YOU CAN EASY WAY TO UNDERSTAND A PROCESS AND ANLYTICAL METHOD OF CLEANING VALIDATION
Cleaning validation is the methodology used to assure that a cleaning process removes residues of the active pharmaceutical ingredients of the product manufactured in a piece of equipment, the cleaning aids utilized in the cleaning process and the microbial attributes.[1] All residues are removed to predetermined levels to ensure the quality of the next product manufactured is not compromised by waste from the previous product and the quality of future products using the equipment, to prevent cross-contamination and as a GMP requirement.
The U.S. Food and Drug Administration (FDA) has strict regulation about the cleaning validation. For example, FDA requires firms to have written general procedures on how cleaning processes will be validated. Also, FDA expects the general validation procedures to address who is responsible for performing and approving the validation study, the acceptance criteria, and when revalidation will be required. FDA also require firms to conduct the validation studies in accordance with the protocols and to document the results of studies.The valuation of cleaning validation is also regulated strictly, which usually mainly covers the aspects of equipment design,cleaning process written, analytical methods and sampling. Each of these processes has their related strict rules and requirements. Regarding to the establishment of limits, FDA does not intend to set acceptance specifications or methods for determining whether a cleaning process is validated. But some limits that have been mentioned by industry include analytical detection levels such as 10 PPM, biological activity levels such as 1/1000 of the normal therapeutic dose and organoleptic levels.[2][3][4]
Cleaning Validation in the context of Active Pharmaceutical Ingredient manufacture may be defined as: "The process of providing documented evidence that the cleaning methods employed within a facility consistently controls potential carryover of product (including intermediates and impurities), cleaning agents and extraneous material into subsequent product to a level which is below predetermined levels".
CLEANING VALIDATION for M.pharm and industry person
1. 1
-: Presented By :- -:Guided By:-
Abhishek Pandey Dr. Bhavna Patel
1st Year M.Pharm (QA) Ms. Shraddha Parmar
Roll No. Q1
Dept. of Pharmaceutical Science
2. INTRODUCTION
2
Cleaning validation is documented evidence with a high degree of assurance that
one can consistently clean a system or a piece of equipment to predetermined and
acceptable limits.
The objectives of good manufacturing practices (GMP) include the prevention of
possible contamination and cross-contamination of pharmaceutical starting
materials and products.
Pharmaceutical products can be contaminated by a variety of substances such as
contaminants associated with microbes, previous products (both active
pharmaceutical ingredients (API) and excipient residues), residues of cleaning
agents, airborne materials, such as dust and particulate matter, lubricants.
Adequate cleaning procedures play an important role in preventing
contamination and cross-contamination. Validation of cleaning methods provides
documented evidence that an approved cleaning procedure will provide clean
equipment, suitable for its intended use.
3. Analytical Methods for Cleaning
Validation
DETECTION LIMITS;
The Food and Drug Administration (FDA) cleaning validation
guidelines call for companies to "determine the specificity and
sensitivity of the analytical method used" .
time was a useful word for analytical methods (referring to the
slope of the working curve);
however, in popular usage, it has been loosely used and has
become synonymous with either "limit of detection" (MD) or
"limit of quantization" (LOQ). The FDA is referring to
LOD/LOQ:
The LOD/LOQ of the analytical method should be at or
(preferably) below the acceptance criterion in the analyzed
sample 3
4. SPECIFICITY
•In terms of method specificity, there is a natural preference for
specific methods.
•After all, if one has a target residue, the best way to measure that
residue is to have an analytical procedure that measures only that
species and excludes all potentially interfering species.
• Specific methods are those methods that target a specific molecule
or species and are designed so that possible interferences are
eliminated.
•Specific methods include HPLC, ion chromatography (K),
SDSPAGE (sodium dodecyl sulfate-polyacrylamide gel
electrophoresis), and atomic absorption (AA).
4
5. NONSPECIFIC METHODS
•Nonspecific methods are usually methods that measure a gross property
that results from contributions from a variety of chemical species.
• Examples of nonspecific methods include conductivity and total
organic carbon (TOC).
•Each provides a measure of an overall property but provides no
information as to the chemical nature of the source of conductance or
organic carbon.
•When a nonspecific method is used for a target residue, it is necessary
to make some assumptions about what that nonspecific property
represents.
•This generally involves expressing the property as if all the measured
property is due to the target species.
5
6. Range:-
Range is a series of values of the measured species or property over
which the analytical
procedure was evaluated.
It is only necessary to assure that the procedure is valid over a range of
expected values.
For example, if the calculated acceptance limit for the analytical sample
is X
ppm, then one might want to evaluate a range from approximately 0.2X
to 1 .OX.
On the other hand, if expected results (perhaps based on prequalification
studies) are to be in the 0.1X to 0.3X range, then validation of a range of
0.05X to 0.5X may be justified. 6
7. 7
LOD / LOQ
LOD is the assay value at which it is still possible to say that the
material is present, but it may be not possible to quantify with a
specific value. LOD is typically estimated by several techniques.
For example, for chromatographic techniques, LOD is estimated at
three times the standard deviation of a baseline response. Values
that are below the LOD are generally reported as < LOD. LOQ is
the lowest assay value for which a reasonable confidence exists
that the value is precise.
There are also rules of thumb for estimating LOQs.
