UV-Visible Spectrophotometry
Calibration of Spectrophotometer
Essentials
Regulations for quality control such as ISO 9000, Good
Laboratory Practice (GLP), Good Manufacturing
Practice (GMP) or standard operating procedures in the
pharmaceutical industry as required by the most
important pharmacopeia's (EP, DAB, USP), require a
regular performance testing of the UV/Vis
spectrophotometers
Performance qualification
Performance qualification Following eight criteria should be
verified with help of standard reference in calibration of uv
spectrophotometer
Wavelength Accuracy
Stray Light
Resolution
Noise
Baseline Flatness
Stability
Photometric Accuracy
Linearity
Wavelength Accuracy
• Wavelength Accuracy Wavelength accuracy is
defined as the deviation of the wavelength
reading at an absorption band or emission
band from the known wavelength of the band.
• The wavelength deviation can cause ignificant
errors in the qualitative and quantitative
results of the UV–Vis measurement.
It is quite obvious that if the spectrophotometer
is not able to maintain an accurate wavelength
scale, the UV absorption profile of the sample
measured by the instrument will be inaccurate.
The true λ max and λ min of the analyte cannot
be characterized accurately .
Protocol for Checking Wavelength
Accuracy
Wavelength accuracy verification is checked by
measuring a known wave- length reference
standard with well-characterized absorption or
emission peaks and comparing the recorded
wavelength of the peak(s) against the value (s)
listed in the certificate of that reference standard.
There are many standards that are commonly used
to verify the wavelength accuracy of a
spectrophotometer
• Spectra of some commonly used wavelength
standards such as a deuterium lamp, mercury
vapour lamp, holmium oxide filter, and
holmium oxide solution (4% holmium oxide in
10% perchloric acid in a 1-cm cell ).
• Acceptance. ±1 nm in the UV range (200 to
380 nm) and ±3 nm in the visible range (380 to
800 nm). Three repeated scans of the same
peak should be within ±0.5nm .
Stray Light
• Stray Light Stray light is defined as the detected
light of any wavelength that is outside the
bandwidth of the wavelength selected.
• The causes for stray light are scattering, higher-
order diffraction of the monochromator , or poor
instrument design.
• Stray light causes a decrease in absorbance and
reduces the linear range of the instrument.
• High-absorbance measurements are affected
more severely by stray light .
• The stray light problem causes a deviation from
linearity at high absorbance
PROTOCOL
For the stray light test, various cut-off filters or
solutions can be used to estimate the stray light
contribution, depending on the wavelengths is
used.
Scan the stray light testing solution in a 1-cm cell
using air as the reference.
The absorption of a solution of potassium
chloride R (12 g/l) between 220 nm and 200 nm
in a light path of 1 cm must rise steeply and at a
wavelength of 198 nm must be greater than 2,
measured against water R as a compensation
liquid
Resolution
• Resolution The resolution of a UV-Vis
spectrophotometer is related to its spectral
bandwidth (SBW). The smaller bandwidth, the
finer the resolution. The SBW depends on the
slit width and the dispersive power of the
monochrometer . Typically, only
spectrophotometers designed for high-
resolution work have a variable slit width.
Spectrophotometers for routine analysis usually
have a fixed slit width.
The resolution of the absorbance measurement
depends on the ratio of the spectral bandwidth
(SBW) of the spectrophotometer to the natural
bandwidth (NBW) of the spectral band to be
measured.
PROTOCOL
The resolution of the absorbance measurement
depends on the ratio of the spectral bandwidth
(SBW) of the spectrophotometer to the natural
bandwidth (NBW) of the spectral band to be
measured
Acceptance : The ratio of the absorbance at λmax
(269 nm) and absorbance at λmin (266 nm) should
be greater than 1.5.
Noise
Noise in the UV-Vis measurement originates
primarily from the light source and electronic
components.
Noise in the measurement affects the accuracy
at both ends of the absorbance scale.
Photon noise from the light source affects the
accuracy of the measurements at low
absorbance.
• Electronic noise from the electronic
components affects the accuracy of the
measurements at high absorbance.
• A high noise level affects the precision of the
measurements and reduces the limit of
detection, thereby rendering the instrument
less sensitive.
PROTOCOL
Air is scanned in the absorbance mode for 10 min.
peak-to-peak noise is recorded at 500 nm.
The root mean square (RMS) noise is then
calculated. Modern spectrophotometers are usually
equipped with the noise estimation function.
Acceptance: The RMS noise should typically be less
than 0.001 AU
Baseline Flatness
Baseline Flatness The intensity of radiation
coming from the light source varies over the
entire UV-Vis range. Most UV-Vis
spectrophotometers have dual light sources.
A deuterium lamp is used for the UV range and a
tungsten lamp is used for the visible range. The
response of the detector also varies over the
spectral range.
The flat baseline test demonstrates the ability of
the instrument to normalize the light intensity
measurement and the spectral output at
different wavelengths throughout the spectral
range .
PROTOCOL
Air is scanned in the absorbance mode. The highest
and lowest deflections in the absorbance unit are
recorded.
Acceptance: The deflection is typically less than
0.01 AU.
Stability
Stability Variations in lamp intensity and electronic
output between the measurements of the
reference and the sample result in instrument drift.
The lamp intensity is a function of the age of the
lamp, temperature fluctuation, and wavelength of
the measurement.
These changes can lead to errors in the value of the
measurements, especially over an extended period
of time.
The resulting error in the measurement may be
positive or negative.
The stability test checks the ability of the
instrument to maintain a steady state over time
so that the effect of the drift on the accuracy of
the measurements is insignificant.
PROTOCOL
Air is scanned in the absorbance mode for 60 min at
a specific wavelength (typically, 340 nm). The
highest and lowest deflections in the absorbance
unit are recorded.
Acceptance: The deflection is typically less than
0.002 AU/h .
Photometric Accuracy
Photometric accuracy is determined by comparing
the difference between the measured absorbance
of the reference standard materials and the
established standard value.
An optically neutral material with little wavelength
dependency for its transmittance/absorbance is
desirable because it eliminates the spectral
bandwidth dependency of measurements.
