CHAPTER 1: What is EMG?
The entire purpose of electrodiagnostic (END) studies is to help figure out whether
there is a problem in the nervous system and if so, where the problem is occurring.
The Most Common END Tests Done Are:
1. Nerve Conduction Studies (NCS)
2. Electromyography (EMG)
• Both NCS and EMG are normally done together.
• Both tests can provide different information.
• NCS provides information on nerves
• EMG provides information on nerves but can indirectly tell us information on the nerves
affecting them.
• Remember they are really an extension of Neurological and Musculoskeletal
examination
Nerve Conduction Studies (NCS)
NCS are done by placing electrodes on the skin and stimulating the nerves through electrical
impulses.
This is the basic setup for a
sensory nerve conduction study.
The machine gives a tracing
of the sensory nerve action
potential
Process of the NCS
• To study motor nerves, electrodes are placed over a muscle that receives its
innervation from the nerves you want to test (stimulate).
• The electrical response of the muscle is then recorded and you can
determine both how fast and how well the nerve responded.
• This can determine whether the problem stems from a nerve or muscle.
NCS are broken down into TWO categories:
1) Motor Nerve Conduction Testing
2) Sensory Nerve Conduction Testing
• NCS can be performed on any accessible nerve including peripheral nerves and
cranial nerves.
Basic Findings are generally two-fold:
1) How fast is the impulse traveling?
2) What does the electrical representation of the nerve stimulation (Action potential
morphology) look like on the screen?
Nerve Conduction Study Terms
• Action Potential: this is the waveform you see on the screen.
• Latency: Time interval between the onset of a stimulus and the onset of action potential
• Conduction Velocity: How fast the fastest part of the impulse travels (sensory or motor conduction
velocity)
• Amplitude: the maximal height of the action potential
• Antidromically: Physiologically opposite direction an impulse travels on a certain nerve.
• Orthodromically: Physiologically normal direction an impulse travels on a certain nerve.
• F-Wave: a compound muscle action potential evoked by antidromically stimulating a motor nerve from a
muscle using an electrical stimulus.
• H-Reflex (Hoffman Reflex): a compound muscle action potential evoked by orthodromically stimulating
sensory fibers, synapsing at the spinal level and returning orthodromically via motor fibers
Electromyography (EMG)
Electromyography (EMG) is the process by which an examiner puts a needle
into a particular muscle and studies the electrical activity of that muscle
• Electrical activity comes from the muscle itself- no shocks are used to
stimulate the muscle.
• Does not actually test nerves, however, does test nerves indirectly by testing
the muscles.
EMG vs NCS
1. You use a needle and put it into the muscle for EMG rather than
electrodes that are placed on the skin for NCS.
2. You don’t use any electrical shocks in EMG; rather you look at the intrinsic
electrical activity of the muscle.
3. You get direct information about the muscles in EMG and indirect
information about the nerves that supply the muscles you test
END of Chapter 1
CHAPTER 2: Why Do Electrodiagnostic
Studies?
Electrodiagnostic testing is an important method for physicians to distinguish
between many nerve and muscle disorders.
• They represent a part of the diagnostic puzzle
• They are sometimes essential in establishing a particular diagnosis and sometimes
they are not useful at all.
In a practical sense, you can consider electrodiagnostic testing in any of the following
circumstances, when a patient is complaining of or has:
1. numbness
2. tingling (parenthesis)
3. pain
4. weakness
5. a limp
6. Muscle atrophy
7. Depressed deep tendon reflexes
8. fatigue
However, the list of signs and symptoms do not automatically lead to electrodiagnostic
testing.
• These tests are an extension of the history and physical examination.
• They are useful in establishing the correct diagnosis, whether someone should have
surgery and prognostic reasons to follow the course of recovery (or deterioration)
from an injury.
Electrodiagnostic studies are useful to:
1) Establish the correct diagnosis.
2) Localize the lesion.
3) Determine treatment even if the diagnosis is already known.
4) Provide information about the prognosis.
The Skilled and Compassionate
Electrodiagnostician
It is important for you to be both technically skilled and able to put the patient at ease:
1. Avoid keeping the patient waiting, as that will increase his or her anxiety.
2. Before you start, explain to the patient what you are going to do.
3. Explain that these tests will be useful in determining the diagnosis.
4. Reassure the patient you will stop the test anytime they request you to.
5. Start with the area of greatest interest- especially if you suspect that the patient
won’t tolerate the entire study.
6. Although not typically used, analgesic or sedating medication can be given.
7. In most instances, it is best not to show patients the needle as many people
associate more pain with a longer needle, when a larger diameter needle would
actually cause more discomfort.
8. Assure the patient you will minimize the length of the exam, doing what is
absolutely necessary to obtain required information.
9. Keep the room warm. Two purposes:
a) The patient will most likely be dressed in a gown and therefore is prone to being cold.
b) The results of your electrodiagnostic test may be affected if the patient’s extremity is cold.
Special Precautions
There are a number of clinical situations that deserve special mention
• These are cases where the electrodiagnostic studies can be safely done, as long as the
physician takes measures to ensure the safety of the patient and the accuracy of the test.
Morbid Obesity: Overweight; it may be difficult (or impossible) to localize specific muscles.
Thin Individuals: important not to insert the needle too far as it can injure other tissues.
Bleeding Disorders: a patient with a known bleeding disorder or those who are on
Anticoagulant medications should be assessed on an individual basis and the risk/ benefit of
the test evaluated.
Blood Precautions: practice safe needle stick procedures.
Contraindications:
• For EMG: Patient with severe bleeding disorders or whose anticoagulation therapy is
out of control.
• For NCS: Patient with cardiac defibrillators implanted.
Complications:
• Infection: bleeding and accidently penetrating into some other part.
Controversy:
• May cause some discomfort (pain)
• Relatively expensive test to perform
About the Machine
The Basic Machine consists of a computer, the associated
hardware and software.
