FLUOROMETRY
BASIC CONCEPTS,
INSTRUMENTATION AND
APPLICATIONS
By
Dr. Basil, B – MBBS (Nigeria),
Department of Chemical Pathology/Metabolic Medicine,
Benue State University Teaching Hospital, Makurdi.
September 2014.
1
INTRODUCTION:
• Luminescence is the emission of light by a substance. It
occurs when an electron returns to the electronic ground
state from an excited state and loses its excess energy as a
photon.
FLUORESCENCE: occurs when a molecule absorbs light at one
wavelength and reemits light at a longer wavelength
• An atom or molecule that fluoresces is termed a
Fluorophore
• Fluorometry is defined as the measurement of the emitted
fluorescence light
• Fluorometric analysis is a very sensitive and widely used
method of quantitative analysis in the chemical and
biological sciences
• Fluorescence starts immediately after the absorption of
light and stops as soon as the incident light is cut off 2
PHOSPHOFLUORESCENCE: This occurs when light
radiation is incident on certain substances and they emit
light continuously even after the incident light is cut off.
• Substances showing phosphorescence are
phosphorescent substances.
Fluorescence and Phosphoflourescence:
• A molecular electronic state in which all of the
electrons are paired are called singlet state.
• In a singlet state molecules are diamagnetic.
• Most of the molecules in their ground state are
paired.
• When such a molecule absorbs uv/visible radiation,
one or more of the paired electron raised to an
excited singlet state /excited triplet state. 3
• The electron spins in the excited state achieved by
absorption of radiation may either be parallel or
antiparallel. Accordingly, this may be a triplet
(parallel) or a singlet (antiparallel) state.
Singlet and Triplet States:
4
BASIC CONCEPTS: Jablonski Diagram
5
• Each molecule contains a series of closely spaced
energy levels
• Absorption of a quantum of light energy by a molecule
causes the transition of an electron from the singlet
ground state to one of the number of possible
vibrational levels of its first singlet state
• Once the molecule is in an excited state, it returns to
its original state in several ways. These are:
– Radiation-less vibrational equilibration
– The fluorescence process from the excited singlet state
– Quenching of the excited singlet state
– Radiation-less crossover to a triplet state
– Quenching of the first triplet state, and
– The phosphorescence process of light emission from the
triplet state. 6
• The vibrational equilibrium before fluorescence
results in some loss of the excitation energy,
• thus, the emitted fluorescence is of less energy
and longer wavelength than the excitation light
• The difference btw the max wavelength of the
excitation light and the max wavelength of the
emitted fluorescence lights is a constant – stokes
shift
• This constant is a measure of energy lost during
the lifetime of the excited state (RVD) before
returning to the ground singlet level (fluorescence
emission).
7
Time Relationship of Fluorescence Emission:
• There is a considerable time delay btw the (1)
absorption of light energy, (2) return to the lowest
excited state, (3) emission of fluorescence light.
8
• The time required for the emitted light to reach 1/e of its
initial intensity , where e is the Naperian base 2.718, is called
the average lifetime of the excited state of the molecule, or
the fluorescence decay time
• The advantage of time-resolved fluorometer (time delay btw
absorption of quanta of light and fluorescence) is the
elimination of background light scattering as a result of
Rayleigh and Raman signals and Short-lived fluorescence
background
• Time-resolved fluorometry depending on how the
fluorescence emission response is measured is categorized
as:
– Pulse flurometry: sample illuminated with as intense brief pulse
of light and the intensity of emission measured as a function of
time with a fast detector
– Phase fluorometry: continuous-wave laser illuminates the
sample, and the emission monitored for impulse and freq
response.
9
Relationship of concentration and fluorescence
intensity:
• Fluorescence intensity is directly proportional to the
concentration of the fluorophore and the excitation
intensity
F = ФIo abc
where F = relative intensity
Ф = fluorescence efficiency (i.e., the
ration btw quanta of light emitted
and quanta of light absorbed)
Io = initial excitation intensity
a = molar absorptivity
b = volume element defined by geometry of
the excitation and emission slits
c = the concentration in mol/L 10
• This relationship holds only for dilute solutions, where
absorbance is less than 2% of the exciting radiation.
Higher than 2%, the fluorescence intensity becomes
nonlinear – inner filter effect.
• The magnitude of fluorescence intensity of a
fluorophore is determined by:
– Its concentration
– The path length
– The intensity of the light source
• Fluorescence measurements are 100 – 1000 times
more sensitive than absorbance measurement due to:
– More intense light source
– Digital filtering techniques
– Sensitive emission photometers
11
• Fluorescence measurements are expressed in “relative”
intensity units because the intensity measured is not an
absolute quantity and its magnitude is defined by the
(instrument-related varriables):
– Instrument slit width
– Detector sensitivity
– Monochromator efficiency
– Excitation intensity
Fluorescence Polarization:
• This is defined by:
P = (Iv – Ih)/(Iv + Ih)
where Iv = intensity of emitted fluorescence light in vertical
plane
Ih = intensity of emitted fluorescence light in horizontal
plane
• Fluorescent polarization is measured by placing a
mechanically or electrically driven polarizer btw the sample
cuvet and detector
12
13
• In normal instrumentation mode, the sample is excited
with polarized light to obtain maximum sensitivity.
• The polarizer analyser is placed to measure the
intensity of the emitted fluorescence both in vertical
and horizontal planes before calculating for P.
• Fluorescence polarization is used to quantitate
analytes by use of the change in fluorescence
depolarization following immunological reactions
• This is done by adding a known quantity of
fluorescent-labelled analyte molecule to a reaction
solution containing an antibody specific to the analyte.
• The labeled analyte binds to the antibody resulting in a
change in its rotational relaxation time and
fluorescence polarization.
14
QUENCHING:
• Decrease in fluorescence intensity due to specific
effects of constituents of the solution e.g.,
concentration, ph, pressure of chemical
substances, temperature, viscosity, etc.
• Types of quenching include:
– Self quenching
– Chemical quenching
– Static quenching
– Collision quenching
Collisional Quenching:
• It reduces fluorescence by collision. where no. of
collisions increased hence quenching takes place.15
Self Quenching/ Concentration Quenching:
16
Chemical Quenching:
• Here decrease in fluorescence intensity due to the
factors like change in ph, presence of oxygen, halides
& heavy metals.
• pH- aniline at pH 5-13 gives fluorescence but at pH <5
&>13 it does not exhibit fluorescence.
• halides like chloride, bromide, iodide & electron
withdrawing groups like NO2, COOH etc. leads to
quenching.
• Heavy metals leads to quenching, because of
collisions of triplet ground state.
Static Quenching:
• This occurs due to complex formation e.g., caffeine
reduces the fluorescence of riboflavin by complex
formation.
17
PHOTODECOMPOSITION:
• Excitation of a weakly fluorescing or dilute solutions
with intense light sources will cause photochemical
decomposition of the analyte (Photobleaching). This
is minimized by:
– Use of the longest feasible wavelength for excitation that
does not introduce light-scattering effects.
– Measure the fluorescence immediately after excitation to
decrease the duration of excitation
– Protect unstable solutions from ambient light by storing
them in dark bottles
– Remove dissolved oxygen from the solution
• Decomposition introduces non-linear response curves
and loss of the majority of the sample fluorescence.18
Factors affecting Fluorescence and
Phosphorescence
Temperature/Viscosity:
• The change in temperature causes the viscosity of the
medium to change which in turn changes the number
of collisions of the molecules of the fluorophore with
solvent molecules.
• A rise in temperature is almost always accompanied by
a decrease in fluorescence.
• Increase in viscosity increases fluorescent intensity
• The increase in the number of collisions between
molecules in turn increases the probability for
deactivation by internal conversion and vibrational
relaxation (Collisional Quenching).
• Temperature of the reaction must be regulated to
within +/- 0.1’C 19
pH:
• Relatively small changes in pH can sometimes cause
substantial changes in the fluorescence intensity and spectral
characteristics of fluorescence.
– For example, serotonin shows a shift in fluorescence
emission maximum from 330 nm at neutral pH to 550 nm in
strong acid without any change in the absorption spectrum.
• In the molecules containing acidic or basic functional groups,
the changes in pH of the medium change the degree of
ionisation of the functional groups. This in turn may affect the
extent of conjugation or the aromaticity of the molecule which
affects its fluorescence.
