Lateral Flow
Achieving 5 x More Sensitivity And Less Variation In
Results Than Covalent Chemistry With Magnetic
Particles
By Anteo Technologies
!
! Page !1
Summary
A new method of conjugating antibodies to particles was compared to covalent conjugation
in a lateral flow assay for human Chorionic Gonadotropin (hCG). Anteo Technologies will be
launching a new Lateral Flow Coupling Kit at AACC 2016 that contains Anteo’s patented
Activation Reagent with all necessary buffers for conjugating antibodies to a range of particles
used for lateral flow applications. Here we demonstrate the performance advantages of using
activated magnetic particles, only requiring the addition of antibody solution and blocking.
Keeping all other variables constant, this kit was directly compared to covalently (EDAC
chemistry) conjugated magnetic particles of the same type using visual detection as the
readout. In this hCG study, this coupling kit achieves five times more sensitivity than the
covalently conjugated magnetic particle based assay using half the amount of antibody.
Comparison of Anteo’s Activated Magnetic Particles and Covalently
Conjugated Magnetic Particles in a Lateral Flow Assay for Human
Chorionic Gonadotropin (hCG)
Introduction
Lateral flow immunoassays (LFIA) or strip
tests are conceptually simple
immunoassays well suited to Point of
Care (PoC) applications requiring little or
no infrastructure support (Wong and Tse
2009, Sharma et al 2015).
In its original conception, the test is easy-to-
use, relatively fast and low cost, giving good
qualitative information where a simple yes/no
answer is sufficient.
However, this same test system has been
evolving to compete with more traditional
laboratory based platforms. To achieve the
requirements of greater assay accuracy and
sensitivity without losing its basic advantages
for PoC conditions, the actual strip test is
now at the core of instruments such as hand
held analyzers comprising sample handling,
reading and analysis modules. In brief, LFIA
is no longer at the cheap and simple end of
the spectrum in terms of diagnostic
technologies (Sharma et al, 2015).
The above trends have resulted in
continuous improvements over various
materials and reagents used in LFIA of which
Anteo Technologies’ Lateral Flow Coupling
Kit is one new development. Historically, gold
colloids have dominated in LFIA with
visualization of conjugated gold colloids as
they were captured on pre-defined lines on
the test strip.
Passive conjugation of antibodies on gold
colloids is conceptually simple but depending
on the antibody, can lead to difficulties such
as poor Ab functionality and loss of colloidal
stability resulting in the formation of
aggregates (Wong and Tse 2009). These
difficulties change with scale, and achieving
good uniformity during scale-up can often be
very challenging.
!
! Page !2
Using 200 nm Magnetic Particles
The use of 200 nm magnetic nanoparticles
activated with Anteo’s reagent for lateral
flow as an alternative to colloidal gold
addresses a number of potential
problematic situations encountered in the
development of new lateral flow assays.
For example, preparation of labeled
conjugates is simplified using magnetic
particles as they are easy to separate during
conjugation, washing and blocking steps,
with the additional benefit of inclusion of a
magnetic label for performing magnetic
immunoassays (MIA) (Shi et al 2015).
Alternatively, when used in conjunction with
sufficiently dark magnetic particles visual
detection alone can produce results rivaling
those of other colored particle types. The kit
contains Anteo’s Activation Reagent that can
be used to activate particle sizes from 200
nm to 1um that become immediately reactive
to antibodies once Anteo’s reagent is applied.!
Conventionally, conjugation of antibodies to
magnetic particles is commonly done by
covalent chemistry using reagents such as 1-
Ethyl-3-(3-dimethylaminopropyl)carbodiimide
(EDC) and N-hydroxysulfosuccinimide (S-
NHS). A reactive ester is formed on the
magnetic particles, which reacts with amino
groups on the protein.
Chemically, the reaction is simple but fraught
with multiple variables to consider when
trying to achieve optimum results. Conditions
that increase stability of the ester during
synthesis make it less reactive to amino
groups on the protein. Increasing ester
reactivity leads to side reactions. While
different approaches can mitigate many such
problems, these methods require pre-
activation of particles which cannot be stored
for any length of time.
The magnetic particles activated by Anteo’s
Activation Reagent are claimed to be stable
in storage and highly reactive to proteins with
other advantages such as ease of use, less
antibodies used, faster conjugation and
reduced aggregation amongst its benefits.
Using a hCG lateral flow assay previously
developed using colloidal gold at Diagnostic
Consulting Network (DCN), the Anteo
coupling kits were compared to covalently
conjugated magnetic particles using the
same assay materials (membrane, pads, and
antibodies) with the exception of the gold
colloids.
!
! Page !3
Method
The hCG assay has previously been
developed using colloidal gold at
Diagnostic Consulting Network (DCN),
Carlsbad, CA for internal use and
demonstration purposes. The same
assay materials (membrane, pads, and
antibodies) were used with Anteo’s
technology and covalently conjugated
magnetic particles.
All materials and tests were performed in
DCN’s laboratories using magnetic particles
conjugated with anti-hCG antibody using
Anteo’s protocol and DCN’s standard
covalent conjugation procedure. hCG
concentrations from 0 to 250mIU/ml were
prepared in 1%BSA, 0.1%Tw-20 1xPBS and
in negative pooled human urine. Tests were
performed using 100ul of sample and test
strips with the covalently conjugated and
Anteo’s conjugated particles dried onto a
conjugate pad. Five replicates for each
concentration were tested and results were
recorded 15 minutes after the addition of the
sample using DCN’s visual grading scale and
an ESE Quant (Qiagen Lake Constance,
Germany) reflectance reader.
The amount of antibody conjugated to the
different particles and the amount of particles
used per test were optimized to produce the
highest sensitivity. Anteo’s conjugate was
tested with an antibody loading of 25ug of
anti-hCG antibody per mg of particles at
0.05% solids. The covalently conjugated
particle was tested with an antibody loading
of 50ug of anti-hCG per mg particles at
0.05% solids.
Results
The results for the visual grades and the
values recorded by the reflectance reader,
including standard deviation and %CV are
shown in Table 1 for each of the two tests.
The limit of detection for the Anteo assay was
~10mIU/ml hCG in buffer and ~25mIU/ml in
urine for the visual and reader based results.
The limit of detection for the covalently
conjugated hCG assay was ~50mIU/mL in
buffer and ~100mIU/mL in urine. As a
reference the limit of the detection for the
optimized colloidal gold assay is at ~2.5mIU/
mL in buffer and ~25mIU/mL in urine.
NOTE: A visual grade of 2 or higher at the
test line is considered positive, while the
cutoff for the reader is at 30mV.
!
! Page !4
Table 1: Visual and Reader results of the Anteo’s technology and covalently conjugated
magnetic particle based hCG assays in buffer.

