Medical Gas Cylinder
R.Srihari
Topics for Discussion
• Definitions
• Medical Gas Cylinder
– Components
– Sizes
– Contents and Pressure
– Testing and Filling
– Color
– Marking, Labelling and Tags
– Rules for safe use of cylinders
– Hazards
Defintions
• Psi: Pounds per square inch
• Psig: Pounds per square inch gauge
GAUGE PRESSURE
It is the difference between the measured
pressure and the atmospheric pressure
Most gauges are measured to read zero
at atmospheric pressure
• Psia: Pounds per square inch absolute
ABSOLUTE PRESSURE
It is based on reference point of zero in a
perfect vacuum
Psia = Psig + atmospheric pressure
Eg: at sea level- atmospheric pressure is
0 but psia is 14.7psia
• Units of Pressure:
– 1 atmosphere = 14.7 psi
760 mmHg
1030 cm H20
1000 mbar
100 KPa
• Critical temperature:
– It is defined as the temperature below which no
gas can be liquefied irrespective of which pressure
is applied
• Critical Pressure:
– Defined as the minimum pressure required to
liquefy a gas at its critical temperature
• Compressed Gas:
– It is defined as any mixture having in a container
an absolute pressure exceeding 40 psi at 70 F
OR
– Regardless of the pressure at 70 F having a
absolute pressure exceeding 104 psi at 130 F
OR
– Any liquid having a vapour pressure exceeding 40
psia at 100 F
• Non-Liquified Gas:
– These are gases that do not liquefy at ambient
temperatures regardless of the pressures applied
– These gases do become liquids at lower
temperatures at which point – called as
CRYOGENIC LIQUIDS
– Eg: Oxygen, Nitrogen, Air, Helium
• Liquified Compressed Gases:
– It is one in which gas becomes liquid at ambient
temperatures at pressures varying from 25-100
psig (172 -10340 Kpa)
– Eg: N20, CO2
Standards for Cylinders in India
• Gas Cylinders Rules(1981)
• Static and Mobile Pressure Vessels(unfired)
rules (1981)
• Bureau of Indian Standards Act(1986)
• All those who produce, supply and transport or
use medical gases must comply with variety of
safety regulations
 put forth and enforced by agencies at
state and central government
• Purity of gases specified by USP and enforced
by FDA
Components
• Consists of:
– Body
– Valve
– Port
– Stem
– Pressure relief devices
• Body:
– Most medical gas cylinders are made of steel with
various alloys added
– In recent years, manufacturers have moved from
traditional steel cylinders towards steel-carbon
fibres cylinders
 Can hold more gas and light in weight
– MRI compatible cylinders are made of aluminium
– Modern cylinders are made of alloy of
MOLYBDENUM + STEEL +/- CHROMIUM
Alloy containing Molybdenum (0.15-0.25%)
Chromium (If +  0.8-1.1%)
Is used to increase strength and to minimise weight
and wall thickness
• Walls of the cylinder vary from 5/64 to ¼ inch
thickness on an average
• Cylinders that have a marking 3AA are made
from steel. The marking 3AL or 3ALM indicates
that the cylinder is made from aluminium
• Cylinder have a flat or a concave base. The other
end may taper into a neck that is fitted with
tapered screw threads that attach to the cylinder
valve
• Valve:
– Cylinders are filled and discharged through a valve
(spindle shaped) attached to the neck
– It is made of bronze or brass which is heavily plated
with nickel or chromium so as to allow rapid
dissipation of heat of compression
– The end which enters the neck of the cylinder is
threaded to fit a corresponding screw thread inside the
neck itself
• A sleeve or washer of soft alloy ( containing high
proportions of lead) completes the gas tight seal
as the valve is screwed into the neck of the
cylinder
• Cylinder valves are of various types.
– Those used on anesthesia machines are ‘flush’ types
which fits with the pin index system on the machine
– For medium and large capacity cylinders – bull nose
valves are used
• Cylinder valves are of 2 types –packed type and
diaphragm type
– Packed type:
• Capable of withstanding high pressures
• A.k.a direct acting valve
• Stem is sealed by resilient packing such as TEFLON which
prevents leaks around the threads
• In large cylinders, the force is transmitted by means of driver
square
• It is opened by 2-3 turns
• Used in most of the cylinders
– Diaphragm type:
• Stem is separated from the seat
• Closure between the cylinder interior and atmosphere is
accomplished by using a seal and a bonnet nut that clamps one or
more circular discs in place
• These discs separate upper and lower stems which may be
permanently attached to the diaphragms
• Upper stem is actuated by manual/automatic means and Lower
stem shuts/permits flow through the valve
• Can be opened fully by ½ to ¾ turns
• Seat does not turn-so less likely to leak
• Generally preferred when pressures are relatively low and no leaks
can be allowed
• However - expensive
• Port:
– It is the point of exit for the gas
It fits into the nipple on the hanger yoke of the
anesthesia machine
– It should be protected in transit by a covering
– When installing a cylinder on anesthesia machine, it is
important for the user not to mistake the port for the
conical depression
– Conical depression is situated on the opposite side of
the port on the cylinder valve and is situated above the
safety relief device
It is present on those cylinders which are designed to
fit on anesthesia machine
– Conical depression is designed to receive the retaining
screw on the yoke of the anesthesia machine
 Screwing the retaining screw into the port may
damage the port
• Stem:
– Each valve contains a stem (spindle/screw-pin) or shaft
that is rotated to open or close the cylinder valve
– It is made up of very hard steel
– To close the valve the stem seals against the seat that is
part of the valve
 when the valve is opened –stem is moved
upwards and allows the gas to flow to the port
• Pressure Relief Devices:
– Every cylinder is fitted with pressure relief devices
whose purpose is to vent the cylinder’s contents to the
atmosphere
 if the pressure of enclosed gas increases to
dangerous levels
– Types:
• Rupture Disc
• Fusible Plug
• Combination of Both
• Pressure Relief valve (spring loaded)
– Rupture Disc:
• When pre-determined pressure is reached the disc
ruptures and allows the gas contents to be discharged
• It is a non- reclosing device held against an orifice
• It protects against excess pressure as a result of high
temperature/overfilling
– Fusible Plug:
• It is thermally operated
• It is a non-reclosing pressure relief device where the plug is
held against the discharge channel
• It provides protection against excess pressure due to high
