8. PULSE WAVE-CONTOUR
• Detected by the use of
an arterially placed
catheter with a pressure
transducer, which can
measure pressure
tracings on a beat-to-
beat basis
8
10. Factors affecting accuracy of PCA
arterial pressure waveform are affected by
• vascular compliance,
• aortic impedance
• Peripheral arterial resistance.
• Technical factors(i.e. eliminating damping or increased tubing
resonance , zeroing)
• severe arrhythmias
• the use of an intra-aortic balloon pump precludes adequate
performance of the devices.
• periods of hemodynamic instability, i.e., rapid changes in vascular
resistance. Is a problem for uncalibrated pulse pressure analysis.
So frequent re-calibration for accurate cardiac output estimation in
these situations is mandatory
10
11. CALIBRATION FOR PCA
• Because vascular impedance varies between patients,
it had to be measured using another modality to initially
calibrate the PCA system
• The calibration method is transpulmonary thermodilution for
PiCCO (need cvp and a-line) and VolumeViewTM
• The LiDCO plus: requires calibration using the transpulmonary
lithium indicator dilution technique, which can be performed
via a peripheral venous line(need a-line)
• FLOTRAC/VIGILEO doesn't need calibration because it estimate
cardiac output by the standard deviation of pulse pressure
sampled during a time window of 20 seconds
11
12. PCA/TPD
• Advantages
– Almost continuous data of CO / SV / SV variation
– Provide information of preload and EVLW
– Provide many other parameters which has potential clinical
utilities
• Disadvantages
– STILL invasive but less than PAC
– arterial pulse signal is affected by many factors
• Arrhythmia
• Damping
• Use of IABP
Need calibration
12
14. Tb
injection
t
dtT
KV)T(T
CO
b
iib
TDa
Transpulmonary thermodilution:
1-Cardiac Output estimation/ calibration
Tb = Blood temperature
Ti = Injectate temperature
Vi = Injectate volume
∫ ∆ Tb
. dt = Area under the thermodilution curve
K = Correction constant, made up of specific weight and
specific heat of blood and injectate
CO Calculation:
Area under the
Thermodilution Curve
After central venous injection of the indicator, the thermistor at the tip of the
arterial catheter measures the downstream temperature changes.
Cardiac output is calculated by analysis of the thermodilution curve using a
modified Stewart-Hamilton algorithm:
14
15. LiDCO
– LiCl: 0.002mmol/l injected
into central vein (peripheral
administration possible as
well)
– Arterial plasma conc.
measured by withdrawing
blood across lithium selective
electrode at 4ml/min
– CO calculated from Li dose
and area under primary
concentration-time curve
before re-circulation
PCV is packed cell volume which may be calculated as hemoglobin concentration (g/dl) / 34
Cardiac Output = (Lithium Dose x 60)/(Area x (1-PCV))
15
16. Advanced Thermodilution Curve Analysis
Transpulmonary thermodilution:
2-Volumetric parameters derivation
Mtt: Mean Transit time
time when half of the indicator
has passed the point of detection in
the artery
DSt: Down Slope time
exponential downslope time of the
thermodilution curve
For the calculations of volumes…
ln Tb
injection
recirculation
MTt
t
e-1
DSt
Tb
…are important.
…and…
All volumetric parameters are obtained by advanced analysis of the
thermodilution curve:
ITTV=CO X MTt
PTV=CO X DSt
16
17. RAEDV
Thermodilution curve
measured with arterial
catheter
CV Bolus Injection
LAEDV LVEDVRVEDV
Right Heart Left
Heart
Lungs
After injection, the indicator passes the following Intrathoracic
compartments:
The intrathoracic compartments can be considered as a series of “mixing
chambers” for the distribution of the injected indicator (intrathoracic thermal
volume).
ITTV
PTV
The largest mixing chamber in this series are the lungs, here the indicator (cold)
has its largest distribution volume (largest thermal volume).
Transpulmonary thermodilution: Volumetric parameters calculation
17
19. Index of Left Ventricular Contractility*
t [s]
P [mm Hg]
dPmx* =dP/dtmax of arterial pressure curve
dPmx* represents left ventricular pressure velocity increase and thus is a
parameter of myocardial contractility
20. Thermodilution Parameters
• Cardiac Output CO
• Global End-Diastolic Volume GEDV
• Intrathoracic Blood Volume ITBV
• Extravascular Lung Water EVLW*
• Pulmonary Vascular Permeability Index PVPI*
• Cardiac Function Index CFI
• Global Ejection Fraction GEF
The PiCCO measures the following parameters:
Pulse Contour Parameters
• Pulse Contour Cardiac Output PCCO
• Arterial Blood Pressure AP
• Heart Rate HR
• Stroke Volume SV
• Stroke Volume Variation SVV
• Pulse Pressure Variation PPV
• Systemic Vascular Resistance SVR
• Index of Left Ventricular Contractility dPmx*
Parameters measured with the PiCCO-Technology
22. What is the current situation?.………..……..………….Cardiac Output!
What is the preload?.……………….....…Global End-Diastolic Volume!
Will volume increase CO?....………...……….Stroke Volume Variation!
What is the afterload?……………..…..Systemic Vascular Resistance!
Are the lungs still dry?...…….……...…..….Extravascular Lung Water!*
What about the contractility?........................ dPmx* LV pressure velocity
Clinical application
23. Decision tree for hemodynamic / volumetric monitoring
CI (l/min/m2)
GEDI (ml/m2)
or ITBI (ml/m2)
ELWI* (ml/kg)
(slowly responding)
>3.0<3.0
>700
>850
<700
<850
>700
>850
<700
<850
ELWI* (ml/kg)
GEDI (ml/m2)
or ITBI (ml/m2)
CFI (1/min)
or GEF (%)
<10 >10 <10 <10 <10>10 >10 >10
V+ V+! V+!V+Cat Cat
OK!
V-
>700
>850
700-800
850-1000
>4.5
>25
>5.5
>30
>4.5
>25
700-800
850-1000
Cat
>5.5
>30
>700
>850
700-800
850-1000
700-800
850-1000
10 10 10 10
V-
V+ = volume loading (! = cautiously) V- = volume contraction Cat = catecholamine / cardiovascular agents
** SVV only applicable in ventilated patients without cardiac arrhythmia
>700
>850
<10Optimise to SVV** (%)<10 <10 <10
R
E
S
U
L
T
S
T
A
R
G
E
T
T
H
E
R
A
P
Y
1.
2. <10 <10 <10 <10
23
24. A protocol for resuscitation of severe burn
patients guided by transpulmonary
thermodilution and lactate levels
A decision tree for the adjustment of fluid and catecholamine therapy according to a
permissive hypovolemia protocol with lower preload targets and lactate measurements to
ensure tissue perfusion is shown
Sánchez et al. Critical Care 2013, 17:R176
26. In conclusion
• Hemodynamic monitoring enable early
detection of change in patient’s conditions
• New techniques provide reasonably good
results and less invasive
• Always correlate the readings / findings
with clinical pictures in order to provide
the best treatment options
26