Introduction
U.S. EPA Method 8260C – Volatile
Organic Compounds (VOCs) by Gas
Chromatography Mass Spectrometry
(GC/MS) is one of the most common
environmental applications for GC/MS.
This method outlines the analysis of
volatile organic compounds in a variety
of solid waste matrices including vari-
ous air sampling trapping media, ground
and surface water, soils, and sediments
among others. The method requires not
only demonstration of laboratory sample preparation and handling competence
but instrument performance as well. The study presented here demonstrates
the PerkinElmer®
Clarus®
SQ 8 GC/MS with purge and trap sample introduction
both meets and exceeds the performance criteria set out in method 8260C and
describes the analytical results and instrumental methodology.
Experimental
The PerkinElmer Clarus SQ 8C GC/MS operating in electron ionization mode
with an Atomx purge and trap sample introduction system (Teledyne Tekmar,
Mason, OH) was used to perform these experiments. The purge and trap
conditions are presented in Table 1 and represent standard conditions for
the analysis of method of VOCs by EPA Method 8260C.
Gas Chromatography/
Mass Spectrometry
a p p l i c a t i o n N o t e
Authors
Ruben Garnica
Dawn May
PerkinElmer, Inc.
Shelton, CT USA
Method 8260C by
Purge and Trap Gas
Chromatography
Mass Spectrometry
using the Clarus SQ 8
2
Table 1. Purge and Trap Instrument Conditions.
Purge and Trap System:	 Tekmar – Atomx
Trap	 Tekmar #9 trap
Sample Size	 5 mL
Purge Parameters:
Valve Oven Temp	 140 °C
Transfer Line Temp	 140 °C
Sample Mount Temp	 90 °C
Water Heater Temp	 90 °C
Sample Vial Temp	 20 °C
Sample Equilibrate Time	 0.00 min
Standby Flow	 10 mL/min
Purge Ready Temp	 40 °C
Condensate Ready Temp	 45 °C
Presweep Time	 0.25 min
Prime Sample Fill Volume	 3.0 mL
Sweep Sample Time	 0.25 min
Sweep Sample Flow	 100 mL/min
Sparge Vessel Heater	 On
Purge Time	 11.00 min
Purge Flow	 40 mL/min
Purge Temp	 40 °C
Condensate Purge Temp	 20 °C
Dry Purge Time	 2.00 min
Dry Purge Flow	 100 mL/min
Dry Purge Temp	 20 °C
Desorb Parameters:	
Water Needle Rinse Volume	 7.0 mL
Sweep Needle Time	 0.50 min
Desorb Preheat Temp	 200 °C
GC Start Signal	 Start of Desorb
Desorb Time	 2.00 min
Drain Flow	 300 mL/min
Desorb Temp	 200 °C
Bake Parameters:	
Number of Water Bake Rinses	 2
Water Bake Rinse Volume	 7.0 mL
Bake Rinse Sweep Time	 0.25 min
Bake Rinse Sweep Flow	 100 mL/min
Bake Rinse Drain Time	 0.40 min
Bake Time	 6.00 min
Bake Flow	 250 mL/min
Bake Temp	 270 °C
Condensate Bake Temp	 200 °C
	
