Perspectives in Blood Flow Measurement By SuhasDeshpande Prof. Druzgalski             EE 506 (Fall 2009) P3
Outline About Blood Flow measurement Measurement techniques Electromagnetic methods Thermal methods Tracer dyes for measurement Ultrasound measurement Plethysmograph probe design Blood flow using thermal analysis Biotelemetry system References
Blood flow measurement Provides vital info about the oxygen and nutrient supply to the various parts of body More difficult and complicated Most of the methods are invasive techniques Blood flow ranges from 0.45 m/s in the arteries to 0.1 m/s in veins Blood pressure measurement provides supplemental info which can be used
Measurement techniques Electromagnetic methods Ultrasonic methods Thermal or using tracers methods Peripheral or indirect methods
Electromagnetic methods Blood being a conductive fluid adheres to the Faraday’s law of electromagnetic induction Flow in a magnetic field results in a generation of electromagnetic force perpendicular to the magnetic field and flow Another technique used is that of Nuclear magnetic resonance [2]
Thermal methods A thermistor or a hot wire anemometer can be implanted inside the body. Blood acts as a coolant for a thermistor . By different designs and calculations based on fluid mechanics and heat transfer flow of blood can be estimated [5]
Tracer dyes for measurement The injection of some tracers helps in imaging and analyzing the blood flow Indocyanine Green is a dye which causes fluorescence and images the blood flow   [6]
Ultrasound measurement Blood flow measurement using implantable CMUT Array (Capacitive Micromachined Ultrasonic Transducer) CMUT consists of a SOC cmos electronics and transducer with cross section 40µm and shaft length 4-10mm [1]
Experimental operation Ultrasonic pulse transmitted at frequency 1/T  Echo is received at time t1 from position 1, t2 from positions 2 The velocity  is calculated as  c(t2-t1)/2T cosѳ [1]
Photoplethysmograph probe design ,[object Object]
  The IR diode (OP240D Optek) and Phototransistor     (LPT85A Siemens) were attached to an ellipsoid plastic sheet 0.5 cm apart
  Probe attached to the skin using an ECG electrode[4]
Blood flow using thermal analysis Arteriola and Venula connected by blood capillaries and AVA Blood flow to capillaries controlled by AVA Temperature regulation is based on blood flow in capillaries Heat transfer model analysis  [7]
Blood flow in fingertip using heat transfer analysis Non Contact sensor placed below consists of an IR detector (Thermopile and resistance temperature sensor) Sensitivity enhanced by a Ge lens Contact sensor consists of thermistor (0.25mm) to transmit the heat   [7]
Blood flow in fingertip using heat transfer analysis First step ambient temperature measured (25s) Second step finger placed on the sensor initial temperature of skin surface measured (35s) Third step measure the thermal release characteristics for correction in second step (15s) Blood flow coefficient is calculated by using the temperature gradient in step 2 [7]
Biotelemetry system The implant module (B) consists of a doppler flow meter µcontroller and a RF link communicator The base station (A) consists of RF link µcontroller and D-A Converter [8]

Más contenido relacionado

La actualidad más candente

Cardiac Pacemakers
Cardiac Pacemakers Cardiac Pacemakers
Cardiac Pacemakers
Anirudhya J
 
Biopotentials
BiopotentialsBiopotentials
Biopotentials
stooty s
 
Transducers for bio medical
Transducers for bio medicalTransducers for bio medical
Transducers for bio medical
SLIET
 

La actualidad más candente (20)

Unit 4 biomedical
Unit 4 biomedicalUnit 4 biomedical
Unit 4 biomedical
 
Blood pH Measurement
Blood pH MeasurementBlood pH Measurement
Blood pH Measurement
 
ECG Elementary
ECG  Elementary ECG  Elementary
ECG Elementary
 
Cardiac pacemaker
Cardiac pacemakerCardiac pacemaker
Cardiac pacemaker
 
Patient monitoring system and biotelemetry
Patient monitoring system and biotelemetryPatient monitoring system and biotelemetry
Patient monitoring system and biotelemetry
 