– For example, aniline shows fluorescence while in acid
solution it does not show fluorescence due to the formation
of anilinium ion.
20
• Therefore, pH control is essential while working with
such molecules and suitable buffers should be
employed for the purpose.
Dissolved Oxygen:
• The paramagnetic substances like dissolved oxygen
and many transition metals with unpaired electrons
dramatically decrease fluorescence and cause
interference in fluorimetric determinations.
• The paramagnetic nature of molecular oxygen
promotes intersystem crossing from singlet to triplet
states in other molecules.
• The longer lifetimes of the triplet states increases the
opportunity for radiationless deactivation to occur.
• Presence of dissolved oxygen influences phospho-
rescence too and causes a large decrease in the
phosphorescence intensity. 21
• It is due to the fact that oxygen at the ground state
gets the energy from an electron in the triplet state
and gets excited.
• This is actually the oxygen emission and not the
phosphorescence. Therefore, it is advisable to make
phosphorescence measurement in the absence of
dissolved oxygen.
Solvent:
• The changes in the “polarity” or hydrogen bonding
ability of the solvent may also significantly affect the
fluorescent behaviour of the analyte.
• The difference in the effect of solvent on the
fluorescence is attributed to the difference in their
ability to stabilise the ground and excited states of the
fluorescent molecule. 22
• Besides solvent polarity, solvent viscosity and solvents with
heavy atoms also affect fluorescence and phosphorescence.
• Increased viscosity increases fluorescence as the
deactivation due to collisions is lowered.
• A higher fluorescence is observed when the solvents do not
contain heavy atoms while phosphorescence increases due
to the presence of heavy atoms in the solvent.
• Some solvents e.g Ethanol, also cause appreciable
fluorescence.
• Other sample matrix, e.g protiens and bilirubin are more
serious contributors to unwanted fluorescence
Adsorption:
• Extreme sensitiveness of the method requires very dilute
solution. Adsorption of the fluorescent substances on the
container wall create serious problems. Hence strong
solutions must be diluted.
• Certain quartz glass and Plastic materials that contain
ultraviolet adsorbers will fluoresce 23
Concentration Effects:
Concentration is proportional to the emitted light energy
absorbed . At the maximum concentration, fluorescence
peaks and may decrease thereafter.
• Concentration quenching
• Inner Filter Effect: refers to the deviation from a linear
relationship btw concentration and fluorescence as the
absorbance of the solution increases above 2% of the exciting
light
– Thus, as the concentration increases, the absorbance of the
excitation intensity increases, and the loss of the light as it travels
through the cuvet increases.
Light effects:
• Type of light: Monochromatic light is essential for the excitation
of fluorescence because the intensity will vary with wavelength.24
• Length of time of exposure: The number of molecules
excited increase with the time of exposure hence
decrease in fluorescence due to Photodecomposition
• Intensity of Incident light: Increase in intensity of light
incident on sample increases fluorescence intensity.
The intensity of light depends upon:
– Light emitted from the lamp.
– Excitation monochromaters.
– Excitation slit width
• Light scattering:
– Rayleigh scattering: Diffusion of radiation in the course
of its passage through a medium containing particles
the size of which is small compared with the
wavelength of the radiation
• occurs with no change of wavelength
– Raman scattering occurs with lengthening of
wavelength.
• It is a property of the solvent and difficult to eliminate at low
fluorophore concentration 25
Nature of Substituent:
• Electron donating group enhances fluorescence –
.g.NH2,OH etc.
• Electron withdrawing groups decrease or destroy
fluorescence. e.g.COOH,NO2, N=N etc.
• High atomic no: atom introduced into the electron
system decreases fluorescence.
Path length:
• The effective path length depends on both the
excitation and emission slit width.
• Use of microcuvette does not reduce the
fluorescence.
• Use of microcell may reduce interferences and
increases the measured fluorescence
26
INSTRUMENTATION:
THE FLUOROMETER
Single-beam and Double-beam Instrument Designs
27
Basic Components:
• Basic components of fluorometers and Spectrofluorometers
include:
– an excitation source, an excitation monochromator, a cuvet, an
emission monochromator, a detector.
• The optical geometry for fluorescence measurement is very
important:
– most commercial fluorometers use the right angle detector
approach because it minimizes the background signal that limits
analytical detection.
– The end-on approach allows the adaptation of a fluorescence
detector to existing 180 degrees absorption instrument.
– The front surface approach has a comparable limit of detection
to right angle detectors and it minimizes the inner filter effect,
but is more susceptible to background light scatter.
• To accommodate these different geometries, the sample
cell is oriented at different angles in relation to the
excitation source and the detector.
28
Basic Components of Fluorometers:
29
Geometrical arrangements of Fluorometres:
30
• If the emission slit is located near the front edge of the
sample cell, the inner filter effect is minimized.
• If the emission slit width is increased, the detector will
be more sensitive, but specificity may decrease.
31
Light Source:
• Mecury Arc Lamp:
– Produce intense line spectrum above 350nm.
– High pressure lamps give lines at 366,405, 436,
546,577,691,734nm.
– Low pressure lamps give additional radiation at 254nm.
• Xenon Arc Lamp:
– Intense radiation by passage of current through an
atmosphere of xenon.
– Spectrum is continuous over the range between over 250-
600nm,peak intensity about 470nm
• Tungsten Lamp:
– Intensity of the lamp is low.
– If excitation is done in the visible region this lamp is used.
– It does not offer UV radiation
• Turnable Dye Lasers:
– Pulsed nitrogen laser as the primary source
– Radiation in the range between 360 and 650 nm is
produced. 32
• Excitation monochromators:- isolates only the
radiation which is absorbed by the molecule.
• Emission monochromaters:- isolates only the radiation
emitted by the molecule.
• Primary filter:- absorbs visible light & transmits UV
light.
• Secondary filter:- absorbs UV radiations & transmits
visible light.
Cuvets:
• Cylindrical or rectangular cells fabricated of silica or
glass used.
• Path length is usually 10mm or 1cm.
• All the surfaces of the sample holder are polished in
fluorimetry.
Monochromators and Filters:
33
Detectors:
• These include:
– Photovoltaic Cell
– Phototube
– Photomultiplier tube: - best and accurate
• Photomultiplier Tube:
– Multiplication of photo electrons by secondary
emission of radiation.
– A photo cathode and series of dynodes are used.
– Each cathode is maintained at 75-100v higher than the
preceding one.
– Over all amplification of 106 is obtained.
34
Single-Beam Fluorometers:
• Tungsten lamp as source of light.
• The primary filter absorbs visible radiation and transmits
uv radiation.
• Emitted radiation measured at 90o by secondary filter.
• Secondary filter absorbs uv radiation and transmits visible
radiation.
Advantages:
• Simple in construction, easy to use and economical
Disadvantages:
• It is not possible to use reference solution & sample
solution at a time.
• Rapid scanning to obtain Exitation & emission spectrum of
the compound is not possible.
35
Double-Beam Fluorometer:
• Two incident beams of light source pass through
primary filters separately and fall on either sample
or reference solution
• The emitted radiation from sample or reference
pass separately through secondary filter.
Advantages:
• Sample and Reference solution can be analysed
simultaneously
Disadvantage:
• Rapid scanning is not possible due to use of filters.
36
APPLICATIONS:
• Determination of Vitamin B1 and B2
• Liquid Chromatography:
– It is an important method of determining compounds as
they appear at the end of chromatogram or capillary
electrophoresis column
• Organic Analysis:
– Qualitative and quantitative analysis of organic aromatic
compounds present in cigarette smoke, air pollutants,
automobile exhausts etc.
• Fluorescent indicators:
– Mainly used in Acid-Base titration e.g:
• Eosin: colourless – green
• Fluorescein: colourless – green
• Quinine sulphate: blue – violet
• Acridine: green – violet. 37
• Fluorometric reagent:
– Aromatic structure with two or more donor functional
groups
• Pharmaceutical Analysis:
38
• Other Applications include:
– Determination of inorganic substances: like Aluminum,
boron, cadmuim with 2-(2OH-phenyl)benzoxazole in
the presence of tartarate, ruthenuim
– Determination of Uranuim salts in Nuclear research
39
REFERENCES:
• Teitz Textbook of Clinical Chemistry and Molecular
diagnosis (5th Edition)
• Dr.B.K.Sharma, Instrumental methods of chemical
analysis.
• Gurdeep R Chatwal, Instrumental methods of
chemical analysis
• http://en.wikipedia.org/wiki/Fluorescence
• http://images.google.co.in/imghp?oe=UTF-
8&hl=en&tab=wi&q=fluorescence
• http://www.bertholdtech.com/ww/en
pub/bioanalytik/biomethods/fluor.cfm 40
THANK YOU
FOR
LISTENING
41

