Bhide A, Acharya G, Bilardo CM, Brezinka C, Cafici D, Hernandez- Andrade
E, Kalache K, Kingdom J, Kiserud T, Lee W, Lees C, Leung KY, Malinger G,
 Mari G, Prefumo F, Sepulveda W and Trudinger B. on behalf of the ISUOG
                      Clinical Standards Committee
Why ?
•The International Society of Ultrasound in Obstetrics and Gynecology
(ISUOG) is a scientific organization that encourages sound clinical practice,
teaching and research for diagnostic imging in women’s healthcare.
•Practice Guidelines and Consensus are intended to reflect what is
considered by ISUOG to be the best practices at the time at which they
were issued.
•Although ISUOG has made every effort to ensure that guidelines are
accurate when issued, neither the Society nor any of its employees or
members accepts any liability for the consequences of any inaccurate or
misleading data, opinions or statements issued by the CSC.
•Guidelines are not intended to establish a legal standard of care because
interpretation of the evidence that underpins the guidelines may be
influenced by individual circumstances and available resources.
•Approved guidelines can be distributed freely with the permission of ISUOG
(info@isuog.org).
Scope
• Summarizes practice guidelines regarding
  how to perform Doppler ultrasonography of
  the feto-placental circulation
• Not intended to
   - define clinical indications,
   - specify proper timing of the Doppler
   examination in pregnancy
   - how to interpret findings
Guidance on safety

•   It is of utmost importance not to expose the embryo and
    fetus to unduly harmful ultrasound energy, particularly in the
    earliest stages of pregnancy.
•   At these stages, Doppler recording, when clinically
    indicated, should be performed at the lowest possible
    energy levels.
•   When performing Doppler imaging, the displayed thermal
    index (TI) should be ≤ 1.0 and exposure time should be kept
    as short as possible, usually no longer than 5–10 min and
    not exceeding 60 min
Pulsed wave Doppler: How can the accuracy of
        Doppler measurements be optimized?

•   Recordings should be obtained during absence of fetal
    breathing and body movements, and if necessary during
    temporary maternal breath hold
•   Colour flow mapping is not mandatory, although it is very
    helpful in the identification of the vessel of interest and in
    defining the direction of blood flow
•   The optimal insonation is complete alignment with the blood
    flow
•   It is advisable to start with relatively wide Doppler gate
    (sample volume)
Pulsed wave Doppler: How can the accuracy of
          Doppler measurements be optimized?
•   Similar to gray-scale imaging, the penetration and resolution of the
    Doppler beam can be optimized by adjusting the frequency (MHz) of the
    Doppler probe
•   By convention, it should be set as low as possible (≤ 50–60 Hz) in order
    to eliminate the low- frequency noise from peripheral blood vessels
•   Doppler horizontal sweep speed should be fast enough to separate
    successive waveforms
•   PRF adjusted so that the waveform fills 75% of Doppler screen
•   Doppler measurements should be reproducible
•   Gain should be adjusted in order to see clearly the Doppler velocity
    waveform, without the presence of artifacts in the background of the
    display
•   It is advisable not to invert the Doppler display on the ultrasound screen
Colour directional Doppler: How can the accuracy
        of Doppler measurements be optimized?
•   The use of colour Doppler increases the total power emitted. Colour
    Doppler resolution increases when the colour box is reduced in size
•   The box should be kept as small as possible to include only the studied
    area
•   The velocity scale or PRF should be adjusted to represent the real colour
    velocity of the studied vessel
•   Gain should be adjusted in order to prevent noise and artifacts
•   The angle of insonation should be adjusted by optimizing the position of
    the ultrasound probe according to the vessel or area studied
Power Doppler and directional power Doppler:
    How can the accuracy of Doppler measurements
                    be optimized?
•    The angle of insonation has less effect on power Doppler signals;
     nevertheless, the same optimization process as for colour directional
     Doppler must be performed
•    There is no aliasing phenomenon using power Doppler; however, an
     inappropriately low PRF may lead to noises and artifacts
•    Gain should be reduced in order to prevent amplification of noise
What equipment is needed for Doppler
     evaluation of the feto-placental circulation?

•   Equipment should have colour flow and spectral wave Doppler capabilities with
    on-screen display of flow velocity scales or pulse repetition frequency (PRF) and
    Doppler ultrasound frequency (in MHz)
•   Mechanical index (MI) and Thermal Index (TI) should be displayed on the
    ultrasound screen
•   Ultrasound system should generate a maximum velocity envelope (MVE) showing
    the whole spectral Doppler waveform
•   MVE should be possible to delineate using automatic or manual waveform traces
•   System software must be able to estimate peak systolic velocity (PSV), end-
    diastolic velocity (EDV) and time-averaged maximum velocity from the MVE and to
    calculate the commonly used Doppler indices i.e. pulsatility (PI) and resistance
    (RI) indices and systolic/diastolic velocity (S/D) ratio. On the tracing, the various
    points included in the calculations should be indicated, to ensure correctly
    calculated indices
What is the appropriate technique for
  obtaining uterine artery Doppler waveforms?