20 patient monitoring
20 patient monitoring20 patient monitoring
20 patient monitoring
 
Vector cardiography
Vector cardiographyVector cardiography
Vector cardiography
 
Cardiac Pacemakers
Cardiac Pacemakers Cardiac Pacemakers
Cardiac Pacemakers
 
Capacitive Type Transducer
Capacitive Type Transducer Capacitive Type Transducer
Capacitive Type Transducer
 
Physiological transducers
Physiological transducersPhysiological transducers
Physiological transducers
 
Medical instrumentation- patient monitoring systems
Medical instrumentation- patient monitoring systems Medical instrumentation- patient monitoring systems
Medical instrumentation- patient monitoring systems
 
Cardiac Pacemaker
Cardiac PacemakerCardiac Pacemaker
Cardiac Pacemaker
 
Application of different medical electrodes
Application of different medical electrodesApplication of different medical electrodes
Application of different medical electrodes
 
Measurement of Blood Pressure
Measurement of Blood PressureMeasurement of Blood Pressure
Measurement of Blood Pressure
 
ECG ARTIFACTS AND PITFALLS
ECG ARTIFACTS AND PITFALLS  ECG ARTIFACTS AND PITFALLS
ECG ARTIFACTS AND PITFALLS
 
Basics of ECG
Basics of ECGBasics of ECG
Basics of ECG
 
Biopotentials
BiopotentialsBiopotentials
Biopotentials
 
Transducers for bio medical
Transducers for bio medicalTransducers for bio medical
Transducers for bio medical
 
Bio Potential and Bio Electrodes
Bio Potential and Bio ElectrodesBio Potential and Bio Electrodes
Bio Potential and Bio Electrodes
 
Medical electronics
Medical electronicsMedical electronics
Medical electronics
 

Destacado

Estimation of body mass distribution using projective image
Estimation of body mass distribution using projective imageEstimation of body mass distribution using projective image
Estimation of body mass distribution using projective image
Suhas Deshpande
 
Trend presentation
Trend presentationTrend presentation
Trend presentation
Joe Spencer
 
OT_Biotelemetry_System_fixed
OT_Biotelemetry_System_fixedOT_Biotelemetry_System_fixed
OT_Biotelemetry_System_fixed
Nikolay Bodunov
 
Biometrics For Security Systems
Biometrics For Security SystemsBiometrics For Security Systems
Biometrics For Security Systems
Suhas Deshpande
 
Wireless telemetry systems
Wireless telemetry systemsWireless telemetry systems
Wireless telemetry systems
Sneha Suluru
 
Biolelemetry1
Biolelemetry1Biolelemetry1
Biolelemetry1
Samuely
 
Power and Electrical Safety in Hospitals
Power and Electrical Safety in HospitalsPower and Electrical Safety in Hospitals
Power and Electrical Safety in Hospitals
Suhas Deshpande
 

Destacado (19)

Blood Flow Meter
Blood Flow MeterBlood Flow Meter
Blood Flow Meter
 
Estimation of body mass distribution using projective image
Estimation of body mass distribution using projective imageEstimation of body mass distribution using projective image
Estimation of body mass distribution using projective image
 
Trend presentation
Trend presentationTrend presentation
Trend presentation
 
OT_Biotelemetry_System_fixed
OT_Biotelemetry_System_fixedOT_Biotelemetry_System_fixed
OT_Biotelemetry_System_fixed
 
The Evolution of Wireless Monitoring in The Life Sciences and Review of Indus...
The Evolution of Wireless Monitoring in The Life Sciences and Review of Indus...The Evolution of Wireless Monitoring in The Life Sciences and Review of Indus...
The Evolution of Wireless Monitoring in The Life Sciences and Review of Indus...
 
