
Pinnacle EDT TrainingPinnacle EDT Training
Pinnacle EDT TrainingSerimRes
 
Flexible Isolator Options
Flexible Isolator OptionsFlexible Isolator Options
Flexible Isolator OptionsTim Cowher
 
Cleaningand sanitizingprocedure
Cleaningand sanitizingprocedure Cleaningand sanitizingprocedure
Cleaningand sanitizingprocedure Ka Jejen
 
Ot infection control rkch
Ot infection control rkchOt infection control rkch
Ot infection control rkchsabahjak
 
Fiat kobelco e27.2 sr mini crawler excavator service repair manual
Fiat kobelco e27.2 sr mini crawler excavator service repair manualFiat kobelco e27.2 sr mini crawler excavator service repair manual
Fiat kobelco e27.2 sr mini crawler excavator service repair manualfujdjffjjskkemem
 
Fiat kobelco e40.2 sr mini crawler excavator service repair manual
Fiat kobelco e40.2 sr mini crawler excavator service repair manualFiat kobelco e40.2 sr mini crawler excavator service repair manual
Fiat kobelco e40.2 sr mini crawler excavator service repair manualfjskksmem
 

La actualidad más candente (10)

soloLAB operation and build maintenance manual
soloLAB operation and build maintenance manualsoloLAB operation and build maintenance manual
soloLAB operation and build maintenance manual
 
ANES 1502 - M13 PPT - Disinfection
ANES 1502 - M13 PPT - DisinfectionANES 1502 - M13 PPT - Disinfection
ANES 1502 - M13 PPT - Disinfection
 
Operation theatre surveillance by Dr.T.V.Rao MD
Operation theatre surveillance by Dr.T.V.Rao MDOperation theatre surveillance by Dr.T.V.Rao MD
Operation theatre surveillance by Dr.T.V.Rao MD
 
Pinnacle EDT Training
Pinnacle EDT TrainingPinnacle EDT Training
Pinnacle EDT Training
 
Flexible Isolator Options
Flexible Isolator OptionsFlexible Isolator Options
Flexible Isolator Options
 
Cleaningand sanitizingprocedure
Cleaningand sanitizingprocedure Cleaningand sanitizingprocedure
Cleaningand sanitizingprocedure
 
Ot infection control rkch
Ot infection control rkchOt infection control rkch
Ot infection control rkch
 
Fiat kobelco e27.2 sr mini crawler excavator service repair manual
Fiat kobelco e27.2 sr mini crawler excavator service repair manualFiat kobelco e27.2 sr mini crawler excavator service repair manual
Fiat kobelco e27.2 sr mini crawler excavator service repair manual
 
Designing of aseptic area
Designing of aseptic areaDesigning of aseptic area
Designing of aseptic area
 
Fiat kobelco e40.2 sr mini crawler excavator service repair manual
Fiat kobelco e40.2 sr mini crawler excavator service repair manualFiat kobelco e40.2 sr mini crawler excavator service repair manual
Fiat kobelco e40.2 sr mini crawler excavator service repair manual
 

Destacado

implant supported complete denture/ cosmetic dentistry training
implant supported complete denture/ cosmetic dentistry trainingimplant supported complete denture/ cosmetic dentistry training
implant supported complete denture/ cosmetic dentistry trainingIndian dental academy
 
Implant Supported Over Denture By Dr pitzak
 Implant Supported Over Denture By Dr pitzak Implant Supported Over Denture By Dr pitzak
Implant Supported Over Denture By Dr pitzakWestern Dental
 
Implants /certified fixed orthodontic courses by Indian dental academy
Implants   /certified fixed orthodontic courses by Indian dental academy Implants   /certified fixed orthodontic courses by Indian dental academy
Implants /certified fixed orthodontic courses by Indian dental academy Indian dental academy
 
Investigations in prosthodontics/ oral surgery courses  
Investigations in prosthodontics/ oral surgery courses  Investigations in prosthodontics/ oral surgery courses  
Investigations in prosthodontics/ oral surgery courses  Indian dental academy
 
S mini implant catalog (2015)
S mini implant catalog (2015)S mini implant catalog (2015)
S mini implant catalog (2015)Kei Lim
 
Implant supported overdentures
Implant supported overdenturesImplant supported overdentures
Implant supported overdenturesMurtaza Kaderi
 
Implant supported overdentures/dental courses
Implant supported overdentures/dental coursesImplant supported overdentures/dental courses
Implant supported overdentures/dental coursesIndian dental academy
 
Complete overdenture(3)
Complete overdenture(3)Complete overdenture(3)
Complete overdenture(3)Ahmed Samir
 
Dental Implants surgical considerations / implant dentistry course/ implant d...
Dental Implants surgical considerations / implant dentistry course/ implant d...Dental Implants surgical considerations / implant dentistry course/ implant d...
Dental Implants surgical considerations / implant dentistry course/ implant d...Indian dental academy
 
types and classification of dental implants
types and classification of dental implantstypes and classification of dental implants
types and classification of dental implantsDesa Ghanavi
 
Attachments in implant retained overdentures/ cosmetic dentistry training
Attachments in implant retained overdentures/ cosmetic dentistry trainingAttachments in implant retained overdentures/ cosmetic dentistry training
Attachments in implant retained overdentures/ cosmetic dentistry trainingIndian dental academy
 
State of the Word 2011
State of the Word 2011State of the Word 2011
State of the Word 2011photomatt
 

Destacado (16)

implant supported complete denture/ cosmetic dentistry training
implant supported complete denture/ cosmetic dentistry trainingimplant supported complete denture/ cosmetic dentistry training
implant supported complete denture/ cosmetic dentistry training
 
Implant Supported Over Denture By Dr pitzak
 Implant Supported Over Denture By Dr pitzak Implant Supported Over Denture By Dr pitzak
Implant Supported Over Denture By Dr pitzak
 
Implants /certified fixed orthodontic courses by Indian dental academy
Implants   /certified fixed orthodontic courses by Indian dental academy Implants   /certified fixed orthodontic courses by Indian dental academy
Implants /certified fixed orthodontic courses by Indian dental academy
 
Investigations in prosthodontics/ oral surgery courses  
Investigations in prosthodontics/ oral surgery courses  Investigations in prosthodontics/ oral surgery courses  
Investigations in prosthodontics/ oral surgery courses  
 
S mini implant catalog (2015)
S mini implant catalog (2015)S mini implant catalog (2015)
S mini implant catalog (2015)
 
Implant supported overdentures
Implant supported overdenturesImplant supported overdentures
Implant supported overdentures
 
Overdentures and attachments part 1
Overdentures and attachments part 1Overdentures and attachments part 1
Overdentures and attachments part 1
 
Implant supported overdentures/dental courses
Implant supported overdentures/dental coursesImplant supported overdentures/dental courses
Implant supported overdentures/dental courses
 
Overdentures
OverdenturesOverdentures
Overdentures
 
Complete overdenture(3)
Complete overdenture(3)Complete overdenture(3)
Complete overdenture(3)
 
Dental Implants surgical considerations / implant dentistry course/ implant d...
Dental Implants surgical considerations / implant dentistry course/ implant d...Dental Implants surgical considerations / implant dentistry course/ implant d...
Dental Implants surgical considerations / implant dentistry course/ implant d...
 
