PHONOCARDIOGRAPHY (PCG)
Ms.P.NISHANTHI
Assistant Professor/ Department of ECE
KIT-Kalaignarkarunanidhi Institute of Technology
PCG
 Graphic record heart sound - Phonogram.
 Recording the sounds connected with the pumping action of heart.
 Sound from heart – phonocardiogram
 Instrument to measure this – phonocardiograph
 Basic function – to pick up the different heart sound,filter the required and
display.
 Two categories –
 heart sound – transient characteristics with short duration.(closing and
opening of valves)
 murmurs – noisy characteristics with long duration.(turbulent blood flow in
heart)
MEDICAL ELECTRONICS - MV/AP/ECE 2
Origin of sound
 Valve closure
 Movement of heart wall
 Valve opening
 Extra cardiac sounds
1. Frequency – 10 to 1000 Hz.
LOW RANGE – 10 – 60 Hz(3rd and 4th)
MEDIUM RANGE – 60 – 150 Hz(1st and 2nd)
HIGH RANGE – 150 – 1000 Hz
2.Amplitude
3.Quality
MEDICAL ELECTRONICS - MV/AP/ECE 3
 1st sound:
Closure of mitral and tricuspid valves.
Freq – 30 to 100 Hz and duration 50 to 100 ms
 2nd sound:
Closure of aortic and pulmonary valves(slight back flow of blood).
Freq – 30 to 100 Hz and duration 25 to 50 ms .
 3rd and 4th low intensity and inaudible.(heard among children).
MEDICAL ELECTRONICS - MV/AP/ECE 4
 3rd sound:
Blood rapid movement into relaxed ventricular chambers
Freq – 10 to 100 Hz and duration 0.04 to 0.08 s.
 4th sound:
atrial contraction.
Freq – 10 to 50 Hz and duration 0.03 to 0.06 s.
MEDICAL ELECTRONICS - MV/AP/ECE 5
 Microphones for PCG:
 Crystal microphone – contains wafer of piezo-electric material, which
generates potentials when subjected to mechanical stresses due to heart
sound. Smaller in size, high sensitivity.
 Dynamic microphone – consists of a moving coil having a fixed magnetic core
inside it. The coil moves with the heart sound and produce a voltage because
of interaction with the magnetic flux.
 New acoustic sensors – polymer based adherent differential-output sensors.
MEDICAL ELECTRONICS - MV/AP/ECE 6
Writing methods
 Light beam galvanometer – expensive and require more power from amplifiers
when used for high freq.
 Direct writing recorders – upper freq 150 Hz.(used for only low freq).
 Technique – envelope detection.
 Fans, Air conditioners and other noise producing gadgets will result in
artefacts on the recordings.
MEDICAL ELECTRONICS - MV/AP/ECE 7
8
Heart Sounds
 S1 – onset of the ventricular contraction
 S2 – closure of the semilunar valves
 S3 – ventricular gallop
 S4 – atrial gallop
 Other – opening snap, ejection sound
 Murmurs
MEDICAL ELECTRONICS - MV/AP/ECE
RECORDING SET - UP
MEDICAL ELECTRONICS - MV/AP/ECE 9
CONDENSER
MICROPHONE
ECG AMP
ECG
ELECTRODE
FILTER
PHONO
AMPLIFIER
FM TAPE
RECORDER
MONITOR
SCOPE
 Amplifiers used to process biopotentials are called
biopotential amplifier.
 Biopotential Signals (e.g., ECG, EMG, EEG, EOG, …
etc.)
 The basic function of biopotential amplifier is to
increase the amplitude of a weak electric signal of
biological origin.
 Biopotential amplifiers typically process voltages, but
in some cases they also process currents.
 The frequency response of typical bioelectric
amplifiers may be from dc (or near dc, i.e., 0.05 Hz)
up to 100 kHz.
Biopotential Amplifier
 Some biopotential amplifiers are ac-coupled, while some are dc-coupled.
 The dc-coupling is required where input signals are clearly dc or changes
very slowly.
 At frequencies as low as 0.05Hz, the ac-coupling should be used instead of
dc-coupling.
 This is to overcome the electrode offset potential.
 Also, the skin-electrode interface generates dc offsets.
 The gain of biopotential amplifiers can be low, medium or high (x10, x100,
x1000, x10000).
Cont…
 Gain factors x1 and x10.
 The unity-gain amplifier is mainly for isolation,
buffering and possibly impedance
transformation between signal source and