All basic electrodiagnostic software allows the clinician to:
• Perform both EMG and NCS
• Collect data.
• Help to analyze results (through automatic calculations
that are usually preprogrammed)
• Store the information.
Data entry through the keyboard.
Information displayed on the screen (EMG has both audio
and visual data).
Recording Electrodes
Surface Electrodes are routine for NCS and come in
• Ring or Disk
a. Disposable- sticky underside
b. Non-disposable- made of stainless steel, silver, gold or etc. Used with adhesive tape and
need the application of conduction gel to reduce impedance and prevent artifacts.
Electrodes of NCS (Surface
electrodes)
Electrodes of EMG
Active: surface (pickup) Active: needle
Reference: surface (recording) *Reference: surface
Ground: surface Ground Surface
* No need for a concentric needle
Needle Electrodes are reserved for EMG but
occasionally for NCS
• Nearly all needle electrodes are disposable and
are used only on one patient.
THREE classes of Needle Electrodes
1) Monopolar Needles: typically less
expensive, less painful and less stable when
compared to Concentric or bipolar needles.
2) Bipolar Needles: more electrically stable
3) Concentric Needles: the reference is the
barrel of the needle, so you do not need a
separate electrode for the reference.
Amplifiers
Amplifiers magnify the signal so that it can be displayed.
Preamplifiers attenuate the biological signal before it gets to the amplifier in
order to:
1. Make sure that the filters have sufficient signal voltage to work with.
2. To make sure that the level of signal voltage is much higher than that of
system noise.
Preamplifiers  Filters  Amplifier
Differential Amplifiers are used in electrodiagnostic studies because of it Common
Mode Rejection Feature.
• Common Mode Rejection (CMR) rejects unwanted signals from the electrodes being
compared that are similar. (i.e. 60 Hz Signal)
• Differential Amplifiers take the electrical impulses from the active electrode and
amplifies them.
• Takes impulses from the reference electrode, inverts them, and amplifies them.
• Combines these two potentials.
• Cancels out any common noise to both electrodes (i.e. 60 Hz artifact present in both electrodes
are cancelled out).
• Differences, however, will be amplified.
CMR ratio is how well an amplifier eliminates similar types of noise
Filters
Filters are used to faithfully reproduce the signal you want while trying to exclude both
High and Low Frequency electrical noise.
• Every signal in both NCS and EMG passes through both a low frequency filter
before being displayed.
• It is important to understand that there is always a trade off when you use filters (the
signal will be altered to some degree)
Type of Filter Other Name Description
Low Frequency
Filter
High Pass Filters out low frequency signals that if present cause a
wandering baseline.
High Frequency
Filter
Low Pass Filters out high frequency signals that if present can cause a
“noisy” baseline especially on sensory studies.
Display System
Display systems for electrodiagnostic
studies are either via a video screen or
paper
Cathode Ray tube is the standard
display, which uses a controlled
electron beam (cathode ray) to excite
phosphorus on the screen that
presents a visual display.
Two Settings on the Display Screen
1. Sweep Speed: the horizontal axis; represents how long an action potential can
last; milliseconds (ms or msec)
2. Sensitivity (gain): Vertical axis; represents response amplitude; Millivolts (mV)
for Motor Studies or microvolts (µV) for Sensory Studies
Initial Motor NCS Settings
Sweep Speed 2-3 msec/division
Sensitivity 5000 µV/ division (5 mV/div)
LFF 10 Hz
HFF 10kHz
Initial Sensory NCS Settings
Sweep Speed 1-2 msec/division (generally 10 divisions are present in a
horizontal screen)
Sensitivity (gain) 20 µV/ division
LFF 2-10 Hz
HFF 2 kHz
Artifacts and Technical Factors
Physiological Factors (Patient- related)
• Stimulus Artifact is an electrically recorded response that is elicited directly from the
stimulator.
• Occurs in all NCS, however, it only becomes a problem when the trailing edge of the recorded artifact
overlaps with the potential being recorded.
• Make sure the ground is between is the recording and stimulating electrodes to minimize stimulus
artifact.
Technical Factors (technically related)
• Filters: faithfully reproduce the signal you want while trying to exclude both high and low
frequencies.
• Electrode Placement: proper electrode placement is a critical part of performing accurate
electrodiagnostic studies.
Stimulation
Supramaximal Stimulation: all measurements made in NCS are done with the
assumption that the strength of the stimulus is high enough to depolarize every axon
in the nerve.
• If Supramaximal stimulation is not achieved at the Distal Site, then you might
mistakenly interpret this recording as signifying axonal loss due to low amplitude.
• If Supramaximal stimulation is not achieved at the Proximal Site, this might appear
to be a conduction block.
• Conduction block is the failure of an Action Potential to be conducted past a particular
point where as conduction is possible below the point of the block.
With Respect to Evoked Potentials the stimulus can be graded as:
• Subthreshold: when the stimulus is less than the threshold.
• Threshold: when the stimulus is sufficient to produce a detectable response.
• Submaximal: is stimuli below the maximal intensity.
• Maximal: is the stimulus intensity after which a further increase in the stimulus
intensity causes no increase in the amplitude.
• Supramaximal: is intensity approximately 20% more than maximal, and used for
Nerve Conduction Studies.
For NCS, too much stimulation may cause co-stimulation of adjacent nerves or may
stimulate nerves farther from the site
The goal is to reach Supramaximal stimulation without applying so much stimulation
that adjacent nerve are also stimulated.
Measurements
The machine will do most of the calculations for you
• However, you still need to measure the distance between stimulation sites and the
recording electrode of a patient’s limb when determining conduction velocity.
Make sure the measurements are done accurately.
• The main problem is if the patient’s limb is moved in different positions, which
changes the distance you are measuring. (Commonly occurs in ulnar nerve studies.)
Remember to follow the course of the nerve, rather than measuring the
shortest distance between the stimulating and recording electrodes.
Sweep Speed & Sensitivity
Both the sweep-speed and sensitivity can affect you NCS results.
• As the sensitivity is increased, the onset latency will decrease
• Record all latency measurement with the same sensitivity and sweep speed.
END of Chapter 2

