                       Noninvasive Cardiac Output Estimation Using a Novel
                                   Photoplethysmogram Index
                                     L. Wang, Emma Pickwell-MacPherson, Y. P. Liang and Y. T. Zhang


    Abstract—Cardiac output (CO) monitoring is essential for                                noninvasive, accurate, reliable and continuous. At present,
indicating the perfusion status of the human cardiovascular                                 no single method meets all these criteria. Intermittent
system under different physiological conditions. However, it is                             thermodilution is widely accepted as the clinical golden
currently limited to hospital use due to the need for either                                standard. This method requires the insertion of a pulmonary
skilled operators or big, expensive measurement devices.
                                                                                            artery catheter (PAC) to obtain one measurement per 3-4
Therefore, in this paper we devise a new CO indicator which
can easily be incorporated into existing wearable devices. To                               minutes [1]. It is too invasive and non-continuous. Two
this end, we propose an index, the inflection and harmonic area                             existing less invasive and continuous methods are
ratio (IHAR), from standard photoplethysmographic (PPG)                                     oesophageal Doppler monitoring and CO2 re-breathing, but
signals, which can be used to continuously monitor CO. We                                   both of these require skilled operators and expensive
evaluate the success of our index by testing on sixteen                                     measurement devices [1]. Amongst the currently used
normotensive subjects before and after bicycle exercise. The
                                                                                            methods, impedance cardiography is probably the only
results showed a strong intra-subject correlation between
IHAR and COimp measured by the bio-impedance method in                                      noninvasive and automatic technique. However, the
fifteen subjects (mean r = 0.82, p<0.01). After least squares                               impedance device is big and expensive, and its accuracy is
linear regression, the precision between COimp and CO                                       often influenced by the change of electrode positions and the
estimated from IHAR (COIHAR) was 1.40 L/min. The total                                      sweat on the skin [2]. Due to the disadvantages mentioned
percentage error of the results was 16.2%, which was well                                   above, these methods are all unquestionably limited to
below the clinical acceptance limit of 30%. The results suggest
                                                                                            bedside use. They are not portable, or wearable, so they are
that IHAR is a promising indicator for wearable and
noninvasive CO monitoring.                                                                  difficult to incorporate into home health care monitoring
                                                                                            systems. To solve this problem, one of the best ways is to
                              I. INTRODUCTION                                               derive a new CO indicator from signals provided by the
                                                                                            existing wearable device. Some preliminary studies have
I   T is well known that heart rate, blood pressure and cardiac
   output (CO) are all essential physiological parameters of
the human cardiovascular system. CO, defined as the blood
                                                                                            implied that the PPG signal could be a candidate for this
                                                                                            application [3]-[6].
                                                                                               The PPG signal, which indicates the blood volume
volume ejected by the heart per minute (unit: L/min, where
                                                                                            changes on site, is an optical signal that could be non-
“L” means liter), is regarded as the ultimate expression of
                                                                                            invasively obtained from body peripheral terminals, such as
cardiovascular performance, since it indicates how well the
                                                                                            ear, finger and toe. The PPG acquisition components,
heart is able to provide enough nutrition and oxygen to the
                                                                                            including a pair of LED emitter and receiver, and related
peripheral organs and tissues. For human beings, in order to
                                                                                            simple circuits, are cheap and small in size, and could be
maintain a normal state of tissue perfusion and oxygen
                                                                                            easily embedded into many existing wearable devices.
delivery condition, the baseline CO should be in the range of
                                                                                               Although PPG is obtained peripherally, many previous
4 L/min to 8 L/min. If CO gets out of this range, it is often a
                                                                                            studies have revealed that the PPG wave contour is primarily
sign of cardiovascular disease, such as hypertension, stroke
                                                                                            influenced by characteristics of the systemic circulation, but
or heart failure. Hence, continuous CO monitoring plays an
                                                                                            not the local perfusion [3]-[6]. Henry Lax and his colleagues
essential role in the evaluation, treatment, and follow-up of
                                                                                            [3] noted that when individuals had cold fingers, this
critically ill patients.
                                                                                            “reduces the over-all amplitude but does not affect the
   Ideally, a technology which measures CO should be
                                                                                            configuration of the pulse wave”. Chowienczyk PJ found in
                                                                                            his experiment that the vasodilation drugs which increased
   Manuscript received April 7, 2009. This work was supported in part by                    the local circulation in the upper arm could not change the
the Hong Kong Innovation and Technology Fund (ITF). The authors are                         corresponding finger PPG wave contour, while, the
grateful to Standard Telecommunication Ltd., Jetfly Technology Ltd.,
Golden Meditech Company Ltd., Bird International Ltd. and Bright Steps                      systematic insertion of vasodilation drugs, such as GTN,
Corporation of their supports to the ITF projects.                                          could change it substantially [4]. More recently, we have
   Y. T. Zhang (phone: 852-2609-8459; e-mail: ytzhang@ee.cuhk.edu.hk),                      successfully utilized a parameter derived from the PPG wave
Emma Pickwell-MacPherson, Y. P. Liang and L. Wang are with the Joint
Research Centre for Biomedical Engineering, Department of Electronic                        to trace the change of total peripheral resistance (TPR) after
Engineering, The Chinese University of Hong Kong, Hong Kong SAR.                            bicycle exercise [5]. McCombie proposed a blind system
   Y. T. Zhang is also with the Institute of Biomedical and Health                          identification method to calculate the cardiac output
Engineering, Shenzhen Institute of Advanced Technology, Chinese
Academy of Sciences, China, and Key Laboratory for Biomedical                               waveform from two PPG signals measured from different
Informatics and Health Engineering, Chinese Academy of Sciences, China.                     body locations, but only a qualitative aortic flow curve was
978-1-4244-3296-7/09/$25.00 ©2009 IEEE                                               1746


        Authorized licensed use limited to: ALBERT MAAREK. Downloaded on January 7, 2010 at 12:38 from IEEE Xplore. Restrictions apply.
obtained [6].                                                                            invasive arterial catheter or a bedside Finapres device for
   In this paper, we propose a novel indicator of CO, the                                acquiring the continuous blood pressure measurement.
inflection and harmonic area ratio (IHAR), which is derived
from PPG signals. The ability of IHAR to trace CO changes                                                                            PPG
was evaluated in a bicycle exercise experiment.                                                         Time domain                                    Frequency domain

                                  ABP                                                                                                                      FFT(f1)




                                                                                                                                          magnitude
                                                                                                                                                             FFT(f2)


                                                                                                        S1                  S2
   CO                                         AC                 TPR

                                                                                                                                                      f1 f2 …         … fn   Hz
                                                                                                                                                            frequency
                                                                                                                  (a)                                         (b)
                                                                                           Fig. 2. (a) is PPG wave in the time domain. S1 and S2 are the areas
                                                                                           under the whole PPG wave and the part of wave after the point of
                Fig. 1. Two-element Windkessel model.                                      inflection, respectively. IPA=S2/S1. (b) is PPG wave in the frequency
                                                                                           domain. FFT ( f n ) is the magnitude at the nth harmonic. NHA =
                                                                                             N                          N
                          II. METHODOLOGY                                                   ∑ FFT ( f ) ∑ FFT ( f ).
                                                                                            n =2
                                                                                                        2
                                                                                                              n
                                                                                                                    n =1
                                                                                                                                 2
                                                                                                                                     n