!
Figure 1: Graph of the results from the Anteo and covalently conjugated magnetic particle hCG
assays in buffer measured by a Qiagen reader.!
!
! Page !5
Figures 3 and 4: Pictures of lateral flow strips run with the Anteo and covalently conjugated
magnetic particles with hCG diluted in buffer.
Figure 3: Anteo magnetic particle test strips run with hCG diluted in buffer
Figure 4: Covalently conjugated magnetic particle test strips run with hCG diluted in buffer.
Table 2: Visual and reader results of the Anteo and covalently conjugated magnetic particle
hCG assays with urine.
!
! Page !6
250mIU/mL 100mIU/mL 50mIU/mL 25mIU/mL 10mIU/mL 5mIU/mL 0mIU/mL
250mIU/mL 100mIU/mL 50mIU/mL 25mIU/mL 10mIU/mL 5mIU/mL 0mIU/mL
!
Figures 5 and 6: Graphs of the reader results from the Anteo and covalently conjugated
magnetic particle hCG assays in urine.
!
! Page !7
Figures 7 and 8: Pictures of the Anteo and covalently conjugated magnetic particle hCG
assays run with hCG diluted in urine.
Figure 7: Anteo magnetic particle test strips run with hCG diluted with urine
Figure 8: Covalently conjugated magnetic particle test strips run with hCG diluted with urine
Conclusions
The direct comparison of the two tests
show:
The Anteo technology based hCG assay is 5
times more sensitive than the covalently
conjugated magnetic particle based test
using the same critical reagents. The
estimated limit of detection for the Anteo
assay is ~10mIU/mL of hCG in buffer and
~25mIU/mL of hCG in urine while the
covalently conjugated hCG assay has limits
of detection of ~50 and 100 mIU/ml of hCG in
buffer and urine respectively. The hCG assay
using the Anteo particles used half the
amount of antibody to achieve better
sensitivity than the covalently conjugated
assay.
The variability of the Anteo technology based
hCG assay is lower than the covalently
conjugated hCG assay. There is also less
non-specific binding with the Anteo particles.!
The Anteo technology based magnetic
particle hCG lateral flow assay produced a
comparable limit of detection to the optimized
colloidal gold reference assay of ~25mIU/mL
in urine. This confirms that particles
conjugated with the Anteo Activation Reagent
can rival traditionally used gold colloids in
lateral flow applications and bring with it
benefits associated with cost, handling and
!
! Page !8
250mIU/mL 100mIU/mL 50mIU/mL 25mIU/mL 10mIU/mL 5mIU/mL 0mIU/mL
250mIU/mL 100mIU/mL 50mIU/mL 25mIU/mL 10mIU/mL 5mIU/mL 0mIU/mL
!
Anteo Products Lateral Flow
Anteo's Lateral Flow Coupling Kit makes conducting
lateral flow tests with magnetic beads easier and
more flexible. Magnetic separation removes the
need to perform centrifugation and filtration
concentration.
Advantages
• An alternative to EDC chemistry the Anteo Kit is
easy to use
• Less hazardous chemicals to work with, Anteo’s
Activation Reagent is water based
• Flexibility: you can bind any antibody subclass to your surface
• Promotes monolayer binding which increases antibody functionality
• Reduces aggregation
• Low background noise
The Anteo Lateral Flow Coupling Kit is launching at AACC 2016. For more information on this
kit email sales@anteotech.com
About Anteo Technologies
Anteo Technologies is transforming the way scientists work by providing new, powerful tools
developed with their patented nanoscale engineered Activation Reagent. This Reagent
significantly advances traditional protein coupling methods, leading to increased assay
sensitivity.
Anteo’s Activation Reagent is designed by using cross-linking metal ions to enable gentle yet
secure multi-point binding of biomolecules to synthetic materials. The Reagent also delivers
solutions to the challenges of establishing highly functional biological interfaces between fragile
biomolecules, and synthetic and often incompatible hydrophilic/hydrophobic materials such as
silica, polystyrene, COC/COP, metals and ceramics.
Water based, Anteo’s Activation Reagent is easy to use. Simply pipette the reagent onto your
surface to activate. Hands on time for surface activation and coupling is 60 minutes.
Find out more at anteotech.com
!
! Page !9
References!
Sharma S, Zapatero-Rodríguez J, Estrela P, O'Kennedy R. Point-of-Care Diagnostics in Low
Resource Settings: Present Status and Future Role of Microfluidics. Biosensors (Basel). 2015
Aug 13;5(3):577-601.
Shi L, Wu F, Wen Y, Zhao F, Xiang J, Ma L. A novel method to detect Listeria monocytogenes
via superparamagnetic lateral flow immunoassay. Anal Bioanal Chem. 2015 Jan;407(2):529-35.
Wong RC and Tse HY. Lateral flow immunoassays, Humana press, DOI
10.1007/978-1-59745-240-3, 2009
!
! Page !10

Más contenido relacionado

La actualidad más candente

MALDI-TOF MS Based Discovery Workflows: A Fully Automated, Bottom-Up Approach
MALDI-TOF MS Based Discovery Workflows: A Fully Automated, Bottom-Up ApproachMALDI-TOF MS Based Discovery Workflows: A Fully Automated, Bottom-Up Approach
MALDI-TOF MS Based Discovery Workflows: A Fully Automated, Bottom-Up ApproachShimadzu Scientific Instruments
 
Synthesis of graphene oxide-TiO2 nanocomposite as an adsorbent for the enrich...
Synthesis of graphene oxide-TiO2 nanocomposite as an adsorbent for the enrich...Synthesis of graphene oxide-TiO2 nanocomposite as an adsorbent for the enrich...
Synthesis of graphene oxide-TiO2 nanocomposite as an adsorbent for the enrich...Nanomedicine Journal (NMJ)
 
A Comparison of Multimodal Chromatographic Resin: Protein Binding & Selectivity
A Comparison of Multimodal Chromatographic Resin: Protein Binding & SelectivityA Comparison of Multimodal Chromatographic Resin: Protein Binding & Selectivity
A Comparison of Multimodal Chromatographic Resin: Protein Binding & SelectivityKBI Biopharma
 
Enhanced electrophoretic resolution of monosulfate GAG disacchride isomers on...
Enhanced electrophoretic resolution of monosulfate GAG disacchride isomers on...Enhanced electrophoretic resolution of monosulfate GAG disacchride isomers on...
Enhanced electrophoretic resolution of monosulfate GAG disacchride isomers on...Yong Zhang
 
Isolation purification technique for Therapeutic Bio molecules : Protein-Peptide
Isolation purification technique for Therapeutic Bio molecules : Protein-PeptideIsolation purification technique for Therapeutic Bio molecules : Protein-Peptide
Isolation purification technique for Therapeutic Bio molecules : Protein-PeptideRohit Gurav
 
Fragment based drug design complementary tool for drug design
Fragment based drug design  complementary tool for drug designFragment based drug design  complementary tool for drug design
Fragment based drug design complementary tool for drug designNIPER hyderabad
 
Nihms809455
Nihms809455Nihms809455
Nihms809455zoyapeer
 
J. Mater. Chem. A, 2015, 3, 20408–20415
J. Mater. Chem. A, 2015, 3, 20408–20415J. Mater. Chem. A, 2015, 3, 20408–20415
J. Mater. Chem. A, 2015, 3, 20408–20415Sedigheh Abedi
 
Molecular Forecaster Services
Molecular Forecaster ServicesMolecular Forecaster Services
Molecular Forecaster Servicestherriee
 
Idealp Pharma Hits & Leads Optimisation Case Study 1
Idealp Pharma Hits & Leads Optimisation Case Study 1Idealp Pharma Hits & Leads Optimisation Case Study 1
Idealp Pharma Hits & Leads Optimisation Case Study 1MehdiChelbi
 

La actualidad más candente (20)

Polysaccharide Hydrogels: a versatile tool for biomedical and pharmaceutical ...
Polysaccharide Hydrogels: a versatile tool for biomedical and pharmaceutical ...Polysaccharide Hydrogels: a versatile tool for biomedical and pharmaceutical ...
Polysaccharide Hydrogels: a versatile tool for biomedical and pharmaceutical ...
 
