Gas chromatography-mass spectrometry (GC-MS) is the synergistic combination of two analytical method to separate and identify different substances within a test sample.
Gas chromatography separates the components of a mixture in time.
Mass spectrometer provides information that aids in the identification and structural elucidation of each component.
Difference Between Search & Browse Methods in Odoo 17
GC-MS Analysis Technique
1. Presented By-
Sudipta Nandi
M. Pharm 1st Sem
Department of Pharmaceutical Sciences
Dr. Harisingh Gour Vishwavidyalaya Sagar, M.P.
(A Central University)
2. Content
INTRODUCTION
HISTORICAL BACKGROUND OF GC-MS
PRINCIPLE
INSTRUMENTATION
• GAS CHROMATOGRAPHY
• INTERFACE
• MASS SPECTROMETER
WORKING
LIMITATIONS
APPLICATION
CONCLUSION
REFERENCES
3. INTRODUCTION
Gas chromatography-mass spectrometry (GC-MS) is the synergistic
combination of two analytical method to separate and identify different
substances within a test sample.
Gas chromatography separates the components of a mixture in time.
Mass spectrometer provides information that aids in the identification and
structural elucidation of each component.
HYPHENATED
TECHNIQUES
5. • Roland Gohlke and Fred mcLafferty introduce use of MS as
detector of GC in 1950s.
• Miniaturized computers has helped in the simplification of
instrument.
• In 1968, the Finnigan Instrument Corporation delivered the first
quadrupole GC-MS.
• By the 2000s computerized GC-MS instruments using quadrupole
technology had become essential.
• In 2005 GC tandem MS-MS have been introduced.
6. The sample solution is injected into the GC inlet where it is
vaporized and swept onto a chromatographic column by the
carrier gas.
The sample flows through column and compounds are
separated by virtue of their relative interaction with the
coating of the column and the carrier gas.
Latter part of column passes through a heated transfer line
and ends at entrance to ion source where compounds eluting
from the column are converted to ions and detected
according to their mass to charge m/z ratio.
7. INSTRUMENTATION LAYOUT
Gas
chromatography
• Carrier gas
• Pneumatic control
• Injector
• Column
• Oven
Interface
• Jet Interface
• Direct Capillary
Infusion Interface
• Watson-Biemann
effusion separator
Mass
spectrometer
• Ion source
• High-vacuum
system
• Mass analyzer
• Ion collector
Data
system
Fig : Principle of GC-MS
8.
9. 1. Carrier Gas :
Served as mobile phase supplied in the steel tank under high pressure.
At pressure of 40-80 psi this passes into flow controllers.
Example- nitrogen, helium; hydrogen and argon can also be used.
Requirements –
Inert
Column requirements
Detectors
Purity- better than 99.995%
Cost effective & available
10. 2. Pneumatic Control :
Gas supply is regulated to the correct pressure and then fed to the required
part of instrument.
Older instruments – manual pressure control via regulators.
Modern GC instruments – Electronic Pneumatic pressure controller.
3. Oven :
Temperature programmable, typically range from 5°C - 400°C but can go as
low as -25°C with cryogenic cooling.
11. 4. Sample Injection Port :
Sample is made to vaporized rapidly before entering to column.
Various kind of injectors :
Packed column injectors
Split injection
Splitless injection
Programmed Split/Splitless injection
Programmed On-Column injector
5. Column :
Two kinds of column used :
Packed column
Capillary column
Gas chromatograph GC-MS utilizes capillary column.
Here, stationary phase has been chemically bonded to the fused silica, e.g.,
DB-5.
12. Packed Column :-
• Less commonly used, having
diameter of 2-3 mm and length of
0.5-10 m.
• Manufactured from steel or glass,
internal wall is treated to avoid
catalytic effect with the sample.
• They can withstand a carrier gas
flow rate within range 10-
40ml/min.
Capillary Column :-
• Consist of long capillary tubing
10-100 m in length.
• Made up from stainless steel &
coil.
• Shorter columns - for fast analysis
• Larger columns – for high
resolution separation
13. 6. Detectors :
Simple and reliable
Sensitive to electronegative groups (halogens)
Largely non-destructive
Limited dynamic range (10²)
Mass sensitive detectors
-Thermal Conductivity Detector (TCD)
-Flame Ionization Detector (FID)
-Electron Capture Detector (ECD)
14. INTERFACE
Pressure incompatibility problem between GC and MS was solved by
inserting an interface.
Mainly three types of interfaces are there commercially available :
Jet/Orifice separator
Direct capillary infusion interface
Watson-Biemann effusion separator
Fig : Direct capillary infusion
Fig : Jet separator Fig : Watson-Biemann effusion
15. MASS SPECTROMETER
“ Mass spectrometry is a technique used for measuring the molecular weight
and determining the molecular formula of an organic compound.”
