OBIETTIVO SU



Tomosintesi digitale della mammella
Digital Breast Tomosynthesis (DBT)

LUIGI PESCARINI, GISELLA GENNARO, ENRICA BALDAN, ELISA-
BETTA BEZZON, COSIMO DI MAGGIO, MANUELA LA GRASSA, ILA-               Operativa di Senologia dell’Istituto Oncologico Veneto (IOV)
RIA POLICO, ALESSANDRO PROIETTI, AIDA TOFFOLI                         in Padova.
  Istituto Oncologico Veneto, IRCCS Padova, via Gattamelata, 64 -
35128 Padova. Dipartimento di Scienze Oncologiche e Chirurgiche
                                                                      Protocollo clinico
dell’Università degli Studi di Padova
                                                                      Come introdotto, il limite principale della mammografia bi-
                                                                      dimensionale sia su pellicola (SFM = Screen Film Mammo-
                                                                      graphy) sia su supporto digitale (FFDM = Full Field Digital
Introduzione                                                          Mammography) è costituito dall’elevata densità mammaria del
                                                                      substrato anatomico derivante dalla sovrapposizione delle di-
                                                                      verse “strutture”.
L    a tomosintesi digitale della mammella (DBT, Digital Bre-
     ast Tomosynthesis) è una tecnica di imaging tridimensio-
nale che permette di ricostruire immagini volumetriche della
                                                                      E’ intuitivo che l’elevato spessore può produrre su una superfi-
                                                                      cie un “effetto di elevata densità”, anche se tra le strutture ghian-
mammella a partire da un numero finito di proiezioni bidi-            dolari esistono piani di grasso più o meno estesi. In presenza di
mensionali a bassa dose, ottenute con angolazioni diverse del         tali condizioni consegue la mancata o scadente visualizzazione
tubo radiogeno. Il principio radiogeometrico della tomosinte-         e la “non percezione” delle lesioni espansive (masse e distor-
si è simile a quello applicato nella vecchia tecnica stratigrafica    sioni) e delle calcificazioni, con errore significativamente più
[1], con la differenza fondamentale che, mentre la stratigrafia       rilevante per le lesioni espansive in relazione al minor contrasto
richiedeva l’acquisizione di esposizioni multiple per ciascuno        intrinseco.
strato che si voleva “mettere a fuoco”, la tomosintesi digitale       In virtù di una “nitida rappresentazione” in assenza di sovrap-
permette di ricostruire un numero arbitrario di piani a parti-        posizioni, la tomosintesi è in grado di “rendere visibili” e/o me-
re dalla stessa sequenza di proiezioni bidimensionali [2]. Ciò è      glio analizzabili nelle forma, nei contorni, nella disposizione e
reso possibile dalla separazione tra il processo di acquisizione e    nel numero le lesioni “non rappresentate o mal rappresentate”
quello di visualizzazione consentita dall’impiego di rivelatori       dalla mammografia.
digitali diretti per cui le stesse proiezioni grezze possono essere   Lo studio in corso presso lo IOV di Padova prevede il con-
processate per ricostruire piani diversi.                             fronto tra le prestazioni cliniche della mammografia digitale
La ricostruzione volumetrica, in linea di principio, consente di      (FFDM) nelle due proiezioni standard (CC cranio-caudale
superare uno dei limiti principali dell’imaging bidimensionale,       e MLO medio-laterale-obliqua) e quelle della tomosintesi
ovvero il mascheramento di lesioni (nel caso della mammella,          (DBT) in un’unica proiezione obliqua, rispettando così il vin-
masse, microcalcificazioni, ecc.), causato dalla sovrapposizione      coli di dose. Infatti, fino a quando non sarà dimostrata almeno
di strutture normali; quindi l’opportunità di dissociare piani        la non-inferiorità clinica della tomosintesi rispetto alla mam-
diversi da parte della tomosintesi fa ritenere possibile una ridu-    mografia, non è ragionevole aumentare la dose per l’esame di
zione del numero di falsi negativi e di falsi positivi dovuti alla    tomosintesi; per questo motivo, la dose per l’acquisizione della
sovrapposizione [3-5].                                                sequenza di proiezioni di tomosintesi è stata calcolata in modo
Di seguito vengono illustrate le prime esperienze di tomosinte-       da non superare quella di una mammografia tradizionale in due
si digitale della mammella sulle immagini prodotte con un pro-        proiezioni (Figura 1).
totipo sviluppato da GE Healthcare ed installato presso l’Unità

                                                                                          N. 55 - 2008   ATTUALITÀ IN SENOLOGIA         25
OBIETTIVO SU

                                                                                   Le sequenze di tomosintesi usate nello studio
                                      FIGURA 1                                     prevedono l’acquisizione di 15 proiezioni a
                                                                                   bassa dose, comprese in un arco di ± 20° ri-
                                                                                   spetto alla perpendicolare al rivelatore.
                                                                                   La dose erogata per una sequenza di tomosin-
                                                                                   tesi è stata calcolata entro un valore limite che
                                                                                   non superi la dose di due proiezioni mammo-
                                                                                   grafiche standard, così che in fase applicativa
                                                                                   la dose di fatto viene frazionata in modo che
                                                                                   per ciascuna proiezione sia 1/15 del totale
                                                                                   [7].
                                                                                   Le proiezioni bidimensionali sono processate
                                                                                   tramite un algoritmo di tipo SART (Simulta-
                                                                                   neous Algebric Reconstruction Technique)
                                                                                   per ricostruire il volume della mammella,
                                                                                   campionato secondo strati contigui di 1 mil-
                                                                                   limetro [8].


                                                                                   Approccio clinico

                                                                                    Alcune osservazioni preliminari sulle possibi-
                                                                                    lità delle tecniche e sul metodo di analisi delle
                                                                                    immagini mettono in evidenza che alla “non
Lo studio è stato condotto dopo aver ottenuto il parere favo-    sovrapposizione della DBT” si contrappone un potenziale “be-
revole del Comitato Etico, l’approvazione del Ministero della    neficio geometrico” connesso alle due proiezioni della FFDM.
Salute per l’uso di un “prototipo” e con il consenso informato   Inoltre se è vero che la DBT può costruire un “numero arbitra-
delle Pazienti eligibili.
                                                                            FIGURA 2
Apparecchiatura

Il sistema per tomosintesi si basa sulla
piattaforma del mammografo digitale
commerciale GE Senographe DS, con
rivelatore flat panel di area 19 x 23
cm2, che impiega Ioduro di Cesio in
ingresso per la conversione dei raggi-X
in luce, con dimensione nominale del
pixel di 100 μm2. La movimentazione
delgantry è stata modificata in modo
da permettere la rotazione del tubo
radiogeno attorno all’asse perpendi-
colare al rivelatore; inoltre il sistema di
compressione è stato reso separato dal
gantry per far sì che il sistema “mam-
mella compressa-rivelatore” rimanga
fermo durante la rotazione del tubo. In
figura 2 è schematizzata l’acquisizione
di un esame di tomosintesi [6].

