SlideShare una empresa de Scribd logo
1 de 13
Descargar para leer sin conexión
Radiochromic films and
Monte Carlo simulations to
analyze abnormal Mosfet
readings in IOERT breast
cancer treatments
G.Taccini1, F. Cavagnetto1, L.Ferri1, R.Bampi1,
M.Guenzi2, G.Iaccario3, A. Soriani3
1 AOU IRCCS San Martino IST, Medical Physics, Genoa, Italy
2 AOU IRCCS San Martino IST,Radiation Oncology, Genoa, Italy
3 Istituto Ragina Elena, IFO, Rome, Italy
IOeRT to treat early-stage breast cancer
Dedicated mobile accelerator: LIAC Sordina, Italy
 Electron beams: 4, 6 ,8, 10 MeV calibrated with Fricke
dosimeters and two plane parallel ion chamber
comparison*
 PMMA cylindrical applicators : 4-10 cm ø for hard
docking beam collimation
 Steel-PTFE shielding disk (3mm+3mm) ; placed
between the deep face of the residual breast and the
pectoralis muscle to minimize thoracic wall irradiation
 Every day dosimetry check to calculate optimized MU
 In vivo dosimetry to check and properly correct MU
Rosi A, Viti V. «Guidelines for quality assurance on ontra-operative radiation therapy» ISTISAN 2003
Report No. 03/1 EN
Cavagnetto F. et al In vivo dosimetry in IORT cancer treatment. 2011; 99
Cavagnetto F. Et al A comparison between several dosimetric methods in IORT. ESTRO 31 PO-0835
Between September 2009 and April 2013: 512 patients.
Eligibility criteria for IORT treatment evaluated at two different decisional steps:
―> first step decision after diagnosis
―> second step decision intrasurgery
Full dose intraoperativly irradiation if: tumor diameter less than 2,5 cm,
infiltrating ductal histology, no mammographic evidence of multifocality,
negative resection margins, no more than 3 axillary nodes, no extensive
intraductal component.
 21 Gy alone for high local relapse risk (pT2-pT1c)
 18 Gy alone for low local relapse risk (max ø 1cm, pT1a-pT1b)
 10 Gy as boost to the tumor bed if not eligible to full dose irradiation
 16 Gy to nipple area complex if nipple sparing mastectomy
Guenzi M. et al A two step selection of breast cancer patients candidates for exclusive
IORT with electrons: a mono-institutional experience. Anticancer Res. (2012) 32(4):1533-6
In vivo dosimetry to monitor exit dose : 293 patients
 MicroMOSFET inside a steril and thin catheter is
fastened to the PTFE face of the shielding disk
–> dose at the deeper part of the target is
checked.
 Treatment splitted in two parts to correct MU
according to detector half treatment reading.
 Dose prescribed at the deeper part of the
target and normalized at 90%: prescribed dose
plus 10% is acceptable otherwise MU are
properly correct.
MicroMOSFET TN-502RDM (Best Medical Canadian):
 Immediate readout
 very small active volume 7” x 6.25” x 1.63”
 Calibrated for each avaiable LIAC energy
Ciocca M et al Real-time in vivo dosimetry using micro-MOSFET
detectors during intraoperative electron beam radiation therapy in
early-stage breast cancer. Radiother Oncol 2006; 78:213-6.
Agostinelli et al. On line optimization of intraoperative electron beam radiotherapy of the
breast. Radiother Oncol 2012; 103: 188-192.)
18% of MOSFET readings at half treatment > 10%
Shielding disk backscattering radiation
MOSFET directional radiation dependance
Air gaps and non homogeneous tissue could create high dose
hot spot in the irradiated glandular tissue due to electrons
nature.
Monte Carlo simulation -> no evidence of backscattering rad
BOLUS
AIR
MOSFET
TEFLON
STEEL
0%
20%
40%
60%
80%
100%
120%
0,0 10,0 20,0 30,0 40,0 50,0 60,0 70,0
D%
mm
experimental set-up
water
Our hipothesis
Monte Carlo simulations
supplied by A Soriani,
G.Iaccario
18% of MOSFET readings at half treatment > 10%
Shielding disk backscattering
MOSFET directional radiation dependance
Air gaps and non homogeneous tissue tissue could create high
dose hot spot in the irradiated glandular tissue due to
electrons nature.
 Micro-MOSFETs in slab water phantom
under full build up conditions irradiated
using electron beams supplied by LIAC.
 Directional dependences of microMOSFET
were found to be within ±2% between 0
and 45 degree LIAC angles.
Chung J.,Suh T. et al Dosimetric Characteristics of Standard and Micro MOSFET Dosimeters as In-vivo Dosimeter for
Clinical Electron Beam J. Korean Phys.Soc. 55,2566
Ciocca M. et al Real-time in vivo dosimetry using micro-MOSFET detectors during intraoperative electron beam
radiation therapy in early-stage breast cancer. Radiother Oncol. 2006 Feb;78(2):213-6. Epub 2005 Dec 15
Our hipothesis
18% of MOSFET readings at half treatment > 10%
Shielding disk backscattering
MOSFET directional radiation dependance
Air gaps and non homogeneous tissue tissue could create high
dose hot spot in the irradiated glandular tissue due to
electrons nature.
Air gaps and non homogeneous tissue simulated with a filled
water glove ; GAFCHROMIC EBT2 radiochromic film irradiation
Our hipothesis
Air gap between two “fingers” overdosages
EBT2
C. Fiandra et al. “Absolute and relative dose measurements
with Gafchromic™ EBT film for high energy electron beams
with different doses per pulse” Med Phys, 35, 2008
 Energy 10 MeV
 Applicator 60 mm straight- end
 500 cGy to R90 according to daily
dosimetry check
 Expected dose ≤ 556 cGy
Epson Expression
10000XL to scan
films
 Picodose Pro®
to extract red
channel
PTW VeriSoft®
to determine
isodoses and
profiles.
7%
9%
11%
13%
15%
17%
19%
21%
-10 -8 -6 -4 -2 0 2 4 6 8 10
DoseDifference%
profile(mm)
% Difference between measured and expected dose
% Difference between measured and expected dose
7%
8%
9%
10%
11%
12%
13%
14%
15%
16%
17%
-10 -8 -6 -4 -2 0 2 4 6 8 10
DoseDifference%
Profile(mm)
Key: gap size
Our purpose: confirm overdosages with Monte Carlo simulations
to see overdosages
the key is : gap size
Reproducible
set up
 Various heights: 0.5-1 cm
 Various widths: 0.2-0.5 cm
Results normilized to the
expected dose at the
appropriate depth.
Monte Carlo simulation confirms our experimental results
Peak Height: +13%EGS ncr/BEAMnrcode,
1mm simulation step,
voxel dimension 1x 5mmx
29mm (GAF thickness)
Monte Carlo simulations
supplied by A Soriani,
G.Iaccario
Summary
 September 2009 / april 2013: 512 patients affected by early-stage breast unifocal
cancer treated with IOeRT .
 In vivo dosimetry using micro-MOSFET performed on 293 patients.
 Micro-MOSFET placed in the deeper part of the target to check the dose delivered
during the first half of the treatment and eventually correct the second half.
 Half treatment MOSFET readings, normalized to prescribed dose are higher than 10%in
the 18% of cases .
 Gafchromic EBT2 films under non homogeneous tissue simulated with a glove filled
with water show overdoses at air-water interfaces.
 Experimental set up with air gap of various size and EBT2 led us to evaluate the gap
size effect.
 Monte Carlo simulation confirm overdosages at air/water interfaces.
 In vivo investigation with both micro MOSFET and GafCHROMIC
F. Cavagnetto et al. In vivo dosimetry using MOSFET and radiochromic films in
intraoperative radiotherapy for breast cancer treatments, OC-0487 2nd ESTRO Forum 2013

