MRI-BASIC
PRINCIPLE/TECHNIQUE/READIN
G
Nuclear magnetic resonance
►Dr Isidor Rabi (Nobel in 1944), discovered NMR
(Nuclear Magnetic Resonance) in the late 1930s, but
considered it to be an artefact of his apparatus
►CJ Gorter, coined the term ‘Nuclear Magnetic
Resonance’ in 1942.
►Bloch and Purcell (Nobel Prize in 1952 )if Certain
nuclei are placed in magnetic field ,absorb energy in
electromagnetic spectrum and re emit energy when
regain their original position.
MRI HARDWARE
 Permanent magnets
 Resistive magnets: In a resistive magnet, an
electrical current is passed through a loop of wire
and generates a magnetic field.
 Superconducting magnets: Superconducting
magnets are the ones most widely used in MR
machines at the present time. They also make
use of electricity, but they have a special current
carrying conductor.
 This is cooled down to superconducting
temperature
(about 4° K or -269° C). At this temperature, the
current conducting material loses its resistance
for electricity.
 In MRI radio frequency coils are necessary to
send in the RF pulse to excite the protons, and to
receive the resulting signal. The same or different
coils can be used for transmission of the RF
pulse and receiving the signal.
Necessary Equipment
Magnet Gradient Coil RF Coil
Source: Joe Gati, photos
RF Coil
3T magnet
gradient coil
(inside)
MRI principle
 MRI is based on the principle of nuclear magnetic
resonance (NMR)
 Two basic principles of NMR:
1.Atoms with an odd number of protons or neutrons
have spin
2.A moving electric charge, be it positive or negative,
produces a magnetic field
Protons
 Protons possess a positive charge.
 Like the earth they are constantly turning around
an axis and have their own magnetic field.
Protons
 Normally protons are aligned in a random
fashion. This, however, changes when they are
exposed to a strong external magnetic field.
 Then they are aligned in only two ways, either
parallel or antiparallel to the external magnetic
field
Precession
 Protons perform a wobbling type of motion in a
strong magnetic field ,called precession.
Precession Frequency
 To calculate the precession frequency.
Larmor equation:
 B0 = is the strength of the external magnetic field,
which is given in Tesla (T)
 is the angular frequency, and ‘y’is the so-
called gyromagnetic ratio.
EFFECT OF MAGNETIC FIELD
 The z-axis runs in the direction of the magnetic
field lines,
 The five protons, which "point" down cancel out
the magnetic effects of the same number of
protons, which "point"up
Radio Frequency (RF)
 A short burst of some electromagnetic wave, which is
called a radio frequency (RF) pulse.
 The purpose of this RF pulse is to disturb the protons,
which are peacefully precessing in alignment with the
external magnetic field.
 Not every RF pulse disturbs the alignment of the protons.
For this, we need a special RF pulse, one that can
exchange energy with the protons.
 Only when the RF pulse and the protons have the same
frequency, can protons can pick up some energy from the
radio wave, a phenomenon called Resonance.
Longitudinal magnetization
 the vectors along the z-axis point in the same
direction, and thus add up to a new magnetic sum
vector pointing up. As this magnetization is in
direction along/longitudinal to the external
magnetic field, it is also called longitudinal
magnetization
EFFECT OF RF PULSE ON
MAGNETIZATION
 The radiofrequency pulse exchanges energy with
the protons (a), and some of them are lifted to a
higher level of energy, pointing downward in the
illustration (b).
 In effect the magnetization along the z-axis
decreases, as the protons which point down
"neutralize" the same number of protons pointing
up.
Transversal Magnetization
 When the protons randomly point left/right, back/forth
and so on, they also cancel their magnetic forces in
these directions
 The RF pulse synchronize them - "in phase". They
now point in the same direction at the same time, and
thus their magnetic vectors add up in this direction.
This results in a magnetic vector pointing to the side
to which the precessing protons point, and this is in a
transverse direction.
 This is why it is called transversal magnetization.
 As the transversal magnetic vector moves around
with the precessing protons, it comes towards the
antenna, goes away from it, comes towards it
again and so on, also with the precession
frequency.
 The resulting MR signal therefore also has the
precession.
Longitudinal and transversal
relaxation
 The newly established transverse magnetization
starts to disappear (a process called transversal
relaxation), and the longitudinal magnetization
grows back to its original size (a process called
longitudinal relaxation).
Longitudinal/Spin lattice
relaxation
 After the RF pulse is switched off, protons go
back from their higher to the lower state of
energy, i.e. point up again.
 This energy is just handed over to their
surroundings, the so called lattice. And this is why
this process is not only called longitudinal
relaxation, but also spin-lattice-relaxation
T1-Curve
 If one plots the longitudinal magnetization vs.
time after the RF pulse was switched off, one gets
a so-called T1-curve.
SPIN SPIN RELAXATION
 After the RF pulse is switched off, protons lose
phase coherence, they get out of step thus
transversal magnetization decreases.
 Plot showing transversal magnetization vs. time after
the RF pulse is switched off, one gets a curve as
illustrated,which is called a T2-curve.
Characteristics of T1 and T2
 T1 is about 300 to 2000 msec, and T2 is about 30
to 150 msec.
 It is difficult to pinpoint the end of the longitudinal
and transversal relaxation exactly.
 T1 =63% of the original longitudinal
magnetization is reached.
 T2 =37% of the original value,T1 is longer than
T2
 T1 varies with the magnetic field strength; it is
longer in stronger magnetic fields.
90°pulse
 If after the RF pulse, the number of protons on
the higher energy level equals the number of
protons on the lower energy level, longitudinal
magnetization disappears, and there is only
transversal magnetization due to phase
coherence.
 The magnetic vector seems to have been "tilted“
90° to the side. The corresponding RF pulse is
thus also called a 90° pulse
TE vs TR
 • TE: the time between the 90° pulse and the
echo.
 • TR: the time between two pulse sequences, i.e.
from one 90° pulse to the next.
HOW TR AFFECTS THE SIGNAL
INTENSITY OF TISSUE
 A and B are two tissues with different relaxation
times.
 Frame 0 shows the situation before, frame 1
immediately after a 90° pulse.
 When we wait for a long time (TRlong) the
longitudinal magnetization of both tissues will
have totally recovered (frame 5).
 A second 90° pulse after this time results in the
same amount of transversal magnetization
(frame 6) for both tissues, as was observed
 When we do not wait as long , but send in the
second RF pulse after a shorter time (TRShort),
longitudinal magnetization of tissue B, which has
the longer T1, has not recovered as much as that
of tissue A with the shorter T1.
 The transversal magnetization of the two tissues
after the second RF pulse will then be different
(frame 5). Thus, by changing the time between
successive RF pulses, we can influence and
modify magnetization and the signal intensity of
TR AND SIGNAL INTENSITY
 Brain has a shorter longitudinal relaxation time
than CSF.
 With a short TR the signal intensities of brain and
CSF differ more than after a long TR.
T2*/T2effect/spin echo sequence
 After the RF pulse is switched off, the protons
dephase (a-c).
 The 180°pulse causes them to precess in the
opposite direction and so they rephase again (d-f).
T2* curve
 The 180° pulse refocusses the dephasing protons which results in a
stronger signal, the spin echo after the time TE.
 The protons then dephase again and can be refocussed another time
by a 180° pulse and so on. Thus it is possible to obtain more than
one signal, more than one spin echo.
 The spin echoes, however, differ in intensity due to so-called T2-
effects.
 A curve connecting the spin echo intensities is the T2 curve. If we did
not use the 180° pulse, the signal intensity would decay much faster.
A curve describing the signal intensity in that case is the T*2 (T2
star) curve, The type of pulse
sequence,
that we used in our
experiment, is called a
spin echo sequence,
consisting of a 90°
pulse and a 180°
pulse
 T2-curves for two tissues with different
transversal relaxation times; tissue A has a
shorter T2 than tissue B, thus loses transversal
magnetization faster.
 With a short TE (TEshort) the difference in signal
intensity is less pronounced than after a longer
TE (TElong).
Short vs Long TE/TR
 A TR of less than 500 msec is considered to be
short, a TR of more than 1500 msec to be long
 A short TE is one that is as short as possible, a
long TE also is about 3 times as long.
 A TE of less than 30 msec is considered to be
short, a TE greater than 80 msec to be long.
spin echo pulse
sequence
 1. (90° - TE/2 -180° - TE/2 -> record signal at TE)
 after TR (time from the beginning of one 90°
pulse to the next 90° pulse) follows another pulse
cycle and signal b measurement:
 2. (90° - TE/2 -180° - TE/2 -> record signal at TE)
EFFECT OF TR/TE ON IMAGE
 By combining T1- and T2-curves signal intensity of certain
tissues can be determined for a pulse sequence using TR and
TE as illustrated.
 With a long TR, differences in T1, in longitudinal magnetization
time are not very important any more, as all tissues have
regained their full longitudinal magnetization.
 When we only wait a very short TE then differences in signal
intensity due to differences in T2 have not yet had time to
become pronounced.
 The resulting picture is thus neither T1- nor T2-weighted, but
mostly determined by the proton density of the tissues (for this,
ideally TE should be zero).
T2-weighted picture
 When we wait a long TR and a long TE, differences
in T2 have had time enough to become pronounced,
the resulting picture is T2-weighted
T1-weighted picture
 When we wait a shorter time TR, differences in T1
influence tissue contrast to a larger extent, the picture is
T1-weighted, especially when we also wait a short TE
(when signal differences due to differing T2s have not
had time to become pronounced)
 T2-curves of different tissues can intersect. The signal intensity of
the tissues is reversed choosing a TE beyond the crossing point
(TEC): before this crossing point (e.g. at TE1) tissue A has a higher
signal intensity than tissue B.
 This means that image contrast is still determined by differences in
T1: the tissue A with the shorter T1 has the stronger signal intensity.
 At TEC both tissues have the same signal intensity, and thus cannot
be differentiated.
 After the crossing point (e.g. at TE2) the relative signal intensities are
reversed, and tissue B has the stronger signal
T1W/T2W IMAGES
Useful for: Evaluating
anatomic detail
CSF: Dark
White Matter: White
Gray Matter: Gray
Vessels: Dark
Useful for: Looking at areas of
edema & pathology
CSF: Bright
White Matter: Gray
Gray Matter: Lighter than white
matter
Vessels: Dark
T1 Recovery
Short TR T1 contrast
(T1 Weighted)
 TR 300-600 ms
 TE 10-30 ms
Bright on T1
 Fat, subacute hemorrhage, melanin, protein rich fluid.
 Slowly flowing blood
 Paramagnetic substances(gadolinium,copper,manganese)
 9
Dark on T1
 Edema, tumor, infection, inflammation,
hemorrhage(hyperacute, chronic)