The Clarus SQ 8C GC/MS conditions are presented in Table 2.
The heated transfer line of the concentrator was connected
directly to the injector port of the Clarus 680 GC. A 2 mm
i.d. injector port liner was used inside the capillary injector.
Split flow adjustments will increase or decrease the split
ratio and the resulting sensitivity of the system and allow
the analysts flexibility in optimizing their GC/MS system.
Table 2. Clarus SQ 8C GC/MS conditions.
Gas Chromatograph:	 PerkinElmer Clarus 680
Analytical Column:	 Elite 624 MS
	 (30 m x 0.25 mm i.d. x 1.4 μm)
Injector Temperature:	 220 °C
Carrier Gas:	 Helium @ 1.0 mL/min
Split Flow:	 80 mL/min		
Oven Program:	 Temperature	 Hold Time	 Rate
	 40 °C	 2 min	 10 °C/min
	 100 °C	 0 min	 30 °C/min
	 200 °C	 4 min	 End
Mass Spectrometer:	 PerkinElmer Clarus SQ 8C
GC Transfer Line
Temperature:	 220 °C
Ion Source Temperature:	300 °C
Function Type:	 Full Scan
Solvent Delay:	 0 – 0.50 min
Scan Range:	 m/z 35 – 270
Scan Time:	 0.20 sec
Interscan Delay:	 0.10 sec
Calibration and performance standards were prepared
from commercially available stock standards and diluted in
Class-A volumetric flasks. Calibration standards were made
from a combination of four multi-component standard mixes
including a 76 component 8260 calibration mix, a 6 compo-
nent 502.2 calibration gas mix, a 3 component VOA surrogate
spike mix, and an 8260 internal standard mix. Stock stan-
dards were diluted to an intermediate concentration from
which initial calibration standards were prepared. The cali-
bration levels used in this study are presented in Table 3.
Standards used for the Method Detection Limit (MDL) and
precision experiments were prepared at appropriate con-
centration levels from intermediate standards. The internal
standard mix was added to all samples such that a resulting
concentration of 20 μg/L was achieved per 5 mL of sample.
The tuning standard, 4-bromofluorobenzene, was prepared
at 50 μg/L. Standards, stock standards, and associated QC/
QA samples were stored in a manner consistent with the
guidelines set out in the method.
3
between analytical runs and in this study the injection-to-
injection time, measured to be under 30 minutes, was
limited by the purge and trap system.
Table 4 presents the analytical results of these experiments
including the Retention Time (RT), Average Relative Response
Factors (Avg RRF) and Percent Relative Standard Deviation
(% RSD) of the initial calibration, the Method Detection
Limits (MDL), and the method precision as Percent Recovery
(% Rec) and accuracy as % RSD. All of the analytes presented
meet or exceed the minimum method requirements and
boast excellent detection limits, precision and accuracy.
Method Detection Limits were determined by analyzing
seven replicate samples at a concentration of 0.5 μg/L and
1.0 µg/L. Precision and accuracy were measured by analyzing
four replicates at a concentration of 25 μg/L.
The majority of compounds in this report are calibrated
over the entire concentration range as presented in Table 3.
Compounds whose calibration range varied from the bulk of
analytes are highlighted in Table 5. Alternate calibration
ranges are expected for this method especially for compounds
referred to as “bad actors”. In all cases, however, EPA method
criteria are met. Overall improved results including MDL and
precision and accuracy were obtained by operating the ion
source at an elevated temperature than would regularly be
used. Figure 2 shows a time selection of the seven TICs from
the MDL study and illustrates the excellent repeatability of
the system at a low concentration level. The inset presents
the seven extracted ion chromatograms of naphthalene (EIC)
at m/z = 128. The % RSD of the seven naphthalene EICs
was measured to be 3.8%, which includes the sample
handling performed by the purge and trap.
Table 3. Calibration points employed in this study.
Calibration Level	 Concentration (μg/L)
1	0.5
2	1.0
3	2.0
4	5.0
5	10
6	25
7	50
8	100
9	200
Results and Discussion
The GC conditions were optimized to provide adequate
separation while keeping the analytical runtime as short
as possible. Figure 1 demonstrates a mid-level (25 μg/L)
total ion chromatogram (TIC) from this analysis. All of the
expected analyte separation is achieved while allowing for
a maximum of throughput per unit laboratory time. The
injection-to-injection time is further enhanced by the perfor-
mance of the Clarus 680 GC oven. The oven incorporates
a dual walled design with a unique air flow path to allow
the fastest cooling times of any commercially available GC
oven on the market. Air flow through the Clarus 680 GC
is unidirectional eliminating the mixing of heated exhaust
and cool ambient air allowing for the fastest transport of
heat out of the GC oven. This high speed cooling allows the
user to dramatically decrease the idle time of the instrument
4
Figure 1. TIC of 8260C volatile organic standard mix at 25 μg/L. Numbering references compound as listed in Table 4.
Figure 2. Zoom of the seven MDL samples. TICs overlaid to illustrate the excellent repeatability achieved at low levels. Inset shows the naphthalene EIC at
m/z = 128; the % RSD was measured to be 3.8%.
5
Table 4. Summarized initial calibration results of the analyzed VOCs from 0.5 – 200 μg/L for most compounds.
A hyphen (-) indicates value either not provided by the method or not generated in this work.
A plus sign (+) indicates calibration calculated using linear regression with R2
presented in table.
				 8260C		 8260C
		 	 Avg	 Minimum		 % RSD	 MDL	 Precision	 Accuracy
#	 Compound	 RT	 RRF	 RRF	 % RSD	 Criteria	 (µg/L)	 (% Rec)	 (% RSD)
1	Dichlorodifluoromethane	 1.38	0.44	0.1	 14	 20	0.14	90	 2.9
2	Chloromethane+
	 1.59	 0.72	 0.1	 0.9992	0.990	 0.15	 96	 4.9
3	Vinyl chloride	 1.66	0.79	0.1	 8.2	 20	0.21	95	 1.8
4	Bromomethane	 2.00	 0.26	 0.1	 18	 20	 0.18	 105	 4.1
5	Chloroethane	 2.08	0.40	0.1	 4.9	 20	0.24	99	 1.9
6	Trichlorofluoromethane	 2.29	 0.57	 0.1	 5.9	 20	 0.06	 109	 1.2
7	Diethyl ether	 2.58	0.37	 -	 6.7	 20	0.11	98	 4.6
8	1,1,2-Trichloro-1,2,2-trifluor	2.78	0.38	0.1	 13	 20	0.15	99	 1.1
9	1,1-Dichloroethene	 2.82	 0.44	 0.1	 15	 20	 0.30	 102	 1.8
10	Carbon disulfide	 3.02	1.46	0.1	 5.2	 20	0.05	93	 1.4
11	 Allyl chloride+
	 3.21	 0.17	 -	 0.9940	0.990	 0.05	 76	 2.7
12	Acetonitrile+
	 3.29	 0.20	 -	 0.9999	0.990	 0.38	 96	 5.5
13	Methylene chloride	 3.37	0.51	0.1	 16	 20	0.29	94	 2.5
14	 trans-1,2 Dichloroethene	 3.59	 0.46	 0.1	 9.3	 20	 0.34	 105	 1.3
15	Acrylonitrile	 3.67	0.44	 -	 8.8	 20	0.14	97	 4.6
16	1,1,-Dichloroethane	 4.06	 0.98	 0.2	 9.2	 20	 0.16	 107	 2.3
17	Chloroprene	 4.11	0.70	 -	 11	 20	0.25	89	 1.7
18	2,2,-Dichloropropane+
	 4.63	 0.15	 -	 0.9996	0.990	 -	 56	 7.4
19	cis-1,2,-Dichloroethene	 4.66	 0.53	 0.1	 9.6	 20	 0.24	 106	 1.9
20	 Methyl acrylate	 4.76	 0.92	 -	 4.2	 20	 0.14	 100	 5.3
21	Propionitrile	 4.84	 0.22	 -	 4.4	 20	 0.12	 100	 5.7
22	Tetrahydrofuran	 4.92	0.53	 -	 13	 20	0.15	86	 4.7
23	Bromochloromethane	 4.92	 0.26	 -	 4.5	 20	 0.09	 107	 3.6
24	Methacrylonitrile	 4.96	0.83	 -	 8.9	 20	0.12	90	 4.7
25	Chloroform	 4.99	 0.69	 0.2	 4.2	 20	 0.09	 103	 2.1
26	1,1,1-Trichloroethane	 5.14	 0.55	 0.1	 5.5	 20	 0.20	 107	 0.9
27	 Pentafluorobenzene (ISTD)	 5.22	 	 	 	 	 	 	
28	 Carbon tetrachloride	 5.27	 0.38	 0.1	 12	 20	 0.13	 108	 1.2
29	1,1-Dichloropropene	 5.32	0.43	 -	 5.9	 20	0.25	89	 1.6
30	Isobutyl alcohol	 5.49	0.09	 -	 11.5	 20	0.22	89	 5.6
31	Benzene	 5.53	1.33	0.5	 6.9	 20	0.22	90	 1.6
32	 1,2-dichloroethane-d4 (Surr)	 5.55	 	 	 	 	 	 	
33	1,2-Dichloroethane	 5.64	0.42	0.1	 9.8	 20	0.09	91	 2.0
34	 1,4-Difluorobenzene (ISTD)	 6.00	 	 	 	 	 	 	
35	Trichloroethene	 6.25	 0.32	 0.2	 8.9	 20	 0.28	 103	 2.5
36	1,2-Dichloropropane	 6.56	0.47	0.1	 6	 20	0.15	96	 1.8
37	Methyl methacrylate	 6.67	0.39	 -	 6.2	 20	0.17	91	 3.9
38	Dibromomethane	 6.70	 0.19	 -	 4.9	 20	 0.10	 106	 2.7
39	Bromodichloromethane	 6.87	 0.28	 0.2	 7.4	 20	 0.17	 102	 1.9
40	2-Nitropropane+
	 7.23	 0.06	 -	 0.9996	0.990	 -	 94	 4.7
41	cis-1,3-Dichloropropene	 7.41	 0.47	 0.2	 8.7	 20	 0.08	 90	 1.6
42	 Toluene-d8 (Surr)	 7.66
6
Table 4 continued
				 8260C		 8260C
		 	 Avg	 Minimum		 % RSD	 MDL	 Precision	 Accuracy
#	 Compound	 RT	 RRF	 RRF	 % RSD	 Criteria	 (µg/L)	 (% Rec)	 (% RSD)
43	Toluene	 7.74	 0.67	 0.4	 6.5	 20	 0.22	 100	 1.5
44	trans-1,3-Dichloropropene	 8.11	 0.39	 0.1	 17	 20	 0.11	 93	 2.5
45	 Ethyl methacrylate	 8.16	 0.66	 -	 11	 20	 0.07	 101	 2.6
46	1,1,2-Trichloroethane	 8.31	 0.28	 0.1	 4.3	 20	 0.06	 96	 1.9
47	Tetrachloroethene	 8.35	 0.43	 0.2	 20	 20	 0.17	 123	 4.6
48	1,3-Dichloropropane	 8.50	 0.65	 -	 4.4	 20	 0.06	 94	 3.1
49	Chlorodibromomethane	 8.69	 0.28	 0.1	 12	 20	 0.10	 108	 3.7
50	 Ethylene dibromide	 8.81	 0.40	 0.1	 3.6	 20	 0.07	 111	 2.8
51	 Chlorobenzene-d5 (ISTD)	 9.21	 	 	 	 	 	 	
52	Chlorobenzene	 9.24	 1.08	 0.5	 4.7	 20	 0.23	 102	 1.0
53	Ethylbenzene	 9.31	 1.34	 0.1	 2.2	 20	 0.20	 104	 1.4
54	1,1,1,2-Tetrachloroethane	 9.31	 0.27	 -	 7.8	 20	 0.19	 105	 1.9
55	 p- & m-Xylene	 9.41	 1.09	 0.1	 4.4	 20	 0.23	 101	 1.6
56	o-Xylene	 9.73	 1.15	 0.1	 2.5	 20	 0.19	 103	 2.2
57	Styrene	 9.76	 1.09	 0.3	 3.8	 20	 0.18	 106	 1.8
58	Bromoform+
	 9.91	 0.22	 0.1	 0.9996	0.990	 0.10	 104	 2.8
59	Isopropylbenzene	 10.01	 1.61	 0.1	 2.8	 20	 0.18	 106	 1.8
60	cis-1,4-Dichloro-2-butene+
	 10.12	0.22	 -	 0.9986	0.990	0.09	 68	 2.2
61	 Bromofluorobenzene (Surr)	 10.15	 	 	 	 	 	 	
62	Bromobenzene	 10.25	 0.98	 -	 8.5	 20	 0.15	 98	 2.3
63	1,1,2,2-Tetrachloroethane	 10.28	 1.17	 0.3	 5.9	 20	 0.16	 93	 9.9
64	n-Propylbenzene	 10.30	 3.21	 -	 4.8	 20	 0.20	 101	 1.1
65	1,2,3-Trichloropropane	 10.32	 1.76	 -	 5.8	 20	 0.10	 91	 3.3
66	trans-1,4-Dichloro-2-butene+
	10.33	0.25	 -	 0.9991	0.990	0.05	 71	 2.2
67	2-Chlorotoluene	 10.38	 1.88	 -	 4.4	 20	 0.19	 102	 0.9
68	1,3,5-Trimethylbenzene	 10.43	 3.00	 -	 6.3	 20	 0.21	 104	 1.1
69	4-Chlorotoluene	 10.47	 1.99	 -	 7.6	 20	 0.15	 100	 3.2
70	tert-Butylbenzene	 10.64	 2.57	 -	 9.2	 20	 0.18	 94	 1.3
71	1,2,4-Trimethylbenzene	 10.69	 2.94	 -	 4.8	 20	 0.19	 105	 1.2
72	sec-Butylbenzene	 10.79	 3.48	 -	 5.1	 20	 0.20	 107	 1.7
73	4-Isopropyltoluene	 10.88	 2.98	 -	 4.5	 20	 0.22	 97	 1.0
74	1,3-Dichlorobenzene	 10.89	 1.64	 0.6	 9	 20	 0.22	 98	 0.7
75	 1,4-Dichlorobenzene-d4 (ISTD)	 10.93	 	 	 	 	 	 	
76	1,4-Dichlorobenzene	 10.95	 1.65	 0.5	 9.4	 20	 0.18	 97	 1.4
77	n-Butylbenzene	 11.14	 2.10	 -	 6.5	 20	 0.17	 98	 1.2
78	1,2-Dichlorobenzene	 11.19	 1.60	 0.4	 4.5	 20	 0.12	 102	 1.5
79	1,2-Dibromo-3-chloropropane	 11.69	 0.33	 0.05	 9.8	 20	 0.08	 98	 3.6
80	Nitrobenzene+
	 11.84	 0.06	 -	 0.9999	0.990	 -	 102	 5.4
81	1,2,4-Trichlorobenzene	 12.24	 1.19	 0.2	 9.3	 20	 0.16	 92	 1.3
82	Hexachlorobutadiene	 12.32	 0.50	 -	 16	 20	 0.23	 91	 1.3
83	Naphthalene	 12.46	 3.60	 -	 7.3	 20	 0.06	 102	 2.4
84	1,2,3-Trichlorobenzene	 12.65	 1.13	 0.2	 7.2	 20	 0.09	 94	 1.6
For a complete listing of our global offices, visit www.perkinelmer.com/ContactUs
Copyright ©2012, PerkinElmer, Inc. All rights reserved. PerkinElmer®
is a registered trademark of PerkinElmer, Inc. All other trademarks are the property of their respective owners.
010017B_01
PerkinElmer, Inc.
940 Winter Street
Waltham, MA 02451 USA	
P: (800) 762-4000 or
(+1) 203-925-4602
www.perkinelmer.com
Table 5. Alternate calibration range compounds.
Calibration Concentration Range	 Compound
0.5 – 100 μg/L	 Tetrachloroethene
1.0 – 200 μg/L	 Bromomethane
	 Allyl chloride
	Acetonitrile
	 Isobutyl alcohol
2.0 – 200 μg/L	 2,2-Dichloropropane
	2-Nitropropane
	Nitrobenzene
1.0 – 400 µg/L	 p- & m-Xylene
Design advancements in the Clarus SQ 8 mass spectrometer
allow such a wide operating range with the Clarifi™
detector
providing much of the advantage. The detector’s enhanced
sensitivity enables full scan mass spectrometry methods to
be performed resulting in library searchable spectra at the
lowest possible concentration levels. With the added sensitivity
of SIFI™
(selected ion full ion) or SIM (selected ion monitoring)
data acquisition at even lower limits are possible for specific
compounds.
The voltage setting on the Clarifi detector was important in
developing the method for such a wide concentration range
and deserves mention here. The UltraTune™
(Standard –
DFTPP/BFB) function built into the TurboMass™
v6.0 software
is designed to set the voltage such that the m/z = 69 peak
of the tune gas FC43 is approximately 80% in the tune window.
This setting is appropriate for general instrument analyses
however users are encouraged to fine tune this setting
to suit their specific analytical needs. In our case this was
achieved by analyzing two samples, the high and low
concentration level of our calibration, at voltage settings
below that set during UltraTune. The optimum voltage
setting produced measurable signal at the lowest level while
not saturating the detector at the highest level. In our case
the initial UltraTune setting was 1570V with test measure-
ments performed at -50V and -100V of this value. The
analysis was performed at 1520V. The stability of the Clarus
SQ 8 system allows this setting to be used for extended
periods of time and need only be reproduced periodically.
The UltraTune (Standard – DFTPP/BFB) function also produces
a satisfactory tune which meets the Tune Evaluation require-
ments set out in the method. Figure 3 demonstrates the
passing BFB Tune Evaluation Sample. Figure 4 shows the
mass spectrum utilized in the Tune Evaluation Test and all
of the necessary peak intensity comparisons are achieved.
Conclusion
The analysis of VOCs by purge and trap GC/MS following
EPA Method 8260C is easily achievable over a wide con-
centration range using the Clarus SQ 8 GC/MS system.
Satisfactory analytical results were achieved for a concen-
tration range from 0.5 – 200 μg/L for a majority of analyte
compounds using the full scan approach providing library
searchable spectra at all concentration levels. Excellent
detection limits are reported as are accuracy and precision
measurements. A number of technological advances make
the Clarus SQ 8 GC/MS the ideal systems for laboratories
wishing to perform high throughput and sensitivity analyses
with an ease of operation currently unmatched.
Figure 3. TurboMass 6.0 EPA BFB tune evaluation results.
Figure 4. Mass spectrum of EPA BFB tune evaluation used for test in Figure 3.

