Spectrophotometer
Specification
And
performance
Introduction
 What is spectrophotometer?
A spectrophotometer is a machine which measures the
intensity of light at selective wavelength.
 In clinical chemistry laboratory, we use spectrophotometer to
detect the concentration of a compound of interest in a
solution with the help of Beer-Lambert law by absorption
spectrophotometry.
The Beer-Lambert law-
• The Beer-Lambert law states that, the concentration of a
substance is directly proportional to the amount of radiant
energy absorbed or inversely proportional to the logarithm of
transmitted radiant energy.
Types of spectrophotometers
 Single-beam spectrophotometers.
 Double-beam spectrophotometers
1.double beam in space spectrophotometer.
2.double-beam in time spectrophotometer
In broad heading these specifications on the part of the machine
are-
 The incident radiant energy which reaches the cuvette must
be monochromatic and specific, i.e. of a fixed wavelength.
 Radiant energy should not be transmitted by any other
mechanism, i.e. there should not be any stray light.
 The photometer i.e. the detector & the read out device must
detect the intensity of light accurately.
Diagrammatic representation of some
spectrophotometers
Schematic representations of Single-beam
spectrophotometers
⃝→→→¦→→ →→¦→→→→→→→
Source entrance exit slit
light slit
monochromator cuvet detector read out device
Schematic representations of Double-beam
spectrophotometers.
Specifications
Parts
 1.A light source.
 2.A device to isolate light of a desired wave-lengths.
 3.A cuvette.
 4.A photodetector.
 5.A readout device and
 6.A data system.
1. Light source
Types of light sources used in spectrophotometer are-
 For visible wavelength range (380-750 nm) - like Tungsten
filament bulb, Argon fluoride LASER etc.
 For UV wavelength range i.e. <380 nm- like hydrogen lamp
and deuterium lamp, high pressure mercury lamp, mercury
arc lamp, gallium phosphate light-emitting diode etc.
 For Infra red wavelength range i.e. >750 nm –like CO2 laser,
Silicon light-emitting diode etc.
2.A device to isolate light of a desired wave-lengths.
 These are a device used for isolating and excluding radiant
energy of a desired wavelengths from undesired wavelengths.
Commonly used are-
 Simple filters- these are thin layer of colored glass.
Monochromators – these are-
 1.Prisms- separates white light into a continuous spectrum
because shorter wavelengths are bent or retracted.
 2.Diffraction gratings- these are prepared by depositing thin
layer of aluminum-copper on the surface of a flat glass plate.
Then ruling many small parallel grooves into the metal
coatings.
3.A cuvette.
 A cuvette is a small vessel used to hold a liquid sample to be
analyzed in the light path of a spectrophotometer.
 These are may be rounded, squared or rectangular shaped.
 Constructed from either glass or silica (quartz) or plastic.
 Most popular is 1 cm light path.
4. photodetector
 A photodetector is a device that converts light into an electrical
signal that is proportional to the number of photons striking it’s
photosensitive surface. A photomultiplier tube (PMT) is commonly
used for measuring light.
photomultiplier tube (PMT)
 PNT are members of the class of vacuum tubes, and more
specifically vacuum phototubes, are extremely sensitive detectors of
light in the ultraviolet, visible, and near-infrared ranges of the
electromagnetic spectrum.
 These detectors multiply the current produced by incident light by as
much as 100 million times in multiple dynode stages, enabling (for
example) individual photons to be detected when the incident flux of
light is very low.
photomultiplier tube (PMT) cont…
It consists of
 A cathode or called as photocathode
 Multiple dynodes - made by light-sensitive metal like lead
selenide, gallium arsenide, Indium antimonide etc.
 Anode.
 The whole system is kept in a vacuum container
Schematic diagram of a photomultiplier tube (PMT)
Photodetector cont…
How to work PMT-
 Voltage is applied between the dynodes and each successive
stage.
 The normal incremental increase in voltage at each dynodes is
from 50-100 V larger than that of previous dynode.
 Typically , a conventional PMT tube has approximately 1500 V
applied to it.
 When 1st photon strikes into the 1st photocathode it releases
excited electron , which are then strikes 2nd photocathode and
release more electrons and again these electrons strikes next
dynodes. This way signal multiplication occurs.
Schematic diagram of a photomultiplier tube (PMT
5.A readout device
 Electrical energy from a detector is displayed on some type of
meter or readout system
 Previously analog devices were used but now a days digital
readout devices which provide visual numeric display of
absorbance or converted value of concentrations.
6.A data system.
 Data system is usually used a computer.
 Here data are analyzed and stored.
Performance
 So it is very obvious from the discussions that , to measure the
exact concentration of a compound ,a spectrophotometer
must follow some specifications.
 So the sensitivity of response of a spectrophotometer results
from combined effects lamp output, efficiency of the filters /
monochromator in the transmissions of light and response of
the photometer detector.
Several analytical control checks should be performed