 Low proton density, calcification
 Flow void
T2 Decay
Long TE T2 contrast
(T2 Weighted)
 TR 2000 ms
 TE 70 ms
Bright on T2
 Edema, tumor, Infection, inflammation, subdural
collection
 Met hemoglobin in late sub acute hemorrhage
Dark on T2
 Low proton density,calcification,fibrous tissue
 Paramagnetic substances(deoxy
hemoglobin,methemoglobin(intracellular),ferritin,h
emosiderin,melanin.
 Protein rich fluid
 Flow void
Which scan best defines the
abnormality
 T1 W Images:
 Subacute Hemorrhage
 Fat-containing structures
 Anatomical Details
 T2 W Images:
 Edema
 Demyelination
 Infarction
 Chronic Hemorrhage
 FLAIR Images:
 Edema,
 Demyelination
 Infarction esp. in Periventricular location
Infarct
 Acute : T1W –Isointense hypo intense
T2W-Hyper intense
 Sub acute: T1W-Low signal,increasedsignal in
peripheral region..hemorrhage(metHb)
T2W- High signal
 Chronic:T1W-low signal
T2W-High siignal
Hemorrhage
FLOW VOID-HOW
 Flow effects are responsible for the black
appearance of flowing blood, the signal void in blood
vessels
 Paramagnetic substances
like Gadolinium shorten
the T1 and the T2 of their
surroundings.
 The respective T1- and
T2-curves are shifted
towards the left.
 In effect, that means that
for a certain TR there is
more, for a certain TE
there is less signal
EFFECT OF CONTRAST MATERIAL
When tissue A and B has less
contrast:
 T1-curves for tissue A and B are
very close to each other,
resulting in only a small
difference in signal intensity
between the tissues at TR.
 NOW- the T1-curve of tissue A
is shifted to the left, as contrast
agent entered tissue A but not
tissue B. At the same time TR
there now is a much greater
difference in signal intensity, i.e.
tissue contrast
The inversion recovery
sequence
 The inversion recovery sequence uses a 180° pulse which
inverts the longitudinal magnetization, followed by a 90° pulse
after the time TI.
 The 90° pulse "tilts“ the magnetization into the transversal (x-y-)
plane, so it can be measured/received.
 The tissue in the bottom row goes back to its original longitudinal
magnetization faster, thus has the shorter T1.
 For the time TI, which is illustrated, this results in less transversal
magnetization after the 90° pulse
Short TI inversion-recovery (STIR)
sequence
Longer TE used..both long T1 and T2
…bright
 Substances having
both LONG T1 and
T2 will be bright.
 T1 and T2 of most
pathologic lesion
are prolonged.
 And substances
with short T1 will be
suppressed eg.
hemorrhage, gd-
enhancement.
Fluid-attenuated inversion recovery
(FLAIR)
 First described in 1992 and has become one of
the corner stones of brain MR imaging protocols
 An IR sequence with a long TR and TE and an
inversion time (TI) that is tailored to null the signal
from CSF
 Water bound to complex molecule with in plaque
has relatively shorter T1 than free water with in
ventricle.
 Long inversion time effectively suppress free
water eg./csf.
 Lesion that contain complex, partially bound
water (less mobile)…shorter T1 than free water
appears bright.
 Effective in high lightening lesion eg
demyelination, stroke, Ischemic gliosis and
tumor.
 Useful in diffrentiating acute infarct from cystic
encephalomalacia
 Useful in SAH …removes CSF signal
GRE
GRE
 In a GRE sequence, an RF pulse is applied that
partly flips the NMV into the transverse plane
(variable flip angle).
 Gradients, as opposed to RF pulses, are used to
dephase (negative gradient) and rephase
(positive gradients) transverse magnetization.
 GRE Sequences contd:
 This feature of GRE sequences is exploited- in
detection of hemorrhage, as the iron in Hb becomes
magnetized locally (produces its own local magnetic
field) and thus dephases the spinning nuclei.
 The technique is particularly helpful for diagnosing
hemorrhagic contusions such as those in the brain .
GREFLAIR
Hemorrhage in right parietal lobe
DWI & ADC
Diffusion-weighted MRI
 DWI images is obtained by applying pairs of
opposing and balanced magnetic field gradients (but
of differing durations and amplitudes) around a spin-
echo refocusing pulse of a T2 weighted sequence.
 Stationary water molecules are unaffected by the
paired gradients, and thus retain their signal.
 Non stationary water molecules acquire phase
information from the first gradient, but are not re
phased by the second gradient, leading to an overall
loss of the MR signal
 The normal motion of water molecules within living
tissues is random (brownian motion).
 In acute stroke, there is an alteration of
homeostasis
 Acute stroke causes excess intracellular water
accumulation, or cytotoxic edema, with an overall
decreased rate of water molecular diffusion within
the affected tissue.
 Tissues with a higher rate of diffusion undergo a
greater loss of signal in a given period of time than
do tissues with a lower diffusion rate.
 Therefore, areas of cytotoxic edema, in which the
motion of water molecules is restricted, appear
brighter on diffusion-weighted images because of
lesser signal losses
Apparent Diffusion Coefficient
 Calculated by acquiring two or more images with
a different gradient duration and amplitude .
 To differentiate T2 shine through effects or
artifacts from real ischemic lesions.
ADC
 useful for estimating the lesion age and
distinguishing acute from subacute DWI lesions.
 Acute ischemic lesions can be divided into
 1- hyperacute lesions (low ADC and DWI-
positive)
 2-subacute lesions (normalized ADC).
 3-Chronic lesions can be differentiated from
acute lesions by normalization of ADC and DWI.
Nonischemic causes for decreased
ADC
 Abscess
 Lymphoma and other tumors
 Multiple sclerosis
 Seizures
 Metabolic (Canavans )
Evaluation of acute stroke on DWI
 The DWI and ADC maps show changes in
ischemic brain within minutes to few hours
 The signal intensity of acute stroke on DW
images increase during the first week after
symptom onset and decrease thereafter, but
signal remains hyper intense for a long period
(up to 72 days in the study by Lausberg et al)
 The ADC values decline rapidly after the onset
of ischemia and subsequently increase from
dark to bright 7-10 days later .
 This property may be used to differentiate the
lesion older than 10 days from more acute ones .
 Chronic infarcts are characterized by elevated
diffusion and appear hypo, iso or hyper intense
on DW images and hyperintense on ADC maps
DW MR imaging characteristics of Various Disease Entities
MR Signal Intensity
Disease DW Image ADC Image ADC Cause
Acute Stroke High Low Restricted Cytotoxic edema
Chronic Strokes Variable High Elevated Gliosis
Hypertensive
encephalopathy
Variable High Elevated Vasogenic edema
Arachnoid cyst Low High Elevated Free water
Epidermoid mass High Low Restricted Cellular tumor
Herpes encephalitis High Low Restricted Cytotoxic edema
CJD High Low Restricted Cytotoxic edema
MS acute lesions Variable High Elevated Vasogenic edema
Chronic lesions Variable High Elevated Gliosis
Slice selection and thickness
 There are two ways to determine slice
thickness.
(a). By using certain bandwidth of RF pulse
eg. If frequencies between 64 and 65 mHz,
protons in slice 1 will be influenced by the
RF pulse.
 When the RF pulse only contains
frequencies between 64 mHz and 64.5
mHz, thus has a smaller bandwidth, slice 2,
which is half as thick as slice 1 will be
imaged.
 When there is more difference in magnetic
field strength between the level of the feet
and the head, i.e. The magnetic gradient is
steeper, the resulting slice will be thinner,
even though the RF pulse bandwidth is the
same.
HOW TO READ
THE IMAGING PLANES
- Axial plane: Transverse images represent
"slices" of the body
- Sagittal plane: Images taken
perpendicular to the axial plane which
separate the left and right sides (lateral
view)
- Coronal plane: Images taken
perpendicular to the sagittal plane which
separate the front from the back. (frontal
view)
How do you describe abnormalities
on MR?
Hyperintense (more intense): If an abnormality is bright (white) on MR,
we describe it as hyperintense.
Isointense (the same intensity): If an abnormality is the same intensity
to a reference structure, we describe it as isointense.
Hypointense (less intense): If an abnormality is dark on MR .
IT IS CT OR MRI?
CT: Computed Tomography MRI: Magnetic Resonance Imaging
Imaging Plane: in the axial
plane. The axial data set can then be
used to reconstruct images in other
planes, sagittal and coronal are the
most common.
Windows: "brain window" AND ‘bone
window ‘
White and Black:air within the
sinuses is black, the brain
parenchyma has a gray appearance
and the skull is bright white
Imaging Plane: in any plane, not just
axial
Sequences: different type of image
is referred to a sequence.
White and Black:same structure
may be bright or dark depending on
the type of sequence; CSF for
example is bright on T2, but dark on
T1. The tissue and imaging
characteristics are a lot more
complicated than CT.
Types of MR Images
Useful for: Evaluating anatomic
detail
CSF: Dark
White Matter: White
Gray Matter: Gray
Vessels: Dark
Useful for: Evaluating for BBB breakdown
in the setting of tumor, infection, MS etc.
CSF: Dark
White Matter: White
Gray Matter: Gray
Vessels: Brigh
Useful for: Looking at areas of edema &
pathology
CSF: Bright
White Matter: Gray
Gray Matter: Lighter than white matter
Vessels: Dark
Useful for: Evaluating areas of edema with
CSF subtraction. Edema stands out
because is CSF dark
CSF: Dark
White Matter: Gray
Gray Matter: Lighter than white matter
Vessels: Dark
Useful for: stroke imaging, abscess,
cellular tumors
CSF: Dark
White Matter: Gray
Gray Matter: Lighter than white matter
Fuzzier image than FLAIR
T2* (T2-star, or SWI)
Form of T2-weighted image which is
susceptible to iron or calcium
Blood, bone, calcium appear dark
Area of blood often appears much larger
than reality(“blooming”)
Useful for: Identification of early
hemorrhage
Look for: DARK only
Recognition:
o Like T2 except
o Cranium, scalp are dark or absent
o Dark areas near frontal and temporal
bones
Apparent Diffusion Coefficient (ADC
Map)
Contains actual data relevant to diffusion
image
Areas of restricted diffusion are dark
Useful for:
o Excluding T2-shine through
o Real restricted diffusion is bright on DWI,
dark
on ADC
Look for: DARK only
Recognition
o Images marked ADC
o Grainy dark images
MR ARTIFACTS
Artifacts in MR images refer to pixels that do not faithfully represent the
anatomy being studied.
1 Motion Artifacts- Motion artifacts occur as a result of movement of
tissue during the data acquisition period.
2.Sequence/Protocol-Related Artifacts- results from the specific
measurement process
used to acquire the image.
subtypes
2a. Aliasing
2b. Chemical Shift Artifacts
2c. Phase Cancellation Artifact
2d. Truncation Artifacts
2e. Coherence Artifacts
2f. Magnetic-Susceptibility Difference Artifacts
3. External Artifacts. External artifacts are generated from sources
other than patient tissue.
3a. Magnetic Field Distortions
3b. Measurement Hardware
3c. Noise
refrences
1.MRI: Basic Principles and Applications, Third Edition, by M. A. Brown
and R. C. Semelka.ISBN 0-471-43310-1 © 2003 John Wiley & Sons,
Inc.
2.Schering: MRI made easy,by Hans H. Schild Nationales Druckhaus
Berlin.ISBN 3-921817-41-
3. Bradley’s text book of neurology-6th edition
4.Robert A Ziemarman’s Neuroimaging clinical and physical principle.
5. https://sites.google.com/a/wisc.edu/neuroradiology/image-
acquisition/the-basics
THANK YOU

