SlideShare una empresa de Scribd logo
1 de 41
1. Define defibrillation.
2. Describe Need and history of defibrillation.
3. Describe the principle and mechanism of
defibrillation.
4. Types and classes of defibrillator
5. Describe the Automated external defibrillator
6. Identify the precautions and risk
7. Troubleshooting of defibrillator
Definition:
Defibrillation is a process in which an electronic device
sends an electric shock to the heart to stop an extremely
rapid, irregular heartbeat, and restore the normal heart
rhythm.
Defibrillation is a common treatment for life
threatening cardiac dysrhythmias, ventricular
fibrillation, and pulse less ventricular tachycardia.
Ventricular fibrillation is a serious cardiac emergency
resulting from asynchronous contraction of the heart
muscles.
 Due to ventricular fibrillation, there is an irregular
rapid heart rhythm.

Fig. Ventricular fibrillation

Fig. Normal heart beat
 Ventricular fibrillation can be converted into a more
efficient rhythm by applying a high energy shock
to the heart.

 This sudden surge across the heart causes all muscle
fibres to contract simultaneously.
 Possibly, the fibres may then respond to normal
physiological pace making pulses.

 The instrument for administering the shock is called
a DEFIBRILLATOR.
Defibrillation is performed to correct lifethreatening fibrillations of the heart, which could
result in cardiac arrest. It should be performed
immediately after identifying that the patient is
experiencing a cardiac emergency, has no pulse,
and is unresponsive.
Defibrillation was invented in
by Prevost and
Batelli, two Italian physiologists. They discovered
that electric shocks could convert ventricular
fibrillation to sinus rhythm in dogs.The first case of a
human life saved by defibrillation was reported by
Beck in 1947 .
 Energy storage capacitor is charged at relatively
slow rate from AC line.

 Energy stored in capacitor is then delivered at a
relatively rapid rate to chest of the patient.

 Simple arrangement involve the discharge of
capacitor energy through the patient‟s own
resistance.
Fig: Schematic diagram of a defibrillator
 The discharge resistance which the patient
represents as purely ohmic resistance of 50 to 100Ω
approximately for a typical electrode size of 80cm2.

 This particular waveform Fig is called „ Lown‟
waveform.

 The pulse width of this waveform is generally 10 ms.
current (amps)

defibrillation
occurs

no defibrillation

pulse duration
• minimum defibrillation energy occurs for pulse
durations of 3 - 10 ms (for most pulse shapes).
• pulse amplitude in tens of amperes (few
thousand volts).
• operator selects energy delivered: 50-360
joules, depends on:
–
–
–
–
–

intrinsic characteristics of patient
patient‟s disease
duration of arrhythmia
patient‟s age
type of arrhythmia (more energy required for v.
fib.)
 Fibrillations cause the heart to stop pumping
blood, leading to brain damage.

 Defibrillators deliver a brief electric shock to the
heart, which enables the heart's natural pacemaker
to regain control and establish a normal heart
rhythm.
 Higher voltages are required for external defibrillation

than for internal defibrillation.
 A corrective shock of 750-800 volts is applied within a

tenth of a second.
 That is the same voltage as 500-533 no of AA batteries!
Electrical pattern

ECG tracing
 Occulsion of the

coronary artery leads to
ischemia
 Ischemia leads to
infarct which causes
interruption of normal
cardiac conduction
 Infarct = VF/VT
Ventricular Fibrillation

Ventricular Tachycardia
 Types of Defibrillator electrodes:a) Spoon shaped electrode
• Applied directly to the heart.
b) Paddle type electrode
• Applied against the chest wall
c) Pad type electrode
• Applied directly on chest wall
fig: Electrodes used in defibrillator (a) a spoon shaped internal
electrode that is applied directly to the heart. (b) a paddle type
electrode applied against the anterior chest wall.
Fig.- Pad electrode
Anterior electrode pad

Apex electrode pad

Fig: anterior –apex scheme of electrode placement
Monophasic pulse or waveform

Bi-phasic pulse or waveform
 There are two general classes of waveforms:
a) mono-phasic waveform

