MICROPRECISION CALIBRATION INC.
                                                                                                                  BLOCK 2 LOT 4 CALAMBA PREMIERE INTERNATIONAL PARK
                                                                                                                              BATINO, CALAMBA CITY, LAGUNA, PHILS 4027
                                                                                                                                                         +63495023015

                                                                            Certificate of Calibration
Date: 5/16/2008                            Lab # 935.05                                                                                                                                Certificate #: 315634


    Customer:
    SAMPLE
    85/18 BUMDUREN KRASANG                                                                                             Purchase Order:                        12345
    BURIRAM, , 31160                                                                                                   Work Order:                            N/A

   MPC Control #: US37290511                                                                                          Serial Number:                          US37290511
   Asset ID:      8957                                                                                                Department:                             N/A
   Gage Type:     POWER SENSOR                                                                                        Performed By:                           DANNY BOY B. BUTIAL
   Manufacturer:  HEWLETT PACKARD                                                                                     Received Condition:                    IN TOLERANCE
   Model Number: 8481A                                                                                                Returned Condition:                    IN TOLERANCE
   Size:          10MHZ~18GHZ                                                                                         Cal Date:                              October 31, 2007
   Temp./RH:      23 ° C / 49 %                                                                                       Cal. Interval:                         12 MONTHS
                                                                                                                      Cal. Due Date:                         October 31, 2008


   *Calibration Notes:
   Site loc: MPC-Lab.
   See attached data report.



   Standards Used To Calibrate Equipment
      I.D.                              Description                                        Model                      Serial                         Manufacturer                       Cal. Due Date Traceability #
   AD3312                          SIGNAL GENERATOR                                       83640B                  3844A00758                     HEWLETT PACKARD                           1/16/2009                 227210
    AI2393                       DIRECTIONAL COUPLER                                      11692D                  1212A02566                     HEWLETT PACKARD                           8/29/2008                 139290
   AN1816                              POWER METER                                        E4419B                  GB40208110                              AGILENT                           3/5/2009                 260329
   AN1869                            POWER SPLITTER                                      11667A                        04040                     HEWLETT PACKARD                            5/9/2009                 310631
   AU6515                             POWER SENSOR                                        8481A                   1726A03110                              AGILENT                          1/25/2009           1-1201452485-1
    R1640                        DIRECTIONAL COUPLER                                       778D                        15050                     HEWLETT PACKARD                           8/20/2008                 133363



   Procedures Used In This Event:
              Procedure Name                                                             Description
                848X SERIES, HP                                             POWER SENSOR, COAXIAL (HP)




Calibrating Technician:                                                                                                                       QC Approval:

                                         DANNY BOY B. BUTIAL                                                                                                                Jarrod Trammell
Uncertainty Estimated at 2 to 1
Uncertainties have been estimated at a 95 percent confidence level (k=2). All calibrations performed conform to ISO 17025:2005, ISO 9001:2000, ANSI Z540-1, MPC Quality Manual, Calibration System Rev5, and
customer requirements

All standards used are either traceable to the National Institute of Standards Technology or have intrinsic accuracy. All services performed have used proper manufacturer and industrial service techniques. This
report may not be reproduced in part without written permission of Micro Precision's Quality Assurance Manager.


                                                                                                              Page 1 of 1                                                                                              (CERT, Rev 0)
Hewlett Packard 8481A Power Sensor


   Procedure Reference: 848X SERIES, HP                      Work Order:          n/a



             ID Number:   US37290511                      Serial Number:     US37290511



         Asset Number: VM#8957 / PW#003850               Calibration Date:   Oct. 31, 2007