26      ATTUALITÀ IN SENOLOGIA      2008 - N. 55
OBIETTIVO SU

rio” di sezioni da un numero finito di proiezioni, l’accuratez-         le “performance” cliniche tramite analisi ROC. Tuttavia, per
za d’informazione dipende dall’intervallo di campionamento              mantenere un approccio più clinico, anziché far riferimento
delle fette. Il campionamento per ogni millimetro di spessore           alle curve frutto dell’analisi statistica, nel seguito seguiremo
delle sezioni (slice), ragionevole tra accuratezza e necessità di       uno schema di valutazione delle immagini mammografiche e di
limitare il numero di immagini nell’ottica dell’ottimizzazione,         tomosintesi secondo i 5 punti proposti da Kopans:
può essere un limite a scopo diagnostico, ad esempio per la “va-        1) Riconoscimento: trovare la lesione
lutazione d’insieme” di piccoli dettagli come le microcalcifica-        2) Presenza: è reale?
zioni (Figura 3A).                                                      3) Localizzazione: dove si trova ?
                                                                        4) Interpretazione: cosa è ?
                                                                        5) Decisione: “cosa farne”
                                                                        Di seguito sono sinteticamente riferite le “impressioni” più in-
                                                                        teressanti, mentre i primi risultati complessivi sui casi studiati
                                                                        sono riportati successivamente.

                                                                        1) Riconoscimento
                                                                        Premesso che la densità mammaria può pregiudicare la rileva-
                                                                        bilità delle lesioni, condizione non superabile per entrambe le
                                                                        tecniche (FFDM e SFM), la DBT consente l’individuazione
                                                                        di un maggior numero di lesioni espansive, con una miglior
                                                                        analisi della forma e dei contorni delle masse e dell’ampiezza e
                                                                        dell’architettura delle distorsioni (Figure 4 A e B).


                                                              Fig. 3A

Pertanto sezioni più spesse (slab) sono a volte più vantaggiose:
le immagini slab sono ottenute per compattamento di più slice
(Figura 3B).


 FFDM                                                         DBT
 sinistra                                                  sinistra
 MLO                                                          MLO




                                                                                                                                    Fig. 4A




                                                    slab [8-12]


                                                              Fig. 3B
Su tali premesse è stato preso in considerazione un primo grup-
po di 50 casi sottoponendoli alla valutazione di tre Radiologi
esperti con metodologia “blinded and independent”, per rica-
vare, dal confronto fra le due tecniche, le prime valutazioni sul-
                                                                                                                                    Fig. 4B
                                                                                            N. 55 - 2008   ATTUALITÀ IN SENOLOGIA      27
OBIETTIVO SU

Per le calcificazioni al momento non sono emerse significative          FFDM                                                      DBT
differenze su questo punto. Le immagini slab costituiscono una          RMLO                                                     RMLO

variante tecnica di riferimento nell’analisi della distribuzione,
con particolare riguardo alla “configurazione duttale” (Figura 5)




                                                                                                                                  Fig. 6B

                                                                       3) Localizzazione
                                                                       Le due proiezioni mammografiche ortogonali sono il riferi-
                                                                       mento standard per l’attribuzione del quadrante delle lesioni;
                                                              Fig. 5   è noto anche che le proiezioni CC e MLO non sono spesso
                                                                       sufficienti a definire l’appartenenza anatomo-topografica.
L’ingrandimento elettronico risulta di aiuto all’analisi sia nella
                                                                       Analogamente la tomosintesi ottenuta con mammella in po-
FFDM che nella DBT.
                                                                       sizione obliqua con le attuali ricostruzioni non è in grado di
                                                                       svolgere questo compito. E’ lecito pensare che ulteriori sviluppi
2) Presenza
                                                                       informatici in un prossimo futuro possano raggiungere anche
Il rumore anatomico (densità e sovrapposizione) in mammo-
                                                                       questo obiettivo.
grafia può produrre dei “falsi positivi”. L’analisi mediante sezio-
ni contigue elimina immagini apparentemente sospette soprat-           Attualmente la DBT è solo in grado di stabilire correttamente
tutto nei casi di sovrapposizione di strie con evidenza di aspetti     la sede delle lesioni, tenuto conto delle modalità di posiziona-
distorsivi (Figura 6A). In una prospettiva opposta la DBT può          mento, definendo correttamente la profondità rispetto ai piani
confermare con significatività “segni deboli” di presenza di le-       cutanei di appoggio e di compressione.
sione rivelati dalla mammografia (Figura 6B).                          Ne traggono beneficio in particolare le immagini di lesioni si-
                                                                       tuate in corrispondenza di questi (Figura 7).


                     Case # 149 - FFDM
 FFDM                                                        DBT
 LCC                                                        LMLO




                                                                                                                                   Fig. 7
                                                             Fig. 6A



28      ATTUALITÀ IN SENOLOGIA     2008 - N. 55
OBIETTIVO SU




4) Interpretazione                                                   Primi risultati e considerazioni
Dai punti precedenti appare evidente in generale un maggior
contributo da parte della DBT riguardo la semeiotica delle le-       Vale la pena riassumere in queste prime applicazioni le
sioni mammarie, che sembra ottimale per le distorsioni, buono        impressioni comuni ai tre Radiologi che concordano
per le masse, ma di minore impatto per le calcificazioni.            nell’affermare che la DBT permette un sostanziale mi-
Per queste appare importante un uso adeguato delle slab che          glioramento nel rilevamento e nell’analisi delle lesioni,
consentono una visione d’insieme.                                    influendo tanto sul convincimento della loro presenza
Come noto, solo in caso di rilievi molto significativi (masse a      quanto nella certezza della loro assenza rispetto alla
contorni sfrangiati, calcificazioni irregolari e racemose) gli ap-   FFDM.
porti della sola immagine radiologica possono essere definitivi      A ciò si associa peraltro un tempo di lettura discreta-
per stabilire con accuratezza una diagnosi di natura; di fatto       mente più lungo, correlato al numero d’immagini e al
è più spesso “orientativa” a stabilire l’indicazione agli apporti    contenuto informativo significativamente più ampio
complementari dell’ecografia, delle indagini agobioptiche di         delle stesse da esaminare con attenzione sia in relazione
secondo livello e della Risonanza Magnetica con mezzo di con-        all’analisi di strato quanto di ricomposizione spaziale
trasto.                                                              da parte dell’operatore medico.
In sintesi l’apporto della DBT diventa rilevante quando intro-       La DBT non è in grado allo stato attuale di definire
duca aspetti correlati ad una migliorata analisi che orientano       con sufficiente accuratezza il quadrante nel quale la le-
con convinzione verso la benignità o la malignità, apporto che       sione è situata.
peraltro deve essere contestualizzato nelle procedure già conva-     I primi risultati delle prestazioni di ciascun Radiologo
lidate in Senologia.                                                 relativi ai 50 casi esaminati sono stati accolti ed ela-
                                                                     borati con il metodo ROC. La curva relativa al rileva-
5) Decisione                                                         mento delle lesioni, sul significato statistico della quale
Il punto cruciale della decisione è frutto di tutte le informazio-   peraltro non è possibile esprimere un giudizio definiti-
ni raccolte, non solo delle immagini.                                vo, è per tutti i Radiologi indicativa di “performance”
E’ chiaro che la DBT è particolarmente significativa quando          migliori, comunque non peggiori, per la DBT rispetto
elimina immagini potenzialmente sospette, come illustrato in         alla FFDM.
Fig 6A o invece riconosce lesioni sospette non visibili o insuffi-   E’ infine da segnalare che la tomosintesi di fatto non
cientemente documentate per rumore anatomico alla mammo-             introduce elementi di “semeiotica nuovi”, ma richiede
grafia (Fig 6B).                                                     un periodo di adattamento all’immagine che nella no-
La decisione, analizzati tali contributi, si coniuga inevitabil-     stra esperienza è stato piuttosto breve.
mente con l’interpretazione e procede di conseguenza.




