 According FDA Validation is
“Established documented evidence which provides a high
degree of assurance that a specific process will continually
produce a product meeting its predetermined specifications
and quality attributes.”
 The Validated method should be stability indicating method,
which means the method that can detect the changes in
properties of drug substance or drug product with time
2
 Verification
 Compendial methods-Verification :
• These methods are listed in pharmacopoeia’s
• Use to evaluate defined characteristics of the drug
substance or drug product
• Suitability of Compendial methods must be verified
under actual conditions
• Should be partially validated
 Validation
 Non -Compendial methods-Validation
• Proposed by the applicant for use instead of Regulatory
Analytical Procedure
• Equivalency:
• Equivalency: required for an in-house method, when compendial standard is claimed
3
Validation
Development
Quality Control
4
ICH/USP
GMPs
FDA
5
Verification data is required.
 The compendial methods as published are typically validated based on an API or an
FPP originating from a specific manufacturer.
 Different sources of the same API or FPP may contain impurities and/or degradation
products that were not considered during the development of the monograph.
 The FPP may also contain excipients not present in the FPP(s) for which the
compendial method was developed.
 API
Assay: No validation generally required. Exception: specificity for any specified
impurities not in the monograph.
Purity: Full validation for specified impurities that are not included in the monograph
(specificity, linearity, accuracy, repeatability, intermediate precision,
LOD/LOQ).
 FPP
Assay: Specificity, accuracy and precision (repeatability). Eg. to rule out excipient
interference
Purity: Specificity, accuracy and precision (repeatability) for known imps.
 Full validation for specified impurities that are not included in the
monograph (specificity, linearity, accuracy, repeatability, intermediate
precision, LOD/LOQ)
6
 Full validation is required for an in-house purity/assay method.
 Equivalency must be demonstrated compared to the compendial
method
• Assay: same samples using the two methods;
• Related substances: same samples spiking with specified
impurities at the limits) using the two methods
 If equivalency is not demonstrated, it can not be claimed as
compendial standard
7
Equivalency is essentially two things:
 Ensure the in-house method is not inferior (in which case they
should adopt the compendial method)
 Ensure the compendial test can be met, if applied.
 Two samples are sufficient.
Note that a compendial method becomes an in-house method when
adjustments are made outside the allowable adjustments (Int.Ph,
USP, EP general chapter Chromatography)
8
 Specificity
 Precision
 Accuracy
 Linearity
 Range
 Quantitation Limit
 Detection Limit
 Robustness
 Stability in Analytical Solution
9
USP ICH
 Specificity
 Linearity and Range
 Accuracy
 Precision
 Limit of Detection
 Limit of Quantitation
 Ruggedness
 Robustness
 Specificity
 Linearity
 Range
 Accuracy
 Precision
 Repeatability
 Intermediate Precision
 Reproducibility
 Limit of Detection
 Limit of Quantitation
10
Type of
Analytical
Procedure
Identification
Impurity testing
Assay
Quantitative Limit Tests
Accuracy No Yes No Yes
Precision
Repeatability No Yes No Yes
Interm. Prec. No Yes No Yes
Specificity Yes Yes Yes Yes
LOD No No Yes No
LOQ No Yes No No
Linearity No Yes No Yes
Range No Yes No Yes
May be required, depending on the nature of the specific test
11
Testing a method to demonstrate it is suitable for its intended purpose and
the results obtained are reliable, accurate and reproducible.
Provides confidence that the method will perform properly under intended
conditions.
 Identification tests;
 Quantitative or limit tests for the control of impurities;
 Quantitative tests of the active moiety in samples of drug substance or drug
product or other selected component(s) in the drug product. Including assay,
Content Uniformity, dissolution, content of perseveratives.
Validation of analytical procedures requires:
 Qualified and calibrated instruments
 Documented methods
 Reliable reference standards
 Qualified analysts
 Sample integrity
12
 Ability of an analytical method to measure the analyte free
from interference due to other components.
 An investigation of specificity should be conducted during
the validation of identification tests, the determination of
impurities and assay
 Should be able to discriminate between compounds closely
related in structure.
 Confirmed by obtaining negative results for samples with
spiked related compounds and positive results for samples
with analyte.
 SELECTIVITY
“A measure of the discriminating power of a given analytical
procedure in differentiating between the analyte and other
components in the test sample.”
13
 Impurities/Degradants Available
 Spike with impurities/degradants
 Show resolution and a lack of interference
 Impurities/Degradants Not Available
 Stress Samples
 For assay, Stressed and Unstressed Samples should be compared.
 For impurity test, impurity profiles should be compared.
 Peak homogeneity tests should be performed
using PDA or mass spectrometry to show that the
analyte chromatographic peak is not attributable to
more than one component.
14
 SELECTIVITY & SPECIFICITY
15
Selectivity: Specificity:
If an analytical procedure is able to
separate and resolve the various
components of a mixture and detect
the analyte qualitatively
If the method determines or
measures quantitatively the
component of interest in the sample
matrix without separation
Restricted to qualitative detection of
the components of a sample
Restricted to quantitative
measurement of one or more analyte
Blank solution - to show no interference
Placebo - to demonstrate the lack of interference from excipients
Spiked samples - to show that all known related substances are resolved from each
other
Stressed sample of about 5% to 20% degradation is used to demonstrate the
resolution between degradants and the analyte of interest
Check peak purity of drug substance by photodiode array detector (PDA)
Representative chromatograms should be provided
Temperature
 Humidity
 Acid Hydrolysis
 Base Hydrolysis
 Oxidation
 Light (UV/Vis/Fl)
Intent is to create 10 to 30 % Degradation
16
 The closeness of agreement (degree of scatter)
between a series of measurements obtained from
multiple samplings of the same homogeneous
sample.
Precision… Considered at 3 Levels
 Repeatability
 Intermediate Precision
 Reproducibility
17
 Repeatability
A minimum of nine determinations covering the specified
range for the procedure, e.g.
 three concentrations/three replicates each, or a
minimum of six determinations at 100% of the test
concentration
 Intermediate Precision
Within-laboratory variations
 usually on different days, different analysts and
different equipment.
 Reproducibility
Precision between laboratories
18
 Closeness of agreement between the value which is accepted either
as a conventional true value or an accepted reference value and the
value found
 Accuracy should be assessed using a
minimum of 9 determinations over a
minimum of 3 concentration levels covering
the specified range
 Accuracy is usually reported as percent
recovery by the assay (using the proposed
analytical procedure) of known added amount of
analyte in the sample or as the difference between the mean and
the accepted true value together with the confidence intervals. The
range for the accuracy limit should be within the linear range
19
 Drug Substance
For Assay
 Accuracy can be inferred once precision, linearity and
specificity have been established
For Related Substances/ Residual Solvents
 Accuracy should be assessed on samples spiked with
known amount of impurities
Drug Product
For Assay
 Known quantities of drug substance to be analysed have
been added to synthetic mixture of drug product
components.
For Related Substances/ Residual Solvents
 Accuracy should be assessed on samples spiked with
known amount of impurities
20
Accuracy range…..
Across the range of LOQ-150% of the target concentration
(shelf life limit), 3-5 concentrations, in triplicate each.
(LOQ, 50%, 100%, 150%)
Per cent recovery: in general, within 80-120%, depends on
the level of limit
Accurate &
precise
Accurate &
imprecise
Inaccurate &
precise
Inaccurate &
imprecise 21
 Linearity of an analytical procedure is its ability to obtain test results
which are directly proportional to concentration of analyte in the
sample within a specified range.
 A series of samples should be prepared in which the analyte
concentrations span the claimed range of the procedure. If there is a
linear relationship, test results should be evaluated by appropriate
statistical methods
 Minimum of 5 concentration is recommended within the range
