Safe use of electrosurgery in
gynaecology
Dr Muhamed Al Bellehy MD
Consultant Obstetrics and gynecology
Dr Muhamed Al Bellehy MD
 the two types of electric current:
o direct current (DC), which moves in one direction
(unidirectional),
o AC, which periodically reverses direction
(bidirectional).
Electrosurgery electrocautery
the application of high‐frequency
alternating current (AC) in surgery to
achieve various thermal tissue effects
including cutting, desiccation and
coagulation.
the passive transfer of
heat to the tissue with no
current passing through
it.
 uses AC with high frequency
o AC to avoid electrolysis
o high frequency to avoid the
Faradic effect of nerve and
muscle stimulation.
 uses DC
Faradic effect ceases at frequencies >
100 kHz.
Because frequencies commonly used in
are >500 kHz, which are similar to
radiofrequency (RF), the term RF
electrosurgery is used.
Dr Muhamed Al Bellehy MD 2
Dr Muhamed Al Bellehy MD 3
Dr Muhamed Al Bellehy MD 4
• Surgical diathermy = electrosurgery
Dr Muhamed Al Bellehy MD 5
Dr Muhamed Al Bellehy MD 6
Dr Muhamed Al Bellehy MD 7
Power (Battery) produces difference in
voltage  current
Resistance will decrease the current
Dr Muhamed Al Bellehy MD 8
Battery produces difference in voltage
 current
Resistance
Change
electricity
into
another
form like
heat or
light
Dr Muhamed Al Bellehy MD 9
Increasing the voltage will result in a
higher current
Increasing the resistance will result in a
lower current
Dr Muhamed Al Bellehy MD 10
Dr Muhamed Al Bellehy MD 11
Power is expressed as Whatt = amount
of energy delivered in time unit
Dr Muhamed Al Bellehy MD 12
Electrosurgical unit = voltage
Electrical
switch
Tissue effect= increase
temperature
Dr Muhamed Al Bellehy MD 13
Power
increase temperature
Depends on
Resistance
Duration
Dr Muhamed Al Bellehy MD 14
Maximum amount of
power
In Voltage
Dr Muhamed Al Bellehy MD 15
amount of power for certain effect
Depends on
Resistance
Type of
electrode
Dr Muhamed Al Bellehy MD 16
Voltage vs power controlled generators
Dr Muhamed Al Bellehy MD 17
• Electro surgery uses alternating current not direct current
• direct current will result in electrolysis
Dr Muhamed Al Bellehy MD 18
Electro surgery uses higher frequency to avoid tissue stimulation
Dr Muhamed Al Bellehy MD 19
Dr Muhamed Al Bellehy MD 20
Current flows 100% of
time + > 200 V
Tissue Vaporization Cutting
Current flows 6% of
time
Tissue Cools in between
 form a coagulum
Coagulation +
Some Cutting when > 200 V
Dr Muhamed Al Bellehy MD 21
Cutting
Coagulation +
Some Cutting when > 200 V
Dr Muhamed Al Bellehy MD 22
Coagulation +
Some Cutting when > 200 V
BLEND mode
Activate Current and higher Voltage in a certain way to obtain effect mixed
of cutting and coagulation
Dr Muhamed Al Bellehy MD 23
Dr Muhamed Al Bellehy MD 24
Dr Muhamed Al Bellehy MD 25
Dr Muhamed Al Bellehy MD 26
Monopolar
Dr Muhamed Al Bellehy MD 27
Dr Muhamed Al Bellehy MD 28
Bipolar
• Only used for tissues
that could be grasped
• Effect between two
the blades only
• So- relatively safe
Dr Muhamed Al Bellehy MD 29
Gap between two
blades
No Gap between two
blades
Dr Muhamed Al Bellehy MD 30
Jaw of bipolar forceps
Dr Muhamed Al Bellehy MD 31
Dispersive blade
Active blade
Dr Muhamed Al Bellehy MD 32
Contact area at
Dispersive blade 10000
times > active electrode
Current density at
Dispersive blade 10000
times < active electrode
Temp at Active
electrode 10000 times >
Dispersive blade
Dr Muhamed Al Bellehy MD 33
Dr Muhamed Al Bellehy MD 34
Dr Muhamed Al Bellehy MD 35
Dr Muhamed Al Bellehy MD 36
Increasing the Contact
area of the active
electrode decreases
current density  more
coagulation
decreasing the Contact
area of the active
electrode increases
current density  more
cutting
Dr Muhamed Al Bellehy MD 37
Dr Muhamed Al Bellehy MD 38
AKA electrosection
• Active electrode is held just
lightly against the tissue
• Using the cutting or blend
modes
• When using the blend mode
cutting + hemostasis of small
BVs
Excessive power  sparking and tissue drag
Dr Muhamed Al Bellehy MD 39
AKA Spray
AKA non contact coagulation
• Use spark or fulguration mode
• Active electrode is held a small
distance from the tissue
• Spark of, electricity will pass to
the tissues  minimal tissue
heating + large superficial
coagulum
Dr Muhamed Al Bellehy MD 40
• Use spark or fulguration mode
• Active electrode is held a small
distance from the tissue
• Spark of, electricity will pass to
the tissues  minimal tissue
heating + large superficial
coagulum
Dr Muhamed Al Bellehy MD 41
Dr Muhamed Al Bellehy MD 42
Dr Muhamed Al Bellehy MD 43
• Coagulation mode *bi – mono
polar*
• Active electrode is held in
direct contact with the tissue
• This result in less heat gradualy
delivered and Spark of,
electricity will pass to the
tissues  deep coagulum
Dr Muhamed Al Bellehy MD 44
• Use spark or fulguration mode
• Active electrode is held a small
distance from the tissue
• Spark of, electricity will pass to
the tissues  minimal tissue
heating + large superficial
coagulum
Dr Muhamed Al Bellehy MD 45
• Best with bi polar*
Dr Muhamed Al Bellehy MD 46
Dr Muhamed Al Bellehy MD 47
Cutting mode can coagulate in
power controlled generators
Coagulation mode can cut when >
200 Volte
Dr Muhamed Al Bellehy MD 48
Dr Muhamed Al Bellehy MD 49
Dr Muhamed Al Bellehy MD 50
Dr Muhamed Al Bellehy MD 51
Dr Muhamed Al Bellehy MD 52
General considerations for
• The tip of the electrode
• The body
• The dispersive blade
Dr Muhamed Al Bellehy MD 53
Dr Muhamed Al Bellehy MD 54
Surgical Smoke
• May obscure visualization
• May have health hazards –
contain toxic gases
Dr Muhamed Al Bellehy MD 55
Smoke evacuator may be used
Dr Muhamed Al Bellehy MD 56
Bowel perforation and injury
to perianal region will result in
leakage of methane and
hydrogen  ?? fire
Alcohol disinfectants
not allowed to dry
?? fire
Dr Muhamed Al Bellehy MD 57
Hand held electrodes should
be kept in non conductive
place to avoid danger of
accidental activation
Dr Muhamed Al Bellehy MD 58
Activating the active electrode
when the not in contact with
patient’s body electrical
glove burns
Dr Muhamed Al Bellehy MD 59
Activating the active electrode when the
surgeon is in contact with active electrode and
the patient  electrical pathway through the
surgeon’s hand  electrical glove burns
Dr Muhamed Al Bellehy MD 60
Prolonged Activation of the active electrode 
increased residual heat  tissue injury
Short Activation of the active electrode should
be practiced
Dr Muhamed Al Bellehy MD 61
Prolonged Activation
of the active
electrode 
increased heat 
carbonization + tissue
sticks to the electrode
As the active
electrode keeps
contact with the
tissue 
carbonization of
the sticking tissue
continue further
increase resistance
 decrease
clinical effect
 staff
wrongly
increase the
power
settings
the tip of the electrode should be kept clean !
Dr Muhamed Al Bellehy MD 62
Dr Muhamed Al Bellehy MD 63
Dr Muhamed Al Bellehy MD 64
Grounded Generators
1. The ground is part of the circuit
between the dispersive blade and
the generator
2. Current pass through the
dispersive blade to the ground
Dr Muhamed Al Bellehy MD 65
Dr Muhamed Al Bellehy MD 66
Alternative Grounded objects
Dr Muhamed Al Bellehy MD 67
Alternative
Grounded
objects
Operating Table and
accessories
IV stand
Older ECG electrodes
Dr Muhamed Al Bellehy MD 68
When the contact area
with the grounded object
is small and the
resistance is less than
that at the dispersive
blade
high current density at
the contact area 
alternate site burn
Dr Muhamed Al Bellehy MD 69
Isolated
generators
The ground is not part of the circuit
between the dispersive blade and
the generator
The Current pass through the body and the
current returning through the dispersive
blade can be measured – and must be
almost equal
Dr Muhamed Al Bellehy MD 70
Isolated
generators
When the difference between Current the
passes through the body and the current
returning through the dispersive blade– is
too large  system is deactivated
Dr Muhamed Al Bellehy MD 71
Isolated
generators
When the difference between Current the
passes through the body and the current
returning through the dispersive blade– is
too small  system is not deactivated
When the difference– is too small  not
necessarily indicates a leak
Thus there is a
small possibility
of alternate site
burn
Dr Muhamed Al Bellehy MD 72
Allow for a small leak Thus there is a small possibility
of alternate site burn
Dr Muhamed Al Bellehy MD 73
A visual check before the procedure, to
confirm that the patient is not contact with
a conducting object
Dr Muhamed Al Bellehy MD 74
Electricity may pass through a conducting
implantable  burns
Dr Muhamed Al Bellehy MD 75
The dispersive blade should not be placed
near or over a conducting implantable
Piercings and jewelry near the surgical site
should be removed
Dr Muhamed Al Bellehy MD 76
Short bursts for 6 seconds
Use bipolar
Avoid monopolar over
surrounding tissues
Dr Muhamed Al Bellehy MD 77
Dr Muhamed Al Bellehy MD 78
Dr Muhamed Al Bellehy MD 79
Contact area at
Dispersive blade 10000
times > active electrode
Current density at
Dispersive blade 10000
times < active electrode
Temp at Active
electrode 10000 times >
Dispersive blade
Dr Muhamed Al Bellehy MD 80
When Contact area at
Dispersive blade is
reduced
Current density at
Dispersive blade is
increased
When Contact area at
Dispersive blade is
severely reduced 
circuit blade burn
Dr Muhamed Al Bellehy MD 81
When choosing the site of dispersive blade consider
• Surgical site
• Contact with the table
• Accessories
• implants
Close to surgical site
Dr Muhamed Al Bellehy MD 82
When choosing the
site of dispersive blade
consider
• Choose a well
vascularized
muscular area
Dr Muhamed Al Bellehy MD 83
Dr Muhamed Al Bellehy MD 84
Poor placement of the
Dispersive blade 
contact area is reduced
Decrease tissue effect
Staff increase power
output
Increase chance of
burns
Dr Muhamed Al Bellehy MD 85
Use the largest suitable
blade
Dr Muhamed Al Bellehy MD 86
Dr Muhamed Al Bellehy MD 87
Return electrode
monitoring (REM)
Uses a small amount of
current to measure the
conduction between the
two blades of the split
blade
If high resistance  small
contract area or poor
tissue conduction
Dr Muhamed Al Bellehy MD 88
Return electrode
monitoring (REM)
Uses a small amount of
current to measure the
conduction between the
two blades of the split
blade
If high resistance  small
contract area or poor
tissue conduction
THEN DEACTIVE THE
SYSTEM
Dr Muhamed Al Bellehy MD 89
THEN DEACTIVE THE
SYSTEM
Dr Muhamed Al Bellehy MD 90
2 BLADES SHOULD BE AT SIMIALR
DISTANCE FROM THE SURGICAL SITE
X
Dr Muhamed Al Bellehy MD 91
Dr Muhamed Al Bellehy MD 92
As the current flows in the cords 
electromagnetic field
Dr Muhamed Al Bellehy MD 93
When the cords are bundled
stronger electromagnetic field
Dr Muhamed Al Bellehy MD 94
When the cords are wrapped around an instrument 
electromagnetic field will charge it with electric current
Dr Muhamed Al Bellehy MD 95
Dr Muhamed Al Bellehy MD 96
When the cords are wrapped around an instrument 
electromagnetic field will charge it with electric current
 patient injury if in contact
Dr Muhamed Al Bellehy MD 97
Do not use near any electronic devices or connections
Dr Muhamed Al Bellehy MD 98
Dr Muhamed Al Bellehy MD 99
Dr Muhamed Al Bellehy MD 100
Dr Muhamed Al Bellehy MD 101
Dr Muhamed Al Bellehy MD 102
Dr Muhamed Al Bellehy MD 103
Jaw of bipolar forceps
Dr Muhamed Al Bellehy MD 104
Dr Muhamed Al Bellehy MD 105
View during laparoscopy is limited to a conical area in
front of the endoscope
Dr Muhamed Al Bellehy MD 106
Contact of the instruments with tissues outside filed of
vision  injury
• Bleeding
• Perforation
• Leakage of body fluids
• Ignition of bowel gases
X
• Do not activate the current when instrument is not
visualized or near any organ or other instruments
• Avoid contact of the tip shortly after activation
• Avoid prolonged activation of current to reduce
residual instrument heat
Dr Muhamed Al Bellehy MD 107
Dr Muhamed Al Bellehy MD 108
Dr Muhamed Al Bellehy MD 109
Dr Muhamed Al Bellehy MD 110
Dr Muhamed Al Bellehy MD 111
high voltage causes spark  Direct coupling with out contact

