The principle and performance of capillary electrophoresis
Ljubica Glavaš-Obrovac
Introduction
Capillary electrophoresis (CE) is a relatively new and a very versatile technique (also
known as high-performance capillary electrophoresis) for the separation and analysis of a variety
of molecules – from small inorganic ions to huge biopolymers. Compared to the high-
performance liquid chromatography, and gas chromatography CE offers several advantages,
including highly efficient and fast separation, relatively inexpensive and long lasting capillary
columns, small sample size requirements and low reagent consumption. It can be used for
analysis of polar ionic, polar non-ionic and non-polar non-ionic compounds, as well as high
molecular weight biomolecules and chiral compounds. Further, CE separations are highly
efficient (N>105
to 106
), sensitive in terms of mass and concentration, fast (usually less than 30
min), simple, and require extremely small amounts of solvents and other consumables. This
highly sophisticated, user friendly and very effective CE has become commercially available at
affordable costs (Slide 2).
Capillary electrophoresis theory
When a sample is injected into a capillary tube and an electric field is applied across the
capillary tube, sample’s components migrate as the result of two types of action: electrophoretic
mobility and electroosmotic mobility. Electrophoretic mobility is the solute’s response to the
applied electrical field. Cations move toward the negatively charged cathode, anions move
toward the positively charged anode, and neutral species remain stationary. The other
contribution to a solute’s migration is electroosmotic flow, which occurs when the buffer moves
through the capillary in response to the applied electrical field. Under normal conditions the
buffer moves toward the cathode, sweeping most solutes, including the anions and neutral
species, toward the negatively charged cathode1
(Slide 3).
Electrophoretic mobility
The velocity with which a solute moves in response to the applied electric field (Slide 4) is called
its electrophoretic velocity, νep; it is defined as
vep = μepE
where μep is the solute’s electrophoretic mobility, and E is the magnitude of the applied electrical
field. A solute’s electrophoretic mobility is defined as
μep = q6πηr
where q is the solute’s charge, η is the buffer viscosity, and r is the solute’s radius.
Electrophoretic mobility and, therefore, electrophoretic velocity, increases for more highly
charged solutes and for solutes of smaller size. Because q is positive for a cation and negative for
an anion, these species migrate in opposite directions. Neutral species, for which q is zero, have
an electrophoretic velocity of zero.
Electroosmotic flow (EOF)
The EOF is caused by applying high-voltage to an electrolyte-filled capillary (Slide 5).
This flow occurs when the buffer running through the silica capillary has a pH greater than 3 and
the silanol (SiOH) groups lose a proton to become silanate (SiO-
) ions. The capillary wall then
has a negative charge, which develops a double layer of cations attracted to it. The inner cation
layer is stationary, while the outer layer is free to move along the capillary. The applied electric
field causes the free cations to move toward the cathode creating a powerful bulk flow. The rate
at which the buffer moves through the capillary, its electroosmotic flow velocity, νeof, is a function
of the applied electric field, E, and the buffer’s electroosmotic mobility, μeof.
νeof =μeofE
Electroosmotic mobility is defined as
Μeof = εζ4πη
Electroosmotic flow is defined as
μEOF=ϵ4πηEζ
where ε is the dielectric constant of the solution, η is the viscosity of the solution, E is the field
strength, and ζ is the zeta potential.
Since the electrophoretic mobility is greater than the EOF, negatively charged particles,
which are naturally attracted to the positively charged anode, will separate out as well. The EOF
works best with a large zeta potential between the cation layers, a large diffuse layer of cations to
drag more molecules towards the cathode, low resistance from the surrounding solution, and
buffer with pH of 9 so that all the SiOH groups are ionized. The main benefit of the EOF is a
separation in a single run, of both anions and cations, and also the separation of ions with quite
different charge-to-size ratios in reasonable length of time as well.
Capillary electrophoresis instrumentation
The capillary electrophoresis instrument1, 2
(Slide 6) consists of:
1. Capillary
2. Injection system
3. High voltage source
4. Electrodes and electrode jars
5. Detector
1. Capillary
In CE, the separation occurs in the capillary and this is also the compartment where
injection and detection takes place. Ideally a capillary should be chemically and physically
resistant, precisely machined with narrow internal diameters (d = 20-100 µm; length of 20-100
cm), does not adsorb solutes, transparent to UV-visible radiation, with a high thermal
conductivity, and low cost. Fused silica capillaries with an external protective polyimide coating
meet almost all these requirements and are hereof a standard choice in CE. Capillaries can be
internally ‘uncoated’, having the inner silica surface in direct contact with buffers and solutes,
but, under some experimental conditions, these may adsorb analytes. Alternatively, the capillaries
can be internally coated with a thin layer of polymers (adsorbed or chemically bound) shielding
the silica surface from interactions with solutes. Capillaries can be also filled with gels,
mimicking slab gel electrophoresis, or with particles carrying a stationary phase, reproducing a
chromatographic system. The most common coatings are polyacrylamide, cellulose, polyvinyl
alcohol (PVA), amino acids, amines, surfactants, and poly(vinylpyrrolidinone) or
polyethyleneimine. Polymers well known as liquid chromatographic (C2, Cg, C1g) or gas
chromatographic stationary phases (polyethylene glycol (PEG), phenyl-methyl silicone) can be
used for this purpose too (Slide 7).
2. Injection system
Modern instrumentation uses two injection principles: hydrodynamic or electrokinetic. In the
first case, a pressure difference is generated between the two ends of the capillary, while the
injection end is dipped in the specimen vial. Alternatively, a voltage difference is established
between the samples and the opposite vial, while samples are being injected. Although
hydrodynamic injection is nonselective, electrokinetic injection (potential driven) is selective.
This is because the sample components enter the capillary according to their electrophoretic
mobility, the mobility and concentration of total ions in the sample and the EOF. Electrokinetic
injection allows field-amplified sample stacking, i.e. a highly efficient method for increasing
analytical sensitivity (in terms of concentration), to be accomplished (Slide 8).
3. High voltage source
The high-voltage power supplies used for CE are generally able to give voltages up to 20-30
kV and currents up to 200-300 µA. Separations are generally carried out under constant potential,
but voltage gradients or steps; and constant current separation is sometimes used. Because EOF
in fused silica capillaries is usually directed towards the cathode, the common polarity in a
capillary electropherograph is with the anode at the injection end of the capillary and the cathode
at the opposite end, close to the detector; however, under specific analytical conditions, the
polarity is reversed (Slide 8).
4. Electrodes and electrode jars
Each side of the high voltage power supply is connected to an electrode. These electrodes
help to induce an electric field to initiate the migration of the sample from the anode to the
cathode through the capillary tube (Slide 8).
5. Detector
In order to gain useful information from the separation technique, it is necessary to detect and
measure analytes. Detection may be qualitative and/or quantitative. While the capillary itself is
the detection cell it is linked to the detection device. It is also possible to couple to detectors that
are outside of the separation capillary although this does require a specialized interface. The most
frequently used method of detection involves absorbance of energy as the analytes move through
a focused beam of light. The CE detection, most often based on UV-visible absorption, is carried
out in-capillary, in order to avoid any possible post-separation added volumes. This can cause
unacceptable band spreading (Slide 9). Indeed, the wall of fused silica capillaries, after removing
the polyimide coating, is highly transparent. Thus, picogram amounts of analytes can easily be
detected even by UV absorption, but, due to the limitations in the sample volume (nL) which can
be injected, the concentration sensitivity with UV detectors is limited to 10-5
M. However, this
limit in sensitivity can be overcome by using other detection techniques, e.g. laser-induced
fluorescence (sensitivity up to 10-12
M) or electrochemical detection (sensitivity up to 10-8
M)
and/or by adopting high efficiency sample stacking methods, which can assure improvements in
sensitivity of 2-3 orders of magnitude. By implementation of diode-array or fast-scanning UV
detectors allows on-line recording of the UV spectra of the peaks, improving the information
content of the CE analysis. Laser induced fluorescence, conductometric detection, and mass
spectrometric with electrospray ionization detection modes have recently become commercially