Overdenture
OverdentureOverdenture
Overdenture
 
types and classification of dental implants
types and classification of dental implantstypes and classification of dental implants
types and classification of dental implants
 
Attachments in implant retained overdentures/ cosmetic dentistry training
Attachments in implant retained overdentures/ cosmetic dentistry trainingAttachments in implant retained overdentures/ cosmetic dentistry training
Attachments in implant retained overdentures/ cosmetic dentistry training
 
State of the Word 2011
State of the Word 2011State of the Word 2011
State of the Word 2011
 
Slideshare ppt
Slideshare pptSlideshare ppt
Slideshare ppt
 

Similar a L8019 tm lodi_manual_revc

User guide universal prosthetic kit -151207
User guide   universal prosthetic kit -151207User guide   universal prosthetic kit -151207
User guide universal prosthetic kit -151207Kei Lim
 
Cleaning And Sterilization Protocol - Arsenal Ankle Plating System | Enovis
Cleaning And Sterilization Protocol - Arsenal Ankle Plating System | EnovisCleaning And Sterilization Protocol - Arsenal Ankle Plating System | Enovis
Cleaning And Sterilization Protocol - Arsenal Ankle Plating System | EnovisDJO®
 
CARDIOVASCULAR INSTRUMENTS CATALOGUE
CARDIOVASCULAR INSTRUMENTS CATALOGUE CARDIOVASCULAR INSTRUMENTS CATALOGUE
CARDIOVASCULAR INSTRUMENTS CATALOGUE zackwilliam
 
Chapter 12 complex surgical instruments
Chapter 12  complex surgical instrumentsChapter 12  complex surgical instruments
Chapter 12 complex surgical instrumentslahcmultimedia
 
Chapter 12 complex surgical instruments
Chapter 12  complex surgical instrumentsChapter 12  complex surgical instruments
Chapter 12 complex surgical instrumentslahcmultimedia
 
Infection rbna ystr !
Infection rbna ystr !Infection rbna ystr !
Infection rbna ystr !Amal Magdy
 
ASA DENTAL ITALY Reprocessing
ASA DENTAL ITALY Reprocessing ASA DENTAL ITALY Reprocessing
ASA DENTAL ITALY Reprocessing Shimaa Hamdy
 
FR Kit user guide
FR Kit user guideFR Kit user guide
FR Kit user guideKei Lim
 
Reusable Medical Devices (SOP).docx
Reusable Medical Devices (SOP).docxReusable Medical Devices (SOP).docx
Reusable Medical Devices (SOP).docxanjalatchi
 
Jay 1 neutral user manuel
Jay 1 neutral user manuelJay 1 neutral user manuel
Jay 1 neutral user manuelJeroxxx
 
BASICS TO FOOT CARE ppt FINAL2014.pdf
BASICS TO FOOT CARE ppt FINAL2014.pdfBASICS TO FOOT CARE ppt FINAL2014.pdf
BASICS TO FOOT CARE ppt FINAL2014.pdfMarcAndreo2
 
STAR™ Ankle Instruments, Accessories, and Instrument Set - DJO®
STAR™ Ankle Instruments, Accessories, and Instrument Set - DJO®				STAR™ Ankle Instruments, Accessories, and Instrument Set - DJO®
STAR™ Ankle Instruments, Accessories, and Instrument Set - DJO® DJO®
 
How to Properly Clean and Maintain Your Artery Forceps.pdf
How to Properly Clean and Maintain Your Artery Forceps.pdfHow to Properly Clean and Maintain Your Artery Forceps.pdf
How to Properly Clean and Maintain Your Artery Forceps.pdfnazmedsmssdnbhd
 
ACM user guide
ACM user guideACM user guide
ACM user guideKei Lim
 
CLEANING AND SANITATION STANDARD
CLEANING AND SANITATION  STANDARDCLEANING AND SANITATION  STANDARD
CLEANING AND SANITATION STANDARDicmseafood
 
Practice Guidelines Endocopic disinfection & Reprocessing
Practice Guidelines Endocopic disinfection & ReprocessingPractice Guidelines Endocopic disinfection & Reprocessing
Practice Guidelines Endocopic disinfection & ReprocessingHossam Ghoneim
 

Similar a L8019 tm lodi_manual_revc (20)

User guide universal prosthetic kit -151207
User guide   universal prosthetic kit -151207User guide   universal prosthetic kit -151207
User guide universal prosthetic kit -151207
 
Cleaning And Sterilization Protocol - Arsenal Ankle Plating System | Enovis
Cleaning And Sterilization Protocol - Arsenal Ankle Plating System | EnovisCleaning And Sterilization Protocol - Arsenal Ankle Plating System | Enovis
Cleaning And Sterilization Protocol - Arsenal Ankle Plating System | Enovis
 
CARDIOVASCULAR INSTRUMENTS CATALOGUE
CARDIOVASCULAR INSTRUMENTS CATALOGUE CARDIOVASCULAR INSTRUMENTS CATALOGUE
CARDIOVASCULAR INSTRUMENTS CATALOGUE
 
Chapter 12 complex surgical instruments
Chapter 12  complex surgical instrumentsChapter 12  complex surgical instruments
Chapter 12 complex surgical instruments
 
Chapter 12 complex surgical instruments
Chapter 12  complex surgical instrumentsChapter 12  complex surgical instruments
Chapter 12 complex surgical instruments
 
Infection rbna ystr !
Infection rbna ystr !Infection rbna ystr !
Infection rbna ystr !
 
Jet hand dryer mitsubushi catalog
 Jet hand dryer mitsubushi  catalog  Jet hand dryer mitsubushi  catalog
Jet hand dryer mitsubushi catalog
 
ASA DENTAL ITALY Reprocessing
ASA DENTAL ITALY Reprocessing ASA DENTAL ITALY Reprocessing
ASA DENTAL ITALY Reprocessing
 
NeoBiotech - FR kit user guide
NeoBiotech - FR kit user guideNeoBiotech - FR kit user guide
NeoBiotech - FR kit user guide
 
FR Kit user guide
FR Kit user guideFR Kit user guide
FR Kit user guide
 
Reusable Medical Devices (SOP).docx
Reusable Medical Devices (SOP).docxReusable Medical Devices (SOP).docx
Reusable Medical Devices (SOP).docx
 
Endoscope cleaning
Endoscope cleaningEndoscope cleaning
Endoscope cleaning
 
Jay 1 neutral user manuel
Jay 1 neutral user manuelJay 1 neutral user manuel
Jay 1 neutral user manuel
 
BASICS TO FOOT CARE ppt FINAL2014.pdf
BASICS TO FOOT CARE ppt FINAL2014.pdfBASICS TO FOOT CARE ppt FINAL2014.pdf
BASICS TO FOOT CARE ppt FINAL2014.pdf
 
STAR™ Ankle Instruments, Accessories, and Instrument Set - DJO®
STAR™ Ankle Instruments, Accessories, and Instrument Set - DJO®				STAR™ Ankle Instruments, Accessories, and Instrument Set - DJO®
STAR™ Ankle Instruments, Accessories, and Instrument Set - DJO®
 
How to Properly Clean and Maintain Your Artery Forceps.pdf
How to Properly Clean and Maintain Your Artery Forceps.pdfHow to Properly Clean and Maintain Your Artery Forceps.pdf
How to Properly Clean and Maintain Your Artery Forceps.pdf
 
ACM user guide
ACM user guideACM user guide
ACM user guide
 
ANES 1502 - M13 PPT: Complex Surgical Instruments
ANES 1502 - M13 PPT: Complex Surgical InstrumentsANES 1502 - M13 PPT: Complex Surgical Instruments
ANES 1502 - M13 PPT: Complex Surgical Instruments
 
CLEANING AND SANITATION STANDARD
CLEANING AND SANITATION  STANDARDCLEANING AND SANITATION  STANDARD
CLEANING AND SANITATION STANDARD
 
Practice Guidelines Endocopic disinfection & Reprocessing
Practice Guidelines Endocopic disinfection & ReprocessingPractice Guidelines Endocopic disinfection & Reprocessing
Practice Guidelines Endocopic disinfection & Reprocessing
 

Más de Centric Learning

General dentistry an evaluation and adjustment method article
General dentistry an evaluation and adjustment method articleGeneral dentistry an evaluation and adjustment method article
General dentistry an evaluation and adjustment method articleCentric Learning
 
Preventing anterior open bites final
Preventing anterior open bites finalPreventing anterior open bites final
Preventing anterior open bites finalCentric Learning
 
Pala beta test impression instructions
Pala beta test impression instructionsPala beta test impression instructions
Pala beta test impression instructionsCentric Learning
 
Pala beta test presentation
Pala beta test presentationPala beta test presentation
Pala beta test presentationCentric Learning
 
2567 pala digital_dentures_manual_v3_singles
2567 pala digital_dentures_manual_v3_singles2567 pala digital_dentures_manual_v3_singles
2567 pala digital_dentures_manual_v3_singlesCentric Learning
 
Secchi flyer - Denver, CO 2014-10-17
Secchi flyer - Denver, CO 2014-10-17Secchi flyer - Denver, CO 2014-10-17
Secchi flyer - Denver, CO 2014-10-17Centric Learning
 
Burris flyer - Tunica, MS 11-14-2014
Burris flyer - Tunica, MS 11-14-2014Burris flyer - Tunica, MS 11-14-2014
Burris flyer - Tunica, MS 11-14-2014Centric Learning
 