In general a mass spectrometer consists of
An ion source
High-vacuum system
A mass-selective analyzer
An ion collector
Data system
16. Ionization Method Typical Analytes Sample
Introduction
Mass
Range
Method
Highlights
Chemical Ionization (CI) Relatively small,
volatile
GC or
liquid/solid
probe
Upto
1000
Daltons
Soft method,
molecular ion peak
[M+H] +
Electron Impact
Ionization (EI)
Relatively small,
volatile
GC or
liquid/solid
probe
Upto
1000
Daltons
Hard method,
versatile, provides
structure info
Electrospray Ionization
(ESI)
Peptides, proteins,
nonvolatile
Liquid
chromatography
Upto
200000
Daltons
Soft method, ions
often multiply
charged
Fast Atom Bombardment
(FAB)
Carbohydrates,
organometallics,
peptides,
nonvolatile
Sample mixed in
viscous matrix
Upto
6000
Daltons
Harder than ESI or
MALDI
Matrix Assisted Laser
Desorption Ionization
(MALDI)
Peptides, proteins,
nucleotides
Sample mixed in
solid matrix
Upto
500000
Daltons
Soft method, very
high mass
17. Mass Selective Analyzer
• They deflects ions down a curved tubes in a magnetic fields based on their
kinetic energy determined by the mass, charge and velocity.
• The magnetic field is scanned to measure different ions.
Mass Analyzers
Quadrupole Ion Trap Time of flight
18. Quadrupole Analyzer :
• Also known as ‘Hewlett-Packard’ or ‘Mass Selective Detector’.
• In quadrupole mass analyser a set of four rods are arranged parallel to the
direction.
• Only m/z is been determined and stable oscillation takes place.
• Ions travels in quadrupole axis with cork screw type of trajectory.
• It functions as a mass filter
19. Ion Trap Mass Analyzer :
The ion trap mass analyzer operates by similar principles where it consists
of circular ring electrode and two end caps that form a chamber.
AC or DC power along RF potential is applied between the cups and the
ring electrode.
Ions entering into
chamber
Trapped by
electromagnetic fields
Ions oscillate in concentric
trajectories
Resonant Ejection
20. Time-of-flight Analyzer :
The time-of-flight (TOF) analyser uses an electric field to accelerate the
ions through the same potential.
Then it measures the time take to reach the detector.
If the particles all have the same charge, the kinetic energies will be
identical and their velocities will depend upon only on their masses.
21. Mass spectrometers employ computer control of same functions and also
use a computerised display and output.
It is used to identify and measure the concentration of one or more analytes
in a complex mixture
Advantages –
Amount of data generated is very large
It stores every sec for upto 90 min.
Data Handling
MASS
CHROMATOGRAMS
SELECTED ION
MONITORING
22. GC TRACE OF A THREE COMPONENT MIXTURE.
THE MASS SPECTROMETER GIVES A SPECTRUM FOR
EACH COMPONENT
24. LIMITATIONS
• Only compounds with vapour pressure exceeding about 10-10 torr can be
analysed by gas chromatography-mass spectrometry (GC-MS).
• Determining positional substitution on aromatic ring is often difficult.
• Certain isomeric compounds cannot be distinguished by mass spectrometry.
• Non-volatile matrices require additional preparation (extraction,
outgassing, etc.)
• Atmospheric gases are challenging (CO2, N2, O2, Ar, CO, H2O).
25. Applications of GC-MS
Petrochemical and hydrocarbon analysis
Geochemical research
Forensic (arsons, explosives, drugs, unknowns)
Environmental analysis
Pesticide analysis, food safety and quality
Clinical toxicology
Food, beverage and perfume analysis
26. CONCLUSION
As we can see, GC-MS is used both in qualitative and quantitative
measurement.
Being hyphenated technique, the cost of GC-MS becomes very high.
Nowadays, cost is also cheaper than before as quadrupole and ion trap
instruments are used, which does not only save space but also save
manufacturing cost.
GC-MS is an analytical method of increased sensitivity and reliability even
in very small quantities (ng).
27. Braun R., Introduction To Instrumental Analysis, Second
Edition, PharmaMed Press, Hyderabad, Page no. 251-270.
Chatwal G.R., Anand S.K., Instrumental method of Chemical
Analysis, Himalaya Publishing House, Fifth Edition-2012,
New Delhi, Page no. 420-449.
Gohlke R.S.(1959), Analytical Chemistry, 36, Page no. 759-
764.