   According to the two-element Windkessel model shown                                      However, as shown in equation (1), blood pressure
in Fig. 1, CO could be calculated from the analysis of                                   measurement is not a necessity for obtaining CO, if proper
continuous arterial blood pressure waveform by a so called                               surrogates of MBP and TPR can be derived from other
pulse contour method. In this model, the cardiovascular                                  signals. According to wave reflection theory, arterial blood
system is analogous to a current source connected with a                                 pulse could be divided into two waves: a first wave
two-element circuit. CO, which is mimicked by the mean                                   produced by heart pumping and a second wave produced by
amount of current passing through the TPR (total peripheral                              pulse wave reflection. Therefore, the inflection point area
resistance), equals the mean pressure (mean arterial blood                               ratio (IPA), the area ratio of the second and first peak in the
pressure, MBP) divided by the TPR:                                                       PPG wave (see Fig. 2 (a)), is mainly influenced by the
                          CO = MBP TPR                   (1).                            strength of pulse wave reflection. Pulse wave reflection
In this method, if a continuous arterial pressure waveform is                            results from the impedance mismatch between different parts
obtained, MBP could be calculated beat to beat. TPR is                                   in the arterial system, e.g., the compliance and resistance
firstly initialized by a pair of calibration CO and MBP data,                            mismatches between the big, elastic arteries and the small
and its value of the current beat is calculated from MBP and                             arteries. Studies have shown that approximate 90% of the
estimated CO of the previous beat, iteratively. The main                                 TPR is located in the small arteries. Hence, if the small
shortcoming of such technique is that it needs either an                                 arteries contract, the TPR will change, which will change the




     5-min.              5-min. 60W                                                           5-min. 60W                                                 5-min. 60W
      rest             bicycle exercise                                                     bicycle exercise                                           bicycle exercise
                                                                                                        ……                                                         ……

                                                                                                                    ……                                              ……
            1              2       3      4       5       6      7        8      9                 10        11                      17                   18 19              25

        3-min.                                    6-min.                                                      6-min.                                            6-min.
      Resting                            Post-exercise 1                                           Post-exercise 2                                        Post-exercise 3
                          represented BP measurement by a registered nurse
                          numbers 1-25 indicate the datasets of simultaneous measurements of CO and PPG
                                                                     Fig. 3. Experiment procedure.
                                                                                  1747


     Authorized licensed use limited to: ALBERT MAAREK. Downloaded on January 7, 2010 at 12:38 from IEEE Xplore. Restrictions apply.
impedance mismatch, influence the strength of the wave                                                               IV. DATA ANALYSIS
reflection, and further change IPA. In our previous                                         For each subject, 25 datasets were obtained, including 1
investigation [5], IPA has shown to be a very good indicator                             dataset at rest and 8 datasets in each post exercise phase (see
of TPR when it changes after exercise. The pulse wave                                    Fig. 3). The 8 datasets after each exercise were obtained
reflection not only changes the shape of the PPG signal in                               from the 6-min continuous data divided by the intermittent
the time domain, but also in the frequency domain. Our                                   BP measurements. As shown in Fig. 3, BP was also
previous study [7] has proved that the normalized harmonic                               measured by a nurse every other dataset, e.g., during datasets
area (NHA), a parameter derived from frequency domain                                    2, 4, 6, 8 for post exercise phase 1.
analysis (see Fig. 2 (b)), is related to pulse wave reflection                              For offline analysis, each dataset of PPG signals was re-
and is strongly correlated with systolic and diastolic blood                             sampled to 100 Hz and divided into beat-to-beat waves. The
pressure. Therefore, IHAR, defined as NHA divided by IPA,                                start point of each beat is defined as the beginning of its
is proposed as a potential CO indicator and expressed as                                 systolic rise, which is also the end of the diastolic fall wave
follows:                                                                                 of the previous beat. Single PPG wave was transferred to the
                       ⎛     N
                                         ⎞                                               frequency domain by 512 points FFT (ensuring a resolution
                       ⎜ ∑ FFT 2 ( f n ) ⎟
                       ⎜1 − n=2          ⎟                                               of more than 10 points per Hz). Then, beat-to-beat IHAR
                       ⎜     N           ⎟                                               was calculated using equation (2). Meanwhile, COimp was
                       ⎜ ∑ FFT ( f n ) ⎟
                                      2
                                                                                         obtained by the impedance cardiograph device, beat by beat.
              IHAR = ⎝ n=1               ⎠              (2),                             “Beat by beat” here meant that COimp was acquired and
                               ⎛ S2 ⎞                                                    saved in a “beat to beat” format, but was actually calculated
                               ⎜ ⎟
                               ⎜S ⎟
                               ⎝ 1⎠                                                      from the rolling average of eight impedance wave and five
where S1 and S2 are the areas under the whole PPG wave and                               ECG wave in the algorithm. As a result, it was impossible in
the part of wave after the point of inflection in the time                               this study to compare IHAR with COimp in a beat-to-beat
domain, respectively (see Fig. 2(a)), and FFT 2 ( f n ) is the                           way, but only to compare the means of each dataset. For CO
                                                                                         estimation, all datasets of IHAR and COimp were firstly used
square of the magnitude at the nth harmonic of the PPG wave                              to produce CO = α ⋅ IHAR + β by the least squares linear
in the frequency domain (see Fig. 2(b)).
                                                                                         regression method, and then, the COIHAR results were
   The performance of IHAR for CO estimation was
                                                                                         obtained from IHAR by the above equation.
evaluated in a bicycle exercise experiment. An impedance
cardiograph device, Physio Flow PF-05 (Manatec
                                                                                                                           V. RESULTS
Biomedical, Macheren, France), was used as the benchmark
of CO measurement. The validity of this device has been                                     As shown in Table I, the hemodynamic parameters,
established in both normotensive adults and cardiovascular                               including heart rate, systolic blood pressure, diastolic blood
patients versus the direct Fick method (lab golden standard                              pressure and cardiac output, have significantly changed after
of CO measurement) both at rest and exercise conditions [2]                              each 5 min of bicycle exercise, compared with those
[8].                                                                                     obtained at rest (p<0.05).
                                                                                            The intra-subject correlation coefficients (r) between
                    III. EXPERIMENT PROTOCOL                                             IHAR and COimp are listed in Table II. For all subjects,
                                                                                         IHAR was positively correlated to COimp. Except for subject
   16 normotensive volunteers (8 males and 8 females), aged
                                                                                         13 (p=0.054), there was a strong correlation between IHAR
from 22 to 34 years, participated in this study. Informed
                                                                                         and COimp in the other 15 subjects (mean r = 0.82, p<0.01).
consent was obtained from each participant. The
experimental procedures are depicted in Fig. 3. Upon arrival,                                                                  TABLE I
                                                                                              CHANGES OF HEMODYNAMIC PARAMETERS AFTER 1 , 2                     , AND 3RD
                                                                                                                                                      ST   ND
subjects were asked to rest for 5 minutes. Then, the systolic
                                                                                                       EXERCISE COMPARED WITH THOSE AT REST
blood pressure (SBP) and diastolic blood pressure (DBP)
                                                                                                                                    1st         2nd               3rd
were measured from the left arm by a registered nurse. After                                                        At Rest
                                                                                                                                  Exercise    Exercise          Exercise
this, PPG and cardiac output were recorded simultaneously                                       Heart Rate
                                                                                                                    76±10        92±17﹡      95±17﹡         100±16﹡
for 1 min. Then, the subjects were asked to ride the bicycle                                     (beat/s)
                                                                                               SBP (mmHg)           106±12       116±14﹡     117±14﹡        118±15﹡
(Corival V2, Lode VB, Groningen, Netherland) for 5
                                                                                               DBP (mmHg)           61±7         59±10﹡      60±10﹡         61±10﹡
minutes at 60 W. Immediately after exercise, 6-min of PPG                                       CO (L/min)          6.3±1.2      8.5±1.8﹡    8.7±1.9﹡       9.2±2.0﹡
and CO data were continuously acquired, meanwhile, at time                                    ﹡
                                                                                                  significant difference with the results obtained at rest (p<0.01).
points after exercise of 0 min, 1.5 min, 3 min, 4.5 min and 6
min, blood pressure (BP) was measured. Then, the same 5
                                                                                            The dynamic range of CO estimated from IHAR (COIHAR)
min exercise and post-exercise data acquisition procedures
                                                                                         was very similar to that of COimp (COIHAR: 4.02 L/min-14.72
were repeated twice. PPG was recorded on the right index
                                                                                         L/min; and COimp: 3.66 L/min-14.48 L/min). As shown in
finger by an in-house designed acquisition system. CO was
                                                                                         Table II, the standard deviation (SD) of the estimation
measured by the Physio Flow PF-05 impedance cardiograph
                                                                                         residue is 0.71 L/min. This value does not vary much with
device.