Más contenido relacionado

La actualidad más candente (20)

Fluorimetry
FluorimetryFluorimetry
Fluorimetry
 
Beer lamberts law
Beer lamberts lawBeer lamberts law
Beer lamberts law
 
NEPHLOMETRY and TURBIDIMETRY PPT 1
NEPHLOMETRY and TURBIDIMETRY PPT 1NEPHLOMETRY and TURBIDIMETRY PPT 1
NEPHLOMETRY and TURBIDIMETRY PPT 1
 
uv -visible spectroscopy
 uv -visible spectroscopy uv -visible spectroscopy
uv -visible spectroscopy
 
Spectrophotometry : Instruments & Applications
Spectrophotometry : Instruments & ApplicationsSpectrophotometry : Instruments & Applications
Spectrophotometry : Instruments & Applications
 
Mass spectrometry
Mass spectrometry Mass spectrometry
Mass spectrometry
 
Spectrofluorimetry
SpectrofluorimetrySpectrofluorimetry
Spectrofluorimetry
 
spectrofluorimetry ppt
spectrofluorimetry pptspectrofluorimetry ppt
spectrofluorimetry ppt
 
HPLC
HPLCHPLC
HPLC
 
Flame photometer (Atomic Emission Spectroscopy) Flame emission spectroscopy
Flame photometer (Atomic Emission Spectroscopy) Flame emission spectroscopyFlame photometer (Atomic Emission Spectroscopy) Flame emission spectroscopy
Flame photometer (Atomic Emission Spectroscopy) Flame emission spectroscopy
 
Spectrofluorimetry
SpectrofluorimetrySpectrofluorimetry
Spectrofluorimetry
 
Flame photometry
Flame photometryFlame photometry
Flame photometry
 
FLAME EMISSION SPECTROSCOPY
FLAME EMISSION SPECTROSCOPY FLAME EMISSION SPECTROSCOPY
FLAME EMISSION SPECTROSCOPY
 
Nmr spectroscopy
Nmr spectroscopyNmr spectroscopy
Nmr spectroscopy
 
Gas Chromatography
Gas ChromatographyGas Chromatography
Gas Chromatography
 
Fluorescence spectrometry
Fluorescence spectrometryFluorescence spectrometry
Fluorescence spectrometry
 
Ion exchange chromatography
Ion exchange chromatography Ion exchange chromatography
Ion exchange chromatography
 
Flame photometry
 Flame photometry Flame photometry
Flame photometry
 
PAPER ELECTROPHORESIS.
PAPER ELECTROPHORESIS.PAPER ELECTROPHORESIS.
PAPER ELECTROPHORESIS.
 
Fluorimetry
FluorimetryFluorimetry
Fluorimetry
 

Similar a Fluorometry

2. Fluorimetry.pdf
2. Fluorimetry.pdf2. Fluorimetry.pdf
2. Fluorimetry.pdfJibrilAliSe
 
Spectrofluorimetry or fluorimetry (www.Redicals.com)
Spectrofluorimetry or fluorimetry (www.Redicals.com)Spectrofluorimetry or fluorimetry (www.Redicals.com)
Spectrofluorimetry or fluorimetry (www.Redicals.com)Goa App
 
Spectrofluorimetry (www.redicals.com)
Spectrofluorimetry (www.redicals.com)Spectrofluorimetry (www.redicals.com)
Spectrofluorimetry (www.redicals.com)Goa App
 
fluorimetry and phosphorimetry m pharmacy notes scop satara
fluorimetry and phosphorimetry m pharmacy notes scop satarafluorimetry and phosphorimetry m pharmacy notes scop satara
fluorimetry and phosphorimetry m pharmacy notes scop sataranikhil salunkhe
 
SPECTROFLUORIMETRY.pptx instrumentation, principle
SPECTROFLUORIMETRY.pptx instrumentation, principleSPECTROFLUORIMETRY.pptx instrumentation, principle
SPECTROFLUORIMETRY.pptx instrumentation, principleDr. Vijaya Barge
 