•

Energy delivered in one direction through the
patient‟s heart

a) Biphasic waveform
•

Energy delivered in both direction through the
patient‟s heart
Fig:- Generation of bi-phasic waveform
 The biphasic waveform is preferred over
monophasic waveform to defibrillate. Why?????
• A monophasic type, give a high-energy shock,
up to 360 to 400 joules due to which increased
cardiac injury and in burns the chest around the
shock pad sites.
• A biphasic type, give two sequential lowerenergy shocks of 120 - 200 joules, with each
shock moving in an opposite polarity between
the pads.
Internal

External

28
Internal defibrillator

a)
•

Electrodes placed directly to the heart

•

e.g..-Pacemaker

b) External defibrillator
•

Electrodes placed directly on the heart

•

e.g..-AED
• For each minute elapsing between onset of
ventricular fibrillation and first defibrillation,
survival decreases by 10%.
• defibrillators should be portable, battery operated,
small size.
• energy in defibrillators usually stored in large
capacitors.
• total energy stored in capacitor:

WC

1
CVC2
2

Vc = capacitor voltage
standby
power
supply

charge

discharge

gate

patient

switch is under
operator control

energy
storage

timing
circuitry

applies shock about 20 ms after
QRS complex, avoids T-wave

ECG
monitor
 AED is a portable electronic device that automatically diagnoses the ventricular fibrillation
in a patient.
Automatic refers to the ability to autonomously
analyse the patient's condition.
 AED is a type of external defibrillation process.
 AEDs require self-adhesive electrodes instead of hand
held paddles.

 The AED uses voice prompts, lights and text
tell the rescuer what steps have to take next.

messages to
 Turned on or opened AED.
 AED will instruct the user to:-

• Connect the electrodes (pads) to the patient.
• Avoid touching the patient to avoid false
readings by the unit.

• The AED examine the electrical output from
the heart and determine the patient is in a
shock able rhythm or not
When device determined that shock is warranted, it
will charge its internal capacitor in preparation to
deliver the shock.
 When charged, the device instructs the user to ensure
no one is touching the victim and then to press a red
button to deliver the shock.
 Many AED units have an 'event memory' which
store the ECG of the patient along with details of the
time the unit was activated and the number and
strength of any shocks delivered.
The paddles used in the procedure should not be
placed:• on a woman's breasts
• over an internal pacemaker patients.
 Before the paddle is used, a gel must be applied to
the patient's skin
• Skin burns from the defibrillator paddles are the
most common complication of defibrillation.
• Other risks include injury to the heart muscle,
abnormal heart rhythms, and blood clots.
• Attach the external and internal paddles if the monitor
reads, "No paddles."
• Check to ensure that the leads are securely attached if
the monitor reads, "No leads.“

• Connect the unit to AC power if the message
reads, "Low battery."
• Verify that the Energy Select control settings are correct
if the defibrillator does not charge.
• Change the electrodes and make sure that the electrodes
adapter cable is properly connected if you receive a
message of "PACER FAILURE." Restart the pacer.
• Close the recorder door and the paper roll if the monitor
message reads, "Check recorder”.






Willis A Tacker, “External Defibrillators,” in The
Biomedical Engineering Handbook, J. Bronzino (ed)
CRC Press, 1995.
www.google.com
en.wikipedia.org/wiki/Defibrillation
http://www.slideworld.org/viewslides.aspx/defibrillator

Más contenido relacionado

La actualidad más candente

La actualidad más candente (20)

Cardiac catheterization
Cardiac  catheterizationCardiac  catheterization
Cardiac catheterization
 
Defibrillator power point presentation for medical students
Defibrillator power point presentation for medical studentsDefibrillator power point presentation for medical students
Defibrillator power point presentation for medical students
 