Más contenido relacionado

La actualidad más candente

Development and Validation of RP – HPLC Method for the estimation of Oxycloza...
Development and Validation of RP – HPLC Method for the estimation of Oxycloza...Development and Validation of RP – HPLC Method for the estimation of Oxycloza...
Development and Validation of RP – HPLC Method for the estimation of Oxycloza...pharmaindexing
 
Development and validation of RP – HPLC method for the estimation of Tylosin ...
Development and validation of RP – HPLC method for the estimation of Tylosin ...Development and validation of RP – HPLC method for the estimation of Tylosin ...
Development and validation of RP – HPLC method for the estimation of Tylosin ...pharmaindexing
 
Analytical purity method development and validation by gas chromatography of ...
Analytical purity method development and validation by gas chromatography of ...Analytical purity method development and validation by gas chromatography of ...
Analytical purity method development and validation by gas chromatography of ...Alexander Decker
 
From Crude to Fuels – Trace Metals Analysis by ICP-OES for ASTM D7691 and D7111
From Crude to Fuels – Trace Metals Analysis by ICP-OES for ASTM D7691 and D7111From Crude to Fuels – Trace Metals Analysis by ICP-OES for ASTM D7691 and D7111
From Crude to Fuels – Trace Metals Analysis by ICP-OES for ASTM D7691 and D7111Shimadzu Scientific Instruments
 
Tocopherol (Vitamin E) Analysis in Vaping E-Liquid by UHPLC-PDA
Tocopherol (Vitamin E) Analysis in Vaping E-Liquid by UHPLC-PDATocopherol (Vitamin E) Analysis in Vaping E-Liquid by UHPLC-PDA
Tocopherol (Vitamin E) Analysis in Vaping E-Liquid by UHPLC-PDAShimadzu Scientific Instruments
 
Identification and Confirmation of Ten Common Seized Drugs Utilizing a Single...
Identification and Confirmation of Ten Common Seized Drugs Utilizing a Single...Identification and Confirmation of Ten Common Seized Drugs Utilizing a Single...
Identification and Confirmation of Ten Common Seized Drugs Utilizing a Single...Shimadzu Scientific Instruments
 
Development and validation of GC-MS method for analysis of chloropyramine hyd...
Development and validation of GC-MS method for analysis of chloropyramine hyd...Development and validation of GC-MS method for analysis of chloropyramine hyd...
Development and validation of GC-MS method for analysis of chloropyramine hyd...iosrphr_editor
 
The Potency Determination of 15 Cannabinoids using the Hemp Analyzer
The Potency Determination of 15 Cannabinoids using the Hemp Analyzer The Potency Determination of 15 Cannabinoids using the Hemp Analyzer
The Potency Determination of 15 Cannabinoids using the Hemp Analyzer Shimadzu Scientific Instruments
 
Ion Chromatography with Electrolytically Regenerated Suppressor for Water Ana...
Ion Chromatography with Electrolytically Regenerated Suppressor for Water Ana...Ion Chromatography with Electrolytically Regenerated Suppressor for Water Ana...
Ion Chromatography with Electrolytically Regenerated Suppressor for Water Ana...Shimadzu Scientific Instruments
 
Download-manuals-water quality-technicalpapers-inter-labfindingsaqc1stround
 Download-manuals-water quality-technicalpapers-inter-labfindingsaqc1stround Download-manuals-water quality-technicalpapers-inter-labfindingsaqc1stround
Download-manuals-water quality-technicalpapers-inter-labfindingsaqc1stroundhydrologywebsite1
 
Instrumentation of Gas Chromatography
Instrumentation of Gas ChromatographyInstrumentation of Gas Chromatography
Instrumentation of Gas ChromatographyZohaib HUSSAIN
 
APPLICATIONS OF GAS CHROMATOGRAPHY [APPLICATIONS OF GC] BY Prof. Dr. P.RAVISA...
APPLICATIONS OF GAS CHROMATOGRAPHY [APPLICATIONS OF GC] BY Prof. Dr. P.RAVISA...APPLICATIONS OF GAS CHROMATOGRAPHY [APPLICATIONS OF GC] BY Prof. Dr. P.RAVISA...
APPLICATIONS OF GAS CHROMATOGRAPHY [APPLICATIONS OF GC] BY Prof. Dr. P.RAVISA...Dr. Ravi Sankar
 
Acute Toxicity Test Using Daphnia Magna - my thesis presentation
Acute Toxicity Test Using Daphnia Magna - my thesis presentationAcute Toxicity Test Using Daphnia Magna - my thesis presentation
Acute Toxicity Test Using Daphnia Magna - my thesis presentationSandra Clarisa Tarigan
 
Art%3 a10.1007%2fs12161 013-9618-4
Art%3 a10.1007%2fs12161 013-9618-4Art%3 a10.1007%2fs12161 013-9618-4
Art%3 a10.1007%2fs12161 013-9618-4Nádia Paracampo
 
Change of Peptides and Free -Amino Acids Contents during Nanjing Dry-Cured Du...
Change of Peptides and Free -Amino Acids Contents during Nanjing Dry-Cured Du...Change of Peptides and Free -Amino Acids Contents during Nanjing Dry-Cured Du...
Change of Peptides and Free -Amino Acids Contents during Nanjing Dry-Cured Du...Agriculture Journal IJOEAR
 
Automated SPE for Capillary Microsampling Poster
Automated SPE for Capillary Microsampling PosterAutomated SPE for Capillary Microsampling Poster
Automated SPE for Capillary Microsampling PosterRick Youngblood
 

La actualidad más candente (19)

Development and Validation of RP – HPLC Method for the estimation of Oxycloza...
Development and Validation of RP – HPLC Method for the estimation of Oxycloza...Development and Validation of RP – HPLC Method for the estimation of Oxycloza...
Development and Validation of RP – HPLC Method for the estimation of Oxycloza...
 
Development and validation of RP – HPLC method for the estimation of Tylosin ...
Development and validation of RP – HPLC method for the estimation of Tylosin ...Development and validation of RP – HPLC method for the estimation of Tylosin ...
Development and validation of RP – HPLC method for the estimation of Tylosin ...
 
Analytical purity method development and validation by gas chromatography of ...
Analytical purity method development and validation by gas chromatography of ...Analytical purity method development and validation by gas chromatography of ...
Analytical purity method development and validation by gas chromatography of ...
 
Gas chromatography
Gas chromatographyGas chromatography
Gas chromatography
 
From Crude to Fuels – Trace Metals Analysis by ICP-OES for ASTM D7691 and D7111
From Crude to Fuels – Trace Metals Analysis by ICP-OES for ASTM D7691 and D7111From Crude to Fuels – Trace Metals Analysis by ICP-OES for ASTM D7691 and D7111
From Crude to Fuels – Trace Metals Analysis by ICP-OES for ASTM D7691 and D7111
 
Tocopherol (Vitamin E) Analysis in Vaping E-Liquid by UHPLC-PDA
Tocopherol (Vitamin E) Analysis in Vaping E-Liquid by UHPLC-PDATocopherol (Vitamin E) Analysis in Vaping E-Liquid by UHPLC-PDA
Tocopherol (Vitamin E) Analysis in Vaping E-Liquid by UHPLC-PDA
 
Identification and Confirmation of Ten Common Seized Drugs Utilizing a Single...
Identification and Confirmation of Ten Common Seized Drugs Utilizing a Single...Identification and Confirmation of Ten Common Seized Drugs Utilizing a Single...
Identification and Confirmation of Ten Common Seized Drugs Utilizing a Single...
 
Development and validation of GC-MS method for analysis of chloropyramine hyd...
Development and validation of GC-MS method for analysis of chloropyramine hyd...Development and validation of GC-MS method for analysis of chloropyramine hyd...
Development and validation of GC-MS method for analysis of chloropyramine hyd...
 
The Potency Determination of 15 Cannabinoids using the Hemp Analyzer
The Potency Determination of 15 Cannabinoids using the Hemp Analyzer The Potency Determination of 15 Cannabinoids using the Hemp Analyzer
The Potency Determination of 15 Cannabinoids using the Hemp Analyzer
 
Ion Chromatography with Electrolytically Regenerated Suppressor for Water Ana...
Ion Chromatography with Electrolytically Regenerated Suppressor for Water Ana...Ion Chromatography with Electrolytically Regenerated Suppressor for Water Ana...
Ion Chromatography with Electrolytically Regenerated Suppressor for Water Ana...
 
Enhancing Volatile Organic Compounds in Water
Enhancing Volatile Organic Compounds in Water Enhancing Volatile Organic Compounds in Water
Enhancing Volatile Organic Compounds in Water
 
Download-manuals-water quality-technicalpapers-inter-labfindingsaqc1stround
 Download-manuals-water quality-technicalpapers-inter-labfindingsaqc1stround Download-manuals-water quality-technicalpapers-inter-labfindingsaqc1stround
Download-manuals-water quality-technicalpapers-inter-labfindingsaqc1stround
 
Icp working
Icp workingIcp working
Icp working
 
Instrumentation of Gas Chromatography
Instrumentation of Gas ChromatographyInstrumentation of Gas Chromatography
Instrumentation of Gas Chromatography
 
APPLICATIONS OF GAS CHROMATOGRAPHY [APPLICATIONS OF GC] BY Prof. Dr. P.RAVISA...
APPLICATIONS OF GAS CHROMATOGRAPHY [APPLICATIONS OF GC] BY Prof. Dr. P.RAVISA...APPLICATIONS OF GAS CHROMATOGRAPHY [APPLICATIONS OF GC] BY Prof. Dr. P.RAVISA...
APPLICATIONS OF GAS CHROMATOGRAPHY [APPLICATIONS OF GC] BY Prof. Dr. P.RAVISA...
 