PROTOCOL
In that 0.006 % w/v solution of potassium dichromate
(60.06 mg/L) is prepared in 0.005 M sulphuric acid. A
0.005 M sulphuric acid solution as the reference is
scanned and then the potassium dichromate solutions
are scanned at 235, 257, 313, and 350 nm.
Acceptance : Six replicate measurements of the 0.006%
w/v potassium dichromate solution at 235, 257, 313, and
350 nm should be less than 0.5% RSD.
Linearity
Linearity The linear dynamic range of the
measurement is limited by stray light at high
absorbance and by noise at low absorbance.
For routine measurements involving samples
and the related reference chemical standards,
the accuracy of the quantification of the sample
depends on the precision and linearity of the
measurements
PROTOCOL
A series of potassium dichromate solutions of
concentration 20, 40, 60, 80, and 100 mg/L in
0.005 M sulphuric acid can be used to test the
linearity of the system.
First, 0.005 M sulphuric acid as reference is
scanned and then the potassium dichromate
solutions at 235, 257, 313, and 350 nm.
• The absorbance values at various wavelengths
are plotted against the concentration of the
solutions and the correlation coefficients are
calculated.
• Acceptance: Correlation coefficient r = 0.999
Ideally used type of filters
Usually by trial and error following type of filters
are considered ideal for working with all kinds of
spectrophotometric instruments.
For checking Wavelength accuracy, Photometric
accuracy, Stray light behaviour & Resolution Liquid
filters are more reliable whereas for visible ranges
solid filters are preferred to calibrate the instrument
for Wavelength and Photometric accuracy
Holmium oxide liquid filter
(Holmium oxide dissolved in perchloric
acid)
• The exact position of
peaksat 241.15 nm,
287.15 nm,361.5 nm and
536.3 nm is determined
and compared with the
values in the certificate.
• Deviations should not be
greater than ± 1 nm in the
ultra-violet and ± 3 nm in
the visible range
Holmium Oxide Spectra
Slit width= 1nm (or to be kept as low as possible)
Testing the wavelength accuracy in the UV and visible range
between 240 nm and 650 nm for spectrophotometers with a spectral bandwidth (slit)
between 0.1 nm and 3 nm
Holmium Oxide Glass doped Filter and its spectra
(2.1 mm thickness)
Didymium oxide
• Generally available as glass filter is more
reliable for routine calibration made out of
special grade glass.
• Testing the wavelength accuracy in the UV and
Vis region (329 nm to 875 nm) at a spectral
bandwidth to 2 nm.
Spectra of Didymium Glass Filter (spectral bandwidth to 2 nm)
Absorbs at-
329 nm
472 nm
512 nm
681 nm
875 nm
Potassium Dichromate
Potassium dichromate dissolved in perchloric
acid is suitable for testing the photometric
accuracy (absorbance) of spectrophotometers
in the UV spectrum.
The absorption values of the cuvette are
measured at the wavelengths 235 nm, 257 nm,
313 nm and 350 nm and recorded in the
certificate
Typical spectrum of a 0.006 % potassium dichromate solution (spectral
bandwidth of 2 nm)
Absorbs at
235 nm
257 nm
313 nm
350 nm
Potassium dichromate liquid filter
(430 nm)
Neutral density glass filter
Absorbs in visible region (440 nm to 635 nm)
Made of SCHOTT Glasss
Cut-Off filters
Spectra of various cut-off filters
Testing for stray light in the UV spectrum (at wavelengths
from 198 nm to 385 nm, depending on the selected filter)
Application
The photometer is set to a starting wavelength that
is 30 nm above the cut-off wavelength of the filter to
be used. If, for example, potassium chloride is used
then the start is at 230 nm.
Then a scan is carried out down towards shorter
wavelengths and the absorption below the cut-off
wavelength is observed. The transmission that is
measured below the cut-off wavelength is stray light
Calibration standards for testing
the resolution
The resolution for modern photometers
depends directly on the slit width. The smaller
the slit and the associated spectral bandwidth,
the higher the resolution.
The choice of material should be made as per
the standards applicable to the analysis or as
mentioned on the format for certificate of
analysis or the manual.
In order to determine the resolution of a spectrophotometer, the behaviour of the
absorption of a maximum at λmax (269 nm) to absorption of a minimum at λmin (266 nm) is
shown.
From this, one obtains details of the spectral bandwidth of the instrument used.
Wavelength spectrum from 265 nm to 270 nm is scanned and the peak maximum and
peak minimum are determined. Then the relationship between the two is calculated
Two type of standards are used-
1. Toluene in n-Hexane
2. Hexane as reference
Control of Absorbance
(Potassium Dichromate method)
• Dry a quantity of Potassium Dichromate by
heating to constant weight at 1300C .
• Weigh and transfer accurately a quantity not less
than 57.0mg and not more than 63 mg to
10000ml volumetric flask. Dissolve and dilute in
sufficient 0.005M H2SO4 to produce 1000ml.
• Measure the absorbance of Potassium
Dichromate solution at wavelength given in the
table
• Calculate value of {A1%
1cm } for each
wavelengths mentioned.
• Acceptance Criteria: (as per I.P and B.P)
Wavelengths(in nm) A(1%/1cm) Limit
235.0 124.5 129.9-126.2
257.0 144.0 142.8-145.7
313.0 48.6 47.0-50.3
350.0 106.6 105.6-108.2
Resolution power
• Record the spectrum of a 0.02% v/v solution of
toluene in hexane in the range of 260nm to 420
nm (before use check the hexane for
transmittance, using water as blank between
260nm to 420 nm & use only if transmittance is
not less than 97%)
• Acceptance Criteria: The ratio of absorbance at
maximum at about 269 nm to that at the
minimum at about 266 nm is not less than 1.5
Limit of Stray Light
• Prepare a 1.2% w/v solution of potassium
chloride in water.
• Measure the absorbance of the above
solution at solution at 198.0, 199.0, 200.0,
201.0 and 202.0 using water as blank.
• Acceptance Criteria:
Absorbance is greater than 2
Wavelength accuracy, Resolution &
baseline flatness
• These are instruments in-built functions
• The note made here is in accordance to
Shimadzu UV-1700 UV-Vis Spectrophotometer
• Attach printer directly to the instrument
(optional)
• Goto MODE and press F3 key to opt for
maintenance mode.
• Press 1 (…continued)
• Press Start/Stop key.
• After screen changes, ensure that nothing is
kept in the optical path & press the start/stop
key again.
• Prints will come after all the tests are over.
• Acceptance Criteria:
For wavelength accuracy: at 656.1±0.3 nm &
at 486.0±0.3 nm.
Resolution:1.0 nm or less.
Baseline flatness: ±0.002Abs.
Reference
• #1 Indian Pharmacopoeia
• #2 U.S. Pharmacopoeia
• #3 Pharmaguideline
• #4 Hellma GmBH product Catalogue
• #5 Schimadzu UV-1700 Users manual
That was
boooooringgggg!!!