Más contenido relacionado

La actualidad más candente (20)

Nerve conduction studies
Nerve conduction studiesNerve conduction studies
Nerve conduction studies
 
Overview on Motor evoked potential
Overview on Motor evoked potentialOverview on Motor evoked potential
Overview on Motor evoked potential
 
Emg
EmgEmg
Emg
 
Interpretation of NCS and EMG
Interpretation of NCS and EMG Interpretation of NCS and EMG
Interpretation of NCS and EMG
 
Nerves conduction study
Nerves conduction study Nerves conduction study
Nerves conduction study
 
Basics of emg
Basics of emgBasics of emg
Basics of emg
 
EEG & Evoked potentials
EEG & Evoked potentialsEEG & Evoked potentials
EEG & Evoked potentials
 
Overview of Nerve Conduction Study
Overview of Nerve Conduction StudyOverview of Nerve Conduction Study
Overview of Nerve Conduction Study
 
NCV AND EMG
NCV AND EMGNCV AND EMG
NCV AND EMG
 
Nerve Conduction Velocity: NCV
Nerve Conduction Velocity: NCVNerve Conduction Velocity: NCV
Nerve Conduction Velocity: NCV
 
Electromyography (emg) basics
Electromyography (emg)   basicsElectromyography (emg)   basics
Electromyography (emg) basics
 
repetitive nerve stimulation
repetitive nerve stimulationrepetitive nerve stimulation
repetitive nerve stimulation
 
Intraoperative Electromyography (EMG)
Intraoperative Electromyography (EMG)Intraoperative Electromyography (EMG)
Intraoperative Electromyography (EMG)
 
Somato Sensory Evoked Potentials (SSEP) By: Murtaza Syed
Somato Sensory Evoked Potentials (SSEP) By: Murtaza SyedSomato Sensory Evoked Potentials (SSEP) By: Murtaza Syed
Somato Sensory Evoked Potentials (SSEP) By: Murtaza Syed
 
Somatosensory evoked potentials
Somatosensory evoked potentialsSomatosensory evoked potentials
Somatosensory evoked potentials
 
Evolked potential
Evolked potentialEvolked potential
Evolked potential
 
EMG
EMGEMG
EMG
 
Electromyogram
ElectromyogramElectromyogram
Electromyogram
 
Emg biofeedback
Emg biofeedbackEmg biofeedback
Emg biofeedback
 
Evoked potentials (1)
Evoked potentials (1)Evoked potentials (1)
Evoked potentials (1)
 

Similar a Emg and ncs slides chapter 1 and 2

Electromyography.pptx for electrical diagnosed
Electromyography.pptx for electrical diagnosedElectromyography.pptx for electrical diagnosed
Electromyography.pptx for electrical diagnosedAhsanAli479495
 
ELECTROMYOGRAPHY.pptx
ELECTROMYOGRAPHY.pptxELECTROMYOGRAPHY.pptx
ELECTROMYOGRAPHY.pptxBatul Dawoodi
 
Emg changes during fatigue and contraction
Emg changes during fatigue and contractionEmg changes during fatigue and contraction
Emg changes during fatigue and contractionDr Akshay RAj Chandra PT
 
EMG project by BME student, funny project.pptx
EMG project by BME student, funny project.pptxEMG project by BME student, funny project.pptx
EMG project by BME student, funny project.pptxzamericanislam
 
Electro diagnostic tests ppt
Electro diagnostic tests pptElectro diagnostic tests ppt
Electro diagnostic tests pptKanchan Sharma
 
Neuroimaging (MRI, PET, SPECT ,fMRI).pptx
 Neuroimaging (MRI, PET, SPECT ,fMRI).pptx Neuroimaging (MRI, PET, SPECT ,fMRI).pptx
Neuroimaging (MRI, PET, SPECT ,fMRI).pptxMr SACHIN
 
Medical Instrumentation- EMG,ERG,EOG,PCG
Medical Instrumentation- EMG,ERG,EOG,PCGMedical Instrumentation- EMG,ERG,EOG,PCG
Medical Instrumentation- EMG,ERG,EOG,PCGPoornima D
 
L1-_EMGMNCV.power point presentation for ncv
L1-_EMGMNCV.power point presentation for ncvL1-_EMGMNCV.power point presentation for ncv
L1-_EMGMNCV.power point presentation for ncvPranavTrehan2
 
EMG PRESENTATION RAGHU KIMS CME 2023 .pptx
EMG PRESENTATION RAGHU KIMS CME 2023 .pptxEMG PRESENTATION RAGHU KIMS CME 2023 .pptx
EMG PRESENTATION RAGHU KIMS CME 2023 .pptxRaghu Nadh
 
Nerve Conduction Tickets to Assess adminsi .pptx
Nerve Conduction Tickets to Assess adminsi .pptxNerve Conduction Tickets to Assess adminsi .pptx
Nerve Conduction Tickets to Assess adminsi .pptxeddiesonabesamis
 
