BASICS OF ULTRASOUND
 BY : DR. CHANDNIWADHWANI
Resident doctor
Department of Radiology
SSG Hospital Vadodara.
 References:
 Radiographics
 Springer
 “The fundamentals of x-ray and radium physics ”-by Joseph Selman
 Rumack
 Topics to be covered:
1. History of ultrasound
2. Principle
3. Ultrasound wave and its interactions
4. Ultrasound machine and its parts
5. Image display
6. Artifacts and their clinical importance
7. Doppler ultrasound
8. Elastography
9. Recent advances
10. Safety issues
Ultrasound:
 Sound waves with frequency greater than
range of human hearing.(>20,000Hz)
 Audible sound range is 20 to 20,000 Hz.
 The sound with frequency < 15 Hz is called
Infrasound.
 Ultrasound was first used for
clinical purposes in Glasgow in
1956.
 Obstretician Ian Donald and
engineer Tom Brown developed first
prototype systems based on an
instrument used to detect
industrial flaws in ships.
ULTRASOUND MACHINE
ECHOLOCATION
 Wavelength: is the distance between
sucessive wave crests.
 Frequency(f) : is the number of cycles per
second( 1 hertz= 1 cycle/second)
 Velocity (c)= wavelength x frequency
 Velocity of sound depends on density and
compressibility of the medium.
Sound waves travel faster in solids and slower in gases.
Higher in bone and metal and lower in lungs and air
PRINCIPLE OF ULTRASOUND
 Ultrasound works on the principle of
piezoelectric effect.
 When a crystal is subjected to mechanical
pressure, electric voltage is generated and
vice versa.
Ultrasound tissue interaction
1. Reflection
2. Refraction
3. Absorption
4. Attenuation
5. Scattering
REFLECTION
 A reflection of a beam is called ECHO.
 The production and detection of echoes
forms the basis of ultrasound.
 Reflection occurs at the interface between
two materials.
 It depends on the property of material-
“ACOUSTIC IMPEDENCE”
 If two materials have same impedence , no
echo produced.
 If the difference in acoustic impedence is:
 Small –weak echo is produced and most of
the sound waves will continue in second
medium
 Large- strong echo is produced
 Very large- all sound waves will be totally
reflected back. Example: tissue-air interface
99% of beam is reflected back.
SCATTERING
 Not all echoes are reflected back to probe.
 Some of them are scattered in all directions
in a non uniform manner.
 More so with very small objects or rough
surfaces.
 Part of scattering goes back to transducer
and generate images is called
BACKSCATTER.
ULTRASOUND TISSUE INTERACTION
Ultrasound machine:
ULTRASOUND MACHINE
TRANSDUCER:
 It is a device that converts energy from one
form to another.
 ULTRASOUNDTRANSDUCER converts
electric energy into sound energy and sound
energy back into electric energy.
TRANSDUCER DESIGN
1. Matching layer
2. Piezoelectric crystal
3. Backing block
4. Acoustic absorber
5. Metal shield
6. Signal cable
MATCHING LAYER
 It minimizes the acoustic impedence
differences between transducer and the
patient.
 Its impedence is intermediate to that of the
soft tissue and the transducer.
 Its thickness is equal to one-forth of the
wavelength, which is known as quarter wave
matching
 Matching layer is made of perspex or
plexiglass loaded with aluminium powder.
CRYSTAL LAYER
 Molecules of piezoelectric crystal are
polarized, one end is positive and other
negative.
 When high frequency current is applied, it
alternatively thickens and thins in its short
axis, and generates ultrasound waves as a
beam in air infront and back of the crystal
face.
DAMPING BLOCK
 Located on the backside of the crystal , made
up of tungsten particles suspended in epoxy
resin
 It absorbs backward US pulse and attenuates
stray US signals.
 Transducer and damping block are separated
from the casing by an insulator(rubber cork).
Function of damping block:
In B- Mode operation,
 It must stop the vibration within a
microsecond so that the transducer becomes
ready to immediately receive the reflected
echoes from the body
ULTRASOUND BEAM
Ultrasound beam characters:
 An unfocused ultrasound beam leaving a flat
crystal has 2 parts:
1. Initial cylindrical segment(near field or
frensnal zone)
2. Diverging conical portion ( far field or
fraunhofer zone)
 The length of near field and divergence of the
far field depend upon:
A. FREQUENCY: higher the frequency longer
the near fiels and less divergent the far field
 Depth resolution increases with higher
frequencies
B. CRYSTAL DIAMETER: increasing diameter
increases the near field length but worsens
the lateral and depth resolution.
TIME GAIN COMPENSATION
 TGC amplifies the signal proportional to the
time delay between transmission and
detection of US pulses.
 It amplifies and brings the signal in the range
of 40- 50 dB.
 This process compensates for tissue
attenuation and makes all equally reflective
boundaries equal in amplitude irrespective of
depth.
-
APPLICATIONS OF A-MODE:
 Opthalmology-distance measurements
 Echoencephalography
 Echocardiography
 Detecting a cyst in breast
 Studying midline displacement in brain
DOPPLER mode
DOPPLER EFFECT:
 The increasing frequency of a fire engine
siren as it approaches and its decreasing
frequency as it recedes , is known as “doppler
effect”.
 It is caused by the compression of sound
waves ahead of the siren and rarefaction of
sound waves behind the siren.
 This thought made Austrian physicist John
Doppler to discover the doppler effect.
Doppler shift:
Aliasing:
 Artifacts are the errors in images produced by
physical processes that affect ultrasound
beam.
 They are potential pitfalls that might confuse
the examiner.
 Some artifacts provide useful information for
novel interpretation.
REVERBERATION
RING DOWN ARTIFACTS
ACOUSTIC SHADOWING
 Tissues deeper to strongly attenuating
objects like calcification, appear darker
because the intensity of transmitted beam is
lower.
 Example:
 Strong after shadowing due to gall stones.
 Rib shadow
ENHANCEMENT
 Seen as abnormally high brightness.
 Occurs when sound travels through a
medium with attenuation rate lower than
surrounding tissue.
 Example:
 Enhancement of tissues below cyst or ducts.
 Tissues deeper to gall and urinary bladder.
Harmonic imaging:
