New Technology:
核磁共振相容節律器
(MRI Conditional Pacemaker System)
林圀宏
心臟內科
中國醫藥大學附設醫院
2013 心臟電生理相關專業人員教育課程
輻射線 警告標誌
Why MRI Scans are Important to Clinical Medicine
• Fastest Growing Standard of Care in Diagnostic Imaging1
• Superior Soft Tissue Imaging2
 Primary method to evaluate:
 Central Nervous System
 Musculoskeletal System
 Oncological Conditions
 Some Cardiovascular Disorders
• MRI complements CT (which excels when imaging bony structures)
• No radiation risk to patient or healthcare provider
• Since the absence of x-ray radiation, MRI is optimal for follow-up of chronic
diseases that require repeat imaging and for diagnostic imaging in young patients
and women of childbearing age.
1. Kaiser CP. Soaring MRI use draws scrutiny. Diagnostic Imaging Online January 4, 2002. CMP United Business Media: A CMP Healthcare Media Web
Site. Available at: http://www.diagnosticimaging.com/dinews/2002010401.shtml. Accessed October 19, 2004.
2. Duru F, Luechinger R, Scheidegger MB, et al. Pacing in magnetic resonance imaging environment: Clinical and technical considerations on
compatibility. Eur Heart J. January 2001;22(2):113-124.
Cervical spine computed tomography (CT) vs MRI in a patient with neck pain and fever.
Nazarian S et al. Circ Arrhythm Electrophysiol 2013;6:419-
428
Copyright © American Heart Association
Brain computed tomography (CT) vs MRI in a patient with weakness.
Nazarian S et al. Circ Arrhythm Electrophysiol 2013;6:419-
428
Copyright © American Heart Association
Cardiac computed tomography (CT) vs MRI in a patient with facial swelling.
Nazarian S et al. Circ Arrhythm Electrophysiol 2013;6:419-
428
Copyright © American Heart Association
Complication while having MRI
核磁共振 (MRI) 警告標誌
10
Anatomy of an MRI Scanner
Three basic components:
 Static magnet
 Gradient magnets
 RF coil
10
Static
Determines magnetic strength
of scanner:
 1.5 or 3.0 tesla
 1 tesla (T) = 10,000
gauss.
 The earth’s magnetic
field = 0.5 gauss.
 Magnet response in
SJM pacemakers occurs
in response to 8-10
gauss directly over the
pacemaker.
Force
• The most intuitive potential interaction of implanted devices
with an external magnetic field is the possibility for
movement and dislocation of the device because of magnetic
force.
• Current lead designs contain little or no ferromagnetic
components and are not likely to experience force and
torque.
• The potential for movement of a pacemaker or ICD
generator in the MRI environment depends on the
 magnetic field strength
 ferromagnetic properties of the device components
 the implant distance from the magnet bore and the
stability of the implant.
Current induction
• The RF and pulsed gradient magnetic fields of the MRI
scanner may induce electric currents in leads within the field,
if the lead is part of a current loop that is completed through
the body.
• The ratio of lead length versus RF wavelength and lead
conformations, such as loops, are strongly associated with the
extent of current induction.
Gradient Waveform
Amplitude
0.4 ms
500 ms – 750 ms
Pacing Pulse
Time
~ 0.005 ms
Gradient Induced Pulse
Gradient Induced Pulse
Gradient Mechanism
Stimulation hazard gradient-induced high rate pacing
Start of Scan
EKG
Pulse Ox
Canine Test
Unintended Stimulation Clinical Impact
The MRI scanner is pacing the heart
Heating and tissue damage
• Metallic devices and leads can act as an antenna thus
amplifying local radiofrequency energy deposition.
• Fractured leads or lead loop configurations may increase
the potential for heating.
• Epicardial leads that are not cooled by blood flow and
abandoned leads may also be prone to increased heating.
Heating of cardiac tissue
Heating at the lead tip can result in threshold changes.
Malfunction of devices
• CIED may provide unnecessary therapies or fail to provide
necessary therapies when placed in the MRI scanner.
• Pacemakers and ICDs have the potential for receiving
electromagnetic interference (EMI) in the MRI environment,
resulting in:
 radiofrequency noise tracking
 asynchronous pacing
 inhibition of demand pacing
 delivery of ICD therapies
 programming changes
 loss of function.
Figure 2. EMI noise from different MR scan protocols interpreted by the device as ventricular
fibrillation (VF).
Roguin A et al. Circulation 2004;110:475-482
Simultaneous electrocardiography and pulse oximetry while in 3T
bore of magnetic resonance imaging followed by application of
gradient scan leading to inhibition of effective pacing.
Gimbel J R Europace 2009;11:1241-1242
Malfunction of devices
• The static magnetic field of the MRI scanner can also
alter device function by inducing unexpected reed switch
opening or closure.
• In addition, temporary programming changes made to
avoid device interaction with the MRI scanner (such as
disabling of tachycardia therapies) may lead to
catastrophic results if a spontaneous arrhythmia occurs
and is not recognized.
Simultaneous recording of ECG and pulse oximetry during MRI of the brain in a patient with a
Thera pacemaker reprogrammed to D00 80 bpm.