Intraoperative neurophysiologic monitoring of the spine
Intraoperative neurophysiologic monitoring of the spineIntraoperative neurophysiologic monitoring of the spine
Intraoperative neurophysiologic monitoring of the spinespine spine
 

Similar a Emg and ncs slides chapter 1 and 2 (20)

Electromyography.pptx for electrical diagnosed
Electromyography.pptx for electrical diagnosedElectromyography.pptx for electrical diagnosed
Electromyography.pptx for electrical diagnosed
 
ELECTROMYOGRAPHY.pptx
ELECTROMYOGRAPHY.pptxELECTROMYOGRAPHY.pptx
ELECTROMYOGRAPHY.pptx
 
Emg changes during fatigue and contraction
Emg changes during fatigue and contractionEmg changes during fatigue and contraction
Emg changes during fatigue and contraction
 
EMG project by BME student, funny project.pptx
EMG project by BME student, funny project.pptxEMG project by BME student, funny project.pptx
EMG project by BME student, funny project.pptx
 
PPT 1 BIOLOGY EMG ECG.pptx
PPT 1 BIOLOGY EMG ECG.pptxPPT 1 BIOLOGY EMG ECG.pptx
PPT 1 BIOLOGY EMG ECG.pptx
 
Electro diagnostic tests ppt
Electro diagnostic tests pptElectro diagnostic tests ppt
Electro diagnostic tests ppt
 
10321518.ppt
10321518.ppt10321518.ppt
10321518.ppt
 
Neuroimaging (MRI, PET, SPECT ,fMRI).pptx
 Neuroimaging (MRI, PET, SPECT ,fMRI).pptx Neuroimaging (MRI, PET, SPECT ,fMRI).pptx
Neuroimaging (MRI, PET, SPECT ,fMRI).pptx
 
Medical Instrumentation- EMG,ERG,EOG,PCG
Medical Instrumentation- EMG,ERG,EOG,PCGMedical Instrumentation- EMG,ERG,EOG,PCG
Medical Instrumentation- EMG,ERG,EOG,PCG
 
NCV and EMG.pptx
NCV and EMG.pptxNCV and EMG.pptx
NCV and EMG.pptx
 
EEG & EMG new.pptx
EEG & EMG new.pptxEEG & EMG new.pptx
EEG & EMG new.pptx
 
EMG
EMGEMG
EMG
 
L1- EMG+MNCV.ppt
L1- EMG+MNCV.pptL1- EMG+MNCV.ppt
L1- EMG+MNCV.ppt
 
L1-_EMGMNCV.power point presentation for ncv
L1-_EMGMNCV.power point presentation for ncvL1-_EMGMNCV.power point presentation for ncv
L1-_EMGMNCV.power point presentation for ncv
 
EMG PRESENTATION RAGHU KIMS CME 2023 .pptx
EMG PRESENTATION RAGHU KIMS CME 2023 .pptxEMG PRESENTATION RAGHU KIMS CME 2023 .pptx
EMG PRESENTATION RAGHU KIMS CME 2023 .pptx
 
Nerve Conduction Tickets to Assess adminsi .pptx
Nerve Conduction Tickets to Assess adminsi .pptxNerve Conduction Tickets to Assess adminsi .pptx
Nerve Conduction Tickets to Assess adminsi .pptx
 
CNS DIAGNOSTIC TESTS
CNS DIAGNOSTIC TESTSCNS DIAGNOSTIC TESTS
CNS DIAGNOSTIC TESTS
 
Intraoperative neurophysiologic monitoring of the spine
Intraoperative neurophysiologic monitoring of the spineIntraoperative neurophysiologic monitoring of the spine
Intraoperative neurophysiologic monitoring of the spine
 
Ect electrical stimulus and procedure
Ect  electrical stimulus and procedureEct  electrical stimulus and procedure
Ect electrical stimulus and procedure
 
E.M.G and N.C.V.pptx
E.M.G and N.C.V.pptxE.M.G and N.C.V.pptx
E.M.G and N.C.V.pptx
 

Último

Asymmetry in the atmosphere of the ultra-hot Jupiter WASP-76 b
Asymmetry in the atmosphere of the ultra-hot Jupiter WASP-76 bAsymmetry in the atmosphere of the ultra-hot Jupiter WASP-76 b
Asymmetry in the atmosphere of the ultra-hot Jupiter WASP-76 bSérgio Sacani
 
Chemical Tests; flame test, positive and negative ions test Edexcel Internati...
Chemical Tests; flame test, positive and negative ions test Edexcel Internati...Chemical Tests; flame test, positive and negative ions test Edexcel Internati...
Chemical Tests; flame test, positive and negative ions test Edexcel Internati...ssuser79fe74
 
High Class Escorts in Hyderabad ₹7.5k Pick Up & Drop With Cash Payment 969456...
High Class Escorts in Hyderabad ₹7.5k Pick Up & Drop With Cash Payment 969456...High Class Escorts in Hyderabad ₹7.5k Pick Up & Drop With Cash Payment 969456...
High Class Escorts in Hyderabad ₹7.5k Pick Up & Drop With Cash Payment 969456...chandars293
 
GUIDELINES ON SIMILAR BIOLOGICS Regulatory Requirements for Marketing Authori...
GUIDELINES ON SIMILAR BIOLOGICS Regulatory Requirements for Marketing Authori...GUIDELINES ON SIMILAR BIOLOGICS Regulatory Requirements for Marketing Authori...
GUIDELINES ON SIMILAR BIOLOGICS Regulatory Requirements for Marketing Authori...Lokesh Kothari
 
