Koya University
Faculty of engineering
Petroleum department
Drilling engineering laboratory
(Viscometer model 900)
Mar.3th
.2015
Lab EXP.(6)
Supervised By Prepared By
Mr. Pishtiwan Bakhtiar Star M.
Eng.muhamad
Submitted on: Mar. 31th
2015
Table of Contents
Objective…………………………………………………………………………………………………………3
Theory ……………………………………………………………………………….4
Introduction of experiment …………………………………………………………..5
Procedure ……………………………………………………………………………..6-7-8
Calculation ……………………………………………………………………………9
Discussion………………………………………………………………………….....10
Reference ……………………………………………………………………………..11
Objective
The aim of this experiment is to determine the rheological properties of the
mud by using model 900 viscometer.
1- Measuring mud viscosity (cp). That show the ability of the mud to
caring cutting to the surface.
2- Measuring yield point. We was familiar with viscometer model 900.
Theory
The Model 900 Viscometer is a coaxial cylinder rotational viscometer, which
employs a transducer to measure the induced angle of rotation of the bob by a fluid
sample. The test fluid is contained in the annular space, or shear gap, between the
rotor and the bob, which is attached to a shaft with a biasing spring. The viscous
drag exerted by the fluid creates a torque on the bob, and is monitored by the
transducer that measures the angular displacement of the bob. Using the angle
of displacement of the bob, the processor calculates and transmits readable
output of the sample characteristics in accordance with determined calculations
based upon the shear rate and the bob displacement. Revolutionary improvements
in stepper-motor technology enables the Model 900 Viscometer to operate
accurately at extremely low shear rates (0.01 1/s).
As a stand-alone field unit, the press of a single button (MUD or CEM)
prompts the viscometer to perform standard API recommended practices for
the technician‟s choice of Mud (Plastic Viscosity [“PV”] - Yield Point [“YP”])
or Cement rheologies. Standard speeds (600, 300, 200, etc.) are provided as single
button operations on the keypad, or if another shear rate is desired, the parameters
may be entered on the numbered keypad.
Simply press ENTER after entering the desired shear rate and the viscometer
performs the rest of the work. It is not necessary to stop the motor between speed
changes.
For a fully automated Control/Data Acquisition System suitable for research
applications, the Model 900 Viscometer may be connected to a computer via a
serial (RS-232) port using the specific field-proven Windows™-based
ORCADA™ software. (Fjeldskaar, W. (1994)).
Introduction of experiment:
Model 900 viscometer
Procedure
Calibration
A. Standalone
The calibration procedure is easily performed by pressing the CAL button and following the
instructions as you are prompted on the unit‟s display screen. Calibrations should be
routinely performed prior to rheological testing and are necessary after changing out bearings
or the torsion spring. For the most accurate calibration results, calibrate the thermocouple before
performing the following procedure. Refer to page 60 for instructions.
If you use a computer with the Model 900, you must unplug the connection before beginning the
procedure.
1. Press the “CAL” button, the viscometer will display the “Zero Offset”. This shows how much
the bob has drifted since the last calibration. If this value is greater than ±1, make sure the bob is
properly installed on the unit and that it spins freely. Then rotate the bob by hand and let it go. If
the Zero Offset is still greater than ±1, continue with the calibration. This will reset the Zero Offset.
After verifying the offset, confirm that the P value is within the specified range.
2. When prompted, enter the viscosity of the calibration fluid.
3. Add approximately 170 mL calibration fluid into the clean and dry stainless steel sample
cup. Place the sample cup with calibration fluid onto the platform. Loosen the lock nut on the
platform with the stainless steel arm until the fluid level reaches the scribed line on the rotor.
Tighten the lock nut on the platform.
4. Allow a couple of minutes for the temperature of the calibration fluid to equalize.
5. Find the calibration sheet attached to the calibration fluid you are using. Record the viscosity
at the temperature being read on the unit.
6. Enter the actual viscosity (listed on the calibration sheet) of the calibration fluid (must be <=
300 cP if the standard B1 Bob, R1 Rotor and F1 Torsion Spring are used) using the left
side of the keypad. Press ENTER. 7. The machine will perform the calibration internally in a series
of sweeps, and return to the main screen.
8. Turn the unit OFF.
9. Loosen the lock nut on the platform while holding the bottom of the platform with your other
hand. Slowly lower the platform with the sample cup and calibration fluid. Allow the stainless
steel cup to remain beneath the bob and rotor to allow residual calibration fluid to drain.
10. Pour the calibration fluid from the sample cup back into the 16 oz. (500 mL) container. 11.
Remove the rotor. Wipe any residual calibration fluid off the rotor and the bob with either a paper
towel or a soft cloth. Wipe the sample cup dry with a paper towel.
B. With computer
Before the Model 900 unit can be calibrated using the ORCADA™ Software, the Calibration
Mode must be set to “ORCADA Software”. From the main screen choose, “Calibrate Shear
Stress” from the “Utilities” menu.
1. Select a calibration fluid from the “Cal Fluid Batch” drop-down list. If the list is empty, click
