SlideShare una empresa de Scribd logo
1 de 6
Descargar para leer sin conexión
ATOMIC ABSORPTION SPECTROPHOTOMETER
Purpose
For analysis of trace metals and other elements in water and sediment samples. Fully
automated PC controlled true double beam Atomic Absorption Spectrophotometer with
absorption and emission capability with deuterium background correction and should have
the capability of measuring multi elements in rapid sequence in single sample measurement,
fitted with 4 or more lamp position and choice of manual or automatic lamp selection with
following features:
Conditions & Requirements
• Installation, commissioning and training in operational use shall be by supplier.
• At least 5 years of after sales service shall be guaranteed
Specifications
General Features • Automatic wave length and slit selection
• Wide range and high sensitivity photomultiplier for maximum
signal to noise ratio
• Choice of burner type (air/acetylene or nitrous oxide)
• Deuterium background correction. Range 185-425 mm to 2.3
Abs. Electronic modulation with automatic attenuation
• Integrated built-in Flame and Graphite Furnace along with auto-
sampler.
• It should be automatically aligned in the path by using the
software.
Spectrophotometer • Type: PC-controlled
• Optics: hollow cathode lamps, double beam, narrow beam
optics matches flame profile.
• Hollow Cathode Lamp or Electrode Less Discharge (EDL)
Lamp
• All reflective system with quartz over-coating on mirrors.
• Sealed against dust and vapour
• Wave length range: 180 to 900 nm
• Slit 0.2 to 2 nm with automatic and manual choice
Monochromator • Monochromator with computer controlled wavelength selection
and peaking
• The system should be based on abrasion corrected Cnerny-
Turner Grating (holographic diffraction grating) up to 1800
grooves per mm blazed at 240 nm
• Continuously adjustable silts with 0.2 to 2 nm spectral
bandwidth
Sensitivity • At least 0.7 Abs
Absorption Cell • Quartz
Lamp turret • At least 4 lamps with automatic Turret controlled with software
and auto-alignment
Gas supply • 0 to 120 ml/min
Gas Control • Automatic – preset oxidant flow and variable full flow adjusted
using manual flow meter adjustment. Automatic oxidant change
over
• Fully automatic gas box with computer-controlled oxidant
selection, automatic gas sequencing, oxidant and fuel
monitoring and control
• Actuated ignition with air/acetylene. Acetylene flow should be
automatically adjustable when switching to or from nitrous-
oxide/acetylene operation
Flame control • Programmable and manual oxidant and fuel selection and
• Monitoring
• Safety interlocks
• Safe and automatic shutdown in case of power failure
Safety system • Separate ignite/flame off buttons. Safety interlocks to monitor
burner type, burner fitting, liquid trap, pressure relief bung,
flame shield, flame operation, main power, oxidant pressure and
deuterium lamp cover
• Fully interlocked operation preventing ignition if proper burner
head, nebulizer, end cap or burner drain system are not correctly
installed, the level of liquid in the drain vessel is incorrect or gas
pressures are too low
• The interlocks should be with automatically shut down the gases
if a flame is not detected
• In the event of power failure or when the emergency flame off
button is used, the flame should be automatically and safely
extinguished
Spray chamber • Solid poly-propylene or fluorinated HDP or equivalent
• Or fluorinated high density polyethylene for aqueous and
organic solutions
Burner • All titanium or equivalent, for air-acetylene and nitrous oxide-
acetylene flames or air/acetylene burner with Teflon shield.
• Burner system chamber should provide analysis of corrosive
and high solid matrices
• Automatic burner rotation system
• Interchangeable full titanium nebulizer for organic and acidic
purpose and burner head for 50 mm and 100 mm, facilities
should be there to switch from Flame to Graphite within few
seconds without removing the air acetylene or nitrous oxide
burner.
Mixing paddles • Removal twin headed mixing paddles fitted inside the spray
chamber for better mixing and extend operation with high
dissolved solids solution
Impact bead • Externally adjustable glass impact bead (Teflon bead)
Hydride
Generator
• Hydride generator (for AS, Se, Hg at ppb level), automatic and
computer controlled
• Sample should be injected through the peristaltic pump, 3
channel peristaltic pump with heated quartz cell, fast heating
rate, temperature range should be from 100 to 1000 degree
Celsius
• Typical precision 1-2% RSD with sample through put of 50
samples/hour determined in triplicate
Graphite Furnace • Built-in integrated furnace with flame facility to switch from
flame to graphite within few seconds without removing the air
acetylene or nitrous oxide burner
• It should automatically align in optical path by using the
software
• It should have fully automated constant temperature zone with
dynamic feedback temperature control, cooling water
temperature compensation, upto 20 steps per program,
temperature programmable from 40 to 3000o
C with maximum
heating rate up to 2000o
C, choice of two inert gases with
programmable flow rate, solid titanium work head for maximum
corrosion resistance, quartz and windows for high light energy,
comprehensive safety interlocks for gases, water flow, tube and
transformer temperature, DC Power supply used for Graphite
Furnace with strict electromagnetic compatibility
• Spray chamber: solid polypropylene of fluorinated HDP or
equivalent
• Recirculating Water chiller: with flow rates of 1.5-2 L/Min at
175-200 kPa for cooling of Graphite Furnace must be quoted
from manufacturer
• Autosampler for Graphite Furnace: Automatic programmable
sample dispenser for GTA 50 samples in 2 mL vials, 5 25 mL
vessels for bulk solutions, automatic preparation up to 10 points
calibration, hot injection automatic flow through rinsing
• Easy access for graphite tube replacement with built-in tube
usage counter should be available
• Auto sampler should be provided with graphite furnace
Operational
requirement
Background correction:
• Sensitivity: 0.7 absorbance or better for 5 ppm Cu
• Calibration: automatic, using up to 6 standards and re-
calibration with single standard
Power supply • 220 ±10 VAC, 47 to 53 Hz
Typical
performance
• Greater than 0.8 Absorbance with precision of <0.5% RSD from
ten 5 sec integration for 5 mg/l Cu solution
Other
requirements
• PC Compatible for operation control and monitoring of AAS
and necessary modules with latest configuration and software,
core 2 DUO processor, 3 GHz, 2 GB DDR RAM, 48x CD ROM
Drive, 320 GB HDD, 17” TFT Colour Monitor, 104 key board,
optical mouse, Window 7 and licensed MS Office with CD,
serial and parallel ports, DeskJet colour printer with suitable
stabilizer
• Software performing analyses at least as per DIN/ISO/US-EPA,
control of flame, calibration, blank correction, data import,
export, handling and reporting, quality control protocols,
computer based training
• All the methods, parameters related to lamp selection,
calibration, control of flow gases should be through window-
based software. It should have programmable and computer
controlled gas control, both air acetylene/nitrous oxide. System
should have all safety features required for the system. The unit
should be operated with Windows software controlling analyser,
setting-up the method parameters, analysing the samples,
performing calibrations. The verification kit STD for Cu to
check performance periodically to be included in offer.
Accessories • Hollow cathode lamps for Ag, As, Cd, Cr, Cu, Mn, Ni, Hg, Pb,
Se and Zn
• Gas filter station with control box
• Acetylene and nitrous oxide gases with cylinders
• Argon gas with cylinder and regulator required for hydride
system
• Fume hood and ducting with exhaust provision
• Air compressor
• Vent and hood assembly
• Nitrous oxide, acetylene and nitrogen two stage gas regulators
with necessary tubing and connectors, nitrous oxide regulator
should be with heater
• Air compressor with air filter or equivalent operation,
maintenance, service and analytical manuals, including
applications for flame and hydride units
• Spare nebuliser with capillary, flow spoiler, impact bead, O ring
kit for nebuliser, burner gasket kits for air- and nitrous oxide-
acetylene, additional spares and
• Dust covers
• 10 KVA online UPS with 30 minutes back-up
• Colour monitor
• Printer
• Complete operational manual
Consumables • Consumables for two year operation as recommended
considering expiry dates

