Neuromuscular Monitoring
Dr. Mohtasib
• 1958, Christie and Churchill-Davidson described how nerve
stimulators could be used to assess neuromuscular function
objectively during anesthesia
• In Awake patients, muscle power can be evaluated through tests of
voluntary muscle strength
• During anesthesia and recovery from anesthesia this is not possible.
Instead, the clinician uses clinical tests to assess muscle power
directly and to estimate neuromuscular function indirectly (muscle
tone, the feel of the anesthesia bag, an indirect measure of
pulmonary compliance, tidal volume, and inspiratory force).
• Onset of NM Blockade.
• To determine level of muscle relaxation during surgery.
• Assessing patients recovery from blockade to minimize risk
of residual paralysis.
Objectives of NM Monitoring
Why do we Monitor?
Residual post-op NM Blockade
• Functional impairment of pharyngeal and upper
esophageal muscles
• Impaired ability to maintain the airway
• Increased risk for post-op pulmonary complications
• Difficult to exclude clinically significant residual
curarization by clinical evaluation
Principles of Peripheral Nerve Stimulation
• Each muscle fiber to a stimulus follows an all-or-none pattern
• Response of the whole muscle depends on the number of muscle
fibers activated
• Response of the muscle decreases in parallel with the numbers of
fibers blocked
• Reduction in response during constant stimulation reflects degree of
NM Blockade
• For this reason stimulus is supramaximal
Essential features of PNS
• Shape of stimulus should be monophasic and rectangular i.e
Square-wave stimulus.
• 0.2- 0.3 msec duration so it falls within absolute refractory period
of motor unit in the nerve.
• Constant current variable voltage
• Battery powered.
• Digital display of delivered current.
• Audible signal on delivery of stimulus.
• Audible alarm for poor electrode contact.
• Multiple patterns of stimulation (single twitch,train-of-four,
double-burst, post-tetanic count).
• Current intensity : It is the amperage (mA) of the current delivered by the nerve
stimulator(0-80 mA). The intensity reaching the nerve is determined by the voltage
generated by the stimulator and resistance and impedance of the electrodes, skin and
underlying tissues
• Nerve stimulators are constant current and variable voltage delivery devices
• Reduction of temperature increases the tissue resistance (increased impedance) and may
cause reduction in the current delivered to fall below the supramaximal level
• Threshold current : It is the lowest current required to depolarize the most sensitive fibres
in a given nerve bundle to elicit a detectable muscle response.
• Maximal current: Current which generate response in all muscle fibre
• Supramaximal current : It is approximately 25% higher intensity than the current required
to depolarize all fibres in a particular nerve bundle. This is generally attained at current
intensity 2-3 times higher than threshold current.
• Submaximal current : A current intensity that induces firing of only a fraction fibres in a
given nerve bundle. A potential advantage of submaximal current is that it is less painful
than supramaximal current.
• Stimulus frequency : The rate (Hz) at which each impulse is repeated in cycles per second
(Hz).
Types of Peripheral Nerve Stimulation
• Neuromuscular function is monitored by evaluating the muscular response
to supramaximal stimulation of a peripheral motor nerve.
• Two types of stimulation can be used: electrical and magnetic.
• Magnetic nerve stimulation has several advantages over electrical nerve
stimulation.It is less painful and does not require physical contact with the
body.
• However, the equipment required is bulky and heavy, it cannot be used for
train-of-four (TOF) stimulation, and it is difficult to achieve supramaximal
stimulation with this method.
Neuromuscular Monitoring is Essential
• After prolonged infusions of neuromuscular blocking drugs or when long-acting
drugs are used
• When surgery or anaesthesia is prolonged
• When inadequate reversal may have devastating effects, for example, severe
respiratory disease, morbid obesity
• In conditions where administration of a reversal agent may cause harm, for
example, tachyarrhythmias, cardiac failure
• Liver or renal dysfunction, when pharmacokinetics of muscular relaxants may be
altered
• Neuromuscular disorders such as myasthenia gravis or Eaton–Lambert syndrome
• Surgeries requiring profound neuromuscular blockade e.g. neurosurgery, vascular
surgery in vital areas like thoracic cavity.
Patterns of Nerve Stimulation
For evaluation of neuromuscular function five patterns of nerve stimulation are
commonly used in clinical practice
• Single twitch stimulation.
• Train-of-four stimulation.
• Tetanic stimulation.
• Post-tetanic count stimulation.
• Double burst stimulation.
Single-Twitch Stimulation
• A Single supramaximal electrical stimuli are applied to a peripheral motor nerve
at frequencies ranging from 1.0 Hz (once every second) to 0.1 Hz (once every 10
seconds) .
• The response to single-twitch stimulation depends on the frequency at which the
individual stimuli are applied.
• Because 1-Hz stimulation shortens the time necessary to determine
supramaximal stimulation, this frequency is sometimes used during induction of
anesthesia.
Train-of-Four Stimulation
• Four supramaximal stimuli are given every 0.5 second (2 Hz).
• When used continuously, each set (train) of stimuli is normally repeated every
10th to 20th second.
• Each stimulus in the train causes the muscle to contract, and “fade” in the
response provides the basis for evaluation.
• That is, dividing the amplitude of the fourth response by the amplitude of the
first response provides the TOF ratio.
• In the control response (the response obtained before the administration of a
muscle relaxant), all four responses are ideally the same: the TOF ratio is 1.0.
• During a partial nondepolarizing block, the ratio decreases (fades) and is
inversely proportional to the degree of blockade.
• During a partial depolarizing block, no fade occurs in the TOF response;
ideally, the TOF ratio is approximately 1.0.
• Fade in the TOF response after injection of succinylcholine signifies the
development of a phase II block.
• The advantages of TOF stimulation are greatest during nondepolarizing
blockade because the degree of block can be read directly from the TOF
response even though a preoperative value is lacking.
• In addition, TOF stimulation has some advantages over tetanic stimulation:
it is less painful and, unlike tetanic stimulation, does not generally affect
the degree of neuromuscular blockade.
Tetanic Stimulation
• Tetanic stimulation consists of very rapid (e.g., 30-, 50-, or 100-Hz) delivery of
electrical stimuli.
• The most commonly used pattern in clinical practice is 50-Hz stimulation given
for 5 seconds, although some investigators have advocated the use of 50-, 100-,
and even 200-Hz stimulation for 1 second.
• During normal neuromuscular transmission and a pure depolarizing block, the
muscle response to 50-Hz tetanic stimulation for 5 seconds is sustained.
• During a nondepolarizing block and a phase II block after the injection of
succinylcholine, the response will not be sustained (i.e., fade occurs)
• Fade in response to tetanic stimulation is normally considered a presynaptic
event; the traditional explanation is that at the start of tetanic stimulation, large
amounts of acetylcholine are released from immediately available stores in the
nerve terminal.
• As these stores become depleted, the rate of acetylcholine release decreases
until equilibrium between mobilization and synthesis of acetylcholine is
achieved.
• When the “margin of safety” at the postsynaptic membrane (i.e., the number of
free cholinergic receptors) is reduced by nondepolarizing neuromuscular
blocking drugs, a typical reduction in twitch height is seen with a fade during, for
instance, repetitive stimulation.
• In addition to this postsynaptic block, nondepolarizing neuromuscular blocking
drugs may also block presynaptic neuronal-type acetylcholine receptors, thereby
leading to impaired mobilization of acetylcholine within the nerve terminal.
• This effect substantially contributes to fade in the response to tetanic (and TOF)
stimulation.
• Although the degree of fade depends primarily on the degree of neuromuscular
blockade, fade also depends on the frequency (Hz) and the length (seconds) of
stimulation and on how often tetanic stimuli are applied.
• During partial nondepolarizing blockade, tetanic nerve stimulation is followed by
a post-tetanic increase in twitch tension (i.e., post-tetanic facilitation of
transmission).
• This event occurs because the increase in mobilization and synthesis of
acetylcholine caused by tetanic stimulation continues for some time after
discontinuation of stimulation.
• The degree and duration of post-tetanic facilitation depend on the
degree of neuromuscular blockade, with post-tetanic facilitation
usually disappearing within 60 seconds of tetanic stimulation.
• In contrast, post-tetanic twitch potentiation, which sometimes
occurs in mechanical recordings before any neuromuscular blocking
drug has been given, is a muscular phenomenon that is not
accompanied by an increase in the compound muscle action
potential.
• Tetanic stimulation is very painful and therefore not normally
acceptable to an unanesthetized patient.
Post-Tetanic Count Stimulation
• Injection of a nondepolarizing neuromuscular blocking drug in a
dose sufficient to ensure smooth tracheal intubation causes intense
neuromuscular blockade of the peripheral muscles.
• Because no response to TOF and single-twitch stimulation occurs
under these conditions, these modes of stimulation cannot be used
to determine the degree of blockade.
• It is possible, however, to quantify intense neuromuscular blockade
of the peripheral muscles by applying tetanic stimulation (50 Hz for
5 seconds) and observing the post-tetanic response to single-
twitch stimulation given at 1 Hz starting 3 seconds after the end of
tetanic stimulation.
• During intense blockade, there is no response to either tetanic or
post-tetanic stimulation.
• As the intense block dissipates, more and more responses to post-tetanic twitch
stimulation appear. For a given neuromuscular blocking drug, the time until return
of the first response to TOF stimulation is related to the number of post-tetanic
twitch responses present at a given time (i.e., the PTC)
• The PTC method is mainly used to assess the degree of neuromuscular blockade
when there is no reaction to single-twitch or TOF nerve stimulation, as may be the
case after injection of a large dose of a nondepolarizing neuromuscular blocking
drug.
• However, PTC can also be used whenever sudden movements must be eliminated
(e.g., during ophthalmic surgery).
• The necessary level of blockade of the adductor pollicis muscle to ensure paralysis
of the diaphragm depends on the type of anesthesia and, in the intensive care unit,
on the level of sedation.
• To ensure elimination of any bucking or coughing in response to tracheobronchial
stimulation, neuromuscular blockade of the peripheral muscles must be so intense
that no response to post-tetanic twitch stimulation can be elicited (PTC 0)
• The response to PTC stimulation depends primarily on the degree of
neuromuscular blockade.
• It also depends on the :-
1. frequency and duration of tetanic stimulation,
2. the length of time between the end of tetanic stimulation and the
first post-tetanic stimulus,
3. the frequency of the single-twitch stimulation,
4. the duration of single-twitch stimulation before tetanic
stimulation.
• When the PTC method is used, these variables should be kept
constant.
• In addition, because of possible antagonism of neuromuscular
blockade in the hand, tetanic stimulation should not be performed
more often than every 6 minutes.
• Pattern of electrical stimulation and evoked muscle responses to
• train-of-four (TOF) nerve stimulation,
• 50-Hz tetanic nerve stimulation for 5 seconds (TE), and
• 1.0-Hz post-tetanic twitch stimulation (PTS) during four different levels of nondepolarizing neuromuscular
blockade.
• [A] During intense blockade of peripheral muscles no response to any of the forms of stimulation occurs.
• [B/ C] During less pronounced blockade, there is still no response to TOF stimulation, but post-tetanic
facilitation of transmission is present.
• [D] During surgical blockade , the first response to TOF appears and post-tetanic facilitation increases further.
• The post-tetanic count is 1 during very deep block (B), 3 during less deep block (C), and 8 during surgical (or
moderate) block (D).
Double-Burst Stimulation
• DBS consists of two short bursts of 50-Hz tetanic stimulation
separated by 750 msec.
• The duration of each square wave impulse in the burst is 0.2 msec.
• Although the number of impulses in each burst can vary, DBS with
three impulses in each of the two tetanic bursts (DBS3,3) is most
commonly used.
• In nonparalyzed muscle, the response to DBS3,3 is two short muscle
contractions of equal strength.
• In a partly paralyzed muscle, the second response is weaker than the
first (i.e., the response fades).
• DBS was developed with the specific aim of allowing manual (tactile)
detection of small amounts of residual blockade under clinical
conditions,and during recovery and immediately after surgery, tactile
evaluation of the response to DBS3,3 is superior to tactile evaluation
of the response to TOF stimulation.