Más contenido relacionado

La actualidad más candente

Instrumentation of uv spectroscopy
Instrumentation of uv spectroscopyInstrumentation of uv spectroscopy
Instrumentation of uv spectroscopy
Arabinda Changmai
 
Method validation for drug substances and drug product _remodified_2014
Method validation for drug substances and drug product _remodified_2014Method validation for drug substances and drug product _remodified_2014
Method validation for drug substances and drug product _remodified_2014
Ramalingam Badmanaban
 

La actualidad más candente (20)

DETECTORS IN HPLC
DETECTORS IN HPLCDETECTORS IN HPLC
DETECTORS IN HPLC
 
Analytical Method Development and validation of UV-Visible spectroscopy
Analytical Method Development and validation of UV-Visible spectroscopyAnalytical Method Development and validation of UV-Visible spectroscopy
Analytical Method Development and validation of UV-Visible spectroscopy
 
calibration of analytical instruments
calibration of analytical instruments calibration of analytical instruments
calibration of analytical instruments
 
Instrumentation of uv spectroscopy
Instrumentation of uv spectroscopyInstrumentation of uv spectroscopy
Instrumentation of uv spectroscopy
 
Notes* for the subject 'Advanced Pharmaceutical Analysis'
Notes* for the subject 'Advanced Pharmaceutical Analysis'Notes* for the subject 'Advanced Pharmaceutical Analysis'
Notes* for the subject 'Advanced Pharmaceutical Analysis'
 
Detectors hplc
Detectors hplcDetectors hplc
Detectors hplc
 
Spectrophotometric Instruments (Detector) and Application of UV – VIS spectro...
Spectrophotometric Instruments (Detector) and Application of UV – VIS spectro...Spectrophotometric Instruments (Detector) and Application of UV – VIS spectro...
Spectrophotometric Instruments (Detector) and Application of UV – VIS spectro...
 
RADIO IMMUNO ASSAY(RIA) BY P.RAVISANKAR
RADIO IMMUNO ASSAY(RIA) BY P.RAVISANKARRADIO IMMUNO ASSAY(RIA) BY P.RAVISANKAR
RADIO IMMUNO ASSAY(RIA) BY P.RAVISANKAR
 
DETECTORS USED IN GAS CHROMATOGRAPHY AND HPLC BY P.RAVISANKAR.
DETECTORS USED IN GAS CHROMATOGRAPHY AND HPLC BY P.RAVISANKAR.DETECTORS USED IN GAS CHROMATOGRAPHY AND HPLC BY P.RAVISANKAR.
DETECTORS USED IN GAS CHROMATOGRAPHY AND HPLC BY P.RAVISANKAR.
 
Spectroflourimetry
SpectroflourimetrySpectroflourimetry
Spectroflourimetry
 
Gas chromatography
Gas chromatographyGas chromatography
Gas chromatography
 
Qualification of analytical instruments
Qualification of analytical instrumentsQualification of analytical instruments
Qualification of analytical instruments
 
Detectors in GC
Detectors in GCDetectors in GC
Detectors in GC
 
High Performance Thin Layer Chromatography (HPTLC) instrumentation
High Performance Thin Layer Chromatography (HPTLC) instrumentationHigh Performance Thin Layer Chromatography (HPTLC) instrumentation
High Performance Thin Layer Chromatography (HPTLC) instrumentation
 
Method validation for drug substances and drug product _remodified_2014
Method validation for drug substances and drug product _remodified_2014Method validation for drug substances and drug product _remodified_2014
Method validation for drug substances and drug product _remodified_2014
 
FLUORIMETRY
FLUORIMETRYFLUORIMETRY
FLUORIMETRY
 
Derivatisation Techniques in Gas chromatography
Derivatisation Techniques in Gas chromatographyDerivatisation Techniques in Gas chromatography
Derivatisation Techniques in Gas chromatography
 
FLAME EMISSION SPECTROSCOPY
FLAME EMISSION SPECTROSCOPY FLAME EMISSION SPECTROSCOPY
FLAME EMISSION SPECTROSCOPY
 
Analytical method development,validation by uv spectroscopy
Analytical method development,validation by uv spectroscopyAnalytical method development,validation by uv spectroscopy
Analytical method development,validation by uv spectroscopy
 
High Performance Thin Layer Chromatography(HPTLC)
High Performance Thin Layer Chromatography(HPTLC)High Performance Thin Layer Chromatography(HPTLC)
High Performance Thin Layer Chromatography(HPTLC)
 

Similar a Calibration of spectrophotometer

Instruments & Application
Instruments & ApplicationInstruments & Application
Instruments & Application
FirasMfarrej
 

Similar a Calibration of spectrophotometer (20)

Wavelength Accuracy.pptx
Wavelength Accuracy.pptxWavelength Accuracy.pptx
Wavelength Accuracy.pptx
 
Validation of uv visible spectrophotometer.pptx
Validation of uv visible spectrophotometer.pptxValidation of uv visible spectrophotometer.pptx
Validation of uv visible spectrophotometer.pptx
 
