DOPPLER PHYSICS
PRINCIPLE OF DOPPLER ULTRASOUND
• BASIS – ‘ DOPPLER EFFECT’ • DISCOVERY –
Physicist Christian Johann Doppler in 1842 •
‘When a sound source and reflector are moving
towards each other, the sound waves are spaced
closer together and reach the receiver at a
higher frequency than that at which they were
originally emitted ’.
DOPPLER SHIFT
• Doppler shift is the difference between the
transmitted and received frequencies
• Transmitted and received frequencies are in
the MHz range
• Doppler shift frequencies often in audible
range
DOPPLER SHIFT
• WHEN RECEIVED FREQUENCY= TRANSMITTED
FREQUENCY, NO DOPPLER SHIFT
• Relationship between Doppler shift (or just
Doppler) frequency, FD and reflector velocity, v is
FD = 2 f0 v cos theta
c
fo is the ultrasound frequency, or the transmitted
beam frequency.
v is the reflector velocity (m/s; cm/s)
q is the Doppler angle
c is the speed of sound
DOPPLER EQUATION
DOPPLER ULTRASOUND
MODES/INSTRUMENTATIONS
• Continuous wave Doppler
• Pulsed wave Doppler
• Spectral doppler
• Colour doppler
• Power doppler
CONTINUOUS WAVE DOPPLER
• Simplest form
• Uses separate transmit & receive crystals that
continuously transmit and receive ultrasound.
• Able to detect the presence and direction of flow,
but are unable to distinguish signals arising from
vessels at different depths. ( THEREFORE IT LACKS
RANGE RESOLUTION.)
• Portable and inexpensive.
• Uses- CARDIAC SCANNERS( HIGH VELOCITIES IN
AORTA ), bedside to ascertain flow in superficial
vessels.
CONTINUOUSWAVE DOPPLER
PULSED-WAVE (PW) DOPPLER
• Uses brief pulses of ultrasound energy using
only one crystal. • The echo delay time( Te)
can be converted into distance and the DEPTH
of the echo source can be determined • The
sensitive volume from which flow data are
sampled can be controlled in terms of shape,
depth, and position.
PULSED WAVE DOPPLER
• PULSE REPETITION FREQUENCY ( PRF)- No. of
pulses transmitted per second • With increasing
scanning depth, PRF decreases as more time is
needed for the echoes to return • At a minimum,
the PRF must beat at least twice the frequency of
the Doppler signal to construct the signal
successfully. • This sets the upper limit to the
flow velocities to be accurately recorded • Use
higher PRF setting for high flow velocities, low
PRF for slow venous flow.
DUPLEX ULTRASOUND SCANNING
• Duplex ultrasound instruments are real-time
B-mode scanners with built-in Doppler
capabilities.
• B-mode Imager( Outline anatomic structures)
• Pulsed-Doppler ( Flow and movement
patterns)
SPECTRAL DOPPLER
• Ultrasound is emitted in pulse similar to B-
mode • Gate is used to determine the interval
after emission when returning signals are
received and therefore the depth from which
the sample is taken can be assessed. • The
Doppler shift data are displayed in graphic form
as a time-varying plot of the frequency
spectrum of the returning signal.
SPECTRAL ANALYSIS
• shows the Doppler shift spectrum and
direction ,usually displayed as the maximum
velocity rather than the frequency .
• Provides QUANTITATIVE INFORMATION.
SPECTRAL ANALYSIS
• The resulting Doppler frequency spectrum
displays the following : Variation with time of
the Doppler frequencies present in the
volume sampled.
• The envelope of the spectrum, represents the
maximum frequencies present at any given
point in time.
• The width of the spectrum at any point,
indicates the range of frequencies present.
DOPPLER SPECTRUM ASSESSMENT
• Assess the following 1. Presence of flow 2.
Direction of flow 3. Amplitude 4. Window 5.
Pulsatility
DOPPLER SPECTRUM ASSESSMENT
SENSITIVITY IMPROVEMENT
• Increasing the power or gain • Decreasing the
velocity scale • Decreasing the reject or filter •
Slowly increasing the SV size.
DOPPLER SPECTRUM ASSESSMENT
• DIRECTION OF FLOW • Pulsed doppler uses
quadrature phase detection to provide
bidirectional doppler information.
• FLOW CAN EITHER BE • Mono-phasic • Bi-
phasic • Tri-phasic • Bidirectional
DOPPLER SPECTRUM ASSESSMENT
• AMPLITUDE: The spectrum displays the echo
amplitude by varying the brightness of the
display. The amplitude of the echoes are
determined by • Echo intensity • Power • Gain •
Dynamic range
DOPPLER SPECTRUM ASSESSMENT
• WINDOW :
• Received doppler shift consists of a range of
frequencies
• Narrow range of frequencies will result in a
narrow display line.
• The clear area underneath the spectrum is
called the window
SPECTRUM BROADENING
• Loss of spectral window is called SPECTRAL
BROADENING.
• OCCURS-• As the blood decelerates in
diastole • SV placed close to the vessel wall •
In small vessels ( parabolic velocity profile )
SPECTRAL BROADENING
• Tortuous vessels • Low flow states • Excessive
gain/power/dynamic range.
• IT IS HALLMARK OF DISTURBED AND/OR
TURBULENT FLOW.
DOPPLER SPECTRUM ASSESSMENT
• PULSATILITY • Measures the difference
between the maximum and minimum
velocities within the cardiac cycle • Indices are
unit less • All increase in value as the flow
pulsatility increases • Can be measured
without knowledge of the doppler angle
DOPPLER INDICES
• S/D ratio(SDR): systolic velocity / diastolic velocity
• Pulsatility index (PI) (Gosling index): (PSV - EDV)
/ TAV
• Resistive index (RI) (Pourcelot index): (PSV - EDV)
/ PSV
• PSV: peak systolic velocity
• EDV: end diastolic velocity
• TAV: time-averaged velocity
COLOUR DOPPLER
• Based on pulsed doppler technique.
• Doppler shifts- converted to colour and the
moving blood is displayed in colors that
corresponds to its velocity and direction.
COLOUR DOPPLER
• Positive Doppler shifts are encoded as red and
negative shifts are encoded as blue. • Velocity of
the flow is represented in shades of color -
faster – brighter.
COLOUR DOPPLER
COLOUR BOX
• Operator adjustable area within the US image
• Affects Image resolution & quality ( change in
box size)- Frame rate decreases with increasing
box size • As small & superficial as possible .
• Deep color box – slower PRF results in aliasing.
COLOUR DOPPLER
• Advantages : It provides an overall view of the
