SlideShare una empresa de Scribd logo
1 de 63
Instruments for Optical
Spectrometry
• Optical spectroscopy encompasses regions
of the electromagnetic spectrum that are not
strictly within the visible range. However,
the components of instruments used in the
IR, visible, and UV are all very similar.
Instrument Components
• Spectroscopic instruments are usually constructed
from the following components:
• A stable source of radiant energy
• A wavelength selector that permits the selection and
isolation of a restricted wavelength region
• One or more sample containers
• A radiation detector or transducer that converts
radiant energy to a measurable signal – normally
electrical
• A signal processor and readout
Instrument Components
• In emission instruments the source and sample are
combined so that emission occurs from heating or
electrical excitation of the sample.
• In absorption and fluorescence instruments the
analyte is held in a sample container and an
external radiant source is employed.
• In absorption instruments the source, sample, and
detector are located along an axis. In fluorescence
instruments the source is placed at an angle of 90°
to the detector-sample axis.
Optical Materials
• The optical components of a spectrometer must be
optically transparent in the wavelength region
being employed for analysis.
• In IR studies, sodium chloride, potassium
chloride, or silver chloride are used.
• In the visible region, borosilicate glass or plastics
such as polystyrene are adequate.
• In the UV, quartz is used because glass absorbs
too much radiation at these wavelengths.
Spectroscopic Sources
• The source must provide a beam of sufficient intensity
in the wavelength region of interest to allow detection
and measurement. It must be stable and not flicker so
P and Po can be measured accurately. Some
instruments measure these values simultaneously so
stability is less of a factor.
• Both line sources and continuous sources are used in
optical spectroscopy. The continuous source is made up
of a whole range of emitting wavelengths that overlap
while line sources emit discrete bands at one or two
wavelengths.
Tungsten Filament Lamps
• Theses emit in the range 320 – 2500 nm so most
of the emissions are in the IR.
• The lamp is useful for visible spectroscopy but not
for UV because the lamp housing absorbs strongly
at these wavelengths.
• The lamp should have a stable voltage supply so
that its spectral characteristics remain constant – it
tends to vary as its temperature changes.
Tungsten Halogen Lamps
• Tungsten halogen lamps are an improvement over
tungsten filament lamps because they have a
longer lifetime and their emission spectra extend
into the UV.
• The lamp contains a small quantity of iodine
within a quartz envelope that houses the filament.
• The lifetime of the tungsten halogen lamp is more
than double that of the tungsten filament lamp.
• In the tungsten filament lamp, the filament
eventually sublimes.
Tungsten Halogen Lamps
• In a tungsten iodine lamp the tungsten
sublimes but reacts with the iodine to give
WI2, which diffuses back to the filament
where it decomposes and re-deposits
tungsten.
Hydrogen and Deuterium Lamps
• These lamps generate true continua. The process
for hydrogen may be shown as:
• H2 + E → H2* → H' + H'' + hν where E is the
energy required to excite the hydrogen molecule.
• The products, H' and H'' may have a variety of
kinetic energies, which means that the photon
emitted, hν, will also vary in energy – thus the
continuum.
Hydrogen and Deuterium Lamps
• The useful range of these lamps is from 160 – 375
nm. Deuterium is preferred to hydrogen because
of its greater intensity.
• The emission is stimulated by generating an
electrical arc in an envelope filled with the gas. A
regulated power supply is necessary for constant
lamp intensity.
Wavelength Selectors
• Spectroscopic instruments need a wavelength
selector to provide a range of wavelengths to
impinge on the sample. The narrower the range of
wavelengths chosen by the wavelength selector,
the more likely Beer’s law is to be followed. This
also leads to improved sensitivity and selectivity.
• The closest type of radiation to a purely
monochromatic source is a laser.
• With normal optical instruments the wavelength
selector provides a range of wavelengths centred
at a nominal wavelength.
Wavelength Selectors
• The range of wavelengths provided by the
wavelength selector is known as the
effective bandwidth or just bandwidth and
is defined as the width of the band in
wavelength units at half peak height. The
ordinate in this plot is transmittance so the
bandwidth is the range for 0.5T.
Wavelength Selectors
• Bandwidths vary enormously from one
wavelength selector to another. For
instance, a high quality monochromator for
the visible region may have a bandwidth of
a few tenths of a nanometre or less. An
absorption filter operating in the same
region may have a bandwidth of 200 nm or
more.
Wavelength Selectors
• Generally two types of wavelength selector
are used, filters or monochromators.
• Filters are rugged, inexpensive, and easy to
maintain.
• Monochromators are costly but allow
nominal wavelengths to be changed
continuously during irradiation of the
sample.
Radiation Filters
• A filter allows only a restricted band of
wavelengths to pass through it.
• Two types of filter are used in spectroscopic
instrumentation, interference filters and
absorption filters.
• Interference filters allow a greater
transmittance of radiation at the nominal
wavelength than absorption filters.
Monochromators
• Two types of monochromators have been used to
disperse multiwavelength radiation into its various
components. These are prisms and diffraction
gratings.
• Prisms are used infrequently nowadays because
gratings can be made more cheaply. Gratings
scratched on the surface of a mirror lead to more
compact instrumentation.
Monochromators
• In a typical reflective grating
monochromator, radiation from a source
enters the monochromator by a narrow
opening called a slit. The radiation is then
collimated by a concave mirror so that a
parallel beam strikes the grating. Angular
dispersion from the grating then occurs at
the reflective surface.
Monochromators
• Consider two wavelengths of radiation, λ1 and λ2,
being dispersed by this grating. If λ1 < λ2 radiation
of wavelength λ1will be reflected from the grating
at a sharper angle than that of wavelength λ2. This
is known as angular dispersion.
• After dispersion the two beams of radiation are
focused by a concave mirror on to the focal plane
of the instrument where they appear as two
images. By rotating the grating either one of the
images can be focused on the exit slit.
Gratings
• Most gratings used in commercial instruments are
replica gratings obtained by making castings from a
master grating. The master grating is prepared by
scoring many parallel and closely spaced grooves on a
flat polished surface with a sharp diamond tool. In UV-
visible work the grating will have from 300-2000
grooves/mm.
• Preparation of the master grating is tedious and
expensive. Replica gratings are prepared from a liquid
resin casting process that preserves the optical
accuracy of the original master grating. The surface of
the replica is made reflective with a coating of
aluminum, gold, or platinum.
Radiation Detectors and
Transducers
• A detector is a device that indicates the existence
of some physical phenomenon. Examples are the
human eye, photographic film, the mercury level
in a thermometer, etc.
• A transducer is a special type of detector that
converts a physical phenomenon, e.g. light
intensity, pH, mass, etc., into electrical signals that
can be amplified, manipulated, and eventually
turned into numbers that are related to the
magnitude of the original signal.
Properties of Transducers
• The ideal electromagnetic radiation
transducer responds rapidly to low levels of
radiant energy over a broad wavelength
range.
• The electrical signal it produces must be
easily amplified and have a relatively low
noise level.
• It is essential that the electrical signal produced by the
transducer is directly proportional to the beam power,
P.
• G = KP + K'
• G is the electrical response of the detector in units of
current, resistance, or potential
• K is a proportionality constant which measures the
sensitivity of the detector in terms of electrical response
per unit of radiant power
• K' represents the dark current. This is a small constant
response, present even when no radiation strikes the
detector. When K' is reduced to a negligible quantity
by electrical subtraction etc., G = KP
Properties of Transducers
Types of Transducers
• Two general types of transducers are used
in optical spectroscopy. UV-visible
transducers respond to light or photons
Photon Detectors
• Types of photon detectors that are used
commercially are: phototubes,
photomultiplier tubes, silicon photodiodes.
• Phototubes are one of the simpler types to
understand because they work on the
Einstein photo-emissive cathode principle.
Photomultiplier Tubes
• This is similar in design to the phototube but is much
more sensitive.
• The surface of the cathode is similar in composition to
that of the phototube. Electrons are emitted from its
surface and accelerated towards the surface of another
electrode called a dynode. Upon striking the dynode,
each electron produces several additional electrons
which are then accelerated towards another dynode.
This process continues at a number of dynodes until 106
– 107
electrons have been produced for each photon.
• The cascade is collected at the anode where the
resulting electric current is further amplified and
measured.
Silicon Photodiodes
• These are transistor devices used in diode array
detectors. In these devices, over 1000 can be
fabricated side by side on a small silicon chip –
each diode is about 0.02 mm thick.
• When one or more diode array detectors are
placed along the focal plane of a monochromator,
the intensity of all wavelengths can be measured
simultaneously.
Signal Processors and Readouts