readout device.
 Used for measurement of action potentials and
other relatively high-amplitude bioelectric
events.
Low Gain Biopotential Amplifier
 Gain factors x100 and x1000.
 Used for recording of ECG waveforms and
muscles potentials (EMG), etc.
Medium Gain Biopotential Amplifier
 Gain factors over x1000.
 Used in very sensitive measurement such as
recording of brain potentials (EEG).
High Gain Biopotential Amplifier
Body Surface Recording Electrodes
1. Metal Plate Electrodes
(historic)
2. Suction Electrodes
(historic interest)
3. Floating Electrodes
4. Flexible Electrodes
Electrode metal
Electrolyte
Think of the
construction of
electrosurgical
electrode
And, how does
electro-surgery
work?
Commonly Used Biopotential Electrodes
Metal plate electrodes
 Large surface: Ancient,
therefore still used, ECG
 Metal disk with stainless steel;
platinum or gold coated
 EMG, EEG
 smaller diameters
 motion artifacts
 Disposable foam-pad: Cheap!
(a) Metal-plate electrode used for application to limbs.
(b) Metal-disk electrode applied with surgical tape.
(c)Disposable foam-pad electrodes, often used with ECG
Commonly Used Biopotential
Electrodes
Suction electrodes
- No straps or adhesives
required
- precordial (chest) ECG
- can only be used for short
periods
Floating electrodes
- metal disk is recessed
- swimming in the electrolyte gel
- not in contact with the skin
- reduces motion artifact
Suction Electrode
Double-sided
Adhesive-tape
ring
Insulating
package
Metal disk
Electrolyte gel
in recess
(a) (b)
(c)
Snap coated with Ag-AgCl External snap
Plastic cup
Tack
Plastic disk
Foam pad
Capillary loops
Dead cellular material
Germinating layer
Gel-coated sponge
Commonly Used Biopotential
Electrodes
Floating Electrodes
Reusable
Disposable
(a) Carbon-filled silicone rubber electrode.
(b) Flexible thin-film neonatal electrode.
(c) Cross-sectional view of the thin-film
electrode in (b).
Commonly Used Biopotential
Electrodes
Flexible electrodes
- Body contours are often
irregular
- Regularly shaped rigid
electrodes
may not always work.
- Special case : infants
- Material :
- Polymer or nylon with silver
- Carbon filled silicon rubber
(Mylar film)
Internal Electrodes
Needle and wire electrodes for
percutaneous measurement of
biopotentials
(a) Insulated needle electrode.
(b) Coaxial needle electrode.
(c) Bipolar coaxial electrode.
(d) Fine-wire electrode connected
to hypodermic needle, before
being inserted.
(e) Cross-sectional view of skin
and muscle, showing coiled
fine-wire electrode in place.
The latest: BION – implanted electrode for muscle recording/stimulation
Alfred E. Mann Foundation
Fetal ECG Electrodes
Electrodes for detecting fetal electrocardiogram during labor, by means
of intracutaneous needles (a) Suction electrode. (b) Cross-sectional view of
suction electrode in place, showing penetration of probe through epidermis.
(c) Helical electrode, which is attached to fetal skin by corkscrew type action.
MEDICAL APPLICATION
 Rheumatic valvular lesions
Allergic disease in which the heart valves are likely to be damaged. This creates abnormal heart
sound.
 The murmurs of aortic steonosis
The blood is ejected from the left ventricles through a small opening of aortic valve. High
pressure 350 mm of Hg and causes turbulent flow of blood.
 The murmurs of aortic regurgitation
During diastole, blood flow backward from aorta to left ventricles,causing a blowing murmur.
 The murmurs of mitral regurgitation
During systole, blood flow backward through mitral valve.
 The murmurs of mitral steonosis
It produce murmurs when,Blood passes with difficulty from LA into the LV due to pressure
difference.
MEDICAL ELECTRONICS - MV/AP/ECE 22
SPECIAL APPLICATION OF PCG
 Fetal PCG
 Esophageal PCG
 Tracheal PCG - Cannula
MEDICAL ELECTRONICS - MV/AP/ECE 23
Thank you