Gammichia conservative dentistry
Gammichia conservative dentistryGammichia conservative dentistry
Gammichia conservative dentistryCentric Learning
 
Demi ultra rt deck may 2014 under curing and cure technology
Demi ultra rt deck may 2014 under curing and cure technologyDemi ultra rt deck may 2014 under curing and cure technology
Demi ultra rt deck may 2014 under curing and cure technologyCentric Learning
 
Mtm simple lingual flyer owen-2014-09-19_10-10
Mtm simple lingual flyer   owen-2014-09-19_10-10Mtm simple lingual flyer   owen-2014-09-19_10-10
Mtm simple lingual flyer owen-2014-09-19_10-10Centric Learning
 
Article shipley increasing rate of aligner progression
Article shipley increasing rate of aligner progressionArticle shipley increasing rate of aligner progression
Article shipley increasing rate of aligner progressionCentric Learning
 
Article teixeira effects of micro-osteoperforation
Article teixeira effects of micro-osteoperforationArticle teixeira effects of micro-osteoperforation
Article teixeira effects of micro-osteoperforationCentric Learning
 
Pro dentist spring 2014 game changers-graham
Pro dentist spring 2014 game changers-grahamPro dentist spring 2014 game changers-graham
Pro dentist spring 2014 game changers-grahamCentric Learning
 

Más de Centric Learning (20)

Lets growtooth
Lets growtoothLets growtooth
Lets growtooth
 
General dentistry an evaluation and adjustment method article
General dentistry an evaluation and adjustment method articleGeneral dentistry an evaluation and adjustment method article
General dentistry an evaluation and adjustment method article
 
Preventing anterior open bites final
Preventing anterior open bites finalPreventing anterior open bites final
Preventing anterior open bites final
 
Pala beta test impression instructions
Pala beta test impression instructionsPala beta test impression instructions
Pala beta test impression instructions
 
Pala beta test presentation
Pala beta test presentationPala beta test presentation
Pala beta test presentation
 
2567 pala digital_dentures_manual_v3_singles
2567 pala digital_dentures_manual_v3_singles2567 pala digital_dentures_manual_v3_singles
2567 pala digital_dentures_manual_v3_singles
 
Secchi flyer - Denver, CO 2014-10-17
Secchi flyer - Denver, CO 2014-10-17Secchi flyer - Denver, CO 2014-10-17
Secchi flyer - Denver, CO 2014-10-17
 
Burris flyer - Tunica, MS 11-14-2014
Burris flyer - Tunica, MS 11-14-2014Burris flyer - Tunica, MS 11-14-2014
Burris flyer - Tunica, MS 11-14-2014
 
Gammichia conservative dentistry
Gammichia conservative dentistryGammichia conservative dentistry
Gammichia conservative dentistry
 
2 irishdentaljournal
2  irishdentaljournal2  irishdentaljournal
2 irishdentaljournal
 
19, 20 dm_dec_clin_donley
19, 20 dm_dec_clin_donley19, 20 dm_dec_clin_donley
19, 20 dm_dec_clin_donley
 
Sweepstakes rules
Sweepstakes rulesSweepstakes rules
Sweepstakes rules
 
Rr58 kerr demiultra-efile
Rr58 kerr demiultra-efileRr58 kerr demiultra-efile
Rr58 kerr demiultra-efile
 
Demi ultra rt deck may 2014 under curing and cure technology
Demi ultra rt deck may 2014 under curing and cure technologyDemi ultra rt deck may 2014 under curing and cure technology
Demi ultra rt deck may 2014 under curing and cure technology
 
Freeland flyer 2014-08-22
Freeland flyer 2014-08-22Freeland flyer 2014-08-22
Freeland flyer 2014-08-22
 
Mtm simple lingual flyer owen-2014-09-19_10-10
Mtm simple lingual flyer   owen-2014-09-19_10-10Mtm simple lingual flyer   owen-2014-09-19_10-10
Mtm simple lingual flyer owen-2014-09-19_10-10
 
Scagd save the date
Scagd save the dateScagd save the date
Scagd save the date
 
Article shipley increasing rate of aligner progression
Article shipley increasing rate of aligner progressionArticle shipley increasing rate of aligner progression
Article shipley increasing rate of aligner progression
 
Article teixeira effects of micro-osteoperforation
Article teixeira effects of micro-osteoperforationArticle teixeira effects of micro-osteoperforation
Article teixeira effects of micro-osteoperforation
 
Pro dentist spring 2014 game changers-graham
Pro dentist spring 2014 game changers-grahamPro dentist spring 2014 game changers-graham
Pro dentist spring 2014 game changers-graham
 

Último

What's New in Teams Calling, Meetings and Devices March 2024
What's New in Teams Calling, Meetings and Devices March 2024What's New in Teams Calling, Meetings and Devices March 2024
What's New in Teams Calling, Meetings and Devices March 2024Stephanie Beckett
 
Gen AI in Business - Global Trends Report 2024.pdf
Gen AI in Business - Global Trends Report 2024.pdfGen AI in Business - Global Trends Report 2024.pdf
Gen AI in Business - Global Trends Report 2024.pdfAddepto
 
Are Multi-Cloud and Serverless Good or Bad?
Are Multi-Cloud and Serverless Good or Bad?Are Multi-Cloud and Serverless Good or Bad?
Are Multi-Cloud and Serverless Good or Bad?Mattias Andersson
 
Commit 2024 - Secret Management made easy
Commit 2024 - Secret Management made easyCommit 2024 - Secret Management made easy
Commit 2024 - Secret Management made easyAlfredo García Lavilla
 
"Debugging python applications inside k8s environment", Andrii Soldatenko
"Debugging python applications inside k8s environment", Andrii Soldatenko"Debugging python applications inside k8s environment", Andrii Soldatenko
"Debugging python applications inside k8s environment", Andrii SoldatenkoFwdays
 
How to write a Business Continuity Plan
How to write a Business Continuity PlanHow to write a Business Continuity Plan
How to write a Business Continuity PlanDatabarracks
 
Dev Dives: Streamline document processing with UiPath Studio Web
Dev Dives: Streamline document processing with UiPath Studio WebDev Dives: Streamline document processing with UiPath Studio Web
Dev Dives: Streamline document processing with UiPath Studio WebUiPathCommunity
 
Developer Data Modeling Mistakes: From Postgres to NoSQL
Developer Data Modeling Mistakes: From Postgres to NoSQLDeveloper Data Modeling Mistakes: From Postgres to NoSQL
Developer Data Modeling Mistakes: From Postgres to NoSQLScyllaDB
 
TrustArc Webinar - How to Build Consumer Trust Through Data Privacy
TrustArc Webinar - How to Build Consumer Trust Through Data PrivacyTrustArc Webinar - How to Build Consumer Trust Through Data Privacy
TrustArc Webinar - How to Build Consumer Trust Through Data PrivacyTrustArc
 
New from BookNet Canada for 2024: Loan Stars - Tech Forum 2024
New from BookNet Canada for 2024: Loan Stars - Tech Forum 2024New from BookNet Canada for 2024: Loan Stars - Tech Forum 2024
New from BookNet Canada for 2024: Loan Stars - Tech Forum 2024BookNet Canada
 
TeamStation AI System Report LATAM IT Salaries 2024
TeamStation AI System Report LATAM IT Salaries 2024TeamStation AI System Report LATAM IT Salaries 2024
TeamStation AI System Report LATAM IT Salaries 2024Lonnie McRorey
 
SAP Build Work Zone - Overview L2-L3.pptx
SAP Build Work Zone - Overview L2-L3.pptxSAP Build Work Zone - Overview L2-L3.pptx
SAP Build Work Zone - Overview L2-L3.pptxNavinnSomaal
 
Merck Moving Beyond Passwords: FIDO Paris Seminar.pptx
Merck Moving Beyond Passwords: FIDO Paris Seminar.pptxMerck Moving Beyond Passwords: FIDO Paris Seminar.pptx
Merck Moving Beyond Passwords: FIDO Paris Seminar.pptxLoriGlavin3
 
The State of Passkeys with FIDO Alliance.pptx
The State of Passkeys with FIDO Alliance.pptxThe State of Passkeys with FIDO Alliance.pptx
The State of Passkeys with FIDO Alliance.pptxLoriGlavin3
 