Más contenido relacionado

La actualidad más candente

Chapter 4 chromatography_b_2010
Chapter 4 chromatography_b_2010Chapter 4 chromatography_b_2010
Chapter 4 chromatography_b_2010drs_swylam
 
Karl fischer titration
Karl  fischer titrationKarl  fischer titration
Karl fischer titrationJayanPatel10
 
Measurement Uncertainty
Measurement UncertaintyMeasurement Uncertainty
Measurement UncertaintyDenizErtekin2
 
Ensuring the validity of results
Ensuring the validity of resultsEnsuring the validity of results
Ensuring the validity of resultsGH Yeoh
 
RADIOACTIVE TRACER TECHNIQUE ON HEAT EXCHANGER
RADIOACTIVE TRACER TECHNIQUE ON HEAT EXCHANGERRADIOACTIVE TRACER TECHNIQUE ON HEAT EXCHANGER
RADIOACTIVE TRACER TECHNIQUE ON HEAT EXCHANGERBilal Ahmed Khan
 
Headspace analysis
Headspace analysisHeadspace analysis
Headspace analysisAnalysys
 
ISO 17025
ISO 17025 ISO 17025
ISO 17025 Akma Ija
 
Mass transferhshsh
Mass transferhshsh Mass transferhshsh
Mass transferhshsh megr1412
 
What Documentation Required for ISO 17025:2017 Accreditation?
What Documentation Required for ISO 17025:2017 Accreditation?What Documentation Required for ISO 17025:2017 Accreditation?
What Documentation Required for ISO 17025:2017 Accreditation?Global Manager Group
 
Hardgrove grindability index (HGI)
Hardgrove grindability index (HGI)Hardgrove grindability index (HGI)
Hardgrove grindability index (HGI)KOMAL AROOSH
 

La actualidad más candente (20)

Lec05 glp 2007
Lec05 glp 2007Lec05 glp 2007
Lec05 glp 2007
 
Chapter 4 chromatography_b_2010
Chapter 4 chromatography_b_2010Chapter 4 chromatography_b_2010
Chapter 4 chromatography_b_2010
 
The ISO 17025 standard: principles and management requirements
The ISO 17025 standard: principles and management requirementsThe ISO 17025 standard: principles and management requirements
The ISO 17025 standard: principles and management requirements
 
Karl fischer titration
Karl  fischer titrationKarl  fischer titration
Karl fischer titration
 
Improve Analysis Precision for ICP-OES and ICP-MS for Environmental and Geolo...
Improve Analysis Precision for ICP-OES and ICP-MS for Environmental and Geolo...Improve Analysis Precision for ICP-OES and ICP-MS for Environmental and Geolo...
Improve Analysis Precision for ICP-OES and ICP-MS for Environmental and Geolo...
 