to certify that the machine is functioning within
desired specifications-
These includes-
 1.Wavelength accuracy
 2.Absence of stray light
 3.Linearity of the detector response.
 4.Photometric accuracy.
 5.multiple wavelength reading
1.Wavelength accuracy
To ensure that a spectrophotometer uses accurate
wavelength, two specific measures are used.
These are-
 Wavelength calibration and
 Use of spectral bandwidth.
 Wavelength calibration
 In narrow spectral bandwidth instruments, it is done with a
holminum oxide glass.
 Holmium oxide glass shows very sharp absorbance peaks at defined
wavelengths over the range of 280-650 nm.
 Thus the operator may compare the wavelength scale readings that
produce maximum absorbance with established values.
 If compared values do not coincides , a calibration correction table
can be constructed to relate scale readings to true wavelengths.
 The absorbance peaks for holmium oxide glass are typically seen at
279.3, 287.6, 333.8, 360.8. 418.5, 536.4, 637.5nm.
 Solutions of holmium oxide in dilute perchloric acid have also been
recommended and may be used with any spectrophotometer.
Spectral transmittance curve of holmium oxide filter
Wavelength calibration cont…
In broader-bandpass instruments wavelength calibrations is done
by using a didymium filter. The spectral transmittance curve for a
didymium filter -example -
Wavelength calibration cont…
• Didymium has several absorption peaks, the setting should be
verified grossly by visual examination of transmitted light. This
light should appear green at 530 nm.
 Use of spectral bandwidth.
 It is the range of wavelengths that reaches the exit slit of a
monochromator.
 It is usually referred to as the range of wavelengths, transmitted at
a point equal to half the peak intensity transmitted.
 Spectral bandwidth may also be calculated from manufacturer’s
specifications.
 Interference filters with spectral bandwidths of 1 to 2 nm are
available and may be used to check those instruments with a
spectral band-width of 8nm or more.
2.Absence of stray light
 Stray light is the radiant energy reaching the detector that consists
of wavelengths other than those defined by the filter /
monochromator.
 It is defined as a ratio or percentage to the total detected light.
 It is mainly due to scattering and diffraction within the
filter/monochromator.
 The effect of stray light on the performance of a
spectrophotometer is an absorbance error especially in the upper
end of absorbance range of the instrument.
 So in practice most of the spectrometer are equipped with stray
light filters.
 Thus a blue filter is used for wavelength setting below 400nm.
2.Absence of stray light cont…
For example
 when the machine is set for 350nm, most of the light is of
wavelengths of visible range.
 Here the blue filter absorbs most of the visible lights but
transmits only UV portion of the spectrums.
 Red filter is used for wavelengths in the range of 650-800 nm.
2.Absence of stray light cont…
Error due to other sources of unwanted light include-
 Light leaks from the cuvette and
 Fluorescence of the sample.
 Light leaks should be excluded by covering the cell
compartments.
 Light arising from fluorescence can increase the signal to the
detector and cause an apparent decrease in absorbance.
 These sources of light are not included in the usual definition
of stray light.
2.Absence of stray light cont…
How to detect stray light?
 Cutoff filters are satisfactory for detection of stray light.
 These may be of glass which produces a sharp cut in the
spectrum with complete absorbance on one side and high
transmittance to the other side.
 Liquid cutoff filters are satisfactory and convenient in the UV
range, where stray light is usually more of a problem.
 50gm/L aqueous solution of sodium nitrate should shows 0%
transmittance when read against water over 300-385nm.
 Acetone read against water , should show 0% T over the range of
250-320nm.
3.Linearity of the detector response.
 A properly functioning spectrophotometer must exhibit a
linear relationship between the radiant energy absorbed and
the instrument readout.
 Solid glass filters may be used to check the linearity of the
instrument.
 A very common method for certifying linearity of detector
response involve the use of solution of varying concentrations
of compounds which follow Beer’s law.
 Such compounds example ,p-nitrophenol at 450 nm,
oxyhaemoglobin at 415 nm, copper sulphate at 650 nm etc.
4.Photometric accuracy.
 The photometer must be able to measure the intensity of
transmitted light accurately in order to measure the
absorbance perfectly.
 Neutral density filters (SRM 1930) available from NIST are
used to check an instrument’s photometric accuracy.
 In practice solutions of potassium dichromate (K2Cr2O7) are
used for overall checks of photometric accuracy.
5.Multiple wavelength readings
 Background interference due to interfering chromogens can
often be eliminated or minimized by inclusion of blanks or by
reading by absorbance at two or three wavelengths.
 In one approach , termed bichromatic , absorbance is
measured at two wavelengths. One to peak absorbance and
another at a point near the base of the peak to serve as the
base line.
 The difference in absorbance at the two wavelengths is
related to concentrations.
Thank you