Advanced Computer Architecture – An Introduction
Advanced Computer Architecture – An IntroductionAdvanced Computer Architecture – An Introduction
Advanced Computer Architecture – An IntroductionDilum Bandara
 
Passkey Providers and Enabling Portability: FIDO Paris Seminar.pptx
Passkey Providers and Enabling Portability: FIDO Paris Seminar.pptxPasskey Providers and Enabling Portability: FIDO Paris Seminar.pptx
Passkey Providers and Enabling Portability: FIDO Paris Seminar.pptxLoriGlavin3
 
The Role of FIDO in a Cyber Secure Netherlands: FIDO Paris Seminar.pptx
The Role of FIDO in a Cyber Secure Netherlands: FIDO Paris Seminar.pptxThe Role of FIDO in a Cyber Secure Netherlands: FIDO Paris Seminar.pptx
The Role of FIDO in a Cyber Secure Netherlands: FIDO Paris Seminar.pptxLoriGlavin3
 
Use of FIDO in the Payments and Identity Landscape: FIDO Paris Seminar.pptx
Use of FIDO in the Payments and Identity Landscape: FIDO Paris Seminar.pptxUse of FIDO in the Payments and Identity Landscape: FIDO Paris Seminar.pptx
Use of FIDO in the Payments and Identity Landscape: FIDO Paris Seminar.pptxLoriGlavin3
 
Moving Beyond Passwords: FIDO Paris Seminar.pdf
Moving Beyond Passwords: FIDO Paris Seminar.pdfMoving Beyond Passwords: FIDO Paris Seminar.pdf
Moving Beyond Passwords: FIDO Paris Seminar.pdfLoriGlavin3
 

Último (20)

What's New in Teams Calling, Meetings and Devices March 2024
What's New in Teams Calling, Meetings and Devices March 2024What's New in Teams Calling, Meetings and Devices March 2024
What's New in Teams Calling, Meetings and Devices March 2024
 
Gen AI in Business - Global Trends Report 2024.pdf
Gen AI in Business - Global Trends Report 2024.pdfGen AI in Business - Global Trends Report 2024.pdf
Gen AI in Business - Global Trends Report 2024.pdf
 
Are Multi-Cloud and Serverless Good or Bad?
Are Multi-Cloud and Serverless Good or Bad?Are Multi-Cloud and Serverless Good or Bad?
Are Multi-Cloud and Serverless Good or Bad?
 
Commit 2024 - Secret Management made easy
Commit 2024 - Secret Management made easyCommit 2024 - Secret Management made easy
Commit 2024 - Secret Management made easy
 
DMCC Future of Trade Web3 - Special Edition
DMCC Future of Trade Web3 - Special EditionDMCC Future of Trade Web3 - Special Edition
DMCC Future of Trade Web3 - Special Edition
 
"Debugging python applications inside k8s environment", Andrii Soldatenko
"Debugging python applications inside k8s environment", Andrii Soldatenko"Debugging python applications inside k8s environment", Andrii Soldatenko
"Debugging python applications inside k8s environment", Andrii Soldatenko
 
How to write a Business Continuity Plan
How to write a Business Continuity PlanHow to write a Business Continuity Plan
How to write a Business Continuity Plan
 
Dev Dives: Streamline document processing with UiPath Studio Web
Dev Dives: Streamline document processing with UiPath Studio WebDev Dives: Streamline document processing with UiPath Studio Web
Dev Dives: Streamline document processing with UiPath Studio Web
 
Developer Data Modeling Mistakes: From Postgres to NoSQL
Developer Data Modeling Mistakes: From Postgres to NoSQLDeveloper Data Modeling Mistakes: From Postgres to NoSQL
Developer Data Modeling Mistakes: From Postgres to NoSQL
 
TrustArc Webinar - How to Build Consumer Trust Through Data Privacy
TrustArc Webinar - How to Build Consumer Trust Through Data PrivacyTrustArc Webinar - How to Build Consumer Trust Through Data Privacy
TrustArc Webinar - How to Build Consumer Trust Through Data Privacy
 
New from BookNet Canada for 2024: Loan Stars - Tech Forum 2024
New from BookNet Canada for 2024: Loan Stars - Tech Forum 2024New from BookNet Canada for 2024: Loan Stars - Tech Forum 2024
New from BookNet Canada for 2024: Loan Stars - Tech Forum 2024
 
TeamStation AI System Report LATAM IT Salaries 2024
TeamStation AI System Report LATAM IT Salaries 2024TeamStation AI System Report LATAM IT Salaries 2024
TeamStation AI System Report LATAM IT Salaries 2024
 
SAP Build Work Zone - Overview L2-L3.pptx
SAP Build Work Zone - Overview L2-L3.pptxSAP Build Work Zone - Overview L2-L3.pptx
SAP Build Work Zone - Overview L2-L3.pptx
 
Merck Moving Beyond Passwords: FIDO Paris Seminar.pptx
Merck Moving Beyond Passwords: FIDO Paris Seminar.pptxMerck Moving Beyond Passwords: FIDO Paris Seminar.pptx
Merck Moving Beyond Passwords: FIDO Paris Seminar.pptx
 
The State of Passkeys with FIDO Alliance.pptx
The State of Passkeys with FIDO Alliance.pptxThe State of Passkeys with FIDO Alliance.pptx
The State of Passkeys with FIDO Alliance.pptx
 
Advanced Computer Architecture – An Introduction
Advanced Computer Architecture – An IntroductionAdvanced Computer Architecture – An Introduction
Advanced Computer Architecture – An Introduction
 
Passkey Providers and Enabling Portability: FIDO Paris Seminar.pptx
Passkey Providers and Enabling Portability: FIDO Paris Seminar.pptxPasskey Providers and Enabling Portability: FIDO Paris Seminar.pptx
Passkey Providers and Enabling Portability: FIDO Paris Seminar.pptx
 
The Role of FIDO in a Cyber Secure Netherlands: FIDO Paris Seminar.pptx
The Role of FIDO in a Cyber Secure Netherlands: FIDO Paris Seminar.pptxThe Role of FIDO in a Cyber Secure Netherlands: FIDO Paris Seminar.pptx
The Role of FIDO in a Cyber Secure Netherlands: FIDO Paris Seminar.pptx
 
Use of FIDO in the Payments and Identity Landscape: FIDO Paris Seminar.pptx
Use of FIDO in the Payments and Identity Landscape: FIDO Paris Seminar.pptxUse of FIDO in the Payments and Identity Landscape: FIDO Paris Seminar.pptx
Use of FIDO in the Payments and Identity Landscape: FIDO Paris Seminar.pptx
 
Moving Beyond Passwords: FIDO Paris Seminar.pdf
Moving Beyond Passwords: FIDO Paris Seminar.pdfMoving Beyond Passwords: FIDO Paris Seminar.pdf
Moving Beyond Passwords: FIDO Paris Seminar.pdf
 