Recombination DNA Technology (Nucleic Acid Hybridization )
Recombination DNA Technology (Nucleic Acid Hybridization )Recombination DNA Technology (Nucleic Acid Hybridization )
Recombination DNA Technology (Nucleic Acid Hybridization )aarthirajkumar25
 
Kochi ❤CALL GIRL 84099*07087 ❤CALL GIRLS IN Kochi ESCORT SERVICE❤CALL GIRL
Kochi ❤CALL GIRL 84099*07087 ❤CALL GIRLS IN Kochi ESCORT SERVICE❤CALL GIRLKochi ❤CALL GIRL 84099*07087 ❤CALL GIRLS IN Kochi ESCORT SERVICE❤CALL GIRL
Kochi ❤CALL GIRL 84099*07087 ❤CALL GIRLS IN Kochi ESCORT SERVICE❤CALL GIRLkantirani197
 
GBSN - Biochemistry (Unit 1)
GBSN - Biochemistry (Unit 1)GBSN - Biochemistry (Unit 1)
GBSN - Biochemistry (Unit 1)Areesha Ahmad
 
GBSN - Microbiology (Unit 2)
GBSN - Microbiology (Unit 2)GBSN - Microbiology (Unit 2)
GBSN - Microbiology (Unit 2)Areesha Ahmad
 
Nanoparticles synthesis and characterization​ ​
Nanoparticles synthesis and characterization​  ​Nanoparticles synthesis and characterization​  ​
Nanoparticles synthesis and characterization​ ​kaibalyasahoo82800
 
Chemistry 4th semester series (krishna).pdf
Chemistry 4th semester series (krishna).pdfChemistry 4th semester series (krishna).pdf
Chemistry 4th semester series (krishna).pdfSumit Kumar yadav
 
Discovery of an Accretion Streamer and a Slow Wide-angle Outflow around FUOri...
Discovery of an Accretion Streamer and a Slow Wide-angle Outflow around FUOri...Discovery of an Accretion Streamer and a Slow Wide-angle Outflow around FUOri...
Discovery of an Accretion Streamer and a Slow Wide-angle Outflow around FUOri...Sérgio Sacani
 
Pulmonary drug delivery system M.pharm -2nd sem P'ceutics
Pulmonary drug delivery system M.pharm -2nd sem P'ceuticsPulmonary drug delivery system M.pharm -2nd sem P'ceutics
Pulmonary drug delivery system M.pharm -2nd sem P'ceuticssakshisoni2385
 
SCIENCE-4-QUARTER4-WEEK-4-PPT-1 (1).pptx
SCIENCE-4-QUARTER4-WEEK-4-PPT-1 (1).pptxSCIENCE-4-QUARTER4-WEEK-4-PPT-1 (1).pptx
SCIENCE-4-QUARTER4-WEEK-4-PPT-1 (1).pptxRizalinePalanog2
 
Stunning ➥8448380779▻ Call Girls In Panchshil Enclave Delhi NCR
Stunning ➥8448380779▻ Call Girls In Panchshil Enclave Delhi NCRStunning ➥8448380779▻ Call Girls In Panchshil Enclave Delhi NCR
Stunning ➥8448380779▻ Call Girls In Panchshil Enclave Delhi NCRDelhi Call girls
 
9654467111 Call Girls In Raj Nagar Delhi Short 1500 Night 6000
9654467111 Call Girls In Raj Nagar Delhi Short 1500 Night 60009654467111 Call Girls In Raj Nagar Delhi Short 1500 Night 6000
9654467111 Call Girls In Raj Nagar Delhi Short 1500 Night 6000Sapana Sha
 
GBSN - Microbiology (Unit 1)
GBSN - Microbiology (Unit 1)GBSN - Microbiology (Unit 1)
GBSN - Microbiology (Unit 1)Areesha Ahmad
 
Green chemistry and Sustainable development.pptx
Green chemistry  and Sustainable development.pptxGreen chemistry  and Sustainable development.pptx
Green chemistry and Sustainable development.pptxRajatChauhan518211
 
Pests of mustard_Identification_Management_Dr.UPR.pdf
Pests of mustard_Identification_Management_Dr.UPR.pdfPests of mustard_Identification_Management_Dr.UPR.pdf
Pests of mustard_Identification_Management_Dr.UPR.pdfPirithiRaju
 
Recombinant DNA technology (Immunological screening)
Recombinant DNA technology (Immunological screening)Recombinant DNA technology (Immunological screening)
Recombinant DNA technology (Immunological screening)PraveenaKalaiselvan1
 
Biopesticide (2).pptx .This slides helps to know the different types of biop...
Biopesticide (2).pptx  .This slides helps to know the different types of biop...Biopesticide (2).pptx  .This slides helps to know the different types of biop...
Biopesticide (2).pptx .This slides helps to know the different types of biop...RohitNehra6
 

Último (20)

Asymmetry in the atmosphere of the ultra-hot Jupiter WASP-76 b
Asymmetry in the atmosphere of the ultra-hot Jupiter WASP-76 bAsymmetry in the atmosphere of the ultra-hot Jupiter WASP-76 b
Asymmetry in the atmosphere of the ultra-hot Jupiter WASP-76 b
 
Chemical Tests; flame test, positive and negative ions test Edexcel Internati...
Chemical Tests; flame test, positive and negative ions test Edexcel Internati...Chemical Tests; flame test, positive and negative ions test Edexcel Internati...
Chemical Tests; flame test, positive and negative ions test Edexcel Internati...
 
High Class Escorts in Hyderabad ₹7.5k Pick Up & Drop With Cash Payment 969456...
High Class Escorts in Hyderabad ₹7.5k Pick Up & Drop With Cash Payment 969456...High Class Escorts in Hyderabad ₹7.5k Pick Up & Drop With Cash Payment 969456...
High Class Escorts in Hyderabad ₹7.5k Pick Up & Drop With Cash Payment 969456...
 