• The signal processor is an electronic device that
amplifies the electrical signal from the detector. It
may also subtract unwanted signals.
• Signal processors may display the signal in a
number of ways including derivative and
integrated peak areas, etc.
• Digital readout devices are most frequently found
in modern instruments. However, chart recorders
are used a great deal still.
Optical Instrument Design
• Optical spectroscopic instruments vary from
the very sophisticated which are expensive
and used for research, to the simple
inexpensive variety which are used for
routine chemical analysis. It is the nature of
the components that determines the cost and
eventual use.
Common Names of Some
Optical Instruments
• A spectroscope is an instrument for visually identifying
species excited in a flame or by other means. The
instrument has a monochromator and an eyepiece in
the focal plane to observe emission lines. It was used in
early spectroscopic studies.
• A colorimeter is an instrument for absorption
measurements in which the eye serves as the detector.
Comparison standards are used to quantify the amount
of analyze. This is a limited use of the term colorimeter
which is now usually applied to any photometer used
for absorption measurements in the visible region of
the spectrum.
Common Names of Some
Optical Instruments
• A photometer is a photoelectric instrument
that can be used for absorption, emission, or
fluorescence measurements. Most
colorimeters are photometers because they
use filters for wavelength selection.
Common Names of Some
Optical Instruments
• A spectrograph records spectra on a photographic
film or plate, placed along the focal plane of the
monochromator. The spectra appear as a set of
black images of the entrance slit. Spectrographs
are usually used for qualitative analysis.
• A spectrometer is a monochromator equipped with
a fixed slit at the focal plane. When the detection
device is a phototransducer the instrument is
known as a spectrophotometer.
Instrument Types
• Three general types of spectroscopic
instruments are in general use.
• These are single beam, double beam in
space, and double beam in time.
• All three types are used in absorption,
emission, and fluorescence studies.
Single Beam Instruments
• In a single beam instrument used in
absorption studies, a filter or
monochromator is used to select the
wavelength of radiation incident on the
sample holder.
• Matched cells can be interposed alternately
in the radiation beam which passes on to a
detector, usually a photomultiplier tube.
Single Beam Instruments
• To operate a manual instrument the following
adjustments are made:
• (a) The percent transmittance is corrected to zero
by placing a shutter between the source and the
detector and adjusting the meter until it reads zero.
• (b) solvent is placed in the radiation path, and the
shutter opened. The percent transmittance reading
of the meter is adjusted to 100 percent.
Single Beam Instruments
• A logarithmic scale can be used to determine
absorbance; a linear scale is used to determine
transmittance.
• Single beam instruments vary in complexity.
Spectrometers
Double Beam Instruments
• In a double beam instrument, two beams of
radiation are formed from the same source by a V-
shaped mirror called a ‘beam splitter’.
• One beam passes through the reference solution to
a detector while the second simultaneously passes
through a solution containing analyte to a second
matched detector.
Double Beam Instruments
• The two outputs are amplified and their ratio (or
log ratio) is determined electronically and
displayed by the readout device.
Spectrometers
Double Beam Instruments
• A second type of double beam instrument utilizes
a rotating sectored mirror to direct the radiation
from the source, as a pulse, either through a
reference cell or through the sample.
• The pulses of radiation are recombined by another
sectored mirror before detection.
• Pulsed radiation is the basis of many modern
research instruments.
Double Beam Instruments
• Double beam instruments compensate for
anything but very short term fluctuations in
radiant output from the source, as well as for any
drift in the detector and amplifier.
• They also compensate for any wide variation in
source intensity with wavelength.
• This design lends itself particularly well to the
continuous recording of spectra. Most modern
UV-visible recording instruments are based on
this design.
Spectrometer
Time separated double beam
Molecular Absorption
Spectroscopy
• Molecular optical spectroscopy is useful for both
qualitative and quantitative measurements.
• UV-visible spectrophotometry is mainly used for
quantitative work.
Ultraviolet and Visible
Absorption Spectroscopy
• Molecules absorb a finite amount of energy to
place the molecule in an upper electronic state.
Various vibrational and rotational levels are
occupied in this excited electronic state. The
absorption spectrum should reflect this.
• In the gas phase, it is possible to observe
absorption of energy to different vibrational and
rotational energy levels. This is a resolved
spectrum with fairly narrow lines.
Ultraviolet and Visible
Absorption Spectroscopy
• In solution or in condensed phases the
spectrum loses its resolution and becomes a
broad band.
• This is because the molecules of absorbing
species can bump into one another more
easily, thus creating a variety of energies
that cannot be resolved.
Absorption by Organic
Compounds
• The wavelength at which organic molecules
absorb UV-visible radiation depends to a large
extent on how firmly bound the electrons are in
the molecule.
• Electrons in single covalent bonds are tightly held
between the atoms in the bond. Excitation of these
electrons is only possible by using wavelengths of
radiation below 180 nm.
• This is of no use for routine measurements
because quartz and various atmospheric
components absorb at these wavelengths.
Absorption by Organic
Compounds
• Electrons in double or triple bonds are more easily
excited because they are more loosely held than
single bond electrons. Unsaturated molecules have
useful absorption spectra.
• The unsaturated organic functional groups that
absorb UV-visible radiation are known as
chromophores. They can serve as a very rough
guide for identification purposes.
Absorption by Organic
Compounds
• Peak position is influenced by a number of factors
such as:
• Solvent effects
• Other structural details of the molecule, such as
conjugation
• Molecules containing heteroatoms with lone pairs
of nonbonding electrons may also absorb in the
UV-visible region of the spectrum and provide
qualitative information on the molecule.
Absorbing species
• Electronic transitions
π, σ, and n electrons
– d and f electrons
– Charge transfer reactions
∀π, σ, and n (non-bonding) electrons
Sigma and Pi orbitals
Electron transitions
Transitions
∀ σ−>σ∗
– UV photon required, high energy
• Methane at 125 nm
• Ethane at 135 nm
• n-> σ∗
– Saturated compounds with unshared e-
• Absorption between 150 nm to 250 nm
∀ε between 100 and 3000 L cm-1
mol-1
• Shifts to shorter wavelengths with polar solvents
– Minimum accessibility
– Halogens, N, O, S
Transitions
• n->π∗, π−>π∗
– Organic compounds, wavelengths 200 to 700
nm
– Requires unsaturated groups
• n->π∗ low ε (10 to 100)
– Shorter wavelengths
∀π−>π∗ higher ε (1000 to 10000)
Absorption by Inorganic Species
• Typically visible absorption of visible radiation
occurs with compounds of elements in the first
and second row of the transition series. This is
why these compounds are colored.
• The absorption involves transitions between filled
and unfilled d-orbitals with energies that depend
on the ligands attached to the elements.
• The position of the energy transition depends on
the position of the element in the periodic table, its
oxidation state, and the nature of the attached
ligand.
Absorption by Inorganic Species
• The lanthanide elements behave somewhat
differently in that the peaks are fairly
narrow.
• This is because transitions are between
electrons in f-orbitals. They are shielded
from effects of the outer electrons and the
energy levels are thus not distorted.
Qualitative Applications of UV-
visible Spectroscopy
• UV-visible qualitative analysis is useful for
determining the presence of certain gross
structural features of molecules, e.g.
unsaturation.
• However it needs to be supplemented by
other spectroscopic methods such as IR and
NMR to obtain detailed information on
molecular structure.
Solvents
• The most useful information is derived from
spectra obtained from dilute solutions or from a
vapour of the analyte.
• Solvents should be of high purity and be
transparent throughout the spectral range.
• Generally non-polar solvents distort spectra least.
Polar solvents tend to obscure structure.
• Solvents tend to change the position of maximum
absorbance so in any qualitative measurement of
wavelength, the same solvent must be used for all
samples.
Quantitative Applications
• This is one of the most important tools available
for quantitative analysis.
• Important characteristics of the method are as
follows:
• (1) Wide applicability. Most compounds will
either absorb themselves or can be derivatized so
that the derivatives absorb. UV-visible
spectrophotometry is the single most important
analytical technique for quantitative chemical
analysis.
Quantitative Applications
• (2) High sensitivity. Typically measurements can
be made with detection limits as low as 10−5
mol/L.
With procedural modifications these detection
limits can be extended even further.
• (3) Moderate to high selectivity. This means that it
is sometimes, but not too often, possible to make
measurements on the analyte without prior
separation from other absorbing and interfering
compounds.
Quantitative Applications
• (4) Excellent accuracy. Normally the
precision of any given measurement will be
within 1–5%. Errors are reduced to similar
levels and can be made even lower by using
special measuring techniques.
• (5) Ease and convenience. Spectrometers
are not cheap, but are easy to use and lend
themselves to automation.

