Public Health Action International Union Against Tuberculosis and Lung Disease
Health solutions for the poor
VOL 8 no 1  PUBLISHED 21 MARCH 2018
PHA 2018; 8(1): 2–6
© 2018 The Union
AFFILIATIONS
1	 KNCV Tuberculosis
Foundation, Abuja, Nigeria
2	 National Tuberculosis
Leprosy Control
Programme, Abuja,
Nigeria
CORRESPONDENCE
Nkiru Nwokoye
Laboratory, KNCV
Tuberculosis Foundation
564/565 Independence
Avenue
Central Business District
Abuja FCT 900211, Nigeria
e-mail: nkiru.nwokoye@
kncvtbc.org
ACKNOWLEDGEMENTS
The authors thank the
Nigerian National
Tuberculosis Programme and
all partners investing in Xpert
implementation for their
support, including the US
Agency for International
Development (Washington
DC, USA) and AB Associate
(Paris, France) for GXAlert.
Conflicts of interest: none
declared.
KEY WORDS
tuberculosis; diagnosis;
implementation; error
Unsuccessful Xpert® MTB/RIF results: the Nigerian experience
M. Gidado,1 N. Nwokoye,1 P. Nwadike,1 P. Ajiboye,1 R. Eneogu,1 S. Useni,1 J. Onazi,1 A. Lawanson,2
E. Elom,2 A. Tubi,2 J. Kuye2
Early diagnosis and access to effective treatment re-
mains fundamental to successful tuberculosis (TB)
control.1 TB diagnosis using smear microscopy takes
24–48 h, while solid culture takes 4–8 weeks. This diag-
nostic delay is an obstacle to achieving early diagno-
sis.2,3 The introduction of the Xpert® MTB/RIF assay
(Cepheid, Sunnyvale, CA, USA) revolutionised TB diag-
nosis and gave hope to struggling programmes. Today,
diagnostic decisions on both drug-susceptible (DS-TB)
and rifampicin (RMP) resistant TB (RR-TB) can be made
within 2 h of sample collection, thus significantly re-
ducing the lag time for diagnosis and facilitating
prompt clinical decision making, which benefits TB
control.4,5 Following the World Health Organization’s
(WHO’s) endorsement of Xpert, many countries, in-
cluding Nigeria, embraced this technology with opti-
mism.6 It is the first simple TB molecular test that is
robust enough to be introduced outside conventional
laboratory settings.7 However, the right placement of
Xpert machines in the laboratory requires careful con-
sideration of available resources, national capacity
building and accessibility of diagnostic testing.8
Xpert roll-out
Some key considerations that had to be taken into ac-
count when selecting sites for the installation of the
Xpert machines included high TB burden sites, cur-
rent workload of the facility, availability of personnel,
adequate infrastructure and the capacity of the appro-
priate treatment centre. It was thus possible to train a
cadre of staff capable of providing technical assistance
during the early stages of implementation.
Installation of Xpert machines commenced at na-
tional- and state-level laboratories with a relatively sta-
ble power supply that would ensure uninterrupted run-
ning of the Xpert assay. Setting up of air conditioners,
refrigerators and inverters was essential to provide the
ambient temperature required to operate the Xpert ma-
chines and store the cartridges, and to guarantee an un-
interrupted power supply during testing. GxAlert inter-
net data connectivity was also established for the upload
of data onto the central server. A 4-day theory and prac-
tical training course was provided for laboratory staff
and other health care workers at the time of the installa-
tion. The supply and distribution of cartridges to facili-
ties was centrally managed by the National Tuberculosis
and Leprosy Control Programme (NTBLCP).
At facility level, trained laboratory personnel per-
formed the assay according to the protocol and na-
tional standard operating procedures (SOPs). Based on
the SOPs, unsuccessful tests showing errors or invalid
or incomplete results were repeated. The resolution of
error results was dependent on the error codes. Prob-
lematic issues that persisted after troubleshooting were
reported to the Nigerian Cepheid technical centre for
further assistance, either remotely or on-site. Primary
concerns about the assay’s performance were ‘test fail-
ures’ or unsuccessful test outcomes that cause delay or
no diagnosis in some patients. Similar concerns about
high rates of error and invalid results were reported by
other countries during the early phase of implementa-
tion.9–11 The WHO recommended country-specific op-
erational research, as diagnostic tests that perform well
in controlled settings may not always perform opti-
mally in the settings where they are to be used.12
We studied the rate, distribution and causes of un-
successful Xpert test outcomes, with the aim of identi-
fying key areas that need to be strengthened for opti-
mal performance of the assay.
DESIGN AND METHODS
The study retrospectively analysed data uploaded be-
tween January and December 2015 from Xpert facili-
Received 6 September 2017
Accepted 11 January 2018
http://dx.doi.org/10.5588/pha.17.0080
Setting:  Nigeria, a high tuberculosis (TB) burden country.
Objective: To study the rate, distribution and causes of
unsuccessful Xpert® MTB/RIF test outcomes, with the aim
of identifying key areas that need to be strengthened for
optimal performance of the assay.
Design: This was a retrospective analysis of data up-
loaded between January and December 2015 from Xpert
facilities to the central server using GXAlert.
Result: Of 52 219 test results uploaded from 176 Xpert
machines, 22.5% were positive for Mycobacterium tuber-
culosis, 10.8% of which were rifampicin-resistant; 4.7%
of the total number of results were invalid, 4.2% had er-
ror results and 2.1% no result outcomes. Technical errors
were most frequent (69%); these were non-seasonal and
occurred in all geopolitical regions and at all health facil-
ity levels. Temperature-related errors were more preva-
lent in the North-West Region, with peaks in April to
June. Peak periods for temperature and machine mal-
function errors coincided with the periods of low utilisa-
tion of the assay.
Conclusion: The key challenge affecting performance
was poor adherence to standard operating procedures.
Periodic refresher training courses, regular supervision,
preventive maintenance of Xpert machines and proper
storage of cartridges are strategies that could improve
Xpert performance.
Unsuccessful Xpert results 3Public Health Action
ties to the central server using GXAlert. Testing failures on the
Xpert assay were classified as ‘error’, ‘invalid’ or ‘no result’. Identi-
fication and grouping of the error codes into categories were
guided by the error resolution SOPs provided by the Xpert manu-
facturer, in which error codes were classified under the causes and
the steps to take to resolve problems.13,14
By following the SOPs, facility-documented error codes were
identified and categorised into five groups based on causes of er-
ror (temperature-related, technical problems, cartridge malfunc-
tion, electrical connection and machine malfunction errors). Er-
rors due to incorrect temperature, which may be of internal origin
or due to external environmental factors, e.g., ambient tempera-
ture outside of the acceptable range, dirt on filters due to accumu-
lated dust, faulty fan and malfunctioning heater component,
were grouped as Category A errors (codes 1001, 1002, 2014, 1004
and 2022). Category B errors (codes 5006, 5007 and 2008) were
related to technical problems, i.e., mostly human errors due to
non-adherence to the SOPs during sample processing, such as fill-
ing reaction tubes with viscous sputum or incorrect sample vol-
umes, and clogged filters due to the presence of debris in sputum,
compromised probe integrity and module failures. Category C er-
rors (error codes 5011 and 2037) were associated with cartridge
malfunction, caused mostly by inappropriate storage of the car-
tridges. Electrical connection issues that resulted in communica-
tion loss between the module and the software due to incorrect
wiring connections (e.g., poor Ethernet connection between the
computer and Xpert, poor connection between gateway and
modules, and fluctuation in power supply, with codes 2127, 2126,
1007 and 2122) were classed under Category D. Machine mal-
function due to the breakdown of components or module mal-
function was grouped under Category E; the associated codes
were 2005, 2016, 2032, 2025, 4014 and 4017.
Data cleaning and validation were performed before analysis to
improve data quality by exporting GXAlert data to SPSS v16 soft-
ware (Statistical Package for the Social Sciences, Chicago, IL, USA).
Each variable was checked for accuracy and consistency, and fre-
quency tables were generated to check for outliers and errors. Any
issue identified was flagged and confirmed using source data.
As this was a retrospective study using anonymous data, ethics
approval was not sought.15
RESULTS
A total of 52 219 tests were performed using 176 Xpert machines.
Xpert machines were installed at all health facility levels, with the
highest proportion (56.6%) in secondary health care facilities. Ap-
proximately 22.5% of the total tests performed were positive for
Mycobacterium tuberculosis, 10.8% of which were RMP-resistant
and 3.0% RMP-indeterminate. Of the total number of tests con-
ducted, 4.7% were invalid, 4.2% had error results and 2.2% had
no result outcomes, putting the rate of unsuccessful tests at
11.0%. While invalid and ‘no result’ outcome rates remained rela-
tively constant across all health care facility levels, error outcomes
predominated (5.7%) at the primary health care level (Table 1). Of
2173 error results recorded, only 1736 (79.9%) were documented
with codes. About 19.1% of the errors could not be categorised
due to lack of information on associated codes.
In all, 20 different error codes were documented and catego-
rised into five broad groups (Categories A–E). The causes and reso-
lution steps for Categories A and D errors are shown in Table 2.
Category B errors, associated with technical issues, were most fre-
quent (69%), and were distributed throughout all geographical
regions and at all health facility levels. Whereas errors due to in-
correct temperature and poor electrical connections were most
frequent in the northern region, errors due to cartridge malfunc-
tion predominated in the southern region. It should be noted
that errors due to machine malfunction cut across the North and
South geographical regions, and were most prevalent in the
North-West and South-West (Table 3).
Technical problems occurred throughout the year. Although no
temperature-related errors occurred in January, these were recorded
from February to December, with a peak in April, May and June. A
downward trend was observed from July, with a steady decline to
its lowest frequency in September and October, before finally in-
creasingly slightly in November and December. Errors associated
with cartridge malfunction and poor electrical connections re-
mained relatively constant in terms of distribution, and occurred
throughout the year. Conversely, machine malfunction errors were
more prevalent in April, May and June; this, interestingly, coin-
cided with the peak period recorded for temperature-related errors.
DISCUSSION
Findings from this study revealed gross underutilisation of Xpert
machines. This is worrisome, as roll-out of the assay was supposed
to increase their use, given the high level of automation that
makes it user-friendly. During the early phases of implementa-
tion, most of the Xpert machines were placed in secondary- and
tertiary-level health facilities. The main assumption for this pref-
erential placement was that a stable power supply was more read-
ily available at tertiary- and secondary-level facilities than at the
primary level. However, from our findings, it is clear that inter-
ruptions in power supply, synonymous with a ‘no result’ out-
come, were evenly distributed among the three facility levels, in-
dicating that an efficient power supply is not dependent on
facility level. Greater attention should be paid to sustainability of
the back-up power supply to ensure uninterrupted testing, rather
than facility-level placement.
TABLE 1  Proportion of test outcomes by facility level
Level of
health facility
Frequency
n (%)
M. tuberculosis detection
RMP resistance
(subset of M. tuberculosis-positive)
Error
n (%)
Invalid
n (%)
No result
n (%)
Positive
n (%)
Negative
n (%)
Resistant
n (%)
Susceptible
n (%)
Indeterminate
n (%)
Primary 2 036 (3.9) 479 (23.5) 1 293 (63.7) 52 (10.8) 419 (87.5) 8 (1.6) 117 (5.7) 98 (4.8) 36 (1.8)
Secondary 29 548 (56.6) 6719 (22.7) 19 529 (66.0) 668 (9.9) 5 872 (87.4) 179 (2.7) 1 190 (4.0) 1 400 (4.7) 710 (2.4)
Tertiary 20 635 (39.5) 4547 (22.0) 13 851 (67.1) 543 (11.9) 3 839 (84.4) 165 (3.6) 866 (4.2) 971 (4.7) 399 (1.9)
 Total 52 219 (100) 11 745 (22.5) 34 673 (66.4) 1 264 (10.8) 10 130 (86.2) 352 (3.0) 2 173 (4.2) 2 469 (4.7) 1 145 (2.2)
RMP = rifampicin.
Unsuccessful Xpert results 4Public Health Action
In addition to an uninterrupted power supply, the capacity of
the health facility to ensure continued operation after installation
should also be taken into consideration when selecting sites for
Xpert machine placement. This includes maintenance of facility
infrastructure such as air conditioners and refrigerators, which
has not always been the case in most of the public health facili-
ties selected. The NTBLCP and donors should prioritise the imple-
mentation of strategies for improved functionality of Xpert ma-
chines over the installation of new machines, to ensure optimal
performance at all times. The 22.5% TB detection rate and 10.7%
RMP resistance rate recorded, along with the uninterpretable re-
sults, was significant. Optimising the functionality of Xpert ma-
chines and reducing the number of uninterpretable results may
increase the detection of TB and RR-TB. The patient population
screened, which comprised presumptive drug-resistant cases and
patients with human immunodeficiency virus with TB symptoms,
contributed to the high RR-TB detection rate. The main impetus
was the timely detection of multidrug-resistant TB,16–19 as RMP re-
sistance is a surrogate for isoniazid resistance.20,21
The key lesson learnt from the Xpert roll-out was that access to
diagnosis is not the only indicator for measuring successful im-
plementation; the rate of unsuccessful tests (which add no value
to treatment) is also an important indicator of the effectiveness of
Xpert implementation over time. Although Xpert was able to de-
TABLE 2  Interpretation and resolution of error codes*
Error type Error code Possible cause Solution
Technically related
(Category B)
5006 Probe check failed because: Check if:
•	 An incorrect amount of reagent was inserted into
the cartridge
•	 Reagent was correctly added to the cartridge
•	 The reagent was of compromised quality •	 Cartridges were stored correctly
•	 Fluid transfer failed Rerun the test using fresh cartridges
5007 Probe check failed because: Check if:
•	 An incorrect amount of reagent was inserted into
the cartridge
•	 Reagent was added to the cartridge correctly