GUIDELINES ON SIMILAR BIOLOGICS Regulatory Requirements for Marketing Authori...
GUIDELINES ON SIMILAR BIOLOGICS Regulatory Requirements for Marketing Authori...GUIDELINES ON SIMILAR BIOLOGICS Regulatory Requirements for Marketing Authori...
GUIDELINES ON SIMILAR BIOLOGICS Regulatory Requirements for Marketing Authori...
 
Recombination DNA Technology (Nucleic Acid Hybridization )
Recombination DNA Technology (Nucleic Acid Hybridization )Recombination DNA Technology (Nucleic Acid Hybridization )
Recombination DNA Technology (Nucleic Acid Hybridization )
 
Kochi ❤CALL GIRL 84099*07087 ❤CALL GIRLS IN Kochi ESCORT SERVICE❤CALL GIRL
Kochi ❤CALL GIRL 84099*07087 ❤CALL GIRLS IN Kochi ESCORT SERVICE❤CALL GIRLKochi ❤CALL GIRL 84099*07087 ❤CALL GIRLS IN Kochi ESCORT SERVICE❤CALL GIRL
Kochi ❤CALL GIRL 84099*07087 ❤CALL GIRLS IN Kochi ESCORT SERVICE❤CALL GIRL
 
GBSN - Biochemistry (Unit 1)
GBSN - Biochemistry (Unit 1)GBSN - Biochemistry (Unit 1)
GBSN - Biochemistry (Unit 1)
 
GBSN - Microbiology (Unit 2)
GBSN - Microbiology (Unit 2)GBSN - Microbiology (Unit 2)
GBSN - Microbiology (Unit 2)
 
Nanoparticles synthesis and characterization​ ​
Nanoparticles synthesis and characterization​  ​Nanoparticles synthesis and characterization​  ​
Nanoparticles synthesis and characterization​ ​
 
Chemistry 4th semester series (krishna).pdf
Chemistry 4th semester series (krishna).pdfChemistry 4th semester series (krishna).pdf
Chemistry 4th semester series (krishna).pdf
 
Discovery of an Accretion Streamer and a Slow Wide-angle Outflow around FUOri...
Discovery of an Accretion Streamer and a Slow Wide-angle Outflow around FUOri...Discovery of an Accretion Streamer and a Slow Wide-angle Outflow around FUOri...
Discovery of an Accretion Streamer and a Slow Wide-angle Outflow around FUOri...
 
Pulmonary drug delivery system M.pharm -2nd sem P'ceutics
Pulmonary drug delivery system M.pharm -2nd sem P'ceuticsPulmonary drug delivery system M.pharm -2nd sem P'ceutics
Pulmonary drug delivery system M.pharm -2nd sem P'ceutics
 
SCIENCE-4-QUARTER4-WEEK-4-PPT-1 (1).pptx
SCIENCE-4-QUARTER4-WEEK-4-PPT-1 (1).pptxSCIENCE-4-QUARTER4-WEEK-4-PPT-1 (1).pptx
SCIENCE-4-QUARTER4-WEEK-4-PPT-1 (1).pptx
 
Stunning ➥8448380779▻ Call Girls In Panchshil Enclave Delhi NCR
Stunning ➥8448380779▻ Call Girls In Panchshil Enclave Delhi NCRStunning ➥8448380779▻ Call Girls In Panchshil Enclave Delhi NCR
Stunning ➥8448380779▻ Call Girls In Panchshil Enclave Delhi NCR
 
9654467111 Call Girls In Raj Nagar Delhi Short 1500 Night 6000
9654467111 Call Girls In Raj Nagar Delhi Short 1500 Night 60009654467111 Call Girls In Raj Nagar Delhi Short 1500 Night 6000
9654467111 Call Girls In Raj Nagar Delhi Short 1500 Night 6000
 
GBSN - Microbiology (Unit 1)
GBSN - Microbiology (Unit 1)GBSN - Microbiology (Unit 1)
GBSN - Microbiology (Unit 1)
 
Green chemistry and Sustainable development.pptx
Green chemistry  and Sustainable development.pptxGreen chemistry  and Sustainable development.pptx
Green chemistry and Sustainable development.pptx
 
Pests of mustard_Identification_Management_Dr.UPR.pdf
Pests of mustard_Identification_Management_Dr.UPR.pdfPests of mustard_Identification_Management_Dr.UPR.pdf
Pests of mustard_Identification_Management_Dr.UPR.pdf
 
Recombinant DNA technology (Immunological screening)
Recombinant DNA technology (Immunological screening)Recombinant DNA technology (Immunological screening)
Recombinant DNA technology (Immunological screening)
 
Biopesticide (2).pptx .This slides helps to know the different types of biop...
Biopesticide (2).pptx  .This slides helps to know the different types of biop...Biopesticide (2).pptx  .This slides helps to know the different types of biop...
Biopesticide (2).pptx .This slides helps to know the different types of biop...
 