 Technique of US which provides images of
better quality compared to conventional
imaging.
 Advantages:
 Reduces reverberation side lobe artefacts.
 Increased axial and lateral resolution
 Increased signal to noise ratio
 Improved resolution in patients with large
body habitus.
RESOLUTION
 It is the ability to appreciate two closely
placed objects as separate.
 3 types:
1. Axial
2. Lateral
3. temporal
 AXIAL : Ability to resolve objects in the line
of beam. Factors affecting axial resolution
include SPL(spatial pulse length) and
frequency
Lateral resolution
 Resolution at 90 degree to the direction of
beam.
 Factors affecting
 Width of the beam
 Distance from the transducer
 Frequency
 Side and grating lobe levels
Temporal resolution
 Ability to detect moving objects in the field of
view in their true sequence.
 It is determined by:
 Frame rate(number of frames generated per
unit second)
ELASTOGRAPHY
 Also called ‘sonoelastography’(palpation
imaging) is a non invasive technique to
depuct relative stiffness or
displacement(strain) in response to imparted
force.
 Basis of elastography is analogous to manual
palpation.
 Stiff tissue deform less and exhibit less strain
than complaint tissue when same force is
applied.
Used to evaluate tissues:
 Breast
 Lymph nodes
 Prostate
 Liver
 Blood vessels
 Thyroid
 Musculoskeletal structures
 Kidney transplant monitoring
 Cardiac
 Deep vein thrombosis
TECHNIQUES OF ELASTOGRAPHY
1. Quasi-static elastography
2. Dynamic elastography
STRAIN ELASTOGRAPHY
 Uses a uniform compression(applied by the user)
at the surface of body to cause deformation of
tissue.
 Scanner calculates and displays the induced
deformation.
 Limitations:
 Operator dependent
 Limited to superficial organs such as breast ,
thyroid
 Absence of quantification.
DYNAMIC ELASTOGRAPHY
1. Transient elastography
2. ARFI ( acoustic radiation force imaging)
technique
3. Supersonic shear wave elastography
TRANSIENT ELASTOGRAPHY
 Probe (3.5 MHz) contains a vibrator and an
US transducer.
 Principle: mechanical pulse induced at skin
surface by vibrator generates a transient
wave that propagates longitudinally.
 velocity and amplitude of shear wave are
measured.
 This velocity is converted to kPa, and reflects
tissue stiffness.
FIBROSCAN(Transient hepatic
elastography)
 Fibroscan device works by measuring shear
wave velocity.
 A 50 MHz wave is passed into liver from a
small transducer.
 The shear wave velocity can be converted
into liver stiffness which is expressed in kilo
pascals.
Beyond staging of liver
fibrosis;
 Fibroscan also used to evaluate patients with
portal hypertension
 Assess recurrence of disease after liver
transplantation
 Predict survival by evaluation of liver texture
 Also role in evaluation of breast cancer,
prostate cancer, and diseases in which
fibrosis(desmoplastic reaction) play a crucial
role.
 Advantages:
 Gold standard to stage fibrosis
 Non invasive as compared to biopsy
 Easy to use
 Quantification of tissue elasticity
 Rapid , painless
 Good reproducibility
 Limitations:
 Difficult in obese and ascites
 Left sided liver cannot be examined
ACOUSTIC RADIATION FORCE
IMAGING (ARFI)
 Uses a focused ultrasound pulse.
 Provides a estimate of stiffness of deep
tissues, non accessible by external
compression.
 Advantages:
 Less user dependent
 Better resolution than strain elastography
 Better transfer of shear modulus contrast to
image contrast
ULTRAFAST SHEAR WAVE
ELASTOGRAPHY(SWE)
 This technique developed by supersonic
imaging allows generation of multiple shear
waves along same longitudinal axis leading to
propagation of plane shear wave.
 It allows measurement of velocity of this
plane at each point of image in real time.
 Advantages:
 Displayed in real time like conventional US
image.
 Good reproducibility
 Quantitative value of stiffness
 Short acquisition time(~30 ms)
 Limitations:
 Expensive
 Requires a complex software
PRACTICAL APPLICATIONS:
 In breast tissues:
 Helps in characterization of benign/malignant
 Characterization of micro calcification
 Elastography of lymph nodes(metastatic
nodes appear stiffer and larger )
 Monitoring treatment response to neo-
adjuvant chemotherapy
In liver:
 Assessment of fibrosis(increased stiffness)
 Prediction of cirrhosis related
complications(correlation between FibroScan
values and development of esophageal varices)
 Assessment of response to anti-viral
therapy(stiffness falls with response and increases
with relapse)
 Characterization of liver tumors
 Biopsy site from stiffest region
 Much larger volume of liver assessed than biopsy
 Among benign tumors, hepatic adenoma has
greatest stiffness.
 Hemangioma – less shear stiffness than
fibrotic liver
 Cholangiocarcinoma-most stiff
 HCC less stiffer than cholangiocarcinoma.
In thyroid:
 Main indication of elastography in thyroid
disease is:
 Nodule characterization-
Example: papillary cancers are hard, follicular
cancers do not increase stiffness.
 Detect metastatic nodes.
In Prostate:
 Normal: elasticity values below 30 kPa.
 BPH: central and transition zone becomes
hard and stiff, give increased values of
elasticity
 Carcinoma: peripheral zone, elasticity
values>35 kPa
 Therefore 35 kPa is considered cut off value
to d/d benign and malignant prostatic lesions.
Elastography for staging
prostatic cancer
 Allows excellent visualisation of prostatic
capsule as soft rim artifact.
In renal application:
 Assessment of fibrosis
 Characterization of tumors
AIUM statement:
 “No confirmed biological effects on patients
or operators caused by exposure at intensities
typical for diagnostic ultrasound..
 Current data indicate that the
benefits…outweigh the risks.”
Summary:
 Ultrasound> 20,000Hz
 Piezoelectric effect= pulse-echo principle
 Frequency and wavelength are inversely
proportional
 Attenuation and frequency are inversely related.
 Broad bandwidth enables multiHertz probr
 Resolution determines image clarity
 Electronic arrays may be linear or sector
 Shadowing and enhancement are the most useful
artifacts in diagnosis
 Diagnostic ultrasound is safe!
basicsofultrasound-170807183128 (1).pdf