•	 The reagent was of compromised quality •	 Cartridges were stored correctly
•	 Fluid transfer failed
•	 The sample was processed incorrectly in the
cartridge
2008 •	 The filter was clogged Use a new cartridge
•	 Pressure sensor failed Run a cartridge contacting buffer only
Poor electrical connection
(Category D)
2127 Module communication loss was detected because
Ethernet cable between computer and Xpert
machine was not connected properly
Unplug and replug instrument and Ethernet cable,
then restart system
2126 Module was reset because there was an intermittent
power supply failure or power supply cable or
connector failure
Restart system
1007† The power supply voltage was out of range Record the information in the error message. If the
error recurs in multiple runs, call Cepheid
Technical Support
2122† Loose or faulty Ethernet cable between the
computer and the Xpert machine
Verify that the Ethernet cable between the computer
and the Xpert machine is connected properly
* Adapted from Reference 14.
† Explanation provided by Cepheid via e-mail.
TABLE 3  Distribution of different categories of error by geographic region and health facility level
Tests
performed
n (%)
Error type categorised
A (temperature-
related)
n (%)
B (technically
related)
n (%)
C (cartridge
malfunction)
n (%)
D (poor electrical
connection)
n (%)
E (machine
malfunction)
n (%)
Total
n
Geographic region
 South-South 8 138 (15.6) 1 (0.5) 155 (73.8) 32 (15.2) 18 (8.6) 4 (1.9) 210
 South-East 8 013 (15.3) 4 (1.5) 212 (80.9) 16 (6.1) 24 (9.2) 6 (2.3) 262
 South-West 9 288 (17.8) 24 (8.1) 169 (56.9) 54 (18.2) 31 (10.4) 19 (6.4) 297
 North-Central 13 891 (26.6) 14 (2.9) 373 (76.6) 53 (10.9) 40 (8.2) 7 (1.4) 487
 North-East 3 871 (7.4) 2 (1.3) 95 (62.5) 9 (5.9) 43 (28.3) 3 (2.0)
 North-West 9 018 (17.3) 41 (12.5) 187 (57.0) 24 (7.3) 57 (17.4) 19 (5.8) 328
  Total 52 219 (100) 86 (5) 1 191 (68.6) 188 (10.8) 213 (12.3) 58 (3.3) 1 736
Facility level
 Primary 6 (6.9) 52 (59.8) 3 (3.4) 24 (27.6) 2 (2.3) 87
 Secondary 25 (2.5) 709 (70.5) 95 (9.5) 145 (14.4) 31 (3.1) 1 005
 Tertiary 55 (8.5) 430 (66.8) 90 (14) 44 (6.8) 25 (3.9) 644
  Total 86 (5) 1 191 (68.6) 188 (10.8) 213 (12.3) 58 (3.3) 1 736
Unsuccessful Xpert results 5Public Health Action
tect a good number of TB cases, the high rate of unsuccessful test
results clearly showed that early diagnosis, which is fundamental
to TB control, may not be achieved in many individuals if Xpert
is not properly managed.
According to the Xpert manufacturer, a cumulative error rate
of 5% is unacceptable and should be investigated and resolved
immediately. The 5.7% error rate recorded in the primary health-
level facilities in our study indicates suboptimal performance;
strategies identified for the resolution of these errors should
therefore be intensified at the primary level.
In view of the above, we went a step further and studied the
distribution of error categories by geographic region and month.
A high percentage (19.1%) of the errors were not coded, which
suggests that most facilities do not make use of the error resolu-
tion SOPs for troubleshooting. A similar observation was made in
a previous article.22 Most errors involved technical issues and
originated from incorrect sample processing or poor sample qual-
ity. It is to be noted that the high rate of technical errors was
among the issues discussed at the Xpert Early Implementers Meet-
ing held in France in 2012.23 Most of the errors recorded by early
implementers, such as South Africa, were linked to improper pro-
cedures used in specimen collection and sample preparation, as
well as faulty modules and cartridges. The fact that Category B er-
rors cut across geographic regions and facility levels shows that all
cadre of staff working at Xpert sites are implicated. This can be
attributed to the high rate of staff attrition in the health sector,
whereby experienced staff leave in high numbers and are replaced
by new personnel without technical experience. The introduction
of planned capacity building schemes—which include training,
supervision, mentoring and proficiency testing—by the NTBLCP
for staff at Xpert facilities will facilitate skill acquisition and con-
tinued quality improvement. To reduce Category C errors, the
storage temperature for cartridges should be carefully monitored
and maintained from the time of landing at the port until the car-
tridges arrive at the warehouse, to their final destination at Xpert
facilities. This also includes maintaining the correct temperature
during transportation.
Category A errors were more frequent in the Northern region
due to its characteristic hot, dry, dusty climate. In addition to the
installation of air conditioners, special attention should be paid
to routine maintenance of the equipment, such as monthly clean-
ing of exhaust fan filters. Raizada et al. reported that most tem-
perature-associated errors were related to clogging of exhaust fan
filters by dust or the inappropriate positioning of equipment.12 It
is therefore recommended that Xpert operators, especially those
in the Northern region, routinely clean the filters, monitor and
chart ambient room temperature, and check the machine’s inter-
nal temperature whenever a Category A error is flagged. This con-
firms the source of the error and facilitates fast and accurate trou-
bleshooting. Maintaining a clearance space of 10 cm between the
Xpert machine and the wall and other equipment will go a long
way to ensure that the warm air generated during testing is ex-
pelled properly from the machine. The seasonal variations in the
errors indicate that temperature-related errors were most preva-
lent from March to June. This is to be expected, given the high
temperatures associated with these months.
CONCLUSION
The performance of the Xpert assay is dependent on adher-
ence to SOPs by programme personnel. Xpert roll-out should
be followed by quality supervision and mentoring to improve
staff competence in test procedures and routine maintenance
of the machine. The significant rate of unsuccessful tests ob-
served in this study shows that evaluations of the assay should
not be based on the case detection rate alone. Performance, in
terms of the number of interpretable results that informed the
patient’s treatment regimen, is also an indicator that should
be assessed and monitored for improved test outcomes. Fi-
nally, Xpert machines should be installed in those facilities
(private or public) where patients prefer to seek medical help,
and where the machines can be better managed, maintained
and utilised.
References
1	 Ghiasi M, Pande T, Pai M. Advances in tuberculosis diagnostics. Curr Trop
Med Rep 2015; 2: 54–61.
2	 Sharma S K, Kohli M, Yadav R N, et al. Evaluating the diagnostic accuracy of
Xpert MTB/RIF assay in pulmonary tuberculosis. PLOS ONE 2015; 10:
e0141011.
3	 Cruciani M, Scarparo C, Malena M, Bosco O, Serpelloni G, Mengoli C. Me-
ta-analysis of BACTEC 960 and BACTEC 460 TB, with or without solid me-
dia, for detection of mycobacteria. J Clin Microbiol 2004; 42: 2321–2325.
4	 Peralta G, Barry P, Pascopella L. Use of nucleic acid amplification tests in tu-
berculosis patients in California, 2010–2013. Open Forum Infect Dis 2016;
3: ofw230.
5	 Shenoi S V, Escombe A R, Friedland G. Transmission of drug-susceptible and
drug-resistant tuberculosis and the critical importance of airborne infection
control in the era of HIV infection and highly active antiretroviral therapy
rollouts. Clin Infect Dis 2010; 50 (Suppl 3): S231–S237.
6	 World Health Organization. Policy statement: automated real-time nucleic
acid amplification technology for rapid and simultaneous detection of tu-
berculosis and rifampicin resistance: Xpert MTB/RIF. WHO/HTM/TB/2011.4.
Geneva, Switzerland: WHO, 2011.
7	 World Health Organization. Xpert MTB/RIF implementation manual: tech-
nical and operational ‘how-to’ practical considerations. WHO/HTM/
TB/2014.1. Geneva, Switzerland: WHO, 2014.
8	 Piatek A S, Van Cleef M, Alexander H, et al. Xpert for TB diagnosis: planned
and purposeful implementation. Glob Health Sci Pract 2013; 1: 18–23.
9	 Van Rie A, Page-Shipp L, Scott L, Sanne I, Steven W. Xpert MTB/RIF for
point-of-care diagnosis of TB in high-HIV burden, resource limited coun-
tries: hype or hope? Expert Rev Mol Diagn 2010; 10: 937–946.
10	 Williamson D A, Basu I, Bower J, Freeman J I, Henderson G, Robert S A. An
evaluation of the Xpert MTB/RIF assay and detection of false-positive rifam-
picin resistance in Mycobacterium tuberculosis. Diagn Microbiol Infect Dis
2012; 74: 207–209.
11	 Van Rie A, Mellet K, John M A, et al. False-positive rifampicin resistance on
Xpert MTB/RIF: case report and clinical implications. Int J Tuberc Lung Dis
2012; 16: 206–208.
12	 Raizada N, Sachdeva K S, Sreenivas A. et al. Feasibility of decentralised de-
ployment of Xpert MTB/RIF test at lower level of health system in India.
PLOS ONE 2014; 9: e8930.
13	 Cepheid. Xpert MTB/RIF. Two hour detection of MTB and resistance to ri-
fampicin. Sunnyvale, CA, USA: Cepheid, 2009.
14	 Cepheid Xpert Dx System. Operator Manual. Software version 4.0. 300–
7607, Rev. C.17. Sunnyvale, CA, USA: Cepheid, 2010.
15	 Junod V, Elger B. Retrospective research: what are the ethical and legal re-
quirements. Swiss Med Weekly 2010: 140: w13041.
16	 Durovni B, Saraceni V, van den Hof S, et al. Impact of replacing smear mi-
croscopy with Xpert MTB/RIF for diagnosing tuberculosis in Brazil: a
stepped-wedge cluster-randomized trial. PLOS Med 2014; 11: e1001766.
17	 Raizada N, Sachdeva K S, Sreenivas A, et al. Catching the missing million:
experiences in enhancing TB and DR-TB detection by providing upfront
Xpert MTB/RIF testing for people living with HIV in India. PLOS ONE 2015;
10: e0116721.
18	 Sachdeva K S, Raizada N, Sreenivas A, et al. Use of Xpert MTB/RIF in decen-
tralized public health setting and its effect on pulmonary TB and DR-TB case
finding in India. PLOS ONE 2015; 10: e0126065.
19	 Trajman A, Durovni B, Saraceni V, et al. Impact on patient’s treatment out-
comes of Xpert MTB/RIF implementionfor the diagnosis of tuberculosis: fol-
low-up of a stepped-wedge randomized clinical trial. PLOS ONE 2015; 10:
e0123252.
20	 Kalokhe A S, Shafiq M, Lee J C, et al. Multidrug-resistant tuberculosis drug
susceptibility and molecular diagnostic testing: a review of the literature.
Am J Med Sci 2013; 345: 143–148.
21	 Thirumurugan R, Kathirvel M, Vallayyachari K, Surendar K, Samrot A V,
Muthaiahb M. Molecular analysis of rpoB gene mutations in rifampicin resis-
tant Mycobacterium tuberculosis isolates by multiple allele specific polymerase
Unsuccessful Xpert results 6Public Health Action
chain reaction in Puducherry, South India. J Infect Public Health 2015; 8:
619–625.
22	 Gidado M, Onazi J, Nwadike P. et al. Assessment of Gene-Xpert MTB/RIF
program implementation and the challenges for enhanced tuberculosis diag-
nosis in Nigeria. SAARC J Tuberc Lung Dis HIV/AIDS 2015; 12: 1–7. https://
www.nepjol.info/index.php/SAARCTB/article/view/15948 Accessed February
2018.
23	 World Health Organization and Global Laboratory Initiative (GLI). Imple-
menters Meeting on Xpert MTB/RIF roll-out, 18–19 April 2012. Annecy,
France: Stop TB Partnership, 2012.
Public Health Action (PHA)  The voice for operational research.
Published by The Union (www.theunion.org), PHA provides a platform to
fulfil its mission, ‘Health solutions for the poor’. PHA publishes high-quality
scientific research that provides new knowledge to improve the accessibility,
equity, quality and efficiency of health systems and services.
e-ISSN 2220-8372
Editor-in-Chief:  Dermot Maher, MD, Switzerland
Contact:  pha@theunion.org
PHA website:  http://www.theunion.org/what-we-do/journals/pha
Article submission:  http://mc.manuscriptcentral.com/pha
Contexte  :  Le Nigeria, pays lourdement frappé par la tuberculose.
Objectif  :  Etudier le taux, la distribution et les causes de mauvais
résultats du test Xpert dans le but d’identifier les domaines clés qui
doivent être renforcés pour une performance optimale du test.
Schéma  :  Analyse rétrospective des données téléchargées entre
janvier et décembre 2015 depuis les structures équipées de Xpert
vers le serveur central à travers le GXAlert.
Résultats  :  Sur 52 219 tests téléchargés à partir de 176 machines,
22,5% ont été positifs pour Mycobacterium tuberculosis, dont 10,8%
ont été résistants à la rifampicine ; globalement, 4,7% ont été
invalides, 4,2% ont eu des résultats erronés et 2,1% n’ont eu aucun
résultat. Les erreurs d’origine technique ont été les plus fréquentes, à
69%, n’ont pas eu de variation saisonnière et sont survenues dans
toutes les zones géopolitiques et à tous les niveaux des structures de
santé. Les erreurs liées à la température ont été prévalentes dans la
région nord-ouest, avec des pics d’avril à juin. Les périodes de pic en
termes de température et de dysfonction des machines ont coïncidé
avec les périodes de faible utilisation du test.
Conclusion  :  Le problème principal qui a affecté la performance du
test a été l’adhérence médiocre aux procédures opératoires
standardisées. Des révisions périodiques de la formation, une
supervision régulière, une maintenance préventive de la machine à
Xpert et un stockage approprié des cartouches constituent des
stratégies susceptibles d’améliorer la performance du Xpert.
Marco de referencia: Nigeria, un país con alta carga de morbilidad
por tuberculosis.
Objetivo: Estudiar la tasa de resultados fallidos de la prueba Xpert,
su distribución y sus causas con el objeto de reconocer las esferas
prioritarias que precisan fortalecimiento, a fin de obtener un
funcionamiento óptimo de la prueba.
Método: Fue este un análisis retrospectivo de los datos enviados al
servidor central por los establecimientos que practican la prueba
Xpert, mediante el sistema GXAlert, de enero a diciembre del 2015.
Resultados: De 52 219 pruebas realizadas en 176 dispositivos y
subidas al sistema, 22,5% fueron positivas para Mycobacterium
tuberculosis y de ellas el 10,8% presentó resistencia a rifampicina; de
todos los resultados, 4,7% fueron inválidos, 4,2% exhibieron error y
2,1% de las pruebas no comportaban un resultado. El error más
frecuente fue el de tipo técnico (69%), el cual no siguió un carácter
estacional y ocurrió en todas las regiones geopolíticas y en
establecimientos de salud de todos los niveles. Los errores debidos a
la temperatura predominaron en la región noroeste, con períodos de
mayor frecuencia de abril a junio. Los períodos de mayor frecuencia
de errores causados por la temperatura o el disfuncionamiento de los
dispositivos coincidieron con épocas de baja utilización de la prueba.
Conclusión: El principal problema que interfirió con el buen
funcionamiento de la prueba fue el incumplimiento de los
procedimientos normalizados de trabajo. Se podría mejorar la eficacia
de la prueba Xpert mediante estrategias como los cursos periódicos
de actualización, la supervisión constante y el mantenimiento
preventivo de los dispositivos, además del almacenamiento adecuado
de los cartuchos de la prueba.