Más contenido relacionado

La actualidad más candente

La actualidad más candente (20)

MRI physics
MRI physicsMRI physics
MRI physics
 
Basic Pulse Sequences In MRI
Basic Pulse Sequences In MRIBasic Pulse Sequences In MRI
Basic Pulse Sequences In MRI
 
MRI Parameters for Imaging
MRI Parameters for ImagingMRI Parameters for Imaging
MRI Parameters for Imaging
 
Contrast media in CT
Contrast media in CTContrast media in CT
Contrast media in CT
 
Mri physics
Mri physicsMri physics
Mri physics
 
BASIC MRI SEQUENCES
BASIC MRI SEQUENCESBASIC MRI SEQUENCES
BASIC MRI SEQUENCES
 
Introduction to mri
Introduction to mriIntroduction to mri
Introduction to mri
 
Slip ring
Slip ringSlip ring
Slip ring
 
The Physics of MRI
The Physics of MRIThe Physics of MRI
The Physics of MRI
 
Mri basics
Mri basicsMri basics
Mri basics
 
Basic principles of magnetic resonance imaging for beginner - Dr Sanjana Ravi...
Basic principles of magnetic resonance imaging for beginner - Dr Sanjana Ravi...Basic principles of magnetic resonance imaging for beginner - Dr Sanjana Ravi...
Basic principles of magnetic resonance imaging for beginner - Dr Sanjana Ravi...
 
Basic of mri
Basic of mriBasic of mri
Basic of mri
 
MRI INSTRUMENTATION/ HARDWARE
MRI INSTRUMENTATION/ HARDWAREMRI INSTRUMENTATION/ HARDWARE
MRI INSTRUMENTATION/ HARDWARE
 
MRI
MRIMRI
MRI
 
Gradient echo pulse sequence and its application
Gradient echo pulse sequence and its applicationGradient echo pulse sequence and its application
Gradient echo pulse sequence and its application
 
GENERATIONS OF COMPUTED TOMOGRAPHY
GENERATIONS OF COMPUTED TOMOGRAPHY GENERATIONS OF COMPUTED TOMOGRAPHY
GENERATIONS OF COMPUTED TOMOGRAPHY
 
CT Physics
CT PhysicsCT Physics
CT Physics
 
Mri image quality gamal mahdaly
Mri image quality gamal mahdalyMri image quality gamal mahdaly
Mri image quality gamal mahdaly
 
Mri components
Mri componentsMri components
Mri components
 
Mri physics PART-1
Mri physics PART-1Mri physics PART-1
Mri physics PART-1
 

Destacado (20)

MRI sequences
MRI sequencesMRI sequences
MRI sequences
 
Basics Of MRI
Basics Of MRIBasics Of MRI
Basics Of MRI
 
Mri basics
Mri basicsMri basics
Mri basics
 
BASICS OF MRI
BASICS OF MRIBASICS OF MRI
BASICS OF MRI
 
Mri ppt
Mri pptMri ppt
Mri ppt
 
Basics of mri
Basics of mriBasics of mri
Basics of mri
 
Magnetic Resonance Imaging
Magnetic Resonance ImagingMagnetic Resonance Imaging
Magnetic Resonance Imaging
 
MRI basic sequences
MRI basic sequencesMRI basic sequences
MRI basic sequences
 
Basics of MRI
Basics of MRIBasics of MRI
Basics of MRI
 
MRI physics basic concepts
MRI physics basic conceptsMRI physics basic concepts
MRI physics basic concepts
 