temperature but not overfilling
• YIELD TEMPERATURE: Temperature at which fusible
material becomes sufficiently soft to extrude from its holder-
so that cylinder contents are discharged
– Spring loaded pressure relief device:
• It is a reclosing device
• When set pressure is exceeded, the pressure in the
cylinder forced the spring to open the channel for letting
out the gases
Gas flows around the safety valve seat to discharge
channel till excess pressure is relieved
• Handle/ Handwheel:
– It is used to open or close a cylinder valve
– It is turned counter-clockwise to open and clock-wise
to close
 this causes the stem to turn
– A good practice is to attach a handle to each anesthesia
machine or other apparatus for which it may be needed
– Each large cylinder has a permanent attached
handwheel that uses a spring and nut  to hold it
firmly in place
Non-Interchangeable Safety Systems
• With withspread use of cylinders containing
different gases , a potential hazard is
connection of a cylinder to equipment intended
for a different gas
• For safety purposes
– Color coding for each gas
– Pin index safety system
Color coding
GAS SHOULDER BODY
OXYGEN WHITE BLACK
NITROUS OXIDE BLUE BLUE
CYCLOPROPANE ORANGE ORANGE
CARBON DI OXIDE GREY GREY
AIR WHITE GREY
NITROGEN BLACK BLACK
ENTONOX WHITE BLUE
Pin Index Safety System
• The use of PISS began in 1952
• In order to ensure that the correct cylinder is
attached to the appropriate hanger yoke of the
anesthesia machine/workstation
A series of pins on the hanger yoke is made to fit
into the corresponding indentations(pits/holes)
drilled into cylinder valve
• It consists of holes on the cylinder valve positioned in an
arc below the outlet port
• Positions of the cylinder valve are on the circumference of a
circle of 9/16 inch (14.3mm) radius centered on the port
• The port has a diameter of 7mm
• The distance between the centre and lower part of the yoke
is 20.6mm
• Unless the pins and holes are aligned, the port
will not seat
• The indentations on the cylinder valve/yoke block
are counted 1-6 from left to right
• The distance between the centre of the 1st pin and
6th pin should be 16mm
• There are 7 positions for pins and holes
• The pins are 4.75mm in diameter and 6mm
long except for pin number 7 which is slightly
thicker and placed at the centre (between port
3 and 4)
GAS PIN INDEX
AIR 1,5
OXYGEN 2,5
NITROUS OXIDE 3,5
NITROGEN 1,4
O2-CO2 (CO2 <7.5%) 2,6
O2-CO2 (CO2 >7.5%) 1,6
ENTONOX 7
Sizes
• Gas suppliers classify cylinders by using a letter code
with A being the smallest
• Volume and Pressure of gas in a particular size cylinder
vary
• O2 and Air – similar ; CO2 and N20- similar
• Size E is the most commonly used in anesthesia
machine and for patient transport and resuscitation
• Size D cylinders are used for limited supplies of gases
where size and weight considerations are important
CYLINDER SIZE DIMENSIONS
(OUTER
DIAMETER X
LENGTH) IN
INCH
EMPTY
CYLINDER
VOLUME
KG
INTERNAL
VOLUME
L
D 4.5 X 17 5 3
E 4.25 X 26 6.3 5
H 9.25 X 51 53 46-50
O2/AIR N20/CO2
SIZE D – CONTENT IN
LITRES
400 940
SIZE D- PRESSURE IN
PSIG
1900 745/838
SIZE E- CONTENT IN
LITRES
660 1590
SIZE E- PRESSURE IN
PSIG
1900 745/838
SIZE H –CONTENT IN
LITRES
6900 15800
SIZE H- PRESSURE IN
PSIG
2200 745
Contents and Pressure
• In a cylinder containing a non-liquefied gas
Pressure declines as the contents are
withdrawn
Hence pressure can be used to measure
cylinder volume(approximately)
• In a cylinder containing a liquified gas, the pressure
depends on VAPOUR PRESSURE of the liquid and is
not an indication of the amount of gas remaining in the
cylinder as long as the contents are partly in the liquid
state
Pressure remains nearly constant till all liquid has
evaporated
After which pressure declines till cylinder is exhausted
Testing
• A cylinder must be inspected and tested atleast
every 5 years or with special permit upto every 10
years
• The test date must be permanently stamped on the
cylinder
• Each cylinder must pass an internal and external
visual check for corrosion and evidence of
physical impact or distortion
• Cylinders are checked for leaks and retention
of structural strength by testing to minimum of
1.66 times their service pressure
• SERVICE PRESSURE IS DEFINED AS THE
MAXIMUM PRESSURE TO WHICH THE
CYLINDER MAY BE FILLED AT 70 F
• Other tests that are done:
– Tensile test
– Flattening test
– Bend test
– Impact test
– These are carried on atleast one out of every 100
cylinders
Filling
• If a cylinder containing a gas under a safe
pressure at normal temperature is subjected to
higher temperature the pressure may
increase to dangerous levels
• To prevent this, regulations have been drawn
limiting the amount of gas a cylinder may
contain
• Non- liquefied gases – may be allowed an
additional 10% filling
• Liquefied gas containing cylinders:
– Pressures will remain constant as long as there is
liquid in cylinder
 to prevent cylinder being overfilled
Maximum amount of gas allowed is defined by the
filling density(filling ratio) for each gas
• Filling density- percent of ratio of weight of
the gas in a cylinder to the weight of water that
the cylinder would hold at 60 F
– N20 – 68%
– CO2- 68%
Marking/Labelling/Tags
• Important for identification
• To check test date
• In case of flammable gases- Caution/ Danger/
Warning label is needed
• Tags should contain either full/ in use/ empty
Hazards
• Incorrect cylinder
• Incorrect valve
• Incorrect color/labelling
• Inoperable valve/ Damaged valve
• Suffocation/fire explosion
• Overfilled contents in cylinder
• Nitrous oxide theft
• Contaminated contents in cylinders
• Thermal injury
Rules for safe use of cylinders
1. To be handled by trained staff
2. Store cylinders in a cool, clean room with adequate
ventilation
3. Do not drape cylinder with any material during
storage
4. Cylinders are best stored upright in a cylinder stand
5. Keep the valve closed when not in use
6. Identify contents by label
7. Remove wrappings( protective covering) before
use
8. Remove dust/ foreign bodies before connecting
9. ‘Cracking’ to be done to reduce to possibility of
flash fire
10. A sealing washer in good condition should be used
11. Flow control valves should be closed before cylinders
opened
12. Quick opening to be avoided as it can generate heat
leading to flame
13. Valve should be fully opened when in use
14. To be kept away from oil, rubber and combustible
substances
15. Do not expose cylinder to heat or higher temperatures
Thank You