Más contenido relacionado

Similar a basicsofultrasound-170807183128 (1).pdf

Ultrasound diagnostics fin
Ultrasound diagnostics finUltrasound diagnostics fin
Ultrasound diagnostics fin
MUBOSScz
 
Basic Physics Of Transoesophageal Echocardiography For The Workshop2
Basic Physics Of Transoesophageal Echocardiography   For The Workshop2Basic Physics Of Transoesophageal Echocardiography   For The Workshop2
Basic Physics Of Transoesophageal Echocardiography For The Workshop2
Anil Ramaiah
 
Emergency us physics and instrumentation
Emergency us physics  and instrumentationEmergency us physics  and instrumentation
Emergency us physics and instrumentation
drnikahmad
 

Similar a basicsofultrasound-170807183128 (1).pdf (20)

USphysics.pptx
USphysics.pptxUSphysics.pptx
USphysics.pptx
 
Sonography
SonographySonography
Sonography
 
Usg Basics
Usg BasicsUsg Basics
Usg Basics
 
ULTRASOUND
ULTRASOUND ULTRASOUND
ULTRASOUND
 
Physic Of Ultrasound
Physic Of UltrasoundPhysic Of Ultrasound
Physic Of Ultrasound
 
Ocular Ultrasonography
Ocular UltrasonographyOcular Ultrasonography
Ocular Ultrasonography
 
Ultrasound and Doppler physics
Ultrasound and Doppler physicsUltrasound and Doppler physics
Ultrasound and Doppler physics
 