L8019 tm lodi_manual_revc

  • 1.
  • 2. THE LOCATOR OVERDENTURE IMPLANT SYSTEM. ® FOUR DECADES OF ATTACHMENT KNOWLEDGE INCORPORATED INTO NARROW DIAMETER OVERDENTURE IMPLANTS. The LOCATOR Overdenture Implant System (LODI) is comprised of 2.4mm and 2.9mm narrow diameter dental implants (available in 10, 12 and 14mm lengths) with a detachable LOCATOR Attachment that is available in a 2.5mm and 4mm cuff height. The LODI is used to restore masticatory function for the patient and may be suitable for immediate function if sufficient primary stability of the implant is achieved at the time of placement. IMPORTANT: This document contains the most current instructions for use. Please read and retain.
  • 3.
  • 4. TABLE OF CONTENTS LOCATOR® overdenture implant system 2 Indications 2 Contraindications 2 Caution 2 Storage and Handling 2 Single-Use Devices 2 Sterilization 3 Cleaning Instructions for Instruments (and Individually Packaged Replacement Attachments) 3 Surgical Tray Cleaning Instructions 3 Inspection and Maintenance of Cleaned Instruments 3 Steam Sterilization Instructions 4 Torque Indicating Ratchet Wrench Cleaning 5 Warnings and Precautions DRILLING DEPTH CONTROL 6 Drill Laser Depth Markings 6 Drill Stops 6 Final Drill Diameter and Depth for Various Bone Types 7 Drilling Sequence Examples 7 Placement of a 2.4mm x 12mm Implant Flapless Surgical Procedure 8 Placement of a 2.9mm x 12mm Implant Flapless Surgical Procedure LOCATOR MALES & EXTENDED RANGE MALES LOCATOR CORE TOOL PRE-SURGICAL TREATMENT PLANNING IMPLANT PLACEMENT LOCATOR ATTACHMENT PLACEMENT PROCESSING LOCATOR DENTURE CAPS & MALES INTO THE DENTURE 18 20 21 21 22 Direct Technique: Chairside Processing Indirect Technique: Laboratory Processing Lab Steps Bite Records Delivery EXPLANATION OF SYMBOLS ON OUTER PACKAGING LABELS 1
  • 5. LOCATOR OVERDENTURE IMPLANT SYSTEM ® INDICATIONS STERILIZATION The LOCATOR Overdenture Implant System is designed to retain overdentures or partial dentures in the mandible or maxilla. The LOCATOR Overdenture Implant is packaged with the LOCATOR Attachment and together are supplied STERILE (subjected to radiation (gamma) as a means of sterilization). CONTRAINDICATIONS Not appropriate where a totally rigid connection is required. Use of a single implant with divergence of greater than 20 degrees is not recommended. Dental implants should not be used in patients with serious medical problems or in a poor general state of health. Patients with medical problems such as; uncontrolled bleeding disorders, drug or alcohol abuse, weakened immune system, titanium allergy or uncontrollable endocrine disorders should be carefully evaluated prior to treatment. CAUTION All other restorative components, instruments, and replacement LOCATOR Attachments (sold separately) are supplied NON-STERILE. The nylon males may be sterilized using a liquid chemical sterilant following the manufacturer’s recommendations to achieve sterilization. Note: The Liquid Chemical Sterilant must be approved for Sterilization not just high-level disinfection and must be compatible with nylon material. Federal (U.S.A.) law restricts this device to sale by or on the order of a licensed dentist. Storage and Handling The LOCATOR Overdenture Implant System in its undamaged, original packaging is not subject to any special considerations for storage or handling (during transport and storage). Only sterile titanium or stainless instruments/tools should be used to handle and deliver the implant to the surgical site. SINGLE-USE DEVICES The LOCATOR Overdenture Implant System is a single-use device. LOCATOR Overdenture Implant: A previously used LOCATOR Overdenture Implant could contain patient contamination build-up. Therefore, the inadvertent re-use of this device could result in infection leading to lack of integration (of the implant to the bone). LOCATOR Males: The inadvertent re-use of LOCATOR nylon males could cause loss of retention of the overdenture due to wear from previous use or damage during removal with the LOCATOR Core Tool. LOCATOR Attachments: The inadvertent re-use of LOCATOR Attachments could contain patient contamination build-up and subsequent wear of the retention bands. This would result in the device to perform with improper fit and function which would result in loss of retention of the prosthesis. 2
  • 6. LOCATOR OVERDENTURE IMPLANT SYSTEM ® (Continued) CLEANING INSTRUCTIONS FOR INSTRUMENTS (AND INDIVIDUALLY PACKAGED REPLACEMENT Attachments) Disassemble any instruments that can be disassembled according to manufacturers’ instructions. Soak instruments in enzymatic cleaning solution (mixed according to manufacturers’ instructions) by completely submerging them for 20 minutes. Scrub instruments using a soft-bristled, nylon brush until all soil has been removed. Remove the instruments from the enzymatic cleaning solution and rinse in tap water for a minimum of 3 minutes. Make sure to thoroughly flush internal holes/crevices of instruments (such as the tissue punch, drill extender, implant drivers, and disassembled core tool and ratchet torque wrench) that have difficult to reach areas. Note: Use of a syringe or water jet will improve flushing of difficult to reach areas. Place instruments in sonication bath (with enzymatic cleaning solution prepared according to manufacturers’ instructions) making sure that they are completely submerged, and sonicate for 10 minutes. Remove the instruments from the sonication bath, and rinse for 3 minutes making sure to thoroughly flush cleaning solution out of the holes/crevices and/ or difficult to reach areas. Remove excess moisture from the instruments with a clean, absorbent, and non-shedding wipe. Surgical Tray cleaning instructions Rinse the tray and tray insert with tap water. Place the Surgical Tray and Insert in enzymatic cleaning solution (mixed according to manufacturers’ instructions) and wipe off soil with a clean, absorbent, non-shedding wipe. Allow the Surgical Tray and Insert to soak in the cleaning solution for 20 minutes making sure that they are completely submerged. 3 Surgical Tray cleaning instructions (CONTINUED) Remove the Surgical Tray and Insert from the enzymatic cleaning solution and rinse in tap water for a minimum of 3 minutes. Make sure to thoroughly flush each piece to completely remove cleaning solution residue. Remove excess moisture from the Surgical Tray and Insert with a clean, absorbent, and non-shedding wipe. INSPECTION AND MAINTENANCE OF CLEANED INSTRUMENTS Carefully inspect each instrument to ensure that all visible contamination has been removed. If contamination is noted, repeat the cleaning process. Please note that if during inspection of instruments, you see signs of wear, damage, or unrecognizable color change, replace the instrument. Re-assemble multi-part instruments and check them for proper function (LOCATOR Core Tool and the Torque Indicating Ratchet Wrench) Reference the IFU that comes with each of these parts and subsequent sections of this document for the proper assembly process. Steam Sterilization Instructions The validation procedures require the use of FDAcleared sterilizers, sterilization trays, sterilization wraps, biological indicators, chemical indicators and other sterilization accessories labeled for the sterilization cycle recommended. The health care facility should monitor the sterilizer for the facility according to an FDA recognized sterility assurance standard such as ANSI/AAMI ST79:2006. Place all instruments into the surgical tray. For gravity cycle, place Surgical Kit in a 10” x 15” Autoclave Bag, and for Pre-Vacuum Cycle double wrap the kit with autoclave wrap material and secure wrap with autoclave tape. Autoclave Sterilization Parameters are listed below: CYCLE TYPE GRAVITY PRE-VACUUM TEMPERATURE 121ºC / 250ºF 132ºC / 270ºF EXPOSURE TIME 50 MINUTES 4 MINUTES DRYING TIME 15 MINUTES 20 MINUTES
  • 7. LOCATOR OVERDENTURE IMPLANT SYSTEM ® (Continued) TORQUE INDICATING RATCHET WRENCH CLEANING Intended Use: A dental torque wrench for placement and adjustment of dental implants, attachments, attachment screws and prosthetic screws during oral surgery and prosthetic procedures. Scale Unit: Ncm. Sterilzation: Autoclave/steam gravity sterilize for 50 minutes at 121ºc, dry for 15 minutes. For pre-vacuum cycle, autoclave/steam sterilize for 4 minutes at 132ºC with drying time of 20 minutes. WARNING: Device must be autoclaved prior to use. This device must not be cleaned using hydrogen peroxide. Note: Drying times may vary according to load. Cleaning: Press the driver to remove it from the head of the wrench, and remove the head by pressing a finger into the recess and gently pulling the head. The three separated parts are now ready for cleaning using the following procedure: Soak torque wrench parts in enzymatic cleaning solution (mixed according to manufacturer’s instructions) by completely submerging it for 20 minutes. Scrub torque wrench parts using a soft bristled, nylon brush until all soil has been removed. Remove the torque wrench parts from the enzymatic cleaning solution and rinse in tap water for a minimum of 3 minutes. Place torque wrench parts in sonication bath (with the enzymatic cleaning solution prepared according to manufacturer’s instructions) making sure they are completely submered, and sonicate for 10 minutes. Remove the torque wrench parts from the sonication bath, and rinse in water for 3 minutes making sure to thoroughly flush cleaning solution out of the holes/ crevices and/or difficult to reach areas. Remove excess moisture from the torque wrench parts with a clean, absorbent, and non-shedding wipe. After sterilization, attach the head of the wrench to the body by pushing the components together and turning them in opposite directions until there is an audible click. Push the driver into the wrench until there is an audible click. The arrow on the head of the wrench shows the direction in which the wrench is functioning. Turn the wrench in the direction of the arrow until the desired torque is achieved. WARNING: Before each use, make sure that the functionalities are intact and that the first line on the scale aligns with the arrow. The arm of the torque wrench must not go beyond the end of the scale, as this could result in inaccurate readings. If the torque wrench is used as an ordinary wrench, without using the torque scale, then it may not be sujected to a load of more than 80Ncm. WARNING: After overloading or if dropped or in other ways mishandled, the wrench must no longer be used since correct function can no longer be guaranteed. Please refer to the LOCATOR Overdenture Implant System Technique Manual available from the manufacturer or your distributor for detailed surgical procedure instructions. It is also available online at www.zestanchors.com 4
