MRI ARTIFACTS
A.SHAJITHA
B SC RADIOLOGY
MADRAS MEDICAL COLLEG
 What is an artifact??
 An image artifact is a structure not normally
present.
 But visible as a result of a limitation or
malfunction in the hardware or software in the
mri device.
 Some affect the quality of the MRI exam while
others do not affect the diagnostic quality but
may be confused with pathology.
 The knowledge of MRI artifacts and noise
producing factors is important for continuing
maintenance of image quality.
 Sources of artifacts
Hardware issues
Software problems
Physiological phenomena
Physical limitations
 Types of artifacts
 Chemical shift
 Aliasing
 Black boundary
 Gibbs or truncation
 Zipper artifact
 Motion artifacts
Entry slice phenomenon
 Field inhomogenity
 Slice overlapping
 Magic angle
 Moire fringes
 RF overflow
 Susceptibility
 Eddy current artifacts
 Center point artifact
Chemical shift artifact
 In the frequency direction, the MR scanner uses
the frequency of the signal to indicate spatial
resolution
 Since water in organs and muscles resonate at
different frequency than fat.
 The MR scanner mistakes the frequency as a
spatial difference.
 Commonly noticed in vertebral end plates,
abdomen, and orbits where fat and other tissues
form borders.
 Appearance
 Spine -one end plate to appear thicker than
the opposite one.
 Abdomen &orbits -Black border at fat water
interfaces and bright border at opposite
border.
 Solution
fat suppression technique
Use wide receiver bandwidth
Aliasing artifact or
wraparound
 It occurs when the dimensions of the body part
being imaged exceed the FoV.
 The part beyond the FoV is projected to the
other side of the image.
 It is caused by under sampling in the phase
encoded direction.
 Choosing an FoV that is smaller than the area
imaged leads to wraparound or aliasing
artifacts.
 Appearance
 signals from outside the FoV seen in the
imaging volume.
 Solution
 Larger FoV and oversampling
 Use saturation bands outside the FoV.
Black Boundary artifact
 Artificially created black line located at water fat
interfaces.
 This results in sharp delineation of the muscle
fat boundary that is sometimes visually
appealing but not an anatomical structure.
 Occurs at TE when the fat and water spins
located in the same pixel are out of phase,
cancelling each others signal.
 Particularly noticeable in GRE sequences.
 Both in frequency and phase directions.
 Remedy
Use phase TE’s
Fat suppression technique
Increase matrix size and bandwidth.
Gibbs or truncation artifact
 Occurs at high contrast boundaries.
 Due to truncation(omission) of sampled signals.
 Commonly seen at the low signal intensity spinal
cord with high signal intensity CSF on T2WI of the
spine.
 As the signal is sampled, some data is necessarily
omitted in k-space, causing the signal intensity of
a given pixel to vary from its ideal signal intensity.
 Appearance
 Bright and dark lines.
 Solution
 Increase matrix size.
Motion artifacts
 Voluntary/involuntary patient motion.
 Blurness in the image.
 Caused by
 Arterial pulsations
 CSF pulsations
 Swallowing
 Breathing
 Peristalsis
 Physical movement
 Solutions
 Patient education
 Sedation
 Motion correction sequences e.g. HASTE
 Navigator echo
 Respiratory gating
 Cardiac triggering
Entry slice phenomenon
 When unsaturated spins in blood first enter in to
a slice or slices.
 Characterized by bright signal in a blood vessel
at the first slice.
 Confused with thrombosis.
 Gradient echo flow techniques to differentiate.
 Solution
 Use spatial saturation bands before the first and
after the last.
Field inhomogenity
 Contribute to signal non uniformity.
 May arise from RF coil construction or standing
wave effects.
 When the RF is shorter than the dimensions of
