HIGH SPEED CUTTING INSTRUMENTS IN
PROSTHODONTICS
Introduction
In order to perform the intricate and detailed procedures associated with
restorative dentistry, the dentist must have a complete knowledge of the
purpose and application of the many instruments required. During each day of
his clinical experience the dentist operates on vital tissues within the oral
cavity where a millimeter or a fraction there of, is a very significant
dimension. A skillful application of sharp hand and rotary instruments requires
ability and coordination gained only by extensive training.
Before the advent of rotary instruments, removal of tooth tissue was
accomplished with sharp – edged chisels, hatchets, and hoes. These hand
instruments possessed a cutting capability, which was used for clearing away
unsupported and undermined enamel resulting from dental caries. Walls and
floors of the cavity were formed by a planning and lateral scraping action of
these sharp edged instruments. At best, such efforts were crude, time
consuming and often difficult.
The first, rotary instruments for cutting tooth tissue were modified hand
instruments. These, drill or bur heads could be twisted in the fingers to
produce a cutting or abrading action. In 1846 the finger ring was introduced
with a drill socket attached for adapting a series of long bundled burs or drills.
1
This was the primitive application of the rotary principle. The first drill having
flexible cable drive and the first angle hand piece were introduced by Charles
Merry between 1858 and 1862. In 1871, Morrison modified and adapted the
dental foot engine from the Singer Sewing machine. This was followed by the
introduction of the electric dental engine utilizing a cable arm in 1883. In 1910
the endless cord on a jointed arm was made available. The earlier dental hand
pieces were capable of speeds from 4500 to 6000 rpm.
In 1940 the use of diamond abrasive paints became widespread. The
diamond point is compared of a number of small diamond particles bound on a
rotary blank.
In 1945 Dr. G.V. black, published a report on the non mechanical
preparation of cavities and in doing so introduced the air abrasive technique.
The impact of Dr. Black’s revolutionary cutting technique on the dental
profession was considerable. This was the first significant break in the long
established traditional method of cavity preparation. The air abrasive principle
utilized particles of aluminium oxide propelled against the tooth surface by a
carbon dioxide stream under the pressure of 110 psi, and funneled through a
tungsten carbide nozzle with a lumen of 0.018 inch. The penetration of enamel
and dentin was rapid but some what difficult to control.
2
In 1949 Walsh and Symons published their initial findings relating to
the removal of tooth tissue with diamond points at rotational speeds upto
70,000 rpm. This report indicated the use of lighter forces and a resulting
increased cutting efficiency at these higher speeds.
In early 1950, the ball-bearing hand piece was introduced.
In 1963, following the work of Nelson the first fluid turbine type
handpiece was introduced. This instrument was capable of rotational speeds of
approximately 50,000 rpm and was limited to diamond instruments operated at
one speed only. In 1954, air-driven hand pieces were developed. A continuous
belt-driven contra-angle which utilized a friction grip chuck and bur was
introduced, making possible cutting speeds of upto 150,000 rpm.
By 1957, many dentists were using rotational speeds upto 3,00,000
rpm. The introduction of air-bearing hand piece in the early 1950 made
possible greater rotational speeds of approximately 5,00,000 rpm.
In 1953, an ultrasonic method of tooth tissue removal was also
introduced, which used suitably shaped tips vibrating at frequencies ranging
from 2,50,000 to 3,00,000 cycles – per seconds.
This brief historical back ground reveals that the profession has been
searching for a suitable method of tooth tissue removal. Only during last 30
3
years, this hunt has slowed down still the profession is trying to refine the
procedure and instruments.
Review of literature
A search through literature reveals various methods used in the past for
removal of tooth tissue. The continuous development of newer methods till
1960, indicates that the earlier instruments had some disadvantages. Inspite of
the introduction of numerous tooth reduction instruments, and procedures, the
principles and the biologic objectives have remained the same. These are as
follows.
1. The operator should remove the least amount of tooth tissue consistent
with necessary mechanical retention.
2. This should be done with the least barm to the periodontal tissues and the
pulp.
3. It should be done with the least discomfort to the patient.
4. No pathologic reactions should be initiated in the pulp.
Advantages of high speeds
1. Smaller stones can be used at the increased speeds.
2. Less fatigue results both for the patient and operator.
3. Due to high speed, very light pressure is required.
4. Less vibrations are felt by the patient.
4
5. The chairside time for a given preparation is considerably reduced.
6. Trauma to the pulp is reduced.
7. The efficiency and life of the cutting tools is increased.
8. Because of small tools, control is easy.
9. Removal of old amalgam and gold restorations is easy.
Disadvantages of high speeds
1. The increased speed creates increased temperatures in the tooth. Therefore
some method of cooling the tooth more efficiently is required not to injure
the pulp. This necessitates additional equipment.
2. When a dentist changes from the lower speeds, which utilize a pressure in
pounds, to high speeds which need only a pressure in ounces, he must
develop a new technique and retrain himself to a new tactile sense.
3. To operate at high speeds good visibility of the cutting instrument is
necessary to avoid over cutting.
4. Due to the ease with which tooth tissue is removed, caution must be taken
not to injure the proximal enamel of the adjacent healthy tooth and the
gingiva.
5. High speeds result in greater wear on the working parts of the handpiece,
necessitating more frequent repairs and replacements.
6. Unless used properly, high speeds have a tendency to create striations on a
tooth surface.
5
7. The ideal preparation for any type of restoration cannot be accomplished
by using high speed equipment alone. The final exactness and finishing
line can best be established by instruments revolving at moderate speeds.
Types of high speed instruments
Hand piece can be divided into four types depending upon their speeds
as follows.
1. Low speed – upto 10,000 rpm.
2. Intermediate speed – 25,000 to 45,000 rpm.
3. High speeds – 50,000 to 1,00,000 rpm.
4. Ultra high speeds 1,00,000 rpm and over.
Kilpatric has further classified the ultra high speed handpiece into three
classes.
Type I – the gear driven centre-angle handpiece, upto1,25,000 rpm.
Type II – the belt driven contra-angle handpiece upto 2,00,000 rpm.
Type III – turbine driven air contra-angle handpiece upto 3,00,000 rpm and
higher.
Heat generation:
Knowledge of the physics tells us that, whenever there is friction
between two surfaces, heat is generated, which may bring about rise in
6
temperature of either or both the surfaces. The same applies in the tooth
reduction procedures. Here the rotating cutting tools come in contact with the
tooth surface and the heat is generated.
It was not until 1930 that the workers began to investigate the heat rise
in the dental pulp.
There are many factors that influence the rise in temperature which
takes place in cutting operations. The greater the speed of rotation of the
cutting tool, the faster the tool revolves, the higher the resultant temperature. It
has been found that the temperature rise develops within 10-12 seconds, after
the cutting operation is started. Size of the cutting instrument has an important
bearing on heat generation, since, its diameter affects the cutting speed at its
periphery. Larger the size of the cutting tool more the host generation.
A third factor is the pressure applied by the dentist during cutting
operation. As the pressure increases, greater will be the rise in temperature.
Hudson and associates in 1954 conducted a study on temperature
developed in dental cutting instruments from their study they have concluded
that,
1. The temperatures produced by dental burs in cutting human dentin ranged
from 125°F to 275°F. Since these temperatures are above those, said to be
7
tolerated by normal human dentin, it would seem advisable to use some
form of coolant.
2. A significant decrease in time required to accomplish a given operation is
apparent, when high operating speeds are used.
3. The amounts of heat transferred to the tooth from the bur decreases, at
speeds above 12000 rpm, since cutting time at these speeds is reduced and
bur temperature remains.
Substantially constant and there is less heat trauma to the vital structures.
Coolants:
From the study of Hudson and Sweeney, it is evident that the
temperatures reached during tooth reduction procedures are above those said
to be tolerated by normal human dentins. This indicates that, some form of
coolant must be used, during the cutting operations, particularly when high
speeds are used.
Every means should be employed to keep the temperature down as
much as possible during cutting operations. Coolants must employed which, to
be effective, should be applied at the point of contact between the cutting
instrument and the tooth tissue. There are three types of coolants usually
employed in dental practice.
8
1. Water.
2. Spray of air and water
3. Air alone.
Peyton has shown that at speeds ranging from 30000 rpm to 170000
rpm and with an application of four ounces of pressure, a temperature rise
within the tooth of less than 15°C occurred when water or air-water sprays
were employed. He also found that even with a water coolant, excessive
temperatures developed, when large diameter instruments or excessive
pressure were applied with increased operating speeds. This indicates that the
use of a coolant, does not eliminate the danger of excessive temperature rise.
A reduction in concentration of the amount of water used during cutting
procedure shows the significant temperature rise of the dental bur.
The minimum volume of water to be applied was estimated at 1.5 ml
per minute.
The question whether water in spray form should be used at mouth or
temperature seems to have no significance as far as temperature rise in the
tooth was concerned. Tylman is of the opinion that if the water reservoir is
kept at 100°F, it is most comfortable to the patient, less liable to be harmful to
the pulp and still reduces the heat of friction during cutting.
9
There are certain other problems associated with the use of the
highspeed cutting tools. Most of the hand pieces are so designed that a spray
or stream of water is directed from the head of the handpiece directly onto the
cutting operation. Where the water strikes the tooth and the cutting tool
directly, full benefit is obtained from the coolant. Where however, the abrasive
on the cutting tool, is on the surface away from the stream of water, water does
not flood the tooth surface being cut, resulting in excessive temperature rise.
The overcome this difficulty perforated disks have been developed, which
permit the water to go through the openings and lubricate the disk and tooth on
the cutting side. The use of perforated disk results in less temperature rise.
Consequently when disks are non perforated, and when the stream of water
cannot be directed to the cutting contact areas, they should be used at speeds
not exceeding 10,000 rpm.
Another advantage of a water coolant lies in the fact that the tooth
debris from the cutting is removed rapidly, preventing the clogging of the
cutting tools. This results in greater cutting efficiency of the stone. Also, it
prolongs the life and effectiveness of the instrument. It is essential that the
water be in intimate contact with the revolving instrument and the tooth tissue.
To do this more effectively, Nelson recommended the addition of a wetting
agent to the water spray.
10
Because the high speed technique requires a larger quantity of water as
a coolant, there is the problem of removing this water from the mouth. To have
the dentist stop frequently to allow the patient to spit out the excess water is
time consuming. The customary saliva ejector has insufficient removal
capacity.
To solve this problem, Thompson has suggested a washed field
technique.
This technique adapts the suction or vacuum principle. It established
and maintains a powerful but gentle negative pressure of air in the mouth,
close to the field of operation.
Accompanying the air stream, is a flow of isothermal water which is
projected copiously onto the operative field. This water is entrained into the
vacuum air stream, which draws it rapidly across the operative area. The
irrigant pulls away with it tooth cuttings and debris. These are taken into the
vacuum air stream and disposed off in a filter system. A clean, clearly visible
operative field is provided. This technique has the distinct advantages that it
facilitates the use of high speed instruments, maintains visibility during
copious irrigation of the operative field, reduces operating time, improves the
patient’s well being and introduces a new concept of cleanliness. Human
11
tissues are maintained in their natured wet safe pain, trauma and postoperative
complication, which may arise due to ingestion of tooth debris are reduced.
Desiccation of hard and soft tissues is avoided. Heat is eliminated thus
preserved the tissues.
Vibration:
Cutting a tooth may be very annoying and unpleasant to the patient but
still not be painful. In pain there is usually an involvement of the nerve
endings, either by trauma or extreme irritation, resulting in an acute, painful
reaction. Most patients associate the sensation of vibration, noise, pressure and
the slight increase in cutting temperature with the sensation of pain.
Consequently, if the factors of vibration, heat and pressure are reduced to a
minimum, the patient usually experience reduced or no pain.
One mechanical factor that influences vibration is the dental handpiece,
whether it is friction-bearing, ball bearing, high speed belt driven or turbine
ultra speed. When the friction bearing, conventional type of handpiece is used
at a speed of 4500 rpm to 6000 rpm, it is connected by the conventional belt
and pulley system of the dental engine. In this case one may expect a high
order of vibration depending upon the condition of various mechanical parts,
their adjustment and speeds of their operation.
12
Pulleys that are worn, a worn belt, or an improperly adjusted belt will
cause vibrations that are transmitted down to the cutting tool. Similarly hand
piece which do not hold the cutting tool properly, which have worm bearing or
are out of adjustment will also cause vibration.
The investigations of Walsh and Symmoss showed that vibration, when
applied to tooth, produced the most unfavorable response when the frequency
was between 100 cps and 200 cps. When the frequencies were above 1000 cps,
they were generally beyond the upper threshold of perception of the average
patient. It is the lower frequencies, in the range of 100 – 200 cps, that are
usually developed at the lower speeds, especially if the equipment is worm and
maladjusted.
Hudson and Sweeney have reported the importance of having centricity
in the cutting tool. They found that eccentric burs when rotated at 6000 to
10000 rpm produced a lower frequency in the range of 100-200 cps, whereas a