Más contenido relacionado

La actualidad más candente

Vandeemeter equation
Vandeemeter equationVandeemeter equation
Vandeemeter equation
Sonam Gandhi
 
'estimation of quinine sulphate by fluorescence spectroscopy with recordings...
 'estimation of quinine sulphate by fluorescence spectroscopy with recordings... 'estimation of quinine sulphate by fluorescence spectroscopy with recordings...
'estimation of quinine sulphate by fluorescence spectroscopy with recordings...
Priya Bardhan Ray
 
Theories of chromatography
Theories of chromatographyTheories of chromatography
Theories of chromatography
Khalid Hussain
 

La actualidad más candente (20)

INSTRUMENTATION OF UV-VISIBLE SPECTROSCOPY
INSTRUMENTATION OF UV-VISIBLE SPECTROSCOPYINSTRUMENTATION OF UV-VISIBLE SPECTROSCOPY
INSTRUMENTATION OF UV-VISIBLE SPECTROSCOPY
 
Instrumentation of IR spectroscopy
Instrumentation of IR spectroscopyInstrumentation of IR spectroscopy
Instrumentation of IR spectroscopy
 
UV-Visible spectroscopy
UV-Visible spectroscopyUV-Visible spectroscopy
UV-Visible spectroscopy
 
Instrumentation IR Spectroscopy: Detectors
Instrumentation IR Spectroscopy: DetectorsInstrumentation IR Spectroscopy: Detectors
Instrumentation IR Spectroscopy: Detectors
 
Fluorimetry
FluorimetryFluorimetry
Fluorimetry
 
Infrared instrumentation
Infrared instrumentationInfrared instrumentation
Infrared instrumentation
 
Uv visible spectroscopy-instrumentation
Uv visible spectroscopy-instrumentationUv visible spectroscopy-instrumentation
Uv visible spectroscopy-instrumentation
 
Vandeemeter equation
Vandeemeter equationVandeemeter equation
Vandeemeter equation
 
Deviations from Beers law
Deviations from Beers lawDeviations from Beers law
Deviations from Beers law
 
Spectrophotometric Instruments (Detector) and Application of UV – VIS spectro...
Spectrophotometric Instruments (Detector) and Application of UV – VIS spectro...Spectrophotometric Instruments (Detector) and Application of UV – VIS spectro...
Spectrophotometric Instruments (Detector) and Application of UV – VIS spectro...
 