available.
Indirect detection is based on the addition into the running buffer of an ionic additive which is
detectable at trace levels by the detector. This ionic compound, having the same charge and
similar mobility to the analyte(s) of interest, is displaced from the zones where the analyte(s)
migrate to preserve electroneutrality, and this displacement gives rise to ‘reversed peaks’, the
area of which is proportional to the concentration of the given analyte(s). Drawbacks of this
detection mode are sensitivity, which is usually lower than with the corresponding ‘direct’ mode,
and the narrower range of linearity
Capillary electroseparation techniques
The most common CE separation techniques1, 2,3, 4
(Slide 10, Slide 11) are:
1. Capillary zone electrophoresis (CZE)
2. Micellar electrokinetic capillary chromatography (MECC or MEKC)
3. Capillary electrochromatography (CEC)
4. Capillary isoelectric focusing (CIEF)
5. Capillary gel electrophoresis (CGE)
6. Capillary isotachophoresis (CITP)
7. Capillary electrophoretic immunoassay (CEIA)
1. Capillary zone electrophoresis (CZE)
The CZE, also known as free solution capillary electrophoresis, is very quickly, easy and the
most commonly used mode of the CE (Slide 12). The separation technique is based on the
differences in electrophoretic mobility, which is directed proportional to the charge on the
molecule, and inversely proportional to the viscosity of the solvent and radius of the atom.
Usually the end of the capillary containing the sample is the anode and solutes migrate toward the
cathode at a velocity determined by their electrophoretic mobility and the EOF. In the CZE, in
addition to the solutes, the buffer solution usually also moves through the capillary under the
influence of an electric field. Coating the capillary’s walls with a nonionic reagent eliminates the
EOF. In this form of the CZE cations migrate from the anode to the cathode. Anions elute into
the source reservoir and neutral species remain stationary. The CZE provides effective
separations of charged species, including inorganic anions and cations, organic acids and amines,
and large biomolecules such as proteins.
2. Capillary gel electrophoresis (CGE)
In the CGE a capillary is filled with a polymeric gel like cross-linked polyacrylamide,
agarose, or solutions of linear polymers (Slide 13). While the gel is porous, a solute migrates
through the gel with a velocity determined both by its electrophoretic mobility and by its size.
The ability to effect a separation using size is helpful when the solutes have similar
electrophoretic mobilities. For example, fragments of DNA of varying length have similar
charge-to-size ratios, making their separation by CZE difficult. Because the DNA fragments are
of different size they can be separated by the CGE. The capillary used for the CGE is usually
treated to eliminate the EOF, preventing the gel’s extrusion from the capillary tubing. Samples
are injected electrokinetically because the gel provides too much resistance for hydrodynamic
sampling. The primary application of CGE is the separation of large biomolecules, including
DNA fragments, proteins, and oligonucleotides. The main advantages over slab-gel
electrophoresis are a wider range of gel matrix and composition, online detection, improved
quantification and automation.
3. Capillary isoelectric focusing (CIEF)
The CIEF is used to separate biological molecules mainly proteins, based on differences
between the isoelectric points (pI). This technique is performed by filling the capillary with a
mixture of ampholytes and the sample, and then forming a pH gradient (Slide 14). By applying
an electric field across the capillary with a basic solution at the cathode and an acidic solution at
the anode, the ampholytes and solutes migrate until they reach a region where their overall charge
is neutral (pH = pI). The ampholyte and solute zones remain extremely narrow because diffusion
to a zone of different pH results in the generation of charge and subsequent migration back to the
proper zone.
4. Micellar electrokinetic chromatography (MEKC)
By the MEKC is possible to separate neutral and charged solutes as well. Sodium
dodecylsulfate, or SDS, has a long-chain hydrophobic tail and a negatively charged ionic
functional group at its head (Slide 15). When the concentration of SDS is sufficiently large
micelles, with the hydrophobic moieties of the surfactant in the interior and the charged moieties
at the exterior, are formed. Micelles have. The separation of neutral molecules is based on the
hydrophobic interaction of solutes with the micelles. The MEKC selectivity can be controlled by
the choice of surfactant and by the addition of modifiers to the buffer. Because micelles have a
negative charge, they migrate toward the cathode with a velocity less than the EOF velocity.
Neutral species partition themselves between the micelles and the buffer solution in a manner
similar to the partitioning of solutes between the two liquid phases in the HPLC. Hydrophilic
neutrals are insoluble in the micelle’s hydrophobic inner environment and elute as a single band,
as they would in the CZE. Neutral solutes that are extremely hydrophobic are completely soluble
in the micelle and elute with the micelles as a single band. Those neutral species that exist in the
partition equilibrium between the buffer solution and the micelles elute between the completely
hydrophilic and completely hydrophobic neutral species. The MECK is used to separate a wide
variety of samples, including mixtures of pharmaceutical compounds, vitamins, and explosives.
5. Isotachophoresis (ITP)
The ITP uses two different buffer systems where solutes are sandwiched between leading and
trailing electrolytes, creating a steady state in which the solute zones migrate in order of
decreasing mobility. Two unique aspects of the ITP are that all solute zones migrate at the same
velocity and that they all adopt the concentration of the leading electrolyte. The latter aspect
makes ITP a very useful technique for the analysis of dilute solutions. Samples can be
concentrated to many orders of magnitude.
6. Capillary electrochromatography (CEC)
In the CEC the capillary is packed with a stationary phase and when an electric field is
applied the EOF moves the mobile phase through the packed column (slide 16). The separation
selectivity depends on partition between the stationary and mobile phases. This allows high
efficiency separation of neutral compounds within very short analysis times. By the CEC is
enabled separation of closely related compounds, shorter run times, and increased sample
transience. As in liquid chromatography, the CEC separations are carried out by the partition of a
solute between the mobile and stationary phases. The additional separation by motilities resolves
charged and neutral components.
7. Capillary electrophoretic immunoassay (CEIA)
Coupling competitive immunoassays with CE-driven bound/free fraction separation has
recently gained considerable attention, particularly whenever fluorescent tracers are used. Both,
the CZE or the MECC, can be used to separate the bound from the free fraction of the tracer in a
competitive immunoassay, allowing indirect but highly sensitive quantitation of several analytes
in a variety of biological samples. Additionally, since different tracers can be easily spotted in
one electropherogram, multicomponent immunoassays can be simultaneously carried out in the
same experiment. Many commercial immunoassay reagents can be adapted to CEIA.
Applications of the capillary electrophoresis
Capillary electrophoresis has had extensive applications in a variety of scientific domains,
including chemistry, biochemistry, pharmacology, toxicology, forensic toxicology, forensic
biology/biochemistry (DNA profiling, analysis of proteins and other biological compounds), and
biomedicine as well5, 6, 7, 8
.
References
1
Whatley H. Basic principles of capillary electrophoresis. In Clinical and Forensic Applications
of Capillary Electrophoresis. Petersen, J., Mohammad, A.A. (Eds.), Humana Press Inc. pp 21-58
(2001).
2
Harstad, R.K., Johnson, A.C.,Weisenberger, M.M.,Bowser, M.T. Capillary Electrophoresis.
Anal. Chem., 88, 299–319 (2016) DOI: 10.1021/acs.analchem.5b04125
3
Karger, B.L. Guttman, A. DNA Sequencing by Capillary Electrophoresis. Electrophoresis. 30,
S196–S202. (2009) doi: 10.1002/elps.200900218
4
Wuethrich, A., Quirino J.P. Derivatisation for separation and detection in capillary
electrophoresis (2015-2017). Electrophoresis. 39, 82-96 (2018), doi: 10.1002/elps.201700252.
5
Tavares, M. F. M., Jager A., da Silva, C.L.; Morales E.P. Applications of capillary
electrophoresis to the analysis of compounds of clinical, forensic, cosmetological, environmental,
nutritional and pharmaceutical importance. J. Braz. Chem. Soc. 14, 281-290 (2003).
6
Thormann, W. et al. Clinical and Forensic Drug Toxicology. In Clinical and Forensic
Applications of Capillary Electrophoresis. Petersen, J., Mohammad, A. (Eds.), Humana Press
Inc. 397-422 (2001).
7
Sarr, S.O., et al. (2016) Validation of a Capillary Electrophoresis method for analyzing
chlorphena- mine maleate-based drugs using the accuracy profile approach. Am. J Anal. Chem.,
7, 452-459 (2016).
8
Ticova A., Prikryl J. Reproducible preparation of nanospray tips for capillary electrophoresis
coupled to mass spectrometry using 3D printed grinding device, Electrophoresis, 37, 924-930
(2016).