Más contenido relacionado

La actualidad más candente

Spectrophotometry
SpectrophotometrySpectrophotometry
Spectrophotometry
suniu
 
Ion selective electrodes(ise)
Ion selective electrodes(ise)Ion selective electrodes(ise)
Ion selective electrodes(ise)
Dilshad P.A.
 

La actualidad más candente (20)

Spetrophotometer
SpetrophotometerSpetrophotometer
Spetrophotometer
 
colorimetry spectrophotometry by dr.Tasnim
 colorimetry spectrophotometry by dr.Tasnim colorimetry spectrophotometry by dr.Tasnim
colorimetry spectrophotometry by dr.Tasnim
 
Colorimeter
ColorimeterColorimeter
Colorimeter
 
Photometry
PhotometryPhotometry
Photometry
 
Colorimetry
ColorimetryColorimetry
Colorimetry
 
Spectrophotometry
Spectrophotometry Spectrophotometry
Spectrophotometry
 
Spectrophotometry
SpectrophotometrySpectrophotometry
Spectrophotometry
 
Colorimetry & spectrophotometry
Colorimetry & spectrophotometryColorimetry & spectrophotometry
Colorimetry & spectrophotometry
 
electrochemiluminescence by Dr. Anurag Yadav
electrochemiluminescence by Dr. Anurag Yadavelectrochemiluminescence by Dr. Anurag Yadav
electrochemiluminescence by Dr. Anurag Yadav
 
Spectrophotometer
SpectrophotometerSpectrophotometer
Spectrophotometer
 
Nephelometry, turbidometry & Reflectance photometry by Dr. Anurag Yadav
Nephelometry, turbidometry & Reflectance photometry by Dr. Anurag YadavNephelometry, turbidometry & Reflectance photometry by Dr. Anurag Yadav
Nephelometry, turbidometry & Reflectance photometry by Dr. Anurag Yadav
 
Fundamentals of Spectrophotometer
Fundamentals of SpectrophotometerFundamentals of Spectrophotometer
Fundamentals of Spectrophotometer
 
Chemiluminescence, types of luminescent
Chemiluminescence, types of luminescent Chemiluminescence, types of luminescent
Chemiluminescence, types of luminescent
 
Ion selective electrodes(ise)
Ion selective electrodes(ise)Ion selective electrodes(ise)
Ion selective electrodes(ise)
 
Cellulose acetate electrophoresis
Cellulose acetate electrophoresisCellulose acetate electrophoresis
Cellulose acetate electrophoresis
 
colorimeter
colorimetercolorimeter
colorimeter
 
Radioisotope technique and methods
Radioisotope technique and methodsRadioisotope technique and methods
Radioisotope technique and methods
 
Chemiluminescence
ChemiluminescenceChemiluminescence
Chemiluminescence
 
Absorption Spectroscopy
Absorption SpectroscopyAbsorption Spectroscopy
Absorption Spectroscopy
 
Colorimeter
ColorimeterColorimeter
Colorimeter
 

Destacado

Spectrophotometry : Instruments & Applications
Spectrophotometry : Instruments & ApplicationsSpectrophotometry : Instruments & Applications
Spectrophotometry : Instruments & Applications
reyhane mazahernasab
 
Spectrophotometer
SpectrophotometerSpectrophotometer
Spectrophotometer
Asad Leo
 
ATOMIC ABSORPTION SPECTROSCOPY
ATOMIC ABSORPTION SPECTROSCOPYATOMIC ABSORPTION SPECTROSCOPY
ATOMIC ABSORPTION SPECTROSCOPY
Musa Khan
 
Bio chemistrytheory
Bio chemistrytheoryBio chemistrytheory
Bio chemistrytheory
XINYOUWANZ
 