                                                                                      N. 55 - 2008   ATTUALITÀ IN SENOLOGIA   29
OBIETTIVO SU




Conclusioni                                                  imaging. Radiology 205, 399-406 (1997)
                                                             3.Bochud, F.O., Valley, J.F., Verdun, F.R.: Estimation
Le prime esperienze cliniche sembrano indicare un im-        of the noisy component of anatomical background.
                                                             Med. Phys. 26, 1365-1370 (1999)
patto favorevole della DBT nel rilevamento delle lesio-
                                                             4.Burgess, A.E., Jacobson, F.L., Judy, P   .F.: On the
ni mammarie.
                                                             difficulty of detecting tumors in mammograms. In:
Il rilevamento tomografico consente di fatto di apprez-
                                                             Proceedings of the 17th International Conference on
zare isolatamente le lesioni che più “soffrono” della so-    Information Processing, pp. 1-11. Springer-Verlag,
vrapposizione e quindi della confusione dei piani alla       London (2001)
mammografia standard.                                        5.Kontos, D., Bakic, P.R., Maidment, A.D.A.: Analysis
Se anche i primi risultati comparativi, per ora esigui,      of parenchymal texture properties in breast tomo-
saranno confermati, la DBT potrà inserirsi a pieno ti-       synthesis images. In: Giger, M.L., Karssemeijer, N.
tolo nel novero delle tecniche idonee allo studio della      (eds.) Proceedings of SPIE Medical Imaging 2007:
mammella finalizzato al rilevamento di lesioni tanto         Computer-Aided Diagnosis, vol.6514, pp. 651417
nella diagnostica senologica quanto nello screening.         (2007)
Gli attesi sviluppi informatici, in particolare i disposi-   6.Gennaro G, di Maggio C, Baldan E, Bezzon E:
tivi CAD dedicati, e le potenzialità connesse all’impie-     Mammografia digitale 3D. In: “Mammografia Digi-
go dei mezzi di contrasto sono da considerare ulteriori      tale”, Poletto Editore 2007.
                                                             7.Wu, T., Liu, B., Moore, R., Kopans, D.: Optimal ac-
promettenti innovazioni che la tomosintesi dovrebbe
                                                             quisition techniques for digital breast tomosynthesis
proporre ai Radiologi in un futuro non lontano.
                                                             screening. In: Flynn, M.J., Hsieh, J. (eds.) Proce-
                                                             edings of SPIE Medical Imaging 2006: Physics of
                                                             Medical Imaging, vol. 6142, pp. 61425E (2006)
Bibliografia                                                  8.Zhang, Y., Chan, H.P Sahiner, B., Wei, J., Goodsitt,
                                                                                    .,
                                                             M.M., Hadjiiski, L.M., Ge, J., Zhou, C.: A comparati-
1.A Vallebona: Una modalità di tecnica per la disso-         ve study of limited-angle cone-beam reconstruction
ciazione radiografica. Radiologia Medica 17, 1090-            methods for breast tomosynthesis. Med. Phys. 33,
1097 (1931).                                                 3781-3795 (2006).
2.Niklason, L.T., Christian, B.T., Niklason, L.E., Ko-
pans, D.B., Castleberry, D.E., Opsahl-Ong, B.H., Lan-        Figure
dberg, C.E., Slanetz, P.J., Giardino, A.A., Moore, R.,
Albagli, D., DeJoule, M., Fitzgerald, P   .F., Fobare,       Fig. 1 - Le curve indicano la dose ghiandolare me-
D.F., Gianbattista, B.W., Kwasnick, R.F., Jianqiang,         dia (AGD = Average Glandular Dose) per la mam-
L., Lubowski, S.J., Possin, G.E., Richotte, J.F., Wei,       mografia tradizionale (SFM) e la tomosintesi (DBT)
C.Y., Weirth, R.F.: Digital tomosynthesis in breast          in funzione dello spessore mammario (breast thick-



30    ATTUALITÀ IN SENOLOGIA   2008 - N. 55
OBIETTIVO SU




ness): esse appaiono quasi sovrapponibili.                 storsiva di 13 mm causata da sovrapposizione di
Fig. 2 - Principio di funzionamento della tomosin-         strutture. La DBT conferma l’assenza della lesione.
tesi. In fase di acquisizione (sinistra) il tubo ruota     Fig. 6 - B) Nella FFDM è con difficoltà riconoscibile
attorno all’asse perpendicolare al rivelatore e, in un     una massa a contorni sfumati di 5 mm. Con DBT ben
numero finito di posizioni prestabilite, il sistema ac-     riconoscibile nella stessa sede massa isolata a con-
quisisce proiezioni bidimensionali a bassa dose che        torni sfrangiati (carcinoma tubulare).
costituiscono i dati grezzi. Le immagini bidimensio-
nali vengono poi trasferite al sistema di ricostruzio-     Fig. 7 - La massa è riconoscibile con FFDM; con DBT
ne che, applicando un algoritmo di “shift-and-add”,        si localizza nei piani superficiali e appare con net-
ovvero di riallineamento delle strutture e di somma        tezza delimitata da trasparenza attribuibile all’aria
del segnale, ricostruisce il volume della mammella         (nevo cutaneo).
che viene poi campionato ad intervalli di 1 mm, per-
mettendo così di mettere a fuoco strutture che stan-
no su piani diversi.

Fig. 3 - A) Sequenza di slice da 1 mm relativo ad
un caso di microcalcificazioni (carcinoma duttale in
situ).
Fig. 3 - B) Confronto tra FFDM e DBT con immagini
slab ottenute per compattamento di 5 slice.