More Related Content

What's hot

Electrophoresis and various techniques
Electrophoresis and various techniquesElectrophoresis and various techniques
Electrophoresis and various techniquesROHIT
 
Affinity Chromatography.pdf
Affinity Chromatography.pdfAffinity Chromatography.pdf
Affinity Chromatography.pdfMohiniTawade
 
PRINCIPLES of FT-NMR & 13C NMR
PRINCIPLES of FT-NMR & 13C NMRPRINCIPLES of FT-NMR & 13C NMR
PRINCIPLES of FT-NMR & 13C NMRAditya Sharma
 
Mass spectroscopy, Ionization techniques and types of mass analyzers
Mass spectroscopy, Ionization techniques and types of mass analyzers Mass spectroscopy, Ionization techniques and types of mass analyzers
Mass spectroscopy, Ionization techniques and types of mass analyzers Muhammad Asif Shaheeen
 
MASS SPECTROSCOPY & ITS INSTRUMENTATION
MASS SPECTROSCOPY & ITS INSTRUMENTATIONMASS SPECTROSCOPY & ITS INSTRUMENTATION
MASS SPECTROSCOPY & ITS INSTRUMENTATIONROHIT
 
Difference & derivative spectrometry
Difference & derivative spectrometryDifference & derivative spectrometry
Difference & derivative spectrometryaishuanju
 
Mass spectrometry and ionization techniques
Mass spectrometry and ionization techniquesMass spectrometry and ionization techniques
Mass spectrometry and ionization techniquesSurbhi Narang
 
Mc lafferty rearrangement
Mc lafferty rearrangementMc lafferty rearrangement
Mc lafferty rearrangementMahendra G S
 

What's hot (20)

Electrophoresis and various techniques
Electrophoresis and various techniquesElectrophoresis and various techniques
Electrophoresis and various techniques
 
Capillary Electrophoresis
Capillary ElectrophoresisCapillary Electrophoresis
Capillary Electrophoresis
 
Mass spectroscopy
Mass spectroscopyMass spectroscopy
Mass spectroscopy
 
Affinity Chromatography.pdf
Affinity Chromatography.pdfAffinity Chromatography.pdf
Affinity Chromatography.pdf
 
Capillary electrophoresis
Capillary electrophoresisCapillary electrophoresis
Capillary electrophoresis
 
Mass Analyser
Mass AnalyserMass Analyser
Mass Analyser
 
FT NMR
FT NMRFT NMR
FT NMR
 
PRINCIPLES of FT-NMR & 13C NMR
PRINCIPLES of FT-NMR & 13C NMRPRINCIPLES of FT-NMR & 13C NMR
PRINCIPLES of FT-NMR & 13C NMR
 
Factors Affecting Resolution In HPLC.
Factors Affecting Resolution In HPLC.Factors Affecting Resolution In HPLC.
Factors Affecting Resolution In HPLC.
 
FT- NMR
FT- NMRFT- NMR
FT- NMR
 
DETECTORS IN HPLC
DETECTORS IN HPLCDETECTORS IN HPLC
DETECTORS IN HPLC
 
Mass spectroscopy, Ionization techniques and types of mass analyzers
Mass spectroscopy, Ionization techniques and types of mass analyzers Mass spectroscopy, Ionization techniques and types of mass analyzers
Mass spectroscopy, Ionization techniques and types of mass analyzers
 
MASS SPECTROSCOPY & ITS INSTRUMENTATION
MASS SPECTROSCOPY & ITS INSTRUMENTATIONMASS SPECTROSCOPY & ITS INSTRUMENTATION
MASS SPECTROSCOPY & ITS INSTRUMENTATION
 
Ft ir instrumentation
Ft ir instrumentationFt ir instrumentation
Ft ir instrumentation
 
Cosy,nosy
Cosy,nosyCosy,nosy
Cosy,nosy
 
Difference & derivative spectrometry
Difference & derivative spectrometryDifference & derivative spectrometry
Difference & derivative spectrometry
 
Mass spectrometry and ionization techniques
Mass spectrometry and ionization techniquesMass spectrometry and ionization techniques
Mass spectrometry and ionization techniques
 
Qsar
QsarQsar
Qsar
 
Mc lafferty rearrangement
Mc lafferty rearrangementMc lafferty rearrangement
Mc lafferty rearrangement
 
PAPER ELECTROPHORESIS.
PAPER ELECTROPHORESIS.PAPER ELECTROPHORESIS.
PAPER ELECTROPHORESIS.
 

Similar to The principle and performance of capillary electrophoresis

Notes for The principle and performance of capillary electrophoresis
Notes for The principle and performance of capillary electrophoresisNotes for The principle and performance of capillary electrophoresis
Notes for The principle and performance of capillary electrophoresisimprovemed
 
Electrophoresis ppt.
Electrophoresis ppt.Electrophoresis ppt.
Electrophoresis ppt.gulamrafey
 
Capillary electrophoresis
Capillary electrophoresisCapillary electrophoresis
Capillary electrophoresisPratik Parikh
 
Capillary electrophoresis final ppt.
Capillary electrophoresis final ppt.Capillary electrophoresis final ppt.
Capillary electrophoresis final ppt.Sandhya Talla
 
General considerations and method development in ce,
General considerations and method development in ce,General considerations and method development in ce,
General considerations and method development in ce,ChowdaryPavani
 
Capillary Electrophoresis-for teaching.pptx
Capillary Electrophoresis-for teaching.pptxCapillary Electrophoresis-for teaching.pptx
Capillary Electrophoresis-for teaching.pptxHassanGhaziaskar
 
1 electrophoresis 1.pptx
1  electrophoresis 1.pptx1  electrophoresis 1.pptx
1 electrophoresis 1.pptxGokulanTV
 
CP Capillary electrophoresis (1).pdf
CP Capillary electrophoresis (1).pdfCP Capillary electrophoresis (1).pdf
CP Capillary electrophoresis (1).pdfMemerHnYar
 
ELECTROPHORESIS.pptx
ELECTROPHORESIS.pptxELECTROPHORESIS.pptx
ELECTROPHORESIS.pptxRajesh Yadav
 
Experiment 4 Electropolymerized Conducting Polymers. In.docx
Experiment 4 Electropolymerized Conducting Polymers.         In.docxExperiment 4 Electropolymerized Conducting Polymers.         In.docx
Experiment 4 Electropolymerized Conducting Polymers. In.docxgitagrimston
 
Microchip electrophoresis
Microchip electrophoresisMicrochip electrophoresis
Microchip electrophoresisArun Babu
 

Similar to The principle and performance of capillary electrophoresis (20)

Notes for The principle and performance of capillary electrophoresis
Notes for The principle and performance of capillary electrophoresisNotes for The principle and performance of capillary electrophoresis
Notes for The principle and performance of capillary electrophoresis
 
Electrophoresis ppt.
Electrophoresis ppt.Electrophoresis ppt.
Electrophoresis ppt.
 
Capillary electrophoresis
Capillary electrophoresisCapillary electrophoresis
Capillary electrophoresis
 
Electrophoresis
ElectrophoresisElectrophoresis
Electrophoresis
 
Capillary electrophoresis
Capillary electrophoresisCapillary electrophoresis
Capillary electrophoresis
 
Capillary electrophoresis final ppt.
Capillary electrophoresis final ppt.Capillary electrophoresis final ppt.
Capillary electrophoresis final ppt.
 