• Trans-abdominally, the probe is placed longitudinally in the
  lower lateral quadrant of the abdomen angled medially
• Colour flow mapping is useful to identify the uterine artery as
  it appears to cross the external iliac artery
• Sample volume is placed 1 cm downstream to the crossover
  point
• If the uterine artery branches before the intersection of the
  external iliac artery, the sample volume should be placed on
  the main artery just before the bifurcation
Normal waveform from the uterine artery
Abnormal waveform from the uterine artery. Note notch (arrow)
implying increased resistance in the uterine artery
Uterine Artery Doppler: Technique
              (Transvaginal)
•   Women should be asked to empty their bladder and
    placed in the dorsal lithotomy position
•   The probe should be placed into the lateral fornix and the
    uterine artery identified using colour Doppler at the level
    of the internal cervical os
•   The same should then be repeated for the contra-lateral
    uterine artery
•   Reference ranges for uterine artery Doppler indices
    depend on the technique of measurement, so
    appropriate reference ranges should be used for
    transabdominal and transvaginal routes
What is the appropriate technique for obtaining
umbilical artery Doppler waveforms?
• There is a significant difference in Doppler indices
  measured at the fetal end, the free loop and the placental
  end of the umbilical cord
• For the sake of simplicity and consistency, measurements
  should be made in a free cord loop
• In multiple pregnancies, and/or when comparing repeated
  measurements longitudinally, recordings from fixed sites
  (fetal end, placental end or intra-abdominal portion) may be
  more reliable
• Appropriate reference ranges should be used according to
  the site of interrogation
Inappropriate technique: Too low temporal an spatial resolution of the waveforms
Appropriate technique: note better quality spectral Doppler tracing
What is the appropriate technique for obtaining
  fetal middle cerebral artery Doppler waveforms?
• An axial section of the brain, including the thalami and the
  sphenoid bone wings is obtained, and magnified
• Colour flow mapping is used to identify the circle of Willis
  and the proximal middle cerebral artery (MCA)
• The pulsed-wave Doppler gate is then placed at the
  proximal one-third of the MCA close to its origin in the
  internal carotid artery
• The angle between the ultrasound beam and the direction
  of blood flow is kept as close as possible to 0 °, and
  should be less than 20 °
What is the appropriate technique for obtaining
  fetal middle cerebral artery Doppler waveforms?
• Care should be taken to avoid any unnecessary pressure
  on the fetal head
• At least three and fewer than 10 consecutive waveforms
  should be recorded. The highest point of the waveform is
  considered as the PSV (cm/s)
• The PSV can be measured using manual calipers or
  auto-trace. The latter yields significantly lower medians
  than does the former, but more closely approximates
  published medians used in clinical practice
• PI is usually calculated using auto-trace measurement,
  but manual tracing is also acceptable
Acceptable middle cerebral artery Doppler shift waveform. Note insonation
What is the appropriate technique for obtaining fetal
 venous Doppler waveforms?
• The vessel is identified by visualizing this connection by 2D
  imaging either in a mid-sagittal longitudinal plane of the fetal
  trunk or in an oblique transverse plane through the upper
  abdomen
• Colour flow mapping demonstrating the high velocity at the
  narrow entrance of the DV confirms its identification and
  indicates the standard sampling site for Doppler
  measurements
• In early pregnancy and in compromised pregnancies
  particular care has to be taken to reduce the sample volume
  appropriately in order to ensure clean recording of the
  lowest velocity during atrial contraction
What is the appropriate technique for obtaining fetal
venous Doppler waveforms?

• The waveform is usually triphasic, but biphasic and
  non-pulsating recordings, though rarer, may be
  seen in healthy fetuses
• The velocities are relatively high, between 55 and
  90 cm/s for most of the second half of pregnancy,
  but lower in early pregnancy
Ductus venosus Doppler recording with sagittal insonation aligning with
the isthmic portion without angle correction. Low-velocity vessel wall filter
(arrow) does not interfere with a-wave (a), which is far from zero line. High
sweep speed allows detailed visualization of variation in velocity
Which indices to use?
• S/D ratio, Resistance index (RI) and Pulsatility
  Index (PI) are the three well-known indices to
  describe arterial flow velocity waveforms
• All three are highly correlated
• There is no published evidence showing
  superiority of any one index over the others, but PI
  is the most commonly used index in current
  practice
• Pulsatility index for veins (PIV) is most commonly
  used for venous waveforms in the current literature
ISUOG’s website is a great resource for more information:

Más contenido relacionado

La actualidad más candente

Anterpartum fetal surveillance
Anterpartum fetal surveillance   Anterpartum fetal surveillance
Anterpartum fetal surveillance maricar chua
 
Doppler interpretation in pregnancy
Doppler interpretation in pregnancyDoppler interpretation in pregnancy
Doppler interpretation in pregnancyAboubakr Elnashar
 
Doppler in gyneacology Dr. Muhammad Bin Zulfiqar
Doppler in gyneacology Dr. Muhammad Bin ZulfiqarDoppler in gyneacology Dr. Muhammad Bin Zulfiqar
Doppler in gyneacology Dr. Muhammad Bin ZulfiqarDr. Muhammad Bin Zulfiqar
 
Role of ultrasound in emergency obstetrics dr.shreedhar
Role of ultrasound in emergency obstetrics dr.shreedharRole of ultrasound in emergency obstetrics dr.shreedhar
Role of ultrasound in emergency obstetrics dr.shreedharTeleradiology Solutions
 
Routine fetal anatomy scan at 18-23 weeks
Routine fetal anatomy scan at 18-23 weeksRoutine fetal anatomy scan at 18-23 weeks
Routine fetal anatomy scan at 18-23 weeksAli Bendary
 
Isuog practice guidelines performance of first trimester fetal ultrasound scan
Isuog practice guidelines performance of first trimester fetal ultrasound scanIsuog practice guidelines performance of first trimester fetal ultrasound scan
Isuog practice guidelines performance of first trimester fetal ultrasound scankaleemullahabid
 
Cesarean Scar Ectopic Pregnancy Current Management Strategies
Cesarean Scar Ectopic Pregnancy Current Management StrategiesCesarean Scar Ectopic Pregnancy Current Management Strategies
Cesarean Scar Ectopic Pregnancy Current Management StrategiesAboubakr Elnashar
 
The role of uterine artery embolization in gynecology practice
The role of uterine artery embolization in gynecology practiceThe role of uterine artery embolization in gynecology practice
The role of uterine artery embolization in gynecology practiceApollo Hospitals
 
C O L O R D O P P L E R I N F E T A L H Y P O X I A
C O L O R  D O P P L E R  I N  F E T A L  H Y P O X I AC O L O R  D O P P L E R  I N  F E T A L  H Y P O X I A
C O L O R D O P P L E R I N F E T A L H Y P O X I ANARENDRA MALHOTRA
 