the anatomic structures examined standing
waves may result.
 Leads to inhomogeneous fat suppression.
 Solution
 Shimming –allow precise control of overall
homogeneity of RF field.
 Use STIR for Fat Sat.
 Coil- use volume coils, allow space between body
and coil.
 Dielectric –use phased array coils.
Slice overlapping
 Loss of signal due to multi slice, multi angle
acquisition or imperfect slice profile.
 Mechanism : Spin saturation.
 If slices at different angles cross, then spins that
have previously excited, could be excited again.
 Mostly seen in spine imaging e.g. L4-L5 or L5-
S1.
 Appearance
 Band of signal loss crossing horizontally in the
image, usually posteriorly.
 Solution
 Continuous imaging
 Increase slice gap.
Magic angle artifact
 Seen most frequently in tendons and ligaments that
are oriented at a 55 degree to main magnetic field.
 Normally dipolar interaction between water
molecules in ligaments are strong.
 Which implies that T2 relaxation is very fast leading
to signal loss.
 Dipolar interaction go to zero.
 Solution
 Lengthen TE
 Use T1 weighted sequences since T1 relaxation is
unaffected by this.
Moire fringes
 An interference pattern most commonly when
doing gradient echo images.
 One cause is aliasing of one side of the body to
the other results in superimposition.
 Can also be caused by receiver picking up a
stimulated echo.
 Similar to the effect of looking through two
window screens.
 Solution
 Improve shimming
RF overflow artifact
 Non uniform, washed out appearance in an
image.
 Occurs when the signal received from the patient
is too intense to be digitized by analog to digital
converter.
 Auto prescanning usually adjusts the receiver
gain.
 Post processing methods also existing but may
be time consuming.
Magnetic suscepbility artifact
 Distortion in the MR image especially seen in
while imaging with metallic orthopedic hardware
or dental work.
 Magnetic field inhomogenetics introduced by the
metallic object in to homogeneous magnetic
field.
 Greater at high magnetic field strength.
 Worst with long TE and gradient echo
sequences.
 Appearance
Bright and dark areas
 Solution
Larger receiver bandwidth
Gradient echo and echo planar sequences
should be avoided.
The use of echo spin particularly fast spin
echo sequences should be considered.
Eddy current artifact
 Varying magnetic field induces electric current
which distort gradient waveforms.
 When diffusion gradient applied, change in
magnetic field creates electric current .
 Such current creates smaller magnetic field that
Bo.
 Modern gradient coils equipped with active
shielding to avoid these effects of electric
conduction.
 Solution
 Shielding gradients.
 A distorted gradient
waveform is used
Central Point artifact
 A focal dot of increased or decreased signal in
the center of the image.
 Caused by constant offset of the DC voltage in
the amplifiers.
 Solution
 Requires calibration.
 Maintain a constant temperature.
Zipper artifact
 Most of them related to hardware or software
problems.
 RF from some radio transmitters will cause
zipper that are oriented perpendicular to
frequency direction.
 Occur in either frequency or phase encoding
directions.
 Solutions
MR scanner room is shut down.
Remove all electric devices from patient.
Spike artifact
 Caused by one bad data point k-space.
 Shows one data point in k-space, which is out of
ordinary.
 Diagonal lines in the image.
 Solution
Repeat the scan
Annefact artifact
 Due to anatomy within the active volume of the
coil, but outside the Fov.
 Signals from such region give rise to ghost
imaging in the phase encoding direction.
 Most commonly in sagittal spine imaging
especially in T and L spines.
 Always turn off the coils which is not in use.
Mri artifacts
Mri artifacts

