2. Thermography Catheter
Basket
Over the wire catheter system
Pull back device with 0.5 mm/s
1 central blood
thermo couple
5 thermo couples
3.3 F catheter
0.02 o
C accuracy
7. Study Aim
Assessment of
• Temperature recording
• Steerability of catheter system
• QCA of coronary arteries
• Pull back recording, Endosonics device
• Practicability
• Safety
8. Clinical History of
Patients
• 12 patients included
• 9 male, 3 female
• 43 – 73 years
• 8 stable angina
• 4 unstable angina IIIb/IIIc
• 4 diabetes melitus
• 3 previous myocardial infarction
• 9 on aspirin, 9 on statine
27. LAD segment 7 lesion
After drug-
eluting stent
Implantation
29% stenosis
3 mm length
2.7 mm 2
pla-
que area
28. Patient 10
• Follow-up
no troponin elevation
o.o2 and o.o4 ng/ml (ref < 0.1)
myoglobin stable
57 and 59 ug/dl (ref < 90)
C-reactive protein
0.1 and 0.1 mg/dl (ref < 0.5)
29. Patient 11
• 43 years old man
• Anterior AMI August 29
peak CK 3286 U/l
treatment 50 mg tenecteplase
(Prof Jansen, Leverkusen)
• previous TIA April 01
• RF: hypercholesterinemia
smoking
190 cm, 98 kg, 2.3 m 2
BSA
• 2.1 mg/dl CRP
36. LAD segment 6 lesion
26% stenosis
11 mm length
0.7 mm 2
pla-
que area
Reperfusion
no - reflow
VT
UH
37. Patient 11
History of PCI
• Detection of hot spot pre PCI
• LAD 3.0 18 mm drug eluting stent
18 bar, post stent PTCA 4.0 mm 10 bar
• RCA 90% stenosis, additional PCI
3.0 13 mm drug eluting stent
• Lab tests: decline of CK, TnI and
myoglobin 69 to 123 and 86 ug/d (ref 90)
• CRP: 0.7 mg/dl (ref 0.5)
• Further course uneventful
38. Thermography Results
• 92 % success rate
• 35.91 – 36.95 o
C blood temperature
• Up to + 0.14 up to + 0.36 o
C in hot spots
• 4/ 11 temperature increase
• 4/ 11 focal hot spot
• 3/ 4 multifocal hot spots
• 1/ 4 MACE corresponding to CRP increase
• 3/ 4 hot spots but normal CRP
39. Future of Thermography
• Combination with IVUS
• Preinterventional scanning in order
to detect hot spots
for futher protective intervention
- aspiration systems or
- filter devices or
- graft stenting or
- cooling down
40.
41. Patient 9
• 61 year old woman
• Unstable angina type IIIB
• troponin I in the gray zone 0.14 ng/ml
• C-reactive protein 0.1 and 0.2 mg/dl
(ref. 0.1)
• Risk factors: diabetes,
hypertension
hypercholesterinemia
• Dialysis 1992 – 2001
• Renal transplantation
42. RCx Segment 11 lesion
70 % stenosis
14 mm length
19 mm 2
pla-
que area
HS