Final year project on
multi language
body Fat analyzer
1 Meka Yesuf…….......02429/04
2 Nanno Diriba……….02723/04
3 Shimelis Nigusu……02980/04
4 Zelalem Adbar………03512/04
Advisor name
1 Gizeaddis L.(MSc)
2 Hundessa D.(BSc)
Date June,2016
s
Low fat content
heart damage
 shrinkage of internal
organs
 immune system
abnormalities
 disorders of reproductive
system
 osteoarthritis
• j
Clinical background
CVD
hypertension,
 Blood pressure
Back pain
cancer of:-
 gall bladder
prostate
Kidney and breast cancer
High fat content
international
• The WHO estimated that
by 2005, at least 1.6 billion
and 400 million people
aged above 15 years were
overweight and obese
respectively. It further
projected that by 2015,
these statistics will
increase to 2.3 billion for
overweight and 700
million for obesity,
Ethiopia
• In Ethiopia, one
study conducted in Addis
Ababa in 2007 reported that
the prevalence of overweight
and obesity on elementary
school students were 7.6%and
0.9% respectively
Literature review
skin caliper, HDL and LDL.
 Need blood sample
 need laboratory and expert lab technician
 uses sophisticated equipment(chemistry
analyzer)
Body composition depends on age ,weight and
Sex. Existing solution doesn’t consider those
parameter
Complex and expensive
Less accuracy
Existing solution and gaps
Goals
Designing body fat
analyzer which is used
to analyze
 Total body water(TBW)
 Fat free mass(FFM)
 Percentage of body fat
which operate with
English Oromic and
Amharic
Constraint
 Low cost
 Accurate
 Simple to use
 Non invasive
 portable
Goal and constraint
sssssssss
 Lack of awareness and device unavailability
leads to in proportional body composition
which causes the severe chronic disease.
Need statement
 Providing early detection of severe chronic
disease caused by in proportional body
composition using BIA method.
Problem definition
To design multi language device used to
analyze fat free mass ,total body water and
percentage of body fat.
Approaches
• Designing BFA Using two BIA electrodes
• Designing the BFA device by ECG
• Designing BFA Using four BIA electrodes
Approach and design solution
• We picked 8 most important criteria ,the ones that
absolutely must be included
Reliability Cost
Durability User friendly
Portability accuracy
Easy to design sensitivity
• Pugh matrix was used to select the best design and
select the design with highest score body fat analyzer
using four BIA electrode.
Cont….
Design solution
Simulation
Algorithm
User interface
R.N Design part proposed implemented
1 Arduino Uno Arduino Uno Arduino Uno
2 LCD display 16*2 40*2
3 Keypad 4*3 4*3
4 Electrode BIA electrode BIA electrode
5 Op-amp TL-072 NE5532
6 Potentiometer 10k 10k
7 language Single language Multi language
8 measured Body fat Body composition
Prototype iteration and design
change
Final prototype
• FFM(fat free mass)
 indicates muscular strength
 should be high in healthy person
 inversely related to body fat
 developed through exercise and
controlling body fat
Device output
• TBW (total body water)
indicates the status of hydration and dehydration
high TBW= hydration
=good health status
=safe metabolic process
low TBW= dehydration
=low metabolic rate
=causes kidney failure
cont.…..
N
o
Features to be
tested
method Design
specification
Result
1 Accuracy Measured data
compared with
standard value
95 % accurate Almost
accurate
2 Safety Measuring output
current of a device
No any micro and
micro shock on the
user and measured
current should be
not more than
500micro ampere
measusured
value not
more than
500micro
ampere and
no
micro/macro
shock
3 Cost Total cost of every
component was
added
500ETB 570ETB was
total cost a
device so still
it is cost
Test performed and relation with
key criteria
Uncontrolled body composition leads to severe
chronic disease. So designing BIA device is
important for routine monitoring of body
composition and early detection of chronic
disease .
The BIA device is constructed by achieving
the desired constraints.
Conclusion
 making BIA to analyze
• early detection of cancerous tissue
• bone density
• integrating weighting scale and audio
• functionality of kidney based on dehydration
amount.
• body shape abnormalities
manufacturing and sell validation
Future step
• we would like to give a heartfelt thanks to our
adviser Gizeaddis, Hundessa ,Solomon ,
industrialengineerWasihun,Dr.bheehm,Dr.thim
othy,for their valuable support.
Acknowledgement
THANK YOU