Role of ultrasound in the Infertility management
Role of ultrasound in the Infertility  management Role of ultrasound in the Infertility  management
Role of ultrasound in the Infertility management Bharti Gahtori
 
Doppler us in the evaluation of fetal growth
Doppler us in the evaluation of fetal growthDoppler us in the evaluation of fetal growth
Doppler us in the evaluation of fetal growthSumiya Arshad
 
Role of tubal surgery in era of ivf
Role of tubal surgery in era of ivfRole of tubal surgery in era of ivf
Role of tubal surgery in era of ivfSanjay Makwana
 
Color doppler in FGR making sence of waves
Color doppler in FGR making sence of wavesColor doppler in FGR making sence of waves
Color doppler in FGR making sence of wavesNARENDRA MALHOTRA
 

La actualidad más candente (20)

Anterpartum fetal surveillance
Anterpartum fetal surveillance   Anterpartum fetal surveillance
Anterpartum fetal surveillance
 
Doppler interpretation in pregnancy
Doppler interpretation in pregnancyDoppler interpretation in pregnancy
Doppler interpretation in pregnancy
 
Doppler in pregnancy
Doppler in pregnancyDoppler in pregnancy
Doppler in pregnancy
 
Doppler in gyneacology Dr. Muhammad Bin Zulfiqar
Doppler in gyneacology Dr. Muhammad Bin ZulfiqarDoppler in gyneacology Dr. Muhammad Bin Zulfiqar
Doppler in gyneacology Dr. Muhammad Bin Zulfiqar
 
Role of ultrasound in emergency obstetrics dr.shreedhar
Role of ultrasound in emergency obstetrics dr.shreedharRole of ultrasound in emergency obstetrics dr.shreedhar
Role of ultrasound in emergency obstetrics dr.shreedhar
 
First trimester scan
First trimester scanFirst trimester scan
First trimester scan
 
Routine fetal anatomy scan at 18-23 weeks
Routine fetal anatomy scan at 18-23 weeksRoutine fetal anatomy scan at 18-23 weeks
Routine fetal anatomy scan at 18-23 weeks
 
Isuog practice guidelines performance of first trimester fetal ultrasound scan
Isuog practice guidelines performance of first trimester fetal ultrasound scanIsuog practice guidelines performance of first trimester fetal ultrasound scan
Isuog practice guidelines performance of first trimester fetal ultrasound scan
 
SCAR ECTOPIC
SCAR ECTOPICSCAR ECTOPIC
SCAR ECTOPIC
 
Cesarean Scar Ectopic Pregnancy Current Management Strategies
Cesarean Scar Ectopic Pregnancy Current Management StrategiesCesarean Scar Ectopic Pregnancy Current Management Strategies
Cesarean Scar Ectopic Pregnancy Current Management Strategies
 
Early pregnancy ultrasonographic evaluation
Early pregnancy ultrasonographic evaluationEarly pregnancy ultrasonographic evaluation
Early pregnancy ultrasonographic evaluation
 
Nuchal translucency
Nuchal translucencyNuchal translucency
Nuchal translucency
 
The role of uterine artery embolization in gynecology practice
The role of uterine artery embolization in gynecology practiceThe role of uterine artery embolization in gynecology practice
The role of uterine artery embolization in gynecology practice
 
Ductus venosus
Ductus venosusDuctus venosus
Ductus venosus
 
Follicular monitoring
Follicular monitoring Follicular monitoring
Follicular monitoring
 
C O L O R D O P P L E R I N F E T A L H Y P O X I A
C O L O R  D O P P L E R  I N  F E T A L  H Y P O X I AC O L O R  D O P P L E R  I N  F E T A L  H Y P O X I A
C O L O R D O P P L E R I N F E T A L H Y P O X I A
 
Role of ultrasound in the Infertility management
Role of ultrasound in the Infertility  management Role of ultrasound in the Infertility  management
Role of ultrasound in the Infertility management
 
Doppler us in the evaluation of fetal growth
Doppler us in the evaluation of fetal growthDoppler us in the evaluation of fetal growth
Doppler us in the evaluation of fetal growth
 
Role of tubal surgery in era of ivf
Role of tubal surgery in era of ivfRole of tubal surgery in era of ivf
Role of tubal surgery in era of ivf
 
Color doppler in FGR making sence of waves
Color doppler in FGR making sence of wavesColor doppler in FGR making sence of waves
Color doppler in FGR making sence of waves
 

Similar a ISUOG Practice Guidelines: Use of Doppler ultrasonography in obstetrics

Guía ISUOG de Doppler obstétrico
Guía ISUOG de Doppler obstétricoGuía ISUOG de Doppler obstétrico
Guía ISUOG de Doppler obstétricoTony Terrones
 
doppler gynecology pdf doppler en gynécologie
doppler gynecology pdf doppler en gynécologiedoppler gynecology pdf doppler en gynécologie
doppler gynecology pdf doppler en gynécologieRimaBoualleg
 
Its is useful for studying the basic concepts of ultrasound for medical imagi...
Its is useful for studying the basic concepts of ultrasound for medical imagi...Its is useful for studying the basic concepts of ultrasound for medical imagi...
Its is useful for studying the basic concepts of ultrasound for medical imagi...DilshanDillu1
 
Respiration motion management
Respiration motion managementRespiration motion management
Respiration motion managementKiran Ramakrishna
 
pediatric echo measurements technique z score
pediatric echo measurements technique z scorepediatric echo measurements technique z score
pediatric echo measurements technique z scoreshivendra23
 
Ultrasound instrumentation practical applications
Ultrasound instrumentation practical applicationsUltrasound instrumentation practical applications
Ultrasound instrumentation practical applicationsshariq ahmad shah
 