Sommer T et al. Circulation 2006;114:1285-1292
Copyright © American Heart Association
During and/or after the MRI
• no inhibition of pacemaker output or cardiac
arrest
• no sustained ventricular arrhythmias
• no unexpected changes of heart rate
• no electrical resets
• no pacemaker system disturbances and
• no sensation of torque or pain
Variables Affecting Magnitude of Risks
 Length/position of pacing leads
 Patient and device position within machine
 Patient factors / medical history
 MRI scan duration
 Blood flow at lead/tissue interface
 Strength of RF field
 Target anatomy of scan
 Type of imaging MRI sequence
 Pacemaker and Lead Design
MR Conditional Labeling*
MR Safe
MR Conditional
MR Unsafe
An item that has been
demonstrated to pose no known
hazards in a specified MRI
environment with specified
conditions of use.
Device Design Solutions
• Minimize ferromagnetic
content
• Hybrid-case connection
• Hall sensor
• Optimize input circuitry
Solutions to Reduce MRI Risk
• Filter board added to pacemaker hardware
• When the filter detects MRI RF signals, it will not allow the
signals to be sent back down the lead (rectification), which
would result in cardiac stimulation.
.
28
X-ray marker
MRI filter assembly
Solutions to Reduce MRI Risk
• 2 filters are incorporated into the pacing lead to reduce the
risk of heating from RF signals.
PATIENT
MANAGEMENT
Patient Care Pathway
1. Pre-Screening and Scheduling
2. Pre-scan
3. Scan
4. Post-Scan
Pre-screening
• Ways to ID pacemaker system:
1. Patient ID Card
2. X-Ray
3. Call cardiology
Patient ID Card
X-Ray
Radiopaque identifies the system via unique
radiopaque visible under x-ray
1 Location of the device radiopaque symbol
2 Device radiopaque MRI symbol
3 Lead radiopaque MRI symbol
Radiopaque identification available:
1. pulse generator
2. lead
36
MRI Pacemaker Identification
MRI Pacemaker Identification (Con’t)
• To verify that a patient has a MRI pacemaker, you may use X-ray
identification, confirming that the pacemaker and the lead are appropriate.
Radiopaque markers will show if the patient has an appropriate system for an
MRI scan
37
St. Jude Medical
Identifier
Device MRI
Symbol
To program MRI SureScan to On, go to Params
-> Additional Features.
When programmed to On, MRI SureScan operation disables
• Arrhythmia detection,
• Magnet mode,
• All user-defined diagnostics.
On
• For patients who require pacing support, program the device to an
asynchronous pacing mode (DOO, AOO, VOO).
• For patients who do not require pacing support, program the device to the
non-pacing mode (ODO).
Patients and their implanted systems must be screened to meet the
following requirements:
•No previously implanted (active or abandoned) medical devices,
leads, lead extenders, or lead adaptors.
•No broken leads or leads with intermittent electrical contact as
confirmed by lead impedance history.
•A SureScan pacing system that has been implanted for a minimum of
6 weeks.
•A SureScan pacing system implanted in the left or right pectoral
region.
•Pace polarity parameters set to Bipolar for programming MRI
SureScan to On (for Advisa DR MRI).
•Pacing capture thresholds of ≤ 2.0 volts (V) at a pulse width of 0.4
milliseconds (ms).
•A lead impedance value of ≥ 200 ohms (Ω) and ≤ 1500 Ω.
•No diaphragmatic stimulation at a pacing output of 5.0 V and at a
pulse width of 1.0 ms in patients whose device will be
programmed to an asynchronous pacing mode when MRI
SureScan is on.
Cardiology requirements:
Radiology requirements
-45-
Competitive Comparison
SJM Accent MRI™
Static magnet 1.5 T- closed bore
Scan Zones Fully body scan
SAR Restrictions Max SAR 4W/kg
MRI settings
activation without
programmer
Yes- SJM MRI Activator™ device
Wireless Telemetry Yes
Daily Remote
Monitoring
Yes
Preparing Accent MRI Patient for an MRI Scan
 Seamless workflow when a patient needs an MRI
 No programmer needed at MRI center
 No impact to clinic efficiencies
MR
Pacemaker
Establish
MRI Settings Values
Settings
(Follow-up Appt)
Patient
Needs MRI
Scan
MRI Scan
DeviceClinicMRICenter
MRI Center
Arrival
MR Mode
OFF Button
MR Mode
ON Button
 Simple application of the MRI Activator
pre- and post-scan
No programmer necessary for MRI access
= Additional programming step required by competitive systems
48
Pre-Scan
Follow-up
(optional)
Preparing Accent MRI Patient for an MRI Scan
(Con’t)
Traditional MRI patient workflow process
Preparing Accent MRI Patient for an MRI Scan
(Con’t)
St. Jude Medical streamlined MRI patient workflow
Take Home Message
• Implantable pulse generators and defibrillators have traditionally
been considered contraindications to MRI.
• Patients with newer generation devices (MRI conditional device)
can safely undergo MRI, including cardiac MRI.
• But, the introduction of MRI conditional systems has led to a
conceptual shift in clinical decision making —
‘can this patient undergo MRI safely?’
is being superseded by
‘should this patient be implanted with an MRI conditional
device?’.
• Perform MRI carefully in Pacemaker-dependent patients