Más contenido relacionado

La actualidad más candente

Best Practices to Risk Based Data Integrity at Data Integrity Conference, Lon...
Best Practices to Risk Based Data Integrity at Data Integrity Conference, Lon...Best Practices to Risk Based Data Integrity at Data Integrity Conference, Lon...
Best Practices to Risk Based Data Integrity at Data Integrity Conference, Lon...Bhaswat Chakraborty
 
A User’s Perspective: ACMG Guidelines for CNVs in VSClinical
A User’s Perspective: ACMG Guidelines for CNVs in VSClinicalA User’s Perspective: ACMG Guidelines for CNVs in VSClinical
A User’s Perspective: ACMG Guidelines for CNVs in VSClinicalGolden Helix
 
Clinical Trial Requirements Medical Devices 27 dec2018
Clinical Trial Requirements Medical Devices 27 dec2018Clinical Trial Requirements Medical Devices 27 dec2018
Clinical Trial Requirements Medical Devices 27 dec2018Bhaswat Chakraborty
 
Implementation of the Defined Approaches on Skin Sensitisation (OECD GL 497) ...
Implementation of the Defined Approaches on Skin Sensitisation (OECD GL 497) ...Implementation of the Defined Approaches on Skin Sensitisation (OECD GL 497) ...
Implementation of the Defined Approaches on Skin Sensitisation (OECD GL 497) ...OECD Environment
 
Virtual clinical trials
Virtual clinical trialsVirtual clinical trials
Virtual clinical trialsDrSahilKumar
 
Developing Protocols & Procedures for CT Data Integrity
Developing Protocols & Procedures for CT Data Integrity Developing Protocols & Procedures for CT Data Integrity
Developing Protocols & Procedures for CT Data Integrity Bhaswat Chakraborty
 
Family-Based Workflows in VarSeq and VSClinical
Family-Based Workflows in VarSeq and VSClinicalFamily-Based Workflows in VarSeq and VSClinical
Family-Based Workflows in VarSeq and VSClinicalGolden Helix
 
Golden Helix's End-to-End Solution for Clinical Labs
Golden Helix's End-to-End Solution for Clinical LabsGolden Helix's End-to-End Solution for Clinical Labs
Golden Helix's End-to-End Solution for Clinical LabsGolden Helix
 
Creating & Managing Reusable Gene Lists with VSClinical
Creating & Managing Reusable Gene Lists with VSClinicalCreating & Managing Reusable Gene Lists with VSClinical
Creating & Managing Reusable Gene Lists with VSClinicalGolden Helix
 
Roundup of This Year's AACC Meeting in Atlanta
Roundup of This Year's AACC Meeting in AtlantaRoundup of This Year's AACC Meeting in Atlanta
Roundup of This Year's AACC Meeting in AtlantaBruce Carlson
 
Lessons Learned From a DDE Phase 2 CT, 2012
Lessons Learned From a DDE Phase 2 CT, 2012Lessons Learned From a DDE Phase 2 CT, 2012
Lessons Learned From a DDE Phase 2 CT, 2012Vadim Tantsyura
 
VSClinical: First Commercial Product to Integrate the Updated ACMG Guidelines...
VSClinical: First Commercial Product to Integrate the Updated ACMG Guidelines...VSClinical: First Commercial Product to Integrate the Updated ACMG Guidelines...
VSClinical: First Commercial Product to Integrate the Updated ACMG Guidelines...Golden Helix
 
Reduce Turn-Around with Enhanced Cancer Annotations and CancerKB Updates
Reduce Turn-Around with Enhanced Cancer Annotations and CancerKB UpdatesReduce Turn-Around with Enhanced Cancer Annotations and CancerKB Updates
Reduce Turn-Around with Enhanced Cancer Annotations and CancerKB UpdatesGolden Helix
 
Pushing the Innovation Envelope: Strategies for Boosting Productivity and ROI...
Pushing the Innovation Envelope: Strategies for Boosting Productivity and ROI...Pushing the Innovation Envelope: Strategies for Boosting Productivity and ROI...
Pushing the Innovation Envelope: Strategies for Boosting Productivity and ROI...Life Sciences Network marcus evans
 
Understanding The Increasing Impact Of Clinical Research On Obtaining Product...
Understanding The Increasing Impact Of Clinical Research On Obtaining Product...Understanding The Increasing Impact Of Clinical Research On Obtaining Product...
Understanding The Increasing Impact Of Clinical Research On Obtaining Product...Gertwbos
 
Clinical Data Management Process Overview_Katalyst HLS
Clinical Data Management Process Overview_Katalyst HLSClinical Data Management Process Overview_Katalyst HLS
Clinical Data Management Process Overview_Katalyst HLSKatalyst HLS
 
Clinical Evaluation of Medical Devices
Clinical Evaluation of Medical Devices Clinical Evaluation of Medical Devices
Clinical Evaluation of Medical Devices Arete-Zoe, LLC
 
Advanced VSClinical Reports with Scripting and Custom Integrations
Advanced VSClinical Reports with Scripting and Custom IntegrationsAdvanced VSClinical Reports with Scripting and Custom Integrations
Advanced VSClinical Reports with Scripting and Custom IntegrationsGolden Helix
 
Risk and quality management Genius Ribam
Risk and quality management Genius RibamRisk and quality management Genius Ribam
Risk and quality management Genius RibamTomasz Moj
 
Efficient Application of NGS Family-Based Analysis
Efficient Application of NGS Family-Based AnalysisEfficient Application of NGS Family-Based Analysis
Efficient Application of NGS Family-Based AnalysisGolden Helix
 

La actualidad más candente (20)

Best Practices to Risk Based Data Integrity at Data Integrity Conference, Lon...
Best Practices to Risk Based Data Integrity at Data Integrity Conference, Lon...Best Practices to Risk Based Data Integrity at Data Integrity Conference, Lon...
Best Practices to Risk Based Data Integrity at Data Integrity Conference, Lon...
 
A User’s Perspective: ACMG Guidelines for CNVs in VSClinical
A User’s Perspective: ACMG Guidelines for CNVs in VSClinicalA User’s Perspective: ACMG Guidelines for CNVs in VSClinical
A User’s Perspective: ACMG Guidelines for CNVs in VSClinical
 
Clinical Trial Requirements Medical Devices 27 dec2018
Clinical Trial Requirements Medical Devices 27 dec2018Clinical Trial Requirements Medical Devices 27 dec2018
Clinical Trial Requirements Medical Devices 27 dec2018
 
Implementation of the Defined Approaches on Skin Sensitisation (OECD GL 497) ...
Implementation of the Defined Approaches on Skin Sensitisation (OECD GL 497) ...Implementation of the Defined Approaches on Skin Sensitisation (OECD GL 497) ...
Implementation of the Defined Approaches on Skin Sensitisation (OECD GL 497) ...
 
Virtual clinical trials
Virtual clinical trialsVirtual clinical trials
Virtual clinical trials
 
Developing Protocols & Procedures for CT Data Integrity
Developing Protocols & Procedures for CT Data Integrity Developing Protocols & Procedures for CT Data Integrity
Developing Protocols & Procedures for CT Data Integrity
 
Family-Based Workflows in VarSeq and VSClinical
Family-Based Workflows in VarSeq and VSClinicalFamily-Based Workflows in VarSeq and VSClinical
Family-Based Workflows in VarSeq and VSClinical
 
Golden Helix's End-to-End Solution for Clinical Labs
Golden Helix's End-to-End Solution for Clinical LabsGolden Helix's End-to-End Solution for Clinical Labs
Golden Helix's End-to-End Solution for Clinical Labs
 
Creating & Managing Reusable Gene Lists with VSClinical
Creating & Managing Reusable Gene Lists with VSClinicalCreating & Managing Reusable Gene Lists with VSClinical
Creating & Managing Reusable Gene Lists with VSClinical
 
Roundup of This Year's AACC Meeting in Atlanta
Roundup of This Year's AACC Meeting in AtlantaRoundup of This Year's AACC Meeting in Atlanta
Roundup of This Year's AACC Meeting in Atlanta
 
Lessons Learned From a DDE Phase 2 CT, 2012
Lessons Learned From a DDE Phase 2 CT, 2012Lessons Learned From a DDE Phase 2 CT, 2012
Lessons Learned From a DDE Phase 2 CT, 2012
 
VSClinical: First Commercial Product to Integrate the Updated ACMG Guidelines...
VSClinical: First Commercial Product to Integrate the Updated ACMG Guidelines...VSClinical: First Commercial Product to Integrate the Updated ACMG Guidelines...
VSClinical: First Commercial Product to Integrate the Updated ACMG Guidelines...
 