Más contenido relacionado

La actualidad más candente

Future Developments In Radiation Therapy For Prostate Cancer
Future Developments In Radiation Therapy For Prostate CancerFuture Developments In Radiation Therapy For Prostate Cancer
Future Developments In Radiation Therapy For Prostate Cancerfondas vakalis
 
THE RATIONALE AND BENEFITS OF IGRT
THE RATIONALE AND BENEFITS OF IGRTTHE RATIONALE AND BENEFITS OF IGRT
THE RATIONALE AND BENEFITS OF IGRTMelissa McClement
 
Igrt Srt For Lung Cancer
Igrt Srt For Lung CancerIgrt Srt For Lung Cancer
Igrt Srt For Lung Cancerfondas vakalis
 
Imrt Symposia Global Overview
Imrt Symposia Global OverviewImrt Symposia Global Overview
Imrt Symposia Global Overviewfondas vakalis
 
New Techniques in Radiotherapy
New Techniques in RadiotherapyNew Techniques in Radiotherapy
New Techniques in RadiotherapySantam Chakraborty
 
Technical Advances in radiotherapy for Lung (and liver) Cancer
Technical Advances in radiotherapy for Lung (and liver) CancerTechnical Advances in radiotherapy for Lung (and liver) Cancer
Technical Advances in radiotherapy for Lung (and liver) Cancerspa718
 
Forward imrt in breast radiotherapy
Forward imrt in breast radiotherapyForward imrt in breast radiotherapy
Forward imrt in breast radiotherapyDr.Amrita Rakesh
 