Fluorimetry phosphorimetry
Fluorimetry phosphorimetryFluorimetry phosphorimetry
Fluorimetry phosphorimetryEnosh Gummadi
 
Spectrofluorimetry
Spectrofluorimetry Spectrofluorimetry
Spectrofluorimetry kabil07
 
spectrofluorimetry1 - 211209042138 .pptx
spectrofluorimetry1 - 211209042138 .pptxspectrofluorimetry1 - 211209042138 .pptx
spectrofluorimetry1 - 211209042138 .pptxTamilarasiSasivarnam1
 
Fluorometry PPT file.ppt
Fluorometry PPT file.pptFluorometry PPT file.ppt
Fluorometry PPT file.pptAbdulAziz685234
 
2078051_635282755632595000.ppt
2078051_635282755632595000.ppt2078051_635282755632595000.ppt
2078051_635282755632595000.pptArun Nt
 
Payalfluorimetryupload
PayalfluorimetryuploadPayalfluorimetryupload
PayalfluorimetryuploadKhannapayal
 
Spectroscopic methods in inorganic Chemistry: Fluorescence spectroscopy
Spectroscopic methods in inorganic Chemistry: Fluorescence spectroscopy Spectroscopic methods in inorganic Chemistry: Fluorescence spectroscopy
Spectroscopic methods in inorganic Chemistry: Fluorescence spectroscopy Chris Sonntag
 
Floroscence Spectroscop.pptx
Floroscence Spectroscop.pptxFloroscence Spectroscop.pptx
Floroscence Spectroscop.pptxchickuujunagade
 
Factors affecting fluorescence intensity(pharmaceutical analysis)
Factors affecting fluorescence intensity(pharmaceutical analysis)Factors affecting fluorescence intensity(pharmaceutical analysis)
Factors affecting fluorescence intensity(pharmaceutical analysis)MdIrfanUddin2
 
Flourimetry 140618015916-phpapp01 - copy
Flourimetry 140618015916-phpapp01 - copyFlourimetry 140618015916-phpapp01 - copy
Flourimetry 140618015916-phpapp01 - copyMANISH KUMAR
 

Similar a Fluorometry (20)

2. Fluorimetry.pdf
2. Fluorimetry.pdf2. Fluorimetry.pdf
2. Fluorimetry.pdf
 
Spectrofluorimetry or fluorimetry (www.Redicals.com)
Spectrofluorimetry or fluorimetry (www.Redicals.com)Spectrofluorimetry or fluorimetry (www.Redicals.com)
Spectrofluorimetry or fluorimetry (www.Redicals.com)
 
Spectrofluorimetry (www.redicals.com)
Spectrofluorimetry (www.redicals.com)Spectrofluorimetry (www.redicals.com)
Spectrofluorimetry (www.redicals.com)
 
Flourimetry
FlourimetryFlourimetry
Flourimetry
 
fluorimetry and phosphorimetry m pharmacy notes scop satara
fluorimetry and phosphorimetry m pharmacy notes scop satarafluorimetry and phosphorimetry m pharmacy notes scop satara
fluorimetry and phosphorimetry m pharmacy notes scop satara
 
SPECTROFLUORIMETRY.pptx instrumentation, principle
SPECTROFLUORIMETRY.pptx instrumentation, principleSPECTROFLUORIMETRY.pptx instrumentation, principle
SPECTROFLUORIMETRY.pptx instrumentation, principle
 
Spectrofluorimetry
SpectrofluorimetrySpectrofluorimetry
Spectrofluorimetry
 
Fluorimetry phosphorimetry
Fluorimetry phosphorimetryFluorimetry phosphorimetry
Fluorimetry phosphorimetry
 
Spectrofluorimetry
Spectrofluorimetry Spectrofluorimetry
Spectrofluorimetry
 
spectrofluorimetry1 - 211209042138 .pptx
spectrofluorimetry1 - 211209042138 .pptxspectrofluorimetry1 - 211209042138 .pptx
spectrofluorimetry1 - 211209042138 .pptx
 
Spectroflurometer
SpectroflurometerSpectroflurometer
Spectroflurometer
 
Fluorometry PPT file.ppt
Fluorometry PPT file.pptFluorometry PPT file.ppt
Fluorometry PPT file.ppt
 
Flourimetry factors
Flourimetry factorsFlourimetry factors
Flourimetry factors
 
2078051_635282755632595000.ppt
2078051_635282755632595000.ppt2078051_635282755632595000.ppt
2078051_635282755632595000.ppt
 
Payalfluorimetryupload
PayalfluorimetryuploadPayalfluorimetryupload
Payalfluorimetryupload
 
Spectroscopic methods in inorganic Chemistry: Fluorescence spectroscopy
Spectroscopic methods in inorganic Chemistry: Fluorescence spectroscopy Spectroscopic methods in inorganic Chemistry: Fluorescence spectroscopy
Spectroscopic methods in inorganic Chemistry: Fluorescence spectroscopy
 
Floroscence Spectroscop.pptx
Floroscence Spectroscop.pptxFloroscence Spectroscop.pptx
Floroscence Spectroscop.pptx
 
Factors affecting fluorescence intensity(pharmaceutical analysis)
Factors affecting fluorescence intensity(pharmaceutical analysis)Factors affecting fluorescence intensity(pharmaceutical analysis)
Factors affecting fluorescence intensity(pharmaceutical analysis)
 
Flourimetry 140618015916-phpapp01 - copy
Flourimetry 140618015916-phpapp01 - copyFlourimetry 140618015916-phpapp01 - copy
Flourimetry 140618015916-phpapp01 - copy
 
Specrtofluorometer
SpecrtofluorometerSpecrtofluorometer
Specrtofluorometer
 

Más de Basil "Lexi" Bruno

Csf in clinico laboratory diagnoses
Csf in clinico laboratory diagnosesCsf in clinico laboratory diagnoses
Csf in clinico laboratory diagnosesBasil "Lexi" Bruno
 
Diagnostic procedures and dynamic tests
Diagnostic procedures and dynamic testsDiagnostic procedures and dynamic tests
Diagnostic procedures and dynamic testsBasil "Lexi" Bruno
 
An Overview of Streptococcal Infections
An Overview of Streptococcal InfectionsAn Overview of Streptococcal Infections
An Overview of Streptococcal InfectionsBasil "Lexi" Bruno
 
Thyroid functions and disorders presentation
Thyroid functions and disorders presentationThyroid functions and disorders presentation
Thyroid functions and disorders presentationBasil "Lexi" Bruno
 
Atomic absorption spectrophotometry
Atomic absorption spectrophotometryAtomic absorption spectrophotometry
Atomic absorption spectrophotometryBasil "Lexi" Bruno
 
Spectrophotometry: basic concepts, instrumentation and application
Spectrophotometry: basic concepts, instrumentation and applicationSpectrophotometry: basic concepts, instrumentation and application
Spectrophotometry: basic concepts, instrumentation and applicationBasil "Lexi" Bruno
 

Más de Basil "Lexi" Bruno (10)