General considerations and method development in ce,
General considerations and method development in ce,General considerations and method development in ce,
General considerations and method development in ce,
 
Capillary Electrophoresis-for teaching.pptx
Capillary Electrophoresis-for teaching.pptxCapillary Electrophoresis-for teaching.pptx
Capillary Electrophoresis-for teaching.pptx
 
ce ms.pptx
ce ms.pptxce ms.pptx
ce ms.pptx
 
1 electrophoresis 1.pptx
1  electrophoresis 1.pptx1  electrophoresis 1.pptx
1 electrophoresis 1.pptx
 
CP Capillary electrophoresis (1).pdf
CP Capillary electrophoresis (1).pdfCP Capillary electrophoresis (1).pdf
CP Capillary electrophoresis (1).pdf
 
Electrophoresis
ElectrophoresisElectrophoresis
Electrophoresis
 
ELECTROPHORESIS.pptx
ELECTROPHORESIS.pptxELECTROPHORESIS.pptx
ELECTROPHORESIS.pptx
 
Electrophoresis
ElectrophoresisElectrophoresis
Electrophoresis
 
Experiment 4 Electropolymerized Conducting Polymers. In.docx
Experiment 4 Electropolymerized Conducting Polymers.         In.docxExperiment 4 Electropolymerized Conducting Polymers.         In.docx
Experiment 4 Electropolymerized Conducting Polymers. In.docx
 
Pratikshaaa
PratikshaaaPratikshaaa
Pratikshaaa
 
Biochem asgnmnt
Biochem asgnmntBiochem asgnmnt
Biochem asgnmnt
 
Microchip electrophoresis
Microchip electrophoresisMicrochip electrophoresis
Microchip electrophoresis
 
PAPER ELECTROPHORESIS.pptx
PAPER ELECTROPHORESIS.pptxPAPER ELECTROPHORESIS.pptx
PAPER ELECTROPHORESIS.pptx
 
Other electrophoresis
Other electrophoresisOther electrophoresis
Other electrophoresis
 

More from improvemed

2019 2020 predavanje letenje, ronjenje drenjancevic
2019 2020 predavanje letenje, ronjenje drenjancevic2019 2020 predavanje letenje, ronjenje drenjancevic
2019 2020 predavanje letenje, ronjenje drenjancevicimprovemed
 
In vitro models of hepatotoxicity
In vitro models of hepatotoxicityIn vitro models of hepatotoxicity
In vitro models of hepatotoxicityimprovemed
 
Etiology of liver diseases
Etiology of liver diseasesEtiology of liver diseases
Etiology of liver diseasesimprovemed
 
An introduction to experimental epidemiology
An introduction to experimental epidemiology An introduction to experimental epidemiology
An introduction to experimental epidemiology improvemed
 
Genotyping methods of nosocomial infections pathogen
Genotyping methods of nosocomial infections pathogenGenotyping methods of nosocomial infections pathogen
Genotyping methods of nosocomial infections pathogenimprovemed
 
Use of MALDI-TOF in the diagnosis of infectious diseases
Use of MALDI-TOF in the diagnosis of infectious diseasesUse of MALDI-TOF in the diagnosis of infectious diseases
Use of MALDI-TOF in the diagnosis of infectious diseasesimprovemed
 
Molecular microbiology methods
Molecular microbiology methodsMolecular microbiology methods
Molecular microbiology methodsimprovemed
 
Isolated vascular rings
Isolated vascular ringsIsolated vascular rings
Isolated vascular ringsimprovemed
 
Isolated blood vessels
Isolated blood vesselsIsolated blood vessels
Isolated blood vesselsimprovemed
 
Notes for Measuring blood flow and reactivity of the blood vessels in the ski...
Notes for Measuring blood flow and reactivity of the blood vessels in the ski...Notes for Measuring blood flow and reactivity of the blood vessels in the ski...
Notes for Measuring blood flow and reactivity of the blood vessels in the ski...improvemed
 
Notes for STAINING AND ANALYSIS of HISTOLOGICAL PREPARATIONS
Notes for STAINING AND ANALYSIS of HISTOLOGICAL PREPARATIONSNotes for STAINING AND ANALYSIS of HISTOLOGICAL PREPARATIONS
Notes for STAINING AND ANALYSIS of HISTOLOGICAL PREPARATIONSimprovemed
 
Notes for Fixation of tissues and organs for educational and scientific purposes
Notes for Fixation of tissues and organs for educational and scientific purposesNotes for Fixation of tissues and organs for educational and scientific purposes
Notes for Fixation of tissues and organs for educational and scientific purposesimprovemed
 
Notes for The principle and performance of liquid chromatography–mass spectro...
Notes for The principle and performance of liquid chromatography–mass spectro...Notes for The principle and performance of liquid chromatography–mass spectro...
Notes for The principle and performance of liquid chromatography–mass spectro...improvemed
 
Notes for Cell Culture Basic Techniques
Notes for Cell Culture Basic TechniquesNotes for Cell Culture Basic Techniques
Notes for Cell Culture Basic Techniquesimprovemed
 
Systems biology for Medicine' is 'Experimental methods and the big datasets
Systems biology for Medicine' is 'Experimental methods and the big datasetsSystems biology for Medicine' is 'Experimental methods and the big datasets
Systems biology for Medicine' is 'Experimental methods and the big datasetsimprovemed
 
Systems biology for medical students/Systems medicine
Systems biology for medical students/Systems medicineSystems biology for medical students/Systems medicine
Systems biology for medical students/Systems medicineimprovemed
 
Basic course for computer based methods
Basic course for computer based methodsBasic course for computer based methods
Basic course for computer based methodsimprovemed
 

More from improvemed (20)

2019 2020 predavanje letenje, ronjenje drenjancevic
2019 2020 predavanje letenje, ronjenje drenjancevic2019 2020 predavanje letenje, ronjenje drenjancevic
2019 2020 predavanje letenje, ronjenje drenjancevic
 
In vitro models of hepatotoxicity
In vitro models of hepatotoxicityIn vitro models of hepatotoxicity
In vitro models of hepatotoxicity
 
Etiology of liver diseases
Etiology of liver diseasesEtiology of liver diseases
Etiology of liver diseases
 
An introduction to experimental epidemiology
An introduction to experimental epidemiology An introduction to experimental epidemiology
An introduction to experimental epidemiology
 
Genotyping methods of nosocomial infections pathogen
Genotyping methods of nosocomial infections pathogenGenotyping methods of nosocomial infections pathogen
Genotyping methods of nosocomial infections pathogen
 
Use of MALDI-TOF in the diagnosis of infectious diseases
Use of MALDI-TOF in the diagnosis of infectious diseasesUse of MALDI-TOF in the diagnosis of infectious diseases
Use of MALDI-TOF in the diagnosis of infectious diseases
 
Molecular microbiology methods
Molecular microbiology methodsMolecular microbiology methods
Molecular microbiology methods
 
Isolated vascular rings
Isolated vascular ringsIsolated vascular rings
Isolated vascular rings
 
Isolated blood vessels
Isolated blood vesselsIsolated blood vessels
Isolated blood vessels
 
Notes for Measuring blood flow and reactivity of the blood vessels in the ski...
Notes for Measuring blood flow and reactivity of the blood vessels in the ski...Notes for Measuring blood flow and reactivity of the blood vessels in the ski...
Notes for Measuring blood flow and reactivity of the blood vessels in the ski...
 
Notes for STAINING AND ANALYSIS of HISTOLOGICAL PREPARATIONS
Notes for STAINING AND ANALYSIS of HISTOLOGICAL PREPARATIONSNotes for STAINING AND ANALYSIS of HISTOLOGICAL PREPARATIONS
Notes for STAINING AND ANALYSIS of HISTOLOGICAL PREPARATIONS
 
Notes for Fixation of tissues and organs for educational and scientific purposes
Notes for Fixation of tissues and organs for educational and scientific purposesNotes for Fixation of tissues and organs for educational and scientific purposes
Notes for Fixation of tissues and organs for educational and scientific purposes
 
Notes for
Notes for Notes for
Notes for
 
Notes for The principle and performance of liquid chromatography–mass spectro...
Notes for The principle and performance of liquid chromatography–mass spectro...Notes for The principle and performance of liquid chromatography–mass spectro...
Notes for The principle and performance of liquid chromatography–mass spectro...
 