Más contenido relacionado

La actualidad más candente

Magnification(macro and micro radiography), distortion
Magnification(macro and micro radiography), distortionMagnification(macro and micro radiography), distortion
Magnification(macro and micro radiography), distortion
parthajyotidas11
 

La actualidad más candente (20)

Computed Tomography Instrumentation and Detector Configuration
Computed Tomography Instrumentation and Detector ConfigurationComputed Tomography Instrumentation and Detector Configuration
Computed Tomography Instrumentation and Detector Configuration
 
Ct tube and detectors
Ct tube and detectorsCt tube and detectors
Ct tube and detectors
 
CT image quality and image resolution.pptx
CT image quality and image resolution.pptxCT image quality and image resolution.pptx
CT image quality and image resolution.pptx
 
Quality Assurance Programme in Computed Tomography
Quality Assurance Programme in Computed TomographyQuality Assurance Programme in Computed Tomography
Quality Assurance Programme in Computed Tomography
 
MRI Coil and Gradient power-point slide pk
MRI Coil and Gradient power-point slide pkMRI Coil and Gradient power-point slide pk
MRI Coil and Gradient power-point slide pk
 
Post processing of computed tomography
Post processing of computed tomographyPost processing of computed tomography
Post processing of computed tomography
 
Recent Advances In MRI
Recent Advances In MRIRecent Advances In MRI
Recent Advances In MRI
 
Mri safety
Mri safetyMri safety
Mri safety
 
CT Procedure OF Abdomen & Pelvis
CT Procedure OF Abdomen & Pelvis CT Procedure OF Abdomen & Pelvis
CT Procedure OF Abdomen & Pelvis
 
CTDI (Computed Tomography Dose Index
CTDI (Computed Tomography Dose IndexCTDI (Computed Tomography Dose Index
CTDI (Computed Tomography Dose Index
 
Magnification(macro and micro radiography), distortion
Magnification(macro and micro radiography), distortionMagnification(macro and micro radiography), distortion
Magnification(macro and micro radiography), distortion
 
Mammography
MammographyMammography
Mammography
 
MRI Safety Basics
MRI Safety BasicsMRI Safety Basics
MRI Safety Basics
 
Ultrasound transducer doppler ppt pdf pk
Ultrasound transducer doppler ppt pdf pkUltrasound transducer doppler ppt pdf pk
Ultrasound transducer doppler ppt pdf pk
 
Multi slice ct ppt
Multi slice ct pptMulti slice ct ppt
Multi slice ct ppt
 
soft tissue radiography
soft tissue radiographysoft tissue radiography
soft tissue radiography
 
Basic Pulse Sequences In MRI
Basic Pulse Sequences In MRIBasic Pulse Sequences In MRI
Basic Pulse Sequences In MRI
 
Mri hazards & safety Daniel JP
Mri hazards &  safety  Daniel JPMri hazards &  safety  Daniel JP
Mri hazards & safety Daniel JP
 
CT Generations and Artefacts
CT Generations and ArtefactsCT Generations and Artefacts
CT Generations and Artefacts
 
Mri components
Mri componentsMri components
Mri components
 

Destacado

Distortion Artifacts in MRI and their correction
Distortion Artifacts in MRI and their correctionDistortion Artifacts in MRI and their correction
Distortion Artifacts in MRI and their correction
Miami Cancer Institute
 
Special Procedures: TBI, TSET and IORT
Special Procedures: TBI, TSET and IORTSpecial Procedures: TBI, TSET and IORT
Special Procedures: TBI, TSET and IORT
Miami Cancer Institute
 

Destacado (20)

Mri physics-uk
Mri physics-ukMri physics-uk
Mri physics-uk
 
Mr spectroscopy
Mr spectroscopyMr spectroscopy
Mr spectroscopy
 
076 cardiac magnetic resonance imaging
076 cardiac magnetic resonance imaging076 cardiac magnetic resonance imaging
076 cardiac magnetic resonance imaging
 
Factors controlling the beam of x-ray
Factors controlling the beam of x-rayFactors controlling the beam of x-ray
Factors controlling the beam of x-ray
 
Cardiac MRI
Cardiac MRICardiac MRI
Cardiac MRI
 
Nguyen khoi viet cardiac mri for the evaluation of ischemic heart disease jfi...
Nguyen khoi viet cardiac mri for the evaluation of ischemic heart disease jfi...Nguyen khoi viet cardiac mri for the evaluation of ischemic heart disease jfi...
Nguyen khoi viet cardiac mri for the evaluation of ischemic heart disease jfi...
 