true running bur at 10000 rpm produced vibrations in the frequency range
above the upper threshold.
Tamner pointed out that only a part of an eccentric cutting tool is used
as it rotates, thus causing unfavorable impacts and vibrations, which fall into
the most annoying frequency range. The disks and stones that are unmounted
and are screwed onto a mandrel very frequently are eccentric and therefore
13
should not be used in high speed cutting operations. The permanently mounted
instruments are indicated in preference to unmounted type.
Poorly built burs with blades not evenly cut or chipped will likewise
cause vibration. In using carbide burs, it is very important the see that none
have chipped blades.
Correct adjustment of the belt is important in the reduction on and
elimination of vibration. A belt that is too loose increases the vibration pattern
transmitted directly to the tooth.
In the ultra highspeed hand pieces the metal chuck holding the cutting
instrument often is replaced by a rubber or plastic chuck. This lessens the
vibration transmitted to the cutting instrument and facilitates the more rapid
cutting action.
In cutting with a water turbine handpiece at 45,000 rpm the intensity of
vibrations was well tolerates by the patient.
Morrison and Grinnel made the following observations.
The deleterious effects of vibration are two fold in origin.
1. Amplitude.
2. Undesirable modulating frequencies.
14
If we minimize or eliminate these factors, we can then reduce the
undesirable effects of vibration.
Amplitude: The wave of vibration consists of frequency and amplitude.
At conventional speeds, amplitude is greater but frequency is less. At
higher speeds the reverse is true. The greatest harm is caused by the amplitude
of vibration which is the factor, most destructive of instruments and which
causes the most apprehension in the patient and the greatest fatigue in the
dentist.
By increasing operating speeds, the amplitude and its effects are
reduced and a more satisfactory result is attained.
Vibration waves are measured in cycles per second. It has been shown
that rotation of approximately 6000 rpm sets up a vibrational wave of
approximately 100 cps. As the rpm is increased the cps of the fundamental
vibration wave are increased until, at ranges of 100000 rpm, we have a
vibration wave of 1600 cps. It has been demonstrated that at wave of vibration
of over 1300 cps, vibration is practically imperceptible to the patient. The
reason for this is not fully understood, but there are two theories for this
phenomenon.
15
(1) The Wedensky inhibition phenomenon – frequency increased to a point
where vibratory perception diminishes due to failure to perceive vibration.
This is because Stimulation occurs during Refractory Period of Recovery.
(2) Vibratory perception depends upon the product of the amount of
stimulation (i.e. pressure) multiplied by the frequency of stimulation
necessary for a reaction. This is called Chronaxie. As the speeds above
1,00,000 rpm, due to light pressure and high speeds, chronaxie is attained,
which is necessary for reaction.
Thus it can be concluded that, the more the rpm, the less the amplitude,
and the greater the frequency. Vibratory perception will be lost in the ultra
highspeed range of 1,00,000 rpm or more.
Spread of pathogenic organisms by Ultra speed cutting procedures:
Atmospheric contamination through the spread of oral organisms
particularly from air turbine action has been a concern of the dental profession
for some time. Dental procedures tend to expose the operator to infectious
diseases. Recent studies suggest that the extent of aerosol produced by air
turbine may increase the normal hazard. A report involving patients with
pulmonary tuberculosis cultures were demonstrated on all petri dishes exposed
during cutting procedures, with the heaviest concentration being at 2 feet in
distances from the patients mouth. This indicates that the dentist and his
16
assistant are exposed to a serious health hazard when operating with an ultra
speed exposed instrument on patients having such pathogens in their oral flora.
When a patient’s history suggests the existence of tuberculosis, pneumonia,
influenza, infections hepatitis or any infectious diseases including the common
cold, a protective face mask should be worn by both dentist and assistant.
During all ultrahigh speed cutting procedures, protective eye-glasses should be
worm routinely.
SUCK-BACK PHENOMENON- The operation of the turbine is switched
off by closing the compressed-air valve abruptly. Then, owing to its own
kinetic energy, the turbine continues its rotation, so that the turbine starts
operating as an air pump. This causes a negative pressure in the area of the
turbine shaft. The negative pressure sucks air from the environment that
can be contaminated by aerosols of saliva and blood of the patient.
Size of cutting instrument and cutting speeds:
It has been pointed out by Peyton, and Nelson that, the important factor
of increased operating speeds is the instrument surface speed in fact per
minute rather than the revolution per minute of the instrument.
The larger the diameter of the cutting instrument, the slower the speed
required at the spindle. The specific phase in preparation of an abutment
should determine the size of the cutting instrument and the rpm that should be
17
used. Employing superspeed for all operations places unnecessary strain upon
the patient and equipment. If the same effect can be accomplished by using a
larger instrument at a lower speed, but still remaining above the threshold of
perception, this should be done. However, oversized cutting tools should not
be used at super speeds due to the difficulty of instrument control and
accuracy of cutting.
VIBRATION SYNDROME : the perception of vibration, pain, touch
and temperature deteriorates. The negative effect of local vibrations occurs
within the range 5-1400 hz, the most harmful being those below 16 hz.
mechanical vibrations arise because the various machines operating at the
dentist’s workplace contain moving parts. The main source is vibrating
power-driven or air-driven instruments, such as low- and high-speed
handpieces as well as ultrasonic instruments. The
vibrations emitted by these machines travel directly from the handles to the
operator’s hand. These are local vibrations.
Biologic response of dentin and pulp to high speed cutting:
Dentin: As the contents of the dentinal tubules are in direct continuity with the
odontoblasts, and pulp, cutting or grinding the dentin causes a reaction in the
pulp and this may lead to changes in the dentin.
18
An early experimental investigation into the effect of cavity preparation
on the dentin and pulp was carried out by Fish in 1932. He cut cavities in the
teeth of dogs and left the cavities open to the saliva. By sealing dyes into the
pulp chambers of the treated teeth he has shown that one of two reactions is
produced in the dentin.
In some cases there was sclerosis of the cut dentinal tubules which
forms a protective some sealing off the pulp from the injury and underneath
this region, there is a further growth of tubular dentine. These reactions are
produced by the stimulation of the odontoblasts. The other reaction that
resulted was the formation of dead tracts. With this lesion some or all of the
odontoblasts, that are in connection with the cut tubules die. On the pulpal
aspect of these tubules, hyaline mineralized barrier, secondary dentine is laid
down, thereby sealing the lesion from the pulp.
Pulp: The changes in the pulp have been studied by Langeland and Morslard
and Shovelton. They state that the damage to the pulp is to a large extent due
to the heat generated. They have shown that when precautions are taken to
minimize heat production by using burs rotated slowly in a speed reducing
handpieces, the only evidence of pulp damage was a slight reduction of the
odontoblast layer with the displacement of a small number of odontoblasts into
the dentinal tubules. When speeds upto 5,000 rpm were employed, there was
more extensive displacement of odontoblasts associated with marked
19
vacuolization of the odontoblast layer, and local hemorrhages may be seen in
the pulp. As the speed was increased, the changes became more severe. When
tooth reduction was done under a stream or spray of water, the damage to the
pulp was markedly reduced.
Pulp changes associated with tooth reduction using the air abrasive
technique have been studied by Kennedy and using ultrasonic technique by
Mitchell and Jenson. The changes in both the cases are similar to those
produced at the speeds of 5,000 rpm.
The effects on the pulp of using high speed rotary instruments such as
the air turbine have been investigated by Marsland and Shovelton. The
changes found are no severe than those produced at lower operating speeds
provided that adequate cooling of the cutting instrument by water jet or
air/water spray is ensured.
Alterations in the hard tissues of tooth cut by air turbines have been
observed. The enamel over a wide area of crown may show minute cracks and
the dentin shows altered staining reactions as a result a local overheating.
RECENT DEVELOPMENT:
‘ANTI-SUCK BACK’- Planmeca compact dental units, the turbine drive
air is not shut off abruptly but controlled down by allowing the driving air
20
to decrease gradually. The software of the dental unit will keep on
supplying the drive air into the turbine according to carefully chosen
parameters. This way there is no possibility for the build-up of a vacuum
effect that would cause suck-back.
Ceramic bearings- no need of lubrication and more resistant to autoclave
sterlization.
Use of quartz rods instead of fibre-optic.
Easy-to-use push-button bur releases.
Swivel systems.
Titanium handpieces.
Smaller head size.
 ELECTRICAL HIGH SPEED HANDPIECES
advantages are:
More power and torque than air turbine handpieces.
Better bur concentricity.
Less vibration and noise.
Broad, controllable speed ranges.
21
Forward/reverse option are available.
With appropriate attachments, one system can be used for
restorative, prosthodontics, prophies and endodontics.
Disadvantages are :
Heavier than air turbine.
More expensive.
Learning curve may be required.
Attachment heads not as small as the small-head air turbines.
May not be able to fully replace the air turbine.
Infection control concerns.
Discussion
It is for more than 125 years, that rotary instruments have been in use,
for tooth reduction operations, in different forms, from a hand rotary
instrument to ultra sonic instruments, which have the rotational speeds ranging
from very low speeds in case of band rotary instruments to 5,00,000 rpm in
case of air bearing hand piece. These remarkable advances in the instruments
have greatly reduced fatigue in the operator because of the physical case of
manipulation and have considerably increased the comfort to the patient by
22
reducing the actual working time and pressures required for tooth reduction,
thereby minimizing the factors of heat generation and pain. Though high speed
techniques have been a born to the dental profession, they have their can
limitations. It is interesting to not that, in spite of considerable improvements
in tooth reduction procedures and the instruments used for the same, the
principles and biologic objectives have not changed.
These improved methods of tooth tissue removal have a potential to
damage the healthy teeth and surrounding structures, if they are used without
proper understanding of their working and if they are used without taking
proper care. Improper handling of these modern equipment may also be
different to the longevity and working capacity of the instruments themselves.
For successful and efficient use of those cutting tools, certain factors
should be given consideration. Heat that is generated, while the tooth tissue is
being removed must be kept, down to the minimum and at the sametime,
whatever heat is generated, must be eliminated as efficiently and as quickly as
possible by employing coolants, in any one of three forms commonly used i.e.
water, air/water spray or air alone. Simultaneously with the coolant, if water or
air/water spray is used, an efficient mechanism for remove of the water from
the oral cavity must be employed. Otherwise, the clinical procedure is delayed,
if the patient has to spit out the water, every now and then. By eliminating the
water evacuation equipment, we are losing one of the advantages of these high
23
speed instruments i.e. reduced working time for a particular preparation. Use
of efficient coolants, not only eliminate the heat generated, but at the same
time, keeps the operating area clean and free of any debris.
High speed cutting methods have a further advantages in that, they
reduce the annoyance that may be caused to the patient, when low speeds are
used with the modern high speed cutting devices, the vibration produced is of
a frequency that is generally beyond the upper threshold of perception of the
average patient.
Pressures that have to be employed in the use of high speeds are
considerably reduced, in comparison with those needed for low speeds.
Thus, when the factors of pressure, temperature and vibration are kept
within the tolerance limits, the patient comfort is certainly improved.
Size of the cutting tool to be used for particular tooth reduction
procedure is an important consideration, particularly while using high speeds.
Oversized cutting tools should be avoided, as they are difficult to control and
at the same time, the accuracy of tooth preparation on procedure is also
adversely affected.
Biologic reactions of the tooth tissues, particularly dentin and pulp,
should not be over locked, when high speeds are employed for tooth reduction
24
operation. These responses have been studied by a number of people and they
have shown that, the response are not significantly different from those, when
low speeds are used, provided, effective coolants are employed.
Thus it can be concluded that, high speed equipments for tooth
reduction if used with proper understanding and due care, provide definite
advantages over the conventional low speed cutting procedures. This fact
places the high speed devices at definitely a higher level as against their low
speed counterparts.
Conclusions
1. High speed cutting devices, if used with a thorough understanding of their
mechanism and due care to the biologic integrity of teeth and surrounding
structure, are a boon to dentistry.
2. In the process of tooth reduction using high speeds considerable amount of
heat is generated and an effective coolant is a must for preservation of
tooth integrity and patient comfort.
3. Vibration is increased with the increase in speed, but it is beyond the
threshold of prerception of the normal human beings and hence not
harmful.
25
4. Biologic reactions of the dentin and pulp, to high speed cutting, cannot be
overlooked.
Summary
A brief history of rotary instruments has been presented. A critical
evaluation of the high speed cutting devices, as to their advantages,
disadvantages, and precautions to be taken during their use, has been assessed
at length. Biologic reactions of dentin and pulp, to high speed cutting have
been discussed in brief.
26
Contents
I. Introduction
II. Review of Literature
a. Advantages of high speeds
b. Disadvantages of high speeds
c. Types of high speed instruments
d. Heat generation
e. Coolants
f. Vibration
g. Spread of pathogenic organisms
h. Size of cutting instruments and cutting speeds
i. Biologic responses of dentin and pulp to high speed cutting
I. Discussion
II. Conclusion
III. Summary
27