           Temperature:      23.0°C                             Humidity:       49.0%




Section 1 - RHO and Calibration Factors
   Reference Cal Factor: 100.0 %

    Frequency ( Ghz )      Cal Factor     CF Uncertainty Measured RHO RHO Upper Limit
          .01               100.2 %           1.3 %           .048        < .166
          .03               100.1 %           1.3 %           .048        < .166
           .1               100.1 %           1.3 %           .023        < .048
           .3               100.2 %           1.3 %           .020        < .048
            1               98.8 %            1.3 %           .020        < .048
            2               98.8 %            1.4 %           .020        < .083
            3               97.7 %            1.6 %           .021        < .083
            4               96.7 %            1.6 %           .021        < .083
            5               97.2 %            1.7 %           .018        < .083
            6               96.3 %            2.0 %           .020        < .083
            7               97.8 %            2.0 %           .029        < .083
            8               95.1 %            2.1 %           .030        < .083
            9               95.2 %            2.3 %           .030        < .083
           10               92.9 %            2.5 %           .042        < .083
           11               96.0 %            2.6 %           .043        < .083
           12               93.1 %            2.5 %           .027        < .083
          12.4              93.0 %            2.5 %           .022        < .083
           13               94.3 %            2.6 %           .016        < .123
           14               91.6 %            2.7 %           .016        < .123
           15               92.2 %            2.7 %           .030        < .123
           16               93.7 %            2.6 %           .058        < .123
           17               91.9 %            2.8 %           .060        < .123
           18               89.5 %            2.9 %           .054        < .123




                                        Page 1 of 2
ID Number: US37290511




                                                                   Calibration Factor Chart
               100.0 %


                98.0 %


                96.0 %


                94.0 %
Cal Factor %




                92.0 %


                90.0 %


                88.0 %


                86.0 %


                84.0 %
                         .01 .03 .1 .3       1       2   3    4     5    6   7   8   9 10 11 12 12.4 13 14 15 16 17 18
                                                                        Frequency ( Ghz )




                                                                             RHO Chart
               .200

               .180

               .160

               .140

               .120
 RHO




               .100
                                                                                                                         Measured RHO
               .080                                                                                                      Upper Limit

               .060

               .040

               .020

               .000
                      .01 .03 .1 .3 1    2       3   4   5     6    7   8    9 10 11 1212.413 14 15 16 17 18
                                                             Frequency ( Ghz )




                                                                              Page 2 of 2

Más contenido relacionado

La actualidad más candente

ISO/IEC 17025
ISO/IEC 17025 ISO/IEC 17025
ISO/IEC 17025
Akma Ija
 

La actualidad más candente (20)

ISO 17025 Accreditation Detail Review
ISO 17025 Accreditation Detail Review ISO 17025 Accreditation Detail Review
ISO 17025 Accreditation Detail Review
 
iso 15189.pptx
iso 15189.pptxiso 15189.pptx
iso 15189.pptx
 
ISO 17025.pptx
ISO 17025.pptxISO 17025.pptx
ISO 17025.pptx
 
ISO 17025 Standard for Laboratories
 ISO 17025 Standard for Laboratories ISO 17025 Standard for Laboratories
ISO 17025 Standard for Laboratories
 
6.Management system requirements.pptx
6.Management system requirements.pptx6.Management system requirements.pptx
6.Management system requirements.pptx
 
Quality Assurance of Laboratory Test Results based on ISO/IEC 17025
Quality Assurance of Laboratory Test Results based on ISO/IEC 17025Quality Assurance of Laboratory Test Results based on ISO/IEC 17025
Quality Assurance of Laboratory Test Results based on ISO/IEC 17025
 
Estimation of Measurement Uncertainty in Labs: a requirement for ISO 17025 Ac...
Estimation of Measurement Uncertainty in Labs: a requirement for ISO 17025 Ac...Estimation of Measurement Uncertainty in Labs: a requirement for ISO 17025 Ac...
Estimation of Measurement Uncertainty in Labs: a requirement for ISO 17025 Ac...
 