Fig. 4 - A) Confronto tra FFDM e DBT: proiezione
LMLO (Left Medio-Lateral-Oblique). Nella FFDM
non sono riconoscibili aspetti riferibili a masse. Nella
DBT è evidente massa allungata con diametro massi-
mo di 10 mm a contorni netti (lesione benigna).
Fig. 4 - B) Anche in questo caso solo nella DBT è
riconoscibile massa a contorni sfrangiati di 7 mm
(carcinoma duttale infiltrante).

Fig. 5 - Confronto tra FFDM e DBT: le informazioni
riguardo le calcificazioni nella forma, numero e di-
stribuzione sono analoghe per entrambe le tecniche.

Fig. 6 - A) La FFDM in CC evidenzia immagine di-



                                                                           N. 55 - 2008   ATTUALITÀ IN SENOLOGIA   31

Más contenido relacionado

La actualidad más candente

Presentation1.pptx, radilogical imaging of ovarian lesions.
Presentation1.pptx, radilogical imaging of ovarian lesions.Presentation1.pptx, radilogical imaging of ovarian lesions.
Presentation1.pptx, radilogical imaging of ovarian lesions.Abdellah Nazeer
 
Hepatocellular carcinoma—role of interventional radiologist Dr. Muhammad Bin ...
Hepatocellular carcinoma—role of interventional radiologist Dr. Muhammad Bin ...Hepatocellular carcinoma—role of interventional radiologist Dr. Muhammad Bin ...
Hepatocellular carcinoma—role of interventional radiologist Dr. Muhammad Bin ...Dr. Muhammad Bin Zulfiqar
 
Cystic masses of the breast by xiu
Cystic masses of the breast by xiuCystic masses of the breast by xiu
Cystic masses of the breast by xiuXiu Srithammasit
 
Ultrasound breast mass
Ultrasound breast massUltrasound breast mass
Ultrasound breast massREKHAKHARE
 
Malignant ovarian tumors DR NN CHAVAN 19102023 .pptx
Malignant ovarian tumors DR NN CHAVAN 19102023 .pptxMalignant ovarian tumors DR NN CHAVAN 19102023 .pptx
Malignant ovarian tumors DR NN CHAVAN 19102023 .pptxNiranjan Chavan
 
Liver lesions
Liver lesionsLiver lesions
Liver lesionsairwave12
 
Presentation1, radiological imaging of endometrial carcinoma.
Presentation1, radiological imaging of endometrial carcinoma.Presentation1, radiological imaging of endometrial carcinoma.
Presentation1, radiological imaging of endometrial carcinoma.Abdellah Nazeer
 
Diagnostic Imaging of Cholangiocarcinoma
Diagnostic Imaging of CholangiocarcinomaDiagnostic Imaging of Cholangiocarcinoma
Diagnostic Imaging of CholangiocarcinomaMohamed M.A. Zaitoun
 
Hepatic peliosis
Hepatic peliosisHepatic peliosis
Hepatic peliosismacshrestha
 
Recent Update on Management of Ulcerative Colitis
Recent Update on Management of Ulcerative ColitisRecent Update on Management of Ulcerative Colitis
Recent Update on Management of Ulcerative ColitisDr Amit Dangi
 
Gastric Outlet Obstruction
Gastric Outlet ObstructionGastric Outlet Obstruction
Gastric Outlet ObstructionDalitso Phiri
 
Hilar Cholangiocarcinoma
Hilar Cholangiocarcinoma Hilar Cholangiocarcinoma
Hilar Cholangiocarcinoma Dr Harsh Shah
 
Radiological anatomy of liver segments
Radiological anatomy of liver segmentsRadiological anatomy of liver segments
Radiological anatomy of liver segmentsTarun Goyal
 
Gastroesophageal Junction Carcinoma
Gastroesophageal  Junction CarcinomaGastroesophageal  Junction Carcinoma
Gastroesophageal Junction CarcinomaDr.Bhavin Vadodariya
 

La actualidad más candente (20)

Presentation1.pptx, radilogical imaging of ovarian lesions.
Presentation1.pptx, radilogical imaging of ovarian lesions.Presentation1.pptx, radilogical imaging of ovarian lesions.
Presentation1.pptx, radilogical imaging of ovarian lesions.
 
Hepatocellular carcinoma—role of interventional radiologist Dr. Muhammad Bin ...
Hepatocellular carcinoma—role of interventional radiologist Dr. Muhammad Bin ...Hepatocellular carcinoma—role of interventional radiologist Dr. Muhammad Bin ...
Hepatocellular carcinoma—role of interventional radiologist Dr. Muhammad Bin ...
 
Cystic masses of the breast by xiu
Cystic masses of the breast by xiuCystic masses of the breast by xiu
Cystic masses of the breast by xiu
 
Ultrasound breast mass
Ultrasound breast massUltrasound breast mass
Ultrasound breast mass
 
Malignant ovarian tumors DR NN CHAVAN 19102023 .pptx
Malignant ovarian tumors DR NN CHAVAN 19102023 .pptxMalignant ovarian tumors DR NN CHAVAN 19102023 .pptx
Malignant ovarian tumors DR NN CHAVAN 19102023 .pptx
 
Liver lesions
Liver lesionsLiver lesions
Liver lesions
 
Choledochal cyst
Choledochal cystCholedochal cyst
Choledochal cyst
 
Presentation1, radiological imaging of endometrial carcinoma.
Presentation1, radiological imaging of endometrial carcinoma.Presentation1, radiological imaging of endometrial carcinoma.
Presentation1, radiological imaging of endometrial carcinoma.
 
Diagnostic Imaging of Cholangiocarcinoma
Diagnostic Imaging of CholangiocarcinomaDiagnostic Imaging of Cholangiocarcinoma
Diagnostic Imaging of Cholangiocarcinoma
 
Hepatic peliosis
Hepatic peliosisHepatic peliosis
Hepatic peliosis
 
Recent Update on Management of Ulcerative Colitis
Recent Update on Management of Ulcerative ColitisRecent Update on Management of Ulcerative Colitis
Recent Update on Management of Ulcerative Colitis
 
Gastric Outlet Obstruction
Gastric Outlet ObstructionGastric Outlet Obstruction
Gastric Outlet Obstruction
 
Hilar Cholangiocarcinoma
Hilar Cholangiocarcinoma Hilar Cholangiocarcinoma
Hilar Cholangiocarcinoma
 
CT Abdomen Case based presentation
CT Abdomen Case based presentationCT Abdomen Case based presentation
CT Abdomen Case based presentation
 
Prostate carcinoma- imaging
Prostate  carcinoma- imagingProstate  carcinoma- imaging
Prostate carcinoma- imaging
 
Radiological anatomy of liver segments
Radiological anatomy of liver segmentsRadiological anatomy of liver segments
Radiological anatomy of liver segments
 
67 thickened gallbladder wall (
67 thickened gallbladder wall (67 thickened gallbladder wall (
67 thickened gallbladder wall (
 