Más contenido relacionado

La actualidad más candente

EVIDENCE-BASED APPROACH TO ROOT CANAL CLEANING AND SHAPING / /certified fixed...
EVIDENCE-BASED APPROACH TO ROOT CANAL CLEANING AND SHAPING / /certified fixed...EVIDENCE-BASED APPROACH TO ROOT CANAL CLEANING AND SHAPING / /certified fixed...
EVIDENCE-BASED APPROACH TO ROOT CANAL CLEANING AND SHAPING / /certified fixed...Indian dental academy
 
gingival retraction methods
gingival retraction methodsgingival retraction methods
gingival retraction methodsIshmeen Kaur
 
Gingival Tissue Management
Gingival Tissue ManagementGingival Tissue Management
Gingival Tissue Managementshabeel pn
 
Endodontic complications
Endodontic complicationsEndodontic complications
Endodontic complicationsMohammed Rhael
 
Clinical endodontics (treatment)
Clinical endodontics (treatment)Clinical endodontics (treatment)
Clinical endodontics (treatment)splendidlight
 
Rotary cutting inst  /orthodontic courses by Indian dental academy 
Rotary cutting inst  /orthodontic courses by Indian dental academy Rotary cutting inst  /orthodontic courses by Indian dental academy 
Rotary cutting inst  /orthodontic courses by Indian dental academy Indian dental academy
 
Sonic & ultrasonic instruments
Sonic & ultrasonic instruments Sonic & ultrasonic instruments
Sonic & ultrasonic instruments Shikha Arya
 
Fluid control & soft tissue management in fpd
Fluid control & soft tissue management in fpdFluid control & soft tissue management in fpd
Fluid control & soft tissue management in fpdvesta enid lydia
 
Dental Rotary cutting instruments /certified fixed orthodontic courses by Ind...
Dental Rotary cutting instruments /certified fixed orthodontic courses by Ind...Dental Rotary cutting instruments /certified fixed orthodontic courses by Ind...
Dental Rotary cutting instruments /certified fixed orthodontic courses by Ind...Indian dental academy
 
Fluid control and ginigival retraction
Fluid control and ginigival retractionFluid control and ginigival retraction
Fluid control and ginigival retractionSayali Raut
 
Ultrasonics in endodontics
Ultrasonics in endodonticsUltrasonics in endodontics
Ultrasonics in endodonticsDeepesh Mehta
 
Instruments used in endodontic treatment
Instruments used in endodontic treatmentInstruments used in endodontic treatment
Instruments used in endodontic treatmentpansovannara
 
Ultrasonic in dentistry
Ultrasonic in dentistryUltrasonic in dentistry
Ultrasonic in dentistryJeff Joy
 
Removal of broken endodontic instruments
Removal of broken endodontic instrumentsRemoval of broken endodontic instruments
Removal of broken endodontic instrumentsPalaniselvi Kamaraj
 
Fluid control and_tissue_management
Fluid control and_tissue_managementFluid control and_tissue_management
Fluid control and_tissue_managementSelva Arockiam
 
Endodontics. anatomy of root canals. instruments
Endodontics. anatomy of root canals. instrumentsEndodontics. anatomy of root canals. instruments
Endodontics. anatomy of root canals. instrumentsLinda Jenhani
 

La actualidad más candente (20)

Speeds in dentistry
Speeds in dentistrySpeeds in dentistry
Speeds in dentistry
 
EVIDENCE-BASED APPROACH TO ROOT CANAL CLEANING AND SHAPING / /certified fixed...
EVIDENCE-BASED APPROACH TO ROOT CANAL CLEANING AND SHAPING / /certified fixed...EVIDENCE-BASED APPROACH TO ROOT CANAL CLEANING AND SHAPING / /certified fixed...
EVIDENCE-BASED APPROACH TO ROOT CANAL CLEANING AND SHAPING / /certified fixed...
 
gingival retraction methods
gingival retraction methodsgingival retraction methods
gingival retraction methods
 
Gingival Tissue Management
Gingival Tissue ManagementGingival Tissue Management
Gingival Tissue Management
 
Endodontic complications
Endodontic complicationsEndodontic complications
Endodontic complications
 
Clinical endodontics (treatment)
Clinical endodontics (treatment)Clinical endodontics (treatment)
Clinical endodontics (treatment)
 
Rotary cutting inst  /orthodontic courses by Indian dental academy 
Rotary cutting inst  /orthodontic courses by Indian dental academy Rotary cutting inst  /orthodontic courses by Indian dental academy 
Rotary cutting inst  /orthodontic courses by Indian dental academy 
 
Sonic & ultrasonic instruments
Sonic & ultrasonic instruments Sonic & ultrasonic instruments
Sonic & ultrasonic instruments
 
Armamentarium 2
Armamentarium 2Armamentarium 2
Armamentarium 2
 
Fluid control and tissue managemtent
Fluid control and tissue managemtentFluid control and tissue managemtent
Fluid control and tissue managemtent
 
Burs in dentisty ashish
Burs in dentisty ashishBurs in dentisty ashish
Burs in dentisty ashish
 
Fluid control & soft tissue management in fpd
Fluid control & soft tissue management in fpdFluid control & soft tissue management in fpd
Fluid control & soft tissue management in fpd
 
Dental Rotary cutting instruments /certified fixed orthodontic courses by Ind...
Dental Rotary cutting instruments /certified fixed orthodontic courses by Ind...Dental Rotary cutting instruments /certified fixed orthodontic courses by Ind...
Dental Rotary cutting instruments /certified fixed orthodontic courses by Ind...
 