Más contenido relacionado

La actualidad más candente

Basic EP study (part1)
Basic  EP  study (part1)Basic  EP  study (part1)
Basic EP study (part1)SolidaSakhan
 
2021 ESC Guidelines on Cardiac Pacing and CRT
2021 ESC Guidelines on Cardiac Pacing and CRT2021 ESC Guidelines on Cardiac Pacing and CRT
2021 ESC Guidelines on Cardiac Pacing and CRTAyman Azoz
 
A D V A N C E D P A C I N G
A D V A N C E D  P A C I N GA D V A N C E D  P A C I N G
A D V A N C E D P A C I N GSherry Knowles
 
Wide Complex Tachycardia
Wide Complex TachycardiaWide Complex Tachycardia
Wide Complex TachycardiaAhmed Ammar
 
Introduction to Electrophysiology - Supraventricular Tachycardias (1/4 lectures)
Introduction to Electrophysiology - Supraventricular Tachycardias (1/4 lectures)Introduction to Electrophysiology - Supraventricular Tachycardias (1/4 lectures)
Introduction to Electrophysiology - Supraventricular Tachycardias (1/4 lectures)Jose Osorio
 
Ep diagnosis and ablation of avnrt
Ep diagnosis and ablation of avnrtEp diagnosis and ablation of avnrt
Ep diagnosis and ablation of avnrtRahul Chalwade
 
narrow QRS tachycardia diagnostic pacing maneuvers
narrow QRS tachycardia diagnostic pacing maneuversnarrow QRS tachycardia diagnostic pacing maneuvers
narrow QRS tachycardia diagnostic pacing maneuversAhmed Taha
 
Conduction system pacing as resynchronization
Conduction system pacing as resynchronizationConduction system pacing as resynchronization
Conduction system pacing as resynchronizationSergio Pinski
 
Electrophysiology study protocol
Electrophysiology study protocolElectrophysiology study protocol
Electrophysiology study protocolSatyam Rajvanshi
 
Ventricular tachycardia
Ventricular tachycardiaVentricular tachycardia
Ventricular tachycardiasushilrocks5
 
How to follow up a patient with a pacemaker
How to follow up a patient with a pacemakerHow to follow up a patient with a pacemaker
How to follow up a patient with a pacemakerYasmeen Kamal
 
Basics of Electrophysiologic study, part 1 (2020)
Basics of Electrophysiologic study, part 1 (2020)Basics of Electrophysiologic study, part 1 (2020)
Basics of Electrophysiologic study, part 1 (2020)salah_atta
 
Arvd - dr prithvi puwar
Arvd - dr prithvi puwarArvd - dr prithvi puwar
Arvd - dr prithvi puwarPrithvi Puwar
 
Wide complex tacycardia
Wide complex tacycardiaWide complex tacycardia
Wide complex tacycardiaKush Bhagat
 

La actualidad más candente (20)

Basic EP study (part1)
Basic  EP  study (part1)Basic  EP  study (part1)
Basic EP study (part1)
 
2021 ESC Guidelines on Cardiac Pacing and CRT
2021 ESC Guidelines on Cardiac Pacing and CRT2021 ESC Guidelines on Cardiac Pacing and CRT
2021 ESC Guidelines on Cardiac Pacing and CRT
 
Electrophysiology AVRT
Electrophysiology AVRTElectrophysiology AVRT
Electrophysiology AVRT
 
Role of CRT and CRTD in CHF
Role of CRT and CRTD in CHFRole of CRT and CRTD in CHF
Role of CRT and CRTD in CHF
 
A D V A N C E D P A C I N G
A D V A N C E D  P A C I N GA D V A N C E D  P A C I N G
A D V A N C E D P A C I N G
 
9.avnrt chang sl-0324-2
9.avnrt chang sl-0324-29.avnrt chang sl-0324-2
9.avnrt chang sl-0324-2
 
SVT maneuvers
SVT maneuversSVT maneuvers
SVT maneuvers
 
Wide Complex Tachycardia
Wide Complex TachycardiaWide Complex Tachycardia
Wide Complex Tachycardia
 
Essentials of ecg interpretation aphrs
Essentials of ecg interpretation aphrsEssentials of ecg interpretation aphrs
Essentials of ecg interpretation aphrs
 
Introduction to Electrophysiology - Supraventricular Tachycardias (1/4 lectures)
Introduction to Electrophysiology - Supraventricular Tachycardias (1/4 lectures)Introduction to Electrophysiology - Supraventricular Tachycardias (1/4 lectures)
Introduction to Electrophysiology - Supraventricular Tachycardias (1/4 lectures)
 
Ep diagnosis and ablation of avnrt
Ep diagnosis and ablation of avnrtEp diagnosis and ablation of avnrt
Ep diagnosis and ablation of avnrt
 
narrow QRS tachycardia diagnostic pacing maneuvers
narrow QRS tachycardia diagnostic pacing maneuversnarrow QRS tachycardia diagnostic pacing maneuvers
narrow QRS tachycardia diagnostic pacing maneuvers
 
Conduction system pacing as resynchronization
Conduction system pacing as resynchronizationConduction system pacing as resynchronization
Conduction system pacing as resynchronization
 
ARRHYTHMIAS AMENABLE TO RADIOFREQUENCY ABLATION IN CHILDREN
ARRHYTHMIAS AMENABLE TO RADIOFREQUENCY ABLATION IN CHILDRENARRHYTHMIAS AMENABLE TO RADIOFREQUENCY ABLATION IN CHILDREN
ARRHYTHMIAS AMENABLE TO RADIOFREQUENCY ABLATION IN CHILDREN
 
Electrophysiology study protocol
Electrophysiology study protocolElectrophysiology study protocol
Electrophysiology study protocol
 
Ventricular tachycardia
Ventricular tachycardiaVentricular tachycardia
Ventricular tachycardia
 
How to follow up a patient with a pacemaker
How to follow up a patient with a pacemakerHow to follow up a patient with a pacemaker
How to follow up a patient with a pacemaker
 
Basics of Electrophysiologic study, part 1 (2020)
Basics of Electrophysiologic study, part 1 (2020)Basics of Electrophysiologic study, part 1 (2020)
Basics of Electrophysiologic study, part 1 (2020)
 
Arvd - dr prithvi puwar
Arvd - dr prithvi puwarArvd - dr prithvi puwar
Arvd - dr prithvi puwar
 
Wide complex tacycardia
Wide complex tacycardiaWide complex tacycardia
Wide complex tacycardia
 

Similar a New Technology: 核磁共振相容節律器 - ”MRI Standard of Care for Pacemaker”)_20130914中區

MRI safety.pptx
MRI safety.pptxMRI safety.pptx
MRI safety.pptxRivaldohRQ
 
Pacemaker and anaesthesia.pptx
Pacemaker and anaesthesia.pptxPacemaker and anaesthesia.pptx
Pacemaker and anaesthesia.pptxTulikaSingh60
 
Mri safety sujan karki
Mri safety sujan karkiMri safety sujan karki
Mri safety sujan karkiSUJAN KARKI
 