                                                                                  1748


     Authorized licensed use limited to: ALBERT MAAREK. Downloaded on January 7, 2010 at 12:38 from IEEE Xplore. Restrictions apply.
subject variance, which indicates a stable performance of                                                       percentage error (PE = 16.2 %) of our technique is well
IHAR based CO estimation. In many previous CO studies, a                                                        below the clinically acceptable error of 30%.
percentage error (PE) is calculated to indicate the variance of                                                    A potential limitation of this work is the change of PPG
estimation error with respect to reference mean                                                                 wave as a result of vascular aging or cardiovascular diseases,
( PE = 1.96SD MEANCO ), and a PE less than 30% is                                                               e.g., the inflection point may be blurred and difficult to find
                                                                  ref
                                                                                                                as the arterial stiffness increases. Therefore, further
regarded to be clinically acceptable [9] [10]. The PE of the
                                                                                                                validation on subjects with a bigger age range and
results in this study is 16.2 %.
                                                                                                                cardiovascular patients are needed. Another limitation is the
                                                                  TABLE II                                      calibration procedure. Since the main purpose of this study
 THE SUMMARY OF THE INTRA-SUBJECT CORRELATION COEFFICIENT r                                                     is to propose and evaluate a novel CO index, we have not
BETWEEN COIMP AND IHAR AS WELL AS THE REGRESSION COEFFICIENT
                                                                                                                fully investigated the calibration procedure. Only a simple
 AND STANDARD DEVIATION (SD) OF RESIDUE IN THE LEAST SQUARE
                     LINEAR REGRESSION                                                                          least squares linear regression was utilized. However,
                                                          COIPHA = α ⋅ IHAR + β                SD of            individual calibration is an essential procedure for IHAR
    Subject                                      r                                            Residue           based CO estimation and will be investigated further in a
                                                            α             β                   (L/min)           separate study.
     S01                                 0.94﹡               195.1                 3.7          0.83
     S02                                 0.94﹡               189.6                 2.9          0.62
     S03                                 0.78﹡               265.0                 0.9          0.61                                         REFERENCES
     S04                                 0.92﹡               199.7                 2.8          0.64            [1]  L. Mathews and K. RK Singh, “Cardiac output monitoring,” Annals of
     S05                                 0.58﹡               689.0                -3.1          0.87                 Cardiac Anaesthesia, vol. 11, Jan.-Jun. 2008, pp. 56-68.
     S06                                 0.80﹡               272.0                 0.0          0.78            [2] R. Richard, E. Lonsdorfer-Wolf, A. Charloux, et al., “Non-invasive
     S07                                 0.91﹡               347.9                -0.4          0.49                 cardiac output evaluation during a maximal progressive exercise test,
     S08                                                     256.1                 4.8          0.86                 using a new impedance cardiograph device,” Eur J Appl Physiol., vol.
                                         0.79﹡
                                                                                                                     85, Aug. 2001, pp. 202-7.
     S09                                 0.96﹡               172.1                 2.0          0.38
                                                                                                                [3] H. Lax, A. Feinberg and B. M. Cohen, “Studies of the arterial pulse
     S10                                 0.89﹡               319.7                 0.9          0.74                 wave and its modification in the presence of human arteriosclerosis,”
     S11                                 0.83﹡               396.6                -0.3          0.85                 J Chronic Dis., vol. 3, Jun. 1956, pp. 618-31.
     S12                                 0.96﹡               282.3                 2.4          0.43            [4] P. J. Chowienczyk, R. P. Kelly, H. MacCallum, et al.,
     S13                                 0.39                185.9                 4.9          0.78                 “Photoplethysmographic assessment of pulse wave reflection: blunted
     S14                                 0.78﹡               345.1                 0.2          0.90                 response to endothelium-Dependent Beta2-Adrenergic vasodilation in
     S15                                 0.69﹡               283.7                 4.2          0.83                 type II diabetes mellitus,” J Am Coll Cardiol, vol. 34, Aug. 1999, pp.
     S16                                 0.90﹡               245.8                 4.2          0.72                 2007-14.
                                                                                                                [5] L. Wang and Y. T. Zhang, “A novel photoplethysmogram index for
     Mean                                0.82                                                   0.71
﹡
                                                                                                                     total peripheral resistance after bicycle exercise,” Proc. of the 5th
    p<0.01.                                                                                                          International Conference on Ubiquitous Healthcare, Pusan, Korea,
                                                                                                                     Oct. 2008, pp. 175-6.
  The Bland-Altman plot (see Fig. 4) shows the agreement                                                        [6] D. McCombie, A. Reisner and H. Asada, “Identification of vascular
of the estimated CO and the reference CO. The percentage                                                             dynamics and estimation of the cardiac output waveform from
of CO estimations within the 95% confidence interval ([-                                                             wearable PPG sensors,” Proc. of the 27th IEEE International
                                                                                                                     Engineering in Medicine and Biology Conference, EMBC 2005,
1.40 L/min -1.40 L/min]) is 95.9%.                                                                                   Shanghai, China, pp. 3490-3.
                                                                                                                [7] Y. S. Yan and Y. T. Zhang, “Noninvasive estimation of blood
                                                                                                                     pressure using photoplethysmographic signals in the period domain,”
                                        3
                                                                                                                     Proc. of the 27th IEEE International Engineering in Medicine and
        Difference of Reference an d




                                        2                                                                            Biology Conference, EMBC 2005, Shanghai, China, pp. 3583-4.
            E stim ated CO (L /m in )




                                                                                                                [8] A. Charloux, E. Lonsdorfer-Wolf, R. Richard, et al., “A new
                                        1                                                                            impedance cardiograph device for the non-invasive evaluation of
                                                                                                                     cardiac output at rest and during exercise: comparison with the
                                        0
                                                                                              95.9%                  “direct” Fick method,” Eur J Appl Physiol., vol. 82, Jul. 2000, pp.
                                             0                5              10          15             20           313-20.
                                        -1                                                                      [9] M. Biais, K. Nouette-Gaulain, V. Cottenceau, et al., “Cardiac output
                                                                                                                     measurement in patients undergoing liver transplantation: pulmonary
                                        -2                                                                           artery catheter versus uncalibrated arterial pressure waveform
                                                                                                                     analysis,” Anesth Analg., vol. 106, May 2008, pp. 1480-6.
                                        -3                                                                      [10] L. A. Critchley and J. A. Critchely, “A meta-analysis of studies using
                                                     Average of Reference and Estimated CO (L/min)                   bias and precision statistics to compare cardiac output measurement
                                                                                                                     techniques,” J Clin Monit Comput., vol. 15, Feb. 1999, pp. 85-91.