Gas chromatography
Gas chromatographyGas chromatography
Gas chromatography
 
ICP Presentation
ICP PresentationICP Presentation
ICP Presentation
 
Microwave digestion
Microwave digestionMicrowave digestion
Microwave digestion
 
Measurement Uncertainty
Measurement UncertaintyMeasurement Uncertainty
Measurement Uncertainty
 
Pesticide analysis
Pesticide analysisPesticide analysis
Pesticide analysis
 
Ensuring the validity of results
Ensuring the validity of resultsEnsuring the validity of results
Ensuring the validity of results
 
RADIOACTIVE TRACER TECHNIQUE ON HEAT EXCHANGER
RADIOACTIVE TRACER TECHNIQUE ON HEAT EXCHANGERRADIOACTIVE TRACER TECHNIQUE ON HEAT EXCHANGER
RADIOACTIVE TRACER TECHNIQUE ON HEAT EXCHANGER
 
Headspace analysis
Headspace analysisHeadspace analysis
Headspace analysis
 
ISO 17025
ISO 17025 ISO 17025
ISO 17025
 
Gas chromatography GC
Gas chromatography GCGas chromatography GC
Gas chromatography GC
 
Mass transferhshsh
Mass transferhshsh Mass transferhshsh
Mass transferhshsh
 
ISO 17025 Accreditation
ISO 17025 AccreditationISO 17025 Accreditation
ISO 17025 Accreditation
 
Introduction of Fired heaters
Introduction of Fired heatersIntroduction of Fired heaters
Introduction of Fired heaters
 
What Documentation Required for ISO 17025:2017 Accreditation?
What Documentation Required for ISO 17025:2017 Accreditation?What Documentation Required for ISO 17025:2017 Accreditation?
What Documentation Required for ISO 17025:2017 Accreditation?
 
Hardgrove grindability index (HGI)
Hardgrove grindability index (HGI)Hardgrove grindability index (HGI)
Hardgrove grindability index (HGI)
 

Similar a Atomic Absorption Spectrophotometer for Trace Metal Analysis

Flame Graphite Furnace Atomic Absorption Spectrophotometer by ACMAS Technolog...
Flame Graphite Furnace Atomic Absorption Spectrophotometer by ACMAS Technolog...Flame Graphite Furnace Atomic Absorption Spectrophotometer by ACMAS Technolog...
Flame Graphite Furnace Atomic Absorption Spectrophotometer by ACMAS Technolog...Acmas Technologies Pvt. Ltd.
 
Zeeman Atomic Absorption Spectrophotometer by ACMAS Technologies Pvt Ltd.
Zeeman Atomic Absorption Spectrophotometer by ACMAS Technologies Pvt Ltd.Zeeman Atomic Absorption Spectrophotometer by ACMAS Technologies Pvt Ltd.
Zeeman Atomic Absorption Spectrophotometer by ACMAS Technologies Pvt Ltd.Acmas Technologies Pvt. Ltd.
 
MRU SWG 100 Fixed Biogas Analyzer
MRU SWG 100 Fixed Biogas Analyzer MRU SWG 100 Fixed Biogas Analyzer
MRU SWG 100 Fixed Biogas Analyzer Diamond Scientific
 
EXPERIMENTAL AND MEASUREMENT METHODS FINAL WORK
EXPERIMENTAL AND MEASUREMENT METHODS FINAL WORKEXPERIMENTAL AND MEASUREMENT METHODS FINAL WORK
EXPERIMENTAL AND MEASUREMENT METHODS FINAL WORKShaho Mohamedali
 
AccuThermo AW 820 Long Time Rapid Thermal Anneal Equipment
AccuThermo AW 820 Long Time  Rapid Thermal Anneal EquipmentAccuThermo AW 820 Long Time  Rapid Thermal Anneal Equipment
AccuThermo AW 820 Long Time Rapid Thermal Anneal EquipmentPeter Chen
 
Chromatographic Analysis of Natural Gas Liquids, sherry petrochemical
Chromatographic Analysis of Natural Gas Liquids, sherry petrochemicalChromatographic Analysis of Natural Gas Liquids, sherry petrochemical
Chromatographic Analysis of Natural Gas Liquids, sherry petrochemicalsherrylabs
 
Matt Heaters by Tristar Engineering & Chemical Co. Maharashtra
Matt Heaters by Tristar Engineering & Chemical Co. MaharashtraMatt Heaters by Tristar Engineering & Chemical Co. Maharashtra
Matt Heaters by Tristar Engineering & Chemical Co. MaharashtraIndiaMART InterMESH Limited
 
AccuThermo AW 810 Rapid Thermal Anneal Equipment
AccuThermo AW 810 Rapid Thermal Anneal EquipmentAccuThermo AW 810 Rapid Thermal Anneal Equipment
AccuThermo AW 810 Rapid Thermal Anneal EquipmentPeter Chen
 
Duncan cairnie ccm energy solutions hvacr seminar- april 2016
Duncan cairnie ccm energy solutions   hvacr seminar- april 2016Duncan cairnie ccm energy solutions   hvacr seminar- april 2016
Duncan cairnie ccm energy solutions hvacr seminar- april 2016ARAaus
 
Engineering tips for sampling systems [Piramoon & Buhler Technologies], by Po...
Engineering tips for sampling systems [Piramoon & Buhler Technologies], by Po...Engineering tips for sampling systems [Piramoon & Buhler Technologies], by Po...
Engineering tips for sampling systems [Piramoon & Buhler Technologies], by Po...Pouriya Niknam
 
Concept NoteR1.ppt
Concept NoteR1.pptConcept NoteR1.ppt
Concept NoteR1.pptMILIND569137
 
Atomic Absorption Spectrophotometer (AAS) by ACMAS Technologies Pvt Ltd.
Atomic Absorption Spectrophotometer (AAS) by ACMAS Technologies Pvt Ltd.Atomic Absorption Spectrophotometer (AAS) by ACMAS Technologies Pvt Ltd.
Atomic Absorption Spectrophotometer (AAS) by ACMAS Technologies Pvt Ltd.Acmas Technologies Pvt. Ltd.
 