Notes for Cell Culture Basic Techniques
Notes for Cell Culture Basic TechniquesNotes for Cell Culture Basic Techniques
Notes for Cell Culture Basic Techniques
 
Big datasets
Big datasetsBig datasets
Big datasets
 
Systems biology for Medicine' is 'Experimental methods and the big datasets
Systems biology for Medicine' is 'Experimental methods and the big datasetsSystems biology for Medicine' is 'Experimental methods and the big datasets
Systems biology for Medicine' is 'Experimental methods and the big datasets
 
Systems biology for medical students/Systems medicine
Systems biology for medical students/Systems medicineSystems biology for medical students/Systems medicine
Systems biology for medical students/Systems medicine
 
Use cases
Use casesUse cases
Use cases
 
Basic course for computer based methods
Basic course for computer based methodsBasic course for computer based methods
Basic course for computer based methods
 

Recently uploaded

Coimbatore Call Girls in Thudiyalur : 7427069034 High Profile Model Escorts |...
Coimbatore Call Girls in Thudiyalur : 7427069034 High Profile Model Escorts |...Coimbatore Call Girls in Thudiyalur : 7427069034 High Profile Model Escorts |...
Coimbatore Call Girls in Thudiyalur : 7427069034 High Profile Model Escorts |...chennailover
 
💚Call Girls In Amritsar 💯Anvi 📲🔝8725944379🔝Amritsar Call Girl No💰Advance Cash...
💚Call Girls In Amritsar 💯Anvi 📲🔝8725944379🔝Amritsar Call Girl No💰Advance Cash...💚Call Girls In Amritsar 💯Anvi 📲🔝8725944379🔝Amritsar Call Girl No💰Advance Cash...
💚Call Girls In Amritsar 💯Anvi 📲🔝8725944379🔝Amritsar Call Girl No💰Advance Cash...Sheetaleventcompany
 
Call Girls Mysore Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Mysore Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Mysore Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Mysore Just Call 8250077686 Top Class Call Girl Service AvailableDipal Arora
 
Call Girls Service Jaipur {9521753030} ❤️VVIP RIDDHI Call Girl in Jaipur Raja...
Call Girls Service Jaipur {9521753030} ❤️VVIP RIDDHI Call Girl in Jaipur Raja...Call Girls Service Jaipur {9521753030} ❤️VVIP RIDDHI Call Girl in Jaipur Raja...
Call Girls Service Jaipur {9521753030} ❤️VVIP RIDDHI Call Girl in Jaipur Raja...Sheetaleventcompany
 
Most Beautiful Call Girl in Bangalore Contact on Whatsapp
Most Beautiful Call Girl in Bangalore Contact on WhatsappMost Beautiful Call Girl in Bangalore Contact on Whatsapp
Most Beautiful Call Girl in Bangalore Contact on WhatsappInaaya Sharma
 
Top Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any Time
Top Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any TimeTop Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any Time
Top Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any TimeCall Girls Delhi
 
Call Girls Kolkata Kalikapur 💯Call Us 🔝 8005736733 🔝 💃 Top Class Call Girl Se...
Call Girls Kolkata Kalikapur 💯Call Us 🔝 8005736733 🔝 💃 Top Class Call Girl Se...Call Girls Kolkata Kalikapur 💯Call Us 🔝 8005736733 🔝 💃 Top Class Call Girl Se...
Call Girls Kolkata Kalikapur 💯Call Us 🔝 8005736733 🔝 💃 Top Class Call Girl Se...Namrata Singh
 
(Low Rate RASHMI ) Rate Of Call Girls Jaipur ❣ 8445551418 ❣ Elite Models & Ce...
(Low Rate RASHMI ) Rate Of Call Girls Jaipur ❣ 8445551418 ❣ Elite Models & Ce...(Low Rate RASHMI ) Rate Of Call Girls Jaipur ❣ 8445551418 ❣ Elite Models & Ce...
(Low Rate RASHMI ) Rate Of Call Girls Jaipur ❣ 8445551418 ❣ Elite Models & Ce...parulsinha
 
Call Girls Amritsar Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Amritsar Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Amritsar Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Amritsar Just Call 8250077686 Top Class Call Girl Service AvailableDipal Arora
 
Top Rated Hyderabad Call Girls Erragadda ⟟ 9332606886 ⟟ Call Me For Genuine ...
Top Rated  Hyderabad Call Girls Erragadda ⟟ 9332606886 ⟟ Call Me For Genuine ...Top Rated  Hyderabad Call Girls Erragadda ⟟ 9332606886 ⟟ Call Me For Genuine ...
Top Rated Hyderabad Call Girls Erragadda ⟟ 9332606886 ⟟ Call Me For Genuine ...chandars293
 
Dehradun Call Girls Service {8854095900} ❤️VVIP ROCKY Call Girl in Dehradun U...
Dehradun Call Girls Service {8854095900} ❤️VVIP ROCKY Call Girl in Dehradun U...Dehradun Call Girls Service {8854095900} ❤️VVIP ROCKY Call Girl in Dehradun U...
Dehradun Call Girls Service {8854095900} ❤️VVIP ROCKY Call Girl in Dehradun U...Sheetaleventcompany
 
Models Call Girls In Hyderabad 9630942363 Hyderabad Call Girl & Hyderabad Esc...
Models Call Girls In Hyderabad 9630942363 Hyderabad Call Girl & Hyderabad Esc...Models Call Girls In Hyderabad 9630942363 Hyderabad Call Girl & Hyderabad Esc...
Models Call Girls In Hyderabad 9630942363 Hyderabad Call Girl & Hyderabad Esc...GENUINE ESCORT AGENCY
 
VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋
VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋
VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋TANUJA PANDEY
 
Top Rated Call Girls Kerala ☎ 8250092165👄 Delivery in 20 Mins Near Me
Top Rated Call Girls Kerala ☎ 8250092165👄 Delivery in 20 Mins Near MeTop Rated Call Girls Kerala ☎ 8250092165👄 Delivery in 20 Mins Near Me
Top Rated Call Girls Kerala ☎ 8250092165👄 Delivery in 20 Mins Near Mechennailover
 
Call Girls Service Jaipur {9521753030 } ❤️VVIP BHAWNA Call Girl in Jaipur Raj...
Call Girls Service Jaipur {9521753030 } ❤️VVIP BHAWNA Call Girl in Jaipur Raj...Call Girls Service Jaipur {9521753030 } ❤️VVIP BHAWNA Call Girl in Jaipur Raj...
Call Girls Service Jaipur {9521753030 } ❤️VVIP BHAWNA Call Girl in Jaipur Raj...khalifaescort01
 
Call Girls Ahmedabad Just Call 9630942363 Top Class Call Girl Service Available
Call Girls Ahmedabad Just Call 9630942363 Top Class Call Girl Service AvailableCall Girls Ahmedabad Just Call 9630942363 Top Class Call Girl Service Available
Call Girls Ahmedabad Just Call 9630942363 Top Class Call Girl Service AvailableGENUINE ESCORT AGENCY
 
Call Girls Jaipur Just Call 9521753030 Top Class Call Girl Service Available
Call Girls Jaipur Just Call 9521753030 Top Class Call Girl Service AvailableCall Girls Jaipur Just Call 9521753030 Top Class Call Girl Service Available
Call Girls Jaipur Just Call 9521753030 Top Class Call Girl Service AvailableJanvi Singh
 
Call Girls Hyderabad Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Hyderabad Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Hyderabad Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Hyderabad Just Call 8250077686 Top Class Call Girl Service AvailableDipal Arora
 
Russian Call Girls Service Jaipur {8445551418} ❤️PALLAVI VIP Jaipur Call Gir...
Russian Call Girls Service  Jaipur {8445551418} ❤️PALLAVI VIP Jaipur Call Gir...Russian Call Girls Service  Jaipur {8445551418} ❤️PALLAVI VIP Jaipur Call Gir...
Russian Call Girls Service Jaipur {8445551418} ❤️PALLAVI VIP Jaipur Call Gir...parulsinha
 
Call Girls in Lucknow Just Call 👉👉7877925207 Top Class Call Girl Service Avai...
Call Girls in Lucknow Just Call 👉👉7877925207 Top Class Call Girl Service Avai...Call Girls in Lucknow Just Call 👉👉7877925207 Top Class Call Girl Service Avai...
Call Girls in Lucknow Just Call 👉👉7877925207 Top Class Call Girl Service Avai...adilkhan87451
 

Recently uploaded (20)

Coimbatore Call Girls in Thudiyalur : 7427069034 High Profile Model Escorts |...
Coimbatore Call Girls in Thudiyalur : 7427069034 High Profile Model Escorts |...Coimbatore Call Girls in Thudiyalur : 7427069034 High Profile Model Escorts |...
Coimbatore Call Girls in Thudiyalur : 7427069034 High Profile Model Escorts |...
 