Equilibrioception
Equilibrioception Equilibrioception
Equilibrioception
 
Biometrics For Security Systems
Biometrics For Security SystemsBiometrics For Security Systems
Biometrics For Security Systems
 
Telesurgery
TelesurgeryTelesurgery
Telesurgery
 
Telrmetry1
Telrmetry1Telrmetry1
Telrmetry1
 
Wireless telemetry systems
Wireless telemetry systemsWireless telemetry systems
Wireless telemetry systems
 
Electromagnetic flowmeters
Electromagnetic flowmetersElectromagnetic flowmeters
Electromagnetic flowmeters
 
Biolelemetry1
Biolelemetry1Biolelemetry1
Biolelemetry1
 
Power and Electrical Safety in Hospitals
Power and Electrical Safety in HospitalsPower and Electrical Safety in Hospitals
Power and Electrical Safety in Hospitals
 
Cardiac output 2
Cardiac output 2Cardiac output 2
Cardiac output 2
 
Ultrasonic flow measurement
Ultrasonic flow measurementUltrasonic flow measurement
Ultrasonic flow measurement
 
Topics
TopicsTopics
Topics
 
Cardiac output, blood flow, and blood pressure
Cardiac output, blood flow, and blood pressureCardiac output, blood flow, and blood pressure
Cardiac output, blood flow, and blood pressure
 
Data transmission and telemetry
Data transmission and telemetryData transmission and telemetry
Data transmission and telemetry
 
Flow Measurement
Flow MeasurementFlow Measurement
Flow Measurement
 

Similar a Perspectives in blood flow measurement

INVESTIGATING THE USE OF IMPEDANCE PLETHYSMOGRAPHY FOR DETECTING DECREASED BL...
INVESTIGATING THE USE OF IMPEDANCE PLETHYSMOGRAPHY FOR DETECTING DECREASED BL...INVESTIGATING THE USE OF IMPEDANCE PLETHYSMOGRAPHY FOR DETECTING DECREASED BL...
INVESTIGATING THE USE OF IMPEDANCE PLETHYSMOGRAPHY FOR DETECTING DECREASED BL...
Mirza Baig
 
AN ELECTRICAL IMPEDANCE TOMOGRAPHY SYSTEM FOR THYROID GLAND WITH A TINY ELECT...
AN ELECTRICAL IMPEDANCE TOMOGRAPHY SYSTEM FOR THYROID GLAND WITH A TINY ELECT...AN ELECTRICAL IMPEDANCE TOMOGRAPHY SYSTEM FOR THYROID GLAND WITH A TINY ELECT...
AN ELECTRICAL IMPEDANCE TOMOGRAPHY SYSTEM FOR THYROID GLAND WITH A TINY ELECT...
ijbesjournal
 

Similar a Perspectives in blood flow measurement (20)

A COMPARATIVE ANALYSIS OF THERMAL FLOW SENSING IN BIOMEDICAL APPLICATIONS
A COMPARATIVE ANALYSIS OF THERMAL FLOW SENSING IN BIOMEDICAL APPLICATIONSA COMPARATIVE ANALYSIS OF THERMAL FLOW SENSING IN BIOMEDICAL APPLICATIONS
A COMPARATIVE ANALYSIS OF THERMAL FLOW SENSING IN BIOMEDICAL APPLICATIONS
 
INVESTIGATING THE USE OF IMPEDANCE PLETHYSMOGRAPHY FOR DETECTING DECREASED BL...
INVESTIGATING THE USE OF IMPEDANCE PLETHYSMOGRAPHY FOR DETECTING DECREASED BL...INVESTIGATING THE USE OF IMPEDANCE PLETHYSMOGRAPHY FOR DETECTING DECREASED BL...
INVESTIGATING THE USE OF IMPEDANCE PLETHYSMOGRAPHY FOR DETECTING DECREASED BL...
 
