MODULE - III




                                                    STQC
Version 1.0   CERTIFIED CALIBRATION PROFFESSIONAL          MIII/CH1/1
Calibration
   A set of operations, performed in accordance
   with a definite documented procedure, that
   compares the measurements performed by an
   instrument to those made by a more accurate
   instrument or standard, for the purpose of
   detecting and reporting or eliminating by
   adjustment, errors in the instrument tested.
                     As per ISO/IEC Guide 2


                                                     STQC
Version 1.0    CERTIFIED CALIBRATION PROFFESSIONAL          MIII/CH1/2
Calibration
   definition (a)
   operation establishing the relation between quantity values
   provided by measurement standards and the
   corresponding indications of a measuring system, carried
   out under specified conditions and including evaluation of
   measurement uncertainty

   definition (b)
   operation that establishes the relation, obtained by
   reference to one or more measurement standards, that
   exists under specified conditions, between the indication of
   a measuring system and the measurement result that
   would be obtained using the measuring system
                               As per VIM
                                                        STQC
Version 1.0       CERTIFIED CALIBRATION PROFFESSIONAL          MIII/CH1/3
Why to Calibrate ?
 •   Initial calibration assures T&ME with in Spec

 •   Periodic calibration is essential to maintain control of measurement
     process made by TME

 •    Assures common reference i.e Traceability to National / International
     standards

 •   Ensures that equipment has been satisfactorily working since last
     calibrated & assumed that the equipment continues to work satisfactorily
     until next calibrated

 •   Calibration records study enables extending/reducing the calibration
     interval

                                                             STQC
Version 1.0           CERTIFIED CALIBRATION PROFFESSIONAL             MIII/CH1/4
OBJECTIVES OF CALIBRATION
              Maintaining traceability of measurements



                                                         Uncertainty
                                                           limits




                                                         Traceability

      Reference

                                                         measurements

                                                         STQC
Version 1.0        CERTIFIED CALIBRATION PROFFESSIONAL           MIII/CH1/5
Why Calibration & Traceability are intertwined?

   • To measure the “Same Measure”
   • Interchangeability of products
   • Legal requirement & consequence of
     liability
   • To demonstrate metrological confirmation
     of product
   • Consistent control of product/measurement


                                                        STQC
Version 1.0       CERTIFIED CALIBRATION PROFFESSIONAL          MIII/CH1/6
When is calibration not required?
        • When equipment is not in use .
        • When eqpt is used as null detector
        • When eqpt is used as transfer device whose
          measurement or output value is not explicitly
          used.
        • Fundamental / Absolute standards.
        • When instrument is fail-safe
        • When instrument makes measurements or
          provides known outputs which are monitored by
          a calibrated device, meter or gage during use.



                                                         STQC
Version 1.0        CERTIFIED CALIBRATION PROFFESSIONAL          MIII/CH1/7
When to Calibrate ?
   - Calibrate too often it is waste of time and
     money.
   - Calibrate too seldom it is waste of product or
     wrong decisions.
   - Balance between cost to calibrate and cost
     not to calibrate
   - Universally Applicable single best practice
     for establishing & adjusting cal interval not
     available.
   - Refer DAP, RP1
                                                     STQC
Version 1.0    CERTIFIED CALIBRATION PROFFESSIONAL          MIII/CH1/8
Where to Calibrate?
•      Calibration is performed in the I/II/III party
       laboratory under controlled environment.
    (The environment need only to be controlled to the extent required.)


•      Accredited Calibration Laboratory

•      Manufacturer’s / Supplier’s Laboratory
       ensuring valid traceability




                                                           STQC
Version 1.0          CERTIFIED CALIBRATION PROFFESSIONAL          MIII/CH1/9
Where to Calibrate ?contd…
   • Accredited Laboratory ensuring the quality of
     calibration services.

   • In the absence of the above, the services of
     calibration laboratory for the parameter of
     interest may be availed ensuring meeting
     technical requirements of ISO/IEC 17025.



                                                      STQC
Version 1.0     CERTIFIED CALIBRATION PROFFESSIONAL          MIII/CH1/10
Basic Requirements - Location/ Environment

              •   Noise Level
              •   Vibration
              •   EMI/EMC
              •   Cleanliness
              •   Lighting
              •   Power supply
              •   Accessibility

                                                           STQC
Version 1.0          CERTIFIED CALIBRATION PROFFESSIONAL          MIII/CH1/11
Basic Requirements – Location/Environment
       Quiet
        * Away from heavy traffic, factories, workshop etc
        * Effective separation from non compatible areas
        * Noise : less than 60 dbA
        * Vibration :
              Taken care in building design & site selection
              noticeable in galvanometers, mercury column, mass
              balance, dimensional measuring equipment
              0.25μm max displacement amplitude from 0.1Hz to 30Hz and
              0.001 g max 30Hz to 200Hz
          * EMI :
            * taken care in site selection & use of shielding/ filtering
            * field strength to be less than 100μV /m
                                                               STQC
Version 1.0             CERTIFIED CALIBRATION PROFFESSIONAL                MIII/CH1/12
Basic Requirements – Location/Environment
   Clean
        Good house keeping
        Higher degree of particulate control (Dimensional &
        Mass Metrology)
   Good lighting
        Recommended level of illumination is 450 – 700lux
        Fluorescent lamps preferred to minimize heating effect
   Services
        Separate feeder line
        Uninterrupted / Regulated power supply (+- 5%)
        Adequate measures against spikes & transients
        Mains Earthing (< 1 ohm)
        Adequate measures against ground currents
                                                 STQC
Version 1.0       CERTIFIED CALIBRATION PROFFESSIONAL   MIII/CH1/13
Basic Requirements – Location/Environment


        Accessibility


              Standards / equipment
              Firefighting devices
              Quality System Document & Instrument
              Manuals


                                                           STQC
Version 1.0          CERTIFIED CALIBRATION PROFFESSIONAL          MIII/CH1/14
Calibration laboratory
                               - basic requirements
    Location – Permanent/On-site                         Temperature
                                                          Humidity
    Environment                                          Air flow
    Equipment                                            Filtration
                                                          Electromagnetic Screening
    Staff                                                Noise Level
                                                          Vibration
    Management                                           EMI/EMC
   Documents                                              Cleanliness
                   Reference standards                    Lighting
   Records
                   Transfer Standards                     Power supply
   Calibration
                   Working standards                      Accessibility
Training           Connecting leads                       Receiving & Calibration area
Authority          adaptors/accessories
Responsibility     Subsidiary equipment
Technical          Computers & automation
Competence         software

                                                                     STQC
Version 1.0          CERTIFIED CALIBRATION PROFFESSIONAL                       MIII/CH1/15
Basic Requirements – Equipment



       Transfer Standards
       Reference standards (Optional)
       Working standards
       Connecting leads and adaptors
       Subsidiary equipment
       Computers & automation software

                                                         STQC
Version 1.0        CERTIFIED CALIBRATION PROFFESSIONAL          MIII/CH1/16
Basic Requirements – Personnel



       Training
       Authority
       Responsibility
       Technical Competence




                                                         STQC
Version 1.0        CERTIFIED CALIBRATION PROFFESSIONAL          MIII/CH1/17
Basic Requirements – Management

       Quality management & Technical
        operations
       Traceability
       Calibration interval
       Audits/ Management reviews
       Implementation of corrective /preventive
        actions


                                                         STQC
Version 1.0        CERTIFIED CALIBRATION PROFFESSIONAL          MIII/CH1/18
Calibration System


      Why Calibration system ?
      - To meet the commitment of the industry to
        provide quality product.
      - TMI is required to be controlled to provide this
        assurance.
      - Through calibration
      - Shall include all the elements necessary for
        ready detection of equipment deficiencies
        and provide objective evidence of
        conformance in the form of records.

                                                        STQC
Version 1.0       CERTIFIED CALIBRATION PROFFESSIONAL          MIII/CH1/19
Calibration System




          It is defined as a set of inter-related and
          interacting elements which influence the
          calibration process and ultimately
          quality of calibration result.




                                                         STQC
Version 1.0        CERTIFIED CALIBRATION PROFFESSIONAL          MIII/CH1/20
Elements of Calibration System

         •    Qualification
         •    Measurement traceability
         •    Adequacy of calibration procedure
         •    Calibration procedure
         •    Qualification labeling
         •    Sealing for integrity
         •    Calibration interval
         •    Records Maintenance


                                                             STQC
Version 1.0            CERTIFIED CALIBRATION PROFFESSIONAL          MIII/CH1/21
Elements of Calibration System
                    contd…
      • Environmental Controls
      • Handling , Preservation & Storage
      • Non- conforming Measurement Equipment
      • Scheduling / Recall system
      • Personnel requirement
      • Audit requirement
      • Calibration Quality
      • Use of outside service
                                                          STQC
Version 1.0         CERTIFIED CALIBRATION PROFFESSIONAL          MIII/CH1/22
ACCREDITATION REQUIREMENTS

 The laboratory must demonstrate control and competence
 Environmental control
 Traceability of standards
 Calibration procedures
 Competence of staff / Authorization of personnel
 Participation in PT programmes
 In-service checks/ control charts
 Validation of methods
 Internal audit & Management review records


                                                          STQC
Version 1.0         CERTIFIED CALIBRATION PROFFESSIONAL          MIII/CH1/23
STQC
Version 1.0   CERTIFIED CALIBRATION PROFFESSIONAL          MIII/CH1/24
CHAPTER - 1




                                                    STQC
Version 1.0   CERTIFIED CALIBRATION PROFFESSIONAL          MIII/CH1/25
CALIBRATION PROCEDURES


   Procedure

              -Ensures consistency in operation

              -Objective evidence

   Calibration procedure provides

              -Specified performance tolerance and functional criteria

              -Essential information & instruction to determine the integrity of eqpt.

              -Logical & timely sequence of necessary facts to understand and

              verify DUC to its intended performance capability.

                                                                   STQC
Version 1.0              CERTIFIED CALIBRATION PROFFESSIONAL               MIII/CH1/26
CALIBRATION PROCEDURES

   - Guides personnel

              - eqpt. Setup, interconnection, control operations

              - specification interpretation

              - tolerance limits

              - any adjustment, correction

   - Establish uniformity of methods/techniques

   - Shall be written such that personnel shall take minimum interpretation,

     effort and time.

   - Shall be written by keeping performance of DUC in mind.
                                                                   STQC
Version 1.0               CERTIFIED CALIBRATION PROFFESSIONAL             MIII/CH1/27
CALIBRATION PROCEDURES




   Requirements

   Source:

              -Prepared internally

              -Prepared by the manufacturer

              -External agency




                                                               STQC
Version 1.0              CERTIFIED CALIBRATION PROFFESSIONAL          MIII/CH1/28
CALIBRATION PROCEDURES

   Completeness:

         - Sufficient instruction

   Approval:

        - shall be approved and controlled to meet regulatory and contractual

         requirements.

   Software:

         - Validation and control




                                                               STQC
Version 1.0              CERTIFIED CALIBRATION PROFFESSIONAL            MIII/CH1/29
CALIBRATION PROCEDURES

   BROADLY COVERS -

               - Performance requirements of DUC

               - Performance requirements of standard

               - Preliminary operation

               - Calibration process

               - Calibration results

               - Closing operation

               - Storage and handling




                                                        STQC
Version 1.0      CERTIFIED CALIBRATION PROFFESSIONAL           MIII/CH1/30
CALIBRATION PROCEDURES – TYPICAL CONTENT
   • Appropriate identification / document control.

   • Scope

   • Purpose - description & performance requirements of unit being calibrated

   • Parameters or quantities and ranges

   • Calibration equipment/system/setup

   • Reference standards and reference materials required

   • Reference Documents – standard/specifications/manuals

   • Adaptors/Accessories required

   • Environmental conditions required including stabilization period needed.

                                                            STQC
Version 1.0           CERTIFIED CALIBRATION PROFFESSIONAL            MIII/CH1/31
CALIBRATION PROCEDURES – TYPICAL CONTENT
  • Description of the procedure, including –
         Apprising of identification mark, sampling, handling, transporting,

              storing and preparation of items.
         Preliminary Checks to be made before the work is started.
         Checks that the equipment is working properly, and where required ,
         Calibration process & adjustments if any, of the equipment
         The method of recording the observations and results.
         Any safety measures/precautions to be observed.
  • Criteria and / or requirements for approval / acceptance / rejection.
  • Data to be recorded and method of analysis and presentation.
  • The uncertainly or the procedure for estimating uncertainty.
  • Calibration status/labeling
                                                                 STQC
Version 1.0                CERTIFIED CALIBRATION PROFFESSIONAL              MIII/CH1/32
CALIBRATION PROCEDURE CONTROL

       Deviations from instructions authorized & recorded
       Assign responsibility
       Establish document control with a standard format
       Draft, comment and edit procedures/instructions
       Review and authorize procedures
       Print, bind, present and make available to all operational
        staff
       Regular review and amendment as needed
       Archive the superseded methods


                                                          STQC
Version 1.0         CERTIFIED CALIBRATION PROFFESSIONAL          MIII/CH1/33
CALIBRATION PROCEDURES


                                EXERCISE – 1


              DEVELOPMENT & REVIEW OF PROCEDURES




                                                         STQC
Version 1.0        CERTIFIED CALIBRATION PROFFESSIONAL          MIII/CH1/34
CALIBRATION PROCEDURES

   Section II

   -table defining performance specification (minimum use specification)

   required of the standard

   -Listing of specific manufacturer model (representative eqpt.) which are

   applicable and can operate with in the specified requirements. Equipment

   which has no known substitute should be clearly identified and described.

   -Environmental condition

   -Any special jigs and fixtures required



                                                              STQC
Version 1.0            CERTIFIED CALIBRATION PROFFESSIONAL             MIII/CH1/35
CALIBRATION PROCEDURES


   Section III

   -Listing of all preliminary actions including functional checks to be performed

   prior to calibration


   Section IV

   -Listing of instructions and all steps to be performed in measurement process


   Section V

   -A results data sheet blank form with nominal values and tolerance limits.