Range:
This is an expression of the lowest and highest levels of analyte that
have been demonstrated to be determinable for the product
The specified range is normally derived from linearity studies
22
 Assay : 80-120% of the theoretical content of active
 Content Uniformity: 70-130%
 Dissolution: ±20% of limits; eg if limits cover from
20% to 90% l.c. (controlled release),
linearity should cover 0-110% of l.c.
 Impurities: reporting level to 120%
of shelf life limit
 Assay/Purity by a single method:
reporting level of the impurities to
120% of assay limit
23
 Is the amount of analyte in a sample which can be
quantitatively determined with suitable precision and
accuracy
 Approaches (instrumental or non-instrumental):
 visual evaluation (Used for non-instrumental methods)
 signal to noise ratio (10:1 for Quantitation Limit)
 standard deviation of the response and the slope
Quantitation Limit = 10
S
 : SD of Response (The residual SD of regression line or SD of y
intercept of regression lines can be used as SD)
S : Slope of calibration curve
24
Signal
Noise
Peak A
LOD
Peak B
LOQ
Baseline
Ybl
LOD LOQ
Statistical estimate of
LOD & LOQ
LOD = 3.3 Sbl / b LOQ = 10 Sbl / b
Y = b X
+ a
25
 Is the amount of analyte in a sample which can be
detected but not necessarily quantitated as an exact
value
 Approaches (instrumental or non-instrumental):
 visual evaluation (Used for non-instrumental methods)
 signal to noise ratio (3:1 for Detection Limit)
 standard deviation of the response and the slope
Quantitation Limit = 3.3
S
 : SD of Response (The residual SD of regression line or SD of y
intercept of regression lines can be used as SD)
S : Slope of calibration curve
26
 Not required for assay/dissolution methods.
 Applicant should provide
 The method of determination
 The limits
 Chromatograms
 Visually identified
 Established by Repeatability
27
 Measure of its capacity to remain unaffected by small ,
but deliberate variations in method parameters and
provide an indication of its reliability during normal
usage
 If measurements are susceptible to variations in
analytical conditions, than it should be suitably
controlled or a precautionary statement should be
included in procedure
 System suitability parameters should be evaluated to
ensure that validity of analytical procedure is
maintained when ever used
28
 For Liquid Chromatography, typical variations are
 Influence of pH in a mobile phase ( ± 0.2 units)
 Influence of mobile phase composition (± 2.0% Organic phase)
 Different columns[ Lots/Suppliers]
 Temperature [Column Oven] (± 5° C)
 Flow rate ( ± 0.2ml/min)
 System Suitability
 Evaluate the System suitability parameters
 Validation of analytical method must be focused on critical separation
of components which can alter during change of conditions
 These critically separated components should be part of system
suitability
29
 Should be established
 By keeping the sample at RT or below 10°C and
injecting it at regular intervals
 No significant variations or trend should be observed
 Any significant variation in particular area / absorbance
should be compared or calculated against the initial value.
 In case of impurity the % of impurity present should be
verified against the initial .
 The outcome of the result should be mentioned in the
particular regular test procedure.
30
When Re- validation is required !!!
 Method parameters have been changed
 The scope of the method has been changed
 Synthetic methods have been changed
 Impurity profile has been changed (Due to change in
excipient).
 Changes in equipment or suppliers of critical supplies of
the API or final drug product will have the potential to
change their degradation profile and may require the
method to be redeveloped and revalidated.
31
 Integral part of many analytical procedures
 Based on the concept that:
 The equipment, electronics, analytical operations and
samples to be analyzed constitute an integral system
that can be evaluated as such
 Determination (in General): Repeatability, Tailing factor
(T), Capacity factor (k’), Resolution (R), and Theoretical
Plates (N)
32
Parameters Recommendations
K’ In general k’ ≥ 2.0
R
R > 2, between the peak of interest and the
closest potential interferent (degradant,
internal STD, impurity, excipient, etc…..)
T T ≤ 2
N In general N > 2000
Repeatability
RSD ≤ 2.0% (n ≥ 5)
Assay: RSD ≤1% (API) or ≤ 2% (FPP), n ≥ 5
Impurities: in general, RSD ≤ 5% at the limit level,
up to 10% or higher at LOQ, n ≥ 6 33
Allowed HPLC Adjustment
USP (Ref: General Chapter <621>) EP (Ref: General Chapter 2.2.46)
Column Length ±70% ±70%
Internal Diameter
Can be adjusted if linear velocity is
kept constant ±25%
Particle Size Reduction of 50%, no increase Reduction of 50%, no increase
Flow Rate
±50% or more as long as the linear
velocity is kept contant ±50%
Column Temperature ±10C ±10% Max 60C
Injection Volume
Change allowed as long as SST
criteria are met
May be decreased (if LOD and
repeatability ok)
PH ±0.2 Units
±0.2 Units (±1% for neutral
substances)
UV wavelength <±3 nm (in condition) <±3 nm ((in condition))
Conc. Salts in Buffer ±10% ±10%
Composition of mobile phase
Minor Components (<50%) ±30%
or ±10% absolute whichever is
smaller
Minor Components ±30% or ±2%
absolute whichever is larger
34
 Some 483 Observations
 Analytical methods are not adequately
validated
 There was no inadequate method validation
specificity data to demonstrate that each
method was capable of distinguishing the
active ingredient from its impurities and degradation products.
 Specificity studies did not include the minimum stress conditions of
acid and base hydrolysis, oxidation, thermal degradation and
photolysis, degradation schematic for the active ingredient that
identifies the major degradation products was not included for
each product.
35
 Some 483 Observations
 Stress studies conducted as part of method validation do
not target a minimum amount of degradation. … a
standard period of two hours as commonly used for stress
studies with no justification…
 Spreadsheets used to calculate linearity, percent recovery,
and final assay results for the cleaning validation of
…were not validated and the data transcribed from
chromatographs to the spreadsheets were not checked for
accuracy.
36
 FDA Waning Letter
 On addition to the example
of modifying both Compendial methods and
customer supplied methods, we also observed
the use of un-validated in-house methods as
well as un-validated modifications to in-house
methods.
 A statement indicating that the method has not been validated in the
particular formulation was included in the certificate of analysis
for…use of this statement does not absolve…from using valid,
accurate, and reproducible methods.
 OOS accuracy results reported by analyst 3 were never submitted in
the final report. Repeat analysis performed in a different system
passed specifications and these results were submitted in the report.
37
 FDA Warning Letter
 There is no assurance that qualification or maintenance of
the laboratory equipment can consistently produce valid
and accurate analytical results in that numerous examples
of test data were invalidated due to instrument
malfunction.
 Attempts to corroborate data in the validation report with
supporting raw data in the laboratory were difficult and
frustrating for the FDA personnel conducting the
inspection
 Raw data and calculations were not checked by a second
responsible individuals required by your procedures.
Inaccurate calculations were noted in the report 38
 FDA Warning Letter
 The process validation samples were
assays using an HPLC method that had not been
validated. The method validation used for both products,
did not include a protocol that included specification and
acceptance criteria. … The method validation was not
reviewed and approved until during the current inspection.
Lots of both products were released for distribution prior
to completion of the method validation.
 Method validation for the product one product is
inadequate in that the data does not assess all variables,
such as different mobile phase concentrations and
analytes, to demonstrate that the method can sustain
variance.
39
 Data for acceptance, release, stability will only be trustworthy
if the methods used are reliable
 More efficient use of resources in the R&D process, as a
consequence
 Validation should be Protocol driven
 Document each and every information with regard to
std/sample/reagent/impurities/process/ preparation etc.
 Document the details of each experiment with suitable way.
 Review each experiment after its completion and relate its
outcome with other experiments
 Identification and elimination of the need to duplicate studies
to meet different regulatory requirements
40
 From My Side……
 Why is analytical monitoring necessary?
 What is the purpose of analytical validation?
41
42

