2. Outline
Introduction to Inertial Sensors;
Static Evaluation of the Noise;
Sit to Stand Task Evaluation;
Conclusions.
6/16/2012 Biomedical Transducers - Inertial Sensors 2
3. Inertia and Inertial Frame
• Inertial Frame of Reference: is a frame in a
state of constant, rectilinear motion with
respect to one another: an accelerometer at
rest in one would detect zero acceleration;
• Newton’s First Law of Inertia: an observer in a
inertial frame of reference observes a body:
inertia is the natural tendency of that body to
remain immobile or in motion with constant
speed along a straight line;
6/16/2012 Biomedical Transducers - Inertial Sensors 3
4. Inertia and Inertial Frame
• Newton’s Second Law: A force will accelerate
a body, in the direction of the force at a rate
inversely proportional to the mass of the
body;
• Mass is the linear quantification of inertia;
• The laws of Classical Mechanics
(Biomechanics included) are valid and
maintain the same form in all inertial
reference systems.
6/16/2012 Biomedical Transducers - Inertial Sensors 4
5. What is a sensor?
• Instrument capable to transduce a physical
quantity to a measurable electric signal;
• Accuracy vs Precision;
• Inertial sensor: functioning principle based on
inertial phenomena.
6/16/2012 Biomedical Transducers - Inertial Sensors 5
6. Inertial Sensors
• Accelerometers: sense linear acceleration
[m/s^2] along a specific axis;
• Gyroscopes: sense angular velocity axis,
measured in [rad/s];
• Magnetometer: sense the strength of a
magnetic field, measured in [mGauss].
6/16/2012 Biomedical Transducers - Inertial Sensors 6
12. 2.2.2 Evaluate and characterize the
noise in terms of mean and standard
deviation of the ouputs
• Mean() function
• Std() function
6/16/2012 12
Biomedical Transducers - Inertial Sensors
13. The results for XSENS IMU are as follows:
6/16/2012 13
Biomedical Transducers - Inertial Sensors
14. The results for OPAL IMU are as follows:
6/16/2012 14
Biomedical Transducers - Inertial Sensors
15. 2.3 Evaluate the drift effect
• Detrend() function
• Polyfit() function, y=mx+b
6/16/2012 15
Biomedical Transducers - Inertial Sensors
16. The results for XSENS IMU are as follows:
6/16/2012 16
Biomedical Transducers - Inertial Sensors
17. The results for OPAL IMU are as follows:
6/16/2012 17
Biomedical Transducers - Inertial Sensors
18. 2.4 What are the main difference
between the noises on each sensor?
6/16/2012 18
Biomedical Transducers - Inertial Sensors
20. From these plots we can conclude that:
• The Xsens IMU, has overall better
performance with respect to the Opal
IMU;
• The Xsens trend of noise drift is almost
parallel to the time axis and the signals
have lower offsets with respect to the
Opal signals.
6/16/2012 20
Biomedical Transducers - Inertial Sensors
21. 2.5 Does the standard deviation of the noise
correspond to that reported in the data sheet?
• Xsens: As we can see in the tables above, the data reported in the
datasheet and our measured ones, differ from a factor of ±.001; So we
obtain very good measurements in terms of accuracy and precision;
• Opal: In this case we have to convert the data from [μg/»Hz] to [m/s2]
for the linear acceleration Noise and from [°/s/»Hz] to [rad/s] for the
angular velocity, using the bandwidth data B = 50[Hz]. Also in this case
we obtain good measurement in terms of accuracy and precision.
6/16/2012 21
Biomedical Transducers - Inertial Sensors
22. 3. Sit to Stand
• Opal IMU1 placed on the Thigh, in lateral
position;
• Opal IMU2 placed on the Trunk, at L5 height;
• 4 trials with 5 repetitions at different speed;
• f_{sample} = 128[Hz];
6/16/2012 Biomedical Transducers - Inertial Sensors 22
25. Digital Filtering
2
sample
cut
n
f
f
W
6/16/2012 Biomedical Transducers - Inertial Sensors 25
Normalized CutOff Frequency
Because of Noisy signals: Lowpass Filtering needed
[b,a] = butter(order,Wn,type): extract the coefficients;
filtfilt(b,a,input): No Phase Shift, forward + backward
filtering.
28. Results: Table
StS Time mean
[s]
TtS Time mean
[s]
StS Angle
mean [°]
TtS Angle
mean [°]
Trial 1 1.7984 1.4375 94.7484° - 90.1806°
Trial 2 1.391 1.1719 96.3518° - 92.9096°
Trial 3 1.4672 1.3531 75.5568° - 71-7260°
Trial 4 .9906 .09562 71.6656° - 69.1158°
6/16/2012 Biomedical Transducers - Inertial Sensors 28
4 Trials 5 Repetitions StS = Sit to Stand Task TtS = Time to Sit Task
29. Sit to Stand Conclusions
+ Results achievable with only 1 IMU (on the
thigh)
+ Robust algorithm
• Kalman fusion filter to improve the algorithm
6/16/2012 Biomedical Transducers - Inertial Sensors 29