Ultrasound diagnostics fin
Ultrasound diagnostics finUltrasound diagnostics fin
Ultrasound diagnostics fin
 
Ophthalmic ultrasonography
Ophthalmic ultrasonographyOphthalmic ultrasonography
Ophthalmic ultrasonography
 
Physics of echo i.tammi raju
Physics of echo i.tammi rajuPhysics of echo i.tammi raju
Physics of echo i.tammi raju
 
Paper
PaperPaper
Paper
 
Ultrasound Imaging_2023.pdf
Ultrasound Imaging_2023.pdfUltrasound Imaging_2023.pdf
Ultrasound Imaging_2023.pdf
 
ULTRASOUND IMAGING PRINCIPLES
ULTRASOUND IMAGING PRINCIPLESULTRASOUND IMAGING PRINCIPLES
ULTRASOUND IMAGING PRINCIPLES
 
Basic Physics Of Transoesophageal Echocardiography For The Workshop2
Basic Physics Of Transoesophageal Echocardiography   For The Workshop2Basic Physics Of Transoesophageal Echocardiography   For The Workshop2
Basic Physics Of Transoesophageal Echocardiography For The Workshop2
 
Usg applications in anaesthesia dr gs
Usg applications in anaesthesia dr gsUsg applications in anaesthesia dr gs
Usg applications in anaesthesia dr gs
 
Ultrasonography final
Ultrasonography finalUltrasonography final
Ultrasonography final
 
Ultrasound Physics Jr.pptx
Ultrasound Physics  Jr.pptxUltrasound Physics  Jr.pptx
Ultrasound Physics Jr.pptx
 
Emergency us physics and instrumentation
Emergency us physics  and instrumentationEmergency us physics  and instrumentation
Emergency us physics and instrumentation
 
Its is useful for studying the basic concepts of ultrasound for medical imagi...
Its is useful for studying the basic concepts of ultrasound for medical imagi...Its is useful for studying the basic concepts of ultrasound for medical imagi...
Its is useful for studying the basic concepts of ultrasound for medical imagi...
 
PPT 1 Introduction to ultrasound.pptx
PPT 1 Introduction to ultrasound.pptxPPT 1 Introduction to ultrasound.pptx
PPT 1 Introduction to ultrasound.pptx
 

Más de littlealphonsa

peripheralarterialdoppler-150215054032-conversion-gate01.pdf
peripheralarterialdoppler-150215054032-conversion-gate01.pdfperipheralarterialdoppler-150215054032-conversion-gate01.pdf
peripheralarterialdoppler-150215054032-conversion-gate01.pdf
littlealphonsa
 
basicechocardiography-170913182344 (2).pdf
basicechocardiography-170913182344 (2).pdfbasicechocardiography-170913182344 (2).pdf
basicechocardiography-170913182344 (2).pdf
littlealphonsa
 
usgphysics-151102111226-lva1-app6892.pdf
usgphysics-151102111226-lva1-app6892.pdfusgphysics-151102111226-lva1-app6892.pdf
usgphysics-151102111226-lva1-app6892.pdf
littlealphonsa
 

Más de littlealphonsa (6)

peripheralarterialdoppler-150215054032-conversion-gate01.pdf
peripheralarterialdoppler-150215054032-conversion-gate01.pdfperipheralarterialdoppler-150215054032-conversion-gate01.pdf
peripheralarterialdoppler-150215054032-conversion-gate01.pdf
 
basicechocardiography-170913182344 (2).pdf
basicechocardiography-170913182344 (2).pdfbasicechocardiography-170913182344 (2).pdf
basicechocardiography-170913182344 (2).pdf
 
dopplerphysicsnani-170418192156.pdf
dopplerphysicsnani-170418192156.pdfdopplerphysicsnani-170418192156.pdf
dopplerphysicsnani-170418192156.pdf
 
usgphysics-151102111226-lva1-app6892.pdf
usgphysics-151102111226-lva1-app6892.pdfusgphysics-151102111226-lva1-app6892.pdf
usgphysics-151102111226-lva1-app6892.pdf
 
7453654.ppt
7453654.ppt7453654.ppt
7453654.ppt
 
Jagdish Presentation1.pptx
Jagdish Presentation1.pptxJagdish Presentation1.pptx
Jagdish Presentation1.pptx
 

Último

Spellings Wk 3 English CAPS CARES Please Practise
Spellings Wk 3 English CAPS CARES Please PractiseSpellings Wk 3 English CAPS CARES Please Practise
Spellings Wk 3 English CAPS CARES Please Practise
AnaAcapella
 
1029 - Danh muc Sach Giao Khoa 10 . pdf
1029 -  Danh muc Sach Giao Khoa 10 . pdf1029 -  Danh muc Sach Giao Khoa 10 . pdf
1029 - Danh muc Sach Giao Khoa 10 . pdf
QucHHunhnh
 