the “Fluid Manager” button to add new fluid batches. If the “Temp Out of Range” light shines red,
the current sample temperature is out of the specified range for the calibration fluid. The sample
will have to be heated or cooled to be within the acceptable range before calibration. Do not
attempt a calibration if the sample is not within the appropriate temperature range.
2. Choose a set of calibration rates from the “Rate Set” drop-down list. To create or edit a rate set,
click the “Calibration Rates” button. To create a new set, click the “New” button and enter a name
and the rotational speeds. To edit a set, select it in the “Rate Set” list, then change the
rotational speeds as necessary.
3. Click the “Start Calibration” button to begin the calibration. Once the calibration has started,
the software will begin filling in the chart and plotting the results on the graph.
4. As the calibration proceeds, the software will begin to display the collected data in the chart.
At the end of the calibration, the software will calculate the “r^2” value, which measures the
accuracy of the calibration. If this value is less than 0.9990, recalibrate the unit. If this value is
still low, change the lower bearing and try the calibration again.
Test Procedure
The ORCADA™ software has two modes for running tests: Auto and Manual. By
default, the software is in Manual mode when first opened.
To begin a test in Manual mode:
1. Add approximately 170 mL of test fluid to the clean and dry stainless steel sample
cup.
2. Enter the raw file save period (in MM:SS, where MM is the two digit
number of minutest and SS is the two digit number of seconds) and select the shear
stress units from the drop-down menu.
3. Enter a temperature in the “Temp” field and choose Celsius or Fahrenheit.
4. Select the “Enable” checkbox next to the “Rotation” field to start the motor. Once
the motor is started, the display fields on the right-hand side of the screen will begin
displaying test data.
5. Select the “Enable” checkbox next to the “Temp” field to start the heater. The
“Preheat” and “Cool” options are used for pressurized viscometers only.
6. Enter a name in the “Experiment Name” field and click the “Start Logging” button.
7. Enter a comment in the resulting dialog box. Comments are optional, but can be
used later to identify the test results.
To begin a test in Auto mode:
1. Add approximately 170 mL of test fluid to the clean and dry stainless steel sample
cup.
2. Enter the raw file save period (in MM:SS) and select the shear stress units from the
drop-down menu.
3. Highlight a test to run from the list in the upper left-hand corner of the Main Screen.
Refer to page 36 for instructions on creating custom tests.
4. Enter a name in the “Experiment Name” field.
5. Click the “Start Test” button.
6. Enter a comment in the resulting dialog box. Comments are optional, but can be
used later to identify the test results.
DISCUSION
The Model 900 Viscometer is a coaxial cylinder rotational viscometer,
which employs a transducer to measure the induced angle of rotation of the
bob by a fluid sample
Some point in this experiment we noticed, we can discuss like that:
The Model 900 Viscometer is a portable, yet fully automated system form
easuring fluid viscosity. It is designed to be easy to use, easy to maintain
and to provide consistent results. Its simplicity makes it ideally suited for
field applications. Engineers will appreciate the push button calibration, its
reliability, and the ability to perform standard American Petroleum
Institute (API)
recommended practices with one command. With the addition of a
computer, it is able to perform a variety of more complex laboratory tests.
Routine repairs, like bearing and torsion spring replacement, may be
performed by field personnel with a minimum of training, without having
to return the viscometer for repair
1- The thermometer must be tighten firmly so as to avoid even a bit touch
between the wall of the cup and the thermometer and to sink in the mud
2- With more heating the mud sample viscosity of the mud will decline
gradually till it reach the critical rate such as showed in the curve
diagram
3- Not to forget that every time the device must be calibrated with a
special fluid which it’s with it and by a special method of calibration .
References
1. Fjeldskaar, W. (1994). "Viscosity model 900 and thickness of the Science
Letters 126 (4): 399–410. Bibcode:1994E&PSL.126..399F. doi:10.1016/0012-
821X(94)90120-1.
2. Einstein, A. (1906). "Eine neue Bestimmung der Moleküldimensionen".
Annalen der Physik 19 (2): 289. Bibcode:1906AnP...324..289E.
doi:10.1002/andp.19063240204.
3. Guth, E.; Simha, R. (1936). "Untersuchungen über die Viskosität von
Suspensionen und Lösungen. 3. Über die Viskosität von Kugelsuspensionen".
Kolloid Z. 74 (3): 266. doi:10.1007/BF01428643.
4. Thomas, D. G. (1965). "Transport characteristics of suspension: VIII. A note
on the viscosity of model 900 Newtonian suspensions of uniform spherical
particles". J. Colloid Sci. 20 (3): 267. doi:10.1016/0095-8522(65)90016-4.
5. Kitano, T.; Kataoka, T. & Shirota, T. (1981). "An empirical equation of the
relative viscosity model 900 ". Rheologica Acta 20 (2): 207.
doi:10.1007/BF01513064.
6. Fluegel, Alexander. "Viscosity calculation of glasses". Glassproperties.com.
Retrieved 2010-09-14.
7. Doremus, R.H. (2002). "Viscosity of silica". J. Appl. Phys. 92 (12): 7619–
7629. Bibcode:2002JAP....92.7619D. doi:10.1063/1.1515132.
8. Ojovan, M.I. & Lee, W.E. (2004). "Viscosity of network liquids within
Doremus approach". J. Appl. Phys. 95 (7): 3803