  • 8. LOCATOR OVERDENTURE IMPLANT SYSTEM ® (Continued) WARNINGS AND PRECAUTIONS The LOCATOR Overdenture Implant System has not been evaluated for safety and compatibility in the MR environment. The LOCATOR Overdenture Implant System has not been tested for heating or migration in the MR environment. Product (implant/attachment) from damaged sterilized packaging must not be used on patients. In the event that the sterilized packaging for the LOCATOR Overdenture Implant System is damaged, the damaged packaging (with the product) must be returned to the manufacturer and a replacement will be provided (if damage to sterilized packaging is caused by product shipment). The drill extender is to be used with surgical drills only and should not be used in high torque applications. Avoid application of excessive bending load on smaller diameter drills during drilling. Drills will dull based on many factors including bone density, handling, autoclave exposure, etc. Replace drills when wear is noticeable to avoid excessive heat being transferred to surrounding bone during osteotomy preparation. If the LOCATOR Overdenture Implant System is subjected to inappropriate loading conditions, there may be a potential risk of metal fatigue or localized bone failure. The use of other tissue grafting components or parts that are made from dissimilar metals should not be used in or near the implant. Patient evaluation including the determination of the general health, oral hygiene habits and status, motivation toward good dental care, and anatomic acceptability prior to implant surgery is critical. Thorough evaluation of the patient’s medical status and health history is mandatory. Panaromic and periapical radiographs as well as thorough oral inspection and palpation are recommended to determine anatomic landmarks, dental pathology, and adequacy of bone. A cephalogram is suggested for totally edentulous patients. Any oral condition that adversely affects natural teeth, if uncorrected, will have an adverse effect on the implants. 5 Periodontal disease, abnormal bone conditions, severe bruxism, cross-bite situations, and extenuating circumstances (e.g. excessive smoking, medical issues, etc) that may adversely affect the procedure must be evaluated and corrected if necessary, or use of the implant may be contraindicated. Based on the results of the patient’s pre-surgical assessment, the clinician should select and order the appropriate implant (determine correct implant diameter and length based on bone type), restorative parts, and tools. Refer to Drilling Sequence section for further details. The clinician should also determine if the patient is allergic to any of the materials that will be used in the procedure as part of the pre-surgical treatment planning. If during patient evaluation, insufficient bone width, abnormal bone defects or contours are detected, then the placement of the implant may be contraindicated. Patient motivation is a key factor in achieving success with any implant. The patient must be willing to practice the oral hygiene necessary for implant maintenance. The clinician must provide the patient with information regarding proper care and maintenance of the implants. Also, they must inform the patient that conditions such as excessive smoking, improper/lack of maintenance may have adverse effects. The use of this or any surgical implant product requires that the clinician be thoroughly familiar with the product and the method for its use and application. They must also be familiar with all the instruments, and surgical procedures required (as described in this document). The clinician must also use reasonable judgment in deciding when and where to use the product.
  • 9. DRILLING DEPTH CONTROL DRILL LASER DEPTH MARKINGS DRILL STOPS 6mm Depth 8mm Depth 10mm Depth 12mm Depth 14mm Depth FINAL DRILL DIAMETER AND DEPTH FOR VARIOUS BONE TYPES BONE TYPE D1 2.4MM implant diameter Final Drill DRILL DEPTH Diameter 2.1mm Full D2 / D3 D4 2.9MM implant diameter Final Drill DRILL DEPTH Diameter 2.4mm Full 1.6mm Depth 4mm less than implant length 2.1mm Depth 4mm less than implant length 1.6mm Depth 4mm less than implant length 2.1mm Depth 4mm less than implant length Bone type is a general classification. The overall bone quality must be assessed by the clinician through treatment planning and at the time of surgery in order to create the appropriate osteotomy size to achieve the desired insertion torque. 6
  • 10. DRILLING SEQUENCE EXAMPLES PLACEMENT OF A 2.4MM x 12MM IMPLANT FLAPLESS SURGICAL PROCEDURE 2.4MM LASER DEPTH MARKINGS 2.4MM DRILL STOPS 2mm Soft Tissue 1.2mm Pilot Full Depth 14mm Line 1.6mm Full Depth 14mm Line 2.1mm Full Depth 10mm Line Tap Speed Until Stop Ratchet to Full Depth 1.2mm Pilot Full Depth 12mm Drill Stop 1.6mm Full Depth 12mm Drill Stop 2.1mm Full Depth 8mm Drill Stop Tap Speed Until Stop Ratchet to Full Depth 2mm Soft Tissue 1.2mm Pilot 1.6mm Full Depth 4mm Short 14mm Line 10mm Line Tap Speed Until Stop Ratchet to Full Depth 1.2mm Pilot 1.6mm Full Depth 4mm Short 12mm 8mm Drill Stop Drill Stop Tap Speed Until Stop Ratchet to Full Depth 1.2mm Pilot 1.6mm Full Depth 4mm Short 12mm 8mm Drill Stop Drill Stop Tap Speed Until Stop Ratchet to Full Depth 2mm Soft Tissue 7 1.2mm Pilot 1.6mm Full Depth 4mm Short 14mm Line 10mm Line Tap Speed Until Stop Ratchet to Full Depth
  • 11. DRILLING SEQUENCE EXAMPLES (Continued) PLACEMENT OF A 2.9MM x 12MM IMPLANT FLAPLESS SURGICAL PROCEDURE 2.9MM LASER DEPTH MARKINGS 2.9MM DRILL STOPS 2mm Soft Tissue 1.2mm Pilot Full Depth 14mm Line 1.6mm Full Depth 14mm Line 2.4mm Full Depth 14mm Line Tap Speed Until Stop Ratchet to Full Depth 1.2mm Pilot Full Depth 12mm Drill Stop 1.6mm Full Depth 12mm Drill Stop 2.4mm Full Depth 12mm Drill Stop Tap Speed Until Stop Ratchet to Full Depth 1.2mm Pilot Full Depth 12mm Drill Stop 1.6mm Full Depth 12mm Drill Stop 2.1mm 4mm Short 8mm Drill Stop Tap Speed Until Stop Ratchet to Full Depth 1.2mm Pilot Full Depth 12mm Drill Stop 1.6mm Full Depth 12mm Drill Stop 2.1mm 4mm Short 8mm Drill Stop Tap Speed Until Stop Ratchet to Full Depth 2mm Soft Tissue 1.2mm Pilot Full Depth 14mm Line 1.6mm Full Depth 14mm Line 2.1mm 4mm Short 10mm Line Tap Speed Until Stop Ratchet to Full Depth 2mm Soft Tissue 1.2mm Pilot 1.6mm Full Depth Full Depth 14mm Line 14mm Line 2.1mm 4mm Short 10mm Line Tap Speed Until Stop Ratchet to Full Depth 8
  • 12. LOCATOR MALES & EXTENDED RANGE MALES ® LOCATOR MALES EXTENDED RANGE MALES The unique Dual Retention innovation provides the LOCATOR Attachment with a greater retention surface area than ever before available with other attachments. Allows you to restore a non-parallel implant with up to 20 degrees of angulation. This calculates to an extensive 40 degrees of divergence between two implants. 1.5 lbs 3 lbs 5 lbs Dual retention, pivoting action provides resiliency to maximize stability and longevity. 0 lbs 1 lbs 2 lbs 4 lbs Extended Range Male’s pivoting action allows for insertion with up to 40° total divergence. a lodi male processing pack is INCLUDED WITH EACH IMPLANT Each processing pack has what you need to select retention levels and address draw correction; improving ease of denture placement and removal. Denture Cap LODI Male Processing Pack 9 Blue 1.5 lbs Pink 3 lbs Spacer Ring Red 1 lbs
  • 13. LOCATOR CORE TOOL ® LOCATOR CORE TOOL The LOCATOR Implant Attachment System features a Core Tool that contains 3-tools-in-1. This convenient tool is used to carry and place the LOCATOR Attachment onto the implant, remove the LOCATOR Male, and insert the male into the LOCATOR Denture Cap. Insert drivers for various types of torque wrenches are available to achieve 30Ncm of torque. REMOVAL The Male Removal Tool has a sharp edge on the end to catch and remove the male from the Denture Cap. INSERTION The Male Seating Tool is used to seat the LOCATOR Male. PLACEMENT The LOCATOR Abutment Driver with the Abutment Holder Sleeve carries the attachment securely and places it onto the implant. Loosen the Male Removal Tool a full 3 turns counter clockwise (you will see a visible gap). To remove a LOCATOR nylon male from the Denture Cap; simply insert the tip into the cap/male assembly and push straight in to the bottom of the nylon male. Then tilt the tool so that the sharp edge of the tip will grab hold of the male and pull it out of the Denture Cap. To discard the nylon male from the tip on the Core Tool; point the tool down and away from you and tighten the Male Removal Tool clockwise back onto the Core Tool. This will activate the removal pin and dislodge the nylon male from the tip end of the Male Removal Tool. Separate the Male Removal Tool section from the LOCATOR Core Tool and use the Male Seating Tool end of the remaining two sections to place a new nylon male into the empty Denture Cap. 10