Último (20)

SOC 101 Demonstration of Learning Presentation
SOC 101 Demonstration of Learning PresentationSOC 101 Demonstration of Learning Presentation
SOC 101 Demonstration of Learning Presentation
 
Spatium Project Simulation student brief
Spatium Project Simulation student briefSpatium Project Simulation student brief
Spatium Project Simulation student brief
 
Micro-Scholarship, What it is, How can it help me.pdf
Micro-Scholarship, What it is, How can it help me.pdfMicro-Scholarship, What it is, How can it help me.pdf
Micro-Scholarship, What it is, How can it help me.pdf
 
TỔNG ÔN TẬP THI VÀO LỚP 10 MÔN TIẾNG ANH NĂM HỌC 2023 - 2024 CÓ ĐÁP ÁN (NGỮ Â...
TỔNG ÔN TẬP THI VÀO LỚP 10 MÔN TIẾNG ANH NĂM HỌC 2023 - 2024 CÓ ĐÁP ÁN (NGỮ Â...TỔNG ÔN TẬP THI VÀO LỚP 10 MÔN TIẾNG ANH NĂM HỌC 2023 - 2024 CÓ ĐÁP ÁN (NGỮ Â...
TỔNG ÔN TẬP THI VÀO LỚP 10 MÔN TIẾNG ANH NĂM HỌC 2023 - 2024 CÓ ĐÁP ÁN (NGỮ Â...
 
FSB Advising Checklist - Orientation 2024
FSB Advising Checklist - Orientation 2024FSB Advising Checklist - Orientation 2024
FSB Advising Checklist - Orientation 2024
 
Google Gemini An AI Revolution in Education.pptx
Google Gemini An AI Revolution in Education.pptxGoogle Gemini An AI Revolution in Education.pptx
Google Gemini An AI Revolution in Education.pptx
 
SKILL OF INTRODUCING THE LESSON MICRO SKILLS.pptx
SKILL OF INTRODUCING THE LESSON MICRO SKILLS.pptxSKILL OF INTRODUCING THE LESSON MICRO SKILLS.pptx
SKILL OF INTRODUCING THE LESSON MICRO SKILLS.pptx
 
Python Notes for mca i year students osmania university.docx
Python Notes for mca i year students osmania university.docxPython Notes for mca i year students osmania university.docx
Python Notes for mca i year students osmania university.docx
 
Basic Civil Engineering first year Notes- Chapter 4 Building.pptx
Basic Civil Engineering first year Notes- Chapter 4 Building.pptxBasic Civil Engineering first year Notes- Chapter 4 Building.pptx
Basic Civil Engineering first year Notes- Chapter 4 Building.pptx
 
Spellings Wk 3 English CAPS CARES Please Practise
Spellings Wk 3 English CAPS CARES Please PractiseSpellings Wk 3 English CAPS CARES Please Practise
Spellings Wk 3 English CAPS CARES Please Practise
 
Food safety_Challenges food safety laboratories_.pdf
Food safety_Challenges food safety laboratories_.pdfFood safety_Challenges food safety laboratories_.pdf
Food safety_Challenges food safety laboratories_.pdf
 
Explore beautiful and ugly buildings. Mathematics helps us create beautiful d...
Explore beautiful and ugly buildings. Mathematics helps us create beautiful d...Explore beautiful and ugly buildings. Mathematics helps us create beautiful d...
Explore beautiful and ugly buildings. Mathematics helps us create beautiful d...
 
ICT role in 21st century education and it's challenges.
ICT role in 21st century education and it's challenges.ICT role in 21st century education and it's challenges.
ICT role in 21st century education and it's challenges.
 
Kodo Millet PPT made by Ghanshyam bairwa college of Agriculture kumher bhara...
Kodo Millet  PPT made by Ghanshyam bairwa college of Agriculture kumher bhara...Kodo Millet  PPT made by Ghanshyam bairwa college of Agriculture kumher bhara...
Kodo Millet PPT made by Ghanshyam bairwa college of Agriculture kumher bhara...
 
How to Create and Manage Wizard in Odoo 17
How to Create and Manage Wizard in Odoo 17How to Create and Manage Wizard in Odoo 17
How to Create and Manage Wizard in Odoo 17
 
Accessible Digital Futures project (20/03/2024)
Accessible Digital Futures project (20/03/2024)Accessible Digital Futures project (20/03/2024)
Accessible Digital Futures project (20/03/2024)
 
Making communications land - Are they received and understood as intended? we...
Making communications land - Are they received and understood as intended? we...Making communications land - Are they received and understood as intended? we...
Making communications land - Are they received and understood as intended? we...
 
