HPLC .pdf

  1. ❏ INTRODUCTION ❏ PRINCIPLE ❏ INSTRUMENTATION ❏ WORKING ❏ TYPES OF COLUMN HPLC: HIGH PRESSURE/PERFORMANCE LIQUID CHROMATOGRAPHY
  2. INTRODUCTION ❏ Chromatography is the technique by which components or solutes from a mixture are separated depending on the comparative amount of each solute which has been dispersed between a moving fluid stream, called the mobile phase, and a stationary phase. ❏ High performance liquid chromatography or commonly known as HPLC is an analytical technique used to separate, identify or quantify each component in a mixture. ❏ Mixture is separated using the basic principle of column chromatography
  3. HIGH PERFORMANCE LIQUID CHROMATOGRAPHY ➢ High performance liquid chromatography is basically an improvised variety of column chromatography. ➢ Instead of mobile phase being permitted to drip down through a column under the force of gravity, it is forced through a column under high pressures of up to 400 atmospheres, making it a much faster technique. This promotes enhanced separation of the constituents of the mixture. ➢ Another major development over column chromatography is the use of highly sensitive detectors
  4. TYPES OF HPLC ❏ Normal Phase HPLC : This method separates analytes on the basis of polarity. NP-HPLC uses polar stationary phase and non-polar mobile phase. Therefore, the stationary phase is usually silica and typical mobile phases are hexane, methylene chloride, chloroform, diethyl ether, and mixtures of these.Polar samples are thus retained on the polar surface of the column packing longer than less polar materials. ❏ Reverse Phase HPLC : The stationary phase is nonpolar (hydrophobic) in nature, while the mobile phase is a polar liquid, such as mixtures of water and methanol or acetonitrile. It works on the principle of hydrophobic interactions hence the more nonpolar the material is, the longer it will be retained.
  5. ❏ Size-exclusion HPLC : The column is filled with material having precisely controlled pore sizes, and the particles are separated according to its their molecular size. Larger molecules are rapidly washed through the column; smaller molecules penetrate inside the porous of the packing particles and elute later. ❏ Ion-Exchange HPLC : The stationary phase has an ionically charged surface of opposite charge to the sample ions. This technique is used almost exclusively with ionic or ionizable samples.The stronger the charge on the sample, the stronger it will be attracted to the ionic surface and thus, the longer it will take to elute. The mobile phase is an aqueous buffer, where both pH and ionic strength are used to control elution time.
  6. HPLC operate under the same basic principle; separation of a sample into its constituent parts because of the difference in the relative affinities of different molecules for the mobile phase and the stationary phase used in the separation. ADSORPTION CHROMATOGRAPHY The separation principle of HPLC is based on the distribution of the analyte (sample) between a mobile phase (eluent) and a stationary phase (packing material of the column). Depending on the chemical structure of the analyte, the molecules are retarded while passing the stationary phase. The specific intermolecular interactions between the molecules of a sample and the packing material define their time “on-column”. Hence, different constituents of a sample are eluted at different times. Thereby, the separation of the sample ingredients is achieved PRINCIPLE
  7. HPLC INSTRUMENTATION AND WORKING It consists of a solvent reservoir, degasser,injector, pump, column, detector and an integrator and display system
  8. ● The sample is injected through a port in the high pressure liquid carrier stream between the pump and the column. ● The separation takes place in the column which varies from 3- 30 cm in length and 3 mm in diameter. ● Typical flow rates fall between 1-2 ml/min with pressure up to several thousand psi. ● The column effluent passes through a non-destructive detector where a property such as ultraviolet absorbance, refractive index and molecular fluorescence is monitored. ● The signal is then amplified and recorded as a detector response v/s retention time. ● The graph thus obtained is called chromatogram. ● The effluent may be discarded, recycled or saved for any further research or studies in a fraction collector which is synchronized with the detector.
  9. INSTRUMENTATION Solvent Reservoir: Mobile phase contents are stored in a glass reservoir In HPLC, usually a mixture of polar and non-polar liquid constituents is used as the mobile phase, or solvent, where the concentration of constituents varies based on the composition of sample Degasser : Degassing is an important action performed prior to mobile phase delivery into HPLC system. Solvents remain in contact with the atmosphere and therefore essentially consist of an equilibrium level of dissolved air. When solvents are mixed, bubbles are formed due to dissolved air which can interfere with the pump operation and sensitivity of detectors. Also the solubility of analytes in air saturated solvents is lower as compared to pure solvents. Degassing of the solvent before its entry into pump can resolve all these issues.
  10. Pump : The pump is considered as the heart of this chromatography system.A pump aspirates the mobile phase from the solvent reservoir and forces it through the system’s column and detector. Depending on a number of factors including column dimensions, particle size of the stationary phase, the flow rate and composition of the mobile phase, operating pressures of up to 42000 kPa (about 6000 psi) can be generated. Sample Injector: The injector can be a single injection or an automated injection system. An injector for an HPLC system should provide injection of the liquid sample within the range of 0.1-100 mL of volume with high reproducibility and under high pressure (up to 4000 psi).
  11. Column: Columns are usually made of polished stainless steel, are between 50 and 300 mm long and have an internal diameter of between 2 and 5 mm. They are commonly filled with a stationary phase with a particle size of 3–10 µm. Columns with internal diameters of less than 2 mm are often referred to as microbore columns. Ideally the temperature of the mobile phase and the column should be kept constant during an analysis. Detector: The HPLC detector, located at the end of the column detect the analytes as they elute from the chromatographic column. Commonly used detectors are UV-spectroscopy, fluorescence, mass-spectrometric and electrochemical detectors.
  12. Data Collection Devices Signals from the detector may be collected on chart recorders or electronic integrators that vary in complexity and in their ability to process, store and reprocess chromatographic data. The computer integrates the response of the detector to each component and places it into a chromatograph that is easy to read and interpret.
  13. TYPES OF COLUMN There are different types of chromatographic columns on basis of composition and method of separations 1 Normal phase columns 2 Reverse phase columns 3 Ion exchange columns 4 Size exclusion columns
  14. Normal phase columns
  15. Reverse phase columns
  16. Ion exchange Hplc columns
  17. Size Exclusion Hplc Columns