                                                                STQC
Version 1.0               CERTIFIED CALIBRATION PROFFESSIONAL           MIII/CH1/36
CALIBRATION METHODS

        • Standard methods –
              Direct/Indirect/Ratio/Substitution/Transfer/Differential


        • Non-standard methods

        • Laboratory developed methods                           NEEDS
                                                                 VALIDATION


        • Expansion of Scope

        • Customer given method
                                                                STQC
Version 1.0               CERTIFIED CALIBRATION PROFFESSIONAL            MIII/CH1/37
VALIDATION OF CALIBRATION METHODS

              •   Selectivity
              •   Uncertainty of the results
              •   Range
              •   Repeatability
              •   Reproducibility
              •   Linearity
              •   Limit of detection
              •   Limit of quantization
              •   Robustness
              •   Bias
              •   Precision


                                                        STQC
Version 1.0       CERTIFIED CALIBRATION PROFFESSIONAL          MIII/CH1/38
GENERAL CALIBRATION TECHNIQUES


                   • SPANNING
                   • NULLING
                   • ZEROING
                   • LINEARIZATION




                                                         STQC
Version 1.0        CERTIFIED CALIBRATION PROFFESSIONAL          MIII/CH1/39
GENERAL CALIBRATION METHODS
                                        NULLING
              •   If the nominal value of UUC is same as that of reference
                  comparison may be made using Nulling process.

              •   Instruments like Galvanometers, Null Detectors &
                  Comparators work on this principle. The scale indicates
                  the sensitivity of detection in these meters.

              •   Nulling is used for more accurate measurements since
                  Null process can be quite sensitive to minute difference
                  between known & unknown values.

              •   Note that the Null detectors need not be calibrated at all
                  for the measured values & the sensitivity of detection is
                  by design, by scale of indication.




                                                                  STQC
Version 1.0              CERTIFIED CALIBRATION PROFFESSIONAL                 MIII/CH1/40
GENERAL CALIBRATION METHODS
                                   ZEROING
   • a compensation technique for
     offset errors due to device to
     device offset, variation (trim
     errors), mechanical stresses
     (mounting stresses), shifts
     due to temperature and aging.

   • For zero referencing for >0
     measurements

   • adjusting an instrument to
     read zero manually / auto
     zeroing




                                                          STQC
Version 1.0         CERTIFIED CALIBRATION PROFFESSIONAL          MIII/CH1/41
GENERAL CALIBRATION METHODS
                                       SPANNING


              -   is adjusting the instrument to read the known
                  concentration of a gas / material / chemical

              -   Range of parameter which the unit/equipment can
                  measure

              -   Scale of measurement




                                                              STQC
Version 1.0             CERTIFIED CALIBRATION PROFFESSIONAL          MIII/CH1/42
GENERAL CALIBRATION METHODS
                           SPANNING & ZEROING


          - are commonly performed during instrument
            start-up or before calibration.

              Ex.: The monitor’s combustible and toxic
                   sensors are zeroed (adjusted to zero) in a
                   known fresh air environment and the
                   oxygen sensor is spanned to fresh air
                   (21%). It is critical that this only be con-
                   ducted in a known fresh air environment.



                                                             STQC
Version 1.0            CERTIFIED CALIBRATION PROFFESSIONAL          MIII/CH1/43
GENERAL CALIBRATION METHODS
                            LINEARIZATION


         To approximate a nonlinear system by a linear one
                      is called Linearization.

        Ex.: Interfacing an NTC thermister to a computerized data acquisition
         system linearization is used to convert the resistance output of the
                                thermister to voltage.




                                                               STQC
Version 1.0             CERTIFIED CALIBRATION PROFFESSIONAL             MIII/CH1/44
CALIBRATION METHODS SELECTION

 ensure laboratory selects and uses test and/or calibration
 methods that:
       •      meet the needs of the customer
       •      are appropriate for the calibration task
       •      the customer has been informed of the method chosen (if not
              specified)
       •      where appropriate, are based on latest international, regional or
              national standards and where necessary the standard be
              supplemented with additional details to ensure consistent
              approach
       •      have been verified for use in the laboratory, if a standard
              method

Ensure laboratory has competence to choose the appropriate methods

                                                                STQC
Version 1.0               CERTIFIED CALIBRATION PROFFESSIONAL          MIII/CH1/45
CHAPTER - 3




                                                    STQC
Version 1.0   CERTIFIED CALIBRATION PROFFESSIONAL          MIII/CH1/46
Basic Requirements – Environment

       Temperature
       Humidity
       Air flow
       Filtration
       Electromagnetic Screening




                                                         STQC
Version 1.0        CERTIFIED CALIBRATION PROFFESSIONAL          MIII/CH1/47
Basic Requirements – Environment

   Temperature
    * Lab design to take into account heat generation through
        personnel, equipment, light etc
    * Recommended continuous & permanent recording
    * Temperature monitoring to be installed at eqpt. or
    operator level
    * Avoid temp gradient through uniform air flow
    * Precise metrology requires tight temp control &
    measurement of actual temp. to apply meaningful
    corrections
    * Allow artifact / eqpt. to attain thermal equilibrium
    (warm up)

                                                           STQC
Version 1.0          CERTIFIED CALIBRATION PROFFESSIONAL          MIII/CH1/48
Basic Requirements – Environment
      Humidity
       * Both low & high humidity to be avoided : static
       electricity & leakage path
       * 45 + 10% RH recommended for all general lab

      Air flow
       * A well designed system to provide uniform air
            circulation with out stratification to minimize temp
            gradient in & around the lab
       * Floor to ceiling vertical laminar flow of air is
       recommended to minimize temp gradient to + 0.1 0C
       * Air velocity of 0.71 + 0.28 cubic mts /minute
       recommended (3 to 5 ft above the floor)
                                                        STQC
Version 1.0         CERTIFIED CALIBRATION PROFFESSIONAL      MIII/CH1/49
Basic Requirements – Environment

    Filtration
     * Provide filtration to limit particulates
     * High degree of particulate control : optical, mass,
          dimensional metrology etc.
         Electro-magnetic Screening
     * Field strength to be less than 100μV/m
       *       For conducted electromagnetic energy on power lines,
               open circuit voltage to be less than 100mV
       *       Provide adequate shielding/ filtering.



                                                              STQC
Version 1.0             CERTIFIED CALIBRATION PROFFESSIONAL          MIII/CH1/50
Accommodation of Cal Lab.
      Accommodation is the premises where the
      laboratory functions are carried out.
      Considerations :
         • Design and construction materials
         • Space including storage
         • Services and facilities
         • Access, housekeeping, security and
           safety
         • Separation of incompatible activities
         • Area for administrative duties
                                                        STQC
Version 1.0       CERTIFIED CALIBRATION PROFFESSIONAL          MIII/CH1/51
Control of Environmental Parameters


      • Define critical environmental parameters that
        affect quality of calibration result
      • Define nominal values & tolerable limits
      • Monitor environmental parameters
      • Take remedial measures, if required
      • Halt Calibration process, if environment is not
        conducive.

                  Refer RP 6 on Laboratory Design

                                                          STQC
Version 1.0         CERTIFIED CALIBRATION PROFFESSIONAL          MIII/CH1/52
Recommended Env. Conditions
              - as per RPCL by CLAS in PALCAN Handbook published by Council of Canada




                                                                  STQC
Version 1.0               CERTIFIED CALIBRATION PROFFESSIONAL               MIII/CH1/53
ACCOMMODATION & ENVIRONMENT - TYPICAL

   1. Clean & vibration free location for Monitoring &
      Measuring devices.
   2. Temperature : Ambient / 23 ± 1 °C OR 23 ± 2 °c
   3. Humidity : 50 ± 10 %
   4. Illumination level : 250 – 450 (700 lux)
   5. Noise level – Below 60 dBA where meas. are affected
   6. Rigid table for installation of equipment
   7. Noisy vacuum source and Pneumatic cylinder to be
      outside the calibration/inspection room/area


                 REFER NABL SPECIFIC CRITERIA 121 - 126


                                                          STQC
Version 1.0         CERTIFIED CALIBRATION PROFFESSIONAL          MIII/CH1/54
Laboratory Storage Space
    Storage must also be such as to not interfere
   with test or calibration results.
Proper storage is needed for :
 Calibration jobs (current and retained)
 Working       materials/consumables/adaptors/accessories/cables
  etc.
 Laboratory documents
 Hazardous substances
 Staff belongings
 Clerical materials and equipment
 Laboratory records
                   REFER NABL SPECIFIC CRITERIA


                                                        STQC
Version 1.0       CERTIFIED CALIBRATION PROFFESSIONAL          MIII/CH1/55
CHAPTER - 4




                                                    STQC
Version 1.0   CERTIFIED CALIBRATION PROFFESSIONAL          MIII/CH1/56
a) Handling, storage, preservation, transportation, safety &
                 security of Calibration unit/equipment/job
              b) Customer requirements
              c) Laboratory capabilities – availability of personnel, cal
                 facilities, technical competence, cal method,
                 accommodation & environment, traceability, sampling
                 etc.
              d) Calibration – measurement data, computations,
                 uncertainty budget, compliance of results
              e) Reporting results – preparation, approval, issue
              f) Labeling
              g) Post cal service activities




                                                                    STQC
Version 1.0                 CERTIFIED CALIBRATION PROFFESSIONAL              MIII/CH1/57
Calibration Equipment

   •    is available and functioning properly
   •    is capable of achieving required accuracy
   •    complies with relevant specifications
   •    has calibration programs established for key quantities or
        values
   •    is calibrated or checked before being placed into service
   •    is checked and/or calibrated before use (see 5.6 also)
   •    is operated by authorised personnel
   •    has current instructions on use and maintenance
   •    is uniquely identified, where practicable


Record of specification checking and maintained evidence.
                                             STQC
Version 1.0          CERTIFIED CALIBRATION PROFFESSIONAL    MIII/CH1/58
Equipment Records
      •       identity of the equipment and its software
      •       manufacturer’s name, model, and serial number or other
              unique identification
      •       evidence that the equipment complies with the accuracy
              requirements and specifications relevant to the tests or
              calibrations
      •       current location, where appropriate
      •       the manufacturer’s instructions, if available, or reference
              to their location
      •       calibration history and due date of next calibration
      •       the maintenance plan, where appropriate, and maintenance
              carried out to date
      •       any damage/malfunction/modification/repair to the equip.


                                                               STQC
Version 1.0              CERTIFIED CALIBRATION PROFFESSIONAL          MIII/CH1/59
Equipment handling
  ensure procedures for measuring equipment are
  documented and include:
    • safe handling
    • transport
    • storage
    • use
    • planned maintenance –
       Preventive/Breakdown/In-service Checks,
       Intermediate Checks
    • where applicable, that copies of correction
       factors are correctly updated
                                                      STQC
Version 1.0     CERTIFIED CALIBRATION PROFFESSIONAL          MIII/CH1/60
Equipment out-of-service
  ensure equipment subjected to overloading or mishandling,
  giving suspect results, or shown to be defective or outside
  specified limits is taken out of service, and is:

        • isolated or clearly labelled or marked as being out of
          service
        • examined for the effect of the defect or departure from
          specified limits on previous tests and/or calibrations
        • addressed under the “Control of nonconforming work”
          procedure

              Have a procedure for handling such situations
                                                             STQC
Version 1.0            CERTIFIED CALIBRATION PROFFESSIONAL          MIII/CH1/61
Calibration Status
    • Ensure equipment calibration status is identified
    • Calibration labelling based on Cal agency
      Label that indicates date of calibration due date for next
      calibration, person responsible for calibration.
      For items where marking cannot be done it e.g. Weights,
      gauges, the status can be put in the Register/colour
      coding etc.
    • where practicable and where intermediate checks
      are needed to maintain confidence in the
      calibration status that a procedure is documented
      to carry out these checks
                                                         STQC
Version 1.0        CERTIFIED CALIBRATION PROFFESSIONAL          MIII/CH1/62
CALIBRATION AS A PROCESS


              Process A – Preparation of list all test & measuring equipment that
                            need to be under calibration control.

               Process B – Calibration of various test and measuring equipment
                           (in-house calibration), if facility exist.

                 Process C - Calibration of test & measuring equipment (from
                                  external Calibration lab)

                  Process D -Verification of calibration results for compliance to
                               measurement requirements




                                                                        STQC
Version 1.0                  CERTIFIED CALIBRATION PROFFESSIONAL                     MIII/CH1/63
Measuring Equipment – Process Model

                          MEASURING PROCESS

I/P           Sensor         Instrumentation                 Display
                                                                              O/P


Measurand                                                              Result



               Sensor may not be required in every case




                                                               STQC
Version 1.0            CERTIFIED CALIBRATION PROFFESSIONAL              MIII/CH1/64
Measuring Equipment


      •   Measuring instrument/system
      •   Software
      •   Reference standard/material
      •   Auxiliary apparatus/Adaptors/Accessories/Cables etc.
      •   Any combination of the above

              Measuring equipment requires metrological confirmation




                                                              STQC
Version 1.0             CERTIFIED CALIBRATION PROFFESSIONAL          MIII/CH1/65
Leading Instrumental Characteristics
          •   Range
          •   Error
          •   Resolution
          •   Drift
          •   Input Impedance (Loading effect)
          •   Compatibility between sensor & Instrumentation
          •   Input – Output relationship (characteristic)
          •   Hysterisis
          •   Ruggedness
          •   Mounting



                                                            STQC
Version 1.0           CERTIFIED CALIBRATION PROFFESSIONAL          MIII/CH1/66
Checks on Measuring Equipment


      •   Mechanical Zero / Zeroing /Auto zero
      •   Check-up of readings during ascent & descent
      •   Check-up of resolution
      •   Check-up of linearity error, if any
      •   Repeatability of measurements
      •   Reproducibility of measurements




                                                          STQC
Version 1.0         CERTIFIED CALIBRATION PROFFESSIONAL          MIII/CH1/67
Metrological Confirmation of Eqpt.


      • Set of operations required to ensure that
        measuring equipment conforms to the
        requirement for its intended use.