Equipments Required
• The nerve stimulator,
• The stimulating electrodes, and
• The recording equipment.
The Nerve Stimulator
It delivers the stimulus to the electrodes. Although many nerve
stimulators are commercially available, not all meet the basic
requirement for clinical use. The real nerve stimulator should have the
following properties.
• It should deliver monophasic and rectangular waveform pulse of
constant current.
• It should deliver a current up to 60-70 mA but not >80mA.
• It should be battery operated and should have a battery check.
• It should have either built in warning system or current level display
that alerts the user when the current selected is not delivered to the
nerve (in case of increased skin resistance)
• It should have polarity indicators for electrodes.
• It should deliver different modes of stimulation.
The Stimulation Electrodes
• Electrical impulses are transmitted from the stimulator to the nerve
by means of surface or needle electrodes. Surface electrodes are
commonly used in clinical anaesthesia. These are pregelled
silver/sliver chloride electrodes having approximately 7-8 mm
diameter of the conducting area.5
• If tissue resistance prevents the stimulating current from reaching the
nerve (i.e. in the morbidly obese patients), needle electrodes can be
used. A Current = 10mA will usually produce supramaximal
stimulation when subcutaneous needle electrodes are used. Although
specially coated needle electrodes are commercially available,
ordinary steel injection needles can be used.
Assessment of Responses to Nerve Stimulation
Five methods are available:
1. Measurement of the evoked mechanical response of the muscle
(mechanomyography [MMG]),
2. Measurement of the evoked electrical response of the muscle
(electromyography [EMG]),
3. Measurement of acceleration of the muscle response
(acceleromyography [AMG]),
4. Measurement of the evoked electrical response in a piezoelectric film
sensor attached to the muscle (piezoelectric neuromuscular monitor
[PZEMG]
5. Phonomyography [PMG]).
Mechanomyography
• The mechanomyogram (MMG) is the mechanical signal observable from the
surface of a muscle when the muscle is contracted.
• At the onset of muscle contraction, gross changes in the muscle shape cause a
large peak in the MMG.
• Subsequent vibrations are due to oscillations of the muscle fibres at the
resonance frequency of the muscle.
• A requirement for correct and reproducible measurement of evoked tension is
that the muscle contraction be isometric.
• In clinical anesthesia, this condition is most easily achieved by measuring thumb
movement after the application of a resting tension of 200 to 300 g (a preload) to
the thumb.
• When the ulnar nerve is stimulated, the thumb (the adductor pollicis muscle) acts
on a force-displacement transducer.
• The force of contraction is then converted into an electrical signal, which is
amplified, displayed, and recorded.
• The arm and hand should be rigidly fixed, and care should be taken to prevent
overloading of the transducer.
• In addition, the transducer should be placed in correct relation to the thumb
(i.e., the thumb should always apply tension precisely along the length of the
transducer).
• It is important to remember that the response to nerve stimulation depends on
the frequency with which the individual stimuli are applied and that the time
used to achieve a stable control response may influence subsequent
determination of the onset time and duration of blockade.
• Generally, the reaction to supramaximal stimulation increases during the first 8 to
12 minutes after commencement of the stimulation.
Electromyography
• (EMG) is a technique for evaluating and recording the electrical activity
produced by skeletal muscles
• Evoked EMG records the compound action potentials produced by stimulation
of a peripheral nerve. The compound action potential is a high-speed event that
for many years could be picked up only by means of a preamplifier and a
storage oscilloscope.
• The evoked EMG response is most often obtained from muscles innervated by
the ulnar or the median nerves.
• Most often, the evoked EMG response is obtained from the thenar or
hypothenar eminence of the hand or from the first dorsal interosseous muscle
of the hand, preferably with the active electrode over the motor point of the
muscle .
• The signal picked up by the analyzer is processed by an amplifier, a rectifier, and
an electronic integrator. The results are displayed either as a percentage of
control or as a TOF ratio.
• The evoked EMG response is most often obtained from muscles innervated by
the ulnar or the median nerves.
• Most often, the evoked EMG response is obtained from the thenar or hypothenar
eminence of the hand or from the first dorsal interosseous muscle of the hand,
preferably with the active electrode over the motor point of the muscle .
• The signal picked up by the analyzer is processed by an amplifier, a rectifier, and
an electronic integrator. The results are displayed either as a percentage of
control or as a TOF ratio.
• Electrode placement for
stimulation of the ulnar nerve
and for recording of the
compound action potential from
three sites of the hand.
• A, Abductor digiti minimi muscle
(in the hypothenar eminence).
• B, Adductor pollicis muscle (in
the thenar eminence).
• C, First dorsal interosseus
muscle.
• Two new sites for recording the EMG response have been introduced: the larynx
and the diaphragm.
• By using a noninvasive disposable laryngeal electrode attached to the tracheal
tube and placed between the vocal cords, it is possible to monitor the onset of
neuromuscular blockade in the laryngeal muscles.
• In paravertebral surface diaphragmatic EMG, the recording electrodes are placed
on the right of vertebrae T12/L1 o r L1/L2 for monitoring the response of the
right diaphragmatic crux to transcutaneous stimulation of the right phrenic nerve
at the neck.
• Evoked electrical and mechanical responses represent different physiologic
events. Evoked EMG records changes in the electrical activity of one or more
muscles, whereas evoked MMG records changes associated with excitation-
contraction coupling and contraction of the muscle as well.
ADVANTAGES
• Equipment for measuring evoked EMG responses is easier to set up,
• The response reflects only factors influencing neuromuscular transmission, and
• The response can be obtained from muscles not accessible to mechanical
recording.
DISADVANTAGES
• Although high-quality recordings are possible in most patients, the results are
not always reliable. For one thing, improper placement of electrodes may result
in inadequate pickup of the compound EMG signal.
• Direct muscle stimulation sometimes occurs. If muscles close to the stimulating
electrodes are stimulated directly, the recording electrodes may pick up an
electrical signal even though neuromuscular transmission is completely blocked.
• Another difficulty is that the EMG response often does not return to the control
value.
• Finally, the evoked EMG response is very sensitive to electrical interference,
such as that caused by diathermy.
Acceleromyograph
• Is a piezoelectric myograph, used to measure the force produced by a
muscle after it has undergone nerve stimulation.
• Acceleromyographs measure muscle activity using a miniature
piezoelectric transducer that is attached to the stimulated muscle.
• A voltage is created when the muscle accelerates and that acceleration
is proportion to force of contraction.
• Acceleromyographs are more costly than the more common twitch
monitors, but have been shown to better alleviate residual blockade and
associated symptoms of muscle weakness, and to improve overall
quality of recovery.
• When an accelerometer is fixed to
the thumb and the ulnar nerve is
stimulated, an electrical signal is
produced whenever the thumb
moves.
• This signal can be analyzed in a
specially designed analyzer.
• TOF-Watch (Organon, part of
Schering-Plough, Corp.). This
neuromuscular transmission monitor
is based on measurement of
acceleration with a piezoelectric
transducer.
• Transducer is fastened to the thumb
and the stimulating electrodes.
• On the display of the TOF-Watch, the
train-of-four (TOF) ratio is given in
percentage.
• Hand adaptor
(elastic preload)
for the TOF-
Watch
transducer
Piezoelectric Neuromuscular Monitors
• The technique of the piezoelectric
monitor is based on the principle that
stretching or bending a flexible
piezoelectric film (e.g., one attached
to the thumb) in response to nerve
stimulation generates a voltage that is
proportional to the amount of
stretching or bending.
• At least two devices
based on this principle
are available
commercially: the
ParaGraph
Neuromuscular
Blockade Monitor and
the M-NMT
MechanoSensor, which
is a part of the Datex
AS/3 monitoring system
(Datex-Ohmeda,
Helsinki, Finland)
Phonomyography
• Contraction of skeletal muscles
generates intrinsic low-frequency
sounds, which can be recorded with
special microphones.
• What does make PMG interesting,
however, is that in theory the method
can be applied not only to the
adductor pollicis muscle but also to
other muscles of interest such as the
diaphragm, larynx, and eye muscles.
• In addition, the ease of application is
attractive.
• Ulnar nerve: MOST COMMON SITE
• place negative electrode (black) on
wrist in line with the smallest digit 1-
2cm below skin crease
• positive electrode (red) 2-3cms
proximal to the negative electrode
• • Response: Adductor pollicis muscle –
thumb adduction
VARIOUS SITE USED FOR NM MONITORING
• Facial nerve: place negative electrode
(black) by ear lobe and the positive
(red) 2cms from the eyebrow (along
facial nerve inferior and lateral to
eye)
• • Response: Orbicularis occuli muscle
– eyelid twitching
• Posterior tibial nerve: place the
negative electrode (black) over
inferolateral aspect of medial
malleolus (palpate posterior tibial
pulse and place electrode there) and
positive electrode (red) 2-3cm
proximal to the negative electrode
• Response: Flexor hallucis brevis
muscle – plantar flexion of big toe
Recovery
• Return of the fourth response in the TOF heralds the recovery phase.
• During neuromuscular recovery, a reasonably good correlation exists
between the actual TOF ratio measured by MMG and clinical
observation.
• When the TOF ratio is 0.4 or less, the patient is generally unable to lift the head
or arm. Tidal volume may be normal, but vital capacity and inspiratory force will
be reduced.
• When the ratio is 0.6, most patients are able to lift their head for 3 seconds,
open their eyes widely, and stick out their tongue, but vital capacity and
inspiratory force are often still reduced.
• At a TOF ratio of 0.7 to 0.75, the patient can normally cough sufficiently and lift
the head for at least 5 seconds, but grip strength may still be as low as about
60% of control.
• When the ratio is 0.8 and higher, vital capacity and inspiratory force are normal.
The patient may, however, still have diplopia and facial weakness
UNUSUAL CLINICAL SITUATIONS
• Hemiplegia - resistance to non-depolarisers within 2-3 days on affected side,
possibly due to loss of cerebral inhibition. Always monitor hemiplegic patients on
the unaffected side.
• Parkinsons Disease, Multiple Sclerosis, Tetanus, Intracranial Lesions - normal
sensitivity to non-depolarisers.
• Paraplegia and Quadriplegia - increased sensitivity to non-depolarisers. The
difference in response of the NMJ for upper and lower lesions suggest that
extrajunctional chemosensitivity is not involved. (It is responsible for the
hyperkalaemia following suxamethonium). May also happen following burns,
immobility, prolonged administration of NMB's, etc.
• Amyotrophic Lateral Sclerosis, Polio - increased sensitivity to non-depolarisers.
• Peripheral Neuropathies - usually normal response, although patients with
neurofibromatosis may be sensitive.
• Myotonias - usually normal response to non-depolarisers, occasional sensitive
patients.
• Muscular Dystrophies - mostly normal responses except in the "Ocular" type,
which is very sensitive to non-depolarisers. Duchenne may be a risk factor for
MH.
Limitations of Neuromuscular Monitoring
• Despite the important role of NMJ monitoring in anaesthesia practice, it is
necessary to use a multifactorial approach for the following reasons:
• 1. Neuromuscular responses may appear normal despite persistance of
receptor occupancy by NMBs. T4:T1 ratio is one even when 40-50% of the
receptors are occupied.
• 2. Because of wide individual variability in evoked responses, some patients
may exhibit weakness at TOF ratio as high as 0.8 to 0.9.
• 3. The established cut-off values for adequate recovery do not guarantee
adequate ventilatory function or airway protection.
• 4. Increased skin impedence resulting from perioperative hypothermia
limits the appropriate interpretation of evoked responses.
Conclusion
• Many anaesthesiologists do not agree with extensive use of NMJ
monitors and argue that patients can be managed satisfactorily
without the devices. Although not included under the standards for
basic anaesthetic monitoing by the American Society of
Anaesthesiologists, the real value of such monitors lies in the fact that
they guide the optimal management of patients receiving NMBs.
Sources
• Millers Anesthesia 7th Edition
• Understanding Anesthesia Equipment, Dorsch and Dorsch 5th Edition
• McGrath CD, Hunter JM. Monitoring of neuromuscular block.
CEACP.2006;6:7-12
• Dr. D. Padmaja, Dr. Srinivas Mantha. Monitoring of Neuromuscular
Junction. IJA.2002;46(4) : 279-288