Dr Muhamed Al Bellehy MD 112
To avoid this
• As low voltage as possible
• Cutting > coagulation
• Activate the current only when in direct conatct with the
tissues
Dr Muhamed Al Bellehy MD 113
Dr Muhamed Al Bellehy MD 114
Dr Muhamed Al Bellehy MD 115
Dr Muhamed Al Bellehy MD 116
Dr Muhamed Al Bellehy MD 117
Dr Muhamed Al Bellehy MD 118
Dr Muhamed Al Bellehy MD 119
When monopolar current is
used at high voltage 
creates high
electromagnetic field 
charge nearby instruments
with current
Dr Muhamed Al Bellehy MD 120
This does not occur with
bipolar instruments, as the
electromagnetic field has
opposing directions
Dr Muhamed Al Bellehy MD 121
Metal trocars disperses the
charge through the skin
Dr Muhamed Al Bellehy MD 122
Hybrid trocar systems
induces capacitive coupling
Dr Muhamed Al Bellehy MD 123
Dr Muhamed Al Bellehy MD 124
Dr Muhamed Al Bellehy MD 125
Dr Muhamed Al Bellehy MD 126
Dr Muhamed Al Bellehy MD 127
Dr Muhamed Al Bellehy MD 128
Bipolar electro surgery is
preferred
Dr Muhamed Al Bellehy MD 129
Dr Muhamed Al Bellehy MD 130
Dr Muhamed Al Bellehy MD 131
Dr Muhamed Al Bellehy MD 132