Csf in clinico laboratory diagnoses
Csf in clinico laboratory diagnosesCsf in clinico laboratory diagnoses
Csf in clinico laboratory diagnoses
 
Diagnostic procedures and dynamic tests
Diagnostic procedures and dynamic testsDiagnostic procedures and dynamic tests
Diagnostic procedures and dynamic tests
 
Lipids methodology
Lipids methodologyLipids methodology
Lipids methodology
 
Electrochemistry
ElectrochemistryElectrochemistry
Electrochemistry
 
An Overview of Streptococcal Infections
An Overview of Streptococcal InfectionsAn Overview of Streptococcal Infections
An Overview of Streptococcal Infections
 
Thyroid functions and disorders presentation
Thyroid functions and disorders presentationThyroid functions and disorders presentation
Thyroid functions and disorders presentation
 
Laboratory wares
Laboratory waresLaboratory wares
Laboratory wares
 
Electrophoresis
ElectrophoresisElectrophoresis
Electrophoresis
 
Atomic absorption spectrophotometry
Atomic absorption spectrophotometryAtomic absorption spectrophotometry
Atomic absorption spectrophotometry
 
Spectrophotometry: basic concepts, instrumentation and application
Spectrophotometry: basic concepts, instrumentation and applicationSpectrophotometry: basic concepts, instrumentation and application
Spectrophotometry: basic concepts, instrumentation and application
 

Último

(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...
(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...
(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...Taniya Sharma
 
Russian Escorts Girls Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls Delhi
Russian Escorts Girls  Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls DelhiRussian Escorts Girls  Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls Delhi
Russian Escorts Girls Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls DelhiAlinaDevecerski
 
Low Rate Call Girls Kochi Anika 8250192130 Independent Escort Service Kochi
Low Rate Call Girls Kochi Anika 8250192130 Independent Escort Service KochiLow Rate Call Girls Kochi Anika 8250192130 Independent Escort Service Kochi
Low Rate Call Girls Kochi Anika 8250192130 Independent Escort Service KochiSuhani Kapoor
 
Call Girl Number in Vashi Mumbai📲 9833363713 💞 Full Night Enjoy
Call Girl Number in Vashi Mumbai📲 9833363713 💞 Full Night EnjoyCall Girl Number in Vashi Mumbai📲 9833363713 💞 Full Night Enjoy
Call Girl Number in Vashi Mumbai📲 9833363713 💞 Full Night Enjoybabeytanya
 
VIP Russian Call Girls in Varanasi Samaira 8250192130 Independent Escort Serv...
VIP Russian Call Girls in Varanasi Samaira 8250192130 Independent Escort Serv...VIP Russian Call Girls in Varanasi Samaira 8250192130 Independent Escort Serv...
VIP Russian Call Girls in Varanasi Samaira 8250192130 Independent Escort Serv...Neha Kaur
 
Call Girls Varanasi Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Varanasi Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Varanasi Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Varanasi Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...astropune
 
Kesar Bagh Call Girl Price 9548273370 , Lucknow Call Girls Service
Kesar Bagh Call Girl Price 9548273370 , Lucknow Call Girls ServiceKesar Bagh Call Girl Price 9548273370 , Lucknow Call Girls Service
Kesar Bagh Call Girl Price 9548273370 , Lucknow Call Girls Servicemakika9823
 
High Profile Call Girls Coimbatore Saanvi☎️ 8250192130 Independent Escort Se...
High Profile Call Girls Coimbatore Saanvi☎️  8250192130 Independent Escort Se...High Profile Call Girls Coimbatore Saanvi☎️  8250192130 Independent Escort Se...
High Profile Call Girls Coimbatore Saanvi☎️ 8250192130 Independent Escort Se...narwatsonia7
 
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.Artifacts in Nuclear Medicine with Identifying and resolving artifacts.
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.MiadAlsulami
 
Premium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort Service
Premium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort ServicePremium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort Service
Premium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort Servicevidya singh
 
Call Girls Kochi Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Kochi Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Kochi Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Kochi Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Call Girls Bhubaneswar Just Call 9907093804 Top Class Call Girl Service Avail...
Call Girls Bhubaneswar Just Call 9907093804 Top Class Call Girl Service Avail...Call Girls Bhubaneswar Just Call 9907093804 Top Class Call Girl Service Avail...
Call Girls Bhubaneswar Just Call 9907093804 Top Class Call Girl Service Avail...Dipal Arora
 
VIP Call Girls Indore Kirti 💚😋 9256729539 🚀 Indore Escorts
VIP Call Girls Indore Kirti 💚😋  9256729539 🚀 Indore EscortsVIP Call Girls Indore Kirti 💚😋  9256729539 🚀 Indore Escorts
VIP Call Girls Indore Kirti 💚😋 9256729539 🚀 Indore Escortsaditipandeya
 
💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...
💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...
💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...Taniya Sharma
 
Bangalore Call Girls Hebbal Kempapura Number 7001035870 Meetin With Bangalor...
Bangalore Call Girls Hebbal Kempapura Number 7001035870  Meetin With Bangalor...Bangalore Call Girls Hebbal Kempapura Number 7001035870  Meetin With Bangalor...
Bangalore Call Girls Hebbal Kempapura Number 7001035870 Meetin With Bangalor...narwatsonia7
 
Call Girls Darjeeling Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Darjeeling Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Darjeeling Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Darjeeling Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Chandrapur Call girls 8617370543 Provides all area service COD available
Chandrapur Call girls 8617370543 Provides all area service COD availableChandrapur Call girls 8617370543 Provides all area service COD available
Chandrapur Call girls 8617370543 Provides all area service COD availableDipal Arora
 
Top Rated Bangalore Call Girls Mg Road ⟟ 8250192130 ⟟ Call Me For Genuine Sex...
Top Rated Bangalore Call Girls Mg Road ⟟ 8250192130 ⟟ Call Me For Genuine Sex...Top Rated Bangalore Call Girls Mg Road ⟟ 8250192130 ⟟ Call Me For Genuine Sex...
Top Rated Bangalore Call Girls Mg Road ⟟ 8250192130 ⟟ Call Me For Genuine Sex...narwatsonia7
 
(Rocky) Jaipur Call Girl - 9521753030 Escorts Service 50% Off with Cash ON De...
(Rocky) Jaipur Call Girl - 9521753030 Escorts Service 50% Off with Cash ON De...(Rocky) Jaipur Call Girl - 9521753030 Escorts Service 50% Off with Cash ON De...
(Rocky) Jaipur Call Girl - 9521753030 Escorts Service 50% Off with Cash ON De...indiancallgirl4rent
 

Último (20)

(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...
(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...
(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...
 