💚Call Girls In Amritsar 💯Anvi 📲🔝8725944379🔝Amritsar Call Girl No💰Advance Cash...
💚Call Girls In Amritsar 💯Anvi 📲🔝8725944379🔝Amritsar Call Girl No💰Advance Cash...💚Call Girls In Amritsar 💯Anvi 📲🔝8725944379🔝Amritsar Call Girl No💰Advance Cash...
💚Call Girls In Amritsar 💯Anvi 📲🔝8725944379🔝Amritsar Call Girl No💰Advance Cash...
 
Call Girls Mysore Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Mysore Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Mysore Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Mysore Just Call 8250077686 Top Class Call Girl Service Available
 
Call Girls Service Jaipur {9521753030} ❤️VVIP RIDDHI Call Girl in Jaipur Raja...
Call Girls Service Jaipur {9521753030} ❤️VVIP RIDDHI Call Girl in Jaipur Raja...Call Girls Service Jaipur {9521753030} ❤️VVIP RIDDHI Call Girl in Jaipur Raja...
Call Girls Service Jaipur {9521753030} ❤️VVIP RIDDHI Call Girl in Jaipur Raja...
 
Most Beautiful Call Girl in Bangalore Contact on Whatsapp
Most Beautiful Call Girl in Bangalore Contact on WhatsappMost Beautiful Call Girl in Bangalore Contact on Whatsapp
Most Beautiful Call Girl in Bangalore Contact on Whatsapp
 
Top Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any Time
Top Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any TimeTop Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any Time
Top Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any Time
 
Call Girls Kolkata Kalikapur 💯Call Us 🔝 8005736733 🔝 💃 Top Class Call Girl Se...
Call Girls Kolkata Kalikapur 💯Call Us 🔝 8005736733 🔝 💃 Top Class Call Girl Se...Call Girls Kolkata Kalikapur 💯Call Us 🔝 8005736733 🔝 💃 Top Class Call Girl Se...
Call Girls Kolkata Kalikapur 💯Call Us 🔝 8005736733 🔝 💃 Top Class Call Girl Se...
 
(Low Rate RASHMI ) Rate Of Call Girls Jaipur ❣ 8445551418 ❣ Elite Models & Ce...
(Low Rate RASHMI ) Rate Of Call Girls Jaipur ❣ 8445551418 ❣ Elite Models & Ce...(Low Rate RASHMI ) Rate Of Call Girls Jaipur ❣ 8445551418 ❣ Elite Models & Ce...
(Low Rate RASHMI ) Rate Of Call Girls Jaipur ❣ 8445551418 ❣ Elite Models & Ce...
 
Call Girls Amritsar Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Amritsar Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Amritsar Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Amritsar Just Call 8250077686 Top Class Call Girl Service Available
 
Top Rated Hyderabad Call Girls Erragadda ⟟ 9332606886 ⟟ Call Me For Genuine ...
Top Rated  Hyderabad Call Girls Erragadda ⟟ 9332606886 ⟟ Call Me For Genuine ...Top Rated  Hyderabad Call Girls Erragadda ⟟ 9332606886 ⟟ Call Me For Genuine ...
Top Rated Hyderabad Call Girls Erragadda ⟟ 9332606886 ⟟ Call Me For Genuine ...
 
Dehradun Call Girls Service {8854095900} ❤️VVIP ROCKY Call Girl in Dehradun U...
Dehradun Call Girls Service {8854095900} ❤️VVIP ROCKY Call Girl in Dehradun U...Dehradun Call Girls Service {8854095900} ❤️VVIP ROCKY Call Girl in Dehradun U...
Dehradun Call Girls Service {8854095900} ❤️VVIP ROCKY Call Girl in Dehradun U...
 
Models Call Girls In Hyderabad 9630942363 Hyderabad Call Girl & Hyderabad Esc...
Models Call Girls In Hyderabad 9630942363 Hyderabad Call Girl & Hyderabad Esc...Models Call Girls In Hyderabad 9630942363 Hyderabad Call Girl & Hyderabad Esc...
Models Call Girls In Hyderabad 9630942363 Hyderabad Call Girl & Hyderabad Esc...
 
VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋
VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋
VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋
 
Top Rated Call Girls Kerala ☎ 8250092165👄 Delivery in 20 Mins Near Me
Top Rated Call Girls Kerala ☎ 8250092165👄 Delivery in 20 Mins Near MeTop Rated Call Girls Kerala ☎ 8250092165👄 Delivery in 20 Mins Near Me
Top Rated Call Girls Kerala ☎ 8250092165👄 Delivery in 20 Mins Near Me
 
Call Girls Service Jaipur {9521753030 } ❤️VVIP BHAWNA Call Girl in Jaipur Raj...
Call Girls Service Jaipur {9521753030 } ❤️VVIP BHAWNA Call Girl in Jaipur Raj...Call Girls Service Jaipur {9521753030 } ❤️VVIP BHAWNA Call Girl in Jaipur Raj...
Call Girls Service Jaipur {9521753030 } ❤️VVIP BHAWNA Call Girl in Jaipur Raj...
 
Call Girls Ahmedabad Just Call 9630942363 Top Class Call Girl Service Available
Call Girls Ahmedabad Just Call 9630942363 Top Class Call Girl Service AvailableCall Girls Ahmedabad Just Call 9630942363 Top Class Call Girl Service Available
Call Girls Ahmedabad Just Call 9630942363 Top Class Call Girl Service Available
 
Call Girls Jaipur Just Call 9521753030 Top Class Call Girl Service Available
Call Girls Jaipur Just Call 9521753030 Top Class Call Girl Service AvailableCall Girls Jaipur Just Call 9521753030 Top Class Call Girl Service Available
Call Girls Jaipur Just Call 9521753030 Top Class Call Girl Service Available
 
Call Girls Hyderabad Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Hyderabad Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Hyderabad Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Hyderabad Just Call 8250077686 Top Class Call Girl Service Available
 
Russian Call Girls Service Jaipur {8445551418} ❤️PALLAVI VIP Jaipur Call Gir...
Russian Call Girls Service  Jaipur {8445551418} ❤️PALLAVI VIP Jaipur Call Gir...Russian Call Girls Service  Jaipur {8445551418} ❤️PALLAVI VIP Jaipur Call Gir...
Russian Call Girls Service Jaipur {8445551418} ❤️PALLAVI VIP Jaipur Call Gir...
 
Call Girls in Lucknow Just Call 👉👉7877925207 Top Class Call Girl Service Avai...
Call Girls in Lucknow Just Call 👉👉7877925207 Top Class Call Girl Service Avai...Call Girls in Lucknow Just Call 👉👉7877925207 Top Class Call Girl Service Avai...
Call Girls in Lucknow Just Call 👉👉7877925207 Top Class Call Girl Service Avai...
 