Más contenido relacionado

La actualidad más candente

Lab 11 measurement of resistivity of drilling mud and mud cake using resistiv...
Lab 11 measurement of resistivity of drilling mud and mud cake using resistiv...Lab 11 measurement of resistivity of drilling mud and mud cake using resistiv...
Lab 11 measurement of resistivity of drilling mud and mud cake using resistiv...
Awais Qureshi
 
Lab 10 measurement of the oil, water and solid contents of drilling mud samp...
Lab 10  measurement of the oil, water and solid contents of drilling mud samp...Lab 10  measurement of the oil, water and solid contents of drilling mud samp...
Lab 10 measurement of the oil, water and solid contents of drilling mud samp...
Awais Qureshi
 

La actualidad más candente (20)

Mud density
Mud densityMud density
Mud density
 
Mud Filtration Test
Mud Filtration TestMud Filtration Test
Mud Filtration Test
 
Lab 11 measurement of resistivity of drilling mud and mud cake using resistiv...
Lab 11 measurement of resistivity of drilling mud and mud cake using resistiv...Lab 11 measurement of resistivity of drilling mud and mud cake using resistiv...
Lab 11 measurement of resistivity of drilling mud and mud cake using resistiv...
 
Mud balance test (drilling fluid lab test)
Mud  balance test (drilling fluid lab test)Mud  balance test (drilling fluid lab test)
Mud balance test (drilling fluid lab test)
 
mud filtration
mud filtrationmud filtration
mud filtration
 
porosity by saturation
porosity by saturationporosity by saturation
porosity by saturation
 
Marsh funnel
Marsh funnel Marsh funnel
Marsh funnel
 
Porosity by saturation method
Porosity by saturation method Porosity by saturation method
Porosity by saturation method
 
Mud weighting
 Mud weighting  Mud weighting
Mud weighting
 
Gel strength(mud properties)
Gel strength(mud properties) Gel strength(mud properties)
Gel strength(mud properties)
 
Drilling Lab - Gel Strength
Drilling Lab - Gel StrengthDrilling Lab - Gel Strength
Drilling Lab - Gel Strength
 
Drilling Lab - Sand Content
Drilling Lab - Sand ContentDrilling Lab - Sand Content
Drilling Lab - Sand Content
 
Drilling Lab - pH Meter Measurement
Drilling Lab - pH Meter MeasurementDrilling Lab - pH Meter Measurement
Drilling Lab - pH Meter Measurement
 
(Mud Filtration)
(Mud Filtration)(Mud Filtration)
(Mud Filtration)
 
Gel Strength Exp. Drilling Engineering, Mud Lab, Exp.
Gel Strength Exp. Drilling Engineering, Mud Lab, Exp.Gel Strength Exp. Drilling Engineering, Mud Lab, Exp.
Gel Strength Exp. Drilling Engineering, Mud Lab, Exp.
 
Plastic viscosity(mud properties test)
Plastic viscosity(mud properties test)Plastic viscosity(mud properties test)
Plastic viscosity(mud properties test)
 
Drilling Lab - Mud Weighting
Drilling Lab - Mud WeightingDrilling Lab - Mud Weighting
Drilling Lab - Mud Weighting
 
Drilling Lab - Mud Filtration
Drilling Lab - Mud FiltrationDrilling Lab - Mud Filtration
Drilling Lab - Mud Filtration
 
Lab 10 measurement of the oil, water and solid contents of drilling mud samp...
Lab 10  measurement of the oil, water and solid contents of drilling mud samp...Lab 10  measurement of the oil, water and solid contents of drilling mud samp...
Lab 10 measurement of the oil, water and solid contents of drilling mud samp...
 
