Principles of Polarity in EEG
Dr Pramod Krishnan, MD (Int Med), DM Neurology (NIMHANS)
Fellowship in Epilepsy (SCTIMST) (L...
Introduction
• EEG is a graphic representation of the difference in voltage
between two different scalp locations plotted ...
Scalp
Skull
Arachnoid mater
Subarachnoid space
Cerebral cortex
Pia mater
Dura
mater
Amplifier
Electrode
Efferent axon
The ...
Pyramidal cells have unique
biophysical advantages:
1. Firing in synchrony (ie, they
summate in time)
2. Similar radial or...
SINK: At synaptic site of EPSP, positive ions enter the cell to depolarize the local membrane.
SOURCE: More distal portion...
Pyramidal dipole.
• Since the pyramidal cell PSPs
generate a circuit of current
flow throughout the length of
each cell, o...
• Basket cells, though numerous do not produce electrical activity
detectable on scalp recordings, because of their struct...
The recording electrode that detects the highest amplitude signal is overlying
the cortical source.
• Source is in a fissure.
• Source is primarily in a sulci.
• Cortical anatomy and
orientation have been altered
due to de...
Pyramidal cell orientation in the cortex
-μv
+μv
SP1 P2
+ +
+
+
+
+
+
+
+
+ +
++++
+
+
+
+
+
+ +
S: Amplitude of voltage.
P1, P2: Electrodes.
The current that reac...
• Scalp EEG can be imagined as
a constantly changing relief
map in which the map is
composed of hills and valleys
whose he...
Differential amplifier
Input
terminal 1
Input
terminal 2
Fp2
C4
Oz
Physiologic electrical currents are detected using diff...
Common mode rejection
• The advantage of differential amplifiers is that they cancel
external noise.
• The most common sou...
Differential amplifier polarity
• Current flowing through the scalp: positive or negative polarity.
• Amplifier output (In...
-40uV
-70uV
-80uV
-40uV
-40uV
-40uV
+60uV
-40uV
-40 –(-70)=+30uV
-80 –(-40)= -40uV
-40- (-40)= 0
+60-(-40)=+100uV
+50uV
+7...
Derivations and Montages.
• An electrode pair in an amplifier: electrode derivation. Eg Fp1-F3.
• If adjacent electrodes a...
Montages
Spatial filtering
Bipolar
Referential
Bipolar derivation shows the electrical field of the small cortical generator (red)....
Referential Vs Bipolar (longitudinal vs transverse)
• Both are equally important.
• Neither is superior; they are spatial ...
Spatial filters: Montages
Longitudinal bipolar montage
Transverse bipolar montage
Laplacian montage
Average reference mont...
EEG scalp localisation
1. The identification and classification of EEG activity is always
based to some degree on localisa...
The EEG signal
• Localization consists of 2 steps:
1. Localising EEG activity to specific electrode positions on the
scalp...
Localising current fields: Bipolar montage
• The electrode with the highest voltage potential is found by:
1. Instrumental...
Bipolar Vs Referential
• Bipolar montage: localisation is accomplished by analysing phase
orientation and amplitude.
• Ref...
Instrumental phase reversal
Fp1-F3
F3-C3
C3-P3
P3-O1
The electrode common to both channels that reverse, phase localises t...
Bipolar longitudinal montage showing instrumental phase reversal across C3, and also frequent
spikes at P3 suggesting that...
Bipolar longitudinal montage showing instrumental phase reversal across F3, suggesting that F3
is closest to the field max...
Sharp
waves at
F3,C3,
Bipolar longitudinal montage showing instrumental phase reversal across F3, suggesting that F3
is cl...
Instrumental phase reversal with cancellation in an
intervening channel.
Fp1-F3
F3-C3
C3-P3
P3-O1
The intervening channel ...
Bipolar longitudinal montage showing instrumental phase reversal across C3, and also phase
reversal with cancellation at T...
End of chain phenomenon.
Fp1-F3
F3-C3
C3-P3
P3-O1
If no phase reversal is seen, then the field maxima is closest to the el...
End of chain phenomenon with cancellation.
Fp1-F3
F3-C3
C3-P3
P3-O1
The field maxima is localised between the electrodes F...
True phase reversal
• This is rare and is usually seen in BCECTS or in conditions with
altered cortical anatomy.
• It mean...
Referential montage showing frequent spikes with frontal positivity and negativity in the central,
parietal and temporal c...
Referential montage of a 5 year old boy with BCETS: The positivity projects anteriorly and
negativity more posteriorly. An...
Common reference montage
Fp1-A1
F3-A1
C3-A1
P3-A1
The channel with the highest amplitude identifies the input 1 electrode ...
Referential montage showing frequent spikes with maximum amplitude at P3 and C3, localising
the field maxima to these elec...
Referential montage showing frequent spikes with maximum amplitude at T6, and smaller
amplitude at O2, localising the fiel...
Referential montage showing generalised polyspike and wave discharges with maximum
amplitude at F3, F4, Fp1 and Fp2.
Bipolar Referential montage showing generalised polyspike and wave discharges with maximum
amplitude at F3, F4, Fp1 and Fp...
Common reference montage with reference
contamination
Fp1-A1
F3-A1
C3-A1
P3-A1
The reference electrode A1 is within the po...
Common reference montage showing waveforms of similar morphology and amplitude in all the
channels suggesting reference co...
Common reference montage with reference
contamination
Fp1-A1
F3-A1
C3-A1
P3-A1
The reference electrode A1 is contaminated ...
Conclusion
• EEG is largely a graphic representation of summated PSPs of
pyramidal neurons of the cerebral cortex.
• The p...
THANK YOU
Principles of polarity in eeg
Próximo SlideShare
Cargando en…5
×