flow in an organ or structure
• Provides directional information about flow
• Provides velocity information about flow and
shows turbulent type flow
• Limitations :• SEMI QUANTITATIVE • Angle
dependence • Aliasing • Artifacts caused by
the noise • Poor temporal resolution.
SPECTRAL DOPPLER
ADVANTAGES:
• depicts quantitative flow at one site
• allows calculations of velocity and indices
• good temporal resolution
POWER DOPPLER
• Also known as Energy Doppler / Amplitude
Doppler. • In this mode, colour is assigned to the
power /strength/ energy of the Doppler signal
rather than the Doppler frequency shift. • Flow is
usually displayed with one colour .
USES • To detect very slow flow, flow in small
vessels, or where the transducer angling is
awkward. • Power Doppler can be used in
conjunction with contrast agents of varying
brightness
POWER DOPPLER
• Advantages :Angle independent
• No aliasing
• Improved signal to noise ratio
• More sensitive to detect low flow
• Better able to define boundaries
• Limitations : No directional information
• Poor temporal resolution (due to relatively low
PRF ).
• Due to the low PRF, power Doppler is particularly
sensitive to flash artifact
OPTIMISATION OF COLOR FLOW
DOPPLER EXAMINATION
1. Transducer Frequency:
• For superficial structures 7-10MHZ is used.
• For deep abdominal structures– 3MHZ –
5MHZ is optimal.
• Choice of transducer frequency is paramount
because the intensity of the scattered sound
varies in proportion to the 4th power of the
Doppler frequency.
OPTIMISATION OF COLOR FLOW
DOPPLER EXAMINATION
2. Doppler Angle: • Strongest signals of Doppler
results when the motion is parallel to beam. • A
Doppler angle of 900 does not display flow
because no doppler shift detected.
OPTIMISATION OF COLOR FLOW
DOPPLER EXAMINATION
3. Sample Volume. • Ideal sample volume–
2/3rd of the vessel width positioned in the
center of the vessel. • If sample volume is more:
- Spectral broadening • If sample volume is less:
Measured velocity is too low.
OPTIMISATION OF COLOR FLOW
DOPPLER EXAMINATION
4. Wall Filters. • Cut off of the low frequency
noises, a cleaner high velocity blood flow signal
is displayed. • If set too high ,the blood flow is
discarded ;if set low, noise will be more.
5. Doppler Gain. • Controls the amplitude of the
colour display in colour or power Doppler mode
& the spectral display in pulse Doppler mode.
OPTIMISATION OF COLOR FLOW
DOPPLER EXAMINATION
6. Velocity scale. • Controls the range of
frequencies displayed • Too high scale-the
dynamic range is too large and low velocity
signals are missed simulating an area of
thrombosis • If scale is too low the dynamic
range is too small to display the high velocity
signals accurately resulting in aliasing.
DOPPLER ARTIFACTS
• 1.Artifacts caused by technical limitations:
• Aliasing
• Blooming artifact/colour bleed
• Directional ambiguity
• Partial volume artifact
• Absence of flow due to high velocity settings
• 2.Artifacts caused by patient anatomy
• Pseudo flow
• Flash artifact
• Mirror image artifact
• 3.Artifacts caused by machine factors
• Edge artifact-along cortical bone
• Twinkling artifact-beneath calculi
DOPPLER ARTIFACTS
• ALIASING : An inaccurate display of colour or
spectral Doppler velocity when the velocity
range exceeds the scale available to display it.
How to reduce ALIASING ?
• Drop the baseline
• Increase the available velocity range.
• Decrease the Doppler frequency shift by using
a lower insolating frequency or
• Increasing the Doppler angle.
• Power Doppler has no aliasing because it has
no directional or velocity component.
Advantage: - • 1. useful for localizing the highest
velocity region. • 2. It is used in identifying the
abnormal area in TIPSS & in localizing AV
fistulae. Disadvantage:- high velocity may not be
accurately measured.
DOPPLER ARTIFACTS
• Blooming Artifact ( Colour Bleed ).
• Caused by abnormally high gain settings.
• Causes the obscuration of thrombus or plaqu
es in the vessel. -
Also seen with ultrasound contrast agents
DOPPLER ARTIFACTS
• Flash Artifact • Manifests as a color signal due
to transducer or patient motion • It is seen in
left lobe of liver due to cardiac pulsation and
in hypoechoic areas such as cysts and fluid
collections.
DOPPLER ARTIFACTS
• Mirror Image artifact:
• displays objects on both sides of a strong refle
ctor
• The reflectors (diaphragm, pleural surface an
d aortic wall) directs some of the echoes to a
second reflector before it returns them to the
transducer resulting
in multipath reflection. Eg Duplication of sub
clavian artery (pleura reflector).
DOPPLER ARTIFACTS
• Twinkling artifact – Appears as discrete focus
of alternating colors behind echogenic object
e.g: Renal Calculi, bladder calcification and
cholesterol crystals in the gall bladder • FB like
iron fillings, aneurysm coil produce this
artifact.
GUIDELINES FOR AN OPTIMAL
COLOUR FLOW DOPPLER
EXAMINATION.
• The colour flow box should be kept as small &
superficial as possible • Adjust the gain and
filter settings • Adjust the velocity scale(PRF)
and baseline according to the flow conditions.
GUIDELINES FOR AN OPTIMAL
COLOUR FLOW DOPPLER
EXAMINATION
• Obtain an optimal Doppler angle by adjusting
the beam steering and probe position. ( 60 deg
or less) • Adjust the pulsed Doppler sample
volume size(gate) appropriately (2/3rd of the
velocity diameter) • AVOID TRANSDUCER
MOTION.
RECENT INNOVATIONS
IN DOPPLER US TECHNIQUES
• • Doppler US is experiencing technical
innovations that also contribute to improved B-
mode imaging. 1 Extended field of view imaging 2
(a) wideband, wide dynamic range systems-
improve the sensitivity of Doppler US; (b) Via
incorporation of custom-designed integrated
circuitry.- Doppler US processing on even
handheld US scanners. • 3 Contrast agents have
become much more than simple “echo
enhancers” for Doppler instrumentation.
REFERENCES
• 1. Diagnostic Ultrasound, Carol M Rumack and
others, 4th ed.
• 2.Clinical sonography-a practical guide:Roger C
Sanders, Barbara Hall-Terracciano-5th ed.