Más contenido relacionado

La actualidad más candente

Non descent vaginal hysterectomy
Non descent vaginal hysterectomyNon descent vaginal hysterectomy
Non descent vaginal hysterectomyRajni Singh
 
Laparoscopic myomectomy
Laparoscopic myomectomyLaparoscopic myomectomy
Laparoscopic myomectomymagdy abdel
 
Role of tubal surgery in era of ivf
Role of tubal surgery in era of ivfRole of tubal surgery in era of ivf
Role of tubal surgery in era of ivfSanjay Makwana
 
ectopic pregnancy Ultrasound Daignosis.pdf
ectopic pregnancy Ultrasound Daignosis.pdfectopic pregnancy Ultrasound Daignosis.pdf
ectopic pregnancy Ultrasound Daignosis.pdfMuhamedAlBellehy1
 
Applications of ir in obstetrics and gynecology2
Applications of ir in obstetrics and gynecology2Applications of ir in obstetrics and gynecology2
Applications of ir in obstetrics and gynecology2pryce27
 
Manchester repair (Fothergill’s Operation)
Manchester repair (Fothergill’s Operation)Manchester repair (Fothergill’s Operation)
Manchester repair (Fothergill’s Operation)Yapa
 
Recurrent pregnancy loss panel discussion
Recurrent pregnancy loss  panel discussionRecurrent pregnancy loss  panel discussion
Recurrent pregnancy loss panel discussionNiranjan Chavan
 
Laproscopy & hysteroscopy in gynecology no video
Laproscopy & hysteroscopy in gynecology  no videoLaproscopy & hysteroscopy in gynecology  no video
Laproscopy & hysteroscopy in gynecology no videoVivek Kakkad
 

La actualidad más candente (20)

Recurrent endometriosis
Recurrent endometriosisRecurrent endometriosis
Recurrent endometriosis
 
Non descent vaginal hysterectomy
Non descent vaginal hysterectomyNon descent vaginal hysterectomy
Non descent vaginal hysterectomy
 
Laparoscopic myomectomy
Laparoscopic myomectomyLaparoscopic myomectomy
Laparoscopic myomectomy
 
Purandares cervicopexy
Purandares cervicopexyPurandares cervicopexy
Purandares cervicopexy
 
Role of tubal surgery in era of ivf
Role of tubal surgery in era of ivfRole of tubal surgery in era of ivf
Role of tubal surgery in era of ivf
 
electro surgery
electro surgeryelectro surgery
electro surgery
 
ectopic pregnancy Ultrasound Daignosis.pdf
ectopic pregnancy Ultrasound Daignosis.pdfectopic pregnancy Ultrasound Daignosis.pdf
ectopic pregnancy Ultrasound Daignosis.pdf
 
Hyseroscopy myoma resection
Hyseroscopy myoma resectionHyseroscopy myoma resection
Hyseroscopy myoma resection
 
Uterine compression sutures
Uterine compression suturesUterine compression sutures
Uterine compression sutures
 
Shirodkar sling surgery
Shirodkar sling surgeryShirodkar sling surgery
Shirodkar sling surgery
 
Applications of ir in obstetrics and gynecology2
Applications of ir in obstetrics and gynecology2Applications of ir in obstetrics and gynecology2
Applications of ir in obstetrics and gynecology2
 
Manchester repair (Fothergill’s Operation)
Manchester repair (Fothergill’s Operation)Manchester repair (Fothergill’s Operation)
Manchester repair (Fothergill’s Operation)
 
Endometrios-"an overview"
Endometrios-"an overview"Endometrios-"an overview"
Endometrios-"an overview"
 
PPPP00P
PPPP00PPPPP00P
PPPP00P
 
CAESAREAN SCAR DEFECT
CAESAREAN SCAR DEFECT  CAESAREAN SCAR DEFECT
CAESAREAN SCAR DEFECT
 
Energy Modalities for Gynecologic Laparoscopic Surgery
Energy Modalities for Gynecologic Laparoscopic SurgeryEnergy Modalities for Gynecologic Laparoscopic Surgery
Energy Modalities for Gynecologic Laparoscopic Surgery
 
Cervical incompetence
Cervical incompetenceCervical incompetence
Cervical incompetence
 
Recurrent pregnancy loss panel discussion
Recurrent pregnancy loss  panel discussionRecurrent pregnancy loss  panel discussion
Recurrent pregnancy loss panel discussion
 
3 d ultrasound in gynecology presentation
3 d ultrasound in gynecology presentation3 d ultrasound in gynecology presentation
3 d ultrasound in gynecology presentation
 
Laproscopy & hysteroscopy in gynecology no video
Laproscopy & hysteroscopy in gynecology  no videoLaproscopy & hysteroscopy in gynecology  no video
Laproscopy & hysteroscopy in gynecology no video
 

Similar a Safe use of electrosurgery in gynaecology

Diathermy - Electrocautery
Diathermy - ElectrocauteryDiathermy - Electrocautery
Diathermy - ElectrocauteryMr.Harshad Khade
 
Electrosurgery and Energised Dissection
Electrosurgery and Energised Dissection Electrosurgery and Energised Dissection
Electrosurgery and Energised Dissection Dr Jibreel Yousuf
 