Tube Wall Temperature Measurement On Steam Reformers - Best Practices
Tube Wall Temperature Measurement On Steam Reformers - Best PracticesTube Wall Temperature Measurement On Steam Reformers - Best Practices
Tube Wall Temperature Measurement On Steam Reformers - Best PracticesGerard B. Hawkins
 
Lead free reflow oven s-8800
Lead free reflow oven s-8800 Lead free reflow oven s-8800
Lead free reflow oven s-8800 Quella Fan
 
5. symposium pres pres flexi thema-d_22-06-2011
5. symposium pres pres flexi thema-d_22-06-20115. symposium pres pres flexi thema-d_22-06-2011
5. symposium pres pres flexi thema-d_22-06-2011Flexigas_Site
 
Lead free reflow oven s-1000
Lead free reflow oven s-1000Lead free reflow oven s-1000
Lead free reflow oven s-1000Harvey shao
 

Similar a Atomic Absorption Spectrophotometer for Trace Metal Analysis (20)

Specifications gc
Specifications gcSpecifications gc
Specifications gc
 
DMA Direct Mercury Analyzer
DMA  Direct Mercury AnalyzerDMA  Direct Mercury Analyzer
DMA Direct Mercury Analyzer
 
Flame Graphite Furnace Atomic Absorption Spectrophotometer by ACMAS Technolog...
Flame Graphite Furnace Atomic Absorption Spectrophotometer by ACMAS Technolog...Flame Graphite Furnace Atomic Absorption Spectrophotometer by ACMAS Technolog...
Flame Graphite Furnace Atomic Absorption Spectrophotometer by ACMAS Technolog...
 
Zeeman Atomic Absorption Spectrophotometer by ACMAS Technologies Pvt Ltd.
Zeeman Atomic Absorption Spectrophotometer by ACMAS Technologies Pvt Ltd.Zeeman Atomic Absorption Spectrophotometer by ACMAS Technologies Pvt Ltd.
Zeeman Atomic Absorption Spectrophotometer by ACMAS Technologies Pvt Ltd.
 
MRU SWG 100 Fixed Biogas Analyzer
MRU SWG 100 Fixed Biogas Analyzer MRU SWG 100 Fixed Biogas Analyzer
MRU SWG 100 Fixed Biogas Analyzer
 
EXPERIMENTAL AND MEASUREMENT METHODS FINAL WORK
EXPERIMENTAL AND MEASUREMENT METHODS FINAL WORKEXPERIMENTAL AND MEASUREMENT METHODS FINAL WORK
EXPERIMENTAL AND MEASUREMENT METHODS FINAL WORK
 
AccuThermo AW 820 Long Time Rapid Thermal Anneal Equipment
AccuThermo AW 820 Long Time  Rapid Thermal Anneal EquipmentAccuThermo AW 820 Long Time  Rapid Thermal Anneal Equipment
AccuThermo AW 820 Long Time Rapid Thermal Anneal Equipment
 
pH electrodes-pH sensors CPS41
pH electrodes-pH sensors CPS41pH electrodes-pH sensors CPS41
pH electrodes-pH sensors CPS41
 
Chromatographic Analysis of Natural Gas Liquids, sherry petrochemical
Chromatographic Analysis of Natural Gas Liquids, sherry petrochemicalChromatographic Analysis of Natural Gas Liquids, sherry petrochemical
Chromatographic Analysis of Natural Gas Liquids, sherry petrochemical
 
Matt Heaters by Tristar Engineering & Chemical Co. Maharashtra
Matt Heaters by Tristar Engineering & Chemical Co. MaharashtraMatt Heaters by Tristar Engineering & Chemical Co. Maharashtra
Matt Heaters by Tristar Engineering & Chemical Co. Maharashtra
 
AccuThermo AW 810 Rapid Thermal Anneal Equipment
AccuThermo AW 810 Rapid Thermal Anneal EquipmentAccuThermo AW 810 Rapid Thermal Anneal Equipment
AccuThermo AW 810 Rapid Thermal Anneal Equipment
 
Duncan cairnie ccm energy solutions hvacr seminar- april 2016
Duncan cairnie ccm energy solutions   hvacr seminar- april 2016Duncan cairnie ccm energy solutions   hvacr seminar- april 2016
Duncan cairnie ccm energy solutions hvacr seminar- april 2016
 
Engineering tips for sampling systems [Piramoon & Buhler Technologies], by Po...
Engineering tips for sampling systems [Piramoon & Buhler Technologies], by Po...Engineering tips for sampling systems [Piramoon & Buhler Technologies], by Po...
Engineering tips for sampling systems [Piramoon & Buhler Technologies], by Po...
 
Cone calorimeter
Cone calorimeterCone calorimeter
Cone calorimeter
 
Concept NoteR1.ppt
Concept NoteR1.pptConcept NoteR1.ppt
Concept NoteR1.ppt
 
Atomic Absorption Spectrophotometer (AAS) by ACMAS Technologies Pvt Ltd.
Atomic Absorption Spectrophotometer (AAS) by ACMAS Technologies Pvt Ltd.Atomic Absorption Spectrophotometer (AAS) by ACMAS Technologies Pvt Ltd.
Atomic Absorption Spectrophotometer (AAS) by ACMAS Technologies Pvt Ltd.
 