More Related Content

Similar to doppler physics.pptx

Doppler artefact (1)
Doppler artefact (1)Doppler artefact (1)
Doppler artefact (1)KamalEldirawi
 
Doppler Ultrasonography And Advancements in USG
Doppler Ultrasonography And Advancements in USGDoppler Ultrasonography And Advancements in USG
Doppler Ultrasonography And Advancements in USGSudil Paudyal
 
KNOBOLOGY FOR ECHOCARDIOGRAM.pptx
KNOBOLOGY FOR ECHOCARDIOGRAM.pptxKNOBOLOGY FOR ECHOCARDIOGRAM.pptx
KNOBOLOGY FOR ECHOCARDIOGRAM.pptxAparanji Gopidi
 
doppler physics.pptx
doppler physics.pptxdoppler physics.pptx
doppler physics.pptxalma dsouza
 
Knobology optimising echo images - dr hafeesh fazulu - pushpagiri
Knobology   optimising echo images - dr hafeesh fazulu - pushpagiriKnobology   optimising echo images - dr hafeesh fazulu - pushpagiri
Knobology optimising echo images - dr hafeesh fazulu - pushpagiriHafeesh Fazulu
 
Doppler ultrasounds (1)
Doppler ultrasounds (1)Doppler ultrasounds (1)
Doppler ultrasounds (1)KamalEldirawi
 
Doppler Effect - Ultrasound
Doppler Effect - UltrasoundDoppler Effect - Ultrasound
Doppler Effect - UltrasoundVictor Ekpo
 
01 basic principles of ultrasound & basic term
01 basic principles of ultrasound & basic term01 basic principles of ultrasound & basic term
01 basic principles of ultrasound & basic termJuan Alcatruz
 
01 basic principles of ultrasound & basic term
01 basic principles of ultrasound & basic term01 basic principles of ultrasound & basic term
01 basic principles of ultrasound & basic termJuan Alcatruz
 
ULTRASOUND IMAGING PRINCIPLES
ULTRASOUND IMAGING PRINCIPLESULTRASOUND IMAGING PRINCIPLES
ULTRASOUND IMAGING PRINCIPLESINDIA ULTRASOUND
 
Ultrasonography
UltrasonographyUltrasonography
UltrasonographyAnu V
 
Doppler principles [1]
Doppler principles [1]Doppler principles [1]
Doppler principles [1]KamalEldirawi
 
1.Introduction to Ultrasound - Adamu.pptx
1.Introduction to Ultrasound - Adamu.pptx1.Introduction to Ultrasound - Adamu.pptx
1.Introduction to Ultrasound - Adamu.pptxabelllll
 

Similar to doppler physics.pptx (20)

Doppler us (2)
Doppler us (2)Doppler us (2)
Doppler us (2)
 
Doppler artefact (1)
Doppler artefact (1)Doppler artefact (1)
Doppler artefact (1)
 
Doppler Ultrasonography And Advancements in USG
Doppler Ultrasonography And Advancements in USGDoppler Ultrasonography And Advancements in USG
Doppler Ultrasonography And Advancements in USG
 
KNOBOLOGY FOR ECHOCARDIOGRAM.pptx
KNOBOLOGY FOR ECHOCARDIOGRAM.pptxKNOBOLOGY FOR ECHOCARDIOGRAM.pptx
KNOBOLOGY FOR ECHOCARDIOGRAM.pptx
 
doppler physics.pptx
doppler physics.pptxdoppler physics.pptx
doppler physics.pptx
 
Doppler Physics
Doppler PhysicsDoppler Physics
Doppler Physics
 
Knobology optimising echo images - dr hafeesh fazulu - pushpagiri
Knobology   optimising echo images - dr hafeesh fazulu - pushpagiriKnobology   optimising echo images - dr hafeesh fazulu - pushpagiri
Knobology optimising echo images - dr hafeesh fazulu - pushpagiri
 
Doppler ultrasounds (1)
Doppler ultrasounds (1)Doppler ultrasounds (1)
Doppler ultrasounds (1)
 
Doppler Effect
Doppler EffectDoppler Effect
Doppler Effect
 
Doppler Effect - Ultrasound
Doppler Effect - UltrasoundDoppler Effect - Ultrasound
Doppler Effect - Ultrasound
 
01 basic principles of ultrasound & basic term
01 basic principles of ultrasound & basic term01 basic principles of ultrasound & basic term
01 basic principles of ultrasound & basic term
 
01 basic principles of ultrasound & basic term
01 basic principles of ultrasound & basic term01 basic principles of ultrasound & basic term
01 basic principles of ultrasound & basic term
 
Doppler us (3)
Doppler us (3)Doppler us (3)
Doppler us (3)
 
ULTRASOUND IMAGING PRINCIPLES
ULTRASOUND IMAGING PRINCIPLESULTRASOUND IMAGING PRINCIPLES
ULTRASOUND IMAGING PRINCIPLES
 
Ultrasonography
UltrasonographyUltrasonography
Ultrasonography
 
Doppler principles [1]
Doppler principles [1]Doppler principles [1]
Doppler principles [1]
 
Doppler echocardiography
Doppler echocardiographyDoppler echocardiography
Doppler echocardiography
 
Doppler echocardiography
Doppler echocardiographyDoppler echocardiography
Doppler echocardiography
 
Doppler physics
Doppler physics Doppler physics
Doppler physics
 
1.Introduction to Ultrasound - Adamu.pptx
1.Introduction to Ultrasound - Adamu.pptx1.Introduction to Ultrasound - Adamu.pptx
1.Introduction to Ultrasound - Adamu.pptx
 

More from rohanjohnjacob

hydrocephalus and csf disorders powerpoint
hydrocephalus and csf disorders powerpointhydrocephalus and csf disorders powerpoint
hydrocephalus and csf disorders powerpointrohanjohnjacob
 
radiology in dementia powerpoint presentation
radiology in dementia powerpoint presentationradiology in dementia powerpoint presentation
radiology in dementia powerpoint presentationrohanjohnjacob
 
