FDA standards for IVDs. Why does FDA need them?

                       Mehis Pold, M.D.
                            July 9, 2010


What is a standard?

Put at its simplest, a standard is an agreed, repeatable way of
doing something. It is a published document that contains a
technical specification or other precise criteria designed to be
used consistently as a rule, guideline, or definition.

                                                                   1
The U.S. FDA – The United States Food and Drug
Administration

Title 21 CFR Part 820 – Title 21 of Code of Federal
Regulations, Part 820
• Title 21 describes the rules of FDA

• Part 820 contains the quality system regulation (QSR) for
medical devices including In Vitro Diagnostic Devices (IVDs).




                                                                2
Part 820 in a nut shell:
Design, production, and distribution of medical
devices including IVDs must be carried out in a
transparently controlled and documented
environment by an appropriately trained staff in
order to meet the established quality and
performance characteristics.




                                                   3
PART 820 QUALITY SYSTEM REGULATION
Subpart A -- General Provisions
Subpart B -- Quality System Requirements
Subpart C -- Design Controls
Subpart D --Document Controls
Subpart E -- Purchasing Controls
Subpart F -- Identification and Traceability
Subpart G -- Production and Process Controls equipment.
Subpart H -- Acceptance Activities
Subpart I-- Nonconforming Product
Subpart J -- Corrective and Preventive Action
Subpart K -- Labeling and Packaging Control
Subpart L -- Handling, Storage, Distribution, and Installation
Subpart M -- Records
Subpart N -- Servicing
Subpart O -- Statistical Techniques



                                                                 4
HISTORY of IVD STANDARDS:
The Clinical Laboratory Improvement Amendments of
1988: IVDs needed regulations to unify their performance & quality
characteristics in order to better serve the public.
What is and IVD standard? IVD standard is a published document
that contains a technical specification or other precise criteria designed
to be used consistently as a rule, guideline, or definition. They are
created by bringing together the experience and expertise of all
interested parties including the manufacturers, users, and regulators of
IVDs.

Definition of IVD: IVDs are products that employ reagents,
instruments, and systems intended for use in diagnosis of disease or
other conditions, including a determination of the state of health, in
order to cure, mitigate, treat, or prevent disease or its sequelae. Such
products are intended for use in the collection, preparation, and
examination of specimens taken from the human body.
                                                                             5
What happens if you developed a novel technology for
      which FDA simply does not have rules?
                         Are you doomed?
                          Certainly not!
                                 (1) Time
                               (2) Patience
                               (3) Dialogue
Pre-IDE: The "pre-IDE" process is an informal "pre-submission" process. It may involve:
     sending analytical or clinical protocols to FDA for review and comment before
   proceeding with studies, or meeting with FDA to discuss protocols and/or possible
regulatory pathways. The term "pre-IDE" has been used for this process so that FDA can
assign it an official tracking number, just as numbers are assigned to 510(k)s and PMAs.
  This does not mean that manufacturers are required to subsequently submit an IDE
                                         application.
                                                                                     6
R&D                                     Manufacturing                        FDA regulatory filing




   Stable prototype = Fully                  Full Part 820-compliance
   established performance                   possible only if fully established
   characteristics                           performance characteristics exist.
                                             Otherwise there is nothing to
                                             compare the QC results to!




Evaluation of Precision Performance of Quantitative Measurement Methods           EP05-A2
Evaluation of the Linearity of Quantitative Measurement                           EP06-A
Interference Testing in Clinical Chemistry                                        EP07-A2
Method Comparison and Bias Estimation Using Patient Samples                       EP09-A2
User Protocol for Evaluation of Qualitative Test Performance                      EP12-A2
Evaluation of Matrix Effects                                                      EP14-A2


                                                                                                  7
FDA values tradition and clarity. Therefore, Title 21
     CFR 820 compliance must support clarity.
Without the IVD standards, FDA would not know how to properly go about
 the approval process. The IVD standards equip not only the FDA but also
  the inventors and manufacturers with useful technical information that
promotes the quality of clinical tests. Without the IVD standards, we would
    be back to the pre-CLIA 88-period that witnessed the deficiencies in
           laboratory testing due to the lack of clear standards.