              - Brings in the concept of “Calibration”




                                                              STQC
Version 1.0             CERTIFIED CALIBRATION PROFFESSIONAL          MIII/CH1/68
Calibration

      • Comparison of an unknown measurement device
        against an equal or better device (TUR/TAR)
      • Unknown device is called
              –   Equipment under Calibration (EUC)
              –   Unit under Calibration (UUC)
              –   Device under Calibration (DUC)
              –   Device being Calibrated (DBC)
      • Equal or better device is called “Standard”



                                                                STQC
Version 1.0               CERTIFIED CALIBRATION PROFFESSIONAL          MIII/CH1/69
Calibration Process
        •     Identification of need/Customer’s requirements
        •     Availability of specifications for DUC
        •     Serviceability status of DUC
        •     Laboratory Capability
        •     Availability of validated method/competence
        •     Availability of Instruments /Standards
        •     Technical comparison of DUC with Instrument / Standards
        •     Repair / adjustment / recalibrate, if required
        •     Verification of Confirmation
        •     Uncertainty computation
        •     Preparation of Calibration Report/Certificate
        •     Status Identification

                                                                STQC
Version 1.0               CERTIFIED CALIBRATION PROFFESSIONAL          MIII/CH1/70
CALIBRATION JOB FLOW




                                                         STQC
Version 1.0   CERTIFIED CALIBRATION PROFFESSIONAL               MIII/CH1/71
Calibration - Job Flow Chart
                        Ext      Int
                                                                                         Check points
                     Customer Service                                 Y

              N              Y
                    Job                      Job card to SCL
                                                                    Any
                                                                               N
                  accepted
                                                                 Non-confor-         CERT. Cleared
                                                                   mance
                                                Calibrate
                                                                                     TP&C
      Int. DUC
                                       Compile result error/uncert.
      Out of Spec
                                                                                   CERT. Issued
                                                Fix label                           with Item
          R&M

                                         DUC with cal report
     DUC after R&R                                                                 EXT       INT

                                                 TP & C
              CAL
                                                   QA
                                                                                   STQC
Version 1.0                      CERTIFIED CALIBRATION PROFFESSIONAL                        MIII/CH1/72
Calibration v/s Business Processes



                                         Control of outside supplies
                                                and services




                           Evaluati             In-house               Report and record    Delivery /
              Receiving                        Calibrations               processing       Shipping
                             on




                                             In-house repair




                                                Customers




                                                                                     STQC
Version 1.0               CERTIFIED CALIBRATION PROFFESSIONAL                                  MIII/CH1/73
Role, Responsibilities & Authority
       Job descriptions
            • HOL,Technical & Quality Managers
            • Scientists and Technicians
            • Attendants
            • Those who verify, test and check
            • Sampling officers
            • Those who act in Emergency
       Performance Criteria
       Performance appraisal

                                                      STQC
Version 1.0     CERTIFIED CALIBRATION PROFFESSIONAL          MIII/CH1/74
Laboratory Manager/Chief Executive

   • Provide direction and coordinate the departmental activi-ties in
     technical, commercial and service establishment
   • Reviews and analyzes reports, records, and other information for
     planning laboratory’s growth, such as new commitments, status of
     work in progress and problem encountered.
   • Assigns or delegates respon-sibility for, specified work or functional
     activities and disseminates policy to down the line.
   • Gives work directions, resolves problem and set deadlines to ensure
     timely completion of work.
   • Coordinates activities of various departments with related activities
     of other departments to ensure efficiency and economy.
   • Monitors and analyzes costs and prepares budget.
   • Prepares reports and records on department activities for
     management.
   • Evaluate current procedures and practices for accomplishing
     department objectives to develop and implement improved
     procedures and practices.

                                                            STQC
Version 1.0           CERTIFIED CALIBRATION PROFFESSIONAL            MIII/CH1/75
The Role of Quality Manager
     • Assurance of Quality of Calibration Results
     • Ensuring implementation of Management system
     • Document Control activity
     • Planning, conducting, compiling & analyzing Internal audit
       programme
     • Organizing, conducting Management Review &
       maintaining records
     • Monitoring corrective & preventive actions on non-
       conforming cal services ensuring effectiveness
     • Handling, investigating, analyzing & resolution of
       Customer complaints
     • Liaison with internal/external customers, accreditation
       body w.r.t quality matters affecting calibration.
     • Reports to Management
                                                         STQC
Version 1.0        CERTIFIED CALIBRATION PROFFESSIONAL          MIII/CH1/76
The Role of Technical Manager
   • selection, approval and commissioning of test methods
   • Management and training of staff
   • Evaluation and sign-off of staff competency
   • Management of laboratory equipment and facilities
   • Design and implementation of the QC programme
   • Stopping work when QC criteria are exceeded
   • Attention of matters relating to impartiality and
     confidentiality
   • Responsibility for the technical validity of results
   • Liaison with the QM in relation to Quality matters
     affecting testing or calibration.
   • Reports to TM
                                                 STQC
Version 1.0     CERTIFIED CALIBRATION PROFFESSIONAL   MIII/CH1/77
The Role of METROLOGIST
   • Develops and evaluates calibration system that measure
     characteristics of objects, substances, or phenomena, such as
     length, mass, time, temperature, electric current, luminous intensity,
     and derived units of physical or chemical measure
   • Identifies magnitude of error sources contributing to uncertainty of
     results to determine reliability of measurement process in
     quantitative term
   • Redesigns or adjusts measurement capability to minimize errors
   • Develops cal-ibration methods and techniques based on principles
     of measurement science, technical analysis of measurement
     problem, and accuracy and precision requirements
   • Directs engineering, quality and laboratory personnel in design,
     manufacture, evaluation, and calibration of measurement standards,
     instru-ments, and test system to ensure selection of approved
     instrumentation
   • Advises others on methods of resolving measurement problem and
     exchanges information with other metrology personnel through
     participation in govern-ment and industrial standardization
     committees and professional societies
                                                            STQC
Version 1.0           CERTIFIED CALIBRATION PROFFESSIONAL            MIII/CH1/78
The Role of Laboratory In-Charge
• Review/Acceptance/follow-up of jobs received from
  customer service to meet committed PDC for the same.
• Preparation of calibration schedule if required and
  Allocation of work
• Maintenance of lab. environment conditions ensuring
  good house keeping
• Periodic monitoring of observations/ measurements/
  compiled data during execution of the job & responsible
  for the technical validity of results
• Preparations of test/cal procedures & implementation
  after due approvals
• Qualifying label for the T&M equipment calibrated
                                                      STQC
Version 1.0     CERTIFIED CALIBRATION PROFFESSIONAL          MIII/CH1/79
The Role of Laboratory In-Charge
• Maintenance of equipment in the qualifying condition
  along with respective manuals/ technical data & history
  log books ensuring traceability
• Maintaining registers used in the laboratory of the
  laboratory area
• Co-ordination with lab. staff, customer service & other
  support services
• Technical guidance to the laboratory personnel in
  calibration & Implementation of the QC programme
• Attention of matters relating to impartiality and
  confidentiality
• Liaison with the Quality Manager in relation to Quality
  matters affecting calibration.
                                               STQC
Version 1.0      CERTIFIED CALIBRATION PROFFESSIONAL    MIII/CH1/80
Roles of Cal Lab. personnel
     • Responsible for the calibration activity
     • Carrying out Calibration as per standard/validated cal methods
     • Preparation of Cal reports, affixing cal label on equipment calibrated.
     • Building technical competence in the area of service being provided
     • Any cal facility giving intermediate operation/ambiguous results/mal-
      functioning should be immediately reported to LIC/TM
     • Maintain the laboratory facility with appropriate use
     • Maintaining calibration service related records
     • Maintaining records reg. environmental controls
     • Maintaining Calibration/Maintenance manuals of cal facilities & traceability
       records
     • Reporting to LIC/TM
     • Any other as assigned from time to time
                                                                STQC
Version 1.0             CERTIFIED CALIBRATION PROFFESSIONAL               MIII/CH1/81
“Approved Signatory”
  Personnel Technical Competence including:
 Relevant qualifications and/or experience.
 Close involvement in the day-to-day operations.
 Familiarity with test procedures including scientific basis
  and technical limitations.
 Ability to make critical evaluation of test results.
 Knowledge of Quality Management system.
 Knowledge of and commitment to AB and their criteria
  and ISO/IEC 17025.
 Sufficient time in the laboratory to become fully familiar
  with the operating systems of the laboratory.
                                                       STQC
 Version 1.0     CERTIFIED CALIBRATION PROFFESSIONAL          MIII/CH1/82
SCHEDULING


              -     CALIBRATION INTERVAL

              -     CONFIRMATION INTERVAL

              -     RECALL NOTIFICATIONS




                                                        STQC
Version 1.0       CERTIFIED CALIBRATION PROFFESSIONAL          MIII/CH1/83
Calibration Interval / Reports

 Initial Calibration:               Calibration adj.            Final calibration:
                                    Procedure adopted
 Comparison                         To ensure the               To confirm the
 w.r.t                              tracking of the             successfulness
 common reference                   Measurand to that           Of the operation
 Either to know                     of reference
 The deviation/value                                            (cal adj.procedure
 (measurand &                                                   Implementation)
 Measurement                                                    (Report2)
 Uncertainty
 (Report 1)                          Report1-Report2

                                             ?

                                                                 STQC
Version 1.0               CERTIFIED CALIBRATION PROFFESSIONAL              MIII/CH1/84
Within Calibration Interval….



              Effect of drift on uncertainty of
              measurement



                               Uncertainty of measurement



                                           Time



                                                            STQC
Version 1.0           CERTIFIED CALIBRATION PROFFESSIONAL          MIII/CH1/85
Calibration Interval Management


                  Effect of right time cal. Adj.
                 on uncertainty of measurement




                            adjustment carried out


                 Uncertainty of measurement
                                                       STQC
Version 1.0      CERTIFIED CALIBRATION PROFFESSIONAL          MIII/CH1/86
Scheduling Calibration Interval


     • Client agreed
         reported as ‘Client Agreed Calibration Interval - '.

     • Client specified
         reported as 'Client Specified Calibration Interval - '.

     • Manufacturer's recommended
         reported as 'Manufacturer's Recommended Calibration Interval - '.




                                                                   STQC
Version 1.0             CERTIFIED CALIBRATION PROFFESSIONAL               MIII/CH1/87
INITIAL CHOICE OF CI
       • MANUFACTURER’S RECOMMENDATION
       • EXTENT & SEVERITY OF USE
       • MAX. ALLOWED DEVIATION
       • UNCERTAINTY OF MEASUREMENT REQUIRED
       •      INFLUENCE OF ENVIRONMENT
       • ADJUSTMENT/CHANGE REQUIRED
       • PUBLISHED DATA ON SIMILAR DEVICES
       • BY EXPERIENCE
       • HISTORY OF EQUIPMENT
       • QUALITY OF CALIBRATION & COST FACTOR

 T&ME or M&ME ARE TO BE CALIBRATED WHENEVER THEY ARE REPARIED.
                                                            STQC
Version 1.0           CERTIFIED CALIBRATION PROFFESSIONAL          MIII/CH1/88
OPTIMUM INTERVAL
   A BALANCE BETWEEN THE COST TO CALIBRATE &
   THE COST NOT TO CALIBRATE SHOULD BE
   EVALUATED.

   AT PRESENT THERE IS NO UNIVERSALLY
   APPLICABLE SINGLE BEST PRACTICE FOR

   ESTABLISHING AND ADJUSTING CAL INTERVAL.
                                      STQC
Version 1.0     CERTIFIED CALIBRATION PROFFESSIONAL   MIII/CH1/89
RECALL NOTIFICATION

   • INTERNAL EQUIPMENT

   • EXTERNAL EQUIPMENT




                                                       STQC
Version 1.0      CERTIFIED CALIBRATION PROFFESSIONAL          MIII/CH1/90
CHAPTER - 5




                                                    STQC
Version 1.0   CERTIFIED CALIBRATION PROFFESSIONAL          MIII/CH1/91
Manual Calibration
              The traditional method of obtaining parameters, requiring extensive
                                   training, time, and labor
                 • Availability of Calibration facilities / System
                 • Qualified & trained personnel
                 • Availability of suitable environment
                 • Traceability of standards
                 • Device under calibration
                 • Calibration Method (Validated)
                 • Measurement records
                 • Uncertainty computations
                 • Calibration Reporting
                                                                   STQC
Version 1.0                 CERTIFIED CALIBRATION PROFFESSIONAL             MIII/CH1/92
Automated Calibration

      1. Computer Systems or Automated Equipment

      2. Calibration software / Automated SW Program are
              useful to process
              • Large quantity of TME
              • Lower skilled personnel
              • Large volume of data used for capture, processing ,
                recording, reporting, storage or retrieval of calibration data.


     3. Interface / Compatible DUC


                                                                STQC
Version 1.0               CERTIFIED CALIBRATION PROFFESSIONAL          MIII/CH1/93
Examples: Automated Calibration
                - to the extent supported by HW
              1. DMM Calibration batch-wise

              2. Batch temp. probe calibration

              3. Batch Energy Meter Calibration

              4. Calibration of ref. pressure transducers

              5. Calibration of RF Power sensors

              6. AC/DC Transfer standards Calibration

              7. Etc..
                                                               STQC
Version 1.0              CERTIFIED CALIBRATION PROFFESSIONAL          MIII/CH1/94
Software Programme

          • Design Application Software

          • Develop/Construct Application Software

          • Test Application Software

          • Procure Application Software

          • Operate Application Software

          • Maintain Application Software


                                                         STQC
Version 1.0        CERTIFIED CALIBRATION PROFFESSIONAL          MIII/CH1/95
Major Elements of Software Programme


      • Software development

      • Software procurement

      • Software Configuration Management

      • Software operations

      • Documents

      • Record keeping


                                                       STQC
Version 1.0      CERTIFIED CALIBRATION PROFFESSIONAL          MIII/CH1/96
Software Development

      • Quality planning based on requirements

      • Multi-phased Software Life Cycle Process
        – Requirements, design, construction, testing,
          installation, operation, maintenance &
          retirement

      • Software Review, Verification & Validation




                                                         STQC
Version 1.0        CERTIFIED CALIBRATION PROFFESSIONAL          MIII/CH1/97
Software Validation

        • Comparison of manual and
          automated methods

        • Inspection of procedure.