1029 - Danh muc Sach Giao Khoa 10 . pdf
1029 -  Danh muc Sach Giao Khoa 10 . pdf1029 -  Danh muc Sach Giao Khoa 10 . pdf
1029 - Danh muc Sach Giao Khoa 10 . pdf
 
General Principles of Intellectual Property: Concepts of Intellectual Proper...
General Principles of Intellectual Property: Concepts of Intellectual  Proper...General Principles of Intellectual Property: Concepts of Intellectual  Proper...
General Principles of Intellectual Property: Concepts of Intellectual Proper...
 
Introduction to Nonprofit Accounting: The Basics
Introduction to Nonprofit Accounting: The BasicsIntroduction to Nonprofit Accounting: The Basics
Introduction to Nonprofit Accounting: The Basics
 

basicsofultrasound-170807183128 (1).pdf

  • 1. BASICS OF ULTRASOUND  BY : DR. CHANDNIWADHWANI Resident doctor Department of Radiology SSG Hospital Vadodara.  References:  Radiographics  Springer  “The fundamentals of x-ray and radium physics ”-by Joseph Selman  Rumack
  • 2.  Topics to be covered: 1. History of ultrasound 2. Principle 3. Ultrasound wave and its interactions 4. Ultrasound machine and its parts 5. Image display 6. Artifacts and their clinical importance 7. Doppler ultrasound 8. Elastography 9. Recent advances 10. Safety issues
  • 3. Ultrasound:  Sound waves with frequency greater than range of human hearing.(>20,000Hz)  Audible sound range is 20 to 20,000 Hz.  The sound with frequency < 15 Hz is called Infrasound.
  • 4.
  • 5.
  • 6.
  • 7.  Ultrasound was first used for clinical purposes in Glasgow in 1956.  Obstretician Ian Donald and engineer Tom Brown developed first prototype systems based on an instrument used to detect industrial flaws in ships.
  • 8.
  • 9.
  • 10.
  • 12.
  • 13.
  • 14.
  • 15.
  • 16.
  • 18.
  • 19.
  • 20.
  • 21.  Wavelength: is the distance between sucessive wave crests.  Frequency(f) : is the number of cycles per second( 1 hertz= 1 cycle/second)  Velocity (c)= wavelength x frequency  Velocity of sound depends on density and compressibility of the medium.
  • 22. Sound waves travel faster in solids and slower in gases. Higher in bone and metal and lower in lungs and air
  • 23. PRINCIPLE OF ULTRASOUND  Ultrasound works on the principle of piezoelectric effect.  When a crystal is subjected to mechanical pressure, electric voltage is generated and vice versa.
  • 24.
  • 25. Ultrasound tissue interaction 1. Reflection 2. Refraction 3. Absorption 4. Attenuation 5. Scattering
  • 26. REFLECTION  A reflection of a beam is called ECHO.  The production and detection of echoes forms the basis of ultrasound.  Reflection occurs at the interface between two materials.  It depends on the property of material- “ACOUSTIC IMPEDENCE”  If two materials have same impedence , no echo produced.
  • 27.  If the difference in acoustic impedence is:  Small –weak echo is produced and most of the sound waves will continue in second medium  Large- strong echo is produced  Very large- all sound waves will be totally reflected back. Example: tissue-air interface 99% of beam is reflected back.
  • 28.
  • 29.
  • 30.
  • 31.
  • 32.
  • 33.
  • 34. SCATTERING  Not all echoes are reflected back to probe.  Some of them are scattered in all directions in a non uniform manner.  More so with very small objects or rough surfaces.  Part of scattering goes back to transducer and generate images is called BACKSCATTER.
  • 35.
  • 36.
  • 40.
  • 41.
  • 42. TRANSDUCER:  It is a device that converts energy from one form to another.  ULTRASOUNDTRANSDUCER converts electric energy into sound energy and sound energy back into electric energy.
  • 43. TRANSDUCER DESIGN 1. Matching layer 2. Piezoelectric crystal 3. Backing block 4. Acoustic absorber 5. Metal shield 6. Signal cable
  • 44.
  • 45. MATCHING LAYER  It minimizes the acoustic impedence differences between transducer and the patient.  Its impedence is intermediate to that of the soft tissue and the transducer.  Its thickness is equal to one-forth of the wavelength, which is known as quarter wave matching  Matching layer is made of perspex or plexiglass loaded with aluminium powder.
  • 46.
  • 47. CRYSTAL LAYER  Molecules of piezoelectric crystal are polarized, one end is positive and other negative.  When high frequency current is applied, it alternatively thickens and thins in its short axis, and generates ultrasound waves as a beam in air infront and back of the crystal face.
  • 48. DAMPING BLOCK  Located on the backside of the crystal , made up of tungsten particles suspended in epoxy resin  It absorbs backward US pulse and attenuates stray US signals.  Transducer and damping block are separated from the casing by an insulator(rubber cork).