Stomach radiology
Stomach radiology Stomach radiology
Stomach radiology
 
MGB Tips and Ticks
MGB Tips and Ticks MGB Tips and Ticks
MGB Tips and Ticks
 
Gastroesophageal Junction Carcinoma
Gastroesophageal  Junction CarcinomaGastroesophageal  Junction Carcinoma
Gastroesophageal Junction Carcinoma
 

Similar a Tomosintesi Mammella

Modulo A - Cone Beam: risorsa diagnostica e vincoli normativi
Modulo A - Cone Beam: risorsa diagnostica e vincoli normativiModulo A - Cone Beam: risorsa diagnostica e vincoli normativi
Modulo A - Cone Beam: risorsa diagnostica e vincoli normativiaccademiatn
 
Cone beam - Modulo C
Cone beam - Modulo CCone beam - Modulo C
Cone beam - Modulo Caccademiatn
 
Operazioni allo sterno con l’aiuto delle immagini 3 d 2010-11-05
Operazioni allo sterno con l’aiuto delle immagini 3 d   2010-11-05Operazioni allo sterno con l’aiuto delle immagini 3 d   2010-11-05
Operazioni allo sterno con l’aiuto delle immagini 3 d 2010-11-05Ionela
 
Risonanza magnetica multiparametrica della prostata(tecnica d’esame)
Risonanza  magnetica multiparametrica della prostata(tecnica d’esame)Risonanza  magnetica multiparametrica della prostata(tecnica d’esame)
Risonanza magnetica multiparametrica della prostata(tecnica d’esame)giuseppe scappatura
 
Melanoma dalla clinica alla diagnosi computerizzata
Melanoma dalla clinica alla diagnosi computerizzataMelanoma dalla clinica alla diagnosi computerizzata
Melanoma dalla clinica alla diagnosi computerizzataMerqurio
 
Melanoma dalla clinica alla diagnosi computerizzata
Melanoma dalla clinica alla diagnosi computerizzataMelanoma dalla clinica alla diagnosi computerizzata
Melanoma dalla clinica alla diagnosi computerizzataMerqurio
 
Imaging digitale applicato al medicale
Imaging digitale applicato al medicaleImaging digitale applicato al medicale
Imaging digitale applicato al medicalePaolo Gibellini
 
Metodo e apparato per PET in adroterapia
Metodo e apparato per PET in adroterapiaMetodo e apparato per PET in adroterapia
Metodo e apparato per PET in adroterapiaToscana Open Research
 
tecnologia in ambito della senologia
tecnologia in ambito della senologiatecnologia in ambito della senologia
tecnologia in ambito della senologiaAntonio Mangiardi
 
Modulo D - Cone Beam: risorsa diagnostica e vincoli normativi
Modulo D - Cone Beam: risorsa diagnostica e vincoli normativiModulo D - Cone Beam: risorsa diagnostica e vincoli normativi
Modulo D - Cone Beam: risorsa diagnostica e vincoli normativiaccademiatn
 
Magic 5 - Applicazioni della Fisica nel campo della Medicina
Magic 5 - Applicazioni della Fisica nel campo della MedicinaMagic 5 - Applicazioni della Fisica nel campo della Medicina
Magic 5 - Applicazioni della Fisica nel campo della Medicinasilviom
 
Confronto con test di wilcoxon
Confronto con test di wilcoxonConfronto con test di wilcoxon
Confronto con test di wilcoxonElekta
 
DEHCA: UN PROGETTO PER LA PREVENZIONE DEL TUMORE ALLA MAMMELLA ANCHE PER LE G...
DEHCA: UN PROGETTO PER LA PREVENZIONE DEL TUMORE ALLA MAMMELLA ANCHE PER LE G...DEHCA: UN PROGETTO PER LA PREVENZIONE DEL TUMORE ALLA MAMMELLA ANCHE PER LE G...
DEHCA: UN PROGETTO PER LA PREVENZIONE DEL TUMORE ALLA MAMMELLA ANCHE PER LE G...convegnonazionaleaiic
 

Similar a Tomosintesi Mammella (20)

Modulo A - Cone Beam: risorsa diagnostica e vincoli normativi
Modulo A - Cone Beam: risorsa diagnostica e vincoli normativiModulo A - Cone Beam: risorsa diagnostica e vincoli normativi
Modulo A - Cone Beam: risorsa diagnostica e vincoli normativi
 
Cone beam - Modulo C
Cone beam - Modulo CCone beam - Modulo C
Cone beam - Modulo C
 
Colonscopia prevenzione e cura
Colonscopia prevenzione e curaColonscopia prevenzione e cura
Colonscopia prevenzione e cura
 
3 d instrumentarium
3 d instrumentarium3 d instrumentarium
3 d instrumentarium
 
Theragnostics_18.10.12
Theragnostics_18.10.12Theragnostics_18.10.12
Theragnostics_18.10.12
 
Operazioni allo sterno con l’aiuto delle immagini 3 d 2010-11-05
Operazioni allo sterno con l’aiuto delle immagini 3 d   2010-11-05Operazioni allo sterno con l’aiuto delle immagini 3 d   2010-11-05
Operazioni allo sterno con l’aiuto delle immagini 3 d 2010-11-05
 
Risonanza magnetica multiparametrica della prostata(tecnica d’esame)
Risonanza  magnetica multiparametrica della prostata(tecnica d’esame)Risonanza  magnetica multiparametrica della prostata(tecnica d’esame)
Risonanza magnetica multiparametrica della prostata(tecnica d’esame)
 
Melanoma dalla clinica alla diagnosi computerizzata
Melanoma dalla clinica alla diagnosi computerizzataMelanoma dalla clinica alla diagnosi computerizzata
Melanoma dalla clinica alla diagnosi computerizzata
 
Melanoma dalla clinica alla diagnosi computerizzata
Melanoma dalla clinica alla diagnosi computerizzataMelanoma dalla clinica alla diagnosi computerizzata
Melanoma dalla clinica alla diagnosi computerizzata
 
Imaging digitale applicato al medicale
Imaging digitale applicato al medicaleImaging digitale applicato al medicale
Imaging digitale applicato al medicale
 
Metodo e apparato per PET in adroterapia
Metodo e apparato per PET in adroterapiaMetodo e apparato per PET in adroterapia
Metodo e apparato per PET in adroterapia
 
tecnologia in ambito della senologia
tecnologia in ambito della senologiatecnologia in ambito della senologia
tecnologia in ambito della senologia
 
PROGETTO VIRTUAL CLONE
PROGETTO VIRTUAL CLONEPROGETTO VIRTUAL CLONE
PROGETTO VIRTUAL CLONE
 
Modulo D - Cone Beam: risorsa diagnostica e vincoli normativi
Modulo D - Cone Beam: risorsa diagnostica e vincoli normativiModulo D - Cone Beam: risorsa diagnostica e vincoli normativi
Modulo D - Cone Beam: risorsa diagnostica e vincoli normativi
 