                                                                              8
Title 21 CFR 820 CONCLUSIONS:
Title 21 CFR 820-compliance is essential for
    successful FDA filing.

Full Title 21 CFR 820-compliace is enabled by
    complete R&D based on IVD standards
    agreed upon with FDA during pre-IDE.

Full 820-compliance is facilitated by pre-IDE
    process. Especially when IVD stems from
    novel technology.

The U.S. FDA is a partner for IVD companies.
                                                9

Más contenido relacionado

La actualidad más candente

Testo 184 - GxP Dictionary - English
Testo 184 - GxP Dictionary - EnglishTesto 184 - GxP Dictionary - English
Testo 184 - GxP Dictionary - EnglishLothar Zorn
 
Validation of sterile Medical Devices manufacturing processes By J. Havel - H...
Validation of sterile Medical Devices manufacturing processes By J. Havel - H...Validation of sterile Medical Devices manufacturing processes By J. Havel - H...
Validation of sterile Medical Devices manufacturing processes By J. Havel - H...qserveconference2013
 
NABL Certification
NABL CertificationNABL Certification
NABL Certificationb0458
 
medical devices and invitro diagnosis by rahul sagar, m. pharm(dra), bbau luc...
medical devices and invitro diagnosis by rahul sagar, m. pharm(dra), bbau luc...medical devices and invitro diagnosis by rahul sagar, m. pharm(dra), bbau luc...
medical devices and invitro diagnosis by rahul sagar, m. pharm(dra), bbau luc...Brajesh Kumar
 
CONCEPT OF URS, DQ, IQ, OQ, PQ
CONCEPT OF URS, DQ, IQ, OQ, PQCONCEPT OF URS, DQ, IQ, OQ, PQ
CONCEPT OF URS, DQ, IQ, OQ, PQROHIT
 
Basis Validation of Biomanufacturing Processes
Basis Validation of Biomanufacturing ProcessesBasis Validation of Biomanufacturing Processes
Basis Validation of Biomanufacturing Processesangelsalaman
 
1 02 (1)
1 02 (1)1 02 (1)
1 02 (1)jntryn
 
Quali Meddweb
Quali MeddwebQuali Meddweb
Quali MeddwebQualiMedd
 
Scale up and post approval changes(supac)
Scale up and post approval changes(supac)Scale up and post approval changes(supac)
Scale up and post approval changes(supac)bdvfgbdhg
 
Supac - Guidance for Modified Release Dosage Form
Supac - Guidance for Modified Release Dosage FormSupac - Guidance for Modified Release Dosage Form
Supac - Guidance for Modified Release Dosage FormJubiliant Generics Limited
 
Revalidation presentation
Revalidation presentationRevalidation presentation
Revalidation presentationDenny Walthers
 
Scale – Up & Post Approval Changes
Scale – Up & Post Approval ChangesScale – Up & Post Approval Changes
Scale – Up & Post Approval ChangesVAIBHAV SHENDE
 

La actualidad más candente (20)

Gtp Guidance09
Gtp Guidance09Gtp Guidance09
Gtp Guidance09
 
Testo 184 - GxP Dictionary - English
Testo 184 - GxP Dictionary - EnglishTesto 184 - GxP Dictionary - English
Testo 184 - GxP Dictionary - English
 
Validation of sterile Medical Devices manufacturing processes By J. Havel - H...
Validation of sterile Medical Devices manufacturing processes By J. Havel - H...Validation of sterile Medical Devices manufacturing processes By J. Havel - H...
Validation of sterile Medical Devices manufacturing processes By J. Havel - H...
 
NABL Certification
NABL CertificationNABL Certification
NABL Certification
 
medical devices and invitro diagnosis by rahul sagar, m. pharm(dra), bbau luc...
medical devices and invitro diagnosis by rahul sagar, m. pharm(dra), bbau luc...medical devices and invitro diagnosis by rahul sagar, m. pharm(dra), bbau luc...
medical devices and invitro diagnosis by rahul sagar, m. pharm(dra), bbau luc...
 