Más contenido relacionado

La actualidad más candente

analytical method validation and validation of hplc
analytical method validation and validation of hplcanalytical method validation and validation of hplc
analytical method validation and validation of hplcvenkatesh thota
 
IMPURITIES OF NEW DRUG PRODUCTS
IMPURITIES OF NEW DRUG PRODUCTS IMPURITIES OF NEW DRUG PRODUCTS
IMPURITIES OF NEW DRUG PRODUCTS prakash64742
 
Analytical Method Validation as per ICH vs USP
Analytical Method Validation as per ICH vs USPAnalytical Method Validation as per ICH vs USP
Analytical Method Validation as per ICH vs USPKushal Shah
 
Stability testing of biotechnological/ biological products (Q5C )
Stability testing of biotechnological/ biological products (Q5C )Stability testing of biotechnological/ biological products (Q5C )
Stability testing of biotechnological/ biological products (Q5C )UshaKhanal3
 
Analytical method validation
Analytical method validationAnalytical method validation
Analytical method validationSHIVANEE VYAS
 
Related Substances-Method Validation-PPT_slide
Related Substances-Method Validation-PPT_slideRelated Substances-Method Validation-PPT_slide
Related Substances-Method Validation-PPT_slideBhanu Prakash N
 
Analytical Method Validation basics by Dr. A. Amsavel
Analytical Method Validation  basics by Dr. A. AmsavelAnalytical Method Validation  basics by Dr. A. Amsavel
Analytical Method Validation basics by Dr. A. AmsavelDr. Amsavel A
 
Q3C GUIDELINE FOR RESIDUAL SOLVENTS
Q3C  GUIDELINE FOR RESIDUAL SOLVENTSQ3C  GUIDELINE FOR RESIDUAL SOLVENTS
Q3C GUIDELINE FOR RESIDUAL SOLVENTSMuhamad Abdalkader
 
Ich guidelines for stability testing of biotechnological biological products (1)
Ich guidelines for stability testing of biotechnological biological products (1)Ich guidelines for stability testing of biotechnological biological products (1)
Ich guidelines for stability testing of biotechnological biological products (1)Dr Raj kumar Kudari
 
Analytical Method Validation
Analytical Method Validation Analytical Method Validation
Analytical Method Validation Vivek Jain
 
ICH Q3B (R2):Impurities in new drug products
ICH Q3B (R2):Impurities in new drug products ICH Q3B (R2):Impurities in new drug products
ICH Q3B (R2):Impurities in new drug products Vinit Gohel
 
Analytical methods,cleaning validation
Analytical methods,cleaning validationAnalytical methods,cleaning validation
Analytical methods,cleaning validationSai Vivek Kosaraju
 
Analytical method validation as per ich and usp
Analytical method validation as per ich and usp Analytical method validation as per ich and usp
Analytical method validation as per ich and usp shreyas B R
 
Radio Immuno Assay
Radio Immuno Assay Radio Immuno Assay
Radio Immuno Assay Neha Suresh
 
Analytical method- Content, Development, validation, Transfer & Life Cycle Ma...
Analytical method- Content, Development, validation, Transfer & Life Cycle Ma...Analytical method- Content, Development, validation, Transfer & Life Cycle Ma...
Analytical method- Content, Development, validation, Transfer & Life Cycle Ma...Md. Mizanur Rahman Miajee
 

La actualidad más candente (20)

Analytical method validation
Analytical method validation Analytical method validation
Analytical method validation
 
analytical method validation and validation of hplc
analytical method validation and validation of hplcanalytical method validation and validation of hplc
analytical method validation and validation of hplc
 
Analytical Method Validation
Analytical Method ValidationAnalytical Method Validation
Analytical Method Validation
 
IMPURITIES OF NEW DRUG PRODUCTS
IMPURITIES OF NEW DRUG PRODUCTS IMPURITIES OF NEW DRUG PRODUCTS
IMPURITIES OF NEW DRUG PRODUCTS
 
Analytical Method Validation as per ICH vs USP
Analytical Method Validation as per ICH vs USPAnalytical Method Validation as per ICH vs USP
Analytical Method Validation as per ICH vs USP
 
Stability testing of biotechnological/ biological products (Q5C )
Stability testing of biotechnological/ biological products (Q5C )Stability testing of biotechnological/ biological products (Q5C )
Stability testing of biotechnological/ biological products (Q5C )
 
Qualification of glassware
Qualification of glasswareQualification of glassware
Qualification of glassware
 
Analytical method validation
Analytical method validationAnalytical method validation
Analytical method validation
 
Qualification Of HPLC
Qualification Of HPLCQualification Of HPLC
Qualification Of HPLC
 
Related Substances-Method Validation-PPT_slide
Related Substances-Method Validation-PPT_slideRelated Substances-Method Validation-PPT_slide
Related Substances-Method Validation-PPT_slide
 
Analytical Method Validation basics by Dr. A. Amsavel
Analytical Method Validation  basics by Dr. A. AmsavelAnalytical Method Validation  basics by Dr. A. Amsavel
Analytical Method Validation basics by Dr. A. Amsavel
 
Q3C GUIDELINE FOR RESIDUAL SOLVENTS
Q3C  GUIDELINE FOR RESIDUAL SOLVENTSQ3C  GUIDELINE FOR RESIDUAL SOLVENTS
Q3C GUIDELINE FOR RESIDUAL SOLVENTS
 
Ich guidelines for stability testing of biotechnological biological products (1)
Ich guidelines for stability testing of biotechnological biological products (1)Ich guidelines for stability testing of biotechnological biological products (1)
Ich guidelines for stability testing of biotechnological biological products (1)
 
Analytical Method Validation
Analytical Method Validation Analytical Method Validation
Analytical Method Validation
 
ICH Q3B (R2):Impurities in new drug products
ICH Q3B (R2):Impurities in new drug products ICH Q3B (R2):Impurities in new drug products
ICH Q3B (R2):Impurities in new drug products
 
Qualification of Gas Chromatography
Qualification of Gas Chromatography Qualification of Gas Chromatography
Qualification of Gas Chromatography
 
Analytical methods,cleaning validation
Analytical methods,cleaning validationAnalytical methods,cleaning validation
Analytical methods,cleaning validation
 
Analytical method validation as per ich and usp
Analytical method validation as per ich and usp Analytical method validation as per ich and usp
Analytical method validation as per ich and usp
 
Radio Immuno Assay
Radio Immuno Assay Radio Immuno Assay
Radio Immuno Assay
 
Analytical method- Content, Development, validation, Transfer & Life Cycle Ma...
Analytical method- Content, Development, validation, Transfer & Life Cycle Ma...Analytical method- Content, Development, validation, Transfer & Life Cycle Ma...
Analytical method- Content, Development, validation, Transfer & Life Cycle Ma...
 