Radiographic anatomy of lungs.pptx
Radiographic anatomy of lungs.pptxRadiographic anatomy of lungs.pptx
Radiographic anatomy of lungs.pptxrohanjohnjacob
 
upper limb trauma.pptx
upper limb trauma.pptxupper limb trauma.pptx
upper limb trauma.pptxrohanjohnjacob
 
radio cassettes and screens.pptx
radio cassettes and screens.pptxradio cassettes and screens.pptx
radio cassettes and screens.pptxrohanjohnjacob
 
Normal abd xray rjj.pptx
Normal abd xray rjj.pptxNormal abd xray rjj.pptx
Normal abd xray rjj.pptxrohanjohnjacob
 
ADVERSE EFFECTS OF CONTRAST AGENTS ppt.pptx
ADVERSE EFFECTS OF CONTRAST AGENTS ppt.pptxADVERSE EFFECTS OF CONTRAST AGENTS ppt.pptx
ADVERSE EFFECTS OF CONTRAST AGENTS ppt.pptxrohanjohnjacob
 
upper limb trauma.pptx
upper limb trauma.pptxupper limb trauma.pptx
upper limb trauma.pptxrohanjohnjacob
 
linesmediastinalstripes-01-101126132247-phpapp01.pptx
linesmediastinalstripes-01-101126132247-phpapp01.pptxlinesmediastinalstripes-01-101126132247-phpapp01.pptx
linesmediastinalstripes-01-101126132247-phpapp01.pptxrohanjohnjacob
 
CONTRAST AGENTS PPT.pptx
CONTRAST AGENTS PPT.pptxCONTRAST AGENTS PPT.pptx
CONTRAST AGENTS PPT.pptxrohanjohnjacob
 

More from rohanjohnjacob (20)

hydrocephalus and csf disorders powerpoint
hydrocephalus and csf disorders powerpointhydrocephalus and csf disorders powerpoint
hydrocephalus and csf disorders powerpoint
 
radiology in dementia powerpoint presentation
radiology in dementia powerpoint presentationradiology in dementia powerpoint presentation
radiology in dementia powerpoint presentation
 
Radiographic anatomy of lungs.pptx
Radiographic anatomy of lungs.pptxRadiographic anatomy of lungs.pptx
Radiographic anatomy of lungs.pptx
 
barium swallow.pptx
barium swallow.pptxbarium swallow.pptx
barium swallow.pptx
 
esophagus 2.pptx
esophagus 2.pptxesophagus 2.pptx
esophagus 2.pptx
 
upper limb trauma.pptx
upper limb trauma.pptxupper limb trauma.pptx
upper limb trauma.pptx
 
usg artifacts.pptx
usg artifacts.pptxusg artifacts.pptx
usg artifacts.pptx
 
radio cassettes and screens.pptx
radio cassettes and screens.pptxradio cassettes and screens.pptx
radio cassettes and screens.pptx
 
Normal abd xray rjj.pptx
Normal abd xray rjj.pptxNormal abd xray rjj.pptx
Normal abd xray rjj.pptx
 
esophagus.pptx
esophagus.pptxesophagus.pptx
esophagus.pptx
 
grids,films.pptx
grids,films.pptxgrids,films.pptx
grids,films.pptx
 
ADVERSE EFFECTS OF CONTRAST AGENTS ppt.pptx
ADVERSE EFFECTS OF CONTRAST AGENTS ppt.pptxADVERSE EFFECTS OF CONTRAST AGENTS ppt.pptx
ADVERSE EFFECTS OF CONTRAST AGENTS ppt.pptx
 
PET SCANNING.pptx
PET SCANNING.pptxPET SCANNING.pptx
PET SCANNING.pptx
 
MAMMOGRAPHY.pptx
MAMMOGRAPHY.pptxMAMMOGRAPHY.pptx
MAMMOGRAPHY.pptx
 
rad ana of chest.pdf
rad ana of chest.pdfrad ana of chest.pdf
rad ana of chest.pdf
 
upper limb trauma.pptx
upper limb trauma.pptxupper limb trauma.pptx
upper limb trauma.pptx
 
linesmediastinalstripes-01-101126132247-phpapp01.pptx
linesmediastinalstripes-01-101126132247-phpapp01.pptxlinesmediastinalstripes-01-101126132247-phpapp01.pptx
linesmediastinalstripes-01-101126132247-phpapp01.pptx
 
DARK ROOM.pptx
DARK ROOM.pptxDARK ROOM.pptx
DARK ROOM.pptx
 
CHEST XRAYS RJJ.pptx
CHEST XRAYS RJJ.pptxCHEST XRAYS RJJ.pptx
CHEST XRAYS RJJ.pptx
 
CONTRAST AGENTS PPT.pptx
CONTRAST AGENTS PPT.pptxCONTRAST AGENTS PPT.pptx
CONTRAST AGENTS PPT.pptx
 

Recently uploaded

Call Girls Service Jaipur {9521753030} ❤️VVIP RIDDHI Call Girl in Jaipur Raja...
Call Girls Service Jaipur {9521753030} ❤️VVIP RIDDHI Call Girl in Jaipur Raja...Call Girls Service Jaipur {9521753030} ❤️VVIP RIDDHI Call Girl in Jaipur Raja...
Call Girls Service Jaipur {9521753030} ❤️VVIP RIDDHI Call Girl in Jaipur Raja...Sheetaleventcompany
 
Jogeshwari ! Call Girls Service Mumbai - 450+ Call Girl Cash Payment 90042684...
Jogeshwari ! Call Girls Service Mumbai - 450+ Call Girl Cash Payment 90042684...Jogeshwari ! Call Girls Service Mumbai - 450+ Call Girl Cash Payment 90042684...
Jogeshwari ! Call Girls Service Mumbai - 450+ Call Girl Cash Payment 90042684...Anamika Rawat
 
Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...
Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...
Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...aartirawatdelhi
 
All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...
All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...
All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...Arohi Goyal
 
Russian Call Girls Lucknow Just Call 👉👉7877925207 Top Class Call Girl Service...
Russian Call Girls Lucknow Just Call 👉👉7877925207 Top Class Call Girl Service...Russian Call Girls Lucknow Just Call 👉👉7877925207 Top Class Call Girl Service...
Russian Call Girls Lucknow Just Call 👉👉7877925207 Top Class Call Girl Service...adilkhan87451
 