Spectrophotometer.pptx
Spectrophotometer.pptxSpectrophotometer.pptx
Spectrophotometer.pptx
 
QUALIFICATION OF UV-VISIBLE SPECTROPHOTOMETER, FTIR, DSC, HPLC
QUALIFICATION OF UV-VISIBLE SPECTROPHOTOMETER, FTIR, DSC, HPLCQUALIFICATION OF UV-VISIBLE SPECTROPHOTOMETER, FTIR, DSC, HPLC
QUALIFICATION OF UV-VISIBLE SPECTROPHOTOMETER, FTIR, DSC, HPLC
 
Uv vis spec
Uv vis specUv vis spec
Uv vis spec
 
Instrumentation of uv spectrophotometer pranav kumar singh
Instrumentation of uv spectrophotometer   pranav kumar singhInstrumentation of uv spectrophotometer   pranav kumar singh
Instrumentation of uv spectrophotometer pranav kumar singh
 
UV-VIS SPECTRO final.ppt
UV-VIS SPECTRO final.pptUV-VIS SPECTRO final.ppt
UV-VIS SPECTRO final.ppt
 
QUALIFICATION OF UV-VISIBLE SPECTROPHOTOMETER
QUALIFICATION OF UV-VISIBLE SPECTROPHOTOMETER QUALIFICATION OF UV-VISIBLE SPECTROPHOTOMETER
QUALIFICATION OF UV-VISIBLE SPECTROPHOTOMETER
 
Molecular Spectroscopy
Molecular SpectroscopyMolecular Spectroscopy
Molecular Spectroscopy
 
performance and specifications of spectrophotometer
performance and specifications of spectrophotometerperformance and specifications of spectrophotometer
performance and specifications of spectrophotometer
 
Analytical instrumentation
Analytical instrumentationAnalytical instrumentation
Analytical instrumentation
 
Infrared Pyrometers for Industrial Use
Infrared Pyrometers for Industrial UseInfrared Pyrometers for Industrial Use
Infrared Pyrometers for Industrial Use
 
Introduction to UV-based detectors
Introduction to UV-based detectors Introduction to UV-based detectors
Introduction to UV-based detectors
 
Calibration - UV VIS Spectrophotometer, HPLC, Gas Chromatograph, IR spectroph...
Calibration - UV VIS Spectrophotometer, HPLC, Gas Chromatograph, IR spectroph...Calibration - UV VIS Spectrophotometer, HPLC, Gas Chromatograph, IR spectroph...
Calibration - UV VIS Spectrophotometer, HPLC, Gas Chromatograph, IR spectroph...
 
Photometry by Dr. Anurag Yadav
Photometry by Dr. Anurag YadavPhotometry by Dr. Anurag Yadav
Photometry by Dr. Anurag Yadav
 
Instruments & Application
Instruments & ApplicationInstruments & Application
Instruments & Application
 
Qualification of UV VISIBLE SPECTROPHOTOMETER
Qualification of UV VISIBLE SPECTROPHOTOMETERQualification of UV VISIBLE SPECTROPHOTOMETER
Qualification of UV VISIBLE SPECTROPHOTOMETER
 
Spectrophotometry july 2018 nitub
Spectrophotometry july 2018 nitubSpectrophotometry july 2018 nitub
Spectrophotometry july 2018 nitub
 
Colorimetry class
Colorimetry classColorimetry class
Colorimetry class
 
Uv spectroscopy instrumentation, by dr. umesh kumar sharma & amp; shyma m s
Uv spectroscopy instrumentation, by dr. umesh kumar sharma & amp; shyma m sUv spectroscopy instrumentation, by dr. umesh kumar sharma & amp; shyma m s
Uv spectroscopy instrumentation, by dr. umesh kumar sharma & amp; shyma m s
 

Último

Dehradun Call Girls Service {8854095900} ❤️VVIP ROCKY Call Girl in Dehradun U...
Dehradun Call Girls Service {8854095900} ❤️VVIP ROCKY Call Girl in Dehradun U...Dehradun Call Girls Service {8854095900} ❤️VVIP ROCKY Call Girl in Dehradun U...
Dehradun Call Girls Service {8854095900} ❤️VVIP ROCKY Call Girl in Dehradun U...
Sheetaleventcompany
 
Difference Between Skeletal Smooth and Cardiac Muscles
Difference Between Skeletal Smooth and Cardiac MusclesDifference Between Skeletal Smooth and Cardiac Muscles
Difference Between Skeletal Smooth and Cardiac Muscles
MedicoseAcademics
 
Premium Call Girls Nagpur {9xx000xx09} ❤️VVIP POOJA Call Girls in Nagpur Maha...
Premium Call Girls Nagpur {9xx000xx09} ❤️VVIP POOJA Call Girls in Nagpur Maha...Premium Call Girls Nagpur {9xx000xx09} ❤️VVIP POOJA Call Girls in Nagpur Maha...
Premium Call Girls Nagpur {9xx000xx09} ❤️VVIP POOJA Call Girls in Nagpur Maha...
Sheetaleventcompany
 
Cara Menggugurkan Kandungan Dengan Cepat Selesai Dalam 24 Jam Secara Alami Bu...
Cara Menggugurkan Kandungan Dengan Cepat Selesai Dalam 24 Jam Secara Alami Bu...Cara Menggugurkan Kandungan Dengan Cepat Selesai Dalam 24 Jam Secara Alami Bu...
Cara Menggugurkan Kandungan Dengan Cepat Selesai Dalam 24 Jam Secara Alami Bu...
Cara Menggugurkan Kandungan 087776558899
 
❤️Amritsar Escorts Service☎️9815674956☎️ Call Girl service in Amritsar☎️ Amri...
❤️Amritsar Escorts Service☎️9815674956☎️ Call Girl service in Amritsar☎️ Amri...❤️Amritsar Escorts Service☎️9815674956☎️ Call Girl service in Amritsar☎️ Amri...
❤️Amritsar Escorts Service☎️9815674956☎️ Call Girl service in Amritsar☎️ Amri...
Sheetaleventcompany
 