  • 14. PRE-SURGICAL TREATMENT PLANNING Evaluate available bone width for implant positions by using the index finger/thumb technique or a ridge mapping instrument. Measure the gingiva depth at each implant location using a perio probe to determine the proper depth measurement line on the osteotomy preparation drills and to select the LOCATOR® Attachment cuff height of either 2.5mm or 4mm. 3A-3B Radiograph the arch using a panoramic radiograph or CBCT to determine available bone height for implant positions. A radiographic template with measurement balls may be fabricated to assist with determining dimensions. A B 11
  • 15. PRE-SURGICAL TREATMENT PLANNING (Continued) Determine if the patient’s existing denture(s) will be used or if new ones will be fabricated. If a new denture is made, follow the conventional process. Have the patient wear the new denture for 2 weeks prior to implant placement. A surgical guide for implant placement may be fabricated from the patient’s existing or new denture prior to surgery. 12
  • 16. IMPLANT PLACEMENT After patient selection and evaluation protocols have been completed, the number of implants required is determined and discussed with the patient. The patient’s denture is then fabricated or modified, followed by identification of appropriate implant sites. For mandibular placement, the position of the nerve and bone quality must be taken into consideration. For maxillary placement, bone volume, bone quality and sinus location must be taken into consideration. EXAMPLE IS PLACEMENT OF FOUR 2.9MM X 10MM IMPLANTS IN TYPE D1 Bone Using the surgical guide or by free hand, mark the implant osteotomy locations by drilling through the gingiva and into the bone crest approximately 6mm using the 1.2mm Pilot Drill. Note the gingival depth. The recommended drill speed is 1200-1500 rpm. Remove the gingiva cores at each site using the Rotary Tissue Punch. Place the guide pin portion into the pilot holes and rotate to cut away the gingiva. Rotate the Tissue Punch to the laser depth mark indicated from the gingiva depth measurement. The recommended speed is up to a maximum of 800 rpm. 13
  • 17. IMPLANT PLACEMENT (Continued) EXAMPLE IS PLACEMENT OF FOUR 2.9MM X 10MM IMPLANTS IN TYPE D1 Bone • CONTINUED 3A-3B A Place the 1.2mm diameter (small) end of the Direction Indicator into the pilot drill osteotomies to verify the proper angulation. Place the proper length drill stop onto the 1.2mm Pilot Drill according to the desired drill depth. Or, drill to the proper laser depth marking on the drill calculated by adding the desired drill depth and tissue depth. The recommended drill speed is 1200-1500 rpm. Continue osteotomy preparation to the desired depth at each implant site. B 4A-4B Place the proper length drill stop onto the 1.6mm drill according to the desired drill depth. Or, drill to the proper laser depth marking on the drill calculated by adding the desired drill depth and tissue depth. The recommended drill speed is 1200-1500 rpm. Continue osteotomy preparation to the desired depth at each implant site. A B 14
  • 18. IMPLANT PLACEMENT (Continued) EXAMPLE IS PLACEMENT OF FOUR 2.9MM X 10MM IMPLANTS IN TYPE D1 Bone • CONTINUED 5A-5B Place the 1.6mm diameter end of the Directional Indicator into the osteotomies to verify the proper angulation. Place the proper length drill stop onto the 2.4mm drill according to the desired drill depth. Or, drill to the proper laser depth marking on the drill calculated by adding desired drill depth and tissue depth. The recommended drill speed is 1200-1500 rpm. Continue osteotomy preparation to the desired depth at each implant site. A B 6A-6B Remove the implant package from the box and peel back the tyvek seal from the plastic tray. Drop the implant vial on the sterile tray. The contents of the plastic tray are sterile and should only contact components within the sterile field. A B 7A-7B Remove the cap from the implant vial and do not discard. The LOCATOR® Attachment is in the cap. Set the drilling unit speed at 50rpm and the placement torque at 35Ncm. Place the Implant Driver in the handpiece. Insert the Implant Driver onto the hex on the top of the implant and press down to frictionally engage. The bottom of the driver should contact the attachment seating surface and fully engage the entire length of the implant hex. A 15 B
  • 19. IMPLANT PLACEMENT (Continued) EXAMPLE IS PLACEMENT OF FOUR 2.9MM X 10MM IMPLANTS IN TYPE D1 Bone • CONTINUED Remove the implant from the vial in a straight out motion. Carry the implant to the mouth, place it into the osteotomy and tap into place at 50 rpm. Avoid sudden movement and/or touching an external object which may dislodge the implant from the driver. Warning: Discard and do not use an implant that has been dropped in a non-sterile area and replace with a new sterile implant. 10A-10B When the drilling unit stops tapping the implant, place the Torque Indicating Ratchet Wrench Insert into the Torque Ratchet Wrench. Place the Implant Driver (short or long) into the insert. 11 Position the Implant Driver onto the hex on the top of the implant and verify that it is fully engaged. Slowly ratchet the implant to full depth. If final seating torque measures 30Ncm or above, the implant may be immediately loaded at the discretion of the clinician. If the final seating torque measures below 30Ncm, the denture acrylic should be relieved and a soft liner placed around the LOCATOR® Attachments during the integration period. Implant insertion torque should not exceed 70Ncm. If 70Ncm of torque is reached, prior to full seating, the implant should be removed and the osteotomy should be enlarged. 16
  • 20. LOCATOR AttachmenT PLACEMENT ® 1A-1B Open the flip cap on the top of the vial cap and remove the LOCATOR Attachment. Place the Abutment Holder Sleeve onto the LOCATOR Abutment Driver. Place the attachment into the Abutment Holder Sleeve to securely carry it to the mouth. A B 2A-2B Thread the LOCATOR Attachment onto the implant until finger tight. If the implant placement torque was 30Ncm or greater, the attachments may be tightened to the recommended torque level of 30Ncm. If the implant placement torque did not reach 30Ncm, the attachments should only be hand tightened. Place the LOCATOR Attachment Torque Driver Insert into the Torque Ratchet Wrench. Insert the driver into the attachment and verify it is fully engaged. Torque the attachments to 30Ncm. A B If the implant placement torque was 30Ncm or greater, the implants may at the discretion of the clinician be immediately loaded. Continue on with steps for processing the LOCATOR Denture Caps and Males into the denture. If the implant placement torque was less than 30Ncm, relieve the denture acrylic and place a soft liner in the denture around the LOCATOR Attachments during the integration period. 17
  • 21. PROCESSING LOCATOR DENTURE CAPS AND MALES INTO THE DENTURE ® DIRECT TECHNIQUE: CHAIRSIDE PROCESSING (NEW OR EXISTING DENTURE) If the implants were not immediately loaded, the LOCATOR Attachments must be torque tightened. Place the LOCATOR Abutment Torque Driver Insert into the Torque Ratchet Wrench. Insert the driver into the attachment and verify that it is fully engaged. Torque the attachments to 30Ncm. Place a White Block Out Spacer Ring around each attachment and press it down to the tissue. Place a Denture Cap with a Black Processing Male inside of it onto each attachment and press down firmly. Apply fit check marking paste inside of the denture. Insert it into the mouth in position over the Denture Caps to mark the areas where the denture will need to be relieved to allow space for the caps to be picked up. Relieve the areas marked with an acrylic bur. Try in the denture to verify that the Denture Caps are not in contact with the acrylic in any area. Cut lingual/palatal vent windows in the denture to visualize full seating and for an excess acrylic vent. 18
  • 22. PROCESSING LOCATOR DENTURE CAPS AND MALES INTO THE DENTURE (CONTINUED) ® DIRECT TECHNIQUE: CHAIRSIDE PROCESSING (CONTINUED) An autopolymerizing or light cure acrylic resin may be used to pick up the caps. Dry the Denture Caps. Apply a small amount of acrylic around the circumference of each cap. Place acrylic into the relief areas of the denture and seat it over the caps and onto the tissue. Have the patient close into occlusion and hold while the acrylic sets. Disengage the denture from the LOCATOR Attachments and remove from the mouth. Verify that the Denture Caps have been securely picked up in the denture. Fill any voids and polish the denture. 7A-7B See the LOCATOR Core Tool instructions on page 10. Remove the Black Processing Male using the Male Removal Tool. Place the selected final male into each Denture Cap using the Male Insertion Tool. It is recommended to place the least retentive male to begin with. See the male retention chart on page 9. A B Place the denture in the mouth and press down to engage the males on the LOCATOR Attachments. Verify the occlusion. Instruct the patient on how to remove and insert the denture. If the retention is not satisfactory, remove the males and replace with the next level of retention. See the male retention chart on page 9. Instruct the patient on proper home care maintenance and required recall visits. 19
  • 23. PROCESSING LOCATOR DENTURE CAPS AND MALES INTO THE DENTURE (CONTINUED) ® INDIRECT TECHNIQUE: LABORATORY PROCESSING (Appt. 1) If the implants were not immediately loaded, the LOCATOR® Attachments must be torque tightened. Place the LOCATOR Abutment Torque Driver Insert into the Torque Ratchet Wrench. Insert the driver into the attachment and verify that it is fully engaged. Torque the attachments to 30Ncm. A stock or custom impression tray may be used. Ensure on each that there is enough space for the 4mm height of the LOCATOR Impression Copings. 4mm Place a LOCATOR Impression Coping on each attachment and press down firmly. Syringe a medium body impression material around the circumference of each coping. Fill the impression tray and insert it over the copings and onto the tissue. Allow the material to set. Remove the impression and verify that the copings have been securely picked up inside of it. Press the LOCATOR Analogs into each impression coping. Send the impression to the laboratory. 20