Reduce Turn-Around with Enhanced Cancer Annotations and CancerKB Updates
Reduce Turn-Around with Enhanced Cancer Annotations and CancerKB UpdatesReduce Turn-Around with Enhanced Cancer Annotations and CancerKB Updates
Reduce Turn-Around with Enhanced Cancer Annotations and CancerKB Updates
 
Pushing the Innovation Envelope: Strategies for Boosting Productivity and ROI...
Pushing the Innovation Envelope: Strategies for Boosting Productivity and ROI...Pushing the Innovation Envelope: Strategies for Boosting Productivity and ROI...
Pushing the Innovation Envelope: Strategies for Boosting Productivity and ROI...
 
Understanding The Increasing Impact Of Clinical Research On Obtaining Product...
Understanding The Increasing Impact Of Clinical Research On Obtaining Product...Understanding The Increasing Impact Of Clinical Research On Obtaining Product...
Understanding The Increasing Impact Of Clinical Research On Obtaining Product...
 
Clinical Data Management Process Overview_Katalyst HLS
Clinical Data Management Process Overview_Katalyst HLSClinical Data Management Process Overview_Katalyst HLS
Clinical Data Management Process Overview_Katalyst HLS
 
Clinical Evaluation of Medical Devices
Clinical Evaluation of Medical Devices Clinical Evaluation of Medical Devices
Clinical Evaluation of Medical Devices
 
Advanced VSClinical Reports with Scripting and Custom Integrations
Advanced VSClinical Reports with Scripting and Custom IntegrationsAdvanced VSClinical Reports with Scripting and Custom Integrations
Advanced VSClinical Reports with Scripting and Custom Integrations
 
Risk and quality management Genius Ribam
Risk and quality management Genius RibamRisk and quality management Genius Ribam
Risk and quality management Genius Ribam
 
Efficient Application of NGS Family-Based Analysis
Efficient Application of NGS Family-Based AnalysisEfficient Application of NGS Family-Based Analysis
Efficient Application of NGS Family-Based Analysis
 

Similar a Unsuccessful Xpert® MTB/RIF Results in Nigeria

On the ground experiences & challenges of a connected diagnostics GxAlert in ...
On the ground experiences & challenges of a connected diagnostics GxAlert in ...On the ground experiences & challenges of a connected diagnostics GxAlert in ...
On the ground experiences & challenges of a connected diagnostics GxAlert in ...SystemOne
 
Point Of Care Testing
Point Of Care TestingPoint Of Care Testing
Point Of Care TestingLAB IDEA
 
GLI TB Diagnostics Connectivity Guide 2016
GLI TB Diagnostics Connectivity Guide 2016GLI TB Diagnostics Connectivity Guide 2016
GLI TB Diagnostics Connectivity Guide 2016SystemOne
 
From the Archives, 2008:Clinical and Economic Advantages Implantable Defibril...
From the Archives, 2008:Clinical and Economic Advantages Implantable Defibril...From the Archives, 2008:Clinical and Economic Advantages Implantable Defibril...
From the Archives, 2008:Clinical and Economic Advantages Implantable Defibril...David Lee Scher, MD
 
Health informatics - Transforming healthcare delivery in hong kong
Health informatics - Transforming healthcare delivery in hong kongHealth informatics - Transforming healthcare delivery in hong kong
Health informatics - Transforming healthcare delivery in hong kongHealth Informatics New Zealand
 
Racing for results: lessons learnt in improving the efficiency of HIV VL and ...
Racing for results: lessons learnt in improving the efficiency of HIV VL and ...Racing for results: lessons learnt in improving the efficiency of HIV VL and ...
Racing for results: lessons learnt in improving the efficiency of HIV VL and ...SystemOne
 
Getting Right with The Joint Commission's Communication Goal
Getting Right with The Joint Commission's Communication GoalGetting Right with The Joint Commission's Communication Goal
Getting Right with The Joint Commission's Communication GoalSpok
 
Evoulution Summit_November_2015_Sundgren
Evoulution Summit_November_2015_SundgrenEvoulution Summit_November_2015_Sundgren
Evoulution Summit_November_2015_SundgrenMats Sundgren
 
Coronavirus Pandemic Part III (A): Diagnostic Technologies
Coronavirus Pandemic Part III (A): Diagnostic TechnologiesCoronavirus Pandemic Part III (A): Diagnostic Technologies
Coronavirus Pandemic Part III (A): Diagnostic TechnologiesKumaraguru Veerasamy
 
ecoa clinical trials.pdf
ecoa clinical trials.pdfecoa clinical trials.pdf
ecoa clinical trials.pdfprocth2
 
Nichols-Scott-Kite-Pharma-2023.pdf
Nichols-Scott-Kite-Pharma-2023.pdfNichols-Scott-Kite-Pharma-2023.pdf
Nichols-Scott-Kite-Pharma-2023.pdfRichardShi22
 
Post-Market-Surveillance_Lesotho-Case-Study_11Oct0217.pptx
Post-Market-Surveillance_Lesotho-Case-Study_11Oct0217.pptxPost-Market-Surveillance_Lesotho-Case-Study_11Oct0217.pptx
Post-Market-Surveillance_Lesotho-Case-Study_11Oct0217.pptxtribowofauzan
 
Post-Market-Surveillance_Lesotho-Case-Study_11Oct0217.pptx
Post-Market-Surveillance_Lesotho-Case-Study_11Oct0217.pptxPost-Market-Surveillance_Lesotho-Case-Study_11Oct0217.pptx
Post-Market-Surveillance_Lesotho-Case-Study_11Oct0217.pptxbiruktesfaye27
 
Award-Winning Device Completes First Semi-Automated System for HPV Screening
Award-Winning Device Completes First Semi-Automated System for HPV ScreeningAward-Winning Device Completes First Semi-Automated System for HPV Screening
Award-Winning Device Completes First Semi-Automated System for HPV ScreeningInhand Electronics Incorporated
 
FAILURE MODE AND EFFECT ANALYSIS (FMEA) WITH SPECIAL REFERENCE TO ICU, EMERGE...
FAILURE MODE AND EFFECT ANALYSIS (FMEA) WITH SPECIAL REFERENCE TO ICU, EMERGE...FAILURE MODE AND EFFECT ANALYSIS (FMEA) WITH SPECIAL REFERENCE TO ICU, EMERGE...
FAILURE MODE AND EFFECT ANALYSIS (FMEA) WITH SPECIAL REFERENCE TO ICU, EMERGE...IRJET Journal
 
Safety Monitoring and Reporting in Clinical Trials DIA Poster 2015
Safety Monitoring and Reporting in Clinical Trials DIA Poster 2015Safety Monitoring and Reporting in Clinical Trials DIA Poster 2015
Safety Monitoring and Reporting in Clinical Trials DIA Poster 2015KCR
 
EQA: An Essential Tool for Laboratory Quality - Africa LT649
EQA: An Essential Tool for Laboratory Quality - Africa LT649 EQA: An Essential Tool for Laboratory Quality - Africa LT649
EQA: An Essential Tool for Laboratory Quality - Africa LT649 Randox
 
The Use of EDC in Canadian Clinical Trials
The Use of EDC in Canadian Clinical TrialsThe Use of EDC in Canadian Clinical Trials
The Use of EDC in Canadian Clinical TrialsKhaled El Emam
 
DEC 3 2015 Mats Sundgren
DEC 3 2015 Mats SundgrenDEC 3 2015 Mats Sundgren
DEC 3 2015 Mats SundgrenGBX Events
 

Similar a Unsuccessful Xpert® MTB/RIF Results in Nigeria (20)

On the ground experiences & challenges of a connected diagnostics GxAlert in ...
On the ground experiences & challenges of a connected diagnostics GxAlert in ...On the ground experiences & challenges of a connected diagnostics GxAlert in ...
On the ground experiences & challenges of a connected diagnostics GxAlert in ...
 
Point Of Care Testing
Point Of Care TestingPoint Of Care Testing
Point Of Care Testing
 
GLI TB Diagnostics Connectivity Guide 2016
GLI TB Diagnostics Connectivity Guide 2016GLI TB Diagnostics Connectivity Guide 2016
GLI TB Diagnostics Connectivity Guide 2016
 
From the Archives, 2008:Clinical and Economic Advantages Implantable Defibril...
From the Archives, 2008:Clinical and Economic Advantages Implantable Defibril...From the Archives, 2008:Clinical and Economic Advantages Implantable Defibril...
From the Archives, 2008:Clinical and Economic Advantages Implantable Defibril...
 
Health informatics - Transforming healthcare delivery in hong kong
Health informatics - Transforming healthcare delivery in hong kongHealth informatics - Transforming healthcare delivery in hong kong
Health informatics - Transforming healthcare delivery in hong kong
 
Racing for results: lessons learnt in improving the efficiency of HIV VL and ...
Racing for results: lessons learnt in improving the efficiency of HIV VL and ...Racing for results: lessons learnt in improving the efficiency of HIV VL and ...
Racing for results: lessons learnt in improving the efficiency of HIV VL and ...
 
Getting Right with The Joint Commission's Communication Goal
Getting Right with The Joint Commission's Communication GoalGetting Right with The Joint Commission's Communication Goal
Getting Right with The Joint Commission's Communication Goal
 
Telenephrology
TelenephrologyTelenephrology
Telenephrology
 
Evoulution Summit_November_2015_Sundgren
Evoulution Summit_November_2015_SundgrenEvoulution Summit_November_2015_Sundgren
Evoulution Summit_November_2015_Sundgren
 
Coronavirus Pandemic Part III (A): Diagnostic Technologies
Coronavirus Pandemic Part III (A): Diagnostic TechnologiesCoronavirus Pandemic Part III (A): Diagnostic Technologies
Coronavirus Pandemic Part III (A): Diagnostic Technologies
 
ecoa clinical trials.pdf
ecoa clinical trials.pdfecoa clinical trials.pdf
ecoa clinical trials.pdf
 
Nichols-Scott-Kite-Pharma-2023.pdf
Nichols-Scott-Kite-Pharma-2023.pdfNichols-Scott-Kite-Pharma-2023.pdf
Nichols-Scott-Kite-Pharma-2023.pdf
 
Post-Market-Surveillance_Lesotho-Case-Study_11Oct0217.pptx
Post-Market-Surveillance_Lesotho-Case-Study_11Oct0217.pptxPost-Market-Surveillance_Lesotho-Case-Study_11Oct0217.pptx
Post-Market-Surveillance_Lesotho-Case-Study_11Oct0217.pptx
 
Post-Market-Surveillance_Lesotho-Case-Study_11Oct0217.pptx
Post-Market-Surveillance_Lesotho-Case-Study_11Oct0217.pptxPost-Market-Surveillance_Lesotho-Case-Study_11Oct0217.pptx
Post-Market-Surveillance_Lesotho-Case-Study_11Oct0217.pptx
 
Award-Winning Device Completes First Semi-Automated System for HPV Screening
Award-Winning Device Completes First Semi-Automated System for HPV ScreeningAward-Winning Device Completes First Semi-Automated System for HPV Screening
Award-Winning Device Completes First Semi-Automated System for HPV Screening
 
FAILURE MODE AND EFFECT ANALYSIS (FMEA) WITH SPECIAL REFERENCE TO ICU, EMERGE...
FAILURE MODE AND EFFECT ANALYSIS (FMEA) WITH SPECIAL REFERENCE TO ICU, EMERGE...FAILURE MODE AND EFFECT ANALYSIS (FMEA) WITH SPECIAL REFERENCE TO ICU, EMERGE...
FAILURE MODE AND EFFECT ANALYSIS (FMEA) WITH SPECIAL REFERENCE TO ICU, EMERGE...
 