IMRT in Head & Neck Cancer
IMRT in Head & Neck CancerIMRT in Head & Neck Cancer
IMRT in Head & Neck CancerJyotirup Goswami
 
Accelerated partial breast irradiation
Accelerated partial breast irradiationAccelerated partial breast irradiation
Accelerated partial breast irradiationBharti Devnani
 

La actualidad más candente (20)

Mundt Gyn Session
Mundt Gyn SessionMundt Gyn Session
Mundt Gyn Session
 
Iort dr kiran
Iort  dr kiran Iort  dr kiran
Iort dr kiran
 
Future Developments In Radiation Therapy For Prostate Cancer
Future Developments In Radiation Therapy For Prostate CancerFuture Developments In Radiation Therapy For Prostate Cancer
Future Developments In Radiation Therapy For Prostate Cancer
 
THE RATIONALE AND BENEFITS OF IGRT
THE RATIONALE AND BENEFITS OF IGRTTHE RATIONALE AND BENEFITS OF IGRT
THE RATIONALE AND BENEFITS OF IGRT
 
Rt
RtRt
Rt
 
Igrt Srt For Lung Cancer
Igrt Srt For Lung CancerIgrt Srt For Lung Cancer
Igrt Srt For Lung Cancer
 
Future Technological needs in Oncology
Future Technological needs in Oncology Future Technological needs in Oncology
Future Technological needs in Oncology
 
Imrt Symposia Global Overview
Imrt Symposia Global OverviewImrt Symposia Global Overview
Imrt Symposia Global Overview
 
IMPROVISED RADIOTHERAPY TECHNIQUES IN TELE COBALT WITHOUT MLC
IMPROVISED RADIOTHERAPY TECHNIQUES IN TELE COBALT WITHOUT MLCIMPROVISED RADIOTHERAPY TECHNIQUES IN TELE COBALT WITHOUT MLC
IMPROVISED RADIOTHERAPY TECHNIQUES IN TELE COBALT WITHOUT MLC
 
Radiotherapy
RadiotherapyRadiotherapy
Radiotherapy
 
4 D Igrt
4 D Igrt4 D Igrt
4 D Igrt
 
New Techniques in Radiotherapy
New Techniques in RadiotherapyNew Techniques in Radiotherapy
New Techniques in Radiotherapy
 
Modern Radiotherapy
Modern RadiotherapyModern Radiotherapy
Modern Radiotherapy
 
Technical Advances in radiotherapy for Lung (and liver) Cancer
Technical Advances in radiotherapy for Lung (and liver) CancerTechnical Advances in radiotherapy for Lung (and liver) Cancer
Technical Advances in radiotherapy for Lung (and liver) Cancer
 
Lung sbrt ppt
Lung  sbrt pptLung  sbrt ppt
Lung sbrt ppt
 
Forward imrt in breast radiotherapy
Forward imrt in breast radiotherapyForward imrt in breast radiotherapy
Forward imrt in breast radiotherapy
 
IMRT in Head & Neck Cancer
IMRT in Head & Neck CancerIMRT in Head & Neck Cancer
IMRT in Head & Neck Cancer
 
IGRT in lung cancer
IGRT in lung cancerIGRT in lung cancer
IGRT in lung cancer
 
Srs and sbrt 2 dr.kiran
Srs and sbrt 2 dr.kiranSrs and sbrt 2 dr.kiran
Srs and sbrt 2 dr.kiran
 
Accelerated partial breast irradiation
Accelerated partial breast irradiationAccelerated partial breast irradiation
Accelerated partial breast irradiation
 

Similar a IORT_ in vivo dosimetry

Topic -Analysis of Patients specific Quality Assurance of IMRT/VMAT
Topic -Analysis of Patients specific Quality Assurance of IMRT/VMATTopic -Analysis of Patients specific Quality Assurance of IMRT/VMAT
Topic -Analysis of Patients specific Quality Assurance of IMRT/VMATRajeev Kumar Pandit
 
Imrt In Gynecologic Malignancies
Imrt In Gynecologic MalignanciesImrt In Gynecologic Malignancies
Imrt In Gynecologic Malignanciesfondas vakalis
 
SPECT/CT: HOW Much Radiation Dose CT Constitute
SPECT/CT: HOW Much Radiation Dose CT ConstituteSPECT/CT: HOW Much Radiation Dose CT Constitute
SPECT/CT: HOW Much Radiation Dose CT ConstituteShahid Younas
 
Target Tape, Spine (3.9.2015)
Target Tape, Spine (3.9.2015)Target Tape, Spine (3.9.2015)
Target Tape, Spine (3.9.2015)Nick Seto
 