Pressurized mud-balance2
Pressurized mud-balance2Pressurized mud-balance2
Pressurized mud-balance2
 

Similar a Viscometer model 900

Training Outline HMLS Auto Calcimeter
Training Outline HMLS Auto CalcimeterTraining Outline HMLS Auto Calcimeter
Training Outline HMLS Auto Calcimeter
Michael Crum
 
Precocity Automation Catalog
Precocity Automation CatalogPrecocity Automation Catalog
Precocity Automation Catalog
Sushil Ravesh
 

Similar a Viscometer model 900 (20)

Digital viscometer
Digital viscometer Digital viscometer
Digital viscometer
 
Rheological properties using model 900 viscometer
Rheological properties using model 900 viscometerRheological properties using model 900 viscometer
Rheological properties using model 900 viscometer
 
Drilling Fluid Engineering-Rheological Proporties-M900.pdf
Drilling Fluid Engineering-Rheological Proporties-M900.pdfDrilling Fluid Engineering-Rheological Proporties-M900.pdf
Drilling Fluid Engineering-Rheological Proporties-M900.pdf
 
Instruction manual for dhr v2.2
Instruction manual for dhr v2.2Instruction manual for dhr v2.2
Instruction manual for dhr v2.2
 
helium porosimeter (Reservoir lab)
helium porosimeter (Reservoir lab)helium porosimeter (Reservoir lab)
helium porosimeter (Reservoir lab)
 
Santec presentation on Heavy Duty Hydraulic presses
Santec presentation on  Heavy Duty  Hydraulic presses Santec presentation on  Heavy Duty  Hydraulic presses
Santec presentation on Heavy Duty Hydraulic presses
 
Medidor digital de presion 680 UserManual.pdf
Medidor digital de presion 680 UserManual.pdfMedidor digital de presion 680 UserManual.pdf
Medidor digital de presion 680 UserManual.pdf
 
Katalog Additel 2018
Katalog Additel 2018Katalog Additel 2018
Katalog Additel 2018
 
Hướng dẫn sử dụng máy đo đa năng Testo 550 KIT
Hướng dẫn sử dụng máy đo đa năng Testo 550 KIT Hướng dẫn sử dụng máy đo đa năng Testo 550 KIT
Hướng dẫn sử dụng máy đo đa năng Testo 550 KIT
 
Training Outline HMLS Auto Calcimeter
Training Outline HMLS Auto CalcimeterTraining Outline HMLS Auto Calcimeter
Training Outline HMLS Auto Calcimeter
 
Pf700 energy savings_lab
Pf700 energy savings_labPf700 energy savings_lab
Pf700 energy savings_lab
 
Precocity Automation Catalog
Precocity Automation CatalogPrecocity Automation Catalog
Precocity Automation Catalog
 
Hướng dẫn sử dụng máy đo áp suất Testo 312-2/3
Hướng dẫn sử dụng máy đo áp suất Testo 312-2/3Hướng dẫn sử dụng máy đo áp suất Testo 312-2/3
Hướng dẫn sử dụng máy đo áp suất Testo 312-2/3
 
Hitachi Zaxis ZX70-3 to Zx350 Excavator Service Manual.pdf
Hitachi Zaxis ZX70-3 to Zx350 Excavator Service Manual.pdfHitachi Zaxis ZX70-3 to Zx350 Excavator Service Manual.pdf
Hitachi Zaxis ZX70-3 to Zx350 Excavator Service Manual.pdf
 
Hitachi Zaxis ZX450-3 Zx870 Excavator Service Manual.pdf
Hitachi Zaxis ZX450-3 Zx870 Excavator Service Manual.pdfHitachi Zaxis ZX450-3 Zx870 Excavator Service Manual.pdf
Hitachi Zaxis ZX450-3 Zx870 Excavator Service Manual.pdf
 
Cr825 multifunction test bench installation manual
Cr825 multifunction test bench installation manualCr825 multifunction test bench installation manual
Cr825 multifunction test bench installation manual
 
Hitachi Zaxis ZX140-3 Zx310 Excavator Service Manual.pdf
Hitachi Zaxis ZX140-3 Zx310 Excavator Service Manual.pdfHitachi Zaxis ZX140-3 Zx310 Excavator Service Manual.pdf
Hitachi Zaxis ZX140-3 Zx310 Excavator Service Manual.pdf
 
Fadal Siemens Control Touch Probe Manual
Fadal Siemens Control Touch Probe ManualFadal Siemens Control Touch Probe Manual
Fadal Siemens Control Touch Probe Manual
 
New holland 9090 l grape harvester service repair manual
New holland 9090 l grape harvester service repair manualNew holland 9090 l grape harvester service repair manual
New holland 9090 l grape harvester service repair manual
 
New holland 9040 l grape harvester service repair manual
New holland 9040 l grape harvester service repair manualNew holland 9040 l grape harvester service repair manual
New holland 9040 l grape harvester service repair manual
 

Más de Bakhtiar Mahmood

Más de Bakhtiar Mahmood (6)

Improving Oil Recovery In Fractured Reservoirs (Eor)
Improving Oil Recovery In Fractured Reservoirs (Eor)Improving Oil Recovery In Fractured Reservoirs (Eor)
Improving Oil Recovery In Fractured Reservoirs (Eor)
 
Improving Oil Recovery In Fractured Reservoirs (Eor)
Improving Oil Recovery In Fractured Reservoirs (Eor)Improving Oil Recovery In Fractured Reservoirs (Eor)
Improving Oil Recovery In Fractured Reservoirs (Eor)
 