Más contenido relacionado

La actualidad más candente

Bougie, trachlite , laryngeal tube , combitube , i gel ,truview
Bougie, trachlite , laryngeal tube , combitube , i gel ,truviewBougie, trachlite , laryngeal tube , combitube , i gel ,truview
Bougie, trachlite , laryngeal tube , combitube , i gel ,truview
Dhritiman Chakrabarti
 
Breathing circuit's
Breathing circuit'sBreathing circuit's
Breathing circuit's
Imran Sheikh
 
Low pressure system in anaesthesia machine
Low pressure system in anaesthesia machineLow pressure system in anaesthesia machine
Low pressure system in anaesthesia machine
Swadheen Rout
 
Anaesthesia breathing systems
Anaesthesia breathing systemsAnaesthesia breathing systems
Anaesthesia breathing systems
D Nkar
 
Circle system low flow anesthesia
Circle system low flow anesthesiaCircle system low flow anesthesia
Circle system low flow anesthesia
Drgeeta Choudhary
 

La actualidad más candente (20)

SAFETY FEATURES OF ANAESTHESIA MACHINE
SAFETY FEATURES OF ANAESTHESIA MACHINESAFETY FEATURES OF ANAESTHESIA MACHINE
SAFETY FEATURES OF ANAESTHESIA MACHINE
 
Anaesthesia machine 1
Anaesthesia machine 1Anaesthesia machine 1
Anaesthesia machine 1
 
Breathing circuits
Breathing circuitsBreathing circuits
Breathing circuits
 
Mapleson breathing systems
Mapleson breathing systemsMapleson breathing systems
Mapleson breathing systems
 
Anaesthesia Vaporizers
Anaesthesia VaporizersAnaesthesia Vaporizers
Anaesthesia Vaporizers
 
Vaporizers
Vaporizers Vaporizers
Vaporizers
 
Bougie, trachlite , laryngeal tube , combitube , i gel ,truview
Bougie, trachlite , laryngeal tube , combitube , i gel ,truviewBougie, trachlite , laryngeal tube , combitube , i gel ,truview
Bougie, trachlite , laryngeal tube , combitube , i gel ,truview
 