de

Principles of polarity in eeg Slide 1 Principles of polarity in eeg Slide 2 Principles of polarity in eeg Slide 3 Principles of polarity in eeg Slide 4 Principles of polarity in eeg Slide 5 Principles of polarity in eeg Slide 6 Principles of polarity in eeg Slide 7 Principles of polarity in eeg Slide 8 Principles of polarity in eeg Slide 9 Principles of polarity in eeg Slide 10 Principles of polarity in eeg Slide 11 Principles of polarity in eeg Slide 12 Principles of polarity in eeg Slide 13 Principles of polarity in eeg Slide 14 Principles of polarity in eeg Slide 15 Principles of polarity in eeg Slide 16 Principles of polarity in eeg Slide 17 Principles of polarity in eeg Slide 18 Principles of polarity in eeg Slide 19 Principles of polarity in eeg Slide 20 Principles of polarity in eeg Slide 21 Principles of polarity in eeg Slide 22 Principles of polarity in eeg Slide 23 Principles of polarity in eeg Slide 24 Principles of polarity in eeg Slide 25 Principles of polarity in eeg Slide 26 Principles of polarity in eeg Slide 27 Principles of polarity in eeg Slide 28 Principles of polarity in eeg Slide 29 Principles of polarity in eeg Slide 30 Principles of polarity in eeg Slide 31 Principles of polarity in eeg Slide 32 Principles of polarity in eeg Slide 33 Principles of polarity in eeg Slide 34 Principles of polarity in eeg Slide 35 Principles of polarity in eeg Slide 36 Principles of polarity in eeg Slide 37 Principles of polarity in eeg Slide 38 Principles of polarity in eeg Slide 39 Principles of polarity in eeg Slide 40 Principles of polarity in eeg Slide 41 Principles of polarity in eeg Slide 42 Principles of polarity in eeg Slide 43 Principles of polarity in eeg Slide 44 Principles of polarity in eeg Slide 45 Principles of polarity in eeg Slide 46 Principles of polarity in eeg Slide 47
Próximo SlideShare
What to Upload to SlideShare
Siguiente

23 recomendaciones

Compartir

Principles of polarity in eeg

This presentation looks at EEG signal generation, pyramidal cells, recording of EEG, source localisation, polarity, analysis of dipole, derivations, montages,