Electrosurgical Unit.pptx
Electrosurgical Unit.pptxElectrosurgical Unit.pptx
Electrosurgical Unit.pptxNayMyoHtet7
 
Complication laparoscopic electrosurgery
Complication laparoscopic electrosurgeryComplication laparoscopic electrosurgery
Complication laparoscopic electrosurgeryMohammed Abdalla
 
Energy sources in surgery
Energy sources in surgeryEnergy sources in surgery
Energy sources in surgeryAnkur Kajal
 
Energy sources in surgery
Energy sources in surgery Energy sources in surgery
Energy sources in surgery Dr.Shivaram HV
 
Energy devices in surgery
Energy devices in surgeryEnergy devices in surgery
Energy devices in surgeryRam Raksha
 
ENERGY DEVICES IN SURGERY
ENERGY DEVICES IN SURGERY ENERGY DEVICES IN SURGERY
ENERGY DEVICES IN SURGERY Be Akash Sah
 
Electrical surgery.pptx
Electrical surgery.pptxElectrical surgery.pptx
Electrical surgery.pptxAmalAli787032
 
Electro surgery in laparoscopy
Electro surgery  in laparoscopyElectro surgery  in laparoscopy
Electro surgery in laparoscopyMed Elsayed
 
Diathermy and fire hazard in operating theatre
Diathermy and fire hazard in operating theatreDiathermy and fire hazard in operating theatre
Diathermy and fire hazard in operating theatrenational hosp abuja
 
intermiate equipments(1).ppt
intermiate equipments(1).pptintermiate equipments(1).ppt
intermiate equipments(1).pptTirusew1
 

Similar a Safe use of electrosurgery in gynaecology (20)

electrosurgery Part 01.pdf
electrosurgery Part 01.pdfelectrosurgery Part 01.pdf
electrosurgery Part 01.pdf
 
Electrosurgery
ElectrosurgeryElectrosurgery
Electrosurgery
 
Elecrosurgery in hysteroscopy
Elecrosurgery in hysteroscopyElecrosurgery in hysteroscopy
Elecrosurgery in hysteroscopy
 
Energy sources in surgery
Energy sources in surgeryEnergy sources in surgery
Energy sources in surgery
 
Diathermy - Electrocautery
Diathermy - ElectrocauteryDiathermy - Electrocautery
Diathermy - Electrocautery
 
Electrosurgery and Energised Dissection
Electrosurgery and Energised Dissection Electrosurgery and Energised Dissection
Electrosurgery and Energised Dissection
 
Electrosurgery
ElectrosurgeryElectrosurgery
Electrosurgery
 
Electrosurgical Unit.pptx
Electrosurgical Unit.pptxElectrosurgical Unit.pptx
Electrosurgical Unit.pptx
 
Complication laparoscopic electrosurgery
Complication laparoscopic electrosurgeryComplication laparoscopic electrosurgery
Complication laparoscopic electrosurgery
 
Energy sources in surgery
Energy sources in surgeryEnergy sources in surgery
Energy sources in surgery
 
Energy sources in surgery
Energy sources in surgery Energy sources in surgery
Energy sources in surgery
 
Energy devices in surgery
Energy devices in surgeryEnergy devices in surgery
Energy devices in surgery
 
Energy sources in surgery
Energy sources in surgeryEnergy sources in surgery
Energy sources in surgery
 
ENERGY DEVICES IN SURGERY
ENERGY DEVICES IN SURGERY ENERGY DEVICES IN SURGERY
ENERGY DEVICES IN SURGERY
 
Electrical surgery.pptx
Electrical surgery.pptxElectrical surgery.pptx
Electrical surgery.pptx
 
Electro surgery in laparoscopy
Electro surgery  in laparoscopyElectro surgery  in laparoscopy
Electro surgery in laparoscopy
 
Surgical diathermy
Surgical diathermySurgical diathermy
Surgical diathermy
 
Diathermy and fire hazard in operating theatre
Diathermy and fire hazard in operating theatreDiathermy and fire hazard in operating theatre
Diathermy and fire hazard in operating theatre
 
Diathermy
DiathermyDiathermy
Diathermy
 
intermiate equipments(1).ppt
intermiate equipments(1).pptintermiate equipments(1).ppt
intermiate equipments(1).ppt
 

Último

Independent Call Girls Service Mohali Sector 116 | 6367187148 | Call Girl Ser...
Independent Call Girls Service Mohali Sector 116 | 6367187148 | Call Girl Ser...Independent Call Girls Service Mohali Sector 116 | 6367187148 | Call Girl Ser...
Independent Call Girls Service Mohali Sector 116 | 6367187148 | Call Girl Ser...karishmasinghjnh
 
Call Girls Rishikesh Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Rishikesh Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Rishikesh Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Rishikesh Just Call 8250077686 Top Class Call Girl Service AvailableDipal Arora
 
Jogeshwari ! Call Girls Service Mumbai - 450+ Call Girl Cash Payment 90042684...
Jogeshwari ! Call Girls Service Mumbai - 450+ Call Girl Cash Payment 90042684...Jogeshwari ! Call Girls Service Mumbai - 450+ Call Girl Cash Payment 90042684...
Jogeshwari ! Call Girls Service Mumbai - 450+ Call Girl Cash Payment 90042684...Anamika Rawat
 
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 9332606886 𖠋 Will You Mis...
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 9332606886 𖠋 Will You Mis...The Most Attractive Hyderabad Call Girls Kothapet 𖠋 9332606886 𖠋 Will You Mis...
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 9332606886 𖠋 Will You Mis...chandars293
 
8980367676 Call Girls In Ahmedabad Escort Service Available 24×7 In Ahmedabad
8980367676 Call Girls In Ahmedabad Escort Service Available 24×7 In Ahmedabad8980367676 Call Girls In Ahmedabad Escort Service Available 24×7 In Ahmedabad
8980367676 Call Girls In Ahmedabad Escort Service Available 24×7 In AhmedabadGENUINE ESCORT AGENCY
 
Call Girl In Pune 👉 Just CALL ME: 9352988975 💋 Call Out Call Both With High p...
Call Girl In Pune 👉 Just CALL ME: 9352988975 💋 Call Out Call Both With High p...Call Girl In Pune 👉 Just CALL ME: 9352988975 💋 Call Out Call Both With High p...
Call Girl In Pune 👉 Just CALL ME: 9352988975 💋 Call Out Call Both With High p...chetankumar9855
 
Call Girls Jaipur Just Call 9521753030 Top Class Call Girl Service Available
Call Girls Jaipur Just Call 9521753030 Top Class Call Girl Service AvailableCall Girls Jaipur Just Call 9521753030 Top Class Call Girl Service Available
Call Girls Jaipur Just Call 9521753030 Top Class Call Girl Service AvailableJanvi Singh
 