💚Call Girls In Amritsar 💯Anvi 📲🔝8725944379🔝Amritsar Call Girl No💰Advance Cash...
💚Call Girls In Amritsar 💯Anvi 📲🔝8725944379🔝Amritsar Call Girl No💰Advance Cash...💚Call Girls In Amritsar 💯Anvi 📲🔝8725944379🔝Amritsar Call Girl No💰Advance Cash...
💚Call Girls In Amritsar 💯Anvi 📲🔝8725944379🔝Amritsar Call Girl No💰Advance Cash...
Sheetaleventcompany
 
💚Chandigarh Call Girls Service 💯Piya 📲🔝8868886958🔝Call Girls In Chandigarh No...
💚Chandigarh Call Girls Service 💯Piya 📲🔝8868886958🔝Call Girls In Chandigarh No...💚Chandigarh Call Girls Service 💯Piya 📲🔝8868886958🔝Call Girls In Chandigarh No...
💚Chandigarh Call Girls Service 💯Piya 📲🔝8868886958🔝Call Girls In Chandigarh No...
Sheetaleventcompany
 
Goa Call Girl Service 📞9xx000xx09📞Just Call Divya📲 Call Girl In Goa No💰Advanc...
Goa Call Girl Service 📞9xx000xx09📞Just Call Divya📲 Call Girl In Goa No💰Advanc...Goa Call Girl Service 📞9xx000xx09📞Just Call Divya📲 Call Girl In Goa No💰Advanc...
Goa Call Girl Service 📞9xx000xx09📞Just Call Divya📲 Call Girl In Goa No💰Advanc...
Sheetaleventcompany
 
Ahmedabad Call Girls Book Now 8980367676 Top Class Ahmedabad Escort Service A...
Ahmedabad Call Girls Book Now 8980367676 Top Class Ahmedabad Escort Service A...Ahmedabad Call Girls Book Now 8980367676 Top Class Ahmedabad Escort Service A...
Ahmedabad Call Girls Book Now 8980367676 Top Class Ahmedabad Escort Service A...
Genuine Call Girls
 
Gorgeous Call Girls Dehradun {8854095900} ❤️VVIP ROCKY Call Girls in Dehradun...
Gorgeous Call Girls Dehradun {8854095900} ❤️VVIP ROCKY Call Girls in Dehradun...Gorgeous Call Girls Dehradun {8854095900} ❤️VVIP ROCKY Call Girls in Dehradun...
Gorgeous Call Girls Dehradun {8854095900} ❤️VVIP ROCKY Call Girls in Dehradun...
Sheetaleventcompany
 

Último (20)

Call Girl In Chandigarh 📞9809698092📞 Just📲 Call Inaaya Chandigarh Call Girls ...
Call Girl In Chandigarh 📞9809698092📞 Just📲 Call Inaaya Chandigarh Call Girls ...Call Girl In Chandigarh 📞9809698092📞 Just📲 Call Inaaya Chandigarh Call Girls ...
Call Girl In Chandigarh 📞9809698092📞 Just📲 Call Inaaya Chandigarh Call Girls ...
 
Chandigarh Call Girls Service ❤️🍑 9809698092 👄🫦Independent Escort Service Cha...
Chandigarh Call Girls Service ❤️🍑 9809698092 👄🫦Independent Escort Service Cha...Chandigarh Call Girls Service ❤️🍑 9809698092 👄🫦Independent Escort Service Cha...
Chandigarh Call Girls Service ❤️🍑 9809698092 👄🫦Independent Escort Service Cha...
 
❤️Call Girl Service In Chandigarh☎️9814379184☎️ Call Girl in Chandigarh☎️ Cha...
❤️Call Girl Service In Chandigarh☎️9814379184☎️ Call Girl in Chandigarh☎️ Cha...❤️Call Girl Service In Chandigarh☎️9814379184☎️ Call Girl in Chandigarh☎️ Cha...
❤️Call Girl Service In Chandigarh☎️9814379184☎️ Call Girl in Chandigarh☎️ Cha...
 
Kolkata Call Girls Shobhabazar 💯Call Us 🔝 8005736733 🔝 💃 Top Class Call Gir...
Kolkata Call Girls Shobhabazar  💯Call Us 🔝 8005736733 🔝 💃  Top Class Call Gir...Kolkata Call Girls Shobhabazar  💯Call Us 🔝 8005736733 🔝 💃  Top Class Call Gir...
Kolkata Call Girls Shobhabazar 💯Call Us 🔝 8005736733 🔝 💃 Top Class Call Gir...
 
Dehradun Call Girls Service {8854095900} ❤️VVIP ROCKY Call Girl in Dehradun U...
Dehradun Call Girls Service {8854095900} ❤️VVIP ROCKY Call Girl in Dehradun U...Dehradun Call Girls Service {8854095900} ❤️VVIP ROCKY Call Girl in Dehradun U...
Dehradun Call Girls Service {8854095900} ❤️VVIP ROCKY Call Girl in Dehradun U...
 
Difference Between Skeletal Smooth and Cardiac Muscles
Difference Between Skeletal Smooth and Cardiac MusclesDifference Between Skeletal Smooth and Cardiac Muscles
Difference Between Skeletal Smooth and Cardiac Muscles
 
Low Cost Call Girls Bangalore {9179660964} ❤️VVIP NISHA Call Girls in Bangalo...
Low Cost Call Girls Bangalore {9179660964} ❤️VVIP NISHA Call Girls in Bangalo...Low Cost Call Girls Bangalore {9179660964} ❤️VVIP NISHA Call Girls in Bangalo...
Low Cost Call Girls Bangalore {9179660964} ❤️VVIP NISHA Call Girls in Bangalo...
 