Safety Monitoring and Reporting in Clinical Trials DIA Poster 2015
Safety Monitoring and Reporting in Clinical Trials DIA Poster 2015Safety Monitoring and Reporting in Clinical Trials DIA Poster 2015
Safety Monitoring and Reporting in Clinical Trials DIA Poster 2015
 
EQA: An Essential Tool for Laboratory Quality - Africa LT649
EQA: An Essential Tool for Laboratory Quality - Africa LT649 EQA: An Essential Tool for Laboratory Quality - Africa LT649
EQA: An Essential Tool for Laboratory Quality - Africa LT649
 
The Use of EDC in Canadian Clinical Trials
The Use of EDC in Canadian Clinical TrialsThe Use of EDC in Canadian Clinical Trials
The Use of EDC in Canadian Clinical Trials
 
DEC 3 2015 Mats Sundgren
DEC 3 2015 Mats SundgrenDEC 3 2015 Mats Sundgren
DEC 3 2015 Mats Sundgren
 

Más de SystemOne

Feasibility of an SMS intervention to deliver tuberculosis testing results in...
Feasibility of an SMS intervention to deliver tuberculosis testing results in...Feasibility of an SMS intervention to deliver tuberculosis testing results in...
Feasibility of an SMS intervention to deliver tuberculosis testing results in...SystemOne
 
The Role of Connected Diagnostics in Strengthening Regional, National and Con...
The Role of Connected Diagnostics in Strengthening Regional, National and Con...The Role of Connected Diagnostics in Strengthening Regional, National and Con...
The Role of Connected Diagnostics in Strengthening Regional, National and Con...SystemOne
 
GxAlert for Real-time Management and Strengthening of Remote GeneXpert Networ...
GxAlert for Real-time Management and Strengthening of Remote GeneXpert Networ...GxAlert for Real-time Management and Strengthening of Remote GeneXpert Networ...
GxAlert for Real-time Management and Strengthening of Remote GeneXpert Networ...SystemOne
 
GxAlert Papua New Guinea Case Study 072518
GxAlert Papua New Guinea Case Study 072518GxAlert Papua New Guinea Case Study 072518
GxAlert Papua New Guinea Case Study 072518SystemOne
 
Gx alert casestudy bangladesh_062618
Gx alert casestudy bangladesh_062618Gx alert casestudy bangladesh_062618
Gx alert casestudy bangladesh_062618SystemOne
 
The impact of digital technologies on point of care diagnostics in resource l...
The impact of digital technologies on point of care diagnostics in resource l...The impact of digital technologies on point of care diagnostics in resource l...
The impact of digital technologies on point of care diagnostics in resource l...SystemOne
 
Diagnosing at Point of Care in South India: Coordination Work and Frictions
Diagnosing at Point of Care in South India: Coordination Work and Frictions Diagnosing at Point of Care in South India: Coordination Work and Frictions
Diagnosing at Point of Care in South India: Coordination Work and Frictions SystemOne
 
Surrogate endpoints in global health research: still searching for killer app...
Surrogate endpoints in global health research: still searching for killer app...Surrogate endpoints in global health research: still searching for killer app...
Surrogate endpoints in global health research: still searching for killer app...SystemOne
 
CROI 2018 Poster #989
CROI 2018 Poster #989CROI 2018 Poster #989
CROI 2018 Poster #989SystemOne
 
2018: FIND AMR Strategy
2018: FIND AMR Strategy2018: FIND AMR Strategy
2018: FIND AMR StrategySystemOne
 
Implementing rapid testing for tuberculosis in Mozambique
Implementing rapid testing for tuberculosis in MozambiqueImplementing rapid testing for tuberculosis in Mozambique
Implementing rapid testing for tuberculosis in MozambiqueSystemOne
 
Remote monitoring of XpertW MTB/RIF testing in Mozambique: results of program...
Remote monitoring of XpertW MTB/RIF testing in Mozambique: results of program...Remote monitoring of XpertW MTB/RIF testing in Mozambique: results of program...
Remote monitoring of XpertW MTB/RIF testing in Mozambique: results of program...SystemOne
 
Using Connected Diagnostics to Implement The New WHO Shorter MDR Treatment Re...
Using Connected Diagnostics to Implement The New WHO Shorter MDR Treatment Re...Using Connected Diagnostics to Implement The New WHO Shorter MDR Treatment Re...
Using Connected Diagnostics to Implement The New WHO Shorter MDR Treatment Re...SystemOne
 
Continuous Quality Improvement (CQI) Through Connected Diagnostics
Continuous Quality Improvement (CQI) Through Connected DiagnosticsContinuous Quality Improvement (CQI) Through Connected Diagnostics
Continuous Quality Improvement (CQI) Through Connected DiagnosticsSystemOne
 
Poster: Staying in Control of Lab Data in a World of Connected Diagnostics
Poster: Staying in Control of Lab Data in a World of Connected Diagnostics Poster: Staying in Control of Lab Data in a World of Connected Diagnostics
Poster: Staying in Control of Lab Data in a World of Connected Diagnostics SystemOne
 
Introduction of GxAlert in Uganda
Introduction of GxAlert in UgandaIntroduction of GxAlert in Uganda
Introduction of GxAlert in UgandaSystemOne
 
History of Use of GxAlert in Nigeria
History of Use of GxAlert in NigeriaHistory of Use of GxAlert in Nigeria
History of Use of GxAlert in NigeriaSystemOne
 
March 2016 Connectivity of Diagnostic Technologies
March 2016 Connectivity of Diagnostic TechnologiesMarch 2016 Connectivity of Diagnostic Technologies
March 2016 Connectivity of Diagnostic TechnologiesSystemOne
 
Albert ERJ 2016 XPert Lessons
Albert ERJ 2016 XPert LessonsAlbert ERJ 2016 XPert Lessons
Albert ERJ 2016 XPert LessonsSystemOne
 
Effects of GeneXpert MTB/RIF Testing and GxAlert eHealth Platforms on MDR-TB ...
Effects of GeneXpert MTB/RIF Testing and GxAlert eHealth Platforms on MDR-TB ...Effects of GeneXpert MTB/RIF Testing and GxAlert eHealth Platforms on MDR-TB ...
Effects of GeneXpert MTB/RIF Testing and GxAlert eHealth Platforms on MDR-TB ...SystemOne
 

Más de SystemOne (20)

Feasibility of an SMS intervention to deliver tuberculosis testing results in...
Feasibility of an SMS intervention to deliver tuberculosis testing results in...Feasibility of an SMS intervention to deliver tuberculosis testing results in...
Feasibility of an SMS intervention to deliver tuberculosis testing results in...
 
The Role of Connected Diagnostics in Strengthening Regional, National and Con...
The Role of Connected Diagnostics in Strengthening Regional, National and Con...The Role of Connected Diagnostics in Strengthening Regional, National and Con...
The Role of Connected Diagnostics in Strengthening Regional, National and Con...
 
GxAlert for Real-time Management and Strengthening of Remote GeneXpert Networ...
GxAlert for Real-time Management and Strengthening of Remote GeneXpert Networ...GxAlert for Real-time Management and Strengthening of Remote GeneXpert Networ...
GxAlert for Real-time Management and Strengthening of Remote GeneXpert Networ...
 
GxAlert Papua New Guinea Case Study 072518
GxAlert Papua New Guinea Case Study 072518GxAlert Papua New Guinea Case Study 072518
GxAlert Papua New Guinea Case Study 072518
 
Gx alert casestudy bangladesh_062618
Gx alert casestudy bangladesh_062618Gx alert casestudy bangladesh_062618
Gx alert casestudy bangladesh_062618
 
The impact of digital technologies on point of care diagnostics in resource l...
The impact of digital technologies on point of care diagnostics in resource l...The impact of digital technologies on point of care diagnostics in resource l...
The impact of digital technologies on point of care diagnostics in resource l...
 
Diagnosing at Point of Care in South India: Coordination Work and Frictions
Diagnosing at Point of Care in South India: Coordination Work and Frictions Diagnosing at Point of Care in South India: Coordination Work and Frictions
Diagnosing at Point of Care in South India: Coordination Work and Frictions
 
Surrogate endpoints in global health research: still searching for killer app...
Surrogate endpoints in global health research: still searching for killer app...Surrogate endpoints in global health research: still searching for killer app...
Surrogate endpoints in global health research: still searching for killer app...
 
CROI 2018 Poster #989
CROI 2018 Poster #989CROI 2018 Poster #989
CROI 2018 Poster #989
 
2018: FIND AMR Strategy
2018: FIND AMR Strategy2018: FIND AMR Strategy
2018: FIND AMR Strategy
 
Implementing rapid testing for tuberculosis in Mozambique
Implementing rapid testing for tuberculosis in MozambiqueImplementing rapid testing for tuberculosis in Mozambique
Implementing rapid testing for tuberculosis in Mozambique
 
Remote monitoring of XpertW MTB/RIF testing in Mozambique: results of program...
Remote monitoring of XpertW MTB/RIF testing in Mozambique: results of program...Remote monitoring of XpertW MTB/RIF testing in Mozambique: results of program...
Remote monitoring of XpertW MTB/RIF testing in Mozambique: results of program...
 
Using Connected Diagnostics to Implement The New WHO Shorter MDR Treatment Re...
Using Connected Diagnostics to Implement The New WHO Shorter MDR Treatment Re...Using Connected Diagnostics to Implement The New WHO Shorter MDR Treatment Re...
Using Connected Diagnostics to Implement The New WHO Shorter MDR Treatment Re...
 
Continuous Quality Improvement (CQI) Through Connected Diagnostics
Continuous Quality Improvement (CQI) Through Connected DiagnosticsContinuous Quality Improvement (CQI) Through Connected Diagnostics
Continuous Quality Improvement (CQI) Through Connected Diagnostics
 
Poster: Staying in Control of Lab Data in a World of Connected Diagnostics
Poster: Staying in Control of Lab Data in a World of Connected Diagnostics Poster: Staying in Control of Lab Data in a World of Connected Diagnostics
Poster: Staying in Control of Lab Data in a World of Connected Diagnostics
 
Introduction of GxAlert in Uganda
Introduction of GxAlert in UgandaIntroduction of GxAlert in Uganda
Introduction of GxAlert in Uganda
 
History of Use of GxAlert in Nigeria
History of Use of GxAlert in NigeriaHistory of Use of GxAlert in Nigeria
History of Use of GxAlert in Nigeria
 
March 2016 Connectivity of Diagnostic Technologies
March 2016 Connectivity of Diagnostic TechnologiesMarch 2016 Connectivity of Diagnostic Technologies
March 2016 Connectivity of Diagnostic Technologies
 
Albert ERJ 2016 XPert Lessons
Albert ERJ 2016 XPert LessonsAlbert ERJ 2016 XPert Lessons
Albert ERJ 2016 XPert Lessons
 
Effects of GeneXpert MTB/RIF Testing and GxAlert eHealth Platforms on MDR-TB ...
Effects of GeneXpert MTB/RIF Testing and GxAlert eHealth Platforms on MDR-TB ...Effects of GeneXpert MTB/RIF Testing and GxAlert eHealth Platforms on MDR-TB ...
Effects of GeneXpert MTB/RIF Testing and GxAlert eHealth Platforms on MDR-TB ...
 

Último

Call Girls LB Nagar 7001305949 all area service COD available Any Time
Call Girls LB Nagar 7001305949 all area service COD available Any TimeCall Girls LB Nagar 7001305949 all area service COD available Any Time
Call Girls LB Nagar 7001305949 all area service COD available Any Timedelhimodelshub1
 
College Call Girls Mumbai Alia 9910780858 Independent Escort Service Mumbai
College Call Girls Mumbai Alia 9910780858 Independent Escort Service MumbaiCollege Call Girls Mumbai Alia 9910780858 Independent Escort Service Mumbai
College Call Girls Mumbai Alia 9910780858 Independent Escort Service Mumbaisonalikaur4
 
Book Call Girls in Noida Pick Up Drop With Cash Payment 9711199171 Call Girls
Book Call Girls in Noida Pick Up Drop With Cash Payment 9711199171 Call GirlsBook Call Girls in Noida Pick Up Drop With Cash Payment 9711199171 Call Girls
Book Call Girls in Noida Pick Up Drop With Cash Payment 9711199171 Call GirlsCall Girls Noida
 
Call Girls Service Chandigarh Grishma ❤️🍑 9907093804 👄🫦 Independent Escort Se...
Call Girls Service Chandigarh Grishma ❤️🍑 9907093804 👄🫦 Independent Escort Se...Call Girls Service Chandigarh Grishma ❤️🍑 9907093804 👄🫦 Independent Escort Se...
Call Girls Service Chandigarh Grishma ❤️🍑 9907093804 👄🫦 Independent Escort Se...High Profile Call Girls Chandigarh Aarushi
 
Russian Call Girls in Goa Samaira 7001305949 Independent Escort Service Goa
Russian Call Girls in Goa Samaira 7001305949 Independent Escort Service GoaRussian Call Girls in Goa Samaira 7001305949 Independent Escort Service Goa
Russian Call Girls in Goa Samaira 7001305949 Independent Escort Service Goanarwatsonia7
 
Basics of Anatomy- Language of Anatomy.pptx
Basics of Anatomy- Language of Anatomy.pptxBasics of Anatomy- Language of Anatomy.pptx
Basics of Anatomy- Language of Anatomy.pptxAyushGupta813444
 
Kukatpally Call Girls Services 9907093804 High Class Babes Here Call Now
Kukatpally Call Girls Services 9907093804 High Class Babes Here Call NowKukatpally Call Girls Services 9907093804 High Class Babes Here Call Now
Kukatpally Call Girls Services 9907093804 High Class Babes Here Call NowHyderabad Call Girls Services
 
Call Girl Service ITPL - [ Cash on Delivery ] Contact 7001305949 Escorts Service
Call Girl Service ITPL - [ Cash on Delivery ] Contact 7001305949 Escorts ServiceCall Girl Service ITPL - [ Cash on Delivery ] Contact 7001305949 Escorts Service
Call Girl Service ITPL - [ Cash on Delivery ] Contact 7001305949 Escorts Servicenarwatsonia7
 