Russian Call Girls Service Jaipur {8445551418} ❤️PALLAVI VIP Jaipur Call Gir...
Russian Call Girls Service  Jaipur {8445551418} ❤️PALLAVI VIP Jaipur Call Gir...Russian Call Girls Service  Jaipur {8445551418} ❤️PALLAVI VIP Jaipur Call Gir...
Russian Call Girls Service Jaipur {8445551418} ❤️PALLAVI VIP Jaipur Call Gir...parulsinha
 
Premium Bangalore Call Girls Jigani Dail 6378878445 Escort Service For Hot Ma...
Premium Bangalore Call Girls Jigani Dail 6378878445 Escort Service For Hot Ma...Premium Bangalore Call Girls Jigani Dail 6378878445 Escort Service For Hot Ma...
Premium Bangalore Call Girls Jigani Dail 6378878445 Escort Service For Hot Ma...tanya dube
 
Model Call Girls In Chennai WhatsApp Booking 7427069034 call girl service 24 ...
Model Call Girls In Chennai WhatsApp Booking 7427069034 call girl service 24 ...Model Call Girls In Chennai WhatsApp Booking 7427069034 call girl service 24 ...
Model Call Girls In Chennai WhatsApp Booking 7427069034 call girl service 24 ...hotbabesbook
 
Most Beautiful Call Girl in Bangalore Contact on Whatsapp
Most Beautiful Call Girl in Bangalore Contact on WhatsappMost Beautiful Call Girl in Bangalore Contact on Whatsapp
Most Beautiful Call Girl in Bangalore Contact on WhatsappInaaya Sharma
 
Call Girls Service Jaipur {8445551418} ❤️VVIP BHAWNA Call Girl in Jaipur Raja...
Call Girls Service Jaipur {8445551418} ❤️VVIP BHAWNA Call Girl in Jaipur Raja...Call Girls Service Jaipur {8445551418} ❤️VVIP BHAWNA Call Girl in Jaipur Raja...
Call Girls Service Jaipur {8445551418} ❤️VVIP BHAWNA Call Girl in Jaipur Raja...parulsinha
 
Call Girls Kolkata Kalikapur 💯Call Us 🔝 8005736733 🔝 💃 Top Class Call Girl Se...
Call Girls Kolkata Kalikapur 💯Call Us 🔝 8005736733 🔝 💃 Top Class Call Girl Se...Call Girls Kolkata Kalikapur 💯Call Us 🔝 8005736733 🔝 💃 Top Class Call Girl Se...
Call Girls Kolkata Kalikapur 💯Call Us 🔝 8005736733 🔝 💃 Top Class Call Girl Se...Namrata Singh
 
Call Girls Service Jaipur {9521753030 } ❤️VVIP BHAWNA Call Girl in Jaipur Raj...
Call Girls Service Jaipur {9521753030 } ❤️VVIP BHAWNA Call Girl in Jaipur Raj...Call Girls Service Jaipur {9521753030 } ❤️VVIP BHAWNA Call Girl in Jaipur Raj...
Call Girls Service Jaipur {9521753030 } ❤️VVIP BHAWNA Call Girl in Jaipur Raj...khalifaescort01
 
Top Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any Time
Top Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any TimeTop Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any Time
Top Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any TimeCall Girls Delhi
 
Call Girls Ahmedabad Just Call 9630942363 Top Class Call Girl Service Available
Call Girls Ahmedabad Just Call 9630942363 Top Class Call Girl Service AvailableCall Girls Ahmedabad Just Call 9630942363 Top Class Call Girl Service Available
Call Girls Ahmedabad Just Call 9630942363 Top Class Call Girl Service AvailableGENUINE ESCORT AGENCY
 
Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...
Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...
Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...hotbabesbook
 
Low Rate Call Girls Bangalore {7304373326} ❤️VVIP NISHA Call Girls in Bangalo...
Low Rate Call Girls Bangalore {7304373326} ❤️VVIP NISHA Call Girls in Bangalo...Low Rate Call Girls Bangalore {7304373326} ❤️VVIP NISHA Call Girls in Bangalo...
Low Rate Call Girls Bangalore {7304373326} ❤️VVIP NISHA Call Girls in Bangalo...Sheetaleventcompany
 

Último (20)

Independent Call Girls Service Mohali Sector 116 | 6367187148 | Call Girl Ser...
Independent Call Girls Service Mohali Sector 116 | 6367187148 | Call Girl Ser...Independent Call Girls Service Mohali Sector 116 | 6367187148 | Call Girl Ser...
Independent Call Girls Service Mohali Sector 116 | 6367187148 | Call Girl Ser...
 
Call Girls Rishikesh Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Rishikesh Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Rishikesh Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Rishikesh Just Call 8250077686 Top Class Call Girl Service Available
 
Jogeshwari ! Call Girls Service Mumbai - 450+ Call Girl Cash Payment 90042684...
Jogeshwari ! Call Girls Service Mumbai - 450+ Call Girl Cash Payment 90042684...Jogeshwari ! Call Girls Service Mumbai - 450+ Call Girl Cash Payment 90042684...
Jogeshwari ! Call Girls Service Mumbai - 450+ Call Girl Cash Payment 90042684...
 
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 9332606886 𖠋 Will You Mis...
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 9332606886 𖠋 Will You Mis...The Most Attractive Hyderabad Call Girls Kothapet 𖠋 9332606886 𖠋 Will You Mis...
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 9332606886 𖠋 Will You Mis...
 
🌹Attapur⬅️ Vip Call Girls Hyderabad 📱9352852248 Book Well Trand Call Girls In...
🌹Attapur⬅️ Vip Call Girls Hyderabad 📱9352852248 Book Well Trand Call Girls In...🌹Attapur⬅️ Vip Call Girls Hyderabad 📱9352852248 Book Well Trand Call Girls In...
🌹Attapur⬅️ Vip Call Girls Hyderabad 📱9352852248 Book Well Trand Call Girls In...
 
8980367676 Call Girls In Ahmedabad Escort Service Available 24×7 In Ahmedabad
8980367676 Call Girls In Ahmedabad Escort Service Available 24×7 In Ahmedabad8980367676 Call Girls In Ahmedabad Escort Service Available 24×7 In Ahmedabad
8980367676 Call Girls In Ahmedabad Escort Service Available 24×7 In Ahmedabad
 
Call Girls in Gagan Vihar (delhi) call me [🔝 9953056974 🔝] escort service 24X7
Call Girls in Gagan Vihar (delhi) call me [🔝  9953056974 🔝] escort service 24X7Call Girls in Gagan Vihar (delhi) call me [🔝  9953056974 🔝] escort service 24X7
Call Girls in Gagan Vihar (delhi) call me [🔝 9953056974 🔝] escort service 24X7
 
Call Girl In Pune 👉 Just CALL ME: 9352988975 💋 Call Out Call Both With High p...
Call Girl In Pune 👉 Just CALL ME: 9352988975 💋 Call Out Call Both With High p...Call Girl In Pune 👉 Just CALL ME: 9352988975 💋 Call Out Call Both With High p...
Call Girl In Pune 👉 Just CALL ME: 9352988975 💋 Call Out Call Both With High p...
 