Emg and ncs slides chapter 1 and 2

  • 2. CHAPTER 1: What is EMG? The entire purpose of electrodiagnostic (END) studies is to help figure out whether there is a problem in the nervous system and if so, where the problem is occurring.
  • 3. The Most Common END Tests Done Are: 1. Nerve Conduction Studies (NCS) 2. Electromyography (EMG) • Both NCS and EMG are normally done together. • Both tests can provide different information. • NCS provides information on nerves • EMG provides information on nerves but can indirectly tell us information on the nerves affecting them. • Remember they are really an extension of Neurological and Musculoskeletal examination
  • 4. Nerve Conduction Studies (NCS) NCS are done by placing electrodes on the skin and stimulating the nerves through electrical impulses. This is the basic setup for a sensory nerve conduction study. The machine gives a tracing of the sensory nerve action potential
  • 5. Process of the NCS • To study motor nerves, electrodes are placed over a muscle that receives its innervation from the nerves you want to test (stimulate). • The electrical response of the muscle is then recorded and you can determine both how fast and how well the nerve responded. • This can determine whether the problem stems from a nerve or muscle.
  • 6. NCS are broken down into TWO categories: 1) Motor Nerve Conduction Testing 2) Sensory Nerve Conduction Testing • NCS can be performed on any accessible nerve including peripheral nerves and cranial nerves. Basic Findings are generally two-fold: 1) How fast is the impulse traveling? 2) What does the electrical representation of the nerve stimulation (Action potential morphology) look like on the screen?
  • 7. Nerve Conduction Study Terms • Action Potential: this is the waveform you see on the screen. • Latency: Time interval between the onset of a stimulus and the onset of action potential • Conduction Velocity: How fast the fastest part of the impulse travels (sensory or motor conduction velocity) • Amplitude: the maximal height of the action potential • Antidromically: Physiologically opposite direction an impulse travels on a certain nerve. • Orthodromically: Physiologically normal direction an impulse travels on a certain nerve. • F-Wave: a compound muscle action potential evoked by antidromically stimulating a motor nerve from a muscle using an electrical stimulus. • H-Reflex (Hoffman Reflex): a compound muscle action potential evoked by orthodromically stimulating sensory fibers, synapsing at the spinal level and returning orthodromically via motor fibers
  • 8. Electromyography (EMG) Electromyography (EMG) is the process by which an examiner puts a needle into a particular muscle and studies the electrical activity of that muscle • Electrical activity comes from the muscle itself- no shocks are used to stimulate the muscle. • Does not actually test nerves, however, does test nerves indirectly by testing the muscles.
  • 9. EMG vs NCS 1. You use a needle and put it into the muscle for EMG rather than electrodes that are placed on the skin for NCS. 2. You don’t use any electrical shocks in EMG; rather you look at the intrinsic electrical activity of the muscle. 3. You get direct information about the muscles in EMG and indirect information about the nerves that supply the muscles you test
  • 11. CHAPTER 2: Why Do Electrodiagnostic Studies? Electrodiagnostic testing is an important method for physicians to distinguish between many nerve and muscle disorders. • They represent a part of the diagnostic puzzle • They are sometimes essential in establishing a particular diagnosis and sometimes they are not useful at all.
  • 12. In a practical sense, you can consider electrodiagnostic testing in any of the following circumstances, when a patient is complaining of or has: 1. numbness 2. tingling (parenthesis) 3. pain 4. weakness 5. a limp 6. Muscle atrophy 7. Depressed deep tendon reflexes 8. fatigue
  • 13. However, the list of signs and symptoms do not automatically lead to electrodiagnostic testing. • These tests are an extension of the history and physical examination. • They are useful in establishing the correct diagnosis, whether someone should have surgery and prognostic reasons to follow the course of recovery (or deterioration) from an injury. Electrodiagnostic studies are useful to: 1) Establish the correct diagnosis. 2) Localize the lesion. 3) Determine treatment even if the diagnosis is already known. 4) Provide information about the prognosis.
  • 14. The Skilled and Compassionate Electrodiagnostician It is important for you to be both technically skilled and able to put the patient at ease: 1. Avoid keeping the patient waiting, as that will increase his or her anxiety. 2. Before you start, explain to the patient what you are going to do. 3. Explain that these tests will be useful in determining the diagnosis. 4. Reassure the patient you will stop the test anytime they request you to. 5. Start with the area of greatest interest- especially if you suspect that the patient won’t tolerate the entire study.
  • 15. 6. Although not typically used, analgesic or sedating medication can be given. 7. In most instances, it is best not to show patients the needle as many people associate more pain with a longer needle, when a larger diameter needle would actually cause more discomfort. 8. Assure the patient you will minimize the length of the exam, doing what is absolutely necessary to obtain required information. 9. Keep the room warm. Two purposes: a) The patient will most likely be dressed in a gown and therefore is prone to being cold. b) The results of your electrodiagnostic test may be affected if the patient’s extremity is cold.
  • 16. Special Precautions There are a number of clinical situations that deserve special mention • These are cases where the electrodiagnostic studies can be safely done, as long as the physician takes measures to ensure the safety of the patient and the accuracy of the test. Morbid Obesity: Overweight; it may be difficult (or impossible) to localize specific muscles. Thin Individuals: important not to insert the needle too far as it can injure other tissues. Bleeding Disorders: a patient with a known bleeding disorder or those who are on Anticoagulant medications should be assessed on an individual basis and the risk/ benefit of the test evaluated. Blood Precautions: practice safe needle stick procedures.
  • 17. Contraindications: • For EMG: Patient with severe bleeding disorders or whose anticoagulation therapy is out of control. • For NCS: Patient with cardiac defibrillators implanted. Complications: • Infection: bleeding and accidently penetrating into some other part. Controversy: • May cause some discomfort (pain) • Relatively expensive test to perform