Acute Toxicity Test Using Daphnia Magna - my thesis presentation
Acute Toxicity Test Using Daphnia Magna - my thesis presentationAcute Toxicity Test Using Daphnia Magna - my thesis presentation
Acute Toxicity Test Using Daphnia Magna - my thesis presentation
 
Art%3 a10.1007%2fs12161 013-9618-4
Art%3 a10.1007%2fs12161 013-9618-4Art%3 a10.1007%2fs12161 013-9618-4
Art%3 a10.1007%2fs12161 013-9618-4
 
Change of Peptides and Free -Amino Acids Contents during Nanjing Dry-Cured Du...
Change of Peptides and Free -Amino Acids Contents during Nanjing Dry-Cured Du...Change of Peptides and Free -Amino Acids Contents during Nanjing Dry-Cured Du...
Change of Peptides and Free -Amino Acids Contents during Nanjing Dry-Cured Du...
 
Automated SPE for Capillary Microsampling Poster
Automated SPE for Capillary Microsampling PosterAutomated SPE for Capillary Microsampling Poster
Automated SPE for Capillary Microsampling Poster
 

Destacado

Use of gc and gcms in api analysis
Use of gc and gcms in api analysis Use of gc and gcms in api analysis
Use of gc and gcms in api analysis Dipak Shetty
 
Pittcon Powerpoint (2014)
Pittcon Powerpoint (2014)Pittcon Powerpoint (2014)
Pittcon Powerpoint (2014)Chris Dwyer
 
GAS CHROMATOGRAPHY AND MASS SPECTROMETRY (GC-MS) BY P.RAVISANKAR.
GAS CHROMATOGRAPHY AND MASS SPECTROMETRY (GC-MS) BY P.RAVISANKAR.GAS CHROMATOGRAPHY AND MASS SPECTROMETRY (GC-MS) BY P.RAVISANKAR.
GAS CHROMATOGRAPHY AND MASS SPECTROMETRY (GC-MS) BY P.RAVISANKAR.Dr. Ravi Sankar
 
Gas chromatography and its instrumentation
Gas chromatography and its instrumentationGas chromatography and its instrumentation
Gas chromatography and its instrumentationArgha Sen
 

Destacado (6)

Use of gc and gcms in api analysis
Use of gc and gcms in api analysis Use of gc and gcms in api analysis
Use of gc and gcms in api analysis
 
Pittcon Powerpoint (2014)
Pittcon Powerpoint (2014)Pittcon Powerpoint (2014)
Pittcon Powerpoint (2014)
 
Gas Chromatography (GC) By Faizan Akram
Gas Chromatography (GC) By Faizan Akram Gas Chromatography (GC) By Faizan Akram
Gas Chromatography (GC) By Faizan Akram
 
GAS CHROMATOGRAPHY AND MASS SPECTROMETRY (GC-MS) BY P.RAVISANKAR.
GAS CHROMATOGRAPHY AND MASS SPECTROMETRY (GC-MS) BY P.RAVISANKAR.GAS CHROMATOGRAPHY AND MASS SPECTROMETRY (GC-MS) BY P.RAVISANKAR.
GAS CHROMATOGRAPHY AND MASS SPECTROMETRY (GC-MS) BY P.RAVISANKAR.
 
Gas chromatography GC
Gas chromatography GCGas chromatography GC
Gas chromatography GC
 
Gas chromatography and its instrumentation
Gas chromatography and its instrumentationGas chromatography and its instrumentation
Gas chromatography and its instrumentation
 

Similar a Method 8260C by Purge and Trap Gas Chromatography Mass Spectrometry using the Clarus SQ 8

Improved Sensitivity and Dynamic Range Using the Clarus SQ 8 GC/MS System for...
Improved Sensitivity and Dynamic Range Using the Clarus SQ 8 GC/MS System for...Improved Sensitivity and Dynamic Range Using the Clarus SQ 8 GC/MS System for...
Improved Sensitivity and Dynamic Range Using the Clarus SQ 8 GC/MS System for...PerkinElmer, Inc.
 
Analysis of Trace Elements in Water by EPA Method 200.8 using ICP Mass Spectr...
Analysis of Trace Elements in Water by EPA Method 200.8 using ICP Mass Spectr...Analysis of Trace Elements in Water by EPA Method 200.8 using ICP Mass Spectr...
Analysis of Trace Elements in Water by EPA Method 200.8 using ICP Mass Spectr...Shimadzu Scientific Instruments
 
Using THGA and Zeeman Background Correction for Blood-Lead Determination in C...
Using THGA and Zeeman Background Correction for Blood-Lead Determination in C...Using THGA and Zeeman Background Correction for Blood-Lead Determination in C...
Using THGA and Zeeman Background Correction for Blood-Lead Determination in C...PerkinElmer, Inc.
 
Determination of Argon in Ammonia Plant Process Gas Streams by Gas Chromatogr...
Determination of Argon in Ammonia Plant Process Gas Streams by Gas Chromatogr...Determination of Argon in Ammonia Plant Process Gas Streams by Gas Chromatogr...
Determination of Argon in Ammonia Plant Process Gas Streams by Gas Chromatogr...Gerard B. Hawkins
 
Qualification of instrumets
Qualification of instrumetsQualification of instrumets
Qualification of instrumetsChowdaryPavani
 
Cwea Me Hg Presentation Er
Cwea Me Hg Presentation ErCwea Me Hg Presentation Er
Cwea Me Hg Presentation Ernoel_enoki
 
PPT - Electronics Industry Wastewater
PPT - Electronics Industry WastewaterPPT - Electronics Industry Wastewater
PPT - Electronics Industry WastewaterNitin Yadav
 
Improved Sensitivity and Dynamic Range Using the Clarus SQ8 GC/MS for EPA Met...
Improved Sensitivity and Dynamic Range Using the Clarus SQ8 GC/MS for EPA Met...Improved Sensitivity and Dynamic Range Using the Clarus SQ8 GC/MS for EPA Met...
Improved Sensitivity and Dynamic Range Using the Clarus SQ8 GC/MS for EPA Met...PerkinElmer, Inc.
 
Analysis of Quaternary Ammonium Compounds (QACs) as Possible Disinfectant Res...
Analysis of Quaternary Ammonium Compounds (QACs) as Possible Disinfectant Res...Analysis of Quaternary Ammonium Compounds (QACs) as Possible Disinfectant Res...
Analysis of Quaternary Ammonium Compounds (QACs) as Possible Disinfectant Res...PerkinElmer, Inc.
 
Determination of Residual Solvents in Flexible Packaging
Determination of Residual Solvents in Flexible PackagingDetermination of Residual Solvents in Flexible Packaging
Determination of Residual Solvents in Flexible PackagingPerkinElmer, Inc.
 
Analysis of Volatile Organic Compounds (VOCs) in Air Using US EPA Method TO-17
Analysis of Volatile Organic Compounds (VOCs) in Air Using US EPA Method TO-17Analysis of Volatile Organic Compounds (VOCs) in Air Using US EPA Method TO-17
Analysis of Volatile Organic Compounds (VOCs) in Air Using US EPA Method TO-17PerkinElmer, Inc.
 
Analysis of Volatile Organic Compounds (VOCs) in Air Using US EPA Method TO-17
Analysis of Volatile Organic Compounds (VOCs) in Air Using US EPA Method TO-17Analysis of Volatile Organic Compounds (VOCs) in Air Using US EPA Method TO-17
Analysis of Volatile Organic Compounds (VOCs) in Air Using US EPA Method TO-17PerkinElmer, Inc.
 
Determination of Carbon Dioxide, Ethane And Nitrogen in Natural Gas by Gas C...
Determination of Carbon Dioxide, Ethane  And Nitrogen in Natural Gas by Gas C...Determination of Carbon Dioxide, Ethane  And Nitrogen in Natural Gas by Gas C...
Determination of Carbon Dioxide, Ethane And Nitrogen in Natural Gas by Gas C...Gerard B. Hawkins
 
Vitamin D in Serum-MicroLiter
Vitamin D in Serum-MicroLiterVitamin D in Serum-MicroLiter
Vitamin D in Serum-MicroLiterRick Youngblood
 

Similar a Method 8260C by Purge and Trap Gas Chromatography Mass Spectrometry using the Clarus SQ 8 (20)

Improved Sensitivity and Dynamic Range Using the Clarus SQ 8 GC/MS System for...
Improved Sensitivity and Dynamic Range Using the Clarus SQ 8 GC/MS System for...Improved Sensitivity and Dynamic Range Using the Clarus SQ 8 GC/MS System for...
Improved Sensitivity and Dynamic Range Using the Clarus SQ 8 GC/MS System for...
 
Analysis of Trace Elements in Water by EPA Method 200.8 using ICP Mass Spectr...
Analysis of Trace Elements in Water by EPA Method 200.8 using ICP Mass Spectr...Analysis of Trace Elements in Water by EPA Method 200.8 using ICP Mass Spectr...
Analysis of Trace Elements in Water by EPA Method 200.8 using ICP Mass Spectr...
 
Tools of the Trade
Tools of the TradeTools of the Trade
Tools of the Trade
 
Using THGA and Zeeman Background Correction for Blood-Lead Determination in C...
Using THGA and Zeeman Background Correction for Blood-Lead Determination in C...Using THGA and Zeeman Background Correction for Blood-Lead Determination in C...
Using THGA and Zeeman Background Correction for Blood-Lead Determination in C...
 
Determination of Argon in Ammonia Plant Process Gas Streams by Gas Chromatogr...
Determination of Argon in Ammonia Plant Process Gas Streams by Gas Chromatogr...Determination of Argon in Ammonia Plant Process Gas Streams by Gas Chromatogr...
Determination of Argon in Ammonia Plant Process Gas Streams by Gas Chromatogr...
 
GC Agilent .pptx
GC Agilent .pptxGC Agilent .pptx
GC Agilent .pptx
 
Qualification of instrumets
Qualification of instrumetsQualification of instrumets
Qualification of instrumets
 
Cwea Me Hg Presentation Er
Cwea Me Hg Presentation ErCwea Me Hg Presentation Er
Cwea Me Hg Presentation Er
 
PPT - Electronics Industry Wastewater
PPT - Electronics Industry WastewaterPPT - Electronics Industry Wastewater
PPT - Electronics Industry Wastewater
 
Improved Sensitivity and Dynamic Range Using the Clarus SQ8 GC/MS for EPA Met...
Improved Sensitivity and Dynamic Range Using the Clarus SQ8 GC/MS for EPA Met...Improved Sensitivity and Dynamic Range Using the Clarus SQ8 GC/MS for EPA Met...
Improved Sensitivity and Dynamic Range Using the Clarus SQ8 GC/MS for EPA Met...
 
Analysis of Quaternary Ammonium Compounds (QACs) as Possible Disinfectant Res...
Analysis of Quaternary Ammonium Compounds (QACs) as Possible Disinfectant Res...Analysis of Quaternary Ammonium Compounds (QACs) as Possible Disinfectant Res...
Analysis of Quaternary Ammonium Compounds (QACs) as Possible Disinfectant Res...
 