Tube Wall Temperature Measurement On Steam Reformers - Best Practices
Tube Wall Temperature Measurement On Steam Reformers - Best PracticesTube Wall Temperature Measurement On Steam Reformers - Best Practices
Tube Wall Temperature Measurement On Steam Reformers - Best Practices
 
Lead free reflow oven s-8800
Lead free reflow oven s-8800 Lead free reflow oven s-8800
Lead free reflow oven s-8800
 
5. symposium pres pres flexi thema-d_22-06-2011
5. symposium pres pres flexi thema-d_22-06-20115. symposium pres pres flexi thema-d_22-06-2011
5. symposium pres pres flexi thema-d_22-06-2011
 
Lead free reflow oven s-1000
Lead free reflow oven s-1000Lead free reflow oven s-1000
Lead free reflow oven s-1000
 

Más de hydrologyproject0 (20)

Specification purewater
Specification purewaterSpecification purewater
Specification purewater
 
Water purifier ion exchange resin column
Water purifier ion exchange resin columnWater purifier ion exchange resin column
Water purifier ion exchange resin column
 
Water purifier, double distillation unit
Water purifier, double distillation unitWater purifier, double distillation unit
Water purifier, double distillation unit
 
Water purifier, distillation unit (automatic)
Water purifier, distillation unit (automatic)Water purifier, distillation unit (automatic)
Water purifier, distillation unit (automatic)
 
Water geyser (heater)
Water geyser (heater)Water geyser (heater)
Water geyser (heater)
 
Water bath, general purpose
Water bath, general purposeWater bath, general purpose
Water bath, general purpose
 
Water bath, bacteriological
Water bath, bacteriologicalWater bath, bacteriological
Water bath, bacteriological
 
Wash bottle
Wash bottleWash bottle
Wash bottle
 
Volumetric flask
Volumetric flaskVolumetric flask
Volumetric flask
 
Visual accumulation tube
Visual accumulation tubeVisual accumulation tube
Visual accumulation tube
 
Vacuum pump
Vacuum pumpVacuum pump
Vacuum pump
 
Turbidity meter (nephelometer)
Turbidity meter (nephelometer)Turbidity meter (nephelometer)
Turbidity meter (nephelometer)
 
Tool kit
Tool kitTool kit
Tool kit
 
Toc analyser, microprocessor controlled
Toc analyser, microprocessor controlledToc analyser, microprocessor controlled
Toc analyser, microprocessor controlled
 
Tissue grinder
Tissue grinderTissue grinder
Tissue grinder
 
Thermometers, set of
Thermometers, set ofThermometers, set of
Thermometers, set of
 
Test tubes
Test tubesTest tubes
Test tubes
 
Test sieves with shaker
Test sieves with shakerTest sieves with shaker
Test sieves with shaker
 
Stirrer, magnetic (with hot plate)
Stirrer, magnetic (with hot plate)Stirrer, magnetic (with hot plate)
Stirrer, magnetic (with hot plate)
 
Sterliser, autoclave
Sterliser, autoclaveSterliser, autoclave
Sterliser, autoclave
 

Último

Boost the utilization of your HCL environment by reevaluating use cases and f...
Boost the utilization of your HCL environment by reevaluating use cases and f...Boost the utilization of your HCL environment by reevaluating use cases and f...
Boost the utilization of your HCL environment by reevaluating use cases and f...Roland Driesen
 
Best Basmati Rice Manufacturers in India
Best Basmati Rice Manufacturers in IndiaBest Basmati Rice Manufacturers in India
Best Basmati Rice Manufacturers in IndiaShree Krishna Exports
 
Call Girls In DLf Gurgaon ➥99902@11544 ( Best price)100% Genuine Escort In 24...
Call Girls In DLf Gurgaon ➥99902@11544 ( Best price)100% Genuine Escort In 24...Call Girls In DLf Gurgaon ➥99902@11544 ( Best price)100% Genuine Escort In 24...
Call Girls In DLf Gurgaon ➥99902@11544 ( Best price)100% Genuine Escort In 24...lizamodels9
 
A DAY IN THE LIFE OF A SALESMAN / WOMAN
A DAY IN THE LIFE OF A  SALESMAN / WOMANA DAY IN THE LIFE OF A  SALESMAN / WOMAN
A DAY IN THE LIFE OF A SALESMAN / WOMANIlamathiKannappan
 
Lucknow 💋 Escorts in Lucknow - 450+ Call Girl Cash Payment 8923113531 Neha Th...
Lucknow 💋 Escorts in Lucknow - 450+ Call Girl Cash Payment 8923113531 Neha Th...Lucknow 💋 Escorts in Lucknow - 450+ Call Girl Cash Payment 8923113531 Neha Th...
Lucknow 💋 Escorts in Lucknow - 450+ Call Girl Cash Payment 8923113531 Neha Th...anilsa9823
 
Pharma Works Profile of Karan Communications
Pharma Works Profile of Karan CommunicationsPharma Works Profile of Karan Communications
Pharma Works Profile of Karan Communicationskarancommunications
 
Monthly Social Media Update April 2024 pptx.pptx
Monthly Social Media Update April 2024 pptx.pptxMonthly Social Media Update April 2024 pptx.pptx
Monthly Social Media Update April 2024 pptx.pptxAndy Lambert
 
RSA Conference Exhibitor List 2024 - Exhibitors Data
RSA Conference Exhibitor List 2024 - Exhibitors DataRSA Conference Exhibitor List 2024 - Exhibitors Data
RSA Conference Exhibitor List 2024 - Exhibitors DataExhibitors Data
 