Magic 5 - Applicazioni della Fisica nel campo della Medicina
Magic 5 - Applicazioni della Fisica nel campo della MedicinaMagic 5 - Applicazioni della Fisica nel campo della Medicina
Magic 5 - Applicazioni della Fisica nel campo della Medicina
 
Mammografia 3d con tomosintesi
Mammografia 3d con tomosintesiMammografia 3d con tomosintesi
Mammografia 3d con tomosintesi
 
Confronto con test di wilcoxon
Confronto con test di wilcoxonConfronto con test di wilcoxon
Confronto con test di wilcoxon
 
Endoprotesi e trattamento
Endoprotesi e trattamentoEndoprotesi e trattamento
Endoprotesi e trattamento
 
Endoprotesi e trattamento
Endoprotesi e trattamentoEndoprotesi e trattamento
Endoprotesi e trattamento
 
DEHCA: UN PROGETTO PER LA PREVENZIONE DEL TUMORE ALLA MAMMELLA ANCHE PER LE G...
DEHCA: UN PROGETTO PER LA PREVENZIONE DEL TUMORE ALLA MAMMELLA ANCHE PER LE G...DEHCA: UN PROGETTO PER LA PREVENZIONE DEL TUMORE ALLA MAMMELLA ANCHE PER LE G...
DEHCA: UN PROGETTO PER LA PREVENZIONE DEL TUMORE ALLA MAMMELLA ANCHE PER LE G...
 