Basics of Diagnostic Ultrasound
Basics of Diagnostic UltrasoundBasics of Diagnostic Ultrasound
Basics of Diagnostic UltrasoundIbrahim Tayyan
 
Transesophageal echocardiography by Dhaval patel
Transesophageal echocardiography by Dhaval patelTransesophageal echocardiography by Dhaval patel
Transesophageal echocardiography by Dhaval patelDhaval Patel
 
Respiration motion management
Respiration motion managementRespiration motion management
Respiration motion managementKiran Ramakrishna
 
Evaluation and imaging for lung SBRT
Evaluation and imaging for lung SBRTEvaluation and imaging for lung SBRT
Evaluation and imaging for lung SBRTNaveen Mummudi
 
Normal doppler spectral pattern of abdominal and limb vessels final
Normal doppler spectral pattern of abdominal and limb vessels finalNormal doppler spectral pattern of abdominal and limb vessels final
Normal doppler spectral pattern of abdominal and limb vessels finalNipun Gupta
 

Similar a ISUOG Practice Guidelines: Use of Doppler ultrasonography in obstetrics (20)

Guía ISUOG de Doppler obstétrico
Guía ISUOG de Doppler obstétricoGuía ISUOG de Doppler obstétrico
Guía ISUOG de Doppler obstétrico
 
doppler gynecology pdf doppler en gynécologie
doppler gynecology pdf doppler en gynécologiedoppler gynecology pdf doppler en gynécologie
doppler gynecology pdf doppler en gynécologie
 
Doppler physics.pptx
Doppler physics.pptxDoppler physics.pptx
Doppler physics.pptx
 
Its is useful for studying the basic concepts of ultrasound for medical imagi...
Its is useful for studying the basic concepts of ultrasound for medical imagi...Its is useful for studying the basic concepts of ultrasound for medical imagi...
Its is useful for studying the basic concepts of ultrasound for medical imagi...
 
Respiration motion management
Respiration motion managementRespiration motion management
Respiration motion management
 
Obstetric doppler
Obstetric dopplerObstetric doppler
Obstetric doppler
 
pediatric echo measurements technique z score
pediatric echo measurements technique z scorepediatric echo measurements technique z score
pediatric echo measurements technique z score
 
Ultrasound instrumentation practical applications
Ultrasound instrumentation practical applicationsUltrasound instrumentation practical applications
Ultrasound instrumentation practical applications
 
Basics of Diagnostic Ultrasound
Basics of Diagnostic UltrasoundBasics of Diagnostic Ultrasound
Basics of Diagnostic Ultrasound
 
Transesophageal echocardiography by Dhaval patel
Transesophageal echocardiography by Dhaval patelTransesophageal echocardiography by Dhaval patel
Transesophageal echocardiography by Dhaval patel
 
Respiration motion management
Respiration motion managementRespiration motion management
Respiration motion management
 
Ct thorax
Ct thoraxCt thorax
Ct thorax
 
Evaluation and imaging for lung SBRT
Evaluation and imaging for lung SBRTEvaluation and imaging for lung SBRT
Evaluation and imaging for lung SBRT
 
Why doppler ?
Why doppler ?Why doppler ?
Why doppler ?
 
Doppler Physics (3)
Doppler Physics (3)Doppler Physics (3)
Doppler Physics (3)
 
Normal doppler spectral pattern of abdominal and limb vessels final
Normal doppler spectral pattern of abdominal and limb vessels finalNormal doppler spectral pattern of abdominal and limb vessels final
Normal doppler spectral pattern of abdominal and limb vessels final
 
Doppler echocardiography
Doppler echocardiographyDoppler echocardiography
Doppler echocardiography
 
Doppler echocardiography
Doppler echocardiographyDoppler echocardiography
Doppler echocardiography
 
Motion Management in Radiation Therapy
Motion Management in Radiation TherapyMotion Management in Radiation Therapy
Motion Management in Radiation Therapy
 
Dop phys (1)
Dop phys (1)Dop phys (1)
Dop phys (1)
 

Último

Call Girls Bangalore - 450+ Call Girl Cash Payment 💯Call Us 🔝 6378878445 🔝 💃 ...
Call Girls Bangalore - 450+ Call Girl Cash Payment 💯Call Us 🔝 6378878445 🔝 💃 ...Call Girls Bangalore - 450+ Call Girl Cash Payment 💯Call Us 🔝 6378878445 🔝 💃 ...
Call Girls Bangalore - 450+ Call Girl Cash Payment 💯Call Us 🔝 6378878445 🔝 💃 ...gragneelam30
 
Call Girls in Lucknow Just Call 👉👉 8875999948 Top Class Call Girl Service Ava...
Call Girls in Lucknow Just Call 👉👉 8875999948 Top Class Call Girl Service Ava...Call Girls in Lucknow Just Call 👉👉 8875999948 Top Class Call Girl Service Ava...
Call Girls in Lucknow Just Call 👉👉 8875999948 Top Class Call Girl Service Ava...Janvi Singh
 
7 steps How to prevent Thalassemia : Dr Sharda Jain & Vandana Gupta
7 steps How to prevent Thalassemia : Dr Sharda Jain & Vandana Gupta7 steps How to prevent Thalassemia : Dr Sharda Jain & Vandana Gupta
7 steps How to prevent Thalassemia : Dr Sharda Jain & Vandana GuptaLifecare Centre
 
Call Girls Mussoorie Just Call 8854095900 Top Class Call Girl Service Available
Call Girls Mussoorie Just Call 8854095900 Top Class Call Girl Service AvailableCall Girls Mussoorie Just Call 8854095900 Top Class Call Girl Service Available
Call Girls Mussoorie Just Call 8854095900 Top Class Call Girl Service AvailableJanvi Singh
 