For chromatographic procedures, the LOQ can be estimated as 10
times the standard deviation of the baseline noise.
8. Linearity Linearity refers to the characteristic of the relationship of the
measured property to the level of analyte present.
Linearity is an indication that the measured signal is directly
proportional to the concentration of the analyte over the range.
As a general rule for cleaning validation studies, the expectations
are that assays will be linear over the range.
Estimates of linearity can be made by such techniques as
determination ~ (0.99 or better).
8
9. Accuracy
Accuracy refers to the trueness of the measurements to
known values.
This is determined by analyzing known standards. There is
no "magic number" for acceptable accuracy.
However, more accurate methods are preferred over less
accurate methods.
For example, if the acceptance criterion was 20 ppm, a
method with a accuracy of 2- 10 percent, giving a result of
18 ppm, could be considered an acceptable result.
On the other hand, a method with an accuracy of 2- 20
percent, giving a result of 18 ppm, will be suspect in terms
of meeting the acceptance criterion. 9
10. Precision Precision refers to the reproducibility of the method and is often
measured by standard deviation.
Simple precision is the reproducibility of the results in the same
lab over a series of replicate assays using the same operator, the
same equipment, and usually on the same day.
Intermediate precision is the reproducibility of results in the
same lab using different operators, different
pieces of equipment, and generally done on different days.
Ruggedness is interlab reproducibility, involving reproducibility
in different labs.
10
11. CLEANING OF EQUIPMENT STAGE 1:
DETERMINE THE MOST APPROPRIATE CLEANING
PROCEDURE FOR THE EQUIPMENT: –
1. Generate acceptance criteria data for the contaminant.
2. The cleaning method will be determined by the process,
the equipment the cleaning agents and the cleaning
techniques available.
3. All aspects of the cleaning procedure should be clearly
defined in SOPs be they manual / CIP or COP
11
12. DEVELOP AND VALIDATE THE
SAMPLING AND CHOSEN ANALYTICAL
METHODS FOR THE COMPOUND(S)
BEING CLEANED:-
1. Swab
2. Rinse
(determine % recovery, limit of detection, limit of
quantitation, accuracy of method, reproducibility, stability
over time ...etc.)
12
13. EVALUATE EQUIPMENT
SURFACES AND DETERMINE
13
1. Worst case locations to sample (swab sampling).
2. Volume and type of rinse solvent to be employed
(rinse sampling)
3. Equipment surface area (necessary to calculate
carryover into subsequent batches)
14. STAGE 2:DEVELOP A CLEANING VALIDATION PROTOCOL FOR
THE PRODUCT AND THE EQUIPMENT BEING
CLEANED:-
14
That should encompass for example:
1. Introduction
2. Scope
3. Equipment
4. Cleaning procedure
5. Sampling procedures
6. Analytical testing procedure
7. Acceptance/Cleaning limits.
8. Acceptance criteria for the validation.
15. STAGE 3GENERATE INTERIM CLEANING VALIDATION REPORTS
ON A CLEAN BY CLEAN BASIS DETAILING THE
ACCEPTABILITY OF THE CLEANING PROCEDURE FOR
THE EQUIPMENT AND THE PRODUCT:-
This is only required where there is a long period of time
between manufacture of the validation runs (see stage 4 for
reporting requirements).
15
16. STAGE 4:GENERATE A CLEANING VALIDATION REPORT
DETAILING THE ACCEPTABILITY OF THE CLEANING
PROCEDURE FOR THE EQUIPMENT AND THE
PRODUCT:-
The report should give a full detailed background and
introduction to the cleaning Validation study and should
evaluate all data generated with respect to the acceptance
criteria employed for the study. The report should also
indicate the requirement if any for revalidation (period of
time /change control etc.)
16
20. FACILITIES
20
Surrounding:- It incorporates following
area
ROAD IN THE
POT
LAWNS AND
GARDEN
SECURITY
INSTALLATION
LAMP POST TRETMENT
PLANT AREA
RAW WATER
SOURCES LIKE
WELL, TUBE
WELL ETC.
21. FACILITIES
21
For each of above areas the QA person should
design cleaning method and write cleaning SOP
covering following points.
Cleaning equipment used.
Cleaning agent with their concentration.
Frequency of cleaning.
Procedure.
Responsibility for supervising.
Records.
Acceptance criteria.
22. Clean-In-Place (CIP) Method Cleaning of the equipment is performed in place without
disassembling.
Cleaning process may be controlled manually or by an
automated program.
Very consistent and reproducible cleaning method.
Can be validated readily.
Being a closed system visual inspection of all components
is difficult.
22
23. Factors affecting the effectiveness
of the CIP cleaning agents
Temperature of the cleaning solution. Elevating the
temperature of a cleaning solution increases its dirt removal
efficiency. Molecules with high kinetic energy dislodge dirt faster
than slow moving molecules of a cold solution.
Concentration of the cleaning agent. A concentrated cleaning
solution will clean a dirty surface much better than a dilute one
due to the increased surface binding capacity.
Contact time of the cleaning solution. The longer the
detergent contact period, the higher the cleaning efficiency.
After some time, the detergent eventually dissolves the hard
stains/soil from the dirty surface.
Pressure exerted by the cleaning solution (or turbulence).
The turbulence creates an abrasive force that dislodges stubborn
soil from the dirty surface.
23