Determination of Residual Solvents in Flexible Packaging
Determination of Residual Solvents in Flexible PackagingDetermination of Residual Solvents in Flexible Packaging
Determination of Residual Solvents in Flexible Packaging
 
Analysis of Volatile Organic Compounds (VOCs) in Air Using US EPA Method TO-17
Analysis of Volatile Organic Compounds (VOCs) in Air Using US EPA Method TO-17Analysis of Volatile Organic Compounds (VOCs) in Air Using US EPA Method TO-17
Analysis of Volatile Organic Compounds (VOCs) in Air Using US EPA Method TO-17
 
Analysis of Volatile Organic Compounds (VOCs) in Air Using US EPA Method TO-17
Analysis of Volatile Organic Compounds (VOCs) in Air Using US EPA Method TO-17Analysis of Volatile Organic Compounds (VOCs) in Air Using US EPA Method TO-17
Analysis of Volatile Organic Compounds (VOCs) in Air Using US EPA Method TO-17
 
Hydraulic Fracturing and Environmental Testing of Water
Hydraulic Fracturing and Environmental Testing of WaterHydraulic Fracturing and Environmental Testing of Water
Hydraulic Fracturing and Environmental Testing of Water
 
Determination of Carbon Dioxide, Ethane And Nitrogen in Natural Gas by Gas C...
Determination of Carbon Dioxide, Ethane  And Nitrogen in Natural Gas by Gas C...Determination of Carbon Dioxide, Ethane  And Nitrogen in Natural Gas by Gas C...
Determination of Carbon Dioxide, Ethane And Nitrogen in Natural Gas by Gas C...
 
An 0903-0042 en
An 0903-0042 enAn 0903-0042 en
An 0903-0042 en
 
An 0903-0042 en
An 0903-0042 enAn 0903-0042 en
An 0903-0042 en
 
Chromatography: Meeting the Challenges of EU regulations with up-to-date Conf...
Chromatography: Meeting the Challenges of EU regulations with up-to-date Conf...Chromatography: Meeting the Challenges of EU regulations with up-to-date Conf...
Chromatography: Meeting the Challenges of EU regulations with up-to-date Conf...
 
Vitamin D in Serum-MicroLiter
Vitamin D in Serum-MicroLiterVitamin D in Serum-MicroLiter
Vitamin D in Serum-MicroLiter
 

Más de PerkinElmer, Inc.

Detectando a Pré Eclampsia no 1º Trimestre
Detectando a Pré Eclampsia no 1º TrimestreDetectando a Pré Eclampsia no 1º Trimestre
Detectando a Pré Eclampsia no 1º TrimestrePerkinElmer, Inc.
 
As Tecnologias que Definem o Desempenho
As Tecnologias que Definem o DesempenhoAs Tecnologias que Definem o Desempenho
As Tecnologias que Definem o DesempenhoPerkinElmer, Inc.
 
GCMS Portátil: Análises em 5 minutos!
GCMS Portátil: Análises em 5 minutos!GCMS Portátil: Análises em 5 minutos!
GCMS Portátil: Análises em 5 minutos!PerkinElmer, Inc.
 
Nós transformamos complexidade em simplicidade
Nós transformamos complexidade em simplicidadeNós transformamos complexidade em simplicidade
Nós transformamos complexidade em simplicidadePerkinElmer, Inc.
 
Janus G3: Rapidez e Automação que você precisa.
Janus G3: Rapidez e Automação que você precisa.Janus G3: Rapidez e Automação que você precisa.
Janus G3: Rapidez e Automação que você precisa.PerkinElmer, Inc.
 
Como a PerkinElmer contribui para diagnósticos mais rápidos?
Como a PerkinElmer contribui para diagnósticos mais rápidos?Como a PerkinElmer contribui para diagnósticos mais rápidos?
Como a PerkinElmer contribui para diagnósticos mais rápidos?PerkinElmer, Inc.
 
Como a PerkinElmer contribui para diagnósticos mais rápidos?
Como a PerkinElmer contribui para diagnósticos mais rápidos?Como a PerkinElmer contribui para diagnósticos mais rápidos?
Como a PerkinElmer contribui para diagnósticos mais rápidos?PerkinElmer, Inc.
 
O Mercado da Carne no Brasil
O Mercado da Carne no BrasilO Mercado da Carne no Brasil
O Mercado da Carne no BrasilPerkinElmer, Inc.
 
KITS de Preparo para Biblioteca NGS
KITS de Preparo para Biblioteca NGSKITS de Preparo para Biblioteca NGS
KITS de Preparo para Biblioteca NGSPerkinElmer, Inc.
 
Escolha um parceiro e não um fornecedor
Escolha um parceiro e não um fornecedorEscolha um parceiro e não um fornecedor
Escolha um parceiro e não um fornecedorPerkinElmer, Inc.
 
Eletroforese Capilar Automatizada
Eletroforese Capilar Automatizada Eletroforese Capilar Automatizada
Eletroforese Capilar Automatizada PerkinElmer, Inc.
 
Analysis of Trace Elements in Fertilizer with AVIO 200 ICP-OES
Analysis of Trace Elements in Fertilizer with AVIO 200 ICP-OESAnalysis of Trace Elements in Fertilizer with AVIO 200 ICP-OES
Analysis of Trace Elements in Fertilizer with AVIO 200 ICP-OESPerkinElmer, Inc.
 
Analysis of Impurities in Semiconductor-Grade Hydrochloric Acid with the NexI...
Analysis of Impurities in Semiconductor-Grade Hydrochloric Acid with the NexI...Analysis of Impurities in Semiconductor-Grade Hydrochloric Acid with the NexI...
Analysis of Impurities in Semiconductor-Grade Hydrochloric Acid with the NexI...PerkinElmer, Inc.
 
New Research Evaluating Cisplatin Uptake in Ovarian Cancer Cells by Single Ce...
New Research Evaluating Cisplatin Uptake in Ovarian Cancer Cells by Single Ce...New Research Evaluating Cisplatin Uptake in Ovarian Cancer Cells by Single Ce...
New Research Evaluating Cisplatin Uptake in Ovarian Cancer Cells by Single Ce...PerkinElmer, Inc.
 
High Quality DNA Isolation Suitable for Ultra Rapid Sequencing
High Quality DNA Isolation Suitable for Ultra Rapid SequencingHigh Quality DNA Isolation Suitable for Ultra Rapid Sequencing
High Quality DNA Isolation Suitable for Ultra Rapid SequencingPerkinElmer, Inc.
 

Más de PerkinElmer, Inc. (20)

Detectando a Pré Eclampsia no 1º Trimestre
Detectando a Pré Eclampsia no 1º TrimestreDetectando a Pré Eclampsia no 1º Trimestre
Detectando a Pré Eclampsia no 1º Trimestre
 
É A VEZ DO PRIME
É A VEZ DO PRIMEÉ A VEZ DO PRIME
É A VEZ DO PRIME
 
TREC PARA A VIDA
TREC PARA A VIDATREC PARA A VIDA
TREC PARA A VIDA
 
As Tecnologias que Definem o Desempenho
As Tecnologias que Definem o DesempenhoAs Tecnologias que Definem o Desempenho
As Tecnologias que Definem o Desempenho
 
GCMS Portátil: Análises em 5 minutos!
GCMS Portátil: Análises em 5 minutos!GCMS Portátil: Análises em 5 minutos!
GCMS Portátil: Análises em 5 minutos!
 
Nós transformamos complexidade em simplicidade
Nós transformamos complexidade em simplicidadeNós transformamos complexidade em simplicidade
Nós transformamos complexidade em simplicidade
 
Janus G3: Rapidez e Automação que você precisa.
Janus G3: Rapidez e Automação que você precisa.Janus G3: Rapidez e Automação que você precisa.
Janus G3: Rapidez e Automação que você precisa.
 
Como a PerkinElmer contribui para diagnósticos mais rápidos?
Como a PerkinElmer contribui para diagnósticos mais rápidos?Como a PerkinElmer contribui para diagnósticos mais rápidos?
Como a PerkinElmer contribui para diagnósticos mais rápidos?
 
Como a PerkinElmer contribui para diagnósticos mais rápidos?
Como a PerkinElmer contribui para diagnósticos mais rápidos?Como a PerkinElmer contribui para diagnósticos mais rápidos?
Como a PerkinElmer contribui para diagnósticos mais rápidos?
 
Aspirina Vs Pré Eclampsia
Aspirina Vs Pré Eclampsia Aspirina Vs Pré Eclampsia
Aspirina Vs Pré Eclampsia
 
O Mercado da Carne no Brasil
O Mercado da Carne no BrasilO Mercado da Carne no Brasil
O Mercado da Carne no Brasil
 
KITS de Preparo para Biblioteca NGS
KITS de Preparo para Biblioteca NGSKITS de Preparo para Biblioteca NGS
KITS de Preparo para Biblioteca NGS
 
O que é X Frágil?
O que é X Frágil?O que é X Frágil?
O que é X Frágil?
 
Escolha um parceiro e não um fornecedor
Escolha um parceiro e não um fornecedorEscolha um parceiro e não um fornecedor
Escolha um parceiro e não um fornecedor
 
Eletroforese Capilar Automatizada
Eletroforese Capilar Automatizada Eletroforese Capilar Automatizada
Eletroforese Capilar Automatizada
 
Analysis of Trace Elements in Fertilizer with AVIO 200 ICP-OES
Analysis of Trace Elements in Fertilizer with AVIO 200 ICP-OESAnalysis of Trace Elements in Fertilizer with AVIO 200 ICP-OES
Analysis of Trace Elements in Fertilizer with AVIO 200 ICP-OES
 
Analysis of Impurities in Semiconductor-Grade Hydrochloric Acid with the NexI...
Analysis of Impurities in Semiconductor-Grade Hydrochloric Acid with the NexI...Analysis of Impurities in Semiconductor-Grade Hydrochloric Acid with the NexI...
Analysis of Impurities in Semiconductor-Grade Hydrochloric Acid with the NexI...
 
New Research Evaluating Cisplatin Uptake in Ovarian Cancer Cells by Single Ce...
New Research Evaluating Cisplatin Uptake in Ovarian Cancer Cells by Single Ce...New Research Evaluating Cisplatin Uptake in Ovarian Cancer Cells by Single Ce...
New Research Evaluating Cisplatin Uptake in Ovarian Cancer Cells by Single Ce...
 