Fluid control and ginigival retraction
Fluid control and ginigival retractionFluid control and ginigival retraction
Fluid control and ginigival retraction
 
Ultrasonics in endodontics
Ultrasonics in endodonticsUltrasonics in endodontics
Ultrasonics in endodontics
 
Instruments used in endodontic treatment
Instruments used in endodontic treatmentInstruments used in endodontic treatment
Instruments used in endodontic treatment
 
Ultrasonic in dentistry
Ultrasonic in dentistryUltrasonic in dentistry
Ultrasonic in dentistry
 
Removal of broken endodontic instruments
Removal of broken endodontic instrumentsRemoval of broken endodontic instruments
Removal of broken endodontic instruments
 
Fluid control and_tissue_management
Fluid control and_tissue_managementFluid control and_tissue_management
Fluid control and_tissue_management
 
Endodontics. anatomy of root canals. instruments
Endodontics. anatomy of root canals. instrumentsEndodontics. anatomy of root canals. instruments
Endodontics. anatomy of root canals. instruments
 

Destacado

Morarji desai
Morarji desai Morarji desai
Morarji desai sujay762
 
10 most profitable changes to my dental office
10 most profitable changes to my dental office10 most profitable changes to my dental office
10 most profitable changes to my dental officeemailankur
 
Ethics in Dentistry
Ethics in DentistryEthics in Dentistry
Ethics in Dentistryntkhab24
 
Key performance indicators for dentists dental practice management
Key performance indicators for dentists dental practice managementKey performance indicators for dentists dental practice management
Key performance indicators for dentists dental practice managementDentivaNL
 
Creative Concepts Presentation
Creative Concepts PresentationCreative Concepts Presentation
Creative Concepts Presentationacquarus
 
30 Law Practice Management Tips
30 Law Practice Management Tips30 Law Practice Management Tips
30 Law Practice Management Tipslgladm59
 
06 professional ethics & code of ethics
06 professional ethics & code of ethics06 professional ethics & code of ethics
06 professional ethics & code of ethicsNimantha Perera
 
Ethics in the workplace
Ethics in the workplaceEthics in the workplace
Ethics in the workplaceviviandabu
 
Professional Ethics
Professional EthicsProfessional Ethics
Professional EthicsKrish Jagan
 
Introduction of business ethics
Introduction of business ethicsIntroduction of business ethics
Introduction of business ethicsneha16sept
 
Importance of-business-ethics
Importance of-business-ethicsImportance of-business-ethics
Importance of-business-ethicsSyed Arslan
 

Destacado (15)

Skills in pharmaceutical care 2014 N.14
Skills in pharmaceutical care 2014 N.14Skills in pharmaceutical care 2014 N.14
Skills in pharmaceutical care 2014 N.14
 
Morarji desai
Morarji desai Morarji desai
Morarji desai
 
Orange_2_Presentacion_V3
Orange_2_Presentacion_V3Orange_2_Presentacion_V3
Orange_2_Presentacion_V3
 
Creative Concept
Creative ConceptCreative Concept
Creative Concept
 
10 most profitable changes to my dental office
10 most profitable changes to my dental office10 most profitable changes to my dental office
10 most profitable changes to my dental office
 
Ethics in Dentistry
Ethics in DentistryEthics in Dentistry
Ethics in Dentistry
 
Key performance indicators for dentists dental practice management
Key performance indicators for dentists dental practice managementKey performance indicators for dentists dental practice management
Key performance indicators for dentists dental practice management
 
Creative Concepts Presentation
Creative Concepts PresentationCreative Concepts Presentation
Creative Concepts Presentation
 
30 Law Practice Management Tips
30 Law Practice Management Tips30 Law Practice Management Tips
30 Law Practice Management Tips
 
06 professional ethics & code of ethics
06 professional ethics & code of ethics06 professional ethics & code of ethics
06 professional ethics & code of ethics
 
Ethics in the workplace
Ethics in the workplaceEthics in the workplace
Ethics in the workplace
 
Professional Ethics
Professional EthicsProfessional Ethics
Professional Ethics
 
Waste Management
Waste ManagementWaste Management
Waste Management
 
Introduction of business ethics
Introduction of business ethicsIntroduction of business ethics
Introduction of business ethics
 
Importance of-business-ethics
Importance of-business-ethicsImportance of-business-ethics
Importance of-business-ethics
 

Similar a High speed cutting/dental lab technology courses by Indian dental academy

Rotary cutting instruments in dentistry /certified fixed orthodontic courses ...
Rotary cutting instruments in dentistry /certified fixed orthodontic courses ...Rotary cutting instruments in dentistry /certified fixed orthodontic courses ...
Rotary cutting instruments in dentistry /certified fixed orthodontic courses ...Indian dental academy
 
Air abrasion technology in conservative dentistry
Air abrasion technology in conservative dentistryAir abrasion technology in conservative dentistry
Air abrasion technology in conservative dentistrySwathe Siva
 
CLEANING AND SHAPING IN ENDODONTICS
CLEANING AND SHAPING IN ENDODONTICSCLEANING AND SHAPING IN ENDODONTICS
CLEANING AND SHAPING IN ENDODONTICSDr. Aadil Thimwala
 
rotary cutting instruments in conservative dentistry
rotary cutting instruments in conservative dentistryrotary cutting instruments in conservative dentistry
rotary cutting instruments in conservative dentistryVasundhara naik
 
Recent advances in airotor..........pptx
Recent advances in airotor..........pptxRecent advances in airotor..........pptx
Recent advances in airotor..........pptxHimanshu Goyal
 
Drilling sequence in implants
Drilling sequence in implantsDrilling sequence in implants
Drilling sequence in implantsDr.Nazia
 
orthodontic appliance and treatment philosophy.docx
orthodontic appliance and treatment philosophy.docxorthodontic appliance and treatment philosophy.docx
orthodontic appliance and treatment philosophy.docxDr.Mohammed Alruby
 
Rotary instruments in operative dentistry
Rotary instruments in operative dentistryRotary instruments in operative dentistry
Rotary instruments in operative dentistryDr. Arpit Viradiya
 
Expansion in orthodontics
Expansion in orthodonticsExpansion in orthodontics
Expansion in orthodonticsSk Aziz Ikbal
 
MECHANICAL PLAQUE CONTROL.pptx
MECHANICAL PLAQUE CONTROL.pptxMECHANICAL PLAQUE CONTROL.pptx
MECHANICAL PLAQUE CONTROL.pptxMustafaYounis12
 
Ultrasonics in Endodontics.pptx
Ultrasonics in Endodontics.pptxUltrasonics in Endodontics.pptx
Ultrasonics in Endodontics.pptxSiddheshKokitkar
 
Rotary Endodontic Instruments
Rotary Endodontic InstrumentsRotary Endodontic Instruments
Rotary Endodontic InstrumentsDr. Sahana Umesh
 
How to-choose-air-rotor-the-ultimate-guide-converted
How to-choose-air-rotor-the-ultimate-guide-convertedHow to-choose-air-rotor-the-ultimate-guide-converted
How to-choose-air-rotor-the-ultimate-guide-convertedUnicorn Denmart
 

Similar a High speed cutting/dental lab technology courses by Indian dental academy (20)

Rotary cutting instruments in dentistry /certified fixed orthodontic courses ...
Rotary cutting instruments in dentistry /certified fixed orthodontic courses ...Rotary cutting instruments in dentistry /certified fixed orthodontic courses ...
Rotary cutting instruments in dentistry /certified fixed orthodontic courses ...
 
Speeds in dentistry
Speeds in dentistry Speeds in dentistry
Speeds in dentistry
 
Dental driller
Dental drillerDental driller
Dental driller
 
High speeds in dentistry (2)
High speeds in dentistry (2)High speeds in dentistry (2)
High speeds in dentistry (2)
 
High speeds in dentistry
High speeds in dentistryHigh speeds in dentistry
High speeds in dentistry
 
Air abrasion technology in conservative dentistry
Air abrasion technology in conservative dentistryAir abrasion technology in conservative dentistry
Air abrasion technology in conservative dentistry
 
Speed in dentistry
Speed in dentistrySpeed in dentistry
Speed in dentistry
 
CLEANING AND SHAPING IN ENDODONTICS
CLEANING AND SHAPING IN ENDODONTICSCLEANING AND SHAPING IN ENDODONTICS
CLEANING AND SHAPING IN ENDODONTICS
 
rotary cutting instruments in conservative dentistry
rotary cutting instruments in conservative dentistryrotary cutting instruments in conservative dentistry
rotary cutting instruments in conservative dentistry
 
Clear aligner therapy
Clear aligner therapyClear aligner therapy
Clear aligner therapy
 
Recent advances in airotor..........pptx
Recent advances in airotor..........pptxRecent advances in airotor..........pptx
Recent advances in airotor..........pptx
 
Drilling sequence in implants
Drilling sequence in implantsDrilling sequence in implants
Drilling sequence in implants
 
orthodontic appliance and treatment philosophy.docx
orthodontic appliance and treatment philosophy.docxorthodontic appliance and treatment philosophy.docx
orthodontic appliance and treatment philosophy.docx
 
Rotary instruments in operative dentistry
Rotary instruments in operative dentistryRotary instruments in operative dentistry
Rotary instruments in operative dentistry
 
Expansion in orthodontics
Expansion in orthodonticsExpansion in orthodontics
Expansion in orthodontics
 
MECHANICAL PLAQUE CONTROL.pptx
MECHANICAL PLAQUE CONTROL.pptxMECHANICAL PLAQUE CONTROL.pptx
MECHANICAL PLAQUE CONTROL.pptx
 
Interproximal enamel reduction
Interproximal enamel reductionInterproximal enamel reduction
Interproximal enamel reduction
 
Ultrasonics in Endodontics.pptx
Ultrasonics in Endodontics.pptxUltrasonics in Endodontics.pptx
Ultrasonics in Endodontics.pptx
 
Rotary Endodontic Instruments
Rotary Endodontic InstrumentsRotary Endodontic Instruments
Rotary Endodontic Instruments
 
How to-choose-air-rotor-the-ultimate-guide-converted
How to-choose-air-rotor-the-ultimate-guide-convertedHow to-choose-air-rotor-the-ultimate-guide-converted
How to-choose-air-rotor-the-ultimate-guide-converted
 

Más de Indian dental academy

Indian Dentist - relocate to united kingdom
Indian Dentist - relocate to united kingdomIndian Dentist - relocate to united kingdom
Indian Dentist - relocate to united kingdomIndian dental academy
 
1ST, 2ND AND 3RD ORDER BENDS IN STANDARD EDGEWISE APPLIANCE SYSTEM /Fixed ort...
1ST, 2ND AND 3RD ORDER BENDS IN STANDARD EDGEWISE APPLIANCE SYSTEM /Fixed ort...1ST, 2ND AND 3RD ORDER BENDS IN STANDARD EDGEWISE APPLIANCE SYSTEM /Fixed ort...
1ST, 2ND AND 3RD ORDER BENDS IN STANDARD EDGEWISE APPLIANCE SYSTEM /Fixed ort...Indian dental academy
 
Invisalign -invisible aligners course in india
Invisalign -invisible aligners course in india Invisalign -invisible aligners course in india
Invisalign -invisible aligners course in india Indian dental academy
 