Defibrillation
DefibrillationDefibrillation
Defibrillation
 
Pacemaker
PacemakerPacemaker
Pacemaker
 
IABP
IABPIABP
IABP
 
Cardiac monitor ppt
Cardiac monitor pptCardiac monitor ppt
Cardiac monitor ppt
 
Pulse oximeter
Pulse oximeterPulse oximeter
Pulse oximeter
 
PTCA
PTCAPTCA
PTCA
 
ECG
ECGECG
ECG
 
Echocardiogram
EchocardiogramEchocardiogram
Echocardiogram
 
Cardioversion
Cardioversion Cardioversion
Cardioversion
 
Syringe Pump
Syringe PumpSyringe Pump
Syringe Pump
 
INFUSION PUMPS
INFUSION PUMPSINFUSION PUMPS
INFUSION PUMPS
 
Central line
Central line Central line
Central line
 
Procedure of Recording ECG
Procedure of Recording ECGProcedure of Recording ECG
Procedure of Recording ECG
 
Pacemaker powerpoint presentation med surg
Pacemaker powerpoint presentation med surgPacemaker powerpoint presentation med surg
Pacemaker powerpoint presentation med surg
 
Stress test / Treadmill test
Stress test / Treadmill testStress test / Treadmill test
Stress test / Treadmill test
 
Cardiac monitoring
Cardiac monitoringCardiac monitoring
Cardiac monitoring
 
ENDOSCOPY
ENDOSCOPYENDOSCOPY
ENDOSCOPY
 
ACLS (4) (1).pptx
ACLS (4) (1).pptxACLS (4) (1).pptx
ACLS (4) (1).pptx
 

Destacado

defibrillation and cardioversion
 defibrillation and cardioversion defibrillation and cardioversion
defibrillation and cardioversion
Mohammed Awwad
 
Defibrillater
DefibrillaterDefibrillater
Defibrillater
mans4ani
 
Market Research Report : External defibrillator market in india 2014 - Sample
Market Research Report : External defibrillator market in india 2014 - SampleMarket Research Report : External defibrillator market in india 2014 - Sample
Market Research Report : External defibrillator market in india 2014 - Sample
Netscribes, Inc.
 

Destacado (11)

Defibrillation
DefibrillationDefibrillation
Defibrillation
 
defibrillation and cardioversion
 defibrillation and cardioversion defibrillation and cardioversion
defibrillation and cardioversion
 
Nursing case-study-burns-philnursingstudent
Nursing case-study-burns-philnursingstudentNursing case-study-burns-philnursingstudent
Nursing case-study-burns-philnursingstudent
 
Nursing case-study-dhf-dengue-hemorrhagic-fever-philnursingstudent
Nursing case-study-dhf-dengue-hemorrhagic-fever-philnursingstudentNursing case-study-dhf-dengue-hemorrhagic-fever-philnursingstudent
Nursing case-study-dhf-dengue-hemorrhagic-fever-philnursingstudent
 
Cardiac arrest
Cardiac arrest Cardiac arrest
Cardiac arrest
 
Defibrillation presentation
Defibrillation presentationDefibrillation presentation
Defibrillation presentation
 
Global Pharmaceuticals & Biotechnology Industry
Global Pharmaceuticals & Biotechnology IndustryGlobal Pharmaceuticals & Biotechnology Industry
Global Pharmaceuticals & Biotechnology Industry
 
Defibrillater
DefibrillaterDefibrillater
Defibrillater
 
Market Research Report : External defibrillator market in india 2014 - Sample
Market Research Report : External defibrillator market in india 2014 - SampleMarket Research Report : External defibrillator market in india 2014 - Sample
Market Research Report : External defibrillator market in india 2014 - Sample
 
Defibrillation, cardioversion and pacing
Defibrillation, cardioversion and pacingDefibrillation, cardioversion and pacing
Defibrillation, cardioversion and pacing
 
Pre operative care
Pre operative carePre operative care
Pre operative care
 

Similar a Defibrillator (ppt)

Similar a Defibrillator (ppt) (20)