Andheri East ) Call Girls in Mumbai Phone No 9004268417 Elite Escort Service ...
Andheri East ) Call Girls in Mumbai Phone No 9004268417 Elite Escort Service ...Andheri East ) Call Girls in Mumbai Phone No 9004268417 Elite Escort Service ...
Andheri East ) Call Girls in Mumbai Phone No 9004268417 Elite Escort Service ...Anamika Rawat
 
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 9332606886 𖠋 Will You Mis...
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 9332606886 𖠋 Will You Mis...The Most Attractive Hyderabad Call Girls Kothapet 𖠋 9332606886 𖠋 Will You Mis...
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 9332606886 𖠋 Will You Mis...chandars293
 
Call Girls Jaipur Just Call 9521753030 Top Class Call Girl Service Available
Call Girls Jaipur Just Call 9521753030 Top Class Call Girl Service AvailableCall Girls Jaipur Just Call 9521753030 Top Class Call Girl Service Available
Call Girls Jaipur Just Call 9521753030 Top Class Call Girl Service AvailableJanvi Singh
 
Saket * Call Girls in Delhi - Phone 9711199012 Escorts Service at 6k to 50k a...
Saket * Call Girls in Delhi - Phone 9711199012 Escorts Service at 6k to 50k a...Saket * Call Girls in Delhi - Phone 9711199012 Escorts Service at 6k to 50k a...
Saket * Call Girls in Delhi - Phone 9711199012 Escorts Service at 6k to 50k a...BhumiSaxena1
 
Most Beautiful Call Girl in Bangalore Contact on Whatsapp
Most Beautiful Call Girl in Bangalore Contact on WhatsappMost Beautiful Call Girl in Bangalore Contact on Whatsapp
Most Beautiful Call Girl in Bangalore Contact on WhatsappInaaya Sharma
 
Call Girls Ahmedabad Just Call 9630942363 Top Class Call Girl Service Available
Call Girls Ahmedabad Just Call 9630942363 Top Class Call Girl Service AvailableCall Girls Ahmedabad Just Call 9630942363 Top Class Call Girl Service Available
Call Girls Ahmedabad Just Call 9630942363 Top Class Call Girl Service AvailableGENUINE ESCORT AGENCY
 
Coimbatore Call Girls in Thudiyalur : 7427069034 High Profile Model Escorts |...
Coimbatore Call Girls in Thudiyalur : 7427069034 High Profile Model Escorts |...Coimbatore Call Girls in Thudiyalur : 7427069034 High Profile Model Escorts |...
Coimbatore Call Girls in Thudiyalur : 7427069034 High Profile Model Escorts |...chennailover
 
Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...
Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...
Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...vidya singh
 
Premium Call Girls In Jaipur {8445551418} ❤️VVIP SEEMA Call Girl in Jaipur Ra...
Premium Call Girls In Jaipur {8445551418} ❤️VVIP SEEMA Call Girl in Jaipur Ra...Premium Call Girls In Jaipur {8445551418} ❤️VVIP SEEMA Call Girl in Jaipur Ra...
Premium Call Girls In Jaipur {8445551418} ❤️VVIP SEEMA Call Girl in Jaipur Ra...parulsinha
 
Trichy Call Girls Book Now 9630942363 Top Class Trichy Escort Service Available
Trichy Call Girls Book Now 9630942363 Top Class Trichy Escort Service AvailableTrichy Call Girls Book Now 9630942363 Top Class Trichy Escort Service Available
Trichy Call Girls Book Now 9630942363 Top Class Trichy Escort Service AvailableGENUINE ESCORT AGENCY
 
Call Girls Raipur Just Call 9630942363 Top Class Call Girl Service Available
Call Girls Raipur Just Call 9630942363 Top Class Call Girl Service AvailableCall Girls Raipur Just Call 9630942363 Top Class Call Girl Service Available
Call Girls Raipur Just Call 9630942363 Top Class Call Girl Service AvailableGENUINE ESCORT AGENCY
 
Call Girls Hosur Just Call 9630942363 Top Class Call Girl Service Available
Call Girls Hosur Just Call 9630942363 Top Class Call Girl Service AvailableCall Girls Hosur Just Call 9630942363 Top Class Call Girl Service Available
Call Girls Hosur Just Call 9630942363 Top Class Call Girl Service AvailableGENUINE ESCORT AGENCY
 
Premium Bangalore Call Girls Jigani Dail 6378878445 Escort Service For Hot Ma...
Premium Bangalore Call Girls Jigani Dail 6378878445 Escort Service For Hot Ma...Premium Bangalore Call Girls Jigani Dail 6378878445 Escort Service For Hot Ma...
Premium Bangalore Call Girls Jigani Dail 6378878445 Escort Service For Hot Ma...tanya dube
 
Mumbai ] (Call Girls) in Mumbai 10k @ I'm VIP Independent Escorts Girls 98333...
Mumbai ] (Call Girls) in Mumbai 10k @ I'm VIP Independent Escorts Girls 98333...Mumbai ] (Call Girls) in Mumbai 10k @ I'm VIP Independent Escorts Girls 98333...
Mumbai ] (Call Girls) in Mumbai 10k @ I'm VIP Independent Escorts Girls 98333...Ishani Gupta
 

Recently uploaded (20)

Call Girls Service Jaipur {9521753030} ❤️VVIP RIDDHI Call Girl in Jaipur Raja...
Call Girls Service Jaipur {9521753030} ❤️VVIP RIDDHI Call Girl in Jaipur Raja...Call Girls Service Jaipur {9521753030} ❤️VVIP RIDDHI Call Girl in Jaipur Raja...
Call Girls Service Jaipur {9521753030} ❤️VVIP RIDDHI Call Girl in Jaipur Raja...
 
Jogeshwari ! Call Girls Service Mumbai - 450+ Call Girl Cash Payment 90042684...
Jogeshwari ! Call Girls Service Mumbai - 450+ Call Girl Cash Payment 90042684...Jogeshwari ! Call Girls Service Mumbai - 450+ Call Girl Cash Payment 90042684...
Jogeshwari ! Call Girls Service Mumbai - 450+ Call Girl Cash Payment 90042684...
 
Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...
Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...
Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...
 
All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...
All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...
All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...
 
Russian Call Girls Lucknow Just Call 👉👉7877925207 Top Class Call Girl Service...
Russian Call Girls Lucknow Just Call 👉👉7877925207 Top Class Call Girl Service...Russian Call Girls Lucknow Just Call 👉👉7877925207 Top Class Call Girl Service...
Russian Call Girls Lucknow Just Call 👉👉7877925207 Top Class Call Girl Service...
 
Andheri East ) Call Girls in Mumbai Phone No 9004268417 Elite Escort Service ...
Andheri East ) Call Girls in Mumbai Phone No 9004268417 Elite Escort Service ...Andheri East ) Call Girls in Mumbai Phone No 9004268417 Elite Escort Service ...
Andheri East ) Call Girls in Mumbai Phone No 9004268417 Elite Escort Service ...
 
🌹Attapur⬅️ Vip Call Girls Hyderabad 📱9352852248 Book Well Trand Call Girls In...
🌹Attapur⬅️ Vip Call Girls Hyderabad 📱9352852248 Book Well Trand Call Girls In...🌹Attapur⬅️ Vip Call Girls Hyderabad 📱9352852248 Book Well Trand Call Girls In...
🌹Attapur⬅️ Vip Call Girls Hyderabad 📱9352852248 Book Well Trand Call Girls In...
 
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 9332606886 𖠋 Will You Mis...
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 9332606886 𖠋 Will You Mis...The Most Attractive Hyderabad Call Girls Kothapet 𖠋 9332606886 𖠋 Will You Mis...
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 9332606886 𖠋 Will You Mis...
 
Call Girls Jaipur Just Call 9521753030 Top Class Call Girl Service Available
Call Girls Jaipur Just Call 9521753030 Top Class Call Girl Service AvailableCall Girls Jaipur Just Call 9521753030 Top Class Call Girl Service Available
Call Girls Jaipur Just Call 9521753030 Top Class Call Girl Service Available
 
Saket * Call Girls in Delhi - Phone 9711199012 Escorts Service at 6k to 50k a...
Saket * Call Girls in Delhi - Phone 9711199012 Escorts Service at 6k to 50k a...Saket * Call Girls in Delhi - Phone 9711199012 Escorts Service at 6k to 50k a...
Saket * Call Girls in Delhi - Phone 9711199012 Escorts Service at 6k to 50k a...
 
Most Beautiful Call Girl in Bangalore Contact on Whatsapp
Most Beautiful Call Girl in Bangalore Contact on WhatsappMost Beautiful Call Girl in Bangalore Contact on Whatsapp
Most Beautiful Call Girl in Bangalore Contact on Whatsapp
 
Call Girls Ahmedabad Just Call 9630942363 Top Class Call Girl Service Available
Call Girls Ahmedabad Just Call 9630942363 Top Class Call Girl Service AvailableCall Girls Ahmedabad Just Call 9630942363 Top Class Call Girl Service Available
Call Girls Ahmedabad Just Call 9630942363 Top Class Call Girl Service Available
 
Coimbatore Call Girls in Thudiyalur : 7427069034 High Profile Model Escorts |...
Coimbatore Call Girls in Thudiyalur : 7427069034 High Profile Model Escorts |...Coimbatore Call Girls in Thudiyalur : 7427069034 High Profile Model Escorts |...
Coimbatore Call Girls in Thudiyalur : 7427069034 High Profile Model Escorts |...
 
Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...
Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...
Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...
 
Premium Call Girls In Jaipur {8445551418} ❤️VVIP SEEMA Call Girl in Jaipur Ra...
Premium Call Girls In Jaipur {8445551418} ❤️VVIP SEEMA Call Girl in Jaipur Ra...Premium Call Girls In Jaipur {8445551418} ❤️VVIP SEEMA Call Girl in Jaipur Ra...
Premium Call Girls In Jaipur {8445551418} ❤️VVIP SEEMA Call Girl in Jaipur Ra...
 
Trichy Call Girls Book Now 9630942363 Top Class Trichy Escort Service Available
Trichy Call Girls Book Now 9630942363 Top Class Trichy Escort Service AvailableTrichy Call Girls Book Now 9630942363 Top Class Trichy Escort Service Available
Trichy Call Girls Book Now 9630942363 Top Class Trichy Escort Service Available
 
Call Girls Raipur Just Call 9630942363 Top Class Call Girl Service Available
Call Girls Raipur Just Call 9630942363 Top Class Call Girl Service AvailableCall Girls Raipur Just Call 9630942363 Top Class Call Girl Service Available
Call Girls Raipur Just Call 9630942363 Top Class Call Girl Service Available
 
Call Girls Hosur Just Call 9630942363 Top Class Call Girl Service Available
Call Girls Hosur Just Call 9630942363 Top Class Call Girl Service AvailableCall Girls Hosur Just Call 9630942363 Top Class Call Girl Service Available
Call Girls Hosur Just Call 9630942363 Top Class Call Girl Service Available
 
Premium Bangalore Call Girls Jigani Dail 6378878445 Escort Service For Hot Ma...
Premium Bangalore Call Girls Jigani Dail 6378878445 Escort Service For Hot Ma...Premium Bangalore Call Girls Jigani Dail 6378878445 Escort Service For Hot Ma...
Premium Bangalore Call Girls Jigani Dail 6378878445 Escort Service For Hot Ma...
 