Call Girls Rishikesh Just Call 9667172968 Top Class Call Girl Service Available
Call Girls Rishikesh Just Call 9667172968 Top Class Call Girl Service AvailableCall Girls Rishikesh Just Call 9667172968 Top Class Call Girl Service Available
Call Girls Rishikesh Just Call 9667172968 Top Class Call Girl Service Available
 
Premium Call Girls Nagpur {9xx000xx09} ❤️VVIP POOJA Call Girls in Nagpur Maha...
Premium Call Girls Nagpur {9xx000xx09} ❤️VVIP POOJA Call Girls in Nagpur Maha...Premium Call Girls Nagpur {9xx000xx09} ❤️VVIP POOJA Call Girls in Nagpur Maha...
Premium Call Girls Nagpur {9xx000xx09} ❤️VVIP POOJA Call Girls in Nagpur Maha...
 
Call Girls in Lucknow Just Call 👉👉 8875999948 Top Class Call Girl Service Ava...
Call Girls in Lucknow Just Call 👉👉 8875999948 Top Class Call Girl Service Ava...Call Girls in Lucknow Just Call 👉👉 8875999948 Top Class Call Girl Service Ava...
Call Girls in Lucknow Just Call 👉👉 8875999948 Top Class Call Girl Service Ava...
 
Cara Menggugurkan Kandungan Dengan Cepat Selesai Dalam 24 Jam Secara Alami Bu...
Cara Menggugurkan Kandungan Dengan Cepat Selesai Dalam 24 Jam Secara Alami Bu...Cara Menggugurkan Kandungan Dengan Cepat Selesai Dalam 24 Jam Secara Alami Bu...
Cara Menggugurkan Kandungan Dengan Cepat Selesai Dalam 24 Jam Secara Alami Bu...
 
Kolkata Call Girls Naktala 💯Call Us 🔝 8005736733 🔝 💃 Top Class Call Girl Se...
Kolkata Call Girls Naktala  💯Call Us 🔝 8005736733 🔝 💃  Top Class Call Girl Se...Kolkata Call Girls Naktala  💯Call Us 🔝 8005736733 🔝 💃  Top Class Call Girl Se...
Kolkata Call Girls Naktala 💯Call Us 🔝 8005736733 🔝 💃 Top Class Call Girl Se...
 
❤️Amritsar Escorts Service☎️9815674956☎️ Call Girl service in Amritsar☎️ Amri...
❤️Amritsar Escorts Service☎️9815674956☎️ Call Girl service in Amritsar☎️ Amri...❤️Amritsar Escorts Service☎️9815674956☎️ Call Girl service in Amritsar☎️ Amri...
❤️Amritsar Escorts Service☎️9815674956☎️ Call Girl service in Amritsar☎️ Amri...
 
Circulatory Shock, types and stages, compensatory mechanisms
Circulatory Shock, types and stages, compensatory mechanismsCirculatory Shock, types and stages, compensatory mechanisms
Circulatory Shock, types and stages, compensatory mechanisms
 
💚Call Girls In Amritsar 💯Anvi 📲🔝8725944379🔝Amritsar Call Girl No💰Advance Cash...
💚Call Girls In Amritsar 💯Anvi 📲🔝8725944379🔝Amritsar Call Girl No💰Advance Cash...💚Call Girls In Amritsar 💯Anvi 📲🔝8725944379🔝Amritsar Call Girl No💰Advance Cash...
💚Call Girls In Amritsar 💯Anvi 📲🔝8725944379🔝Amritsar Call Girl No💰Advance Cash...
 
💚Chandigarh Call Girls Service 💯Piya 📲🔝8868886958🔝Call Girls In Chandigarh No...
💚Chandigarh Call Girls Service 💯Piya 📲🔝8868886958🔝Call Girls In Chandigarh No...💚Chandigarh Call Girls Service 💯Piya 📲🔝8868886958🔝Call Girls In Chandigarh No...
💚Chandigarh Call Girls Service 💯Piya 📲🔝8868886958🔝Call Girls In Chandigarh No...
 
Goa Call Girl Service 📞9xx000xx09📞Just Call Divya📲 Call Girl In Goa No💰Advanc...
Goa Call Girl Service 📞9xx000xx09📞Just Call Divya📲 Call Girl In Goa No💰Advanc...Goa Call Girl Service 📞9xx000xx09📞Just Call Divya📲 Call Girl In Goa No💰Advanc...
Goa Call Girl Service 📞9xx000xx09📞Just Call Divya📲 Call Girl In Goa No💰Advanc...
 
Ahmedabad Call Girls Book Now 8980367676 Top Class Ahmedabad Escort Service A...
Ahmedabad Call Girls Book Now 8980367676 Top Class Ahmedabad Escort Service A...Ahmedabad Call Girls Book Now 8980367676 Top Class Ahmedabad Escort Service A...
Ahmedabad Call Girls Book Now 8980367676 Top Class Ahmedabad Escort Service A...
 
Call Girls Kathua Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Kathua Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Kathua Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Kathua Just Call 8250077686 Top Class Call Girl Service Available
 
Gorgeous Call Girls Dehradun {8854095900} ❤️VVIP ROCKY Call Girls in Dehradun...
Gorgeous Call Girls Dehradun {8854095900} ❤️VVIP ROCKY Call Girls in Dehradun...Gorgeous Call Girls Dehradun {8854095900} ❤️VVIP ROCKY Call Girls in Dehradun...
Gorgeous Call Girls Dehradun {8854095900} ❤️VVIP ROCKY Call Girls in Dehradun...
 