Computer Tomography (CT Scan)
Computer Tomography (CT Scan)Computer Tomography (CT Scan)
Computer Tomography (CT Scan)
 
MRI Basics by Saleh Asad
MRI  Basics  by Saleh AsadMRI  Basics  by Saleh Asad
MRI Basics by Saleh Asad
 
Mrisequences 130118064505-phpapp02
Mrisequences 130118064505-phpapp02Mrisequences 130118064505-phpapp02
Mrisequences 130118064505-phpapp02
 
Mri basics
Mri basicsMri basics
Mri basics
 
Mri basic sequences
Mri basic sequencesMri basic sequences
Mri basic sequences
 
MRI Physics RV
MRI Physics RVMRI Physics RV
MRI Physics RV
 
Mri physics ii
Mri physics iiMri physics ii
Mri physics ii
 
MRI
MRIMRI
MRI
 
MRI
MRIMRI
MRI
 
Basicunderstandingonmagneticresonanceimagingmri 141231045409-conversion-gate01
Basicunderstandingonmagneticresonanceimagingmri 141231045409-conversion-gate01Basicunderstandingonmagneticresonanceimagingmri 141231045409-conversion-gate01
Basicunderstandingonmagneticresonanceimagingmri 141231045409-conversion-gate01
 

Similar a Mri basic principle and sequences

Similar a Mri basic principle and sequences (20)

MRI PHYSICS FOR RADIOLOGY RESIDENTS INDIA
MRI PHYSICS FOR RADIOLOGY RESIDENTS INDIAMRI PHYSICS FOR RADIOLOGY RESIDENTS INDIA
MRI PHYSICS FOR RADIOLOGY RESIDENTS INDIA
 
Basics of MRI Physics
Basics of MRI PhysicsBasics of MRI Physics
Basics of MRI Physics
 
MRI 3-Marks QUESTIONS AND ANSWERS
MRI 3-Marks QUESTIONS AND ANSWERSMRI 3-Marks QUESTIONS AND ANSWERS
MRI 3-Marks QUESTIONS AND ANSWERS
 
NMR spectroscopy by roshan bodhe
NMR spectroscopy by roshan bodheNMR spectroscopy by roshan bodhe
NMR spectroscopy by roshan bodhe
 
BASIC MRI SEQUENCES
BASIC MRI SEQUENCESBASIC MRI SEQUENCES
BASIC MRI SEQUENCES
 
Basic principles of mri
Basic principles of mriBasic principles of mri
Basic principles of mri
 
Mri basic principles
Mri basic principlesMri basic principles
Mri basic principles
 
mribasicprinciples-19091912334002451.pdf
mribasicprinciples-19091912334002451.pdfmribasicprinciples-19091912334002451.pdf
mribasicprinciples-19091912334002451.pdf
 
mri physics.pptx
mri physics.pptxmri physics.pptx
mri physics.pptx
 
nmr
nmrnmr
nmr
 
MRI PHYSICS.pptx
MRI PHYSICS.pptxMRI PHYSICS.pptx
MRI PHYSICS.pptx
 
Magnetic Resonance Imaging
Magnetic Resonance ImagingMagnetic Resonance Imaging
Magnetic Resonance Imaging
 
Magnetic Resonance Imaging-An Overview
Magnetic Resonance Imaging-An OverviewMagnetic Resonance Imaging-An Overview
Magnetic Resonance Imaging-An Overview
 
Ppt mri brain
Ppt mri brainPpt mri brain
Ppt mri brain
 
Unit 2
Unit 2Unit 2
Unit 2
 
MRI obstetric practice part 1 Basic Physics
MRI obstetric practice part 1 Basic PhysicsMRI obstetric practice part 1 Basic Physics
MRI obstetric practice part 1 Basic Physics
 
Magnetic Resonance Imaging
Magnetic Resonance ImagingMagnetic Resonance Imaging
Magnetic Resonance Imaging
 
Magnetic resonance imaging
Magnetic resonance imagingMagnetic resonance imaging
Magnetic resonance imaging
 
NMR spectroscopy
NMR spectroscopyNMR spectroscopy
NMR spectroscopy
 
08-BAT_Lect_NMR.pdf
08-BAT_Lect_NMR.pdf08-BAT_Lect_NMR.pdf
08-BAT_Lect_NMR.pdf
 

Último

Ahmedabad Call Girls Book Now 9630942363 Top Class Ahmedabad Escort Service A...
Ahmedabad Call Girls Book Now 9630942363 Top Class Ahmedabad Escort Service A...Ahmedabad Call Girls Book Now 9630942363 Top Class Ahmedabad Escort Service A...
Ahmedabad Call Girls Book Now 9630942363 Top Class Ahmedabad Escort Service A...Genuine Call Girls
 
Call Girls Rishikesh Just Call 9667172968 Top Class Call Girl Service Available
Call Girls Rishikesh Just Call 9667172968 Top Class Call Girl Service AvailableCall Girls Rishikesh Just Call 9667172968 Top Class Call Girl Service Available
Call Girls Rishikesh Just Call 9667172968 Top Class Call Girl Service Availableperfect solution
 
Call Girls Kathua Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Kathua Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Kathua Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Kathua Just Call 8250077686 Top Class Call Girl Service AvailableDipal Arora
 
Bhawanipatna Call Girls 📞9332606886 Call Girls in Bhawanipatna Escorts servic...
Bhawanipatna Call Girls 📞9332606886 Call Girls in Bhawanipatna Escorts servic...Bhawanipatna Call Girls 📞9332606886 Call Girls in Bhawanipatna Escorts servic...
Bhawanipatna Call Girls 📞9332606886 Call Girls in Bhawanipatna Escorts servic...Dipal Arora
 
Goa Call Girl Service 📞9xx000xx09📞Just Call Divya📲 Call Girl In Goa No💰Advanc...
Goa Call Girl Service 📞9xx000xx09📞Just Call Divya📲 Call Girl In Goa No💰Advanc...Goa Call Girl Service 📞9xx000xx09📞Just Call Divya📲 Call Girl In Goa No💰Advanc...
Goa Call Girl Service 📞9xx000xx09📞Just Call Divya📲 Call Girl In Goa No💰Advanc...Sheetaleventcompany
 
💰Call Girl In Bangalore☎️63788-78445💰 Call Girl service in Bangalore☎️Bangalo...
💰Call Girl In Bangalore☎️63788-78445💰 Call Girl service in Bangalore☎️Bangalo...💰Call Girl In Bangalore☎️63788-78445💰 Call Girl service in Bangalore☎️Bangalo...
💰Call Girl In Bangalore☎️63788-78445💰 Call Girl service in Bangalore☎️Bangalo...gragneelam30
 
Call Girls Shahdol Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Shahdol Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Shahdol Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Shahdol Just Call 8250077686 Top Class Call Girl Service AvailableDipal Arora
 
Kolkata Call Girls Shobhabazar 💯Call Us 🔝 8005736733 🔝 💃 Top Class Call Gir...
Kolkata Call Girls Shobhabazar  💯Call Us 🔝 8005736733 🔝 💃  Top Class Call Gir...Kolkata Call Girls Shobhabazar  💯Call Us 🔝 8005736733 🔝 💃  Top Class Call Gir...
Kolkata Call Girls Shobhabazar 💯Call Us 🔝 8005736733 🔝 💃 Top Class Call Gir...Namrata Singh
 
7 steps How to prevent Thalassemia : Dr Sharda Jain & Vandana Gupta
7 steps How to prevent Thalassemia : Dr Sharda Jain & Vandana Gupta7 steps How to prevent Thalassemia : Dr Sharda Jain & Vandana Gupta
7 steps How to prevent Thalassemia : Dr Sharda Jain & Vandana GuptaLifecare Centre
 
💚Reliable Call Girls Chandigarh 💯Niamh 📲🔝8868886958🔝Call Girl In Chandigarh N...
💚Reliable Call Girls Chandigarh 💯Niamh 📲🔝8868886958🔝Call Girl In Chandigarh N...💚Reliable Call Girls Chandigarh 💯Niamh 📲🔝8868886958🔝Call Girl In Chandigarh N...
💚Reliable Call Girls Chandigarh 💯Niamh 📲🔝8868886958🔝Call Girl In Chandigarh N...Sheetaleventcompany
 
Premium Call Girls Dehradun {8854095900} ❤️VVIP ANJU Call Girls in Dehradun U...
Premium Call Girls Dehradun {8854095900} ❤️VVIP ANJU Call Girls in Dehradun U...Premium Call Girls Dehradun {8854095900} ❤️VVIP ANJU Call Girls in Dehradun U...
Premium Call Girls Dehradun {8854095900} ❤️VVIP ANJU Call Girls in Dehradun U...Sheetaleventcompany
 