Mumbai ] (Call Girls) in Mumbai 10k @ I'm VIP Independent Escorts Girls 98333...
Mumbai ] (Call Girls) in Mumbai 10k @ I'm VIP Independent Escorts Girls 98333...Mumbai ] (Call Girls) in Mumbai 10k @ I'm VIP Independent Escorts Girls 98333...
Mumbai ] (Call Girls) in Mumbai 10k @ I'm VIP Independent Escorts Girls 98333...
 

doppler physics.pptx

  • 2. PRINCIPLE OF DOPPLER ULTRASOUND • BASIS – ‘ DOPPLER EFFECT’ • DISCOVERY – Physicist Christian Johann Doppler in 1842 • ‘When a sound source and reflector are moving towards each other, the sound waves are spaced closer together and reach the receiver at a higher frequency than that at which they were originally emitted ’.
  • 3. DOPPLER SHIFT • Doppler shift is the difference between the transmitted and received frequencies • Transmitted and received frequencies are in the MHz range • Doppler shift frequencies often in audible range
  • 4. DOPPLER SHIFT • WHEN RECEIVED FREQUENCY= TRANSMITTED FREQUENCY, NO DOPPLER SHIFT
  • 5. • Relationship between Doppler shift (or just Doppler) frequency, FD and reflector velocity, v is FD = 2 f0 v cos theta c fo is the ultrasound frequency, or the transmitted beam frequency. v is the reflector velocity (m/s; cm/s) q is the Doppler angle c is the speed of sound DOPPLER EQUATION
  • 6. DOPPLER ULTRASOUND MODES/INSTRUMENTATIONS • Continuous wave Doppler • Pulsed wave Doppler • Spectral doppler • Colour doppler • Power doppler
  • 7. CONTINUOUS WAVE DOPPLER • Simplest form • Uses separate transmit & receive crystals that continuously transmit and receive ultrasound. • Able to detect the presence and direction of flow, but are unable to distinguish signals arising from vessels at different depths. ( THEREFORE IT LACKS RANGE RESOLUTION.) • Portable and inexpensive. • Uses- CARDIAC SCANNERS( HIGH VELOCITIES IN AORTA ), bedside to ascertain flow in superficial vessels.
  • 9. PULSED-WAVE (PW) DOPPLER • Uses brief pulses of ultrasound energy using only one crystal. • The echo delay time( Te) can be converted into distance and the DEPTH of the echo source can be determined • The sensitive volume from which flow data are sampled can be controlled in terms of shape, depth, and position.
  • 10. PULSED WAVE DOPPLER • PULSE REPETITION FREQUENCY ( PRF)- No. of pulses transmitted per second • With increasing scanning depth, PRF decreases as more time is needed for the echoes to return • At a minimum, the PRF must beat at least twice the frequency of the Doppler signal to construct the signal successfully. • This sets the upper limit to the flow velocities to be accurately recorded • Use higher PRF setting for high flow velocities, low PRF for slow venous flow.
  • 11.
  • 12. DUPLEX ULTRASOUND SCANNING • Duplex ultrasound instruments are real-time B-mode scanners with built-in Doppler capabilities. • B-mode Imager( Outline anatomic structures) • Pulsed-Doppler ( Flow and movement patterns)
  • 13. SPECTRAL DOPPLER • Ultrasound is emitted in pulse similar to B- mode • Gate is used to determine the interval after emission when returning signals are received and therefore the depth from which the sample is taken can be assessed. • The Doppler shift data are displayed in graphic form as a time-varying plot of the frequency spectrum of the returning signal.
  • 14.
  • 15. SPECTRAL ANALYSIS • shows the Doppler shift spectrum and direction ,usually displayed as the maximum velocity rather than the frequency . • Provides QUANTITATIVE INFORMATION.
  • 16. SPECTRAL ANALYSIS • The resulting Doppler frequency spectrum displays the following : Variation with time of the Doppler frequencies present in the volume sampled. • The envelope of the spectrum, represents the maximum frequencies present at any given point in time. • The width of the spectrum at any point, indicates the range of frequencies present.
  • 17. DOPPLER SPECTRUM ASSESSMENT • Assess the following 1. Presence of flow 2. Direction of flow 3. Amplitude 4. Window 5. Pulsatility
  • 18. DOPPLER SPECTRUM ASSESSMENT SENSITIVITY IMPROVEMENT • Increasing the power or gain • Decreasing the velocity scale • Decreasing the reject or filter • Slowly increasing the SV size.
  • 19. DOPPLER SPECTRUM ASSESSMENT • DIRECTION OF FLOW • Pulsed doppler uses quadrature phase detection to provide bidirectional doppler information. • FLOW CAN EITHER BE • Mono-phasic • Bi- phasic • Tri-phasic • Bidirectional
  • 20.
  • 21.
  • 22.
  • 23.
  • 24.
  • 25. DOPPLER SPECTRUM ASSESSMENT • AMPLITUDE: The spectrum displays the echo amplitude by varying the brightness of the display. The amplitude of the echoes are determined by • Echo intensity • Power • Gain • Dynamic range
  • 26. DOPPLER SPECTRUM ASSESSMENT • WINDOW : • Received doppler shift consists of a range of frequencies • Narrow range of frequencies will result in a narrow display line. • The clear area underneath the spectrum is called the window
  • 27. SPECTRUM BROADENING • Loss of spectral window is called SPECTRAL BROADENING. • OCCURS-• As the blood decelerates in diastole • SV placed close to the vessel wall • In small vessels ( parabolic velocity profile )
  • 28. SPECTRAL BROADENING • Tortuous vessels • Low flow states • Excessive gain/power/dynamic range. • IT IS HALLMARK OF DISTURBED AND/OR TURBULENT FLOW.
  • 29.
  • 30. DOPPLER SPECTRUM ASSESSMENT • PULSATILITY • Measures the difference between the maximum and minimum velocities within the cardiac cycle • Indices are unit less • All increase in value as the flow pulsatility increases • Can be measured without knowledge of the doppler angle
  • 31.
  • 32. DOPPLER INDICES • S/D ratio(SDR): systolic velocity / diastolic velocity • Pulsatility index (PI) (Gosling index): (PSV - EDV) / TAV • Resistive index (RI) (Pourcelot index): (PSV - EDV) / PSV • PSV: peak systolic velocity • EDV: end diastolic velocity • TAV: time-averaged velocity
  • 33. COLOUR DOPPLER • Based on pulsed doppler technique. • Doppler shifts- converted to colour and the moving blood is displayed in colors that corresponds to its velocity and direction.
  • 34. COLOUR DOPPLER • Positive Doppler shifts are encoded as red and negative shifts are encoded as blue. • Velocity of the flow is represented in shades of color - faster – brighter.
  • 35.
  • 36. COLOUR DOPPLER COLOUR BOX • Operator adjustable area within the US image • Affects Image resolution & quality ( change in box size)- Frame rate decreases with increasing box size • As small & superficial as possible . • Deep color box – slower PRF results in aliasing.