Más contenido relacionado

La actualidad más candente

Goal directed fluid therapy
Goal directed fluid therapyGoal directed fluid therapy
Goal directed fluid therapythanigai arasu
 
Double Lumen Endobronchial Tubes ppt
Double Lumen Endobronchial Tubes pptDouble Lumen Endobronchial Tubes ppt
Double Lumen Endobronchial Tubes pptImran Sheikh
 
Physics In Anaesthesia
Physics In AnaesthesiaPhysics In Anaesthesia
Physics In AnaesthesiaNARENDRA PATIL
 
Uptake and distribution of inhaled anesthetic
Uptake and distribution of inhaled anestheticUptake and distribution of inhaled anesthetic
Uptake and distribution of inhaled anestheticPrakash Gondode
 
Cerebral physiology and effects of anaesthetic agents
Cerebral physiology and effects of anaesthetic agentsCerebral physiology and effects of anaesthetic agents
Cerebral physiology and effects of anaesthetic agentsRicha Kumar
 
Monitoring depth of anaesthesia
Monitoring depth of anaesthesiaMonitoring depth of anaesthesia
Monitoring depth of anaesthesiadr anurag giri
 
Intro to Hypoxic pulmonary vasoconstriction
Intro to Hypoxic pulmonary vasoconstriction Intro to Hypoxic pulmonary vasoconstriction
Intro to Hypoxic pulmonary vasoconstriction Arun Shetty
 
Ropivacaine vs Bupivacaine
Ropivacaine vs Bupivacaine Ropivacaine vs Bupivacaine
Ropivacaine vs Bupivacaine Reza Aminnejad
 
caudal anesthesia.pdf
caudal anesthesia.pdfcaudal anesthesia.pdf
caudal anesthesia.pdfKhodifadVijay
 
Uses of Ultrasound in Anesthesiology
Uses of Ultrasound in AnesthesiologyUses of Ultrasound in Anesthesiology
Uses of Ultrasound in AnesthesiologySaneesh P J
 
Circle system low flow anesthesia
Circle system low flow anesthesiaCircle system low flow anesthesia
Circle system low flow anesthesiaDrgeeta Choudhary
 
Cardiac output monitoring
Cardiac output monitoringCardiac output monitoring
Cardiac output monitoringmauryaramgopal
 
NEUROMUSCULAR MONITORING
NEUROMUSCULAR MONITORINGNEUROMUSCULAR MONITORING
NEUROMUSCULAR MONITORINGAshish Gupta
 
Total Intravenous Anaesthesia
Total Intravenous AnaesthesiaTotal Intravenous Anaesthesia
Total Intravenous AnaesthesiaBrijesh Savidhan
 
Low flow Anaesthesia & Gas Monitoring
Low flow Anaesthesia & Gas MonitoringLow flow Anaesthesia & Gas Monitoring
Low flow Anaesthesia & Gas MonitoringKalpesh Shah
 
Monitoring depth of anaesthesia
Monitoring depth of anaesthesiaMonitoring depth of anaesthesia
Monitoring depth of anaesthesiaShamita Roy
 

La actualidad más candente (20)

Goal directed fluid therapy
Goal directed fluid therapyGoal directed fluid therapy
Goal directed fluid therapy
 
ASRA Guidelines
ASRA GuidelinesASRA Guidelines
ASRA Guidelines
 
Double Lumen Endobronchial Tubes ppt
Double Lumen Endobronchial Tubes pptDouble Lumen Endobronchial Tubes ppt
Double Lumen Endobronchial Tubes ppt
 
Physics In Anaesthesia
Physics In AnaesthesiaPhysics In Anaesthesia
Physics In Anaesthesia
 