Más contenido relacionado

La actualidad más candente

La actualidad más candente (20)

Electrical safety
Electrical safetyElectrical safety
Electrical safety
 
Isolation amplifier
Isolation amplifierIsolation amplifier
Isolation amplifier
 
Application of different medical electrodes
Application of different medical electrodesApplication of different medical electrodes
Application of different medical electrodes
 
EEG (ELECTROENCEPALOGRAM)
EEG (ELECTROENCEPALOGRAM)EEG (ELECTROENCEPALOGRAM)
EEG (ELECTROENCEPALOGRAM)
 
Instrumentation amplifier in ECG
Instrumentation amplifier  in ECG Instrumentation amplifier  in ECG
Instrumentation amplifier in ECG
 
EEG
EEGEEG
EEG
 
Biomedical signal processing
Biomedical signal processingBiomedical signal processing
Biomedical signal processing
 
Presentation paper
Presentation paperPresentation paper
Presentation paper
 
Biopotentials
BiopotentialsBiopotentials
Biopotentials
 
ECG
ECGECG
ECG
 
Electrical Hazards and Patient Safety in Biomedical Equipment
Electrical Hazards and Patient Safety in Biomedical EquipmentElectrical Hazards and Patient Safety in Biomedical Equipment
Electrical Hazards and Patient Safety in Biomedical Equipment
 
Rheological Properties of blood
Rheological Properties of bloodRheological Properties of blood
Rheological Properties of blood
 
Biopotential electrodes
Biopotential electrodesBiopotential electrodes
Biopotential electrodes
 
10 and 20 electrode placement
10 and 20 electrode placement10 and 20 electrode placement
10 and 20 electrode placement
 
Biotelemetry
BiotelemetryBiotelemetry
Biotelemetry
 
Patient monitoring system and biotelemetry
Patient monitoring system and biotelemetryPatient monitoring system and biotelemetry
Patient monitoring system and biotelemetry
 
medical electronics 5 module ppt
medical electronics 5 module pptmedical electronics 5 module ppt
medical electronics 5 module ppt
 
Biomedical Instrumentation
Biomedical InstrumentationBiomedical Instrumentation
Biomedical Instrumentation
 
ECG Noise cancelling
ECG Noise cancelling ECG Noise cancelling
ECG Noise cancelling
 
Biomedical Instrumentation
Biomedical InstrumentationBiomedical Instrumentation
Biomedical Instrumentation
 

Similar a Phonocardiography(PCG)

BIO MEDICAL RECORDERS
BIO MEDICAL RECORDERS BIO MEDICAL RECORDERS
BIO MEDICAL RECORDERS gowri R
 
Lecturing 1
Lecturing 1Lecturing 1
Lecturing 1gowri R
 
various type of artifacts in ECG signal & how it's remove
various type of artifacts in ECG signal & how it's removevarious type of artifacts in ECG signal & how it's remove
various type of artifacts in ECG signal & how it's removeManish Kumar
 
EEG Lecture- Introduction.pptx
EEG Lecture- Introduction.pptxEEG Lecture- Introduction.pptx
EEG Lecture- Introduction.pptxmunnam37
 
Electro surgery in laparoscopy
Electro surgery  in laparoscopyElectro surgery  in laparoscopy
Electro surgery in laparoscopyMed Elsayed
 
Biosignals andthermometry fin
Biosignals andthermometry finBiosignals andthermometry fin
Biosignals andthermometry finMUBOSScz
 
Electrophysiological tests
Electrophysiological testsElectrophysiological tests
Electrophysiological testsReshma Peter
 