Spectrophotometry.ppt 18.07.08
Spectrophotometry.ppt 18.07.08Spectrophotometry.ppt 18.07.08
Spectrophotometry.ppt 18.07.08
Amna Azam
 

Destacado (18)

Colorimeter and spectrophotometer, Mass Spectrometer
Colorimeter and spectrophotometer, Mass SpectrometerColorimeter and spectrophotometer, Mass Spectrometer
Colorimeter and spectrophotometer, Mass Spectrometer
 
Spectrophotometry : Instruments & Applications
Spectrophotometry : Instruments & ApplicationsSpectrophotometry : Instruments & Applications
Spectrophotometry : Instruments & Applications
 
spectrophotometer
spectrophotometerspectrophotometer
spectrophotometer
 
Chapter 19
Chapter 19Chapter 19
Chapter 19
 
Spectrophotometer-MiniScan EZ by HunterLab
Spectrophotometer-MiniScan EZ by HunterLabSpectrophotometer-MiniScan EZ by HunterLab
Spectrophotometer-MiniScan EZ by HunterLab
 
Spectrophotometer
SpectrophotometerSpectrophotometer
Spectrophotometer
 
Spectrophotometry
SpectrophotometrySpectrophotometry
Spectrophotometry
 
Spectrophotometer
SpectrophotometerSpectrophotometer
Spectrophotometer
 
Spectrophotometer
SpectrophotometerSpectrophotometer
Spectrophotometer
 
ATOMIC ABSORPTION SPECTROSCOPY
ATOMIC ABSORPTION SPECTROSCOPYATOMIC ABSORPTION SPECTROSCOPY
ATOMIC ABSORPTION SPECTROSCOPY
 
UV visible spectroscopy
UV visible spectroscopyUV visible spectroscopy
UV visible spectroscopy
 
Ft ir instrument
Ft ir instrumentFt ir instrument
Ft ir instrument
 
Ft ir
Ft irFt ir
Ft ir
 
Bio chemistrytheory
Bio chemistrytheoryBio chemistrytheory
Bio chemistrytheory
 
Infrared absorption spectroscopy
Infrared absorption spectroscopyInfrared absorption spectroscopy
Infrared absorption spectroscopy
 
SPECTROSCOPY
SPECTROSCOPYSPECTROSCOPY
SPECTROSCOPY
 
Spectrophotometry.ppt 18.07.08
Spectrophotometry.ppt 18.07.08Spectrophotometry.ppt 18.07.08
Spectrophotometry.ppt 18.07.08
 
IR and FT-IR Spectroscopy
IR and FT-IR SpectroscopyIR and FT-IR Spectroscopy
IR and FT-IR Spectroscopy
 

Similar a performance and specifications of spectrophotometer

Similar a performance and specifications of spectrophotometer (20)

Spectrophotometer introducing
Spectrophotometer  introducingSpectrophotometer  introducing
Spectrophotometer introducing
 
Spectroscopy.pptx
Spectroscopy.pptxSpectroscopy.pptx
Spectroscopy.pptx
 
Ultra violet spectrometer
Ultra violet spectrometerUltra violet spectrometer
Ultra violet spectrometer
 
UV-VIS SPECTROSCOPY
UV-VIS SPECTROSCOPYUV-VIS SPECTROSCOPY
UV-VIS SPECTROSCOPY
 
Instrumentation of uv visible spectroscopy
Instrumentation of uv visible spectroscopyInstrumentation of uv visible spectroscopy
Instrumentation of uv visible spectroscopy
 
Ultraviolet and visible spectroscopy
Ultraviolet and visible  spectroscopyUltraviolet and visible  spectroscopy
Ultraviolet and visible spectroscopy
 
INSTRUMENTATIONS UV-VIS.pdf
INSTRUMENTATIONS UV-VIS.pdfINSTRUMENTATIONS UV-VIS.pdf
INSTRUMENTATIONS UV-VIS.pdf
 
UV SPECTROSCOPY ppt.pptx
UV SPECTROSCOPY ppt.pptxUV SPECTROSCOPY ppt.pptx
UV SPECTROSCOPY ppt.pptx
 
UV SPECTROSCOPY ppt.pptx
UV SPECTROSCOPY ppt.pptxUV SPECTROSCOPY ppt.pptx
UV SPECTROSCOPY ppt.pptx
 
UV SPECTROSCOPY AND INSTRUMENTATION
UV SPECTROSCOPY AND INSTRUMENTATION UV SPECTROSCOPY AND INSTRUMENTATION
UV SPECTROSCOPY AND INSTRUMENTATION
 