Uptake and distribution of inhaled anesthetic
Uptake and distribution of inhaled anestheticUptake and distribution of inhaled anesthetic
Uptake and distribution of inhaled anesthetic
 
Cerebral physiology and effects of anaesthetic agents
Cerebral physiology and effects of anaesthetic agentsCerebral physiology and effects of anaesthetic agents
Cerebral physiology and effects of anaesthetic agents
 
Monitoring depth of anaesthesia
Monitoring depth of anaesthesiaMonitoring depth of anaesthesia
Monitoring depth of anaesthesia
 
One lung ventilation
One lung ventilationOne lung ventilation
One lung ventilation
 
Intro to Hypoxic pulmonary vasoconstriction
Intro to Hypoxic pulmonary vasoconstriction Intro to Hypoxic pulmonary vasoconstriction
Intro to Hypoxic pulmonary vasoconstriction
 
Ropivacaine vs Bupivacaine
Ropivacaine vs Bupivacaine Ropivacaine vs Bupivacaine
Ropivacaine vs Bupivacaine
 
caudal anesthesia.pdf
caudal anesthesia.pdfcaudal anesthesia.pdf
caudal anesthesia.pdf
 
Uses of Ultrasound in Anesthesiology
Uses of Ultrasound in AnesthesiologyUses of Ultrasound in Anesthesiology
Uses of Ultrasound in Anesthesiology
 
Circle system low flow anesthesia
Circle system low flow anesthesiaCircle system low flow anesthesia
Circle system low flow anesthesia
 
Cardiac output monitoring
Cardiac output monitoringCardiac output monitoring
Cardiac output monitoring
 
NEUROMUSCULAR MONITORING
NEUROMUSCULAR MONITORINGNEUROMUSCULAR MONITORING
NEUROMUSCULAR MONITORING
 
NON OPERATING ROOM ANAESTHESIA
NON OPERATING ROOM ANAESTHESIANON OPERATING ROOM ANAESTHESIA
NON OPERATING ROOM ANAESTHESIA
 
Total Intravenous Anaesthesia
Total Intravenous AnaesthesiaTotal Intravenous Anaesthesia
Total Intravenous Anaesthesia
 
Low flow Anaesthesia & Gas Monitoring
Low flow Anaesthesia & Gas MonitoringLow flow Anaesthesia & Gas Monitoring
Low flow Anaesthesia & Gas Monitoring
 
Inhalational Agents
Inhalational AgentsInhalational Agents
Inhalational Agents
 
Monitoring depth of anaesthesia
Monitoring depth of anaesthesiaMonitoring depth of anaesthesia
Monitoring depth of anaesthesia
 

Similar a Neuromuscular Monitoring

Neuromuscular monitoring
Neuromuscular monitoringNeuromuscular monitoring
Neuromuscular monitoringprateek gupta
 
Neuromuscular Monitoring by Dr Rajeev Ujjwal
Neuromuscular Monitoring by Dr Rajeev UjjwalNeuromuscular Monitoring by Dr Rajeev Ujjwal
Neuromuscular Monitoring by Dr Rajeev Ujjwaldrrajeevujjwal
 
Neuromuscular monitoring
Neuromuscular monitoring Neuromuscular monitoring
Neuromuscular monitoring Tess Jose
 
Neuromuscular Junction Monitoring & Assessment
Neuromuscular Junction Monitoring & AssessmentNeuromuscular Junction Monitoring & Assessment
Neuromuscular Junction Monitoring & AssessmentJaydeep2409
 
Neuro muscular monitoring
Neuro muscular monitoringNeuro muscular monitoring
Neuro muscular monitoringkitubhaimbbs
 
018 neuromuscular monitoring
018 neuromuscular monitoring018 neuromuscular monitoring
018 neuromuscular monitoringbothyshiri
 
nmmonitoring-141107064516-conversion-gate02.pdf
nmmonitoring-141107064516-conversion-gate02.pdfnmmonitoring-141107064516-conversion-gate02.pdf
nmmonitoring-141107064516-conversion-gate02.pdfMANJUNATHNREDDY
 
Neuromuscular monitoring presentation .pptx
Neuromuscular monitoring presentation .pptxNeuromuscular monitoring presentation .pptx
Neuromuscular monitoring presentation .pptxAkhilGamingYt
 
Nmt monitoring2011 dr emad
Nmt monitoring2011 dr emadNmt monitoring2011 dr emad
Nmt monitoring2011 dr emadHossam atef
 
Nueromuscular monitoring
Nueromuscular monitoringNueromuscular monitoring
Nueromuscular monitoringShiv Sunder
 
Evaluation trmors dystoniamyoclonus
Evaluation trmors dystoniamyoclonusEvaluation trmors dystoniamyoclonus
Evaluation trmors dystoniamyoclonusNeurologyKota
 
Electromyography (EMG) Basics
Electromyography (EMG) BasicsElectromyography (EMG) Basics
Electromyography (EMG) BasicsMUHAMMED AJMAL P
 
Different types of low frequency currents.pdf
Different types of low frequency currents.pdfDifferent types of low frequency currents.pdf
Different types of low frequency currents.pdfeyobkaseye
 

Similar a Neuromuscular Monitoring (20)

Neuromuscular monitoring
Neuromuscular monitoringNeuromuscular monitoring
Neuromuscular monitoring
 
Neuromuscular Monitoring by Dr Rajeev Ujjwal
Neuromuscular Monitoring by Dr Rajeev UjjwalNeuromuscular Monitoring by Dr Rajeev Ujjwal
Neuromuscular Monitoring by Dr Rajeev Ujjwal
 
Neuromuscular monitoring
Neuromuscular monitoring Neuromuscular monitoring
Neuromuscular monitoring
 
Neuromuscular Junction Monitoring & Assessment
Neuromuscular Junction Monitoring & AssessmentNeuromuscular Junction Monitoring & Assessment
Neuromuscular Junction Monitoring & Assessment
 
Neuro muscular monitoring
Neuro muscular monitoringNeuro muscular monitoring
Neuro muscular monitoring
 
018 neuromuscular monitoring
018 neuromuscular monitoring018 neuromuscular monitoring
018 neuromuscular monitoring
 
nmmonitoring-141107064516-conversion-gate02.pdf
nmmonitoring-141107064516-conversion-gate02.pdfnmmonitoring-141107064516-conversion-gate02.pdf
nmmonitoring-141107064516-conversion-gate02.pdf
 
Neuromuscular monitoring presentation .pptx
Neuromuscular monitoring presentation .pptxNeuromuscular monitoring presentation .pptx
Neuromuscular monitoring presentation .pptx
 
Neuromuscular monitoring
Neuromuscular monitoringNeuromuscular monitoring
Neuromuscular monitoring
 
Neuro muscular monitoring
Neuro muscular monitoringNeuro muscular monitoring
Neuro muscular monitoring
 
Nmt monitoring2011 dr emad
Nmt monitoring2011 dr emadNmt monitoring2011 dr emad
Nmt monitoring2011 dr emad
 
Ect electrical stimulus and procedure
Ect  electrical stimulus and procedureEct  electrical stimulus and procedure
Ect electrical stimulus and procedure
 
Nueromuscular monitoring
Nueromuscular monitoringNueromuscular monitoring
Nueromuscular monitoring
 
4344196.ppt
4344196.ppt4344196.ppt
4344196.ppt
 
Basics of emg
Basics of emgBasics of emg
Basics of emg
 
ECT
ECTECT
ECT
 
Evaluation trmors dystoniamyoclonus
Evaluation trmors dystoniamyoclonusEvaluation trmors dystoniamyoclonus
Evaluation trmors dystoniamyoclonus
 
Electromyography (EMG) Basics
Electromyography (EMG) BasicsElectromyography (EMG) Basics
Electromyography (EMG) Basics
 
Role of ECT in psychiatry
Role of ECT in psychiatryRole of ECT in psychiatry
Role of ECT in psychiatry
 
Different types of low frequency currents.pdf
Different types of low frequency currents.pdfDifferent types of low frequency currents.pdf
Different types of low frequency currents.pdf
 

Último

Call Girls Hyderabad Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Hyderabad Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Hyderabad Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Hyderabad Just Call 8250077686 Top Class Call Girl Service AvailableDipal Arora
 
Saket * Call Girls in Delhi - Phone 9711199012 Escorts Service at 6k to 50k a...
Saket * Call Girls in Delhi - Phone 9711199012 Escorts Service at 6k to 50k a...Saket * Call Girls in Delhi - Phone 9711199012 Escorts Service at 6k to 50k a...
Saket * Call Girls in Delhi - Phone 9711199012 Escorts Service at 6k to 50k a...BhumiSaxena1
 
Most Beautiful Call Girl in Bangalore Contact on Whatsapp
Most Beautiful Call Girl in Bangalore Contact on WhatsappMost Beautiful Call Girl in Bangalore Contact on Whatsapp
Most Beautiful Call Girl in Bangalore Contact on WhatsappInaaya Sharma
 
Call Girls Service Jaipur {9521753030} ❤️VVIP RIDDHI Call Girl in Jaipur Raja...
Call Girls Service Jaipur {9521753030} ❤️VVIP RIDDHI Call Girl in Jaipur Raja...Call Girls Service Jaipur {9521753030} ❤️VVIP RIDDHI Call Girl in Jaipur Raja...
Call Girls Service Jaipur {9521753030} ❤️VVIP RIDDHI Call Girl in Jaipur Raja...Sheetaleventcompany
 