Audiolibros relacionados

Gratis con una prueba de 30 días de Scribd

Ver todo

Principles of polarity in eeg

  1. 1. Principles of Polarity in EEG Dr Pramod Krishnan, MD (Int Med), DM Neurology (NIMHANS) Fellowship in Epilepsy (SCTIMST) (LMU, Munich) World Sleep Federation Certified Sleep Medicine Specialist. Consultant Neurologist and Epileptologist Head of the Department of Neurology, Manipal Hospital, Bengaluru.
  2. 2. Introduction • EEG is a graphic representation of the difference in voltage between two different scalp locations plotted over time. • The scalp EEG signal generated by cerebral neurons is modified by: 1. Electrical conductive properties of the tissues between the electrical source and the recording electrode on the scalp. 2. Conductive properties of the electrode. 3. Orientation of the cortical generator to the recording electrode.
  3. 3. Scalp Skull Arachnoid mater Subarachnoid space Cerebral cortex Pia mater Dura mater Amplifier Electrode Efferent axon The spontaneous EEG signal in the routine scalp recorded EEG arises from cortical pyramidal cell post synaptic potentials. Pyramidal cells Synapses
  4. 4. Pyramidal cells have unique biophysical advantages: 1. Firing in synchrony (ie, they summate in time) 2. Similar radial orientation (ie they are aimed in the same direction towards the cortical surface). So, summation is better. 3. Long duration of potentials making them ideal for producing a large enough signal to be conducted through the skull and recordable at the scalp. Pyramidal cell of the cortex
  5. 5. SINK: At synaptic site of EPSP, positive ions enter the cell to depolarize the local membrane. SOURCE: More distal portion of cell, current flows out of the cell passively completing a closed circuit. •Primary current: Current flow within the cell (intracellular). (MEG Generation) •Secondary or return current: Current flow outside the cell (extracellular). (EEG generation) Sink Source + + + + + + + +
  6. 6. Pyramidal dipole. • Since the pyramidal cell PSPs generate a circuit of current flow throughout the length of each cell, one end of the pyramidal cell has the opposite polarity from the other. • A μvolt EEG signal is recorded when numerous pyramidal cells are synchronously activated and many small dipoles are combined.
  7. 7. • Basket cells, though numerous do not produce electrical activity detectable on scalp recordings, because of their structure/orientation. • However, pyramidal cells, because of their orientation and synchronous firing, show electrical activity strong enough to be recorded by scalp electrodes.
  8. 8. The recording electrode that detects the highest amplitude signal is overlying the cortical source.
  9. 9. • Source is in a fissure. • Source is primarily in a sulci. • Cortical anatomy and orientation have been altered due to developmental malformation, trauma, surgery. • Skull thickness is altered (surgical defects, thickening due to Pagets disease etc). Exceptions to this simple model
  10. 10. Pyramidal cell orientation in the cortex
  11. 11. -μv +μv SP1 P2 + + + + + + + + + + + ++++ + + + + + + + S: Amplitude of voltage. P1, P2: Electrodes. The current that reaches the electrode from the cortical source is equivalent to the area contained by the lines that begin at the cortex and converge on the electrode. For P2, the source projects both negativity from the outer cortex (area between solid lines) and positivity from the inner cortex (area between dashed lines). Therefore the voltage of the signal is reduced. Volume conduction theory for a cortical source that occupies a gyrus
  12. 12. • Scalp EEG can be imagined as a constantly changing relief map in which the map is composed of hills and valleys whose heights and depths represent a particular voltage. • The more electrodes are used, the more accurately they will map the actual contours of the scalp potential.
  13. 13. Differential amplifier Input terminal 1 Input terminal 2 Fp2 C4 Oz Physiologic electrical currents are detected using differential amplifiers. Input 1- Input 2= Recorded EEG signal (hence differential)
  14. 14. Common mode rejection • The advantage of differential amplifiers is that they cancel external noise. • The most common source of noise are other electrical devices. • Eg, When current is supplied at 50-60 cycles/sec, it causes electrical interference which equally contaminates input 1 and 2. This gets cancelled out by the differential amplifier. • Because differential amplifier is used, the absolute voltage value at electrodes attached to either Input 1 or 2 is never known, only the difference between them is known.