Call Girls Kukatpally 7001305949 all area service COD available Any Time
Call Girls Kukatpally 7001305949 all area service COD available Any TimeCall Girls Kukatpally 7001305949 all area service COD available Any Time
Call Girls Kukatpally 7001305949 all area service COD available Any Timedelhimodelshub1
 
Gurgaon Sector 90 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few ...
Gurgaon Sector 90 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few ...Gurgaon Sector 90 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few ...
Gurgaon Sector 90 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few ...ggsonu500
 
hyderabad call girl.pdfRussian Call Girls in Hyderabad Amrita 9907093804 Inde...
hyderabad call girl.pdfRussian Call Girls in Hyderabad Amrita 9907093804 Inde...hyderabad call girl.pdfRussian Call Girls in Hyderabad Amrita 9907093804 Inde...
hyderabad call girl.pdfRussian Call Girls in Hyderabad Amrita 9907093804 Inde...delhimodelshub1
 
Hi,Fi Call Girl In Whitefield - [ Cash on Delivery ] Contact 7001305949 Escor...
Hi,Fi Call Girl In Whitefield - [ Cash on Delivery ] Contact 7001305949 Escor...Hi,Fi Call Girl In Whitefield - [ Cash on Delivery ] Contact 7001305949 Escor...
Hi,Fi Call Girl In Whitefield - [ Cash on Delivery ] Contact 7001305949 Escor...narwatsonia7
 
Housewife Call Girls Nandini Layout - Phone No 7001305949 For Ultimate Sexual...
Housewife Call Girls Nandini Layout - Phone No 7001305949 For Ultimate Sexual...Housewife Call Girls Nandini Layout - Phone No 7001305949 For Ultimate Sexual...
Housewife Call Girls Nandini Layout - Phone No 7001305949 For Ultimate Sexual...narwatsonia7
 
2025 Inpatient Prospective Payment System (IPPS) Proposed Rule
2025 Inpatient Prospective Payment System (IPPS) Proposed Rule2025 Inpatient Prospective Payment System (IPPS) Proposed Rule
2025 Inpatient Prospective Payment System (IPPS) Proposed RuleShelby Lewis
 
Call Girl Chandigarh Mallika ❤️🍑 9907093804 👄🫦 Independent Escort Service Cha...
Call Girl Chandigarh Mallika ❤️🍑 9907093804 👄🫦 Independent Escort Service Cha...Call Girl Chandigarh Mallika ❤️🍑 9907093804 👄🫦 Independent Escort Service Cha...
Call Girl Chandigarh Mallika ❤️🍑 9907093804 👄🫦 Independent Escort Service Cha...High Profile Call Girls Chandigarh Aarushi
 
EMS and Extrication: Coordinating Critical Care
EMS and Extrication: Coordinating Critical CareEMS and Extrication: Coordinating Critical Care
EMS and Extrication: Coordinating Critical CareRommie Duckworth
 

Último (20)

Call Girls LB Nagar 7001305949 all area service COD available Any Time
Call Girls LB Nagar 7001305949 all area service COD available Any TimeCall Girls LB Nagar 7001305949 all area service COD available Any Time
Call Girls LB Nagar 7001305949 all area service COD available Any Time
 
College Call Girls Mumbai Alia 9910780858 Independent Escort Service Mumbai
College Call Girls Mumbai Alia 9910780858 Independent Escort Service MumbaiCollege Call Girls Mumbai Alia 9910780858 Independent Escort Service Mumbai
College Call Girls Mumbai Alia 9910780858 Independent Escort Service Mumbai
 
Call Girl Guwahati Aashi 👉 7001305949 👈 🔝 Independent Escort Service Guwahati
Call Girl Guwahati Aashi 👉 7001305949 👈 🔝 Independent Escort Service GuwahatiCall Girl Guwahati Aashi 👉 7001305949 👈 🔝 Independent Escort Service Guwahati
Call Girl Guwahati Aashi 👉 7001305949 👈 🔝 Independent Escort Service Guwahati
 
Book Call Girls in Noida Pick Up Drop With Cash Payment 9711199171 Call Girls
Book Call Girls in Noida Pick Up Drop With Cash Payment 9711199171 Call GirlsBook Call Girls in Noida Pick Up Drop With Cash Payment 9711199171 Call Girls
Book Call Girls in Noida Pick Up Drop With Cash Payment 9711199171 Call Girls
 
Call Girl Lucknow Gauri 🔝 8923113531 🔝 🎶 Independent Escort Service Lucknow
Call Girl Lucknow Gauri 🔝 8923113531  🔝 🎶 Independent Escort Service LucknowCall Girl Lucknow Gauri 🔝 8923113531  🔝 🎶 Independent Escort Service Lucknow
Call Girl Lucknow Gauri 🔝 8923113531 🔝 🎶 Independent Escort Service Lucknow
 
Call Girls Service Chandigarh Grishma ❤️🍑 9907093804 👄🫦 Independent Escort Se...
Call Girls Service Chandigarh Grishma ❤️🍑 9907093804 👄🫦 Independent Escort Se...Call Girls Service Chandigarh Grishma ❤️🍑 9907093804 👄🫦 Independent Escort Se...
Call Girls Service Chandigarh Grishma ❤️🍑 9907093804 👄🫦 Independent Escort Se...
 
Russian Call Girls in Goa Samaira 7001305949 Independent Escort Service Goa
Russian Call Girls in Goa Samaira 7001305949 Independent Escort Service GoaRussian Call Girls in Goa Samaira 7001305949 Independent Escort Service Goa
Russian Call Girls in Goa Samaira 7001305949 Independent Escort Service Goa
 
Call Girls Guwahati Aaradhya 👉 7001305949👈 🎶 Independent Escort Service Guwahati
Call Girls Guwahati Aaradhya 👉 7001305949👈 🎶 Independent Escort Service GuwahatiCall Girls Guwahati Aaradhya 👉 7001305949👈 🎶 Independent Escort Service Guwahati
Call Girls Guwahati Aaradhya 👉 7001305949👈 🎶 Independent Escort Service Guwahati
 
Basics of Anatomy- Language of Anatomy.pptx
Basics of Anatomy- Language of Anatomy.pptxBasics of Anatomy- Language of Anatomy.pptx
Basics of Anatomy- Language of Anatomy.pptx
 
Kukatpally Call Girls Services 9907093804 High Class Babes Here Call Now
Kukatpally Call Girls Services 9907093804 High Class Babes Here Call NowKukatpally Call Girls Services 9907093804 High Class Babes Here Call Now
Kukatpally Call Girls Services 9907093804 High Class Babes Here Call Now
 
Call Girl Service ITPL - [ Cash on Delivery ] Contact 7001305949 Escorts Service
Call Girl Service ITPL - [ Cash on Delivery ] Contact 7001305949 Escorts ServiceCall Girl Service ITPL - [ Cash on Delivery ] Contact 7001305949 Escorts Service
Call Girl Service ITPL - [ Cash on Delivery ] Contact 7001305949 Escorts Service
 
Call Girls Kukatpally 7001305949 all area service COD available Any Time
Call Girls Kukatpally 7001305949 all area service COD available Any TimeCall Girls Kukatpally 7001305949 all area service COD available Any Time
Call Girls Kukatpally 7001305949 all area service COD available Any Time
 
Gurgaon Sector 90 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few ...
Gurgaon Sector 90 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few ...Gurgaon Sector 90 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few ...
Gurgaon Sector 90 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few ...
 
hyderabad call girl.pdfRussian Call Girls in Hyderabad Amrita 9907093804 Inde...
hyderabad call girl.pdfRussian Call Girls in Hyderabad Amrita 9907093804 Inde...hyderabad call girl.pdfRussian Call Girls in Hyderabad Amrita 9907093804 Inde...
hyderabad call girl.pdfRussian Call Girls in Hyderabad Amrita 9907093804 Inde...
 
Hi,Fi Call Girl In Whitefield - [ Cash on Delivery ] Contact 7001305949 Escor...
Hi,Fi Call Girl In Whitefield - [ Cash on Delivery ] Contact 7001305949 Escor...Hi,Fi Call Girl In Whitefield - [ Cash on Delivery ] Contact 7001305949 Escor...
Hi,Fi Call Girl In Whitefield - [ Cash on Delivery ] Contact 7001305949 Escor...
 
Housewife Call Girls Nandini Layout - Phone No 7001305949 For Ultimate Sexual...
Housewife Call Girls Nandini Layout - Phone No 7001305949 For Ultimate Sexual...Housewife Call Girls Nandini Layout - Phone No 7001305949 For Ultimate Sexual...
Housewife Call Girls Nandini Layout - Phone No 7001305949 For Ultimate Sexual...
 
2025 Inpatient Prospective Payment System (IPPS) Proposed Rule
2025 Inpatient Prospective Payment System (IPPS) Proposed Rule2025 Inpatient Prospective Payment System (IPPS) Proposed Rule
2025 Inpatient Prospective Payment System (IPPS) Proposed Rule
 
Call Girl Chandigarh Mallika ❤️🍑 9907093804 👄🫦 Independent Escort Service Cha...
Call Girl Chandigarh Mallika ❤️🍑 9907093804 👄🫦 Independent Escort Service Cha...Call Girl Chandigarh Mallika ❤️🍑 9907093804 👄🫦 Independent Escort Service Cha...
Call Girl Chandigarh Mallika ❤️🍑 9907093804 👄🫦 Independent Escort Service Cha...
 
College Call Girls Dehradun Kavya 🔝 7001305949 🔝 📍 Independent Escort Service...
College Call Girls Dehradun Kavya 🔝 7001305949 🔝 📍 Independent Escort Service...College Call Girls Dehradun Kavya 🔝 7001305949 🔝 📍 Independent Escort Service...
College Call Girls Dehradun Kavya 🔝 7001305949 🔝 📍 Independent Escort Service...
 
EMS and Extrication: Coordinating Critical Care
EMS and Extrication: Coordinating Critical CareEMS and Extrication: Coordinating Critical Care
EMS and Extrication: Coordinating Critical Care
 