CONCEPT OF URS, DQ, IQ, OQ, PQ
CONCEPT OF URS, DQ, IQ, OQ, PQCONCEPT OF URS, DQ, IQ, OQ, PQ
CONCEPT OF URS, DQ, IQ, OQ, PQ
 
Basis Validation of Biomanufacturing Processes
Basis Validation of Biomanufacturing ProcessesBasis Validation of Biomanufacturing Processes
Basis Validation of Biomanufacturing Processes
 
About Compliance Insight Inc.
About Compliance Insight Inc.About Compliance Insight Inc.
About Compliance Insight Inc.
 
Stability Dosage-Development-Presentation
Stability Dosage-Development-PresentationStability Dosage-Development-Presentation
Stability Dosage-Development-Presentation
 
2 the regulations
2 the regulations2 the regulations
2 the regulations
 
1 02 (1)
1 02 (1)1 02 (1)
1 02 (1)
 
aaaQA_Dir. SteveCVQ216
aaaQA_Dir. SteveCVQ216aaaQA_Dir. SteveCVQ216
aaaQA_Dir. SteveCVQ216
 
Quali Meddweb
Quali MeddwebQuali Meddweb
Quali Meddweb
 
Qualification of Gas Chromatography
Qualification of Gas Chromatography Qualification of Gas Chromatography
Qualification of Gas Chromatography
 
Scale up and post approval changes(supac)
Scale up and post approval changes(supac)Scale up and post approval changes(supac)
Scale up and post approval changes(supac)
 
Supac
SupacSupac
Supac
 
Supac - Guidance for Modified Release Dosage Form
Supac - Guidance for Modified Release Dosage FormSupac - Guidance for Modified Release Dosage Form
Supac - Guidance for Modified Release Dosage Form
 
Revalidation presentation
Revalidation presentationRevalidation presentation
Revalidation presentation
 
Validation of water supply system
Validation of water supply systemValidation of water supply system
Validation of water supply system
 
Scale – Up & Post Approval Changes
Scale – Up & Post Approval ChangesScale – Up & Post Approval Changes
Scale – Up & Post Approval Changes
 

Similar a Why Does FDA Need Standards For In Vitro Diagnostic Devices

CFR Part 820 Transition Update.pdf
CFR Part 820 Transition Update.pdfCFR Part 820 Transition Update.pdf
CFR Part 820 Transition Update.pdfMinShikar
 
(IDE)and(IVD),QMS,21 CFR part820 , 801)
(IDE)and(IVD),QMS,21 CFR part820  , 801)(IDE)and(IVD),QMS,21 CFR part820  , 801)
(IDE)and(IVD),QMS,21 CFR part820 , 801)chahattyagi200
 
Marlene Garcia Swider ASQ, PDA, SoCalBio Presentation Oct. 6
Marlene Garcia Swider ASQ, PDA, SoCalBio Presentation Oct. 6Marlene Garcia Swider ASQ, PDA, SoCalBio Presentation Oct. 6
Marlene Garcia Swider ASQ, PDA, SoCalBio Presentation Oct. 6Rx Research Services CRO
 
Topic - (IDE)and(IVD),21CFR part 820,801
Topic - (IDE)and(IVD),21CFR part 820,801Topic - (IDE)and(IVD),21CFR part 820,801
Topic - (IDE)and(IVD),21CFR part 820,801chahattyagi200
 
Medical Device Development Firm Glossary of terms and abbreviations.pdf
Medical Device Development Firm Glossary of terms and abbreviations.pdfMedical Device Development Firm Glossary of terms and abbreviations.pdf
Medical Device Development Firm Glossary of terms and abbreviations.pdfAlexander Sprauve
 
FDA Regulations and Medical Device Pathways to Market
FDA Regulations and Medical Device Pathways to MarketFDA Regulations and Medical Device Pathways to Market
FDA Regulations and Medical Device Pathways to MarketMethodSense, Inc.
 
FDA Regulations and Medical Device Pathways to Market
FDA Regulations and Medical Device Pathways to MarketFDA Regulations and Medical Device Pathways to Market
FDA Regulations and Medical Device Pathways to MarketMethodSense, Inc.
 