Breathing circuit's
Breathing circuit'sBreathing circuit's
Breathing circuit's
 
Safety features in anesthesia machine
Safety features in anesthesia machineSafety features in anesthesia machine
Safety features in anesthesia machine
 
Facemask , oral and nasal airways
Facemask , oral and nasal airwaysFacemask , oral and nasal airways
Facemask , oral and nasal airways
 
Medical gas Cylinder
Medical gas CylinderMedical gas Cylinder
Medical gas Cylinder
 
Low pressure system in anaesthesia machine
Low pressure system in anaesthesia machineLow pressure system in anaesthesia machine
Low pressure system in anaesthesia machine
 
Airway equipments for Anaesthesia
Airway equipments for AnaesthesiaAirway equipments for Anaesthesia
Airway equipments for Anaesthesia
 
Anaesthesia breathing systems
Anaesthesia breathing systemsAnaesthesia breathing systems
Anaesthesia breathing systems
 
Anesthesia machine
Anesthesia machineAnesthesia machine
Anesthesia machine
 
Circle system low flow anesthesia
Circle system low flow anesthesiaCircle system low flow anesthesia
Circle system low flow anesthesia
 
Medical Gas Cylinders
 Medical Gas Cylinders Medical Gas Cylinders
Medical Gas Cylinders
 
CIRCLE SYSTEM.pptx
CIRCLE SYSTEM.pptxCIRCLE SYSTEM.pptx
CIRCLE SYSTEM.pptx
 
Mapleson circuits
Mapleson circuitsMapleson circuits
Mapleson circuits
 
The basic anaesthesia machine
The basic anaesthesia machineThe basic anaesthesia machine
The basic anaesthesia machine
 

Similar a Medical gas cylinders

oxygen cylinder- Zanzibar.pptx
oxygen cylinder- Zanzibar.pptxoxygen cylinder- Zanzibar.pptx
oxygen cylinder- Zanzibar.pptx
WahidiAzani
 

Similar a Medical gas cylinders (20)

HIGH PRESSURE SYSTEM.pptx
HIGH PRESSURE SYSTEM.pptxHIGH PRESSURE SYSTEM.pptx
HIGH PRESSURE SYSTEM.pptx
 
Safe Handling of Compressed Gases by UCSD
Safe Handling of Compressed Gases by UCSDSafe Handling of Compressed Gases by UCSD
Safe Handling of Compressed Gases by UCSD
 
oxygen cylinder- Zanzibar.pptx
oxygen cylinder- Zanzibar.pptxoxygen cylinder- Zanzibar.pptx
oxygen cylinder- Zanzibar.pptx
 
Oxygen MANUFACTRE STORAGE PREPERATION AND CLINICAL ASPECT
Oxygen  MANUFACTRE STORAGE PREPERATION AND CLINICAL ASPECTOxygen  MANUFACTRE STORAGE PREPERATION AND CLINICAL ASPECT
Oxygen MANUFACTRE STORAGE PREPERATION AND CLINICAL ASPECT
 
Session_3_Oxygen_Cylinder-_USAID_RISE.pptx
Session_3_Oxygen_Cylinder-_USAID_RISE.pptxSession_3_Oxygen_Cylinder-_USAID_RISE.pptx
Session_3_Oxygen_Cylinder-_USAID_RISE.pptx
 
anaesthesia machine ppt 2-1.pptx
anaesthesia machine ppt 2-1.pptxanaesthesia machine ppt 2-1.pptx
anaesthesia machine ppt 2-1.pptx
 
8. Reciprocating compressors.pdf
8. Reciprocating compressors.pdf8. Reciprocating compressors.pdf
8. Reciprocating compressors.pdf
 
Presentation on aneasthetic machine edited 21st july (2)
Presentation on aneasthetic machine edited 21st july (2)Presentation on aneasthetic machine edited 21st july (2)
Presentation on aneasthetic machine edited 21st july (2)
 
compressor maintenance, type of compressor.
compressor maintenance, type of compressor.compressor maintenance, type of compressor.
compressor maintenance, type of compressor.
 
Compressed+gas+safety
Compressed+gas+safetyCompressed+gas+safety
Compressed+gas+safety
 
Handling of Gases
Handling of GasesHandling of Gases
Handling of Gases
 
ANESTHESIA MACHINE PPT .pptx
ANESTHESIA MACHINE PPT .pptxANESTHESIA MACHINE PPT .pptx
ANESTHESIA MACHINE PPT .pptx
 
anaesthesia machine(1).pptx
anaesthesia machine(1).pptxanaesthesia machine(1).pptx
anaesthesia machine(1).pptx
 
Basic Detailed Valves.pdf
Basic Detailed Valves.pdfBasic Detailed Valves.pdf
Basic Detailed Valves.pdf
 
valves.pdf
valves.pdfvalves.pdf
valves.pdf
 
compressed-gas-safety.ppt
compressed-gas-safety.pptcompressed-gas-safety.ppt
compressed-gas-safety.ppt
 
compressed-gas-safety.ppt
compressed-gas-safety.pptcompressed-gas-safety.ppt
compressed-gas-safety.ppt
 