Bionic Arm Using EMG Processing
Bionic Arm Using EMG ProcessingBionic Arm Using EMG Processing
Bionic Arm Using EMG Processingbiomedicz
 
A versatile low cost Electronic Stethoscope design
A versatile low cost Electronic Stethoscope designA versatile low cost Electronic Stethoscope design
A versatile low cost Electronic Stethoscope designiosrjce
 

Similar a Phonocardiography(PCG) (20)

BIO MEDICAL RECORDERS
BIO MEDICAL RECORDERS BIO MEDICAL RECORDERS
BIO MEDICAL RECORDERS
 
Lecturing 1
Lecturing 1Lecturing 1
Lecturing 1
 
Factors
FactorsFactors
Factors
 
various type of artifacts in ECG signal & how it's remove
various type of artifacts in ECG signal & how it's removevarious type of artifacts in ECG signal & how it's remove
various type of artifacts in ECG signal & how it's remove
 
ECG (1).pptx
ECG (1).pptxECG (1).pptx
ECG (1).pptx
 
ECG (1).pptx
ECG (1).pptxECG (1).pptx
ECG (1).pptx
 
IEEE_2011_Conference
IEEE_2011_ConferenceIEEE_2011_Conference
IEEE_2011_Conference
 
EEG Lecture- Introduction.pptx
EEG Lecture- Introduction.pptxEEG Lecture- Introduction.pptx
EEG Lecture- Introduction.pptx
 
Electro surgery in laparoscopy
Electro surgery  in laparoscopyElectro surgery  in laparoscopy
Electro surgery in laparoscopy
 
ecochG.pptx
ecochG.pptxecochG.pptx
ecochG.pptx
 
Biosignals andthermometry fin
Biosignals andthermometry finBiosignals andthermometry fin
Biosignals andthermometry fin
 
Elecrosurgery in hysteroscopy
Elecrosurgery in hysteroscopyElecrosurgery in hysteroscopy
Elecrosurgery in hysteroscopy
 
Electrophysiological tests
Electrophysiological testsElectrophysiological tests
Electrophysiological tests
 
Cochlear Implant
Cochlear ImplantCochlear Implant
Cochlear Implant
 
Presentation 21 4-11electrotherapy
Presentation 21 4-11electrotherapyPresentation 21 4-11electrotherapy
Presentation 21 4-11electrotherapy
 
Bionic Arm Using EMG Processing
Bionic Arm Using EMG ProcessingBionic Arm Using EMG Processing
Bionic Arm Using EMG Processing
 
report
reportreport
report
 
PCG.pdf
PCG.pdfPCG.pdf
PCG.pdf
 
J010614854
J010614854J010614854
J010614854
 
A versatile low cost Electronic Stethoscope design
A versatile low cost Electronic Stethoscope designA versatile low cost Electronic Stethoscope design
A versatile low cost Electronic Stethoscope design
 

Último

complete construction, environmental and economics information of biomass com...
complete construction, environmental and economics information of biomass com...complete construction, environmental and economics information of biomass com...
complete construction, environmental and economics information of biomass com...asadnawaz62
 
UNIT III ANALOG ELECTRONICS (BASIC ELECTRONICS)
UNIT III ANALOG ELECTRONICS (BASIC ELECTRONICS)UNIT III ANALOG ELECTRONICS (BASIC ELECTRONICS)
UNIT III ANALOG ELECTRONICS (BASIC ELECTRONICS)Dr SOUNDIRARAJ N
 
Transport layer issues and challenges - Guide
Transport layer issues and challenges - GuideTransport layer issues and challenges - Guide
Transport layer issues and challenges - GuideGOPINATHS437943
 
Class 1 | NFPA 72 | Overview Fire Alarm System
Class 1 | NFPA 72 | Overview Fire Alarm SystemClass 1 | NFPA 72 | Overview Fire Alarm System
Class 1 | NFPA 72 | Overview Fire Alarm Systemirfanmechengr
 
Past, Present and Future of Generative AI
Past, Present and Future of Generative AIPast, Present and Future of Generative AI
Past, Present and Future of Generative AIabhishek36461
 