Defibrillatorppt 131028115457-phpapp01
Defibrillatorppt 131028115457-phpapp01Defibrillatorppt 131028115457-phpapp01
Defibrillatorppt 131028115457-phpapp01
 
defibrillator.pptx
defibrillator.pptxdefibrillator.pptx
defibrillator.pptx
 
DEFIBRILLATION.pptx
DEFIBRILLATION.pptxDEFIBRILLATION.pptx
DEFIBRILLATION.pptx
 
Defibrillator
Defibrillator Defibrillator
Defibrillator
 
Defibrillators
DefibrillatorsDefibrillators
Defibrillators
 
Cardiovertor
CardiovertorCardiovertor
Cardiovertor
 
Defibrillators
DefibrillatorsDefibrillators
Defibrillators
 
Defibrillation
DefibrillationDefibrillation
Defibrillation
 
Defibrillator prakash
Defibrillator prakashDefibrillator prakash
Defibrillator prakash
 
Bio-Medical Therapeutic of in Pacemaker& Respiratory
Bio-Medical Therapeutic of in  Pacemaker& RespiratoryBio-Medical Therapeutic of in  Pacemaker& Respiratory
Bio-Medical Therapeutic of in Pacemaker& Respiratory
 
Defibrilator
DefibrilatorDefibrilator
Defibrilator
 
D & C 2012
D & C 2012D & C 2012
D & C 2012
 
Defibrillator.ppt
Defibrillator.pptDefibrillator.ppt
Defibrillator.ppt
 
Defibrillator.ppt
Defibrillator.pptDefibrillator.ppt
Defibrillator.ppt
 
Defibrillator.ppt
Defibrillator.pptDefibrillator.ppt
Defibrillator.ppt
 
Defibrillator.ppt
Defibrillator.pptDefibrillator.ppt
Defibrillator.ppt
 
Defibrillator ppt.pptx
Defibrillator ppt.pptxDefibrillator ppt.pptx
Defibrillator ppt.pptx
 
Dc shock
Dc shock  Dc shock
Dc shock
 
Medical Electronics - Hints for Slow Learner
Medical Electronics - Hints for Slow Learner Medical Electronics - Hints for Slow Learner
Medical Electronics - Hints for Slow Learner
 
Defibrillator.pptx
Defibrillator.pptxDefibrillator.pptx
Defibrillator.pptx
 

Último

Spellings Wk 3 English CAPS CARES Please Practise
Spellings Wk 3 English CAPS CARES Please PractiseSpellings Wk 3 English CAPS CARES Please Practise
Spellings Wk 3 English CAPS CARES Please Practise
AnaAcapella
 
Salient Features of India constitution especially power and functions
Salient Features of India constitution especially power and functionsSalient Features of India constitution especially power and functions
Salient Features of India constitution especially power and functions
KarakKing
 
Jual Obat Aborsi Hongkong ( Asli No.1 ) 085657271886 Obat Penggugur Kandungan...
Jual Obat Aborsi Hongkong ( Asli No.1 ) 085657271886 Obat Penggugur Kandungan...Jual Obat Aborsi Hongkong ( Asli No.1 ) 085657271886 Obat Penggugur Kandungan...
Jual Obat Aborsi Hongkong ( Asli No.1 ) 085657271886 Obat Penggugur Kandungan...
ZurliaSoop
 

Último (20)

Accessible Digital Futures project (20/03/2024)
Accessible Digital Futures project (20/03/2024)Accessible Digital Futures project (20/03/2024)
Accessible Digital Futures project (20/03/2024)
 
TỔNG ÔN TẬP THI VÀO LỚP 10 MÔN TIẾNG ANH NĂM HỌC 2023 - 2024 CÓ ĐÁP ÁN (NGỮ Â...
TỔNG ÔN TẬP THI VÀO LỚP 10 MÔN TIẾNG ANH NĂM HỌC 2023 - 2024 CÓ ĐÁP ÁN (NGỮ Â...TỔNG ÔN TẬP THI VÀO LỚP 10 MÔN TIẾNG ANH NĂM HỌC 2023 - 2024 CÓ ĐÁP ÁN (NGỮ Â...
TỔNG ÔN TẬP THI VÀO LỚP 10 MÔN TIẾNG ANH NĂM HỌC 2023 - 2024 CÓ ĐÁP ÁN (NGỮ Â...
 