Safe Use of Cylinder Gases by Consultnet Limited
Safe Use of Cylinder Gases by Consultnet LimitedSafe Use of Cylinder Gases by Consultnet Limited
Safe Use of Cylinder Gases by Consultnet Limited
 
Centrifugal compressor
Centrifugal compressor Centrifugal compressor
Centrifugal compressor
 
Centrifugal Details.pdf
Centrifugal Details.pdfCentrifugal Details.pdf
Centrifugal Details.pdf
 

Más de rajkumarsrihari

Oxygen dissociation curve
Oxygen dissociation curveOxygen dissociation curve
Oxygen dissociation curve
rajkumarsrihari
 

Más de rajkumarsrihari (11)

Anesthetic considerations for endocrine diseases – an overview
Anesthetic considerations for endocrine diseases – an overviewAnesthetic considerations for endocrine diseases – an overview
Anesthetic considerations for endocrine diseases – an overview
 
Perioperative management of a patient with diabetes mellitus
Perioperative management of a patient with diabetes mellitusPerioperative management of a patient with diabetes mellitus
Perioperative management of a patient with diabetes mellitus
 
Thyrotoxicosis and myxedema-Anesthetic implications
Thyrotoxicosis and myxedema-Anesthetic implicationsThyrotoxicosis and myxedema-Anesthetic implications
Thyrotoxicosis and myxedema-Anesthetic implications
 
Pulmonary artery catheter
Pulmonary artery catheterPulmonary artery catheter
Pulmonary artery catheter
 
Malignant hyperthermia and dantrolene sodium
Malignant hyperthermia and dantrolene sodiumMalignant hyperthermia and dantrolene sodium
Malignant hyperthermia and dantrolene sodium
 
Rhabdomyolysis
RhabdomyolysisRhabdomyolysis
Rhabdomyolysis
 
Sensory and motor innervation of upper airway
Sensory and motor innervation of upper airwaySensory and motor innervation of upper airway
Sensory and motor innervation of upper airway
 
Blood components and preparation
Blood components and preparationBlood components and preparation
Blood components and preparation
 
Pain physiology
Pain physiologyPain physiology
Pain physiology
 
Meningomyelocoele
MeningomyelocoeleMeningomyelocoele
Meningomyelocoele
 
Oxygen dissociation curve
Oxygen dissociation curveOxygen dissociation curve
Oxygen dissociation curve
 

Último

Nanded Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Nanded Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetNanded Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Nanded Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Call Girls Service
 
palanpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
palanpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meetpalanpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
palanpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Call Girls Service
 
Malda Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Malda Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetMalda Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Malda Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Call Girls Service
 
Hubli Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Hubli Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetHubli Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Hubli Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Call Girls Service
 
bhopal Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
bhopal Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meetbhopal Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
bhopal Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Call Girls Service
 
Premium Call Girls Bangalore {7304373326} ❤️VVIP POOJA Call Girls in Bangalor...
Premium Call Girls Bangalore {7304373326} ❤️VVIP POOJA Call Girls in Bangalor...Premium Call Girls Bangalore {7304373326} ❤️VVIP POOJA Call Girls in Bangalor...
Premium Call Girls Bangalore {7304373326} ❤️VVIP POOJA Call Girls in Bangalor...
Sheetaleventcompany
 
Jalna Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Jalna Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetJalna Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Jalna Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Call Girls Service
 
Mangalore Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Mangalore Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetMangalore Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Mangalore Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Call Girls Service
 
Tirupati Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Tirupati Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetTirupati Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Tirupati Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Call Girls Service
 
Bhagalpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Bhagalpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetBhagalpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Bhagalpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Call Girls Service
 
nagpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
nagpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meetnagpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
nagpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Call Girls Service
 
Muzaffarpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Muzaffarpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetMuzaffarpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Muzaffarpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Call Girls Service
 
raisen Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
raisen Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meetraisen Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
raisen Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Call Girls Service
 
Thoothukudi Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Thoothukudi Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetThoothukudi Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Thoothukudi Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Call Girls Service
 
Erode Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Erode Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetErode Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Erode Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Call Girls Service
 
Sambalpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Sambalpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetSambalpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Sambalpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Call Girls Service
 

Último (20)

Kolkata Call Girls Miss Inaaya ❤️ at @30% discount Everyday Call girl
Kolkata Call Girls Miss Inaaya ❤️ at @30% discount Everyday Call girlKolkata Call Girls Miss Inaaya ❤️ at @30% discount Everyday Call girl
Kolkata Call Girls Miss Inaaya ❤️ at @30% discount Everyday Call girl
 
Nanded Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Nanded Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetNanded Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Nanded Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
 
Jaipur Call Girls 9257276172 Call Girl in Jaipur Rajasthan
Jaipur Call Girls 9257276172 Call Girl in Jaipur RajasthanJaipur Call Girls 9257276172 Call Girl in Jaipur Rajasthan
Jaipur Call Girls 9257276172 Call Girl in Jaipur Rajasthan
 
palanpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
palanpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meetpalanpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
palanpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
 
Malda Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Malda Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetMalda Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Malda Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
 
Hubli Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Hubli Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetHubli Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Hubli Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
 
bhopal Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
bhopal Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meetbhopal Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
bhopal Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
 
Premium Call Girls Bangalore {7304373326} ❤️VVIP POOJA Call Girls in Bangalor...
Premium Call Girls Bangalore {7304373326} ❤️VVIP POOJA Call Girls in Bangalor...Premium Call Girls Bangalore {7304373326} ❤️VVIP POOJA Call Girls in Bangalor...
Premium Call Girls Bangalore {7304373326} ❤️VVIP POOJA Call Girls in Bangalor...
 