                                             Fig. 4. Bland-Altman plot of cardiac output.


                                                            VI. CONCLUSION
   In this study, we have tested the performance of a new CO
index, IHAR, in 16 normotensive subjects through a bicycle
exercise study. The strong intra-subject correlation suggests
that IHAR can successfully trace the CO changes over a
wide dynamic range before and after exercise. The small

                                                                                                             1749


        Authorized licensed use limited to: ALBERT MAAREK. Downloaded on January 7, 2010 at 12:38 from IEEE Xplore. Restrictions apply.

Más contenido relacionado

Similar a Ptg and cardiac output.jsp

IRJET - Sleep Apnea Detection using Physiological Signals
IRJET -  	  Sleep Apnea Detection using Physiological SignalsIRJET -  	  Sleep Apnea Detection using Physiological Signals
IRJET - Sleep Apnea Detection using Physiological SignalsIRJET Journal
 
Noninvasive blood glucose monitoring system based on near-infrared method
Noninvasive blood glucose monitoring system based on near-infrared method  Noninvasive blood glucose monitoring system based on near-infrared method
Noninvasive blood glucose monitoring system based on near-infrared method IJECEIAES
 
Design and development of electro optical system for acquisition of ppg signa...
Design and development of electro optical system for acquisition of ppg signa...Design and development of electro optical system for acquisition of ppg signa...
Design and development of electro optical system for acquisition of ppg signa...eSAT Publishing House
 
Merged document 12
Merged document 12Merged document 12
Merged document 12V!vEk@nAnD S
 
Monitoring Modality in anesthesia
Monitoring Modality  in anesthesiaMonitoring Modality  in anesthesia
Monitoring Modality in anesthesiaDr.RMLIMS lucknow
 
Analysis of cardiorespiratory coupling in athletes in rest using causal approach
Analysis of cardiorespiratory coupling in athletes in rest using causal approachAnalysis of cardiorespiratory coupling in athletes in rest using causal approach
Analysis of cardiorespiratory coupling in athletes in rest using causal approachMarcel Młyńczak
 
MONITORINGpulseoxy44554545654543354454.pptx
MONITORINGpulseoxy44554545654543354454.pptxMONITORINGpulseoxy44554545654543354454.pptx
MONITORINGpulseoxy44554545654543354454.pptxrijjorajoo
 
V k bairagi research paper
V k bairagi research paperV k bairagi research paper
V k bairagi research paperaissmsblogs
 
Mesuring ph with FOS (Fibre Optic Sensors)
Mesuring ph with FOS (Fibre Optic Sensors)Mesuring ph with FOS (Fibre Optic Sensors)
Mesuring ph with FOS (Fibre Optic Sensors)Subhrajit Sadhukhan
 
Azriel Perel - Techniques for the Future - IFAD 2011
Azriel Perel - Techniques for the Future - IFAD 2011Azriel Perel - Techniques for the Future - IFAD 2011
Azriel Perel - Techniques for the Future - IFAD 2011International Fluid Academy
 
5_6280601286203146342 exact science.pptx
5_6280601286203146342 exact science.pptx5_6280601286203146342 exact science.pptx
5_6280601286203146342 exact science.pptxMinaz Patel
 

Similar a Ptg and cardiac output.jsp (20)

U34129131
U34129131U34129131
U34129131
 
IRJET - Sleep Apnea Detection using Physiological Signals
IRJET -  	  Sleep Apnea Detection using Physiological SignalsIRJET -  	  Sleep Apnea Detection using Physiological Signals
IRJET - Sleep Apnea Detection using Physiological Signals
 
Icp monitoring seminar
Icp monitoring seminarIcp monitoring seminar
Icp monitoring seminar
 
Noninvasive blood glucose monitoring system based on near-infrared method
Noninvasive blood glucose monitoring system based on near-infrared method  Noninvasive blood glucose monitoring system based on near-infrared method
Noninvasive blood glucose monitoring system based on near-infrared method
 
Design and development of electro optical system for acquisition of ppg signa...
Design and development of electro optical system for acquisition of ppg signa...Design and development of electro optical system for acquisition of ppg signa...
Design and development of electro optical system for acquisition of ppg signa...
 
Merged document 12
Merged document 12Merged document 12
Merged document 12
 
Monitoring Modality in anesthesia
Monitoring Modality  in anesthesiaMonitoring Modality  in anesthesia
Monitoring Modality in anesthesia
 
main.pdf
main.pdfmain.pdf
main.pdf
 
Analysis of cardiorespiratory coupling in athletes in rest using causal approach
Analysis of cardiorespiratory coupling in athletes in rest using causal approachAnalysis of cardiorespiratory coupling in athletes in rest using causal approach
Analysis of cardiorespiratory coupling in athletes in rest using causal approach
 
Hhh
HhhHhh
Hhh
 
MONITORINGpulseoxy44554545654543354454.pptx
MONITORINGpulseoxy44554545654543354454.pptxMONITORINGpulseoxy44554545654543354454.pptx
MONITORINGpulseoxy44554545654543354454.pptx
 
V k bairagi research paper
V k bairagi research paperV k bairagi research paper
V k bairagi research paper
 
Mesuring ph with FOS (Fibre Optic Sensors)
Mesuring ph with FOS (Fibre Optic Sensors)Mesuring ph with FOS (Fibre Optic Sensors)
Mesuring ph with FOS (Fibre Optic Sensors)
 
40120140507005
4012014050700540120140507005
40120140507005
 
40120140507005
4012014050700540120140507005
40120140507005
 
PsterPresentation2016
PsterPresentation2016PsterPresentation2016
PsterPresentation2016
 
Azriel Perel - Techniques for the Future - IFAD 2011
Azriel Perel - Techniques for the Future - IFAD 2011Azriel Perel - Techniques for the Future - IFAD 2011
Azriel Perel - Techniques for the Future - IFAD 2011
 
Respiratory monitoring
Respiratory monitoringRespiratory monitoring
Respiratory monitoring
 
Pulse oximeter
Pulse  oximeterPulse  oximeter
Pulse oximeter
 
5_6280601286203146342 exact science.pptx
5_6280601286203146342 exact science.pptx5_6280601286203146342 exact science.pptx
5_6280601286203146342 exact science.pptx
 

Más de ES-Teck India

EIS Technology: New marker using bioimpedance technology in screening for ADHD
EIS Technology: New marker using bioimpedance technology in screening for ADHDEIS Technology: New marker using bioimpedance technology in screening for ADHD
EIS Technology: New marker using bioimpedance technology in screening for ADHDES-Teck India
 
ESO system, ES-BC system and EIS-GS system - Miami University study
ESO system, ES-BC system and EIS-GS system - Miami University study ESO system, ES-BC system and EIS-GS system - Miami University study
ESO system, ES-BC system and EIS-GS system - Miami University study ES-Teck India
 
EIS technology : bioimpedance application in selective serotonin reuptake
EIS technology : bioimpedance application in selective serotonin reuptakeEIS technology : bioimpedance application in selective serotonin reuptake
EIS technology : bioimpedance application in selective serotonin reuptakeES-Teck India
 
EIS Technology: bioimpedance chronoamperometry in adjunct to screen the prost...
EIS Technology: bioimpedance chronoamperometry in adjunct to screen the prost...EIS Technology: bioimpedance chronoamperometry in adjunct to screen the prost...
EIS Technology: bioimpedance chronoamperometry in adjunct to screen the prost...ES-Teck India
 
