1. Quality control and applications of
Radiopharmaceuticals
Presented by
Presented by
Savita patil
Savita patil
M.Pharm IIIrd sem
M.Pharm IIIrd sem
(Pharmaceutics)
(Pharmaceutics)
Govt. college of pharmacy,
Govt. college of pharmacy,
Powerpoint Templates Aurangabad.
Aurangabad.
Page 1
3. Introduction
Radiopharmaceuticals are unique medicinal formulations
containing radioisotopes which are used in major clinical areas for
diagnosis and/or therapy.
A radiopharmaceutical contains a radioactive moiety with a
biodistribution moiety.
After administration radiophamaceutical localizes in a specific
organ/ tissue depending on biodistributing moiety.
Powerpoint Templates
Page 3
4. QC tests of
QC tests of
radiopharmaceuticals
radiopharmaceuticals
QC of radiopharmaceuticals includes:
The purity
Potency
Biologic safety
Sterility and
Efficacy of the radiopharmaceuticals.
The tests should be faster and effective since
radionuclide
have short half life.
Powerpoint Templates
Page 4
5. The QC tests for most of the radiopharmaceuticals are
mentioned in pharmacopoeia (EP, BP, etc)
The quality control system should include a procedure which
describes measures to be taken if unsatisfactory test results are
obtained.
Powerpoint Templates
Page 5
7. Parameters of each test
Parameters of each test
Powerpoint Templates
Page 7
8. A. Physical tests
A. Physical tests
1.pH and ionic strength:
Ideally pH should be 7.4, but it can vary from 2-9.
Correct ionic strength achieved by addition of an acid or
alkali.
2. Physical characteristics:
Recognize the color and state of a radiopharmaceuticals
Colloidal or aggregated preparations : Identify size by
microscope with haemocytometer.
Powerpoint Templates
Page 8
9. Importance of particle size determination
Importance of particle size determination
Size of particles will determine the site where the
radiopharmaceutical will get localized.
E.g. 99mTc MAA, Tc-labeled albumin microspheres : 10
to 100 µm
If>100µm: cause pulmonary embolism
If <10 µm: Deposits in the reticuloendothelial cells.
Powerpoint Templates
Page 9
10. B. Radiochemical tests
B. Radiochemical tests
1. Radionuclidic Purity:
% of the total radioactivity that is present in the
form of the stated specified radionuclide
Determined by measuring the half-lives and
characteristic radiations emitted by individual
radionuclides.
γ-ray emission radionuclides : checked with
multi-channel analyzer
Pure β emission radionuclides : checked with β-
spectrometer or a liquid scintillation counter
Powerpoint Templates
Page 10
11. 2. Radiochemical purity:
It is the proportion of the radionuclide present in the stated chemical form.
Radiochemical purity can be determined any separation technique, and
measurement of radioactivity.
% RCP= radioactivity in the desired chemical/ radioactivity of total
radiopharmaceutical x 100
Radiochemical impurities may arise from decomposition due to:
(1) action of solvent
(2) change in temperature or pH
(3) light
(4) presence of oxidizing or reducing agents
(5) radiolysis
Powerpoint Templates
Page 11
13. TLC of radiopharmaceuticals
Radiopharmaceutical Stationary phase Mobile phase
14
C-urea cellulose butanol-water-acetic acid (12:5:3)
123/131
I-hippuran Silica gel chloroform-acetic acid (9:1)
131
I-iodocholesterol Silica gel chloroform-ethanol (1:1)
111
In-octreotide Silica gel 0.1 M citrate buffer pH 5
123
I-iomazenil Silica gel ethyl acetate-ammonia
(200:1)
Powerpoint Templates
Page 13
14. Radioactivity
Necessary to determine since radioactivity will determine the radiation
dose to each patient.
It is expressed in SI unit Bequrrels (Bq) or Curie, a non SI unit.
Where 1 Bq=1 disintegration/sec
Activity determinations are carried out by isotope dose calibrator.
Calibration of dose calibrator Should be done by a reference source,
e.g. Co 57,Ra-226 or Cs-137
Powerpoint Templates
Page 14
16. 1. Geiger Muller counter
Principle: Property to cause ionization of a gas,e.g. Argon.
Beta particles enter the G-M tube through Mica window
Argon gas is ionized
Argon ions are attracted to negative electrode.
Process continues and net result is to ionize whole vol. of tube equivalent to
current
Amplification Digital/ audio output
Powerpoint Templates
Page 16
18. Conti……………….
Advantage:
It provides a high output
Disadvantage:
Efficiencies are low for low energy Beta emitters e.g. 3H.
Powerpoint Templates
Page 18
19. 2. Liquid scintillation counter
Principle: Property of radiation to cause fluorescence.