Dehradun Call Girls 8854095900 Call Girl in Dehradun Uttrakhand
Dehradun Call Girls 8854095900 Call Girl in Dehradun  UttrakhandDehradun Call Girls 8854095900 Call Girl in Dehradun  Uttrakhand
Dehradun Call Girls 8854095900 Call Girl in Dehradun Uttrakhand
 
Jalna Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Jalna Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetJalna Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Jalna Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
 
Mangalore Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Mangalore Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetMangalore Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Mangalore Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
 
Tirupati Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Tirupati Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetTirupati Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Tirupati Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
 
Bhagalpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Bhagalpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetBhagalpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Bhagalpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
 
nagpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
nagpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meetnagpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
nagpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
 
(Deeksha) 💓 9920725232 💓High Profile Call Girls Navi Mumbai You Can Get The S...
(Deeksha) 💓 9920725232 💓High Profile Call Girls Navi Mumbai You Can Get The S...(Deeksha) 💓 9920725232 💓High Profile Call Girls Navi Mumbai You Can Get The S...
(Deeksha) 💓 9920725232 💓High Profile Call Girls Navi Mumbai You Can Get The S...
 
Muzaffarpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Muzaffarpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetMuzaffarpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Muzaffarpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
 
raisen Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
raisen Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meetraisen Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
raisen Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
 
Thoothukudi Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Thoothukudi Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetThoothukudi Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Thoothukudi Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
 
Erode Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Erode Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetErode Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Erode Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
 
Sambalpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Sambalpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetSambalpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Sambalpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
 