Less computational approach to detect QRS complexes in ECG rhythms
Less computational approach to detect QRS complexes in ECG rhythmsLess computational approach to detect QRS complexes in ECG rhythms
Less computational approach to detect QRS complexes in ECG rhythms
 
emitechtalk.pptx
emitechtalk.pptxemitechtalk.pptx
emitechtalk.pptx
 
Flow cytometry
Flow cytometryFlow cytometry
Flow cytometry
 
Electrical Capacitance Volume Tomography
Electrical Capacitance Volume Tomography Electrical Capacitance Volume Tomography
Electrical Capacitance Volume Tomography
 
Desktop Tomography System using Planar ECT Device
Desktop Tomography System using Planar ECT DeviceDesktop Tomography System using Planar ECT Device
Desktop Tomography System using Planar ECT Device
 
40120140502011
4012014050201140120140502011
40120140502011
 
Flow rate
Flow rateFlow rate
Flow rate
 
Non invasive
Non invasiveNon invasive
Non invasive
 
Ultrasound Machine-A Revolution In Medical Imaging
Ultrasound Machine-A Revolution In Medical ImagingUltrasound Machine-A Revolution In Medical Imaging
Ultrasound Machine-A Revolution In Medical Imaging
 
A Simple and Robust Algorithm for the Detection of QRS Complexes
A Simple and Robust Algorithm for the Detection of QRS ComplexesA Simple and Robust Algorithm for the Detection of QRS Complexes
A Simple and Robust Algorithm for the Detection of QRS Complexes
 
A simple portable ecg monitor with iot
A simple portable ecg monitor with iotA simple portable ecg monitor with iot
A simple portable ecg monitor with iot
 
Exposure to energy audit equipment
Exposure to energy audit equipmentExposure to energy audit equipment
Exposure to energy audit equipment
 
Research Inventy : International Journal of Engineering and Science
Research Inventy : International Journal of Engineering and ScienceResearch Inventy : International Journal of Engineering and Science
Research Inventy : International Journal of Engineering and Science
 
Measurement of Pulse rate and SPo2 using Pulse Oximeter developed using LabVIEW
Measurement of Pulse rate and SPo2 using Pulse Oximeter developed using LabVIEWMeasurement of Pulse rate and SPo2 using Pulse Oximeter developed using LabVIEW
Measurement of Pulse rate and SPo2 using Pulse Oximeter developed using LabVIEW
 
Concentration measurements of bubbles in a water column using an optical tomo...
Concentration measurements of bubbles in a water column using an optical tomo...Concentration measurements of bubbles in a water column using an optical tomo...
Concentration measurements of bubbles in a water column using an optical tomo...
 
AN ELECTRICAL IMPEDANCE TOMOGRAPHY SYSTEM FOR THYROID GLAND WITH A TINY ELECT...
AN ELECTRICAL IMPEDANCE TOMOGRAPHY SYSTEM FOR THYROID GLAND WITH A TINY ELECT...AN ELECTRICAL IMPEDANCE TOMOGRAPHY SYSTEM FOR THYROID GLAND WITH A TINY ELECT...
AN ELECTRICAL IMPEDANCE TOMOGRAPHY SYSTEM FOR THYROID GLAND WITH A TINY ELECT...
 
ii unit
 ii unit ii unit
ii unit
 
A Wireless Physiological Monitoring System for Hyperbaric Oxygen Chamber
A Wireless Physiological Monitoring System for Hyperbaric Oxygen ChamberA Wireless Physiological Monitoring System for Hyperbaric Oxygen Chamber
A Wireless Physiological Monitoring System for Hyperbaric Oxygen Chamber
 