Call Girls Jaipur Just Call 9521753030 Top Class Call Girl Service Available
Call Girls Jaipur Just Call 9521753030 Top Class Call Girl Service AvailableCall Girls Jaipur Just Call 9521753030 Top Class Call Girl Service Available
Call Girls Jaipur Just Call 9521753030 Top Class Call Girl Service Available
 
Russian Call Girls Service Jaipur {8445551418} ❤️PALLAVI VIP Jaipur Call Gir...
Russian Call Girls Service  Jaipur {8445551418} ❤️PALLAVI VIP Jaipur Call Gir...Russian Call Girls Service  Jaipur {8445551418} ❤️PALLAVI VIP Jaipur Call Gir...
Russian Call Girls Service Jaipur {8445551418} ❤️PALLAVI VIP Jaipur Call Gir...
 
Premium Bangalore Call Girls Jigani Dail 6378878445 Escort Service For Hot Ma...
Premium Bangalore Call Girls Jigani Dail 6378878445 Escort Service For Hot Ma...Premium Bangalore Call Girls Jigani Dail 6378878445 Escort Service For Hot Ma...
Premium Bangalore Call Girls Jigani Dail 6378878445 Escort Service For Hot Ma...
 
Model Call Girls In Chennai WhatsApp Booking 7427069034 call girl service 24 ...
Model Call Girls In Chennai WhatsApp Booking 7427069034 call girl service 24 ...Model Call Girls In Chennai WhatsApp Booking 7427069034 call girl service 24 ...
Model Call Girls In Chennai WhatsApp Booking 7427069034 call girl service 24 ...
 
Most Beautiful Call Girl in Bangalore Contact on Whatsapp
Most Beautiful Call Girl in Bangalore Contact on WhatsappMost Beautiful Call Girl in Bangalore Contact on Whatsapp
Most Beautiful Call Girl in Bangalore Contact on Whatsapp
 
Call Girls Service Jaipur {8445551418} ❤️VVIP BHAWNA Call Girl in Jaipur Raja...
Call Girls Service Jaipur {8445551418} ❤️VVIP BHAWNA Call Girl in Jaipur Raja...Call Girls Service Jaipur {8445551418} ❤️VVIP BHAWNA Call Girl in Jaipur Raja...
Call Girls Service Jaipur {8445551418} ❤️VVIP BHAWNA Call Girl in Jaipur Raja...
 
Call Girls Kolkata Kalikapur 💯Call Us 🔝 8005736733 🔝 💃 Top Class Call Girl Se...
Call Girls Kolkata Kalikapur 💯Call Us 🔝 8005736733 🔝 💃 Top Class Call Girl Se...Call Girls Kolkata Kalikapur 💯Call Us 🔝 8005736733 🔝 💃 Top Class Call Girl Se...
Call Girls Kolkata Kalikapur 💯Call Us 🔝 8005736733 🔝 💃 Top Class Call Girl Se...
 
Call Girls Service Jaipur {9521753030 } ❤️VVIP BHAWNA Call Girl in Jaipur Raj...
Call Girls Service Jaipur {9521753030 } ❤️VVIP BHAWNA Call Girl in Jaipur Raj...Call Girls Service Jaipur {9521753030 } ❤️VVIP BHAWNA Call Girl in Jaipur Raj...
Call Girls Service Jaipur {9521753030 } ❤️VVIP BHAWNA Call Girl in Jaipur Raj...
 
Top Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any Time
Top Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any TimeTop Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any Time
Top Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any Time
 
Call Girls Ahmedabad Just Call 9630942363 Top Class Call Girl Service Available
Call Girls Ahmedabad Just Call 9630942363 Top Class Call Girl Service AvailableCall Girls Ahmedabad Just Call 9630942363 Top Class Call Girl Service Available
Call Girls Ahmedabad Just Call 9630942363 Top Class Call Girl Service Available
 
Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...
Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...
Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...
 
Low Rate Call Girls Bangalore {7304373326} ❤️VVIP NISHA Call Girls in Bangalo...
Low Rate Call Girls Bangalore {7304373326} ❤️VVIP NISHA Call Girls in Bangalo...Low Rate Call Girls Bangalore {7304373326} ❤️VVIP NISHA Call Girls in Bangalo...
Low Rate Call Girls Bangalore {7304373326} ❤️VVIP NISHA Call Girls in Bangalo...
 