UV-VISIBLE SPECTROSCOPY INSTRUMENTATION.pptx
UV-VISIBLE SPECTROSCOPY INSTRUMENTATION.pptxUV-VISIBLE SPECTROSCOPY INSTRUMENTATION.pptx
UV-VISIBLE SPECTROSCOPY INSTRUMENTATION.pptx
 
Uv vis spectroscopy- part-3
Uv vis spectroscopy- part-3Uv vis spectroscopy- part-3
Uv vis spectroscopy- part-3
 
Instrumentation of uv spectrophotometer pranav kumar singh
Instrumentation of uv spectrophotometer   pranav kumar singhInstrumentation of uv spectrophotometer   pranav kumar singh
Instrumentation of uv spectrophotometer pranav kumar singh
 
Uv/visible spectroscopy
Uv/visible spectroscopyUv/visible spectroscopy
Uv/visible spectroscopy
 
Uv
UvUv
Uv
 
uvvisiblespectroscopy.pptxhahahahahabhhs
uvvisiblespectroscopy.pptxhahahahahabhhsuvvisiblespectroscopy.pptxhahahahahabhhs
uvvisiblespectroscopy.pptxhahahahahabhhs
 
uvvisiblespectroscopy.pptxhahahahahabhhs
uvvisiblespectroscopy.pptxhahahahahabhhsuvvisiblespectroscopy.pptxhahahahahabhhs
uvvisiblespectroscopy.pptxhahahahahabhhs
 
PHOTOMETRY.pptx
PHOTOMETRY.pptxPHOTOMETRY.pptx
PHOTOMETRY.pptx
 
U.V Spectroscopy
U.V SpectroscopyU.V Spectroscopy
U.V Spectroscopy
 
Spectroscopy الشيت الثالث.pptx
Spectroscopy الشيت الثالث.pptxSpectroscopy الشيت الثالث.pptx
Spectroscopy الشيت الثالث.pptx
 

Último

1029 - Danh muc Sach Giao Khoa 10 . pdf
1029 -  Danh muc Sach Giao Khoa 10 . pdf1029 -  Danh muc Sach Giao Khoa 10 . pdf
1029 - Danh muc Sach Giao Khoa 10 . pdf
QucHHunhnh
 
1029-Danh muc Sach Giao Khoa khoi 6.pdf
1029-Danh muc Sach Giao Khoa khoi  6.pdf1029-Danh muc Sach Giao Khoa khoi  6.pdf
1029-Danh muc Sach Giao Khoa khoi 6.pdf
QucHHunhnh
 
Activity 01 - Artificial Culture (1).pdf
Activity 01 - Artificial Culture (1).pdfActivity 01 - Artificial Culture (1).pdf
Activity 01 - Artificial Culture (1).pdf
ciinovamais
 
Russian Escort Service in Delhi 11k Hotel Foreigner Russian Call Girls in Delhi
Russian Escort Service in Delhi 11k Hotel Foreigner Russian Call Girls in DelhiRussian Escort Service in Delhi 11k Hotel Foreigner Russian Call Girls in Delhi
Russian Escort Service in Delhi 11k Hotel Foreigner Russian Call Girls in Delhi
kauryashika82
 

Último (20)

Measures of Central Tendency: Mean, Median and Mode
Measures of Central Tendency: Mean, Median and ModeMeasures of Central Tendency: Mean, Median and Mode
Measures of Central Tendency: Mean, Median and Mode
 
How to Give a Domain for a Field in Odoo 17
How to Give a Domain for a Field in Odoo 17How to Give a Domain for a Field in Odoo 17
How to Give a Domain for a Field in Odoo 17
 
Introduction to Nonprofit Accounting: The Basics
Introduction to Nonprofit Accounting: The BasicsIntroduction to Nonprofit Accounting: The Basics
Introduction to Nonprofit Accounting: The Basics
 
1029 - Danh muc Sach Giao Khoa 10 . pdf
1029 -  Danh muc Sach Giao Khoa 10 . pdf1029 -  Danh muc Sach Giao Khoa 10 . pdf
1029 - Danh muc Sach Giao Khoa 10 . pdf
 
PROCESS RECORDING FORMAT.docx
PROCESS      RECORDING        FORMAT.docxPROCESS      RECORDING        FORMAT.docx
PROCESS RECORDING FORMAT.docx
 
Measures of Dispersion and Variability: Range, QD, AD and SD
Measures of Dispersion and Variability: Range, QD, AD and SDMeasures of Dispersion and Variability: Range, QD, AD and SD
Measures of Dispersion and Variability: Range, QD, AD and SD
 
Sociology 101 Demonstration of Learning Exhibit
Sociology 101 Demonstration of Learning ExhibitSociology 101 Demonstration of Learning Exhibit
Sociology 101 Demonstration of Learning Exhibit
 
This PowerPoint helps students to consider the concept of infinity.
This PowerPoint helps students to consider the concept of infinity.This PowerPoint helps students to consider the concept of infinity.
This PowerPoint helps students to consider the concept of infinity.
 