Tomosintesi Mammella

  • 1. OBIETTIVO SU Tomosintesi digitale della mammella Digital Breast Tomosynthesis (DBT) LUIGI PESCARINI, GISELLA GENNARO, ENRICA BALDAN, ELISA- BETTA BEZZON, COSIMO DI MAGGIO, MANUELA LA GRASSA, ILA- Operativa di Senologia dell’Istituto Oncologico Veneto (IOV) RIA POLICO, ALESSANDRO PROIETTI, AIDA TOFFOLI in Padova. Istituto Oncologico Veneto, IRCCS Padova, via Gattamelata, 64 - 35128 Padova. Dipartimento di Scienze Oncologiche e Chirurgiche Protocollo clinico dell’Università degli Studi di Padova Come introdotto, il limite principale della mammografia bi- dimensionale sia su pellicola (SFM = Screen Film Mammo- graphy) sia su supporto digitale (FFDM = Full Field Digital Introduzione Mammography) è costituito dall’elevata densità mammaria del substrato anatomico derivante dalla sovrapposizione delle di- verse “strutture”. L a tomosintesi digitale della mammella (DBT, Digital Bre- ast Tomosynthesis) è una tecnica di imaging tridimensio- nale che permette di ricostruire immagini volumetriche della E’ intuitivo che l’elevato spessore può produrre su una superfi- cie un “effetto di elevata densità”, anche se tra le strutture ghian- mammella a partire da un numero finito di proiezioni bidi- dolari esistono piani di grasso più o meno estesi. In presenza di mensionali a bassa dose, ottenute con angolazioni diverse del tali condizioni consegue la mancata o scadente visualizzazione tubo radiogeno. Il principio radiogeometrico della tomosinte- e la “non percezione” delle lesioni espansive (masse e distor- si è simile a quello applicato nella vecchia tecnica stratigrafica sioni) e delle calcificazioni, con errore significativamente più [1], con la differenza fondamentale che, mentre la stratigrafia rilevante per le lesioni espansive in relazione al minor contrasto richiedeva l’acquisizione di esposizioni multiple per ciascuno intrinseco. strato che si voleva “mettere a fuoco”, la tomosintesi digitale In virtù di una “nitida rappresentazione” in assenza di sovrap- permette di ricostruire un numero arbitrario di piani a parti- posizioni, la tomosintesi è in grado di “rendere visibili” e/o me- re dalla stessa sequenza di proiezioni bidimensionali [2]. Ciò è glio analizzabili nelle forma, nei contorni, nella disposizione e reso possibile dalla separazione tra il processo di acquisizione e nel numero le lesioni “non rappresentate o mal rappresentate” quello di visualizzazione consentita dall’impiego di rivelatori dalla mammografia. digitali diretti per cui le stesse proiezioni grezze possono essere Lo studio in corso presso lo IOV di Padova prevede il con- processate per ricostruire piani diversi. fronto tra le prestazioni cliniche della mammografia digitale La ricostruzione volumetrica, in linea di principio, consente di (FFDM) nelle due proiezioni standard (CC cranio-caudale superare uno dei limiti principali dell’imaging bidimensionale, e MLO medio-laterale-obliqua) e quelle della tomosintesi ovvero il mascheramento di lesioni (nel caso della mammella, (DBT) in un’unica proiezione obliqua, rispettando così il vin- masse, microcalcificazioni, ecc.), causato dalla sovrapposizione coli di dose. Infatti, fino a quando non sarà dimostrata almeno di strutture normali; quindi l’opportunità di dissociare piani la non-inferiorità clinica della tomosintesi rispetto alla mam- diversi da parte della tomosintesi fa ritenere possibile una ridu- mografia, non è ragionevole aumentare la dose per l’esame di zione del numero di falsi negativi e di falsi positivi dovuti alla tomosintesi; per questo motivo, la dose per l’acquisizione della sovrapposizione [3-5]. sequenza di proiezioni di tomosintesi è stata calcolata in modo Di seguito vengono illustrate le prime esperienze di tomosinte- da non superare quella di una mammografia tradizionale in due si digitale della mammella sulle immagini prodotte con un pro- proiezioni (Figura 1). totipo sviluppato da GE Healthcare ed installato presso l’Unità N. 55 - 2008 ATTUALITÀ IN SENOLOGIA 25
  • 2. OBIETTIVO SU Le sequenze di tomosintesi usate nello studio FIGURA 1 prevedono l’acquisizione di 15 proiezioni a bassa dose, comprese in un arco di ± 20° ri- spetto alla perpendicolare al rivelatore. La dose erogata per una sequenza di tomosin- tesi è stata calcolata entro un valore limite che non superi la dose di due proiezioni mammo- grafiche standard, così che in fase applicativa la dose di fatto viene frazionata in modo che per ciascuna proiezione sia 1/15 del totale [7]. Le proiezioni bidimensionali sono processate tramite un algoritmo di tipo SART (Simulta- neous Algebric Reconstruction Technique) per ricostruire il volume della mammella, campionato secondo strati contigui di 1 mil- limetro [8]. Approccio clinico Alcune osservazioni preliminari sulle possibi- lità delle tecniche e sul metodo di analisi delle immagini mettono in evidenza che alla “non Lo studio è stato condotto dopo aver ottenuto il parere favo- sovrapposizione della DBT” si contrappone un potenziale “be- revole del Comitato Etico, l’approvazione del Ministero della neficio geometrico” connesso alle due proiezioni della FFDM. Salute per l’uso di un “prototipo” e con il consenso informato Inoltre se è vero che la DBT può costruire un “numero arbitra- delle Pazienti eligibili. FIGURA 2 Apparecchiatura Il sistema per tomosintesi si basa sulla piattaforma del mammografo digitale commerciale GE Senographe DS, con rivelatore flat panel di area 19 x 23 cm2, che impiega Ioduro di Cesio in ingresso per la conversione dei raggi-X in luce, con dimensione nominale del pixel di 100 μm2. La movimentazione delgantry è stata modificata in modo da permettere la rotazione del tubo radiogeno attorno all’asse perpendi- colare al rivelatore; inoltre il sistema di compressione è stato reso separato dal gantry per far sì che il sistema “mam- mella compressa-rivelatore” rimanga fermo durante la rotazione del tubo. In figura 2 è schematizzata l’acquisizione di un esame di tomosintesi [6]. 26 ATTUALITÀ IN SENOLOGIA 2008 - N. 55
  • 3. OBIETTIVO SU rio” di sezioni da un numero finito di proiezioni, l’accuratez- le “performance” cliniche tramite analisi ROC. Tuttavia, per za d’informazione dipende dall’intervallo di campionamento mantenere un approccio più clinico, anziché far riferimento delle fette. Il campionamento per ogni millimetro di spessore alle curve frutto dell’analisi statistica, nel seguito seguiremo delle sezioni (slice), ragionevole tra accuratezza e necessità di uno schema di valutazione delle immagini mammografiche e di limitare il numero di immagini nell’ottica dell’ottimizzazione, tomosintesi secondo i 5 punti proposti da Kopans: può essere un limite a scopo diagnostico, ad esempio per la “va- 1) Riconoscimento: trovare la lesione lutazione d’insieme” di piccoli dettagli come le microcalcifica- 2) Presenza: è reale? zioni (Figura 3A). 3) Localizzazione: dove si trova ? 4) Interpretazione: cosa è ? 5) Decisione: “cosa farne” Di seguito sono sinteticamente riferite le “impressioni” più in- teressanti, mentre i primi risultati complessivi sui casi studiati sono riportati successivamente. 1) Riconoscimento Premesso che la densità mammaria può pregiudicare la rileva- bilità delle lesioni, condizione non superabile per entrambe le tecniche (FFDM e SFM), la DBT consente l’individuazione di un maggior numero di lesioni espansive, con una miglior analisi della forma e dei contorni delle masse e dell’ampiezza e dell’architettura delle distorsioni (Figure 4 A e B). Fig. 3A Pertanto sezioni più spesse (slab) sono a volte più vantaggiose: le immagini slab sono ottenute per compattamento di più slice (Figura 3B). FFDM DBT sinistra sinistra MLO MLO Fig. 4A slab [8-12] Fig. 3B Su tali premesse è stato preso in considerazione un primo grup- po di 50 casi sottoponendoli alla valutazione di tre Radiologi esperti con metodologia “blinded and independent”, per rica- vare, dal confronto fra le due tecniche, le prime valutazioni sul- Fig. 4B N. 55 - 2008 ATTUALITÀ IN SENOLOGIA 27