Russian Escorts Girls Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls Delhi
Russian Escorts Girls  Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls DelhiRussian Escorts Girls  Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls Delhi
Russian Escorts Girls Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls Delhi
 
Low Rate Call Girls Kochi Anika 8250192130 Independent Escort Service Kochi
Low Rate Call Girls Kochi Anika 8250192130 Independent Escort Service KochiLow Rate Call Girls Kochi Anika 8250192130 Independent Escort Service Kochi
Low Rate Call Girls Kochi Anika 8250192130 Independent Escort Service Kochi
 
Call Girl Number in Vashi Mumbai📲 9833363713 💞 Full Night Enjoy
Call Girl Number in Vashi Mumbai📲 9833363713 💞 Full Night EnjoyCall Girl Number in Vashi Mumbai📲 9833363713 💞 Full Night Enjoy
Call Girl Number in Vashi Mumbai📲 9833363713 💞 Full Night Enjoy
 
VIP Russian Call Girls in Varanasi Samaira 8250192130 Independent Escort Serv...
VIP Russian Call Girls in Varanasi Samaira 8250192130 Independent Escort Serv...VIP Russian Call Girls in Varanasi Samaira 8250192130 Independent Escort Serv...
VIP Russian Call Girls in Varanasi Samaira 8250192130 Independent Escort Serv...
 
Call Girls Varanasi Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Varanasi Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Varanasi Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Varanasi Just Call 9907093804 Top Class Call Girl Service Available
 
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...
 
Kesar Bagh Call Girl Price 9548273370 , Lucknow Call Girls Service
Kesar Bagh Call Girl Price 9548273370 , Lucknow Call Girls ServiceKesar Bagh Call Girl Price 9548273370 , Lucknow Call Girls Service
Kesar Bagh Call Girl Price 9548273370 , Lucknow Call Girls Service
 
High Profile Call Girls Coimbatore Saanvi☎️ 8250192130 Independent Escort Se...
High Profile Call Girls Coimbatore Saanvi☎️  8250192130 Independent Escort Se...High Profile Call Girls Coimbatore Saanvi☎️  8250192130 Independent Escort Se...
High Profile Call Girls Coimbatore Saanvi☎️ 8250192130 Independent Escort Se...
 
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.Artifacts in Nuclear Medicine with Identifying and resolving artifacts.
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.
 
Premium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort Service
Premium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort ServicePremium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort Service
Premium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort Service
 
Call Girls Kochi Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Kochi Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Kochi Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Kochi Just Call 9907093804 Top Class Call Girl Service Available
 
Call Girls Bhubaneswar Just Call 9907093804 Top Class Call Girl Service Avail...
Call Girls Bhubaneswar Just Call 9907093804 Top Class Call Girl Service Avail...Call Girls Bhubaneswar Just Call 9907093804 Top Class Call Girl Service Avail...
Call Girls Bhubaneswar Just Call 9907093804 Top Class Call Girl Service Avail...
 
VIP Call Girls Indore Kirti 💚😋 9256729539 🚀 Indore Escorts
VIP Call Girls Indore Kirti 💚😋  9256729539 🚀 Indore EscortsVIP Call Girls Indore Kirti 💚😋  9256729539 🚀 Indore Escorts
VIP Call Girls Indore Kirti 💚😋 9256729539 🚀 Indore Escorts
 
💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...
💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...
💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...
 
Bangalore Call Girls Hebbal Kempapura Number 7001035870 Meetin With Bangalor...
Bangalore Call Girls Hebbal Kempapura Number 7001035870  Meetin With Bangalor...Bangalore Call Girls Hebbal Kempapura Number 7001035870  Meetin With Bangalor...
Bangalore Call Girls Hebbal Kempapura Number 7001035870 Meetin With Bangalor...
 
Call Girls Darjeeling Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Darjeeling Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Darjeeling Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Darjeeling Just Call 9907093804 Top Class Call Girl Service Available
 
Chandrapur Call girls 8617370543 Provides all area service COD available
Chandrapur Call girls 8617370543 Provides all area service COD availableChandrapur Call girls 8617370543 Provides all area service COD available
Chandrapur Call girls 8617370543 Provides all area service COD available
 
Top Rated Bangalore Call Girls Mg Road ⟟ 8250192130 ⟟ Call Me For Genuine Sex...
Top Rated Bangalore Call Girls Mg Road ⟟ 8250192130 ⟟ Call Me For Genuine Sex...Top Rated Bangalore Call Girls Mg Road ⟟ 8250192130 ⟟ Call Me For Genuine Sex...
Top Rated Bangalore Call Girls Mg Road ⟟ 8250192130 ⟟ Call Me For Genuine Sex...
 
(Rocky) Jaipur Call Girl - 9521753030 Escorts Service 50% Off with Cash ON De...
(Rocky) Jaipur Call Girl - 9521753030 Escorts Service 50% Off with Cash ON De...(Rocky) Jaipur Call Girl - 9521753030 Escorts Service 50% Off with Cash ON De...
(Rocky) Jaipur Call Girl - 9521753030 Escorts Service 50% Off with Cash ON De...
 