Coimbatore Call Girls in Thudiyalur : 7427069034 High Profile Model Escorts |...
Coimbatore Call Girls in Thudiyalur : 7427069034 High Profile Model Escorts |...Coimbatore Call Girls in Thudiyalur : 7427069034 High Profile Model Escorts |...
Coimbatore Call Girls in Thudiyalur : 7427069034 High Profile Model Escorts |...chennailover
 
Call Girls Mysore Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Mysore Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Mysore Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Mysore Just Call 8250077686 Top Class Call Girl Service AvailableDipal Arora
 
Kollam call girls Mallu aunty service 7877702510
Kollam call girls Mallu aunty service 7877702510Kollam call girls Mallu aunty service 7877702510
Kollam call girls Mallu aunty service 7877702510Vipesco
 
Call Girl In Pune 👉 Just CALL ME: 9352988975 💋 Call Out Call Both With High p...
Call Girl In Pune 👉 Just CALL ME: 9352988975 💋 Call Out Call Both With High p...Call Girl In Pune 👉 Just CALL ME: 9352988975 💋 Call Out Call Both With High p...
Call Girl In Pune 👉 Just CALL ME: 9352988975 💋 Call Out Call Both With High p...chetankumar9855
 
Call Girls Madurai Just Call 9630942363 Top Class Call Girl Service Available
Call Girls Madurai Just Call 9630942363 Top Class Call Girl Service AvailableCall Girls Madurai Just Call 9630942363 Top Class Call Girl Service Available
Call Girls Madurai Just Call 9630942363 Top Class Call Girl Service AvailableGENUINE ESCORT AGENCY
 
Call Girls Service Jaipur {9521753030 } ❤️VVIP BHAWNA Call Girl in Jaipur Raj...
Call Girls Service Jaipur {9521753030 } ❤️VVIP BHAWNA Call Girl in Jaipur Raj...Call Girls Service Jaipur {9521753030 } ❤️VVIP BHAWNA Call Girl in Jaipur Raj...
Call Girls Service Jaipur {9521753030 } ❤️VVIP BHAWNA Call Girl in Jaipur Raj...khalifaescort01
 
Russian Call Girls Service Jaipur {8445551418} ❤️PALLAVI VIP Jaipur Call Gir...
Russian Call Girls Service  Jaipur {8445551418} ❤️PALLAVI VIP Jaipur Call Gir...Russian Call Girls Service  Jaipur {8445551418} ❤️PALLAVI VIP Jaipur Call Gir...
Russian Call Girls Service Jaipur {8445551418} ❤️PALLAVI VIP Jaipur Call Gir...parulsinha
 
Low Rate Call Girls Bangalore {7304373326} ❤️VVIP NISHA Call Girls in Bangalo...
Low Rate Call Girls Bangalore {7304373326} ❤️VVIP NISHA Call Girls in Bangalo...Low Rate Call Girls Bangalore {7304373326} ❤️VVIP NISHA Call Girls in Bangalo...
Low Rate Call Girls Bangalore {7304373326} ❤️VVIP NISHA Call Girls in Bangalo...Sheetaleventcompany
 
VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋
VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋
VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋TANUJA PANDEY
 
Independent Call Girls In Jaipur { 8445551418 } ✔ ANIKA MEHTA ✔ Get High Prof...
Independent Call Girls In Jaipur { 8445551418 } ✔ ANIKA MEHTA ✔ Get High Prof...Independent Call Girls In Jaipur { 8445551418 } ✔ ANIKA MEHTA ✔ Get High Prof...
Independent Call Girls In Jaipur { 8445551418 } ✔ ANIKA MEHTA ✔ Get High Prof...parulsinha
 
Models Call Girls In Hyderabad 9630942363 Hyderabad Call Girl & Hyderabad Esc...
Models Call Girls In Hyderabad 9630942363 Hyderabad Call Girl & Hyderabad Esc...Models Call Girls In Hyderabad 9630942363 Hyderabad Call Girl & Hyderabad Esc...
Models Call Girls In Hyderabad 9630942363 Hyderabad Call Girl & Hyderabad Esc...GENUINE ESCORT AGENCY
 
Top Rated Hyderabad Call Girls Chintal ⟟ 9332606886 ⟟ Call Me For Genuine Se...
Top Rated  Hyderabad Call Girls Chintal ⟟ 9332606886 ⟟ Call Me For Genuine Se...Top Rated  Hyderabad Call Girls Chintal ⟟ 9332606886 ⟟ Call Me For Genuine Se...
Top Rated Hyderabad Call Girls Chintal ⟟ 9332606886 ⟟ Call Me For Genuine Se...chandars293
 
💚Call Girls In Amritsar 💯Anvi 📲🔝8725944379🔝Amritsar Call Girl No💰Advance Cash...
💚Call Girls In Amritsar 💯Anvi 📲🔝8725944379🔝Amritsar Call Girl No💰Advance Cash...💚Call Girls In Amritsar 💯Anvi 📲🔝8725944379🔝Amritsar Call Girl No💰Advance Cash...
💚Call Girls In Amritsar 💯Anvi 📲🔝8725944379🔝Amritsar Call Girl No💰Advance Cash...Sheetaleventcompany
 
Call Girls Kolkata Kalikapur 💯Call Us 🔝 8005736733 🔝 💃 Top Class Call Girl Se...
Call Girls Kolkata Kalikapur 💯Call Us 🔝 8005736733 🔝 💃 Top Class Call Girl Se...Call Girls Kolkata Kalikapur 💯Call Us 🔝 8005736733 🔝 💃 Top Class Call Girl Se...
Call Girls Kolkata Kalikapur 💯Call Us 🔝 8005736733 🔝 💃 Top Class Call Girl Se...Namrata Singh
 

Último (20)

Call Girls Hyderabad Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Hyderabad Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Hyderabad Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Hyderabad Just Call 8250077686 Top Class Call Girl Service Available
 
Saket * Call Girls in Delhi - Phone 9711199012 Escorts Service at 6k to 50k a...
Saket * Call Girls in Delhi - Phone 9711199012 Escorts Service at 6k to 50k a...Saket * Call Girls in Delhi - Phone 9711199012 Escorts Service at 6k to 50k a...
Saket * Call Girls in Delhi - Phone 9711199012 Escorts Service at 6k to 50k a...
 
Most Beautiful Call Girl in Bangalore Contact on Whatsapp
Most Beautiful Call Girl in Bangalore Contact on WhatsappMost Beautiful Call Girl in Bangalore Contact on Whatsapp
Most Beautiful Call Girl in Bangalore Contact on Whatsapp
 
Call Girls Service Jaipur {9521753030} ❤️VVIP RIDDHI Call Girl in Jaipur Raja...
Call Girls Service Jaipur {9521753030} ❤️VVIP RIDDHI Call Girl in Jaipur Raja...Call Girls Service Jaipur {9521753030} ❤️VVIP RIDDHI Call Girl in Jaipur Raja...
Call Girls Service Jaipur {9521753030} ❤️VVIP RIDDHI Call Girl in Jaipur Raja...
 
Coimbatore Call Girls in Thudiyalur : 7427069034 High Profile Model Escorts |...
Coimbatore Call Girls in Thudiyalur : 7427069034 High Profile Model Escorts |...Coimbatore Call Girls in Thudiyalur : 7427069034 High Profile Model Escorts |...
Coimbatore Call Girls in Thudiyalur : 7427069034 High Profile Model Escorts |...
 
Call Girls in Gagan Vihar (delhi) call me [🔝 9953056974 🔝] escort service 24X7
Call Girls in Gagan Vihar (delhi) call me [🔝  9953056974 🔝] escort service 24X7Call Girls in Gagan Vihar (delhi) call me [🔝  9953056974 🔝] escort service 24X7
Call Girls in Gagan Vihar (delhi) call me [🔝 9953056974 🔝] escort service 24X7
 
Call Girls Mysore Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Mysore Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Mysore Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Mysore Just Call 8250077686 Top Class Call Girl Service Available
 
Kollam call girls Mallu aunty service 7877702510
Kollam call girls Mallu aunty service 7877702510Kollam call girls Mallu aunty service 7877702510
Kollam call girls Mallu aunty service 7877702510
 
🌹Attapur⬅️ Vip Call Girls Hyderabad 📱9352852248 Book Well Trand Call Girls In...
🌹Attapur⬅️ Vip Call Girls Hyderabad 📱9352852248 Book Well Trand Call Girls In...🌹Attapur⬅️ Vip Call Girls Hyderabad 📱9352852248 Book Well Trand Call Girls In...
🌹Attapur⬅️ Vip Call Girls Hyderabad 📱9352852248 Book Well Trand Call Girls In...
 
Call Girl In Pune 👉 Just CALL ME: 9352988975 💋 Call Out Call Both With High p...
Call Girl In Pune 👉 Just CALL ME: 9352988975 💋 Call Out Call Both With High p...Call Girl In Pune 👉 Just CALL ME: 9352988975 💋 Call Out Call Both With High p...
Call Girl In Pune 👉 Just CALL ME: 9352988975 💋 Call Out Call Both With High p...
 