Unsuccessful Xpert® MTB/RIF Results in Nigeria

  • 1. Public Health Action International Union Against Tuberculosis and Lung Disease Health solutions for the poor VOL 8 no 1  PUBLISHED 21 MARCH 2018 PHA 2018; 8(1): 2–6 © 2018 The Union AFFILIATIONS 1 KNCV Tuberculosis Foundation, Abuja, Nigeria 2 National Tuberculosis Leprosy Control Programme, Abuja, Nigeria CORRESPONDENCE Nkiru Nwokoye Laboratory, KNCV Tuberculosis Foundation 564/565 Independence Avenue Central Business District Abuja FCT 900211, Nigeria e-mail: nkiru.nwokoye@ kncvtbc.org ACKNOWLEDGEMENTS The authors thank the Nigerian National Tuberculosis Programme and all partners investing in Xpert implementation for their support, including the US Agency for International Development (Washington DC, USA) and AB Associate (Paris, France) for GXAlert. Conflicts of interest: none declared. KEY WORDS tuberculosis; diagnosis; implementation; error Unsuccessful Xpert® MTB/RIF results: the Nigerian experience M. Gidado,1 N. Nwokoye,1 P. Nwadike,1 P. Ajiboye,1 R. Eneogu,1 S. Useni,1 J. Onazi,1 A. Lawanson,2 E. Elom,2 A. Tubi,2 J. Kuye2 Early diagnosis and access to effective treatment re- mains fundamental to successful tuberculosis (TB) control.1 TB diagnosis using smear microscopy takes 24–48 h, while solid culture takes 4–8 weeks. This diag- nostic delay is an obstacle to achieving early diagno- sis.2,3 The introduction of the Xpert® MTB/RIF assay (Cepheid, Sunnyvale, CA, USA) revolutionised TB diag- nosis and gave hope to struggling programmes. Today, diagnostic decisions on both drug-susceptible (DS-TB) and rifampicin (RMP) resistant TB (RR-TB) can be made within 2 h of sample collection, thus significantly re- ducing the lag time for diagnosis and facilitating prompt clinical decision making, which benefits TB control.4,5 Following the World Health Organization’s (WHO’s) endorsement of Xpert, many countries, in- cluding Nigeria, embraced this technology with opti- mism.6 It is the first simple TB molecular test that is robust enough to be introduced outside conventional laboratory settings.7 However, the right placement of Xpert machines in the laboratory requires careful con- sideration of available resources, national capacity building and accessibility of diagnostic testing.8 Xpert roll-out Some key considerations that had to be taken into ac- count when selecting sites for the installation of the Xpert machines included high TB burden sites, cur- rent workload of the facility, availability of personnel, adequate infrastructure and the capacity of the appro- priate treatment centre. It was thus possible to train a cadre of staff capable of providing technical assistance during the early stages of implementation. Installation of Xpert machines commenced at na- tional- and state-level laboratories with a relatively sta- ble power supply that would ensure uninterrupted run- ning of the Xpert assay. Setting up of air conditioners, refrigerators and inverters was essential to provide the ambient temperature required to operate the Xpert ma- chines and store the cartridges, and to guarantee an un- interrupted power supply during testing. GxAlert inter- net data connectivity was also established for the upload of data onto the central server. A 4-day theory and prac- tical training course was provided for laboratory staff and other health care workers at the time of the installa- tion. The supply and distribution of cartridges to facili- ties was centrally managed by the National Tuberculosis and Leprosy Control Programme (NTBLCP). At facility level, trained laboratory personnel per- formed the assay according to the protocol and na- tional standard operating procedures (SOPs). Based on the SOPs, unsuccessful tests showing errors or invalid or incomplete results were repeated. The resolution of error results was dependent on the error codes. Prob- lematic issues that persisted after troubleshooting were reported to the Nigerian Cepheid technical centre for further assistance, either remotely or on-site. Primary concerns about the assay’s performance were ‘test fail- ures’ or unsuccessful test outcomes that cause delay or no diagnosis in some patients. Similar concerns about high rates of error and invalid results were reported by other countries during the early phase of implementa- tion.9–11 The WHO recommended country-specific op- erational research, as diagnostic tests that perform well in controlled settings may not always perform opti- mally in the settings where they are to be used.12 We studied the rate, distribution and causes of un- successful Xpert test outcomes, with the aim of identi- fying key areas that need to be strengthened for opti- mal performance of the assay. DESIGN AND METHODS The study retrospectively analysed data uploaded be- tween January and December 2015 from Xpert facili- Received 6 September 2017 Accepted 11 January 2018 http://dx.doi.org/10.5588/pha.17.0080 Setting:  Nigeria, a high tuberculosis (TB) burden country. Objective: To study the rate, distribution and causes of unsuccessful Xpert® MTB/RIF test outcomes, with the aim of identifying key areas that need to be strengthened for optimal performance of the assay. Design: This was a retrospective analysis of data up- loaded between January and December 2015 from Xpert facilities to the central server using GXAlert. Result: Of 52 219 test results uploaded from 176 Xpert machines, 22.5% were positive for Mycobacterium tuber- culosis, 10.8% of which were rifampicin-resistant; 4.7% of the total number of results were invalid, 4.2% had er- ror results and 2.1% no result outcomes. Technical errors were most frequent (69%); these were non-seasonal and occurred in all geopolitical regions and at all health facil- ity levels. Temperature-related errors were more preva- lent in the North-West Region, with peaks in April to June. Peak periods for temperature and machine mal- function errors coincided with the periods of low utilisa- tion of the assay. Conclusion: The key challenge affecting performance was poor adherence to standard operating procedures. Periodic refresher training courses, regular supervision, preventive maintenance of Xpert machines and proper storage of cartridges are strategies that could improve Xpert performance.
  • 2. Unsuccessful Xpert results 3Public Health Action ties to the central server using GXAlert. Testing failures on the Xpert assay were classified as ‘error’, ‘invalid’ or ‘no result’. Identi- fication and grouping of the error codes into categories were guided by the error resolution SOPs provided by the Xpert manu- facturer, in which error codes were classified under the causes and the steps to take to resolve problems.13,14 By following the SOPs, facility-documented error codes were identified and categorised into five groups based on causes of er- ror (temperature-related, technical problems, cartridge malfunc- tion, electrical connection and machine malfunction errors). Er- rors due to incorrect temperature, which may be of internal origin or due to external environmental factors, e.g., ambient tempera- ture outside of the acceptable range, dirt on filters due to accumu- lated dust, faulty fan and malfunctioning heater component, were grouped as Category A errors (codes 1001, 1002, 2014, 1004 and 2022). Category B errors (codes 5006, 5007 and 2008) were related to technical problems, i.e., mostly human errors due to non-adherence to the SOPs during sample processing, such as fill- ing reaction tubes with viscous sputum or incorrect sample vol- umes, and clogged filters due to the presence of debris in sputum, compromised probe integrity and module failures. Category C er- rors (error codes 5011 and 2037) were associated with cartridge malfunction, caused mostly by inappropriate storage of the car- tridges. Electrical connection issues that resulted in communica- tion loss between the module and the software due to incorrect wiring connections (e.g., poor Ethernet connection between the computer and Xpert, poor connection between gateway and modules, and fluctuation in power supply, with codes 2127, 2126, 1007 and 2122) were classed under Category D. Machine mal- function due to the breakdown of components or module mal- function was grouped under Category E; the associated codes were 2005, 2016, 2032, 2025, 4014 and 4017. Data cleaning and validation were performed before analysis to improve data quality by exporting GXAlert data to SPSS v16 soft- ware (Statistical Package for the Social Sciences, Chicago, IL, USA). Each variable was checked for accuracy and consistency, and fre- quency tables were generated to check for outliers and errors. Any issue identified was flagged and confirmed using source data. As this was a retrospective study using anonymous data, ethics approval was not sought.15 RESULTS A total of 52 219 tests were performed using 176 Xpert machines. Xpert machines were installed at all health facility levels, with the highest proportion (56.6%) in secondary health care facilities. Ap- proximately 22.5% of the total tests performed were positive for Mycobacterium tuberculosis, 10.8% of which were RMP-resistant and 3.0% RMP-indeterminate. Of the total number of tests con- ducted, 4.7% were invalid, 4.2% had error results and 2.2% had no result outcomes, putting the rate of unsuccessful tests at 11.0%. While invalid and ‘no result’ outcome rates remained rela- tively constant across all health care facility levels, error outcomes predominated (5.7%) at the primary health care level (Table 1). Of 2173 error results recorded, only 1736 (79.9%) were documented with codes. About 19.1% of the errors could not be categorised due to lack of information on associated codes. In all, 20 different error codes were documented and catego- rised into five broad groups (Categories A–E). The causes and reso- lution steps for Categories A and D errors are shown in Table 2. Category B errors, associated with technical issues, were most fre- quent (69%), and were distributed throughout all geographical regions and at all health facility levels. Whereas errors due to in- correct temperature and poor electrical connections were most frequent in the northern region, errors due to cartridge malfunc- tion predominated in the southern region. It should be noted that errors due to machine malfunction cut across the North and South geographical regions, and were most prevalent in the North-West and South-West (Table 3). Technical problems occurred throughout the year. Although no temperature-related errors occurred in January, these were recorded from February to December, with a peak in April, May and June. A downward trend was observed from July, with a steady decline to its lowest frequency in September and October, before finally in- creasingly slightly in November and December. Errors associated with cartridge malfunction and poor electrical connections re- mained relatively constant in terms of distribution, and occurred throughout the year. Conversely, machine malfunction errors were more prevalent in April, May and June; this, interestingly, coin- cided with the peak period recorded for temperature-related errors. DISCUSSION Findings from this study revealed gross underutilisation of Xpert machines. This is worrisome, as roll-out of the assay was supposed to increase their use, given the high level of automation that makes it user-friendly. During the early phases of implementa- tion, most of the Xpert machines were placed in secondary- and tertiary-level health facilities. The main assumption for this pref- erential placement was that a stable power supply was more read- ily available at tertiary- and secondary-level facilities than at the primary level. However, from our findings, it is clear that inter- ruptions in power supply, synonymous with a ‘no result’ out- come, were evenly distributed among the three facility levels, in- dicating that an efficient power supply is not dependent on facility level. Greater attention should be paid to sustainability of the back-up power supply to ensure uninterrupted testing, rather than facility-level placement. TABLE 1  Proportion of test outcomes by facility level Level of health facility Frequency n (%) M. tuberculosis detection RMP resistance (subset of M. tuberculosis-positive) Error n (%) Invalid n (%) No result n (%) Positive n (%) Negative n (%) Resistant n (%) Susceptible n (%) Indeterminate n (%) Primary 2 036 (3.9) 479 (23.5) 1 293 (63.7) 52 (10.8) 419 (87.5) 8 (1.6) 117 (5.7) 98 (4.8) 36 (1.8) Secondary 29 548 (56.6) 6719 (22.7) 19 529 (66.0) 668 (9.9) 5 872 (87.4) 179 (2.7) 1 190 (4.0) 1 400 (4.7) 710 (2.4) Tertiary 20 635 (39.5) 4547 (22.0) 13 851 (67.1) 543 (11.9) 3 839 (84.4) 165 (3.6) 866 (4.2) 971 (4.7) 399 (1.9)  Total 52 219 (100) 11 745 (22.5) 34 673 (66.4) 1 264 (10.8) 10 130 (86.2) 352 (3.0) 2 173 (4.2) 2 469 (4.7) 1 145 (2.2) RMP = rifampicin.