Mri safety – the dos and donts
Mri safety – the dos and dontsMri safety – the dos and donts
Mri safety – the dos and dontsWan Najwa Zaini
 
Magnetic resonance imaging
Magnetic resonance imagingMagnetic resonance imaging
Magnetic resonance imagingvishal gupta
 
mri safety the dos and do nots-1.pptx
mri safety the dos and do nots-1.pptxmri safety the dos and do nots-1.pptx
mri safety the dos and do nots-1.pptxVanshikaGarg76
 
Mri hazards & safety Daniel JP
Mri hazards &  safety  Daniel JPMri hazards &  safety  Daniel JP
Mri hazards & safety Daniel JPalmasmkm
 
Emergency Department ICD and pacemaker issues
Emergency Department ICD and pacemaker issuesEmergency Department ICD and pacemaker issues
Emergency Department ICD and pacemaker issuesSCGH ED CME
 
Recent advancement in MRI
Recent advancement in MRIRecent advancement in MRI
Recent advancement in MRIYehya Yaqoob
 
Surgery in patients with icd and pacemakers
Surgery in patients with icd and pacemakersSurgery in patients with icd and pacemakers
Surgery in patients with icd and pacemakersAbhilash Cheriyan
 
ANESTHESIA for MRI procedures
ANESTHESIA for MRI proceduresANESTHESIA for MRI procedures
ANESTHESIA for MRI proceduresprateek gupta
 
Advanced medical imaging mri safety 2013
Advanced medical imaging mri safety 2013Advanced medical imaging mri safety 2013
Advanced medical imaging mri safety 2013eklundc
 
ANESTHESIA FOR MRI AND CT SCANs suite room
ANESTHESIA FOR MRI AND CT SCANs suite roomANESTHESIA FOR MRI AND CT SCANs suite room
ANESTHESIA FOR MRI AND CT SCANs suite roomZIKRULLAH MALLICK
 
MRI_presentation 3.pptx
MRI_presentation 3.pptxMRI_presentation 3.pptx
MRI_presentation 3.pptxAlissiaPanzera
 
Ultrasound Machine-A Revolution In Medical Imaging
Ultrasound Machine-A Revolution In Medical ImagingUltrasound Machine-A Revolution In Medical Imaging
Ultrasound Machine-A Revolution In Medical ImagingRAVI KANT
 

Similar a New Technology: 核磁共振相容節律器 - ”MRI Standard of Care for Pacemaker”)_20130914中區 (20)

MRI safety.pptx
MRI safety.pptxMRI safety.pptx
MRI safety.pptx
 
Pacemaker and anaesthesia.pptx
Pacemaker and anaesthesia.pptxPacemaker and anaesthesia.pptx
Pacemaker and anaesthesia.pptx
 
Mri safety sujan karki
Mri safety sujan karkiMri safety sujan karki
Mri safety sujan karki
 
Mri safety – the dos and donts
Mri safety – the dos and dontsMri safety – the dos and donts
Mri safety – the dos and donts
 
BASIC OF MRI
BASIC OF MRI BASIC OF MRI
BASIC OF MRI
 
Magnetic resonance imaging
Magnetic resonance imagingMagnetic resonance imaging
Magnetic resonance imaging
 
MRI SAFETY
MRI SAFETYMRI SAFETY
MRI SAFETY
 
mri safety the dos and do nots-1.pptx
mri safety the dos and do nots-1.pptxmri safety the dos and do nots-1.pptx
mri safety the dos and do nots-1.pptx
 
Mri hazards & safety Daniel JP
Mri hazards &  safety  Daniel JPMri hazards &  safety  Daniel JP
Mri hazards & safety Daniel JP
 
Emergency Department ICD and pacemaker issues
Emergency Department ICD and pacemaker issuesEmergency Department ICD and pacemaker issues
Emergency Department ICD and pacemaker issues
 
Recent advancement in MRI
Recent advancement in MRIRecent advancement in MRI
Recent advancement in MRI
 
MRI safety
MRI safety MRI safety
MRI safety
 
Surgery in patients with icd and pacemakers
Surgery in patients with icd and pacemakersSurgery in patients with icd and pacemakers
Surgery in patients with icd and pacemakers
 
ANESTHESIA for MRI procedures
ANESTHESIA for MRI proceduresANESTHESIA for MRI procedures
ANESTHESIA for MRI procedures
 
MRI Safety
MRI SafetyMRI Safety
MRI Safety
 
Advanced medical imaging mri safety 2013
Advanced medical imaging mri safety 2013Advanced medical imaging mri safety 2013
Advanced medical imaging mri safety 2013
 
ANESTHESIA FOR MRI AND CT SCANs suite room
ANESTHESIA FOR MRI AND CT SCANs suite roomANESTHESIA FOR MRI AND CT SCANs suite room
ANESTHESIA FOR MRI AND CT SCANs suite room
 
MRI_presentation 3.pptx
MRI_presentation 3.pptxMRI_presentation 3.pptx
MRI_presentation 3.pptx
 
Introduction to mri
Introduction to mriIntroduction to mri
Introduction to mri
 
Ultrasound Machine-A Revolution In Medical Imaging
Ultrasound Machine-A Revolution In Medical ImagingUltrasound Machine-A Revolution In Medical Imaging
Ultrasound Machine-A Revolution In Medical Imaging
 

Más de Taiwan Heart Rhythm Society

The clinical application of intracardiac echocardiography in cardiac
The clinical application of intracardiac echocardiography in cardiacThe clinical application of intracardiac echocardiography in cardiac
The clinical application of intracardiac echocardiography in cardiacTaiwan Heart Rhythm Society
 
Oral anticoagulants in patients with atrial fibrillation
Oral anticoagulants in patients with atrial fibrillationOral anticoagulants in patients with atrial fibrillation
Oral anticoagulants in patients with atrial fibrillationTaiwan Heart Rhythm Society
 