'estimation of quinine sulphate by fluorescence spectroscopy with recordings...
 'estimation of quinine sulphate by fluorescence spectroscopy with recordings... 'estimation of quinine sulphate by fluorescence spectroscopy with recordings...
'estimation of quinine sulphate by fluorescence spectroscopy with recordings...
 
Sampling techniques for ir
Sampling techniques for irSampling techniques for ir
Sampling techniques for ir
 
Theories of chromatography
Theories of chromatographyTheories of chromatography
Theories of chromatography
 
Thin Layer Chroatography
Thin Layer ChroatographyThin Layer Chroatography
Thin Layer Chroatography
 
infrared spectroscopy and factors effecting the IR
infrared spectroscopy and factors effecting the IRinfrared spectroscopy and factors effecting the IR
infrared spectroscopy and factors effecting the IR
 
Detectors
DetectorsDetectors
Detectors
 
Fluorimetry
FluorimetryFluorimetry
Fluorimetry
 
Derivative spectroscopy
Derivative spectroscopyDerivative spectroscopy
Derivative spectroscopy
 
Principles of fluorimetry by ANN
Principles of fluorimetry by ANNPrinciples of fluorimetry by ANN
Principles of fluorimetry by ANN
 
FTIR spectroscopy
FTIR spectroscopyFTIR spectroscopy
FTIR spectroscopy
 

Destacado

MolecularSpectroscopyPortfolio
MolecularSpectroscopyPortfolioMolecularSpectroscopyPortfolio
MolecularSpectroscopyPortfolio
Keegan McHose
 
spectroscpy slides sidra
spectroscpy slides sidraspectroscpy slides sidra
spectroscpy slides sidra
Dr Sidra Khalid
 
Uv – Visible Spectrophotometer.HARIS
Uv – Visible Spectrophotometer.HARISUv – Visible Spectrophotometer.HARIS
Uv – Visible Spectrophotometer.HARIS
HARIS.P
 

Destacado (20)

Ppt spectroscopy
Ppt spectroscopyPpt spectroscopy
Ppt spectroscopy
 
Presentation1
Presentation1Presentation1
Presentation1
 
Uv visible spectrphotometery
Uv visible spectrphotometeryUv visible spectrphotometery
Uv visible spectrphotometery
 
spectroscopy sidra
spectroscopy sidraspectroscopy sidra
spectroscopy sidra
 
Absorbance and emission
Absorbance and emissionAbsorbance and emission
Absorbance and emission
 
MolecularSpectroscopyPortfolio
MolecularSpectroscopyPortfolioMolecularSpectroscopyPortfolio
MolecularSpectroscopyPortfolio
 
Basics of uv spectroscopy
Basics of uv spectroscopyBasics of uv spectroscopy
Basics of uv spectroscopy
 
Ultravoilet spectroscopy
Ultravoilet spectroscopyUltravoilet spectroscopy
Ultravoilet spectroscopy
 
spectroscpy slides sidra
spectroscpy slides sidraspectroscpy slides sidra
spectroscpy slides sidra
 
UV / VIS
UV / VIS UV / VIS
UV / VIS
 
Uv-visible spectroscopy
Uv-visible spectroscopyUv-visible spectroscopy
Uv-visible spectroscopy
 
CHM260 - UV-VIS
CHM260 - UV-VISCHM260 - UV-VIS
CHM260 - UV-VIS
 
Spectroscopy uv vis
Spectroscopy uv visSpectroscopy uv vis
Spectroscopy uv vis
 
UV VISIBLE REFLECTANCESPECTROSCOPY AND APPLICATION
UV VISIBLE REFLECTANCESPECTROSCOPY AND APPLICATION UV VISIBLE REFLECTANCESPECTROSCOPY AND APPLICATION
UV VISIBLE REFLECTANCESPECTROSCOPY AND APPLICATION
 
Wade12 ir ms
Wade12 ir msWade12 ir ms
Wade12 ir ms
 
Uv – Visible Spectrophotometer.HARIS
Uv – Visible Spectrophotometer.HARISUv – Visible Spectrophotometer.HARIS
Uv – Visible Spectrophotometer.HARIS
 
U.V Spectroscopy.
U.V Spectroscopy.U.V Spectroscopy.
U.V Spectroscopy.
 
UV visible Spectroscoy - How to used UV Spectrophotometer
UV visible Spectroscoy - How to used UV SpectrophotometerUV visible Spectroscoy - How to used UV Spectrophotometer
UV visible Spectroscoy - How to used UV Spectrophotometer
 
Colorimeter and spectrophotometer, Mass Spectrometer
Colorimeter and spectrophotometer, Mass SpectrometerColorimeter and spectrophotometer, Mass Spectrometer
Colorimeter and spectrophotometer, Mass Spectrometer
 
Chapter 01 - Principal Accounting (Warren Reeve Fess)
Chapter 01 - Principal Accounting (Warren Reeve Fess)Chapter 01 - Principal Accounting (Warren Reeve Fess)
Chapter 01 - Principal Accounting (Warren Reeve Fess)
 

Similar a Uv vis-instrument-1

Instrumentation in photonics
Instrumentation in photonicsInstrumentation in photonics
Instrumentation in photonics
karoline Enoch
 
uv visible spectrometery ppt
uv visible spectrometery pptuv visible spectrometery ppt
uv visible spectrometery ppt
rajesh kumar
 

Similar a Uv vis-instrument-1 (20)

Uv spectroscopy instrumentation, by dr. umesh kumar sharma & amp; shyma m s
Uv spectroscopy instrumentation, by dr. umesh kumar sharma & amp; shyma m sUv spectroscopy instrumentation, by dr. umesh kumar sharma & amp; shyma m s
Uv spectroscopy instrumentation, by dr. umesh kumar sharma & amp; shyma m s
 
FOURIER TRANSFORM - INFRARED SPECTROSCOPY
FOURIER TRANSFORM - INFRARED SPECTROSCOPYFOURIER TRANSFORM - INFRARED SPECTROSCOPY
FOURIER TRANSFORM - INFRARED SPECTROSCOPY
 
Instrumentation of uv visible spectrophotometer
Instrumentation of uv visible spectrophotometerInstrumentation of uv visible spectrophotometer
Instrumentation of uv visible spectrophotometer
 
Spectrophotometry
Spectrophotometry Spectrophotometry
Spectrophotometry
 
Analytical instrumentation
Analytical instrumentationAnalytical instrumentation
Analytical instrumentation
 