Software and Systems Engineering Standards: Verification and Validation of Sy...
Software and Systems Engineering Standards: Verification and Validation of Sy...Software and Systems Engineering Standards: Verification and Validation of Sy...
Software and Systems Engineering Standards: Verification and Validation of Sy...VICTOR MAESTRE RAMIREZ
 
Crushers to screens in aggregate production
Crushers to screens in aggregate productionCrushers to screens in aggregate production
Crushers to screens in aggregate productionChinnuNinan
 
Comparative study of High-rise Building Using ETABS,SAP200 and SAFE., SAFE an...
Comparative study of High-rise Building Using ETABS,SAP200 and SAFE., SAFE an...Comparative study of High-rise Building Using ETABS,SAP200 and SAFE., SAFE an...
Comparative study of High-rise Building Using ETABS,SAP200 and SAFE., SAFE an...Erbil Polytechnic University
 
Risk Management in Engineering Construction Project
Risk Management in Engineering Construction ProjectRisk Management in Engineering Construction Project
Risk Management in Engineering Construction ProjectErbil Polytechnic University
 
Industrial Safety Unit-IV workplace health and safety.ppt
Industrial Safety Unit-IV workplace health and safety.pptIndustrial Safety Unit-IV workplace health and safety.ppt
Industrial Safety Unit-IV workplace health and safety.pptNarmatha D
 
Research Methodology for Engineering pdf
Research Methodology for Engineering pdfResearch Methodology for Engineering pdf
Research Methodology for Engineering pdfCaalaaAbdulkerim
 
TechTAC® CFD Report Summary: A Comparison of Two Types of Tubing Anchor Catchers
TechTAC® CFD Report Summary: A Comparison of Two Types of Tubing Anchor CatchersTechTAC® CFD Report Summary: A Comparison of Two Types of Tubing Anchor Catchers
TechTAC® CFD Report Summary: A Comparison of Two Types of Tubing Anchor Catcherssdickerson1
 
Arduino_CSE ece ppt for working and principal of arduino.ppt
Arduino_CSE ece ppt for working and principal of arduino.pptArduino_CSE ece ppt for working and principal of arduino.ppt
Arduino_CSE ece ppt for working and principal of arduino.pptSAURABHKUMAR892774
 
Input Output Management in Operating System
Input Output Management in Operating SystemInput Output Management in Operating System
Input Output Management in Operating SystemRashmi Bhat
 
Instrumentation, measurement and control of bio process parameters ( Temperat...
Instrumentation, measurement and control of bio process parameters ( Temperat...Instrumentation, measurement and control of bio process parameters ( Temperat...
Instrumentation, measurement and control of bio process parameters ( Temperat...121011101441
 
Earthing details of Electrical Substation
Earthing details of Electrical SubstationEarthing details of Electrical Substation
Earthing details of Electrical Substationstephanwindworld
 
"Exploring the Essential Functions and Design Considerations of Spillways in ...
"Exploring the Essential Functions and Design Considerations of Spillways in ..."Exploring the Essential Functions and Design Considerations of Spillways in ...
"Exploring the Essential Functions and Design Considerations of Spillways in ...Erbil Polytechnic University
 

Último (20)

complete construction, environmental and economics information of biomass com...
complete construction, environmental and economics information of biomass com...complete construction, environmental and economics information of biomass com...
complete construction, environmental and economics information of biomass com...
 
UNIT III ANALOG ELECTRONICS (BASIC ELECTRONICS)
UNIT III ANALOG ELECTRONICS (BASIC ELECTRONICS)UNIT III ANALOG ELECTRONICS (BASIC ELECTRONICS)
UNIT III ANALOG ELECTRONICS (BASIC ELECTRONICS)
 
Transport layer issues and challenges - Guide
Transport layer issues and challenges - GuideTransport layer issues and challenges - Guide
Transport layer issues and challenges - Guide
 
Class 1 | NFPA 72 | Overview Fire Alarm System
Class 1 | NFPA 72 | Overview Fire Alarm SystemClass 1 | NFPA 72 | Overview Fire Alarm System
Class 1 | NFPA 72 | Overview Fire Alarm System
 
Past, Present and Future of Generative AI
Past, Present and Future of Generative AIPast, Present and Future of Generative AI
Past, Present and Future of Generative AI
 
Software and Systems Engineering Standards: Verification and Validation of Sy...
Software and Systems Engineering Standards: Verification and Validation of Sy...Software and Systems Engineering Standards: Verification and Validation of Sy...
Software and Systems Engineering Standards: Verification and Validation of Sy...
 