  15. 15. Differential amplifier polarity • Current flowing through the scalp: positive or negative polarity. • Amplifier output (Input 1- Input 2) is assigned a polarity (positive or negative). • Input 1 more negative than Input 2: upward deflection • Input 1= Input 2: no deflection (flat line) • Input 1 less negative than Input 2: downward deflection
  16. 16. -40uV -70uV -80uV -40uV -40uV -40uV +60uV -40uV -40 –(-70)=+30uV -80 –(-40)= -40uV -40- (-40)= 0 +60-(-40)=+100uV +50uV +70uV +50-(+70)=-20uV 1 1 1 1 1 2 2 2 2 2
  17. 17. Derivations and Montages. • An electrode pair in an amplifier: electrode derivation. Eg Fp1-F3. • If adjacent electrodes are paired, it is called bipolar. Eg Fp1-F3 • If the pair includes a referential electrode, it is called referential derivation. Eg Fp1-A1. • A display that consists of a combination of derivations is referred to as a montage. • Accordingly montages can be referential or bipolar, to visualise widespread vs localised fields respectively. • Bipolar montages can be longitudinal or transverse.
  18. 18. Montages
  19. 19. Spatial filtering Bipolar Referential Bipolar derivation shows the electrical field of the small cortical generator (red). Larger generator (green) is cancelled out as both the electrodes are placed over it. In the referential derivation, the larger field is clearly seen as one electrode is outside the two fields, but the field of the smaller generator is not well made out.
  20. 20. Referential Vs Bipolar (longitudinal vs transverse) • Both are equally important. • Neither is superior; they are spatial filters that emphasize different aspects of the scalp topography of EEG voltage fields. • Longitudinal and transverse bipolar montages are both used because scalp voltage fields can be asymmetric along either anterior to posterior or transverse axis. • To assess such asymmetries, we need to use both a longitudinal and a transverse bipolar montage, or by using one bipolar and reference montage.
  21. 21. Spatial filters: Montages Longitudinal bipolar montage Transverse bipolar montage Laplacian montage Average reference montage Common reference montage Maximumfilteringofwidespreadfields Maximumenhancementofhighlylocalisedfields
  22. 22. EEG scalp localisation 1. The identification and classification of EEG activity is always based to some degree on localisation. 2. Localisation can help determine if an activity is cerebral in origin or an artefact. 3. Clinical correlation of an EEG requires cortical localization of the EEG findings. 4. Localization helps in correctly interpreting an EEG pattern the reader has not encountered before.
  23. 23. The EEG signal • Localization consists of 2 steps: 1. Localising EEG activity to specific electrode positions on the scalp 2. Relating the scalp localization to the likely source in the underlying cerebral cortex.
  24. 24. Localising current fields: Bipolar montage • The electrode with the highest voltage potential is found by: 1. Instrumental phase reversal. 2. Instrumental phase reversal with cancellation in an intervening derivation. 3. End of chain phenomenon or the absence of phase reversal. 4. End of chain phenomenon with cancellation. 5. True phase reversal (double phase reversal).
  25. 25. Bipolar Vs Referential • Bipolar montage: localisation is accomplished by analysing phase orientation and amplitude. • Referential montage: localisation is mainly by analysing amplitude. • Phase refers to the upward or downward deflection of the waveform from the baseline. • Phase reversal has occurred if the phase has changed direction from up to down or from down to up.
  26. 26. Instrumental phase reversal Fp1-F3 F3-C3 C3-P3 P3-O1 The electrode common to both channels that reverse, phase localises the maximum field potential to that common electrode (e.g. F3). Since it is the way the electrodes are arranged with the input 2 electrode becoming Input 1 electrode in the next channel that causes the phase reversal, it is called instrumental phase reversal.
  27. 27. Bipolar longitudinal montage showing instrumental phase reversal across C3, and also frequent spikes at P3 suggesting that C3 and P3 are closest to the field maxima.
  28. 28. Bipolar longitudinal montage showing instrumental phase reversal across F3, suggesting that F3 is closest to the field maxima.
  29. 29. Sharp waves at F3,C3, Bipolar longitudinal montage showing instrumental phase reversal across F3, suggesting that F3 is closest to the field maxima.
  30. 30. Instrumental phase reversal with cancellation in an intervening channel. Fp1-F3 F3-C3 C3-P3 P3-O1 The intervening channel with cancellation (F3-C3) localises the field maximum between the electrodes in that channel.
  31. 31. Bipolar longitudinal montage showing instrumental phase reversal across C3, and also phase reversal with cancellation at T3-T5.