Igrt In Gynecologic Malignancies
Igrt In Gynecologic MalignanciesIgrt In Gynecologic Malignancies
Igrt In Gynecologic Malignanciesfondas vakalis
 
Stereotactic body radiotherapy
Stereotactic body radiotherapyStereotactic body radiotherapy
Stereotactic body radiotherapyNanditha Nukala
 
Brachytherapy IGABT
Brachytherapy IGABTBrachytherapy IGABT
Brachytherapy IGABTPuneet Seth
 
A low cost and portable microwave imaging system for breast tumor detection u...
A low cost and portable microwave imaging system for breast tumor detection u...A low cost and portable microwave imaging system for breast tumor detection u...
A low cost and portable microwave imaging system for breast tumor detection u...rsfdtd
 
s41598-019-51620-z.pdf
s41598-019-51620-z.pdfs41598-019-51620-z.pdf
s41598-019-51620-z.pdfAsit Panda
 
Method and apparatus for pet in hadrontherapy
Method and apparatus for pet in hadrontherapyMethod and apparatus for pet in hadrontherapy
Method and apparatus for pet in hadrontherapyToscana Open Research
 
Real-time model for adaptive radiation therapy a biomechanical approach using...
Real-time model for adaptive radiation therapy a biomechanical approach using...Real-time model for adaptive radiation therapy a biomechanical approach using...
Real-time model for adaptive radiation therapy a biomechanical approach using...Az.Ospedaliero-Universitaria di Modena
 
Virtual Navigator Real-Time Ultrasound Fusion Imaging with Positron Emission ...
Virtual Navigator Real-Time Ultrasound Fusion Imaging with Positron Emission ...Virtual Navigator Real-Time Ultrasound Fusion Imaging with Positron Emission ...
Virtual Navigator Real-Time Ultrasound Fusion Imaging with Positron Emission ...rosopeplaton
 
IRJET- A Review Paper on Cancer Biology using Biotecnology and Nanotechnology
IRJET- A Review Paper on Cancer Biology using Biotecnology and NanotechnologyIRJET- A Review Paper on Cancer Biology using Biotecnology and Nanotechnology
IRJET- A Review Paper on Cancer Biology using Biotecnology and NanotechnologyIRJET Journal
 
The role and techniques of MRI in gamma knife radiosurgery: a technologist’s ...
The role and techniques of MRI in gamma knife radiosurgery: a technologist’s ...The role and techniques of MRI in gamma knife radiosurgery: a technologist’s ...
The role and techniques of MRI in gamma knife radiosurgery: a technologist’s ...Holly Chun
 
MR Guided Focused Ultrasound (MRgFUS) Overview of Present and Future Clinical...
MR Guided Focused Ultrasound (MRgFUS) Overview of Present and Future Clinical...MR Guided Focused Ultrasound (MRgFUS) Overview of Present and Future Clinical...
MR Guided Focused Ultrasound (MRgFUS) Overview of Present and Future Clinical...u.surgery
 

Similar a IORT_ in vivo dosimetry (20)

Topic -Analysis of Patients specific Quality Assurance of IMRT/VMAT
Topic -Analysis of Patients specific Quality Assurance of IMRT/VMATTopic -Analysis of Patients specific Quality Assurance of IMRT/VMAT
Topic -Analysis of Patients specific Quality Assurance of IMRT/VMAT
 
Apbi
ApbiApbi
Apbi
 
F04433538
F04433538F04433538
F04433538
 
Imrt In Gynecologic Malignancies
Imrt In Gynecologic MalignanciesImrt In Gynecologic Malignancies
Imrt In Gynecologic Malignancies
 
SPECT/CT: HOW Much Radiation Dose CT Constitute
SPECT/CT: HOW Much Radiation Dose CT ConstituteSPECT/CT: HOW Much Radiation Dose CT Constitute
SPECT/CT: HOW Much Radiation Dose CT Constitute
 
Target Tape, Spine (3.9.2015)
Target Tape, Spine (3.9.2015)Target Tape, Spine (3.9.2015)
Target Tape, Spine (3.9.2015)
 
Igrt In Gynecologic Malignancies
Igrt In Gynecologic MalignanciesIgrt In Gynecologic Malignancies
Igrt In Gynecologic Malignancies
 
Stereotactic body radiotherapy
Stereotactic body radiotherapyStereotactic body radiotherapy
Stereotactic body radiotherapy
 
Brachytherapy IGABT
Brachytherapy IGABTBrachytherapy IGABT
Brachytherapy IGABT
 
A low cost and portable microwave imaging system for breast tumor detection u...
A low cost and portable microwave imaging system for breast tumor detection u...A low cost and portable microwave imaging system for breast tumor detection u...
A low cost and portable microwave imaging system for breast tumor detection u...
 
