1. How will Labguard®
watch over your quality?
BIOMERIEUX PERFORMANCE SOLUTIONS™ • Quality & Compliance Services
2. EVALUATION
l Pre-installation evaluation with our experts
to assess the installation constraints of your
infrastructure
l Complete configuration of the network
SERVICE AGREEMENT
l Specialists dedicated to Labguard®
systems
l International hotline for assistance
l Remote engineering, applications, software support
l Contract for adjustments and preventive maintenance
QUALIFICATION
GUIDELINES (IQ, OQ, PQ)
bioMérieux offers a qualification package for
the Labguard system. Our qualification package
includes the following elements:
l Risk analysis with risk scenario identification.
l Qualification protocol
l Test sheets (for each qualification phase)
l Phase conclusion sheets
TRAINING
l Complete training on the equipment for users and
administrators
l Dedicated training for metrology
METROLOGY SERVICES
bioMérieux offers a full range of metrology services adapted to
your needs and regulatory requirements:
l Temperature reference sensor
l Calibration/adjustment characterization kit
l Calibration/adjustment characterization license
l External or internal calibration adjustments sensors
l ISO 17025 accredited or traceable metrology available
LABGUARD®
3D SERVICES
l Record all equipment needs in the Labguard®
software
l Configuration of monitoring (set points, alarm thresholds ...)
and metrological parameters
2
Please ask your bioMérieux
contact for more information.
bioMerie
ux will assist you as a partner
using
the full range of dedic
atedservices.
bioMérieux provides
a full range of services
made to ensure
the reliability of protocols
and equipment for
the long term so that
quality assurance
can be improved.
3. MOBILITY
METRO
In today’s competitive environment with increasing
workloads, there is a heightened demand for
cost containment, while still maintaining or
improving quality.
Environmental monitoring is an essential part
of regulatory compliance, quality and laboratory
safety. Adapted to all your specific needs, we have
design a comprehensive environmental monitoring
program: the Labguard®
3D Solution.
• Designed for mobile applications
• Interactive alarm memory integrated
• Compatible with Labguard 3D software
• Complete range of tools
• bioMérieux quality services
• ISO or Traceable metrology
4. HARDWARE
SOFTWARE
OLOGY
• Unique innovative design
• High performance battery
• High-tech with visual alarms
• True web server
• Flexibility and scalability
• 24/7 real-time monitoring
5. HARDWARE
HARDWARE
The innovative hardware from Labguard®
communicates
with the user. It has universal sensor connectivity,
up to 4 different inputs.
The benefits of Labguard 3D Hardware are:
IIIIIIIII Interactivity - multifunctional display easy to use
IIIIIIIII Universal - automatic detection of sensor -
until 4 inputs in the same transmitter
IIIIIIIII Scalable - radio Ethernet transmitter - battery,
power supply Power Over Ethernet connection
IIIIIIIII Metrology - certificate embedded in the sensor
4 parameters on
the same transmitter
4 differents chambers
monitered on the same transmitter
3
6. Reference 416010 416013 416017 416011 416014 416016 416012 416015 416018
Communication
Radio Band 868 Mhz 915 Mhz 433 Mhz 868 Mhz 915 Mhz 433 Mhz 868 Mhz 915 Mhz 433 Mhz
Power (up to) 25 mW 25 mW 10 mW 25 mW 25 mW 10 mW 25 mW 25 mW 10 mW
Type Radio frequency
Connectivity USB / Radio frequency
Transmission Rate 9600 bps
Display No Yes Yes
Channel
Number 1 2 4
Connectivity M8 Male 4 Points
Transmitter