Kolkata Call Girls Naktala 💯Call Us 🔝 8005736733 🔝 💃 Top Class Call Girl Se...
Kolkata Call Girls Naktala  💯Call Us 🔝 8005736733 🔝 💃  Top Class Call Girl Se...Kolkata Call Girls Naktala  💯Call Us 🔝 8005736733 🔝 💃  Top Class Call Girl Se...
Kolkata Call Girls Naktala 💯Call Us 🔝 8005736733 🔝 💃 Top Class Call Girl Se...Namrata Singh
 
Call Girls Rishikesh Just Call 9667172968 Top Class Call Girl Service Available
Call Girls Rishikesh Just Call 9667172968 Top Class Call Girl Service AvailableCall Girls Rishikesh Just Call 9667172968 Top Class Call Girl Service Available
Call Girls Rishikesh Just Call 9667172968 Top Class Call Girl Service Availableperfect solution
 
❤️Chandigarh Escorts Service☎️9814379184☎️ Call Girl service in Chandigarh☎️ ...
❤️Chandigarh Escorts Service☎️9814379184☎️ Call Girl service in Chandigarh☎️ ...❤️Chandigarh Escorts Service☎️9814379184☎️ Call Girl service in Chandigarh☎️ ...
❤️Chandigarh Escorts Service☎️9814379184☎️ Call Girl service in Chandigarh☎️ ...Sheetaleventcompany
 
(RIYA)🎄Airhostess Call Girl Jaipur Call Now 8445551418 Premium Collection Of ...
(RIYA)🎄Airhostess Call Girl Jaipur Call Now 8445551418 Premium Collection Of ...(RIYA)🎄Airhostess Call Girl Jaipur Call Now 8445551418 Premium Collection Of ...
(RIYA)🎄Airhostess Call Girl Jaipur Call Now 8445551418 Premium Collection Of ...TanyaAhuja34
 
Circulatory Shock, types and stages, compensatory mechanisms
Circulatory Shock, types and stages, compensatory mechanismsCirculatory Shock, types and stages, compensatory mechanisms
Circulatory Shock, types and stages, compensatory mechanismsMedicoseAcademics
 
Race Course Road } Book Call Girls in Bangalore | Whatsapp No 6378878445 VIP ...
Race Course Road } Book Call Girls in Bangalore | Whatsapp No 6378878445 VIP ...Race Course Road } Book Call Girls in Bangalore | Whatsapp No 6378878445 VIP ...
Race Course Road } Book Call Girls in Bangalore | Whatsapp No 6378878445 VIP ...dishamehta3332
 
Chandigarh Call Girls Service ❤️🍑 9809698092 👄🫦Independent Escort Service Cha...
Chandigarh Call Girls Service ❤️🍑 9809698092 👄🫦Independent Escort Service Cha...Chandigarh Call Girls Service ❤️🍑 9809698092 👄🫦Independent Escort Service Cha...
Chandigarh Call Girls Service ❤️🍑 9809698092 👄🫦Independent Escort Service Cha...Sheetaleventcompany
 
👉 Chennai Sexy Aunty’s WhatsApp Number 👉📞 7427069034 👉📞 Just📲 Call Ruhi Colle...
👉 Chennai Sexy Aunty’s WhatsApp Number 👉📞 7427069034 👉📞 Just📲 Call Ruhi Colle...👉 Chennai Sexy Aunty’s WhatsApp Number 👉📞 7427069034 👉📞 Just📲 Call Ruhi Colle...
👉 Chennai Sexy Aunty’s WhatsApp Number 👉📞 7427069034 👉📞 Just📲 Call Ruhi Colle...rajnisinghkjn
 
Call Girl In Indore 📞9235973566📞 Just📲 Call Inaaya Indore Call Girls Service ...
Call Girl In Indore 📞9235973566📞 Just📲 Call Inaaya Indore Call Girls Service ...Call Girl In Indore 📞9235973566📞 Just📲 Call Inaaya Indore Call Girls Service ...
Call Girl In Indore 📞9235973566📞 Just📲 Call Inaaya Indore Call Girls Service ...Sheetaleventcompany
 
Call girls Service Phullen / 9332606886 Genuine Call girls with real Photos a...
Call girls Service Phullen / 9332606886 Genuine Call girls with real Photos a...Call girls Service Phullen / 9332606886 Genuine Call girls with real Photos a...
Call girls Service Phullen / 9332606886 Genuine Call girls with real Photos a...call girls hydrabad
 
💰Call Girl In Bangalore☎️63788-78445💰 Call Girl service in Bangalore☎️Bangalo...
💰Call Girl In Bangalore☎️63788-78445💰 Call Girl service in Bangalore☎️Bangalo...💰Call Girl In Bangalore☎️63788-78445💰 Call Girl service in Bangalore☎️Bangalo...
💰Call Girl In Bangalore☎️63788-78445💰 Call Girl service in Bangalore☎️Bangalo...gragneelam30
 
Call Girls Shahdol Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Shahdol Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Shahdol Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Shahdol Just Call 8250077686 Top Class Call Girl Service AvailableDipal Arora
 
Call Girl in Chennai | Whatsapp No 📞 7427069034 📞 VIP Escorts Service Availab...
Call Girl in Chennai | Whatsapp No 📞 7427069034 📞 VIP Escorts Service Availab...Call Girl in Chennai | Whatsapp No 📞 7427069034 📞 VIP Escorts Service Availab...
Call Girl in Chennai | Whatsapp No 📞 7427069034 📞 VIP Escorts Service Availab...amritaverma53
 
Chandigarh Call Girls Service ❤️🍑 9809698092 👄🫦Independent Escort Service Cha...
Chandigarh Call Girls Service ❤️🍑 9809698092 👄🫦Independent Escort Service Cha...Chandigarh Call Girls Service ❤️🍑 9809698092 👄🫦Independent Escort Service Cha...
Chandigarh Call Girls Service ❤️🍑 9809698092 👄🫦Independent Escort Service Cha...Sheetaleventcompany
 