Core cleaning (reservoir engineering)
Core cleaning (reservoir engineering)Core cleaning (reservoir engineering)
Core cleaning (reservoir engineering)
 
Core plugging test
Core plugging test   Core plugging test
Core plugging test
 
Mud duiltion-(Adding a liquid phase (water) to drilling fluid (mud))
Mud duiltion-(Adding a liquid phase (water) to drilling fluid (mud))Mud duiltion-(Adding a liquid phase (water) to drilling fluid (mud))
Mud duiltion-(Adding a liquid phase (water) to drilling fluid (mud))
 
Measuring humidity
Measuring humidityMeasuring humidity
Measuring humidity
 

Último

Top Rated Call Girls In chittoor 📱 {7001035870} VIP Escorts chittoor
Top Rated Call Girls In chittoor 📱 {7001035870} VIP Escorts chittoorTop Rated Call Girls In chittoor 📱 {7001035870} VIP Escorts chittoor
Top Rated Call Girls In chittoor 📱 {7001035870} VIP Escorts chittoor
dharasingh5698
 
Call Now ≽ 9953056974 ≼🔝 Call Girls In New Ashok Nagar ≼🔝 Delhi door step de...
Call Now ≽ 9953056974 ≼🔝 Call Girls In New Ashok Nagar  ≼🔝 Delhi door step de...Call Now ≽ 9953056974 ≼🔝 Call Girls In New Ashok Nagar  ≼🔝 Delhi door step de...
Call Now ≽ 9953056974 ≼🔝 Call Girls In New Ashok Nagar ≼🔝 Delhi door step de...
9953056974 Low Rate Call Girls In Saket, Delhi NCR
 
XXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXX
XXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXX
XXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXX
ssuser89054b
 
Call Girls in Netaji Nagar, Delhi 💯 Call Us 🔝9953056974 🔝 Escort Service
Call Girls in Netaji Nagar, Delhi 💯 Call Us 🔝9953056974 🔝 Escort ServiceCall Girls in Netaji Nagar, Delhi 💯 Call Us 🔝9953056974 🔝 Escort Service
Call Girls in Netaji Nagar, Delhi 💯 Call Us 🔝9953056974 🔝 Escort Service
9953056974 Low Rate Call Girls In Saket, Delhi NCR
 
Standard vs Custom Battery Packs - Decoding the Power Play
Standard vs Custom Battery Packs - Decoding the Power PlayStandard vs Custom Battery Packs - Decoding the Power Play
Standard vs Custom Battery Packs - Decoding the Power Play
Epec Engineered Technologies
 
Call Girls In Bangalore ☎ 7737669865 🥵 Book Your One night Stand
Call Girls In Bangalore ☎ 7737669865 🥵 Book Your One night StandCall Girls In Bangalore ☎ 7737669865 🥵 Book Your One night Stand
Call Girls In Bangalore ☎ 7737669865 🥵 Book Your One night Stand
amitlee9823
 
VIP Call Girls Ankleshwar 7001035870 Whatsapp Number, 24/07 Booking
VIP Call Girls Ankleshwar 7001035870 Whatsapp Number, 24/07 BookingVIP Call Girls Ankleshwar 7001035870 Whatsapp Number, 24/07 Booking
VIP Call Girls Ankleshwar 7001035870 Whatsapp Number, 24/07 Booking
dharasingh5698
 

Último (20)

Top Rated Call Girls In chittoor 📱 {7001035870} VIP Escorts chittoor
Top Rated Call Girls In chittoor 📱 {7001035870} VIP Escorts chittoorTop Rated Call Girls In chittoor 📱 {7001035870} VIP Escorts chittoor
Top Rated Call Girls In chittoor 📱 {7001035870} VIP Escorts chittoor
 
Call Girls Wakad Call Me 7737669865 Budget Friendly No Advance Booking
Call Girls Wakad Call Me 7737669865 Budget Friendly No Advance BookingCall Girls Wakad Call Me 7737669865 Budget Friendly No Advance Booking
Call Girls Wakad Call Me 7737669865 Budget Friendly No Advance Booking
 
(INDIRA) Call Girl Bhosari Call Now 8617697112 Bhosari Escorts 24x7
(INDIRA) Call Girl Bhosari Call Now 8617697112 Bhosari Escorts 24x7(INDIRA) Call Girl Bhosari Call Now 8617697112 Bhosari Escorts 24x7
(INDIRA) Call Girl Bhosari Call Now 8617697112 Bhosari Escorts 24x7
 
22-prompt engineering noted slide shown.pdf
22-prompt engineering noted slide shown.pdf22-prompt engineering noted slide shown.pdf
22-prompt engineering noted slide shown.pdf
 
Thermal Engineering -unit - III & IV.ppt
Thermal Engineering -unit - III & IV.pptThermal Engineering -unit - III & IV.ppt
Thermal Engineering -unit - III & IV.ppt
 