Medical gas cylinders

  • 2. Topics for Discussion • Definitions • Medical Gas Cylinder – Components – Sizes – Contents and Pressure – Testing and Filling – Color – Marking, Labelling and Tags – Rules for safe use of cylinders – Hazards
  • 3. Defintions • Psi: Pounds per square inch • Psig: Pounds per square inch gauge GAUGE PRESSURE It is the difference between the measured pressure and the atmospheric pressure Most gauges are measured to read zero at atmospheric pressure
  • 4. • Psia: Pounds per square inch absolute ABSOLUTE PRESSURE It is based on reference point of zero in a perfect vacuum Psia = Psig + atmospheric pressure Eg: at sea level- atmospheric pressure is 0 but psia is 14.7psia
  • 5. • Units of Pressure: – 1 atmosphere = 14.7 psi 760 mmHg 1030 cm H20 1000 mbar 100 KPa
  • 6. • Critical temperature: – It is defined as the temperature below which no gas can be liquefied irrespective of which pressure is applied • Critical Pressure: – Defined as the minimum pressure required to liquefy a gas at its critical temperature
  • 7. • Compressed Gas: – It is defined as any mixture having in a container an absolute pressure exceeding 40 psi at 70 F OR – Regardless of the pressure at 70 F having a absolute pressure exceeding 104 psi at 130 F OR – Any liquid having a vapour pressure exceeding 40 psia at 100 F
  • 8. • Non-Liquified Gas: – These are gases that do not liquefy at ambient temperatures regardless of the pressures applied – These gases do become liquids at lower temperatures at which point – called as CRYOGENIC LIQUIDS – Eg: Oxygen, Nitrogen, Air, Helium
  • 9. • Liquified Compressed Gases: – It is one in which gas becomes liquid at ambient temperatures at pressures varying from 25-100 psig (172 -10340 Kpa) – Eg: N20, CO2
  • 10. Standards for Cylinders in India • Gas Cylinders Rules(1981) • Static and Mobile Pressure Vessels(unfired) rules (1981) • Bureau of Indian Standards Act(1986)
  • 11. • All those who produce, supply and transport or use medical gases must comply with variety of safety regulations  put forth and enforced by agencies at state and central government • Purity of gases specified by USP and enforced by FDA
  • 12. Components • Consists of: – Body – Valve – Port – Stem – Pressure relief devices
  • 13. • Body: – Most medical gas cylinders are made of steel with various alloys added – In recent years, manufacturers have moved from traditional steel cylinders towards steel-carbon fibres cylinders  Can hold more gas and light in weight
  • 14. – MRI compatible cylinders are made of aluminium – Modern cylinders are made of alloy of MOLYBDENUM + STEEL +/- CHROMIUM Alloy containing Molybdenum (0.15-0.25%) Chromium (If +  0.8-1.1%) Is used to increase strength and to minimise weight and wall thickness
  • 15. • Walls of the cylinder vary from 5/64 to ¼ inch thickness on an average • Cylinders that have a marking 3AA are made from steel. The marking 3AL or 3ALM indicates that the cylinder is made from aluminium • Cylinder have a flat or a concave base. The other end may taper into a neck that is fitted with tapered screw threads that attach to the cylinder valve
  • 16.
  • 17. • Valve: – Cylinders are filled and discharged through a valve (spindle shaped) attached to the neck – It is made of bronze or brass which is heavily plated with nickel or chromium so as to allow rapid dissipation of heat of compression – The end which enters the neck of the cylinder is threaded to fit a corresponding screw thread inside the neck itself
  • 18. • A sleeve or washer of soft alloy ( containing high proportions of lead) completes the gas tight seal as the valve is screwed into the neck of the cylinder • Cylinder valves are of various types. – Those used on anesthesia machines are ‘flush’ types which fits with the pin index system on the machine – For medium and large capacity cylinders – bull nose valves are used
  • 19.
  • 20.
  • 21. • Cylinder valves are of 2 types –packed type and diaphragm type – Packed type: • Capable of withstanding high pressures • A.k.a direct acting valve • Stem is sealed by resilient packing such as TEFLON which prevents leaks around the threads • In large cylinders, the force is transmitted by means of driver square • It is opened by 2-3 turns • Used in most of the cylinders
  • 22.
  • 23. – Diaphragm type: • Stem is separated from the seat • Closure between the cylinder interior and atmosphere is accomplished by using a seal and a bonnet nut that clamps one or more circular discs in place • These discs separate upper and lower stems which may be permanently attached to the diaphragms • Upper stem is actuated by manual/automatic means and Lower stem shuts/permits flow through the valve • Can be opened fully by ½ to ¾ turns • Seat does not turn-so less likely to leak • Generally preferred when pressures are relatively low and no leaks can be allowed • However - expensive
  • 24.
  • 25. • Port: – It is the point of exit for the gas It fits into the nipple on the hanger yoke of the anesthesia machine – It should be protected in transit by a covering – When installing a cylinder on anesthesia machine, it is important for the user not to mistake the port for the conical depression
  • 26.
  • 27. – Conical depression is situated on the opposite side of the port on the cylinder valve and is situated above the safety relief device It is present on those cylinders which are designed to fit on anesthesia machine – Conical depression is designed to receive the retaining screw on the yoke of the anesthesia machine  Screwing the retaining screw into the port may damage the port
  • 28.
  • 29. • Stem: – Each valve contains a stem (spindle/screw-pin) or shaft that is rotated to open or close the cylinder valve – It is made up of very hard steel – To close the valve the stem seals against the seat that is part of the valve  when the valve is opened –stem is moved upwards and allows the gas to flow to the port
  • 30. • Pressure Relief Devices: – Every cylinder is fitted with pressure relief devices whose purpose is to vent the cylinder’s contents to the atmosphere  if the pressure of enclosed gas increases to dangerous levels – Types: • Rupture Disc • Fusible Plug • Combination of Both • Pressure Relief valve (spring loaded)
  • 31. – Rupture Disc: • When pre-determined pressure is reached the disc ruptures and allows the gas contents to be discharged • It is a non- reclosing device held against an orifice • It protects against excess pressure as a result of high temperature/overfilling
  • 32.