  • 49. Function of damping block: In B- Mode operation,  It must stop the vibration within a microsecond so that the transducer becomes ready to immediately receive the reflected echoes from the body
  • 50.
  • 52.
  • 53.
  • 54.
  • 55.
  • 56.
  • 57. Ultrasound beam characters:  An unfocused ultrasound beam leaving a flat crystal has 2 parts: 1. Initial cylindrical segment(near field or frensnal zone) 2. Diverging conical portion ( far field or fraunhofer zone)
  • 58.  The length of near field and divergence of the far field depend upon: A. FREQUENCY: higher the frequency longer the near fiels and less divergent the far field  Depth resolution increases with higher frequencies B. CRYSTAL DIAMETER: increasing diameter increases the near field length but worsens the lateral and depth resolution.
  • 59.
  • 60.
  • 61.
  • 62.
  • 63.
  • 64. TIME GAIN COMPENSATION  TGC amplifies the signal proportional to the time delay between transmission and detection of US pulses.  It amplifies and brings the signal in the range of 40- 50 dB.  This process compensates for tissue attenuation and makes all equally reflective boundaries equal in amplitude irrespective of depth.
  • 65.
  • 66.
  • 67. -
  • 68.
  • 69.
  • 70.
  • 71.
  • 72.
  • 73.
  • 74.
  • 75.
  • 76.
  • 77.
  • 78.
  • 79. APPLICATIONS OF A-MODE:  Opthalmology-distance measurements  Echoencephalography  Echocardiography  Detecting a cyst in breast  Studying midline displacement in brain
  • 80.
  • 81.
  • 82.
  • 83.
  • 84.
  • 85.
  • 87. DOPPLER EFFECT:  The increasing frequency of a fire engine siren as it approaches and its decreasing frequency as it recedes , is known as “doppler effect”.  It is caused by the compression of sound waves ahead of the siren and rarefaction of sound waves behind the siren.  This thought made Austrian physicist John Doppler to discover the doppler effect.
  • 90.
  • 91.  Artifacts are the errors in images produced by physical processes that affect ultrasound beam.  They are potential pitfalls that might confuse the examiner.  Some artifacts provide useful information for novel interpretation.
  • 94.
  • 95.
  • 96.
  • 97.
  • 98.
  • 99.
  • 100. ACOUSTIC SHADOWING  Tissues deeper to strongly attenuating objects like calcification, appear darker because the intensity of transmitted beam is lower.  Example:  Strong after shadowing due to gall stones.  Rib shadow
  • 101.
  • 102. ENHANCEMENT  Seen as abnormally high brightness.  Occurs when sound travels through a medium with attenuation rate lower than surrounding tissue.  Example:  Enhancement of tissues below cyst or ducts.  Tissues deeper to gall and urinary bladder.
  • 103.
  • 104.
  • 105.
  • 106. Harmonic imaging:  Technique of US which provides images of better quality compared to conventional imaging.  Advantages:  Reduces reverberation side lobe artefacts.  Increased axial and lateral resolution  Increased signal to noise ratio  Improved resolution in patients with large body habitus.
  • 107. RESOLUTION  It is the ability to appreciate two closely placed objects as separate.  3 types: 1. Axial 2. Lateral 3. temporal  AXIAL : Ability to resolve objects in the line of beam. Factors affecting axial resolution include SPL(spatial pulse length) and frequency
  • 108. Lateral resolution  Resolution at 90 degree to the direction of beam.  Factors affecting  Width of the beam  Distance from the transducer  Frequency  Side and grating lobe levels
  • 109. Temporal resolution  Ability to detect moving objects in the field of view in their true sequence.  It is determined by:  Frame rate(number of frames generated per unit second)
  • 110. ELASTOGRAPHY  Also called ‘sonoelastography’(palpation imaging) is a non invasive technique to depuct relative stiffness or displacement(strain) in response to imparted force.  Basis of elastography is analogous to manual palpation.  Stiff tissue deform less and exhibit less strain than complaint tissue when same force is applied.
  • 111. Used to evaluate tissues:  Breast  Lymph nodes  Prostate  Liver  Blood vessels  Thyroid  Musculoskeletal structures  Kidney transplant monitoring  Cardiac  Deep vein thrombosis
  • 112. TECHNIQUES OF ELASTOGRAPHY 1. Quasi-static elastography 2. Dynamic elastography
  • 113. STRAIN ELASTOGRAPHY  Uses a uniform compression(applied by the user) at the surface of body to cause deformation of tissue.  Scanner calculates and displays the induced deformation.  Limitations:  Operator dependent  Limited to superficial organs such as breast , thyroid  Absence of quantification.
