2. Background
• While goal-directed therapy in severe Sepsis/septic shock has
shown to improve outcome, CO and DO2 is only part of the goal
directed therapy, there are many others. Rivers et al NEJM 2001;
345.
• PACs are regarded amongst many as the GOLD standard for
measuring CO/LVEDP and even whilst using our GOLD standard
we have no conclusive data for improvement in patient outcome.
Shah et al, JAMA 2005 (5051 patients) and PAC-man study, Harvey
et al, Lancet 2005. Given this, one does question the benefit of any
CO monitoring, no matter how close the correlation to our GOLD
standard.
• Infact Edwards Lifesciences at the beginning of all their
sales/presentation always have a disclaimer ‘There is NO data to
suggest that care is improved or outcomes differ with the utilization
of monitors that assess cardiac function’.
3. Techniques for monitoring CO
• Invasive
• PA catheter (Dye, thermodilution intermittent/continuous)
• Fick Principle
• Low-invasive
• TOE
• Oesophageal doppler
• Arterial pulse contour analysis (PiCCO, LiDCO, Computer algorithm)
• True Non-Invasive
• TTE
• Transthoracic bio-impedence
• Clinical Examination!!
4. Pulmonary Artery Catheter
Quick summary
PAC is designed to measure:
• Intra-cardiac pressures
RAP/RVP/PAP/PAOP
• CO/SV and indirectly many
SVR/LVSW etc
• True SvO2
Contra-indications:
• Tricuspid or pulmonary valve
mechanical prosthesis
• Right heart mass (thrombus
and/or tumor)
• Tricuspid or pulmonary valve
endocarditis
• Many relative contraindications
reducing accuracy
6. PA Catheter Complications
• Infection • Thrombocytopenia
• Air emboli • Catheter knotting
• Thrombosis • Ventricular
Dysrhythmias
• PA infarction
• Hemothorax
• PA rupture
• Pneumothorax
• Balloon rupture
• Cardiac Tamponade
7. Fick Principle
• Fick Principle: measure volume displacement
• First proposed in 1870
• “The total uptake or release of a substance by an
organ is the product of the blood flow through that
organ and the arteriovenous concentration
difference of the substance.”
• CO = O2 consumption (ml/min) art –
mixed venous O2 conc. (ml/l)
• Limited by cumbersome equipment, sampling errors,
need for invasive monitoring and steady-state
haemodynamic and metabolic conditions
9. TOE
• CO can be obtained via either:
• SV= CSA x VTI, VTI= Vmean x t
(using PW doppler flow through either AV/PV/LVOT)
• SV= EDV – ESV
• CO measurements using CW doppler have been validated as a
clinical alternative to thermodilution.
• Useful tool for regional as well as overall cardiac function. RV and
LV discrepancies. Visual tool as well as quantative.
• Recent data to suggest management benefit in general ICU setting,
L.E. Orme et al, Br. J. Anaesth., March 2009; 102: 340 – 344 .
• Operator dependent; time consuming; no beat to beat reactivity;
accuracy in tachycardias; training required; cost of equipment.
10. Oesophageal Doppler
• Ultrasonic doppler –Continuous or pulse wave doppler
• Placed parallel to direction of blood flow
• Velocity of blood measured by change in frequency of of the
reflected ultrasound wave
• 4 MHz continuous or 5 MHz pulsed wave
• Transoesphageal doppler
• Measures descending thoracic blood flow
• User dependent, requires anaesthetised pt currently
• Makes certain anatomical and mathematical assumptions
• Suprasternal doppler
• non-invasively measures asc. Aorta flow
• Limited by probe position, aortic valve abnormalites & median
sternotomy
11. Oesophageal Doppler principles
• Principle of stroke volume
calculation from aortic
velocity (VAo)
measurements.
• The area under the
maximum aortic velocity
envelope (VTI) represents
the stroke distance
14. Oesophageal Doppler
• Disadvantages
• Interference by NG tube / diathermy
• Dislodged by movement
• Contraindicated in oesophageal surgery
• Advantages
• Min invasive
• Real time measurement
• Rapid insertion
• Minimal skill level
• Good trend monitor
15. Pulse contour analysis
• 3 main types:
• Transpulmonary thermodilution + pulse contour
analysis (PiCCO)
• Lithium dilution (LiDCO)
• Pure Pulse contour analysis using computer algorithm
(Flowtrac/Vigileo)
16. PiCCO
• Pulse contour analysis with intermittent thermodilution
measurement.