The principle and performance of capillary electrophoresis

  • 1. The principle and performance of capillary electrophoresis Ljubica Glavaš-Obrovac Introduction Capillary electrophoresis (CE) is a relatively new and a very versatile technique (also known as high-performance capillary electrophoresis) for the separation and analysis of a variety of molecules – from small inorganic ions to huge biopolymers. Compared to the high- performance liquid chromatography, and gas chromatography CE offers several advantages, including highly efficient and fast separation, relatively inexpensive and long lasting capillary columns, small sample size requirements and low reagent consumption. It can be used for analysis of polar ionic, polar non-ionic and non-polar non-ionic compounds, as well as high molecular weight biomolecules and chiral compounds. Further, CE separations are highly efficient (N>105 to 106 ), sensitive in terms of mass and concentration, fast (usually less than 30 min), simple, and require extremely small amounts of solvents and other consumables. This highly sophisticated, user friendly and very effective CE has become commercially available at affordable costs (Slide 2). Capillary electrophoresis theory When a sample is injected into a capillary tube and an electric field is applied across the capillary tube, sample’s components migrate as the result of two types of action: electrophoretic mobility and electroosmotic mobility. Electrophoretic mobility is the solute’s response to the applied electrical field. Cations move toward the negatively charged cathode, anions move toward the positively charged anode, and neutral species remain stationary. The other contribution to a solute’s migration is electroosmotic flow, which occurs when the buffer moves through the capillary in response to the applied electrical field. Under normal conditions the buffer moves toward the cathode, sweeping most solutes, including the anions and neutral species, toward the negatively charged cathode1 (Slide 3). Electrophoretic mobility The velocity with which a solute moves in response to the applied electric field (Slide 4) is called its electrophoretic velocity, νep; it is defined as vep = μepE where μep is the solute’s electrophoretic mobility, and E is the magnitude of the applied electrical field. A solute’s electrophoretic mobility is defined as μep = q6πηr where q is the solute’s charge, η is the buffer viscosity, and r is the solute’s radius.
  • 2. Electrophoretic mobility and, therefore, electrophoretic velocity, increases for more highly charged solutes and for solutes of smaller size. Because q is positive for a cation and negative for an anion, these species migrate in opposite directions. Neutral species, for which q is zero, have an electrophoretic velocity of zero. Electroosmotic flow (EOF) The EOF is caused by applying high-voltage to an electrolyte-filled capillary (Slide 5). This flow occurs when the buffer running through the silica capillary has a pH greater than 3 and the silanol (SiOH) groups lose a proton to become silanate (SiO- ) ions. The capillary wall then has a negative charge, which develops a double layer of cations attracted to it. The inner cation layer is stationary, while the outer layer is free to move along the capillary. The applied electric field causes the free cations to move toward the cathode creating a powerful bulk flow. The rate at which the buffer moves through the capillary, its electroosmotic flow velocity, νeof, is a function of the applied electric field, E, and the buffer’s electroosmotic mobility, μeof. νeof =μeofE Electroosmotic mobility is defined as Μeof = εζ4πη Electroosmotic flow is defined as μEOF=ϵ4πηEζ where ε is the dielectric constant of the solution, η is the viscosity of the solution, E is the field strength, and ζ is the zeta potential. Since the electrophoretic mobility is greater than the EOF, negatively charged particles, which are naturally attracted to the positively charged anode, will separate out as well. The EOF works best with a large zeta potential between the cation layers, a large diffuse layer of cations to drag more molecules towards the cathode, low resistance from the surrounding solution, and buffer with pH of 9 so that all the SiOH groups are ionized. The main benefit of the EOF is a separation in a single run, of both anions and cations, and also the separation of ions with quite different charge-to-size ratios in reasonable length of time as well. Capillary electrophoresis instrumentation The capillary electrophoresis instrument1, 2 (Slide 6) consists of: 1. Capillary 2. Injection system 3. High voltage source 4. Electrodes and electrode jars
  • 3. 5. Detector 1. Capillary In CE, the separation occurs in the capillary and this is also the compartment where injection and detection takes place. Ideally a capillary should be chemically and physically resistant, precisely machined with narrow internal diameters (d = 20-100 µm; length of 20-100 cm), does not adsorb solutes, transparent to UV-visible radiation, with a high thermal conductivity, and low cost. Fused silica capillaries with an external protective polyimide coating meet almost all these requirements and are hereof a standard choice in CE. Capillaries can be internally ‘uncoated’, having the inner silica surface in direct contact with buffers and solutes, but, under some experimental conditions, these may adsorb analytes. Alternatively, the capillaries can be internally coated with a thin layer of polymers (adsorbed or chemically bound) shielding the silica surface from interactions with solutes. Capillaries can be also filled with gels, mimicking slab gel electrophoresis, or with particles carrying a stationary phase, reproducing a chromatographic system. The most common coatings are polyacrylamide, cellulose, polyvinyl alcohol (PVA), amino acids, amines, surfactants, and poly(vinylpyrrolidinone) or polyethyleneimine. Polymers well known as liquid chromatographic (C2, Cg, C1g) or gas chromatographic stationary phases (polyethylene glycol (PEG), phenyl-methyl silicone) can be used for this purpose too (Slide 7). 2. Injection system Modern instrumentation uses two injection principles: hydrodynamic or electrokinetic. In the first case, a pressure difference is generated between the two ends of the capillary, while the injection end is dipped in the specimen vial. Alternatively, a voltage difference is established between the samples and the opposite vial, while samples are being injected. Although hydrodynamic injection is nonselective, electrokinetic injection (potential driven) is selective. This is because the sample components enter the capillary according to their electrophoretic mobility, the mobility and concentration of total ions in the sample and the EOF. Electrokinetic injection allows field-amplified sample stacking, i.e. a highly efficient method for increasing analytical sensitivity (in terms of concentration), to be accomplished (Slide 8). 3. High voltage source The high-voltage power supplies used for CE are generally able to give voltages up to 20-30 kV and currents up to 200-300 µA. Separations are generally carried out under constant potential, but voltage gradients or steps; and constant current separation is sometimes used. Because EOF in fused silica capillaries is usually directed towards the cathode, the common polarity in a capillary electropherograph is with the anode at the injection end of the capillary and the cathode at the opposite end, close to the detector; however, under specific analytical conditions, the polarity is reversed (Slide 8). 4. Electrodes and electrode jars
  • 4. Each side of the high voltage power supply is connected to an electrode. These electrodes help to induce an electric field to initiate the migration of the sample from the anode to the cathode through the capillary tube (Slide 8). 5. Detector In order to gain useful information from the separation technique, it is necessary to detect and measure analytes. Detection may be qualitative and/or quantitative. While the capillary itself is the detection cell it is linked to the detection device. It is also possible to couple to detectors that are outside of the separation capillary although this does require a specialized interface. The most frequently used method of detection involves absorbance of energy as the analytes move through a focused beam of light. The CE detection, most often based on UV-visible absorption, is carried out in-capillary, in order to avoid any possible post-separation added volumes. This can cause unacceptable band spreading (Slide 9). Indeed, the wall of fused silica capillaries, after removing the polyimide coating, is highly transparent. Thus, picogram amounts of analytes can easily be detected even by UV absorption, but, due to the limitations in the sample volume (nL) which can be injected, the concentration sensitivity with UV detectors is limited to 10-5 M. However, this limit in sensitivity can be overcome by using other detection techniques, e.g. laser-induced fluorescence (sensitivity up to 10-12 M) or electrochemical detection (sensitivity up to 10-8 M) and/or by adopting high efficiency sample stacking methods, which can assure improvements in sensitivity of 2-3 orders of magnitude. By implementation of diode-array or fast-scanning UV detectors allows on-line recording of the UV spectra of the peaks, improving the information content of the CE analysis. Laser induced fluorescence, conductometric detection, and mass spectrometric with electrospray ionization detection modes have recently become commercially available. Indirect detection is based on the addition into the running buffer of an ionic additive which is detectable at trace levels by the detector. This ionic compound, having the same charge and similar mobility to the analyte(s) of interest, is displaced from the zones where the analyte(s) migrate to preserve electroneutrality, and this displacement gives rise to ‘reversed peaks’, the area of which is proportional to the concentration of the given analyte(s). Drawbacks of this detection mode are sensitivity, which is usually lower than with the corresponding ‘direct’ mode, and the narrower range of linearity Capillary electroseparation techniques The most common CE separation techniques1, 2,3, 4 (Slide 10, Slide 11) are: 1. Capillary zone electrophoresis (CZE) 2. Micellar electrokinetic capillary chromatography (MECC or MEKC) 3. Capillary electrochromatography (CEC) 4. Capillary isoelectric focusing (CIEF)