Más de Taiwan Heart Rhythm Society (20)

Arrhythmia news 045.pdf
Arrhythmia news 045.pdfArrhythmia news 045.pdf
Arrhythmia news 045.pdf
 
photo.pptx
photo.pptxphoto.pptx
photo.pptx
 
Arrhythmia news no.44
Arrhythmia news no.44Arrhythmia news no.44
Arrhythmia news no.44
 
Thrs arrhythmia news
Thrs arrhythmia newsThrs arrhythmia news
Thrs arrhythmia news
 
Arrhythmia news 042
Arrhythmia news 042Arrhythmia news 042
Arrhythmia news 042
 
Picture
PicturePicture
Picture
 
Arrhythmia news no.41
Arrhythmia news no.41Arrhythmia news no.41
Arrhythmia news no.41
 
Arrhythmia news no.40
Arrhythmia news no.40Arrhythmia news no.40
Arrhythmia news no.40
 
Arrhythmia news 039
Arrhythmia news 039Arrhythmia news 039
Arrhythmia news 039
 
Challenging and Unknown ECGs (2)
Challenging and Unknown ECGs (2)Challenging and Unknown ECGs (2)
Challenging and Unknown ECGs (2)
 
Arrhythmia news 038
Arrhythmia news 038Arrhythmia news 038
Arrhythmia news 038
 
Photos
PhotosPhotos
Photos
 
Arrhythmia news 037
Arrhythmia news 037Arrhythmia news 037
Arrhythmia news 037
 
Arrhythmia news no.36
Arrhythmia news no.36Arrhythmia news no.36
Arrhythmia news no.36
 
The clinical application of intracardiac echocardiography in cardiac
The clinical application of intracardiac echocardiography in cardiacThe clinical application of intracardiac echocardiography in cardiac
The clinical application of intracardiac echocardiography in cardiac
 
Comprehensive management
Comprehensive managementComprehensive management
Comprehensive management
 
Arrhythmia news 035
Arrhythmia news 035Arrhythmia news 035
Arrhythmia news 035
 
Oral anticoagulants in patients with atrial fibrillation
Oral anticoagulants in patients with atrial fibrillationOral anticoagulants in patients with atrial fibrillation
Oral anticoagulants in patients with atrial fibrillation
 
THRS allied professional training course
THRS allied professional training courseTHRS allied professional training course
THRS allied professional training course
 
Pictures
PicturesPictures
Pictures
 

Último

See the 2,456 pharmacies on the National E-Pharmacy Platform
See the 2,456 pharmacies on the National E-Pharmacy PlatformSee the 2,456 pharmacies on the National E-Pharmacy Platform
See the 2,456 pharmacies on the National E-Pharmacy PlatformKweku Zurek
 
Low Rate Call Girls Mumbai Suman 9910780858 Independent Escort Service Mumbai
Low Rate Call Girls Mumbai Suman 9910780858 Independent Escort Service MumbaiLow Rate Call Girls Mumbai Suman 9910780858 Independent Escort Service Mumbai
Low Rate Call Girls Mumbai Suman 9910780858 Independent Escort Service Mumbaisonalikaur4
 
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original Photos
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original PhotosCall Girl Service Bidadi - For 7001305949 Cheap & Best with original Photos
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original Photosnarwatsonia7
 
Housewife Call Girls Bangalore - Call 7001305949 Rs-3500 with A/C Room Cash o...
Housewife Call Girls Bangalore - Call 7001305949 Rs-3500 with A/C Room Cash o...Housewife Call Girls Bangalore - Call 7001305949 Rs-3500 with A/C Room Cash o...
Housewife Call Girls Bangalore - Call 7001305949 Rs-3500 with A/C Room Cash o...narwatsonia7
 
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy Girls
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy GirlsCall Girls In Andheri East Call 9920874524 Book Hot And Sexy Girls
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy Girlsnehamumbai
 
Call Girls Hsr Layout Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hsr Layout Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Hsr Layout Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hsr Layout Just Call 7001305949 Top Class Call Girl Service Availablenarwatsonia7
 
Ahmedabad Call Girls CG Road 🔝9907093804 Short 1500 💋 Night 6000
Ahmedabad Call Girls CG Road 🔝9907093804  Short 1500  💋 Night 6000Ahmedabad Call Girls CG Road 🔝9907093804  Short 1500  💋 Night 6000
Ahmedabad Call Girls CG Road 🔝9907093804 Short 1500 💋 Night 6000aliya bhat
 
Call Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
Call Girl Bangalore Nandini 7001305949 Independent Escort Service BangaloreCall Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
Call Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalorenarwatsonia7
 
Call Girl Lucknow Mallika 7001305949 Independent Escort Service Lucknow
Call Girl Lucknow Mallika 7001305949 Independent Escort Service LucknowCall Girl Lucknow Mallika 7001305949 Independent Escort Service Lucknow
Call Girl Lucknow Mallika 7001305949 Independent Escort Service Lucknownarwatsonia7
 
Call Girls Frazer Town Just Call 7001305949 Top Class Call Girl Service Avail...
Call Girls Frazer Town Just Call 7001305949 Top Class Call Girl Service Avail...Call Girls Frazer Town Just Call 7001305949 Top Class Call Girl Service Avail...
Call Girls Frazer Town Just Call 7001305949 Top Class Call Girl Service Avail...narwatsonia7
 
Housewife Call Girls Hsr Layout - Call 7001305949 Rs-3500 with A/C Room Cash ...
Housewife Call Girls Hsr Layout - Call 7001305949 Rs-3500 with A/C Room Cash ...Housewife Call Girls Hsr Layout - Call 7001305949 Rs-3500 with A/C Room Cash ...
Housewife Call Girls Hsr Layout - Call 7001305949 Rs-3500 with A/C Room Cash ...narwatsonia7
 
VIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service Mumbai
VIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service MumbaiVIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service Mumbai
VIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service Mumbaisonalikaur4
 