Dimension (HxWxD) 123 X 107 X 31 mm
Weight 190 g
Comsumption
Connectivity Lithium Battery / USB / Sector
Battery 2 Lithium Battery Li AA 3,6 V, 2600 mAh
Sector 5VDC, 1A with power supply
Standby 5 µA
Transmission 50 mA
Battery Lifetime Until 2 years
Internal Memory Data stored in the sensors
Protection Index IP55
Software Compatibility All Labguard®
Software Pack From Fuchsia Version
Included Lithium Battery, Wall Mount
l No display one input available
l 360° - projected visual alarms
l 3 radio frequencies available
l Automatic detection of new sensors
l Real time or data logging
communication
l Alert for battery lost communication
l Simple action buttons
l Lithium battery, USB or power supply
l Wall mounted kit
l Display with 2 universal
inputs
l All type of sensors in
the same transmitter
l Plug play sensors
l Display with 4 universal
inputs
l Worldwide innovation
Conformities
CE 868 Mhz FCC 915 Mhz IC 915 Mhz
DIRECTIVES
RTTE 1999/5/CE
STANDARDS RADIO EMC
CFR 47 FCC
Part 15
STANDARDS RADIO EMC
RSS-Gen /
RSS-210
Issue 8 : DEC10
Low Voltage 2006/95/CE
EMC 2004/108/CE
STANDARDS
EN 301 489-3 V 1.4.1 : 2002
EN 300 220-1 V2.4.1 : 2012
EN 300 220-2 V2.4.1 : 2012
EN 61010-1 : 2010
EN 61326-1 : 2006
EN 62479 : 2010
Radio Transmitter IIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIII
4
7. l Display with 4 universal inputs
l Power Over Ethernet / USB or Power supply
l 360° - projected visual alarms
l Automatic detection of sensors
l Real time or data logging communication
l Simple action buttons
l Wall mounted kit
Ethernet Transmitter IIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIII
Reference 416019
Communication
Type Ethernet
Connectivity Ethernet/USB
Transmission rate 10/100Mbps
Interface Ethernet Connecteur RJ45
Network protocol TCP/IP
Port used 5807
Display Yes
Channel
Number 4
Connectivity M8 Male 4 Points
Transmitter
Dimension (HxWxD) 135 x 93 x 33mm
Weight 195g
Consumption
Connectivity Ethernet/USB/sector
Power Over Ethernet Power Classe 1 (3,84W max)
Internal memory Data stored in the sensors
Software compatibility All Labguard®
Software Pack From Fuchsia Version
Included Lithium battery, Wall mount, RJ45 3m cable
Conformities
CE FCC IC
DIRECTIVES
Low Voltage 2006/95/CE
STANDARDS EMC
CFR 47 FCC
Part 15
STANDARDS EMC
ICES 003 :
2012
EMC 2004/108/CE
STANDARDS
EN 61010-1 : 2010
EN 61326-1 : 2006
EN 62311 : 2008
HARDWARE
Name of channel
Level of radio
communication
Number
of channel
Alarm status Data measurement
unit
Level of battery
5
8. l Coloured visual alarms
l USB or Ethernet available
connection
l USB - Ethernet adapters
included - Power-over Ethernet
l Wall mounted kit
Radio Receiver IIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIII
Reference 416035 416036 416037
Communication
Radio Band 868 Mhz 915 Mhz 433 Mhz
Power (up to) 25 mW 25 mW 10 mW
Connectivity Ethernet/USB/Radio frequence
Port used 5807
Network protocol TCP/IP
IP address allocation DHCP or fixed
Receiver
Dimension (HxWxD) 55 x 103 x 31 mm
Weight 90 g
Consumption
Sector 5VDC, 1A with power supply POE
Power Over Ethernet Power Classe 1 (3,84W max)
Connectivity Ethernet/USB
Consumption 300 mA
Software compatibility All Labguard®
Software Pack From Fuchsia Version
Included Wall mount, RJ45 USB 2m cable
6
Conformities
CE FCC IC
DIRECTIVES
RTTE 1999/5/CE
STANDARDS RADIO EMC
CFR 47 FCC
Part 15
STANDARDS RADIO EMC
RSS-Gen /
RSS-210
Low Voltage 2006/95/CE
EMC 2004/108/CE
STANDARDS
EN 301 489-3 : 2002 V 1.4.1
EN 300 220-2 V2.4.1
EN 61010-1 : 2010
EN 61326-1 : 2006
9. l Numeric sensor with unique ID