  • 5. 5. Capillary gel electrophoresis (CGE) 6. Capillary isotachophoresis (CITP) 7. Capillary electrophoretic immunoassay (CEIA) 1. Capillary zone electrophoresis (CZE) The CZE, also known as free solution capillary electrophoresis, is very quickly, easy and the most commonly used mode of the CE (Slide 12). The separation technique is based on the differences in electrophoretic mobility, which is directed proportional to the charge on the molecule, and inversely proportional to the viscosity of the solvent and radius of the atom. Usually the end of the capillary containing the sample is the anode and solutes migrate toward the cathode at a velocity determined by their electrophoretic mobility and the EOF. In the CZE, in addition to the solutes, the buffer solution usually also moves through the capillary under the influence of an electric field. Coating the capillary’s walls with a nonionic reagent eliminates the EOF. In this form of the CZE cations migrate from the anode to the cathode. Anions elute into the source reservoir and neutral species remain stationary. The CZE provides effective separations of charged species, including inorganic anions and cations, organic acids and amines, and large biomolecules such as proteins. 2. Capillary gel electrophoresis (CGE) In the CGE a capillary is filled with a polymeric gel like cross-linked polyacrylamide, agarose, or solutions of linear polymers (Slide 13). While the gel is porous, a solute migrates through the gel with a velocity determined both by its electrophoretic mobility and by its size. The ability to effect a separation using size is helpful when the solutes have similar electrophoretic mobilities. For example, fragments of DNA of varying length have similar charge-to-size ratios, making their separation by CZE difficult. Because the DNA fragments are of different size they can be separated by the CGE. The capillary used for the CGE is usually treated to eliminate the EOF, preventing the gel’s extrusion from the capillary tubing. Samples are injected electrokinetically because the gel provides too much resistance for hydrodynamic sampling. The primary application of CGE is the separation of large biomolecules, including DNA fragments, proteins, and oligonucleotides. The main advantages over slab-gel electrophoresis are a wider range of gel matrix and composition, online detection, improved quantification and automation. 3. Capillary isoelectric focusing (CIEF) The CIEF is used to separate biological molecules mainly proteins, based on differences between the isoelectric points (pI). This technique is performed by filling the capillary with a mixture of ampholytes and the sample, and then forming a pH gradient (Slide 14). By applying an electric field across the capillary with a basic solution at the cathode and an acidic solution at the anode, the ampholytes and solutes migrate until they reach a region where their overall charge is neutral (pH = pI). The ampholyte and solute zones remain extremely narrow because diffusion
  • 6. to a zone of different pH results in the generation of charge and subsequent migration back to the proper zone. 4. Micellar electrokinetic chromatography (MEKC) By the MEKC is possible to separate neutral and charged solutes as well. Sodium dodecylsulfate, or SDS, has a long-chain hydrophobic tail and a negatively charged ionic functional group at its head (Slide 15). When the concentration of SDS is sufficiently large micelles, with the hydrophobic moieties of the surfactant in the interior and the charged moieties at the exterior, are formed. Micelles have. The separation of neutral molecules is based on the hydrophobic interaction of solutes with the micelles. The MEKC selectivity can be controlled by the choice of surfactant and by the addition of modifiers to the buffer. Because micelles have a negative charge, they migrate toward the cathode with a velocity less than the EOF velocity. Neutral species partition themselves between the micelles and the buffer solution in a manner similar to the partitioning of solutes between the two liquid phases in the HPLC. Hydrophilic neutrals are insoluble in the micelle’s hydrophobic inner environment and elute as a single band, as they would in the CZE. Neutral solutes that are extremely hydrophobic are completely soluble in the micelle and elute with the micelles as a single band. Those neutral species that exist in the partition equilibrium between the buffer solution and the micelles elute between the completely hydrophilic and completely hydrophobic neutral species. The MECK is used to separate a wide variety of samples, including mixtures of pharmaceutical compounds, vitamins, and explosives. 5. Isotachophoresis (ITP) The ITP uses two different buffer systems where solutes are sandwiched between leading and trailing electrolytes, creating a steady state in which the solute zones migrate in order of decreasing mobility. Two unique aspects of the ITP are that all solute zones migrate at the same velocity and that they all adopt the concentration of the leading electrolyte. The latter aspect makes ITP a very useful technique for the analysis of dilute solutions. Samples can be concentrated to many orders of magnitude. 6. Capillary electrochromatography (CEC) In the CEC the capillary is packed with a stationary phase and when an electric field is applied the EOF moves the mobile phase through the packed column (slide 16). The separation selectivity depends on partition between the stationary and mobile phases. This allows high efficiency separation of neutral compounds within very short analysis times. By the CEC is enabled separation of closely related compounds, shorter run times, and increased sample transience. As in liquid chromatography, the CEC separations are carried out by the partition of a solute between the mobile and stationary phases. The additional separation by motilities resolves charged and neutral components.
  • 7. 7. Capillary electrophoretic immunoassay (CEIA) Coupling competitive immunoassays with CE-driven bound/free fraction separation has recently gained considerable attention, particularly whenever fluorescent tracers are used. Both, the CZE or the MECC, can be used to separate the bound from the free fraction of the tracer in a competitive immunoassay, allowing indirect but highly sensitive quantitation of several analytes in a variety of biological samples. Additionally, since different tracers can be easily spotted in one electropherogram, multicomponent immunoassays can be simultaneously carried out in the same experiment. Many commercial immunoassay reagents can be adapted to CEIA. Applications of the capillary electrophoresis Capillary electrophoresis has had extensive applications in a variety of scientific domains, including chemistry, biochemistry, pharmacology, toxicology, forensic toxicology, forensic biology/biochemistry (DNA profiling, analysis of proteins and other biological compounds), and biomedicine as well5, 6, 7, 8 . References 1 Whatley H. Basic principles of capillary electrophoresis. In Clinical and Forensic Applications of Capillary Electrophoresis. Petersen, J., Mohammad, A.A. (Eds.), Humana Press Inc. pp 21-58 (2001). 2 Harstad, R.K., Johnson, A.C.,Weisenberger, M.M.,Bowser, M.T. Capillary Electrophoresis. Anal. Chem., 88, 299–319 (2016) DOI: 10.1021/acs.analchem.5b04125 3 Karger, B.L. Guttman, A. DNA Sequencing by Capillary Electrophoresis. Electrophoresis. 30, S196–S202. (2009) doi: 10.1002/elps.200900218 4 Wuethrich, A., Quirino J.P. Derivatisation for separation and detection in capillary electrophoresis (2015-2017). Electrophoresis. 39, 82-96 (2018), doi: 10.1002/elps.201700252. 5 Tavares, M. F. M., Jager A., da Silva, C.L.; Morales E.P. Applications of capillary electrophoresis to the analysis of compounds of clinical, forensic, cosmetological, environmental, nutritional and pharmaceutical importance. J. Braz. Chem. Soc. 14, 281-290 (2003). 6 Thormann, W. et al. Clinical and Forensic Drug Toxicology. In Clinical and Forensic Applications of Capillary Electrophoresis. Petersen, J., Mohammad, A. (Eds.), Humana Press Inc. 397-422 (2001). 7 Sarr, S.O., et al. (2016) Validation of a Capillary Electrophoresis method for analyzing chlorphena- mine maleate-based drugs using the accuracy profile approach. Am. J Anal. Chem., 7, 452-459 (2016). 8 Ticova A., Prikryl J. Reproducible preparation of nanospray tips for capillary electrophoresis coupled to mass spectrometry using 3D printed grinding device, Electrophoresis, 37, 924-930 (2016).