Kolkata Call Girls Services 9907093804 @24x7 High Class Babes Here Call Now
Kolkata Call Girls Services 9907093804 @24x7 High Class Babes Here Call NowKolkata Call Girls Services 9907093804 @24x7 High Class Babes Here Call Now
Kolkata Call Girls Services 9907093804 @24x7 High Class Babes Here Call NowNehru place Escorts
 
Dwarka Sector 6 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few Cl...
Dwarka Sector 6 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few Cl...Dwarka Sector 6 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few Cl...
Dwarka Sector 6 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few Cl...rajnisinghkjn
 
College Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort Service
College Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort ServiceCollege Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort Service
College Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort ServiceNehru place Escorts
 
Call Girls Jayanagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jayanagar Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Jayanagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jayanagar Just Call 7001305949 Top Class Call Girl Service Availablenarwatsonia7
 
97111 47426 Call Girls In Delhi MUNIRKAA
97111 47426 Call Girls In Delhi MUNIRKAA97111 47426 Call Girls In Delhi MUNIRKAA
97111 47426 Call Girls In Delhi MUNIRKAAjennyeacort
 
Call Girls Service Noida Maya 9711199012 Independent Escort Service Noida
Call Girls Service Noida Maya 9711199012 Independent Escort Service NoidaCall Girls Service Noida Maya 9711199012 Independent Escort Service Noida
Call Girls Service Noida Maya 9711199012 Independent Escort Service NoidaPooja Gupta
 
Book Call Girls in Yelahanka - For 7001305949 Cheap & Best with original Photos
Book Call Girls in Yelahanka - For 7001305949 Cheap & Best with original PhotosBook Call Girls in Yelahanka - For 7001305949 Cheap & Best with original Photos
Book Call Girls in Yelahanka - For 7001305949 Cheap & Best with original Photosnarwatsonia7
 

Último (20)

sauth delhi call girls in Bhajanpura 🔝 9953056974 🔝 escort Service
sauth delhi call girls in Bhajanpura 🔝 9953056974 🔝 escort Servicesauth delhi call girls in Bhajanpura 🔝 9953056974 🔝 escort Service
sauth delhi call girls in Bhajanpura 🔝 9953056974 🔝 escort Service
 
See the 2,456 pharmacies on the National E-Pharmacy Platform
See the 2,456 pharmacies on the National E-Pharmacy PlatformSee the 2,456 pharmacies on the National E-Pharmacy Platform
See the 2,456 pharmacies on the National E-Pharmacy Platform
 
Low Rate Call Girls Mumbai Suman 9910780858 Independent Escort Service Mumbai
Low Rate Call Girls Mumbai Suman 9910780858 Independent Escort Service MumbaiLow Rate Call Girls Mumbai Suman 9910780858 Independent Escort Service Mumbai
Low Rate Call Girls Mumbai Suman 9910780858 Independent Escort Service Mumbai
 
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original Photos
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original PhotosCall Girl Service Bidadi - For 7001305949 Cheap & Best with original Photos
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original Photos
 
Housewife Call Girls Bangalore - Call 7001305949 Rs-3500 with A/C Room Cash o...
Housewife Call Girls Bangalore - Call 7001305949 Rs-3500 with A/C Room Cash o...Housewife Call Girls Bangalore - Call 7001305949 Rs-3500 with A/C Room Cash o...
Housewife Call Girls Bangalore - Call 7001305949 Rs-3500 with A/C Room Cash o...
 
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy Girls
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy GirlsCall Girls In Andheri East Call 9920874524 Book Hot And Sexy Girls
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy Girls
 
Call Girls Hsr Layout Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hsr Layout Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Hsr Layout Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hsr Layout Just Call 7001305949 Top Class Call Girl Service Available
 
Ahmedabad Call Girls CG Road 🔝9907093804 Short 1500 💋 Night 6000
Ahmedabad Call Girls CG Road 🔝9907093804  Short 1500  💋 Night 6000Ahmedabad Call Girls CG Road 🔝9907093804  Short 1500  💋 Night 6000
Ahmedabad Call Girls CG Road 🔝9907093804 Short 1500 💋 Night 6000
 
Call Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
Call Girl Bangalore Nandini 7001305949 Independent Escort Service BangaloreCall Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
Call Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
 
Call Girl Lucknow Mallika 7001305949 Independent Escort Service Lucknow
Call Girl Lucknow Mallika 7001305949 Independent Escort Service LucknowCall Girl Lucknow Mallika 7001305949 Independent Escort Service Lucknow
Call Girl Lucknow Mallika 7001305949 Independent Escort Service Lucknow
 
Call Girls Frazer Town Just Call 7001305949 Top Class Call Girl Service Avail...
Call Girls Frazer Town Just Call 7001305949 Top Class Call Girl Service Avail...Call Girls Frazer Town Just Call 7001305949 Top Class Call Girl Service Avail...
Call Girls Frazer Town Just Call 7001305949 Top Class Call Girl Service Avail...
 
Housewife Call Girls Hsr Layout - Call 7001305949 Rs-3500 with A/C Room Cash ...
Housewife Call Girls Hsr Layout - Call 7001305949 Rs-3500 with A/C Room Cash ...Housewife Call Girls Hsr Layout - Call 7001305949 Rs-3500 with A/C Room Cash ...
Housewife Call Girls Hsr Layout - Call 7001305949 Rs-3500 with A/C Room Cash ...
 
VIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service Mumbai
VIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service MumbaiVIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service Mumbai
VIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service Mumbai
 
Kolkata Call Girls Services 9907093804 @24x7 High Class Babes Here Call Now
Kolkata Call Girls Services 9907093804 @24x7 High Class Babes Here Call NowKolkata Call Girls Services 9907093804 @24x7 High Class Babes Here Call Now
Kolkata Call Girls Services 9907093804 @24x7 High Class Babes Here Call Now
 
Dwarka Sector 6 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few Cl...
Dwarka Sector 6 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few Cl...Dwarka Sector 6 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few Cl...
Dwarka Sector 6 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few Cl...
 
College Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort Service
College Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort ServiceCollege Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort Service
College Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort Service
 
Call Girls Jayanagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jayanagar Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Jayanagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jayanagar Just Call 7001305949 Top Class Call Girl Service Available
 
97111 47426 Call Girls In Delhi MUNIRKAA
97111 47426 Call Girls In Delhi MUNIRKAA97111 47426 Call Girls In Delhi MUNIRKAA
97111 47426 Call Girls In Delhi MUNIRKAA
 
Call Girls Service Noida Maya 9711199012 Independent Escort Service Noida
Call Girls Service Noida Maya 9711199012 Independent Escort Service NoidaCall Girls Service Noida Maya 9711199012 Independent Escort Service Noida
Call Girls Service Noida Maya 9711199012 Independent Escort Service Noida
 
Book Call Girls in Yelahanka - For 7001305949 Cheap & Best with original Photos
Book Call Girls in Yelahanka - For 7001305949 Cheap & Best with original PhotosBook Call Girls in Yelahanka - For 7001305949 Cheap & Best with original Photos
Book Call Girls in Yelahanka - For 7001305949 Cheap & Best with original Photos
 

New Technology: 核磁共振相容節律器 - ”MRI Standard of Care for Pacemaker”)_20130914中區

  • 1. New Technology: 核磁共振相容節律器 (MRI Conditional Pacemaker System) 林圀宏 心臟內科 中國醫藥大學附設醫院 2013 心臟電生理相關專業人員教育課程
  • 3. Why MRI Scans are Important to Clinical Medicine • Fastest Growing Standard of Care in Diagnostic Imaging1 • Superior Soft Tissue Imaging2  Primary method to evaluate:  Central Nervous System  Musculoskeletal System  Oncological Conditions  Some Cardiovascular Disorders • MRI complements CT (which excels when imaging bony structures) • No radiation risk to patient or healthcare provider • Since the absence of x-ray radiation, MRI is optimal for follow-up of chronic diseases that require repeat imaging and for diagnostic imaging in young patients and women of childbearing age. 1. Kaiser CP. Soaring MRI use draws scrutiny. Diagnostic Imaging Online January 4, 2002. CMP United Business Media: A CMP Healthcare Media Web Site. Available at: http://www.diagnosticimaging.com/dinews/2002010401.shtml. Accessed October 19, 2004. 2. Duru F, Luechinger R, Scheidegger MB, et al. Pacing in magnetic resonance imaging environment: Clinical and technical considerations on compatibility. Eur Heart J. January 2001;22(2):113-124.
  • 4. Cervical spine computed tomography (CT) vs MRI in a patient with neck pain and fever. Nazarian S et al. Circ Arrhythm Electrophysiol 2013;6:419- 428 Copyright © American Heart Association
  • 5. Brain computed tomography (CT) vs MRI in a patient with weakness. Nazarian S et al. Circ Arrhythm Electrophysiol 2013;6:419- 428 Copyright © American Heart Association
  • 6. Cardiac computed tomography (CT) vs MRI in a patient with facial swelling. Nazarian S et al. Circ Arrhythm Electrophysiol 2013;6:419- 428 Copyright © American Heart Association
  • 7.
  • 10. 10 Anatomy of an MRI Scanner Three basic components:  Static magnet  Gradient magnets  RF coil 10 Static Determines magnetic strength of scanner:  1.5 or 3.0 tesla  1 tesla (T) = 10,000 gauss.  The earth’s magnetic field = 0.5 gauss.  Magnet response in SJM pacemakers occurs in response to 8-10 gauss directly over the pacemaker.
  • 11.
  • 12.
  • 13. Force • The most intuitive potential interaction of implanted devices with an external magnetic field is the possibility for movement and dislocation of the device because of magnetic force. • Current lead designs contain little or no ferromagnetic components and are not likely to experience force and torque. • The potential for movement of a pacemaker or ICD generator in the MRI environment depends on the  magnetic field strength  ferromagnetic properties of the device components  the implant distance from the magnet bore and the stability of the implant.
  • 14. Current induction • The RF and pulsed gradient magnetic fields of the MRI scanner may induce electric currents in leads within the field, if the lead is part of a current loop that is completed through the body. • The ratio of lead length versus RF wavelength and lead conformations, such as loops, are strongly associated with the extent of current induction.
  • 15. Gradient Waveform Amplitude 0.4 ms 500 ms – 750 ms Pacing Pulse Time ~ 0.005 ms Gradient Induced Pulse Gradient Induced Pulse Gradient Mechanism
  • 16. Stimulation hazard gradient-induced high rate pacing Start of Scan EKG Pulse Ox Canine Test Unintended Stimulation Clinical Impact The MRI scanner is pacing the heart
  • 17. Heating and tissue damage • Metallic devices and leads can act as an antenna thus amplifying local radiofrequency energy deposition. • Fractured leads or lead loop configurations may increase the potential for heating. • Epicardial leads that are not cooled by blood flow and abandoned leads may also be prone to increased heating.
  • 18. Heating of cardiac tissue Heating at the lead tip can result in threshold changes.
  • 19. Malfunction of devices • CIED may provide unnecessary therapies or fail to provide necessary therapies when placed in the MRI scanner. • Pacemakers and ICDs have the potential for receiving electromagnetic interference (EMI) in the MRI environment, resulting in:  radiofrequency noise tracking  asynchronous pacing  inhibition of demand pacing  delivery of ICD therapies  programming changes  loss of function.
  • 20. Figure 2. EMI noise from different MR scan protocols interpreted by the device as ventricular fibrillation (VF). Roguin A et al. Circulation 2004;110:475-482