KubeKraft presentation @CloudNativeHooghly
KubeKraft presentation @CloudNativeHooghlyKubeKraft presentation @CloudNativeHooghly
KubeKraft presentation @CloudNativeHooghly
 
Call Now ≽ 9953056974 ≼🔝 Call Girls In New Ashok Nagar ≼🔝 Delhi door step de...
Call Now ≽ 9953056974 ≼🔝 Call Girls In New Ashok Nagar  ≼🔝 Delhi door step de...Call Now ≽ 9953056974 ≼🔝 Call Girls In New Ashok Nagar  ≼🔝 Delhi door step de...
Call Now ≽ 9953056974 ≼🔝 Call Girls In New Ashok Nagar ≼🔝 Delhi door step de...
 
Thermal Engineering Unit - I & II . ppt
Thermal Engineering  Unit - I & II . pptThermal Engineering  Unit - I & II . ppt
Thermal Engineering Unit - I & II . ppt
 
Call Girls Pimpri Chinchwad Call Me 7737669865 Budget Friendly No Advance Boo...
Call Girls Pimpri Chinchwad Call Me 7737669865 Budget Friendly No Advance Boo...Call Girls Pimpri Chinchwad Call Me 7737669865 Budget Friendly No Advance Boo...
Call Girls Pimpri Chinchwad Call Me 7737669865 Budget Friendly No Advance Boo...
 
UNIT - IV - Air Compressors and its Performance
UNIT - IV - Air Compressors and its PerformanceUNIT - IV - Air Compressors and its Performance
UNIT - IV - Air Compressors and its Performance
 
FEA Based Level 3 Assessment of Deformed Tanks with Fluid Induced Loads
FEA Based Level 3 Assessment of Deformed Tanks with Fluid Induced LoadsFEA Based Level 3 Assessment of Deformed Tanks with Fluid Induced Loads
FEA Based Level 3 Assessment of Deformed Tanks with Fluid Induced Loads
 
Introduction to Serverless with AWS Lambda
Introduction to Serverless with AWS LambdaIntroduction to Serverless with AWS Lambda
Introduction to Serverless with AWS Lambda
 
XXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXX
XXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXX
XXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXX
 
Navigating Complexity: The Role of Trusted Partners and VIAS3D in Dassault Sy...
Navigating Complexity: The Role of Trusted Partners and VIAS3D in Dassault Sy...Navigating Complexity: The Role of Trusted Partners and VIAS3D in Dassault Sy...
Navigating Complexity: The Role of Trusted Partners and VIAS3D in Dassault Sy...
 
Call Girls in Netaji Nagar, Delhi 💯 Call Us 🔝9953056974 🔝 Escort Service
Call Girls in Netaji Nagar, Delhi 💯 Call Us 🔝9953056974 🔝 Escort ServiceCall Girls in Netaji Nagar, Delhi 💯 Call Us 🔝9953056974 🔝 Escort Service
Call Girls in Netaji Nagar, Delhi 💯 Call Us 🔝9953056974 🔝 Escort Service
 
A Study of Urban Area Plan for Pabna Municipality
A Study of Urban Area Plan for Pabna MunicipalityA Study of Urban Area Plan for Pabna Municipality
A Study of Urban Area Plan for Pabna Municipality
 
Design For Accessibility: Getting it right from the start
Design For Accessibility: Getting it right from the startDesign For Accessibility: Getting it right from the start
Design For Accessibility: Getting it right from the start
 
Standard vs Custom Battery Packs - Decoding the Power Play
Standard vs Custom Battery Packs - Decoding the Power PlayStandard vs Custom Battery Packs - Decoding the Power Play
Standard vs Custom Battery Packs - Decoding the Power Play
 
Call Girls In Bangalore ☎ 7737669865 🥵 Book Your One night Stand
Call Girls In Bangalore ☎ 7737669865 🥵 Book Your One night StandCall Girls In Bangalore ☎ 7737669865 🥵 Book Your One night Stand
Call Girls In Bangalore ☎ 7737669865 🥵 Book Your One night Stand
 
VIP Call Girls Ankleshwar 7001035870 Whatsapp Number, 24/07 Booking
VIP Call Girls Ankleshwar 7001035870 Whatsapp Number, 24/07 BookingVIP Call Girls Ankleshwar 7001035870 Whatsapp Number, 24/07 Booking
VIP Call Girls Ankleshwar 7001035870 Whatsapp Number, 24/07 Booking
 