MALDI-TOF MS Based Discovery Workflows: A Fully Automated, Bottom-Up Approach
MALDI-TOF MS Based Discovery Workflows: A Fully Automated, Bottom-Up ApproachMALDI-TOF MS Based Discovery Workflows: A Fully Automated, Bottom-Up Approach
MALDI-TOF MS Based Discovery Workflows: A Fully Automated, Bottom-Up Approach
 
Molecular Imprinting
Molecular ImprintingMolecular Imprinting
Molecular Imprinting
 
Niranjan seminar
Niranjan seminarNiranjan seminar
Niranjan seminar
 
Final PPT
Final PPTFinal PPT
Final PPT
 
IInd research paper
IInd research paperIInd research paper
IInd research paper
 
Bioinspired strategies for bone regeneration.
Bioinspired strategies for bone regeneration.Bioinspired strategies for bone regeneration.
Bioinspired strategies for bone regeneration.
 
Synthesis of graphene oxide-TiO2 nanocomposite as an adsorbent for the enrich...
Synthesis of graphene oxide-TiO2 nanocomposite as an adsorbent for the enrich...Synthesis of graphene oxide-TiO2 nanocomposite as an adsorbent for the enrich...
Synthesis of graphene oxide-TiO2 nanocomposite as an adsorbent for the enrich...
 
Synthesis, characterization and biological evaluation of benzimidazoles and i...
Synthesis, characterization and biological evaluation of benzimidazoles and i...Synthesis, characterization and biological evaluation of benzimidazoles and i...
Synthesis, characterization and biological evaluation of benzimidazoles and i...
 
A Comparison of Multimodal Chromatographic Resin: Protein Binding & Selectivity
A Comparison of Multimodal Chromatographic Resin: Protein Binding & SelectivityA Comparison of Multimodal Chromatographic Resin: Protein Binding & Selectivity
A Comparison of Multimodal Chromatographic Resin: Protein Binding & Selectivity
 
Enhanced electrophoretic resolution of monosulfate GAG disacchride isomers on...
Enhanced electrophoretic resolution of monosulfate GAG disacchride isomers on...Enhanced electrophoretic resolution of monosulfate GAG disacchride isomers on...
Enhanced electrophoretic resolution of monosulfate GAG disacchride isomers on...
 
Isolation purification technique for Therapeutic Bio molecules : Protein-Peptide
Isolation purification technique for Therapeutic Bio molecules : Protein-PeptideIsolation purification technique for Therapeutic Bio molecules : Protein-Peptide
Isolation purification technique for Therapeutic Bio molecules : Protein-Peptide
 
Fragment based drug design complementary tool for drug design
Fragment based drug design  complementary tool for drug designFragment based drug design  complementary tool for drug design
Fragment based drug design complementary tool for drug design
 
Poster presentation
Poster presentationPoster presentation
Poster presentation
 
Nihms809455
Nihms809455Nihms809455
Nihms809455
 
J. Mater. Chem. A, 2015, 3, 20408–20415
J. Mater. Chem. A, 2015, 3, 20408–20415J. Mater. Chem. A, 2015, 3, 20408–20415
J. Mater. Chem. A, 2015, 3, 20408–20415
 
Tracer techniques
Tracer techniquesTracer techniques
Tracer techniques
 
Tracer technique
Tracer techniqueTracer technique
Tracer technique
 
Molecular Forecaster Services
Molecular Forecaster ServicesMolecular Forecaster Services
Molecular Forecaster Services
 
Idealp Pharma Hits & Leads Optimisation Case Study 1
Idealp Pharma Hits & Leads Optimisation Case Study 1Idealp Pharma Hits & Leads Optimisation Case Study 1
Idealp Pharma Hits & Leads Optimisation Case Study 1
 

Destacado

Anteo Diagnostics Quarterly Report - (ASX : ADO)
Anteo Diagnostics Quarterly Report - (ASX : ADO)Anteo Diagnostics Quarterly Report - (ASX : ADO)
Anteo Diagnostics Quarterly Report - (ASX : ADO)Matt Sanderson
 
Anteo Diagnostics acquires Diasource
Anteo Diagnostics acquires Diasource Anteo Diagnostics acquires Diasource
Anteo Diagnostics acquires Diasource Matt Sanderson
 
Anteo Technologies Investor Presentation
Anteo Technologies Investor PresentationAnteo Technologies Investor Presentation
Anteo Technologies Investor PresentationMatt Sanderson
 
Anteo Diagnostics & Atomo Troponin Lateral Flow test
Anteo Diagnostics & Atomo Troponin Lateral Flow testAnteo Diagnostics & Atomo Troponin Lateral Flow test
Anteo Diagnostics & Atomo Troponin Lateral Flow testMatt Sanderson
 
Anteo Diagnostics to acquire Global Diagnostics Company - Diasource
Anteo Diagnostics to acquire Global Diagnostics Company - DiasourceAnteo Diagnostics to acquire Global Diagnostics Company - Diasource
Anteo Diagnostics to acquire Global Diagnostics Company - DiasourceMatt Sanderson
 
Novel methods in Bioseparations
Novel methods in BioseparationsNovel methods in Bioseparations
Novel methods in BioseparationsMatt Sanderson
 
Multiplex energy metabolism
Multiplex energy metabolismMultiplex energy metabolism
Multiplex energy metabolismMatt Sanderson
 
Anteo Diagnostics Research Report November 2015
Anteo Diagnostics Research Report November 2015Anteo Diagnostics Research Report November 2015
Anteo Diagnostics Research Report November 2015Matt Sanderson
 
Silicon battery Presentation - Lithium Ion
Silicon battery Presentation - Lithium IonSilicon battery Presentation - Lithium Ion
Silicon battery Presentation - Lithium IonMatt Sanderson
 
Multiplexing and Arraying in Lateral Flow Assays
Multiplexing and Arraying in Lateral Flow AssaysMultiplexing and Arraying in Lateral Flow Assays
Multiplexing and Arraying in Lateral Flow AssaysBrendan O'Farrell
 
ASX Australian Biotech Market cap list
ASX Australian Biotech Market cap listASX Australian Biotech Market cap list
ASX Australian Biotech Market cap listMatt Sanderson
 
Achieving 5 x More Sensitivity And Less Variation In Results Than Covalent Ch...
Achieving 5 x More Sensitivity And Less Variation In Results Than Covalent Ch...Achieving 5 x More Sensitivity And Less Variation In Results Than Covalent Ch...
Achieving 5 x More Sensitivity And Less Variation In Results Than Covalent Ch...Anteo Technologies
 

Destacado (13)

Anteo Diagnostics Quarterly Report - (ASX : ADO)
Anteo Diagnostics Quarterly Report - (ASX : ADO)Anteo Diagnostics Quarterly Report - (ASX : ADO)
Anteo Diagnostics Quarterly Report - (ASX : ADO)
 
Anteo Diagnostics acquires Diasource
Anteo Diagnostics acquires Diasource Anteo Diagnostics acquires Diasource
Anteo Diagnostics acquires Diasource
 
Anteo Technologies Investor Presentation
Anteo Technologies Investor PresentationAnteo Technologies Investor Presentation
Anteo Technologies Investor Presentation
 
Anteo Diagnostics & Atomo Troponin Lateral Flow test
Anteo Diagnostics & Atomo Troponin Lateral Flow testAnteo Diagnostics & Atomo Troponin Lateral Flow test
Anteo Diagnostics & Atomo Troponin Lateral Flow test
 
Anteo Diagnostics to acquire Global Diagnostics Company - Diasource
Anteo Diagnostics to acquire Global Diagnostics Company - DiasourceAnteo Diagnostics to acquire Global Diagnostics Company - Diasource
Anteo Diagnostics to acquire Global Diagnostics Company - Diasource
 