Premium Call Girls Nagpur {9xx000xx09} ❤️VVIP POOJA Call Girls in Nagpur Maha...
Premium Call Girls Nagpur {9xx000xx09} ❤️VVIP POOJA Call Girls in Nagpur Maha...Premium Call Girls Nagpur {9xx000xx09} ❤️VVIP POOJA Call Girls in Nagpur Maha...
Premium Call Girls Nagpur {9xx000xx09} ❤️VVIP POOJA Call Girls in Nagpur Maha...Sheetaleventcompany
 
Call Girls in Lucknow Just Call 👉👉8630512678 Top Class Call Girl Service Avai...
Call Girls in Lucknow Just Call 👉👉8630512678 Top Class Call Girl Service Avai...Call Girls in Lucknow Just Call 👉👉8630512678 Top Class Call Girl Service Avai...
Call Girls in Lucknow Just Call 👉👉8630512678 Top Class Call Girl Service Avai...soniyagrag336
 
Low Cost Call Girls Bangalore {9179660964} ❤️VVIP NISHA Call Girls in Bangalo...
Low Cost Call Girls Bangalore {9179660964} ❤️VVIP NISHA Call Girls in Bangalo...Low Cost Call Girls Bangalore {9179660964} ❤️VVIP NISHA Call Girls in Bangalo...
Low Cost Call Girls Bangalore {9179660964} ❤️VVIP NISHA Call Girls in Bangalo...Sheetaleventcompany
 
❤️Call Girl Service In Chandigarh☎️9814379184☎️ Call Girl in Chandigarh☎️ Cha...
❤️Call Girl Service In Chandigarh☎️9814379184☎️ Call Girl in Chandigarh☎️ Cha...❤️Call Girl Service In Chandigarh☎️9814379184☎️ Call Girl in Chandigarh☎️ Cha...
❤️Call Girl Service In Chandigarh☎️9814379184☎️ Call Girl in Chandigarh☎️ Cha...Sheetaleventcompany
 
(RIYA)🎄Airhostess Call Girl Jaipur Call Now 8445551418 Premium Collection Of ...
(RIYA)🎄Airhostess Call Girl Jaipur Call Now 8445551418 Premium Collection Of ...(RIYA)🎄Airhostess Call Girl Jaipur Call Now 8445551418 Premium Collection Of ...
(RIYA)🎄Airhostess Call Girl Jaipur Call Now 8445551418 Premium Collection Of ...TanyaAhuja34
 
Kolkata Call Girls Service ❤️🍑 9xx000xx09 👄🫦 Independent Escort Service Kolka...
Kolkata Call Girls Service ❤️🍑 9xx000xx09 👄🫦 Independent Escort Service Kolka...Kolkata Call Girls Service ❤️🍑 9xx000xx09 👄🫦 Independent Escort Service Kolka...
Kolkata Call Girls Service ❤️🍑 9xx000xx09 👄🫦 Independent Escort Service Kolka...Sheetaleventcompany
 
ANATOMY AND PHYSIOLOGY OF RESPIRATORY SYSTEM.pptx
ANATOMY AND PHYSIOLOGY OF RESPIRATORY SYSTEM.pptxANATOMY AND PHYSIOLOGY OF RESPIRATORY SYSTEM.pptx
ANATOMY AND PHYSIOLOGY OF RESPIRATORY SYSTEM.pptxSwetaba Besh
 
❤️Amritsar Escorts Service☎️9815674956☎️ Call Girl service in Amritsar☎️ Amri...
❤️Amritsar Escorts Service☎️9815674956☎️ Call Girl service in Amritsar☎️ Amri...❤️Amritsar Escorts Service☎️9815674956☎️ Call Girl service in Amritsar☎️ Amri...
❤️Amritsar Escorts Service☎️9815674956☎️ Call Girl service in Amritsar☎️ Amri...Sheetaleventcompany
 
Chennai ❣️ Call Girl 6378878445 Call Girls in Chennai Escort service book now
Chennai ❣️ Call Girl 6378878445 Call Girls in Chennai Escort service book nowChennai ❣️ Call Girl 6378878445 Call Girls in Chennai Escort service book now
Chennai ❣️ Call Girl 6378878445 Call Girls in Chennai Escort service book nowtanudubay92
 

Último (20)

Ahmedabad Call Girls Book Now 9630942363 Top Class Ahmedabad Escort Service A...
Ahmedabad Call Girls Book Now 9630942363 Top Class Ahmedabad Escort Service A...Ahmedabad Call Girls Book Now 9630942363 Top Class Ahmedabad Escort Service A...
Ahmedabad Call Girls Book Now 9630942363 Top Class Ahmedabad Escort Service A...
 
Call Girls Rishikesh Just Call 9667172968 Top Class Call Girl Service Available
Call Girls Rishikesh Just Call 9667172968 Top Class Call Girl Service AvailableCall Girls Rishikesh Just Call 9667172968 Top Class Call Girl Service Available
Call Girls Rishikesh Just Call 9667172968 Top Class Call Girl Service Available
 
Call Girls Kathua Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Kathua Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Kathua Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Kathua Just Call 8250077686 Top Class Call Girl Service Available
 
Bhawanipatna Call Girls 📞9332606886 Call Girls in Bhawanipatna Escorts servic...
Bhawanipatna Call Girls 📞9332606886 Call Girls in Bhawanipatna Escorts servic...Bhawanipatna Call Girls 📞9332606886 Call Girls in Bhawanipatna Escorts servic...
Bhawanipatna Call Girls 📞9332606886 Call Girls in Bhawanipatna Escorts servic...
 
Goa Call Girl Service 📞9xx000xx09📞Just Call Divya📲 Call Girl In Goa No💰Advanc...
Goa Call Girl Service 📞9xx000xx09📞Just Call Divya📲 Call Girl In Goa No💰Advanc...Goa Call Girl Service 📞9xx000xx09📞Just Call Divya📲 Call Girl In Goa No💰Advanc...
Goa Call Girl Service 📞9xx000xx09📞Just Call Divya📲 Call Girl In Goa No💰Advanc...
 
💰Call Girl In Bangalore☎️63788-78445💰 Call Girl service in Bangalore☎️Bangalo...
💰Call Girl In Bangalore☎️63788-78445💰 Call Girl service in Bangalore☎️Bangalo...💰Call Girl In Bangalore☎️63788-78445💰 Call Girl service in Bangalore☎️Bangalo...
💰Call Girl In Bangalore☎️63788-78445💰 Call Girl service in Bangalore☎️Bangalo...
 
Call Girls Shahdol Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Shahdol Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Shahdol Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Shahdol Just Call 8250077686 Top Class Call Girl Service Available
 
Kolkata Call Girls Shobhabazar 💯Call Us 🔝 8005736733 🔝 💃 Top Class Call Gir...
Kolkata Call Girls Shobhabazar  💯Call Us 🔝 8005736733 🔝 💃  Top Class Call Gir...Kolkata Call Girls Shobhabazar  💯Call Us 🔝 8005736733 🔝 💃  Top Class Call Gir...
Kolkata Call Girls Shobhabazar 💯Call Us 🔝 8005736733 🔝 💃 Top Class Call Gir...
 
7 steps How to prevent Thalassemia : Dr Sharda Jain & Vandana Gupta
7 steps How to prevent Thalassemia : Dr Sharda Jain & Vandana Gupta7 steps How to prevent Thalassemia : Dr Sharda Jain & Vandana Gupta
7 steps How to prevent Thalassemia : Dr Sharda Jain & Vandana Gupta
 
💚Reliable Call Girls Chandigarh 💯Niamh 📲🔝8868886958🔝Call Girl In Chandigarh N...
💚Reliable Call Girls Chandigarh 💯Niamh 📲🔝8868886958🔝Call Girl In Chandigarh N...💚Reliable Call Girls Chandigarh 💯Niamh 📲🔝8868886958🔝Call Girl In Chandigarh N...
💚Reliable Call Girls Chandigarh 💯Niamh 📲🔝8868886958🔝Call Girl In Chandigarh N...
 
Premium Call Girls Dehradun {8854095900} ❤️VVIP ANJU Call Girls in Dehradun U...
Premium Call Girls Dehradun {8854095900} ❤️VVIP ANJU Call Girls in Dehradun U...Premium Call Girls Dehradun {8854095900} ❤️VVIP ANJU Call Girls in Dehradun U...
Premium Call Girls Dehradun {8854095900} ❤️VVIP ANJU Call Girls in Dehradun U...
 