Z Score,T Score, Percential Rank and Box Plot Graph
Z Score,T Score, Percential Rank and Box Plot GraphZ Score,T Score, Percential Rank and Box Plot Graph
Z Score,T Score, Percential Rank and Box Plot Graph
 
On National Teacher Day, meet the 2024-25 Kenan Fellows
On National Teacher Day, meet the 2024-25 Kenan FellowsOn National Teacher Day, meet the 2024-25 Kenan Fellows
On National Teacher Day, meet the 2024-25 Kenan Fellows
 
1029-Danh muc Sach Giao Khoa khoi 6.pdf
1029-Danh muc Sach Giao Khoa khoi  6.pdf1029-Danh muc Sach Giao Khoa khoi  6.pdf
1029-Danh muc Sach Giao Khoa khoi 6.pdf
 
General Principles of Intellectual Property: Concepts of Intellectual Proper...
General Principles of Intellectual Property: Concepts of Intellectual  Proper...General Principles of Intellectual Property: Concepts of Intellectual  Proper...
General Principles of Intellectual Property: Concepts of Intellectual Proper...
 
ICT Role in 21st Century Education & its Challenges.pptx
ICT Role in 21st Century Education & its Challenges.pptxICT Role in 21st Century Education & its Challenges.pptx
ICT Role in 21st Century Education & its Challenges.pptx
 
Activity 01 - Artificial Culture (1).pdf
Activity 01 - Artificial Culture (1).pdfActivity 01 - Artificial Culture (1).pdf
Activity 01 - Artificial Culture (1).pdf
 
Role Of Transgenic Animal In Target Validation-1.pptx
Role Of Transgenic Animal In Target Validation-1.pptxRole Of Transgenic Animal In Target Validation-1.pptx
Role Of Transgenic Animal In Target Validation-1.pptx
 
ICT role in 21st century education and it's challenges.
ICT role in 21st century education and it's challenges.ICT role in 21st century education and it's challenges.
ICT role in 21st century education and it's challenges.
 
microwave assisted reaction. General introduction
microwave assisted reaction. General introductionmicrowave assisted reaction. General introduction
microwave assisted reaction. General introduction
 
Russian Escort Service in Delhi 11k Hotel Foreigner Russian Call Girls in Delhi
Russian Escort Service in Delhi 11k Hotel Foreigner Russian Call Girls in DelhiRussian Escort Service in Delhi 11k Hotel Foreigner Russian Call Girls in Delhi
Russian Escort Service in Delhi 11k Hotel Foreigner Russian Call Girls in Delhi
 
Class 11th Physics NEET formula sheet pdf
Class 11th Physics NEET formula sheet pdfClass 11th Physics NEET formula sheet pdf
Class 11th Physics NEET formula sheet pdf
 
Ecological Succession. ( ECOSYSTEM, B. Pharmacy, 1st Year, Sem-II, Environmen...
Ecological Succession. ( ECOSYSTEM, B. Pharmacy, 1st Year, Sem-II, Environmen...Ecological Succession. ( ECOSYSTEM, B. Pharmacy, 1st Year, Sem-II, Environmen...
Ecological Succession. ( ECOSYSTEM, B. Pharmacy, 1st Year, Sem-II, Environmen...
 