        • Inter-laboratory comparison




                                                      STQC
Version 1.0     CERTIFIED CALIBRATION PROFFESSIONAL          MIII/CH1/98
Software Procurement

              • Requirements & Activities

              • Applicable Guidelines

              • Dedicated/designed/custom made




                                                         STQC
Version 1.0        CERTIFIED CALIBRATION PROFFESSIONAL          MIII/CH1/99
Software Configuration Management

                   • Identification

                   • Change Control

                   • Status Accounting




                                                          STQC
Version 1.0         CERTIFIED CALIBRATION PROFFESSIONAL          MIII/CH1/100
Software Operations

              • Access Control

              • Backup & Recovery

              • Problem Reporting & Corrective Action




                                                         STQC
Version 1.0        CERTIFIED CALIBRATION PROFFESSIONAL          MIII/CH1/101
Documentation/Records

      • Software Design & Development Procedure
      • Plan for Software QA
      • Software Requirements Documentation
      • Software Design & Implementation Documentation
      • Software verification & validation Documentation
      • User Documentation
      • Standards / Specifications applicable


                                                        STQC
Version 1.0       CERTIFIED CALIBRATION PROFFESSIONAL          MIII/CH1/102
Automated Calibration
     SAMPLE OF AUTOMATED AND MANUAL SYSTEM EQUIPMENT COSTS
                                 (Fluke analysis)
     Equipment                          Automated           Manual
     SPRT                               $ 3,000             $ 3,000
     Recording thermometer              $ 7,000
     Nano volt meter / Mohm             $ 3,500
     Ice point reference for T/Cs       $ 600
     Dual Dry-Block –35 ºC to 670 ºC    $ 8,900             $ 8,900
     Computer                           $ 1,500             $ 1,500
     Automated software                 $ 1,300
     Report generation software         $ 500
     Total                              $ 21,700            $ 18,000

     Automated batch calibration systems for industrial sensors may have
     batch/volume efficiency advantages, in time savings and increased
     productivity for calibration laboratories that find themselves with
     increasing temperature probe workloads.
                                                                      STQC
Version 1.0           CERTIFIED CALIBRATION PROFFESSIONAL                    MIII/CH1/103
Internet Enabled Calibration
              Recommendations as per NPL, UK Report CMSC 28/03

              TECHNIQUES                             SECURITY ISSUES

   •    userid/password protection           •         Authentication
                                             •         Access Control
   • Algorithms for encryption of
     sensitive measurement data
                                             •         Confidentiality
                                             •         Integrity
   • Message authentication                  •         Non-repudiation
     code for certificate,                   •         Availability
     Measurement data                        •         Audit
   • MySQL vulnerabilities                   •         Privacy
                                             •         Certification
   • Backups and fire safes                  •         General Protection

                                                               STQC
Version 1.0            CERTIFIED CALIBRATION PROFFESSIONAL             MIII/CH1/104
STQC
Version 1.0   CERTIFIED CALIBRATION PROFFESSIONAL          MIII/CH1/105
CHAPTER - 6




                                                    STQC
Version 1.0   CERTIFIED CALIBRATION PROFFESSIONAL          MIII/CH1/106
Calibration Results

      - Based on following technical records

      •       Original Observations

      •       Calculations/Computations and data transfers

      •       Calibration Report Format




                                                          STQC
Version 1.0         CERTIFIED CALIBRATION PROFFESSIONAL          MIII/CH1/107
Reporting Calibration Results

          Report and Certificate format
       • Appropriate to each type of
              test/calibration
       • For easy understanding / to
              avoid misuse

                                                          STQC
Version 1.0         CERTIFIED CALIBRATION PROFFESSIONAL          MIII/CH1/108
Calibration Reports / Certificates
   • Calibration Report
       A calibration report usually provides a set of results that show the performance
       of the instrument being calibrated in comparison to a reference instrument or
       standard. Measurements are recorded for each function, range and points
       tested, and the uncertainty of each measurement is reported.

   • Calibration Certificate
       A calibration certificate usually states the compliance of an instrument with the
       manufacturer's specification. Measurements made by the instrument being
       calibrated are compared to the manufacturer's specification. If the instrument
       is within specification, a statement that the instrument meets the
       manufacturer's specification at the points tested is made in the conformance
       certificate. Any points that are outside specification, or not able to be
       determined, are reported as exceptions. Assessment of meeting the
       manufacturer's specification takes into account the measurement uncertainty.
       A statement of conformance to the manufacturer's specification may also be
       given in calibration reports.

                                                                     STQC
Version 1.0              CERTIFIED CALIBRATION PROFFESSIONAL                  MIII/CH1/109
Calibration Reports / Certificates


      • Accurate, clear, unambiguous & objective
      • No recommendation for Calibration interval,
        unless asked for
      • Confidentiality in electronic transmission
      • Specified & approved format
      • Specific statement for amendment
      • Must include information requested by customer

               Sample report/certificate formats refer NCSL, RP-11


                                                              STQC
Version 1.0            CERTIFIED CALIBRATION PROFFESSIONAL           MIII/CH1/110
Contents of A Calibration Report
               a) A title -Test report/Calibration certificate

               b) Name and Address of laboratory

               c) Unique identification of the report/certificate

               d) Name and address of customer

               e) Identification of method used

               f) Unambiguous identification of item

               g) Date of receipt / performance of Test / Calibration

               h) Reference to sampling plan / procedure
                                                                 STQC
Version 1.0             CERTIFIED CALIBRATION PROFFESSIONAL             MIII/CH1/111
Contents of Calibration Report contd..
I) Test/Calibration results and units of measurement

j) Name, function, Signature of person authorizing the report/certificate

k) Statement-result related to the item Tested/Calibrated

l) Calibration Results with Units

m) Name(s), function(s) and signature(s) or equivalent identification of
   person(s) authorizing the report or certificate;

n) Statement to the effect that the results relate only to the items tested
   or calibrated, where relevant;

o) Subcontracted results clearly identified.

                                                           STQC
Version 1.0          CERTIFIED CALIBRATION PROFFESSIONAL          MIII/CH1/112
Calibration Reporting
      Calibration Result Interpretation where necessary
      ensure calibration certificates include the following:

              • Calibration conditions
              • Uncertainty of measurement
              • Measurement traceability statement
              • Statement of compliance with an identified
              metrological specification or clause
              • Compliance data records
              • Repairs/adjustments if any
              • Calibration Interval
              • Estimated Uncertainty

                                                             STQC
Version 1.0            CERTIFIED CALIBRATION PROFFESSIONAL          MIII/CH1/113
STQC
Version 1.0   CERTIFIED CALIBRATION PROFFESSIONAL          MIII/CH1/114
Amendment/Supplement to Report

     • Identification or reference to the original
       report identity.
     • when a complete new test report or
       calibration certificate is required, it must
       be uniquely identified and include a
       reference to the original it replaces




                                                         STQC
Version 1.0        CERTIFIED CALIBRATION PROFFESSIONAL          MIII/CH1/115
Acceptance of Calibration Report

   • Do you accept them or not?
   • How do you know if they are okay to
     accept?
   • What are the related issues? What are the
     steps to determine what is okay and what
     is not?



                                                         STQC
Version 1.0        CERTIFIED CALIBRATION PROFFESSIONAL          MIII/CH1/116
Acceptance of Calibration Report
     •   evaluate the "scope of accreditation." The laboratory scope, which is
         published with the accreditation or recognition certificate, defines the
         parameters and associated uncertainties that have been assessed by
         the accreditation bodies.
     •   An accreditation body logo
     •   evaluate standards to ensure suitability & Forward traceability report
         5.6.2 Show unbroken chain of traceability from an instrument to
         national standards.
     •   Rights to revoke acceptance of certificates for cause (bad
         measurement results) & seek clarifications
     •   Calibration certificate contents as per 5.10 of ISO/IEC 17025.
     •   Due for calibration 5.6.1 – determine instrument exceeded cal interval.
     •   Reverse Traceability report 4.9.1- Identify all the instruments that
         are affected if corrective action is neccessiated


                                                              STQC
Version 1.0            CERTIFIED CALIBRATION PROFFESSIONAL             MIII/CH1/117
HANDLING DEVIATIONS
                    - FROM CALIBRATION QUALITY LEVEL

              1. Interval adjustment
              2. Procedure Changes
              3. Specification Revision
              4. Derating or limited Calibration
              5. Removal of equipment from services
              6. Instrument Repair & Recalibration
              7. Upgradation of equipment
              8. Replacement

                                                            STQC
Version 1.0           CERTIFIED CALIBRATION PROFFESSIONAL          MIII/CH1/118
CALIBRATION LABELLING

    Valid Calibration: when all instrument parameters have been calibrated &
    meets the manufacturer’s specifications. If not affix following type of labels.

                  LABELING LIMITAION OF USE (Refer NCSL RP- 6)




    Labeling location, content, color & size depends on Calibration agency
                                                                      STQC
Version 1.0              CERTIFIED CALIBRATION PROFFESSIONAL                   MIII/CH1/119
External Calibration

 When Cal Service is subcontracted, ensure:
Approval from customer, preferably in writing,
 when the lab intends to subcontract.
Ensure and demonstrate subcontractors
 competence through evaluation & review
Laboratory’s responsibility for subcontractors
 work.
Maintain a register of all subcontracting work

                                                     STQC
Version 1.0    CERTIFIED CALIBRATION PROFFESSIONAL          MIII/CH1/120
External Calibration
 When Cal Services is availed, ensure:

 Credibility of Cal agency through accreditation & scope of accreditation
  covering your requirements

 Gain confidence in their services through visits/audits on:
      –   type of calibration service
      –   applicable documents, technical reference, or standard
      –   equipment specification
      –   service performance requirements
      –   physical, environmental, and data-format requirements
      –   safety and biohazard requirements
      –   confidentiality and security requirements
      –   certificate of calibration requirements




                                                                   STQC
Version 1.0              CERTIFIED CALIBRATION PROFFESSIONAL              MIII/CH1/121
CHAPTER - 7




                                                    STQC
Version 1.0   CERTIFIED CALIBRATION PROFFESSIONAL          MIII/CH1/122
REQUIREMENTS FOR DOCUMENTING RECORDS


1. ISO/IEC 17025:2005, clause 4.12, paragraph 4.12.2.1, states: "The laboratory
   shall retain records of original observations, derived data. . . to estab-lish an
   audit trail . . . for a defined period. The records for each calibration shall contain
   sufficient information to . . . enable the test or calibration to be repeated."
   Paragraph 4.12.2.2 states: "Observations, data and calculations shall be
   recorded at the time they are made. . . ."
2. ISO 9001:2000, clause 7.6.on Control of Monitoring & Measuring equipment.
3. ISO 14001:2004, Clause 4.5.2 on equipment used for environmental
   maesurements.
4. ISO 10012:2003, clause 6.2.3 states: "Records containing information
   required for the operation of the measurement management system shall
    be maintained. Documented procedures shall ensure the identification, storage,
    protec-tion, retrieval, retention time and disposition of records."
5. NCSLI RP-6, clause 5.6, Records, states: "Records should be maintained for all
   test and measuring equipment that is included in the calibration system


                                                                     STQC
Version 1.0             CERTIFIED CALIBRATION PROFFESSIONAL                    MIII/CH1/123
Why Calibration Records are necessary?

      • Understanding the relationship of historical data
        and stability of a standard
      • As Found, and when needed, As Left readings
        giving reliability of equipment
      • Log book of Instruments logging Calibration
        history on maintainability
      • Trend of measurements from periodic Calibration
        Reports
      • Authentication of Calibration Data/Results

                                                         STQC
Version 1.0        CERTIFIED CALIBRATION PROFFESSIONAL          MIII/CH1/124
Calibration Records
              •   Service Request Forms
              •   Work Orders/Job Orders
              •   Planning / Job allocation Records
              •   Observations/data computations/Measurement sheets, Uncertainty
                  Budget records
              •   Preliminary performance/verification records
              •   Calibration records of equipment used for cal. Control Charts on
                  trends of calibration standards
              •   Calibration Report / Certificates
              •   Replicate/ Retesting records
              •   Training records
              •   Proficiency/ Inter-laboratory Comparison records
              •   Nonconforming calibration, Customer complaints, CA/PA records
              •   Calibration System Audit records
              •   Repair/Adjustment/Maintenance Records              STQC
Version 1.0                  CERTIFIED CALIBRATION PROFFESSIONAL            MIII/CH1/125
Management of Calibration Records



      •   Legibility & Indelibility of records
      •   Identification of records
      •   Storage of records
      •   Maintenance/Preservation of records
      •   Retention & Disposition of records
      •   Amendment of records with authentication




                                                          STQC
Version 1.0         CERTIFIED CALIBRATION PROFFESSIONAL          MIII/CH1/126
STQC
Version 1.0   CERTIFIED CALIBRATION PROFFESSIONAL          MIII/CH1/127

More Related Content

What's hot

Revalidation presentation
Revalidation presentationRevalidation presentation
Revalidation presentationDenny Walthers
 
introduction to validation
introduction to validation introduction to validation
introduction to validation Sachin G
 
Calabration principles-chapter1
Calabration principles-chapter1Calabration principles-chapter1
Calabration principles-chapter1Complaint2015
 
Measurement Traceability
Measurement TraceabilityMeasurement Traceability
Measurement Traceabilitysigmatest2011
 
CLEANING VALIDATION for M.pharm and industry person
CLEANING VALIDATION  for M.pharm and industry personCLEANING VALIDATION  for M.pharm and industry person
CLEANING VALIDATION for M.pharm and industry personabhishek pandey
 
Batch Review And Batch Release.pptx
Batch Review And Batch Release.pptxBatch Review And Batch Release.pptx
Batch Review And Batch Release.pptxAbhishekJadhav189260
 
Cleaning validation
Cleaning validationCleaning validation
Cleaning validationVinay Jain
 
Uncertainity in Measurement
Uncertainity in MeasurementUncertainity in Measurement
Uncertainity in MeasurementThe Apprentiice
 
qualification of FTIR and DSC
qualification of FTIR and DSCqualification of FTIR and DSC
qualification of FTIR and DSCAmansharma1378
 
Qualification of Autoclave.pptx
Qualification of Autoclave.pptxQualification of Autoclave.pptx
Qualification of Autoclave.pptxDhruvi50
 
Qualification of Dry Heat Sterilizer or Tunnel.pptx
Qualification of Dry Heat Sterilizer or Tunnel.pptxQualification of Dry Heat Sterilizer or Tunnel.pptx
Qualification of Dry Heat Sterilizer or Tunnel.pptxDhruvi50
 
Validation Protocol
Validation ProtocolValidation Protocol
Validation ProtocolSagar Savale
 
Calibration of analytical balance
Calibration of analytical balanceCalibration of analytical balance
Calibration of analytical balanceSubrat Acharya
 

What's hot (20)

Calibration
CalibrationCalibration
Calibration
 
Cleaning validation
Cleaning validationCleaning validation
Cleaning validation
 
Revalidation presentation
Revalidation presentationRevalidation presentation
Revalidation presentation
 
introduction to validation
introduction to validation introduction to validation
introduction to validation
 