Exclusive Call Girls Bangalore {7304373326} ❤️VVIP POOJA Call Girls in Bangal...
Exclusive Call Girls Bangalore {7304373326} ❤️VVIP POOJA Call Girls in Bangal...Exclusive Call Girls Bangalore {7304373326} ❤️VVIP POOJA Call Girls in Bangal...
Exclusive Call Girls Bangalore {7304373326} ❤️VVIP POOJA Call Girls in Bangal...Sheetaleventcompany
 
Cardiac Output, Venous Return, and Their Regulation
Cardiac Output, Venous Return, and Their RegulationCardiac Output, Venous Return, and Their Regulation
Cardiac Output, Venous Return, and Their RegulationMedicoseAcademics
 

Último (20)

Call Girls Bangalore - 450+ Call Girl Cash Payment 💯Call Us 🔝 6378878445 🔝 💃 ...
Call Girls Bangalore - 450+ Call Girl Cash Payment 💯Call Us 🔝 6378878445 🔝 💃 ...Call Girls Bangalore - 450+ Call Girl Cash Payment 💯Call Us 🔝 6378878445 🔝 💃 ...
Call Girls Bangalore - 450+ Call Girl Cash Payment 💯Call Us 🔝 6378878445 🔝 💃 ...
 
Call Girls in Lucknow Just Call 👉👉 8875999948 Top Class Call Girl Service Ava...
Call Girls in Lucknow Just Call 👉👉 8875999948 Top Class Call Girl Service Ava...Call Girls in Lucknow Just Call 👉👉 8875999948 Top Class Call Girl Service Ava...
Call Girls in Lucknow Just Call 👉👉 8875999948 Top Class Call Girl Service Ava...
 
7 steps How to prevent Thalassemia : Dr Sharda Jain & Vandana Gupta
7 steps How to prevent Thalassemia : Dr Sharda Jain & Vandana Gupta7 steps How to prevent Thalassemia : Dr Sharda Jain & Vandana Gupta
7 steps How to prevent Thalassemia : Dr Sharda Jain & Vandana Gupta
 
Call Girls Mussoorie Just Call 8854095900 Top Class Call Girl Service Available
Call Girls Mussoorie Just Call 8854095900 Top Class Call Girl Service AvailableCall Girls Mussoorie Just Call 8854095900 Top Class Call Girl Service Available
Call Girls Mussoorie Just Call 8854095900 Top Class Call Girl Service Available
 
Kolkata Call Girls Naktala 💯Call Us 🔝 8005736733 🔝 💃 Top Class Call Girl Se...
Kolkata Call Girls Naktala  💯Call Us 🔝 8005736733 🔝 💃  Top Class Call Girl Se...Kolkata Call Girls Naktala  💯Call Us 🔝 8005736733 🔝 💃  Top Class Call Girl Se...
Kolkata Call Girls Naktala 💯Call Us 🔝 8005736733 🔝 💃 Top Class Call Girl Se...
 
Call Girls Rishikesh Just Call 9667172968 Top Class Call Girl Service Available
Call Girls Rishikesh Just Call 9667172968 Top Class Call Girl Service AvailableCall Girls Rishikesh Just Call 9667172968 Top Class Call Girl Service Available
Call Girls Rishikesh Just Call 9667172968 Top Class Call Girl Service Available
 
❤️Chandigarh Escorts Service☎️9814379184☎️ Call Girl service in Chandigarh☎️ ...
❤️Chandigarh Escorts Service☎️9814379184☎️ Call Girl service in Chandigarh☎️ ...❤️Chandigarh Escorts Service☎️9814379184☎️ Call Girl service in Chandigarh☎️ ...
❤️Chandigarh Escorts Service☎️9814379184☎️ Call Girl service in Chandigarh☎️ ...
 
(RIYA)🎄Airhostess Call Girl Jaipur Call Now 8445551418 Premium Collection Of ...
(RIYA)🎄Airhostess Call Girl Jaipur Call Now 8445551418 Premium Collection Of ...(RIYA)🎄Airhostess Call Girl Jaipur Call Now 8445551418 Premium Collection Of ...
(RIYA)🎄Airhostess Call Girl Jaipur Call Now 8445551418 Premium Collection Of ...
 
Circulatory Shock, types and stages, compensatory mechanisms
Circulatory Shock, types and stages, compensatory mechanismsCirculatory Shock, types and stages, compensatory mechanisms
Circulatory Shock, types and stages, compensatory mechanisms
 
Race Course Road } Book Call Girls in Bangalore | Whatsapp No 6378878445 VIP ...
Race Course Road } Book Call Girls in Bangalore | Whatsapp No 6378878445 VIP ...Race Course Road } Book Call Girls in Bangalore | Whatsapp No 6378878445 VIP ...
Race Course Road } Book Call Girls in Bangalore | Whatsapp No 6378878445 VIP ...
 
Chandigarh Call Girls Service ❤️🍑 9809698092 👄🫦Independent Escort Service Cha...
Chandigarh Call Girls Service ❤️🍑 9809698092 👄🫦Independent Escort Service Cha...Chandigarh Call Girls Service ❤️🍑 9809698092 👄🫦Independent Escort Service Cha...
Chandigarh Call Girls Service ❤️🍑 9809698092 👄🫦Independent Escort Service Cha...
 
👉 Chennai Sexy Aunty’s WhatsApp Number 👉📞 7427069034 👉📞 Just📲 Call Ruhi Colle...
👉 Chennai Sexy Aunty’s WhatsApp Number 👉📞 7427069034 👉📞 Just📲 Call Ruhi Colle...👉 Chennai Sexy Aunty’s WhatsApp Number 👉📞 7427069034 👉📞 Just📲 Call Ruhi Colle...
👉 Chennai Sexy Aunty’s WhatsApp Number 👉📞 7427069034 👉📞 Just📲 Call Ruhi Colle...
 