Cardiac MRI principle
Cardiac MRI principleCardiac MRI principle
Cardiac MRI principle
 
Mri artifacts gamal mahdaly
Mri artifacts gamal mahdalyMri artifacts gamal mahdaly
Mri artifacts gamal mahdaly
 
What is spatial Resolution
What is spatial ResolutionWhat is spatial Resolution
What is spatial Resolution
 
Contrast based artifacts
Contrast based artifactsContrast based artifacts
Contrast based artifacts
 
Distortion Artifacts in MRI and their correction
Distortion Artifacts in MRI and their correctionDistortion Artifacts in MRI and their correction
Distortion Artifacts in MRI and their correction
 
Gre&SE
Gre&SEGre&SE
Gre&SE
 
Cardiac mri
Cardiac mriCardiac mri
Cardiac mri
 
Cardiac MRI
Cardiac MRICardiac MRI
Cardiac MRI
 
Fluroscopy
Fluroscopy Fluroscopy
Fluroscopy
 
Mr fluoroscopy
Mr fluoroscopyMr fluoroscopy
Mr fluoroscopy
 
Dark room equipments and entrance
Dark room equipments and entranceDark room equipments and entrance
Dark room equipments and entrance
 
Special Procedures: TBI, TSET and IORT
Special Procedures: TBI, TSET and IORTSpecial Procedures: TBI, TSET and IORT
Special Procedures: TBI, TSET and IORT
 
Radiographic factor
Radiographic factorRadiographic factor
Radiographic factor
 
Advances in Brachytherapy Treatment Planning and Delivery
Advances in Brachytherapy Treatment Planning and DeliveryAdvances in Brachytherapy Treatment Planning and Delivery
Advances in Brachytherapy Treatment Planning and Delivery
 

Similar a Mri artifacts

artifactsusgctmriradiologypk-190830131406.pptx
artifactsusgctmriradiologypk-190830131406.pptxartifactsusgctmriradiologypk-190830131406.pptx
artifactsusgctmriradiologypk-190830131406.pptx
Shubham661884
 
ppMRIARTIFACTS.pptx
ppMRIARTIFACTS.pptxppMRIARTIFACTS.pptx
ppMRIARTIFACTS.pptx
Shubham661884
 
MRIARTIFACTS.pptx
MRIARTIFACTS.pptxMRIARTIFACTS.pptx
MRIARTIFACTS.pptx
Shubham661884
 
Laser and its use in veterinary practice
Laser and its use in veterinary practiceLaser and its use in veterinary practice
Laser and its use in veterinary practice
Manzoor Bhat
 
LASER SURGERY VSR.pptx
LASER SURGERY VSR.pptxLASER SURGERY VSR.pptx
LASER SURGERY VSR.pptx
DrKanteshkumarMJ
 
1 diagnostic imaging
1 diagnostic imaging1 diagnostic imaging
1 diagnostic imaging
Simba Syed
 

Similar a Mri artifacts (20)

MRI Artifacts
MRI ArtifactsMRI Artifacts
MRI Artifacts
 
Artifacts and Quality Control
Artifacts and Quality ControlArtifacts and Quality Control
Artifacts and Quality Control
 
MRI Artefacts
MRI ArtefactsMRI Artefacts
MRI Artefacts
 
Navneet.mr hardware & artifacts related to it
Navneet.mr hardware & artifacts related to itNavneet.mr hardware & artifacts related to it
Navneet.mr hardware & artifacts related to it
 
mri artefacts.pptx
mri artefacts.pptxmri artefacts.pptx
mri artefacts.pptx
 
artifactsusgctmriradiologypk-190830131406.pptx
artifactsusgctmriradiologypk-190830131406.pptxartifactsusgctmriradiologypk-190830131406.pptx
artifactsusgctmriradiologypk-190830131406.pptx
 
ppMRIARTIFACTS.pptx
ppMRIARTIFACTS.pptxppMRIARTIFACTS.pptx
ppMRIARTIFACTS.pptx
 
ppMRIARTIFACTS.pptx
ppMRIARTIFACTS.pptxppMRIARTIFACTS.pptx
ppMRIARTIFACTS.pptx
 
MRI Artifact Radiology
MRI Artifact RadiologyMRI Artifact Radiology
MRI Artifact Radiology
 