Fluorometry

  • 1. FLUOROMETRY BASIC CONCEPTS, INSTRUMENTATION AND APPLICATIONS By Dr. Basil, B – MBBS (Nigeria), Department of Chemical Pathology/Metabolic Medicine, Benue State University Teaching Hospital, Makurdi. September 2014. 1
  • 2. INTRODUCTION: • Luminescence is the emission of light by a substance. It occurs when an electron returns to the electronic ground state from an excited state and loses its excess energy as a photon. FLUORESCENCE: occurs when a molecule absorbs light at one wavelength and reemits light at a longer wavelength • An atom or molecule that fluoresces is termed a Fluorophore • Fluorometry is defined as the measurement of the emitted fluorescence light • Fluorometric analysis is a very sensitive and widely used method of quantitative analysis in the chemical and biological sciences • Fluorescence starts immediately after the absorption of light and stops as soon as the incident light is cut off 2
  • 3. PHOSPHOFLUORESCENCE: This occurs when light radiation is incident on certain substances and they emit light continuously even after the incident light is cut off. • Substances showing phosphorescence are phosphorescent substances. Fluorescence and Phosphoflourescence: • A molecular electronic state in which all of the electrons are paired are called singlet state. • In a singlet state molecules are diamagnetic. • Most of the molecules in their ground state are paired. • When such a molecule absorbs uv/visible radiation, one or more of the paired electron raised to an excited singlet state /excited triplet state. 3
  • 4. • The electron spins in the excited state achieved by absorption of radiation may either be parallel or antiparallel. Accordingly, this may be a triplet (parallel) or a singlet (antiparallel) state. Singlet and Triplet States: 4
  • 6. • Each molecule contains a series of closely spaced energy levels • Absorption of a quantum of light energy by a molecule causes the transition of an electron from the singlet ground state to one of the number of possible vibrational levels of its first singlet state • Once the molecule is in an excited state, it returns to its original state in several ways. These are: – Radiation-less vibrational equilibration – The fluorescence process from the excited singlet state – Quenching of the excited singlet state – Radiation-less crossover to a triplet state – Quenching of the first triplet state, and – The phosphorescence process of light emission from the triplet state. 6
  • 7. • The vibrational equilibrium before fluorescence results in some loss of the excitation energy, • thus, the emitted fluorescence is of less energy and longer wavelength than the excitation light • The difference btw the max wavelength of the excitation light and the max wavelength of the emitted fluorescence lights is a constant – stokes shift • This constant is a measure of energy lost during the lifetime of the excited state (RVD) before returning to the ground singlet level (fluorescence emission). 7
  • 8. Time Relationship of Fluorescence Emission: • There is a considerable time delay btw the (1) absorption of light energy, (2) return to the lowest excited state, (3) emission of fluorescence light. 8
  • 9. • The time required for the emitted light to reach 1/e of its initial intensity , where e is the Naperian base 2.718, is called the average lifetime of the excited state of the molecule, or the fluorescence decay time • The advantage of time-resolved fluorometer (time delay btw absorption of quanta of light and fluorescence) is the elimination of background light scattering as a result of Rayleigh and Raman signals and Short-lived fluorescence background • Time-resolved fluorometry depending on how the fluorescence emission response is measured is categorized as: – Pulse flurometry: sample illuminated with as intense brief pulse of light and the intensity of emission measured as a function of time with a fast detector – Phase fluorometry: continuous-wave laser illuminates the sample, and the emission monitored for impulse and freq response. 9
  • 10. Relationship of concentration and fluorescence intensity: • Fluorescence intensity is directly proportional to the concentration of the fluorophore and the excitation intensity F = ФIo abc where F = relative intensity Ф = fluorescence efficiency (i.e., the ration btw quanta of light emitted and quanta of light absorbed) Io = initial excitation intensity a = molar absorptivity b = volume element defined by geometry of the excitation and emission slits c = the concentration in mol/L 10
  • 11. • This relationship holds only for dilute solutions, where absorbance is less than 2% of the exciting radiation. Higher than 2%, the fluorescence intensity becomes nonlinear – inner filter effect. • The magnitude of fluorescence intensity of a fluorophore is determined by: – Its concentration – The path length – The intensity of the light source • Fluorescence measurements are 100 – 1000 times more sensitive than absorbance measurement due to: – More intense light source – Digital filtering techniques – Sensitive emission photometers 11
  • 12. • Fluorescence measurements are expressed in “relative” intensity units because the intensity measured is not an absolute quantity and its magnitude is defined by the (instrument-related varriables): – Instrument slit width – Detector sensitivity – Monochromator efficiency – Excitation intensity Fluorescence Polarization: • This is defined by: P = (Iv – Ih)/(Iv + Ih) where Iv = intensity of emitted fluorescence light in vertical plane Ih = intensity of emitted fluorescence light in horizontal plane • Fluorescent polarization is measured by placing a mechanically or electrically driven polarizer btw the sample cuvet and detector 12
  • 13. 13
  • 14. • In normal instrumentation mode, the sample is excited with polarized light to obtain maximum sensitivity. • The polarizer analyser is placed to measure the intensity of the emitted fluorescence both in vertical and horizontal planes before calculating for P. • Fluorescence polarization is used to quantitate analytes by use of the change in fluorescence depolarization following immunological reactions • This is done by adding a known quantity of fluorescent-labelled analyte molecule to a reaction solution containing an antibody specific to the analyte. • The labeled analyte binds to the antibody resulting in a change in its rotational relaxation time and fluorescence polarization. 14
  • 15. QUENCHING: • Decrease in fluorescence intensity due to specific effects of constituents of the solution e.g., concentration, ph, pressure of chemical substances, temperature, viscosity, etc. • Types of quenching include: – Self quenching – Chemical quenching – Static quenching – Collision quenching Collisional Quenching: • It reduces fluorescence by collision. where no. of collisions increased hence quenching takes place.15
  • 17. Chemical Quenching: • Here decrease in fluorescence intensity due to the factors like change in ph, presence of oxygen, halides & heavy metals. • pH- aniline at pH 5-13 gives fluorescence but at pH <5 &>13 it does not exhibit fluorescence. • halides like chloride, bromide, iodide & electron withdrawing groups like NO2, COOH etc. leads to quenching. • Heavy metals leads to quenching, because of collisions of triplet ground state. Static Quenching: • This occurs due to complex formation e.g., caffeine reduces the fluorescence of riboflavin by complex formation. 17
  • 18. PHOTODECOMPOSITION: • Excitation of a weakly fluorescing or dilute solutions with intense light sources will cause photochemical decomposition of the analyte (Photobleaching). This is minimized by: – Use of the longest feasible wavelength for excitation that does not introduce light-scattering effects. – Measure the fluorescence immediately after excitation to decrease the duration of excitation – Protect unstable solutions from ambient light by storing them in dark bottles – Remove dissolved oxygen from the solution • Decomposition introduces non-linear response curves and loss of the majority of the sample fluorescence.18
  • 19. Factors affecting Fluorescence and Phosphorescence Temperature/Viscosity: • The change in temperature causes the viscosity of the medium to change which in turn changes the number of collisions of the molecules of the fluorophore with solvent molecules. • A rise in temperature is almost always accompanied by a decrease in fluorescence. • Increase in viscosity increases fluorescent intensity • The increase in the number of collisions between molecules in turn increases the probability for deactivation by internal conversion and vibrational relaxation (Collisional Quenching). • Temperature of the reaction must be regulated to within +/- 0.1’C 19
  • 20. pH: • Relatively small changes in pH can sometimes cause substantial changes in the fluorescence intensity and spectral characteristics of fluorescence. – For example, serotonin shows a shift in fluorescence emission maximum from 330 nm at neutral pH to 550 nm in strong acid without any change in the absorption spectrum. • In the molecules containing acidic or basic functional groups, the changes in pH of the medium change the degree of ionisation of the functional groups. This in turn may affect the extent of conjugation or the aromaticity of the molecule which affects its fluorescence. – For example, aniline shows fluorescence while in acid solution it does not show fluorescence due to the formation of anilinium ion. 20