Similar a Analytical Method Validation

To perform Analytical method validation of Paracetamol Tablets by UV-spectrop...
To perform Analytical method validation of Paracetamol Tablets by UV-spectrop...To perform Analytical method validation of Paracetamol Tablets by UV-spectrop...
To perform Analytical method validation of Paracetamol Tablets by UV-spectrop...Aakashdeep Raval
 
Analytical method validation
Analytical method validationAnalytical method validation
Analytical method validationSai Praveen Reddy
 
Analytical method validation
Analytical method validationAnalytical method validation
Analytical method validationSUBHASISH DAS
 
Analytical method validation
Analytical method validationAnalytical method validation
Analytical method validationAJAYKUMAR4872
 
analytical method validation.pptx
analytical method validation.pptxanalytical method validation.pptx
analytical method validation.pptxLaxmidharSahoo11
 
ANALYTICAL METHOD VALIDATION (PHARMACEUTICAL INDUSTRY).pdf
ANALYTICAL METHOD VALIDATION (PHARMACEUTICAL INDUSTRY).pdfANALYTICAL METHOD VALIDATION (PHARMACEUTICAL INDUSTRY).pdf
ANALYTICAL METHOD VALIDATION (PHARMACEUTICAL INDUSTRY).pdfBALASUNDARESAN M
 
Validation of Analytical Procedures.pdf
Validation of Analytical Procedures.pdfValidation of Analytical Procedures.pdf
Validation of Analytical Procedures.pdfpharma344
 
Ich guidelines for validation final
Ich guidelines for validation finalIch guidelines for validation final
Ich guidelines for validation finalsumel ashique
 
All you know About Analytical method validation
All you  know About Analytical method validation All you  know About Analytical method validation
All you know About Analytical method validation CynthiaCynthiaPLegro
 
Analytical method Validation
Analytical method ValidationAnalytical method Validation
Analytical method Validationprashik shimpi
 
Validation of Analytical Methods.pdf
Validation of Analytical Methods.pdfValidation of Analytical Methods.pdf
Validation of Analytical Methods.pdfanjaneyulu49
 
Ich guidelines on validation for analytical method/equipments
Ich guidelines on validation for analytical method/equipmentsIch guidelines on validation for analytical method/equipments
Ich guidelines on validation for analytical method/equipmentssakshi singh
 
validation of analytical procedure USFDA Guidline
validation of analytical procedure USFDA Guidlinevalidation of analytical procedure USFDA Guidline
validation of analytical procedure USFDA GuidlineArchana Chavhan
 
Bioanalytical method validation guidelines for ligand binding assays
Bioanalytical method validation guidelines for ligand binding assaysBioanalytical method validation guidelines for ligand binding assays
Bioanalytical method validation guidelines for ligand binding assaysSHANE_LOBO145
 
Analytical Methods Validation| PV M Pharmacy QA
Analytical Methods Validation| PV M Pharmacy QAAnalytical Methods Validation| PV M Pharmacy QA
Analytical Methods Validation| PV M Pharmacy QAGhanshyam Nawale
 
INSTRUMENTAL ANALYSIS INTRODUCTION
INSTRUMENTAL ANALYSIS INTRODUCTIONINSTRUMENTAL ANALYSIS INTRODUCTION
INSTRUMENTAL ANALYSIS INTRODUCTIONHamunyare Ndwabe
 

Similar a Analytical Method Validation (20)

To perform Analytical method validation of Paracetamol Tablets by UV-spectrop...
To perform Analytical method validation of Paracetamol Tablets by UV-spectrop...To perform Analytical method validation of Paracetamol Tablets by UV-spectrop...
To perform Analytical method validation of Paracetamol Tablets by UV-spectrop...
 
Validation parameters
Validation parametersValidation parameters
Validation parameters
 
Validation methods.pptx
Validation methods.pptxValidation methods.pptx
Validation methods.pptx
 
Analytical method validation
Analytical method validationAnalytical method validation
Analytical method validation
 
Validation my ppt
Validation my pptValidation my ppt
Validation my ppt
 
Analytical method validation
Analytical method validationAnalytical method validation
Analytical method validation
 
Analytical method validation
Analytical method validationAnalytical method validation
Analytical method validation
 
analytical method validation.pptx
analytical method validation.pptxanalytical method validation.pptx
analytical method validation.pptx
 
ANALYTICAL METHOD VALIDATION (PHARMACEUTICAL INDUSTRY).pdf
ANALYTICAL METHOD VALIDATION (PHARMACEUTICAL INDUSTRY).pdfANALYTICAL METHOD VALIDATION (PHARMACEUTICAL INDUSTRY).pdf
ANALYTICAL METHOD VALIDATION (PHARMACEUTICAL INDUSTRY).pdf
 
Validation of Analytical Procedures.pdf
Validation of Analytical Procedures.pdfValidation of Analytical Procedures.pdf
Validation of Analytical Procedures.pdf
 
Ich guidelines for validation final
Ich guidelines for validation finalIch guidelines for validation final
Ich guidelines for validation final
 
Validation
Validation Validation
Validation
 
All you know About Analytical method validation
All you  know About Analytical method validation All you  know About Analytical method validation
All you know About Analytical method validation
 
Analytical method Validation
Analytical method ValidationAnalytical method Validation
Analytical method Validation
 
Validation of Analytical Methods.pdf
Validation of Analytical Methods.pdfValidation of Analytical Methods.pdf
Validation of Analytical Methods.pdf
 
Ich guidelines on validation for analytical method/equipments
Ich guidelines on validation for analytical method/equipmentsIch guidelines on validation for analytical method/equipments
Ich guidelines on validation for analytical method/equipments
 
validation of analytical procedure USFDA Guidline
validation of analytical procedure USFDA Guidlinevalidation of analytical procedure USFDA Guidline
validation of analytical procedure USFDA Guidline
 
Bioanalytical method validation guidelines for ligand binding assays
Bioanalytical method validation guidelines for ligand binding assaysBioanalytical method validation guidelines for ligand binding assays
Bioanalytical method validation guidelines for ligand binding assays
 
Analytical Methods Validation| PV M Pharmacy QA
Analytical Methods Validation| PV M Pharmacy QAAnalytical Methods Validation| PV M Pharmacy QA
Analytical Methods Validation| PV M Pharmacy QA
 
INSTRUMENTAL ANALYSIS INTRODUCTION
INSTRUMENTAL ANALYSIS INTRODUCTIONINSTRUMENTAL ANALYSIS INTRODUCTION
INSTRUMENTAL ANALYSIS INTRODUCTION
 

Último

Rishikesh Call Girls Service 6398383382 Real Russian Girls Looking Models
Rishikesh Call Girls Service 6398383382 Real Russian Girls Looking ModelsRishikesh Call Girls Service 6398383382 Real Russian Girls Looking Models
Rishikesh Call Girls Service 6398383382 Real Russian Girls Looking ModelsRupali Sharma
 
Escorts Lahore || 🔞 03274100048 || Escort service in Lahore
Escorts Lahore || 🔞 03274100048 || Escort service in LahoreEscorts Lahore || 🔞 03274100048 || Escort service in Lahore
Escorts Lahore || 🔞 03274100048 || Escort service in LahoreDeny Daniel
 
Top 20 Famous Indian Female Pornstars Name List 2024
Top 20 Famous Indian Female Pornstars Name List 2024Top 20 Famous Indian Female Pornstars Name List 2024
Top 20 Famous Indian Female Pornstars Name List 2024Sheetaleventcompany
 