MRIARTIFACTS.pptx
MRIARTIFACTS.pptxMRIARTIFACTS.pptx
MRIARTIFACTS.pptx
 
ICU Echo, now in 3D - by Janin
ICU Echo, now in 3D - by JaninICU Echo, now in 3D - by Janin
ICU Echo, now in 3D - by Janin
 
Laser and its use in veterinary practice
Laser and its use in veterinary practiceLaser and its use in veterinary practice
Laser and its use in veterinary practice
 
LASER SURGERY VSR.pptx
LASER SURGERY VSR.pptxLASER SURGERY VSR.pptx
LASER SURGERY VSR.pptx
 
Artifacts
ArtifactsArtifacts
Artifacts
 
CT and MRI
CT and MRICT and MRI
CT and MRI
 
Mri artifacts
Mri artifactsMri artifacts
Mri artifacts
 
Artifacts in MRI
Artifacts in MRIArtifacts in MRI
Artifacts in MRI
 
MRI 3-Marks QUESTIONS AND ANSWERS
MRI 3-Marks QUESTIONS AND ANSWERSMRI 3-Marks QUESTIONS AND ANSWERS
MRI 3-Marks QUESTIONS AND ANSWERS
 
Introduction to mri
Introduction to mriIntroduction to mri
Introduction to mri
 
1 diagnostic imaging
1 diagnostic imaging1 diagnostic imaging
1 diagnostic imaging
 

Último

Activity 01 - Artificial Culture (1).pdf
Activity 01 - Artificial Culture (1).pdfActivity 01 - Artificial Culture (1).pdf
Activity 01 - Artificial Culture (1).pdf
ciinovamais
 
Russian Escort Service in Delhi 11k Hotel Foreigner Russian Call Girls in Delhi
Russian Escort Service in Delhi 11k Hotel Foreigner Russian Call Girls in DelhiRussian Escort Service in Delhi 11k Hotel Foreigner Russian Call Girls in Delhi
Russian Escort Service in Delhi 11k Hotel Foreigner Russian Call Girls in Delhi
kauryashika82
 
Spellings Wk 3 English CAPS CARES Please Practise
Spellings Wk 3 English CAPS CARES Please PractiseSpellings Wk 3 English CAPS CARES Please Practise
Spellings Wk 3 English CAPS CARES Please Practise
AnaAcapella
 
1029-Danh muc Sach Giao Khoa khoi 6.pdf
1029-Danh muc Sach Giao Khoa khoi  6.pdf1029-Danh muc Sach Giao Khoa khoi  6.pdf
1029-Danh muc Sach Giao Khoa khoi 6.pdf
QucHHunhnh
 

Último (20)

Activity 01 - Artificial Culture (1).pdf
Activity 01 - Artificial Culture (1).pdfActivity 01 - Artificial Culture (1).pdf
Activity 01 - Artificial Culture (1).pdf
 
Explore beautiful and ugly buildings. Mathematics helps us create beautiful d...
Explore beautiful and ugly buildings. Mathematics helps us create beautiful d...Explore beautiful and ugly buildings. Mathematics helps us create beautiful d...
Explore beautiful and ugly buildings. Mathematics helps us create beautiful d...
 