Call Girl In Indore 📞9235973566📞 Just📲 Call Inaaya Indore Call Girls Service ...
Call Girl In Indore 📞9235973566📞 Just📲 Call Inaaya Indore Call Girls Service ...Call Girl In Indore 📞9235973566📞 Just📲 Call Inaaya Indore Call Girls Service ...
Call Girl In Indore 📞9235973566📞 Just📲 Call Inaaya Indore Call Girls Service ...
 
Call girls Service Phullen / 9332606886 Genuine Call girls with real Photos a...
Call girls Service Phullen / 9332606886 Genuine Call girls with real Photos a...Call girls Service Phullen / 9332606886 Genuine Call girls with real Photos a...
Call girls Service Phullen / 9332606886 Genuine Call girls with real Photos a...
 
💰Call Girl In Bangalore☎️63788-78445💰 Call Girl service in Bangalore☎️Bangalo...
💰Call Girl In Bangalore☎️63788-78445💰 Call Girl service in Bangalore☎️Bangalo...💰Call Girl In Bangalore☎️63788-78445💰 Call Girl service in Bangalore☎️Bangalo...
💰Call Girl In Bangalore☎️63788-78445💰 Call Girl service in Bangalore☎️Bangalo...
 
Call Girls Shahdol Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Shahdol Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Shahdol Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Shahdol Just Call 8250077686 Top Class Call Girl Service Available
 
Call Girl in Chennai | Whatsapp No 📞 7427069034 📞 VIP Escorts Service Availab...
Call Girl in Chennai | Whatsapp No 📞 7427069034 📞 VIP Escorts Service Availab...Call Girl in Chennai | Whatsapp No 📞 7427069034 📞 VIP Escorts Service Availab...
Call Girl in Chennai | Whatsapp No 📞 7427069034 📞 VIP Escorts Service Availab...
 
Chandigarh Call Girls Service ❤️🍑 9809698092 👄🫦Independent Escort Service Cha...
Chandigarh Call Girls Service ❤️🍑 9809698092 👄🫦Independent Escort Service Cha...Chandigarh Call Girls Service ❤️🍑 9809698092 👄🫦Independent Escort Service Cha...
Chandigarh Call Girls Service ❤️🍑 9809698092 👄🫦Independent Escort Service Cha...
 
Exclusive Call Girls Bangalore {7304373326} ❤️VVIP POOJA Call Girls in Bangal...
Exclusive Call Girls Bangalore {7304373326} ❤️VVIP POOJA Call Girls in Bangal...Exclusive Call Girls Bangalore {7304373326} ❤️VVIP POOJA Call Girls in Bangal...
Exclusive Call Girls Bangalore {7304373326} ❤️VVIP POOJA Call Girls in Bangal...
 
Cardiac Output, Venous Return, and Their Regulation
Cardiac Output, Venous Return, and Their RegulationCardiac Output, Venous Return, and Their Regulation
Cardiac Output, Venous Return, and Their Regulation
 

