3. Why MRI Scans are Important to Clinical Medicine
• Fastest Growing Standard of Care in Diagnostic Imaging1
• Superior Soft Tissue Imaging2
Primary method to evaluate:
Central Nervous System
Musculoskeletal System
Oncological Conditions
Some Cardiovascular Disorders
• MRI complements CT (which excels when imaging bony structures)
• No radiation risk to patient or healthcare provider
• Since the absence of x-ray radiation, MRI is optimal for follow-up of chronic
diseases that require repeat imaging and for diagnostic imaging in young patients
and women of childbearing age.
1. Kaiser CP. Soaring MRI use draws scrutiny. Diagnostic Imaging Online January 4, 2002. CMP United Business Media: A CMP Healthcare Media Web
Site. Available at: http://www.diagnosticimaging.com/dinews/2002010401.shtml. Accessed October 19, 2004.
2. Duru F, Luechinger R, Scheidegger MB, et al. Pacing in magnetic resonance imaging environment: Clinical and technical considerations on
compatibility. Eur Heart J. January 2001;22(2):113-124.
10. 10
Anatomy of an MRI Scanner
Three basic components:
Static magnet
Gradient magnets
RF coil
10
Static
Determines magnetic strength
of scanner:
1.5 or 3.0 tesla
1 tesla (T) = 10,000
gauss.
The earth’s magnetic
field = 0.5 gauss.
Magnet response in
SJM pacemakers occurs
in response to 8-10
gauss directly over the
pacemaker.
11.
12.
13. Force
• The most intuitive potential interaction of implanted devices
with an external magnetic field is the possibility for
movement and dislocation of the device because of magnetic
force.
• Current lead designs contain little or no ferromagnetic
components and are not likely to experience force and
torque.
• The potential for movement of a pacemaker or ICD
generator in the MRI environment depends on the
magnetic field strength
ferromagnetic properties of the device components
the implant distance from the magnet bore and the
stability of the implant.
14. Current induction
• The RF and pulsed gradient magnetic fields of the MRI
scanner may induce electric currents in leads within the field,
if the lead is part of a current loop that is completed through
the body.
• The ratio of lead length versus RF wavelength and lead
conformations, such as loops, are strongly associated with the
extent of current induction.
16. Stimulation hazard gradient-induced high rate pacing
Start of Scan
EKG
Pulse Ox
Canine Test
Unintended Stimulation Clinical Impact
The MRI scanner is pacing the heart
17. Heating and tissue damage
• Metallic devices and leads can act as an antenna thus
amplifying local radiofrequency energy deposition.
• Fractured leads or lead loop configurations may increase
the potential for heating.
• Epicardial leads that are not cooled by blood flow and
abandoned leads may also be prone to increased heating.
18. Heating of cardiac tissue
Heating at the lead tip can result in threshold changes.
19. Malfunction of devices
• CIED may provide unnecessary therapies or fail to provide
necessary therapies when placed in the MRI scanner.
• Pacemakers and ICDs have the potential for receiving
electromagnetic interference (EMI) in the MRI environment,
resulting in:
radiofrequency noise tracking
asynchronous pacing
inhibition of demand pacing
delivery of ICD therapies
programming changes
loss of function.
20. Figure 2. EMI noise from different MR scan protocols interpreted by the device as ventricular
fibrillation (VF).
Roguin A et al. Circulation 2004;110:475-482
21. Simultaneous electrocardiography and pulse oximetry while in 3T
bore of magnetic resonance imaging followed by application of
gradient scan leading to inhibition of effective pacing.
Gimbel J R Europace 2009;11:1241-1242
22. Malfunction of devices
• The static magnetic field of the MRI scanner can also
alter device function by inducing unexpected reed switch
opening or closure.
• In addition, temporary programming changes made to
avoid device interaction with the MRI scanner (such as
disabling of tachycardia therapies) may lead to
catastrophic results if a spontaneous arrhythmia occurs
and is not recognized.