UGC NET Paper 1 Mathematical Reasoning & Aptitude.pdf
UGC NET Paper 1 Mathematical Reasoning & Aptitude.pdfUGC NET Paper 1 Mathematical Reasoning & Aptitude.pdf
UGC NET Paper 1 Mathematical Reasoning & Aptitude.pdf
 
Key note speaker Neum_Admir Softic_ENG.pdf
Key note speaker Neum_Admir Softic_ENG.pdfKey note speaker Neum_Admir Softic_ENG.pdf
Key note speaker Neum_Admir Softic_ENG.pdf
 
Holdier Curriculum Vitae (April 2024).pdf
Holdier Curriculum Vitae (April 2024).pdfHoldier Curriculum Vitae (April 2024).pdf
Holdier Curriculum Vitae (April 2024).pdf
 
ComPTIA Overview | Comptia Security+ Book SY0-701
ComPTIA Overview | Comptia Security+ Book SY0-701ComPTIA Overview | Comptia Security+ Book SY0-701
ComPTIA Overview | Comptia Security+ Book SY0-701
 
Spatium Project Simulation student brief
Spatium Project Simulation student briefSpatium Project Simulation student brief
Spatium Project Simulation student brief
 
Unit-IV; Professional Sales Representative (PSR).pptx
Unit-IV; Professional Sales Representative (PSR).pptxUnit-IV; Professional Sales Representative (PSR).pptx
Unit-IV; Professional Sales Representative (PSR).pptx
 
Dyslexia AI Workshop for Slideshare.pptx
Dyslexia AI Workshop for Slideshare.pptxDyslexia AI Workshop for Slideshare.pptx
Dyslexia AI Workshop for Slideshare.pptx
 
psychiatric nursing HISTORY COLLECTION .docx
psychiatric  nursing HISTORY  COLLECTION  .docxpsychiatric  nursing HISTORY  COLLECTION  .docx
psychiatric nursing HISTORY COLLECTION .docx
 
This PowerPoint helps students to consider the concept of infinity.
This PowerPoint helps students to consider the concept of infinity.This PowerPoint helps students to consider the concept of infinity.
This PowerPoint helps students to consider the concept of infinity.
 
Unit-V; Pricing (Pharma Marketing Management).pptx
Unit-V; Pricing (Pharma Marketing Management).pptxUnit-V; Pricing (Pharma Marketing Management).pptx
Unit-V; Pricing (Pharma Marketing Management).pptx
 
Russian Escort Service in Delhi 11k Hotel Foreigner Russian Call Girls in Delhi
Russian Escort Service in Delhi 11k Hotel Foreigner Russian Call Girls in DelhiRussian Escort Service in Delhi 11k Hotel Foreigner Russian Call Girls in Delhi
Russian Escort Service in Delhi 11k Hotel Foreigner Russian Call Girls in Delhi
 
Unit-IV- Pharma. Marketing Channels.pptx
Unit-IV- Pharma. Marketing Channels.pptxUnit-IV- Pharma. Marketing Channels.pptx
Unit-IV- Pharma. Marketing Channels.pptx
 
Basic Civil Engineering first year Notes- Chapter 4 Building.pptx
Basic Civil Engineering first year Notes- Chapter 4 Building.pptxBasic Civil Engineering first year Notes- Chapter 4 Building.pptx
Basic Civil Engineering first year Notes- Chapter 4 Building.pptx
 
Making communications land - Are they received and understood as intended? we...
Making communications land - Are they received and understood as intended? we...Making communications land - Are they received and understood as intended? we...
Making communications land - Are they received and understood as intended? we...
 