Call Girls Madurai Just Call 9630942363 Top Class Call Girl Service Available
Call Girls Madurai Just Call 9630942363 Top Class Call Girl Service AvailableCall Girls Madurai Just Call 9630942363 Top Class Call Girl Service Available
Call Girls Madurai Just Call 9630942363 Top Class Call Girl Service Available
 
Call Girls Service Jaipur {9521753030 } ❤️VVIP BHAWNA Call Girl in Jaipur Raj...
Call Girls Service Jaipur {9521753030 } ❤️VVIP BHAWNA Call Girl in Jaipur Raj...Call Girls Service Jaipur {9521753030 } ❤️VVIP BHAWNA Call Girl in Jaipur Raj...
Call Girls Service Jaipur {9521753030 } ❤️VVIP BHAWNA Call Girl in Jaipur Raj...
 
Russian Call Girls Service Jaipur {8445551418} ❤️PALLAVI VIP Jaipur Call Gir...
Russian Call Girls Service  Jaipur {8445551418} ❤️PALLAVI VIP Jaipur Call Gir...Russian Call Girls Service  Jaipur {8445551418} ❤️PALLAVI VIP Jaipur Call Gir...
Russian Call Girls Service Jaipur {8445551418} ❤️PALLAVI VIP Jaipur Call Gir...
 
Low Rate Call Girls Bangalore {7304373326} ❤️VVIP NISHA Call Girls in Bangalo...
Low Rate Call Girls Bangalore {7304373326} ❤️VVIP NISHA Call Girls in Bangalo...Low Rate Call Girls Bangalore {7304373326} ❤️VVIP NISHA Call Girls in Bangalo...
Low Rate Call Girls Bangalore {7304373326} ❤️VVIP NISHA Call Girls in Bangalo...
 
VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋
VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋
VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋
 
Independent Call Girls In Jaipur { 8445551418 } ✔ ANIKA MEHTA ✔ Get High Prof...
Independent Call Girls In Jaipur { 8445551418 } ✔ ANIKA MEHTA ✔ Get High Prof...Independent Call Girls In Jaipur { 8445551418 } ✔ ANIKA MEHTA ✔ Get High Prof...
Independent Call Girls In Jaipur { 8445551418 } ✔ ANIKA MEHTA ✔ Get High Prof...
 
Models Call Girls In Hyderabad 9630942363 Hyderabad Call Girl & Hyderabad Esc...
Models Call Girls In Hyderabad 9630942363 Hyderabad Call Girl & Hyderabad Esc...Models Call Girls In Hyderabad 9630942363 Hyderabad Call Girl & Hyderabad Esc...
Models Call Girls In Hyderabad 9630942363 Hyderabad Call Girl & Hyderabad Esc...
 
Top Rated Hyderabad Call Girls Chintal ⟟ 9332606886 ⟟ Call Me For Genuine Se...
Top Rated  Hyderabad Call Girls Chintal ⟟ 9332606886 ⟟ Call Me For Genuine Se...Top Rated  Hyderabad Call Girls Chintal ⟟ 9332606886 ⟟ Call Me For Genuine Se...
Top Rated Hyderabad Call Girls Chintal ⟟ 9332606886 ⟟ Call Me For Genuine Se...
 
💚Call Girls In Amritsar 💯Anvi 📲🔝8725944379🔝Amritsar Call Girl No💰Advance Cash...
💚Call Girls In Amritsar 💯Anvi 📲🔝8725944379🔝Amritsar Call Girl No💰Advance Cash...💚Call Girls In Amritsar 💯Anvi 📲🔝8725944379🔝Amritsar Call Girl No💰Advance Cash...
💚Call Girls In Amritsar 💯Anvi 📲🔝8725944379🔝Amritsar Call Girl No💰Advance Cash...
 
Call Girls Kolkata Kalikapur 💯Call Us 🔝 8005736733 🔝 💃 Top Class Call Girl Se...
Call Girls Kolkata Kalikapur 💯Call Us 🔝 8005736733 🔝 💃 Top Class Call Girl Se...Call Girls Kolkata Kalikapur 💯Call Us 🔝 8005736733 🔝 💃 Top Class Call Girl Se...
Call Girls Kolkata Kalikapur 💯Call Us 🔝 8005736733 🔝 💃 Top Class Call Girl Se...
 