Invisible aligners for your orthodontics pratice
Invisible aligners for your orthodontics praticeInvisible aligners for your orthodontics pratice
Invisible aligners for your orthodontics praticeIndian dental academy
 
Development of muscles of mastication / dental implant courses
Development of muscles of mastication / dental implant coursesDevelopment of muscles of mastication / dental implant courses
Development of muscles of mastication / dental implant coursesIndian dental academy
 
Corticosteriods uses in dentistry/ oral surgery courses  
Corticosteriods uses in dentistry/ oral surgery courses  Corticosteriods uses in dentistry/ oral surgery courses  
Corticosteriods uses in dentistry/ oral surgery courses  Indian dental academy
 
Cytotoxicity of silicone materials used in maxillofacial prosthesis / dental ...
Cytotoxicity of silicone materials used in maxillofacial prosthesis / dental ...Cytotoxicity of silicone materials used in maxillofacial prosthesis / dental ...
Cytotoxicity of silicone materials used in maxillofacial prosthesis / dental ...Indian dental academy
 
Diagnosis and treatment planning in completely endntulous arches/dental courses
Diagnosis and treatment planning in completely endntulous arches/dental coursesDiagnosis and treatment planning in completely endntulous arches/dental courses
Diagnosis and treatment planning in completely endntulous arches/dental coursesIndian dental academy
 
Properties of Denture base materials /rotary endodontic courses
Properties of Denture base materials /rotary endodontic coursesProperties of Denture base materials /rotary endodontic courses
Properties of Denture base materials /rotary endodontic coursesIndian dental academy
 
Use of modified tooth forms in complete denture occlusion / dental implant...
Use of modified  tooth forms  in  complete denture occlusion / dental implant...Use of modified  tooth forms  in  complete denture occlusion / dental implant...
Use of modified tooth forms in complete denture occlusion / dental implant...Indian dental academy
 
Dental luting cements / oral surgery courses  
Dental   luting cements / oral surgery courses  Dental   luting cements / oral surgery courses  
Dental luting cements / oral surgery courses  Indian dental academy
 
Dental casting alloys/ oral surgery courses  
Dental casting alloys/ oral surgery courses  Dental casting alloys/ oral surgery courses  
Dental casting alloys/ oral surgery courses  Indian dental academy
 
Dental casting investment materials/endodontic courses
Dental casting investment materials/endodontic coursesDental casting investment materials/endodontic courses
Dental casting investment materials/endodontic coursesIndian dental academy
 
Dental casting waxes/ oral surgery courses  
Dental casting waxes/ oral surgery courses  Dental casting waxes/ oral surgery courses  
Dental casting waxes/ oral surgery courses  Indian dental academy
 
Dental ceramics/prosthodontic courses
Dental ceramics/prosthodontic coursesDental ceramics/prosthodontic courses
Dental ceramics/prosthodontic coursesIndian dental academy
 
Dental implant/ oral surgery courses  
Dental implant/ oral surgery courses  Dental implant/ oral surgery courses  
Dental implant/ oral surgery courses  Indian dental academy
 
Dental perspective/cosmetic dentistry courses
Dental perspective/cosmetic dentistry coursesDental perspective/cosmetic dentistry courses
Dental perspective/cosmetic dentistry coursesIndian dental academy
 
Dental tissues and their replacements/ oral surgery courses  
Dental tissues and their replacements/ oral surgery courses  Dental tissues and their replacements/ oral surgery courses  
Dental tissues and their replacements/ oral surgery courses  Indian dental academy
 

Más de Indian dental academy (20)

Indian Dentist - relocate to united kingdom
Indian Dentist - relocate to united kingdomIndian Dentist - relocate to united kingdom
Indian Dentist - relocate to united kingdom
 
1ST, 2ND AND 3RD ORDER BENDS IN STANDARD EDGEWISE APPLIANCE SYSTEM /Fixed ort...
1ST, 2ND AND 3RD ORDER BENDS IN STANDARD EDGEWISE APPLIANCE SYSTEM /Fixed ort...1ST, 2ND AND 3RD ORDER BENDS IN STANDARD EDGEWISE APPLIANCE SYSTEM /Fixed ort...
1ST, 2ND AND 3RD ORDER BENDS IN STANDARD EDGEWISE APPLIANCE SYSTEM /Fixed ort...
 
Invisalign -invisible aligners course in india
Invisalign -invisible aligners course in india Invisalign -invisible aligners course in india
Invisalign -invisible aligners course in india
 
Invisible aligners for your orthodontics pratice
Invisible aligners for your orthodontics praticeInvisible aligners for your orthodontics pratice
Invisible aligners for your orthodontics pratice
 
online fixed orthodontics course
online fixed orthodontics courseonline fixed orthodontics course
online fixed orthodontics course
 
online orthodontics course
online orthodontics courseonline orthodontics course
online orthodontics course
 
Development of muscles of mastication / dental implant courses
Development of muscles of mastication / dental implant coursesDevelopment of muscles of mastication / dental implant courses
Development of muscles of mastication / dental implant courses
 
Corticosteriods uses in dentistry/ oral surgery courses  
Corticosteriods uses in dentistry/ oral surgery courses  Corticosteriods uses in dentistry/ oral surgery courses  
Corticosteriods uses in dentistry/ oral surgery courses  
 
Cytotoxicity of silicone materials used in maxillofacial prosthesis / dental ...
Cytotoxicity of silicone materials used in maxillofacial prosthesis / dental ...Cytotoxicity of silicone materials used in maxillofacial prosthesis / dental ...
Cytotoxicity of silicone materials used in maxillofacial prosthesis / dental ...
 
Diagnosis and treatment planning in completely endntulous arches/dental courses
Diagnosis and treatment planning in completely endntulous arches/dental coursesDiagnosis and treatment planning in completely endntulous arches/dental courses
Diagnosis and treatment planning in completely endntulous arches/dental courses
 
Properties of Denture base materials /rotary endodontic courses
Properties of Denture base materials /rotary endodontic coursesProperties of Denture base materials /rotary endodontic courses
Properties of Denture base materials /rotary endodontic courses
 
Use of modified tooth forms in complete denture occlusion / dental implant...
Use of modified  tooth forms  in  complete denture occlusion / dental implant...Use of modified  tooth forms  in  complete denture occlusion / dental implant...
Use of modified tooth forms in complete denture occlusion / dental implant...
 
Dental luting cements / oral surgery courses  
Dental   luting cements / oral surgery courses  Dental   luting cements / oral surgery courses  
Dental luting cements / oral surgery courses  
 
Dental casting alloys/ oral surgery courses  
Dental casting alloys/ oral surgery courses  Dental casting alloys/ oral surgery courses  
Dental casting alloys/ oral surgery courses  
 
Dental casting investment materials/endodontic courses
Dental casting investment materials/endodontic coursesDental casting investment materials/endodontic courses
Dental casting investment materials/endodontic courses
 
Dental casting waxes/ oral surgery courses  
Dental casting waxes/ oral surgery courses  Dental casting waxes/ oral surgery courses  
Dental casting waxes/ oral surgery courses  
 
Dental ceramics/prosthodontic courses
Dental ceramics/prosthodontic coursesDental ceramics/prosthodontic courses
Dental ceramics/prosthodontic courses
 
Dental implant/ oral surgery courses  
Dental implant/ oral surgery courses  Dental implant/ oral surgery courses  
Dental implant/ oral surgery courses  
 
Dental perspective/cosmetic dentistry courses
Dental perspective/cosmetic dentistry coursesDental perspective/cosmetic dentistry courses
Dental perspective/cosmetic dentistry courses
 
Dental tissues and their replacements/ oral surgery courses  
Dental tissues and their replacements/ oral surgery courses  Dental tissues and their replacements/ oral surgery courses  
Dental tissues and their replacements/ oral surgery courses  
 

Último

NEONATAL RESPIRATORY CARE FROM A PHYSIO POV.pptx
NEONATAL RESPIRATORY CARE FROM A PHYSIO POV.pptxNEONATAL RESPIRATORY CARE FROM A PHYSIO POV.pptx
NEONATAL RESPIRATORY CARE FROM A PHYSIO POV.pptxHanineHassan2
 
Identifying Signs of Mental Health Presentation (1).pptx
Identifying Signs of Mental Health Presentation (1).pptxIdentifying Signs of Mental Health Presentation (1).pptx
Identifying Signs of Mental Health Presentation (1).pptxsandhulove46637
 
Understanding Warts and Moles: Differences, Types, and Common Locations
Understanding Warts and Moles: Differences, Types, and Common LocationsUnderstanding Warts and Moles: Differences, Types, and Common Locations
Understanding Warts and Moles: Differences, Types, and Common LocationsNeha Sharma
 
Eating Disorders in Athletes I Sports Psychology
Eating Disorders in Athletes I Sports PsychologyEating Disorders in Athletes I Sports Psychology
Eating Disorders in Athletes I Sports Psychologyshantisphysio
 
Diseases of the Respiratory System (J00-J99),.pptx
Diseases of the Respiratory System (J00-J99),.pptxDiseases of the Respiratory System (J00-J99),.pptx
Diseases of the Respiratory System (J00-J99),.pptxEMADABATHINI PRABHU TEJA
 
person with disability and pwd act ppt.pptx
person with disability and pwd act ppt.pptxperson with disability and pwd act ppt.pptx
person with disability and pwd act ppt.pptxMUKESH PADMANABHAN
 
21 NEMT Trends & Statistics to Know in 2024
21 NEMT Trends & Statistics to Know in 202421 NEMT Trends & Statistics to Know in 2024
21 NEMT Trends & Statistics to Know in 2024Traumasoft LLC
 
CECT NECK NECK ANGIOGRAPHY CAROTID ANGIOGRAPHY
CECT NECK NECK ANGIOGRAPHY CAROTID ANGIOGRAPHYCECT NECK NECK ANGIOGRAPHY CAROTID ANGIOGRAPHY
CECT NECK NECK ANGIOGRAPHY CAROTID ANGIOGRAPHYRMC
 
EYE CANCER.pptx prepared by Neha kewat digital learning
EYE CANCER.pptx prepared by  Neha kewat digital learningEYE CANCER.pptx prepared by  Neha kewat digital learning
EYE CANCER.pptx prepared by Neha kewat digital learningNehaKewat
 
2024 Compliatric Webianr Series - Contracts and MOUs from a HRSA Perspective.pdf
2024 Compliatric Webianr Series - Contracts and MOUs from a HRSA Perspective.pdf2024 Compliatric Webianr Series - Contracts and MOUs from a HRSA Perspective.pdf
2024 Compliatric Webianr Series - Contracts and MOUs from a HRSA Perspective.pdfCompliatric Where Compliance Happens
 