004 301 training quality system management regulations, gmp (2)
004 301 training quality system management regulations, gmp (2)004 301 training quality system management regulations, gmp (2)
004 301 training quality system management regulations, gmp (2)ZARRAR MEHMOOD
 
medical regulation.pdf
medical regulation.pdfmedical regulation.pdf
medical regulation.pdfssuser0e9b33
 
Ruud Cox - Testing in a Medical Device Context - EuroSTAR 2012
Ruud Cox - Testing in a Medical Device Context - EuroSTAR 2012Ruud Cox - Testing in a Medical Device Context - EuroSTAR 2012
Ruud Cox - Testing in a Medical Device Context - EuroSTAR 2012TEST Huddle
 
Handling deviations & unexpected results during method validation
Handling deviations & unexpected results during method validationHandling deviations & unexpected results during method validation
Handling deviations & unexpected results during method validationInstitute of Validation Technology
 
Quality System Requirements 21 CFR Part 820 and Labelling Requirements for Me...
Quality System Requirements 21 CFR Part 820 and Labelling Requirements for Me...Quality System Requirements 21 CFR Part 820 and Labelling Requirements for Me...
Quality System Requirements 21 CFR Part 820 and Labelling Requirements for Me...Swapnil Fernandes
 
validation and verification of medical device.pptx
validation and verification of medical device.pptxvalidation and verification of medical device.pptx
validation and verification of medical device.pptxGargiVaghela
 
Developing Product Requirements For Medical Devices
Developing Product Requirements For Medical DevicesDeveloping Product Requirements For Medical Devices
Developing Product Requirements For Medical DevicesWalt Maclay
 

Similar a Why Does FDA Need Standards For In Vitro Diagnostic Devices (20)

CFR Part 820 Transition Update.pdf
CFR Part 820 Transition Update.pdfCFR Part 820 Transition Update.pdf
CFR Part 820 Transition Update.pdf
 
(IDE)and(IVD),QMS,21 CFR part820 , 801)
(IDE)and(IVD),QMS,21 CFR part820  , 801)(IDE)and(IVD),QMS,21 CFR part820  , 801)
(IDE)and(IVD),QMS,21 CFR part820 , 801)
 
Marlene Garcia Swider ASQ, PDA, SoCalBio Presentation Oct. 6
Marlene Garcia Swider ASQ, PDA, SoCalBio Presentation Oct. 6Marlene Garcia Swider ASQ, PDA, SoCalBio Presentation Oct. 6
Marlene Garcia Swider ASQ, PDA, SoCalBio Presentation Oct. 6
 
Topic - (IDE)and(IVD),21CFR part 820,801
Topic - (IDE)and(IVD),21CFR part 820,801Topic - (IDE)and(IVD),21CFR part 820,801
Topic - (IDE)and(IVD),21CFR part 820,801
 
Medical Device Development Firm Glossary of terms and abbreviations.pdf
Medical Device Development Firm Glossary of terms and abbreviations.pdfMedical Device Development Firm Glossary of terms and abbreviations.pdf
Medical Device Development Firm Glossary of terms and abbreviations.pdf
 
FDA Regulations and Medical Device Pathways to Market
FDA Regulations and Medical Device Pathways to MarketFDA Regulations and Medical Device Pathways to Market
FDA Regulations and Medical Device Pathways to Market
 
FDA Regulations and Medical Device Pathways to Market
FDA Regulations and Medical Device Pathways to MarketFDA Regulations and Medical Device Pathways to Market
FDA Regulations and Medical Device Pathways to Market
 
004 301 training quality system management regulations, gmp (2)
004 301 training quality system management regulations, gmp (2)004 301 training quality system management regulations, gmp (2)
004 301 training quality system management regulations, gmp (2)
 
TwinSPIN_Lecture.ppt
TwinSPIN_Lecture.pptTwinSPIN_Lecture.ppt
TwinSPIN_Lecture.ppt
 
Quality Systems
Quality SystemsQuality Systems
Quality Systems
 
medical regulation.pdf
medical regulation.pdfmedical regulation.pdf
medical regulation.pdf
 
