4. Gas-Filled Detectors I ; Ionization chamber region P ; Proportional region GM ; Geigur-Mueller region
5.
6. Ionization Chamber HV + - Negative ion Positive ion 1234 Electrometer The response is proportional to ionization rate (activity, exposure rate)
7.
8.
9. Dose calibrators SC97 Proportionality between the number of photons emitted and the ionization current Well-shaped ionization chamber filled with a gas of high atomic number (e.g. Xenon) and kept under pressure
10.
11. ACTIVITY MEASUREMENT Setting Measured activity Tc-99m 1.00 Co-57 1.19 In-111 2.35 Tl-201 1.76 Ga-67 1.12 I-123 2.19 I-131 1.43 Measured activity/True activity of Tc.99m if the indicated settings are used
12. Geometric efficiency The quotient: number of photons reaching the detector over the number of photons emitted from the sample Increasing geometric efficiency
16. Sealed sources for calibration of activity meters 1.9 5.27 y 1173, 1332 Co-60 7.4 30 y 662 Cs-137 9.3 10.7 y 81, 356 Ba-133 185 271 d 122 Co-57 Activity (MBq) Half-life Photon energy (keV) Radionuclide
17. Measurement of precision and accuracy Source (sealed) : Cs-137 or Co-57 Procedure: Select settings for the radionuclide and adjust background. Insert source in holder and make 10 measurements. Data analysis : To assess precision , calculate for each source (i) the percentage difference between the measured activity A i and their mean A mv . (+/-5%) To assess accuracy , calculate the percentage difference between the mean activity and the certified activity. (+/- 10%).
18. Measurement of reproducibility Measure the activity of a sealed reference source e.g. every morning. Use Tc-99m settings.
19.
20. Measurement of linearity Use a radionuclide with short half-life e.g. Tc-99m Make repeated measurements during several half-lives.
21.
22.
23.
24.
25. Examples of use of sample counters RIA 125 I Kidney clearance 51 Cr Vitamin B12 deficiency 57 Co, 58 Co Ferrokinetic studies 59 Fe Total body water 3 H Blood volume 125 I, 51 Cr, 99m Tc Biomedical research 3 H, 14 C
26. Gamma counter HV Ampl. PHA Timer Scaler Rate- meter Gain Base Window Voltage Detector Sample Lead shield PM-tube
28. Pulse height analyzer UL LL Time Pulse height (V) The pulse height analyzer allows only pulses of a certain height (energy) to be counted. counted not counted
41. Thank you Pawitra Masa-at 4937092 SIRS/M March 23, 2007
Notas del editor
This image shows the importance to use the correct setting of the instrument. It should be read as follows; If the sample to measure contains Tc99m and the setting of the activity meter is for e.g. I-123 then the measured activity will be 2.19 times the correct value.
The image should be used to point out the importance of reproducible geometry .
This is an example of an automatic sample holder which guaranties reproducible geometry
Another way to document the result is by using the first measurement as standard and then calculate the expected activity at different times and hence plot the measured activity vs the calculated activity. The measurements can also be performed using specially designed attenuators, meaning that all the measurements can be made within a very short time.
It is important to explain why there is a proportionality between the photon energy absorbed in the detector and the pulse height
Explain the origin of the scattered photons (detector and sample) and the completely absorbed photons.
Why use a limited window (B) instead of one that covers the whole pulse height distribution (A)?