s41598-019-51620-z.pdf
s41598-019-51620-z.pdfs41598-019-51620-z.pdf
s41598-019-51620-z.pdf
 
Radiation therapy in gynecologic cancer 17-03-15
Radiation therapy in gynecologic cancer 17-03-15Radiation therapy in gynecologic cancer 17-03-15
Radiation therapy in gynecologic cancer 17-03-15
 
Radiation Therapy in Oral Cancer 2018
Radiation Therapy in Oral Cancer 2018Radiation Therapy in Oral Cancer 2018
Radiation Therapy in Oral Cancer 2018
 
Nanotechnology in Cancer - Dr. Cote
Nanotechnology in Cancer - Dr. CoteNanotechnology in Cancer - Dr. Cote
Nanotechnology in Cancer - Dr. Cote
 
Method and apparatus for pet in hadrontherapy
Method and apparatus for pet in hadrontherapyMethod and apparatus for pet in hadrontherapy
Method and apparatus for pet in hadrontherapy
 
Real-time model for adaptive radiation therapy a biomechanical approach using...
Real-time model for adaptive radiation therapy a biomechanical approach using...Real-time model for adaptive radiation therapy a biomechanical approach using...
Real-time model for adaptive radiation therapy a biomechanical approach using...
 
Virtual Navigator Real-Time Ultrasound Fusion Imaging with Positron Emission ...
Virtual Navigator Real-Time Ultrasound Fusion Imaging with Positron Emission ...Virtual Navigator Real-Time Ultrasound Fusion Imaging with Positron Emission ...
Virtual Navigator Real-Time Ultrasound Fusion Imaging with Positron Emission ...
 
IRJET- A Review Paper on Cancer Biology using Biotecnology and Nanotechnology
IRJET- A Review Paper on Cancer Biology using Biotecnology and NanotechnologyIRJET- A Review Paper on Cancer Biology using Biotecnology and Nanotechnology
IRJET- A Review Paper on Cancer Biology using Biotecnology and Nanotechnology
 
The role and techniques of MRI in gamma knife radiosurgery: a technologist’s ...
The role and techniques of MRI in gamma knife radiosurgery: a technologist’s ...The role and techniques of MRI in gamma knife radiosurgery: a technologist’s ...
The role and techniques of MRI in gamma knife radiosurgery: a technologist’s ...
 
MR Guided Focused Ultrasound (MRgFUS) Overview of Present and Future Clinical...
MR Guided Focused Ultrasound (MRgFUS) Overview of Present and Future Clinical...MR Guided Focused Ultrasound (MRgFUS) Overview of Present and Future Clinical...
MR Guided Focused Ultrasound (MRgFUS) Overview of Present and Future Clinical...
 

Último

High Profile Call Girls Coimbatore Saanvi☎️ 8250192130 Independent Escort Se...
High Profile Call Girls Coimbatore Saanvi☎️  8250192130 Independent Escort Se...High Profile Call Girls Coimbatore Saanvi☎️  8250192130 Independent Escort Se...
High Profile Call Girls Coimbatore Saanvi☎️ 8250192130 Independent Escort Se...narwatsonia7
 
Vip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls Available
Vip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls AvailableVip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls Available
Vip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls AvailableNehru place Escorts
 
Bangalore Call Girls Nelamangala Number 7001035870 Meetin With Bangalore Esc...
Bangalore Call Girls Nelamangala Number 7001035870  Meetin With Bangalore Esc...Bangalore Call Girls Nelamangala Number 7001035870  Meetin With Bangalore Esc...
Bangalore Call Girls Nelamangala Number 7001035870 Meetin With Bangalore Esc...narwatsonia7
 
Call Girls Varanasi Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Varanasi Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Varanasi Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Varanasi Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...
Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...
Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...aartirawatdelhi
 
Call Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on Delivery
Call Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on DeliveryCall Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on Delivery
Call Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on Deliverynehamumbai
 
💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...
💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...
💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...Taniya Sharma
 
Call Girls Nagpur Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Nagpur Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Nagpur Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Nagpur Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 6297143586 𖠋 Will You Mis...
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 6297143586 𖠋 Will You Mis...The Most Attractive Hyderabad Call Girls Kothapet 𖠋 6297143586 𖠋 Will You Mis...
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 6297143586 𖠋 Will You Mis...chandars293
 
VIP Call Girls Tirunelveli Aaradhya 8250192130 Independent Escort Service Tir...
VIP Call Girls Tirunelveli Aaradhya 8250192130 Independent Escort Service Tir...VIP Call Girls Tirunelveli Aaradhya 8250192130 Independent Escort Service Tir...
VIP Call Girls Tirunelveli Aaradhya 8250192130 Independent Escort Service Tir...narwatsonia7
 
