Common to see variability of 1 to 2 counts.
If variability is more than 10 counts, contact WL support.
If offset value is more than 20 counts from cal sheet value, contact WL support.
NTU offsets cannot be changed in 1.26 firmware or earlier versions as there are no equivalent commands
WQM Host 1.52 need to Enable ECO Setup under Help Drop down menu.
Additional supplies on top of: computer, power supply and test cable.
If pressure cannot be reset, there is an issue with pressure sensor. Either too much noise, broken sensor…
Should hear BLIS pump clicking.
If not, need to check if unit is powered up and plugged in.
Check for intake for blockage.
If it’s not empty by the 3rd run, it’s not working properly.
Performing cast once moored package is in place will give baseline and also prior to package recovery.