Crushers to screens in aggregate production
Crushers to screens in aggregate productionCrushers to screens in aggregate production
Crushers to screens in aggregate production
 
Designing pile caps according to ACI 318-19.pptx
Designing pile caps according to ACI 318-19.pptxDesigning pile caps according to ACI 318-19.pptx
Designing pile caps according to ACI 318-19.pptx
 
Comparative study of High-rise Building Using ETABS,SAP200 and SAFE., SAFE an...
Comparative study of High-rise Building Using ETABS,SAP200 and SAFE., SAFE an...Comparative study of High-rise Building Using ETABS,SAP200 and SAFE., SAFE an...
Comparative study of High-rise Building Using ETABS,SAP200 and SAFE., SAFE an...
 
young call girls in Green Park🔝 9953056974 🔝 escort Service
young call girls in Green Park🔝 9953056974 🔝 escort Serviceyoung call girls in Green Park🔝 9953056974 🔝 escort Service
young call girls in Green Park🔝 9953056974 🔝 escort Service
 
Risk Management in Engineering Construction Project
Risk Management in Engineering Construction ProjectRisk Management in Engineering Construction Project
Risk Management in Engineering Construction Project
 
Industrial Safety Unit-IV workplace health and safety.ppt
Industrial Safety Unit-IV workplace health and safety.pptIndustrial Safety Unit-IV workplace health and safety.ppt
Industrial Safety Unit-IV workplace health and safety.ppt
 
Research Methodology for Engineering pdf
Research Methodology for Engineering pdfResearch Methodology for Engineering pdf
Research Methodology for Engineering pdf
 
TechTAC® CFD Report Summary: A Comparison of Two Types of Tubing Anchor Catchers
TechTAC® CFD Report Summary: A Comparison of Two Types of Tubing Anchor CatchersTechTAC® CFD Report Summary: A Comparison of Two Types of Tubing Anchor Catchers
TechTAC® CFD Report Summary: A Comparison of Two Types of Tubing Anchor Catchers
 
POWER SYSTEMS-1 Complete notes examples
POWER SYSTEMS-1 Complete notes  examplesPOWER SYSTEMS-1 Complete notes  examples
POWER SYSTEMS-1 Complete notes examples
 
Arduino_CSE ece ppt for working and principal of arduino.ppt
Arduino_CSE ece ppt for working and principal of arduino.pptArduino_CSE ece ppt for working and principal of arduino.ppt
Arduino_CSE ece ppt for working and principal of arduino.ppt
 
Input Output Management in Operating System
Input Output Management in Operating SystemInput Output Management in Operating System
Input Output Management in Operating System
 
Instrumentation, measurement and control of bio process parameters ( Temperat...
Instrumentation, measurement and control of bio process parameters ( Temperat...Instrumentation, measurement and control of bio process parameters ( Temperat...
Instrumentation, measurement and control of bio process parameters ( Temperat...
 
Earthing details of Electrical Substation
Earthing details of Electrical SubstationEarthing details of Electrical Substation
Earthing details of Electrical Substation
 
"Exploring the Essential Functions and Design Considerations of Spillways in ...
"Exploring the Essential Functions and Design Considerations of Spillways in ..."Exploring the Essential Functions and Design Considerations of Spillways in ...
"Exploring the Essential Functions and Design Considerations of Spillways in ...
 