  • 37.
  • 38. COLOUR DOPPLER • Advantages : It provides an overall view of the flow in an organ or structure • Provides directional information about flow • Provides velocity information about flow and shows turbulent type flow • Limitations :• SEMI QUANTITATIVE • Angle dependence • Aliasing • Artifacts caused by the noise • Poor temporal resolution.
  • 39. SPECTRAL DOPPLER ADVANTAGES: • depicts quantitative flow at one site • allows calculations of velocity and indices • good temporal resolution
  • 40.
  • 41.
  • 42. POWER DOPPLER • Also known as Energy Doppler / Amplitude Doppler. • In this mode, colour is assigned to the power /strength/ energy of the Doppler signal rather than the Doppler frequency shift. • Flow is usually displayed with one colour . USES • To detect very slow flow, flow in small vessels, or where the transducer angling is awkward. • Power Doppler can be used in conjunction with contrast agents of varying brightness
  • 43.
  • 44. POWER DOPPLER • Advantages :Angle independent • No aliasing • Improved signal to noise ratio • More sensitive to detect low flow • Better able to define boundaries • Limitations : No directional information • Poor temporal resolution (due to relatively low PRF ). • Due to the low PRF, power Doppler is particularly sensitive to flash artifact
  • 45. OPTIMISATION OF COLOR FLOW DOPPLER EXAMINATION 1. Transducer Frequency: • For superficial structures 7-10MHZ is used. • For deep abdominal structures– 3MHZ – 5MHZ is optimal. • Choice of transducer frequency is paramount because the intensity of the scattered sound varies in proportion to the 4th power of the Doppler frequency.
  • 46. OPTIMISATION OF COLOR FLOW DOPPLER EXAMINATION 2. Doppler Angle: • Strongest signals of Doppler results when the motion is parallel to beam. • A Doppler angle of 900 does not display flow because no doppler shift detected.
  • 47. OPTIMISATION OF COLOR FLOW DOPPLER EXAMINATION 3. Sample Volume. • Ideal sample volume– 2/3rd of the vessel width positioned in the center of the vessel. • If sample volume is more: - Spectral broadening • If sample volume is less: Measured velocity is too low.
  • 48. OPTIMISATION OF COLOR FLOW DOPPLER EXAMINATION 4. Wall Filters. • Cut off of the low frequency noises, a cleaner high velocity blood flow signal is displayed. • If set too high ,the blood flow is discarded ;if set low, noise will be more. 5. Doppler Gain. • Controls the amplitude of the colour display in colour or power Doppler mode & the spectral display in pulse Doppler mode.
  • 49. OPTIMISATION OF COLOR FLOW DOPPLER EXAMINATION 6. Velocity scale. • Controls the range of frequencies displayed • Too high scale-the dynamic range is too large and low velocity signals are missed simulating an area of thrombosis • If scale is too low the dynamic range is too small to display the high velocity signals accurately resulting in aliasing.
  • 50. DOPPLER ARTIFACTS • 1.Artifacts caused by technical limitations: • Aliasing • Blooming artifact/colour bleed • Directional ambiguity • Partial volume artifact • Absence of flow due to high velocity settings • 2.Artifacts caused by patient anatomy • Pseudo flow • Flash artifact • Mirror image artifact • 3.Artifacts caused by machine factors • Edge artifact-along cortical bone • Twinkling artifact-beneath calculi
  • 51. DOPPLER ARTIFACTS • ALIASING : An inaccurate display of colour or spectral Doppler velocity when the velocity range exceeds the scale available to display it.
  • 52.
  • 53. How to reduce ALIASING ? • Drop the baseline • Increase the available velocity range. • Decrease the Doppler frequency shift by using a lower insolating frequency or • Increasing the Doppler angle.
  • 54. • Power Doppler has no aliasing because it has no directional or velocity component. Advantage: - • 1. useful for localizing the highest velocity region. • 2. It is used in identifying the abnormal area in TIPSS & in localizing AV fistulae. Disadvantage:- high velocity may not be accurately measured.
  • 55. DOPPLER ARTIFACTS • Blooming Artifact ( Colour Bleed ). • Caused by abnormally high gain settings. • Causes the obscuration of thrombus or plaqu es in the vessel. - Also seen with ultrasound contrast agents
  • 56.
  • 57. DOPPLER ARTIFACTS • Flash Artifact • Manifests as a color signal due to transducer or patient motion • It is seen in left lobe of liver due to cardiac pulsation and in hypoechoic areas such as cysts and fluid collections.
  • 58.
  • 59. DOPPLER ARTIFACTS • Mirror Image artifact: • displays objects on both sides of a strong refle ctor • The reflectors (diaphragm, pleural surface an d aortic wall) directs some of the echoes to a second reflector before it returns them to the transducer resulting in multipath reflection. Eg Duplication of sub clavian artery (pleura reflector).
  • 60.
  • 61. DOPPLER ARTIFACTS • Twinkling artifact – Appears as discrete focus of alternating colors behind echogenic object e.g: Renal Calculi, bladder calcification and cholesterol crystals in the gall bladder • FB like iron fillings, aneurysm coil produce this artifact.
  • 62.
  • 63. GUIDELINES FOR AN OPTIMAL COLOUR FLOW DOPPLER EXAMINATION. • The colour flow box should be kept as small & superficial as possible • Adjust the gain and filter settings • Adjust the velocity scale(PRF) and baseline according to the flow conditions.
  • 64. GUIDELINES FOR AN OPTIMAL COLOUR FLOW DOPPLER EXAMINATION • Obtain an optimal Doppler angle by adjusting the beam steering and probe position. ( 60 deg or less) • Adjust the pulsed Doppler sample volume size(gate) appropriately (2/3rd of the velocity diameter) • AVOID TRANSDUCER MOTION.
  • 65. RECENT INNOVATIONS IN DOPPLER US TECHNIQUES • • Doppler US is experiencing technical innovations that also contribute to improved B- mode imaging. 1 Extended field of view imaging 2 (a) wideband, wide dynamic range systems- improve the sensitivity of Doppler US; (b) Via incorporation of custom-designed integrated circuitry.- Doppler US processing on even handheld US scanners. • 3 Contrast agents have become much more than simple “echo enhancers” for Doppler instrumentation.
  • 66. REFERENCES • 1. Diagnostic Ultrasound, Carol M Rumack and others, 4th ed. • 2.Clinical sonography-a practical guide:Roger C Sanders, Barbara Hall-Terracciano-5th ed.