Safe use of electrosurgery in gynaecology

  • 1. Safe use of electrosurgery in gynaecology Dr Muhamed Al Bellehy MD Consultant Obstetrics and gynecology Dr Muhamed Al Bellehy MD
  • 2.  the two types of electric current: o direct current (DC), which moves in one direction (unidirectional), o AC, which periodically reverses direction (bidirectional). Electrosurgery electrocautery the application of high‐frequency alternating current (AC) in surgery to achieve various thermal tissue effects including cutting, desiccation and coagulation. the passive transfer of heat to the tissue with no current passing through it.  uses AC with high frequency o AC to avoid electrolysis o high frequency to avoid the Faradic effect of nerve and muscle stimulation.  uses DC Faradic effect ceases at frequencies > 100 kHz. Because frequencies commonly used in are >500 kHz, which are similar to radiofrequency (RF), the term RF electrosurgery is used. Dr Muhamed Al Bellehy MD 2
  • 3. Dr Muhamed Al Bellehy MD 3
  • 4. Dr Muhamed Al Bellehy MD 4
  • 5. • Surgical diathermy = electrosurgery Dr Muhamed Al Bellehy MD 5
  • 6. Dr Muhamed Al Bellehy MD 6
  • 7. Dr Muhamed Al Bellehy MD 7
  • 8. Power (Battery) produces difference in voltage  current Resistance will decrease the current Dr Muhamed Al Bellehy MD 8
  • 9. Battery produces difference in voltage  current Resistance Change electricity into another form like heat or light Dr Muhamed Al Bellehy MD 9
  • 10. Increasing the voltage will result in a higher current Increasing the resistance will result in a lower current Dr Muhamed Al Bellehy MD 10
  • 11. Dr Muhamed Al Bellehy MD 11
  • 12. Power is expressed as Whatt = amount of energy delivered in time unit Dr Muhamed Al Bellehy MD 12
  • 13. Electrosurgical unit = voltage Electrical switch Tissue effect= increase temperature Dr Muhamed Al Bellehy MD 13
  • 15. Maximum amount of power In Voltage Dr Muhamed Al Bellehy MD 15
  • 16. amount of power for certain effect Depends on Resistance Type of electrode Dr Muhamed Al Bellehy MD 16
  • 17. Voltage vs power controlled generators Dr Muhamed Al Bellehy MD 17
  • 18. • Electro surgery uses alternating current not direct current • direct current will result in electrolysis Dr Muhamed Al Bellehy MD 18
  • 19. Electro surgery uses higher frequency to avoid tissue stimulation Dr Muhamed Al Bellehy MD 19
  • 20. Dr Muhamed Al Bellehy MD 20
  • 21. Current flows 100% of time + > 200 V Tissue Vaporization Cutting Current flows 6% of time Tissue Cools in between  form a coagulum Coagulation + Some Cutting when > 200 V Dr Muhamed Al Bellehy MD 21
  • 22. Cutting Coagulation + Some Cutting when > 200 V Dr Muhamed Al Bellehy MD 22
  • 23. Coagulation + Some Cutting when > 200 V BLEND mode Activate Current and higher Voltage in a certain way to obtain effect mixed of cutting and coagulation Dr Muhamed Al Bellehy MD 23
  • 24. Dr Muhamed Al Bellehy MD 24
  • 25. Dr Muhamed Al Bellehy MD 25
  • 26. Dr Muhamed Al Bellehy MD 26
  • 27. Monopolar Dr Muhamed Al Bellehy MD 27
  • 28. Dr Muhamed Al Bellehy MD 28
  • 29. Bipolar • Only used for tissues that could be grasped • Effect between two the blades only • So- relatively safe Dr Muhamed Al Bellehy MD 29
  • 30. Gap between two blades No Gap between two blades Dr Muhamed Al Bellehy MD 30
  • 31. Jaw of bipolar forceps Dr Muhamed Al Bellehy MD 31
  • 32. Dispersive blade Active blade Dr Muhamed Al Bellehy MD 32
  • 33. Contact area at Dispersive blade 10000 times > active electrode Current density at Dispersive blade 10000 times < active electrode Temp at Active electrode 10000 times > Dispersive blade Dr Muhamed Al Bellehy MD 33
  • 34. Dr Muhamed Al Bellehy MD 34
  • 35. Dr Muhamed Al Bellehy MD 35
  • 36. Dr Muhamed Al Bellehy MD 36
  • 37. Increasing the Contact area of the active electrode decreases current density  more coagulation decreasing the Contact area of the active electrode increases current density  more cutting Dr Muhamed Al Bellehy MD 37
  • 38. Dr Muhamed Al Bellehy MD 38
  • 39. AKA electrosection • Active electrode is held just lightly against the tissue • Using the cutting or blend modes • When using the blend mode cutting + hemostasis of small BVs Excessive power  sparking and tissue drag Dr Muhamed Al Bellehy MD 39
  • 40. AKA Spray AKA non contact coagulation • Use spark or fulguration mode • Active electrode is held a small distance from the tissue • Spark of, electricity will pass to the tissues  minimal tissue heating + large superficial coagulum Dr Muhamed Al Bellehy MD 40
  • 41. • Use spark or fulguration mode • Active electrode is held a small distance from the tissue • Spark of, electricity will pass to the tissues  minimal tissue heating + large superficial coagulum Dr Muhamed Al Bellehy MD 41
  • 42. Dr Muhamed Al Bellehy MD 42
  • 43. Dr Muhamed Al Bellehy MD 43
  • 44. • Coagulation mode *bi – mono polar* • Active electrode is held in direct contact with the tissue • This result in less heat gradualy delivered and Spark of, electricity will pass to the tissues  deep coagulum Dr Muhamed Al Bellehy MD 44
  • 45. • Use spark or fulguration mode • Active electrode is held a small distance from the tissue • Spark of, electricity will pass to the tissues  minimal tissue heating + large superficial coagulum Dr Muhamed Al Bellehy MD 45
  • 46. • Best with bi polar* Dr Muhamed Al Bellehy MD 46
  • 47. Dr Muhamed Al Bellehy MD 47
  • 48. Cutting mode can coagulate in power controlled generators Coagulation mode can cut when > 200 Volte Dr Muhamed Al Bellehy MD 48
  • 49. Dr Muhamed Al Bellehy MD 49
  • 50. Dr Muhamed Al Bellehy MD 50
  • 51. Dr Muhamed Al Bellehy MD 51
  • 52. Dr Muhamed Al Bellehy MD 52
  • 53. General considerations for • The tip of the electrode • The body • The dispersive blade Dr Muhamed Al Bellehy MD 53
  • 54. Dr Muhamed Al Bellehy MD 54
  • 55. Surgical Smoke • May obscure visualization • May have health hazards – contain toxic gases Dr Muhamed Al Bellehy MD 55
  • 56. Smoke evacuator may be used Dr Muhamed Al Bellehy MD 56
  • 57. Bowel perforation and injury to perianal region will result in leakage of methane and hydrogen  ?? fire Alcohol disinfectants not allowed to dry ?? fire Dr Muhamed Al Bellehy MD 57
  • 58. Hand held electrodes should be kept in non conductive place to avoid danger of accidental activation Dr Muhamed Al Bellehy MD 58
  • 59. Activating the active electrode when the not in contact with patient’s body electrical glove burns Dr Muhamed Al Bellehy MD 59
  • 60. Activating the active electrode when the surgeon is in contact with active electrode and the patient  electrical pathway through the surgeon’s hand  electrical glove burns Dr Muhamed Al Bellehy MD 60
  • 61. Prolonged Activation of the active electrode  increased residual heat  tissue injury Short Activation of the active electrode should be practiced Dr Muhamed Al Bellehy MD 61
  • 62. Prolonged Activation of the active electrode  increased heat  carbonization + tissue sticks to the electrode As the active electrode keeps contact with the tissue  carbonization of the sticking tissue continue further increase resistance  decrease clinical effect  staff wrongly increase the power settings the tip of the electrode should be kept clean ! Dr Muhamed Al Bellehy MD 62