"ANATOMY AND PHYSIOLOGY OF THE SKIN".pdf
"ANATOMY AND PHYSIOLOGY OF THE SKIN".pdf"ANATOMY AND PHYSIOLOGY OF THE SKIN".pdf
"ANATOMY AND PHYSIOLOGY OF THE SKIN".pdfDolisha Warbi
 
Health literacies in marginalised communities LILAC 24.pptx
Health literacies in marginalised communities LILAC 24.pptxHealth literacies in marginalised communities LILAC 24.pptx
Health literacies in marginalised communities LILAC 24.pptxPamela McKinney
 
Assisted Living Care Residency - PapayaCare
Assisted Living Care Residency - PapayaCareAssisted Living Care Residency - PapayaCare
Assisted Living Care Residency - PapayaCareratilalthakkar704
 
Hematinics and Erythropoietin- Pharmacology of Hematinics
Hematinics and Erythropoietin- Pharmacology of HematinicsHematinics and Erythropoietin- Pharmacology of Hematinics
Hematinics and Erythropoietin- Pharmacology of Hematinicsnetraangadi2
 
Pharmacovigilance audits inspections.pptx
Pharmacovigilance audits inspections.pptxPharmacovigilance audits inspections.pptx
Pharmacovigilance audits inspections.pptxCliniminds India
 
Introduction to Evaluation and Skin Benefits
Introduction to Evaluation and Skin BenefitsIntroduction to Evaluation and Skin Benefits
Introduction to Evaluation and Skin Benefitssahilgabhane29
 
Empathy Is a Stress Response - Choose Compassion instead
Empathy Is a Stress Response - Choose Compassion insteadEmpathy Is a Stress Response - Choose Compassion instead
Empathy Is a Stress Response - Choose Compassion insteadAlex Clapson
 

Último (20)

NEONATAL RESPIRATORY CARE FROM A PHYSIO POV.pptx
NEONATAL RESPIRATORY CARE FROM A PHYSIO POV.pptxNEONATAL RESPIRATORY CARE FROM A PHYSIO POV.pptx
NEONATAL RESPIRATORY CARE FROM A PHYSIO POV.pptx
 
Identifying Signs of Mental Health Presentation (1).pptx
Identifying Signs of Mental Health Presentation (1).pptxIdentifying Signs of Mental Health Presentation (1).pptx
Identifying Signs of Mental Health Presentation (1).pptx
 
Understanding Warts and Moles: Differences, Types, and Common Locations
Understanding Warts and Moles: Differences, Types, and Common LocationsUnderstanding Warts and Moles: Differences, Types, and Common Locations
Understanding Warts and Moles: Differences, Types, and Common Locations
 
SCOPE OF CRITICAL CARE ORGANIZATION
SCOPE OF CRITICAL CARE ORGANIZATIONSCOPE OF CRITICAL CARE ORGANIZATION
SCOPE OF CRITICAL CARE ORGANIZATION
 
Eating Disorders in Athletes I Sports Psychology
Eating Disorders in Athletes I Sports PsychologyEating Disorders in Athletes I Sports Psychology
Eating Disorders in Athletes I Sports Psychology
 
Diseases of the Respiratory System (J00-J99),.pptx
Diseases of the Respiratory System (J00-J99),.pptxDiseases of the Respiratory System (J00-J99),.pptx
Diseases of the Respiratory System (J00-J99),.pptx
 
person with disability and pwd act ppt.pptx
person with disability and pwd act ppt.pptxperson with disability and pwd act ppt.pptx
person with disability and pwd act ppt.pptx
 
21 NEMT Trends & Statistics to Know in 2024
21 NEMT Trends & Statistics to Know in 202421 NEMT Trends & Statistics to Know in 2024
21 NEMT Trends & Statistics to Know in 2024
 
CECT NECK NECK ANGIOGRAPHY CAROTID ANGIOGRAPHY
CECT NECK NECK ANGIOGRAPHY CAROTID ANGIOGRAPHYCECT NECK NECK ANGIOGRAPHY CAROTID ANGIOGRAPHY
CECT NECK NECK ANGIOGRAPHY CAROTID ANGIOGRAPHY
 
EYE CANCER.pptx prepared by Neha kewat digital learning
EYE CANCER.pptx prepared by  Neha kewat digital learningEYE CANCER.pptx prepared by  Neha kewat digital learning
EYE CANCER.pptx prepared by Neha kewat digital learning
 
Annual Training
Annual TrainingAnnual Training
Annual Training
 
2024 Compliatric Webianr Series - Contracts and MOUs from a HRSA Perspective.pdf
2024 Compliatric Webianr Series - Contracts and MOUs from a HRSA Perspective.pdf2024 Compliatric Webianr Series - Contracts and MOUs from a HRSA Perspective.pdf
2024 Compliatric Webianr Series - Contracts and MOUs from a HRSA Perspective.pdf
 
"ANATOMY AND PHYSIOLOGY OF THE SKIN".pdf
"ANATOMY AND PHYSIOLOGY OF THE SKIN".pdf"ANATOMY AND PHYSIOLOGY OF THE SKIN".pdf
"ANATOMY AND PHYSIOLOGY OF THE SKIN".pdf
 
Painting Rats White Angers Them to No End
Painting Rats White Angers Them to No EndPainting Rats White Angers Them to No End
Painting Rats White Angers Them to No End
 
Health literacies in marginalised communities LILAC 24.pptx
Health literacies in marginalised communities LILAC 24.pptxHealth literacies in marginalised communities LILAC 24.pptx
Health literacies in marginalised communities LILAC 24.pptx
 
Assisted Living Care Residency - PapayaCare
Assisted Living Care Residency - PapayaCareAssisted Living Care Residency - PapayaCare
Assisted Living Care Residency - PapayaCare
 
Hematinics and Erythropoietin- Pharmacology of Hematinics
Hematinics and Erythropoietin- Pharmacology of HematinicsHematinics and Erythropoietin- Pharmacology of Hematinics
Hematinics and Erythropoietin- Pharmacology of Hematinics
 
Pharmacovigilance audits inspections.pptx
Pharmacovigilance audits inspections.pptxPharmacovigilance audits inspections.pptx
Pharmacovigilance audits inspections.pptx
 
Introduction to Evaluation and Skin Benefits
Introduction to Evaluation and Skin BenefitsIntroduction to Evaluation and Skin Benefits
Introduction to Evaluation and Skin Benefits
 
Empathy Is a Stress Response - Choose Compassion instead
Empathy Is a Stress Response - Choose Compassion insteadEmpathy Is a Stress Response - Choose Compassion instead
Empathy Is a Stress Response - Choose Compassion instead
 