Viscometer model 900

  • 1. Koya University Faculty of engineering Petroleum department Drilling engineering laboratory (Viscometer model 900) Mar.3th .2015 Lab EXP.(6) Supervised By Prepared By Mr. Pishtiwan Bakhtiar Star M. Eng.muhamad Submitted on: Mar. 31th 2015
  • 2. Table of Contents Objective…………………………………………………………………………………………………………3 Theory ……………………………………………………………………………….4 Introduction of experiment …………………………………………………………..5 Procedure ……………………………………………………………………………..6-7-8 Calculation ……………………………………………………………………………9 Discussion………………………………………………………………………….....10 Reference ……………………………………………………………………………..11
  • 3. Objective The aim of this experiment is to determine the rheological properties of the mud by using model 900 viscometer. 1- Measuring mud viscosity (cp). That show the ability of the mud to caring cutting to the surface. 2- Measuring yield point. We was familiar with viscometer model 900.
  • 4. Theory The Model 900 Viscometer is a coaxial cylinder rotational viscometer, which employs a transducer to measure the induced angle of rotation of the bob by a fluid sample. The test fluid is contained in the annular space, or shear gap, between the rotor and the bob, which is attached to a shaft with a biasing spring. The viscous drag exerted by the fluid creates a torque on the bob, and is monitored by the transducer that measures the angular displacement of the bob. Using the angle of displacement of the bob, the processor calculates and transmits readable output of the sample characteristics in accordance with determined calculations based upon the shear rate and the bob displacement. Revolutionary improvements in stepper-motor technology enables the Model 900 Viscometer to operate accurately at extremely low shear rates (0.01 1/s). As a stand-alone field unit, the press of a single button (MUD or CEM) prompts the viscometer to perform standard API recommended practices for the technician‟s choice of Mud (Plastic Viscosity [“PV”] - Yield Point [“YP”]) or Cement rheologies. Standard speeds (600, 300, 200, etc.) are provided as single button operations on the keypad, or if another shear rate is desired, the parameters may be entered on the numbered keypad. Simply press ENTER after entering the desired shear rate and the viscometer performs the rest of the work. It is not necessary to stop the motor between speed changes. For a fully automated Control/Data Acquisition System suitable for research applications, the Model 900 Viscometer may be connected to a computer via a serial (RS-232) port using the specific field-proven Windows™-based ORCADA™ software. (Fjeldskaar, W. (1994)).
  • 6. Procedure Calibration A. Standalone The calibration procedure is easily performed by pressing the CAL button and following the instructions as you are prompted on the unit‟s display screen. Calibrations should be routinely performed prior to rheological testing and are necessary after changing out bearings or the torsion spring. For the most accurate calibration results, calibrate the thermocouple before performing the following procedure. Refer to page 60 for instructions. If you use a computer with the Model 900, you must unplug the connection before beginning the procedure. 1. Press the “CAL” button, the viscometer will display the “Zero Offset”. This shows how much the bob has drifted since the last calibration. If this value is greater than ±1, make sure the bob is properly installed on the unit and that it spins freely. Then rotate the bob by hand and let it go. If the Zero Offset is still greater than ±1, continue with the calibration. This will reset the Zero Offset. After verifying the offset, confirm that the P value is within the specified range. 2. When prompted, enter the viscosity of the calibration fluid. 3. Add approximately 170 mL calibration fluid into the clean and dry stainless steel sample cup. Place the sample cup with calibration fluid onto the platform. Loosen the lock nut on the platform with the stainless steel arm until the fluid level reaches the scribed line on the rotor. Tighten the lock nut on the platform. 4. Allow a couple of minutes for the temperature of the calibration fluid to equalize. 5. Find the calibration sheet attached to the calibration fluid you are using. Record the viscosity at the temperature being read on the unit. 6. Enter the actual viscosity (listed on the calibration sheet) of the calibration fluid (must be <= 300 cP if the standard B1 Bob, R1 Rotor and F1 Torsion Spring are used) using the left side of the keypad. Press ENTER. 7. The machine will perform the calibration internally in a series of sweeps, and return to the main screen. 8. Turn the unit OFF. 9. Loosen the lock nut on the platform while holding the bottom of the platform with your other hand. Slowly lower the platform with the sample cup and calibration fluid. Allow the stainless steel cup to remain beneath the bob and rotor to allow residual calibration fluid to drain. 10. Pour the calibration fluid from the sample cup back into the 16 oz. (500 mL) container. 11. Remove the rotor. Wipe any residual calibration fluid off the rotor and the bob with either a paper towel or a soft cloth. Wipe the sample cup dry with a paper towel.