performance and specifications of spectrophotometer

  • 2. Introduction  What is spectrophotometer? A spectrophotometer is a machine which measures the intensity of light at selective wavelength.  In clinical chemistry laboratory, we use spectrophotometer to detect the concentration of a compound of interest in a solution with the help of Beer-Lambert law by absorption spectrophotometry.
  • 3. The Beer-Lambert law- • The Beer-Lambert law states that, the concentration of a substance is directly proportional to the amount of radiant energy absorbed or inversely proportional to the logarithm of transmitted radiant energy.
  • 4. Types of spectrophotometers  Single-beam spectrophotometers.  Double-beam spectrophotometers 1.double beam in space spectrophotometer. 2.double-beam in time spectrophotometer
  • 5. In broad heading these specifications on the part of the machine are-  The incident radiant energy which reaches the cuvette must be monochromatic and specific, i.e. of a fixed wavelength.  Radiant energy should not be transmitted by any other mechanism, i.e. there should not be any stray light.  The photometer i.e. the detector & the read out device must detect the intensity of light accurately.
  • 6. Diagrammatic representation of some spectrophotometers
  • 7. Schematic representations of Single-beam spectrophotometers ⃝→→→¦→→ →→¦→→→→→→→ Source entrance exit slit light slit monochromator cuvet detector read out device
  • 8.
  • 9. Schematic representations of Double-beam spectrophotometers.
  • 10.
  • 12. Parts  1.A light source.  2.A device to isolate light of a desired wave-lengths.  3.A cuvette.  4.A photodetector.  5.A readout device and  6.A data system.
  • 13. 1. Light source Types of light sources used in spectrophotometer are-  For visible wavelength range (380-750 nm) - like Tungsten filament bulb, Argon fluoride LASER etc.  For UV wavelength range i.e. <380 nm- like hydrogen lamp and deuterium lamp, high pressure mercury lamp, mercury arc lamp, gallium phosphate light-emitting diode etc.  For Infra red wavelength range i.e. >750 nm –like CO2 laser, Silicon light-emitting diode etc.
  • 14. 2.A device to isolate light of a desired wave-lengths.  These are a device used for isolating and excluding radiant energy of a desired wavelengths from undesired wavelengths. Commonly used are-  Simple filters- these are thin layer of colored glass. Monochromators – these are-  1.Prisms- separates white light into a continuous spectrum because shorter wavelengths are bent or retracted.  2.Diffraction gratings- these are prepared by depositing thin layer of aluminum-copper on the surface of a flat glass plate. Then ruling many small parallel grooves into the metal coatings.
  • 15. 3.A cuvette.  A cuvette is a small vessel used to hold a liquid sample to be analyzed in the light path of a spectrophotometer.  These are may be rounded, squared or rectangular shaped.  Constructed from either glass or silica (quartz) or plastic.  Most popular is 1 cm light path.
  • 16. 4. photodetector  A photodetector is a device that converts light into an electrical signal that is proportional to the number of photons striking it’s photosensitive surface. A photomultiplier tube (PMT) is commonly used for measuring light. photomultiplier tube (PMT)  PNT are members of the class of vacuum tubes, and more specifically vacuum phototubes, are extremely sensitive detectors of light in the ultraviolet, visible, and near-infrared ranges of the electromagnetic spectrum.  These detectors multiply the current produced by incident light by as much as 100 million times in multiple dynode stages, enabling (for example) individual photons to be detected when the incident flux of light is very low.
  • 17. photomultiplier tube (PMT) cont… It consists of  A cathode or called as photocathode  Multiple dynodes - made by light-sensitive metal like lead selenide, gallium arsenide, Indium antimonide etc.  Anode.  The whole system is kept in a vacuum container
  • 18. Schematic diagram of a photomultiplier tube (PMT)
  • 19. Photodetector cont… How to work PMT-  Voltage is applied between the dynodes and each successive stage.  The normal incremental increase in voltage at each dynodes is from 50-100 V larger than that of previous dynode.  Typically , a conventional PMT tube has approximately 1500 V applied to it.  When 1st photon strikes into the 1st photocathode it releases excited electron , which are then strikes 2nd photocathode and release more electrons and again these electrons strikes next dynodes. This way signal multiplication occurs.
  • 20. Schematic diagram of a photomultiplier tube (PMT
  • 21.
  • 22. 5.A readout device  Electrical energy from a detector is displayed on some type of meter or readout system  Previously analog devices were used but now a days digital readout devices which provide visual numeric display of absorbance or converted value of concentrations.
  • 23. 6.A data system.  Data system is usually used a computer.  Here data are analyzed and stored.
  • 25.  So it is very obvious from the discussions that , to measure the exact concentration of a compound ,a spectrophotometer must follow some specifications.  So the sensitivity of response of a spectrophotometer results from combined effects lamp output, efficiency of the filters / monochromator in the transmissions of light and response of the photometer detector.