Calibration of spectrophotometer

  • 2. Essentials Regulations for quality control such as ISO 9000, Good Laboratory Practice (GLP), Good Manufacturing Practice (GMP) or standard operating procedures in the pharmaceutical industry as required by the most important pharmacopeia's (EP, DAB, USP), require a regular performance testing of the UV/Vis spectrophotometers
  • 3. Performance qualification Performance qualification Following eight criteria should be verified with help of standard reference in calibration of uv spectrophotometer Wavelength Accuracy Stray Light Resolution Noise Baseline Flatness Stability Photometric Accuracy Linearity
  • 4. Wavelength Accuracy • Wavelength Accuracy Wavelength accuracy is defined as the deviation of the wavelength reading at an absorption band or emission band from the known wavelength of the band. • The wavelength deviation can cause ignificant errors in the qualitative and quantitative results of the UV–Vis measurement.
  • 5. It is quite obvious that if the spectrophotometer is not able to maintain an accurate wavelength scale, the UV absorption profile of the sample measured by the instrument will be inaccurate. The true λ max and λ min of the analyte cannot be characterized accurately .
  • 6. Protocol for Checking Wavelength Accuracy Wavelength accuracy verification is checked by measuring a known wave- length reference standard with well-characterized absorption or emission peaks and comparing the recorded wavelength of the peak(s) against the value (s) listed in the certificate of that reference standard. There are many standards that are commonly used to verify the wavelength accuracy of a spectrophotometer
  • 7. • Spectra of some commonly used wavelength standards such as a deuterium lamp, mercury vapour lamp, holmium oxide filter, and holmium oxide solution (4% holmium oxide in 10% perchloric acid in a 1-cm cell ). • Acceptance. ±1 nm in the UV range (200 to 380 nm) and ±3 nm in the visible range (380 to 800 nm). Three repeated scans of the same peak should be within ±0.5nm .
  • 8. Stray Light • Stray Light Stray light is defined as the detected light of any wavelength that is outside the bandwidth of the wavelength selected. • The causes for stray light are scattering, higher- order diffraction of the monochromator , or poor instrument design. • Stray light causes a decrease in absorbance and reduces the linear range of the instrument. • High-absorbance measurements are affected more severely by stray light . • The stray light problem causes a deviation from linearity at high absorbance
  • 9. PROTOCOL For the stray light test, various cut-off filters or solutions can be used to estimate the stray light contribution, depending on the wavelengths is used. Scan the stray light testing solution in a 1-cm cell using air as the reference.
  • 10. The absorption of a solution of potassium chloride R (12 g/l) between 220 nm and 200 nm in a light path of 1 cm must rise steeply and at a wavelength of 198 nm must be greater than 2, measured against water R as a compensation liquid
  • 11. Resolution • Resolution The resolution of a UV-Vis spectrophotometer is related to its spectral bandwidth (SBW). The smaller bandwidth, the finer the resolution. The SBW depends on the slit width and the dispersive power of the monochrometer . Typically, only spectrophotometers designed for high- resolution work have a variable slit width.
  • 12. Spectrophotometers for routine analysis usually have a fixed slit width. The resolution of the absorbance measurement depends on the ratio of the spectral bandwidth (SBW) of the spectrophotometer to the natural bandwidth (NBW) of the spectral band to be measured.
  • 13. PROTOCOL The resolution of the absorbance measurement depends on the ratio of the spectral bandwidth (SBW) of the spectrophotometer to the natural bandwidth (NBW) of the spectral band to be measured Acceptance : The ratio of the absorbance at λmax (269 nm) and absorbance at λmin (266 nm) should be greater than 1.5.
  • 14. Noise Noise in the UV-Vis measurement originates primarily from the light source and electronic components. Noise in the measurement affects the accuracy at both ends of the absorbance scale. Photon noise from the light source affects the accuracy of the measurements at low absorbance.
  • 15. • Electronic noise from the electronic components affects the accuracy of the measurements at high absorbance. • A high noise level affects the precision of the measurements and reduces the limit of detection, thereby rendering the instrument less sensitive.
  • 16. PROTOCOL Air is scanned in the absorbance mode for 10 min. peak-to-peak noise is recorded at 500 nm. The root mean square (RMS) noise is then calculated. Modern spectrophotometers are usually equipped with the noise estimation function. Acceptance: The RMS noise should typically be less than 0.001 AU
  • 17. Baseline Flatness Baseline Flatness The intensity of radiation coming from the light source varies over the entire UV-Vis range. Most UV-Vis spectrophotometers have dual light sources. A deuterium lamp is used for the UV range and a tungsten lamp is used for the visible range. The response of the detector also varies over the spectral range.
  • 18. The flat baseline test demonstrates the ability of the instrument to normalize the light intensity measurement and the spectral output at different wavelengths throughout the spectral range .
  • 19. PROTOCOL Air is scanned in the absorbance mode. The highest and lowest deflections in the absorbance unit are recorded. Acceptance: The deflection is typically less than 0.01 AU.
  • 20. Stability Stability Variations in lamp intensity and electronic output between the measurements of the reference and the sample result in instrument drift. The lamp intensity is a function of the age of the lamp, temperature fluctuation, and wavelength of the measurement. These changes can lead to errors in the value of the measurements, especially over an extended period of time.
  • 21. The resulting error in the measurement may be positive or negative. The stability test checks the ability of the instrument to maintain a steady state over time so that the effect of the drift on the accuracy of the measurements is insignificant.
  • 22. PROTOCOL Air is scanned in the absorbance mode for 60 min at a specific wavelength (typically, 340 nm). The highest and lowest deflections in the absorbance unit are recorded. Acceptance: The deflection is typically less than 0.002 AU/h .
  • 23. Photometric Accuracy Photometric accuracy is determined by comparing the difference between the measured absorbance of the reference standard materials and the established standard value. An optically neutral material with little wavelength dependency for its transmittance/absorbance is desirable because it eliminates the spectral bandwidth dependency of measurements.