  • 21. Simultaneous electrocardiography and pulse oximetry while in 3T bore of magnetic resonance imaging followed by application of gradient scan leading to inhibition of effective pacing. Gimbel J R Europace 2009;11:1241-1242
  • 22. Malfunction of devices • The static magnetic field of the MRI scanner can also alter device function by inducing unexpected reed switch opening or closure. • In addition, temporary programming changes made to avoid device interaction with the MRI scanner (such as disabling of tachycardia therapies) may lead to catastrophic results if a spontaneous arrhythmia occurs and is not recognized.
  • 23. Simultaneous recording of ECG and pulse oximetry during MRI of the brain in a patient with a Thera pacemaker reprogrammed to D00 80 bpm. Sommer T et al. Circulation 2006;114:1285-1292 Copyright © American Heart Association
  • 24. During and/or after the MRI • no inhibition of pacemaker output or cardiac arrest • no sustained ventricular arrhythmias • no unexpected changes of heart rate • no electrical resets • no pacemaker system disturbances and • no sensation of torque or pain
  • 25. Variables Affecting Magnitude of Risks  Length/position of pacing leads  Patient and device position within machine  Patient factors / medical history  MRI scan duration  Blood flow at lead/tissue interface  Strength of RF field  Target anatomy of scan  Type of imaging MRI sequence  Pacemaker and Lead Design
  • 26. MR Conditional Labeling* MR Safe MR Conditional MR Unsafe An item that has been demonstrated to pose no known hazards in a specified MRI environment with specified conditions of use.
  • 27. Device Design Solutions • Minimize ferromagnetic content • Hybrid-case connection • Hall sensor • Optimize input circuitry
  • 28. Solutions to Reduce MRI Risk • Filter board added to pacemaker hardware • When the filter detects MRI RF signals, it will not allow the signals to be sent back down the lead (rectification), which would result in cardiac stimulation. . 28 X-ray marker MRI filter assembly
  • 29. Solutions to Reduce MRI Risk • 2 filters are incorporated into the pacing lead to reduce the risk of heating from RF signals.
  • 31. Patient Care Pathway 1. Pre-Screening and Scheduling 2. Pre-scan 3. Scan 4. Post-Scan
  • 32. Pre-screening • Ways to ID pacemaker system: 1. Patient ID Card 2. X-Ray 3. Call cardiology
  • 34. X-Ray Radiopaque identifies the system via unique radiopaque visible under x-ray 1 Location of the device radiopaque symbol 2 Device radiopaque MRI symbol 3 Lead radiopaque MRI symbol
  • 35. Radiopaque identification available: 1. pulse generator 2. lead
  • 37. MRI Pacemaker Identification (Con’t) • To verify that a patient has a MRI pacemaker, you may use X-ray identification, confirming that the pacemaker and the lead are appropriate. Radiopaque markers will show if the patient has an appropriate system for an MRI scan 37 St. Jude Medical Identifier Device MRI Symbol
  • 38. To program MRI SureScan to On, go to Params -> Additional Features.
  • 39. When programmed to On, MRI SureScan operation disables • Arrhythmia detection, • Magnet mode, • All user-defined diagnostics. On
  • 40.
  • 41. • For patients who require pacing support, program the device to an asynchronous pacing mode (DOO, AOO, VOO). • For patients who do not require pacing support, program the device to the non-pacing mode (ODO).
  • 42.
  • 43. Patients and their implanted systems must be screened to meet the following requirements: •No previously implanted (active or abandoned) medical devices, leads, lead extenders, or lead adaptors. •No broken leads or leads with intermittent electrical contact as confirmed by lead impedance history. •A SureScan pacing system that has been implanted for a minimum of 6 weeks. •A SureScan pacing system implanted in the left or right pectoral region. •Pace polarity parameters set to Bipolar for programming MRI SureScan to On (for Advisa DR MRI). •Pacing capture thresholds of ≤ 2.0 volts (V) at a pulse width of 0.4 milliseconds (ms). •A lead impedance value of ≥ 200 ohms (Ω) and ≤ 1500 Ω. •No diaphragmatic stimulation at a pacing output of 5.0 V and at a pulse width of 1.0 ms in patients whose device will be programmed to an asynchronous pacing mode when MRI SureScan is on. Cardiology requirements:
  • 45. -45- Competitive Comparison SJM Accent MRI™ Static magnet 1.5 T- closed bore Scan Zones Fully body scan SAR Restrictions Max SAR 4W/kg MRI settings activation without programmer Yes- SJM MRI Activator™ device Wireless Telemetry Yes Daily Remote Monitoring Yes
  • 46. Preparing Accent MRI Patient for an MRI Scan
  • 47.  Seamless workflow when a patient needs an MRI  No programmer needed at MRI center  No impact to clinic efficiencies MR Pacemaker Establish MRI Settings Values Settings (Follow-up Appt) Patient Needs MRI Scan MRI Scan DeviceClinicMRICenter MRI Center Arrival MR Mode OFF Button MR Mode ON Button  Simple application of the MRI Activator pre- and post-scan No programmer necessary for MRI access = Additional programming step required by competitive systems 48 Pre-Scan Follow-up (optional)
  • 48. Preparing Accent MRI Patient for an MRI Scan (Con’t) Traditional MRI patient workflow process
  • 49. Preparing Accent MRI Patient for an MRI Scan (Con’t) St. Jude Medical streamlined MRI patient workflow
  • 50. Take Home Message • Implantable pulse generators and defibrillators have traditionally been considered contraindications to MRI. • Patients with newer generation devices (MRI conditional device) can safely undergo MRI, including cardiac MRI. • But, the introduction of MRI conditional systems has led to a conceptual shift in clinical decision making — ‘can this patient undergo MRI safely?’ is being superseded by ‘should this patient be implanted with an MRI conditional device?’. • Perform MRI carefully in Pacemaker-dependent patients