  • 7. B. With computer Before the Model 900 unit can be calibrated using the ORCADA™ Software, the Calibration Mode must be set to “ORCADA Software”. From the main screen choose, “Calibrate Shear Stress” from the “Utilities” menu. 1. Select a calibration fluid from the “Cal Fluid Batch” drop-down list. If the list is empty, click the “Fluid Manager” button to add new fluid batches. If the “Temp Out of Range” light shines red, the current sample temperature is out of the specified range for the calibration fluid. The sample will have to be heated or cooled to be within the acceptable range before calibration. Do not attempt a calibration if the sample is not within the appropriate temperature range. 2. Choose a set of calibration rates from the “Rate Set” drop-down list. To create or edit a rate set, click the “Calibration Rates” button. To create a new set, click the “New” button and enter a name and the rotational speeds. To edit a set, select it in the “Rate Set” list, then change the rotational speeds as necessary. 3. Click the “Start Calibration” button to begin the calibration. Once the calibration has started, the software will begin filling in the chart and plotting the results on the graph. 4. As the calibration proceeds, the software will begin to display the collected data in the chart. At the end of the calibration, the software will calculate the “r^2” value, which measures the accuracy of the calibration. If this value is less than 0.9990, recalibrate the unit. If this value is still low, change the lower bearing and try the calibration again.
  • 8. Test Procedure The ORCADA™ software has two modes for running tests: Auto and Manual. By default, the software is in Manual mode when first opened. To begin a test in Manual mode: 1. Add approximately 170 mL of test fluid to the clean and dry stainless steel sample cup. 2. Enter the raw file save period (in MM:SS, where MM is the two digit number of minutest and SS is the two digit number of seconds) and select the shear stress units from the drop-down menu. 3. Enter a temperature in the “Temp” field and choose Celsius or Fahrenheit. 4. Select the “Enable” checkbox next to the “Rotation” field to start the motor. Once the motor is started, the display fields on the right-hand side of the screen will begin displaying test data. 5. Select the “Enable” checkbox next to the “Temp” field to start the heater. The “Preheat” and “Cool” options are used for pressurized viscometers only. 6. Enter a name in the “Experiment Name” field and click the “Start Logging” button. 7. Enter a comment in the resulting dialog box. Comments are optional, but can be used later to identify the test results. To begin a test in Auto mode: 1. Add approximately 170 mL of test fluid to the clean and dry stainless steel sample cup. 2. Enter the raw file save period (in MM:SS) and select the shear stress units from the drop-down menu. 3. Highlight a test to run from the list in the upper left-hand corner of the Main Screen. Refer to page 36 for instructions on creating custom tests. 4. Enter a name in the “Experiment Name” field. 5. Click the “Start Test” button. 6. Enter a comment in the resulting dialog box. Comments are optional, but can be used later to identify the test results.
  • 9. DISCUSION The Model 900 Viscometer is a coaxial cylinder rotational viscometer, which employs a transducer to measure the induced angle of rotation of the bob by a fluid sample Some point in this experiment we noticed, we can discuss like that: The Model 900 Viscometer is a portable, yet fully automated system form easuring fluid viscosity. It is designed to be easy to use, easy to maintain and to provide consistent results. Its simplicity makes it ideally suited for field applications. Engineers will appreciate the push button calibration, its reliability, and the ability to perform standard American Petroleum Institute (API) recommended practices with one command. With the addition of a computer, it is able to perform a variety of more complex laboratory tests. Routine repairs, like bearing and torsion spring replacement, may be performed by field personnel with a minimum of training, without having to return the viscometer for repair 1- The thermometer must be tighten firmly so as to avoid even a bit touch between the wall of the cup and the thermometer and to sink in the mud 2- With more heating the mud sample viscosity of the mud will decline gradually till it reach the critical rate such as showed in the curve diagram 3- Not to forget that every time the device must be calibrated with a special fluid which it’s with it and by a special method of calibration .
  • 10. References 1. Fjeldskaar, W. (1994). "Viscosity model 900 and thickness of the Science Letters 126 (4): 399–410. Bibcode:1994E&PSL.126..399F. doi:10.1016/0012- 821X(94)90120-1. 2. Einstein, A. (1906). "Eine neue Bestimmung der Moleküldimensionen". Annalen der Physik 19 (2): 289. Bibcode:1906AnP...324..289E. doi:10.1002/andp.19063240204. 3. Guth, E.; Simha, R. (1936). "Untersuchungen über die Viskosität von Suspensionen und Lösungen. 3. Über die Viskosität von Kugelsuspensionen". Kolloid Z. 74 (3): 266. doi:10.1007/BF01428643. 4. Thomas, D. G. (1965). "Transport characteristics of suspension: VIII. A note on the viscosity of model 900 Newtonian suspensions of uniform spherical particles". J. Colloid Sci. 20 (3): 267. doi:10.1016/0095-8522(65)90016-4. 5. Kitano, T.; Kataoka, T. & Shirota, T. (1981). "An empirical equation of the relative viscosity model 900 ". Rheologica Acta 20 (2): 207. doi:10.1007/BF01513064. 6. Fluegel, Alexander. "Viscosity calculation of glasses". Glassproperties.com. Retrieved 2010-09-14. 7. Doremus, R.H. (2002). "Viscosity of silica". J. Appl. Phys. 92 (12): 7619– 7629. Bibcode:2002JAP....92.7619D. doi:10.1063/1.1515132. 8. Ojovan, M.I. & Lee, W.E. (2004). "Viscosity of network liquids within Doremus approach". J. Appl. Phys. 95 (7): 3803