  • 18. About the Machine The Basic Machine consists of a computer, the associated hardware and software. All basic electrodiagnostic software allows the clinician to: • Perform both EMG and NCS • Collect data. • Help to analyze results (through automatic calculations that are usually preprogrammed) • Store the information. Data entry through the keyboard. Information displayed on the screen (EMG has both audio and visual data).
  • 19. Recording Electrodes Surface Electrodes are routine for NCS and come in • Ring or Disk a. Disposable- sticky underside b. Non-disposable- made of stainless steel, silver, gold or etc. Used with adhesive tape and need the application of conduction gel to reduce impedance and prevent artifacts. Electrodes of NCS (Surface electrodes) Electrodes of EMG Active: surface (pickup) Active: needle Reference: surface (recording) *Reference: surface Ground: surface Ground Surface * No need for a concentric needle
  • 20. Needle Electrodes are reserved for EMG but occasionally for NCS • Nearly all needle electrodes are disposable and are used only on one patient. THREE classes of Needle Electrodes 1) Monopolar Needles: typically less expensive, less painful and less stable when compared to Concentric or bipolar needles. 2) Bipolar Needles: more electrically stable 3) Concentric Needles: the reference is the barrel of the needle, so you do not need a separate electrode for the reference.
  • 21. Amplifiers Amplifiers magnify the signal so that it can be displayed. Preamplifiers attenuate the biological signal before it gets to the amplifier in order to: 1. Make sure that the filters have sufficient signal voltage to work with. 2. To make sure that the level of signal voltage is much higher than that of system noise. Preamplifiers  Filters  Amplifier
  • 22. Differential Amplifiers are used in electrodiagnostic studies because of it Common Mode Rejection Feature. • Common Mode Rejection (CMR) rejects unwanted signals from the electrodes being compared that are similar. (i.e. 60 Hz Signal) • Differential Amplifiers take the electrical impulses from the active electrode and amplifies them. • Takes impulses from the reference electrode, inverts them, and amplifies them. • Combines these two potentials. • Cancels out any common noise to both electrodes (i.e. 60 Hz artifact present in both electrodes are cancelled out). • Differences, however, will be amplified. CMR ratio is how well an amplifier eliminates similar types of noise
  • 23. Filters Filters are used to faithfully reproduce the signal you want while trying to exclude both High and Low Frequency electrical noise. • Every signal in both NCS and EMG passes through both a low frequency filter before being displayed. • It is important to understand that there is always a trade off when you use filters (the signal will be altered to some degree) Type of Filter Other Name Description Low Frequency Filter High Pass Filters out low frequency signals that if present cause a wandering baseline. High Frequency Filter Low Pass Filters out high frequency signals that if present can cause a “noisy” baseline especially on sensory studies.
  • 24. Display System Display systems for electrodiagnostic studies are either via a video screen or paper Cathode Ray tube is the standard display, which uses a controlled electron beam (cathode ray) to excite phosphorus on the screen that presents a visual display.
  • 25. Two Settings on the Display Screen 1. Sweep Speed: the horizontal axis; represents how long an action potential can last; milliseconds (ms or msec) 2. Sensitivity (gain): Vertical axis; represents response amplitude; Millivolts (mV) for Motor Studies or microvolts (µV) for Sensory Studies
  • 26. Initial Motor NCS Settings Sweep Speed 2-3 msec/division Sensitivity 5000 µV/ division (5 mV/div) LFF 10 Hz HFF 10kHz Initial Sensory NCS Settings Sweep Speed 1-2 msec/division (generally 10 divisions are present in a horizontal screen) Sensitivity (gain) 20 µV/ division LFF 2-10 Hz HFF 2 kHz
  • 27. Artifacts and Technical Factors Physiological Factors (Patient- related) • Stimulus Artifact is an electrically recorded response that is elicited directly from the stimulator. • Occurs in all NCS, however, it only becomes a problem when the trailing edge of the recorded artifact overlaps with the potential being recorded. • Make sure the ground is between is the recording and stimulating electrodes to minimize stimulus artifact. Technical Factors (technically related) • Filters: faithfully reproduce the signal you want while trying to exclude both high and low frequencies. • Electrode Placement: proper electrode placement is a critical part of performing accurate electrodiagnostic studies.
  • 28. Stimulation Supramaximal Stimulation: all measurements made in NCS are done with the assumption that the strength of the stimulus is high enough to depolarize every axon in the nerve. • If Supramaximal stimulation is not achieved at the Distal Site, then you might mistakenly interpret this recording as signifying axonal loss due to low amplitude. • If Supramaximal stimulation is not achieved at the Proximal Site, this might appear to be a conduction block. • Conduction block is the failure of an Action Potential to be conducted past a particular point where as conduction is possible below the point of the block.
  • 29. With Respect to Evoked Potentials the stimulus can be graded as: • Subthreshold: when the stimulus is less than the threshold. • Threshold: when the stimulus is sufficient to produce a detectable response. • Submaximal: is stimuli below the maximal intensity. • Maximal: is the stimulus intensity after which a further increase in the stimulus intensity causes no increase in the amplitude. • Supramaximal: is intensity approximately 20% more than maximal, and used for Nerve Conduction Studies. For NCS, too much stimulation may cause co-stimulation of adjacent nerves or may stimulate nerves farther from the site The goal is to reach Supramaximal stimulation without applying so much stimulation that adjacent nerve are also stimulated.
  • 30. Measurements The machine will do most of the calculations for you • However, you still need to measure the distance between stimulation sites and the recording electrode of a patient’s limb when determining conduction velocity. Make sure the measurements are done accurately. • The main problem is if the patient’s limb is moved in different positions, which changes the distance you are measuring. (Commonly occurs in ulnar nerve studies.) Remember to follow the course of the nerve, rather than measuring the shortest distance between the stimulating and recording electrodes.
  • 31. Sweep Speed & Sensitivity Both the sweep-speed and sensitivity can affect you NCS results. • As the sensitivity is increased, the onset latency will decrease • Record all latency measurement with the same sensitivity and sweep speed.