Phonocardiography(PCG)

  • 1. PHONOCARDIOGRAPHY (PCG) Ms.P.NISHANTHI Assistant Professor/ Department of ECE KIT-Kalaignarkarunanidhi Institute of Technology
  • 2. PCG  Graphic record heart sound - Phonogram.  Recording the sounds connected with the pumping action of heart.  Sound from heart – phonocardiogram  Instrument to measure this – phonocardiograph  Basic function – to pick up the different heart sound,filter the required and display.  Two categories –  heart sound – transient characteristics with short duration.(closing and opening of valves)  murmurs – noisy characteristics with long duration.(turbulent blood flow in heart) MEDICAL ELECTRONICS - MV/AP/ECE 2
  • 3. Origin of sound  Valve closure  Movement of heart wall  Valve opening  Extra cardiac sounds 1. Frequency – 10 to 1000 Hz. LOW RANGE – 10 – 60 Hz(3rd and 4th) MEDIUM RANGE – 60 – 150 Hz(1st and 2nd) HIGH RANGE – 150 – 1000 Hz 2.Amplitude 3.Quality MEDICAL ELECTRONICS - MV/AP/ECE 3
  • 4.  1st sound: Closure of mitral and tricuspid valves. Freq – 30 to 100 Hz and duration 50 to 100 ms  2nd sound: Closure of aortic and pulmonary valves(slight back flow of blood). Freq – 30 to 100 Hz and duration 25 to 50 ms .  3rd and 4th low intensity and inaudible.(heard among children). MEDICAL ELECTRONICS - MV/AP/ECE 4
  • 5.  3rd sound: Blood rapid movement into relaxed ventricular chambers Freq – 10 to 100 Hz and duration 0.04 to 0.08 s.  4th sound: atrial contraction. Freq – 10 to 50 Hz and duration 0.03 to 0.06 s. MEDICAL ELECTRONICS - MV/AP/ECE 5
  • 6.  Microphones for PCG:  Crystal microphone – contains wafer of piezo-electric material, which generates potentials when subjected to mechanical stresses due to heart sound. Smaller in size, high sensitivity.  Dynamic microphone – consists of a moving coil having a fixed magnetic core inside it. The coil moves with the heart sound and produce a voltage because of interaction with the magnetic flux.  New acoustic sensors – polymer based adherent differential-output sensors. MEDICAL ELECTRONICS - MV/AP/ECE 6
  • 7. Writing methods  Light beam galvanometer – expensive and require more power from amplifiers when used for high freq.  Direct writing recorders – upper freq 150 Hz.(used for only low freq).  Technique – envelope detection.  Fans, Air conditioners and other noise producing gadgets will result in artefacts on the recordings. MEDICAL ELECTRONICS - MV/AP/ECE 7
  • 8. 8 Heart Sounds  S1 – onset of the ventricular contraction  S2 – closure of the semilunar valves  S3 – ventricular gallop  S4 – atrial gallop  Other – opening snap, ejection sound  Murmurs MEDICAL ELECTRONICS - MV/AP/ECE
  • 9. RECORDING SET - UP MEDICAL ELECTRONICS - MV/AP/ECE 9 CONDENSER MICROPHONE ECG AMP ECG ELECTRODE FILTER PHONO AMPLIFIER FM TAPE RECORDER MONITOR SCOPE
  • 10.  Amplifiers used to process biopotentials are called biopotential amplifier.  Biopotential Signals (e.g., ECG, EMG, EEG, EOG, … etc.)  The basic function of biopotential amplifier is to increase the amplitude of a weak electric signal of biological origin.  Biopotential amplifiers typically process voltages, but in some cases they also process currents.  The frequency response of typical bioelectric amplifiers may be from dc (or near dc, i.e., 0.05 Hz) up to 100 kHz. Biopotential Amplifier
  • 11.  Some biopotential amplifiers are ac-coupled, while some are dc-coupled.  The dc-coupling is required where input signals are clearly dc or changes very slowly.  At frequencies as low as 0.05Hz, the ac-coupling should be used instead of dc-coupling.  This is to overcome the electrode offset potential.  Also, the skin-electrode interface generates dc offsets.  The gain of biopotential amplifiers can be low, medium or high (x10, x100, x1000, x10000). Cont…