Neuromuscular Monitoring

  • 2. • 1958, Christie and Churchill-Davidson described how nerve stimulators could be used to assess neuromuscular function objectively during anesthesia • In Awake patients, muscle power can be evaluated through tests of voluntary muscle strength • During anesthesia and recovery from anesthesia this is not possible. Instead, the clinician uses clinical tests to assess muscle power directly and to estimate neuromuscular function indirectly (muscle tone, the feel of the anesthesia bag, an indirect measure of pulmonary compliance, tidal volume, and inspiratory force).
  • 3. • Onset of NM Blockade. • To determine level of muscle relaxation during surgery. • Assessing patients recovery from blockade to minimize risk of residual paralysis. Objectives of NM Monitoring
  • 4. Why do we Monitor? Residual post-op NM Blockade • Functional impairment of pharyngeal and upper esophageal muscles • Impaired ability to maintain the airway • Increased risk for post-op pulmonary complications • Difficult to exclude clinically significant residual curarization by clinical evaluation
  • 5. Principles of Peripheral Nerve Stimulation • Each muscle fiber to a stimulus follows an all-or-none pattern • Response of the whole muscle depends on the number of muscle fibers activated • Response of the muscle decreases in parallel with the numbers of fibers blocked • Reduction in response during constant stimulation reflects degree of NM Blockade • For this reason stimulus is supramaximal
  • 6. Essential features of PNS • Shape of stimulus should be monophasic and rectangular i.e Square-wave stimulus. • 0.2- 0.3 msec duration so it falls within absolute refractory period of motor unit in the nerve. • Constant current variable voltage • Battery powered. • Digital display of delivered current. • Audible signal on delivery of stimulus. • Audible alarm for poor electrode contact. • Multiple patterns of stimulation (single twitch,train-of-four, double-burst, post-tetanic count).
  • 7. • Current intensity : It is the amperage (mA) of the current delivered by the nerve stimulator(0-80 mA). The intensity reaching the nerve is determined by the voltage generated by the stimulator and resistance and impedance of the electrodes, skin and underlying tissues • Nerve stimulators are constant current and variable voltage delivery devices • Reduction of temperature increases the tissue resistance (increased impedance) and may cause reduction in the current delivered to fall below the supramaximal level • Threshold current : It is the lowest current required to depolarize the most sensitive fibres in a given nerve bundle to elicit a detectable muscle response. • Maximal current: Current which generate response in all muscle fibre • Supramaximal current : It is approximately 25% higher intensity than the current required to depolarize all fibres in a particular nerve bundle. This is generally attained at current intensity 2-3 times higher than threshold current. • Submaximal current : A current intensity that induces firing of only a fraction fibres in a given nerve bundle. A potential advantage of submaximal current is that it is less painful than supramaximal current. • Stimulus frequency : The rate (Hz) at which each impulse is repeated in cycles per second (Hz).
  • 8. Types of Peripheral Nerve Stimulation • Neuromuscular function is monitored by evaluating the muscular response to supramaximal stimulation of a peripheral motor nerve. • Two types of stimulation can be used: electrical and magnetic. • Magnetic nerve stimulation has several advantages over electrical nerve stimulation.It is less painful and does not require physical contact with the body. • However, the equipment required is bulky and heavy, it cannot be used for train-of-four (TOF) stimulation, and it is difficult to achieve supramaximal stimulation with this method.
  • 9. Neuromuscular Monitoring is Essential • After prolonged infusions of neuromuscular blocking drugs or when long-acting drugs are used • When surgery or anaesthesia is prolonged • When inadequate reversal may have devastating effects, for example, severe respiratory disease, morbid obesity • In conditions where administration of a reversal agent may cause harm, for example, tachyarrhythmias, cardiac failure • Liver or renal dysfunction, when pharmacokinetics of muscular relaxants may be altered • Neuromuscular disorders such as myasthenia gravis or Eaton–Lambert syndrome • Surgeries requiring profound neuromuscular blockade e.g. neurosurgery, vascular surgery in vital areas like thoracic cavity.
  • 10. Patterns of Nerve Stimulation For evaluation of neuromuscular function five patterns of nerve stimulation are commonly used in clinical practice • Single twitch stimulation. • Train-of-four stimulation. • Tetanic stimulation. • Post-tetanic count stimulation. • Double burst stimulation.
  • 11. Single-Twitch Stimulation • A Single supramaximal electrical stimuli are applied to a peripheral motor nerve at frequencies ranging from 1.0 Hz (once every second) to 0.1 Hz (once every 10 seconds) . • The response to single-twitch stimulation depends on the frequency at which the individual stimuli are applied. • Because 1-Hz stimulation shortens the time necessary to determine supramaximal stimulation, this frequency is sometimes used during induction of anesthesia.
  • 12.
  • 13. Train-of-Four Stimulation • Four supramaximal stimuli are given every 0.5 second (2 Hz). • When used continuously, each set (train) of stimuli is normally repeated every 10th to 20th second. • Each stimulus in the train causes the muscle to contract, and “fade” in the response provides the basis for evaluation. • That is, dividing the amplitude of the fourth response by the amplitude of the first response provides the TOF ratio. • In the control response (the response obtained before the administration of a muscle relaxant), all four responses are ideally the same: the TOF ratio is 1.0.
  • 14. • During a partial nondepolarizing block, the ratio decreases (fades) and is inversely proportional to the degree of blockade. • During a partial depolarizing block, no fade occurs in the TOF response; ideally, the TOF ratio is approximately 1.0. • Fade in the TOF response after injection of succinylcholine signifies the development of a phase II block. • The advantages of TOF stimulation are greatest during nondepolarizing blockade because the degree of block can be read directly from the TOF response even though a preoperative value is lacking. • In addition, TOF stimulation has some advantages over tetanic stimulation: it is less painful and, unlike tetanic stimulation, does not generally affect the degree of neuromuscular blockade.
  • 15.
  • 16.
  • 17. Tetanic Stimulation • Tetanic stimulation consists of very rapid (e.g., 30-, 50-, or 100-Hz) delivery of electrical stimuli. • The most commonly used pattern in clinical practice is 50-Hz stimulation given for 5 seconds, although some investigators have advocated the use of 50-, 100-, and even 200-Hz stimulation for 1 second. • During normal neuromuscular transmission and a pure depolarizing block, the muscle response to 50-Hz tetanic stimulation for 5 seconds is sustained. • During a nondepolarizing block and a phase II block after the injection of succinylcholine, the response will not be sustained (i.e., fade occurs) • Fade in response to tetanic stimulation is normally considered a presynaptic event; the traditional explanation is that at the start of tetanic stimulation, large amounts of acetylcholine are released from immediately available stores in the nerve terminal. • As these stores become depleted, the rate of acetylcholine release decreases until equilibrium between mobilization and synthesis of acetylcholine is achieved.
  • 18. • When the “margin of safety” at the postsynaptic membrane (i.e., the number of free cholinergic receptors) is reduced by nondepolarizing neuromuscular blocking drugs, a typical reduction in twitch height is seen with a fade during, for instance, repetitive stimulation. • In addition to this postsynaptic block, nondepolarizing neuromuscular blocking drugs may also block presynaptic neuronal-type acetylcholine receptors, thereby leading to impaired mobilization of acetylcholine within the nerve terminal. • This effect substantially contributes to fade in the response to tetanic (and TOF) stimulation. • Although the degree of fade depends primarily on the degree of neuromuscular blockade, fade also depends on the frequency (Hz) and the length (seconds) of stimulation and on how often tetanic stimuli are applied. • During partial nondepolarizing blockade, tetanic nerve stimulation is followed by a post-tetanic increase in twitch tension (i.e., post-tetanic facilitation of transmission). • This event occurs because the increase in mobilization and synthesis of acetylcholine caused by tetanic stimulation continues for some time after discontinuation of stimulation.
  • 19. • The degree and duration of post-tetanic facilitation depend on the degree of neuromuscular blockade, with post-tetanic facilitation usually disappearing within 60 seconds of tetanic stimulation. • In contrast, post-tetanic twitch potentiation, which sometimes occurs in mechanical recordings before any neuromuscular blocking drug has been given, is a muscular phenomenon that is not accompanied by an increase in the compound muscle action potential. • Tetanic stimulation is very painful and therefore not normally acceptable to an unanesthetized patient.
  • 20.
  • 21. Post-Tetanic Count Stimulation • Injection of a nondepolarizing neuromuscular blocking drug in a dose sufficient to ensure smooth tracheal intubation causes intense neuromuscular blockade of the peripheral muscles. • Because no response to TOF and single-twitch stimulation occurs under these conditions, these modes of stimulation cannot be used to determine the degree of blockade. • It is possible, however, to quantify intense neuromuscular blockade of the peripheral muscles by applying tetanic stimulation (50 Hz for 5 seconds) and observing the post-tetanic response to single- twitch stimulation given at 1 Hz starting 3 seconds after the end of tetanic stimulation. • During intense blockade, there is no response to either tetanic or post-tetanic stimulation.
  • 22. • As the intense block dissipates, more and more responses to post-tetanic twitch stimulation appear. For a given neuromuscular blocking drug, the time until return of the first response to TOF stimulation is related to the number of post-tetanic twitch responses present at a given time (i.e., the PTC) • The PTC method is mainly used to assess the degree of neuromuscular blockade when there is no reaction to single-twitch or TOF nerve stimulation, as may be the case after injection of a large dose of a nondepolarizing neuromuscular blocking drug. • However, PTC can also be used whenever sudden movements must be eliminated (e.g., during ophthalmic surgery). • The necessary level of blockade of the adductor pollicis muscle to ensure paralysis of the diaphragm depends on the type of anesthesia and, in the intensive care unit, on the level of sedation. • To ensure elimination of any bucking or coughing in response to tracheobronchial stimulation, neuromuscular blockade of the peripheral muscles must be so intense that no response to post-tetanic twitch stimulation can be elicited (PTC 0) • The response to PTC stimulation depends primarily on the degree of neuromuscular blockade.
  • 23. • It also depends on the :- 1. frequency and duration of tetanic stimulation, 2. the length of time between the end of tetanic stimulation and the first post-tetanic stimulus, 3. the frequency of the single-twitch stimulation, 4. the duration of single-twitch stimulation before tetanic stimulation. • When the PTC method is used, these variables should be kept constant. • In addition, because of possible antagonism of neuromuscular blockade in the hand, tetanic stimulation should not be performed more often than every 6 minutes.
  • 24. • Pattern of electrical stimulation and evoked muscle responses to • train-of-four (TOF) nerve stimulation, • 50-Hz tetanic nerve stimulation for 5 seconds (TE), and • 1.0-Hz post-tetanic twitch stimulation (PTS) during four different levels of nondepolarizing neuromuscular blockade. • [A] During intense blockade of peripheral muscles no response to any of the forms of stimulation occurs. • [B/ C] During less pronounced blockade, there is still no response to TOF stimulation, but post-tetanic facilitation of transmission is present. • [D] During surgical blockade , the first response to TOF appears and post-tetanic facilitation increases further. • The post-tetanic count is 1 during very deep block (B), 3 during less deep block (C), and 8 during surgical (or moderate) block (D).
  • 25. Double-Burst Stimulation • DBS consists of two short bursts of 50-Hz tetanic stimulation separated by 750 msec. • The duration of each square wave impulse in the burst is 0.2 msec. • Although the number of impulses in each burst can vary, DBS with three impulses in each of the two tetanic bursts (DBS3,3) is most commonly used.
  • 26. • In nonparalyzed muscle, the response to DBS3,3 is two short muscle contractions of equal strength. • In a partly paralyzed muscle, the second response is weaker than the first (i.e., the response fades). • DBS was developed with the specific aim of allowing manual (tactile) detection of small amounts of residual blockade under clinical conditions,and during recovery and immediately after surgery, tactile evaluation of the response to DBS3,3 is superior to tactile evaluation of the response to TOF stimulation.