CG Clarus® All In One
CG Clarus® All In OneCG Clarus® All In One
CG Clarus® All In One
 
High Quality DNA Isolation Suitable for Ultra Rapid Sequencing
High Quality DNA Isolation Suitable for Ultra Rapid SequencingHigh Quality DNA Isolation Suitable for Ultra Rapid Sequencing
High Quality DNA Isolation Suitable for Ultra Rapid Sequencing
 

Último

VIP Model Call Girls Viman Nagar ( Pune ) Call ON 8005736733 Starting From 5K...
VIP Model Call Girls Viman Nagar ( Pune ) Call ON 8005736733 Starting From 5K...VIP Model Call Girls Viman Nagar ( Pune ) Call ON 8005736733 Starting From 5K...
VIP Model Call Girls Viman Nagar ( Pune ) Call ON 8005736733 Starting From 5K...SUHANI PANDEY
 
Get Premium Attur Layout Call Girls (8005736733) 24x7 Rate 15999 with A/c Roo...
Get Premium Attur Layout Call Girls (8005736733) 24x7 Rate 15999 with A/c Roo...Get Premium Attur Layout Call Girls (8005736733) 24x7 Rate 15999 with A/c Roo...
Get Premium Attur Layout Call Girls (8005736733) 24x7 Rate 15999 with A/c Roo...MOHANI PANDEY
 
Call Girls Service Pune ₹7.5k Pick Up & Drop With Cash Payment 8005736733 Cal...
Call Girls Service Pune ₹7.5k Pick Up & Drop With Cash Payment 8005736733 Cal...Call Girls Service Pune ₹7.5k Pick Up & Drop With Cash Payment 8005736733 Cal...
Call Girls Service Pune ₹7.5k Pick Up & Drop With Cash Payment 8005736733 Cal...SUHANI PANDEY
 
VIP Model Call Girls Charholi Budruk ( Pune ) Call ON 8005736733 Starting Fro...
VIP Model Call Girls Charholi Budruk ( Pune ) Call ON 8005736733 Starting Fro...VIP Model Call Girls Charholi Budruk ( Pune ) Call ON 8005736733 Starting Fro...
VIP Model Call Girls Charholi Budruk ( Pune ) Call ON 8005736733 Starting Fro...SUHANI PANDEY
 
Kondhwa ( Call Girls ) Pune 6297143586 Hot Model With Sexy Bhabi Ready For ...
Kondhwa ( Call Girls ) Pune  6297143586  Hot Model With Sexy Bhabi Ready For ...Kondhwa ( Call Girls ) Pune  6297143586  Hot Model With Sexy Bhabi Ready For ...
Kondhwa ( Call Girls ) Pune 6297143586 Hot Model With Sexy Bhabi Ready For ...tanu pandey
 
RA 7942:vThe Philippine Mining Act of 1995
RA 7942:vThe Philippine Mining Act of 1995RA 7942:vThe Philippine Mining Act of 1995
RA 7942:vThe Philippine Mining Act of 1995garthraymundo123
 
celebrity 💋 Kanpur Escorts Just Dail 8250092165 service available anytime 24 ...
celebrity 💋 Kanpur Escorts Just Dail 8250092165 service available anytime 24 ...celebrity 💋 Kanpur Escorts Just Dail 8250092165 service available anytime 24 ...
celebrity 💋 Kanpur Escorts Just Dail 8250092165 service available anytime 24 ...Call Girls in Nagpur High Profile
 
Get Premium Hoskote Call Girls (8005736733) 24x7 Rate 15999 with A/c Room Cas...
Get Premium Hoskote Call Girls (8005736733) 24x7 Rate 15999 with A/c Room Cas...Get Premium Hoskote Call Girls (8005736733) 24x7 Rate 15999 with A/c Room Cas...
Get Premium Hoskote Call Girls (8005736733) 24x7 Rate 15999 with A/c Room Cas...MOHANI PANDEY
 
Introduction to heat waves and Heatwaves in Bangladesh.pptx
Introduction to heat waves and Heatwaves in Bangladesh.pptxIntroduction to heat waves and Heatwaves in Bangladesh.pptx
Introduction to heat waves and Heatwaves in Bangladesh.pptxNazmusSakibNS
 
Call Now ☎️🔝 9332606886 🔝 Call Girls ❤ Service In Muzaffarpur Female Escorts ...
Call Now ☎️🔝 9332606886 🔝 Call Girls ❤ Service In Muzaffarpur Female Escorts ...Call Now ☎️🔝 9332606886 🔝 Call Girls ❤ Service In Muzaffarpur Female Escorts ...
Call Now ☎️🔝 9332606886 🔝 Call Girls ❤ Service In Muzaffarpur Female Escorts ...Anamikakaur10
 
Call Girls Budhwar Peth Call Me 7737669865 Budget Friendly No Advance Booking
Call Girls Budhwar Peth Call Me 7737669865 Budget Friendly No Advance BookingCall Girls Budhwar Peth Call Me 7737669865 Budget Friendly No Advance Booking
Call Girls Budhwar Peth Call Me 7737669865 Budget Friendly No Advance Bookingroncy bisnoi
 
RATING SYSTEMS- IGBC, GRIHA, LEED--.pptx
RATING  SYSTEMS- IGBC, GRIHA, LEED--.pptxRATING  SYSTEMS- IGBC, GRIHA, LEED--.pptx
RATING SYSTEMS- IGBC, GRIHA, LEED--.pptxJIT KUMAR GUPTA
 
Alandi Road ( Call Girls ) Pune 6297143586 Hot Model With Sexy Bhabi Ready ...
Alandi Road ( Call Girls ) Pune  6297143586  Hot Model With Sexy Bhabi Ready ...Alandi Road ( Call Girls ) Pune  6297143586  Hot Model With Sexy Bhabi Ready ...
Alandi Road ( Call Girls ) Pune 6297143586 Hot Model With Sexy Bhabi Ready ...tanu pandey
 
Plastic Bank Beyond EPR - Sustainability Programs
Plastic Bank Beyond EPR - Sustainability ProgramsPlastic Bank Beyond EPR - Sustainability Programs
Plastic Bank Beyond EPR - Sustainability Programsitadmin50
 
DENR EPR Law Compliance Updates April 2024
DENR EPR Law Compliance Updates April 2024DENR EPR Law Compliance Updates April 2024
DENR EPR Law Compliance Updates April 2024itadmin50
 

Último (20)

VIP Model Call Girls Viman Nagar ( Pune ) Call ON 8005736733 Starting From 5K...
VIP Model Call Girls Viman Nagar ( Pune ) Call ON 8005736733 Starting From 5K...VIP Model Call Girls Viman Nagar ( Pune ) Call ON 8005736733 Starting From 5K...
VIP Model Call Girls Viman Nagar ( Pune ) Call ON 8005736733 Starting From 5K...
 
Green Marketing
Green MarketingGreen Marketing
Green Marketing
 
Get Premium Attur Layout Call Girls (8005736733) 24x7 Rate 15999 with A/c Roo...
Get Premium Attur Layout Call Girls (8005736733) 24x7 Rate 15999 with A/c Roo...Get Premium Attur Layout Call Girls (8005736733) 24x7 Rate 15999 with A/c Roo...
Get Premium Attur Layout Call Girls (8005736733) 24x7 Rate 15999 with A/c Roo...
 
Call Girls Service Pune ₹7.5k Pick Up & Drop With Cash Payment 8005736733 Cal...
Call Girls Service Pune ₹7.5k Pick Up & Drop With Cash Payment 8005736733 Cal...Call Girls Service Pune ₹7.5k Pick Up & Drop With Cash Payment 8005736733 Cal...
Call Girls Service Pune ₹7.5k Pick Up & Drop With Cash Payment 8005736733 Cal...
 
VIP Model Call Girls Charholi Budruk ( Pune ) Call ON 8005736733 Starting Fro...
VIP Model Call Girls Charholi Budruk ( Pune ) Call ON 8005736733 Starting Fro...VIP Model Call Girls Charholi Budruk ( Pune ) Call ON 8005736733 Starting Fro...
VIP Model Call Girls Charholi Budruk ( Pune ) Call ON 8005736733 Starting Fro...
 
Kondhwa ( Call Girls ) Pune 6297143586 Hot Model With Sexy Bhabi Ready For ...
Kondhwa ( Call Girls ) Pune  6297143586  Hot Model With Sexy Bhabi Ready For ...Kondhwa ( Call Girls ) Pune  6297143586  Hot Model With Sexy Bhabi Ready For ...
Kondhwa ( Call Girls ) Pune 6297143586 Hot Model With Sexy Bhabi Ready For ...
 
RA 7942:vThe Philippine Mining Act of 1995
RA 7942:vThe Philippine Mining Act of 1995RA 7942:vThe Philippine Mining Act of 1995
RA 7942:vThe Philippine Mining Act of 1995
 
(Anamika) VIP Call Girls Jammu Call Now 8617697112 Jammu Escorts 24x7
(Anamika) VIP Call Girls Jammu Call Now 8617697112 Jammu Escorts 24x7(Anamika) VIP Call Girls Jammu Call Now 8617697112 Jammu Escorts 24x7
(Anamika) VIP Call Girls Jammu Call Now 8617697112 Jammu Escorts 24x7
 
celebrity 💋 Kanpur Escorts Just Dail 8250092165 service available anytime 24 ...
celebrity 💋 Kanpur Escorts Just Dail 8250092165 service available anytime 24 ...celebrity 💋 Kanpur Escorts Just Dail 8250092165 service available anytime 24 ...
celebrity 💋 Kanpur Escorts Just Dail 8250092165 service available anytime 24 ...
 
Get Premium Hoskote Call Girls (8005736733) 24x7 Rate 15999 with A/c Room Cas...
Get Premium Hoskote Call Girls (8005736733) 24x7 Rate 15999 with A/c Room Cas...Get Premium Hoskote Call Girls (8005736733) 24x7 Rate 15999 with A/c Room Cas...
Get Premium Hoskote Call Girls (8005736733) 24x7 Rate 15999 with A/c Room Cas...
 
Deforestation
DeforestationDeforestation
Deforestation
 
Introduction to heat waves and Heatwaves in Bangladesh.pptx
Introduction to heat waves and Heatwaves in Bangladesh.pptxIntroduction to heat waves and Heatwaves in Bangladesh.pptx
Introduction to heat waves and Heatwaves in Bangladesh.pptx
 
Call Now ☎️🔝 9332606886 🔝 Call Girls ❤ Service In Muzaffarpur Female Escorts ...
Call Now ☎️🔝 9332606886 🔝 Call Girls ❤ Service In Muzaffarpur Female Escorts ...Call Now ☎️🔝 9332606886 🔝 Call Girls ❤ Service In Muzaffarpur Female Escorts ...
Call Now ☎️🔝 9332606886 🔝 Call Girls ❤ Service In Muzaffarpur Female Escorts ...
 