• Enables continuous hemodynamic monitoring using:
– femoral or brachial artery catheter
– central venous catheter
• Adult or pediatric patients who have or may develop
pulmonary edema or ARDS are likely candidates
18. PiCCO
• Advantages
• CCO
• Continuous Volume responsiveness
• Short response time 12 secs
• Paediatric application >2kg
• Disadvantages
• Needs CVP and proximal arterial line
• Needs calibration
19. LiDCO
• Minimally invasive
• Safe, non-toxic rapidly redistributed and no 1st pass
effect
• LiCl: 0.002mmol/l injected into central vein (peripheral
administration possible as well)
• Arterial plasma conc. measured by withdrawing blood
across lithium selective electrode at 4ml/min
• CO calculated from Li dose and area under primary
concentration-time curve before re-circulation
21. Advantages
• SAFE
• Central/peripheral venous and arterial catheters
• Injectate is an isotonic (150 mM) solution of lithium
chloride
• 0.15 -0.30 mmol for an average adult
• patient weight (>40kg) and absence of renal dysfunction or
dialysis
• ACCURATE
• SIMPLE TO USE
22. Disadvantages
• NMBA interfer with calibration
• AF
• iABP
• People on lithium therapy
• Renal impairment
• Weight <40kg
• Pregnancy 1st trimester
• Intra cardiac shunts
23. Flowtrac/Vigilleo
• Continuously computes stroke volume from arterial pressure signal
• Requires NO manual calibration
• Demographic Data
• Arterial waveform analysis
• Quantification of SVV
• Aortic pulse pressure is proportional to SV and is inversely related
to aortic compliance.
• If compliance (and resistance) is constant a bigger SV will mean a greater PP.
• Disposable Transducer
• Latex-Free
• c.£120 per set up
24. Trending Stroke Volume
Systolic
press.
PP ∝ SV
Diastolic
press.
• Arterial pressure is sampled at 100 Hz
• Changes in stroke volume will result in corresponding changes in the
pulse pressure
• A robust “whole waveform” measure of the pulse pressure is
achieved by taking the standard deviation of the sampled points of
each beat
• sd(AP) ∝ Pulse Pressure ∝ Stroke Volume
• SV estimates are calculated every 20 sec
25. The effect of compliance on PP:
Age, gender and BSA factors
• Younger • Older
vs.
• Male vs • Female
• Higher BSA
.
vs. • Lower BSA
For the same
volume ➔
• Compliance inversely affects PP
• The algorithm compensates for the
effects of compliance on PP based on age,
gender and BSA
26. Effect of vascular tone
• The algorithm looks for characteristic changes in the arterial
pressure waveform that reflect changes in tone (i.e., MAP,
Skewness, Kurtosis)
• Those changes are included in the continuous calculation.
MAP Skewness Kurtosis
27. Error Sources
ICO CCO FloTrac
sensor
Comp constant ------ Patient data
Age
Height & weight
Gender
Clinician technique ------ AS/AI
Injectate Volume
Dysrhythmia
Injectate Temperature
Injection Timing
Catheter migration Catheter migration Sensor height
Ventilator timing Sequential compression Aortic balloon pump
device
Patient temperature shifts Patient temperature shifts Patient arm movement
Infusions & drips Infusions & drips Line bubbles
catheter whip (fem)
Valve regurgitation Valve regurgitation Pressure dampening
(extreme vasopressors)
28. Validation?
• ‘Cardiac Output Determination From the Arterial Pressure
Wave: Clinical Testing of a Novel Algorithm That Does Not
Require Calibration’ Journal of Cardiothorac and Vasc Anesth
2007.
• ‘Uncalibrated pulse contour-derived stoke volume variation
predict fluid responsiveness in mechanically ventilated patients
undergoing liver transplantation’ BJA 2008.
• Still awaiting rigorous validation studies.
30. TTE
• Non invasive
• Similar method of calculation of CO to TOE
• In addition to the limitation of TOE, obtaining
echocardiographic windows in mechanically
ventilated patients can be difficult.
31. Transthoracic bio-impedence (Impedence
Cardiography/ Impedance plethysmograpghy)
• 4 dual sensors with 8 lead wires
placed on neck and chest
• Current transmitted and seeks
path of least resistance: blood
filled aorta
• Baseline impedance (resistance)
to signal is measured
• With each heartbeat, blood
volume and velocity in the aorta
change
• Corresponding change in
impedance is measured
• Baseline and changes in
impedance are used to measure
and calculate hemodynamic
parameters
32. Parameters
• Measured • Calculated
– Stroke Volume / Index
– Heart Rate
– Cardiac Output / Index
– Non Invasive BP – SVR / SVRI
– Velocity Index – LCW / LCWI
– Acceleration Index – Systolic Time Ratio
– Heather Index
– Thoracic Fluid Content – Ejection Time Ratio
– Pre-Ejection Time
– LV Ejection Time
34. Validation?
• Ziegler D, et al. “Comparison of cardiac output
measurements by TEB vs. intermittent bolus
thermodilution in mechanically ventilated
patients” Chest. 1999; Vol. 16, No. 4 (suppl. 2):281S.