Introduction to Nonprofit Accounting: The Basics
Introduction to Nonprofit Accounting: The BasicsIntroduction to Nonprofit Accounting: The Basics
Introduction to Nonprofit Accounting: The Basics
 
Spellings Wk 3 English CAPS CARES Please Practise
Spellings Wk 3 English CAPS CARES Please PractiseSpellings Wk 3 English CAPS CARES Please Practise
Spellings Wk 3 English CAPS CARES Please Practise
 
Application orientated numerical on hev.ppt
Application orientated numerical on hev.pptApplication orientated numerical on hev.ppt
Application orientated numerical on hev.ppt
 
Salient Features of India constitution especially power and functions
Salient Features of India constitution especially power and functionsSalient Features of India constitution especially power and functions
Salient Features of India constitution especially power and functions
 
Unit-V; Pricing (Pharma Marketing Management).pptx
Unit-V; Pricing (Pharma Marketing Management).pptxUnit-V; Pricing (Pharma Marketing Management).pptx
Unit-V; Pricing (Pharma Marketing Management).pptx
 
Sociology 101 Demonstration of Learning Exhibit
Sociology 101 Demonstration of Learning ExhibitSociology 101 Demonstration of Learning Exhibit
Sociology 101 Demonstration of Learning Exhibit
 
Unit-IV; Professional Sales Representative (PSR).pptx
Unit-IV; Professional Sales Representative (PSR).pptxUnit-IV; Professional Sales Representative (PSR).pptx
Unit-IV; Professional Sales Representative (PSR).pptx
 
Jual Obat Aborsi Hongkong ( Asli No.1 ) 085657271886 Obat Penggugur Kandungan...
Jual Obat Aborsi Hongkong ( Asli No.1 ) 085657271886 Obat Penggugur Kandungan...Jual Obat Aborsi Hongkong ( Asli No.1 ) 085657271886 Obat Penggugur Kandungan...
Jual Obat Aborsi Hongkong ( Asli No.1 ) 085657271886 Obat Penggugur Kandungan...
 
Making communications land - Are they received and understood as intended? we...
Making communications land - Are they received and understood as intended? we...Making communications land - Are they received and understood as intended? we...
Making communications land - Are they received and understood as intended? we...
 
Mixin Classes in Odoo 17 How to Extend Models Using Mixin Classes
Mixin Classes in Odoo 17  How to Extend Models Using Mixin ClassesMixin Classes in Odoo 17  How to Extend Models Using Mixin Classes
Mixin Classes in Odoo 17 How to Extend Models Using Mixin Classes
 
Micro-Scholarship, What it is, How can it help me.pdf
Micro-Scholarship, What it is, How can it help me.pdfMicro-Scholarship, What it is, How can it help me.pdf
Micro-Scholarship, What it is, How can it help me.pdf
 
Single or Multiple melodic lines structure
Single or Multiple melodic lines structureSingle or Multiple melodic lines structure
Single or Multiple melodic lines structure
 
Google Gemini An AI Revolution in Education.pptx
Google Gemini An AI Revolution in Education.pptxGoogle Gemini An AI Revolution in Education.pptx
Google Gemini An AI Revolution in Education.pptx
 
UGC NET Paper 1 Mathematical Reasoning & Aptitude.pdf
UGC NET Paper 1 Mathematical Reasoning & Aptitude.pdfUGC NET Paper 1 Mathematical Reasoning & Aptitude.pdf
UGC NET Paper 1 Mathematical Reasoning & Aptitude.pdf
 
ICT Role in 21st Century Education & its Challenges.pptx
ICT Role in 21st Century Education & its Challenges.pptxICT Role in 21st Century Education & its Challenges.pptx
ICT Role in 21st Century Education & its Challenges.pptx
 
Dyslexia AI Workshop for Slideshare.pptx
Dyslexia AI Workshop for Slideshare.pptxDyslexia AI Workshop for Slideshare.pptx
Dyslexia AI Workshop for Slideshare.pptx
 