Summary of clinical investigations es teck complex system
Summary of clinical investigations es teck complex systemSummary of clinical investigations es teck complex system
Summary of clinical investigations es teck complex systemES-Teck India
 
Guidelines heart rate_variability_ft_1996[1]
Guidelines heart rate_variability_ft_1996[1]Guidelines heart rate_variability_ft_1996[1]
Guidelines heart rate_variability_ft_1996[1]ES-Teck India
 
Arterioscler thromb vasc_biol_22_147
Arterioscler thromb vasc_biol_22_147Arterioscler thromb vasc_biol_22_147
Arterioscler thromb vasc_biol_22_147ES-Teck India
 
Serotonin ssri effects
Serotonin ssri effectsSerotonin ssri effects
Serotonin ssri effectsES-Teck India
 
Serotonin and blood flow
Serotonin and blood flowSerotonin and blood flow
Serotonin and blood flowES-Teck India
 

Más de ES-Teck India (20)

EIS Technology: New marker using bioimpedance technology in screening for ADHD
EIS Technology: New marker using bioimpedance technology in screening for ADHDEIS Technology: New marker using bioimpedance technology in screening for ADHD
EIS Technology: New marker using bioimpedance technology in screening for ADHD
 
ESO system, ES-BC system and EIS-GS system - Miami University study
ESO system, ES-BC system and EIS-GS system - Miami University study ESO system, ES-BC system and EIS-GS system - Miami University study
ESO system, ES-BC system and EIS-GS system - Miami University study
 
Es teck india
Es teck indiaEs teck india
Es teck india
 
EIS technology : bioimpedance application in selective serotonin reuptake
EIS technology : bioimpedance application in selective serotonin reuptakeEIS technology : bioimpedance application in selective serotonin reuptake
EIS technology : bioimpedance application in selective serotonin reuptake
 
EIS Technology: bioimpedance chronoamperometry in adjunct to screen the prost...
EIS Technology: bioimpedance chronoamperometry in adjunct to screen the prost...EIS Technology: bioimpedance chronoamperometry in adjunct to screen the prost...
EIS Technology: bioimpedance chronoamperometry in adjunct to screen the prost...
 
Summary of clinical investigations es teck complex system
Summary of clinical investigations es teck complex systemSummary of clinical investigations es teck complex system
Summary of clinical investigations es teck complex system
 
Stptg 4
Stptg 4Stptg 4
Stptg 4
 
Stptg 2
Stptg 2Stptg 2
Stptg 2
 
Sdptg1
Sdptg1Sdptg1
Sdptg1
 
Sdptg 3
Sdptg 3Sdptg 3
Sdptg 3
 
Pulseox[1]
Pulseox[1]Pulseox[1]
Pulseox[1]
 
Pulse wave indice
Pulse wave indicePulse wave indice
Pulse wave indice
 
Pulse wave analysis
Pulse wave analysisPulse wave analysis
Pulse wave analysis
 
Guidelines heart rate_variability_ft_1996[1]
Guidelines heart rate_variability_ft_1996[1]Guidelines heart rate_variability_ft_1996[1]
Guidelines heart rate_variability_ft_1996[1]
 
Arterioscler thromb vasc_biol_22_147
Arterioscler thromb vasc_biol_22_147Arterioscler thromb vasc_biol_22_147
Arterioscler thromb vasc_biol_22_147
 
Angitensin and ptg
Angitensin and ptgAngitensin and ptg
Angitensin and ptg
 
About pulseoximetry
About pulseoximetryAbout pulseoximetry
About pulseoximetry
 
Serotonin ssri effects
Serotonin ssri effectsSerotonin ssri effects
Serotonin ssri effects
 
Serotonin and blood flow
Serotonin and blood flowSerotonin and blood flow
Serotonin and blood flow
 
Oxidative stress
Oxidative stressOxidative stress
Oxidative stress
 

Último

Call Girls Gwalior Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Gwalior Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Gwalior Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Gwalior Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
(Rocky) Jaipur Call Girl - 09521753030 Escorts Service 50% Off with Cash ON D...
(Rocky) Jaipur Call Girl - 09521753030 Escorts Service 50% Off with Cash ON D...(Rocky) Jaipur Call Girl - 09521753030 Escorts Service 50% Off with Cash ON D...
(Rocky) Jaipur Call Girl - 09521753030 Escorts Service 50% Off with Cash ON D...indiancallgirl4rent
 
Call Girls Bangalore Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Bangalore Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Bangalore Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Bangalore Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...
All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...
All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...Arohi Goyal
 
Russian Escorts Girls Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls Delhi
Russian Escorts Girls  Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls DelhiRussian Escorts Girls  Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls Delhi
Russian Escorts Girls Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls DelhiAlinaDevecerski
 
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...astropune
 
VIP Call Girls Indore Kirti 💚😋 9256729539 🚀 Indore Escorts
VIP Call Girls Indore Kirti 💚😋  9256729539 🚀 Indore EscortsVIP Call Girls Indore Kirti 💚😋  9256729539 🚀 Indore Escorts
VIP Call Girls Indore Kirti 💚😋 9256729539 🚀 Indore Escortsaditipandeya
 
Call Girls Tirupati Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Tirupati Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Tirupati Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Tirupati Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Top Rated Bangalore Call Girls Ramamurthy Nagar ⟟ 9332606886 ⟟ Call Me For G...
Top Rated Bangalore Call Girls Ramamurthy Nagar ⟟  9332606886 ⟟ Call Me For G...Top Rated Bangalore Call Girls Ramamurthy Nagar ⟟  9332606886 ⟟ Call Me For G...
Top Rated Bangalore Call Girls Ramamurthy Nagar ⟟ 9332606886 ⟟ Call Me For G...narwatsonia7
 
Call Girls Faridabad Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Faridabad Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Faridabad Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Faridabad Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...
Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...
Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...Call Girls in Nagpur High Profile
 
💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...
💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...
💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...Taniya Sharma
 
Lucknow Call girls - 8800925952 - 24x7 service with hotel room
Lucknow Call girls - 8800925952 - 24x7 service with hotel roomLucknow Call girls - 8800925952 - 24x7 service with hotel room
Lucknow Call girls - 8800925952 - 24x7 service with hotel roomdiscovermytutordmt
 
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...Dipal Arora
 
VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋
VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋
VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋TANUJA PANDEY
 
Call Girls Visakhapatnam Just Call 9907093804 Top Class Call Girl Service Ava...
Call Girls Visakhapatnam Just Call 9907093804 Top Class Call Girl Service Ava...Call Girls Visakhapatnam Just Call 9907093804 Top Class Call Girl Service Ava...
Call Girls Visakhapatnam Just Call 9907093804 Top Class Call Girl Service Ava...Dipal Arora
 
(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...
(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...
(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...Taniya Sharma
 
Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...
Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...
Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...hotbabesbook
 
Call Girls Jabalpur Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Jabalpur Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Jabalpur Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Jabalpur Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...
Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...
Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...vidya singh
 

Último (20)

Call Girls Gwalior Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Gwalior Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Gwalior Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Gwalior Just Call 9907093804 Top Class Call Girl Service Available
 
(Rocky) Jaipur Call Girl - 09521753030 Escorts Service 50% Off with Cash ON D...
(Rocky) Jaipur Call Girl - 09521753030 Escorts Service 50% Off with Cash ON D...(Rocky) Jaipur Call Girl - 09521753030 Escorts Service 50% Off with Cash ON D...
(Rocky) Jaipur Call Girl - 09521753030 Escorts Service 50% Off with Cash ON D...
 