Advantage: can be used for low energy beta emitters, e.g. 3H and 14C.
Problem :
1. sample preparation: Requirement to have a clear solution or
dispersion in non aqueous solvent.
2. variable quenching: may reduce efficiency.
Powerpoint Templates
Page 19
20. Method
Sample is dissolved in solvent with a scintillant material, e.g. Diphenyloxazole
Ionizing radiation(B particles) transfers energy from solvent to scintillant
molecule
Molecules are excited to higher state and latter on fluoresce with emission of
light
Detected by photomultiplier
Powerpoint Templates
Page 20
22. Isotope dilution analysis:
• General two methods:
• 1.Direct method: Determination of an inactive compound by
dilution with a radioactive compound.
• 2.Reverse method: Involves the estimation of radionuclide by
dilution with an inactive nuclide
Powerpoint Templates
Page 22
23. C. Biological tests
1. Sterility:
1. Sterility:
Sterility: It is complete absence of live micro organisms and its spores.
Sterility: It is complete absence of live micro organisms and its spores.
Methods used:
Methods used:
Procedure: incubate radiopharmaceutical with growth medium
Procedure: incubate radiopharmaceutical with growth medium
containing 14C glucose.
containing 14C glucose.
Bacteria (if any) will metabolize glucose releasing 14CO2.2.
Bacteria (if any) will metabolize glucose releasing 14CO
Detection of amounts of radioactivity greater than background indicates
Detection of amounts of radioactivity greater than background indicates
bacterial growth.
bacterial growth.
Powerpoint Templates
Page 23
24. Cont…
2. Incubate the radiopharmaceutical sample in fluid thioglycolate
2. Incubate the radiopharmaceutical sample in fluid thioglycolate
medium at 30to 35ºC for 7 to 14 days For detection of anaerobic bacteria,
medium at 30to 35ºC for 7 to 14 days For detection of anaerobic bacteria,
or
or
3. Soyabean-casein digest medium for incubation at20 to 25ºC for 7 to 14
3. Soyabean-casein digest medium for incubation at20 to 25ºC for 7 to 14
days. For detection of fungi and aerobic bacteria.
days. For detection of fungi and aerobic bacteria.
And growth of micro organisms is seen.
And growth of micro organisms is seen.
Powerpoint Templates
Page 24
25. 2. Pyrogen testing
LAL(Limulus ambeocyte lysate) test can be used.
LAL (Limulus ambeocyte lysate) test can be used.
Methodinvolves incubation of product with lysate of circulating ambeocytes
Method involves incubation of product with lysate of circulating ambeocytes
of aahorseshoe crab (Limulus polyphenus) . .
of horseshoe crab (Limulus polyphenus)
Presence of pyrogens will cause the lysate to form the gel.
Presence of pyrogens will cause the lysate to form the gel.
Powerpoint Templates
Page 25
26. 3. Biodistribution
Importantto determine that the agent does localize in the organ of
Important to determine that the agent does localize in the organ of
interest.
interest.
Procedure:
Procedure:
Administration of radiopharmaceutical in animals
Administration of radiopharmaceutical in animals
Ascertain the relative quantity of radioactivity in various organs.
Ascertain the relative quantity of radioactivity in various organs.
Powerpoint Templates
Page 26
27. Other QC tests
Apart from above mentioned tests the final product container should be
checked for proper labeling prior to release of product.
Written procedure should be for dealing with products failing to meet the
required standard.
investigation measures taken to prevent future events
documented.
Release can only be effected if:
The product complies with the specifications
The product has been prepared according to Good Radiopharmacy Practice
Powerpoint Templates
Page 27
28. Labelling :
• the name of the product and the name of the radionuclide;
• any product identification code;
• the name of the manufacturer;
• an identification number (batch number);
The label on the • for liquid preparations, the total radioactivity in the container, or the
outer package radioactive
• concentration per millilitre, at a stated date and, if necessary, hour,
should include : and the volume of liquid in container;
• for solid preparations, such as freeze-dried preparations, the total
radioactivity at a
• stated date and, if necessary, hour;
• for capsules, the radioactivity of each capsule at a stated date and, if
necessary, hour and the number of capsules in the container;
• where relevant, the international symbol for radioactivity.
Powerpoint Templates
Page 28
31. Therapeutic Applicatios
Of
Radiopharmaceuticals
Radiolabelled Molecules Disease
Chromic Phosphate P32 For Lung, Ovarian, Uterine, And
Prostate Cancers
Sodium Iodide I 131 Thyroid Cancer
Samarium Sm 153 Cancerous Bone Tissue
Sodium Phosphate P 32 Cancerous Bone Tissue And Other
Types Of Cancers
Strontium Chloride Sr 89 Cancerous Bone Tissue
Powerpoint Templates
Page 31