👉Bangalore Call Girl Service👉📞 7304373326 👉📞 Just📲 Call Rajveer Call Girls Se...
👉Bangalore Call Girl Service👉📞 7304373326 👉📞 Just📲 Call Rajveer Call Girls Se...👉Bangalore Call Girl Service👉📞 7304373326 👉📞 Just📲 Call Rajveer Call Girls Se...
👉Bangalore Call Girl Service👉📞 7304373326 👉📞 Just📲 Call Rajveer Call Girls Se...Sheetaleventcompany
 
Call Girls Service Chandigarh Sexy Video ❤️🍑 8511114078 👄🫦 Independent Escort...
Call Girls Service Chandigarh Sexy Video ❤️🍑 8511114078 👄🫦 Independent Escort...Call Girls Service Chandigarh Sexy Video ❤️🍑 8511114078 👄🫦 Independent Escort...
Call Girls Service Chandigarh Sexy Video ❤️🍑 8511114078 👄🫦 Independent Escort...Sheetaleventcompany
 
Sexy Call Girl Palani Arshi 💚9058824046💚 Palani Escort Service
Sexy Call Girl Palani Arshi 💚9058824046💚 Palani Escort ServiceSexy Call Girl Palani Arshi 💚9058824046💚 Palani Escort Service
Sexy Call Girl Palani Arshi 💚9058824046💚 Palani Escort Servicejaanseema653
 
Kolkata Call Girls Miss Inaaya ❤️ at @30% discount Everyday Call girl
Kolkata Call Girls Miss Inaaya ❤️ at @30% discount Everyday Call girlKolkata Call Girls Miss Inaaya ❤️ at @30% discount Everyday Call girl
Kolkata Call Girls Miss Inaaya ❤️ at @30% discount Everyday Call girlonly4webmaster01
 
Independent Call Girls Hyderabad 💋 9352988975 💋 Genuine WhatsApp Number for R...
Independent Call Girls Hyderabad 💋 9352988975 💋 Genuine WhatsApp Number for R...Independent Call Girls Hyderabad 💋 9352988975 💋 Genuine WhatsApp Number for R...
Independent Call Girls Hyderabad 💋 9352988975 💋 Genuine WhatsApp Number for R...Ahmedabad Call Girls
 
AECS Layout Escorts (Bangalore) 9352852248 Women seeking Men Real Service
AECS Layout Escorts (Bangalore) 9352852248 Women seeking Men Real ServiceAECS Layout Escorts (Bangalore) 9352852248 Women seeking Men Real Service
AECS Layout Escorts (Bangalore) 9352852248 Women seeking Men Real ServiceAhmedabad Call Girls
 
🍑👄Ludhiana Escorts Service☎️98157-77685🍑👄 Call Girl service in Ludhiana☎️Ludh...
🍑👄Ludhiana Escorts Service☎️98157-77685🍑👄 Call Girl service in Ludhiana☎️Ludh...🍑👄Ludhiana Escorts Service☎️98157-77685🍑👄 Call Girl service in Ludhiana☎️Ludh...
🍑👄Ludhiana Escorts Service☎️98157-77685🍑👄 Call Girl service in Ludhiana☎️Ludh...dilpreetentertainmen
 
(Deeksha) 💓 9920725232 💓High Profile Call Girls Navi Mumbai You Can Get The S...
(Deeksha) 💓 9920725232 💓High Profile Call Girls Navi Mumbai You Can Get The S...(Deeksha) 💓 9920725232 💓High Profile Call Girls Navi Mumbai You Can Get The S...
(Deeksha) 💓 9920725232 💓High Profile Call Girls Navi Mumbai You Can Get The S...Ahmedabad Call Girls
 
Best Lahore Escorts 😮‍💨03250114445 || VIP escorts in Lahore
Best Lahore Escorts 😮‍💨03250114445 || VIP escorts in LahoreBest Lahore Escorts 😮‍💨03250114445 || VIP escorts in Lahore
Best Lahore Escorts 😮‍💨03250114445 || VIP escorts in LahoreDeny Daniel
 
Sexy Call Girl Tiruvannamalai Arshi 💚9058824046💚 Tiruvannamalai Escort Service
Sexy Call Girl Tiruvannamalai Arshi 💚9058824046💚 Tiruvannamalai Escort ServiceSexy Call Girl Tiruvannamalai Arshi 💚9058824046💚 Tiruvannamalai Escort Service
Sexy Call Girl Tiruvannamalai Arshi 💚9058824046💚 Tiruvannamalai Escort Servicejaanseema653
 
Sexy Call Girl Villupuram Arshi 💚9058824046💚 Villupuram Escort Service
Sexy Call Girl Villupuram Arshi 💚9058824046💚 Villupuram Escort ServiceSexy Call Girl Villupuram Arshi 💚9058824046💚 Villupuram Escort Service
Sexy Call Girl Villupuram Arshi 💚9058824046💚 Villupuram Escort Servicejaanseema653
 
visakhapatnam Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
visakhapatnam Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meetvisakhapatnam Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
visakhapatnam Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetCall Girls Chandigarh
 
💚Chandigarh Call Girls Service 💯Jiya 📲🔝8868886958🔝Call Girls In Chandigarh No...
💚Chandigarh Call Girls Service 💯Jiya 📲🔝8868886958🔝Call Girls In Chandigarh No...💚Chandigarh Call Girls Service 💯Jiya 📲🔝8868886958🔝Call Girls In Chandigarh No...
💚Chandigarh Call Girls Service 💯Jiya 📲🔝8868886958🔝Call Girls In Chandigarh No...Sheetaleventcompany
 
(Big Boobs Indian Girls) 💓 9257276172 💓High Profile Call Girls Jaipur You Can...
(Big Boobs Indian Girls) 💓 9257276172 💓High Profile Call Girls Jaipur You Can...(Big Boobs Indian Girls) 💓 9257276172 💓High Profile Call Girls Jaipur You Can...
(Big Boobs Indian Girls) 💓 9257276172 💓High Profile Call Girls Jaipur You Can...Joya Singh
 
Call Girl in Bangalore 9632137771 {LowPrice} ❤️ (Navya) Bangalore Call Girls ...
Call Girl in Bangalore 9632137771 {LowPrice} ❤️ (Navya) Bangalore Call Girls ...Call Girl in Bangalore 9632137771 {LowPrice} ❤️ (Navya) Bangalore Call Girls ...
Call Girl in Bangalore 9632137771 {LowPrice} ❤️ (Navya) Bangalore Call Girls ...mahaiklolahd
 
Jaipur Call Girls 9257276172 Call Girl in Jaipur Rajasthan
Jaipur Call Girls 9257276172 Call Girl in Jaipur RajasthanJaipur Call Girls 9257276172 Call Girl in Jaipur Rajasthan
Jaipur Call Girls 9257276172 Call Girl in Jaipur Rajasthanindiancallgirl4rent
 
Indore Call Girl Service 📞9235973566📞Just Call Inaaya📲 Call Girls In Indore N...
Indore Call Girl Service 📞9235973566📞Just Call Inaaya📲 Call Girls In Indore N...Indore Call Girl Service 📞9235973566📞Just Call Inaaya📲 Call Girls In Indore N...
Indore Call Girl Service 📞9235973566📞Just Call Inaaya📲 Call Girls In Indore N...Sheetaleventcompany
 

Último (20)

Rishikesh Call Girls Service 6398383382 Real Russian Girls Looking Models
Rishikesh Call Girls Service 6398383382 Real Russian Girls Looking ModelsRishikesh Call Girls Service 6398383382 Real Russian Girls Looking Models
Rishikesh Call Girls Service 6398383382 Real Russian Girls Looking Models
 
Escorts Lahore || 🔞 03274100048 || Escort service in Lahore
Escorts Lahore || 🔞 03274100048 || Escort service in LahoreEscorts Lahore || 🔞 03274100048 || Escort service in Lahore
Escorts Lahore || 🔞 03274100048 || Escort service in Lahore
 
Top 20 Famous Indian Female Pornstars Name List 2024
Top 20 Famous Indian Female Pornstars Name List 2024Top 20 Famous Indian Female Pornstars Name List 2024
Top 20 Famous Indian Female Pornstars Name List 2024
 
👉Bangalore Call Girl Service👉📞 7304373326 👉📞 Just📲 Call Rajveer Call Girls Se...
👉Bangalore Call Girl Service👉📞 7304373326 👉📞 Just📲 Call Rajveer Call Girls Se...👉Bangalore Call Girl Service👉📞 7304373326 👉📞 Just📲 Call Rajveer Call Girls Se...
👉Bangalore Call Girl Service👉📞 7304373326 👉📞 Just📲 Call Rajveer Call Girls Se...
 