HMCS Max Bernays Pre-Deployment Brief (May 2024).pptx
HMCS Max Bernays Pre-Deployment Brief (May 2024).pptxHMCS Max Bernays Pre-Deployment Brief (May 2024).pptx
HMCS Max Bernays Pre-Deployment Brief (May 2024).pptx
 
Holdier Curriculum Vitae (April 2024).pdf
Holdier Curriculum Vitae (April 2024).pdfHoldier Curriculum Vitae (April 2024).pdf
Holdier Curriculum Vitae (April 2024).pdf
 

Defibrillator (ppt)

  • 1.
  • 2. 1. Define defibrillation. 2. Describe Need and history of defibrillation. 3. Describe the principle and mechanism of defibrillation. 4. Types and classes of defibrillator 5. Describe the Automated external defibrillator 6. Identify the precautions and risk 7. Troubleshooting of defibrillator
  • 3.
  • 4. Definition: Defibrillation is a process in which an electronic device sends an electric shock to the heart to stop an extremely rapid, irregular heartbeat, and restore the normal heart rhythm. Defibrillation is a common treatment for life threatening cardiac dysrhythmias, ventricular fibrillation, and pulse less ventricular tachycardia.
  • 5. Ventricular fibrillation is a serious cardiac emergency resulting from asynchronous contraction of the heart muscles.  Due to ventricular fibrillation, there is an irregular rapid heart rhythm. Fig. Ventricular fibrillation Fig. Normal heart beat
  • 6.  Ventricular fibrillation can be converted into a more efficient rhythm by applying a high energy shock to the heart.  This sudden surge across the heart causes all muscle fibres to contract simultaneously.  Possibly, the fibres may then respond to normal physiological pace making pulses.  The instrument for administering the shock is called a DEFIBRILLATOR.
  • 7. Defibrillation is performed to correct lifethreatening fibrillations of the heart, which could result in cardiac arrest. It should be performed immediately after identifying that the patient is experiencing a cardiac emergency, has no pulse, and is unresponsive.
  • 8. Defibrillation was invented in by Prevost and Batelli, two Italian physiologists. They discovered that electric shocks could convert ventricular fibrillation to sinus rhythm in dogs.The first case of a human life saved by defibrillation was reported by Beck in 1947 .
  • 9.  Energy storage capacitor is charged at relatively slow rate from AC line.  Energy stored in capacitor is then delivered at a relatively rapid rate to chest of the patient.  Simple arrangement involve the discharge of capacitor energy through the patient‟s own resistance.
  • 10. Fig: Schematic diagram of a defibrillator
  • 11.  The discharge resistance which the patient represents as purely ohmic resistance of 50 to 100Ω approximately for a typical electrode size of 80cm2.  This particular waveform Fig is called „ Lown‟ waveform.  The pulse width of this waveform is generally 10 ms.
  • 13. • minimum defibrillation energy occurs for pulse durations of 3 - 10 ms (for most pulse shapes). • pulse amplitude in tens of amperes (few thousand volts).
  • 14. • operator selects energy delivered: 50-360 joules, depends on: – – – – – intrinsic characteristics of patient patient‟s disease duration of arrhythmia patient‟s age type of arrhythmia (more energy required for v. fib.)
  • 15.  Fibrillations cause the heart to stop pumping blood, leading to brain damage.  Defibrillators deliver a brief electric shock to the heart, which enables the heart's natural pacemaker to regain control and establish a normal heart rhythm.
  • 16.  Higher voltages are required for external defibrillation than for internal defibrillation.  A corrective shock of 750-800 volts is applied within a tenth of a second.  That is the same voltage as 500-533 no of AA batteries!
  • 18.  Occulsion of the coronary artery leads to ischemia  Ischemia leads to infarct which causes interruption of normal cardiac conduction  Infarct = VF/VT
  • 20.  Types of Defibrillator electrodes:a) Spoon shaped electrode • Applied directly to the heart. b) Paddle type electrode • Applied against the chest wall c) Pad type electrode • Applied directly on chest wall
  • 21. fig: Electrodes used in defibrillator (a) a spoon shaped internal electrode that is applied directly to the heart. (b) a paddle type electrode applied against the anterior chest wall.