ISO/IEC 17025
ISO/IEC 17025 ISO/IEC 17025
ISO/IEC 17025
 
Measure uncertainty
Measure uncertaintyMeasure uncertainty
Measure uncertainty
 
iso 15189 short.pptx
iso 15189 short.pptxiso 15189 short.pptx
iso 15189 short.pptx
 
Second Party Audit and External Third Party Audit
Second Party Audit and External Third Party AuditSecond Party Audit and External Third Party Audit
Second Party Audit and External Third Party Audit
 
Improving Laboratory Performance Through Quality Control - The role of EQA in...
Improving Laboratory Performance Through Quality Control - The role of EQA in...Improving Laboratory Performance Through Quality Control - The role of EQA in...
Improving Laboratory Performance Through Quality Control - The role of EQA in...
 
What Documentation Required for ISO 17025:2017 Accreditation?
What Documentation Required for ISO 17025:2017 Accreditation?What Documentation Required for ISO 17025:2017 Accreditation?
What Documentation Required for ISO 17025:2017 Accreditation?
 
List of ISO 17025:2017 documents for calibration laboratory
List of ISO 17025:2017 documents for calibration laboratoryList of ISO 17025:2017 documents for calibration laboratory
List of ISO 17025:2017 documents for calibration laboratory
 
Different Approaches in Estimating Measurement Uncertainty
Different Approaches in Estimating Measurement UncertaintyDifferent Approaches in Estimating Measurement Uncertainty
Different Approaches in Estimating Measurement Uncertainty
 
Handling of OOS Dr.A. Amsavel
Handling of OOS  Dr.A. AmsavelHandling of OOS  Dr.A. Amsavel
Handling of OOS Dr.A. Amsavel
 
How to Perform a Successful Internal Quality Audit
How to Perform a Successful Internal Quality AuditHow to Perform a Successful Internal Quality Audit
How to Perform a Successful Internal Quality Audit
 
Method verification
Method verificationMethod verification
Method verification
 
Quality manual
Quality manualQuality manual
Quality manual
 
Designing an appropriate qc design procedure for your lab 5 mar15
Designing an appropriate qc design procedure for your lab 5 mar15Designing an appropriate qc design procedure for your lab 5 mar15
Designing an appropriate qc design procedure for your lab 5 mar15
 

Similar a 17025 ph sample certificate

Nitrogen plant Genset Report
Nitrogen plant Genset ReportNitrogen plant Genset Report
Nitrogen plant Genset Report
Osvaldo Zumba
 
Cinap alternadores 2012
Cinap alternadores 2012Cinap alternadores 2012
Cinap alternadores 2012
mapple2012
 
1. va1 lici-00 eg-i-m4d-das-0001.positive displacement flowmeters
1. va1 lici-00 eg-i-m4d-das-0001.positive displacement flowmeters1. va1 lici-00 eg-i-m4d-das-0001.positive displacement flowmeters
1. va1 lici-00 eg-i-m4d-das-0001.positive displacement flowmeters
Khang Trang
 
JCB 3CX-2 Sitemastr BACKOHE LOADER Parts Catalogue Manual (Serial Number 0029...
JCB 3CX-2 Sitemastr BACKOHE LOADER Parts Catalogue Manual (Serial Number 0029...JCB 3CX-2 Sitemastr BACKOHE LOADER Parts Catalogue Manual (Serial Number 0029...
JCB 3CX-2 Sitemastr BACKOHE LOADER Parts Catalogue Manual (Serial Number 0029...
fujsked9dik
 
JCB 3CX-2 Sitemastr BACKOHE LOADER Parts Catalogue Manual (Serial Number 0029...
JCB 3CX-2 Sitemastr BACKOHE LOADER Parts Catalogue Manual (Serial Number 0029...JCB 3CX-2 Sitemastr BACKOHE LOADER Parts Catalogue Manual (Serial Number 0029...
JCB 3CX-2 Sitemastr BACKOHE LOADER Parts Catalogue Manual (Serial Number 0029...
jdjskdkdmmd
 
JCB 3CX 20 BACKOHE LOADER Parts Catalogue Manual (Serial Number 00400000-0043...
JCB 3CX 20 BACKOHE LOADER Parts Catalogue Manual (Serial Number 00400000-0043...JCB 3CX 20 BACKOHE LOADER Parts Catalogue Manual (Serial Number 00400000-0043...
JCB 3CX 20 BACKOHE LOADER Parts Catalogue Manual (Serial Number 00400000-0043...
jdjskdkdmmd
 