Call Girls Service Chandigarh Sexy Video ❤️🍑 8511114078 👄🫦 Independent Escort...
Call Girls Service Chandigarh Sexy Video ❤️🍑 8511114078 👄🫦 Independent Escort...Call Girls Service Chandigarh Sexy Video ❤️🍑 8511114078 👄🫦 Independent Escort...
Call Girls Service Chandigarh Sexy Video ❤️🍑 8511114078 👄🫦 Independent Escort...
 
Sexy Call Girl Palani Arshi 💚9058824046💚 Palani Escort Service
Sexy Call Girl Palani Arshi 💚9058824046💚 Palani Escort ServiceSexy Call Girl Palani Arshi 💚9058824046💚 Palani Escort Service
Sexy Call Girl Palani Arshi 💚9058824046💚 Palani Escort Service
 
Kolkata Call Girls Miss Inaaya ❤️ at @30% discount Everyday Call girl
Kolkata Call Girls Miss Inaaya ❤️ at @30% discount Everyday Call girlKolkata Call Girls Miss Inaaya ❤️ at @30% discount Everyday Call girl
Kolkata Call Girls Miss Inaaya ❤️ at @30% discount Everyday Call girl
 
Independent Call Girls Hyderabad 💋 9352988975 💋 Genuine WhatsApp Number for R...
Independent Call Girls Hyderabad 💋 9352988975 💋 Genuine WhatsApp Number for R...Independent Call Girls Hyderabad 💋 9352988975 💋 Genuine WhatsApp Number for R...
Independent Call Girls Hyderabad 💋 9352988975 💋 Genuine WhatsApp Number for R...
 
AECS Layout Escorts (Bangalore) 9352852248 Women seeking Men Real Service
AECS Layout Escorts (Bangalore) 9352852248 Women seeking Men Real ServiceAECS Layout Escorts (Bangalore) 9352852248 Women seeking Men Real Service
AECS Layout Escorts (Bangalore) 9352852248 Women seeking Men Real Service
 
🍑👄Ludhiana Escorts Service☎️98157-77685🍑👄 Call Girl service in Ludhiana☎️Ludh...
🍑👄Ludhiana Escorts Service☎️98157-77685🍑👄 Call Girl service in Ludhiana☎️Ludh...🍑👄Ludhiana Escorts Service☎️98157-77685🍑👄 Call Girl service in Ludhiana☎️Ludh...
🍑👄Ludhiana Escorts Service☎️98157-77685🍑👄 Call Girl service in Ludhiana☎️Ludh...
 
(Deeksha) 💓 9920725232 💓High Profile Call Girls Navi Mumbai You Can Get The S...
(Deeksha) 💓 9920725232 💓High Profile Call Girls Navi Mumbai You Can Get The S...(Deeksha) 💓 9920725232 💓High Profile Call Girls Navi Mumbai You Can Get The S...
(Deeksha) 💓 9920725232 💓High Profile Call Girls Navi Mumbai You Can Get The S...
 
Best Lahore Escorts 😮‍💨03250114445 || VIP escorts in Lahore
Best Lahore Escorts 😮‍💨03250114445 || VIP escorts in LahoreBest Lahore Escorts 😮‍💨03250114445 || VIP escorts in Lahore
Best Lahore Escorts 😮‍💨03250114445 || VIP escorts in Lahore
 
Sexy Call Girl Tiruvannamalai Arshi 💚9058824046💚 Tiruvannamalai Escort Service
Sexy Call Girl Tiruvannamalai Arshi 💚9058824046💚 Tiruvannamalai Escort ServiceSexy Call Girl Tiruvannamalai Arshi 💚9058824046💚 Tiruvannamalai Escort Service
Sexy Call Girl Tiruvannamalai Arshi 💚9058824046💚 Tiruvannamalai Escort Service
 
Sexy Call Girl Villupuram Arshi 💚9058824046💚 Villupuram Escort Service
Sexy Call Girl Villupuram Arshi 💚9058824046💚 Villupuram Escort ServiceSexy Call Girl Villupuram Arshi 💚9058824046💚 Villupuram Escort Service
Sexy Call Girl Villupuram Arshi 💚9058824046💚 Villupuram Escort Service
 
visakhapatnam Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
visakhapatnam Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meetvisakhapatnam Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
visakhapatnam Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
 
💚Chandigarh Call Girls Service 💯Jiya 📲🔝8868886958🔝Call Girls In Chandigarh No...
💚Chandigarh Call Girls Service 💯Jiya 📲🔝8868886958🔝Call Girls In Chandigarh No...💚Chandigarh Call Girls Service 💯Jiya 📲🔝8868886958🔝Call Girls In Chandigarh No...
💚Chandigarh Call Girls Service 💯Jiya 📲🔝8868886958🔝Call Girls In Chandigarh No...
 
(Big Boobs Indian Girls) 💓 9257276172 💓High Profile Call Girls Jaipur You Can...
(Big Boobs Indian Girls) 💓 9257276172 💓High Profile Call Girls Jaipur You Can...(Big Boobs Indian Girls) 💓 9257276172 💓High Profile Call Girls Jaipur You Can...
(Big Boobs Indian Girls) 💓 9257276172 💓High Profile Call Girls Jaipur You Can...
 
Call Girl in Bangalore 9632137771 {LowPrice} ❤️ (Navya) Bangalore Call Girls ...
Call Girl in Bangalore 9632137771 {LowPrice} ❤️ (Navya) Bangalore Call Girls ...Call Girl in Bangalore 9632137771 {LowPrice} ❤️ (Navya) Bangalore Call Girls ...
Call Girl in Bangalore 9632137771 {LowPrice} ❤️ (Navya) Bangalore Call Girls ...
 
Jaipur Call Girls 9257276172 Call Girl in Jaipur Rajasthan
Jaipur Call Girls 9257276172 Call Girl in Jaipur RajasthanJaipur Call Girls 9257276172 Call Girl in Jaipur Rajasthan
Jaipur Call Girls 9257276172 Call Girl in Jaipur Rajasthan
 
Indore Call Girl Service 📞9235973566📞Just Call Inaaya📲 Call Girls In Indore N...
Indore Call Girl Service 📞9235973566📞Just Call Inaaya📲 Call Girls In Indore N...Indore Call Girl Service 📞9235973566📞Just Call Inaaya📲 Call Girls In Indore N...
Indore Call Girl Service 📞9235973566📞Just Call Inaaya📲 Call Girls In Indore N...
 