UV Instrumentation
UV Instrumentation UV Instrumentation
UV Instrumentation
 
Spectrophotometer.ppt (T.Y. Bsc).pptx
Spectrophotometer.ppt (T.Y. Bsc).pptxSpectrophotometer.ppt (T.Y. Bsc).pptx
Spectrophotometer.ppt (T.Y. Bsc).pptx
 
Photo transistors
Photo transistorsPhoto transistors
Photo transistors
 
UV SPECTROSCOPY AND INSTRUMENTATION
UV SPECTROSCOPY AND INSTRUMENTATION UV SPECTROSCOPY AND INSTRUMENTATION
UV SPECTROSCOPY AND INSTRUMENTATION
 
Ultra violet spectrometer
Ultra violet spectrometerUltra violet spectrometer
Ultra violet spectrometer
 
Spectroscopy
Spectroscopy   Spectroscopy
Spectroscopy
 
Instrumentation of uv visible spectroscopy
Instrumentation of uv visible spectroscopyInstrumentation of uv visible spectroscopy
Instrumentation of uv visible spectroscopy
 
Instrumentation in photonics
Instrumentation in photonicsInstrumentation in photonics
Instrumentation in photonics
 
Spectrophotometry.pptx
Spectrophotometry.pptxSpectrophotometry.pptx
Spectrophotometry.pptx
 
Instrumentation of uv spectrophotometer pranav kumar singh
Instrumentation of uv spectrophotometer   pranav kumar singhInstrumentation of uv spectrophotometer   pranav kumar singh
Instrumentation of uv spectrophotometer pranav kumar singh
 
UV-VIS SPECTROSCOPY
UV-VIS SPECTROSCOPYUV-VIS SPECTROSCOPY
UV-VIS SPECTROSCOPY
 
uv visible spectrometery ppt
uv visible spectrometery pptuv visible spectrometery ppt
uv visible spectrometery ppt
 
Uv,vis,nmr,mass,ir
Uv,vis,nmr,mass,irUv,vis,nmr,mass,ir
Uv,vis,nmr,mass,ir
 
Infrared spectroscopy
Infrared spectroscopyInfrared spectroscopy
Infrared spectroscopy
 
IR Spectroscopy
IR SpectroscopyIR Spectroscopy
IR Spectroscopy
 

Más de MOHIT CHAUDHARY

Más de MOHIT CHAUDHARY (17)

Strategies for effective oral presentations
Strategies for effective oral presentationsStrategies for effective oral presentations
Strategies for effective oral presentations
 
14.proposal writing
14.proposal writing14.proposal writing
14.proposal writing
 
4. non verbal communication
4. non verbal communication4. non verbal communication
4. non verbal communication
 
Report writing
Report writingReport writing
Report writing
 
Example tenders
Example tendersExample tenders
Example tenders
 
17.business letters
17.business letters17.business letters
17.business letters
 
8.ethical issues in communication
8.ethical issues in communication8.ethical issues in communication
8.ethical issues in communication
 
9.written communication
9.written communication9.written communication
9.written communication
 
10.the memorandum
10.the memorandum10.the memorandum
10.the memorandum
 
5.formal informal comm channels
5.formal informal comm channels5.formal informal comm channels
5.formal informal comm channels
 
Sample report 1
Sample report 1Sample report 1
Sample report 1
 
Interviews
InterviewsInterviews
Interviews
 
Type a and b personalities
Type a and b personalitiesType a and b personalities
Type a and b personalities
 
how miscommunication can lead to worst situation
how miscommunication can lead to worst situationhow miscommunication can lead to worst situation
how miscommunication can lead to worst situation
 
3.verbal and non verbal communication
3.verbal and non verbal communication3.verbal and non verbal communication
3.verbal and non verbal communication
 
7. effective listening
7. effective listening7. effective listening
7. effective listening
 
2.barriers in communication
2.barriers in communication2.barriers in communication
2.barriers in communication
 

Último

The basics of sentences session 3pptx.pptx
The basics of sentences session 3pptx.pptxThe basics of sentences session 3pptx.pptx
The basics of sentences session 3pptx.pptx
heathfieldcps1
 

Último (20)

This PowerPoint helps students to consider the concept of infinity.
This PowerPoint helps students to consider the concept of infinity.This PowerPoint helps students to consider the concept of infinity.
This PowerPoint helps students to consider the concept of infinity.
 
Food safety_Challenges food safety laboratories_.pdf
Food safety_Challenges food safety laboratories_.pdfFood safety_Challenges food safety laboratories_.pdf
Food safety_Challenges food safety laboratories_.pdf
 
FSB Advising Checklist - Orientation 2024
FSB Advising Checklist - Orientation 2024FSB Advising Checklist - Orientation 2024
FSB Advising Checklist - Orientation 2024
 
Micro-Scholarship, What it is, How can it help me.pdf
Micro-Scholarship, What it is, How can it help me.pdfMicro-Scholarship, What it is, How can it help me.pdf
Micro-Scholarship, What it is, How can it help me.pdf
 
The basics of sentences session 3pptx.pptx
The basics of sentences session 3pptx.pptxThe basics of sentences session 3pptx.pptx
The basics of sentences session 3pptx.pptx
 
HMCS Max Bernays Pre-Deployment Brief (May 2024).pptx
HMCS Max Bernays Pre-Deployment Brief (May 2024).pptxHMCS Max Bernays Pre-Deployment Brief (May 2024).pptx
HMCS Max Bernays Pre-Deployment Brief (May 2024).pptx
 
Application orientated numerical on hev.ppt
Application orientated numerical on hev.pptApplication orientated numerical on hev.ppt
Application orientated numerical on hev.ppt
 
Google Gemini An AI Revolution in Education.pptx
Google Gemini An AI Revolution in Education.pptxGoogle Gemini An AI Revolution in Education.pptx
Google Gemini An AI Revolution in Education.pptx
 
General Principles of Intellectual Property: Concepts of Intellectual Proper...
General Principles of Intellectual Property: Concepts of Intellectual  Proper...General Principles of Intellectual Property: Concepts of Intellectual  Proper...
General Principles of Intellectual Property: Concepts of Intellectual Proper...
 