More Related Content

Similar to final powerpoint

Technology Assessment/Outcome & Cost-Effectiveness Analysis 2016
Technology Assessment/Outcome & Cost-Effectiveness Analysis 2016Technology Assessment/Outcome & Cost-Effectiveness Analysis 2016
Technology Assessment/Outcome & Cost-Effectiveness Analysis 2016
evadew1
 
Wearable Sensors for Cardiac Rehabilitation
Wearable Sensors for Cardiac RehabilitationWearable Sensors for Cardiac Rehabilitation
Wearable Sensors for Cardiac Rehabilitation
Ashot Melik-Martirosian
 
CDISC journey in Leukemia studies using IWCLL 2008
CDISC journey in Leukemia studies using IWCLL 2008CDISC journey in Leukemia studies using IWCLL 2008
CDISC journey in Leukemia studies using IWCLL 2008
Kevin Lee
 

Similar to final powerpoint (20)

Development of platform techniques to produce cheap, eco-friendly and easy-to...
Development of platform techniques to produce cheap, eco-friendly and easy-to...Development of platform techniques to produce cheap, eco-friendly and easy-to...
Development of platform techniques to produce cheap, eco-friendly and easy-to...
 
2010 06 - LOINC-ICF
2010 06 - LOINC-ICF2010 06 - LOINC-ICF
2010 06 - LOINC-ICF
 
Nanosensors and printed electronics integration in future lab on a chip biose...
Nanosensors and printed electronics integration in future lab on a chip biose...Nanosensors and printed electronics integration in future lab on a chip biose...
Nanosensors and printed electronics integration in future lab on a chip biose...
 
8. #ifad2019 review of recent monitoring trials (edwards)
8. #ifad2019 review of recent monitoring trials (edwards)8. #ifad2019 review of recent monitoring trials (edwards)
8. #ifad2019 review of recent monitoring trials (edwards)
 
Technology Assessment/Outcome & Cost-Effectiveness Analysis 2016
Technology Assessment/Outcome & Cost-Effectiveness Analysis 2016Technology Assessment/Outcome & Cost-Effectiveness Analysis 2016
Technology Assessment/Outcome & Cost-Effectiveness Analysis 2016
 
Global Med Solutione
Global Med SolutioneGlobal Med Solutione
Global Med Solutione
 
Technology Assessment, Outcomes Research and Economic Analyses
Technology Assessment, Outcomes Research and Economic AnalysesTechnology Assessment, Outcomes Research and Economic Analyses
Technology Assessment, Outcomes Research and Economic Analyses
 
Big Innovation Conversation: E-Sepsis
Big Innovation Conversation: E-SepsisBig Innovation Conversation: E-Sepsis
Big Innovation Conversation: E-Sepsis
 
ASCO 2018 Recap :: June 2018 Webinar
ASCO 2018 Recap :: June 2018 WebinarASCO 2018 Recap :: June 2018 Webinar
ASCO 2018 Recap :: June 2018 Webinar
 
Dr Emmanuel Nsutebu - ECO 14: Patient Safety
Dr Emmanuel Nsutebu - ECO 14: Patient SafetyDr Emmanuel Nsutebu - ECO 14: Patient Safety
Dr Emmanuel Nsutebu - ECO 14: Patient Safety
 
Improving communication culture in the UK
Improving communication culture in the UKImproving communication culture in the UK
Improving communication culture in the UK
 
PPG, ECG and Blood Pressure Circuitry
PPG, ECG and Blood Pressure CircuitryPPG, ECG and Blood Pressure Circuitry
PPG, ECG and Blood Pressure Circuitry
 
Assessment of some cardiac biomarkers in adult hiv
Assessment of some cardiac biomarkers in adult hivAssessment of some cardiac biomarkers in adult hiv
Assessment of some cardiac biomarkers in adult hiv
 
Beyond the medical curve presentation
Beyond the medical curve presentationBeyond the medical curve presentation
Beyond the medical curve presentation
 
Enhanced Recovery Canada Presentation
Enhanced Recovery Canada PresentationEnhanced Recovery Canada Presentation
Enhanced Recovery Canada Presentation
 
Webinar - Surviving Sepsis: State of the Art
Webinar - Surviving Sepsis: State of the ArtWebinar - Surviving Sepsis: State of the Art
Webinar - Surviving Sepsis: State of the Art
 
Wearable Sensors for Cardiac Rehabilitation
Wearable Sensors for Cardiac RehabilitationWearable Sensors for Cardiac Rehabilitation
Wearable Sensors for Cardiac Rehabilitation
 
CDISC journey in Leukemia studies using IWCLL 2008
CDISC journey in Leukemia studies using IWCLL 2008CDISC journey in Leukemia studies using IWCLL 2008
CDISC journey in Leukemia studies using IWCLL 2008
 