Event mailer assignment progress report .pdf
Event mailer assignment progress report .pdfEvent mailer assignment progress report .pdf
Event mailer assignment progress report .pdftbatkhuu1
 
Cracking the Cultural Competence Code.pptx
Cracking the Cultural Competence Code.pptxCracking the Cultural Competence Code.pptx
Cracking the Cultural Competence Code.pptxWorkforce Group
 
Progress Report - Oracle Database Analyst Summit
Progress  Report - Oracle Database Analyst SummitProgress  Report - Oracle Database Analyst Summit
Progress Report - Oracle Database Analyst SummitHolger Mueller
 
KYC-Verified Accounts: Helping Companies Handle Challenging Regulatory Enviro...
KYC-Verified Accounts: Helping Companies Handle Challenging Regulatory Enviro...KYC-Verified Accounts: Helping Companies Handle Challenging Regulatory Enviro...
KYC-Verified Accounts: Helping Companies Handle Challenging Regulatory Enviro...Any kyc Account
 
Call Girls Pune Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Pune Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Pune Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Pune Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Ensure the security of your HCL environment by applying the Zero Trust princi...
Ensure the security of your HCL environment by applying the Zero Trust princi...Ensure the security of your HCL environment by applying the Zero Trust princi...
Ensure the security of your HCL environment by applying the Zero Trust princi...Roland Driesen
 
Call Girls Jp Nagar Just Call 👗 7737669865 👗 Top Class Call Girl Service Bang...
Call Girls Jp Nagar Just Call 👗 7737669865 👗 Top Class Call Girl Service Bang...Call Girls Jp Nagar Just Call 👗 7737669865 👗 Top Class Call Girl Service Bang...
Call Girls Jp Nagar Just Call 👗 7737669865 👗 Top Class Call Girl Service Bang...amitlee9823
 
Call Girls In Holiday Inn Express Gurugram➥99902@11544 ( Best price)100% Genu...
Call Girls In Holiday Inn Express Gurugram➥99902@11544 ( Best price)100% Genu...Call Girls In Holiday Inn Express Gurugram➥99902@11544 ( Best price)100% Genu...
Call Girls In Holiday Inn Express Gurugram➥99902@11544 ( Best price)100% Genu...lizamodels9
 
Creating Low-Code Loan Applications using the Trisotech Mortgage Feature Set
Creating Low-Code Loan Applications using the Trisotech Mortgage Feature SetCreating Low-Code Loan Applications using the Trisotech Mortgage Feature Set
Creating Low-Code Loan Applications using the Trisotech Mortgage Feature SetDenis Gagné
 
A305_A2_file_Batkhuu progress report.pdf
A305_A2_file_Batkhuu progress report.pdfA305_A2_file_Batkhuu progress report.pdf
A305_A2_file_Batkhuu progress report.pdftbatkhuu1
 
Call Girls Navi Mumbai Just Call 9907093804 Top Class Call Girl Service Avail...
Call Girls Navi Mumbai Just Call 9907093804 Top Class Call Girl Service Avail...Call Girls Navi Mumbai Just Call 9907093804 Top Class Call Girl Service Avail...
Call Girls Navi Mumbai Just Call 9907093804 Top Class Call Girl Service Avail...Dipal Arora
 

Último (20)

Boost the utilization of your HCL environment by reevaluating use cases and f...
Boost the utilization of your HCL environment by reevaluating use cases and f...Boost the utilization of your HCL environment by reevaluating use cases and f...
Boost the utilization of your HCL environment by reevaluating use cases and f...
 
Best Basmati Rice Manufacturers in India
Best Basmati Rice Manufacturers in IndiaBest Basmati Rice Manufacturers in India
Best Basmati Rice Manufacturers in India
 
Call Girls In DLf Gurgaon ➥99902@11544 ( Best price)100% Genuine Escort In 24...
Call Girls In DLf Gurgaon ➥99902@11544 ( Best price)100% Genuine Escort In 24...Call Girls In DLf Gurgaon ➥99902@11544 ( Best price)100% Genuine Escort In 24...
Call Girls In DLf Gurgaon ➥99902@11544 ( Best price)100% Genuine Escort In 24...
 
A DAY IN THE LIFE OF A SALESMAN / WOMAN
A DAY IN THE LIFE OF A  SALESMAN / WOMANA DAY IN THE LIFE OF A  SALESMAN / WOMAN
A DAY IN THE LIFE OF A SALESMAN / WOMAN
 
Lucknow 💋 Escorts in Lucknow - 450+ Call Girl Cash Payment 8923113531 Neha Th...
Lucknow 💋 Escorts in Lucknow - 450+ Call Girl Cash Payment 8923113531 Neha Th...Lucknow 💋 Escorts in Lucknow - 450+ Call Girl Cash Payment 8923113531 Neha Th...
Lucknow 💋 Escorts in Lucknow - 450+ Call Girl Cash Payment 8923113531 Neha Th...
 