Calabration principles-chapter1
Calabration principles-chapter1Calabration principles-chapter1
Calabration principles-chapter1
 
Measurement Traceability
Measurement TraceabilityMeasurement Traceability
Measurement Traceability
 
CLEANING VALIDATION for M.pharm and industry person
CLEANING VALIDATION  for M.pharm and industry personCLEANING VALIDATION  for M.pharm and industry person
CLEANING VALIDATION for M.pharm and industry person
 
Qualification of HPLC
Qualification of HPLCQualification of HPLC
Qualification of HPLC
 
validation of autoclave
validation of autoclavevalidation of autoclave
validation of autoclave
 
Batch Review And Batch Release.pptx
Batch Review And Batch Release.pptxBatch Review And Batch Release.pptx
Batch Review And Batch Release.pptx
 
Cleaning validation
Cleaning validationCleaning validation
Cleaning validation
 
Qms
QmsQms
Qms
 
Uncertainity in Measurement
Uncertainity in MeasurementUncertainity in Measurement
Uncertainity in Measurement
 
qualification of FTIR and DSC
qualification of FTIR and DSCqualification of FTIR and DSC
qualification of FTIR and DSC
 
Qualification of Autoclave.pptx
Qualification of Autoclave.pptxQualification of Autoclave.pptx
Qualification of Autoclave.pptx
 
Qualification of Dry Heat Sterilizer or Tunnel.pptx
Qualification of Dry Heat Sterilizer or Tunnel.pptxQualification of Dry Heat Sterilizer or Tunnel.pptx
Qualification of Dry Heat Sterilizer or Tunnel.pptx
 
ISO / IEC 17025
ISO / IEC 17025ISO / IEC 17025
ISO / IEC 17025
 
Validation Protocol
Validation ProtocolValidation Protocol
Validation Protocol
 
Quality audit
Quality auditQuality audit
Quality audit
 
Calibration of analytical balance
Calibration of analytical balanceCalibration of analytical balance
Calibration of analytical balance
 

Similar to Calibration systems

qualification of instrument(UV & FTIR) BY Bhumi Suratiya, M.Pharm sem 2.pptx
qualification of instrument(UV & FTIR) BY Bhumi Suratiya, M.Pharm sem 2.pptxqualification of instrument(UV & FTIR) BY Bhumi Suratiya, M.Pharm sem 2.pptx
qualification of instrument(UV & FTIR) BY Bhumi Suratiya, M.Pharm sem 2.pptxBhumiSuratiya
 
Environmental monitoring - nonviable particle testing
Environmental monitoring - nonviable particle testingEnvironmental monitoring - nonviable particle testing
Environmental monitoring - nonviable particle testingJerry Fahrni
 
Qal 991 particulate monitor datasheet
Qal 991 particulate monitor datasheetQal 991 particulate monitor datasheet
Qal 991 particulate monitor datasheeta1-cbiss
 
qualification of analytical instruments..M pharmacy 1st year.validation
qualification of analytical instruments..M pharmacy 1st year.validationqualification of analytical instruments..M pharmacy 1st year.validation
qualification of analytical instruments..M pharmacy 1st year.validationSohailPattan
 
Metrology manager resume
Metrology manager resumeMetrology manager resume
Metrology manager resumeKarthick G
 
Qualification of analytical instruments
Qualification of analytical instruments Qualification of analytical instruments
Qualification of analytical instruments PUNEET NIRMAL
 
Kolkata 2010 7_waree_calibrating_ff_transmitters (1)
Kolkata 2010 7_waree_calibrating_ff_transmitters (1)Kolkata 2010 7_waree_calibrating_ff_transmitters (1)
Kolkata 2010 7_waree_calibrating_ff_transmitters (1)Ashok Kumar Barla
 
Inline freeness transmitter
Inline freeness transmitterInline freeness transmitter
Inline freeness transmitterMiia Moffitt
 
EPM In-Line Verification 24-04-2014
EPM In-Line Verification 24-04-2014EPM In-Line Verification 24-04-2014
EPM In-Line Verification 24-04-2014Bert Roos
 

Similar to Calibration systems (20)

Calibration
CalibrationCalibration
Calibration
 
Presentation for APPON
Presentation for APPONPresentation for APPON
Presentation for APPON
 
qualification of instrument(UV & FTIR) BY Bhumi Suratiya, M.Pharm sem 2.pptx
qualification of instrument(UV & FTIR) BY Bhumi Suratiya, M.Pharm sem 2.pptxqualification of instrument(UV & FTIR) BY Bhumi Suratiya, M.Pharm sem 2.pptx
qualification of instrument(UV & FTIR) BY Bhumi Suratiya, M.Pharm sem 2.pptx
 
Standards - Accuracy, Precision & Traceability
Standards - Accuracy, Precision & Traceability Standards - Accuracy, Precision & Traceability
Standards - Accuracy, Precision & Traceability
 
Environmental monitoring - nonviable particle testing
Environmental monitoring - nonviable particle testingEnvironmental monitoring - nonviable particle testing
Environmental monitoring - nonviable particle testing
 
calibration.pptx
calibration.pptxcalibration.pptx
calibration.pptx
 
Qal 991 particulate monitor datasheet
Qal 991 particulate monitor datasheetQal 991 particulate monitor datasheet
Qal 991 particulate monitor datasheet
 
qualification of analytical instruments..M pharmacy 1st year.validation
qualification of analytical instruments..M pharmacy 1st year.validationqualification of analytical instruments..M pharmacy 1st year.validation
qualification of analytical instruments..M pharmacy 1st year.validation
 
Sop 1 20140911
Sop 1 20140911Sop 1 20140911
Sop 1 20140911
 
Calibraciones
CalibracionesCalibraciones
Calibraciones
 
Qualification of high performance liquid chromatography
Qualification of high performance liquid chromatographyQualification of high performance liquid chromatography
Qualification of high performance liquid chromatography
 
Metrology manager resume
Metrology manager resumeMetrology manager resume
Metrology manager resume
 
Qualification of analytical instruments
Qualification of analytical instruments Qualification of analytical instruments
Qualification of analytical instruments
 
Kolkata 2010 7_waree_calibrating_ff_transmitters (1)
Kolkata 2010 7_waree_calibrating_ff_transmitters (1)Kolkata 2010 7_waree_calibrating_ff_transmitters (1)
Kolkata 2010 7_waree_calibrating_ff_transmitters (1)
 
Inline freeness transmitter
Inline freeness transmitterInline freeness transmitter
Inline freeness transmitter
 
Calibration
CalibrationCalibration
Calibration
 
EPM In-Line Verification 24-04-2014
EPM In-Line Verification 24-04-2014EPM In-Line Verification 24-04-2014
EPM In-Line Verification 24-04-2014
 
Anab Guidance Scopes Accreditation
Anab Guidance Scopes AccreditationAnab Guidance Scopes Accreditation
Anab Guidance Scopes Accreditation
 
Overview of Instrument Calibration
Overview of Instrument CalibrationOverview of Instrument Calibration
Overview of Instrument Calibration
 
TPI - Example Report.
TPI - Example Report.TPI - Example Report.
TPI - Example Report.
 

Recently uploaded

What Could Cause A VW Tiguan's Radiator Fan To Stop Working
What Could Cause A VW Tiguan's Radiator Fan To Stop WorkingWhat Could Cause A VW Tiguan's Radiator Fan To Stop Working
What Could Cause A VW Tiguan's Radiator Fan To Stop WorkingEscondido German Auto
 
907MTAMount Coventry University Bachelor's Diploma in Engineering
907MTAMount Coventry University Bachelor's Diploma in Engineering907MTAMount Coventry University Bachelor's Diploma in Engineering
907MTAMount Coventry University Bachelor's Diploma in EngineeringFi sss
 
Independent Andheri Call Girls 9833363713
Independent Andheri Call Girls 9833363713Independent Andheri Call Girls 9833363713
Independent Andheri Call Girls 9833363713Komal Khan
 
(Griffith毕业证)格里菲斯大学毕业证毕业证成绩单修改留信学历认证原版一比一
(Griffith毕业证)格里菲斯大学毕业证毕业证成绩单修改留信学历认证原版一比一(Griffith毕业证)格里菲斯大学毕业证毕业证成绩单修改留信学历认证原版一比一
(Griffith毕业证)格里菲斯大学毕业证毕业证成绩单修改留信学历认证原版一比一ejgeojhg
 
(办理学位证)墨尔本大学毕业证(Unimelb毕业证书)成绩单留信学历认证原版一模一样
(办理学位证)墨尔本大学毕业证(Unimelb毕业证书)成绩单留信学历认证原版一模一样(办理学位证)墨尔本大学毕业证(Unimelb毕业证书)成绩单留信学历认证原版一模一样
(办理学位证)墨尔本大学毕业证(Unimelb毕业证书)成绩单留信学历认证原版一模一样whjjkkk
 
2024 TOP 10 most fuel-efficient vehicles according to the US agency
2024 TOP 10 most fuel-efficient vehicles according to the US agency2024 TOP 10 most fuel-efficient vehicles according to the US agency
2024 TOP 10 most fuel-efficient vehicles according to the US agencyHyundai Motor Group
 
UNIT-III-TRANSMISSION SYSTEMS REAR AXLES
UNIT-III-TRANSMISSION SYSTEMS REAR AXLESUNIT-III-TRANSMISSION SYSTEMS REAR AXLES
UNIT-III-TRANSMISSION SYSTEMS REAR AXLESDineshKumar4165
 
Call Girl Service Global Village Dubai +971509430017 Independent Call Girls G...
Call Girl Service Global Village Dubai +971509430017 Independent Call Girls G...Call Girl Service Global Village Dubai +971509430017 Independent Call Girls G...
Call Girl Service Global Village Dubai +971509430017 Independent Call Girls G...kexey39068
 
UNOSAFE ELEVATOR PRIVATE LTD BANGALORE BROUCHER
UNOSAFE ELEVATOR PRIVATE LTD BANGALORE BROUCHERUNOSAFE ELEVATOR PRIVATE LTD BANGALORE BROUCHER
UNOSAFE ELEVATOR PRIVATE LTD BANGALORE BROUCHERunosafeads
 
定制(MSVU毕业证书)圣文森山大学毕业证毕业证成绩单原版一比一定制
定制(MSVU毕业证书)圣文森山大学毕业证毕业证成绩单原版一比一定制定制(MSVU毕业证书)圣文森山大学毕业证毕业证成绩单原版一比一定制
定制(MSVU毕业证书)圣文森山大学毕业证毕业证成绩单原版一比一定制miwhjk
 
Not Sure About VW EGR Valve Health Look For These Symptoms
Not Sure About VW EGR Valve Health Look For These SymptomsNot Sure About VW EGR Valve Health Look For These Symptoms
Not Sure About VW EGR Valve Health Look For These SymptomsFifth Gear Automotive
 
Equity & Freight Electrification by Jose Miguel Acosta Cordova
Equity & Freight Electrification by Jose Miguel Acosta CordovaEquity & Freight Electrification by Jose Miguel Acosta Cordova
Equity & Freight Electrification by Jose Miguel Acosta CordovaForth
 
Dubai Call Girls Services Call 09900000000
Dubai Call Girls Services Call 09900000000Dubai Call Girls Services Call 09900000000
Dubai Call Girls Services Call 09900000000Komal Khan
 
原版1:1定制宾州州立大学毕业证(PSU毕业证)#文凭成绩单#真实留信学历认证永久存档
原版1:1定制宾州州立大学毕业证(PSU毕业证)#文凭成绩单#真实留信学历认证永久存档原版1:1定制宾州州立大学毕业证(PSU毕业证)#文凭成绩单#真实留信学历认证永久存档
原版1:1定制宾州州立大学毕业证(PSU毕业证)#文凭成绩单#真实留信学历认证永久存档208367051
 
原版1:1定制阳光海岸大学毕业证(JCU毕业证)#文凭成绩单#真实留信学历认证永久存档
原版1:1定制阳光海岸大学毕业证(JCU毕业证)#文凭成绩单#真实留信学历认证永久存档原版1:1定制阳光海岸大学毕业证(JCU毕业证)#文凭成绩单#真实留信学历认证永久存档
原版1:1定制阳光海岸大学毕业证(JCU毕业证)#文凭成绩单#真实留信学历认证永久存档208367051
 
15 Disadvantages of Automated Farming: Balancing Efficiency with Environment ...
15 Disadvantages of Automated Farming: Balancing Efficiency with Environment ...15 Disadvantages of Automated Farming: Balancing Efficiency with Environment ...
15 Disadvantages of Automated Farming: Balancing Efficiency with Environment ...CIOWomenMagazine
 
办理学位证(MLU文凭证书)哈勒 维滕贝格大学毕业证成绩单原版一模一样
办理学位证(MLU文凭证书)哈勒 维滕贝格大学毕业证成绩单原版一模一样办理学位证(MLU文凭证书)哈勒 维滕贝格大学毕业证成绩单原版一模一样
办理学位证(MLU文凭证书)哈勒 维滕贝格大学毕业证成绩单原版一模一样umasea
 
UNIT-V-ELECTRIC AND HYBRID VEHICLES.pptx
UNIT-V-ELECTRIC AND HYBRID VEHICLES.pptxUNIT-V-ELECTRIC AND HYBRID VEHICLES.pptx
UNIT-V-ELECTRIC AND HYBRID VEHICLES.pptxDineshKumar4165
 

Recently uploaded (20)

What Could Cause A VW Tiguan's Radiator Fan To Stop Working
What Could Cause A VW Tiguan's Radiator Fan To Stop WorkingWhat Could Cause A VW Tiguan's Radiator Fan To Stop Working
What Could Cause A VW Tiguan's Radiator Fan To Stop Working
 
907MTAMount Coventry University Bachelor's Diploma in Engineering
907MTAMount Coventry University Bachelor's Diploma in Engineering907MTAMount Coventry University Bachelor's Diploma in Engineering
907MTAMount Coventry University Bachelor's Diploma in Engineering
 
Independent Andheri Call Girls 9833363713
Independent Andheri Call Girls 9833363713Independent Andheri Call Girls 9833363713
Independent Andheri Call Girls 9833363713
 
(Griffith毕业证)格里菲斯大学毕业证毕业证成绩单修改留信学历认证原版一比一
(Griffith毕业证)格里菲斯大学毕业证毕业证成绩单修改留信学历认证原版一比一(Griffith毕业证)格里菲斯大学毕业证毕业证成绩单修改留信学历认证原版一比一
(Griffith毕业证)格里菲斯大学毕业证毕业证成绩单修改留信学历认证原版一比一
 
(办理学位证)墨尔本大学毕业证(Unimelb毕业证书)成绩单留信学历认证原版一模一样
(办理学位证)墨尔本大学毕业证(Unimelb毕业证书)成绩单留信学历认证原版一模一样(办理学位证)墨尔本大学毕业证(Unimelb毕业证书)成绩单留信学历认证原版一模一样
(办理学位证)墨尔本大学毕业证(Unimelb毕业证书)成绩单留信学历认证原版一模一样
 
2024 TOP 10 most fuel-efficient vehicles according to the US agency
2024 TOP 10 most fuel-efficient vehicles according to the US agency2024 TOP 10 most fuel-efficient vehicles according to the US agency
2024 TOP 10 most fuel-efficient vehicles according to the US agency
 
UNIT-III-TRANSMISSION SYSTEMS REAR AXLES
UNIT-III-TRANSMISSION SYSTEMS REAR AXLESUNIT-III-TRANSMISSION SYSTEMS REAR AXLES
UNIT-III-TRANSMISSION SYSTEMS REAR AXLES
 
Call Girl Service Global Village Dubai +971509430017 Independent Call Girls G...
Call Girl Service Global Village Dubai +971509430017 Independent Call Girls G...Call Girl Service Global Village Dubai +971509430017 Independent Call Girls G...
Call Girl Service Global Village Dubai +971509430017 Independent Call Girls G...
 