  • 26. Several analytical control checks should be performed to certify that the machine is functioning within desired specifications- These includes-  1.Wavelength accuracy  2.Absence of stray light  3.Linearity of the detector response.  4.Photometric accuracy.  5.multiple wavelength reading
  • 27. 1.Wavelength accuracy To ensure that a spectrophotometer uses accurate wavelength, two specific measures are used. These are-  Wavelength calibration and  Use of spectral bandwidth.
  • 28.  Wavelength calibration  In narrow spectral bandwidth instruments, it is done with a holminum oxide glass.  Holmium oxide glass shows very sharp absorbance peaks at defined wavelengths over the range of 280-650 nm.  Thus the operator may compare the wavelength scale readings that produce maximum absorbance with established values.  If compared values do not coincides , a calibration correction table can be constructed to relate scale readings to true wavelengths.  The absorbance peaks for holmium oxide glass are typically seen at 279.3, 287.6, 333.8, 360.8. 418.5, 536.4, 637.5nm.  Solutions of holmium oxide in dilute perchloric acid have also been recommended and may be used with any spectrophotometer.
  • 29. Spectral transmittance curve of holmium oxide filter
  • 30. Wavelength calibration cont… In broader-bandpass instruments wavelength calibrations is done by using a didymium filter. The spectral transmittance curve for a didymium filter -example -
  • 31. Wavelength calibration cont… • Didymium has several absorption peaks, the setting should be verified grossly by visual examination of transmitted light. This light should appear green at 530 nm.
  • 32.  Use of spectral bandwidth.  It is the range of wavelengths that reaches the exit slit of a monochromator.  It is usually referred to as the range of wavelengths, transmitted at a point equal to half the peak intensity transmitted.  Spectral bandwidth may also be calculated from manufacturer’s specifications.  Interference filters with spectral bandwidths of 1 to 2 nm are available and may be used to check those instruments with a spectral band-width of 8nm or more.
  • 33. 2.Absence of stray light  Stray light is the radiant energy reaching the detector that consists of wavelengths other than those defined by the filter / monochromator.  It is defined as a ratio or percentage to the total detected light.  It is mainly due to scattering and diffraction within the filter/monochromator.  The effect of stray light on the performance of a spectrophotometer is an absorbance error especially in the upper end of absorbance range of the instrument.  So in practice most of the spectrometer are equipped with stray light filters.  Thus a blue filter is used for wavelength setting below 400nm.
  • 34. 2.Absence of stray light cont… For example  when the machine is set for 350nm, most of the light is of wavelengths of visible range.  Here the blue filter absorbs most of the visible lights but transmits only UV portion of the spectrums.  Red filter is used for wavelengths in the range of 650-800 nm.
  • 35. 2.Absence of stray light cont… Error due to other sources of unwanted light include-  Light leaks from the cuvette and  Fluorescence of the sample.  Light leaks should be excluded by covering the cell compartments.  Light arising from fluorescence can increase the signal to the detector and cause an apparent decrease in absorbance.  These sources of light are not included in the usual definition of stray light.
  • 36. 2.Absence of stray light cont… How to detect stray light?  Cutoff filters are satisfactory for detection of stray light.  These may be of glass which produces a sharp cut in the spectrum with complete absorbance on one side and high transmittance to the other side.  Liquid cutoff filters are satisfactory and convenient in the UV range, where stray light is usually more of a problem.  50gm/L aqueous solution of sodium nitrate should shows 0% transmittance when read against water over 300-385nm.  Acetone read against water , should show 0% T over the range of 250-320nm.
  • 37. 3.Linearity of the detector response.  A properly functioning spectrophotometer must exhibit a linear relationship between the radiant energy absorbed and the instrument readout.  Solid glass filters may be used to check the linearity of the instrument.  A very common method for certifying linearity of detector response involve the use of solution of varying concentrations of compounds which follow Beer’s law.  Such compounds example ,p-nitrophenol at 450 nm, oxyhaemoglobin at 415 nm, copper sulphate at 650 nm etc.
  • 38. 4.Photometric accuracy.  The photometer must be able to measure the intensity of transmitted light accurately in order to measure the absorbance perfectly.  Neutral density filters (SRM 1930) available from NIST are used to check an instrument’s photometric accuracy.  In practice solutions of potassium dichromate (K2Cr2O7) are used for overall checks of photometric accuracy.
  • 39. 5.Multiple wavelength readings  Background interference due to interfering chromogens can often be eliminated or minimized by inclusion of blanks or by reading by absorbance at two or three wavelengths.  In one approach , termed bichromatic , absorbance is measured at two wavelengths. One to peak absorbance and another at a point near the base of the peak to serve as the base line.  The difference in absorbance at the two wavelengths is related to concentrations.