Premium Call Girls Nagpur {9xx000xx09} ❤️VVIP POOJA Call Girls in Nagpur Maha...
Premium Call Girls Nagpur {9xx000xx09} ❤️VVIP POOJA Call Girls in Nagpur Maha...Premium Call Girls Nagpur {9xx000xx09} ❤️VVIP POOJA Call Girls in Nagpur Maha...
Premium Call Girls Nagpur {9xx000xx09} ❤️VVIP POOJA Call Girls in Nagpur Maha...
 
Call Girls in Lucknow Just Call 👉👉8630512678 Top Class Call Girl Service Avai...
Call Girls in Lucknow Just Call 👉👉8630512678 Top Class Call Girl Service Avai...Call Girls in Lucknow Just Call 👉👉8630512678 Top Class Call Girl Service Avai...
Call Girls in Lucknow Just Call 👉👉8630512678 Top Class Call Girl Service Avai...
 
Low Cost Call Girls Bangalore {9179660964} ❤️VVIP NISHA Call Girls in Bangalo...
Low Cost Call Girls Bangalore {9179660964} ❤️VVIP NISHA Call Girls in Bangalo...Low Cost Call Girls Bangalore {9179660964} ❤️VVIP NISHA Call Girls in Bangalo...
Low Cost Call Girls Bangalore {9179660964} ❤️VVIP NISHA Call Girls in Bangalo...
 
❤️Call Girl Service In Chandigarh☎️9814379184☎️ Call Girl in Chandigarh☎️ Cha...
❤️Call Girl Service In Chandigarh☎️9814379184☎️ Call Girl in Chandigarh☎️ Cha...❤️Call Girl Service In Chandigarh☎️9814379184☎️ Call Girl in Chandigarh☎️ Cha...
❤️Call Girl Service In Chandigarh☎️9814379184☎️ Call Girl in Chandigarh☎️ Cha...
 
(RIYA)🎄Airhostess Call Girl Jaipur Call Now 8445551418 Premium Collection Of ...
(RIYA)🎄Airhostess Call Girl Jaipur Call Now 8445551418 Premium Collection Of ...(RIYA)🎄Airhostess Call Girl Jaipur Call Now 8445551418 Premium Collection Of ...
(RIYA)🎄Airhostess Call Girl Jaipur Call Now 8445551418 Premium Collection Of ...
 
Kolkata Call Girls Service ❤️🍑 9xx000xx09 👄🫦 Independent Escort Service Kolka...
Kolkata Call Girls Service ❤️🍑 9xx000xx09 👄🫦 Independent Escort Service Kolka...Kolkata Call Girls Service ❤️🍑 9xx000xx09 👄🫦 Independent Escort Service Kolka...
Kolkata Call Girls Service ❤️🍑 9xx000xx09 👄🫦 Independent Escort Service Kolka...
 
ANATOMY AND PHYSIOLOGY OF RESPIRATORY SYSTEM.pptx
ANATOMY AND PHYSIOLOGY OF RESPIRATORY SYSTEM.pptxANATOMY AND PHYSIOLOGY OF RESPIRATORY SYSTEM.pptx
ANATOMY AND PHYSIOLOGY OF RESPIRATORY SYSTEM.pptx
 
❤️Amritsar Escorts Service☎️9815674956☎️ Call Girl service in Amritsar☎️ Amri...
❤️Amritsar Escorts Service☎️9815674956☎️ Call Girl service in Amritsar☎️ Amri...❤️Amritsar Escorts Service☎️9815674956☎️ Call Girl service in Amritsar☎️ Amri...
❤️Amritsar Escorts Service☎️9815674956☎️ Call Girl service in Amritsar☎️ Amri...
 
Chennai ❣️ Call Girl 6378878445 Call Girls in Chennai Escort service book now
Chennai ❣️ Call Girl 6378878445 Call Girls in Chennai Escort service book nowChennai ❣️ Call Girl 6378878445 Call Girls in Chennai Escort service book now
Chennai ❣️ Call Girl 6378878445 Call Girls in Chennai Escort service book now
 