UNOSAFE ELEVATOR PRIVATE LTD BANGALORE BROUCHER
UNOSAFE ELEVATOR PRIVATE LTD BANGALORE BROUCHERUNOSAFE ELEVATOR PRIVATE LTD BANGALORE BROUCHER
UNOSAFE ELEVATOR PRIVATE LTD BANGALORE BROUCHER
 
Indian Downtown Call Girls # 00971528903066 # Indian Call Girls In Downtown D...
Indian Downtown Call Girls # 00971528903066 # Indian Call Girls In Downtown D...Indian Downtown Call Girls # 00971528903066 # Indian Call Girls In Downtown D...
Indian Downtown Call Girls # 00971528903066 # Indian Call Girls In Downtown D...
 
定制(MSVU毕业证书)圣文森山大学毕业证毕业证成绩单原版一比一定制
定制(MSVU毕业证书)圣文森山大学毕业证毕业证成绩单原版一比一定制定制(MSVU毕业证书)圣文森山大学毕业证毕业证成绩单原版一比一定制
定制(MSVU毕业证书)圣文森山大学毕业证毕业证成绩单原版一比一定制
 
Not Sure About VW EGR Valve Health Look For These Symptoms
Not Sure About VW EGR Valve Health Look For These SymptomsNot Sure About VW EGR Valve Health Look For These Symptoms
Not Sure About VW EGR Valve Health Look For These Symptoms
 
Equity & Freight Electrification by Jose Miguel Acosta Cordova
Equity & Freight Electrification by Jose Miguel Acosta CordovaEquity & Freight Electrification by Jose Miguel Acosta Cordova
Equity & Freight Electrification by Jose Miguel Acosta Cordova
 
Hot Sexy call girls in Pira Garhi🔝 9953056974 🔝 escort Service
Hot Sexy call girls in Pira Garhi🔝 9953056974 🔝 escort ServiceHot Sexy call girls in Pira Garhi🔝 9953056974 🔝 escort Service
Hot Sexy call girls in Pira Garhi🔝 9953056974 🔝 escort Service
 
Dubai Call Girls Services Call 09900000000
Dubai Call Girls Services Call 09900000000Dubai Call Girls Services Call 09900000000
Dubai Call Girls Services Call 09900000000
 
原版1:1定制宾州州立大学毕业证(PSU毕业证)#文凭成绩单#真实留信学历认证永久存档
原版1:1定制宾州州立大学毕业证(PSU毕业证)#文凭成绩单#真实留信学历认证永久存档原版1:1定制宾州州立大学毕业证(PSU毕业证)#文凭成绩单#真实留信学历认证永久存档
原版1:1定制宾州州立大学毕业证(PSU毕业证)#文凭成绩单#真实留信学历认证永久存档
 
原版1:1定制阳光海岸大学毕业证(JCU毕业证)#文凭成绩单#真实留信学历认证永久存档
原版1:1定制阳光海岸大学毕业证(JCU毕业证)#文凭成绩单#真实留信学历认证永久存档原版1:1定制阳光海岸大学毕业证(JCU毕业证)#文凭成绩单#真实留信学历认证永久存档
原版1:1定制阳光海岸大学毕业证(JCU毕业证)#文凭成绩单#真实留信学历认证永久存档
 
15 Disadvantages of Automated Farming: Balancing Efficiency with Environment ...
15 Disadvantages of Automated Farming: Balancing Efficiency with Environment ...15 Disadvantages of Automated Farming: Balancing Efficiency with Environment ...
15 Disadvantages of Automated Farming: Balancing Efficiency with Environment ...
 
办理学位证(MLU文凭证书)哈勒 维滕贝格大学毕业证成绩单原版一模一样
办理学位证(MLU文凭证书)哈勒 维滕贝格大学毕业证成绩单原版一模一样办理学位证(MLU文凭证书)哈勒 维滕贝格大学毕业证成绩单原版一模一样
办理学位证(MLU文凭证书)哈勒 维滕贝格大学毕业证成绩单原版一模一样
 
UNIT-V-ELECTRIC AND HYBRID VEHICLES.pptx
UNIT-V-ELECTRIC AND HYBRID VEHICLES.pptxUNIT-V-ELECTRIC AND HYBRID VEHICLES.pptx
UNIT-V-ELECTRIC AND HYBRID VEHICLES.pptx
 