Ruud Cox - Testing in a Medical Device Context - EuroSTAR 2012
Ruud Cox - Testing in a Medical Device Context - EuroSTAR 2012Ruud Cox - Testing in a Medical Device Context - EuroSTAR 2012
Ruud Cox - Testing in a Medical Device Context - EuroSTAR 2012
 
Medical devices CHINA
Medical devices CHINA Medical devices CHINA
Medical devices CHINA
 
Handling deviations & unexpected results during method validation
Handling deviations & unexpected results during method validationHandling deviations & unexpected results during method validation
Handling deviations & unexpected results during method validation
 
Quality System Requirements 21 CFR Part 820 and Labelling Requirements for Me...
Quality System Requirements 21 CFR Part 820 and Labelling Requirements for Me...Quality System Requirements 21 CFR Part 820 and Labelling Requirements for Me...
Quality System Requirements 21 CFR Part 820 and Labelling Requirements for Me...
 
Medical Device
Medical DeviceMedical Device
Medical Device
 
Validation_Tutorial.ppt
Validation_Tutorial.pptValidation_Tutorial.ppt
Validation_Tutorial.ppt
 
Validation
ValidationValidation
Validation
 
validation and verification of medical device.pptx
validation and verification of medical device.pptxvalidation and verification of medical device.pptx
validation and verification of medical device.pptx
 
Developing Product Requirements For Medical Devices
Developing Product Requirements For Medical DevicesDeveloping Product Requirements For Medical Devices
Developing Product Requirements For Medical Devices
 

Más de Mehis Pold

Rare events or not i want to know about them
Rare events or not i want to know about themRare events or not i want to know about them
Rare events or not i want to know about themMehis Pold
 
COSMIC ALK-FUSIONS
COSMIC ALK-FUSIONSCOSMIC ALK-FUSIONS
COSMIC ALK-FUSIONSMehis Pold
 
Hidden value in medical genetics databases. Splice the silence!
Hidden value in medical genetics databases. Splice the silence!Hidden value in medical genetics databases. Splice the silence!
Hidden value in medical genetics databases. Splice the silence!Mehis Pold
 
Why do the silent mutations matter?
Why do the silent mutations matter?Why do the silent mutations matter?
Why do the silent mutations matter?Mehis Pold
 
Development of multivariate classifiers in cancer
Development of multivariate classifiers in cancerDevelopment of multivariate classifiers in cancer
Development of multivariate classifiers in cancerMehis Pold
 
Art Of Breast Cancer
Art Of Breast CancerArt Of Breast Cancer
Art Of Breast CancerMehis Pold
 
Endometriosis gene-expression, meta-analysis
Endometriosis gene-expression, meta-analysisEndometriosis gene-expression, meta-analysis
Endometriosis gene-expression, meta-analysisMehis Pold
 

Más de Mehis Pold (7)

Rare events or not i want to know about them
Rare events or not i want to know about themRare events or not i want to know about them
Rare events or not i want to know about them
 
COSMIC ALK-FUSIONS
COSMIC ALK-FUSIONSCOSMIC ALK-FUSIONS
COSMIC ALK-FUSIONS
 
Hidden value in medical genetics databases. Splice the silence!
Hidden value in medical genetics databases. Splice the silence!Hidden value in medical genetics databases. Splice the silence!
Hidden value in medical genetics databases. Splice the silence!
 
Why do the silent mutations matter?
Why do the silent mutations matter?Why do the silent mutations matter?
Why do the silent mutations matter?
 
Development of multivariate classifiers in cancer
Development of multivariate classifiers in cancerDevelopment of multivariate classifiers in cancer
Development of multivariate classifiers in cancer
 
Art Of Breast Cancer
Art Of Breast CancerArt Of Breast Cancer
Art Of Breast Cancer
 
Endometriosis gene-expression, meta-analysis
Endometriosis gene-expression, meta-analysisEndometriosis gene-expression, meta-analysis
Endometriosis gene-expression, meta-analysis
 