Production Certifications
Production CertificationsProduction Certifications
Production Certifications
Mark Aguilar
 

Similar a 17025 ph sample certificate (18)

Nitrogen plant Genset Report
Nitrogen plant Genset ReportNitrogen plant Genset Report
Nitrogen plant Genset Report
 
Actuators Repuestos equipo pesado en Panama - RARS
Actuators Repuestos equipo pesado en Panama - RARSActuators Repuestos equipo pesado en Panama - RARS
Actuators Repuestos equipo pesado en Panama - RARS
 
Cinap alternadores 2012
Cinap alternadores 2012Cinap alternadores 2012
Cinap alternadores 2012
 
30009609 102-ee1
30009609 102-ee130009609 102-ee1
30009609 102-ee1
 
Pioneer xr-a370 670
Pioneer xr-a370 670Pioneer xr-a370 670
Pioneer xr-a370 670
 
2015.09 ATTF 1186C Incubator (Paulo).pdf
2015.09 ATTF 1186C Incubator (Paulo).pdf2015.09 ATTF 1186C Incubator (Paulo).pdf
2015.09 ATTF 1186C Incubator (Paulo).pdf
 
1. va1 lici-00 eg-i-m4d-das-0001.positive displacement flowmeters
1. va1 lici-00 eg-i-m4d-das-0001.positive displacement flowmeters1. va1 lici-00 eg-i-m4d-das-0001.positive displacement flowmeters
1. va1 lici-00 eg-i-m4d-das-0001.positive displacement flowmeters
 
JCB 3CX-2 Sitemastr BACKOHE LOADER Parts Catalogue Manual (Serial Number 0029...
JCB 3CX-2 Sitemastr BACKOHE LOADER Parts Catalogue Manual (Serial Number 0029...JCB 3CX-2 Sitemastr BACKOHE LOADER Parts Catalogue Manual (Serial Number 0029...
JCB 3CX-2 Sitemastr BACKOHE LOADER Parts Catalogue Manual (Serial Number 0029...
 
JCB 3CX-2 Sitemastr BACKOHE LOADER Parts Catalogue Manual (Serial Number 0029...
JCB 3CX-2 Sitemastr BACKOHE LOADER Parts Catalogue Manual (Serial Number 0029...JCB 3CX-2 Sitemastr BACKOHE LOADER Parts Catalogue Manual (Serial Number 0029...
JCB 3CX-2 Sitemastr BACKOHE LOADER Parts Catalogue Manual (Serial Number 0029...
 
John Deere 200lc Excavator Operation And Tests Repair Manual.pdf
John Deere 200lc Excavator Operation And Tests Repair Manual.pdfJohn Deere 200lc Excavator Operation And Tests Repair Manual.pdf
John Deere 200lc Excavator Operation And Tests Repair Manual.pdf
 
EAPL TIMER & CONTROLLER PRICE LIST wef 01-08-2022.pdf
EAPL TIMER & CONTROLLER PRICE LIST wef 01-08-2022.pdfEAPL TIMER & CONTROLLER PRICE LIST wef 01-08-2022.pdf
EAPL TIMER & CONTROLLER PRICE LIST wef 01-08-2022.pdf
 
Hagie sts10 parts manual 2010
Hagie sts10 parts manual 2010Hagie sts10 parts manual 2010
Hagie sts10 parts manual 2010
 
Product catalog 2012- Lubrication & Workshop Equipment
Product catalog 2012- Lubrication & Workshop EquipmentProduct catalog 2012- Lubrication & Workshop Equipment
Product catalog 2012- Lubrication & Workshop Equipment
 
JCB 3CX 20 BACKOHE LOADER Parts Catalogue Manual (Serial Number 00400000-0043...
JCB 3CX 20 BACKOHE LOADER Parts Catalogue Manual (Serial Number 00400000-0043...JCB 3CX 20 BACKOHE LOADER Parts Catalogue Manual (Serial Number 00400000-0043...
JCB 3CX 20 BACKOHE LOADER Parts Catalogue Manual (Serial Number 00400000-0043...
 