Towards a code of practice for AI in AT.pptx
Towards a code of practice for AI in AT.pptxTowards a code of practice for AI in AT.pptx
Towards a code of practice for AI in AT.pptx
 
Sociology 101 Demonstration of Learning Exhibit
Sociology 101 Demonstration of Learning ExhibitSociology 101 Demonstration of Learning Exhibit
Sociology 101 Demonstration of Learning Exhibit
 
Single or Multiple melodic lines structure
Single or Multiple melodic lines structureSingle or Multiple melodic lines structure
Single or Multiple melodic lines structure
 
80 ĐỀ THI THỬ TUYỂN SINH TIẾNG ANH VÀO 10 SỞ GD – ĐT THÀNH PHỐ HỒ CHÍ MINH NĂ...
80 ĐỀ THI THỬ TUYỂN SINH TIẾNG ANH VÀO 10 SỞ GD – ĐT THÀNH PHỐ HỒ CHÍ MINH NĂ...80 ĐỀ THI THỬ TUYỂN SINH TIẾNG ANH VÀO 10 SỞ GD – ĐT THÀNH PHỐ HỒ CHÍ MINH NĂ...
80 ĐỀ THI THỬ TUYỂN SINH TIẾNG ANH VÀO 10 SỞ GD – ĐT THÀNH PHỐ HỒ CHÍ MINH NĂ...
 
2024-NATIONAL-LEARNING-CAMP-AND-OTHER.pptx
2024-NATIONAL-LEARNING-CAMP-AND-OTHER.pptx2024-NATIONAL-LEARNING-CAMP-AND-OTHER.pptx
2024-NATIONAL-LEARNING-CAMP-AND-OTHER.pptx
 
UGC NET Paper 1 Mathematical Reasoning & Aptitude.pdf
UGC NET Paper 1 Mathematical Reasoning & Aptitude.pdfUGC NET Paper 1 Mathematical Reasoning & Aptitude.pdf
UGC NET Paper 1 Mathematical Reasoning & Aptitude.pdf
 
Mehran University Newsletter Vol-X, Issue-I, 2024
Mehran University Newsletter Vol-X, Issue-I, 2024Mehran University Newsletter Vol-X, Issue-I, 2024
Mehran University Newsletter Vol-X, Issue-I, 2024
 
On_Translating_a_Tamil_Poem_by_A_K_Ramanujan.pptx
On_Translating_a_Tamil_Poem_by_A_K_Ramanujan.pptxOn_Translating_a_Tamil_Poem_by_A_K_Ramanujan.pptx
On_Translating_a_Tamil_Poem_by_A_K_Ramanujan.pptx
 
SOC 101 Demonstration of Learning Presentation
SOC 101 Demonstration of Learning PresentationSOC 101 Demonstration of Learning Presentation
SOC 101 Demonstration of Learning Presentation
 
Unit 3 Emotional Intelligence and Spiritual Intelligence.pdf
Unit 3 Emotional Intelligence and Spiritual Intelligence.pdfUnit 3 Emotional Intelligence and Spiritual Intelligence.pdf
Unit 3 Emotional Intelligence and Spiritual Intelligence.pdf
 
ICT role in 21st century education and it's challenges.
ICT role in 21st century education and it's challenges.ICT role in 21st century education and it's challenges.
ICT role in 21st century education and it's challenges.
 