Why Does FDA Need Standards For In Vitro Diagnostic Devices

  • 1. FDA standards for IVDs. Why does FDA need them? Mehis Pold, M.D. July 9, 2010 What is a standard? Put at its simplest, a standard is an agreed, repeatable way of doing something. It is a published document that contains a technical specification or other precise criteria designed to be used consistently as a rule, guideline, or definition. 1
  • 2. The U.S. FDA – The United States Food and Drug Administration Title 21 CFR Part 820 – Title 21 of Code of Federal Regulations, Part 820 • Title 21 describes the rules of FDA • Part 820 contains the quality system regulation (QSR) for medical devices including In Vitro Diagnostic Devices (IVDs). 2
  • 3. Part 820 in a nut shell: Design, production, and distribution of medical devices including IVDs must be carried out in a transparently controlled and documented environment by an appropriately trained staff in order to meet the established quality and performance characteristics. 3
  • 4. PART 820 QUALITY SYSTEM REGULATION Subpart A -- General Provisions Subpart B -- Quality System Requirements Subpart C -- Design Controls Subpart D --Document Controls Subpart E -- Purchasing Controls Subpart F -- Identification and Traceability Subpart G -- Production and Process Controls equipment. Subpart H -- Acceptance Activities Subpart I-- Nonconforming Product Subpart J -- Corrective and Preventive Action Subpart K -- Labeling and Packaging Control Subpart L -- Handling, Storage, Distribution, and Installation Subpart M -- Records Subpart N -- Servicing Subpart O -- Statistical Techniques 4
  • 5. HISTORY of IVD STANDARDS: The Clinical Laboratory Improvement Amendments of 1988: IVDs needed regulations to unify their performance & quality characteristics in order to better serve the public. What is and IVD standard? IVD standard is a published document that contains a technical specification or other precise criteria designed to be used consistently as a rule, guideline, or definition. They are created by bringing together the experience and expertise of all interested parties including the manufacturers, users, and regulators of IVDs. Definition of IVD: IVDs are products that employ reagents, instruments, and systems intended for use in diagnosis of disease or other conditions, including a determination of the state of health, in order to cure, mitigate, treat, or prevent disease or its sequelae. Such products are intended for use in the collection, preparation, and examination of specimens taken from the human body. 5
  • 6. What happens if you developed a novel technology for which FDA simply does not have rules? Are you doomed? Certainly not! (1) Time (2) Patience (3) Dialogue Pre-IDE: The "pre-IDE" process is an informal "pre-submission" process. It may involve: sending analytical or clinical protocols to FDA for review and comment before proceeding with studies, or meeting with FDA to discuss protocols and/or possible regulatory pathways. The term "pre-IDE" has been used for this process so that FDA can assign it an official tracking number, just as numbers are assigned to 510(k)s and PMAs. This does not mean that manufacturers are required to subsequently submit an IDE application. 6
  • 7. R&D Manufacturing FDA regulatory filing Stable prototype = Fully Full Part 820-compliance established performance possible only if fully established characteristics performance characteristics exist. Otherwise there is nothing to compare the QC results to! Evaluation of Precision Performance of Quantitative Measurement Methods EP05-A2 Evaluation of the Linearity of Quantitative Measurement EP06-A Interference Testing in Clinical Chemistry EP07-A2 Method Comparison and Bias Estimation Using Patient Samples EP09-A2 User Protocol for Evaluation of Qualitative Test Performance EP12-A2 Evaluation of Matrix Effects EP14-A2 7
  • 8. FDA values tradition and clarity. Therefore, Title 21 CFR 820 compliance must support clarity. Without the IVD standards, FDA would not know how to properly go about the approval process. The IVD standards equip not only the FDA but also the inventors and manufacturers with useful technical information that promotes the quality of clinical tests. Without the IVD standards, we would be back to the pre-CLIA 88-period that witnessed the deficiencies in laboratory testing due to the lack of clear standards. 8
  • 9. Title 21 CFR 820 CONCLUSIONS: Title 21 CFR 820-compliance is essential for successful FDA filing. Full Title 21 CFR 820-compliace is enabled by complete R&D based on IVD standards agreed upon with FDA during pre-IDE. Full 820-compliance is facilitated by pre-IDE process. Especially when IVD stems from novel technology. The U.S. FDA is a partner for IVD companies. 9