Datasheet
DatasheetDatasheet
Datasheet
 
Computer Sale in Bangalore
Computer Sale in BangaloreComputer Sale in Bangalore
Computer Sale in Bangalore
 
Production Certifications
Production CertificationsProduction Certifications
Production Certifications
 
04 ret670 test report
04 ret670 test report04 ret670 test report
04 ret670 test report
 

Último

Cloud Frontiers: A Deep Dive into Serverless Spatial Data and FME
Cloud Frontiers:  A Deep Dive into Serverless Spatial Data and FMECloud Frontiers:  A Deep Dive into Serverless Spatial Data and FME
Cloud Frontiers: A Deep Dive into Serverless Spatial Data and FME
Safe Software
 
Architecting Cloud Native Applications
Architecting Cloud Native ApplicationsArchitecting Cloud Native Applications
Architecting Cloud Native Applications
WSO2
 
Why Teams call analytics are critical to your entire business
Why Teams call analytics are critical to your entire businessWhy Teams call analytics are critical to your entire business
Why Teams call analytics are critical to your entire business
panagenda
 
Finding Java's Hidden Performance Traps @ DevoxxUK 2024
Finding Java's Hidden Performance Traps @ DevoxxUK 2024Finding Java's Hidden Performance Traps @ DevoxxUK 2024
Finding Java's Hidden Performance Traps @ DevoxxUK 2024
Victor Rentea
 

Último (20)

EMPOWERMENT TECHNOLOGY GRADE 11 QUARTER 2 REVIEWER
EMPOWERMENT TECHNOLOGY GRADE 11 QUARTER 2 REVIEWEREMPOWERMENT TECHNOLOGY GRADE 11 QUARTER 2 REVIEWER
EMPOWERMENT TECHNOLOGY GRADE 11 QUARTER 2 REVIEWER
 
Corporate and higher education May webinar.pptx
Corporate and higher education May webinar.pptxCorporate and higher education May webinar.pptx
Corporate and higher education May webinar.pptx
 
ProductAnonymous-April2024-WinProductDiscovery-MelissaKlemke
ProductAnonymous-April2024-WinProductDiscovery-MelissaKlemkeProductAnonymous-April2024-WinProductDiscovery-MelissaKlemke
ProductAnonymous-April2024-WinProductDiscovery-MelissaKlemke
 
Strategize a Smooth Tenant-to-tenant Migration and Copilot Takeoff
Strategize a Smooth Tenant-to-tenant Migration and Copilot TakeoffStrategize a Smooth Tenant-to-tenant Migration and Copilot Takeoff
Strategize a Smooth Tenant-to-tenant Migration and Copilot Takeoff
 
Artificial Intelligence Chap.5 : Uncertainty
Artificial Intelligence Chap.5 : UncertaintyArtificial Intelligence Chap.5 : Uncertainty
Artificial Intelligence Chap.5 : Uncertainty
 
FWD Group - Insurer Innovation Award 2024
FWD Group - Insurer Innovation Award 2024FWD Group - Insurer Innovation Award 2024
FWD Group - Insurer Innovation Award 2024
 
Web Form Automation for Bonterra Impact Management (fka Social Solutions Apri...
Web Form Automation for Bonterra Impact Management (fka Social Solutions Apri...Web Form Automation for Bonterra Impact Management (fka Social Solutions Apri...
Web Form Automation for Bonterra Impact Management (fka Social Solutions Apri...
 