Calibration systems

  • 1. MODULE - III STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/1
  • 2. Calibration A set of operations, performed in accordance with a definite documented procedure, that compares the measurements performed by an instrument to those made by a more accurate instrument or standard, for the purpose of detecting and reporting or eliminating by adjustment, errors in the instrument tested. As per ISO/IEC Guide 2 STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/2
  • 3. Calibration definition (a) operation establishing the relation between quantity values provided by measurement standards and the corresponding indications of a measuring system, carried out under specified conditions and including evaluation of measurement uncertainty definition (b) operation that establishes the relation, obtained by reference to one or more measurement standards, that exists under specified conditions, between the indication of a measuring system and the measurement result that would be obtained using the measuring system As per VIM STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/3
  • 4. Why to Calibrate ? • Initial calibration assures T&ME with in Spec • Periodic calibration is essential to maintain control of measurement process made by TME • Assures common reference i.e Traceability to National / International standards • Ensures that equipment has been satisfactorily working since last calibrated & assumed that the equipment continues to work satisfactorily until next calibrated • Calibration records study enables extending/reducing the calibration interval STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/4
  • 5. OBJECTIVES OF CALIBRATION Maintaining traceability of measurements Uncertainty limits Traceability Reference measurements STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/5
  • 6. Why Calibration & Traceability are intertwined? • To measure the “Same Measure” • Interchangeability of products • Legal requirement & consequence of liability • To demonstrate metrological confirmation of product • Consistent control of product/measurement STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/6
  • 7. When is calibration not required? • When equipment is not in use . • When eqpt is used as null detector • When eqpt is used as transfer device whose measurement or output value is not explicitly used. • Fundamental / Absolute standards. • When instrument is fail-safe • When instrument makes measurements or provides known outputs which are monitored by a calibrated device, meter or gage during use. STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/7
  • 8. When to Calibrate ? - Calibrate too often it is waste of time and money. - Calibrate too seldom it is waste of product or wrong decisions. - Balance between cost to calibrate and cost not to calibrate - Universally Applicable single best practice for establishing & adjusting cal interval not available. - Refer DAP, RP1 STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/8
  • 9. Where to Calibrate? • Calibration is performed in the I/II/III party laboratory under controlled environment. (The environment need only to be controlled to the extent required.) • Accredited Calibration Laboratory • Manufacturer’s / Supplier’s Laboratory ensuring valid traceability STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/9
  • 10. Where to Calibrate ?contd… • Accredited Laboratory ensuring the quality of calibration services. • In the absence of the above, the services of calibration laboratory for the parameter of interest may be availed ensuring meeting technical requirements of ISO/IEC 17025. STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/10
  • 11. Basic Requirements - Location/ Environment • Noise Level • Vibration • EMI/EMC • Cleanliness • Lighting • Power supply • Accessibility STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/11
  • 12. Basic Requirements – Location/Environment Quiet * Away from heavy traffic, factories, workshop etc * Effective separation from non compatible areas * Noise : less than 60 dbA * Vibration : Taken care in building design & site selection noticeable in galvanometers, mercury column, mass balance, dimensional measuring equipment 0.25μm max displacement amplitude from 0.1Hz to 30Hz and 0.001 g max 30Hz to 200Hz * EMI : * taken care in site selection & use of shielding/ filtering * field strength to be less than 100μV /m STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/12
  • 13. Basic Requirements – Location/Environment  Clean Good house keeping Higher degree of particulate control (Dimensional & Mass Metrology)  Good lighting Recommended level of illumination is 450 – 700lux Fluorescent lamps preferred to minimize heating effect  Services Separate feeder line Uninterrupted / Regulated power supply (+- 5%) Adequate measures against spikes & transients Mains Earthing (< 1 ohm) Adequate measures against ground currents STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/13
  • 14. Basic Requirements – Location/Environment  Accessibility Standards / equipment Firefighting devices Quality System Document & Instrument Manuals STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/14
  • 15. Calibration laboratory - basic requirements  Location – Permanent/On-site  Temperature  Humidity  Environment  Air flow  Equipment  Filtration  Electromagnetic Screening  Staff  Noise Level  Vibration  Management  EMI/EMC Documents  Cleanliness Reference standards  Lighting Records Transfer Standards  Power supply Calibration Working standards  Accessibility Training Connecting leads  Receiving & Calibration area Authority adaptors/accessories Responsibility Subsidiary equipment Technical Computers & automation Competence software STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/15
  • 16. Basic Requirements – Equipment  Transfer Standards  Reference standards (Optional)  Working standards  Connecting leads and adaptors  Subsidiary equipment  Computers & automation software STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/16
  • 17. Basic Requirements – Personnel  Training  Authority  Responsibility  Technical Competence STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/17
  • 18. Basic Requirements – Management  Quality management & Technical operations  Traceability  Calibration interval  Audits/ Management reviews  Implementation of corrective /preventive actions STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/18
  • 19. Calibration System Why Calibration system ? - To meet the commitment of the industry to provide quality product. - TMI is required to be controlled to provide this assurance. - Through calibration - Shall include all the elements necessary for ready detection of equipment deficiencies and provide objective evidence of conformance in the form of records. STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/19
  • 20. Calibration System It is defined as a set of inter-related and interacting elements which influence the calibration process and ultimately quality of calibration result. STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/20
  • 21. Elements of Calibration System • Qualification • Measurement traceability • Adequacy of calibration procedure • Calibration procedure • Qualification labeling • Sealing for integrity • Calibration interval • Records Maintenance STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/21
  • 22. Elements of Calibration System contd… • Environmental Controls • Handling , Preservation & Storage • Non- conforming Measurement Equipment • Scheduling / Recall system • Personnel requirement • Audit requirement • Calibration Quality • Use of outside service STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/22
  • 23. ACCREDITATION REQUIREMENTS The laboratory must demonstrate control and competence  Environmental control  Traceability of standards  Calibration procedures  Competence of staff / Authorization of personnel  Participation in PT programmes  In-service checks/ control charts  Validation of methods  Internal audit & Management review records STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/23
  • 24. STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/24
  • 25. CHAPTER - 1 STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/25
  • 26. CALIBRATION PROCEDURES Procedure -Ensures consistency in operation -Objective evidence Calibration procedure provides -Specified performance tolerance and functional criteria -Essential information & instruction to determine the integrity of eqpt. -Logical & timely sequence of necessary facts to understand and verify DUC to its intended performance capability. STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/26
  • 27. CALIBRATION PROCEDURES - Guides personnel - eqpt. Setup, interconnection, control operations - specification interpretation - tolerance limits - any adjustment, correction - Establish uniformity of methods/techniques - Shall be written such that personnel shall take minimum interpretation, effort and time. - Shall be written by keeping performance of DUC in mind. STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/27
  • 28. CALIBRATION PROCEDURES Requirements Source: -Prepared internally -Prepared by the manufacturer -External agency STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/28
  • 29. CALIBRATION PROCEDURES Completeness: - Sufficient instruction Approval: - shall be approved and controlled to meet regulatory and contractual requirements. Software: - Validation and control STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/29
  • 30. CALIBRATION PROCEDURES BROADLY COVERS - - Performance requirements of DUC - Performance requirements of standard - Preliminary operation - Calibration process - Calibration results - Closing operation - Storage and handling STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/30
  • 31. CALIBRATION PROCEDURES – TYPICAL CONTENT • Appropriate identification / document control. • Scope • Purpose - description & performance requirements of unit being calibrated • Parameters or quantities and ranges • Calibration equipment/system/setup • Reference standards and reference materials required • Reference Documents – standard/specifications/manuals • Adaptors/Accessories required • Environmental conditions required including stabilization period needed. STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/31
  • 32. CALIBRATION PROCEDURES – TYPICAL CONTENT • Description of the procedure, including –  Apprising of identification mark, sampling, handling, transporting, storing and preparation of items.  Preliminary Checks to be made before the work is started.  Checks that the equipment is working properly, and where required ,  Calibration process & adjustments if any, of the equipment  The method of recording the observations and results.  Any safety measures/precautions to be observed. • Criteria and / or requirements for approval / acceptance / rejection. • Data to be recorded and method of analysis and presentation. • The uncertainly or the procedure for estimating uncertainty. • Calibration status/labeling STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/32
  • 33. CALIBRATION PROCEDURE CONTROL  Deviations from instructions authorized & recorded  Assign responsibility  Establish document control with a standard format  Draft, comment and edit procedures/instructions  Review and authorize procedures  Print, bind, present and make available to all operational staff  Regular review and amendment as needed  Archive the superseded methods STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/33
  • 34. CALIBRATION PROCEDURES EXERCISE – 1 DEVELOPMENT & REVIEW OF PROCEDURES STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/34
  • 35. CALIBRATION PROCEDURES Section II -table defining performance specification (minimum use specification) required of the standard -Listing of specific manufacturer model (representative eqpt.) which are applicable and can operate with in the specified requirements. Equipment which has no known substitute should be clearly identified and described. -Environmental condition -Any special jigs and fixtures required STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/35
  • 36. CALIBRATION PROCEDURES Section III -Listing of all preliminary actions including functional checks to be performed prior to calibration Section IV -Listing of instructions and all steps to be performed in measurement process Section V -A results data sheet blank form with nominal values and tolerance limits. STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/36
  • 37. CALIBRATION METHODS • Standard methods – Direct/Indirect/Ratio/Substitution/Transfer/Differential • Non-standard methods • Laboratory developed methods NEEDS VALIDATION • Expansion of Scope • Customer given method STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/37
  • 38. VALIDATION OF CALIBRATION METHODS • Selectivity • Uncertainty of the results • Range • Repeatability • Reproducibility • Linearity • Limit of detection • Limit of quantization • Robustness • Bias • Precision STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/38
  • 39. GENERAL CALIBRATION TECHNIQUES • SPANNING • NULLING • ZEROING • LINEARIZATION STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/39
  • 40. GENERAL CALIBRATION METHODS NULLING • If the nominal value of UUC is same as that of reference comparison may be made using Nulling process. • Instruments like Galvanometers, Null Detectors & Comparators work on this principle. The scale indicates the sensitivity of detection in these meters. • Nulling is used for more accurate measurements since Null process can be quite sensitive to minute difference between known & unknown values. • Note that the Null detectors need not be calibrated at all for the measured values & the sensitivity of detection is by design, by scale of indication. STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/40
  • 41. GENERAL CALIBRATION METHODS ZEROING • a compensation technique for offset errors due to device to device offset, variation (trim errors), mechanical stresses (mounting stresses), shifts due to temperature and aging. • For zero referencing for >0 measurements • adjusting an instrument to read zero manually / auto zeroing STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/41
  • 42. GENERAL CALIBRATION METHODS SPANNING - is adjusting the instrument to read the known concentration of a gas / material / chemical - Range of parameter which the unit/equipment can measure - Scale of measurement STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/42
  • 43. GENERAL CALIBRATION METHODS SPANNING & ZEROING - are commonly performed during instrument start-up or before calibration. Ex.: The monitor’s combustible and toxic sensors are zeroed (adjusted to zero) in a known fresh air environment and the oxygen sensor is spanned to fresh air (21%). It is critical that this only be con- ducted in a known fresh air environment. STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/43
  • 44. GENERAL CALIBRATION METHODS LINEARIZATION To approximate a nonlinear system by a linear one is called Linearization. Ex.: Interfacing an NTC thermister to a computerized data acquisition system linearization is used to convert the resistance output of the thermister to voltage. STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/44
  • 45. CALIBRATION METHODS SELECTION ensure laboratory selects and uses test and/or calibration methods that: • meet the needs of the customer • are appropriate for the calibration task • the customer has been informed of the method chosen (if not specified) • where appropriate, are based on latest international, regional or national standards and where necessary the standard be supplemented with additional details to ensure consistent approach • have been verified for use in the laboratory, if a standard method Ensure laboratory has competence to choose the appropriate methods STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/45
  • 46. CHAPTER - 3 STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/46
  • 47. Basic Requirements – Environment  Temperature  Humidity  Air flow  Filtration  Electromagnetic Screening STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/47
  • 48. Basic Requirements – Environment  Temperature * Lab design to take into account heat generation through personnel, equipment, light etc * Recommended continuous & permanent recording * Temperature monitoring to be installed at eqpt. or operator level * Avoid temp gradient through uniform air flow * Precise metrology requires tight temp control & measurement of actual temp. to apply meaningful corrections * Allow artifact / eqpt. to attain thermal equilibrium (warm up) STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/48
  • 49. Basic Requirements – Environment  Humidity * Both low & high humidity to be avoided : static electricity & leakage path * 45 + 10% RH recommended for all general lab  Air flow * A well designed system to provide uniform air circulation with out stratification to minimize temp gradient in & around the lab * Floor to ceiling vertical laminar flow of air is recommended to minimize temp gradient to + 0.1 0C * Air velocity of 0.71 + 0.28 cubic mts /minute recommended (3 to 5 ft above the floor) STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/49
  • 50. Basic Requirements – Environment  Filtration * Provide filtration to limit particulates * High degree of particulate control : optical, mass, dimensional metrology etc.  Electro-magnetic Screening * Field strength to be less than 100μV/m * For conducted electromagnetic energy on power lines, open circuit voltage to be less than 100mV * Provide adequate shielding/ filtering. STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/50
  • 51. Accommodation of Cal Lab. Accommodation is the premises where the laboratory functions are carried out. Considerations : • Design and construction materials • Space including storage • Services and facilities • Access, housekeeping, security and safety • Separation of incompatible activities • Area for administrative duties STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/51
  • 52. Control of Environmental Parameters • Define critical environmental parameters that affect quality of calibration result • Define nominal values & tolerable limits • Monitor environmental parameters • Take remedial measures, if required • Halt Calibration process, if environment is not conducive. Refer RP 6 on Laboratory Design STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/52
  • 53. Recommended Env. Conditions - as per RPCL by CLAS in PALCAN Handbook published by Council of Canada STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/53
  • 54. ACCOMMODATION & ENVIRONMENT - TYPICAL 1. Clean & vibration free location for Monitoring & Measuring devices. 2. Temperature : Ambient / 23 ± 1 °C OR 23 ± 2 °c 3. Humidity : 50 ± 10 % 4. Illumination level : 250 – 450 (700 lux) 5. Noise level – Below 60 dBA where meas. are affected 6. Rigid table for installation of equipment 7. Noisy vacuum source and Pneumatic cylinder to be outside the calibration/inspection room/area REFER NABL SPECIFIC CRITERIA 121 - 126 STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/54
  • 55. Laboratory Storage Space Storage must also be such as to not interfere with test or calibration results. Proper storage is needed for :  Calibration jobs (current and retained)  Working materials/consumables/adaptors/accessories/cables etc.  Laboratory documents  Hazardous substances  Staff belongings  Clerical materials and equipment  Laboratory records REFER NABL SPECIFIC CRITERIA STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/55
  • 56. CHAPTER - 4 STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/56
  • 57. a) Handling, storage, preservation, transportation, safety & security of Calibration unit/equipment/job b) Customer requirements c) Laboratory capabilities – availability of personnel, cal facilities, technical competence, cal method, accommodation & environment, traceability, sampling etc. d) Calibration – measurement data, computations, uncertainty budget, compliance of results e) Reporting results – preparation, approval, issue f) Labeling g) Post cal service activities STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/57
  • 58. Calibration Equipment • is available and functioning properly • is capable of achieving required accuracy • complies with relevant specifications • has calibration programs established for key quantities or values • is calibrated or checked before being placed into service • is checked and/or calibrated before use (see 5.6 also) • is operated by authorised personnel • has current instructions on use and maintenance • is uniquely identified, where practicable Record of specification checking and maintained evidence. STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/58
  • 59. Equipment Records • identity of the equipment and its software • manufacturer’s name, model, and serial number or other unique identification • evidence that the equipment complies with the accuracy requirements and specifications relevant to the tests or calibrations • current location, where appropriate • the manufacturer’s instructions, if available, or reference to their location • calibration history and due date of next calibration • the maintenance plan, where appropriate, and maintenance carried out to date • any damage/malfunction/modification/repair to the equip. STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/59
  • 60. Equipment handling ensure procedures for measuring equipment are documented and include: • safe handling • transport • storage • use • planned maintenance – Preventive/Breakdown/In-service Checks, Intermediate Checks • where applicable, that copies of correction factors are correctly updated STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/60
  • 61. Equipment out-of-service ensure equipment subjected to overloading or mishandling, giving suspect results, or shown to be defective or outside specified limits is taken out of service, and is: • isolated or clearly labelled or marked as being out of service • examined for the effect of the defect or departure from specified limits on previous tests and/or calibrations • addressed under the “Control of nonconforming work” procedure Have a procedure for handling such situations STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/61
  • 62. Calibration Status • Ensure equipment calibration status is identified • Calibration labelling based on Cal agency Label that indicates date of calibration due date for next calibration, person responsible for calibration. For items where marking cannot be done it e.g. Weights, gauges, the status can be put in the Register/colour coding etc. • where practicable and where intermediate checks are needed to maintain confidence in the calibration status that a procedure is documented to carry out these checks STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/62
  • 63. CALIBRATION AS A PROCESS Process A – Preparation of list all test & measuring equipment that need to be under calibration control. Process B – Calibration of various test and measuring equipment (in-house calibration), if facility exist. Process C - Calibration of test & measuring equipment (from external Calibration lab) Process D -Verification of calibration results for compliance to measurement requirements STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/63
  • 64. Measuring Equipment – Process Model MEASURING PROCESS I/P Sensor Instrumentation Display O/P Measurand Result Sensor may not be required in every case STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/64
  • 65. Measuring Equipment • Measuring instrument/system • Software • Reference standard/material • Auxiliary apparatus/Adaptors/Accessories/Cables etc. • Any combination of the above Measuring equipment requires metrological confirmation STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/65
  • 66. Leading Instrumental Characteristics • Range • Error • Resolution • Drift • Input Impedance (Loading effect) • Compatibility between sensor & Instrumentation • Input – Output relationship (characteristic) • Hysterisis • Ruggedness • Mounting STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/66
  • 67. Checks on Measuring Equipment • Mechanical Zero / Zeroing /Auto zero • Check-up of readings during ascent & descent • Check-up of resolution • Check-up of linearity error, if any • Repeatability of measurements • Reproducibility of measurements STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/67