Mri basic principle and sequences

  • 1. PRESENTER: DR PAWAN KUMAR MRI-BASIC PRINCIPLE/TECHNIQUE/READIN G
  • 2. Nuclear magnetic resonance ►Dr Isidor Rabi (Nobel in 1944), discovered NMR (Nuclear Magnetic Resonance) in the late 1930s, but considered it to be an artefact of his apparatus ►CJ Gorter, coined the term ‘Nuclear Magnetic Resonance’ in 1942. ►Bloch and Purcell (Nobel Prize in 1952 )if Certain nuclei are placed in magnetic field ,absorb energy in electromagnetic spectrum and re emit energy when regain their original position.
  • 3.
  • 4.
  • 5. MRI HARDWARE  Permanent magnets  Resistive magnets: In a resistive magnet, an electrical current is passed through a loop of wire and generates a magnetic field.  Superconducting magnets: Superconducting magnets are the ones most widely used in MR machines at the present time. They also make use of electricity, but they have a special current carrying conductor.  This is cooled down to superconducting temperature (about 4° K or -269° C). At this temperature, the current conducting material loses its resistance for electricity.
  • 6.  In MRI radio frequency coils are necessary to send in the RF pulse to excite the protons, and to receive the resulting signal. The same or different coils can be used for transmission of the RF pulse and receiving the signal.
  • 7. Necessary Equipment Magnet Gradient Coil RF Coil Source: Joe Gati, photos RF Coil 3T magnet gradient coil (inside)
  • 8.
  • 9. MRI principle  MRI is based on the principle of nuclear magnetic resonance (NMR)  Two basic principles of NMR: 1.Atoms with an odd number of protons or neutrons have spin 2.A moving electric charge, be it positive or negative, produces a magnetic field
  • 10. Protons  Protons possess a positive charge.  Like the earth they are constantly turning around an axis and have their own magnetic field.
  • 11. Protons  Normally protons are aligned in a random fashion. This, however, changes when they are exposed to a strong external magnetic field.  Then they are aligned in only two ways, either parallel or antiparallel to the external magnetic field
  • 12. Precession  Protons perform a wobbling type of motion in a strong magnetic field ,called precession.
  • 13. Precession Frequency  To calculate the precession frequency. Larmor equation:  B0 = is the strength of the external magnetic field, which is given in Tesla (T)  is the angular frequency, and ‘y’is the so- called gyromagnetic ratio.
  • 14. EFFECT OF MAGNETIC FIELD  The z-axis runs in the direction of the magnetic field lines,  The five protons, which "point" down cancel out the magnetic effects of the same number of protons, which "point"up
  • 15. Radio Frequency (RF)  A short burst of some electromagnetic wave, which is called a radio frequency (RF) pulse.  The purpose of this RF pulse is to disturb the protons, which are peacefully precessing in alignment with the external magnetic field.  Not every RF pulse disturbs the alignment of the protons. For this, we need a special RF pulse, one that can exchange energy with the protons.  Only when the RF pulse and the protons have the same frequency, can protons can pick up some energy from the radio wave, a phenomenon called Resonance.
  • 16. Longitudinal magnetization  the vectors along the z-axis point in the same direction, and thus add up to a new magnetic sum vector pointing up. As this magnetization is in direction along/longitudinal to the external magnetic field, it is also called longitudinal magnetization
  • 17. EFFECT OF RF PULSE ON MAGNETIZATION  The radiofrequency pulse exchanges energy with the protons (a), and some of them are lifted to a higher level of energy, pointing downward in the illustration (b).  In effect the magnetization along the z-axis decreases, as the protons which point down "neutralize" the same number of protons pointing up.
  • 18. Transversal Magnetization  When the protons randomly point left/right, back/forth and so on, they also cancel their magnetic forces in these directions  The RF pulse synchronize them - "in phase". They now point in the same direction at the same time, and thus their magnetic vectors add up in this direction. This results in a magnetic vector pointing to the side to which the precessing protons point, and this is in a transverse direction.  This is why it is called transversal magnetization.
  • 19.  As the transversal magnetic vector moves around with the precessing protons, it comes towards the antenna, goes away from it, comes towards it again and so on, also with the precession frequency.  The resulting MR signal therefore also has the precession.
  • 20. Longitudinal and transversal relaxation  The newly established transverse magnetization starts to disappear (a process called transversal relaxation), and the longitudinal magnetization grows back to its original size (a process called longitudinal relaxation).
  • 21. Longitudinal/Spin lattice relaxation  After the RF pulse is switched off, protons go back from their higher to the lower state of energy, i.e. point up again.  This energy is just handed over to their surroundings, the so called lattice. And this is why this process is not only called longitudinal relaxation, but also spin-lattice-relaxation
  • 22. T1-Curve  If one plots the longitudinal magnetization vs. time after the RF pulse was switched off, one gets a so-called T1-curve.
  • 23. SPIN SPIN RELAXATION  After the RF pulse is switched off, protons lose phase coherence, they get out of step thus transversal magnetization decreases.
  • 24.  Plot showing transversal magnetization vs. time after the RF pulse is switched off, one gets a curve as illustrated,which is called a T2-curve.
  • 25. Characteristics of T1 and T2  T1 is about 300 to 2000 msec, and T2 is about 30 to 150 msec.  It is difficult to pinpoint the end of the longitudinal and transversal relaxation exactly.  T1 =63% of the original longitudinal magnetization is reached.  T2 =37% of the original value,T1 is longer than T2  T1 varies with the magnetic field strength; it is longer in stronger magnetic fields.
  • 26. 90°pulse  If after the RF pulse, the number of protons on the higher energy level equals the number of protons on the lower energy level, longitudinal magnetization disappears, and there is only transversal magnetization due to phase coherence.  The magnetic vector seems to have been "tilted“ 90° to the side. The corresponding RF pulse is thus also called a 90° pulse
  • 27. TE vs TR  • TE: the time between the 90° pulse and the echo.  • TR: the time between two pulse sequences, i.e. from one 90° pulse to the next.
  • 28. HOW TR AFFECTS THE SIGNAL INTENSITY OF TISSUE
  • 29.  A and B are two tissues with different relaxation times.  Frame 0 shows the situation before, frame 1 immediately after a 90° pulse.  When we wait for a long time (TRlong) the longitudinal magnetization of both tissues will have totally recovered (frame 5).  A second 90° pulse after this time results in the same amount of transversal magnetization (frame 6) for both tissues, as was observed
  • 30.  When we do not wait as long , but send in the second RF pulse after a shorter time (TRShort), longitudinal magnetization of tissue B, which has the longer T1, has not recovered as much as that of tissue A with the shorter T1.  The transversal magnetization of the two tissues after the second RF pulse will then be different (frame 5). Thus, by changing the time between successive RF pulses, we can influence and modify magnetization and the signal intensity of
  • 31. TR AND SIGNAL INTENSITY  Brain has a shorter longitudinal relaxation time than CSF.  With a short TR the signal intensities of brain and CSF differ more than after a long TR.
  • 32. T2*/T2effect/spin echo sequence  After the RF pulse is switched off, the protons dephase (a-c).  The 180°pulse causes them to precess in the opposite direction and so they rephase again (d-f).
  • 33. T2* curve  The 180° pulse refocusses the dephasing protons which results in a stronger signal, the spin echo after the time TE.  The protons then dephase again and can be refocussed another time by a 180° pulse and so on. Thus it is possible to obtain more than one signal, more than one spin echo.  The spin echoes, however, differ in intensity due to so-called T2- effects.  A curve connecting the spin echo intensities is the T2 curve. If we did not use the 180° pulse, the signal intensity would decay much faster. A curve describing the signal intensity in that case is the T*2 (T2 star) curve, The type of pulse sequence, that we used in our experiment, is called a spin echo sequence, consisting of a 90° pulse and a 180° pulse
  • 34.  T2-curves for two tissues with different transversal relaxation times; tissue A has a shorter T2 than tissue B, thus loses transversal magnetization faster.  With a short TE (TEshort) the difference in signal intensity is less pronounced than after a longer TE (TElong).
  • 35. Short vs Long TE/TR  A TR of less than 500 msec is considered to be short, a TR of more than 1500 msec to be long  A short TE is one that is as short as possible, a long TE also is about 3 times as long.  A TE of less than 30 msec is considered to be short, a TE greater than 80 msec to be long.
  • 36. spin echo pulse sequence  1. (90° - TE/2 -180° - TE/2 -> record signal at TE)  after TR (time from the beginning of one 90° pulse to the next 90° pulse) follows another pulse cycle and signal b measurement:  2. (90° - TE/2 -180° - TE/2 -> record signal at TE)
  • 37. EFFECT OF TR/TE ON IMAGE  By combining T1- and T2-curves signal intensity of certain tissues can be determined for a pulse sequence using TR and TE as illustrated.  With a long TR, differences in T1, in longitudinal magnetization time are not very important any more, as all tissues have regained their full longitudinal magnetization.  When we only wait a very short TE then differences in signal intensity due to differences in T2 have not yet had time to become pronounced.  The resulting picture is thus neither T1- nor T2-weighted, but mostly determined by the proton density of the tissues (for this, ideally TE should be zero).
  • 38. T2-weighted picture  When we wait a long TR and a long TE, differences in T2 have had time enough to become pronounced, the resulting picture is T2-weighted
  • 39. T1-weighted picture  When we wait a shorter time TR, differences in T1 influence tissue contrast to a larger extent, the picture is T1-weighted, especially when we also wait a short TE (when signal differences due to differing T2s have not had time to become pronounced)
  • 40.  T2-curves of different tissues can intersect. The signal intensity of the tissues is reversed choosing a TE beyond the crossing point (TEC): before this crossing point (e.g. at TE1) tissue A has a higher signal intensity than tissue B.  This means that image contrast is still determined by differences in T1: the tissue A with the shorter T1 has the stronger signal intensity.  At TEC both tissues have the same signal intensity, and thus cannot be differentiated.  After the crossing point (e.g. at TE2) the relative signal intensities are reversed, and tissue B has the stronger signal
  • 42. Useful for: Evaluating anatomic detail CSF: Dark White Matter: White Gray Matter: Gray Vessels: Dark Useful for: Looking at areas of edema & pathology CSF: Bright White Matter: Gray Gray Matter: Lighter than white matter Vessels: Dark
  • 43.
  • 44. T1 Recovery Short TR T1 contrast (T1 Weighted)  TR 300-600 ms  TE 10-30 ms
  • 45. Bright on T1  Fat, subacute hemorrhage, melanin, protein rich fluid.  Slowly flowing blood  Paramagnetic substances(gadolinium,copper,manganese)  9