Novel methods in Bioseparations
Novel methods in BioseparationsNovel methods in Bioseparations
Novel methods in Bioseparations
 
Multiplex energy metabolism
Multiplex energy metabolismMultiplex energy metabolism
Multiplex energy metabolism
 
Anteo Diagnostics Research Report November 2015
Anteo Diagnostics Research Report November 2015Anteo Diagnostics Research Report November 2015
Anteo Diagnostics Research Report November 2015
 
DCN Company Profile 2015
DCN Company Profile 2015DCN Company Profile 2015
DCN Company Profile 2015
 
Silicon battery Presentation - Lithium Ion
Silicon battery Presentation - Lithium IonSilicon battery Presentation - Lithium Ion
Silicon battery Presentation - Lithium Ion
 
Multiplexing and Arraying in Lateral Flow Assays
Multiplexing and Arraying in Lateral Flow AssaysMultiplexing and Arraying in Lateral Flow Assays
Multiplexing and Arraying in Lateral Flow Assays
 
ASX Australian Biotech Market cap list
ASX Australian Biotech Market cap listASX Australian Biotech Market cap list
ASX Australian Biotech Market cap list
 
Achieving 5 x More Sensitivity And Less Variation In Results Than Covalent Ch...
Achieving 5 x More Sensitivity And Less Variation In Results Than Covalent Ch...Achieving 5 x More Sensitivity And Less Variation In Results Than Covalent Ch...
Achieving 5 x More Sensitivity And Less Variation In Results Than Covalent Ch...
 

Similar a 5x More Sensitive Lateral Flow Using Activated Magnetic Particles

Lateral Flow: Making Magnetic Particles A Viable And Easier To Use Replacemen...
Lateral Flow: Making Magnetic Particles A Viable And Easier To Use Replacemen...Lateral Flow: Making Magnetic Particles A Viable And Easier To Use Replacemen...
Lateral Flow: Making Magnetic Particles A Viable And Easier To Use Replacemen...Anteo Technologies
 
Antibody Coupling - Single Coupling Approach To Bind Antibodies Diverse Speci...
Antibody Coupling - Single Coupling Approach To Bind Antibodies Diverse Speci...Antibody Coupling - Single Coupling Approach To Bind Antibodies Diverse Speci...
Antibody Coupling - Single Coupling Approach To Bind Antibodies Diverse Speci...Anteo Technologies
 
biotechnology of aminophenol PhD defenseppt.ppt
biotechnology of aminophenol PhD defenseppt.pptbiotechnology of aminophenol PhD defenseppt.ppt
biotechnology of aminophenol PhD defenseppt.pptmisgana18
 
Design and development of nanomaterials for biomolecular detection and cancer...
Design and development of nanomaterials for biomolecular detection and cancer...Design and development of nanomaterials for biomolecular detection and cancer...
Design and development of nanomaterials for biomolecular detection and cancer...Arun kumar
 
Enzyme based Biosensor for pesticide Detection
Enzyme based Biosensor for pesticide DetectionEnzyme based Biosensor for pesticide Detection
Enzyme based Biosensor for pesticide DetectionSubhasis Sarkar
 
ELECTROCHEMICAL SENSORS
ELECTROCHEMICAL SENSORS ELECTROCHEMICAL SENSORS
ELECTROCHEMICAL SENSORS Lavaraj Devkota
 
laser-induced fluorescence orUV detectors for analysis of EPO and Tobramycin ...
laser-induced fluorescence orUV detectors for analysis of EPO and Tobramycin ...laser-induced fluorescence orUV detectors for analysis of EPO and Tobramycin ...
laser-induced fluorescence orUV detectors for analysis of EPO and Tobramycin ...Wael Ebied
 
Elisa and Electrophoresis
Elisa   and   Electrophoresis Elisa   and   Electrophoresis
Elisa and Electrophoresis PrernaBurathoki
 
Kott_Applied spectrocopy_2014_68_1108
Kott_Applied spectrocopy_2014_68_1108Kott_Applied spectrocopy_2014_68_1108
Kott_Applied spectrocopy_2014_68_1108Laila Kott
 
Kott applied spectrocopy 2014_68_1108
Kott applied spectrocopy 2014_68_1108Kott applied spectrocopy 2014_68_1108
Kott applied spectrocopy 2014_68_1108Laila Kott
 
research article (Recovered)
research article (Recovered)research article (Recovered)
research article (Recovered)Taylor Bogan
 
Counter immunoelectrophoresis
Counter immunoelectrophoresisCounter immunoelectrophoresis
Counter immunoelectrophoresisAhmedRiyadh17
 
New approaches for stable isotope ratio measurements iaea tecdoc 1247
New approaches for stable isotope ratio measurements   iaea tecdoc 1247New approaches for stable isotope ratio measurements   iaea tecdoc 1247
New approaches for stable isotope ratio measurements iaea tecdoc 1247Mahbubul Hassan
 
Two dimensional gel electrophoresis
Two dimensional gel electrophoresisTwo dimensional gel electrophoresis
Two dimensional gel electrophoresisShyam K Uthaman
 
Breathsafe- a biosensor designed for detecting H1N1
Breathsafe- a biosensor designed for detecting H1N1Breathsafe- a biosensor designed for detecting H1N1
Breathsafe- a biosensor designed for detecting H1N1Gayathri Vijayakumar
 
Designing A Whole New Ultra Poreux Cellulose Based Material
Designing A Whole New Ultra Poreux Cellulose Based MaterialDesigning A Whole New Ultra Poreux Cellulose Based Material
Designing A Whole New Ultra Poreux Cellulose Based MaterialTiffany Graham
 
Immunological Assays;RIA and ELISA.pjsjsptx
Immunological Assays;RIA and ELISA.pjsjsptxImmunological Assays;RIA and ELISA.pjsjsptx
Immunological Assays;RIA and ELISA.pjsjsptxanshikabhatnagar1299
 

Similar a 5x More Sensitive Lateral Flow Using Activated Magnetic Particles (20)

Lateral Flow: Making Magnetic Particles A Viable And Easier To Use Replacemen...
Lateral Flow: Making Magnetic Particles A Viable And Easier To Use Replacemen...Lateral Flow: Making Magnetic Particles A Viable And Easier To Use Replacemen...
Lateral Flow: Making Magnetic Particles A Viable And Easier To Use Replacemen...
 
Antibody Coupling - Single Coupling Approach To Bind Antibodies Diverse Speci...
Antibody Coupling - Single Coupling Approach To Bind Antibodies Diverse Speci...Antibody Coupling - Single Coupling Approach To Bind Antibodies Diverse Speci...
Antibody Coupling - Single Coupling Approach To Bind Antibodies Diverse Speci...
 
Abstract
AbstractAbstract
Abstract
 
biotechnology of aminophenol PhD defenseppt.ppt
biotechnology of aminophenol PhD defenseppt.pptbiotechnology of aminophenol PhD defenseppt.ppt
biotechnology of aminophenol PhD defenseppt.ppt
 
Design and development of nanomaterials for biomolecular detection and cancer...
Design and development of nanomaterials for biomolecular detection and cancer...Design and development of nanomaterials for biomolecular detection and cancer...
Design and development of nanomaterials for biomolecular detection and cancer...
 