  • 24. PROCESSING LOCATOR DENTURE CAPS AND MALES INTO THE DENTURE (CONTINUED) ® INDIRECT TECHNIQUE: LABORATORY PROCESSING (CONTINUED) LAB STEP 1 Verify that the analogs are secure in the impression copings. Pour die stone into the impression to fabricate the master cast. Fabricate the baseplate and wax rim on the cast for the bite registration. The Denture Caps with Black Processing Males may be processed into the baseplate to provide stabilization during record making. Bite Records (Appt. 2) Place the wax rim into the mouth and make the bite records. Take an impression of the opposing arch and pour the cast. Select a shade for the denture teeth. Lab Step 2 Articulate the master cast with the opposing using the bite records. Set the denture teeth on the baseplate and fabricate a wax denture for try in. 21
  • 25. PROCESSING LOCATOR DENTURE CAPS AND MALES INTO THE DENTURE (CONTINUED) ® Denture Try In (Appt. 3) Place the wax denture into the mouth and verify the esthetics, phonetics and occlusion. Lab Step 3 Wax and flask the denture for processing. Separate the flask and boil away all wax. Place the Denture Caps with Black Processing Males on the analogs and press down firmly. Place the cast back into the flask and verify that there is no contact with the teeth. Close the flask and process the denture. Remove the denture from the flask, finish and polish. 2A-2B See the LOCATOR Core Tool instructions on page 10. Remove the Black Processing Male using the Male Removal Tool.Place the selected final male into each Denture Cap using the Male Insertion Tool. It is recommended to place the least retentive male to begin with. See the male retention chart on page 9. A B DELIVERY (Appt. 4) Place the denture in the mouth and press down to engage the males on the LOCATOR Attachments. Verify the occlusion. Instruct the patient on how to remove and insert the denture. If the retention is not satisfactory, remove the males and replace with the next level of retention. See the male retention chart on page 9. Instruct the patient on proper home care maintenance and required recall visits. 22
  • 26. EXPLANATION OF SYMBOLS ON OUTER PACKAGING LABELS SYMBOL STANDARD ISO 980 EXPLANATION OF SYMBOL Symbol for “DO NOT REUSE” LOCATION OF SYMBOL ON THE FOLLOWING LABELS: For single-use. Use only once. Symbol for “BATCH CODE” ISO 980 1) L9127-XXXX 2) L7409-XXXX 3) L9108-XXXX ON THE FOLLOWING LABELS: This symbol shall be accompanied by the manufacturer’s batch code or lot code. The batch/lot code shall be adjacent to the symbol. 4) L9127-XXXX 5) L7409-XXXX 6) L9108-XXXX 7) L9115-07420 8) L9912-07362 9) L9109-XXXX Symbol for “MANUFACTURER” ISO 980 ISO 980 ON THE FOLLOWING LABELS: This symbol shall be accompanied by the manufacturer name (Zest Anchors) and address (2061 Wineridge Place, Escondido, CA 92029); adjacent to the symbol. 10) L9127-XXXX 11) L7409-XXXX 12) L9108-XXXX 13) L9115-07420 14) L9912-07362 15) L9109-XXXX Symbol for “STERILIZED USING IRRADIATION” ON THE FOLLOWING LABELS: NOTE: Refers to Implant/Attachment sterilized packaging only. Symbol for “CATALOGUE NUMBER” ISO 980 23 16) L9127-XXXX 17) L7409-XXXX ON THE FOLLOWING LABELS: The product catalogue number shall be after or below the symbol adjacent to it. 18) L9127-XXXX 19) L7409-XXXX 20) L9108-XXXX
  • 27. NOTES ________________________________________________________________________________________________________________________________________________________________________________________ ________________________________________________________________________________________________________________________________________________________________________________________ ________________________________________________________________________________________________________________________________________________________________________________________ ________________________________________________________________________________________________________________________________________________________________________________________ ________________________________________________________________________________________________________________________________________________________________________________________ ________________________________________________________________________________________________________________________________________________________________________________________ ________________________________________________________________________________________________________________________________________________________________________________________ ________________________________________________________________________________________________________________________________________________________________________________________ ________________________________________________________________________________________________________________________________________________________________________________________ ________________________________________________________________________________________________________________________________________________________________________________________ ________________________________________________________________________________________________________________________________________________________________________________________ ________________________________________________________________________________________________________________________________________________________________________________________ ________________________________________________________________________________________________________________________________________________________________________________________ ________________________________________________________________________________________________________________________________________________________________________________________ ________________________________________________________________________________________________________________________________________________________________________________________ ________________________________________________________________________________________________________________________________________________________________________________________ ________________________________________________________________________________________________________________________________________________________________________________________ ________________________________________________________________________________________________________________________________________________________________________________________ ________________________________________________________________________________________________________________________________________________________________________________________ ________________________________________________________________________________________________________________________________________________________________________________________ ________________________________________________________________________________________________________________________________________________________________________________________ ________________________________________________________________________________________________________________________________________________________________________________________ ________________________________________________________________________________________________________________________________________________________________________________________ ________________________________________________________________________________________________________________________________________________________________________________________ ________________________________________________________________________________________________________________________________________________________________________________________ ________________________________________________________________________________________________________________________________________________________________________________________ ________________________________________________________________________________________________________________________________________________________________________________________ ________________________________________________________________________________________________________________________________________________________________________________________ ________________________________________________________________________________________________________________________________________________________________________________________ ________________________________________________________________________________________________________________________________________________________________________________________ ________________________________________________________________________________________________________________________________________________________________________________________ 24
  • 28. ZEST ANCHORS 2061 Wineridge Place Escondido, CA 92029 USA Tel: 1.855.868.LODI (5634) Fax: 760.743.7975 Email: sales@zestanchors.com www.zestanchors.com Distributed by: ©2012 ZEST Anchors LLC. All rights reserved. ZEST and LOCATOR are registered trademarks of ZEST IP Holdings, LLC. L8019-TM REV C 2/2013