Uv vis-instrument-1

  • 1. Instruments for Optical Spectrometry • Optical spectroscopy encompasses regions of the electromagnetic spectrum that are not strictly within the visible range. However, the components of instruments used in the IR, visible, and UV are all very similar.
  • 2. Instrument Components • Spectroscopic instruments are usually constructed from the following components: • A stable source of radiant energy • A wavelength selector that permits the selection and isolation of a restricted wavelength region • One or more sample containers • A radiation detector or transducer that converts radiant energy to a measurable signal – normally electrical • A signal processor and readout
  • 3. Instrument Components • In emission instruments the source and sample are combined so that emission occurs from heating or electrical excitation of the sample. • In absorption and fluorescence instruments the analyte is held in a sample container and an external radiant source is employed. • In absorption instruments the source, sample, and detector are located along an axis. In fluorescence instruments the source is placed at an angle of 90° to the detector-sample axis.
  • 4. Optical Materials • The optical components of a spectrometer must be optically transparent in the wavelength region being employed for analysis. • In IR studies, sodium chloride, potassium chloride, or silver chloride are used. • In the visible region, borosilicate glass or plastics such as polystyrene are adequate. • In the UV, quartz is used because glass absorbs too much radiation at these wavelengths.
  • 5. Spectroscopic Sources • The source must provide a beam of sufficient intensity in the wavelength region of interest to allow detection and measurement. It must be stable and not flicker so P and Po can be measured accurately. Some instruments measure these values simultaneously so stability is less of a factor. • Both line sources and continuous sources are used in optical spectroscopy. The continuous source is made up of a whole range of emitting wavelengths that overlap while line sources emit discrete bands at one or two wavelengths.
  • 6. Tungsten Filament Lamps • Theses emit in the range 320 – 2500 nm so most of the emissions are in the IR. • The lamp is useful for visible spectroscopy but not for UV because the lamp housing absorbs strongly at these wavelengths. • The lamp should have a stable voltage supply so that its spectral characteristics remain constant – it tends to vary as its temperature changes.
  • 7. Tungsten Halogen Lamps • Tungsten halogen lamps are an improvement over tungsten filament lamps because they have a longer lifetime and their emission spectra extend into the UV. • The lamp contains a small quantity of iodine within a quartz envelope that houses the filament. • The lifetime of the tungsten halogen lamp is more than double that of the tungsten filament lamp. • In the tungsten filament lamp, the filament eventually sublimes.
  • 8. Tungsten Halogen Lamps • In a tungsten iodine lamp the tungsten sublimes but reacts with the iodine to give WI2, which diffuses back to the filament where it decomposes and re-deposits tungsten.
  • 9. Hydrogen and Deuterium Lamps • These lamps generate true continua. The process for hydrogen may be shown as: • H2 + E → H2* → H' + H'' + hν where E is the energy required to excite the hydrogen molecule. • The products, H' and H'' may have a variety of kinetic energies, which means that the photon emitted, hν, will also vary in energy – thus the continuum.
  • 10. Hydrogen and Deuterium Lamps • The useful range of these lamps is from 160 – 375 nm. Deuterium is preferred to hydrogen because of its greater intensity. • The emission is stimulated by generating an electrical arc in an envelope filled with the gas. A regulated power supply is necessary for constant lamp intensity.
  • 11. Wavelength Selectors • Spectroscopic instruments need a wavelength selector to provide a range of wavelengths to impinge on the sample. The narrower the range of wavelengths chosen by the wavelength selector, the more likely Beer’s law is to be followed. This also leads to improved sensitivity and selectivity. • The closest type of radiation to a purely monochromatic source is a laser. • With normal optical instruments the wavelength selector provides a range of wavelengths centred at a nominal wavelength.
  • 12. Wavelength Selectors • The range of wavelengths provided by the wavelength selector is known as the effective bandwidth or just bandwidth and is defined as the width of the band in wavelength units at half peak height. The ordinate in this plot is transmittance so the bandwidth is the range for 0.5T.
  • 13. Wavelength Selectors • Bandwidths vary enormously from one wavelength selector to another. For instance, a high quality monochromator for the visible region may have a bandwidth of a few tenths of a nanometre or less. An absorption filter operating in the same region may have a bandwidth of 200 nm or more.
  • 14. Wavelength Selectors • Generally two types of wavelength selector are used, filters or monochromators. • Filters are rugged, inexpensive, and easy to maintain. • Monochromators are costly but allow nominal wavelengths to be changed continuously during irradiation of the sample.
  • 15. Radiation Filters • A filter allows only a restricted band of wavelengths to pass through it. • Two types of filter are used in spectroscopic instrumentation, interference filters and absorption filters. • Interference filters allow a greater transmittance of radiation at the nominal wavelength than absorption filters.
  • 16. Monochromators • Two types of monochromators have been used to disperse multiwavelength radiation into its various components. These are prisms and diffraction gratings. • Prisms are used infrequently nowadays because gratings can be made more cheaply. Gratings scratched on the surface of a mirror lead to more compact instrumentation.
  • 17. Monochromators • In a typical reflective grating monochromator, radiation from a source enters the monochromator by a narrow opening called a slit. The radiation is then collimated by a concave mirror so that a parallel beam strikes the grating. Angular dispersion from the grating then occurs at the reflective surface.
  • 18. Monochromators • Consider two wavelengths of radiation, λ1 and λ2, being dispersed by this grating. If λ1 < λ2 radiation of wavelength λ1will be reflected from the grating at a sharper angle than that of wavelength λ2. This is known as angular dispersion. • After dispersion the two beams of radiation are focused by a concave mirror on to the focal plane of the instrument where they appear as two images. By rotating the grating either one of the images can be focused on the exit slit.
  • 19. Gratings • Most gratings used in commercial instruments are replica gratings obtained by making castings from a master grating. The master grating is prepared by scoring many parallel and closely spaced grooves on a flat polished surface with a sharp diamond tool. In UV- visible work the grating will have from 300-2000 grooves/mm. • Preparation of the master grating is tedious and expensive. Replica gratings are prepared from a liquid resin casting process that preserves the optical accuracy of the original master grating. The surface of the replica is made reflective with a coating of aluminum, gold, or platinum.
  • 20. Radiation Detectors and Transducers • A detector is a device that indicates the existence of some physical phenomenon. Examples are the human eye, photographic film, the mercury level in a thermometer, etc. • A transducer is a special type of detector that converts a physical phenomenon, e.g. light intensity, pH, mass, etc., into electrical signals that can be amplified, manipulated, and eventually turned into numbers that are related to the magnitude of the original signal.
  • 21. Properties of Transducers • The ideal electromagnetic radiation transducer responds rapidly to low levels of radiant energy over a broad wavelength range. • The electrical signal it produces must be easily amplified and have a relatively low noise level.
  • 22. • It is essential that the electrical signal produced by the transducer is directly proportional to the beam power, P. • G = KP + K' • G is the electrical response of the detector in units of current, resistance, or potential • K is a proportionality constant which measures the sensitivity of the detector in terms of electrical response per unit of radiant power • K' represents the dark current. This is a small constant response, present even when no radiation strikes the detector. When K' is reduced to a negligible quantity by electrical subtraction etc., G = KP Properties of Transducers
  • 23. Types of Transducers • Two general types of transducers are used in optical spectroscopy. UV-visible transducers respond to light or photons
  • 24. Photon Detectors • Types of photon detectors that are used commercially are: phototubes, photomultiplier tubes, silicon photodiodes. • Phototubes are one of the simpler types to understand because they work on the Einstein photo-emissive cathode principle.
  • 25.
  • 26. Photomultiplier Tubes • This is similar in design to the phototube but is much more sensitive. • The surface of the cathode is similar in composition to that of the phototube. Electrons are emitted from its surface and accelerated towards the surface of another electrode called a dynode. Upon striking the dynode, each electron produces several additional electrons which are then accelerated towards another dynode. This process continues at a number of dynodes until 106 – 107 electrons have been produced for each photon. • The cascade is collected at the anode where the resulting electric current is further amplified and measured.
  • 27.
  • 28. Silicon Photodiodes • These are transistor devices used in diode array detectors. In these devices, over 1000 can be fabricated side by side on a small silicon chip – each diode is about 0.02 mm thick. • When one or more diode array detectors are placed along the focal plane of a monochromator, the intensity of all wavelengths can be measured simultaneously.
  • 29.