Enzyme based Biosensor for pesticide Detection
Enzyme based Biosensor for pesticide DetectionEnzyme based Biosensor for pesticide Detection
Enzyme based Biosensor for pesticide Detection
 
ELECTROCHEMICAL SENSORS
ELECTROCHEMICAL SENSORS ELECTROCHEMICAL SENSORS
ELECTROCHEMICAL SENSORS
 
laser-induced fluorescence orUV detectors for analysis of EPO and Tobramycin ...
laser-induced fluorescence orUV detectors for analysis of EPO and Tobramycin ...laser-induced fluorescence orUV detectors for analysis of EPO and Tobramycin ...
laser-induced fluorescence orUV detectors for analysis of EPO and Tobramycin ...
 
Nanobiosensors
NanobiosensorsNanobiosensors
Nanobiosensors
 
Elisa and Electrophoresis
Elisa   and   Electrophoresis Elisa   and   Electrophoresis
Elisa and Electrophoresis
 
Kott_Applied spectrocopy_2014_68_1108
Kott_Applied spectrocopy_2014_68_1108Kott_Applied spectrocopy_2014_68_1108
Kott_Applied spectrocopy_2014_68_1108
 
Kott applied spectrocopy 2014_68_1108
Kott applied spectrocopy 2014_68_1108Kott applied spectrocopy 2014_68_1108
Kott applied spectrocopy 2014_68_1108
 
RBC Poster Final
RBC Poster FinalRBC Poster Final
RBC Poster Final
 
research article (Recovered)
research article (Recovered)research article (Recovered)
research article (Recovered)
 
Counter immunoelectrophoresis
Counter immunoelectrophoresisCounter immunoelectrophoresis
Counter immunoelectrophoresis
 
New approaches for stable isotope ratio measurements iaea tecdoc 1247
New approaches for stable isotope ratio measurements   iaea tecdoc 1247New approaches for stable isotope ratio measurements   iaea tecdoc 1247
New approaches for stable isotope ratio measurements iaea tecdoc 1247
 
Two dimensional gel electrophoresis
Two dimensional gel electrophoresisTwo dimensional gel electrophoresis
Two dimensional gel electrophoresis
 
Breathsafe- a biosensor designed for detecting H1N1
Breathsafe- a biosensor designed for detecting H1N1Breathsafe- a biosensor designed for detecting H1N1
Breathsafe- a biosensor designed for detecting H1N1
 
Designing A Whole New Ultra Poreux Cellulose Based Material
Designing A Whole New Ultra Poreux Cellulose Based MaterialDesigning A Whole New Ultra Poreux Cellulose Based Material
Designing A Whole New Ultra Poreux Cellulose Based Material
 
Immunological Assays;RIA and ELISA.pjsjsptx
Immunological Assays;RIA and ELISA.pjsjsptxImmunological Assays;RIA and ELISA.pjsjsptx
Immunological Assays;RIA and ELISA.pjsjsptx
 

Más de Matt Sanderson

A Novel Chemistry to Couple Antibodies and Challenging Proteins to Luminex M...
A Novel Chemistry to  Couple Antibodies and Challenging Proteins to Luminex M...A Novel Chemistry to  Couple Antibodies and Challenging Proteins to Luminex M...
A Novel Chemistry to Couple Antibodies and Challenging Proteins to Luminex M...Matt Sanderson
 
Anteo Technologies - New York presentation 2015
Anteo Technologies - New York presentation 2015Anteo Technologies - New York presentation 2015
Anteo Technologies - New York presentation 2015Matt Sanderson
 
High Sensitivity Troponin
High Sensitivity TroponinHigh Sensitivity Troponin
High Sensitivity TroponinMatt Sanderson
 
Aus Biotech 2014 - Anteo Diagnostics
Aus Biotech 2014 - Anteo DiagnosticsAus Biotech 2014 - Anteo Diagnostics
Aus Biotech 2014 - Anteo DiagnosticsMatt Sanderson
 
Anteo Diagnostics Microcap Presentation October 2013
Anteo Diagnostics Microcap Presentation October 2013Anteo Diagnostics Microcap Presentation October 2013
Anteo Diagnostics Microcap Presentation October 2013Matt Sanderson
 
Anteo Diagnostics ASX Spotlight Presentation March 2014
Anteo Diagnostics ASX Spotlight Presentation March 2014Anteo Diagnostics ASX Spotlight Presentation March 2014
Anteo Diagnostics ASX Spotlight Presentation March 2014Matt Sanderson
 

Más de Matt Sanderson (7)

A Novel Chemistry to Couple Antibodies and Challenging Proteins to Luminex M...
A Novel Chemistry to  Couple Antibodies and Challenging Proteins to Luminex M...A Novel Chemistry to  Couple Antibodies and Challenging Proteins to Luminex M...
A Novel Chemistry to Couple Antibodies and Challenging Proteins to Luminex M...
 
Anteo Technologies - New York presentation 2015
Anteo Technologies - New York presentation 2015Anteo Technologies - New York presentation 2015
Anteo Technologies - New York presentation 2015
 
Top 10 IVD Companies
Top 10 IVD CompaniesTop 10 IVD Companies
Top 10 IVD Companies
 
High Sensitivity Troponin
High Sensitivity TroponinHigh Sensitivity Troponin
High Sensitivity Troponin
 
Aus Biotech 2014 - Anteo Diagnostics
Aus Biotech 2014 - Anteo DiagnosticsAus Biotech 2014 - Anteo Diagnostics
Aus Biotech 2014 - Anteo Diagnostics
 
Anteo Diagnostics Microcap Presentation October 2013
Anteo Diagnostics Microcap Presentation October 2013Anteo Diagnostics Microcap Presentation October 2013
Anteo Diagnostics Microcap Presentation October 2013
 
Anteo Diagnostics ASX Spotlight Presentation March 2014
Anteo Diagnostics ASX Spotlight Presentation March 2014Anteo Diagnostics ASX Spotlight Presentation March 2014
Anteo Diagnostics ASX Spotlight Presentation March 2014
 

Último

Russian Call Girls Hyderabad Saloni 9907093804 Independent Escort Service Hyd...
Russian Call Girls Hyderabad Saloni 9907093804 Independent Escort Service Hyd...Russian Call Girls Hyderabad Saloni 9907093804 Independent Escort Service Hyd...
Russian Call Girls Hyderabad Saloni 9907093804 Independent Escort Service Hyd...delhimodelshub1
 
Call Girl Bangalore Aashi 7001305949 Independent Escort Service Bangalore
Call Girl Bangalore Aashi 7001305949 Independent Escort Service BangaloreCall Girl Bangalore Aashi 7001305949 Independent Escort Service Bangalore
Call Girl Bangalore Aashi 7001305949 Independent Escort Service Bangalorenarwatsonia7
 
SARS (SEVERE ACUTE RESPIRATORY SYNDROME).pdf
SARS (SEVERE ACUTE RESPIRATORY SYNDROME).pdfSARS (SEVERE ACUTE RESPIRATORY SYNDROME).pdf
SARS (SEVERE ACUTE RESPIRATORY SYNDROME).pdfDolisha Warbi
 
Call Girls in Adil Nagar 7001305949 Free Delivery at Your Door Model
Call Girls in Adil Nagar 7001305949 Free Delivery at Your Door ModelCall Girls in Adil Nagar 7001305949 Free Delivery at Your Door Model
Call Girls in Adil Nagar 7001305949 Free Delivery at Your Door ModelCall Girls Lucknow
 