  • 21. • Therefore, pH control is essential while working with such molecules and suitable buffers should be employed for the purpose. Dissolved Oxygen: • The paramagnetic substances like dissolved oxygen and many transition metals with unpaired electrons dramatically decrease fluorescence and cause interference in fluorimetric determinations. • The paramagnetic nature of molecular oxygen promotes intersystem crossing from singlet to triplet states in other molecules. • The longer lifetimes of the triplet states increases the opportunity for radiationless deactivation to occur. • Presence of dissolved oxygen influences phospho- rescence too and causes a large decrease in the phosphorescence intensity. 21
  • 22. • It is due to the fact that oxygen at the ground state gets the energy from an electron in the triplet state and gets excited. • This is actually the oxygen emission and not the phosphorescence. Therefore, it is advisable to make phosphorescence measurement in the absence of dissolved oxygen. Solvent: • The changes in the “polarity” or hydrogen bonding ability of the solvent may also significantly affect the fluorescent behaviour of the analyte. • The difference in the effect of solvent on the fluorescence is attributed to the difference in their ability to stabilise the ground and excited states of the fluorescent molecule. 22
  • 23. • Besides solvent polarity, solvent viscosity and solvents with heavy atoms also affect fluorescence and phosphorescence. • Increased viscosity increases fluorescence as the deactivation due to collisions is lowered. • A higher fluorescence is observed when the solvents do not contain heavy atoms while phosphorescence increases due to the presence of heavy atoms in the solvent. • Some solvents e.g Ethanol, also cause appreciable fluorescence. • Other sample matrix, e.g protiens and bilirubin are more serious contributors to unwanted fluorescence Adsorption: • Extreme sensitiveness of the method requires very dilute solution. Adsorption of the fluorescent substances on the container wall create serious problems. Hence strong solutions must be diluted. • Certain quartz glass and Plastic materials that contain ultraviolet adsorbers will fluoresce 23
  • 24. Concentration Effects: Concentration is proportional to the emitted light energy absorbed . At the maximum concentration, fluorescence peaks and may decrease thereafter. • Concentration quenching • Inner Filter Effect: refers to the deviation from a linear relationship btw concentration and fluorescence as the absorbance of the solution increases above 2% of the exciting light – Thus, as the concentration increases, the absorbance of the excitation intensity increases, and the loss of the light as it travels through the cuvet increases. Light effects: • Type of light: Monochromatic light is essential for the excitation of fluorescence because the intensity will vary with wavelength.24
  • 25. • Length of time of exposure: The number of molecules excited increase with the time of exposure hence decrease in fluorescence due to Photodecomposition • Intensity of Incident light: Increase in intensity of light incident on sample increases fluorescence intensity. The intensity of light depends upon: – Light emitted from the lamp. – Excitation monochromaters. – Excitation slit width • Light scattering: – Rayleigh scattering: Diffusion of radiation in the course of its passage through a medium containing particles the size of which is small compared with the wavelength of the radiation • occurs with no change of wavelength – Raman scattering occurs with lengthening of wavelength. • It is a property of the solvent and difficult to eliminate at low fluorophore concentration 25
  • 26. Nature of Substituent: • Electron donating group enhances fluorescence – .g.NH2,OH etc. • Electron withdrawing groups decrease or destroy fluorescence. e.g.COOH,NO2, N=N etc. • High atomic no: atom introduced into the electron system decreases fluorescence. Path length: • The effective path length depends on both the excitation and emission slit width. • Use of microcuvette does not reduce the fluorescence. • Use of microcell may reduce interferences and increases the measured fluorescence 26
  • 27. INSTRUMENTATION: THE FLUOROMETER Single-beam and Double-beam Instrument Designs 27
  • 28. Basic Components: • Basic components of fluorometers and Spectrofluorometers include: – an excitation source, an excitation monochromator, a cuvet, an emission monochromator, a detector. • The optical geometry for fluorescence measurement is very important: – most commercial fluorometers use the right angle detector approach because it minimizes the background signal that limits analytical detection. – The end-on approach allows the adaptation of a fluorescence detector to existing 180 degrees absorption instrument. – The front surface approach has a comparable limit of detection to right angle detectors and it minimizes the inner filter effect, but is more susceptible to background light scatter. • To accommodate these different geometries, the sample cell is oriented at different angles in relation to the excitation source and the detector. 28
  • 29. Basic Components of Fluorometers: 29
  • 30. Geometrical arrangements of Fluorometres: 30
  • 31. • If the emission slit is located near the front edge of the sample cell, the inner filter effect is minimized. • If the emission slit width is increased, the detector will be more sensitive, but specificity may decrease. 31
  • 32. Light Source: • Mecury Arc Lamp: – Produce intense line spectrum above 350nm. – High pressure lamps give lines at 366,405, 436, 546,577,691,734nm. – Low pressure lamps give additional radiation at 254nm. • Xenon Arc Lamp: – Intense radiation by passage of current through an atmosphere of xenon. – Spectrum is continuous over the range between over 250- 600nm,peak intensity about 470nm • Tungsten Lamp: – Intensity of the lamp is low. – If excitation is done in the visible region this lamp is used. – It does not offer UV radiation • Turnable Dye Lasers: – Pulsed nitrogen laser as the primary source – Radiation in the range between 360 and 650 nm is produced. 32
  • 33. • Excitation monochromators:- isolates only the radiation which is absorbed by the molecule. • Emission monochromaters:- isolates only the radiation emitted by the molecule. • Primary filter:- absorbs visible light & transmits UV light. • Secondary filter:- absorbs UV radiations & transmits visible light. Cuvets: • Cylindrical or rectangular cells fabricated of silica or glass used. • Path length is usually 10mm or 1cm. • All the surfaces of the sample holder are polished in fluorimetry. Monochromators and Filters: 33
  • 34. Detectors: • These include: – Photovoltaic Cell – Phototube – Photomultiplier tube: - best and accurate • Photomultiplier Tube: – Multiplication of photo electrons by secondary emission of radiation. – A photo cathode and series of dynodes are used. – Each cathode is maintained at 75-100v higher than the preceding one. – Over all amplification of 106 is obtained. 34
  • 35. Single-Beam Fluorometers: • Tungsten lamp as source of light. • The primary filter absorbs visible radiation and transmits uv radiation. • Emitted radiation measured at 90o by secondary filter. • Secondary filter absorbs uv radiation and transmits visible radiation. Advantages: • Simple in construction, easy to use and economical Disadvantages: • It is not possible to use reference solution & sample solution at a time. • Rapid scanning to obtain Exitation & emission spectrum of the compound is not possible. 35
  • 36. Double-Beam Fluorometer: • Two incident beams of light source pass through primary filters separately and fall on either sample or reference solution • The emitted radiation from sample or reference pass separately through secondary filter. Advantages: • Sample and Reference solution can be analysed simultaneously Disadvantage: • Rapid scanning is not possible due to use of filters. 36
  • 37. APPLICATIONS: • Determination of Vitamin B1 and B2 • Liquid Chromatography: – It is an important method of determining compounds as they appear at the end of chromatogram or capillary electrophoresis column • Organic Analysis: – Qualitative and quantitative analysis of organic aromatic compounds present in cigarette smoke, air pollutants, automobile exhausts etc. • Fluorescent indicators: – Mainly used in Acid-Base titration e.g: • Eosin: colourless – green • Fluorescein: colourless – green • Quinine sulphate: blue – violet • Acridine: green – violet. 37
  • 38. • Fluorometric reagent: – Aromatic structure with two or more donor functional groups • Pharmaceutical Analysis: 38
  • 39. • Other Applications include: – Determination of inorganic substances: like Aluminum, boron, cadmuim with 2-(2OH-phenyl)benzoxazole in the presence of tartarate, ruthenuim – Determination of Uranuim salts in Nuclear research 39
  • 40. REFERENCES: • Teitz Textbook of Clinical Chemistry and Molecular diagnosis (5th Edition) • Dr.B.K.Sharma, Instrumental methods of chemical analysis. • Gurdeep R Chatwal, Instrumental methods of chemical analysis • http://en.wikipedia.org/wiki/Fluorescence • http://images.google.co.in/imghp?oe=UTF- 8&hl=en&tab=wi&q=fluorescence • http://www.bertholdtech.com/ww/en pub/bioanalytik/biomethods/fluor.cfm 40