Call Girls Bangalore Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Bangalore Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Bangalore Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Bangalore Just Call 9907093804 Top Class Call Girl Service Available
 
All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...
All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...
All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...
 
Russian Escorts Girls Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls Delhi
Russian Escorts Girls  Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls DelhiRussian Escorts Girls  Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls Delhi
Russian Escorts Girls Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls Delhi
 
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...
 
VIP Call Girls Indore Kirti 💚😋 9256729539 🚀 Indore Escorts
VIP Call Girls Indore Kirti 💚😋  9256729539 🚀 Indore EscortsVIP Call Girls Indore Kirti 💚😋  9256729539 🚀 Indore Escorts
VIP Call Girls Indore Kirti 💚😋 9256729539 🚀 Indore Escorts
 
Call Girls Tirupati Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Tirupati Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Tirupati Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Tirupati Just Call 9907093804 Top Class Call Girl Service Available
 
Top Rated Bangalore Call Girls Ramamurthy Nagar ⟟ 9332606886 ⟟ Call Me For G...
Top Rated Bangalore Call Girls Ramamurthy Nagar ⟟  9332606886 ⟟ Call Me For G...Top Rated Bangalore Call Girls Ramamurthy Nagar ⟟  9332606886 ⟟ Call Me For G...
Top Rated Bangalore Call Girls Ramamurthy Nagar ⟟ 9332606886 ⟟ Call Me For G...
 
Call Girls Faridabad Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Faridabad Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Faridabad Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Faridabad Just Call 9907093804 Top Class Call Girl Service Available
 
Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...
Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...
Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...
 
💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...
💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...
💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...
 
Lucknow Call girls - 8800925952 - 24x7 service with hotel room
Lucknow Call girls - 8800925952 - 24x7 service with hotel roomLucknow Call girls - 8800925952 - 24x7 service with hotel room
Lucknow Call girls - 8800925952 - 24x7 service with hotel room
 
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
 
VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋
VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋
VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋
 
Call Girls Visakhapatnam Just Call 9907093804 Top Class Call Girl Service Ava...
Call Girls Visakhapatnam Just Call 9907093804 Top Class Call Girl Service Ava...Call Girls Visakhapatnam Just Call 9907093804 Top Class Call Girl Service Ava...
Call Girls Visakhapatnam Just Call 9907093804 Top Class Call Girl Service Ava...
 
(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...
(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...
(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...
 
Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...
Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...
Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...
 
Call Girls Jabalpur Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Jabalpur Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Jabalpur Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Jabalpur Just Call 9907093804 Top Class Call Girl Service Available
 
Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...
Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...
Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...
 