l Internal memory (3800 data)
l Metrology certificate inside
l Calibration ISO 17025 available
l Fully plug play sensor
l Large range of sensors
Sensors IIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIII
HARDWARE
Sensor Temp Sensor PT100 Sensor Humidity Sensor Humidity Sensor Co2
Sensor Co2
Reference 416038 416042 416039 416045 416043 416044
Application Temperature High Temperature Humidity Temperature Humidity Temperature CO2
CO2
Operating range -30° to +80°C -90° to +130°C
20 to 80 % HR
0° to 60°C
0 to 100 % HR
-40° to +80°C
0 to 10% CO2
-20° to +60°C
0 to 20% CO2
-20° to +60°C
Dimensions of sensor L = 45 mm L = 30 mm L = 55 mm L = 80 mm L = 100 mm L = 100 mm
Diameter of sensor 10 mm 3 mm 12 mm 12 mm 18 mm 18 mm
Filter N/A N/A PTFE 1µm PTFE 0,2 µm PTFE 0,2 µm PTFE 0,2 µm
Length of cable 2 m 2,5 m 2 m 2 m 2 m 2 m
Flat cable YES NO YES YES NO NO
Size of cable 5 mm x 1,5 mm ∅ 2 mm 5 mm x 1,5 mm 5 mm x 1,5 mm ∅ 7 mm ∅ 7 mm
Best accuracy ±0,20°C* ±0,15°C*
±0,30°C*
±5% RH
±0,20°C*
±2.5% RH
1,5% of range
+2% of reading
1,5% of range
+2% of reading
Resolution 0,06°C 0,01°C 0,1% RH 0,1% RH 0,1% CO2
0,1% CO2
Power supply N/A N/A N/A N/A** Included Included
Protection index IP68 IP68 N/A IP65 IP65 IP65
Sensor differential
pressure
Adapter
Thermocouple K
Adapter input
for dry contact
Adapter relay
for dry contact
Adapter
PT100
Adapter
0-10 V
Adapter
0-20 mA
Reference 416046 416052 416047 416050 416051 416041 416040
Application
Differential
pressure
Very High
Temperature
IN IN/OUT
High
Temperature
Signal Signal
Operating range Adjustable -200° to +1 200°C N/A
2 A to 30 V DC/AC
0,3 A to 230 V DC/AC
-200° to
+300°C
0 - 10 V 0-20 mA
Dimensions of sensor
L x W x H =
90 x 71,5 x 36 mm
N/A N/A N/A N/A N/A N/A
Length of cable N/A 0,3 m 2,5 m 2,5 m 0,3 m 2,5 m 2,5 m
Flat cable NO YES NO NO NO NO NO
Size of cable ∅ 3,8 mm 5 mm x 1,5 mm ∅ 3,8 mm ∅ 4 mm 5 mm x 1,5 mm ∅ 3,8 mm ∅ 3,8 mm
Best accuracy
±0,5% Pa
+1% of reading
±1°C N/A N/A ±0,5°C N/A N/A
Resolution 0,01% Pa 0,05°C N/A N/A 0,01°C 0,001 0,001
Power supply Included N/A N/A N/A N/A N/A N/A
Protection index IP54 N/A N/A N/A N/A N/A N/A
* bioMérieux calibration. ** Power supply transmitter recommended. 7
Temperature
sensor
PT 100 Humidity
sensor
10. MOBILITY
l Radio communication
l 360° - projected visual alarms
acquisition button
l Alarms provided in real time to the user
l Replaceable battery
l Temperature or/ humidity monitoring
l Waterproof
Mobiguard for temperature humidityMobiguard for temperature
Mobiguard IIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIII
Reference 416026 416027 416028 416029 416030 416031
Communication
Type Radio frequency
Radio Band 868 Mhz 915 Mhz 433 Mhz 868 Mhz 915 Mhz 433 Mhz
Transmission rate 9600 bps
Power (up to) 25mW 25mW 10mW 25mW 25mW 10mW
Sensors range
Application Temperature Temperature / Humidity
Temperature -30° +60°C -30° +60°C
Humidity N/A 20 ~ 80 % RH
Best accuracy* ±0.20°C ≤ 0.5°C* ±5 % RH
Protection index IP67 IP44
Material
Dimension (HxWxD) 61 x 94 x 23 mm
Weight 110 g
Consumption
Type 2 Battery Li AAA 1,5 V, 1200 mA
Standby 5 µA
Transmission 50 mA
Battery lifetime 1 year
Internal memory Around 3800 measurements
Software compatibility All Labguard®
Software Pack From Fuchsia Version
Included Lithium battery, Wall mount
8
Conformities
CE FCC IC
DIRECTIVES
RTTE 1999/5/CE
STANDARDS RADIO EMC
CFR 47 FCC
Part 15
STANDARDS RADIO EMC
RSS-Gen /
RSS-210
Low Voltage 2006/95/CE
EMC 2004/108/CE
STANDARDS
EN 301 489-3 : 2002 V 1.4.1
EN 300 220-2 V2.4.1
EN 61010-1 : 2010
EN 61326-1 : 2006
* bioMérieux calibration.