Call Girls Budhwar Peth Call Me 7737669865 Budget Friendly No Advance Booking
Call Girls Budhwar Peth Call Me 7737669865 Budget Friendly No Advance BookingCall Girls Budhwar Peth Call Me 7737669865 Budget Friendly No Advance Booking
Call Girls Budhwar Peth Call Me 7737669865 Budget Friendly No Advance Booking
 
RATING SYSTEMS- IGBC, GRIHA, LEED--.pptx
RATING  SYSTEMS- IGBC, GRIHA, LEED--.pptxRATING  SYSTEMS- IGBC, GRIHA, LEED--.pptx
RATING SYSTEMS- IGBC, GRIHA, LEED--.pptx
 
Alandi Road ( Call Girls ) Pune 6297143586 Hot Model With Sexy Bhabi Ready ...
Alandi Road ( Call Girls ) Pune  6297143586  Hot Model With Sexy Bhabi Ready ...Alandi Road ( Call Girls ) Pune  6297143586  Hot Model With Sexy Bhabi Ready ...
Alandi Road ( Call Girls ) Pune 6297143586 Hot Model With Sexy Bhabi Ready ...
 
Water Pollution
Water Pollution Water Pollution
Water Pollution
 
(INDIRA) Call Girl Katra Call Now 8617697112 Katra Escorts 24x7
(INDIRA) Call Girl Katra Call Now 8617697112 Katra Escorts 24x7(INDIRA) Call Girl Katra Call Now 8617697112 Katra Escorts 24x7
(INDIRA) Call Girl Katra Call Now 8617697112 Katra Escorts 24x7
 
Plastic Bank Beyond EPR - Sustainability Programs
Plastic Bank Beyond EPR - Sustainability ProgramsPlastic Bank Beyond EPR - Sustainability Programs
Plastic Bank Beyond EPR - Sustainability Programs
 
DENR EPR Law Compliance Updates April 2024
DENR EPR Law Compliance Updates April 2024DENR EPR Law Compliance Updates April 2024
DENR EPR Law Compliance Updates April 2024
 

Method 8260C by Purge and Trap Gas Chromatography Mass Spectrometry using the Clarus SQ 8