Introduction to Multilingual Retrieval Augmented Generation (RAG)
Introduction to Multilingual Retrieval Augmented Generation (RAG)Introduction to Multilingual Retrieval Augmented Generation (RAG)
Introduction to Multilingual Retrieval Augmented Generation (RAG)
 
Cloud Frontiers: A Deep Dive into Serverless Spatial Data and FME
Cloud Frontiers:  A Deep Dive into Serverless Spatial Data and FMECloud Frontiers:  A Deep Dive into Serverless Spatial Data and FME
Cloud Frontiers: A Deep Dive into Serverless Spatial Data and FME
 
Exploring Multimodal Embeddings with Milvus
Exploring Multimodal Embeddings with MilvusExploring Multimodal Embeddings with Milvus
Exploring Multimodal Embeddings with Milvus
 
Biography Of Angeliki Cooney | Senior Vice President Life Sciences | Albany, ...
Biography Of Angeliki Cooney | Senior Vice President Life Sciences | Albany, ...Biography Of Angeliki Cooney | Senior Vice President Life Sciences | Albany, ...
Biography Of Angeliki Cooney | Senior Vice President Life Sciences | Albany, ...
 
Six Myths about Ontologies: The Basics of Formal Ontology
Six Myths about Ontologies: The Basics of Formal OntologySix Myths about Ontologies: The Basics of Formal Ontology
Six Myths about Ontologies: The Basics of Formal Ontology
 
MS Copilot expands with MS Graph connectors
MS Copilot expands with MS Graph connectorsMS Copilot expands with MS Graph connectors
MS Copilot expands with MS Graph connectors
 
"I see eyes in my soup": How Delivery Hero implemented the safety system for ...
"I see eyes in my soup": How Delivery Hero implemented the safety system for ..."I see eyes in my soup": How Delivery Hero implemented the safety system for ...
"I see eyes in my soup": How Delivery Hero implemented the safety system for ...
 
Apidays New York 2024 - Accelerating FinTech Innovation by Vasa Krishnan, Fin...
Apidays New York 2024 - Accelerating FinTech Innovation by Vasa Krishnan, Fin...Apidays New York 2024 - Accelerating FinTech Innovation by Vasa Krishnan, Fin...
Apidays New York 2024 - Accelerating FinTech Innovation by Vasa Krishnan, Fin...
 
Architecting Cloud Native Applications
Architecting Cloud Native ApplicationsArchitecting Cloud Native Applications
Architecting Cloud Native Applications
 
DBX First Quarter 2024 Investor Presentation
DBX First Quarter 2024 Investor PresentationDBX First Quarter 2024 Investor Presentation
DBX First Quarter 2024 Investor Presentation
 
Why Teams call analytics are critical to your entire business
Why Teams call analytics are critical to your entire businessWhy Teams call analytics are critical to your entire business
Why Teams call analytics are critical to your entire business
 
Understanding the FAA Part 107 License ..
Understanding the FAA Part 107 License ..Understanding the FAA Part 107 License ..
Understanding the FAA Part 107 License ..
 
Finding Java's Hidden Performance Traps @ DevoxxUK 2024
Finding Java's Hidden Performance Traps @ DevoxxUK 2024Finding Java's Hidden Performance Traps @ DevoxxUK 2024
Finding Java's Hidden Performance Traps @ DevoxxUK 2024
 