High speed cutting/dental lab technology courses by Indian dental academy

  • 1. HIGH SPEED CUTTING INSTRUMENTS IN PROSTHODONTICS Introduction In order to perform the intricate and detailed procedures associated with restorative dentistry, the dentist must have a complete knowledge of the purpose and application of the many instruments required. During each day of his clinical experience the dentist operates on vital tissues within the oral cavity where a millimeter or a fraction there of, is a very significant dimension. A skillful application of sharp hand and rotary instruments requires ability and coordination gained only by extensive training. Before the advent of rotary instruments, removal of tooth tissue was accomplished with sharp – edged chisels, hatchets, and hoes. These hand instruments possessed a cutting capability, which was used for clearing away unsupported and undermined enamel resulting from dental caries. Walls and floors of the cavity were formed by a planning and lateral scraping action of these sharp edged instruments. At best, such efforts were crude, time consuming and often difficult. The first, rotary instruments for cutting tooth tissue were modified hand instruments. These, drill or bur heads could be twisted in the fingers to produce a cutting or abrading action. In 1846 the finger ring was introduced with a drill socket attached for adapting a series of long bundled burs or drills. 1
  • 2. This was the primitive application of the rotary principle. The first drill having flexible cable drive and the first angle hand piece were introduced by Charles Merry between 1858 and 1862. In 1871, Morrison modified and adapted the dental foot engine from the Singer Sewing machine. This was followed by the introduction of the electric dental engine utilizing a cable arm in 1883. In 1910 the endless cord on a jointed arm was made available. The earlier dental hand pieces were capable of speeds from 4500 to 6000 rpm. In 1940 the use of diamond abrasive paints became widespread. The diamond point is compared of a number of small diamond particles bound on a rotary blank. In 1945 Dr. G.V. black, published a report on the non mechanical preparation of cavities and in doing so introduced the air abrasive technique. The impact of Dr. Black’s revolutionary cutting technique on the dental profession was considerable. This was the first significant break in the long established traditional method of cavity preparation. The air abrasive principle utilized particles of aluminium oxide propelled against the tooth surface by a carbon dioxide stream under the pressure of 110 psi, and funneled through a tungsten carbide nozzle with a lumen of 0.018 inch. The penetration of enamel and dentin was rapid but some what difficult to control. 2
  • 3. In 1949 Walsh and Symons published their initial findings relating to the removal of tooth tissue with diamond points at rotational speeds upto 70,000 rpm. This report indicated the use of lighter forces and a resulting increased cutting efficiency at these higher speeds. In early 1950, the ball-bearing hand piece was introduced. In 1963, following the work of Nelson the first fluid turbine type handpiece was introduced. This instrument was capable of rotational speeds of approximately 50,000 rpm and was limited to diamond instruments operated at one speed only. In 1954, air-driven hand pieces were developed. A continuous belt-driven contra-angle which utilized a friction grip chuck and bur was introduced, making possible cutting speeds of upto 150,000 rpm. By 1957, many dentists were using rotational speeds upto 3,00,000 rpm. The introduction of air-bearing hand piece in the early 1950 made possible greater rotational speeds of approximately 5,00,000 rpm. In 1953, an ultrasonic method of tooth tissue removal was also introduced, which used suitably shaped tips vibrating at frequencies ranging from 2,50,000 to 3,00,000 cycles – per seconds. This brief historical back ground reveals that the profession has been searching for a suitable method of tooth tissue removal. Only during last 30 3
  • 4. years, this hunt has slowed down still the profession is trying to refine the procedure and instruments. Review of literature A search through literature reveals various methods used in the past for removal of tooth tissue. The continuous development of newer methods till 1960, indicates that the earlier instruments had some disadvantages. Inspite of the introduction of numerous tooth reduction instruments, and procedures, the principles and the biologic objectives have remained the same. These are as follows. 1. The operator should remove the least amount of tooth tissue consistent with necessary mechanical retention. 2. This should be done with the least barm to the periodontal tissues and the pulp. 3. It should be done with the least discomfort to the patient. 4. No pathologic reactions should be initiated in the pulp. Advantages of high speeds 1. Smaller stones can be used at the increased speeds. 2. Less fatigue results both for the patient and operator. 3. Due to high speed, very light pressure is required. 4. Less vibrations are felt by the patient. 4
  • 5. 5. The chairside time for a given preparation is considerably reduced. 6. Trauma to the pulp is reduced. 7. The efficiency and life of the cutting tools is increased. 8. Because of small tools, control is easy. 9. Removal of old amalgam and gold restorations is easy. Disadvantages of high speeds 1. The increased speed creates increased temperatures in the tooth. Therefore some method of cooling the tooth more efficiently is required not to injure the pulp. This necessitates additional equipment. 2. When a dentist changes from the lower speeds, which utilize a pressure in pounds, to high speeds which need only a pressure in ounces, he must develop a new technique and retrain himself to a new tactile sense. 3. To operate at high speeds good visibility of the cutting instrument is necessary to avoid over cutting. 4. Due to the ease with which tooth tissue is removed, caution must be taken not to injure the proximal enamel of the adjacent healthy tooth and the gingiva. 5. High speeds result in greater wear on the working parts of the handpiece, necessitating more frequent repairs and replacements. 6. Unless used properly, high speeds have a tendency to create striations on a tooth surface. 5
  • 6. 7. The ideal preparation for any type of restoration cannot be accomplished by using high speed equipment alone. The final exactness and finishing line can best be established by instruments revolving at moderate speeds. Types of high speed instruments Hand piece can be divided into four types depending upon their speeds as follows. 1. Low speed – upto 10,000 rpm. 2. Intermediate speed – 25,000 to 45,000 rpm. 3. High speeds – 50,000 to 1,00,000 rpm. 4. Ultra high speeds 1,00,000 rpm and over. Kilpatric has further classified the ultra high speed handpiece into three classes. Type I – the gear driven centre-angle handpiece, upto1,25,000 rpm. Type II – the belt driven contra-angle handpiece upto 2,00,000 rpm. Type III – turbine driven air contra-angle handpiece upto 3,00,000 rpm and higher. Heat generation: Knowledge of the physics tells us that, whenever there is friction between two surfaces, heat is generated, which may bring about rise in 6
  • 7. temperature of either or both the surfaces. The same applies in the tooth reduction procedures. Here the rotating cutting tools come in contact with the tooth surface and the heat is generated. It was not until 1930 that the workers began to investigate the heat rise in the dental pulp. There are many factors that influence the rise in temperature which takes place in cutting operations. The greater the speed of rotation of the cutting tool, the faster the tool revolves, the higher the resultant temperature. It has been found that the temperature rise develops within 10-12 seconds, after the cutting operation is started. Size of the cutting instrument has an important bearing on heat generation, since, its diameter affects the cutting speed at its periphery. Larger the size of the cutting tool more the host generation. A third factor is the pressure applied by the dentist during cutting operation. As the pressure increases, greater will be the rise in temperature. Hudson and associates in 1954 conducted a study on temperature developed in dental cutting instruments from their study they have concluded that, 1. The temperatures produced by dental burs in cutting human dentin ranged from 125°F to 275°F. Since these temperatures are above those, said to be 7
  • 8. tolerated by normal human dentin, it would seem advisable to use some form of coolant. 2. A significant decrease in time required to accomplish a given operation is apparent, when high operating speeds are used. 3. The amounts of heat transferred to the tooth from the bur decreases, at speeds above 12000 rpm, since cutting time at these speeds is reduced and bur temperature remains. Substantially constant and there is less heat trauma to the vital structures. Coolants: From the study of Hudson and Sweeney, it is evident that the temperatures reached during tooth reduction procedures are above those said to be tolerated by normal human dentins. This indicates that, some form of coolant must be used, during the cutting operations, particularly when high speeds are used. Every means should be employed to keep the temperature down as much as possible during cutting operations. Coolants must employed which, to be effective, should be applied at the point of contact between the cutting instrument and the tooth tissue. There are three types of coolants usually employed in dental practice. 8
  • 9. 1. Water. 2. Spray of air and water 3. Air alone. Peyton has shown that at speeds ranging from 30000 rpm to 170000 rpm and with an application of four ounces of pressure, a temperature rise within the tooth of less than 15°C occurred when water or air-water sprays were employed. He also found that even with a water coolant, excessive temperatures developed, when large diameter instruments or excessive pressure were applied with increased operating speeds. This indicates that the use of a coolant, does not eliminate the danger of excessive temperature rise. A reduction in concentration of the amount of water used during cutting procedure shows the significant temperature rise of the dental bur. The minimum volume of water to be applied was estimated at 1.5 ml per minute. The question whether water in spray form should be used at mouth or temperature seems to have no significance as far as temperature rise in the tooth was concerned. Tylman is of the opinion that if the water reservoir is kept at 100°F, it is most comfortable to the patient, less liable to be harmful to the pulp and still reduces the heat of friction during cutting. 9
  • 10. There are certain other problems associated with the use of the highspeed cutting tools. Most of the hand pieces are so designed that a spray or stream of water is directed from the head of the handpiece directly onto the cutting operation. Where the water strikes the tooth and the cutting tool directly, full benefit is obtained from the coolant. Where however, the abrasive on the cutting tool, is on the surface away from the stream of water, water does not flood the tooth surface being cut, resulting in excessive temperature rise. The overcome this difficulty perforated disks have been developed, which permit the water to go through the openings and lubricate the disk and tooth on the cutting side. The use of perforated disk results in less temperature rise. Consequently when disks are non perforated, and when the stream of water cannot be directed to the cutting contact areas, they should be used at speeds not exceeding 10,000 rpm. Another advantage of a water coolant lies in the fact that the tooth debris from the cutting is removed rapidly, preventing the clogging of the cutting tools. This results in greater cutting efficiency of the stone. Also, it prolongs the life and effectiveness of the instrument. It is essential that the water be in intimate contact with the revolving instrument and the tooth tissue. To do this more effectively, Nelson recommended the addition of a wetting agent to the water spray. 10
  • 11. Because the high speed technique requires a larger quantity of water as a coolant, there is the problem of removing this water from the mouth. To have the dentist stop frequently to allow the patient to spit out the excess water is time consuming. The customary saliva ejector has insufficient removal capacity. To solve this problem, Thompson has suggested a washed field technique. This technique adapts the suction or vacuum principle. It established and maintains a powerful but gentle negative pressure of air in the mouth, close to the field of operation. Accompanying the air stream, is a flow of isothermal water which is projected copiously onto the operative field. This water is entrained into the vacuum air stream, which draws it rapidly across the operative area. The irrigant pulls away with it tooth cuttings and debris. These are taken into the vacuum air stream and disposed off in a filter system. A clean, clearly visible operative field is provided. This technique has the distinct advantages that it facilitates the use of high speed instruments, maintains visibility during copious irrigation of the operative field, reduces operating time, improves the patient’s well being and introduces a new concept of cleanliness. Human 11
  • 12. tissues are maintained in their natured wet safe pain, trauma and postoperative complication, which may arise due to ingestion of tooth debris are reduced. Desiccation of hard and soft tissues is avoided. Heat is eliminated thus preserved the tissues. Vibration: Cutting a tooth may be very annoying and unpleasant to the patient but still not be painful. In pain there is usually an involvement of the nerve endings, either by trauma or extreme irritation, resulting in an acute, painful reaction. Most patients associate the sensation of vibration, noise, pressure and the slight increase in cutting temperature with the sensation of pain. Consequently, if the factors of vibration, heat and pressure are reduced to a minimum, the patient usually experience reduced or no pain. One mechanical factor that influences vibration is the dental handpiece, whether it is friction-bearing, ball bearing, high speed belt driven or turbine ultra speed. When the friction bearing, conventional type of handpiece is used at a speed of 4500 rpm to 6000 rpm, it is connected by the conventional belt and pulley system of the dental engine. In this case one may expect a high order of vibration depending upon the condition of various mechanical parts, their adjustment and speeds of their operation. 12
  • 13. Pulleys that are worn, a worn belt, or an improperly adjusted belt will cause vibrations that are transmitted down to the cutting tool. Similarly hand piece which do not hold the cutting tool properly, which have worm bearing or are out of adjustment will also cause vibration. The investigations of Walsh and Symmoss showed that vibration, when applied to tooth, produced the most unfavorable response when the frequency was between 100 cps and 200 cps. When the frequencies were above 1000 cps, they were generally beyond the upper threshold of perception of the average patient. It is the lower frequencies, in the range of 100 – 200 cps, that are usually developed at the lower speeds, especially if the equipment is worm and maladjusted. Hudson and Sweeney have reported the importance of having centricity in the cutting tool. They found that eccentric burs when rotated at 6000 to 10000 rpm produced a lower frequency in the range of 100-200 cps, whereas a true running bur at 10000 rpm produced vibrations in the frequency range above the upper threshold. Tamner pointed out that only a part of an eccentric cutting tool is used as it rotates, thus causing unfavorable impacts and vibrations, which fall into the most annoying frequency range. The disks and stones that are unmounted and are screwed onto a mandrel very frequently are eccentric and therefore 13