  • 1. Introduction U.S. EPA Method 8260C – Volatile Organic Compounds (VOCs) by Gas Chromatography Mass Spectrometry (GC/MS) is one of the most common environmental applications for GC/MS. This method outlines the analysis of volatile organic compounds in a variety of solid waste matrices including vari- ous air sampling trapping media, ground and surface water, soils, and sediments among others. The method requires not only demonstration of laboratory sample preparation and handling competence but instrument performance as well. The study presented here demonstrates the PerkinElmer® Clarus® SQ 8 GC/MS with purge and trap sample introduction both meets and exceeds the performance criteria set out in method 8260C and describes the analytical results and instrumental methodology. Experimental The PerkinElmer Clarus SQ 8C GC/MS operating in electron ionization mode with an Atomx purge and trap sample introduction system (Teledyne Tekmar, Mason, OH) was used to perform these experiments. The purge and trap conditions are presented in Table 1 and represent standard conditions for the analysis of method of VOCs by EPA Method 8260C. Gas Chromatography/ Mass Spectrometry a p p l i c a t i o n N o t e Authors Ruben Garnica Dawn May PerkinElmer, Inc. Shelton, CT USA Method 8260C by Purge and Trap Gas Chromatography Mass Spectrometry using the Clarus SQ 8
  • 2. 2 Table 1. Purge and Trap Instrument Conditions. Purge and Trap System: Tekmar – Atomx Trap Tekmar #9 trap Sample Size 5 mL Purge Parameters: Valve Oven Temp 140 °C Transfer Line Temp 140 °C Sample Mount Temp 90 °C Water Heater Temp 90 °C Sample Vial Temp 20 °C Sample Equilibrate Time 0.00 min Standby Flow 10 mL/min Purge Ready Temp 40 °C Condensate Ready Temp 45 °C Presweep Time 0.25 min Prime Sample Fill Volume 3.0 mL Sweep Sample Time 0.25 min Sweep Sample Flow 100 mL/min Sparge Vessel Heater On Purge Time 11.00 min Purge Flow 40 mL/min Purge Temp 40 °C Condensate Purge Temp 20 °C Dry Purge Time 2.00 min Dry Purge Flow 100 mL/min Dry Purge Temp 20 °C Desorb Parameters: Water Needle Rinse Volume 7.0 mL Sweep Needle Time 0.50 min Desorb Preheat Temp 200 °C GC Start Signal Start of Desorb Desorb Time 2.00 min Drain Flow 300 mL/min Desorb Temp 200 °C Bake Parameters: Number of Water Bake Rinses 2 Water Bake Rinse Volume 7.0 mL Bake Rinse Sweep Time 0.25 min Bake Rinse Sweep Flow 100 mL/min Bake Rinse Drain Time 0.40 min Bake Time 6.00 min Bake Flow 250 mL/min Bake Temp 270 °C Condensate Bake Temp 200 °C The Clarus SQ 8C GC/MS conditions are presented in Table 2. The heated transfer line of the concentrator was connected directly to the injector port of the Clarus 680 GC. A 2 mm i.d. injector port liner was used inside the capillary injector. Split flow adjustments will increase or decrease the split ratio and the resulting sensitivity of the system and allow the analysts flexibility in optimizing their GC/MS system. Table 2. Clarus SQ 8C GC/MS conditions. Gas Chromatograph: PerkinElmer Clarus 680 Analytical Column: Elite 624 MS (30 m x 0.25 mm i.d. x 1.4 μm) Injector Temperature: 220 °C Carrier Gas: Helium @ 1.0 mL/min Split Flow: 80 mL/min Oven Program: Temperature Hold Time Rate 40 °C 2 min 10 °C/min 100 °C 0 min 30 °C/min 200 °C 4 min End Mass Spectrometer: PerkinElmer Clarus SQ 8C GC Transfer Line Temperature: 220 °C Ion Source Temperature: 300 °C Function Type: Full Scan Solvent Delay: 0 – 0.50 min Scan Range: m/z 35 – 270 Scan Time: 0.20 sec Interscan Delay: 0.10 sec Calibration and performance standards were prepared from commercially available stock standards and diluted in Class-A volumetric flasks. Calibration standards were made from a combination of four multi-component standard mixes including a 76 component 8260 calibration mix, a 6 compo- nent 502.2 calibration gas mix, a 3 component VOA surrogate spike mix, and an 8260 internal standard mix. Stock stan- dards were diluted to an intermediate concentration from which initial calibration standards were prepared. The cali- bration levels used in this study are presented in Table 3. Standards used for the Method Detection Limit (MDL) and precision experiments were prepared at appropriate con- centration levels from intermediate standards. The internal standard mix was added to all samples such that a resulting concentration of 20 μg/L was achieved per 5 mL of sample. The tuning standard, 4-bromofluorobenzene, was prepared at 50 μg/L. Standards, stock standards, and associated QC/ QA samples were stored in a manner consistent with the guidelines set out in the method.
  • 3. 3 between analytical runs and in this study the injection-to- injection time, measured to be under 30 minutes, was limited by the purge and trap system. Table 4 presents the analytical results of these experiments including the Retention Time (RT), Average Relative Response Factors (Avg RRF) and Percent Relative Standard Deviation (% RSD) of the initial calibration, the Method Detection Limits (MDL), and the method precision as Percent Recovery (% Rec) and accuracy as % RSD. All of the analytes presented meet or exceed the minimum method requirements and boast excellent detection limits, precision and accuracy. Method Detection Limits were determined by analyzing seven replicate samples at a concentration of 0.5 μg/L and 1.0 µg/L. Precision and accuracy were measured by analyzing four replicates at a concentration of 25 μg/L. The majority of compounds in this report are calibrated over the entire concentration range as presented in Table 3. Compounds whose calibration range varied from the bulk of analytes are highlighted in Table 5. Alternate calibration ranges are expected for this method especially for compounds referred to as “bad actors”. In all cases, however, EPA method criteria are met. Overall improved results including MDL and precision and accuracy were obtained by operating the ion source at an elevated temperature than would regularly be used. Figure 2 shows a time selection of the seven TICs from the MDL study and illustrates the excellent repeatability of the system at a low concentration level. The inset presents the seven extracted ion chromatograms of naphthalene (EIC) at m/z = 128. The % RSD of the seven naphthalene EICs was measured to be 3.8%, which includes the sample handling performed by the purge and trap. Table 3. Calibration points employed in this study. Calibration Level Concentration (μg/L) 1 0.5 2 1.0 3 2.0 4 5.0 5 10 6 25 7 50 8 100 9 200 Results and Discussion The GC conditions were optimized to provide adequate separation while keeping the analytical runtime as short as possible. Figure 1 demonstrates a mid-level (25 μg/L) total ion chromatogram (TIC) from this analysis. All of the expected analyte separation is achieved while allowing for a maximum of throughput per unit laboratory time. The injection-to-injection time is further enhanced by the perfor- mance of the Clarus 680 GC oven. The oven incorporates a dual walled design with a unique air flow path to allow the fastest cooling times of any commercially available GC oven on the market. Air flow through the Clarus 680 GC is unidirectional eliminating the mixing of heated exhaust and cool ambient air allowing for the fastest transport of heat out of the GC oven. This high speed cooling allows the user to dramatically decrease the idle time of the instrument
  • 4. 4 Figure 1. TIC of 8260C volatile organic standard mix at 25 μg/L. Numbering references compound as listed in Table 4. Figure 2. Zoom of the seven MDL samples. TICs overlaid to illustrate the excellent repeatability achieved at low levels. Inset shows the naphthalene EIC at m/z = 128; the % RSD was measured to be 3.8%.
  • 5. 5 Table 4. Summarized initial calibration results of the analyzed VOCs from 0.5 – 200 μg/L for most compounds. A hyphen (-) indicates value either not provided by the method or not generated in this work. A plus sign (+) indicates calibration calculated using linear regression with R2 presented in table. 8260C 8260C Avg Minimum % RSD MDL Precision Accuracy # Compound RT RRF RRF % RSD Criteria (µg/L) (% Rec) (% RSD) 1 Dichlorodifluoromethane 1.38 0.44 0.1 14 20 0.14 90 2.9 2 Chloromethane+ 1.59 0.72 0.1 0.9992 0.990 0.15 96 4.9 3 Vinyl chloride 1.66 0.79 0.1 8.2 20 0.21 95 1.8 4 Bromomethane 2.00 0.26 0.1 18 20 0.18 105 4.1 5 Chloroethane 2.08 0.40 0.1 4.9 20 0.24 99 1.9 6 Trichlorofluoromethane 2.29 0.57 0.1 5.9 20 0.06 109 1.2 7 Diethyl ether 2.58 0.37 - 6.7 20 0.11 98 4.6 8 1,1,2-Trichloro-1,2,2-trifluor 2.78 0.38 0.1 13 20 0.15 99 1.1 9 1,1-Dichloroethene 2.82 0.44 0.1 15 20 0.30 102 1.8 10 Carbon disulfide 3.02 1.46 0.1 5.2 20 0.05 93 1.4 11 Allyl chloride+ 3.21 0.17 - 0.9940 0.990 0.05 76 2.7 12 Acetonitrile+ 3.29 0.20 - 0.9999 0.990 0.38 96 5.5 13 Methylene chloride 3.37 0.51 0.1 16 20 0.29 94 2.5 14 trans-1,2 Dichloroethene 3.59 0.46 0.1 9.3 20 0.34 105 1.3 15 Acrylonitrile 3.67 0.44 - 8.8 20 0.14 97 4.6 16 1,1,-Dichloroethane 4.06 0.98 0.2 9.2 20 0.16 107 2.3 17 Chloroprene 4.11 0.70 - 11 20 0.25 89 1.7 18 2,2,-Dichloropropane+ 4.63 0.15 - 0.9996 0.990 - 56 7.4 19 cis-1,2,-Dichloroethene 4.66 0.53 0.1 9.6 20 0.24 106 1.9 20 Methyl acrylate 4.76 0.92 - 4.2 20 0.14 100 5.3 21 Propionitrile 4.84 0.22 - 4.4 20 0.12 100 5.7 22 Tetrahydrofuran 4.92 0.53 - 13 20 0.15 86 4.7 23 Bromochloromethane 4.92 0.26 - 4.5 20 0.09 107 3.6 24 Methacrylonitrile 4.96 0.83 - 8.9 20 0.12 90 4.7 25 Chloroform 4.99 0.69 0.2 4.2 20 0.09 103 2.1 26 1,1,1-Trichloroethane 5.14 0.55 0.1 5.5 20 0.20 107 0.9 27 Pentafluorobenzene (ISTD) 5.22 28 Carbon tetrachloride 5.27 0.38 0.1 12 20 0.13 108 1.2 29 1,1-Dichloropropene 5.32 0.43 - 5.9 20 0.25 89 1.6 30 Isobutyl alcohol 5.49 0.09 - 11.5 20 0.22 89 5.6 31 Benzene 5.53 1.33 0.5 6.9 20 0.22 90 1.6 32 1,2-dichloroethane-d4 (Surr) 5.55 33 1,2-Dichloroethane 5.64 0.42 0.1 9.8 20 0.09 91 2.0 34 1,4-Difluorobenzene (ISTD) 6.00 35 Trichloroethene 6.25 0.32 0.2 8.9 20 0.28 103 2.5 36 1,2-Dichloropropane 6.56 0.47 0.1 6 20 0.15 96 1.8 37 Methyl methacrylate 6.67 0.39 - 6.2 20 0.17 91 3.9 38 Dibromomethane 6.70 0.19 - 4.9 20 0.10 106 2.7 39 Bromodichloromethane 6.87 0.28 0.2 7.4 20 0.17 102 1.9 40 2-Nitropropane+ 7.23 0.06 - 0.9996 0.990 - 94 4.7 41 cis-1,3-Dichloropropene 7.41 0.47 0.2 8.7 20 0.08 90 1.6 42 Toluene-d8 (Surr) 7.66
  • 6. 6 Table 4 continued 8260C 8260C Avg Minimum % RSD MDL Precision Accuracy # Compound RT RRF RRF % RSD Criteria (µg/L) (% Rec) (% RSD) 43 Toluene 7.74 0.67 0.4 6.5 20 0.22 100 1.5 44 trans-1,3-Dichloropropene 8.11 0.39 0.1 17 20 0.11 93 2.5 45 Ethyl methacrylate 8.16 0.66 - 11 20 0.07 101 2.6 46 1,1,2-Trichloroethane 8.31 0.28 0.1 4.3 20 0.06 96 1.9 47 Tetrachloroethene 8.35 0.43 0.2 20 20 0.17 123 4.6 48 1,3-Dichloropropane 8.50 0.65 - 4.4 20 0.06 94 3.1 49 Chlorodibromomethane 8.69 0.28 0.1 12 20 0.10 108 3.7 50 Ethylene dibromide 8.81 0.40 0.1 3.6 20 0.07 111 2.8 51 Chlorobenzene-d5 (ISTD) 9.21 52 Chlorobenzene 9.24 1.08 0.5 4.7 20 0.23 102 1.0 53 Ethylbenzene 9.31 1.34 0.1 2.2 20 0.20 104 1.4 54 1,1,1,2-Tetrachloroethane 9.31 0.27 - 7.8 20 0.19 105 1.9 55 p- & m-Xylene 9.41 1.09 0.1 4.4 20 0.23 101 1.6 56 o-Xylene 9.73 1.15 0.1 2.5 20 0.19 103 2.2 57 Styrene 9.76 1.09 0.3 3.8 20 0.18 106 1.8 58 Bromoform+ 9.91 0.22 0.1 0.9996 0.990 0.10 104 2.8 59 Isopropylbenzene 10.01 1.61 0.1 2.8 20 0.18 106 1.8 60 cis-1,4-Dichloro-2-butene+ 10.12 0.22 - 0.9986 0.990 0.09 68 2.2 61 Bromofluorobenzene (Surr) 10.15 62 Bromobenzene 10.25 0.98 - 8.5 20 0.15 98 2.3 63 1,1,2,2-Tetrachloroethane 10.28 1.17 0.3 5.9 20 0.16 93 9.9 64 n-Propylbenzene 10.30 3.21 - 4.8 20 0.20 101 1.1 65 1,2,3-Trichloropropane 10.32 1.76 - 5.8 20 0.10 91 3.3 66 trans-1,4-Dichloro-2-butene+ 10.33 0.25 - 0.9991 0.990 0.05 71 2.2 67 2-Chlorotoluene 10.38 1.88 - 4.4 20 0.19 102 0.9 68 1,3,5-Trimethylbenzene 10.43 3.00 - 6.3 20 0.21 104 1.1 69 4-Chlorotoluene 10.47 1.99 - 7.6 20 0.15 100 3.2 70 tert-Butylbenzene 10.64 2.57 - 9.2 20 0.18 94 1.3 71 1,2,4-Trimethylbenzene 10.69 2.94 - 4.8 20 0.19 105 1.2 72 sec-Butylbenzene 10.79 3.48 - 5.1 20 0.20 107 1.7 73 4-Isopropyltoluene 10.88 2.98 - 4.5 20 0.22 97 1.0 74 1,3-Dichlorobenzene 10.89 1.64 0.6 9 20 0.22 98 0.7 75 1,4-Dichlorobenzene-d4 (ISTD) 10.93 76 1,4-Dichlorobenzene 10.95 1.65 0.5 9.4 20 0.18 97 1.4 77 n-Butylbenzene 11.14 2.10 - 6.5 20 0.17 98 1.2 78 1,2-Dichlorobenzene 11.19 1.60 0.4 4.5 20 0.12 102 1.5 79 1,2-Dibromo-3-chloropropane 11.69 0.33 0.05 9.8 20 0.08 98 3.6 80 Nitrobenzene+ 11.84 0.06 - 0.9999 0.990 - 102 5.4 81 1,2,4-Trichlorobenzene 12.24 1.19 0.2 9.3 20 0.16 92 1.3 82 Hexachlorobutadiene 12.32 0.50 - 16 20 0.23 91 1.3 83 Naphthalene 12.46 3.60 - 7.3 20 0.06 102 2.4 84 1,2,3-Trichlorobenzene 12.65 1.13 0.2 7.2 20 0.09 94 1.6
  • 7. For a complete listing of our global offices, visit www.perkinelmer.com/ContactUs Copyright ©2012, PerkinElmer, Inc. All rights reserved. PerkinElmer® is a registered trademark of PerkinElmer, Inc. All other trademarks are the property of their respective owners. 010017B_01 PerkinElmer, Inc. 940 Winter Street Waltham, MA 02451 USA P: (800) 762-4000 or (+1) 203-925-4602 www.perkinelmer.com Table 5. Alternate calibration range compounds. Calibration Concentration Range Compound 0.5 – 100 μg/L Tetrachloroethene 1.0 – 200 μg/L Bromomethane Allyl chloride Acetonitrile Isobutyl alcohol 2.0 – 200 μg/L 2,2-Dichloropropane 2-Nitropropane Nitrobenzene 1.0 – 400 µg/L p- & m-Xylene Design advancements in the Clarus SQ 8 mass spectrometer allow such a wide operating range with the Clarifi™ detector providing much of the advantage. The detector’s enhanced sensitivity enables full scan mass spectrometry methods to be performed resulting in library searchable spectra at the lowest possible concentration levels. With the added sensitivity of SIFI™ (selected ion full ion) or SIM (selected ion monitoring) data acquisition at even lower limits are possible for specific compounds. The voltage setting on the Clarifi detector was important in developing the method for such a wide concentration range and deserves mention here. The UltraTune™ (Standard – DFTPP/BFB) function built into the TurboMass™ v6.0 software is designed to set the voltage such that the m/z = 69 peak of the tune gas FC43 is approximately 80% in the tune window. This setting is appropriate for general instrument analyses however users are encouraged to fine tune this setting to suit their specific analytical needs. In our case this was achieved by analyzing two samples, the high and low concentration level of our calibration, at voltage settings below that set during UltraTune. The optimum voltage setting produced measurable signal at the lowest level while not saturating the detector at the highest level. In our case the initial UltraTune setting was 1570V with test measure- ments performed at -50V and -100V of this value. The analysis was performed at 1520V. The stability of the Clarus SQ 8 system allows this setting to be used for extended periods of time and need only be reproduced periodically. The UltraTune (Standard – DFTPP/BFB) function also produces a satisfactory tune which meets the Tune Evaluation require- ments set out in the method. Figure 3 demonstrates the passing BFB Tune Evaluation Sample. Figure 4 shows the mass spectrum utilized in the Tune Evaluation Test and all of the necessary peak intensity comparisons are achieved. Conclusion The analysis of VOCs by purge and trap GC/MS following EPA Method 8260C is easily achievable over a wide con- centration range using the Clarus SQ 8 GC/MS system. Satisfactory analytical results were achieved for a concen- tration range from 0.5 – 200 μg/L for a majority of analyte compounds using the full scan approach providing library searchable spectra at all concentration levels. Excellent detection limits are reported as are accuracy and precision measurements. A number of technological advances make the Clarus SQ 8 GC/MS the ideal systems for laboratories wishing to perform high throughput and sensitivity analyses with an ease of operation currently unmatched. Figure 3. TurboMass 6.0 EPA BFB tune evaluation results. Figure 4. Mass spectrum of EPA BFB tune evaluation used for test in Figure 3.