Chandrapur Call girls 8617370543 Provides all area service COD available
Chandrapur Call girls 8617370543 Provides all area service COD availableChandrapur Call girls 8617370543 Provides all area service COD available
Chandrapur Call girls 8617370543 Provides all area service COD availableDipal Arora
 
Top Rated Bangalore Call Girls Richmond Circle ⟟ 8250192130 ⟟ Call Me For Gen...
Top Rated Bangalore Call Girls Richmond Circle ⟟ 8250192130 ⟟ Call Me For Gen...Top Rated Bangalore Call Girls Richmond Circle ⟟ 8250192130 ⟟ Call Me For Gen...
Top Rated Bangalore Call Girls Richmond Circle ⟟ 8250192130 ⟟ Call Me For Gen...narwatsonia7
 
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.Artifacts in Nuclear Medicine with Identifying and resolving artifacts.
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.MiadAlsulami
 
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...astropune
 
VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...
VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...
VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...Garima Khatri
 
(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...
(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...
(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...Taniya Sharma
 
Call Girls Darjeeling Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Darjeeling Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Darjeeling Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Darjeeling Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Call Girl Number in Vashi Mumbai📲 9833363713 💞 Full Night Enjoy
Call Girl Number in Vashi Mumbai📲 9833363713 💞 Full Night EnjoyCall Girl Number in Vashi Mumbai📲 9833363713 💞 Full Night Enjoy
Call Girl Number in Vashi Mumbai📲 9833363713 💞 Full Night Enjoybabeytanya
 
Call Girls Siliguri Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Siliguri Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Siliguri Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Siliguri Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Call Girls Bhubaneswar Just Call 9907093804 Top Class Call Girl Service Avail...
Call Girls Bhubaneswar Just Call 9907093804 Top Class Call Girl Service Avail...Call Girls Bhubaneswar Just Call 9907093804 Top Class Call Girl Service Avail...
Call Girls Bhubaneswar Just Call 9907093804 Top Class Call Girl Service Avail...Dipal Arora
 

Último (20)

High Profile Call Girls Coimbatore Saanvi☎️ 8250192130 Independent Escort Se...
High Profile Call Girls Coimbatore Saanvi☎️  8250192130 Independent Escort Se...High Profile Call Girls Coimbatore Saanvi☎️  8250192130 Independent Escort Se...
High Profile Call Girls Coimbatore Saanvi☎️ 8250192130 Independent Escort Se...
 
Vip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls Available
Vip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls AvailableVip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls Available
Vip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls Available
 
Bangalore Call Girls Nelamangala Number 7001035870 Meetin With Bangalore Esc...
Bangalore Call Girls Nelamangala Number 7001035870  Meetin With Bangalore Esc...Bangalore Call Girls Nelamangala Number 7001035870  Meetin With Bangalore Esc...
Bangalore Call Girls Nelamangala Number 7001035870 Meetin With Bangalore Esc...
 
Call Girls Varanasi Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Varanasi Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Varanasi Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Varanasi Just Call 9907093804 Top Class Call Girl Service Available
 
Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...
Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...
Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...
 
Call Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on Delivery
Call Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on DeliveryCall Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on Delivery
Call Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on Delivery
 
💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...
💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...
💎VVIP Kolkata Call Girls Parganas🩱7001035870🩱Independent Girl ( Ac Rooms Avai...
 
Call Girls Nagpur Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Nagpur Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Nagpur Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Nagpur Just Call 9907093804 Top Class Call Girl Service Available
 
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 6297143586 𖠋 Will You Mis...
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 6297143586 𖠋 Will You Mis...The Most Attractive Hyderabad Call Girls Kothapet 𖠋 6297143586 𖠋 Will You Mis...
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 6297143586 𖠋 Will You Mis...
 
VIP Call Girls Tirunelveli Aaradhya 8250192130 Independent Escort Service Tir...
VIP Call Girls Tirunelveli Aaradhya 8250192130 Independent Escort Service Tir...VIP Call Girls Tirunelveli Aaradhya 8250192130 Independent Escort Service Tir...
VIP Call Girls Tirunelveli Aaradhya 8250192130 Independent Escort Service Tir...
 
Chandrapur Call girls 8617370543 Provides all area service COD available
Chandrapur Call girls 8617370543 Provides all area service COD availableChandrapur Call girls 8617370543 Provides all area service COD available
Chandrapur Call girls 8617370543 Provides all area service COD available
 
Top Rated Bangalore Call Girls Richmond Circle ⟟ 8250192130 ⟟ Call Me For Gen...
Top Rated Bangalore Call Girls Richmond Circle ⟟ 8250192130 ⟟ Call Me For Gen...Top Rated Bangalore Call Girls Richmond Circle ⟟ 8250192130 ⟟ Call Me For Gen...
Top Rated Bangalore Call Girls Richmond Circle ⟟ 8250192130 ⟟ Call Me For Gen...
 