24. During and/or after the MRI
• no inhibition of pacemaker output or cardiac
arrest
• no sustained ventricular arrhythmias
• no unexpected changes of heart rate
• no electrical resets
• no pacemaker system disturbances and
• no sensation of torque or pain
25. Variables Affecting Magnitude of Risks
Length/position of pacing leads
Patient and device position within machine
Patient factors / medical history
MRI scan duration
Blood flow at lead/tissue interface
Strength of RF field
Target anatomy of scan
Type of imaging MRI sequence
Pacemaker and Lead Design
26. MR Conditional Labeling*
MR Safe
MR Conditional
MR Unsafe
An item that has been
demonstrated to pose no known
hazards in a specified MRI
environment with specified
conditions of use.
28. Solutions to Reduce MRI Risk
• Filter board added to pacemaker hardware
• When the filter detects MRI RF signals, it will not allow the
signals to be sent back down the lead (rectification), which
would result in cardiac stimulation.
.
28
X-ray marker
MRI filter assembly
29. Solutions to Reduce MRI Risk
• 2 filters are incorporated into the pacing lead to reduce the
risk of heating from RF signals.
34. X-Ray
Radiopaque identifies the system via unique
radiopaque visible under x-ray
1 Location of the device radiopaque symbol
2 Device radiopaque MRI symbol
3 Lead radiopaque MRI symbol
37. MRI Pacemaker Identification (Con’t)
• To verify that a patient has a MRI pacemaker, you may use X-ray
identification, confirming that the pacemaker and the lead are appropriate.
Radiopaque markers will show if the patient has an appropriate system for an
MRI scan
37
St. Jude Medical
Identifier
Device MRI
Symbol
38. To program MRI SureScan to On, go to Params
-> Additional Features.
39. When programmed to On, MRI SureScan operation disables
• Arrhythmia detection,
• Magnet mode,
• All user-defined diagnostics.
On
40.
41. • For patients who require pacing support, program the device to an
asynchronous pacing mode (DOO, AOO, VOO).
• For patients who do not require pacing support, program the device to the
non-pacing mode (ODO).
42.
43. Patients and their implanted systems must be screened to meet the
following requirements:
•No previously implanted (active or abandoned) medical devices,
leads, lead extenders, or lead adaptors.
•No broken leads or leads with intermittent electrical contact as
confirmed by lead impedance history.
•A SureScan pacing system that has been implanted for a minimum of
6 weeks.
•A SureScan pacing system implanted in the left or right pectoral
region.
•Pace polarity parameters set to Bipolar for programming MRI
SureScan to On (for Advisa DR MRI).
•Pacing capture thresholds of ≤ 2.0 volts (V) at a pulse width of 0.4
milliseconds (ms).
•A lead impedance value of ≥ 200 ohms (Ω) and ≤ 1500 Ω.
•No diaphragmatic stimulation at a pacing output of 5.0 V and at a
pulse width of 1.0 ms in patients whose device will be
programmed to an asynchronous pacing mode when MRI
SureScan is on.
Cardiology requirements:
47. Seamless workflow when a patient needs an MRI
No programmer needed at MRI center
No impact to clinic efficiencies
MR
Pacemaker
Establish
MRI Settings Values
Settings
(Follow-up Appt)
Patient
Needs MRI
Scan
MRI Scan
DeviceClinicMRICenter
MRI Center
Arrival
MR Mode
OFF Button
MR Mode
ON Button
Simple application of the MRI Activator
pre- and post-scan
No programmer necessary for MRI access
= Additional programming step required by competitive systems
48
Pre-Scan
Follow-up
(optional)
48. Preparing Accent MRI Patient for an MRI Scan
(Con’t)
Traditional MRI patient workflow process
49. Preparing Accent MRI Patient for an MRI Scan
(Con’t)
St. Jude Medical streamlined MRI patient workflow
50. Take Home Message
• Implantable pulse generators and defibrillators have traditionally
been considered contraindications to MRI.
• Patients with newer generation devices (MRI conditional device)
can safely undergo MRI, including cardiac MRI.
• But, the introduction of MRI conditional systems has led to a
conceptual shift in clinical decision making —
‘can this patient undergo MRI safely?’
is being superseded by
‘should this patient be implanted with an MRI conditional
device?’.
• Perform MRI carefully in Pacemaker-dependent patients