Pharma Works Profile of Karan Communications
Pharma Works Profile of Karan CommunicationsPharma Works Profile of Karan Communications
Pharma Works Profile of Karan Communications
 
Monthly Social Media Update April 2024 pptx.pptx
Monthly Social Media Update April 2024 pptx.pptxMonthly Social Media Update April 2024 pptx.pptx
Monthly Social Media Update April 2024 pptx.pptx
 
RSA Conference Exhibitor List 2024 - Exhibitors Data
RSA Conference Exhibitor List 2024 - Exhibitors DataRSA Conference Exhibitor List 2024 - Exhibitors Data
RSA Conference Exhibitor List 2024 - Exhibitors Data
 
Event mailer assignment progress report .pdf
Event mailer assignment progress report .pdfEvent mailer assignment progress report .pdf
Event mailer assignment progress report .pdf
 
Cracking the Cultural Competence Code.pptx
Cracking the Cultural Competence Code.pptxCracking the Cultural Competence Code.pptx
Cracking the Cultural Competence Code.pptx
 
unwanted pregnancy Kit [+918133066128] Abortion Pills IN Dubai UAE Abudhabi
unwanted pregnancy Kit [+918133066128] Abortion Pills IN Dubai UAE Abudhabiunwanted pregnancy Kit [+918133066128] Abortion Pills IN Dubai UAE Abudhabi
unwanted pregnancy Kit [+918133066128] Abortion Pills IN Dubai UAE Abudhabi
 
Progress Report - Oracle Database Analyst Summit
Progress  Report - Oracle Database Analyst SummitProgress  Report - Oracle Database Analyst Summit
Progress Report - Oracle Database Analyst Summit
 
KYC-Verified Accounts: Helping Companies Handle Challenging Regulatory Enviro...
KYC-Verified Accounts: Helping Companies Handle Challenging Regulatory Enviro...KYC-Verified Accounts: Helping Companies Handle Challenging Regulatory Enviro...
KYC-Verified Accounts: Helping Companies Handle Challenging Regulatory Enviro...
 
Call Girls Pune Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Pune Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Pune Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Pune Just Call 9907093804 Top Class Call Girl Service Available
 
Ensure the security of your HCL environment by applying the Zero Trust princi...
Ensure the security of your HCL environment by applying the Zero Trust princi...Ensure the security of your HCL environment by applying the Zero Trust princi...
Ensure the security of your HCL environment by applying the Zero Trust princi...
 
Call Girls Jp Nagar Just Call 👗 7737669865 👗 Top Class Call Girl Service Bang...
Call Girls Jp Nagar Just Call 👗 7737669865 👗 Top Class Call Girl Service Bang...Call Girls Jp Nagar Just Call 👗 7737669865 👗 Top Class Call Girl Service Bang...
Call Girls Jp Nagar Just Call 👗 7737669865 👗 Top Class Call Girl Service Bang...
 
Call Girls In Holiday Inn Express Gurugram➥99902@11544 ( Best price)100% Genu...
Call Girls In Holiday Inn Express Gurugram➥99902@11544 ( Best price)100% Genu...Call Girls In Holiday Inn Express Gurugram➥99902@11544 ( Best price)100% Genu...
Call Girls In Holiday Inn Express Gurugram➥99902@11544 ( Best price)100% Genu...
 
Creating Low-Code Loan Applications using the Trisotech Mortgage Feature Set
Creating Low-Code Loan Applications using the Trisotech Mortgage Feature SetCreating Low-Code Loan Applications using the Trisotech Mortgage Feature Set
Creating Low-Code Loan Applications using the Trisotech Mortgage Feature Set
 
A305_A2_file_Batkhuu progress report.pdf
A305_A2_file_Batkhuu progress report.pdfA305_A2_file_Batkhuu progress report.pdf
A305_A2_file_Batkhuu progress report.pdf
 
Call Girls Navi Mumbai Just Call 9907093804 Top Class Call Girl Service Avail...
Call Girls Navi Mumbai Just Call 9907093804 Top Class Call Girl Service Avail...Call Girls Navi Mumbai Just Call 9907093804 Top Class Call Girl Service Avail...
Call Girls Navi Mumbai Just Call 9907093804 Top Class Call Girl Service Avail...
 