Spellings Wk 3 English CAPS CARES Please Practise
Spellings Wk 3 English CAPS CARES Please PractiseSpellings Wk 3 English CAPS CARES Please Practise
Spellings Wk 3 English CAPS CARES Please Practise
 
1029-Danh muc Sach Giao Khoa khoi 6.pdf
1029-Danh muc Sach Giao Khoa khoi  6.pdf1029-Danh muc Sach Giao Khoa khoi  6.pdf
1029-Danh muc Sach Giao Khoa khoi 6.pdf
 
Accessible Digital Futures project (20/03/2024)
Accessible Digital Futures project (20/03/2024)Accessible Digital Futures project (20/03/2024)
Accessible Digital Futures project (20/03/2024)
 
On National Teacher Day, meet the 2024-25 Kenan Fellows
On National Teacher Day, meet the 2024-25 Kenan FellowsOn National Teacher Day, meet the 2024-25 Kenan Fellows
On National Teacher Day, meet the 2024-25 Kenan Fellows
 

Mri artifacts

  • 1. MRI ARTIFACTS A.SHAJITHA B SC RADIOLOGY MADRAS MEDICAL COLLEG
  • 2.  What is an artifact??  An image artifact is a structure not normally present.  But visible as a result of a limitation or malfunction in the hardware or software in the mri device.  Some affect the quality of the MRI exam while others do not affect the diagnostic quality but may be confused with pathology.  The knowledge of MRI artifacts and noise producing factors is important for continuing maintenance of image quality.
  • 3.  Sources of artifacts Hardware issues Software problems Physiological phenomena Physical limitations
  • 4.  Types of artifacts  Chemical shift  Aliasing  Black boundary  Gibbs or truncation  Zipper artifact  Motion artifacts Entry slice phenomenon  Field inhomogenity  Slice overlapping
  • 5.  Magic angle  Moire fringes  RF overflow  Susceptibility  Eddy current artifacts  Center point artifact
  • 6. Chemical shift artifact  In the frequency direction, the MR scanner uses the frequency of the signal to indicate spatial resolution  Since water in organs and muscles resonate at different frequency than fat.  The MR scanner mistakes the frequency as a spatial difference.  Commonly noticed in vertebral end plates, abdomen, and orbits where fat and other tissues form borders.
  • 7.  Appearance  Spine -one end plate to appear thicker than the opposite one.  Abdomen &orbits -Black border at fat water interfaces and bright border at opposite border.  Solution fat suppression technique Use wide receiver bandwidth
  • 8.
  • 9.
  • 10. Aliasing artifact or wraparound  It occurs when the dimensions of the body part being imaged exceed the FoV.  The part beyond the FoV is projected to the other side of the image.  It is caused by under sampling in the phase encoded direction.  Choosing an FoV that is smaller than the area imaged leads to wraparound or aliasing artifacts.
  • 11.  Appearance  signals from outside the FoV seen in the imaging volume.  Solution  Larger FoV and oversampling  Use saturation bands outside the FoV.
  • 12.
  • 13.
  • 14. Black Boundary artifact  Artificially created black line located at water fat interfaces.  This results in sharp delineation of the muscle fat boundary that is sometimes visually appealing but not an anatomical structure.  Occurs at TE when the fat and water spins located in the same pixel are out of phase, cancelling each others signal.  Particularly noticeable in GRE sequences.  Both in frequency and phase directions.
  • 15.  Remedy Use phase TE’s Fat suppression technique Increase matrix size and bandwidth.
  • 16. Gibbs or truncation artifact  Occurs at high contrast boundaries.  Due to truncation(omission) of sampled signals.  Commonly seen at the low signal intensity spinal cord with high signal intensity CSF on T2WI of the spine.  As the signal is sampled, some data is necessarily omitted in k-space, causing the signal intensity of a given pixel to vary from its ideal signal intensity.  Appearance  Bright and dark lines.  Solution  Increase matrix size.
  • 17.
  • 18. Motion artifacts  Voluntary/involuntary patient motion.  Blurness in the image.  Caused by  Arterial pulsations  CSF pulsations  Swallowing  Breathing  Peristalsis  Physical movement
  • 19.  Solutions  Patient education  Sedation  Motion correction sequences e.g. HASTE  Navigator echo  Respiratory gating  Cardiac triggering
  • 20.