  • 4. OBIETTIVO SU Per le calcificazioni al momento non sono emerse significative FFDM DBT differenze su questo punto. Le immagini slab costituiscono una RMLO RMLO variante tecnica di riferimento nell’analisi della distribuzione, con particolare riguardo alla “configurazione duttale” (Figura 5) Fig. 6B 3) Localizzazione Le due proiezioni mammografiche ortogonali sono il riferi- mento standard per l’attribuzione del quadrante delle lesioni; Fig. 5 è noto anche che le proiezioni CC e MLO non sono spesso sufficienti a definire l’appartenenza anatomo-topografica. L’ingrandimento elettronico risulta di aiuto all’analisi sia nella Analogamente la tomosintesi ottenuta con mammella in po- FFDM che nella DBT. sizione obliqua con le attuali ricostruzioni non è in grado di svolgere questo compito. E’ lecito pensare che ulteriori sviluppi 2) Presenza informatici in un prossimo futuro possano raggiungere anche Il rumore anatomico (densità e sovrapposizione) in mammo- questo obiettivo. grafia può produrre dei “falsi positivi”. L’analisi mediante sezio- ni contigue elimina immagini apparentemente sospette soprat- Attualmente la DBT è solo in grado di stabilire correttamente tutto nei casi di sovrapposizione di strie con evidenza di aspetti la sede delle lesioni, tenuto conto delle modalità di posiziona- distorsivi (Figura 6A). In una prospettiva opposta la DBT può mento, definendo correttamente la profondità rispetto ai piani confermare con significatività “segni deboli” di presenza di le- cutanei di appoggio e di compressione. sione rivelati dalla mammografia (Figura 6B). Ne traggono beneficio in particolare le immagini di lesioni si- tuate in corrispondenza di questi (Figura 7). Case # 149 - FFDM FFDM DBT LCC LMLO Fig. 7 Fig. 6A 28 ATTUALITÀ IN SENOLOGIA 2008 - N. 55
  • 5. OBIETTIVO SU 4) Interpretazione Primi risultati e considerazioni Dai punti precedenti appare evidente in generale un maggior contributo da parte della DBT riguardo la semeiotica delle le- Vale la pena riassumere in queste prime applicazioni le sioni mammarie, che sembra ottimale per le distorsioni, buono impressioni comuni ai tre Radiologi che concordano per le masse, ma di minore impatto per le calcificazioni. nell’affermare che la DBT permette un sostanziale mi- Per queste appare importante un uso adeguato delle slab che glioramento nel rilevamento e nell’analisi delle lesioni, consentono una visione d’insieme. influendo tanto sul convincimento della loro presenza Come noto, solo in caso di rilievi molto significativi (masse a quanto nella certezza della loro assenza rispetto alla contorni sfrangiati, calcificazioni irregolari e racemose) gli ap- FFDM. porti della sola immagine radiologica possono essere definitivi A ciò si associa peraltro un tempo di lettura discreta- per stabilire con accuratezza una diagnosi di natura; di fatto mente più lungo, correlato al numero d’immagini e al è più spesso “orientativa” a stabilire l’indicazione agli apporti contenuto informativo significativamente più ampio complementari dell’ecografia, delle indagini agobioptiche di delle stesse da esaminare con attenzione sia in relazione secondo livello e della Risonanza Magnetica con mezzo di con- all’analisi di strato quanto di ricomposizione spaziale trasto. da parte dell’operatore medico. In sintesi l’apporto della DBT diventa rilevante quando intro- La DBT non è in grado allo stato attuale di definire duca aspetti correlati ad una migliorata analisi che orientano con sufficiente accuratezza il quadrante nel quale la le- con convinzione verso la benignità o la malignità, apporto che sione è situata. peraltro deve essere contestualizzato nelle procedure già conva- I primi risultati delle prestazioni di ciascun Radiologo lidate in Senologia. relativi ai 50 casi esaminati sono stati accolti ed ela- borati con il metodo ROC. La curva relativa al rileva- 5) Decisione mento delle lesioni, sul significato statistico della quale Il punto cruciale della decisione è frutto di tutte le informazio- peraltro non è possibile esprimere un giudizio definiti- ni raccolte, non solo delle immagini. vo, è per tutti i Radiologi indicativa di “performance” E’ chiaro che la DBT è particolarmente significativa quando migliori, comunque non peggiori, per la DBT rispetto elimina immagini potenzialmente sospette, come illustrato in alla FFDM. Fig 6A o invece riconosce lesioni sospette non visibili o insuffi- E’ infine da segnalare che la tomosintesi di fatto non cientemente documentate per rumore anatomico alla mammo- introduce elementi di “semeiotica nuovi”, ma richiede grafia (Fig 6B). un periodo di adattamento all’immagine che nella no- La decisione, analizzati tali contributi, si coniuga inevitabil- stra esperienza è stato piuttosto breve. mente con l’interpretazione e procede di conseguenza. N. 55 - 2008 ATTUALITÀ IN SENOLOGIA 29
  • 6. OBIETTIVO SU Conclusioni imaging. Radiology 205, 399-406 (1997) 3.Bochud, F.O., Valley, J.F., Verdun, F.R.: Estimation Le prime esperienze cliniche sembrano indicare un im- of the noisy component of anatomical background. Med. Phys. 26, 1365-1370 (1999) patto favorevole della DBT nel rilevamento delle lesio- 4.Burgess, A.E., Jacobson, F.L., Judy, P .F.: On the ni mammarie. difficulty of detecting tumors in mammograms. In: Il rilevamento tomografico consente di fatto di apprez- Proceedings of the 17th International Conference on zare isolatamente le lesioni che più “soffrono” della so- Information Processing, pp. 1-11. Springer-Verlag, vrapposizione e quindi della confusione dei piani alla London (2001) mammografia standard. 5.Kontos, D., Bakic, P.R., Maidment, A.D.A.: Analysis Se anche i primi risultati comparativi, per ora esigui, of parenchymal texture properties in breast tomo- saranno confermati, la DBT potrà inserirsi a pieno ti- synthesis images. In: Giger, M.L., Karssemeijer, N. tolo nel novero delle tecniche idonee allo studio della (eds.) Proceedings of SPIE Medical Imaging 2007: mammella finalizzato al rilevamento di lesioni tanto Computer-Aided Diagnosis, vol.6514, pp. 651417 nella diagnostica senologica quanto nello screening. (2007) Gli attesi sviluppi informatici, in particolare i disposi- 6.Gennaro G, di Maggio C, Baldan E, Bezzon E: tivi CAD dedicati, e le potenzialità connesse all’impie- Mammografia digitale 3D. In: “Mammografia Digi- go dei mezzi di contrasto sono da considerare ulteriori tale”, Poletto Editore 2007. 7.Wu, T., Liu, B., Moore, R., Kopans, D.: Optimal ac- promettenti innovazioni che la tomosintesi dovrebbe quisition techniques for digital breast tomosynthesis proporre ai Radiologi in un futuro non lontano. screening. In: Flynn, M.J., Hsieh, J. (eds.) Proce- edings of SPIE Medical Imaging 2006: Physics of Medical Imaging, vol. 6142, pp. 61425E (2006) Bibliografia 8.Zhang, Y., Chan, H.P Sahiner, B., Wei, J., Goodsitt, ., M.M., Hadjiiski, L.M., Ge, J., Zhou, C.: A comparati- 1.A Vallebona: Una modalità di tecnica per la disso- ve study of limited-angle cone-beam reconstruction ciazione radiografica. Radiologia Medica 17, 1090- methods for breast tomosynthesis. Med. Phys. 33, 1097 (1931). 3781-3795 (2006). 2.Niklason, L.T., Christian, B.T., Niklason, L.E., Ko- pans, D.B., Castleberry, D.E., Opsahl-Ong, B.H., Lan- Figure dberg, C.E., Slanetz, P.J., Giardino, A.A., Moore, R., Albagli, D., DeJoule, M., Fitzgerald, P .F., Fobare, Fig. 1 - Le curve indicano la dose ghiandolare me- D.F., Gianbattista, B.W., Kwasnick, R.F., Jianqiang, dia (AGD = Average Glandular Dose) per la mam- L., Lubowski, S.J., Possin, G.E., Richotte, J.F., Wei, mografia tradizionale (SFM) e la tomosintesi (DBT) C.Y., Weirth, R.F.: Digital tomosynthesis in breast in funzione dello spessore mammario (breast thick- 30 ATTUALITÀ IN SENOLOGIA 2008 - N. 55
  • 7. OBIETTIVO SU ness): esse appaiono quasi sovrapponibili. storsiva di 13 mm causata da sovrapposizione di Fig. 2 - Principio di funzionamento della tomosin- strutture. La DBT conferma l’assenza della lesione. tesi. In fase di acquisizione (sinistra) il tubo ruota Fig. 6 - B) Nella FFDM è con difficoltà riconoscibile attorno all’asse perpendicolare al rivelatore e, in un una massa a contorni sfumati di 5 mm. Con DBT ben numero finito di posizioni prestabilite, il sistema ac- riconoscibile nella stessa sede massa isolata a con- quisisce proiezioni bidimensionali a bassa dose che torni sfrangiati (carcinoma tubulare). costituiscono i dati grezzi. Le immagini bidimensio- nali vengono poi trasferite al sistema di ricostruzio- Fig. 7 - La massa è riconoscibile con FFDM; con DBT ne che, applicando un algoritmo di “shift-and-add”, si localizza nei piani superficiali e appare con net- ovvero di riallineamento delle strutture e di somma tezza delimitata da trasparenza attribuibile all’aria del segnale, ricostruisce il volume della mammella (nevo cutaneo). che viene poi campionato ad intervalli di 1 mm, per- mettendo così di mettere a fuoco strutture che stan- no su piani diversi. Fig. 3 - A) Sequenza di slice da 1 mm relativo ad un caso di microcalcificazioni (carcinoma duttale in situ). Fig. 3 - B) Confronto tra FFDM e DBT con immagini slab ottenute per compattamento di 5 slice. Fig. 4 - A) Confronto tra FFDM e DBT: proiezione LMLO (Left Medio-Lateral-Oblique). Nella FFDM non sono riconoscibili aspetti riferibili a masse. Nella DBT è evidente massa allungata con diametro massi- mo di 10 mm a contorni netti (lesione benigna). Fig. 4 - B) Anche in questo caso solo nella DBT è riconoscibile massa a contorni sfrangiati di 7 mm (carcinoma duttale infiltrante). Fig. 5 - Confronto tra FFDM e DBT: le informazioni riguardo le calcificazioni nella forma, numero e di- stribuzione sono analoghe per entrambe le tecniche. Fig. 6 - A) La FFDM in CC evidenzia immagine di- N. 55 - 2008 ATTUALITÀ IN SENOLOGIA 31