Atomic Absorption Spectrophotometer for Trace Metal Analysis

  • 1. ATOMIC ABSORPTION SPECTROPHOTOMETER Purpose For analysis of trace metals and other elements in water and sediment samples. Fully automated PC controlled true double beam Atomic Absorption Spectrophotometer with absorption and emission capability with deuterium background correction and should have the capability of measuring multi elements in rapid sequence in single sample measurement, fitted with 4 or more lamp position and choice of manual or automatic lamp selection with following features: Conditions & Requirements • Installation, commissioning and training in operational use shall be by supplier. • At least 5 years of after sales service shall be guaranteed Specifications General Features • Automatic wave length and slit selection • Wide range and high sensitivity photomultiplier for maximum signal to noise ratio • Choice of burner type (air/acetylene or nitrous oxide) • Deuterium background correction. Range 185-425 mm to 2.3 Abs. Electronic modulation with automatic attenuation • Integrated built-in Flame and Graphite Furnace along with auto- sampler. • It should be automatically aligned in the path by using the software. Spectrophotometer • Type: PC-controlled • Optics: hollow cathode lamps, double beam, narrow beam optics matches flame profile. • Hollow Cathode Lamp or Electrode Less Discharge (EDL) Lamp • All reflective system with quartz over-coating on mirrors. • Sealed against dust and vapour • Wave length range: 180 to 900 nm • Slit 0.2 to 2 nm with automatic and manual choice Monochromator • Monochromator with computer controlled wavelength selection and peaking
  • 2. • The system should be based on abrasion corrected Cnerny- Turner Grating (holographic diffraction grating) up to 1800 grooves per mm blazed at 240 nm • Continuously adjustable silts with 0.2 to 2 nm spectral bandwidth Sensitivity • At least 0.7 Abs Absorption Cell • Quartz Lamp turret • At least 4 lamps with automatic Turret controlled with software and auto-alignment Gas supply • 0 to 120 ml/min Gas Control • Automatic – preset oxidant flow and variable full flow adjusted using manual flow meter adjustment. Automatic oxidant change over • Fully automatic gas box with computer-controlled oxidant selection, automatic gas sequencing, oxidant and fuel monitoring and control • Actuated ignition with air/acetylene. Acetylene flow should be automatically adjustable when switching to or from nitrous- oxide/acetylene operation Flame control • Programmable and manual oxidant and fuel selection and • Monitoring • Safety interlocks • Safe and automatic shutdown in case of power failure Safety system • Separate ignite/flame off buttons. Safety interlocks to monitor burner type, burner fitting, liquid trap, pressure relief bung, flame shield, flame operation, main power, oxidant pressure and deuterium lamp cover • Fully interlocked operation preventing ignition if proper burner head, nebulizer, end cap or burner drain system are not correctly installed, the level of liquid in the drain vessel is incorrect or gas pressures are too low • The interlocks should be with automatically shut down the gases
  • 3. if a flame is not detected • In the event of power failure or when the emergency flame off button is used, the flame should be automatically and safely extinguished Spray chamber • Solid poly-propylene or fluorinated HDP or equivalent • Or fluorinated high density polyethylene for aqueous and organic solutions Burner • All titanium or equivalent, for air-acetylene and nitrous oxide- acetylene flames or air/acetylene burner with Teflon shield. • Burner system chamber should provide analysis of corrosive and high solid matrices • Automatic burner rotation system • Interchangeable full titanium nebulizer for organic and acidic purpose and burner head for 50 mm and 100 mm, facilities should be there to switch from Flame to Graphite within few seconds without removing the air acetylene or nitrous oxide burner. Mixing paddles • Removal twin headed mixing paddles fitted inside the spray chamber for better mixing and extend operation with high dissolved solids solution Impact bead • Externally adjustable glass impact bead (Teflon bead) Hydride Generator • Hydride generator (for AS, Se, Hg at ppb level), automatic and computer controlled • Sample should be injected through the peristaltic pump, 3 channel peristaltic pump with heated quartz cell, fast heating rate, temperature range should be from 100 to 1000 degree Celsius • Typical precision 1-2% RSD with sample through put of 50 samples/hour determined in triplicate Graphite Furnace • Built-in integrated furnace with flame facility to switch from flame to graphite within few seconds without removing the air acetylene or nitrous oxide burner • It should automatically align in optical path by using the
  • 4. software • It should have fully automated constant temperature zone with dynamic feedback temperature control, cooling water temperature compensation, upto 20 steps per program, temperature programmable from 40 to 3000o C with maximum heating rate up to 2000o C, choice of two inert gases with programmable flow rate, solid titanium work head for maximum corrosion resistance, quartz and windows for high light energy, comprehensive safety interlocks for gases, water flow, tube and transformer temperature, DC Power supply used for Graphite Furnace with strict electromagnetic compatibility • Spray chamber: solid polypropylene of fluorinated HDP or equivalent • Recirculating Water chiller: with flow rates of 1.5-2 L/Min at 175-200 kPa for cooling of Graphite Furnace must be quoted from manufacturer • Autosampler for Graphite Furnace: Automatic programmable sample dispenser for GTA 50 samples in 2 mL vials, 5 25 mL vessels for bulk solutions, automatic preparation up to 10 points calibration, hot injection automatic flow through rinsing • Easy access for graphite tube replacement with built-in tube usage counter should be available • Auto sampler should be provided with graphite furnace Operational requirement Background correction: • Sensitivity: 0.7 absorbance or better for 5 ppm Cu • Calibration: automatic, using up to 6 standards and re- calibration with single standard Power supply • 220 ±10 VAC, 47 to 53 Hz Typical performance • Greater than 0.8 Absorbance with precision of <0.5% RSD from ten 5 sec integration for 5 mg/l Cu solution Other requirements • PC Compatible for operation control and monitoring of AAS and necessary modules with latest configuration and software, core 2 DUO processor, 3 GHz, 2 GB DDR RAM, 48x CD ROM Drive, 320 GB HDD, 17” TFT Colour Monitor, 104 key board, optical mouse, Window 7 and licensed MS Office with CD, serial and parallel ports, DeskJet colour printer with suitable
  • 5. stabilizer • Software performing analyses at least as per DIN/ISO/US-EPA, control of flame, calibration, blank correction, data import, export, handling and reporting, quality control protocols, computer based training • All the methods, parameters related to lamp selection, calibration, control of flow gases should be through window- based software. It should have programmable and computer controlled gas control, both air acetylene/nitrous oxide. System should have all safety features required for the system. The unit should be operated with Windows software controlling analyser, setting-up the method parameters, analysing the samples, performing calibrations. The verification kit STD for Cu to check performance periodically to be included in offer. Accessories • Hollow cathode lamps for Ag, As, Cd, Cr, Cu, Mn, Ni, Hg, Pb, Se and Zn • Gas filter station with control box • Acetylene and nitrous oxide gases with cylinders • Argon gas with cylinder and regulator required for hydride system • Fume hood and ducting with exhaust provision • Air compressor • Vent and hood assembly • Nitrous oxide, acetylene and nitrogen two stage gas regulators with necessary tubing and connectors, nitrous oxide regulator should be with heater • Air compressor with air filter or equivalent operation, maintenance, service and analytical manuals, including applications for flame and hydride units • Spare nebuliser with capillary, flow spoiler, impact bead, O ring kit for nebuliser, burner gasket kits for air- and nitrous oxide- acetylene, additional spares and • Dust covers
  • 6. • 10 KVA online UPS with 30 minutes back-up • Colour monitor • Printer • Complete operational manual Consumables • Consumables for two year operation as recommended considering expiry dates