  32. 32. End of chain phenomenon. Fp1-F3 F3-C3 C3-P3 P3-O1 If no phase reversal is seen, then the field maxima is closest to the electrode with the maximum amplitude. It is localising to the electrode at the end of the electrode chain and does not suggest the source is deep in the brain.
  33. 33. End of chain phenomenon with cancellation. Fp1-F3 F3-C3 C3-P3 P3-O1 The field maxima is localised between the electrodes Fp1 and F3 at the end of the bipolar chain. The rate of change of voltage (amplitude) between two electrodes is usually greatest near the field maxima and least at a distance from it. The presence of a higher amplitude in channel 2 than in 3 localises the field maxima to the front of the head.
  34. 34. True phase reversal • This is rare and is usually seen in BCECTS or in conditions with altered cortical anatomy. • It means that two different polarities are simultaneously present in adjacent cortical areas. • The epileptiform discharges arise at an angle that is parallel to the cortical surface (tangential dipole), allowing both the negativity from the outer cortex and positivity from the deeper cortex to be seen simultaneously.
  35. 35. Referential montage showing frequent spikes with frontal positivity and negativity in the central, parietal and temporal channels, consistent with BCECTS (tangential dipole).
  36. 36. Referential montage of a 5 year old boy with BCETS: The positivity projects anteriorly and negativity more posteriorly. Anterior positivity is lower in amplitude than the posterior negativity. This would produce two instrumental phase reversals in a longitudinal bipolar chain.
  37. 37. Common reference montage Fp1-A1 F3-A1 C3-A1 P3-A1 The channel with the highest amplitude identifies the input 1 electrode closest to the field maxima (F3). In this example, reference electrode A1 is outside the scalp field.
  38. 38. Referential montage showing frequent spikes with maximum amplitude at P3 and C3, localising the field maxima to these electrodes.
  39. 39. Referential montage showing frequent spikes with maximum amplitude at T6, and smaller amplitude at O2, localising the field maxima to the T6 electrode.
  40. 40. Referential montage showing generalised polyspike and wave discharges with maximum amplitude at F3, F4, Fp1 and Fp2.
  41. 41. Bipolar Referential montage showing generalised polyspike and wave discharges with maximum amplitude at F3, F4, Fp1 and Fp2. Few leading focal spikes (frontal) are seen
  42. 42. Common reference montage with reference contamination Fp1-A1 F3-A1 C3-A1 P3-A1 The reference electrode A1 is within the positive scalp field, whereas the other scalp electrodes are outside. Therefore, the deflection is each channel is the same, and is upward, because for each channel, input 1 is more negative than input 2. +
  43. 43. Common reference montage showing waveforms of similar morphology and amplitude in all the channels suggesting reference contamination. Changing the reference to another location may eliminate this abnormality.
  44. 44. Common reference montage with reference contamination Fp1-A1 F3-A1 C3-A1 P3-A1 The reference electrode A1 is contaminated with activity also detected by the Input 1 electrodes. A field with a single polarity will not produce waveforms of opposite phase in a referential recording between adjacent channels unless the reference is contaminated. Two phase reversals are pathognomonic of reference electrode contamination.
  45. 45. Conclusion • EEG is largely a graphic representation of summated PSPs of pyramidal neurons of the cerebral cortex. • The pyramidal dipole is recorded on the scalp, mainly a radial dipole, but sometimes it is a tangential dipole. • Bipolar montages identify localised fields, referential montages record widespread fields. • Surface negativity is recorded as upward deflection in input 1. • For source localisation, phase reversal with or without cancellation is used in bipolar montage, and maximum amplitude is used in referential montages.
  46. 46. THANK YOU
  • MichelleEvans757321

    Nov. 23, 2021
  • SameeraGhachi

    Sep. 11, 2021
  • rakeshkumar3383

    Aug. 27, 2021
  • archanasharma0411

    Jul. 4, 2021
  • MaryamFarahmandfar

    Apr. 13, 2021
  • FathimaSN

    Mar. 19, 2021
  • atulmj28

    Mar. 14, 2021
  • MaryamAzizizadeh

    Nov. 28, 2020
  • RvLz

    Sep. 16, 2020
  • hammoda_m

    Aug. 2, 2020
  • jenithasherlin

    Jul. 31, 2020
  • ZhangJianlei

    May. 31, 2020
  • JenniferPatterson63

    Apr. 7, 2020
  • ssuser7e3e44

    Feb. 11, 2020
  • VaishaliBaghi

    Feb. 4, 2020
  • Rachita31

    Feb. 3, 2020
  • ArashKahrom

    Jan. 15, 2020
  • WaltGram

    Jan. 6, 2020
  • drpramodkrishnan

    Dec. 4, 2019
  • VasileTibre1

    Dec. 1, 2019

This presentation looks at EEG signal generation, pyramidal cells, recording of EEG, source localisation, polarity, analysis of dipole, derivations, montages,