11. MOBILITY
Tomkey data logger
l USB key format (connected
directly to a PC via its
USB port)
l Data display display (last
measurements recorded,
high alarm, low alarm)
l Interchangeable lithium
battery waterproof
Tomprobe data logger
l Comes in a button cell format
l Can be connected to a PC via a USB adapter
Others solutions IIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIII
Reference 416049 MD30101 MD31011 MD31121 MD31020
Name Tomkey Tomprobe Tomprobe Tomclave Interface USB
Application Temperature Temperature High Temperature Capsule Interface
Memory (values) 8000 2048 2048 N/A N/A
Temperature -20°C to +60°C -20°C to +85°C 0°C to +125°C N/A N/A
Display YES NO NO N/A N/A
Reference MD6188
Name Mobility Manager license
Application
Standalone availability for Labguard Mobiguard,
Tomkey Tomprobe
9
The Labguard®
3D mobility range offers a secure, robust and innovative solution
for continuous monitoring during transport with radio communication (Mobiguard).
In response to your needs, we propose other autonomous dataloggers with
an integrated USB port. The benefits of our mobility solutions are:
IIIIIIIII Mobility - designed for transport datalogging
IIIIIIIII Traceability - fully compatible with Labguard®
Software and Mobility Manager
IIIIIIIII Compliance - calibration ISO17025 available
IIIIIIIII Flexibility - no installation needed
12. SOFTWARE
Labguard®
3D software offers the most advanced monitoring,
recording and alarm features for temperatures and other
environmental parameters. Labguard 3D software has been
designed to fulfill demanding monitoring requirements of
any installations. The benefits of Labguard 3D Software are:
IIIIIIIII Universal - one single web-based software license
IIIIIIIII Flexible - unlimited connections for multi-site deployment
IIIIIIIII Interactive - records and monitors all physical parameters
IIIIIIIII Compliant - helps you meet accreditation and requirements
of all major regulatory bodies (21CFR Part 11 features)
10
Alarm notification
by widget
Full web - based
installation
Notification
via email
13. SOFTWARE
User group management: create and manage groups within the software solution: user groups, monitored
system groups, service groups within an establishment.
Audit trail: monitor every action carried out within the Labguard®
software solution (data can then be exported in
Excel or PDF format).
Mobility manager license: design for transport management within the Labguard software. It can also be used to
manage sample collection tours.
Alert manager license: allows alarm notifications through SMS and Phone. Includes a scheduler. To be used with
the GPRS modem.
21 CFR part 11 license: allows Labguard software to be compliant with FDA guidelines regarding electronic records
and signature. Mandatory for pharmaceutical industries.
LDAP license (Lightweight Directory Access Protocol): allows to synchronize Labguard software with customer IS
user’s directory , providing a «single sign on» where one password for a user is shared. User information’s are automa-
tically updated from your company/database.
Modbus license: communication protocol used for Building Management System (BMS) including HVAC (Heating,
Ventilation, Air Conditioning). Allows Labguard software to communicate data to the customer’s BMS infrastructure.
Calibration license: allows to carry out Labguard sensors calibration and adjustment.