  • 30. Signal Processors and Readouts • The signal processor is an electronic device that amplifies the electrical signal from the detector. It may also subtract unwanted signals. • Signal processors may display the signal in a number of ways including derivative and integrated peak areas, etc. • Digital readout devices are most frequently found in modern instruments. However, chart recorders are used a great deal still.
  • 31. Optical Instrument Design • Optical spectroscopic instruments vary from the very sophisticated which are expensive and used for research, to the simple inexpensive variety which are used for routine chemical analysis. It is the nature of the components that determines the cost and eventual use.
  • 32. Common Names of Some Optical Instruments • A spectroscope is an instrument for visually identifying species excited in a flame or by other means. The instrument has a monochromator and an eyepiece in the focal plane to observe emission lines. It was used in early spectroscopic studies. • A colorimeter is an instrument for absorption measurements in which the eye serves as the detector. Comparison standards are used to quantify the amount of analyze. This is a limited use of the term colorimeter which is now usually applied to any photometer used for absorption measurements in the visible region of the spectrum.
  • 33. Common Names of Some Optical Instruments • A photometer is a photoelectric instrument that can be used for absorption, emission, or fluorescence measurements. Most colorimeters are photometers because they use filters for wavelength selection.
  • 34. Common Names of Some Optical Instruments • A spectrograph records spectra on a photographic film or plate, placed along the focal plane of the monochromator. The spectra appear as a set of black images of the entrance slit. Spectrographs are usually used for qualitative analysis. • A spectrometer is a monochromator equipped with a fixed slit at the focal plane. When the detection device is a phototransducer the instrument is known as a spectrophotometer.
  • 35. Instrument Types • Three general types of spectroscopic instruments are in general use. • These are single beam, double beam in space, and double beam in time. • All three types are used in absorption, emission, and fluorescence studies.
  • 36. Single Beam Instruments • In a single beam instrument used in absorption studies, a filter or monochromator is used to select the wavelength of radiation incident on the sample holder. • Matched cells can be interposed alternately in the radiation beam which passes on to a detector, usually a photomultiplier tube.
  • 37. Single Beam Instruments • To operate a manual instrument the following adjustments are made: • (a) The percent transmittance is corrected to zero by placing a shutter between the source and the detector and adjusting the meter until it reads zero. • (b) solvent is placed in the radiation path, and the shutter opened. The percent transmittance reading of the meter is adjusted to 100 percent.
  • 38. Single Beam Instruments • A logarithmic scale can be used to determine absorbance; a linear scale is used to determine transmittance. • Single beam instruments vary in complexity.
  • 40. Double Beam Instruments • In a double beam instrument, two beams of radiation are formed from the same source by a V- shaped mirror called a ‘beam splitter’. • One beam passes through the reference solution to a detector while the second simultaneously passes through a solution containing analyte to a second matched detector.
  • 41. Double Beam Instruments • The two outputs are amplified and their ratio (or log ratio) is determined electronically and displayed by the readout device.
  • 43. Double Beam Instruments • A second type of double beam instrument utilizes a rotating sectored mirror to direct the radiation from the source, as a pulse, either through a reference cell or through the sample. • The pulses of radiation are recombined by another sectored mirror before detection. • Pulsed radiation is the basis of many modern research instruments.
  • 44. Double Beam Instruments • Double beam instruments compensate for anything but very short term fluctuations in radiant output from the source, as well as for any drift in the detector and amplifier. • They also compensate for any wide variation in source intensity with wavelength. • This design lends itself particularly well to the continuous recording of spectra. Most modern UV-visible recording instruments are based on this design.
  • 46. Molecular Absorption Spectroscopy • Molecular optical spectroscopy is useful for both qualitative and quantitative measurements. • UV-visible spectrophotometry is mainly used for quantitative work.
  • 47. Ultraviolet and Visible Absorption Spectroscopy • Molecules absorb a finite amount of energy to place the molecule in an upper electronic state. Various vibrational and rotational levels are occupied in this excited electronic state. The absorption spectrum should reflect this. • In the gas phase, it is possible to observe absorption of energy to different vibrational and rotational energy levels. This is a resolved spectrum with fairly narrow lines.
  • 48. Ultraviolet and Visible Absorption Spectroscopy • In solution or in condensed phases the spectrum loses its resolution and becomes a broad band. • This is because the molecules of absorbing species can bump into one another more easily, thus creating a variety of energies that cannot be resolved.
  • 49. Absorption by Organic Compounds • The wavelength at which organic molecules absorb UV-visible radiation depends to a large extent on how firmly bound the electrons are in the molecule. • Electrons in single covalent bonds are tightly held between the atoms in the bond. Excitation of these electrons is only possible by using wavelengths of radiation below 180 nm. • This is of no use for routine measurements because quartz and various atmospheric components absorb at these wavelengths.
  • 50. Absorption by Organic Compounds • Electrons in double or triple bonds are more easily excited because they are more loosely held than single bond electrons. Unsaturated molecules have useful absorption spectra. • The unsaturated organic functional groups that absorb UV-visible radiation are known as chromophores. They can serve as a very rough guide for identification purposes.
  • 51. Absorption by Organic Compounds • Peak position is influenced by a number of factors such as: • Solvent effects • Other structural details of the molecule, such as conjugation • Molecules containing heteroatoms with lone pairs of nonbonding electrons may also absorb in the UV-visible region of the spectrum and provide qualitative information on the molecule.
  • 52. Absorbing species • Electronic transitions π, σ, and n electrons – d and f electrons – Charge transfer reactions ∀π, σ, and n (non-bonding) electrons
  • 53. Sigma and Pi orbitals
  • 55. Transitions ∀ σ−>σ∗ – UV photon required, high energy • Methane at 125 nm • Ethane at 135 nm • n-> σ∗ – Saturated compounds with unshared e- • Absorption between 150 nm to 250 nm ∀ε between 100 and 3000 L cm-1 mol-1 • Shifts to shorter wavelengths with polar solvents – Minimum accessibility – Halogens, N, O, S
  • 56. Transitions • n->π∗, π−>π∗ – Organic compounds, wavelengths 200 to 700 nm – Requires unsaturated groups • n->π∗ low ε (10 to 100) – Shorter wavelengths ∀π−>π∗ higher ε (1000 to 10000)
  • 57. Absorption by Inorganic Species • Typically visible absorption of visible radiation occurs with compounds of elements in the first and second row of the transition series. This is why these compounds are colored. • The absorption involves transitions between filled and unfilled d-orbitals with energies that depend on the ligands attached to the elements. • The position of the energy transition depends on the position of the element in the periodic table, its oxidation state, and the nature of the attached ligand.
  • 58. Absorption by Inorganic Species • The lanthanide elements behave somewhat differently in that the peaks are fairly narrow. • This is because transitions are between electrons in f-orbitals. They are shielded from effects of the outer electrons and the energy levels are thus not distorted.
  • 59. Qualitative Applications of UV- visible Spectroscopy • UV-visible qualitative analysis is useful for determining the presence of certain gross structural features of molecules, e.g. unsaturation. • However it needs to be supplemented by other spectroscopic methods such as IR and NMR to obtain detailed information on molecular structure.
  • 60. Solvents • The most useful information is derived from spectra obtained from dilute solutions or from a vapour of the analyte. • Solvents should be of high purity and be transparent throughout the spectral range. • Generally non-polar solvents distort spectra least. Polar solvents tend to obscure structure. • Solvents tend to change the position of maximum absorbance so in any qualitative measurement of wavelength, the same solvent must be used for all samples.
  • 61. Quantitative Applications • This is one of the most important tools available for quantitative analysis. • Important characteristics of the method are as follows: • (1) Wide applicability. Most compounds will either absorb themselves or can be derivatized so that the derivatives absorb. UV-visible spectrophotometry is the single most important analytical technique for quantitative chemical analysis.
  • 62. Quantitative Applications • (2) High sensitivity. Typically measurements can be made with detection limits as low as 10−5 mol/L. With procedural modifications these detection limits can be extended even further. • (3) Moderate to high selectivity. This means that it is sometimes, but not too often, possible to make measurements on the analyte without prior separation from other absorbing and interfering compounds.
  • 63. Quantitative Applications • (4) Excellent accuracy. Normally the precision of any given measurement will be within 1–5%. Errors are reduced to similar levels and can be made even lower by using special measuring techniques. • (5) Ease and convenience. Spectrometers are not cheap, but are easy to use and lend themselves to automation.