Air-Hostess Call Girls Shanti Nagar - Call 7001305949 Rs-3500 with A/C Room C...
Air-Hostess Call Girls Shanti Nagar - Call 7001305949 Rs-3500 with A/C Room C...Air-Hostess Call Girls Shanti Nagar - Call 7001305949 Rs-3500 with A/C Room C...
Air-Hostess Call Girls Shanti Nagar - Call 7001305949 Rs-3500 with A/C Room C...narwatsonia7
 
Russian Call Girls Ajmeri Gate | 9711199171 | High Profile -New Model -Availa...
Russian Call Girls Ajmeri Gate | 9711199171 | High Profile -New Model -Availa...Russian Call Girls Ajmeri Gate | 9711199171 | High Profile -New Model -Availa...
Russian Call Girls Ajmeri Gate | 9711199171 | High Profile -New Model -Availa...sandeepkumar69420
 
Call Girls Dwarka 9999965857 Cheap & Best with original Photos
Call Girls Dwarka 9999965857 Cheap & Best with original PhotosCall Girls Dwarka 9999965857 Cheap & Best with original Photos
Call Girls Dwarka 9999965857 Cheap & Best with original Photosparshadkalavatidevi7
 
Russian Call Girl Chandapura Dommasandra Road - 7001305949 Escorts Service 50...
Russian Call Girl Chandapura Dommasandra Road - 7001305949 Escorts Service 50...Russian Call Girl Chandapura Dommasandra Road - 7001305949 Escorts Service 50...
Russian Call Girl Chandapura Dommasandra Road - 7001305949 Escorts Service 50...narwatsonia7
 
Pregnancy and Breastfeeding Dental Considerations.pptx
Pregnancy and Breastfeeding Dental Considerations.pptxPregnancy and Breastfeeding Dental Considerations.pptx
Pregnancy and Breastfeeding Dental Considerations.pptxcrosalofton
 
Gurgaon DLF Phase 5 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Fe...
Gurgaon DLF Phase 5 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Fe...Gurgaon DLF Phase 5 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Fe...
Gurgaon DLF Phase 5 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Fe...ggsonu500
 
Disaster Management Cycle (DMC)| Ms. Pooja Sharma , Department of Hospital A...
Disaster Management Cycle (DMC)| Ms. Pooja Sharma , Department of  Hospital A...Disaster Management Cycle (DMC)| Ms. Pooja Sharma , Department of  Hospital A...
Disaster Management Cycle (DMC)| Ms. Pooja Sharma , Department of Hospital A...Era University , Lucknow
 
MVP Health Care City of Schenectady Presentation
MVP Health Care City of Schenectady PresentationMVP Health Care City of Schenectady Presentation
MVP Health Care City of Schenectady PresentationMVP Health Care
 
Models Call Girls Electronic City | 7001305949 At Low Cost Cash Payment Booking
Models Call Girls Electronic City | 7001305949 At Low Cost Cash Payment BookingModels Call Girls Electronic City | 7001305949 At Low Cost Cash Payment Booking
Models Call Girls Electronic City | 7001305949 At Low Cost Cash Payment Bookingnarwatsonia7
 
Call Girls Gurgaon Vani 9999965857 Independent Escort Service Gurgaon
Call Girls Gurgaon Vani 9999965857 Independent Escort Service GurgaonCall Girls Gurgaon Vani 9999965857 Independent Escort Service Gurgaon
Call Girls Gurgaon Vani 9999965857 Independent Escort Service Gurgaonnitachopra
 
2024 HCAT Healthcare Technology Insights
2024 HCAT Healthcare Technology Insights2024 HCAT Healthcare Technology Insights
2024 HCAT Healthcare Technology InsightsHealth Catalyst
 
Gurgaon iffco chowk 🔝 Call Girls Service 🔝 ( 8264348440 ) unlimited hard sex ...
Gurgaon iffco chowk 🔝 Call Girls Service 🔝 ( 8264348440 ) unlimited hard sex ...Gurgaon iffco chowk 🔝 Call Girls Service 🔝 ( 8264348440 ) unlimited hard sex ...
Gurgaon iffco chowk 🔝 Call Girls Service 🔝 ( 8264348440 ) unlimited hard sex ...callgirlsinsaket2024
 
Call Girl Service ITPL - [ Cash on Delivery ] Contact 7001305949 Escorts Service
Call Girl Service ITPL - [ Cash on Delivery ] Contact 7001305949 Escorts ServiceCall Girl Service ITPL - [ Cash on Delivery ] Contact 7001305949 Escorts Service
Call Girl Service ITPL - [ Cash on Delivery ] Contact 7001305949 Escorts Servicenarwatsonia7
 
EMS and Extrication: Coordinating Critical Care
EMS and Extrication: Coordinating Critical CareEMS and Extrication: Coordinating Critical Care
EMS and Extrication: Coordinating Critical CareRommie Duckworth
 
Russian Escorts Delhi | 9711199171 | all area service available
Russian Escorts Delhi | 9711199171 | all area service availableRussian Escorts Delhi | 9711199171 | all area service available
Russian Escorts Delhi | 9711199171 | all area service availablesandeepkumar69420
 

Último (20)

Russian Call Girls Hyderabad Saloni 9907093804 Independent Escort Service Hyd...
Russian Call Girls Hyderabad Saloni 9907093804 Independent Escort Service Hyd...Russian Call Girls Hyderabad Saloni 9907093804 Independent Escort Service Hyd...
Russian Call Girls Hyderabad Saloni 9907093804 Independent Escort Service Hyd...
 
Call Girl Bangalore Aashi 7001305949 Independent Escort Service Bangalore
Call Girl Bangalore Aashi 7001305949 Independent Escort Service BangaloreCall Girl Bangalore Aashi 7001305949 Independent Escort Service Bangalore
Call Girl Bangalore Aashi 7001305949 Independent Escort Service Bangalore
 
SARS (SEVERE ACUTE RESPIRATORY SYNDROME).pdf
SARS (SEVERE ACUTE RESPIRATORY SYNDROME).pdfSARS (SEVERE ACUTE RESPIRATORY SYNDROME).pdf
SARS (SEVERE ACUTE RESPIRATORY SYNDROME).pdf
 
Call Girls in Adil Nagar 7001305949 Free Delivery at Your Door Model
Call Girls in Adil Nagar 7001305949 Free Delivery at Your Door ModelCall Girls in Adil Nagar 7001305949 Free Delivery at Your Door Model
Call Girls in Adil Nagar 7001305949 Free Delivery at Your Door Model
 
Air-Hostess Call Girls Shanti Nagar - Call 7001305949 Rs-3500 with A/C Room C...
Air-Hostess Call Girls Shanti Nagar - Call 7001305949 Rs-3500 with A/C Room C...Air-Hostess Call Girls Shanti Nagar - Call 7001305949 Rs-3500 with A/C Room C...
Air-Hostess Call Girls Shanti Nagar - Call 7001305949 Rs-3500 with A/C Room C...
 
Russian Call Girls Ajmeri Gate | 9711199171 | High Profile -New Model -Availa...
Russian Call Girls Ajmeri Gate | 9711199171 | High Profile -New Model -Availa...Russian Call Girls Ajmeri Gate | 9711199171 | High Profile -New Model -Availa...
Russian Call Girls Ajmeri Gate | 9711199171 | High Profile -New Model -Availa...
 
Call Girls Dwarka 9999965857 Cheap & Best with original Photos
Call Girls Dwarka 9999965857 Cheap & Best with original PhotosCall Girls Dwarka 9999965857 Cheap & Best with original Photos
Call Girls Dwarka 9999965857 Cheap & Best with original Photos
 
Russian Call Girl Chandapura Dommasandra Road - 7001305949 Escorts Service 50...
Russian Call Girl Chandapura Dommasandra Road - 7001305949 Escorts Service 50...Russian Call Girl Chandapura Dommasandra Road - 7001305949 Escorts Service 50...
Russian Call Girl Chandapura Dommasandra Road - 7001305949 Escorts Service 50...
 