  • 114.
  • 115.
  • 116. DYNAMIC ELASTOGRAPHY 1. Transient elastography 2. ARFI ( acoustic radiation force imaging) technique 3. Supersonic shear wave elastography
  • 117. TRANSIENT ELASTOGRAPHY  Probe (3.5 MHz) contains a vibrator and an US transducer.  Principle: mechanical pulse induced at skin surface by vibrator generates a transient wave that propagates longitudinally.  velocity and amplitude of shear wave are measured.  This velocity is converted to kPa, and reflects tissue stiffness.
  • 118.
  • 119. FIBROSCAN(Transient hepatic elastography)  Fibroscan device works by measuring shear wave velocity.  A 50 MHz wave is passed into liver from a small transducer.  The shear wave velocity can be converted into liver stiffness which is expressed in kilo pascals.
  • 120. Beyond staging of liver fibrosis;  Fibroscan also used to evaluate patients with portal hypertension  Assess recurrence of disease after liver transplantation  Predict survival by evaluation of liver texture  Also role in evaluation of breast cancer, prostate cancer, and diseases in which fibrosis(desmoplastic reaction) play a crucial role.
  • 121.  Advantages:  Gold standard to stage fibrosis  Non invasive as compared to biopsy  Easy to use  Quantification of tissue elasticity  Rapid , painless  Good reproducibility  Limitations:  Difficult in obese and ascites  Left sided liver cannot be examined
  • 122. ACOUSTIC RADIATION FORCE IMAGING (ARFI)  Uses a focused ultrasound pulse.  Provides a estimate of stiffness of deep tissues, non accessible by external compression.  Advantages:  Less user dependent  Better resolution than strain elastography  Better transfer of shear modulus contrast to image contrast
  • 123. ULTRAFAST SHEAR WAVE ELASTOGRAPHY(SWE)  This technique developed by supersonic imaging allows generation of multiple shear waves along same longitudinal axis leading to propagation of plane shear wave.  It allows measurement of velocity of this plane at each point of image in real time.
  • 124.
  • 125.
  • 126.  Advantages:  Displayed in real time like conventional US image.  Good reproducibility  Quantitative value of stiffness  Short acquisition time(~30 ms)  Limitations:  Expensive  Requires a complex software
  • 127.
  • 128.
  • 129. PRACTICAL APPLICATIONS:  In breast tissues:  Helps in characterization of benign/malignant  Characterization of micro calcification  Elastography of lymph nodes(metastatic nodes appear stiffer and larger )  Monitoring treatment response to neo- adjuvant chemotherapy
  • 130.
  • 131.
  • 132. In liver:  Assessment of fibrosis(increased stiffness)  Prediction of cirrhosis related complications(correlation between FibroScan values and development of esophageal varices)  Assessment of response to anti-viral therapy(stiffness falls with response and increases with relapse)  Characterization of liver tumors  Biopsy site from stiffest region  Much larger volume of liver assessed than biopsy
  • 133.  Among benign tumors, hepatic adenoma has greatest stiffness.  Hemangioma – less shear stiffness than fibrotic liver  Cholangiocarcinoma-most stiff  HCC less stiffer than cholangiocarcinoma.
  • 134.
  • 135. In thyroid:  Main indication of elastography in thyroid disease is:  Nodule characterization- Example: papillary cancers are hard, follicular cancers do not increase stiffness.  Detect metastatic nodes.
  • 136.
  • 137. In Prostate:  Normal: elasticity values below 30 kPa.  BPH: central and transition zone becomes hard and stiff, give increased values of elasticity  Carcinoma: peripheral zone, elasticity values>35 kPa  Therefore 35 kPa is considered cut off value to d/d benign and malignant prostatic lesions.
  • 138. Elastography for staging prostatic cancer  Allows excellent visualisation of prostatic capsule as soft rim artifact.
  • 139.
  • 140. In renal application:  Assessment of fibrosis  Characterization of tumors
  • 141.
  • 142.
  • 143. AIUM statement:  “No confirmed biological effects on patients or operators caused by exposure at intensities typical for diagnostic ultrasound..  Current data indicate that the benefits…outweigh the risks.”
  • 144. Summary:  Ultrasound> 20,000Hz  Piezoelectric effect= pulse-echo principle  Frequency and wavelength are inversely proportional  Attenuation and frequency are inversely related.  Broad bandwidth enables multiHertz probr  Resolution determines image clarity  Electronic arrays may be linear or sector  Shadowing and enhancement are the most useful artifacts in diagnosis  Diagnostic ultrasound is safe!