Ptg and cardiac output.jsp

  • 1. 31st Annual International Conference of the IEEE EMBS Minneapolis, Minnesota, USA, September 2-6, 2009 Noninvasive Cardiac Output Estimation Using a Novel Photoplethysmogram Index L. Wang, Emma Pickwell-MacPherson, Y. P. Liang and Y. T. Zhang Abstract—Cardiac output (CO) monitoring is essential for noninvasive, accurate, reliable and continuous. At present, indicating the perfusion status of the human cardiovascular no single method meets all these criteria. Intermittent system under different physiological conditions. However, it is thermodilution is widely accepted as the clinical golden currently limited to hospital use due to the need for either standard. This method requires the insertion of a pulmonary skilled operators or big, expensive measurement devices. artery catheter (PAC) to obtain one measurement per 3-4 Therefore, in this paper we devise a new CO indicator which can easily be incorporated into existing wearable devices. To minutes [1]. It is too invasive and non-continuous. Two this end, we propose an index, the inflection and harmonic area existing less invasive and continuous methods are ratio (IHAR), from standard photoplethysmographic (PPG) oesophageal Doppler monitoring and CO2 re-breathing, but signals, which can be used to continuously monitor CO. We both of these require skilled operators and expensive evaluate the success of our index by testing on sixteen measurement devices [1]. Amongst the currently used normotensive subjects before and after bicycle exercise. The methods, impedance cardiography is probably the only results showed a strong intra-subject correlation between IHAR and COimp measured by the bio-impedance method in noninvasive and automatic technique. However, the fifteen subjects (mean r = 0.82, p<0.01). After least squares impedance device is big and expensive, and its accuracy is linear regression, the precision between COimp and CO often influenced by the change of electrode positions and the estimated from IHAR (COIHAR) was 1.40 L/min. The total sweat on the skin [2]. Due to the disadvantages mentioned percentage error of the results was 16.2%, which was well above, these methods are all unquestionably limited to below the clinical acceptance limit of 30%. The results suggest bedside use. They are not portable, or wearable, so they are that IHAR is a promising indicator for wearable and noninvasive CO monitoring. difficult to incorporate into home health care monitoring systems. To solve this problem, one of the best ways is to I. INTRODUCTION derive a new CO indicator from signals provided by the existing wearable device. Some preliminary studies have I T is well known that heart rate, blood pressure and cardiac output (CO) are all essential physiological parameters of the human cardiovascular system. CO, defined as the blood implied that the PPG signal could be a candidate for this application [3]-[6]. The PPG signal, which indicates the blood volume volume ejected by the heart per minute (unit: L/min, where changes on site, is an optical signal that could be non- “L” means liter), is regarded as the ultimate expression of invasively obtained from body peripheral terminals, such as cardiovascular performance, since it indicates how well the ear, finger and toe. The PPG acquisition components, heart is able to provide enough nutrition and oxygen to the including a pair of LED emitter and receiver, and related peripheral organs and tissues. For human beings, in order to simple circuits, are cheap and small in size, and could be maintain a normal state of tissue perfusion and oxygen easily embedded into many existing wearable devices. delivery condition, the baseline CO should be in the range of Although PPG is obtained peripherally, many previous 4 L/min to 8 L/min. If CO gets out of this range, it is often a studies have revealed that the PPG wave contour is primarily sign of cardiovascular disease, such as hypertension, stroke influenced by characteristics of the systemic circulation, but or heart failure. Hence, continuous CO monitoring plays an not the local perfusion [3]-[6]. Henry Lax and his colleagues essential role in the evaluation, treatment, and follow-up of [3] noted that when individuals had cold fingers, this critically ill patients. “reduces the over-all amplitude but does not affect the Ideally, a technology which measures CO should be configuration of the pulse wave”. Chowienczyk PJ found in his experiment that the vasodilation drugs which increased Manuscript received April 7, 2009. This work was supported in part by the local circulation in the upper arm could not change the the Hong Kong Innovation and Technology Fund (ITF). The authors are corresponding finger PPG wave contour, while, the grateful to Standard Telecommunication Ltd., Jetfly Technology Ltd., Golden Meditech Company Ltd., Bird International Ltd. and Bright Steps systematic insertion of vasodilation drugs, such as GTN, Corporation of their supports to the ITF projects. could change it substantially [4]. More recently, we have Y. T. Zhang (phone: 852-2609-8459; e-mail: ytzhang@ee.cuhk.edu.hk), successfully utilized a parameter derived from the PPG wave Emma Pickwell-MacPherson, Y. P. Liang and L. Wang are with the Joint Research Centre for Biomedical Engineering, Department of Electronic to trace the change of total peripheral resistance (TPR) after Engineering, The Chinese University of Hong Kong, Hong Kong SAR. bicycle exercise [5]. McCombie proposed a blind system Y. T. Zhang is also with the Institute of Biomedical and Health identification method to calculate the cardiac output Engineering, Shenzhen Institute of Advanced Technology, Chinese Academy of Sciences, China, and Key Laboratory for Biomedical waveform from two PPG signals measured from different Informatics and Health Engineering, Chinese Academy of Sciences, China. body locations, but only a qualitative aortic flow curve was 978-1-4244-3296-7/09/$25.00 ©2009 IEEE 1746 Authorized licensed use limited to: ALBERT MAAREK. Downloaded on January 7, 2010 at 12:38 from IEEE Xplore. Restrictions apply.
  • 2. obtained [6]. invasive arterial catheter or a bedside Finapres device for In this paper, we propose a novel indicator of CO, the acquiring the continuous blood pressure measurement. inflection and harmonic area ratio (IHAR), which is derived from PPG signals. The ability of IHAR to trace CO changes PPG was evaluated in a bicycle exercise experiment. Time domain Frequency domain ABP FFT(f1) magnitude FFT(f2) S1 S2 CO AC TPR f1 f2 … … fn Hz frequency (a) (b) Fig. 2. (a) is PPG wave in the time domain. S1 and S2 are the areas under the whole PPG wave and the part of wave after the point of Fig. 1. Two-element Windkessel model. inflection, respectively. IPA=S2/S1. (b) is PPG wave in the frequency domain. FFT ( f n ) is the magnitude at the nth harmonic. NHA = N N II. METHODOLOGY ∑ FFT ( f ) ∑ FFT ( f ). n =2 2 n n =1 2 n According to the two-element Windkessel model shown However, as shown in equation (1), blood pressure in Fig. 1, CO could be calculated from the analysis of measurement is not a necessity for obtaining CO, if proper continuous arterial blood pressure waveform by a so called surrogates of MBP and TPR can be derived from other pulse contour method. In this model, the cardiovascular signals. According to wave reflection theory, arterial blood system is analogous to a current source connected with a pulse could be divided into two waves: a first wave two-element circuit. CO, which is mimicked by the mean produced by heart pumping and a second wave produced by amount of current passing through the TPR (total peripheral pulse wave reflection. Therefore, the inflection point area resistance), equals the mean pressure (mean arterial blood ratio (IPA), the area ratio of the second and first peak in the pressure, MBP) divided by the TPR: PPG wave (see Fig. 2 (a)), is mainly influenced by the CO = MBP TPR (1). strength of pulse wave reflection. Pulse wave reflection In this method, if a continuous arterial pressure waveform is results from the impedance mismatch between different parts obtained, MBP could be calculated beat to beat. TPR is in the arterial system, e.g., the compliance and resistance firstly initialized by a pair of calibration CO and MBP data, mismatches between the big, elastic arteries and the small and its value of the current beat is calculated from MBP and arteries. Studies have shown that approximate 90% of the estimated CO of the previous beat, iteratively. The main TPR is located in the small arteries. Hence, if the small shortcoming of such technique is that it needs either an arteries contract, the TPR will change, which will change the 5-min. 5-min. 60W 5-min. 60W 5-min. 60W rest bicycle exercise bicycle exercise bicycle exercise …… …… …… …… 1 2 3 4 5 6 7 8 9 10 11 17 18 19 25 3-min. 6-min. 6-min. 6-min. Resting Post-exercise 1 Post-exercise 2 Post-exercise 3 represented BP measurement by a registered nurse numbers 1-25 indicate the datasets of simultaneous measurements of CO and PPG Fig. 3. Experiment procedure. 1747 Authorized licensed use limited to: ALBERT MAAREK. Downloaded on January 7, 2010 at 12:38 from IEEE Xplore. Restrictions apply.