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.Artifacts in Nuclear Medicine with Identifying and resolving artifacts.
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.
 
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...
 
VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...
VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...
VIP Mumbai Call Girls Hiranandani Gardens Just Call 9920874524 with A/C Room ...
 
(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...
(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...
(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...
 
Call Girls Darjeeling Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Darjeeling Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Darjeeling Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Darjeeling Just Call 9907093804 Top Class Call Girl Service Available
 
Call Girl Number in Vashi Mumbai📲 9833363713 💞 Full Night Enjoy
Call Girl Number in Vashi Mumbai📲 9833363713 💞 Full Night EnjoyCall Girl Number in Vashi Mumbai📲 9833363713 💞 Full Night Enjoy
Call Girl Number in Vashi Mumbai📲 9833363713 💞 Full Night Enjoy
 
Call Girls Siliguri Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Siliguri Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Siliguri Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Siliguri Just Call 9907093804 Top Class Call Girl Service Available
 
Call Girls Bhubaneswar Just Call 9907093804 Top Class Call Girl Service Avail...
Call Girls Bhubaneswar Just Call 9907093804 Top Class Call Girl Service Avail...Call Girls Bhubaneswar Just Call 9907093804 Top Class Call Girl Service Avail...
Call Girls Bhubaneswar Just Call 9907093804 Top Class Call Girl Service Avail...
 