Analytical Method Validation

  • 1. 1
  • 2.  According FDA Validation is “Established documented evidence which provides a high degree of assurance that a specific process will continually produce a product meeting its predetermined specifications and quality attributes.”  The Validated method should be stability indicating method, which means the method that can detect the changes in properties of drug substance or drug product with time 2
  • 3.  Verification  Compendial methods-Verification : • These methods are listed in pharmacopoeia’s • Use to evaluate defined characteristics of the drug substance or drug product • Suitability of Compendial methods must be verified under actual conditions • Should be partially validated  Validation  Non -Compendial methods-Validation • Proposed by the applicant for use instead of Regulatory Analytical Procedure • Equivalency: • Equivalency: required for an in-house method, when compendial standard is claimed 3
  • 6. Verification data is required.  The compendial methods as published are typically validated based on an API or an FPP originating from a specific manufacturer.  Different sources of the same API or FPP may contain impurities and/or degradation products that were not considered during the development of the monograph.  The FPP may also contain excipients not present in the FPP(s) for which the compendial method was developed.  API Assay: No validation generally required. Exception: specificity for any specified impurities not in the monograph. Purity: Full validation for specified impurities that are not included in the monograph (specificity, linearity, accuracy, repeatability, intermediate precision, LOD/LOQ).  FPP Assay: Specificity, accuracy and precision (repeatability). Eg. to rule out excipient interference Purity: Specificity, accuracy and precision (repeatability) for known imps.  Full validation for specified impurities that are not included in the monograph (specificity, linearity, accuracy, repeatability, intermediate precision, LOD/LOQ) 6
  • 7.  Full validation is required for an in-house purity/assay method.  Equivalency must be demonstrated compared to the compendial method • Assay: same samples using the two methods; • Related substances: same samples spiking with specified impurities at the limits) using the two methods  If equivalency is not demonstrated, it can not be claimed as compendial standard 7
  • 8. Equivalency is essentially two things:  Ensure the in-house method is not inferior (in which case they should adopt the compendial method)  Ensure the compendial test can be met, if applied.  Two samples are sufficient. Note that a compendial method becomes an in-house method when adjustments are made outside the allowable adjustments (Int.Ph, USP, EP general chapter Chromatography) 8
  • 9.  Specificity  Precision  Accuracy  Linearity  Range  Quantitation Limit  Detection Limit  Robustness  Stability in Analytical Solution 9
  • 10. USP ICH  Specificity  Linearity and Range  Accuracy  Precision  Limit of Detection  Limit of Quantitation  Ruggedness  Robustness  Specificity  Linearity  Range  Accuracy  Precision  Repeatability  Intermediate Precision  Reproducibility  Limit of Detection  Limit of Quantitation 10
  • 11. Type of Analytical Procedure Identification Impurity testing Assay Quantitative Limit Tests Accuracy No Yes No Yes Precision Repeatability No Yes No Yes Interm. Prec. No Yes No Yes Specificity Yes Yes Yes Yes LOD No No Yes No LOQ No Yes No No Linearity No Yes No Yes Range No Yes No Yes May be required, depending on the nature of the specific test 11
  • 12. Testing a method to demonstrate it is suitable for its intended purpose and the results obtained are reliable, accurate and reproducible. Provides confidence that the method will perform properly under intended conditions.  Identification tests;  Quantitative or limit tests for the control of impurities;  Quantitative tests of the active moiety in samples of drug substance or drug product or other selected component(s) in the drug product. Including assay, Content Uniformity, dissolution, content of perseveratives. Validation of analytical procedures requires:  Qualified and calibrated instruments  Documented methods  Reliable reference standards  Qualified analysts  Sample integrity 12
  • 13.  Ability of an analytical method to measure the analyte free from interference due to other components.  An investigation of specificity should be conducted during the validation of identification tests, the determination of impurities and assay  Should be able to discriminate between compounds closely related in structure.  Confirmed by obtaining negative results for samples with spiked related compounds and positive results for samples with analyte.  SELECTIVITY “A measure of the discriminating power of a given analytical procedure in differentiating between the analyte and other components in the test sample.” 13
  • 14.  Impurities/Degradants Available  Spike with impurities/degradants  Show resolution and a lack of interference  Impurities/Degradants Not Available  Stress Samples  For assay, Stressed and Unstressed Samples should be compared.  For impurity test, impurity profiles should be compared.  Peak homogeneity tests should be performed using PDA or mass spectrometry to show that the analyte chromatographic peak is not attributable to more than one component. 14
  • 15.  SELECTIVITY & SPECIFICITY 15 Selectivity: Specificity: If an analytical procedure is able to separate and resolve the various components of a mixture and detect the analyte qualitatively If the method determines or measures quantitatively the component of interest in the sample matrix without separation Restricted to qualitative detection of the components of a sample Restricted to quantitative measurement of one or more analyte Blank solution - to show no interference Placebo - to demonstrate the lack of interference from excipients Spiked samples - to show that all known related substances are resolved from each other Stressed sample of about 5% to 20% degradation is used to demonstrate the resolution between degradants and the analyte of interest Check peak purity of drug substance by photodiode array detector (PDA) Representative chromatograms should be provided
  • 16. Temperature  Humidity  Acid Hydrolysis  Base Hydrolysis  Oxidation  Light (UV/Vis/Fl) Intent is to create 10 to 30 % Degradation 16
  • 17.  The closeness of agreement (degree of scatter) between a series of measurements obtained from multiple samplings of the same homogeneous sample. Precision… Considered at 3 Levels  Repeatability  Intermediate Precision  Reproducibility 17
  • 18.  Repeatability A minimum of nine determinations covering the specified range for the procedure, e.g.  three concentrations/three replicates each, or a minimum of six determinations at 100% of the test concentration  Intermediate Precision Within-laboratory variations  usually on different days, different analysts and different equipment.  Reproducibility Precision between laboratories 18
  • 19.  Closeness of agreement between the value which is accepted either as a conventional true value or an accepted reference value and the value found  Accuracy should be assessed using a minimum of 9 determinations over a minimum of 3 concentration levels covering the specified range  Accuracy is usually reported as percent recovery by the assay (using the proposed analytical procedure) of known added amount of analyte in the sample or as the difference between the mean and the accepted true value together with the confidence intervals. The range for the accuracy limit should be within the linear range 19
  • 20.  Drug Substance For Assay  Accuracy can be inferred once precision, linearity and specificity have been established For Related Substances/ Residual Solvents  Accuracy should be assessed on samples spiked with known amount of impurities Drug Product For Assay  Known quantities of drug substance to be analysed have been added to synthetic mixture of drug product components. For Related Substances/ Residual Solvents  Accuracy should be assessed on samples spiked with known amount of impurities 20
  • 21. Accuracy range….. Across the range of LOQ-150% of the target concentration (shelf life limit), 3-5 concentrations, in triplicate each. (LOQ, 50%, 100%, 150%) Per cent recovery: in general, within 80-120%, depends on the level of limit Accurate & precise Accurate & imprecise Inaccurate & precise Inaccurate & imprecise 21
  • 22.  Linearity of an analytical procedure is its ability to obtain test results which are directly proportional to concentration of analyte in the sample within a specified range.  A series of samples should be prepared in which the analyte concentrations span the claimed range of the procedure. If there is a linear relationship, test results should be evaluated by appropriate statistical methods  Minimum of 5 concentration is recommended within the range Range: This is an expression of the lowest and highest levels of analyte that have been demonstrated to be determinable for the product The specified range is normally derived from linearity studies 22