  • 3. impedance mismatch, influence the strength of the wave IV. DATA ANALYSIS reflection, and further change IPA. In our previous For each subject, 25 datasets were obtained, including 1 investigation [5], IPA has shown to be a very good indicator dataset at rest and 8 datasets in each post exercise phase (see of TPR when it changes after exercise. The pulse wave Fig. 3). The 8 datasets after each exercise were obtained reflection not only changes the shape of the PPG signal in from the 6-min continuous data divided by the intermittent the time domain, but also in the frequency domain. Our BP measurements. As shown in Fig. 3, BP was also previous study [7] has proved that the normalized harmonic measured by a nurse every other dataset, e.g., during datasets area (NHA), a parameter derived from frequency domain 2, 4, 6, 8 for post exercise phase 1. analysis (see Fig. 2 (b)), is related to pulse wave reflection For offline analysis, each dataset of PPG signals was re- and is strongly correlated with systolic and diastolic blood sampled to 100 Hz and divided into beat-to-beat waves. The pressure. Therefore, IHAR, defined as NHA divided by IPA, start point of each beat is defined as the beginning of its is proposed as a potential CO indicator and expressed as systolic rise, which is also the end of the diastolic fall wave follows: of the previous beat. Single PPG wave was transferred to the ⎛ N ⎞ frequency domain by 512 points FFT (ensuring a resolution ⎜ ∑ FFT 2 ( f n ) ⎟ ⎜1 − n=2 ⎟ of more than 10 points per Hz). Then, beat-to-beat IHAR ⎜ N ⎟ was calculated using equation (2). Meanwhile, COimp was ⎜ ∑ FFT ( f n ) ⎟ 2 obtained by the impedance cardiograph device, beat by beat. IHAR = ⎝ n=1 ⎠ (2), “Beat by beat” here meant that COimp was acquired and ⎛ S2 ⎞ saved in a “beat to beat” format, but was actually calculated ⎜ ⎟ ⎜S ⎟ ⎝ 1⎠ from the rolling average of eight impedance wave and five where S1 and S2 are the areas under the whole PPG wave and ECG wave in the algorithm. As a result, it was impossible in the part of wave after the point of inflection in the time this study to compare IHAR with COimp in a beat-to-beat domain, respectively (see Fig. 2(a)), and FFT 2 ( f n ) is the way, but only to compare the means of each dataset. For CO estimation, all datasets of IHAR and COimp were firstly used square of the magnitude at the nth harmonic of the PPG wave to produce CO = α ⋅ IHAR + β by the least squares linear in the frequency domain (see Fig. 2(b)). regression method, and then, the COIHAR results were The performance of IHAR for CO estimation was obtained from IHAR by the above equation. evaluated in a bicycle exercise experiment. An impedance cardiograph device, Physio Flow PF-05 (Manatec V. RESULTS Biomedical, Macheren, France), was used as the benchmark of CO measurement. The validity of this device has been As shown in Table I, the hemodynamic parameters, established in both normotensive adults and cardiovascular including heart rate, systolic blood pressure, diastolic blood patients versus the direct Fick method (lab golden standard pressure and cardiac output, have significantly changed after of CO measurement) both at rest and exercise conditions [2] each 5 min of bicycle exercise, compared with those [8]. obtained at rest (p<0.05). The intra-subject correlation coefficients (r) between III. EXPERIMENT PROTOCOL IHAR and COimp are listed in Table II. For all subjects, IHAR was positively correlated to COimp. Except for subject 16 normotensive volunteers (8 males and 8 females), aged 13 (p=0.054), there was a strong correlation between IHAR from 22 to 34 years, participated in this study. Informed and COimp in the other 15 subjects (mean r = 0.82, p<0.01). consent was obtained from each participant. The experimental procedures are depicted in Fig. 3. Upon arrival, TABLE I CHANGES OF HEMODYNAMIC PARAMETERS AFTER 1 , 2 , AND 3RD ST ND subjects were asked to rest for 5 minutes. Then, the systolic EXERCISE COMPARED WITH THOSE AT REST blood pressure (SBP) and diastolic blood pressure (DBP) 1st 2nd 3rd were measured from the left arm by a registered nurse. After At Rest Exercise Exercise Exercise this, PPG and cardiac output were recorded simultaneously Heart Rate 76±10 92±17﹡ 95±17﹡ 100±16﹡ for 1 min. Then, the subjects were asked to ride the bicycle (beat/s) SBP (mmHg) 106±12 116±14﹡ 117±14﹡ 118±15﹡ (Corival V2, Lode VB, Groningen, Netherland) for 5 DBP (mmHg) 61±7 59±10﹡ 60±10﹡ 61±10﹡ minutes at 60 W. Immediately after exercise, 6-min of PPG CO (L/min) 6.3±1.2 8.5±1.8﹡ 8.7±1.9﹡ 9.2±2.0﹡ and CO data were continuously acquired, meanwhile, at time ﹡ significant difference with the results obtained at rest (p<0.01). points after exercise of 0 min, 1.5 min, 3 min, 4.5 min and 6 min, blood pressure (BP) was measured. Then, the same 5 The dynamic range of CO estimated from IHAR (COIHAR) min exercise and post-exercise data acquisition procedures was very similar to that of COimp (COIHAR: 4.02 L/min-14.72 were repeated twice. PPG was recorded on the right index L/min; and COimp: 3.66 L/min-14.48 L/min). As shown in finger by an in-house designed acquisition system. CO was Table II, the standard deviation (SD) of the estimation measured by the Physio Flow PF-05 impedance cardiograph residue is 0.71 L/min. This value does not vary much with device. 1748 Authorized licensed use limited to: ALBERT MAAREK. Downloaded on January 7, 2010 at 12:38 from IEEE Xplore. Restrictions apply.
  • 4. subject variance, which indicates a stable performance of percentage error (PE = 16.2 %) of our technique is well IHAR based CO estimation. In many previous CO studies, a below the clinically acceptable error of 30%. percentage error (PE) is calculated to indicate the variance of A potential limitation of this work is the change of PPG estimation error with respect to reference mean wave as a result of vascular aging or cardiovascular diseases, ( PE = 1.96SD MEANCO ), and a PE less than 30% is e.g., the inflection point may be blurred and difficult to find ref as the arterial stiffness increases. Therefore, further regarded to be clinically acceptable [9] [10]. The PE of the validation on subjects with a bigger age range and results in this study is 16.2 %. cardiovascular patients are needed. Another limitation is the TABLE II calibration procedure. Since the main purpose of this study THE SUMMARY OF THE INTRA-SUBJECT CORRELATION COEFFICIENT r is to propose and evaluate a novel CO index, we have not BETWEEN COIMP AND IHAR AS WELL AS THE REGRESSION COEFFICIENT fully investigated the calibration procedure. Only a simple AND STANDARD DEVIATION (SD) OF RESIDUE IN THE LEAST SQUARE LINEAR REGRESSION least squares linear regression was utilized. However, COIPHA = α ⋅ IHAR + β SD of individual calibration is an essential procedure for IHAR Subject r Residue based CO estimation and will be investigated further in a α β (L/min) separate study. S01 0.94﹡ 195.1 3.7 0.83 S02 0.94﹡ 189.6 2.9 0.62 S03 0.78﹡ 265.0 0.9 0.61 REFERENCES S04 0.92﹡ 199.7 2.8 0.64 [1] L. Mathews and K. RK Singh, “Cardiac output monitoring,” Annals of S05 0.58﹡ 689.0 -3.1 0.87 Cardiac Anaesthesia, vol. 11, Jan.-Jun. 2008, pp. 56-68. S06 0.80﹡ 272.0 0.0 0.78 [2] R. Richard, E. Lonsdorfer-Wolf, A. Charloux, et al., “Non-invasive S07 0.91﹡ 347.9 -0.4 0.49 cardiac output evaluation during a maximal progressive exercise test, S08 256.1 4.8 0.86 using a new impedance cardiograph device,” Eur J Appl Physiol., vol. 0.79﹡ 85, Aug. 2001, pp. 202-7. S09 0.96﹡ 172.1 2.0 0.38 [3] H. Lax, A. Feinberg and B. M. Cohen, “Studies of the arterial pulse S10 0.89﹡ 319.7 0.9 0.74 wave and its modification in the presence of human arteriosclerosis,” S11 0.83﹡ 396.6 -0.3 0.85 J Chronic Dis., vol. 3, Jun. 1956, pp. 618-31. S12 0.96﹡ 282.3 2.4 0.43 [4] P. J. Chowienczyk, R. P. Kelly, H. MacCallum, et al., S13 0.39 185.9 4.9 0.78 “Photoplethysmographic assessment of pulse wave reflection: blunted S14 0.78﹡ 345.1 0.2 0.90 response to endothelium-Dependent Beta2-Adrenergic vasodilation in S15 0.69﹡ 283.7 4.2 0.83 type II diabetes mellitus,” J Am Coll Cardiol, vol. 34, Aug. 1999, pp. S16 0.90﹡ 245.8 4.2 0.72 2007-14. [5] L. Wang and Y. T. Zhang, “A novel photoplethysmogram index for Mean 0.82 0.71 ﹡ total peripheral resistance after bicycle exercise,” Proc. of the 5th p<0.01. International Conference on Ubiquitous Healthcare, Pusan, Korea, Oct. 2008, pp. 175-6. The Bland-Altman plot (see Fig. 4) shows the agreement [6] D. McCombie, A. Reisner and H. Asada, “Identification of vascular of the estimated CO and the reference CO. The percentage dynamics and estimation of the cardiac output waveform from of CO estimations within the 95% confidence interval ([- wearable PPG sensors,” Proc. of the 27th IEEE International Engineering in Medicine and Biology Conference, EMBC 2005, 1.40 L/min -1.40 L/min]) is 95.9%. Shanghai, China, pp. 3490-3. [7] Y. S. Yan and Y. T. Zhang, “Noninvasive estimation of blood pressure using photoplethysmographic signals in the period domain,” 3 Proc. of the 27th IEEE International Engineering in Medicine and Difference of Reference an d 2 Biology Conference, EMBC 2005, Shanghai, China, pp. 3583-4. E stim ated CO (L /m in ) [8] A. Charloux, E. Lonsdorfer-Wolf, R. Richard, et al., “A new 1 impedance cardiograph device for the non-invasive evaluation of cardiac output at rest and during exercise: comparison with the 0 95.9% “direct” Fick method,” Eur J Appl Physiol., vol. 82, Jul. 2000, pp. 0 5 10 15 20 313-20. -1 [9] M. Biais, K. Nouette-Gaulain, V. Cottenceau, et al., “Cardiac output measurement in patients undergoing liver transplantation: pulmonary -2 artery catheter versus uncalibrated arterial pressure waveform analysis,” Anesth Analg., vol. 106, May 2008, pp. 1480-6. -3 [10] L. A. Critchley and J. A. Critchely, “A meta-analysis of studies using Average of Reference and Estimated CO (L/min) bias and precision statistics to compare cardiac output measurement techniques,” J Clin Monit Comput., vol. 15, Feb. 1999, pp. 85-91. Fig. 4. Bland-Altman plot of cardiac output. VI. CONCLUSION In this study, we have tested the performance of a new CO index, IHAR, in 16 normotensive subjects through a bicycle exercise study. The strong intra-subject correlation suggests that IHAR can successfully trace the CO changes over a wide dynamic range before and after exercise. The small 1749 Authorized licensed use limited to: ALBERT MAAREK. Downloaded on January 7, 2010 at 12:38 from IEEE Xplore. Restrictions apply.