IORT_ in vivo dosimetry

  • 1. Radiochromic films and Monte Carlo simulations to analyze abnormal Mosfet readings in IOERT breast cancer treatments G.Taccini1, F. Cavagnetto1, L.Ferri1, R.Bampi1, M.Guenzi2, G.Iaccario3, A. Soriani3 1 AOU IRCCS San Martino IST, Medical Physics, Genoa, Italy 2 AOU IRCCS San Martino IST,Radiation Oncology, Genoa, Italy 3 Istituto Ragina Elena, IFO, Rome, Italy
  • 2. IOeRT to treat early-stage breast cancer Dedicated mobile accelerator: LIAC Sordina, Italy  Electron beams: 4, 6 ,8, 10 MeV calibrated with Fricke dosimeters and two plane parallel ion chamber comparison*  PMMA cylindrical applicators : 4-10 cm ø for hard docking beam collimation  Steel-PTFE shielding disk (3mm+3mm) ; placed between the deep face of the residual breast and the pectoralis muscle to minimize thoracic wall irradiation  Every day dosimetry check to calculate optimized MU  In vivo dosimetry to check and properly correct MU Rosi A, Viti V. «Guidelines for quality assurance on ontra-operative radiation therapy» ISTISAN 2003 Report No. 03/1 EN Cavagnetto F. et al In vivo dosimetry in IORT cancer treatment. 2011; 99 Cavagnetto F. Et al A comparison between several dosimetric methods in IORT. ESTRO 31 PO-0835
  • 3. Between September 2009 and April 2013: 512 patients. Eligibility criteria for IORT treatment evaluated at two different decisional steps: ―> first step decision after diagnosis ―> second step decision intrasurgery Full dose intraoperativly irradiation if: tumor diameter less than 2,5 cm, infiltrating ductal histology, no mammographic evidence of multifocality, negative resection margins, no more than 3 axillary nodes, no extensive intraductal component.  21 Gy alone for high local relapse risk (pT2-pT1c)  18 Gy alone for low local relapse risk (max ø 1cm, pT1a-pT1b)  10 Gy as boost to the tumor bed if not eligible to full dose irradiation  16 Gy to nipple area complex if nipple sparing mastectomy Guenzi M. et al A two step selection of breast cancer patients candidates for exclusive IORT with electrons: a mono-institutional experience. Anticancer Res. (2012) 32(4):1533-6
  • 4. In vivo dosimetry to monitor exit dose : 293 patients  MicroMOSFET inside a steril and thin catheter is fastened to the PTFE face of the shielding disk –> dose at the deeper part of the target is checked.  Treatment splitted in two parts to correct MU according to detector half treatment reading.  Dose prescribed at the deeper part of the target and normalized at 90%: prescribed dose plus 10% is acceptable otherwise MU are properly correct. MicroMOSFET TN-502RDM (Best Medical Canadian):  Immediate readout  very small active volume 7” x 6.25” x 1.63”  Calibrated for each avaiable LIAC energy Ciocca M et al Real-time in vivo dosimetry using micro-MOSFET detectors during intraoperative electron beam radiation therapy in early-stage breast cancer. Radiother Oncol 2006; 78:213-6. Agostinelli et al. On line optimization of intraoperative electron beam radiotherapy of the breast. Radiother Oncol 2012; 103: 188-192.)
  • 5. 18% of MOSFET readings at half treatment > 10% Shielding disk backscattering radiation MOSFET directional radiation dependance Air gaps and non homogeneous tissue could create high dose hot spot in the irradiated glandular tissue due to electrons nature. Monte Carlo simulation -> no evidence of backscattering rad BOLUS AIR MOSFET TEFLON STEEL 0% 20% 40% 60% 80% 100% 120% 0,0 10,0 20,0 30,0 40,0 50,0 60,0 70,0 D% mm experimental set-up water Our hipothesis Monte Carlo simulations supplied by A Soriani, G.Iaccario
  • 6. 18% of MOSFET readings at half treatment > 10% Shielding disk backscattering MOSFET directional radiation dependance Air gaps and non homogeneous tissue tissue could create high dose hot spot in the irradiated glandular tissue due to electrons nature.  Micro-MOSFETs in slab water phantom under full build up conditions irradiated using electron beams supplied by LIAC.  Directional dependences of microMOSFET were found to be within ±2% between 0 and 45 degree LIAC angles. Chung J.,Suh T. et al Dosimetric Characteristics of Standard and Micro MOSFET Dosimeters as In-vivo Dosimeter for Clinical Electron Beam J. Korean Phys.Soc. 55,2566 Ciocca M. et al Real-time in vivo dosimetry using micro-MOSFET detectors during intraoperative electron beam radiation therapy in early-stage breast cancer. Radiother Oncol. 2006 Feb;78(2):213-6. Epub 2005 Dec 15 Our hipothesis
  • 7. 18% of MOSFET readings at half treatment > 10% Shielding disk backscattering MOSFET directional radiation dependance Air gaps and non homogeneous tissue tissue could create high dose hot spot in the irradiated glandular tissue due to electrons nature. Air gaps and non homogeneous tissue simulated with a filled water glove ; GAFCHROMIC EBT2 radiochromic film irradiation Our hipothesis
  • 8. Air gap between two “fingers” overdosages EBT2 C. Fiandra et al. “Absolute and relative dose measurements with Gafchromic™ EBT film for high energy electron beams with different doses per pulse” Med Phys, 35, 2008  Energy 10 MeV  Applicator 60 mm straight- end  500 cGy to R90 according to daily dosimetry check  Expected dose ≤ 556 cGy Epson Expression 10000XL to scan films  Picodose Pro® to extract red channel PTW VeriSoft® to determine isodoses and profiles.
  • 9. 7% 9% 11% 13% 15% 17% 19% 21% -10 -8 -6 -4 -2 0 2 4 6 8 10 DoseDifference% profile(mm) % Difference between measured and expected dose
  • 10. % Difference between measured and expected dose 7% 8% 9% 10% 11% 12% 13% 14% 15% 16% 17% -10 -8 -6 -4 -2 0 2 4 6 8 10 DoseDifference% Profile(mm) Key: gap size Our purpose: confirm overdosages with Monte Carlo simulations
  • 11. to see overdosages the key is : gap size Reproducible set up  Various heights: 0.5-1 cm  Various widths: 0.2-0.5 cm Results normilized to the expected dose at the appropriate depth.
  • 12. Monte Carlo simulation confirms our experimental results Peak Height: +13%EGS ncr/BEAMnrcode, 1mm simulation step, voxel dimension 1x 5mmx 29mm (GAF thickness) Monte Carlo simulations supplied by A Soriani, G.Iaccario
  • 13. Summary  September 2009 / april 2013: 512 patients affected by early-stage breast unifocal cancer treated with IOeRT .  In vivo dosimetry using micro-MOSFET performed on 293 patients.  Micro-MOSFET placed in the deeper part of the target to check the dose delivered during the first half of the treatment and eventually correct the second half.  Half treatment MOSFET readings, normalized to prescribed dose are higher than 10%in the 18% of cases .  Gafchromic EBT2 films under non homogeneous tissue simulated with a glove filled with water show overdoses at air-water interfaces.  Experimental set up with air gap of various size and EBT2 led us to evaluate the gap size effect.  Monte Carlo simulation confirm overdosages at air/water interfaces.  In vivo investigation with both micro MOSFET and GafCHROMIC F. Cavagnetto et al. In vivo dosimetry using MOSFET and radiochromic films in intraoperative radiotherapy for breast cancer treatments, OC-0487 2nd ESTRO Forum 2013