Characterization license: allows to carry out temperature characterization (mapping) of thermostatic chambers
using FDX 15140 or IEC 60068-3-X standards, with up to 15 temperature points.
License
WELCOME
License
EXECUTIVE
License
INTEGRAL
Reference MD6111 MD6171 MD6179
Channel 40 Unlimited Unlimited
User Group Managment l l l
Audit trail l l l
MD6188 Mobility manager license l l l
MD6109 Alert manager license l l l
MD6180 21 CFR Part 11 license l l l
MD6181 LDAP license l l l
MD6186 Modbus license l l l
MD6107 Calibration license l l l
MD6108 Characterization license l l l
MD6105 MASTER LABGUARD SOFTWARE CD + book jacket + mouse pad
l Included l Optional l Non applicable
11
14. METROLOGY
A complete range of Metrology equipment
for thermostatic chambers with internal instrumentation
and services is part of the Labguard®
3D range.
The benefits of bioMérieux Metrology solutions are:
IIIIIIIII bioMérieux has an internal laboratory accredited by COFRAC* to
provide calibration services ISO17025 in temperature (n°2-5717)
Jan. 2014
IIIIIIIII Efficient, simple easy to use thanks to the Labguard®
software
compatibility
IIIIIIIII Used by the majority of our clients that need to be compliant with
the ISO standards
* Member of ILAC Organization (International Laboratory Accreditation Cooperation).
l A complete solution to calibrate, check and adjust all Labguard®
range monitoring and transport sensors.
l Labguard®
software automatically configures and manages the calibration bath.
l Automate your calibrations via the bath control and automatic stage stability detection systems.
l Manage calibration certificate validity periods.
Calibration kit: composed by temperature
reference sensor with high accuracy to calibrate
and adjust temperature sensors within the range
of -90°C to +200°C.
Calibration bath: used to calibrate
and adjust several temperature sensors
at the same time, with several different
temperature settings.
Range: -30°C to +130°C
(silicone oil included).
Calibration kit IIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIII
Reference 415431 MD4001
Designation Calibration kit Calibration bath
12
★★★★
★★★★
15. METROLOGY
l Manage all the thermostatic chambers and mapping traceability (from 9 to 15 sensors).
l Centralize your mapping data in the Labguard®
software solution, for data sharing.
l Comply with the FDX 15-140 and IEC 60068-3-X standards.
l Carry out mapping operations on a laptop disconnected from the main system,
with data synchronization when the laptop is reconnected to the system.
Temperature calibration services available for Labguard 3D®
Characterization (mapping) kit IIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIII
Metrology Services IIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIII
Reference 415432 415002
Designation Mapping kit (PT100 only) PT100 sensor for Mapping kit (x1)
Reference 2300847 2300848 417493 2300843 2300844 2300849
Type of allocation Point on request Point on request Point on request
Characterization kit
(9 or 15 probes)
Characterization kit
(9 or 15 probes)
Reference kit
(Standard sensor)
Calibration points
-40°C 1Pt
+150°C
-40°C 1Pt
+150°C
-40°C 1Pt
+150°C
20°/+5°/+44°C
-80°/-20°/
+5°/+44°/+110°C
5 points chosen
within the range
-40°C/+150°C
(By default
-30°/0°/+20°/
+44°/+125C)
416038 Sensor Temp -30° to +80°C l
416042 Sensor PT100 -90° to +130°C l l
416051 Adapter PT100 -200° to +200°C l l
416026 Mobiguard 868 -30° to +40°C l
416027 Mobiguard 915 -30° to +40°C l
416028 Mobiguard 433 -30° to +40°C l
415431 Calibration Kit l l l
415432 Characterization Kit l l l
Reference 2300845 2300846 2300834 2300814 2300835 2300841 2300842
Type of allocation All applications Deep freezer Freezer Refrigerator Ambiant Incubator Sterilization
Calibration points
-30°/+5°/
+23°/+44°C
-80°C
-30°/-25°/
-20°C
+2°/+5°/
+8°C
+18°/+23°/
+28°C
+30°/+37°/
+44°C
+40º/+121ºC
416038 Sensor Temp -30° to +80°C l l l l l
416042 Sensor PT100 -90° to +130°C l l l l l l l
416051 Adapter PT100 -200° to +200°C l l l l l l l
416026 Mobiguard 868 -30° to +40°C l l l l l
416027 Mobiguard 915 -30° to +40°C l l l l l
416028 Mobiguard 433 -30° to +40°C l l l l l
Temperature reference sensor is a high accuracy reference sensor to calibrate and adjust
temperature sensors within the range of - 80°C to + 200°C.