  • 33. – Fusible Plug: • It is thermally operated • It is a non-reclosing pressure relief device where the plug is held against the discharge channel • It provides protection against excess pressure due to high temperature but not overfilling • YIELD TEMPERATURE: Temperature at which fusible material becomes sufficiently soft to extrude from its holder- so that cylinder contents are discharged
  • 34. – Spring loaded pressure relief device: • It is a reclosing device • When set pressure is exceeded, the pressure in the cylinder forced the spring to open the channel for letting out the gases Gas flows around the safety valve seat to discharge channel till excess pressure is relieved
  • 35. • Handle/ Handwheel: – It is used to open or close a cylinder valve – It is turned counter-clockwise to open and clock-wise to close  this causes the stem to turn – A good practice is to attach a handle to each anesthesia machine or other apparatus for which it may be needed – Each large cylinder has a permanent attached handwheel that uses a spring and nut  to hold it firmly in place
  • 36.
  • 37.
  • 38. Non-Interchangeable Safety Systems • With withspread use of cylinders containing different gases , a potential hazard is connection of a cylinder to equipment intended for a different gas • For safety purposes – Color coding for each gas – Pin index safety system
  • 39. Color coding GAS SHOULDER BODY OXYGEN WHITE BLACK NITROUS OXIDE BLUE BLUE CYCLOPROPANE ORANGE ORANGE CARBON DI OXIDE GREY GREY AIR WHITE GREY NITROGEN BLACK BLACK ENTONOX WHITE BLUE
  • 40. Pin Index Safety System • The use of PISS began in 1952 • In order to ensure that the correct cylinder is attached to the appropriate hanger yoke of the anesthesia machine/workstation A series of pins on the hanger yoke is made to fit into the corresponding indentations(pits/holes) drilled into cylinder valve
  • 41. • It consists of holes on the cylinder valve positioned in an arc below the outlet port • Positions of the cylinder valve are on the circumference of a circle of 9/16 inch (14.3mm) radius centered on the port • The port has a diameter of 7mm • The distance between the centre and lower part of the yoke is 20.6mm
  • 42.
  • 43. • Unless the pins and holes are aligned, the port will not seat • The indentations on the cylinder valve/yoke block are counted 1-6 from left to right • The distance between the centre of the 1st pin and 6th pin should be 16mm • There are 7 positions for pins and holes
  • 44. • The pins are 4.75mm in diameter and 6mm long except for pin number 7 which is slightly thicker and placed at the centre (between port 3 and 4)
  • 45. GAS PIN INDEX AIR 1,5 OXYGEN 2,5 NITROUS OXIDE 3,5 NITROGEN 1,4 O2-CO2 (CO2 <7.5%) 2,6 O2-CO2 (CO2 >7.5%) 1,6 ENTONOX 7
  • 46. Sizes • Gas suppliers classify cylinders by using a letter code with A being the smallest • Volume and Pressure of gas in a particular size cylinder vary • O2 and Air – similar ; CO2 and N20- similar • Size E is the most commonly used in anesthesia machine and for patient transport and resuscitation • Size D cylinders are used for limited supplies of gases where size and weight considerations are important
  • 47. CYLINDER SIZE DIMENSIONS (OUTER DIAMETER X LENGTH) IN INCH EMPTY CYLINDER VOLUME KG INTERNAL VOLUME L D 4.5 X 17 5 3 E 4.25 X 26 6.3 5 H 9.25 X 51 53 46-50
  • 48. O2/AIR N20/CO2 SIZE D – CONTENT IN LITRES 400 940 SIZE D- PRESSURE IN PSIG 1900 745/838 SIZE E- CONTENT IN LITRES 660 1590 SIZE E- PRESSURE IN PSIG 1900 745/838 SIZE H –CONTENT IN LITRES 6900 15800 SIZE H- PRESSURE IN PSIG 2200 745
  • 49. Contents and Pressure • In a cylinder containing a non-liquefied gas Pressure declines as the contents are withdrawn Hence pressure can be used to measure cylinder volume(approximately)
  • 50.
  • 51. • In a cylinder containing a liquified gas, the pressure depends on VAPOUR PRESSURE of the liquid and is not an indication of the amount of gas remaining in the cylinder as long as the contents are partly in the liquid state Pressure remains nearly constant till all liquid has evaporated After which pressure declines till cylinder is exhausted
  • 52.
  • 53. Testing • A cylinder must be inspected and tested atleast every 5 years or with special permit upto every 10 years • The test date must be permanently stamped on the cylinder • Each cylinder must pass an internal and external visual check for corrosion and evidence of physical impact or distortion
  • 54. • Cylinders are checked for leaks and retention of structural strength by testing to minimum of 1.66 times their service pressure • SERVICE PRESSURE IS DEFINED AS THE MAXIMUM PRESSURE TO WHICH THE CYLINDER MAY BE FILLED AT 70 F
  • 55. • Other tests that are done: – Tensile test – Flattening test – Bend test – Impact test – These are carried on atleast one out of every 100 cylinders
  • 56. Filling • If a cylinder containing a gas under a safe pressure at normal temperature is subjected to higher temperature the pressure may increase to dangerous levels • To prevent this, regulations have been drawn limiting the amount of gas a cylinder may contain
  • 57. • Non- liquefied gases – may be allowed an additional 10% filling • Liquefied gas containing cylinders: – Pressures will remain constant as long as there is liquid in cylinder  to prevent cylinder being overfilled Maximum amount of gas allowed is defined by the filling density(filling ratio) for each gas
  • 58. • Filling density- percent of ratio of weight of the gas in a cylinder to the weight of water that the cylinder would hold at 60 F – N20 – 68% – CO2- 68%
  • 59. Marking/Labelling/Tags • Important for identification • To check test date • In case of flammable gases- Caution/ Danger/ Warning label is needed • Tags should contain either full/ in use/ empty
  • 60.
  • 61. Hazards • Incorrect cylinder • Incorrect valve • Incorrect color/labelling • Inoperable valve/ Damaged valve • Suffocation/fire explosion • Overfilled contents in cylinder • Nitrous oxide theft • Contaminated contents in cylinders • Thermal injury
  • 62. Rules for safe use of cylinders 1. To be handled by trained staff 2. Store cylinders in a cool, clean room with adequate ventilation 3. Do not drape cylinder with any material during storage 4. Cylinders are best stored upright in a cylinder stand 5. Keep the valve closed when not in use
  • 63. 6. Identify contents by label 7. Remove wrappings( protective covering) before use 8. Remove dust/ foreign bodies before connecting 9. ‘Cracking’ to be done to reduce to possibility of flash fire
  • 64. 10. A sealing washer in good condition should be used 11. Flow control valves should be closed before cylinders opened 12. Quick opening to be avoided as it can generate heat leading to flame 13. Valve should be fully opened when in use 14. To be kept away from oil, rubber and combustible substances 15. Do not expose cylinder to heat or higher temperatures