  • 3. Unsuccessful Xpert results 4Public Health Action In addition to an uninterrupted power supply, the capacity of the health facility to ensure continued operation after installation should also be taken into consideration when selecting sites for Xpert machine placement. This includes maintenance of facility infrastructure such as air conditioners and refrigerators, which has not always been the case in most of the public health facili- ties selected. The NTBLCP and donors should prioritise the imple- mentation of strategies for improved functionality of Xpert ma- chines over the installation of new machines, to ensure optimal performance at all times. The 22.5% TB detection rate and 10.7% RMP resistance rate recorded, along with the uninterpretable re- sults, was significant. Optimising the functionality of Xpert ma- chines and reducing the number of uninterpretable results may increase the detection of TB and RR-TB. The patient population screened, which comprised presumptive drug-resistant cases and patients with human immunodeficiency virus with TB symptoms, contributed to the high RR-TB detection rate. The main impetus was the timely detection of multidrug-resistant TB,16–19 as RMP re- sistance is a surrogate for isoniazid resistance.20,21 The key lesson learnt from the Xpert roll-out was that access to diagnosis is not the only indicator for measuring successful im- plementation; the rate of unsuccessful tests (which add no value to treatment) is also an important indicator of the effectiveness of Xpert implementation over time. Although Xpert was able to de- TABLE 2  Interpretation and resolution of error codes* Error type Error code Possible cause Solution Technically related (Category B) 5006 Probe check failed because: Check if: • An incorrect amount of reagent was inserted into the cartridge • Reagent was correctly added to the cartridge • The reagent was of compromised quality • Cartridges were stored correctly • Fluid transfer failed Rerun the test using fresh cartridges 5007 Probe check failed because: Check if: • An incorrect amount of reagent was inserted into the cartridge • Reagent was added to the cartridge correctly • The reagent was of compromised quality • Cartridges were stored correctly • Fluid transfer failed • The sample was processed incorrectly in the cartridge 2008 • The filter was clogged Use a new cartridge • Pressure sensor failed Run a cartridge contacting buffer only Poor electrical connection (Category D) 2127 Module communication loss was detected because Ethernet cable between computer and Xpert machine was not connected properly Unplug and replug instrument and Ethernet cable, then restart system 2126 Module was reset because there was an intermittent power supply failure or power supply cable or connector failure Restart system 1007† The power supply voltage was out of range Record the information in the error message. If the error recurs in multiple runs, call Cepheid Technical Support 2122† Loose or faulty Ethernet cable between the computer and the Xpert machine Verify that the Ethernet cable between the computer and the Xpert machine is connected properly * Adapted from Reference 14. † Explanation provided by Cepheid via e-mail. TABLE 3  Distribution of different categories of error by geographic region and health facility level Tests performed n (%) Error type categorised A (temperature- related) n (%) B (technically related) n (%) C (cartridge malfunction) n (%) D (poor electrical connection) n (%) E (machine malfunction) n (%) Total n Geographic region  South-South 8 138 (15.6) 1 (0.5) 155 (73.8) 32 (15.2) 18 (8.6) 4 (1.9) 210  South-East 8 013 (15.3) 4 (1.5) 212 (80.9) 16 (6.1) 24 (9.2) 6 (2.3) 262  South-West 9 288 (17.8) 24 (8.1) 169 (56.9) 54 (18.2) 31 (10.4) 19 (6.4) 297  North-Central 13 891 (26.6) 14 (2.9) 373 (76.6) 53 (10.9) 40 (8.2) 7 (1.4) 487  North-East 3 871 (7.4) 2 (1.3) 95 (62.5) 9 (5.9) 43 (28.3) 3 (2.0)  North-West 9 018 (17.3) 41 (12.5) 187 (57.0) 24 (7.3) 57 (17.4) 19 (5.8) 328   Total 52 219 (100) 86 (5) 1 191 (68.6) 188 (10.8) 213 (12.3) 58 (3.3) 1 736 Facility level  Primary 6 (6.9) 52 (59.8) 3 (3.4) 24 (27.6) 2 (2.3) 87  Secondary 25 (2.5) 709 (70.5) 95 (9.5) 145 (14.4) 31 (3.1) 1 005  Tertiary 55 (8.5) 430 (66.8) 90 (14) 44 (6.8) 25 (3.9) 644   Total 86 (5) 1 191 (68.6) 188 (10.8) 213 (12.3) 58 (3.3) 1 736
  • 4. Unsuccessful Xpert results 5Public Health Action tect a good number of TB cases, the high rate of unsuccessful test results clearly showed that early diagnosis, which is fundamental to TB control, may not be achieved in many individuals if Xpert is not properly managed. According to the Xpert manufacturer, a cumulative error rate of 5% is unacceptable and should be investigated and resolved immediately. The 5.7% error rate recorded in the primary health- level facilities in our study indicates suboptimal performance; strategies identified for the resolution of these errors should therefore be intensified at the primary level. In view of the above, we went a step further and studied the distribution of error categories by geographic region and month. A high percentage (19.1%) of the errors were not coded, which suggests that most facilities do not make use of the error resolu- tion SOPs for troubleshooting. A similar observation was made in a previous article.22 Most errors involved technical issues and originated from incorrect sample processing or poor sample qual- ity. It is to be noted that the high rate of technical errors was among the issues discussed at the Xpert Early Implementers Meet- ing held in France in 2012.23 Most of the errors recorded by early implementers, such as South Africa, were linked to improper pro- cedures used in specimen collection and sample preparation, as well as faulty modules and cartridges. The fact that Category B er- rors cut across geographic regions and facility levels shows that all cadre of staff working at Xpert sites are implicated. This can be attributed to the high rate of staff attrition in the health sector, whereby experienced staff leave in high numbers and are replaced by new personnel without technical experience. The introduction of planned capacity building schemes—which include training, supervision, mentoring and proficiency testing—by the NTBLCP for staff at Xpert facilities will facilitate skill acquisition and con- tinued quality improvement. To reduce Category C errors, the storage temperature for cartridges should be carefully monitored and maintained from the time of landing at the port until the car- tridges arrive at the warehouse, to their final destination at Xpert facilities. This also includes maintaining the correct temperature during transportation. Category A errors were more frequent in the Northern region due to its characteristic hot, dry, dusty climate. In addition to the installation of air conditioners, special attention should be paid to routine maintenance of the equipment, such as monthly clean- ing of exhaust fan filters. Raizada et al. reported that most tem- perature-associated errors were related to clogging of exhaust fan filters by dust or the inappropriate positioning of equipment.12 It is therefore recommended that Xpert operators, especially those in the Northern region, routinely clean the filters, monitor and chart ambient room temperature, and check the machine’s inter- nal temperature whenever a Category A error is flagged. This con- firms the source of the error and facilitates fast and accurate trou- bleshooting. Maintaining a clearance space of 10 cm between the Xpert machine and the wall and other equipment will go a long way to ensure that the warm air generated during testing is ex- pelled properly from the machine. The seasonal variations in the errors indicate that temperature-related errors were most preva- lent from March to June. This is to be expected, given the high temperatures associated with these months. CONCLUSION The performance of the Xpert assay is dependent on adher- ence to SOPs by programme personnel. Xpert roll-out should be followed by quality supervision and mentoring to improve staff competence in test procedures and routine maintenance of the machine. The significant rate of unsuccessful tests ob- served in this study shows that evaluations of the assay should not be based on the case detection rate alone. Performance, in terms of the number of interpretable results that informed the patient’s treatment regimen, is also an indicator that should be assessed and monitored for improved test outcomes. Fi- nally, Xpert machines should be installed in those facilities (private or public) where patients prefer to seek medical help, and where the machines can be better managed, maintained and utilised. References 1 Ghiasi M, Pande T, Pai M. Advances in tuberculosis diagnostics. Curr Trop Med Rep 2015; 2: 54–61. 2 Sharma S K, Kohli M, Yadav R N, et al. Evaluating the diagnostic accuracy of Xpert MTB/RIF assay in pulmonary tuberculosis. PLOS ONE 2015; 10: e0141011. 3 Cruciani M, Scarparo C, Malena M, Bosco O, Serpelloni G, Mengoli C. Me- ta-analysis of BACTEC 960 and BACTEC 460 TB, with or without solid me- dia, for detection of mycobacteria. J Clin Microbiol 2004; 42: 2321–2325. 4 Peralta G, Barry P, Pascopella L. Use of nucleic acid amplification tests in tu- berculosis patients in California, 2010–2013. Open Forum Infect Dis 2016; 3: ofw230. 5 Shenoi S V, Escombe A R, Friedland G. Transmission of drug-susceptible and drug-resistant tuberculosis and the critical importance of airborne infection control in the era of HIV infection and highly active antiretroviral therapy rollouts. Clin Infect Dis 2010; 50 (Suppl 3): S231–S237. 6 World Health Organization. Policy statement: automated real-time nucleic acid amplification technology for rapid and simultaneous detection of tu- berculosis and rifampicin resistance: Xpert MTB/RIF. WHO/HTM/TB/2011.4. Geneva, Switzerland: WHO, 2011. 7 World Health Organization. Xpert MTB/RIF implementation manual: tech- nical and operational ‘how-to’ practical considerations. WHO/HTM/ TB/2014.1. Geneva, Switzerland: WHO, 2014. 8 Piatek A S, Van Cleef M, Alexander H, et al. Xpert for TB diagnosis: planned and purposeful implementation. Glob Health Sci Pract 2013; 1: 18–23. 9 Van Rie A, Page-Shipp L, Scott L, Sanne I, Steven W. Xpert MTB/RIF for point-of-care diagnosis of TB in high-HIV burden, resource limited coun- tries: hype or hope? Expert Rev Mol Diagn 2010; 10: 937–946. 10 Williamson D A, Basu I, Bower J, Freeman J I, Henderson G, Robert S A. An evaluation of the Xpert MTB/RIF assay and detection of false-positive rifam- picin resistance in Mycobacterium tuberculosis. Diagn Microbiol Infect Dis 2012; 74: 207–209. 11 Van Rie A, Mellet K, John M A, et al. False-positive rifampicin resistance on Xpert MTB/RIF: case report and clinical implications. Int J Tuberc Lung Dis 2012; 16: 206–208. 12 Raizada N, Sachdeva K S, Sreenivas A. et al. Feasibility of decentralised de- ployment of Xpert MTB/RIF test at lower level of health system in India. PLOS ONE 2014; 9: e8930. 13 Cepheid. Xpert MTB/RIF. Two hour detection of MTB and resistance to ri- fampicin. Sunnyvale, CA, USA: Cepheid, 2009. 14 Cepheid Xpert Dx System. Operator Manual. Software version 4.0. 300– 7607, Rev. C.17. Sunnyvale, CA, USA: Cepheid, 2010. 15 Junod V, Elger B. Retrospective research: what are the ethical and legal re- quirements. Swiss Med Weekly 2010: 140: w13041. 16 Durovni B, Saraceni V, van den Hof S, et al. Impact of replacing smear mi- croscopy with Xpert MTB/RIF for diagnosing tuberculosis in Brazil: a stepped-wedge cluster-randomized trial. PLOS Med 2014; 11: e1001766. 17 Raizada N, Sachdeva K S, Sreenivas A, et al. Catching the missing million: experiences in enhancing TB and DR-TB detection by providing upfront Xpert MTB/RIF testing for people living with HIV in India. PLOS ONE 2015; 10: e0116721. 18 Sachdeva K S, Raizada N, Sreenivas A, et al. Use of Xpert MTB/RIF in decen- tralized public health setting and its effect on pulmonary TB and DR-TB case finding in India. PLOS ONE 2015; 10: e0126065. 19 Trajman A, Durovni B, Saraceni V, et al. Impact on patient’s treatment out- comes of Xpert MTB/RIF implementionfor the diagnosis of tuberculosis: fol- low-up of a stepped-wedge randomized clinical trial. PLOS ONE 2015; 10: e0123252. 20 Kalokhe A S, Shafiq M, Lee J C, et al. Multidrug-resistant tuberculosis drug susceptibility and molecular diagnostic testing: a review of the literature. Am J Med Sci 2013; 345: 143–148. 21 Thirumurugan R, Kathirvel M, Vallayyachari K, Surendar K, Samrot A V, Muthaiahb M. Molecular analysis of rpoB gene mutations in rifampicin resis- tant Mycobacterium tuberculosis isolates by multiple allele specific polymerase
  • 5. Unsuccessful Xpert results 6Public Health Action chain reaction in Puducherry, South India. J Infect Public Health 2015; 8: 619–625. 22 Gidado M, Onazi J, Nwadike P. et al. Assessment of Gene-Xpert MTB/RIF program implementation and the challenges for enhanced tuberculosis diag- nosis in Nigeria. SAARC J Tuberc Lung Dis HIV/AIDS 2015; 12: 1–7. https:// www.nepjol.info/index.php/SAARCTB/article/view/15948 Accessed February 2018. 23 World Health Organization and Global Laboratory Initiative (GLI). Imple- menters Meeting on Xpert MTB/RIF roll-out, 18–19 April 2012. Annecy, France: Stop TB Partnership, 2012. Public Health Action (PHA)  The voice for operational research. Published by The Union (www.theunion.org), PHA provides a platform to fulfil its mission, ‘Health solutions for the poor’. PHA publishes high-quality scientific research that provides new knowledge to improve the accessibility, equity, quality and efficiency of health systems and services. e-ISSN 2220-8372 Editor-in-Chief:  Dermot Maher, MD, Switzerland Contact:  pha@theunion.org PHA website:  http://www.theunion.org/what-we-do/journals/pha Article submission:  http://mc.manuscriptcentral.com/pha Contexte  :  Le Nigeria, pays lourdement frappé par la tuberculose. Objectif  :  Etudier le taux, la distribution et les causes de mauvais résultats du test Xpert dans le but d’identifier les domaines clés qui doivent être renforcés pour une performance optimale du test. Schéma  :  Analyse rétrospective des données téléchargées entre janvier et décembre 2015 depuis les structures équipées de Xpert vers le serveur central à travers le GXAlert. Résultats  :  Sur 52 219 tests téléchargés à partir de 176 machines, 22,5% ont été positifs pour Mycobacterium tuberculosis, dont 10,8% ont été résistants à la rifampicine ; globalement, 4,7% ont été invalides, 4,2% ont eu des résultats erronés et 2,1% n’ont eu aucun résultat. Les erreurs d’origine technique ont été les plus fréquentes, à 69%, n’ont pas eu de variation saisonnière et sont survenues dans toutes les zones géopolitiques et à tous les niveaux des structures de santé. Les erreurs liées à la température ont été prévalentes dans la région nord-ouest, avec des pics d’avril à juin. Les périodes de pic en termes de température et de dysfonction des machines ont coïncidé avec les périodes de faible utilisation du test. Conclusion  :  Le problème principal qui a affecté la performance du test a été l’adhérence médiocre aux procédures opératoires standardisées. Des révisions périodiques de la formation, une supervision régulière, une maintenance préventive de la machine à Xpert et un stockage approprié des cartouches constituent des stratégies susceptibles d’améliorer la performance du Xpert. Marco de referencia: Nigeria, un país con alta carga de morbilidad por tuberculosis. Objetivo: Estudiar la tasa de resultados fallidos de la prueba Xpert, su distribución y sus causas con el objeto de reconocer las esferas prioritarias que precisan fortalecimiento, a fin de obtener un funcionamiento óptimo de la prueba. Método: Fue este un análisis retrospectivo de los datos enviados al servidor central por los establecimientos que practican la prueba Xpert, mediante el sistema GXAlert, de enero a diciembre del 2015. Resultados: De 52 219 pruebas realizadas en 176 dispositivos y subidas al sistema, 22,5% fueron positivas para Mycobacterium tuberculosis y de ellas el 10,8% presentó resistencia a rifampicina; de todos los resultados, 4,7% fueron inválidos, 4,2% exhibieron error y 2,1% de las pruebas no comportaban un resultado. El error más frecuente fue el de tipo técnico (69%), el cual no siguió un carácter estacional y ocurrió en todas las regiones geopolíticas y en establecimientos de salud de todos los niveles. Los errores debidos a la temperatura predominaron en la región noroeste, con períodos de mayor frecuencia de abril a junio. Los períodos de mayor frecuencia de errores causados por la temperatura o el disfuncionamiento de los dispositivos coincidieron con épocas de baja utilización de la prueba. Conclusión: El principal problema que interfirió con el buen funcionamiento de la prueba fue el incumplimiento de los procedimientos normalizados de trabajo. Se podría mejorar la eficacia de la prueba Xpert mediante estrategias como los cursos periódicos de actualización, la supervisión constante y el mantenimiento preventivo de los dispositivos, además del almacenamiento adecuado de los cartuchos de la prueba.