  • 23.  Assay : 80-120% of the theoretical content of active  Content Uniformity: 70-130%  Dissolution: ±20% of limits; eg if limits cover from 20% to 90% l.c. (controlled release), linearity should cover 0-110% of l.c.  Impurities: reporting level to 120% of shelf life limit  Assay/Purity by a single method: reporting level of the impurities to 120% of assay limit 23
  • 24.  Is the amount of analyte in a sample which can be quantitatively determined with suitable precision and accuracy  Approaches (instrumental or non-instrumental):  visual evaluation (Used for non-instrumental methods)  signal to noise ratio (10:1 for Quantitation Limit)  standard deviation of the response and the slope Quantitation Limit = 10 S  : SD of Response (The residual SD of regression line or SD of y intercept of regression lines can be used as SD) S : Slope of calibration curve 24
  • 25. Signal Noise Peak A LOD Peak B LOQ Baseline Ybl LOD LOQ Statistical estimate of LOD & LOQ LOD = 3.3 Sbl / b LOQ = 10 Sbl / b Y = b X + a 25
  • 26.  Is the amount of analyte in a sample which can be detected but not necessarily quantitated as an exact value  Approaches (instrumental or non-instrumental):  visual evaluation (Used for non-instrumental methods)  signal to noise ratio (3:1 for Detection Limit)  standard deviation of the response and the slope Quantitation Limit = 3.3 S  : SD of Response (The residual SD of regression line or SD of y intercept of regression lines can be used as SD) S : Slope of calibration curve 26
  • 27.  Not required for assay/dissolution methods.  Applicant should provide  The method of determination  The limits  Chromatograms  Visually identified  Established by Repeatability 27
  • 28.  Measure of its capacity to remain unaffected by small , but deliberate variations in method parameters and provide an indication of its reliability during normal usage  If measurements are susceptible to variations in analytical conditions, than it should be suitably controlled or a precautionary statement should be included in procedure  System suitability parameters should be evaluated to ensure that validity of analytical procedure is maintained when ever used 28
  • 29.  For Liquid Chromatography, typical variations are  Influence of pH in a mobile phase ( ± 0.2 units)  Influence of mobile phase composition (± 2.0% Organic phase)  Different columns[ Lots/Suppliers]  Temperature [Column Oven] (± 5° C)  Flow rate ( ± 0.2ml/min)  System Suitability  Evaluate the System suitability parameters  Validation of analytical method must be focused on critical separation of components which can alter during change of conditions  These critically separated components should be part of system suitability 29
  • 30.  Should be established  By keeping the sample at RT or below 10°C and injecting it at regular intervals  No significant variations or trend should be observed  Any significant variation in particular area / absorbance should be compared or calculated against the initial value.  In case of impurity the % of impurity present should be verified against the initial .  The outcome of the result should be mentioned in the particular regular test procedure. 30
  • 31. When Re- validation is required !!!  Method parameters have been changed  The scope of the method has been changed  Synthetic methods have been changed  Impurity profile has been changed (Due to change in excipient).  Changes in equipment or suppliers of critical supplies of the API or final drug product will have the potential to change their degradation profile and may require the method to be redeveloped and revalidated. 31
  • 32.  Integral part of many analytical procedures  Based on the concept that:  The equipment, electronics, analytical operations and samples to be analyzed constitute an integral system that can be evaluated as such  Determination (in General): Repeatability, Tailing factor (T), Capacity factor (k’), Resolution (R), and Theoretical Plates (N) 32
  • 33. Parameters Recommendations K’ In general k’ ≥ 2.0 R R > 2, between the peak of interest and the closest potential interferent (degradant, internal STD, impurity, excipient, etc…..) T T ≤ 2 N In general N > 2000 Repeatability RSD ≤ 2.0% (n ≥ 5) Assay: RSD ≤1% (API) or ≤ 2% (FPP), n ≥ 5 Impurities: in general, RSD ≤ 5% at the limit level, up to 10% or higher at LOQ, n ≥ 6 33
  • 34. Allowed HPLC Adjustment USP (Ref: General Chapter <621>) EP (Ref: General Chapter 2.2.46) Column Length ±70% ±70% Internal Diameter Can be adjusted if linear velocity is kept constant ±25% Particle Size Reduction of 50%, no increase Reduction of 50%, no increase Flow Rate ±50% or more as long as the linear velocity is kept contant ±50% Column Temperature ±10C ±10% Max 60C Injection Volume Change allowed as long as SST criteria are met May be decreased (if LOD and repeatability ok) PH ±0.2 Units ±0.2 Units (±1% for neutral substances) UV wavelength <±3 nm (in condition) <±3 nm ((in condition)) Conc. Salts in Buffer ±10% ±10% Composition of mobile phase Minor Components (<50%) ±30% or ±10% absolute whichever is smaller Minor Components ±30% or ±2% absolute whichever is larger 34
  • 35.  Some 483 Observations  Analytical methods are not adequately validated  There was no inadequate method validation specificity data to demonstrate that each method was capable of distinguishing the active ingredient from its impurities and degradation products.  Specificity studies did not include the minimum stress conditions of acid and base hydrolysis, oxidation, thermal degradation and photolysis, degradation schematic for the active ingredient that identifies the major degradation products was not included for each product. 35
  • 36.  Some 483 Observations  Stress studies conducted as part of method validation do not target a minimum amount of degradation. … a standard period of two hours as commonly used for stress studies with no justification…  Spreadsheets used to calculate linearity, percent recovery, and final assay results for the cleaning validation of …were not validated and the data transcribed from chromatographs to the spreadsheets were not checked for accuracy. 36
  • 37.  FDA Waning Letter  On addition to the example of modifying both Compendial methods and customer supplied methods, we also observed the use of un-validated in-house methods as well as un-validated modifications to in-house methods.  A statement indicating that the method has not been validated in the particular formulation was included in the certificate of analysis for…use of this statement does not absolve…from using valid, accurate, and reproducible methods.  OOS accuracy results reported by analyst 3 were never submitted in the final report. Repeat analysis performed in a different system passed specifications and these results were submitted in the report. 37
  • 38.  FDA Warning Letter  There is no assurance that qualification or maintenance of the laboratory equipment can consistently produce valid and accurate analytical results in that numerous examples of test data were invalidated due to instrument malfunction.  Attempts to corroborate data in the validation report with supporting raw data in the laboratory were difficult and frustrating for the FDA personnel conducting the inspection  Raw data and calculations were not checked by a second responsible individuals required by your procedures. Inaccurate calculations were noted in the report 38
  • 39.  FDA Warning Letter  The process validation samples were assays using an HPLC method that had not been validated. The method validation used for both products, did not include a protocol that included specification and acceptance criteria. … The method validation was not reviewed and approved until during the current inspection. Lots of both products were released for distribution prior to completion of the method validation.  Method validation for the product one product is inadequate in that the data does not assess all variables, such as different mobile phase concentrations and analytes, to demonstrate that the method can sustain variance. 39
  • 40.  Data for acceptance, release, stability will only be trustworthy if the methods used are reliable  More efficient use of resources in the R&D process, as a consequence  Validation should be Protocol driven  Document each and every information with regard to std/sample/reagent/impurities/process/ preparation etc.  Document the details of each experiment with suitable way.  Review each experiment after its completion and relate its outcome with other experiments  Identification and elimination of the need to duplicate studies to meet different regulatory requirements 40
  • 41.  From My Side……  Why is analytical monitoring necessary?  What is the purpose of analytical validation? 41