  • 46. Dark on T1  Edema, tumor, infection, inflammation, hemorrhage(hyperacute, chronic)  Low proton density, calcification  Flow void
  • 47. T2 Decay Long TE T2 contrast (T2 Weighted)  TR 2000 ms  TE 70 ms
  • 48. Bright on T2  Edema, tumor, Infection, inflammation, subdural collection  Met hemoglobin in late sub acute hemorrhage
  • 49. Dark on T2  Low proton density,calcification,fibrous tissue  Paramagnetic substances(deoxy hemoglobin,methemoglobin(intracellular),ferritin,h emosiderin,melanin.  Protein rich fluid  Flow void
  • 50. Which scan best defines the abnormality  T1 W Images:  Subacute Hemorrhage  Fat-containing structures  Anatomical Details  T2 W Images:  Edema  Demyelination  Infarction  Chronic Hemorrhage  FLAIR Images:  Edema,  Demyelination  Infarction esp. in Periventricular location
  • 51. Infarct  Acute : T1W –Isointense hypo intense T2W-Hyper intense  Sub acute: T1W-Low signal,increasedsignal in peripheral region..hemorrhage(metHb) T2W- High signal  Chronic:T1W-low signal T2W-High siignal
  • 53. FLOW VOID-HOW  Flow effects are responsible for the black appearance of flowing blood, the signal void in blood vessels
  • 54.  Paramagnetic substances like Gadolinium shorten the T1 and the T2 of their surroundings.  The respective T1- and T2-curves are shifted towards the left.  In effect, that means that for a certain TR there is more, for a certain TE there is less signal EFFECT OF CONTRAST MATERIAL
  • 55. When tissue A and B has less contrast:  T1-curves for tissue A and B are very close to each other, resulting in only a small difference in signal intensity between the tissues at TR.  NOW- the T1-curve of tissue A is shifted to the left, as contrast agent entered tissue A but not tissue B. At the same time TR there now is a much greater difference in signal intensity, i.e. tissue contrast
  • 56. The inversion recovery sequence  The inversion recovery sequence uses a 180° pulse which inverts the longitudinal magnetization, followed by a 90° pulse after the time TI.  The 90° pulse "tilts“ the magnetization into the transversal (x-y-) plane, so it can be measured/received.  The tissue in the bottom row goes back to its original longitudinal magnetization faster, thus has the shorter T1.  For the time TI, which is illustrated, this results in less transversal magnetization after the 90° pulse
  • 57. Short TI inversion-recovery (STIR) sequence Longer TE used..both long T1 and T2 …bright
  • 58.  Substances having both LONG T1 and T2 will be bright.  T1 and T2 of most pathologic lesion are prolonged.  And substances with short T1 will be suppressed eg. hemorrhage, gd- enhancement.
  • 59. Fluid-attenuated inversion recovery (FLAIR)  First described in 1992 and has become one of the corner stones of brain MR imaging protocols  An IR sequence with a long TR and TE and an inversion time (TI) that is tailored to null the signal from CSF
  • 60.
  • 61.  Water bound to complex molecule with in plaque has relatively shorter T1 than free water with in ventricle.  Long inversion time effectively suppress free water eg./csf.  Lesion that contain complex, partially bound water (less mobile)…shorter T1 than free water appears bright.  Effective in high lightening lesion eg demyelination, stroke, Ischemic gliosis and tumor.
  • 62.  Useful in diffrentiating acute infarct from cystic encephalomalacia  Useful in SAH …removes CSF signal
  • 63. GRE
  • 64. GRE  In a GRE sequence, an RF pulse is applied that partly flips the NMV into the transverse plane (variable flip angle).  Gradients, as opposed to RF pulses, are used to dephase (negative gradient) and rephase (positive gradients) transverse magnetization.
  • 65.
  • 66.  GRE Sequences contd:  This feature of GRE sequences is exploited- in detection of hemorrhage, as the iron in Hb becomes magnetized locally (produces its own local magnetic field) and thus dephases the spinning nuclei.  The technique is particularly helpful for diagnosing hemorrhagic contusions such as those in the brain .
  • 69. Diffusion-weighted MRI  DWI images is obtained by applying pairs of opposing and balanced magnetic field gradients (but of differing durations and amplitudes) around a spin- echo refocusing pulse of a T2 weighted sequence.  Stationary water molecules are unaffected by the paired gradients, and thus retain their signal.  Non stationary water molecules acquire phase information from the first gradient, but are not re phased by the second gradient, leading to an overall loss of the MR signal
  • 70.  The normal motion of water molecules within living tissues is random (brownian motion).  In acute stroke, there is an alteration of homeostasis  Acute stroke causes excess intracellular water accumulation, or cytotoxic edema, with an overall decreased rate of water molecular diffusion within the affected tissue.  Tissues with a higher rate of diffusion undergo a greater loss of signal in a given period of time than do tissues with a lower diffusion rate.  Therefore, areas of cytotoxic edema, in which the motion of water molecules is restricted, appear brighter on diffusion-weighted images because of lesser signal losses
  • 71.
  • 72. Apparent Diffusion Coefficient  Calculated by acquiring two or more images with a different gradient duration and amplitude .  To differentiate T2 shine through effects or artifacts from real ischemic lesions.
  • 73. ADC  useful for estimating the lesion age and distinguishing acute from subacute DWI lesions.  Acute ischemic lesions can be divided into  1- hyperacute lesions (low ADC and DWI- positive)  2-subacute lesions (normalized ADC).  3-Chronic lesions can be differentiated from acute lesions by normalization of ADC and DWI.
  • 74. Nonischemic causes for decreased ADC  Abscess  Lymphoma and other tumors  Multiple sclerosis  Seizures  Metabolic (Canavans )
  • 75. Evaluation of acute stroke on DWI  The DWI and ADC maps show changes in ischemic brain within minutes to few hours  The signal intensity of acute stroke on DW images increase during the first week after symptom onset and decrease thereafter, but signal remains hyper intense for a long period (up to 72 days in the study by Lausberg et al)  The ADC values decline rapidly after the onset of ischemia and subsequently increase from dark to bright 7-10 days later .
  • 76.  This property may be used to differentiate the lesion older than 10 days from more acute ones .  Chronic infarcts are characterized by elevated diffusion and appear hypo, iso or hyper intense on DW images and hyperintense on ADC maps
  • 77.
  • 78. DW MR imaging characteristics of Various Disease Entities MR Signal Intensity Disease DW Image ADC Image ADC Cause Acute Stroke High Low Restricted Cytotoxic edema Chronic Strokes Variable High Elevated Gliosis Hypertensive encephalopathy Variable High Elevated Vasogenic edema Arachnoid cyst Low High Elevated Free water Epidermoid mass High Low Restricted Cellular tumor Herpes encephalitis High Low Restricted Cytotoxic edema CJD High Low Restricted Cytotoxic edema MS acute lesions Variable High Elevated Vasogenic edema Chronic lesions Variable High Elevated Gliosis
  • 79. Slice selection and thickness  There are two ways to determine slice thickness. (a). By using certain bandwidth of RF pulse eg. If frequencies between 64 and 65 mHz, protons in slice 1 will be influenced by the RF pulse.  When the RF pulse only contains frequencies between 64 mHz and 64.5 mHz, thus has a smaller bandwidth, slice 2, which is half as thick as slice 1 will be imaged.  When there is more difference in magnetic field strength between the level of the feet and the head, i.e. The magnetic gradient is steeper, the resulting slice will be thinner, even though the RF pulse bandwidth is the same.
  • 81. THE IMAGING PLANES - Axial plane: Transverse images represent "slices" of the body - Sagittal plane: Images taken perpendicular to the axial plane which separate the left and right sides (lateral view) - Coronal plane: Images taken perpendicular to the sagittal plane which separate the front from the back. (frontal view)
  • 82. How do you describe abnormalities on MR? Hyperintense (more intense): If an abnormality is bright (white) on MR, we describe it as hyperintense. Isointense (the same intensity): If an abnormality is the same intensity to a reference structure, we describe it as isointense. Hypointense (less intense): If an abnormality is dark on MR .
  • 83. IT IS CT OR MRI? CT: Computed Tomography MRI: Magnetic Resonance Imaging Imaging Plane: in the axial plane. The axial data set can then be used to reconstruct images in other planes, sagittal and coronal are the most common. Windows: "brain window" AND ‘bone window ‘ White and Black:air within the sinuses is black, the brain parenchyma has a gray appearance and the skull is bright white Imaging Plane: in any plane, not just axial Sequences: different type of image is referred to a sequence. White and Black:same structure may be bright or dark depending on the type of sequence; CSF for example is bright on T2, but dark on T1. The tissue and imaging characteristics are a lot more complicated than CT.
  • 84. Types of MR Images Useful for: Evaluating anatomic detail CSF: Dark White Matter: White Gray Matter: Gray Vessels: Dark Useful for: Evaluating for BBB breakdown in the setting of tumor, infection, MS etc. CSF: Dark White Matter: White Gray Matter: Gray Vessels: Brigh
  • 85. Useful for: Looking at areas of edema & pathology CSF: Bright White Matter: Gray Gray Matter: Lighter than white matter Vessels: Dark Useful for: Evaluating areas of edema with CSF subtraction. Edema stands out because is CSF dark CSF: Dark White Matter: Gray Gray Matter: Lighter than white matter Vessels: Dark
  • 86. Useful for: stroke imaging, abscess, cellular tumors CSF: Dark White Matter: Gray Gray Matter: Lighter than white matter Fuzzier image than FLAIR T2* (T2-star, or SWI) Form of T2-weighted image which is susceptible to iron or calcium Blood, bone, calcium appear dark Area of blood often appears much larger than reality(“blooming”) Useful for: Identification of early hemorrhage Look for: DARK only Recognition: o Like T2 except o Cranium, scalp are dark or absent o Dark areas near frontal and temporal bones
  • 87. Apparent Diffusion Coefficient (ADC Map) Contains actual data relevant to diffusion image Areas of restricted diffusion are dark Useful for: o Excluding T2-shine through o Real restricted diffusion is bright on DWI, dark on ADC Look for: DARK only Recognition o Images marked ADC o Grainy dark images
  • 88. MR ARTIFACTS Artifacts in MR images refer to pixels that do not faithfully represent the anatomy being studied. 1 Motion Artifacts- Motion artifacts occur as a result of movement of tissue during the data acquisition period. 2.Sequence/Protocol-Related Artifacts- results from the specific measurement process used to acquire the image. subtypes 2a. Aliasing 2b. Chemical Shift Artifacts 2c. Phase Cancellation Artifact 2d. Truncation Artifacts 2e. Coherence Artifacts 2f. Magnetic-Susceptibility Difference Artifacts 3. External Artifacts. External artifacts are generated from sources other than patient tissue. 3a. Magnetic Field Distortions 3b. Measurement Hardware 3c. Noise
  • 89. refrences 1.MRI: Basic Principles and Applications, Third Edition, by M. A. Brown and R. C. Semelka.ISBN 0-471-43310-1 © 2003 John Wiley & Sons, Inc. 2.Schering: MRI made easy,by Hans H. Schild Nationales Druckhaus Berlin.ISBN 3-921817-41- 3. Bradley’s text book of neurology-6th edition 4.Robert A Ziemarman’s Neuroimaging clinical and physical principle. 5. https://sites.google.com/a/wisc.edu/neuroradiology/image- acquisition/the-basics