  • 24. PROTOCOL In that 0.006 % w/v solution of potassium dichromate (60.06 mg/L) is prepared in 0.005 M sulphuric acid. A 0.005 M sulphuric acid solution as the reference is scanned and then the potassium dichromate solutions are scanned at 235, 257, 313, and 350 nm. Acceptance : Six replicate measurements of the 0.006% w/v potassium dichromate solution at 235, 257, 313, and 350 nm should be less than 0.5% RSD.
  • 25. Linearity Linearity The linear dynamic range of the measurement is limited by stray light at high absorbance and by noise at low absorbance. For routine measurements involving samples and the related reference chemical standards, the accuracy of the quantification of the sample depends on the precision and linearity of the measurements
  • 26. PROTOCOL A series of potassium dichromate solutions of concentration 20, 40, 60, 80, and 100 mg/L in 0.005 M sulphuric acid can be used to test the linearity of the system. First, 0.005 M sulphuric acid as reference is scanned and then the potassium dichromate solutions at 235, 257, 313, and 350 nm.
  • 27. • The absorbance values at various wavelengths are plotted against the concentration of the solutions and the correlation coefficients are calculated. • Acceptance: Correlation coefficient r = 0.999
  • 28.
  • 29. Ideally used type of filters Usually by trial and error following type of filters are considered ideal for working with all kinds of spectrophotometric instruments. For checking Wavelength accuracy, Photometric accuracy, Stray light behaviour & Resolution Liquid filters are more reliable whereas for visible ranges solid filters are preferred to calibrate the instrument for Wavelength and Photometric accuracy
  • 30. Holmium oxide liquid filter (Holmium oxide dissolved in perchloric acid) • The exact position of peaksat 241.15 nm, 287.15 nm,361.5 nm and 536.3 nm is determined and compared with the values in the certificate. • Deviations should not be greater than ± 1 nm in the ultra-violet and ± 3 nm in the visible range
  • 31. Holmium Oxide Spectra Slit width= 1nm (or to be kept as low as possible) Testing the wavelength accuracy in the UV and visible range between 240 nm and 650 nm for spectrophotometers with a spectral bandwidth (slit) between 0.1 nm and 3 nm
  • 32. Holmium Oxide Glass doped Filter and its spectra (2.1 mm thickness)
  • 33. Didymium oxide • Generally available as glass filter is more reliable for routine calibration made out of special grade glass. • Testing the wavelength accuracy in the UV and Vis region (329 nm to 875 nm) at a spectral bandwidth to 2 nm.
  • 34. Spectra of Didymium Glass Filter (spectral bandwidth to 2 nm) Absorbs at- 329 nm 472 nm 512 nm 681 nm 875 nm
  • 35. Potassium Dichromate Potassium dichromate dissolved in perchloric acid is suitable for testing the photometric accuracy (absorbance) of spectrophotometers in the UV spectrum. The absorption values of the cuvette are measured at the wavelengths 235 nm, 257 nm, 313 nm and 350 nm and recorded in the certificate
  • 36. Typical spectrum of a 0.006 % potassium dichromate solution (spectral bandwidth of 2 nm) Absorbs at 235 nm 257 nm 313 nm 350 nm
  • 37. Potassium dichromate liquid filter (430 nm)
  • 38. Neutral density glass filter Absorbs in visible region (440 nm to 635 nm) Made of SCHOTT Glasss
  • 39. Cut-Off filters Spectra of various cut-off filters Testing for stray light in the UV spectrum (at wavelengths from 198 nm to 385 nm, depending on the selected filter)
  • 40. Application The photometer is set to a starting wavelength that is 30 nm above the cut-off wavelength of the filter to be used. If, for example, potassium chloride is used then the start is at 230 nm. Then a scan is carried out down towards shorter wavelengths and the absorption below the cut-off wavelength is observed. The transmission that is measured below the cut-off wavelength is stray light
  • 41. Calibration standards for testing the resolution The resolution for modern photometers depends directly on the slit width. The smaller the slit and the associated spectral bandwidth, the higher the resolution. The choice of material should be made as per the standards applicable to the analysis or as mentioned on the format for certificate of analysis or the manual.
  • 42. In order to determine the resolution of a spectrophotometer, the behaviour of the absorption of a maximum at λmax (269 nm) to absorption of a minimum at λmin (266 nm) is shown. From this, one obtains details of the spectral bandwidth of the instrument used. Wavelength spectrum from 265 nm to 270 nm is scanned and the peak maximum and peak minimum are determined. Then the relationship between the two is calculated
  • 43. Two type of standards are used- 1. Toluene in n-Hexane 2. Hexane as reference
  • 44.
  • 45. Control of Absorbance (Potassium Dichromate method) • Dry a quantity of Potassium Dichromate by heating to constant weight at 1300C . • Weigh and transfer accurately a quantity not less than 57.0mg and not more than 63 mg to 10000ml volumetric flask. Dissolve and dilute in sufficient 0.005M H2SO4 to produce 1000ml. • Measure the absorbance of Potassium Dichromate solution at wavelength given in the table
  • 46. • Calculate value of {A1% 1cm } for each wavelengths mentioned. • Acceptance Criteria: (as per I.P and B.P) Wavelengths(in nm) A(1%/1cm) Limit 235.0 124.5 129.9-126.2 257.0 144.0 142.8-145.7 313.0 48.6 47.0-50.3 350.0 106.6 105.6-108.2
  • 47. Resolution power • Record the spectrum of a 0.02% v/v solution of toluene in hexane in the range of 260nm to 420 nm (before use check the hexane for transmittance, using water as blank between 260nm to 420 nm & use only if transmittance is not less than 97%) • Acceptance Criteria: The ratio of absorbance at maximum at about 269 nm to that at the minimum at about 266 nm is not less than 1.5
  • 48. Limit of Stray Light • Prepare a 1.2% w/v solution of potassium chloride in water. • Measure the absorbance of the above solution at solution at 198.0, 199.0, 200.0, 201.0 and 202.0 using water as blank. • Acceptance Criteria: Absorbance is greater than 2
  • 49. Wavelength accuracy, Resolution & baseline flatness • These are instruments in-built functions • The note made here is in accordance to Shimadzu UV-1700 UV-Vis Spectrophotometer • Attach printer directly to the instrument (optional) • Goto MODE and press F3 key to opt for maintenance mode. • Press 1 (…continued)
  • 50. • Press Start/Stop key. • After screen changes, ensure that nothing is kept in the optical path & press the start/stop key again. • Prints will come after all the tests are over. • Acceptance Criteria: For wavelength accuracy: at 656.1±0.3 nm & at 486.0±0.3 nm. Resolution:1.0 nm or less. Baseline flatness: ±0.002Abs.
  • 51.
  • 52. Reference • #1 Indian Pharmacopoeia • #2 U.S. Pharmacopoeia • #3 Pharmaguideline • #4 Hellma GmBH product Catalogue • #5 Schimadzu UV-1700 Users manual