  • 63. Dr Muhamed Al Bellehy MD 63
  • 64. Dr Muhamed Al Bellehy MD 64
  • 65. Grounded Generators 1. The ground is part of the circuit between the dispersive blade and the generator 2. Current pass through the dispersive blade to the ground Dr Muhamed Al Bellehy MD 65
  • 66. Dr Muhamed Al Bellehy MD 66
  • 67. Alternative Grounded objects Dr Muhamed Al Bellehy MD 67
  • 68. Alternative Grounded objects Operating Table and accessories IV stand Older ECG electrodes Dr Muhamed Al Bellehy MD 68
  • 69. When the contact area with the grounded object is small and the resistance is less than that at the dispersive blade high current density at the contact area  alternate site burn Dr Muhamed Al Bellehy MD 69
  • 70. Isolated generators The ground is not part of the circuit between the dispersive blade and the generator The Current pass through the body and the current returning through the dispersive blade can be measured – and must be almost equal Dr Muhamed Al Bellehy MD 70
  • 71. Isolated generators When the difference between Current the passes through the body and the current returning through the dispersive blade– is too large  system is deactivated Dr Muhamed Al Bellehy MD 71
  • 72. Isolated generators When the difference between Current the passes through the body and the current returning through the dispersive blade– is too small  system is not deactivated When the difference– is too small  not necessarily indicates a leak Thus there is a small possibility of alternate site burn Dr Muhamed Al Bellehy MD 72
  • 73. Allow for a small leak Thus there is a small possibility of alternate site burn Dr Muhamed Al Bellehy MD 73
  • 74. A visual check before the procedure, to confirm that the patient is not contact with a conducting object Dr Muhamed Al Bellehy MD 74
  • 75. Electricity may pass through a conducting implantable  burns Dr Muhamed Al Bellehy MD 75
  • 76. The dispersive blade should not be placed near or over a conducting implantable Piercings and jewelry near the surgical site should be removed Dr Muhamed Al Bellehy MD 76
  • 77. Short bursts for 6 seconds Use bipolar Avoid monopolar over surrounding tissues Dr Muhamed Al Bellehy MD 77
  • 78. Dr Muhamed Al Bellehy MD 78
  • 79. Dr Muhamed Al Bellehy MD 79
  • 80. Contact area at Dispersive blade 10000 times > active electrode Current density at Dispersive blade 10000 times < active electrode Temp at Active electrode 10000 times > Dispersive blade Dr Muhamed Al Bellehy MD 80
  • 81. When Contact area at Dispersive blade is reduced Current density at Dispersive blade is increased When Contact area at Dispersive blade is severely reduced  circuit blade burn Dr Muhamed Al Bellehy MD 81
  • 82. When choosing the site of dispersive blade consider • Surgical site • Contact with the table • Accessories • implants Close to surgical site Dr Muhamed Al Bellehy MD 82
  • 83. When choosing the site of dispersive blade consider • Choose a well vascularized muscular area Dr Muhamed Al Bellehy MD 83
  • 84. Dr Muhamed Al Bellehy MD 84
  • 85. Poor placement of the Dispersive blade  contact area is reduced Decrease tissue effect Staff increase power output Increase chance of burns Dr Muhamed Al Bellehy MD 85
  • 86. Use the largest suitable blade Dr Muhamed Al Bellehy MD 86
  • 87. Dr Muhamed Al Bellehy MD 87
  • 88. Return electrode monitoring (REM) Uses a small amount of current to measure the conduction between the two blades of the split blade If high resistance  small contract area or poor tissue conduction Dr Muhamed Al Bellehy MD 88
  • 89. Return electrode monitoring (REM) Uses a small amount of current to measure the conduction between the two blades of the split blade If high resistance  small contract area or poor tissue conduction THEN DEACTIVE THE SYSTEM Dr Muhamed Al Bellehy MD 89
  • 90. THEN DEACTIVE THE SYSTEM Dr Muhamed Al Bellehy MD 90
  • 91. 2 BLADES SHOULD BE AT SIMIALR DISTANCE FROM THE SURGICAL SITE X Dr Muhamed Al Bellehy MD 91
  • 92. Dr Muhamed Al Bellehy MD 92
  • 93. As the current flows in the cords  electromagnetic field Dr Muhamed Al Bellehy MD 93
  • 94. When the cords are bundled stronger electromagnetic field Dr Muhamed Al Bellehy MD 94
  • 95. When the cords are wrapped around an instrument  electromagnetic field will charge it with electric current Dr Muhamed Al Bellehy MD 95
  • 96. Dr Muhamed Al Bellehy MD 96
  • 97. When the cords are wrapped around an instrument  electromagnetic field will charge it with electric current  patient injury if in contact Dr Muhamed Al Bellehy MD 97
  • 98. Do not use near any electronic devices or connections Dr Muhamed Al Bellehy MD 98
  • 99. Dr Muhamed Al Bellehy MD 99
  • 100. Dr Muhamed Al Bellehy MD 100
  • 101. Dr Muhamed Al Bellehy MD 101
  • 102. Dr Muhamed Al Bellehy MD 102
  • 103. Dr Muhamed Al Bellehy MD 103
  • 104. Jaw of bipolar forceps Dr Muhamed Al Bellehy MD 104
  • 105. Dr Muhamed Al Bellehy MD 105
  • 106. View during laparoscopy is limited to a conical area in front of the endoscope Dr Muhamed Al Bellehy MD 106
  • 107. Contact of the instruments with tissues outside filed of vision  injury • Bleeding • Perforation • Leakage of body fluids • Ignition of bowel gases X • Do not activate the current when instrument is not visualized or near any organ or other instruments • Avoid contact of the tip shortly after activation • Avoid prolonged activation of current to reduce residual instrument heat Dr Muhamed Al Bellehy MD 107
  • 108. Dr Muhamed Al Bellehy MD 108
  • 109. Dr Muhamed Al Bellehy MD 109
  • 110. Dr Muhamed Al Bellehy MD 110
  • 111. Dr Muhamed Al Bellehy MD 111
  • 112. high voltage causes spark  Direct coupling with out contact  Dr Muhamed Al Bellehy MD 112
  • 113. To avoid this • As low voltage as possible • Cutting > coagulation • Activate the current only when in direct conatct with the tissues Dr Muhamed Al Bellehy MD 113
  • 114. Dr Muhamed Al Bellehy MD 114
  • 115. Dr Muhamed Al Bellehy MD 115
  • 116. Dr Muhamed Al Bellehy MD 116
  • 117. Dr Muhamed Al Bellehy MD 117
  • 118. Dr Muhamed Al Bellehy MD 118
  • 119. Dr Muhamed Al Bellehy MD 119
  • 120. When monopolar current is used at high voltage  creates high electromagnetic field  charge nearby instruments with current Dr Muhamed Al Bellehy MD 120
  • 121. This does not occur with bipolar instruments, as the electromagnetic field has opposing directions Dr Muhamed Al Bellehy MD 121
  • 122. Metal trocars disperses the charge through the skin Dr Muhamed Al Bellehy MD 122
  • 123. Hybrid trocar systems induces capacitive coupling Dr Muhamed Al Bellehy MD 123
  • 124. Dr Muhamed Al Bellehy MD 124
  • 125. Dr Muhamed Al Bellehy MD 125
  • 126. Dr Muhamed Al Bellehy MD 126
  • 127. Dr Muhamed Al Bellehy MD 127
  • 128. Dr Muhamed Al Bellehy MD 128
  • 129. Bipolar electro surgery is preferred Dr Muhamed Al Bellehy MD 129
  • 130. Dr Muhamed Al Bellehy MD 130
  • 131. Dr Muhamed Al Bellehy MD 131
  • 132. Dr Muhamed Al Bellehy MD 132