Perspectives in blood flow measurement

  • 1. Perspectives in Blood Flow Measurement By SuhasDeshpande Prof. Druzgalski EE 506 (Fall 2009) P3
  • 2. Outline About Blood Flow measurement Measurement techniques Electromagnetic methods Thermal methods Tracer dyes for measurement Ultrasound measurement Plethysmograph probe design Blood flow using thermal analysis Biotelemetry system References
  • 3. Blood flow measurement Provides vital info about the oxygen and nutrient supply to the various parts of body More difficult and complicated Most of the methods are invasive techniques Blood flow ranges from 0.45 m/s in the arteries to 0.1 m/s in veins Blood pressure measurement provides supplemental info which can be used
  • 4. Measurement techniques Electromagnetic methods Ultrasonic methods Thermal or using tracers methods Peripheral or indirect methods
  • 5. Electromagnetic methods Blood being a conductive fluid adheres to the Faraday’s law of electromagnetic induction Flow in a magnetic field results in a generation of electromagnetic force perpendicular to the magnetic field and flow Another technique used is that of Nuclear magnetic resonance [2]
  • 6. Thermal methods A thermistor or a hot wire anemometer can be implanted inside the body. Blood acts as a coolant for a thermistor . By different designs and calculations based on fluid mechanics and heat transfer flow of blood can be estimated [5]
  • 7. Tracer dyes for measurement The injection of some tracers helps in imaging and analyzing the blood flow Indocyanine Green is a dye which causes fluorescence and images the blood flow [6]
  • 8. Ultrasound measurement Blood flow measurement using implantable CMUT Array (Capacitive Micromachined Ultrasonic Transducer) CMUT consists of a SOC cmos electronics and transducer with cross section 40µm and shaft length 4-10mm [1]
  • 9. Experimental operation Ultrasonic pulse transmitted at frequency 1/T Echo is received at time t1 from position 1, t2 from positions 2 The velocity is calculated as c(t2-t1)/2T cosѳ [1]
  • 10.
  • 11. The IR diode (OP240D Optek) and Phototransistor (LPT85A Siemens) were attached to an ellipsoid plastic sheet 0.5 cm apart
  • 12. Probe attached to the skin using an ECG electrode[4]
  • 13. Blood flow using thermal analysis Arteriola and Venula connected by blood capillaries and AVA Blood flow to capillaries controlled by AVA Temperature regulation is based on blood flow in capillaries Heat transfer model analysis [7]
  • 14. Blood flow in fingertip using heat transfer analysis Non Contact sensor placed below consists of an IR detector (Thermopile and resistance temperature sensor) Sensitivity enhanced by a Ge lens Contact sensor consists of thermistor (0.25mm) to transmit the heat [7]
  • 15. Blood flow in fingertip using heat transfer analysis First step ambient temperature measured (25s) Second step finger placed on the sensor initial temperature of skin surface measured (35s) Third step measure the thermal release characteristics for correction in second step (15s) Blood flow coefficient is calculated by using the temperature gradient in step 2 [7]
  • 16. Biotelemetry system The implant module (B) consists of a doppler flow meter µcontroller and a RF link communicator The base station (A) consists of RF link µcontroller and D-A Converter [8]
  • 17.
  • 18. References [1] M. Wang, "The initial doppler blood flow measurement using an implantable CMUT array," Proceedings - IEEE Ultrasonics Symposium, pp. 2442, 2007. [2] C. Elkins, "Magnetic resonance velocimetry: Applications of magnetic resonance imaging in the measurement of fluid motion," Exp. Fluids, vol. 43, pp. 823, 2007. [3] K. Nagata. (2009), “Heat transfer analysis for peripheral blood flow measurement system”. Review of scientific instruments 80(6), [4] D. Zikic, "An improved reflective photoplethysmograph probe design for detection of an arterial blood flow," Journal of Medical Engineering Technology, vol. 32, pp. 23, 2008. [5] D. Wei, "Optimal design of a thermistor probe for surface measurement of cerebral blood flow," IEEE Trans. Biomed. Eng., vol. 37, pp. 1159, 1990. [6] M. Miwa, "ICG fluorescence imaging and its medical applications," Proceedings of SPIE--the International Society for Optical Engineering, vol. 7160, 2009. [7] K. Nagata. (2009), “Heat transfer analysis for peripheral blood flow measurement system”. Review of scientific instruments 80(6), [8] A. Gra¨ns, “A fully implantable multi-channel biotelemetry system for measurement of blood flow and temperature: a first evaluation in the green sturgeon,” Hydrobiologia (2009)