Pregnancy and Breastfeeding Dental Considerations.pptx
Pregnancy and Breastfeeding Dental Considerations.pptxPregnancy and Breastfeeding Dental Considerations.pptx
Pregnancy and Breastfeeding Dental Considerations.pptx
 
Gurgaon DLF Phase 5 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Fe...
Gurgaon DLF Phase 5 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Fe...Gurgaon DLF Phase 5 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Fe...
Gurgaon DLF Phase 5 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Fe...
 
Disaster Management Cycle (DMC)| Ms. Pooja Sharma , Department of Hospital A...
Disaster Management Cycle (DMC)| Ms. Pooja Sharma , Department of  Hospital A...Disaster Management Cycle (DMC)| Ms. Pooja Sharma , Department of  Hospital A...
Disaster Management Cycle (DMC)| Ms. Pooja Sharma , Department of Hospital A...
 
MVP Health Care City of Schenectady Presentation
MVP Health Care City of Schenectady PresentationMVP Health Care City of Schenectady Presentation
MVP Health Care City of Schenectady Presentation
 
Russian Call Girls South Delhi 9711199171 discount on your booking
Russian Call Girls South Delhi 9711199171 discount on your bookingRussian Call Girls South Delhi 9711199171 discount on your booking
Russian Call Girls South Delhi 9711199171 discount on your booking
 
Models Call Girls Electronic City | 7001305949 At Low Cost Cash Payment Booking
Models Call Girls Electronic City | 7001305949 At Low Cost Cash Payment BookingModels Call Girls Electronic City | 7001305949 At Low Cost Cash Payment Booking
Models Call Girls Electronic City | 7001305949 At Low Cost Cash Payment Booking
 
Call Girls Gurgaon Vani 9999965857 Independent Escort Service Gurgaon
Call Girls Gurgaon Vani 9999965857 Independent Escort Service GurgaonCall Girls Gurgaon Vani 9999965857 Independent Escort Service Gurgaon
Call Girls Gurgaon Vani 9999965857 Independent Escort Service Gurgaon
 
2024 HCAT Healthcare Technology Insights
2024 HCAT Healthcare Technology Insights2024 HCAT Healthcare Technology Insights
2024 HCAT Healthcare Technology Insights
 
Gurgaon iffco chowk 🔝 Call Girls Service 🔝 ( 8264348440 ) unlimited hard sex ...
Gurgaon iffco chowk 🔝 Call Girls Service 🔝 ( 8264348440 ) unlimited hard sex ...Gurgaon iffco chowk 🔝 Call Girls Service 🔝 ( 8264348440 ) unlimited hard sex ...
Gurgaon iffco chowk 🔝 Call Girls Service 🔝 ( 8264348440 ) unlimited hard sex ...
 
Call Girl Service ITPL - [ Cash on Delivery ] Contact 7001305949 Escorts Service
Call Girl Service ITPL - [ Cash on Delivery ] Contact 7001305949 Escorts ServiceCall Girl Service ITPL - [ Cash on Delivery ] Contact 7001305949 Escorts Service
Call Girl Service ITPL - [ Cash on Delivery ] Contact 7001305949 Escorts Service
 
EMS and Extrication: Coordinating Critical Care
EMS and Extrication: Coordinating Critical CareEMS and Extrication: Coordinating Critical Care
EMS and Extrication: Coordinating Critical Care
 
Russian Escorts Delhi | 9711199171 | all area service available
Russian Escorts Delhi | 9711199171 | all area service availableRussian Escorts Delhi | 9711199171 | all area service available
Russian Escorts Delhi | 9711199171 | all area service available
 

5x More Sensitive Lateral Flow Using Activated Magnetic Particles

  • 1. Lateral Flow Achieving 5 x More Sensitivity And Less Variation In Results Than Covalent Chemistry With Magnetic Particles By Anteo Technologies ! ! Page !1
  • 2. Summary A new method of conjugating antibodies to particles was compared to covalent conjugation in a lateral flow assay for human Chorionic Gonadotropin (hCG). Anteo Technologies will be launching a new Lateral Flow Coupling Kit at AACC 2016 that contains Anteo’s patented Activation Reagent with all necessary buffers for conjugating antibodies to a range of particles used for lateral flow applications. Here we demonstrate the performance advantages of using activated magnetic particles, only requiring the addition of antibody solution and blocking. Keeping all other variables constant, this kit was directly compared to covalently (EDAC chemistry) conjugated magnetic particles of the same type using visual detection as the readout. In this hCG study, this coupling kit achieves five times more sensitivity than the covalently conjugated magnetic particle based assay using half the amount of antibody. Comparison of Anteo’s Activated Magnetic Particles and Covalently Conjugated Magnetic Particles in a Lateral Flow Assay for Human Chorionic Gonadotropin (hCG) Introduction Lateral flow immunoassays (LFIA) or strip tests are conceptually simple immunoassays well suited to Point of Care (PoC) applications requiring little or no infrastructure support (Wong and Tse 2009, Sharma et al 2015). In its original conception, the test is easy-to- use, relatively fast and low cost, giving good qualitative information where a simple yes/no answer is sufficient. However, this same test system has been evolving to compete with more traditional laboratory based platforms. To achieve the requirements of greater assay accuracy and sensitivity without losing its basic advantages for PoC conditions, the actual strip test is now at the core of instruments such as hand held analyzers comprising sample handling, reading and analysis modules. In brief, LFIA is no longer at the cheap and simple end of the spectrum in terms of diagnostic technologies (Sharma et al, 2015). The above trends have resulted in continuous improvements over various materials and reagents used in LFIA of which Anteo Technologies’ Lateral Flow Coupling Kit is one new development. Historically, gold colloids have dominated in LFIA with visualization of conjugated gold colloids as they were captured on pre-defined lines on the test strip. Passive conjugation of antibodies on gold colloids is conceptually simple but depending on the antibody, can lead to difficulties such as poor Ab functionality and loss of colloidal stability resulting in the formation of aggregates (Wong and Tse 2009). These difficulties change with scale, and achieving good uniformity during scale-up can often be very challenging. ! ! Page !2
  • 3. Using 200 nm Magnetic Particles The use of 200 nm magnetic nanoparticles activated with Anteo’s reagent for lateral flow as an alternative to colloidal gold addresses a number of potential problematic situations encountered in the development of new lateral flow assays. For example, preparation of labeled conjugates is simplified using magnetic particles as they are easy to separate during conjugation, washing and blocking steps, with the additional benefit of inclusion of a magnetic label for performing magnetic immunoassays (MIA) (Shi et al 2015). Alternatively, when used in conjunction with sufficiently dark magnetic particles visual detection alone can produce results rivaling those of other colored particle types. The kit contains Anteo’s Activation Reagent that can be used to activate particle sizes from 200 nm to 1um that become immediately reactive to antibodies once Anteo’s reagent is applied.! Conventionally, conjugation of antibodies to magnetic particles is commonly done by covalent chemistry using reagents such as 1- Ethyl-3-(3-dimethylaminopropyl)carbodiimide (EDC) and N-hydroxysulfosuccinimide (S- NHS). A reactive ester is formed on the magnetic particles, which reacts with amino groups on the protein. Chemically, the reaction is simple but fraught with multiple variables to consider when trying to achieve optimum results. Conditions that increase stability of the ester during synthesis make it less reactive to amino groups on the protein. Increasing ester reactivity leads to side reactions. While different approaches can mitigate many such problems, these methods require pre- activation of particles which cannot be stored for any length of time. The magnetic particles activated by Anteo’s Activation Reagent are claimed to be stable in storage and highly reactive to proteins with other advantages such as ease of use, less antibodies used, faster conjugation and reduced aggregation amongst its benefits. Using a hCG lateral flow assay previously developed using colloidal gold at Diagnostic Consulting Network (DCN), the Anteo coupling kits were compared to covalently conjugated magnetic particles using the same assay materials (membrane, pads, and antibodies) with the exception of the gold colloids. ! ! Page !3
  • 4. Method The hCG assay has previously been developed using colloidal gold at Diagnostic Consulting Network (DCN), Carlsbad, CA for internal use and demonstration purposes. The same assay materials (membrane, pads, and antibodies) were used with Anteo’s technology and covalently conjugated magnetic particles. All materials and tests were performed in DCN’s laboratories using magnetic particles conjugated with anti-hCG antibody using Anteo’s protocol and DCN’s standard covalent conjugation procedure. hCG concentrations from 0 to 250mIU/ml were prepared in 1%BSA, 0.1%Tw-20 1xPBS and in negative pooled human urine. Tests were performed using 100ul of sample and test strips with the covalently conjugated and Anteo’s conjugated particles dried onto a conjugate pad. Five replicates for each concentration were tested and results were recorded 15 minutes after the addition of the sample using DCN’s visual grading scale and an ESE Quant (Qiagen Lake Constance, Germany) reflectance reader. The amount of antibody conjugated to the different particles and the amount of particles used per test were optimized to produce the highest sensitivity. Anteo’s conjugate was tested with an antibody loading of 25ug of anti-hCG antibody per mg of particles at 0.05% solids. The covalently conjugated particle was tested with an antibody loading of 50ug of anti-hCG per mg particles at 0.05% solids. Results The results for the visual grades and the values recorded by the reflectance reader, including standard deviation and %CV are shown in Table 1 for each of the two tests. The limit of detection for the Anteo assay was ~10mIU/ml hCG in buffer and ~25mIU/ml in urine for the visual and reader based results. The limit of detection for the covalently conjugated hCG assay was ~50mIU/mL in buffer and ~100mIU/mL in urine. As a reference the limit of the detection for the optimized colloidal gold assay is at ~2.5mIU/ mL in buffer and ~25mIU/mL in urine. NOTE: A visual grade of 2 or higher at the test line is considered positive, while the cutoff for the reader is at 30mV. ! ! Page !4
  • 5. Table 1: Visual and Reader results of the Anteo’s technology and covalently conjugated magnetic particle based hCG assays in buffer.
 ! Figure 1: Graph of the results from the Anteo and covalently conjugated magnetic particle hCG assays in buffer measured by a Qiagen reader.! ! ! Page !5
  • 6. Figures 3 and 4: Pictures of lateral flow strips run with the Anteo and covalently conjugated magnetic particles with hCG diluted in buffer. Figure 3: Anteo magnetic particle test strips run with hCG diluted in buffer Figure 4: Covalently conjugated magnetic particle test strips run with hCG diluted in buffer. Table 2: Visual and reader results of the Anteo and covalently conjugated magnetic particle hCG assays with urine. ! ! Page !6 250mIU/mL 100mIU/mL 50mIU/mL 25mIU/mL 10mIU/mL 5mIU/mL 0mIU/mL 250mIU/mL 100mIU/mL 50mIU/mL 25mIU/mL 10mIU/mL 5mIU/mL 0mIU/mL
  • 7. ! Figures 5 and 6: Graphs of the reader results from the Anteo and covalently conjugated magnetic particle hCG assays in urine. ! ! Page !7
  • 8. Figures 7 and 8: Pictures of the Anteo and covalently conjugated magnetic particle hCG assays run with hCG diluted in urine. Figure 7: Anteo magnetic particle test strips run with hCG diluted with urine Figure 8: Covalently conjugated magnetic particle test strips run with hCG diluted with urine Conclusions The direct comparison of the two tests show: The Anteo technology based hCG assay is 5 times more sensitive than the covalently conjugated magnetic particle based test using the same critical reagents. The estimated limit of detection for the Anteo assay is ~10mIU/mL of hCG in buffer and ~25mIU/mL of hCG in urine while the covalently conjugated hCG assay has limits of detection of ~50 and 100 mIU/ml of hCG in buffer and urine respectively. The hCG assay using the Anteo particles used half the amount of antibody to achieve better sensitivity than the covalently conjugated assay. The variability of the Anteo technology based hCG assay is lower than the covalently conjugated hCG assay. There is also less non-specific binding with the Anteo particles.! The Anteo technology based magnetic particle hCG lateral flow assay produced a comparable limit of detection to the optimized colloidal gold reference assay of ~25mIU/mL in urine. This confirms that particles conjugated with the Anteo Activation Reagent can rival traditionally used gold colloids in lateral flow applications and bring with it benefits associated with cost, handling and ! ! Page !8 250mIU/mL 100mIU/mL 50mIU/mL 25mIU/mL 10mIU/mL 5mIU/mL 0mIU/mL 250mIU/mL 100mIU/mL 50mIU/mL 25mIU/mL 10mIU/mL 5mIU/mL 0mIU/mL
  • 9. ! Anteo Products Lateral Flow Anteo's Lateral Flow Coupling Kit makes conducting lateral flow tests with magnetic beads easier and more flexible. Magnetic separation removes the need to perform centrifugation and filtration concentration. Advantages • An alternative to EDC chemistry the Anteo Kit is easy to use • Less hazardous chemicals to work with, Anteo’s Activation Reagent is water based • Flexibility: you can bind any antibody subclass to your surface • Promotes monolayer binding which increases antibody functionality • Reduces aggregation • Low background noise The Anteo Lateral Flow Coupling Kit is launching at AACC 2016. For more information on this kit email sales@anteotech.com About Anteo Technologies Anteo Technologies is transforming the way scientists work by providing new, powerful tools developed with their patented nanoscale engineered Activation Reagent. This Reagent significantly advances traditional protein coupling methods, leading to increased assay sensitivity. Anteo’s Activation Reagent is designed by using cross-linking metal ions to enable gentle yet secure multi-point binding of biomolecules to synthetic materials. The Reagent also delivers solutions to the challenges of establishing highly functional biological interfaces between fragile biomolecules, and synthetic and often incompatible hydrophilic/hydrophobic materials such as silica, polystyrene, COC/COP, metals and ceramics. Water based, Anteo’s Activation Reagent is easy to use. Simply pipette the reagent onto your surface to activate. Hands on time for surface activation and coupling is 60 minutes. Find out more at anteotech.com ! ! Page !9
  • 10. References! Sharma S, Zapatero-Rodríguez J, Estrela P, O'Kennedy R. Point-of-Care Diagnostics in Low Resource Settings: Present Status and Future Role of Microfluidics. Biosensors (Basel). 2015 Aug 13;5(3):577-601. Shi L, Wu F, Wen Y, Zhao F, Xiang J, Ma L. A novel method to detect Listeria monocytogenes via superparamagnetic lateral flow immunoassay. Anal Bioanal Chem. 2015 Jan;407(2):529-35. Wong RC and Tse HY. Lateral flow immunoassays, Humana press, DOI 10.1007/978-1-59745-240-3, 2009 ! ! Page !10