  • 23. Anterior electrode pad Apex electrode pad Fig: anterior –apex scheme of electrode placement
  • 24. Monophasic pulse or waveform Bi-phasic pulse or waveform
  • 25.  There are two general classes of waveforms: a) mono-phasic waveform • Energy delivered in one direction through the patient‟s heart a) Biphasic waveform • Energy delivered in both direction through the patient‟s heart
  • 26. Fig:- Generation of bi-phasic waveform
  • 27.  The biphasic waveform is preferred over monophasic waveform to defibrillate. Why????? • A monophasic type, give a high-energy shock, up to 360 to 400 joules due to which increased cardiac injury and in burns the chest around the shock pad sites. • A biphasic type, give two sequential lowerenergy shocks of 120 - 200 joules, with each shock moving in an opposite polarity between the pads.
  • 29. Internal defibrillator a) • Electrodes placed directly to the heart • e.g..-Pacemaker b) External defibrillator • Electrodes placed directly on the heart • e.g..-AED
  • 30. • For each minute elapsing between onset of ventricular fibrillation and first defibrillation, survival decreases by 10%. • defibrillators should be portable, battery operated, small size. • energy in defibrillators usually stored in large capacitors. • total energy stored in capacitor: WC 1 CVC2 2 Vc = capacitor voltage
  • 31. standby power supply charge discharge gate patient switch is under operator control energy storage timing circuitry applies shock about 20 ms after QRS complex, avoids T-wave ECG monitor
  • 32.
  • 33.  AED is a portable electronic device that automatically diagnoses the ventricular fibrillation in a patient. Automatic refers to the ability to autonomously analyse the patient's condition.  AED is a type of external defibrillation process.
  • 34.  AEDs require self-adhesive electrodes instead of hand held paddles.  The AED uses voice prompts, lights and text tell the rescuer what steps have to take next. messages to
  • 35.  Turned on or opened AED.  AED will instruct the user to:- • Connect the electrodes (pads) to the patient. • Avoid touching the patient to avoid false readings by the unit. • The AED examine the electrical output from the heart and determine the patient is in a shock able rhythm or not
  • 36. When device determined that shock is warranted, it will charge its internal capacitor in preparation to deliver the shock.  When charged, the device instructs the user to ensure no one is touching the victim and then to press a red button to deliver the shock.  Many AED units have an 'event memory' which store the ECG of the patient along with details of the time the unit was activated and the number and strength of any shocks delivered.
  • 37. The paddles used in the procedure should not be placed:• on a woman's breasts • over an internal pacemaker patients.  Before the paddle is used, a gel must be applied to the patient's skin
  • 38. • Skin burns from the defibrillator paddles are the most common complication of defibrillation. • Other risks include injury to the heart muscle, abnormal heart rhythms, and blood clots.
  • 39. • Attach the external and internal paddles if the monitor reads, "No paddles." • Check to ensure that the leads are securely attached if the monitor reads, "No leads.“ • Connect the unit to AC power if the message reads, "Low battery." • Verify that the Energy Select control settings are correct if the defibrillator does not charge.
  • 40. • Change the electrodes and make sure that the electrodes adapter cable is properly connected if you receive a message of "PACER FAILURE." Restart the pacer. • Close the recorder door and the paper roll if the monitor message reads, "Check recorder”.
  • 41.     Willis A Tacker, “External Defibrillators,” in The Biomedical Engineering Handbook, J. Bronzino (ed) CRC Press, 1995. www.google.com en.wikipedia.org/wiki/Defibrillation http://www.slideworld.org/viewslides.aspx/defibrillator

Notas del editor

  1. Explain the cardiac conduction pathways with the electrical impulse originating in the SA node (right atrium) which travels to the AV node (division of the right and left ventricles), down along the right and left bundle branches to the Purkinje fibers.Intrinsic heart rates at the SA node are 80 bpm, the AV node is 60 bpm, and the ventricular rate (Purkinje fibers) is 40 bpm.