  • 68. Metrological Confirmation of Eqpt. • Set of operations required to ensure that measuring equipment conforms to the requirement for its intended use. - Brings in the concept of “Calibration” STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/68
  • 69. Calibration • Comparison of an unknown measurement device against an equal or better device (TUR/TAR) • Unknown device is called – Equipment under Calibration (EUC) – Unit under Calibration (UUC) – Device under Calibration (DUC) – Device being Calibrated (DBC) • Equal or better device is called “Standard” STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/69
  • 70. Calibration Process • Identification of need/Customer’s requirements • Availability of specifications for DUC • Serviceability status of DUC • Laboratory Capability • Availability of validated method/competence • Availability of Instruments /Standards • Technical comparison of DUC with Instrument / Standards • Repair / adjustment / recalibrate, if required • Verification of Confirmation • Uncertainty computation • Preparation of Calibration Report/Certificate • Status Identification STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/70
  • 71. CALIBRATION JOB FLOW STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/71
  • 72. Calibration - Job Flow Chart Ext Int Check points Customer Service Y N Y Job Job card to SCL Any N accepted Non-confor- CERT. Cleared mance Calibrate TP&C Int. DUC Compile result error/uncert. Out of Spec CERT. Issued Fix label with Item R&M DUC with cal report DUC after R&R EXT INT TP & C CAL QA STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/72
  • 73. Calibration v/s Business Processes Control of outside supplies and services Evaluati In-house Report and record Delivery / Receiving Calibrations processing Shipping on In-house repair Customers STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/73
  • 74. Role, Responsibilities & Authority  Job descriptions • HOL,Technical & Quality Managers • Scientists and Technicians • Attendants • Those who verify, test and check • Sampling officers • Those who act in Emergency  Performance Criteria  Performance appraisal STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/74
  • 75. Laboratory Manager/Chief Executive • Provide direction and coordinate the departmental activi-ties in technical, commercial and service establishment • Reviews and analyzes reports, records, and other information for planning laboratory’s growth, such as new commitments, status of work in progress and problem encountered. • Assigns or delegates respon-sibility for, specified work or functional activities and disseminates policy to down the line. • Gives work directions, resolves problem and set deadlines to ensure timely completion of work. • Coordinates activities of various departments with related activities of other departments to ensure efficiency and economy. • Monitors and analyzes costs and prepares budget. • Prepares reports and records on department activities for management. • Evaluate current procedures and practices for accomplishing department objectives to develop and implement improved procedures and practices. STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/75
  • 76. The Role of Quality Manager • Assurance of Quality of Calibration Results • Ensuring implementation of Management system • Document Control activity • Planning, conducting, compiling & analyzing Internal audit programme • Organizing, conducting Management Review & maintaining records • Monitoring corrective & preventive actions on non- conforming cal services ensuring effectiveness • Handling, investigating, analyzing & resolution of Customer complaints • Liaison with internal/external customers, accreditation body w.r.t quality matters affecting calibration. • Reports to Management STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/76
  • 77. The Role of Technical Manager • selection, approval and commissioning of test methods • Management and training of staff • Evaluation and sign-off of staff competency • Management of laboratory equipment and facilities • Design and implementation of the QC programme • Stopping work when QC criteria are exceeded • Attention of matters relating to impartiality and confidentiality • Responsibility for the technical validity of results • Liaison with the QM in relation to Quality matters affecting testing or calibration. • Reports to TM STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/77
  • 78. The Role of METROLOGIST • Develops and evaluates calibration system that measure characteristics of objects, substances, or phenomena, such as length, mass, time, temperature, electric current, luminous intensity, and derived units of physical or chemical measure • Identifies magnitude of error sources contributing to uncertainty of results to determine reliability of measurement process in quantitative term • Redesigns or adjusts measurement capability to minimize errors • Develops cal-ibration methods and techniques based on principles of measurement science, technical analysis of measurement problem, and accuracy and precision requirements • Directs engineering, quality and laboratory personnel in design, manufacture, evaluation, and calibration of measurement standards, instru-ments, and test system to ensure selection of approved instrumentation • Advises others on methods of resolving measurement problem and exchanges information with other metrology personnel through participation in govern-ment and industrial standardization committees and professional societies STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/78
  • 79. The Role of Laboratory In-Charge • Review/Acceptance/follow-up of jobs received from customer service to meet committed PDC for the same. • Preparation of calibration schedule if required and Allocation of work • Maintenance of lab. environment conditions ensuring good house keeping • Periodic monitoring of observations/ measurements/ compiled data during execution of the job & responsible for the technical validity of results • Preparations of test/cal procedures & implementation after due approvals • Qualifying label for the T&M equipment calibrated STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/79
  • 80. The Role of Laboratory In-Charge • Maintenance of equipment in the qualifying condition along with respective manuals/ technical data & history log books ensuring traceability • Maintaining registers used in the laboratory of the laboratory area • Co-ordination with lab. staff, customer service & other support services • Technical guidance to the laboratory personnel in calibration & Implementation of the QC programme • Attention of matters relating to impartiality and confidentiality • Liaison with the Quality Manager in relation to Quality matters affecting calibration. STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/80
  • 81. Roles of Cal Lab. personnel • Responsible for the calibration activity • Carrying out Calibration as per standard/validated cal methods • Preparation of Cal reports, affixing cal label on equipment calibrated. • Building technical competence in the area of service being provided • Any cal facility giving intermediate operation/ambiguous results/mal- functioning should be immediately reported to LIC/TM • Maintain the laboratory facility with appropriate use • Maintaining calibration service related records • Maintaining records reg. environmental controls • Maintaining Calibration/Maintenance manuals of cal facilities & traceability records • Reporting to LIC/TM • Any other as assigned from time to time STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/81
  • 82. “Approved Signatory” Personnel Technical Competence including:  Relevant qualifications and/or experience.  Close involvement in the day-to-day operations.  Familiarity with test procedures including scientific basis and technical limitations.  Ability to make critical evaluation of test results.  Knowledge of Quality Management system.  Knowledge of and commitment to AB and their criteria and ISO/IEC 17025.  Sufficient time in the laboratory to become fully familiar with the operating systems of the laboratory. STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/82
  • 83. SCHEDULING - CALIBRATION INTERVAL - CONFIRMATION INTERVAL - RECALL NOTIFICATIONS STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/83
  • 84. Calibration Interval / Reports Initial Calibration: Calibration adj. Final calibration: Procedure adopted Comparison To ensure the To confirm the w.r.t tracking of the successfulness common reference Measurand to that Of the operation Either to know of reference The deviation/value (cal adj.procedure (measurand & Implementation) Measurement (Report2) Uncertainty (Report 1) Report1-Report2 ? STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/84
  • 85. Within Calibration Interval…. Effect of drift on uncertainty of measurement Uncertainty of measurement Time STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/85
  • 86. Calibration Interval Management Effect of right time cal. Adj. on uncertainty of measurement adjustment carried out Uncertainty of measurement STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/86
  • 87. Scheduling Calibration Interval • Client agreed reported as ‘Client Agreed Calibration Interval - '. • Client specified reported as 'Client Specified Calibration Interval - '. • Manufacturer's recommended reported as 'Manufacturer's Recommended Calibration Interval - '. STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/87
  • 88. INITIAL CHOICE OF CI • MANUFACTURER’S RECOMMENDATION • EXTENT & SEVERITY OF USE • MAX. ALLOWED DEVIATION • UNCERTAINTY OF MEASUREMENT REQUIRED • INFLUENCE OF ENVIRONMENT • ADJUSTMENT/CHANGE REQUIRED • PUBLISHED DATA ON SIMILAR DEVICES • BY EXPERIENCE • HISTORY OF EQUIPMENT • QUALITY OF CALIBRATION & COST FACTOR T&ME or M&ME ARE TO BE CALIBRATED WHENEVER THEY ARE REPARIED. STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/88
  • 89. OPTIMUM INTERVAL A BALANCE BETWEEN THE COST TO CALIBRATE & THE COST NOT TO CALIBRATE SHOULD BE EVALUATED. AT PRESENT THERE IS NO UNIVERSALLY APPLICABLE SINGLE BEST PRACTICE FOR ESTABLISHING AND ADJUSTING CAL INTERVAL. STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/89
  • 90. RECALL NOTIFICATION • INTERNAL EQUIPMENT • EXTERNAL EQUIPMENT STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/90
  • 91. CHAPTER - 5 STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/91
  • 92. Manual Calibration The traditional method of obtaining parameters, requiring extensive training, time, and labor • Availability of Calibration facilities / System • Qualified & trained personnel • Availability of suitable environment • Traceability of standards • Device under calibration • Calibration Method (Validated) • Measurement records • Uncertainty computations • Calibration Reporting STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/92
  • 93. Automated Calibration 1. Computer Systems or Automated Equipment 2. Calibration software / Automated SW Program are useful to process • Large quantity of TME • Lower skilled personnel • Large volume of data used for capture, processing , recording, reporting, storage or retrieval of calibration data. 3. Interface / Compatible DUC STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/93
  • 94. Examples: Automated Calibration - to the extent supported by HW 1. DMM Calibration batch-wise 2. Batch temp. probe calibration 3. Batch Energy Meter Calibration 4. Calibration of ref. pressure transducers 5. Calibration of RF Power sensors 6. AC/DC Transfer standards Calibration 7. Etc.. STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/94
  • 95. Software Programme • Design Application Software • Develop/Construct Application Software • Test Application Software • Procure Application Software • Operate Application Software • Maintain Application Software STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/95
  • 96. Major Elements of Software Programme • Software development • Software procurement • Software Configuration Management • Software operations • Documents • Record keeping STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/96
  • 97. Software Development • Quality planning based on requirements • Multi-phased Software Life Cycle Process – Requirements, design, construction, testing, installation, operation, maintenance & retirement • Software Review, Verification & Validation STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/97
  • 98. Software Validation • Comparison of manual and automated methods • Inspection of procedure. • Inter-laboratory comparison STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/98
  • 99. Software Procurement • Requirements & Activities • Applicable Guidelines • Dedicated/designed/custom made STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/99
  • 100. Software Configuration Management • Identification • Change Control • Status Accounting STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/100
  • 101. Software Operations • Access Control • Backup & Recovery • Problem Reporting & Corrective Action STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/101
  • 102. Documentation/Records • Software Design & Development Procedure • Plan for Software QA • Software Requirements Documentation • Software Design & Implementation Documentation • Software verification & validation Documentation • User Documentation • Standards / Specifications applicable STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/102
  • 103. Automated Calibration SAMPLE OF AUTOMATED AND MANUAL SYSTEM EQUIPMENT COSTS (Fluke analysis) Equipment Automated Manual SPRT $ 3,000 $ 3,000 Recording thermometer $ 7,000 Nano volt meter / Mohm $ 3,500 Ice point reference for T/Cs $ 600 Dual Dry-Block –35 ºC to 670 ºC $ 8,900 $ 8,900 Computer $ 1,500 $ 1,500 Automated software $ 1,300 Report generation software $ 500 Total $ 21,700 $ 18,000 Automated batch calibration systems for industrial sensors may have batch/volume efficiency advantages, in time savings and increased productivity for calibration laboratories that find themselves with increasing temperature probe workloads. STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/103
  • 104. Internet Enabled Calibration Recommendations as per NPL, UK Report CMSC 28/03 TECHNIQUES SECURITY ISSUES • userid/password protection • Authentication • Access Control • Algorithms for encryption of sensitive measurement data • Confidentiality • Integrity • Message authentication • Non-repudiation code for certificate, • Availability Measurement data • Audit • MySQL vulnerabilities • Privacy • Certification • Backups and fire safes • General Protection STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/104
  • 105. STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/105
  • 106. CHAPTER - 6 STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/106
  • 107. Calibration Results - Based on following technical records • Original Observations • Calculations/Computations and data transfers • Calibration Report Format STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/107
  • 108. Reporting Calibration Results Report and Certificate format • Appropriate to each type of test/calibration • For easy understanding / to avoid misuse STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/108
  • 109. Calibration Reports / Certificates • Calibration Report A calibration report usually provides a set of results that show the performance of the instrument being calibrated in comparison to a reference instrument or standard. Measurements are recorded for each function, range and points tested, and the uncertainty of each measurement is reported. • Calibration Certificate A calibration certificate usually states the compliance of an instrument with the manufacturer's specification. Measurements made by the instrument being calibrated are compared to the manufacturer's specification. If the instrument is within specification, a statement that the instrument meets the manufacturer's specification at the points tested is made in the conformance certificate. Any points that are outside specification, or not able to be determined, are reported as exceptions. Assessment of meeting the manufacturer's specification takes into account the measurement uncertainty. A statement of conformance to the manufacturer's specification may also be given in calibration reports. STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/109
  • 110. Calibration Reports / Certificates • Accurate, clear, unambiguous & objective • No recommendation for Calibration interval, unless asked for • Confidentiality in electronic transmission • Specified & approved format • Specific statement for amendment • Must include information requested by customer Sample report/certificate formats refer NCSL, RP-11 STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/110
  • 111. Contents of A Calibration Report a) A title -Test report/Calibration certificate b) Name and Address of laboratory c) Unique identification of the report/certificate d) Name and address of customer e) Identification of method used f) Unambiguous identification of item g) Date of receipt / performance of Test / Calibration h) Reference to sampling plan / procedure STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/111
  • 112. Contents of Calibration Report contd.. I) Test/Calibration results and units of measurement j) Name, function, Signature of person authorizing the report/certificate k) Statement-result related to the item Tested/Calibrated l) Calibration Results with Units m) Name(s), function(s) and signature(s) or equivalent identification of person(s) authorizing the report or certificate; n) Statement to the effect that the results relate only to the items tested or calibrated, where relevant; o) Subcontracted results clearly identified. STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/112
  • 113. Calibration Reporting Calibration Result Interpretation where necessary ensure calibration certificates include the following: • Calibration conditions • Uncertainty of measurement • Measurement traceability statement • Statement of compliance with an identified metrological specification or clause • Compliance data records • Repairs/adjustments if any • Calibration Interval • Estimated Uncertainty STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/113
  • 114. STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/114
  • 115. Amendment/Supplement to Report • Identification or reference to the original report identity. • when a complete new test report or calibration certificate is required, it must be uniquely identified and include a reference to the original it replaces STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/115
  • 116. Acceptance of Calibration Report • Do you accept them or not? • How do you know if they are okay to accept? • What are the related issues? What are the steps to determine what is okay and what is not? STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/116
  • 117. Acceptance of Calibration Report • evaluate the "scope of accreditation." The laboratory scope, which is published with the accreditation or recognition certificate, defines the parameters and associated uncertainties that have been assessed by the accreditation bodies. • An accreditation body logo • evaluate standards to ensure suitability & Forward traceability report 5.6.2 Show unbroken chain of traceability from an instrument to national standards. • Rights to revoke acceptance of certificates for cause (bad measurement results) & seek clarifications • Calibration certificate contents as per 5.10 of ISO/IEC 17025. • Due for calibration 5.6.1 – determine instrument exceeded cal interval. • Reverse Traceability report 4.9.1- Identify all the instruments that are affected if corrective action is neccessiated STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/117
  • 118. HANDLING DEVIATIONS - FROM CALIBRATION QUALITY LEVEL 1. Interval adjustment 2. Procedure Changes 3. Specification Revision 4. Derating or limited Calibration 5. Removal of equipment from services 6. Instrument Repair & Recalibration 7. Upgradation of equipment 8. Replacement STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/118
  • 119. CALIBRATION LABELLING Valid Calibration: when all instrument parameters have been calibrated & meets the manufacturer’s specifications. If not affix following type of labels. LABELING LIMITAION OF USE (Refer NCSL RP- 6) Labeling location, content, color & size depends on Calibration agency STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/119
  • 120. External Calibration When Cal Service is subcontracted, ensure: Approval from customer, preferably in writing, when the lab intends to subcontract. Ensure and demonstrate subcontractors competence through evaluation & review Laboratory’s responsibility for subcontractors work. Maintain a register of all subcontracting work STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/120
  • 121. External Calibration When Cal Services is availed, ensure:  Credibility of Cal agency through accreditation & scope of accreditation covering your requirements  Gain confidence in their services through visits/audits on: – type of calibration service – applicable documents, technical reference, or standard – equipment specification – service performance requirements – physical, environmental, and data-format requirements – safety and biohazard requirements – confidentiality and security requirements – certificate of calibration requirements STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/121
  • 122. CHAPTER - 7 STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/122
  • 123. REQUIREMENTS FOR DOCUMENTING RECORDS 1. ISO/IEC 17025:2005, clause 4.12, paragraph 4.12.2.1, states: "The laboratory shall retain records of original observations, derived data. . . to estab-lish an audit trail . . . for a defined period. The records for each calibration shall contain sufficient information to . . . enable the test or calibration to be repeated." Paragraph 4.12.2.2 states: "Observations, data and calculations shall be recorded at the time they are made. . . ." 2. ISO 9001:2000, clause 7.6.on Control of Monitoring & Measuring equipment. 3. ISO 14001:2004, Clause 4.5.2 on equipment used for environmental maesurements. 4. ISO 10012:2003, clause 6.2.3 states: "Records containing information required for the operation of the measurement management system shall be maintained. Documented procedures shall ensure the identification, storage, protec-tion, retrieval, retention time and disposition of records." 5. NCSLI RP-6, clause 5.6, Records, states: "Records should be maintained for all test and measuring equipment that is included in the calibration system STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/123
  • 124. Why Calibration Records are necessary? • Understanding the relationship of historical data and stability of a standard • As Found, and when needed, As Left readings giving reliability of equipment • Log book of Instruments logging Calibration history on maintainability • Trend of measurements from periodic Calibration Reports • Authentication of Calibration Data/Results STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/124
  • 125. Calibration Records • Service Request Forms • Work Orders/Job Orders • Planning / Job allocation Records • Observations/data computations/Measurement sheets, Uncertainty Budget records • Preliminary performance/verification records • Calibration records of equipment used for cal. Control Charts on trends of calibration standards • Calibration Report / Certificates • Replicate/ Retesting records • Training records • Proficiency/ Inter-laboratory Comparison records • Nonconforming calibration, Customer complaints, CA/PA records • Calibration System Audit records • Repair/Adjustment/Maintenance Records STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/125
  • 126. Management of Calibration Records • Legibility & Indelibility of records • Identification of records • Storage of records • Maintenance/Preservation of records • Retention & Disposition of records • Amendment of records with authentication STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/126
  • 127. STQC Version 1.0 CERTIFIED CALIBRATION PROFFESSIONAL MIII/CH1/127