• Scott Sageman, et al. “Equivalence of
Bioimpedance and Thermodilution in Measuring
Cardiac Index After Cardiac Surgery” J
Cardiothoracic & Vasc. Anesth. 16:8-14, 2002
35. Conclusions
• Correctly powered and vigorous validation studies still awaited
for ICG and Flotrac/vigileo.
• PAC/LiDCO/PiCCO/Oesophageal doppler/TOE all validated
for select patient groups. Knowing limitations of each
monitoring system vital when deciding which to employ.
• Absolute numbers vs trends.
• No outcome improvement data with use of CO monitoring.
• Clinical examination.
• The information obtained is only as useful as the person
interpreting it and using it to modify patient Rx.
High PEEP,Tamponade, Pneumothorax, reduced myocardial compliance, Mitral stenosis, Pulmonary occlusive disease, severe MR
Arterial pressure is sampled at 100 times per second (100 Hz). Each sample is a pressure data point measured in mm Hg. This data is analyzed and updated every 20 seconds, utilizing the 2000 data points (20 sec x 100 Hz) collected. The standard deviation (sd(AP)) of these data points is proportional to the pulse pressure, which is proportional to the stroke volume. This method is robust in its assessment of pulse pressure because it looks at the whole waveform. sd(AP) = a measure of variance The Stroke Volume is averaged and displayed every 20 seconds. The user has the capability to choose between a 20 second calculation or 5 minute moving average option.
There is a direct relationship (i.e., the shape of the trend shown above) between arterial pressure and large vessel compliance as it relates to a human’s age or gender. That is, a male will typically have a more compliant aorta than a female of the same age, and a younger person will have a more compliant aorta than an elderly person. This relationship was quantified and mathematically modeled by Langewouters. Through development of the algorithm, it has also been found that there is a relationship between aortic compliance and BSA. For example, a larger person (higher BSA) will typically have more compliant vessels than a smaller person (lower BSA). As part of the assessment for vascular tone, estimates of aortic compliance based on the above principles are important but not all encompassing. Therefore, in addition to Langewouter’s mathematical model, further waveform analysis is conducted by the algorithm to take into account patient specific, real time effects of vascular tone on the waveform.
The aortic compliance function (i.e., age, gender, BSA function) within X helps to provide a “ballpark estimate” of the patient’s likely vascular tone. Further analysis of the waveform shape is also a significant factor to providing vascular tone estimates essential to the SV calculation. This method allows for reliable calculation of key flow parameters without manual calibration. X (pronounced “khi”) = a symbol for dynamic polynomial functions, a function that continuously adjusts to multiple changing variables Depending upon the state of the patient’s vasculature, the arterial waveform will take a shape that can be characterized mathematically. The statistical tools used, in addition to the aforementioned standard deviation, are the mean, skewness and kurtosis. Mean , clinically known as MAP, can provide an indication of the increase or decrease in resistance. Skewness , or symmetry of the data, is also often associated with the stiffness of the patient’s vasculature. Kurtosis , a measure of how peaked or flat the data distribution is, provides an indication of the nature of vasculature as well. For example, larger vessels will typically have a “flat” distribution as compared to peripheral vasculature. Skewness : The angle or slope exhibited on the rise of the waveform. Kurtosis : How flat and wide the waveform is If the shape changes, the mathematical calculation will change, providing a patient-specific, real time assessment of shifts in vascular tone. The greater the magnitude increase in tone, as calculated as a decrease in X, the lesser the weight pulse pressure estimates (i.e., sd(AP)) will have in the SV calculation. Furthermore, the effect of differences in arterial sites on the arterial waveform are neutralized within the X function. Therefore, FloTrac can be used with any existing peripheral arterial catheter. This is possible through the analysis of arterial waveform characteristics that are specific to different points within the vascular tree. Two key variables in this portion of the function are the standard deviation (pulsatility) and the kurtosis (peaked/flat nature) of arterial pressure.
Additional sources of error are noted and should be taken into consideration when comparing these two very different technologies.