  • 12.  Gain factors x1 and x10.  The unity-gain amplifier is mainly for isolation, buffering and possibly impedance transformation between signal source and readout device.  Used for measurement of action potentials and other relatively high-amplitude bioelectric events. Low Gain Biopotential Amplifier
  • 13.  Gain factors x100 and x1000.  Used for recording of ECG waveforms and muscles potentials (EMG), etc. Medium Gain Biopotential Amplifier
  • 14.  Gain factors over x1000.  Used in very sensitive measurement such as recording of brain potentials (EEG). High Gain Biopotential Amplifier
  • 15. Body Surface Recording Electrodes 1. Metal Plate Electrodes (historic) 2. Suction Electrodes (historic interest) 3. Floating Electrodes 4. Flexible Electrodes Electrode metal Electrolyte Think of the construction of electrosurgical electrode And, how does electro-surgery work?
  • 16. Commonly Used Biopotential Electrodes Metal plate electrodes  Large surface: Ancient, therefore still used, ECG  Metal disk with stainless steel; platinum or gold coated  EMG, EEG  smaller diameters  motion artifacts  Disposable foam-pad: Cheap! (a) Metal-plate electrode used for application to limbs. (b) Metal-disk electrode applied with surgical tape. (c)Disposable foam-pad electrodes, often used with ECG
  • 17. Commonly Used Biopotential Electrodes Suction electrodes - No straps or adhesives required - precordial (chest) ECG - can only be used for short periods Floating electrodes - metal disk is recessed - swimming in the electrolyte gel - not in contact with the skin - reduces motion artifact Suction Electrode
  • 18. Double-sided Adhesive-tape ring Insulating package Metal disk Electrolyte gel in recess (a) (b) (c) Snap coated with Ag-AgCl External snap Plastic cup Tack Plastic disk Foam pad Capillary loops Dead cellular material Germinating layer Gel-coated sponge Commonly Used Biopotential Electrodes Floating Electrodes Reusable Disposable
  • 19. (a) Carbon-filled silicone rubber electrode. (b) Flexible thin-film neonatal electrode. (c) Cross-sectional view of the thin-film electrode in (b). Commonly Used Biopotential Electrodes Flexible electrodes - Body contours are often irregular - Regularly shaped rigid electrodes may not always work. - Special case : infants - Material : - Polymer or nylon with silver - Carbon filled silicon rubber (Mylar film)
  • 20. Internal Electrodes Needle and wire electrodes for percutaneous measurement of biopotentials (a) Insulated needle electrode. (b) Coaxial needle electrode. (c) Bipolar coaxial electrode. (d) Fine-wire electrode connected to hypodermic needle, before being inserted. (e) Cross-sectional view of skin and muscle, showing coiled fine-wire electrode in place. The latest: BION – implanted electrode for muscle recording/stimulation Alfred E. Mann Foundation
  • 21. Fetal ECG Electrodes Electrodes for detecting fetal electrocardiogram during labor, by means of intracutaneous needles (a) Suction electrode. (b) Cross-sectional view of suction electrode in place, showing penetration of probe through epidermis. (c) Helical electrode, which is attached to fetal skin by corkscrew type action.
  • 22. MEDICAL APPLICATION  Rheumatic valvular lesions Allergic disease in which the heart valves are likely to be damaged. This creates abnormal heart sound.  The murmurs of aortic steonosis The blood is ejected from the left ventricles through a small opening of aortic valve. High pressure 350 mm of Hg and causes turbulent flow of blood.  The murmurs of aortic regurgitation During diastole, blood flow backward from aorta to left ventricles,causing a blowing murmur.  The murmurs of mitral regurgitation During systole, blood flow backward through mitral valve.  The murmurs of mitral steonosis It produce murmurs when,Blood passes with difficulty from LA into the LV due to pressure difference. MEDICAL ELECTRONICS - MV/AP/ECE 22
  • 23. SPECIAL APPLICATION OF PCG  Fetal PCG  Esophageal PCG  Tracheal PCG - Cannula MEDICAL ELECTRONICS - MV/AP/ECE 23