  • 21. Entry slice phenomenon  When unsaturated spins in blood first enter in to a slice or slices.  Characterized by bright signal in a blood vessel at the first slice.  Confused with thrombosis.  Gradient echo flow techniques to differentiate.  Solution  Use spatial saturation bands before the first and after the last.
  • 22.
  • 23. Field inhomogenity  Contribute to signal non uniformity.  May arise from RF coil construction or standing wave effects.  When the RF is shorter than the dimensions of the anatomic structures examined standing waves may result.  Leads to inhomogeneous fat suppression.
  • 24.  Solution  Shimming –allow precise control of overall homogeneity of RF field.  Use STIR for Fat Sat.  Coil- use volume coils, allow space between body and coil.  Dielectric –use phased array coils.
  • 25.
  • 26. Slice overlapping  Loss of signal due to multi slice, multi angle acquisition or imperfect slice profile.  Mechanism : Spin saturation.  If slices at different angles cross, then spins that have previously excited, could be excited again.  Mostly seen in spine imaging e.g. L4-L5 or L5- S1.
  • 27.  Appearance  Band of signal loss crossing horizontally in the image, usually posteriorly.  Solution  Continuous imaging  Increase slice gap.
  • 28.
  • 29. Magic angle artifact  Seen most frequently in tendons and ligaments that are oriented at a 55 degree to main magnetic field.  Normally dipolar interaction between water molecules in ligaments are strong.  Which implies that T2 relaxation is very fast leading to signal loss.  Dipolar interaction go to zero.  Solution  Lengthen TE  Use T1 weighted sequences since T1 relaxation is unaffected by this.
  • 30.
  • 31. Moire fringes  An interference pattern most commonly when doing gradient echo images.  One cause is aliasing of one side of the body to the other results in superimposition.  Can also be caused by receiver picking up a stimulated echo.  Similar to the effect of looking through two window screens.
  • 33. RF overflow artifact  Non uniform, washed out appearance in an image.  Occurs when the signal received from the patient is too intense to be digitized by analog to digital converter.  Auto prescanning usually adjusts the receiver gain.  Post processing methods also existing but may be time consuming.
  • 34.
  • 35. Magnetic suscepbility artifact  Distortion in the MR image especially seen in while imaging with metallic orthopedic hardware or dental work.  Magnetic field inhomogenetics introduced by the metallic object in to homogeneous magnetic field.  Greater at high magnetic field strength.  Worst with long TE and gradient echo sequences.
  • 36.  Appearance Bright and dark areas  Solution Larger receiver bandwidth Gradient echo and echo planar sequences should be avoided. The use of echo spin particularly fast spin echo sequences should be considered.
  • 37.
  • 38. Eddy current artifact  Varying magnetic field induces electric current which distort gradient waveforms.  When diffusion gradient applied, change in magnetic field creates electric current .  Such current creates smaller magnetic field that Bo.  Modern gradient coils equipped with active shielding to avoid these effects of electric conduction.
  • 39.  Solution  Shielding gradients.  A distorted gradient waveform is used
  • 40. Central Point artifact  A focal dot of increased or decreased signal in the center of the image.  Caused by constant offset of the DC voltage in the amplifiers.  Solution  Requires calibration.  Maintain a constant temperature.
  • 41.
  • 42. Zipper artifact  Most of them related to hardware or software problems.  RF from some radio transmitters will cause zipper that are oriented perpendicular to frequency direction.  Occur in either frequency or phase encoding directions.  Solutions MR scanner room is shut down. Remove all electric devices from patient.
  • 43.
  • 44. Spike artifact  Caused by one bad data point k-space.  Shows one data point in k-space, which is out of ordinary.  Diagonal lines in the image.  Solution Repeat the scan
  • 45.
  • 46. Annefact artifact  Due to anatomy within the active volume of the coil, but outside the Fov.  Signals from such region give rise to ghost imaging in the phase encoding direction.  Most commonly in sagittal spine imaging especially in T and L spines.  Always turn off the coils which is not in use.