  • 27. Equipments Required • The nerve stimulator, • The stimulating electrodes, and • The recording equipment.
  • 28. The Nerve Stimulator It delivers the stimulus to the electrodes. Although many nerve stimulators are commercially available, not all meet the basic requirement for clinical use. The real nerve stimulator should have the following properties. • It should deliver monophasic and rectangular waveform pulse of constant current. • It should deliver a current up to 60-70 mA but not >80mA. • It should be battery operated and should have a battery check. • It should have either built in warning system or current level display that alerts the user when the current selected is not delivered to the nerve (in case of increased skin resistance) • It should have polarity indicators for electrodes. • It should deliver different modes of stimulation.
  • 29. The Stimulation Electrodes • Electrical impulses are transmitted from the stimulator to the nerve by means of surface or needle electrodes. Surface electrodes are commonly used in clinical anaesthesia. These are pregelled silver/sliver chloride electrodes having approximately 7-8 mm diameter of the conducting area.5 • If tissue resistance prevents the stimulating current from reaching the nerve (i.e. in the morbidly obese patients), needle electrodes can be used. A Current = 10mA will usually produce supramaximal stimulation when subcutaneous needle electrodes are used. Although specially coated needle electrodes are commercially available, ordinary steel injection needles can be used.
  • 30. Assessment of Responses to Nerve Stimulation Five methods are available: 1. Measurement of the evoked mechanical response of the muscle (mechanomyography [MMG]), 2. Measurement of the evoked electrical response of the muscle (electromyography [EMG]), 3. Measurement of acceleration of the muscle response (acceleromyography [AMG]), 4. Measurement of the evoked electrical response in a piezoelectric film sensor attached to the muscle (piezoelectric neuromuscular monitor [PZEMG] 5. Phonomyography [PMG]).
  • 31. Mechanomyography • The mechanomyogram (MMG) is the mechanical signal observable from the surface of a muscle when the muscle is contracted. • At the onset of muscle contraction, gross changes in the muscle shape cause a large peak in the MMG. • Subsequent vibrations are due to oscillations of the muscle fibres at the resonance frequency of the muscle. • A requirement for correct and reproducible measurement of evoked tension is that the muscle contraction be isometric. • In clinical anesthesia, this condition is most easily achieved by measuring thumb movement after the application of a resting tension of 200 to 300 g (a preload) to the thumb. • When the ulnar nerve is stimulated, the thumb (the adductor pollicis muscle) acts on a force-displacement transducer.
  • 32. • The force of contraction is then converted into an electrical signal, which is amplified, displayed, and recorded. • The arm and hand should be rigidly fixed, and care should be taken to prevent overloading of the transducer. • In addition, the transducer should be placed in correct relation to the thumb (i.e., the thumb should always apply tension precisely along the length of the transducer). • It is important to remember that the response to nerve stimulation depends on the frequency with which the individual stimuli are applied and that the time used to achieve a stable control response may influence subsequent determination of the onset time and duration of blockade. • Generally, the reaction to supramaximal stimulation increases during the first 8 to 12 minutes after commencement of the stimulation.
  • 33. Electromyography • (EMG) is a technique for evaluating and recording the electrical activity produced by skeletal muscles • Evoked EMG records the compound action potentials produced by stimulation of a peripheral nerve. The compound action potential is a high-speed event that for many years could be picked up only by means of a preamplifier and a storage oscilloscope. • The evoked EMG response is most often obtained from muscles innervated by the ulnar or the median nerves. • Most often, the evoked EMG response is obtained from the thenar or hypothenar eminence of the hand or from the first dorsal interosseous muscle of the hand, preferably with the active electrode over the motor point of the muscle . • The signal picked up by the analyzer is processed by an amplifier, a rectifier, and an electronic integrator. The results are displayed either as a percentage of control or as a TOF ratio.
  • 34. • The evoked EMG response is most often obtained from muscles innervated by the ulnar or the median nerves. • Most often, the evoked EMG response is obtained from the thenar or hypothenar eminence of the hand or from the first dorsal interosseous muscle of the hand, preferably with the active electrode over the motor point of the muscle . • The signal picked up by the analyzer is processed by an amplifier, a rectifier, and an electronic integrator. The results are displayed either as a percentage of control or as a TOF ratio.
  • 35. • Electrode placement for stimulation of the ulnar nerve and for recording of the compound action potential from three sites of the hand. • A, Abductor digiti minimi muscle (in the hypothenar eminence). • B, Adductor pollicis muscle (in the thenar eminence). • C, First dorsal interosseus muscle.
  • 36. • Two new sites for recording the EMG response have been introduced: the larynx and the diaphragm. • By using a noninvasive disposable laryngeal electrode attached to the tracheal tube and placed between the vocal cords, it is possible to monitor the onset of neuromuscular blockade in the laryngeal muscles. • In paravertebral surface diaphragmatic EMG, the recording electrodes are placed on the right of vertebrae T12/L1 o r L1/L2 for monitoring the response of the right diaphragmatic crux to transcutaneous stimulation of the right phrenic nerve at the neck. • Evoked electrical and mechanical responses represent different physiologic events. Evoked EMG records changes in the electrical activity of one or more muscles, whereas evoked MMG records changes associated with excitation- contraction coupling and contraction of the muscle as well.
  • 37. ADVANTAGES • Equipment for measuring evoked EMG responses is easier to set up, • The response reflects only factors influencing neuromuscular transmission, and • The response can be obtained from muscles not accessible to mechanical recording. DISADVANTAGES • Although high-quality recordings are possible in most patients, the results are not always reliable. For one thing, improper placement of electrodes may result in inadequate pickup of the compound EMG signal. • Direct muscle stimulation sometimes occurs. If muscles close to the stimulating electrodes are stimulated directly, the recording electrodes may pick up an electrical signal even though neuromuscular transmission is completely blocked. • Another difficulty is that the EMG response often does not return to the control value. • Finally, the evoked EMG response is very sensitive to electrical interference, such as that caused by diathermy.
  • 38. Acceleromyograph • Is a piezoelectric myograph, used to measure the force produced by a muscle after it has undergone nerve stimulation. • Acceleromyographs measure muscle activity using a miniature piezoelectric transducer that is attached to the stimulated muscle. • A voltage is created when the muscle accelerates and that acceleration is proportion to force of contraction. • Acceleromyographs are more costly than the more common twitch monitors, but have been shown to better alleviate residual blockade and associated symptoms of muscle weakness, and to improve overall quality of recovery.
  • 39. • When an accelerometer is fixed to the thumb and the ulnar nerve is stimulated, an electrical signal is produced whenever the thumb moves. • This signal can be analyzed in a specially designed analyzer. • TOF-Watch (Organon, part of Schering-Plough, Corp.). This neuromuscular transmission monitor is based on measurement of acceleration with a piezoelectric transducer. • Transducer is fastened to the thumb and the stimulating electrodes. • On the display of the TOF-Watch, the train-of-four (TOF) ratio is given in percentage.
  • 40. • Hand adaptor (elastic preload) for the TOF- Watch transducer
  • 41. Piezoelectric Neuromuscular Monitors • The technique of the piezoelectric monitor is based on the principle that stretching or bending a flexible piezoelectric film (e.g., one attached to the thumb) in response to nerve stimulation generates a voltage that is proportional to the amount of stretching or bending.
  • 42. • At least two devices based on this principle are available commercially: the ParaGraph Neuromuscular Blockade Monitor and the M-NMT MechanoSensor, which is a part of the Datex AS/3 monitoring system (Datex-Ohmeda, Helsinki, Finland)
  • 43. Phonomyography • Contraction of skeletal muscles generates intrinsic low-frequency sounds, which can be recorded with special microphones. • What does make PMG interesting, however, is that in theory the method can be applied not only to the adductor pollicis muscle but also to other muscles of interest such as the diaphragm, larynx, and eye muscles. • In addition, the ease of application is attractive.
  • 44. • Ulnar nerve: MOST COMMON SITE • place negative electrode (black) on wrist in line with the smallest digit 1- 2cm below skin crease • positive electrode (red) 2-3cms proximal to the negative electrode • • Response: Adductor pollicis muscle – thumb adduction VARIOUS SITE USED FOR NM MONITORING
  • 45. • Facial nerve: place negative electrode (black) by ear lobe and the positive (red) 2cms from the eyebrow (along facial nerve inferior and lateral to eye) • • Response: Orbicularis occuli muscle – eyelid twitching
  • 46. • Posterior tibial nerve: place the negative electrode (black) over inferolateral aspect of medial malleolus (palpate posterior tibial pulse and place electrode there) and positive electrode (red) 2-3cm proximal to the negative electrode • Response: Flexor hallucis brevis muscle – plantar flexion of big toe
  • 47. Recovery • Return of the fourth response in the TOF heralds the recovery phase. • During neuromuscular recovery, a reasonably good correlation exists between the actual TOF ratio measured by MMG and clinical observation.
  • 48. • When the TOF ratio is 0.4 or less, the patient is generally unable to lift the head or arm. Tidal volume may be normal, but vital capacity and inspiratory force will be reduced. • When the ratio is 0.6, most patients are able to lift their head for 3 seconds, open their eyes widely, and stick out their tongue, but vital capacity and inspiratory force are often still reduced. • At a TOF ratio of 0.7 to 0.75, the patient can normally cough sufficiently and lift the head for at least 5 seconds, but grip strength may still be as low as about 60% of control. • When the ratio is 0.8 and higher, vital capacity and inspiratory force are normal. The patient may, however, still have diplopia and facial weakness
  • 49. UNUSUAL CLINICAL SITUATIONS • Hemiplegia - resistance to non-depolarisers within 2-3 days on affected side, possibly due to loss of cerebral inhibition. Always monitor hemiplegic patients on the unaffected side. • Parkinsons Disease, Multiple Sclerosis, Tetanus, Intracranial Lesions - normal sensitivity to non-depolarisers. • Paraplegia and Quadriplegia - increased sensitivity to non-depolarisers. The difference in response of the NMJ for upper and lower lesions suggest that extrajunctional chemosensitivity is not involved. (It is responsible for the hyperkalaemia following suxamethonium). May also happen following burns, immobility, prolonged administration of NMB's, etc. • Amyotrophic Lateral Sclerosis, Polio - increased sensitivity to non-depolarisers. • Peripheral Neuropathies - usually normal response, although patients with neurofibromatosis may be sensitive. • Myotonias - usually normal response to non-depolarisers, occasional sensitive patients. • Muscular Dystrophies - mostly normal responses except in the "Ocular" type, which is very sensitive to non-depolarisers. Duchenne may be a risk factor for MH.
  • 50. Limitations of Neuromuscular Monitoring • Despite the important role of NMJ monitoring in anaesthesia practice, it is necessary to use a multifactorial approach for the following reasons: • 1. Neuromuscular responses may appear normal despite persistance of receptor occupancy by NMBs. T4:T1 ratio is one even when 40-50% of the receptors are occupied. • 2. Because of wide individual variability in evoked responses, some patients may exhibit weakness at TOF ratio as high as 0.8 to 0.9. • 3. The established cut-off values for adequate recovery do not guarantee adequate ventilatory function or airway protection. • 4. Increased skin impedence resulting from perioperative hypothermia limits the appropriate interpretation of evoked responses.
  • 51. Conclusion • Many anaesthesiologists do not agree with extensive use of NMJ monitors and argue that patients can be managed satisfactorily without the devices. Although not included under the standards for basic anaesthetic monitoing by the American Society of Anaesthesiologists, the real value of such monitors lies in the fact that they guide the optimal management of patients receiving NMBs.
  • 52. Sources • Millers Anesthesia 7th Edition • Understanding Anesthesia Equipment, Dorsch and Dorsch 5th Edition • McGrath CD, Hunter JM. Monitoring of neuromuscular block. CEACP.2006;6:7-12 • Dr. D. Padmaja, Dr. Srinivas Mantha. Monitoring of Neuromuscular Junction. IJA.2002;46(4) : 279-288