°C/°F
13
METROLOGY ISO 17025
16. Humidity calibration services available for Labguard 3D®
C02
calibration services available for Labguard 3D
Important: references are presented as example. We can provide more calibrations for non-ISO standards
(named traceable).
Please advise your bioMérieux contact for the complete offer with more information.
Reference 417496 417495
Type of allocation Ambiant 1 point on request
Calibration points 55% at 23°C 5% 1Pt HR 95%
416039 Sensor Temp/Humidity 20 to 80% l l
416045 Sensor Temp/Humidity 0 to 100% l l
416029 Mobiguard 868 TEMP/HUM l l
416030 Mobiguard 915 TEMP/HUM l l
416031 Mobiguard 433 TEMP/HUM l l
Reference 417494
Type of allocation 1 point (at 37°C)
Calibration points 0,5%, 5% or 10%
416043 Sensor CO2
0 to 10% l
416044 Sensor CO2
0 to 20% l
Reference 416058 416059 416060 416061 416065 416066
Version DE SP IT PT FR EN
14
TRACEABLE METROLOGY
Qualification Guide IIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIII
Reference 2300856 2300850 2300851 2300852 2300853 2300854 2300857 2300855
Type of allocation
Global all
applications
Freezer Refrigerator Ambiant Incubator Sterilization
Point on
request
Characterization
Kit (9 or
15 probes)
Calibration points
-30º/+5º/
+23º/+44ºC
-30º/-25º/
-20ºC
+2º/+5º/
+8ºC
+18º/+23º/
+28ºC
+30º/+37º/
+44ºC
+40º/+121ºC
-40ºC 1Pt
+150ºC
-20º/+5º/
+44ºC
416038 Sensor Temp -30° to +80°C l l l l l l
416026 Mobiguard 868 -30° to +40°C l l l l l l
416027 Mobiguard 915 -30° to +40°C l l l l l l
416028 Mobiguard 433 -30° to +40°C l l l l l l
416042 Sensor PT100 -90° to +130°C l l l l l l l
416051 Adapter PT100 -200° to +200°C l l l l l l l
415432 Characterization Kit l
C02
17. ACCESSORIES
ACCESSORIES
Reference Name Number of unity
416067 Battery LI Transmitter 2
416068 Battery LI Mobiguard 2
416069 Adapter Relay for Receiver 1
416070 Cable µUSB/USB 1
416072 Cable Ethernet 1
416073 Sensor Mounting Clamp 20
416074 Cap Protect USB Connector 3
416075 Cap Protect Sensor Connector 10
415002 Probe PT100 -200° to +200°C 1
416053 Power Supply EU 1
416054 Power Supply US 1
416055 Power Supply UK 1
416056 Power Supply AU/CN 1
416020 Transmitter Radio Wall Mount 1
416021 Transmitter Ethernet Wall Mount 1
416022 Receiver Wall Mount 1
416023 Mobiguard Wall Mount 1
416057 Cable Extension for Sensor 1M 1
MD12301 Probe PT100 Nitro Tank 1
MD12313 Probe TCK 0 to +1100°C 1
MD13411 Probe PT100 Calib Kit 1
MD4001 Calibration Bath TPM 165S 1
MD4002 Silicone Oil 1 Liter (-30C° to +130°C) 1
MD4003 Tripod For Bath TPM 165S 1
MD4004 Perforated Cap For Bath TPM 165S 1
MD4005 Calibration Bath Interface 1
MD5041 GPRS Modem (SMS + Voice) 1