17025 ph sample certificate

  • 1. MICROPRECISION CALIBRATION INC. BLOCK 2 LOT 4 CALAMBA PREMIERE INTERNATIONAL PARK BATINO, CALAMBA CITY, LAGUNA, PHILS 4027 +63495023015 Certificate of Calibration Date: 5/16/2008 Lab # 935.05 Certificate #: 315634 Customer: SAMPLE 85/18 BUMDUREN KRASANG Purchase Order: 12345 BURIRAM, , 31160 Work Order: N/A MPC Control #: US37290511 Serial Number: US37290511 Asset ID: 8957 Department: N/A Gage Type: POWER SENSOR Performed By: DANNY BOY B. BUTIAL Manufacturer: HEWLETT PACKARD Received Condition: IN TOLERANCE Model Number: 8481A Returned Condition: IN TOLERANCE Size: 10MHZ~18GHZ Cal Date: October 31, 2007 Temp./RH: 23 ° C / 49 % Cal. Interval: 12 MONTHS Cal. Due Date: October 31, 2008 *Calibration Notes: Site loc: MPC-Lab. See attached data report. Standards Used To Calibrate Equipment I.D. Description Model Serial Manufacturer Cal. Due Date Traceability # AD3312 SIGNAL GENERATOR 83640B 3844A00758 HEWLETT PACKARD 1/16/2009 227210 AI2393 DIRECTIONAL COUPLER 11692D 1212A02566 HEWLETT PACKARD 8/29/2008 139290 AN1816 POWER METER E4419B GB40208110 AGILENT 3/5/2009 260329 AN1869 POWER SPLITTER 11667A 04040 HEWLETT PACKARD 5/9/2009 310631 AU6515 POWER SENSOR 8481A 1726A03110 AGILENT 1/25/2009 1-1201452485-1 R1640 DIRECTIONAL COUPLER 778D 15050 HEWLETT PACKARD 8/20/2008 133363 Procedures Used In This Event: Procedure Name Description 848X SERIES, HP POWER SENSOR, COAXIAL (HP) Calibrating Technician: QC Approval: DANNY BOY B. BUTIAL Jarrod Trammell Uncertainty Estimated at 2 to 1 Uncertainties have been estimated at a 95 percent confidence level (k=2). All calibrations performed conform to ISO 17025:2005, ISO 9001:2000, ANSI Z540-1, MPC Quality Manual, Calibration System Rev5, and customer requirements All standards used are either traceable to the National Institute of Standards Technology or have intrinsic accuracy. All services performed have used proper manufacturer and industrial service techniques. This report may not be reproduced in part without written permission of Micro Precision's Quality Assurance Manager. Page 1 of 1 (CERT, Rev 0)
  • 2. Hewlett Packard 8481A Power Sensor Procedure Reference: 848X SERIES, HP Work Order: n/a ID Number: US37290511 Serial Number: US37290511 Asset Number: VM#8957 / PW#003850 Calibration Date: Oct. 31, 2007 Temperature: 23.0°C Humidity: 49.0% Section 1 - RHO and Calibration Factors Reference Cal Factor: 100.0 % Frequency ( Ghz ) Cal Factor CF Uncertainty Measured RHO RHO Upper Limit .01 100.2 % 1.3 % .048 < .166 .03 100.1 % 1.3 % .048 < .166 .1 100.1 % 1.3 % .023 < .048 .3 100.2 % 1.3 % .020 < .048 1 98.8 % 1.3 % .020 < .048 2 98.8 % 1.4 % .020 < .083 3 97.7 % 1.6 % .021 < .083 4 96.7 % 1.6 % .021 < .083 5 97.2 % 1.7 % .018 < .083 6 96.3 % 2.0 % .020 < .083 7 97.8 % 2.0 % .029 < .083 8 95.1 % 2.1 % .030 < .083 9 95.2 % 2.3 % .030 < .083 10 92.9 % 2.5 % .042 < .083 11 96.0 % 2.6 % .043 < .083 12 93.1 % 2.5 % .027 < .083 12.4 93.0 % 2.5 % .022 < .083 13 94.3 % 2.6 % .016 < .123 14 91.6 % 2.7 % .016 < .123 15 92.2 % 2.7 % .030 < .123 16 93.7 % 2.6 % .058 < .123 17 91.9 % 2.8 % .060 < .123 18 89.5 % 2.9 % .054 < .123 Page 1 of 2
  • 3. ID Number: US37290511 Calibration Factor Chart 100.0 % 98.0 % 96.0 % 94.0 % Cal Factor % 92.0 % 90.0 % 88.0 % 86.0 % 84.0 % .01 .03 .1 .3 1 2 3 4 5 6 7 8 9 10 11 12 12.4 13 14 15 16 17 18 Frequency ( Ghz ) RHO Chart .200 .180 .160 .140 .120 RHO .100 Measured RHO .080 Upper Limit .060 .040 .020 .000 .01 .03 .1 .3 1 2 3 4 5 6 7 8 9 10 11 1212.413 14 15 16 17 18 Frequency ( Ghz ) Page 2 of 2