PREVENTION OF NCD'S.pptx
PREVENTION OF NCD'S.pptxPREVENTION OF NCD'S.pptx
PREVENTION OF NCD'S.pptx
 
Diagnosis of chronic disease through biosensors copy
Diagnosis of chronic disease through biosensors   copyDiagnosis of chronic disease through biosensors   copy
Diagnosis of chronic disease through biosensors copy
 

final powerpoint

  • 1. Final year project on multi language body Fat analyzer 1 Meka Yesuf…….......02429/04 2 Nanno Diriba……….02723/04 3 Shimelis Nigusu……02980/04 4 Zelalem Adbar………03512/04 Advisor name 1 Gizeaddis L.(MSc) 2 Hundessa D.(BSc) Date June,2016
  • 2. s Low fat content heart damage  shrinkage of internal organs  immune system abnormalities  disorders of reproductive system  osteoarthritis • j Clinical background
  • 3. CVD hypertension,  Blood pressure Back pain cancer of:-  gall bladder prostate Kidney and breast cancer High fat content
  • 4. international • The WHO estimated that by 2005, at least 1.6 billion and 400 million people aged above 15 years were overweight and obese respectively. It further projected that by 2015, these statistics will increase to 2.3 billion for overweight and 700 million for obesity, Ethiopia • In Ethiopia, one study conducted in Addis Ababa in 2007 reported that the prevalence of overweight and obesity on elementary school students were 7.6%and 0.9% respectively Literature review
  • 5. skin caliper, HDL and LDL.  Need blood sample  need laboratory and expert lab technician  uses sophisticated equipment(chemistry analyzer) Body composition depends on age ,weight and Sex. Existing solution doesn’t consider those parameter Complex and expensive Less accuracy Existing solution and gaps
  • 6. Goals Designing body fat analyzer which is used to analyze  Total body water(TBW)  Fat free mass(FFM)  Percentage of body fat which operate with English Oromic and Amharic Constraint  Low cost  Accurate  Simple to use  Non invasive  portable Goal and constraint
  • 7. sssssssss  Lack of awareness and device unavailability leads to in proportional body composition which causes the severe chronic disease. Need statement  Providing early detection of severe chronic disease caused by in proportional body composition using BIA method. Problem definition
  • 8. To design multi language device used to analyze fat free mass ,total body water and percentage of body fat. Approaches • Designing BFA Using two BIA electrodes • Designing the BFA device by ECG • Designing BFA Using four BIA electrodes Approach and design solution
  • 9. • We picked 8 most important criteria ,the ones that absolutely must be included Reliability Cost Durability User friendly Portability accuracy Easy to design sensitivity • Pugh matrix was used to select the best design and select the design with highest score body fat analyzer using four BIA electrode. Cont….
  • 14. R.N Design part proposed implemented 1 Arduino Uno Arduino Uno Arduino Uno 2 LCD display 16*2 40*2 3 Keypad 4*3 4*3 4 Electrode BIA electrode BIA electrode 5 Op-amp TL-072 NE5532 6 Potentiometer 10k 10k 7 language Single language Multi language 8 measured Body fat Body composition Prototype iteration and design change
  • 16. • FFM(fat free mass)  indicates muscular strength  should be high in healthy person  inversely related to body fat  developed through exercise and controlling body fat Device output
  • 17. • TBW (total body water) indicates the status of hydration and dehydration high TBW= hydration =good health status =safe metabolic process low TBW= dehydration =low metabolic rate =causes kidney failure cont.…..
  • 18. N o Features to be tested method Design specification Result 1 Accuracy Measured data compared with standard value 95 % accurate Almost accurate 2 Safety Measuring output current of a device No any micro and micro shock on the user and measured current should be not more than 500micro ampere measusured value not more than 500micro ampere and no micro/macro shock 3 Cost Total cost of every component was added 500ETB 570ETB was total cost a device so still it is cost Test performed and relation with key criteria
  • 19. Uncontrolled body composition leads to severe chronic disease. So designing BIA device is important for routine monitoring of body composition and early detection of chronic disease . The BIA device is constructed by achieving the desired constraints. Conclusion
  • 20.  making BIA to analyze • early detection of cancerous tissue • bone density • integrating weighting scale and audio • functionality of kidney based on dehydration amount. • body shape abnormalities manufacturing and sell validation Future step
  • 21. • we would like to give a heartfelt thanks to our adviser Gizeaddis, Hundessa ,Solomon , industrialengineerWasihun,Dr.bheehm,Dr.thim othy,for their valuable support. Acknowledgement