  • 14. should not be used in high speed cutting operations. The permanently mounted instruments are indicated in preference to unmounted type. Poorly built burs with blades not evenly cut or chipped will likewise cause vibration. In using carbide burs, it is very important the see that none have chipped blades. Correct adjustment of the belt is important in the reduction on and elimination of vibration. A belt that is too loose increases the vibration pattern transmitted directly to the tooth. In the ultra highspeed hand pieces the metal chuck holding the cutting instrument often is replaced by a rubber or plastic chuck. This lessens the vibration transmitted to the cutting instrument and facilitates the more rapid cutting action. In cutting with a water turbine handpiece at 45,000 rpm the intensity of vibrations was well tolerates by the patient. Morrison and Grinnel made the following observations. The deleterious effects of vibration are two fold in origin. 1. Amplitude. 2. Undesirable modulating frequencies. 14
  • 15. If we minimize or eliminate these factors, we can then reduce the undesirable effects of vibration. Amplitude: The wave of vibration consists of frequency and amplitude. At conventional speeds, amplitude is greater but frequency is less. At higher speeds the reverse is true. The greatest harm is caused by the amplitude of vibration which is the factor, most destructive of instruments and which causes the most apprehension in the patient and the greatest fatigue in the dentist. By increasing operating speeds, the amplitude and its effects are reduced and a more satisfactory result is attained. Vibration waves are measured in cycles per second. It has been shown that rotation of approximately 6000 rpm sets up a vibrational wave of approximately 100 cps. As the rpm is increased the cps of the fundamental vibration wave are increased until, at ranges of 100000 rpm, we have a vibration wave of 1600 cps. It has been demonstrated that at wave of vibration of over 1300 cps, vibration is practically imperceptible to the patient. The reason for this is not fully understood, but there are two theories for this phenomenon. 15
  • 16. (1) The Wedensky inhibition phenomenon – frequency increased to a point where vibratory perception diminishes due to failure to perceive vibration. This is because Stimulation occurs during Refractory Period of Recovery. (2) Vibratory perception depends upon the product of the amount of stimulation (i.e. pressure) multiplied by the frequency of stimulation necessary for a reaction. This is called Chronaxie. As the speeds above 1,00,000 rpm, due to light pressure and high speeds, chronaxie is attained, which is necessary for reaction. Thus it can be concluded that, the more the rpm, the less the amplitude, and the greater the frequency. Vibratory perception will be lost in the ultra highspeed range of 1,00,000 rpm or more. Spread of pathogenic organisms by Ultra speed cutting procedures: Atmospheric contamination through the spread of oral organisms particularly from air turbine action has been a concern of the dental profession for some time. Dental procedures tend to expose the operator to infectious diseases. Recent studies suggest that the extent of aerosol produced by air turbine may increase the normal hazard. A report involving patients with pulmonary tuberculosis cultures were demonstrated on all petri dishes exposed during cutting procedures, with the heaviest concentration being at 2 feet in distances from the patients mouth. This indicates that the dentist and his 16
  • 17. assistant are exposed to a serious health hazard when operating with an ultra speed exposed instrument on patients having such pathogens in their oral flora. When a patient’s history suggests the existence of tuberculosis, pneumonia, influenza, infections hepatitis or any infectious diseases including the common cold, a protective face mask should be worn by both dentist and assistant. During all ultrahigh speed cutting procedures, protective eye-glasses should be worm routinely. SUCK-BACK PHENOMENON- The operation of the turbine is switched off by closing the compressed-air valve abruptly. Then, owing to its own kinetic energy, the turbine continues its rotation, so that the turbine starts operating as an air pump. This causes a negative pressure in the area of the turbine shaft. The negative pressure sucks air from the environment that can be contaminated by aerosols of saliva and blood of the patient. Size of cutting instrument and cutting speeds: It has been pointed out by Peyton, and Nelson that, the important factor of increased operating speeds is the instrument surface speed in fact per minute rather than the revolution per minute of the instrument. The larger the diameter of the cutting instrument, the slower the speed required at the spindle. The specific phase in preparation of an abutment should determine the size of the cutting instrument and the rpm that should be 17
  • 18. used. Employing superspeed for all operations places unnecessary strain upon the patient and equipment. If the same effect can be accomplished by using a larger instrument at a lower speed, but still remaining above the threshold of perception, this should be done. However, oversized cutting tools should not be used at super speeds due to the difficulty of instrument control and accuracy of cutting. VIBRATION SYNDROME : the perception of vibration, pain, touch and temperature deteriorates. The negative effect of local vibrations occurs within the range 5-1400 hz, the most harmful being those below 16 hz. mechanical vibrations arise because the various machines operating at the dentist’s workplace contain moving parts. The main source is vibrating power-driven or air-driven instruments, such as low- and high-speed handpieces as well as ultrasonic instruments. The vibrations emitted by these machines travel directly from the handles to the operator’s hand. These are local vibrations. Biologic response of dentin and pulp to high speed cutting: Dentin: As the contents of the dentinal tubules are in direct continuity with the odontoblasts, and pulp, cutting or grinding the dentin causes a reaction in the pulp and this may lead to changes in the dentin. 18
  • 19. An early experimental investigation into the effect of cavity preparation on the dentin and pulp was carried out by Fish in 1932. He cut cavities in the teeth of dogs and left the cavities open to the saliva. By sealing dyes into the pulp chambers of the treated teeth he has shown that one of two reactions is produced in the dentin. In some cases there was sclerosis of the cut dentinal tubules which forms a protective some sealing off the pulp from the injury and underneath this region, there is a further growth of tubular dentine. These reactions are produced by the stimulation of the odontoblasts. The other reaction that resulted was the formation of dead tracts. With this lesion some or all of the odontoblasts, that are in connection with the cut tubules die. On the pulpal aspect of these tubules, hyaline mineralized barrier, secondary dentine is laid down, thereby sealing the lesion from the pulp. Pulp: The changes in the pulp have been studied by Langeland and Morslard and Shovelton. They state that the damage to the pulp is to a large extent due to the heat generated. They have shown that when precautions are taken to minimize heat production by using burs rotated slowly in a speed reducing handpieces, the only evidence of pulp damage was a slight reduction of the odontoblast layer with the displacement of a small number of odontoblasts into the dentinal tubules. When speeds upto 5,000 rpm were employed, there was more extensive displacement of odontoblasts associated with marked 19
  • 20. vacuolization of the odontoblast layer, and local hemorrhages may be seen in the pulp. As the speed was increased, the changes became more severe. When tooth reduction was done under a stream or spray of water, the damage to the pulp was markedly reduced. Pulp changes associated with tooth reduction using the air abrasive technique have been studied by Kennedy and using ultrasonic technique by Mitchell and Jenson. The changes in both the cases are similar to those produced at the speeds of 5,000 rpm. The effects on the pulp of using high speed rotary instruments such as the air turbine have been investigated by Marsland and Shovelton. The changes found are no severe than those produced at lower operating speeds provided that adequate cooling of the cutting instrument by water jet or air/water spray is ensured. Alterations in the hard tissues of tooth cut by air turbines have been observed. The enamel over a wide area of crown may show minute cracks and the dentin shows altered staining reactions as a result a local overheating. RECENT DEVELOPMENT: ‘ANTI-SUCK BACK’- Planmeca compact dental units, the turbine drive air is not shut off abruptly but controlled down by allowing the driving air 20
  • 21. to decrease gradually. The software of the dental unit will keep on supplying the drive air into the turbine according to carefully chosen parameters. This way there is no possibility for the build-up of a vacuum effect that would cause suck-back. Ceramic bearings- no need of lubrication and more resistant to autoclave sterlization. Use of quartz rods instead of fibre-optic. Easy-to-use push-button bur releases. Swivel systems. Titanium handpieces. Smaller head size.  ELECTRICAL HIGH SPEED HANDPIECES advantages are: More power and torque than air turbine handpieces. Better bur concentricity. Less vibration and noise. Broad, controllable speed ranges. 21
  • 22. Forward/reverse option are available. With appropriate attachments, one system can be used for restorative, prosthodontics, prophies and endodontics. Disadvantages are : Heavier than air turbine. More expensive. Learning curve may be required. Attachment heads not as small as the small-head air turbines. May not be able to fully replace the air turbine. Infection control concerns. Discussion It is for more than 125 years, that rotary instruments have been in use, for tooth reduction operations, in different forms, from a hand rotary instrument to ultra sonic instruments, which have the rotational speeds ranging from very low speeds in case of band rotary instruments to 5,00,000 rpm in case of air bearing hand piece. These remarkable advances in the instruments have greatly reduced fatigue in the operator because of the physical case of manipulation and have considerably increased the comfort to the patient by 22
  • 23. reducing the actual working time and pressures required for tooth reduction, thereby minimizing the factors of heat generation and pain. Though high speed techniques have been a born to the dental profession, they have their can limitations. It is interesting to not that, in spite of considerable improvements in tooth reduction procedures and the instruments used for the same, the principles and biologic objectives have not changed. These improved methods of tooth tissue removal have a potential to damage the healthy teeth and surrounding structures, if they are used without proper understanding of their working and if they are used without taking proper care. Improper handling of these modern equipment may also be different to the longevity and working capacity of the instruments themselves. For successful and efficient use of those cutting tools, certain factors should be given consideration. Heat that is generated, while the tooth tissue is being removed must be kept, down to the minimum and at the sametime, whatever heat is generated, must be eliminated as efficiently and as quickly as possible by employing coolants, in any one of three forms commonly used i.e. water, air/water spray or air alone. Simultaneously with the coolant, if water or air/water spray is used, an efficient mechanism for remove of the water from the oral cavity must be employed. Otherwise, the clinical procedure is delayed, if the patient has to spit out the water, every now and then. By eliminating the water evacuation equipment, we are losing one of the advantages of these high 23
  • 24. speed instruments i.e. reduced working time for a particular preparation. Use of efficient coolants, not only eliminate the heat generated, but at the same time, keeps the operating area clean and free of any debris. High speed cutting methods have a further advantages in that, they reduce the annoyance that may be caused to the patient, when low speeds are used with the modern high speed cutting devices, the vibration produced is of a frequency that is generally beyond the upper threshold of perception of the average patient. Pressures that have to be employed in the use of high speeds are considerably reduced, in comparison with those needed for low speeds. Thus, when the factors of pressure, temperature and vibration are kept within the tolerance limits, the patient comfort is certainly improved. Size of the cutting tool to be used for particular tooth reduction procedure is an important consideration, particularly while using high speeds. Oversized cutting tools should be avoided, as they are difficult to control and at the same time, the accuracy of tooth preparation on procedure is also adversely affected. Biologic reactions of the tooth tissues, particularly dentin and pulp, should not be over locked, when high speeds are employed for tooth reduction 24
  • 25. operation. These responses have been studied by a number of people and they have shown that, the response are not significantly different from those, when low speeds are used, provided, effective coolants are employed. Thus it can be concluded that, high speed equipments for tooth reduction if used with proper understanding and due care, provide definite advantages over the conventional low speed cutting procedures. This fact places the high speed devices at definitely a higher level as against their low speed counterparts. Conclusions 1. High speed cutting devices, if used with a thorough understanding of their mechanism and due care to the biologic integrity of teeth and surrounding structure, are a boon to dentistry. 2. In the process of tooth reduction using high speeds considerable amount of heat is generated and an effective coolant is a must for preservation of tooth integrity and patient comfort. 3. Vibration is increased with the increase in speed, but it is beyond the threshold of prerception of the normal human beings and hence not harmful. 25
  • 26. 4. Biologic reactions of the dentin and pulp, to high speed cutting, cannot be overlooked. Summary A brief history of rotary instruments has been presented. A critical evaluation of the high speed cutting devices, as to their advantages, disadvantages, and precautions to be taken during their use, has been assessed at length. Biologic reactions of dentin and pulp, to high speed cutting have been discussed in brief. 26
  • 27. Contents I. Introduction II. Review of Literature a. Advantages of high speeds b. Disadvantages of high speeds c. Types of high speed instruments d. Heat generation e. Coolants f. Vibration g. Spread of pathogenic organisms h. Size of cutting instruments and cutting speeds i. Biologic responses of dentin and pulp to high speed cutting I. Discussion II. Conclusion III. Summary 27