ISUOG Practice Guidelines: Use of Doppler ultrasonography in obstetrics

  • 1. Bhide A, Acharya G, Bilardo CM, Brezinka C, Cafici D, Hernandez- Andrade E, Kalache K, Kingdom J, Kiserud T, Lee W, Lees C, Leung KY, Malinger G, Mari G, Prefumo F, Sepulveda W and Trudinger B. on behalf of the ISUOG Clinical Standards Committee
  • 2. Why ? •The International Society of Ultrasound in Obstetrics and Gynecology (ISUOG) is a scientific organization that encourages sound clinical practice, teaching and research for diagnostic imging in women’s healthcare. •Practice Guidelines and Consensus are intended to reflect what is considered by ISUOG to be the best practices at the time at which they were issued. •Although ISUOG has made every effort to ensure that guidelines are accurate when issued, neither the Society nor any of its employees or members accepts any liability for the consequences of any inaccurate or misleading data, opinions or statements issued by the CSC. •Guidelines are not intended to establish a legal standard of care because interpretation of the evidence that underpins the guidelines may be influenced by individual circumstances and available resources. •Approved guidelines can be distributed freely with the permission of ISUOG (info@isuog.org).
  • 3. Scope • Summarizes practice guidelines regarding how to perform Doppler ultrasonography of the feto-placental circulation • Not intended to - define clinical indications, - specify proper timing of the Doppler examination in pregnancy - how to interpret findings
  • 4. Guidance on safety • It is of utmost importance not to expose the embryo and fetus to unduly harmful ultrasound energy, particularly in the earliest stages of pregnancy. • At these stages, Doppler recording, when clinically indicated, should be performed at the lowest possible energy levels. • When performing Doppler imaging, the displayed thermal index (TI) should be ≤ 1.0 and exposure time should be kept as short as possible, usually no longer than 5–10 min and not exceeding 60 min
  • 5. Pulsed wave Doppler: How can the accuracy of Doppler measurements be optimized? • Recordings should be obtained during absence of fetal breathing and body movements, and if necessary during temporary maternal breath hold • Colour flow mapping is not mandatory, although it is very helpful in the identification of the vessel of interest and in defining the direction of blood flow • The optimal insonation is complete alignment with the blood flow • It is advisable to start with relatively wide Doppler gate (sample volume)
  • 6. Pulsed wave Doppler: How can the accuracy of Doppler measurements be optimized? • Similar to gray-scale imaging, the penetration and resolution of the Doppler beam can be optimized by adjusting the frequency (MHz) of the Doppler probe • By convention, it should be set as low as possible (≤ 50–60 Hz) in order to eliminate the low- frequency noise from peripheral blood vessels • Doppler horizontal sweep speed should be fast enough to separate successive waveforms • PRF adjusted so that the waveform fills 75% of Doppler screen • Doppler measurements should be reproducible • Gain should be adjusted in order to see clearly the Doppler velocity waveform, without the presence of artifacts in the background of the display • It is advisable not to invert the Doppler display on the ultrasound screen
  • 7. Colour directional Doppler: How can the accuracy of Doppler measurements be optimized? • The use of colour Doppler increases the total power emitted. Colour Doppler resolution increases when the colour box is reduced in size • The box should be kept as small as possible to include only the studied area • The velocity scale or PRF should be adjusted to represent the real colour velocity of the studied vessel • Gain should be adjusted in order to prevent noise and artifacts • The angle of insonation should be adjusted by optimizing the position of the ultrasound probe according to the vessel or area studied
  • 8. Power Doppler and directional power Doppler: How can the accuracy of Doppler measurements be optimized? • The angle of insonation has less effect on power Doppler signals; nevertheless, the same optimization process as for colour directional Doppler must be performed • There is no aliasing phenomenon using power Doppler; however, an inappropriately low PRF may lead to noises and artifacts • Gain should be reduced in order to prevent amplification of noise
  • 9. What equipment is needed for Doppler evaluation of the feto-placental circulation? • Equipment should have colour flow and spectral wave Doppler capabilities with on-screen display of flow velocity scales or pulse repetition frequency (PRF) and Doppler ultrasound frequency (in MHz) • Mechanical index (MI) and Thermal Index (TI) should be displayed on the ultrasound screen • Ultrasound system should generate a maximum velocity envelope (MVE) showing the whole spectral Doppler waveform • MVE should be possible to delineate using automatic or manual waveform traces • System software must be able to estimate peak systolic velocity (PSV), end- diastolic velocity (EDV) and time-averaged maximum velocity from the MVE and to calculate the commonly used Doppler indices i.e. pulsatility (PI) and resistance (RI) indices and systolic/diastolic velocity (S/D) ratio. On the tracing, the various points included in the calculations should be indicated, to ensure correctly calculated indices
  • 10. What is the appropriate technique for obtaining uterine artery Doppler waveforms? • Trans-abdominally, the probe is placed longitudinally in the lower lateral quadrant of the abdomen angled medially • Colour flow mapping is useful to identify the uterine artery as it appears to cross the external iliac artery • Sample volume is placed 1 cm downstream to the crossover point • If the uterine artery branches before the intersection of the external iliac artery, the sample volume should be placed on the main artery just before the bifurcation
  • 11. Normal waveform from the uterine artery
  • 12. Abnormal waveform from the uterine artery. Note notch (arrow) implying increased resistance in the uterine artery
  • 13. Uterine Artery Doppler: Technique (Transvaginal) • Women should be asked to empty their bladder and placed in the dorsal lithotomy position • The probe should be placed into the lateral fornix and the uterine artery identified using colour Doppler at the level of the internal cervical os • The same should then be repeated for the contra-lateral uterine artery • Reference ranges for uterine artery Doppler indices depend on the technique of measurement, so appropriate reference ranges should be used for transabdominal and transvaginal routes
  • 14. What is the appropriate technique for obtaining umbilical artery Doppler waveforms? • There is a significant difference in Doppler indices measured at the fetal end, the free loop and the placental end of the umbilical cord • For the sake of simplicity and consistency, measurements should be made in a free cord loop • In multiple pregnancies, and/or when comparing repeated measurements longitudinally, recordings from fixed sites (fetal end, placental end or intra-abdominal portion) may be more reliable • Appropriate reference ranges should be used according to the site of interrogation
  • 15. Inappropriate technique: Too low temporal an spatial resolution of the waveforms
  • 16. Appropriate technique: note better quality spectral Doppler tracing
  • 17. What is the appropriate technique for obtaining fetal middle cerebral artery Doppler waveforms? • An axial section of the brain, including the thalami and the sphenoid bone wings is obtained, and magnified • Colour flow mapping is used to identify the circle of Willis and the proximal middle cerebral artery (MCA) • The pulsed-wave Doppler gate is then placed at the proximal one-third of the MCA close to its origin in the internal carotid artery • The angle between the ultrasound beam and the direction of blood flow is kept as close as possible to 0 °, and should be less than 20 °
  • 18. What is the appropriate technique for obtaining fetal middle cerebral artery Doppler waveforms? • Care should be taken to avoid any unnecessary pressure on the fetal head • At least three and fewer than 10 consecutive waveforms should be recorded. The highest point of the waveform is considered as the PSV (cm/s) • The PSV can be measured using manual calipers or auto-trace. The latter yields significantly lower medians than does the former, but more closely approximates published medians used in clinical practice • PI is usually calculated using auto-trace measurement, but manual tracing is also acceptable
  • 19. Acceptable middle cerebral artery Doppler shift waveform. Note insonation
  • 20. What is the appropriate technique for obtaining fetal venous Doppler waveforms? • The vessel is identified by visualizing this connection by 2D imaging either in a mid-sagittal longitudinal plane of the fetal trunk or in an oblique transverse plane through the upper abdomen • Colour flow mapping demonstrating the high velocity at the narrow entrance of the DV confirms its identification and indicates the standard sampling site for Doppler measurements • In early pregnancy and in compromised pregnancies particular care has to be taken to reduce the sample volume appropriately in order to ensure clean recording of the lowest velocity during atrial contraction
  • 21. What is the appropriate technique for obtaining fetal venous Doppler waveforms? • The waveform is usually triphasic, but biphasic and non-pulsating recordings, though rarer, may be seen in healthy fetuses • The velocities are relatively high, between 55 and 90 cm/s for most of the second half of pregnancy, but lower in early pregnancy
  • 22. Ductus venosus Doppler recording with sagittal insonation aligning with the isthmic portion without angle correction. Low-velocity vessel wall filter (arrow) does not interfere with a-wave (a), which is far from zero line. High sweep speed allows detailed visualization of variation in velocity
  • 23. Which indices to use? • S/D ratio, Resistance index (RI) and Pulsatility Index (PI) are the three well-known indices to describe arterial flow velocity waveforms • All three are highly correlated • There is no published evidence showing superiority of any one index over the others, but PI is the most commonly used index in current practice • Pulsatility index for veins (PIV) is most commonly used for venous waveforms in the current literature
  • 24. ISUOG’s website is a great resource for more information: